Centrifuge User Guide
Centrifuge User Guide
E-mail:satis@nuve.com.tr
WARRANTY CERTIFICATE
1.      Nüve warrants that the equipment delivered is free from defects during material and
        workmanship. This warranty is provided for a period of two years. The warranty period
        begins from the delivery date.
2.      Warranty does not apply to parts normally consumed during operation or general
        maintenance or any adjustments described in the operating instructions provided with
        the equipment.
3.      Nüve does not accept any liability in the case where the goods are not used in
        accordance with their proper intent.
4.      The warranty may not be claimed for damages occurred during the shipment, for
        damages resulting from improper handling or use, the defects in maintenance,
        negligence, bad functioning of auxiliary equipment, in the case of force majeure or
        accident and incorrect power supply.
5.      In the event of failure, Nüve shall be under no liability for any injury, or any loss or
        damage as the result of the failure other than the guarantee conditions.
•      BEFORE OPERATING THE INSTRUMENT THIS MANUAL SHOULD BE READ
       CAREFULLY.
•      THE VALIDITY OF THE GUARANTEE IS SUBJECT TO THE OBSERVATION OF
       THE INSTRUCTIONS AND PRECAUTIONS DESCRIBED IN THIS MANUAL.
•      INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF NÜVE. IT
       MAY NOT BE DUPLICATED OR DISTRIBUTED WITHOUT PERMISSION.
                                                                                                   1
DEAR NUVE USER
We would like to take this opportunity to thank you for preferring this Nüve product. Please
read the operating instructions carefully and keep them handy for future reference.
Please detain the packing material until you see that the unit is in good condition and it is
operating properly. If an external or internal damage is observed, contact the transportation
company immediately and report the damage. According to ICC regulations, this
responsibility belongs to the customer.
In case of a problem contact your Nüve agent for an authorized service or maintenance.
The validity of the guarantee is subject to compliance with the instructions and precautions
described in this manual.
Nüve reserves the right to improve or change the design of its products without any
obligation to modify previously manufactured products.
Information contained in this document is the property of Nüve. It may not be duplicated or
distributed without its permission.
                                                                                                   2
TABLE OF CONTENTS
SECTION 1 – INTRODUCTION ............................................................................................ 4
 1.1. USE AND FUNCTION ............................................................................................. 4
SECTION 2 – TECHNICAL SPECIFICATIONS ..................................................................... 5
 2.1. TECHNICAL SPECIFICATIONS TABLE ................................................................. 5
 2.2. ACCESSORIES FOR NF 1200/1200R .................................................................... 6
 2.3. GENERAL PRESENTATION .................................................................................. 7
 2.4. ROTOR SELECTION TABLE .................................................................................. 8
 2.5. PRECAUTIONS AND USAGE LIMITATIONS ......................................................... 9
SECTION 3 – SYMBOLS .....................................................................................................10
SECTION 4 – INSTALLATION .............................................................................................10
 4.1. LIFTING AND TRANSPORT ..................................................................................10
 4.2. UNPACKING ..........................................................................................................10
 4.3. POSITIONING........................................................................................................11
 4.4. MAIN SUPPLY .......................................................................................................11
SECTION 5 – OPERATING UNIT ........................................................................................11
 5.1. OPERATING ..........................................................................................................11
 5.2. DISPLAY AND CONTROL PANEL.........................................................................12
 5.3. MAKING A NEW PROGRAM .................................................................................14
   5.3.1 SELECTING THE MEMORIZED PROGRAM .....................................................15
   5.3.2 PULSE MODE ....................................................................................................16
 5.4    END OF OPERATION............................................................................................17
 5.5    SAFETY INTERLOCK SYSTEM ............................................................................17
 5.6    IMBALANCE DETECTION SYSTEM .....................................................................17
 5.7    MANUAL LID OPENING ........................................................................................18
SECTION 6 – OPERATING PRINCIPLES............................................................................18
 6.1    PREPARATION OF THE ROTOR TO RUN ...........................................................18
 6.2    LOADING ...............................................................................................................18
 6.3    ROTOR INSTALLATION ........................................................................................19
 6.4    DRIVE SYSTEM ....................................................................................................20
SECTION 7 – CLEANING AND PERIODIC MAINTENANCE ...............................................20
 7.1    PERIODIC MAINTENANCE ...................................................................................20
 7.2    STERILIZATION ....................................................................................................21
 7.3    CORROSION INFORMATION ...............................................................................21
   7.3.1 CHEMICAL CORROSION ..................................................................................21
   7.3.2 STRESS CORROSION ......................................................................................21
 7.4    CLEANING .............................................................................................................22
 7.5    ELECTRICITY ........................................................................................................22
SECTION 8 – DISPOSAL MANAGEMENT CONCEPT ........................................................22
BÖLÜM 9 – TROUBLESHOOTING ......................................................................................23
SECTION 10 – ELECTRICAL CIRCUIT DIAGRAMS ...........................................................24
SECTION 11 – WARNING LABEL .......................................................................................26
                                                                                                                                    3
SECTION 1 – INTRODUCTION
The NF 1200 / NF 1200 R centrifuges are designed to separate liquids of different densities
by applying centrifugal force. Different types of rotors, which are swing-out and angle rotors,
can be fitted to the centrifuges.
