THE PROBE LEVEL MONITOR
THE PROBE
LEVEL MONITOR
 Instruction Manual PL-517   January 2001
33455170
Rev. 2.2
Introduction
    Note:
    The Probe is to be used only in the manner outlined in this
    instruction manual.
    The Probe is an ultrasonic level monitor combining sensor and
    electronics in a single package. It is designed to measure liquid
                                                                  ®
    levels in closed vessels. The process part (sensor) is Tefzel ,
    allowing The Probe to be used in a wide variety of industries. The
    optional sanitary version affords quick removal and ease of cleaning
    as demanded by the food, beverage and pharmaceutical industries.
    The sensor houses the ultrasonic transducer and temperature
    sensing element. The Probe emits a series of ultrasonic pulses from
    the transducer. Each pulse is reflected as an echo from the material
    and sensed by the transducer. The echo is processed by The Probe
    using Milltronic's proven `Sonic Intelligence' techniques. Filtering is
    applied to help discriminate between the true echo from the
    material, and false echoes from acoustical and electrical noises and
    agitator blades in motion. The time for the pulse to travel to the
    material and back is temperature compensated and then converted
    into distance for display, mA output and relay actuation.
    The Probe also offers a digital communications option for data
                                           ®
    collection and remote access using Hart field communication protocol.
Installation
Environmental
    The Probe should be mounted in an area that is within the
    temperature range specified and that is suitable to the housing
    rating and materials of construction. The front lid should be
    accessible to allow programming, wiring and display viewing.
    It is advisable to keep The Probe away from high voltage or current
    runs, contactors and SCR control drives.
Location
      Locate The Probe so that it will       The Probe's sound path should not
         have a clear sound path              intersect the fill path rough walls,
     perpendicular to the liquid surface              seams, rungs, etc
                                                                        fill
                                  sanitary
                                  ferrule
                                             pipe
                                           rungs
                                       seams
Page 2                          The Probe                               PL-517
Mounting
    Note:
    Mount The Probe so that the face of the sensor is at least 25 cm
    above the highest anticipated level.
Threaded
    The Probe is available in three thread types: 2" NPT, 2" BSP or PF2.
                             hinged lid
                                                   7mm               84mm
                                                   (0.3”)            (3.3”)
                                                                     max.
                   117mm
                    (4.6”)
                                cable entry
                                ‘knock out’
                                                  279mm                       electronics
                                                   (11”)
                                 87mm
                                 (3.4”)
                                  Ø                                            mounting
                                                                               thread
                                                            74mm
                                                            (2.9”)             sensor
          61mm                                                       53mm
          (2.4”)                                                     (2.1”)
                                                                     Ø max
    Note:
    Before inserting The Probe into its mounting hole, insure that the
    threads are of the same type to avoid damaging The Probe threads.
Flange Adapter (optional)
    The Probe can be fitted with the optional 75 mm (3") flange adapter
    for mating to 3" ANSI, DIN 65PN10 and JIS 10K3B flanges.
     optional                                 The Probe
      flange
     adapter
PL-517                              The Probe                                 Page 3
Sanitary
    Notes:
    •    Mount The Probe so that the face of the sensor is at least 25
         cm above the highest anticipated level.
    •    The Sanitary Probe 3A approval is for chemical clean-in-place
         applications to 60°C (140°F) only. Ensure your cleaning
                                             ®               ®
         chemicals are compatible with Tefzel or Kynar Flex ,
         depending on your Probe type.
                              hinged lid     7mm              84mm
                                             (0.3”)           (3.3”)
                                                              max.
                  117mm
                   (4.6”)
                            cable entry
                            ‘knock out’
                                                                          electronics
                                            279mm
                                             (11”)
                              87mm
                              (3.4”)                                       mounting
                               Ø                                           thread
                               integral            76mm
                               sealing             (2.9”)                  sensor
                               ring
         61mm                                                 53mm
         (2.4”)                                               (2.1”)
                                                              Ø max
•   mount The Probe onto the top of the tank's sanitary ferrule
•   secure mating by surrounding the joint with the clamp
•   tighten adjusting wing nut
                                                            Sanitary Ferrule,
                                       The Probe
                                                            4” FDA Approved
                                                            (US Food and Drug
                                                              Administration)
                                           clamp*
                                             adjusting
                                             wing nut
                                             ferrule          119 mm (4.68”)
                                                               97 mm (3.83”)
                                                tank
                                                         dimensions are approximate
    *supplied with the Probe
    Note:
    Inside of sanitary ferrule must be smooth, free of burrs, seams or
    ridges.
