XC-A30 ESR Analyzer
Service Manual
  CARETIUM MEDICAL INSTRUMENTS Co., Ltd.
Address: Future Plaza , Building A, 10th Floor, Unit 1002, Nanshan , Shenzhen, China
                        Tel: 86 755 25273714          Fax : 86 755 25273096
                     Http://www.caretium.com info@caretium.com
                               XC-A30 ESR Analyzer Service Manual
                                                         Contents
1、General........................................................................................................................... 1
      1.1     Internal structure: .................................................................................................1
2、Sample collection and anticoagulant .............................................................................. 1
      2.1     Vein collection ......................................................................................................1
      2.2     ESR tube ..............................................................................................................1
3、Sample preparation and mixing ...................................................................................... 1
4、Precautions .................................................................................................................... 2
5、Maintenance................................................................................................................... 2
6、Troubleshooting.............................................................................................................. 3
7、Disassembly of component and replacement................................................................. 5
      7.1     Cover disassembly ...............................................................................................5
      7.2     LCD screen disassembly......................................................................................5
      7.3     Main-board disassembly.......................................................................................6
      7.4     Transformer replacement .....................................................................................7
      7.5     Fuse replacement.................................................................................................8
      7.6     Fan disassembly ..................................................................................................8
      7.7     Optical detection board disassembly....................................................................9
      7.8     Printer disassembly ............................................................................................ 10
      7.9     Separation of the optical detection board pin and the socket ............................. 11
      7.10      Regulator block disassembly............................................................................ 11
8、Main board instruction .................................................................................................. 12
                                  XC-A30 ESR Analyzer Service Manual                                                                2
1、General
    XC-A30 automated Erythrocyte Sedimentation Rate analyzer is a high-speed and
easy-to-use instrument. The measurement is base on Westergen by optical obstructed
principle.
1.1     Internal structure:
    Power switch, power socket, fan, RS-232 interface, bar code scanning interface,
grounding pole, voltage regulator components, LCD display, touch screen, printer, printer
driver, mainboard, optical detection board、step motor.
2、Sample collection and anticoagulant
2.1     Vein collection
          Requirement: The blood collection should be finished within 30 seconds without
      blood clot and hemolysis. Otherwise it will affect the test results or repeatability.
2.2     ESR tube
          Requirement: Please use ESR tube provided or certified by Caretium.
          At the present, there are two kinds of ESR tube in the market::
          1) The ESR tube is a vacuum tube with 0.32ml anti-coagulant.
          2) The ESR tube isn’t a vacuum tube and the operator needs to add anticoagulant
               manually.
3、Sample preparation and mixing
            It requires 1.28ml whole blood for each sample.
            To ensure injection of proper volume of blood into the ESR tube, there is a
       indicating line on the ESR tube. The blood should be around indicating line ±
       2mm.The sample and anti-coagulant must be fully mixed before measuring.
          (Note: Air bubbles should be excluded from the anti-coagulant blood.)
                           XC-A30 ESR Analyzer Service Manual                                 1
      Make the tube upside down slowly for at least 5 times, so that the anticoagulant
  and blood mix well.
4、Precautions
  1、 Do not use the ESR tube if there is breakage or dust around it.
  2、 Check if there is anti-coagulant in the ESR tube before measurement; make sure
      the volume of anti-coagulant is 0.32ml.
  3、 The timing from sample collection to measurement should be less than 30 minutes.
  4、 The upside-down mixing should be slow. Air bubbles should be avoided; otherwise
      it will affect the test results.
  5、There should not be any clot inside the blood; otherwise it will affect the test results
      or repeatability.
  6、Don’t insert the ESR tube when the instrument initialization fails.
  7、Don’t take out the ESR tube during the measuring; otherwise, it will affect the test
      results. Reinject sample to another ESR tube if retest is required.
  8、Do not insert the tube into the measurement hole during the initialization; otherwise,
      the initialization for the hole may fail..
  9、Please take the ESR tube out vertically after measurement is finished,
5、Maintenance
  1、There is no special requirement for maintenance. Keep the operation environment
      dry and clean. Keep the ESR tube hole away from the dust. Avoid direct sunlight to
      the instrument. Please shut down the instrument when the measurement is
      finished and cover the ESR tube hole.
  2、Insert and taking out the ESR tube carefully to avoid the tube being damaged which
      will lead to the blood flowing into the instrument and the test result uncorrectly.
  3、Clean the instrument surface carefully, so that there is no liquid flowing into the
      holes.
  5、 Clean the outside surrounding of the holes surface with 5% sodium hypochlorite to
      protect the operator.
                       XC-A30 ESR Analyzer Service Manual                                   2
6、Troubleshooting
      Symptom                               Cause                                  Solution
                             Check if the fuse of power               Replace it if necessary.
                             socket is broken.
  LCD doesn’t display        Open the cover; check if the             Re-connect it if necessary.
                             connection between the LCD
                             and the main-board is reliable.
                             Check        the   power        supply   Add a voltage regulator if
                             stability.                               necessary.
                             Open the cover; check if the             Re-connect it if necessary.
     LCD displays            connection between the LCD
    incompletion or          and the main-board is reliable.
          incorrect          Open the cover, check if the             Replace the LCD display.
                             assembly       position    of    LCD
                             correctly.
                             The fuses are broken.                    Replace the fuse.
                             Open the cover, check if the             Re-connect it if necessary.
Printer    doesn’t    work
                             connection between the printer,
after power-on.
                             printer board and the
                             Mainboard reliable.
                             Check if the connection with             Re-connect them.
 No print-out after the      the printer, printer board and           Replace printer or printer
     test finished.          main-board reliable;                     board if necessary; make
                                                                      sure the setting of printer in
                                                                      the instrument is “On”
The result deviation is      Check the power supply.                  If   necessary    add    voltage
much too big.                                                         regulator.
