Statfax 4200
Statfax 4200
MICROPLATE READER
OPERATOR’S MANUAL
© Awareness Technology Inc. 2008 Stat Fax® 4200 Operator’s Manual Revised 2018 Rev. K
TABLE OF CONTENTS
1. Introduction ...................................................................................................... 1
1.1 Applications .............................................................................................................................. 1
1.1.1 Intended Use ..................................................................................................................... 1
1.1.2 Summary of the Instrument ............................................................................................... 1
1.1.3 Principles of Operation ...................................................................................................... 2
1.2 Warning Markings Inscriptions d’avertissement ...................................................................... 3
1.2.1 Safety Symbols Le Symboles de Sûreté ........................................................................... 3
1.2.2 Safety Terms Terminologie de Sûreté .............................................................................. 3
1.3 Safety Precautions ................................................................................................................... 4
1.3.1 Disposal ............................................................................................................................. 5
1.4 Specifications ........................................................................................................................... 6
1.5 Installation ................................................................................................................................ 8
1.5.1 General .............................................................................................................................. 8
1.5.2 Installation/Preparation...................................................................................................... 8
1.5.3 Load Paper ........................................................................................................................ 9
1.5.4 Touch Screen Description ............................................................................................... 10
1.6 Parts and Controls ................................................................................................................. 10
1.7 Checkout Procedure .............................................................................................................. 11
4. Troubleshooting............................................................................................. 52
4.1 Messages and Flags .............................................................................................................. 52
4.2 Error Messages ...................................................................................................................... 55
5. References...................................................................................................... 56
7. Information ..................................................................................................... 58
1.1 Applications
The Stat Fax® 4200 may be used for in-vitro diagnostics. It is a user-programmable open
system with selectable plate formatting, alphanumeric test naming, automatic interpretation
options, duplicate well options, curve plotting and editing, flags and error messages from
which the user can create a menu of pre-programmed assays for instant recall.
Stat Fax® 4200 is intended to be used as a stand-alone instrument without a PC but it has
a connection in the event a user wishes to download results. For a PC-controlled
microplate reader ask about the ChroMate TM Model 4300.
• Six-filter VIS model (405, 450, 492, 545, 600, and 630 nm)
Substitute filters in the range of 405 nm - 750 nm are available on special order models.
The Stat Fax® 4200 accepts all standard microplates. Trays containing microstrips may also
be read whether they are 12 wells in length, 8 wells or even partial strips.
User-programmable Memory
The Stat Fax® 4200 Microplate Reader’s software allows the operator to create, edit and
store assays. Standard curves are also stored in this memory. Test protocols remain
stored until either changed or deleted by the user.
Light from a lamp installed in a sealed light-tight compartment is filtered using a rotating
filter wheel. The filtered light is then divided into 8 channels using a fiber optic bundle.
The plate carrier precisely positions each row of 8 wells into the optical path for reading.
Light passing through the 8 wells is directed to 8 photo-detectors.
The filter wheel moves to a second position to read the 8 wells at a second wavelength.
Using bichromatic differential absorbance values corrects for optical imperfections in the
plastic wells and removes the effects of meniscus and turbidity.
The photo-detectors generate electronic signals which are amplified and interpreted by the
instrument circuitry.
DANGER Le “de marque: Indique le risque immédiat de dommages (assessible tandis que vous
DANGER” lisez la marque)
1.3.1 Disposal
Dispose of according to local regulations.
Before the instrument is removed from the laboratory for disposal or servicing, it must be
decontaminated.
CAUTION! L’ATTENTION!
Caution: Treat all components during use and disposal as you would any
biohazardous material.
Photometric Accuracy: ............................................... +/- (1% +/- 0.010) 0.0 through 1.5 ABS
Standard Wavelengths: .............................................. 405, 450, 492, and 630nm. 6-filter: 405,
450, 492, 545, 600, and 630nm. Filter
options from 405nm to 700nm.