Page 4                            The Probe                              PL-517
Interconnection
    Notes:
    •    Installation shall only be performed by qualified personnel and
         in accordance with local governing regulations.
    •    Separate cables and conduits may be required to conform to
         standard instrumentation wiring practices, or electrical codes.
     A.     with lid closed, remove cable entry `knock out' on either
     side as required
     B.    open lid by loosening the lid screw
     C.    run cable to The Probe
     D.    connect mA output, power supply and relay wiring7
     E.    close lid. Tighten screw to 1.1 to 1.7 N-m ( 10 to 15 in-lb)
           of torque
    Note:
    Non-metallic enclosure does not provide grounding between conduit
    connections. The use of approved watertight conduit hubs/glands is
    required for IP65 / NEMA 4 / CSA Type 4 (outdoor) applications.
      mA Output: 4 - 20 mA non-isolated
          into 750 ohms max at 24 Vdc
              Power Supply: 18 to 30 V dc
               Connect to protective earth
                 Relay: dry contact, closed
                  unpowered or alarm state,
         refer to specifications on page 19.
                All field wiring must have insulation suitable for at least
                250 V.
                dc terminals shall be supplied from an SELV source in
                accordance with IEC-1010-1 Annex H.
PL-517                              The Probe                          Page 5
Operation
Start Up
•    With The Probe correctly installed (or aimed at a wall 0.25 to 5 m
     away), apply power.
•    The Probe starts up displaying the following:
                             program mode
                                         alphanumeric
           ‘20’ key
                                                                  units
            ‘4’ key
                               LOE / fault
                      operation status
•    It then defaults to the Run mode, which is the measurement reading
     of the distance from the transducer face to the material level in the
     units indicated:
•    If the default display differs form that shown, refer to Operation
     Status on page 7.
Calibration
     The calibration of the mA output may be done such that its span will
     be either proportional or inversely proportional to the material level.
             proportional span               inversely proportional span
             high level = 20mA                    high level = 4 mA
              low level = 4 mA                    low level = 20mA
     Note:
     The 4 and 20 mA levels may be calibrated in any order.
Calibration: Reference Method
     Adjust the material level (or target) to the desired distance from the
     sensor face. Press the “4” or “20” key (as appropriate) to view the
     stored distance associated with that mA output value. Press the key
     a second time to set the new distance reference. After viewing or
     calibrating, Probe operation automatically reverts to the Run mode
     (6 sec).
     The calibration value is referenced from the face of The Probe
     sensor, in the units displayed.
Page 6                           The Probe                            PL-517
4 mA calibration
                                         4 mA
                                         calibration
                                                             calibration invalid
                                         new 4mA
                                         calibration   if
                                                                   retry
20 mA calibration
                                         20 mA
                                         calibration
                                                             calibration invalid
                                         new 20 mA
                                         calibration    if
                                                                   retry
    Note:
    Calibration bypasses the measurement response rate.
Operation Status
    The graphic portion of the display gives the user a visual indication
    of The Probe's operating status. Viewing the graphic can assist the
    user in properly locating and installing The Probe to achieve
    optimum performance.
             GOOD                  WAITING              LOE / FAULT
    The logo will change from full to partial to indicate operation status.
    After the `Waiting' period, the `?' icon will appear for an `LOE /
    FAULT' indication. When a valid echo is again received, a `Good'
    indication will resume. Refer to Troubleshooting on page 15.
PL-517                         The Probe                               Page 7
Adjustments
    There are several operating adjustments that can be made to The
    Probe.
    To access the operating adjustments, simultaneously press the "4"
    and "20" keys until the desired adjustment is obtained. A viewing
    sequence of the stored value is automatically initiated. During this
    time the value can be changed by pressing either the "4" or "20"
    key. After viewing or changing, operation automatically reverts to the
    Run mode (6 sec).
                                                    4 mA calibration,
                                                    scrolling
                                                    20 mA calibration,
                                                    scrolling
                                                    blanking
                                                    speed of response
                                                    alarm
                                                    fail-safe
                                                    fail-safe timer
                                                    units
Page 8                        The Probe                           PL-517
Calibration, Scrolling Method
    The 4 and 20 mA calibration values can be selected where
    reference levels, either from the material in the vessel or from a
    target, cannot be provided. This method can also be used to trim the
    output levels obtained by the Reference Method (see page 6).