The microprocessor control system of the centrifuges enables the user to observe and to
control the speed (RPM or RCF), the acceleration and braking rates, run time, program
number and temperature of the bowl for NF 1200 R. The microprocessor control system also
displays the RPM or RCF value during the run.
The ventilation system of the NF 1200 limits temperature in the bowl between the ambient
temperature + 3 °C and ambient temperature + 12 °C depending on the rotor type. The NF
1200 R is equipped with a cooling system which provides to operate between -9 °C and + 40
° C.
It ensures safe processes by means of the locking system which does not allow the
centrifuge to operate if the lid is open and which does not allow the rotor to spin if the lid is
not closed.
The lid locking system which does not allow the centrifuge to operate if the lid is open, which
also does not allow the lid to be opened while the rotor is spinning and the possibility to open
the lid by pushing only one key when the program ends, provide safe and easy working
conditions. There are audible and visible alarms to inform the operator when the lid is open,
when the program ends and when any error conditions occur. In case of a power failure, the
lid could be opened manually by using a manual lid opening tool
                                                                                                    4
SECTION 2 – TECHNICAL SPECIFICATIONS
                                  NF 1200                                     NF 1200 R
                                RA 280 rotor: 4100 rpm, MP 200 rotor: 4100 rpm,
                      RS 600 rotor: 9000 rpm, RS 100 rotor: 12000 rpm, RS24 rotor: 14000
   Max Speed
                                                    rpm,
                                               RS 30 rotor: 13500 rpm
                     RA 280 rotor: 3082 x g, MP 200 rotor: 2349 x g, RS 600 rotor: 8965 x g,
       Max. RCF
                      RS 100 rotor: 14167, RS24 rotor : 18405 x g, RS 30 rotor: 20170 x g
                          RA 280 rotor: 4x280 ml, MP 200 rotor: 2x3 microtitre plates,
  Max. Capacity      RS 600 rotor: 6x100 ml, RS 100 rotor: 10x10 ml, RS 24 rotor: 24x1.5/2
                                           ml, RS 30 rotor: 30x2 ml
 Control System                             Microprocessor controlled
        Display                                 High Visibility Digital
 No. of Memories                                          10
  Speed Range                                     500-14000 rpm
   Speed Step                                          10 rpm
 Speed Accuracy                                       ± 20 rpm
   Timer Range                              01-99 min + Hold Position
       Timer Step                                       1 min
Acceleration Rate                              0: Slowest 9: Fastest
   Braking Rate                                0: Slowest 9: Fastest
  Power supply                                      230 V, 50 Hz
   Temp. range                                                              -9° C / 40 ° C
Refrigerant Liquid                                                              R134a
 Temp. Accuracy                                                                  1°C
      Temp.             (at maximum speed) ambient
                                                                          +4°C at max. speed
   Performance                  temp+12°C
   Max. Power                      500 W                                        800 W
 Ext. Dimension                455 x 545 x 405                             770 x 545 x 410
  Packing Dim.                 600 x 530 x 470                             870 x 650 x 550
   Net / Gross
                                  42 / 47 kg                                  75 / 80 kg
    Weight
                                                                                               5
2.2.   ACCESSORIES FOR NF 1 200/1200R
                                                                               6
2.3.    GENERAL PRESENTATION
Figure 1
Figure 2
                                                                                         7
2.4.   ROTOR SELECTION TABLE
                                                                   Max.   Max.RCF
                                              Max.Tube   Radius             xg
 Rotor       Description        Capacity                          Speed
                                              Dia.(mm)    (mm)
                                                                  (rpm)
           Swing-out Rotor       4x280ml                  164     4,100    3,082
                                  380 ml
           Set of 4 buckets                      76
                                 nominal
            Set of 4 inserts     1x280ml         65
            Set of 4 inserts   1x250ml flat      62
            Set of 4 inserts   1x200ml flat      55
 RA280      Set of 4 inserts     1x100ml         39
                                  2x50ml
            Set of 4 inserts                     30
                                  conical
            Set of 4 inserts     4x25ml          25
                                  5x15ml
            Set of 4 inserts                     17
                                  conical
            Set of 4 inserts     12x15ml         17
            Set of 4 inserts    12x5/7ml        13.5
            Set of 4 inserts   19x5ml RIA        13
                                  2x3
            Microtitre Plate
MP200                           microtitre                125     4,100    2,349
                Rotor
                                 plates
                                                                                    8
Angle Rotor Accessories;
                                                          Max.