Electronic
Other
1.5.1 General
Carefully unpack the instrument, removing it from its plastic bag. Report any damage to
your freight carrier at once. Retain the original packing material for future use in the event
that the instrument is shipped to another location or returned for service. Items packed with
the instrument include:
1.5.2 Installation/Preparation
Instrument Placement Place the instrument on a flat working surface capable of safely
and Use supporting the weight of the instrument, approximately 5.4 kg (12lbs).
A clearance of at least 8cm (3”) around the instrument is required to
assure optimal ventilation.
Assure Clean Power The circuit used should be substantially free of large voltage
Availability transients (Kilovolt amp loads) such as large pumps, large
centrifuges, refrigerators and freezers, air conditioners, large
autoclaves, ovens, and dryers. The instrument may fail to operate
normally if the power supply is interrupted. If this occurs, turn the
instrument off for a moment. When the instrument is turned back on,
it will resume normal operation, but data that was not stored in
nonvolatile memory will be lost. If power fluctuations or loss are
frequent, an Uninterruptible Power Supply (UPS) is recommended.
Power Switch Position The ON/OFF switch is on the back of the instrument. The power
switch should be in the OFF position (down) before connecting the
power cord to the power supply.
For 110-120 V used in the United States, use a UL listed cord set
consisting of an 18 AWG, Type SPT-1 two conductor cord
maximum 3 meters (10 feet) in length, rated 7 A, 125 V, with a
polarized parallel blade type attachment plug.
For 220-240 V used inside the United States, use a UL listed cord as
above, except rated 250V.
For other locations, use the power cord certified for the country of
use.
Printer paper
Figure 1.5.3-2 Printer paper installed
compartment
release button Pull up at least 1 inch of paper and then
A press the compartment cover down until it
snaps closed. Turn the power switch OFF
and back ON. The wake up routine will
follow. The printer will print several lines.
Figure 1.5.3-1 Paper compartment lid
Locate the roll of thermal printer paper P/N Wait until it has stopped. If there is no
150003. printing, the internal printer is disabled
(reference Section 2.3.2 Settings).
Press down on the printer paper compartment
release button (A) and the paper compartment lid NOTE: The printer manufacturer strongly
will open. recommends the use of thermal paper
part number 150003 to reduce lint and
Place the paper roll in the well such that the
extend printer life. Contact your dealer for
leading edge of the paper feeds upward from the
replacement rolls.
front of the printer as shown.
The following terms are used in this manual to describe parts and controls of the Stat Fax® 4200
microplate reader. More details on the operation of each feature are provided in Section 2-
Operating Procedures.
A) Door open
B
B) Printer and printer
paper compartment C
C) Display
Checkout Procedure
Visually confirm the following items:
√ Power supply is connected to the unit and the power plug is inserted into an AC
outlet.
√ Power switch is set to OFF.
The instrument is now ready for power-up. Confirm that the instrument responds as
described.
√ Set the power switch to the ON position.
The printer will print out the
instrument model, serial
AWARENESS
number, laboratory name (if TECHNOLOGY
defined under the Settings INC
menu Laboratory Name option;
the default is no laboratory Stat Fax
name), firmware version,
revision level, date and time 4200
(Figure 1.7-2).
Loading …
Click Tray Out and make sure
Figure 1.7-1 Power up Screen Display
door opens and microplate tray
extends out.
• Be sure to run a sufficient number of controls in each assay. If controls are not within
their acceptable limits or if incomplete, disregard test results.
Review the following questions before beginning a test (refer to the reagent kit insert for required
information):
• Tray In – the plate carrier transports the plate into the and the door closes
• Tray Out – the door opens and the plate carrier extends out for the ease of
inserting a test plate on the carrier
The status line displays the Lamp status (OFF or ON) and the time. The printer button (PRN) has
two options: 1. to advance the paper, 2. to reprint the last 64 lines of data. NOTE: PRN does not
print graphs, only data.