    To change the stored calibration value, obtain the `c 4' or `c 20'
    display. The calibration value may be increased by pressing the "20"
    key or decreased by pressing the "4" key. When the display has
    scrolled to the desired value, stop pressing the key. The display
    automatically reverts to the Run mode (6 sec).
4 mA calibration
                                                    4 mA calibration
                                                    initiated
                                                    view stored 4 mA
                                                    calibration value
                                                    i.e. 4.50 m
                                                    press to increase
                                                    to new calibration
                                                    value i.e. 4.60 m
                                                    new calibration
                                                    value
20 mA calibration
                                                    20 mA calibration
                                                    initiated
                                                    view stored 20 mA
                                                    calibration value
                                                    i.e. 0.50 m
                                                    press to decrease
                                                    to new calibration
                                                    value i.e. 0.45 m
                                                    new calibration
                                                    value
    Note:
    For faster scrolling, hold the key depressed during the calibration
    adjustment and release when desired value is obtained.
PL-517                        The Probe                           Page 9
Blanking
    Blanking is used to ignore the zone in front of the transducer where
    false echoes are at a level that interfere with the processing of the
    true echo. It is measured outward from the sensor face. The
    minimum recommended blanking value is 0.25 m (0.82 ft) but can
    be increased in order to extend the blanking.
       The Probe
                                                               blanking
    To change the stored blanking value, obtain the `bL' display. The
    blanking value may be increased by pressing the "20" key or
    subsequently decreased by pressing the "4" key. When the display
    has scrolled to the desired value, stop pressing the key. The display
    automatically returns to the Run mode (6 sec).
                                                    blanking
                                    3 sec
                                                    stored blanking
                                                    value i.e. 0.25 m
                                                    increase blanking
                                                    i.e. 0.36 m
                                                    oops - too far!
                                                    decrease to
                                                    desired blanking
                                                    value i.e. 0.35 m
                                                    new blanking value
                                    6 sec
    Note:
    For faster scrolling, hold the key depressed during the blanking
    adjustment and release when desired value is obtained.
Page 10                       The Probe                            PL-517
Speed of Response
    The speed of response adjustment allows the user to collectively set
    a number of operating parameters.
      measurement        is the limit to which The Probe will be able to
         response:       keep up with rates of change.
                         If The Probe measurement cannot keep up
                         with the rate of level change, set the
                         adjustment from `1' to `2'. If The Probe still
                         cannot keep up with the rate of level change,
                         set the adjustment option to `3'. Avoid choosing
                         an option that is too fast for your application.
            agitator     discriminates between agitator blades in
     discrimination      motion, and the material (target) surface.
                   :
               filter:   discriminates between false echoes from
                         acoustical and electrical noise and the material
                         (target) surface.
            fail-safe    establishes the `Waiting' period from the time a
               timer:    loss of echo or operating fault condition starts
                         until the fail-safe default is effected. Adjusting
                         the speed of response will set the fail-safe
                         timer to the default values in the chart. If a
                         different response is required,adjust the ‘FSt’
                         option (see page 13).
            measurement           agitator                       fail-safe
    SP       response          discrimination        filter        timer
                1 m/min
    1*                                on              on          10 min
              (3.3 ft/min)
                5 m/min
    2                                 on              on          3 min
             (16.4 ft/min)
    3         immediate               off             off         3 min
              0.03 m/min
    4                                 on              on          10 min
              (0.1 ft/min)
    * = factory setting.
    To change the speed of response, obtain the `SP' display. Scroll
    forward through the options (1-2-3) by pressing the "20" key. Scroll
    backward through the options (3-2-1) by pressing the "4" key. When
    the desired option is displayed, stop pressing the key. The display
    will automatically return to the Run mode (6 sec).
                                                      speed of response
                                      3 sec
                                                      current option
                                                      i.e. 1 m/min
                                                      option 2
                                                      i.e. 5 m/min
                                                      option 2
                                                      selected
                                      6 sec
PL-517                          The Probe                            Page 11
Alarm
    The alarm adjustment assigns one of the following functions to the
    relay.
    O = loss of echo / fault alarm (factory setting)
          The relay is energized with the contacts open during normal
          operation (see Fail-safe on page 13).
    Ø = process alarm
          The non zero value entered becomes the alarm setpoint,
          referenced to The Probe's sensor face. The relay de-
          energizes and the contacts close when the material is within
          the alarm zone. There is a hysterisis equivalent to 5% of the
          empty calibration distance.