                                            Max.Tube                Max.Speed
Rotor        Description     Capacity                    Radius                   Max RCFxg
                                            Dia(mm)                   (rpm)
                                                          (mm)
             Angle Rotor      6x100ml           38          99         9,000          8,965
              Set of 6
                              1x50ml            30
              adaptors
RS600         Set of 6        1x50 ml
                                               30.5
              adaptors        conical
              Set of 6
                             1x30/38ml          26
              adaptors
              Set of 6        1x15 ml
                                                17
              adaptors        conical
             Angle Rotor      10x10ml           16          88        12,000         14,167
RS100         Set of 10
                               1x6ml            13
              adaptors
              Set of 10
                             1x1.5/2ml          11
              adaptors
             Angle Rotor     24x1.5/2ml         11          84        14,000         18,405
              Adaptor       1x500/800µl         8
    RS24                      1x200 µl
               Adaptor                         6.5
                                PCR
                           1x250/400/70
               Adaptor                          6
                                0 µl
             Angle Rotor       30x2ml           11          99        13.500         20.170
              Adaptor       1x500/800µl         8
    RS30                      1x200 µl
               Adaptor                         6.5
                                PCR
                           1x250/400/70
               Adaptor                          6
                                0 µl
•      Do not use the device for any purpose other than the usage purpose.
•      Prior to first use, the user’s manual should be read and the device is only to be used by
       authorized and trained personnel. Only authorized technical personnel handle the
       product in case of any failure.
•      The working bench should be durable to the device weight and vibration isolated.
•      Ensure that the rotor is placed correctly prior to usage.
•      According to the standard IEC 61010-2-020, anyone and any hazardous materials
       should not be in the 300 mm safety zone while centrifuge is running.
•      Do not move the device while it is running.
•      Do not open the lid while rotor is spinning.
•      Apply the manual lid opening procedure in the case of power cut or in the case of any
       error.
•      Use only the the spare parts, rotors and accessories which are supplied by NUVE.
•      Load the rotor according to the explanations in the user’s manual.
•      Start the device after ensuring the rotor is loaded correctly.
•      Do not use the centrifuge in areas which are in explosive danger.
                                                                                                   9
•      Do not centrifuge the explosive, flammable, radioactive, corrosive materials and the
       materials which may react with each other.
•      The centrifuge and the rotor are not microbiologically leak-proof. Use tubes with leak-
       proof covers, if hazardous, toxic and pathogenic microorganisms are centrifuged.
•      Do not use corrosive materials which may be harmful for the device integrity, rotor and
       accessories.
•      Do not use rotors and accessories with corrosion and mechanical damages.
•      Mains supply should be appropriate to power of the device and grounded.
•      Use tubes whose sizes are suitable to the rotor and accessories.
•      Tubes which are used in the centrifuge should not be deformed by the effect of the
       centrifuge force.
•      Use glass tubes to balance, if glass tubes are used. Use plastics tubes to balance, if
       plastics tubes are used.
•      Do not start the device unless tubes are in balance.
•      Imbalance loading may cause mixing the samples, broken tubes, and damages on the
       rotor and motor shaft.
SECTION 3 – SYMBOLS
SECTION 4 – INSTALLATION
Because of the heavy weight of the centrifuge, all lifting and transport must be carried out
using proper handling equipment. The centrifuge should be lifted from underneath and never
be turned over.