Select Run Test from the Main display. Tests programmed and saved on the instrument
will display sorted by their test number. Each of the test parameters, including the mode,
wavelengths, standards, units, and the ranges are all stored for reuse.
Blanks and standards (including entire standard curves) that have been read are also
saved. When the test is recalled, the user has the option of using the previous saved
curve or reading a new one.
1 – TSH
Cancel
2 – T3
SAMPLE
3 – T3 Uptake
4 – T4 <<
5 – T4 Total
6 – FSH v >>
7 – hCG
Select
8 – LH
9 – Testosterone
By #
Feature Function
Arrows Use the ^ UP and V DOWN arrow keys to highlight
selection; use the >> side arrows to advance to
the next screen; use << arrows for previous
screen
Select The highlighted test is executed once the Select
key is pressed
By # Allows the user to enter a specific test number to
recall.
Cancel Returns to the main power on display screen
Unit Settings
Arrow Keys: Utilize the
Adjust Date and Time ^ arrow keys to scroll up and
down the list of options or
Printer Setup press the option directly
v from the list; the option will
Lamp Control
highlight
Plate Assignment Select
Sound Settings
Done Select: The high-lighted
Laboratory Name
option is executed once the
External Output Select key is pressed
Feature Function
Adjust Date and Provides access to the Set Time option Hours, Minutes, Seconds
Time as well as Month, Day, Year
Printer Setup Enables/disables the internal printer. Graphs can be printed when
the internal printer is set to OFF. Allows the contrast to be
adjusted light to dark. It is advisable to print at contrast level set at
3. Higher t levels may degrade after several lengthy printouts.
Text height options are small, medium and large.
Lamp Control Lamp idle feature will automatically turn off the lamp; the default
setting is 600 seconds (10 minutes). Lamp Warm up range is
between 45 and 240 seconds (45 seconds is the default setting).
Plate Assignment Offers plate format options such as 8 way or 12 way; duplicates
side by side.
Sound Settings Provides access to and status of touch screen sound settings (ON
or OFF), and sound volume (1=low to 8=high). Sound options are:
start up sound, key press & release and other sounds (i.e. sound
when the Quit button is selected).
Laboratory Name Allows a laboratory named to be entered; laboratory name will
print out (reference Checkout Procedure in Section 1.8).
External Output Allows external output to be turned ON or OFF. Reference the
External Output Section 2.3.2.1 for setup instructions.
Plate Print Format User can select either a 1 to 12 or H to A printout of plate data.
• Power on the instrument and set the External Output to “ON” by: (a) selecting
Settings from the main display, (b) highlight External Output on the Unit Settings
display, (c) press the Select button, (d) on the External Output Setup display
select the ON button and ensure that the Status field displays ON, (e) select the
Save button, and then, (f) select the Done button.
• Connect the type B end of the USB cable to the type B port of the
instrument.
• With the personal computer (PC) powered on, connect the type A end of the USB
cable to a USB port on the PC.
Type A
Select Manage Tests on the Main display screen and the following options will display:
Delete Test
Feature Function
Create Test Allows user to:
• Select Axes
• Controls management (Enable; name; number of replicates;
input low/high ranges; input action to take (i.e. warn, continue
or end test); record Lot Number; input expiration date; location).
2.3.3.1 Multi-Test
With the Multi-Test option, any stored assay regardless of mode or wavelength may
be used. The read direction (8 or 12-way) will be the instrument was set to read in the
Settings menu. With the 8-way format, two, three or six different test modes may be
run in the same plate; with the12-way format up to four tests may be run.
If any test is selected to run in the multi-test plate that has a stored curve, the user
will be prompted to accept, print or discard the curve.
The instrument will prompt the user to select the next test. To read less than a full
plate, select “Cancel” after choosing the last test to be run.
Print Plate will enable the location and absorbance reading of each well to be printed,
as shown on the plate map.
In order to print the test information and results, after each test is run, select
“Results”.