                             The optical detection board in           Avoid placing the
                             direct sunlight.                         instrument in direct
                                                                      sunslight.
                             Check if the outer sutface of            Clean it or replace a new
                             ESR tube is clean.                       tube and re-test the sample.
                             The blood position is too high           Re-add sample and re-test it.
                             or too low.
                             There are bubbles inside the             Re-add sample and re-test it.
                             blood.
                          XC-A30 ESR Analyzer Service Manual                                             3
                              There is clot inside the blood.    Re-add sample and re-test it.
                              The deviation of environment       Adjust environment
                              temperature is too big.            temperature; re-add sample
                                                                 and re-test it.
                              The instrument does not have       Calibrate temperature;
                              temperature calibration            re-add sample and re-test it.
                              The screen displays “Error”        Remove dust around the
                                                                 optical detection board or
                                                                 main-board.
                              The optical detection board is     Avoid placing the
                              in direct sunlight.                instrument in direct
                                                                 sunlight.
                              The position of optical            Fixing or adjust it
  Initialization failured
                              detection board get
                              excursion or get loose
                              The connection of the              Re-joint or replace it.
                              components of optical
                              detection board is unreliable or
                              damaged.
Displaying tube               The components of optical          Remove the dust or replace
inserted, but no tube         detection board are covered by     the optical detection
inserted                      dust or damaged.                   board.
                              There is dust or greasy dirt       Remove the dust or greasy
  The motor gives big
                              around the pillar which is for     dirt.
   noise or shaking
                              the board going up and down
                              Dust around the optical
                              detection board.                   Remove the dust.
                              The position of optical            Fix     or   adjust   the   board
The stepper motor
                              detection board get excursion      position.
keeps running and
                              or get loose.
gives big shaking
                              The connection of the              Re-joint or replace it.
noise.
                              components of optical
                              detection board is unreliable or
                              damaged.
                            XC-A30 ESR Analyzer Service Manual                                       4
7、Disassembly of component and replacement
7.1   Cover disassembly
        a) Three screws need disassembly in the rear panel. As shown in figure 1:
                                              Figure 1
        b)   Eight screws need disassembly in the bottom panel. As shown in figure 2
                                               Figure 2
7.2   LCD screen disassembly
       Open the instrument cover.
      a) Four screws need assembly on the LCD cover. As shown in figure 3 :
                       XC-A30 ESR Analyzer Service Manual                              5
                                            Figure 3
      b) Then cut off the band that the blue arrowhead pointing to in the figure 3. Four
        screws need disassembly. As shown in figure 4:
                                             Figure 4
7.3   Main-board disassembly
    Open the cover,
    Separate the main-board's voltage regulator lead-wire with the voltage regulator
components lead-wire that the blue arrowhead pointing to in the figure 5.
     Do not need to remove the main-board holder; the main-board is fixed on the holder
with four screws ( as shown in figure 5)
                       XC-A30 ESR Analyzer Service Manual                              6
                                             Figure 5
7.4    Transformer replacement
      Open the cover.
      Prepare a screwdriver, a plier or M7.0 spanner.
      Method:
        a) There are two screw caps in the bottom board as shown in Figure 6:
                                           Figure 6
         b) There are two screws in the bottom panel as shown in Figure 7:
                                           Figure 7
                         XC-A30 ESR Analyzer Service Manual                     7
         c) Cut off the tape pointed by the black arrowhead in the figure 8 and soldering iron
            the lead-wire. Take off the power plug in the main board and then the voltage
            regulator can be disassembled
                                          Figure 8
7.5     Fuse replacement
      No need to open the cover.
      Take out the fuse with screwdriver. As shown in figure 9
      The fuse specification: F250V/0.5A
                                          Figure 9
7.6     Fan disassembly
   Open the cover, prepare a screwdriver , a plier or M5.5 spanner.
  a) Place the screwdriver on the screw as shown in figure 10:
                          XC-A30 ESR Analyzer Service Manual                                 8
                                    Figure 10
  b) Disassemble the nuts with a plier or a spanner as shown in figure 11, and take off the
     fan..
                                        Figure 11
      Requirement:1)The lead-wire position is better to be as shown in Figure 11 in the
                    assembly of a new fan
                  2)No need to tighten the screws too hard, avoiding the fan distortion.
7.7   Optical detection board disassembly
         Loose the screws as shown in figure 12.
                                    Figure 12
                       XC-A30 ESR Analyzer Service Manual                                9
7.8   Printer disassembly
        Open the cover, pull out the pin of printer data line and power code.
        a) Loose the screws as shown in figure 13, then take off the printer components
                                           Figure 13
       b)   Loose the screw as shown in figure 14.
                                           Figure 14
       c) Separation of printer and power board as shown in figure 15.
                                           Figure 15
                       XC-A30 ESR Analyzer Service Manual                                 10
  7.9    Separation of the optical detection board pin and the socket
            Press the pin and socket by hands as shown in figure 16.
                                            Figure 16
  7.10    Voltage regulator block disassembly
         Loose the screw as shown in figure 17
7805 Voltage
 regulator
7812 Voltage
regulator
                                                 Figure 17
                           XC-A30 ESR Analyzer Service Manual           11
       8、Main board instruction
Voltage regulator
 lead wire
                           J102 printer power code         Printer data line     LCD data line
                                                                                                 J104 main
                                                                                                 testing motor
                                                                                                  2803 integration
                                                                                                  Optical
                                                                                                  detection board
                                                                                                  2803 integration
               Grounding                                                                          J106 RS-232 and
                                                                                                  barcode interface
                                   Transforme        Fan     J301 touch screen     SD2200EP
                             XC-A30 ESR Analyzer Service Manual                                    12