4.2. UNPACKING
Remove the packing cardboard box and the second nylon packing around the centrifuge.
The below written are provided with the instrument, please check them;
•    1 ea. user’s manual
•    1ea. Warranty certification
•    1 ea. power cable
•    1 ea. manual lid opening tool
•    1 ea. 24 socket wrench
•    1 ea. Wrench
•    Lubricant oil
                                                                                                 10
4.3.    POSITIONING
The centrifuge requires 230 V, 50 / 60 Hz. Please make sure that the supplied mains match
the required power ratings. If no, provide an extra line to support.
5.1. OPERATING
•   Open the lid and check that there is no sample in the instrument.
•   Turn the centrifuges on by using On/Off switch.
•   See that the display and control panel activates.
•   Learn the functions of the control panel (See Part 5.2).
•   Set the values and start the operation (See Part 5.3)
                                                                                                  11
5.2.    DISPLAY AND CONTROL PANEL
Display of speed (500-14.000 rpm) or display of RCF (0-18.405 x g) during the run, rotor type
during programming and error code.
Display of the run time from 1 min. to 99 min. and hold position (H).
Display of the acceleration and brake rates from 0 to 10, for NF 1200 R it also displays the
bowl temperature during programming and the run.
The 4 leds light up consecutively during the run to indicate the spinning of the rotor.
This has two functional move. Turn clockwise and opposite clockwise, reach program menu,
set value and “READY” situation. By turning the Encoder button increase or decrease values
during programming stage. When Led is “READY” situation, by pushing Encoder button start
device. Stops the run by starting the braking phase manually.
07-LID Key
                                                                                                12
08- LID Indicator Led
It turns on while the acceleration rate is being programmed and while the rotor is
accelerating.
This indicator turns on while the breaking rate is being programmed and while the rotor is
breaking.
It indicates that the cooling system functions. During programming, it indicates that the
temperature value is being set on Acc/Br/Temp Display (3).
PULSE mode on if pushing button by during READY situation. It is pushed to see the RPM or
RCF value on the speed display during the run, to pass among the displays and to store the
programmed values.
                                                                                             13
5.3.   MAKING A NEW PROGRAM
Determine the rotor type, the speed and time, the acceleration and braking rates.
                                                                                    14
        ROTOR TYPE                            DISLPAY                   MAX. (RPM)   MAX. (RCF)
      RA 280 Swing – Out                         4.280                       4,100     3,082
     MP 200 Swing – Out                          PLAtE                       4,100     2,349
         RS 600 Angle                            6.100                       9,000     8,965
         RS 100 Angle                            10.10                      12,000     14,167
         RS 24 Angle                           24.     1.5                  14,000     18,406
         RS 30 Angle                           30.     1.5                   13500     19968
                    The correct rotor type should be selected in order to see the correct
                    RCF values during the run.
                     Pre Cooling;
                     While operating the NF 1200 R centrifuge at sub-ambient
                     temperatures, in order to get the maximum performance, without
                     loading the rotor run the centrifuge at 1.500 rpm, at a temperature
                     which is 3° C lower than the required temperature and at hold position
                     until you see that the temperature inside the bowl drops to the required
                     temperature. Then, stop the centrifuge, load your samples and start
                     centrifugation.
• Load the rotor with samples by paying attention to the dynamic and static balances.
                                                                                                  15
•       Close the lid, see that the lid open warning led (8) turns off.
•       See that the READY led (16) turns on and push Encoder button (6).
•       The rotor starts accelerating according to the set value and the acceleration led turns
        on. The set speed or RCF value is displayed during the set time duration. (RPM or
        RCF screen will appear in every 3 seconds). For the operations whose speeds are
        higher than 2.000 rpm, the rotor accelerates according to the set acceleration value up
        to 2.000 rpm and then continues accelerating with the highest acceleration value, 9 to
        reach the set speed. As the speed reaches the set value, “-“ sign appears on the
        acc/brk/temp display (3).
•       The elapsed time is counted down as the centrifugation starts.
•       When the time display shows “00”, the program ends, the breaking led turns on and the
        set break rate is shown at acc/brk/temp display (3). The rotor brakes with the highest
        break rate, 9 until it slows down to 2.000 rpm and continues braking according to the
        set braking value.
•       The elapsed time until the rotor stops is counted and is shown at the time display. This
        is the time that passes until the rotor stops.