If all tests in the multi-test plate have the same filter configuration, then the plate is
read only one time. The user can select “Print Plate” and “Results” for each test.
If different filters are used for any other test(s), the plate will automatically be read
again.
2) Select “Read Plate” and the first test will run. When that test is done, use the
“Results” button to printout out data. (Print Plate can be used to verify well and
well absorbance reading location.)
2) Reading of the Standards in the second test, TSH, results in an invalid curve.
3) The next test in the plate will not run. Multi-test mode is ended.
Utility Menu
Change Access Level Arrow Keys: Utilize the arrow
^ keys to scroll up and down
Show Version Details
the list of options or press the
option directly from the list;
Erase ALL Tests v the option will be highlighted.
Feature Function
Change Access Level Display prompts to ‘Enter Pass Code’. Contact
Technical Support for pass code.
Show Version Details Displays current firmware version, model, serial
number, build date, current time.
Erase All Tests Display will prompt “Please Confirm. Erase All Tests?”
press OK to clear all tests or Cancel.
Recall Calibration Allows user to restore the original factory settings,
including calibration.
Print Unit Settings Prints all information about an individual instrument’s
calibration settings and other information.
Filter Voltages Displays real time filter wheel voltages; provides
option to print them out. Acceptable range is from 2 to
10 v.
Check Wheel Speed Displays results as the wheel speed is tested.
Do not place the tabbed ends of strips in row 1; they should be in row 12. Be sure to place
the strips in the order in which Blanks, Calibrators and Samples are to be read.
When using a Blank, it should always be located in Well A1. If replicate blanks are used,
they will be placed on the plate map as the next well from A1, to B1 if reading 8 way strips,
or below if reading 12 way strips.
For best results, do not fill wells completely; 200-250µl depending on well total volume is
the maximum fill recommended.
Use caution when attaching labels to plates so they do not jam in reader or interfere with
read path.
The microplate may be formatted for 8 way or 12 way placement of Blanks, Controls and
Samples. Selecting 8 way or 12 way is done on the Settings screen, Plate Alignment
option.
The default location is the well or wells adjacent to the Blank(s). To use a different
location, while enabling the use of Controls, select the Location field (Figure 2.4.2.1-
1).
Figure 2.4.2.1-1
In the example below, a cutoff mode assay is being run and strips are to be read in
the 8 way read direction. Duplicate Positive Controls (PC) and a single Negative
Controls (NC) are being used in the assay (Figure 2.4.2.1-2).
The Negative control is to be read in well A5. This is done by selecting that box on
the screen. The new location is presented in the Location box. .
H G F E D C B A H G F E D C B A
NC PC PC 1 PC PC 1
2 2
3 3
4 4
5 Location NC 5 Location
6 6 A5
7 7
8 Select 8 Select
9 9
10 10
11 Default 11 Default
12 Cancel 12 Cancel
When using replicates, only the location of one Control is to be selected. The
replicates will follow according to the Plate Alignment chosen in Settings.
NOTE: If the
H G F E D C B A
Controls or their NC 1
Replicates overlap 2
due to the plate 3
setup, an error flag 4
will display to warn PC PC 5 Location
the operator: 6 A5
7
“Enabled control 8 Select
location already 9
identified!” 10
11 Default
12 Cancel
Figure 2.4.2.1-4
The example below (Figure 2.4.3.-1) shows the layout of a Point to Point mode assay that
has duplicate Blanks, 5 Standards, using all three Controls read in duplicate.
H G F E D C B A P1
S6 S5 S4 S3 S2 S1 B2 B1 1
2 1 C3 C3 C2 C2 C1 C1 2
10 9 8 7 6 5 4 3 3
4
5
6
7
8
9
10
11
12
QUIT # Samples Read Plate
Figure 2.4.3-1
If fewer than a full plate of wells is to be read, select the ‘#Samples’ button. Once the
plate has been read (via Read Plate), absorbance readings will be on the display for each
well.