                         The Probe
                                                               blanking
                alarm
              setpoint
                                                               alarm zone
                  hysterisis
    To change the alarm function or setpoint, obtain the `AL' display.
    The setpoint may be increased by pressing the "20" key or
    subsequently decreased by pressing the "4" key. When the display
    has scrolled to the desired value, stop pressing the key. The display
    automatically returns to the Run mode (6 sec).
                                                       alarm
                                         3 sec
                                                       stored function
                                                       i.e. LOE / fault
                                                       adjust setpoint
                                                       i.e. 1.36 m
                                                       oops - too far!
                                                       decrease to
                                                       desired setpoint
                                                       i.e. 1.35 m
                                                       new alarm value
                                         6 sec
    Note:
    For faster scrolling, hold the key depressed during the setpoint
    adjustment and release when the desired value is obtained.
Page 12                              The Probe                        PL-517
Fail-Safe
    In the event a loss of echo or fault condition exceeds the `Waiting'
    period (see Speed of Response on page 11 or Fail-safe Timer
    below), the ` ? ' icon appears and one of the following fail-safe
    defaults is immediately effected.
                                      p                i
     FLS         default           mA              mA             reading
     1             full             22              4               hold
     2           empty               4              22              hold
     3*           hold             hold            hold             hold
    p=    proportional span
    i=    inversely proportional span
    *=    factory default
    To change the fail-safe default obtain the `FLS' display. Scroll
    forward through the options (1-2-3) by pressing the "20" key. Scroll
    backward through the options (3-2-1) by pressing the "4" key. When
    the desired option is displayed, stop pressing the key. The display
    will automatically return to the Run mode (6 sec).
                                                      fail-safe
                                     3 sec
                                                      current option
                                                      i.e. full
                                                      option 2
                                                      i.e. empty
                                                      option 2
                                                      selected
                                     6 sec
Fail-Safe Timer
    The fail-safe timer allows the user to vary the ‘waiting’ period from
    the time a loss of echo or operating fault condition begins, until the
    fail-safe default is effected. The ‘waiting’ period is adjustable from 1
    to 15 minutes, in 1 minute increments.
    The fail-safe timer value will default to settings determined by the
    speed of response (see page 11). If a different value is desired, the
    fail-safe timer should be adjusted after the speed of response is set.
    To change the fail-safe timer, obtain the ‘FSt’ display. Increase the
    ‘waiting’ period by pressing the “20” key, and decrease it by
    pressing the “4” key, stopping when the desired value is displayed.
    The display automatically reverts to the Run mode (6 sec).
PL-517                         The Probe                           Page 13
    Units
    The units of the measurement reading can be selected as follows:
    1 = metres, m (factory setting)
    2 = feet, ft
    The selected units are also applicable to the `Blanking' and `Alarm'
    adjustments.
    To change the units obtain the `Un' display. Scroll forward through
    the options ( 1 - 2 ) by pressing the "20" key. Scroll backward
    through the options ( 2 - 1 ) by pressing the "4" key. When the
    desired option is displayed, stop pressing the key. The display will
    automatically return to the Run mode (6 sec).
                                                    units
                                      3 sec
                                                    current option
                                                    i.e. m
                                                    option 2
                                                    i.e. ft
                                                    option 2
                                                    selected
                                      6 sec
Hart® Communication
General
    It is assumed that the user is knowledged in the theory and
                            ®
    application of the HART field communication protocol. If not, contact
                                           ®
    the HART Communication Foundation . ( 512-794-0369 U.S.A.)
    Expanded Device Type Code:
•   manufacturer identification code = 88
•   manufacturer device type code = 200
•   expanded device type code = 22728
    Physical Layer Information :
•   field device category = C
•   capacitance number (CN) = 1
    Miscellaneous :
•   multidrop mode is supported
•   burst mode is not supported
Online Menu
              ®
    The HART Probe communication online menu in the Supplement
    section is configured for a Rosemount model 275 hand held
    communicator programmed with the Milltronics device description.
Page 14                       The Probe                          PL-517
Supplement
Loading vs. Supply Voltage
Troubleshooting
          The echo is not reliable and The Probe is
          waiting for a valid echo before updating the
          measurement.
          Probable causes are:
          •   material or object in contact with sensor
              face
          •   The Probe is too close to the fill point
          •   The Probe is not perpendicular to the liquid
              surface
          •   change in level too fast
          •   measurement out of range
          •   foam on liquid surface
          •   high level of vibration in the mounting
              structure
          •   level inside the blanking zone
          The `Waiting' period has expired. Investigate
          the probable causes listed above.