•       When the speed display shows “0”, the “end” expression appears. The user is warned
        by an intermittently sounding alarm.
•       Push the LID button (7) to open the lid.
•       You may leave the centrifuge at stand-by position.
•       Load the rotor with samples by paying attention to the dynamic and static balances.
•       Set the all parameters for program (See 5.3)
•       Close the lid, see that the lid open warning led (8) turns off.
•       See that the READY led (16) turns on and push Encoder button (6).
•       You can reach the desired speed by pressing and holding the Encoder button (6) within
        the RPM speed limits you have set.
•       In the PULSE mode, the rotor operates according to the set acceleration and braking
        values you set.
•       While operating in Pulse mode, press the Encoder button (6) and start counting from
        the moment the rotor starts to accelerate. When you take your hand out of the Encoder
        button, it will stop counting by braking.
                      You may display the RPM or RCF value on the speed display during
                      the run by pushing the function button.
                                                                                                   16
                     At hold position, the rotor keeps spinning until the Encoder button (6) is
                     pushed. After the Encoder button has been pushed, the rotor brakes
                     according to the set brake rate.
                     The centrifuge waits at the stand-by position with the program values of
                     the last program.
                     Please open the lid after every operation, otherwise the centrifuge
                     cannot be run again.
At the end of the cooling operation with the NF 1200 R centrifuges, dry with a non-wool soft
cloth or paper towel at the centrifuge cell. (When the end of cooling operation with NF1200,
has a puddle at the centrifuge cell. So dry to water with paper towel and soft cloth.)
The safety interlock system prevents opening of the lid when the rotor is spinning.
The centrifuge does not operate until the lid is closed and the lid remains locked until the
rotor stops spinning. The “Lid Open” indicator turns on to warn the user if the lid is not closed
properly.
The “Err 1” imbalance error is displayed and the imbalance indicator (9) turns on. The lid
cannot be opened until the rotor stops spinning. The centrifuge can only be started after the
lid has been opened and the rotor has been re-loaded correctly.
To avoid facing imbalance problem, please make sure to insert the tubes correctly.
                                                                                                    17
5.7   MANUAL LID OPENING
In case of power cut or any breakdown, the centrifuge can be opened manually to access the
samples.
                      Before opening the lid manually, make sure that the rotor already
                      stopped spinning. Upon opening the lid, lift it by hand and observe the
                      rotor. If the rotor is still spinning, close the lid and wait approximately
                      10 minutes before repeating the operation. This operation must be
                      carried by someone who is informed of the danger and of the
                      precautions which must be undertaken.
Chemical corrosion or mechanical corrosion may do severe damage to the rotor and the
centrifuge. Particles which are stuck inside the inserts cause the breakage of tubes and lead
to major imbalance please check to make sure that no particles are left on the rotor.
The central hole of the rotor and the motor shaft should also be clean and dry before all
centrifugal operations and they must be kept in that way all the time.
6.2 LOADING
The most important condition of an efficient centrifugation is to balance the tubes properly.
Loading must be done by meeting the requisites of static and dynamic balance.
Static Balance: This balance states that diametrically oppositely replaced weights are
almost the same. In application, the liquid level in the tubes should be at the same height to
balance the load.
Dynamic Balance: This balance states that diametrically oppositely replaced center of the
gravity of tubes are symmetrical with respect to spinning axis of the rotor.
•     Although in hospitals the samples having almost the same densities are centrifuged, in
      industry, samples having different densities may be centrifuged. In this case, the
      dynamic balance becomes more important factor than the static balance is.
                                                                                                    18
•     If the number of tubes to be centrifuged is less than the capacity of the rotor, the tubes
      must be placed oppositely. If an odd number of tubes is centrifuged, a water filled tube
      at the same weight should be used for balancing.
                     Imbalance of the rotor may cause major damage to the rotor and
                     centrifuge.
                     Balance the rotor with glass tubes if you use glass tubes for
                     centrifugation.
                     Balance the rotor with plastic tubes if you use plastic tubes for
                     centrifugation.
Figure 5
                                                                                                   19
                                             Figure 6
•     The rotor is driven by a three phase asynchronous motor. The microprocessor control
      system assures the correct drive speed.
•     The force applied to the rotor is directly related to the shape of the rotor, the swing-out
      rotor receives more load than the angle rotor does. Longer radius and more
      accessories increase the load of the rotor and decrease the spinning speed.