There is an option of ‘Print Plate’ to print out the plate as seen on the display. This is
especially useful when running Absorbance mode.
Select the Results button and the test results will be printed. The message “Run another
plate of samples” will appear. The user can choose to read more or to Quit.
Conc
IU/mL
e
g/dL
mmol/L
umol/L
ppm
ppb
Customize
TEST DEFINITION
Name Exit
Mode Absorbance
^
Offset Absorbance feature is only needed when both of these conditions occur:
Monochromatic reading
No blank well
It is a mathematical blank to correct for the differential between air and liquid with a
meniscus.
To determine the user-entered value, pipette a volume of wash buffer into a well that is
equal to the total volume in test wells at the read time.
For example, if the well has 100µl of Substrate Part A per 100µl of Substrate Part B, plus
50µl of stop solution, the total volume in the test wells at the read time is 250µl. Therefore
read the absorbance of 250µl of colorless wash buffer to determine the correct value for
the Offset Absorbance. Typical values are from .250 to .400A.
Lamp Warmup has a default setting of 45 seconds. This can also be adjusted from the
Lamp Control feature. The range is 45 to 240 seconds.
Mode Function
Absorbance See Section 2.5.1 - Absorbance mode reads and prints the
monochromatic or bichromatic differential absorbance at the user-
selected wavelengths. Blanking is optional. Most Assays require a mode
other than Absorbance Mode. In this mode, no calculations are made -
only absorbance values are reported.
Factor See Section 2.5.2 – In Factor Mode, the endpoint absorbance readings
will be multiplied by a user-entered factor to calculate a result.
Single Standard See Section 2.5.3 – Single Standard mode reads a calibrator, and then
calculates concentrations based on a single-point standard curve
passing through the point (0,0). A blank is required to determine the
(0,0) point. A factor (equal to the concentration of calibrator ÷ the
absorbance of calibrator) is generated in this mode, and then multiplied
by subsequent absorbance readings to determine concentrations.
Point to Point See Section 2.5.4 – The plate reader accepts a number of calibrators
and calculates concentrations based on the point-to-point calibration
curve. The resulting calibrator curve is a series of lines connecting the
calibrator points, which may be entered in ascending or descending
order of absorbance. The direction of slope between the first and
second calibrators determines the direction of the curve. If the direction
of the curve changes direction, the curve will be flagged as being
“invalid” and no interpretations will be printed.
Unknown samples are calculated as follows:
The unknown sample’s absorbance is read and compared to the
absorbance of the calibrator.
An unknown Sample, with absorbance higher than the calibrator with
the highest absorbance value, is calculated using a line that passes
through the two Calibrator points with the highest absorbance value. An
unknown Sample with absorbance lower than the lowest Calibrator
absorbance is calculated from the line that passes through the two
Calibrators with the lowest absorbance values.
Mode options are located under Manage Tests on the Main display screen. Select Create
Test and the Test Definition screen will display; select the Mode field and the list of Mode
options will display.
Each of the seven (7) modes follow the same basic pattern of options, with other variables
available for selection dependent upon the mode being used.
SELECT MODE
Absorbance Cancel
Factor
Single Standard ^
Point to Point
Regression v
Cubic Spline
Cut Off Select
Available Modes
CAUTION!
CAUTION! Verify mode selection matches with the test kit package
insert or manufacturer’s instructions before reporting test results! Always
include normal and abnormal controls.
Absorbance mode will read and print sample absorbance values at user selected
wavelengths. All of the other modes that follow function the same way but use the
measured absorbance to calculate final results. Press Manage Tests, press Create Test,
and the Test Definition window will display.
TEST DEFINITION
Name Exit
Mode Absorbance
^
SELECT FILTERS
SELECT FILTERS None Cancel
405 Cancel 405 ^
450 ^ 450
492 v 492 v
630 Select 630 Select
If Differential Filter is set to "None", then the Offset Absorbance field is displayed.