          Refer to Speed of Response on page 11 or Fail-
          safe Timer on page 13 for duration of `Waiting'
          periods.
PL-517               The Probe                            Page 15
Hart ® Communication Online Menu
    Universal Commands Supported:
    0, 1, 2, 3, 6, 11, 12, 13, 14, 15, 16, 17, 18, 19.
    Common Practice Commands Supported:
    35, 36, 37, 38, 40, 41, 42, 44, 45, 46, 48, 49, 59.
    Transmitter Specific Commands Supported:
    128 / 129       read / write      materials of construction
    130 / 131       read / write      blanking distance
    132 / 133       read / write      speed of response
    134 / 135       read / write      alarm selection
    136 / 137       read / write      fail-safe alarm level
       Not available with generic device description.
Page 16                            The Probe                      PL-517
PL-517   The Probe   Page 17
Maintenance
    The Probe requires no maintenance or cleaning.
Tefzel®
          ®
    Tefzel is a fluoropolymer inert to most chemicals. For exposure to
    specific environments, check with chemical compatibility charts
    before installing and operating The Probe in your application.
          ®
    Tefzel is a registered trade mark of DuPont.
Patents
    Instrument Housing Design:
•   Canada:          70345
•   U.S.A.:          07/858/707
•   Germany:         M92022723
•   U.K.:            2021748
•   France:          921873
•   Japan:           966217
    Electronics / Sensor:
•   U.S.A.:           5,267,219
                      5,339,292
•   U.K.:             2,260,059
•   patent applications in U.K., Canada, Europe, Africa, Australia
Page 18                       The Probe                          PL-517
Specifications
Power:
  { 18 to 30 V dc, 0.2 A max
Environmental:
  { location:            { indoor / outdoor
  { altitude:            { 2000 m max.
  {ambient               { continuous: -40 to 60°C (-40* to 140°F)
    temperature                          -20°C (-5°F) if metal mounting
  { relative humidity:   { suitable for outdoor (Type 4X / NEMA 4X / IP65
                          enclosure)
   { installation category: { II
   { pollution degree: { 4
   { process pressure { vented to atmosphere
Range:
  { series 861:           { 0.25 to 5 m ( 0.8 to 16.4 ft )(liquids only)
  { series 862:           { 0.25 to 8 m ( 0.8 to 26.2 ft )
Beam Angle:
  { 10° at -3 dB boundary
Memory:
  { non-volatile EEPROM, no battery required
Programming:
   { 2 tactile keys
Temperature Compensation:
  { built-in to compensate over the operating range.
Display:
   { liquid crystal
   { three 9 mm (0.35") digits for reading of distance between sensor face and
     material
   { multisegment graphic for operation status
Output
  { mA:                  { range:       4 – 20 mA
                         { span:        proportional or inversely proportional
                         { accuracy: 0.25% of full scale
                         { resolution: 3 mm (0.125")
                         { loading:     750 ohms max at 24 V dc supply
   { Relay:              { 1 normally closed contact rated at 5 A at 2 50 V ac non-
                          inductive or 24 V dc
                    { fault on power, application or device failure
   { Communication: { HART® protocol 5.2 (optional)
Construction:
  { combined sensor and electronics package
  { sensor housing: { material:      { Tefzel® or Kynar-Flex ®
                     { mounting:
                       { threaded: { 2" NPT, 2" BSP PF2
                         { flanged: { flange adapter, threaded Probe to 3"
                                       ANSI, DIN 65PN10 and JIS 10K3B
                         { sanitary: { 4" FDA approved sanitary ferrule with
                           (series     integral sealing ring c/w 304 stainless
                           861 only) steel clamp
  { electronics      { material:     { PVC
    housing:         { access:       { hinged lid
                                     { 22 mm (0.87") dia. `knock out' for
                                       conduit entrance, 2 places
                                     { 6 screw terminal block for 2.5 mm 2
                                                                   2
                                       (14 ga) solid wire / 1.5 mm (16 ga)
                                       stranded wire max
Enclosure Rating:
  { IP 65 / NEMA 4 / CSA Type 4
Weight:
  { 1.7 Kg (3.7 lb)
Approvals:
  { CE*, FM CSA NRTL/C
    * EMC performance available on request.
PL-517                               The Probe                             Page 19