•     The centrifuge does not allow the rotor to spin at a speed which it cannot resist
      mechanically.
•     Please make that sure the correct type of rotor is selected during programming.
SECTION 7 – CLEANING AND PERIODIC MAINTENANCE
                                                                                                    20
7.2   STERILIZATION
•     Apply alcohol, for example %70 ethanol or isoprophanol, for 10 minutes against
      bacteria and viruses.
•     The rotors and buckets may be autoclaved at 121°C and under 215 kPa pressure for
      20 minutes but please do not forget to remove all accessories.
•     Do not use formaldehyde for the sterilization process.
•     Phenol is a corrosive substance and should never be used.
•     Glutaraldehyde is a toxic substance and increases the rate of fatty acid in the body.
7.3   CORROSION INFORMATIO N
•     Nuve rotors which are made of aluminum are designed to spin at proper RCFs for
      many years. When used properly, their resistance to corrosion and their life span
      increases and the imbalance problems decrease.
•     All accessories should be checked thoroughly and regularly as almost all laboratories
      already have the conditions which lead to corrosion easily.
•     Chemical vapors in the laboratory environment which dissolve in the water on the rotor
      (in refrigerated centrifuges)
•     Corrosive liquids which overflow from overfilled and unsealed tubes. (the liquids which
      spread out during centrifugation)
•     Contaminated and non-cleaned buckets, tubes and bottles.
                      If the centrifuged samples are corrosive, only rinsing with water is not
                      sufficient. The residuals dissolve in the water and humidity on the rotor
                      and in the buckets.
                     Some particles may stick to the tubes, buckets and adapters. These
                     particles crash and do harm to anodized surface during centrifugation
                     and ease the occurrence of corrosion.
The corrosive material is pushed against the aluminum alloy by the centrifugation “g” force
during the centrifugation. This situation causes the stress corrosion to occur more quickly
                                                                                                    21
than the chemical corrosion does. Microscopic cracks occur under the force of the
centrifugation.
Every centrifugation causes the aluminum rotor to be attacked by the chemical more and
more and eventually micro-cracks decrease the resistance of the rotor against the
centrifugation force. Fortunately, no crash occurs just after the first micro-cracks have
appeared as the rotors are manufactured according to the high safety limits.
The corrosion of the small amount of corrosive materials does not result in severe cracks but
weakens the mechanical resistance of the rotor in the course of time.
7.4 CLEANING
Centrifuged at high voltages are present behind the panels. These panels are electrically
disconnecting Do not open the centrifuge.
SECTION 8 – DISPOSAL MANAGEMENT CONCEPT
The currently valid local regulations governing disposal must be observed. It is in the
responsibility of the user to arrange proper disposal of the individual components.
The instruments and electronic accessories (without batteries, power packs etc.) must be
disposed off according to the regulations for the disposal of electronic components.
Batteries, power packs and similar power source have to be dismounted from
electric/electronic parts and disposed off in accordance with applicable local regulations.
                                                                                                22
BÖLÜM 9 – TROUBLESHOOTING
Check that,
Err 1
• Imbalance error. Wrong loading.
•   To avoid facing imbalance problem, please make sure to insert the tubes correctly. (See 5.6)
Err 2
•   Motor overheat. Please switch off the centrifuge for at least 20 minutes
Err 3
•   Speed failure. The optical sensor could have failed. Please observe whether the
    “running” indicator leds are on or not. If it is not on, the optical sensor is defective. If it is
    on, the motor could have failed. Contact to your service if either occurs.
•   This error code occurs when the mains voltage drops below the lower limit.
Err 4
•   Lid failure. This failure occurs when lid is opened during the centrifugation.
Err 5
• The temperature sensor endings are broken. This error code appears only on NF 1200R.
Err 6
• Motor failure. The motor driver unit (inverter) is defective. Switch off the centrifuge and
    switch on again after the motor has stopped. If you still see the error code, contact to the
    service.
Err 7
• “Lid Open” switch are broken on the Lid Board.
Err 8
• “Lid Close” switch are broken on the Lid Board.
Con.
•   The communication between the display & main PCB and the Display PCB fails.
                                                                                                         23
SECTION 10 – ELECTRICAL CIRCUIT DIAGRAMS
                                           24
25
SECTION 11 – WARNING LABEL
EARTHED FUSES
           NF 1200R                   NF 1200
         FUSES ( 2x10A)            FUSES ( 2x4A)
26