To operate in Factor Mode, press Manage Tests and then press Create Test. A Test
Definition window will display. Press the Mode field and the Select Mode screen will
display. Use the arrow keys to advance through the list to the desired mode and press
Select or use the stylus to highlight Factor.
SELECT MODE
Absorbance Cancel
Factor
Single Standard ^
Point to Point
Regression v
Cubic Spline
Cut Off Select
NOTE: The Factor comes from the assay package insert or from a previous assay in the
Single Standard Mode.
Selecting any of the Control fields will open a Test Definition Control screen with options to
enable the Control (Yes or No), input the number of Replicates, Name the Control, input a
Low and a High Range, set the Action to Take (Warn, Continue or End Test), input a Lot
Number, input an Expiration Date, and input a Location. Press the Done button when
finished with editing fields.
Interpretation Normal ^
Normal Range Low v
Normal Range High
Valid Range Low Edit
Valid Range High Done
SELECT MODE
Absorbance Cancel
Factor
Single Standard ^
Point to Point
Regression v
Cubic Spline
Cut Off Select
TEST DEFINITION
Name Exit
Mode Single Standard
^
When Controls are read in duplicate, the mean of the duplicates can be calculated and that
one value used, or, each Control can be required to fall within the ranges individually. The
instrument will prompt ‘Check Mean?’ once duplicate Controls are selected.
The Single Standard Mode Test Definition Final page will display and allow editing of
Interpretation Mode, Units, Decimals, and number of Sample Replicates. Use the arrow
keys to advance through the options; press the Edit button to change the values of the
selections on the screen. Press Save to save the selections.
In Point to Point and Cubic Spline modes, the standard concentrations must be read in
either increasing or decreasing order.
All of the Test Definition screens, use of Controls, and items related to Interpretations work
the same for each type of multi-point. Use the arrow keys to scroll through the mode
selections to the desired selection and press Select or use the stylus to highlight
selections.
MULTI-POINT
MODES
TEST DEFINITION
Name Exit
Mode Point to Point
^
The Test Definition 2nd page includes use of % Absorbance, number of Standards,
number of Standard replicates, selection of Axes, and the use of Controls.
Note: Point-to-Point and Regression require at least two Standards, Cubic Spline mode
requires at least three.
% Absorbance assigns a value of 100% to the first Standard, which must be entered from
darkest to lightest.
The %Abs value is displayed in the Interpretation field of the report. This mode does not
support ln, logit or log scales for the absorbance axis. Upon recall of a % Absorbance
mode test, the user will be asked whether to reread the highest value Standard.
Select the Standards field and the Standards Test Definition screen will open.
Standards Exit
Number of Standards 3 ^
Std1 Concentration 10.0 v
Std2 Concentration 20.0
Std3 Concentration 30.0 Edit
Done
Once the curve has been read, select the ‘Edit’ option before the ‘Accept’ or ‘Print’
options. Using the stylus, select the duplicate standard to be deleted.
As shown in the example below, Duplicate Standards are labeled as ‘Dupl I’ and ‘Dupl II’.
SELECT AXES
Y=Abs X=Conc Cancel
Y=Ln(1000*Abs) X=Conc ^
Y=Abs X=Ln(Conc) v
Y=Ln(1000*Abs) X=Ln(Conc) Select
Y=Logit(Abs) X=Log(Conc)
Calculation Explanation
“Y= ABS, X=CONC” Both the absorbance data (y) and the
concentration data (x) are linear.
“Y=Ln (1000*ABS), X=Conc” The natural log of the absorbance is plotted
against the concentration. The absorbance
values are multiplied by 1000 before taking the
logs.
“Y=ABS, X=Ln(Conc)” Absorbance is plotted against the natural log of
the concentration.
“Y= Ln (1000*ABS), X=Ln(Conc)” The natural log of the absorbance is plotted
against the natural log of the concentration.
“Y= Logit(Abs), X= Log(Conc)” Select to calculate unknowns using the
equation: Abs Logit = Ln [(sample/0 cal) / 1-
(sample/0 cal)]
Refer carefully to the Reagent Kit product insert for correct selection.
Point-to-Point and Regression modes require at least two standards, Cubic Spline
requires at least three standards.
When Controls are read in duplicate, the mean of the duplicates can be calculated
and that one value used, or, each Control can be required to fall within the ranges
individually. The instrument will prompt ‘Check Mean?’ once duplicate Controls are
selected.
Interpretation Normal ^
Normal Range Low v
Normal Range High
Valid Range Low Edit
Valid Range High Done
Test Definition –Normal Interpretation
In the Cut Off Mode, ‘X ‘is multiplied by the mean of the negative controls. ‘Y’ is multiplied
by the mean of the positive controls, and each of these two numbers is added to ‘F’ to
arrive at the Cut Off Value (COV). The ‘X’, ‘Y’ and ‘F’ factors can be positive or negative
numbers, one or zero, or decimals such as 0.5 to fit the test requirements provided in the
Kit package insert or as instructed by the Kit manufacturer. In this mode be careful to enter
all QC parameters completely and correctly. They are usually provided as Absorbance
limits.
To operate in Cut Off Mode, select Manage Tests and then select Create Test. The Test
Definition screen will display. Select the Mode field and the Select Mode screen will display
the list of modes. Use the arrow keys to scroll through the mode selections to the desired
selection and press Select or use the stylus to highlight Cut Off.
SELECT MODE
Absorbance Cancel
Factor
Single Standard ^
Point to Point
Regression v
Cubic Spline
Cut Off Select
TEST DEFINITION
Name Exit
Mode Cut Off
^
When Cutoff Controls are read in multiples, the mean of the replicates
can be calculated and that value used, or each Control can be required
to fall within the ranges set. “Check Mean” will display once replicate
Controls are selected.
mean
Note: When setting the Controls, if there are Replicates, the Location will be the starting
location. If the Controls or their Replicates overlap due to the plate setup, an error flag will
display while running the test to warn the operator. (Reference Section 2.4.2.1)
Selecting the ‘COV =’ field will open the Select COV Formula screen. In addition to the
universal equation, Stat Fax® 4200 offers three more optional formulas.
Where COV is the Cutoff Absorbance Value, mNC is the mean of the Negative Controls;
mPC is the mean of the Positive Controls. X, Y, and F are user-entered coefficients that
can have any positive or negative numerical value including zero and 1.
Highlight the desired formula from the list and press the Select button.
(mPC / mNC):
Enter the lowest Backup to be used in case
acceptable ratio for controls read are too high or
the mean of the TEST DEFINITION Cutoff Quality too low.
positive controls to ^
the mean of the Minimum COV v
negative controls Maximum COV
Note: ‘Reversed’ field is ‘YES’
Edit
when Low Absorbances are
(mPC/mNC)>= Done
(mPC – mNC): Enter Positive and High Absorbances
(mPC-mNC)>=
the lowest are Negative.
acceptable difference Reversed
for the mean
absorbance of the
positive controls
minus the mean
absorbance of the
negative controls.
The Blank is not subtracted in the Reverse Cutoff option. Pay close attention to “<” less
than and “>” greater than symbols when entering control limits in this case.
The Cut Off Mode Test Definition final page displays Interpretation Mode, Units, Decimals,
and number of Sample Replicates.
3.1 Cleaning
3.1.1 Exterior
CAUTION: Solvents such as acetone or thinner will damage the instrument! Use only water
and recommended cleaners! Avoid abrasive cleaners. The display area is liquid-resistant,
but is easily scratched.
The exterior of the instrument may be cleaned with a soft cloth using plain water. If
needed, a mild all-purpose (nonabrasive) cleaner may be used. A 10% solution of
chlorine bleach (5.25% Sodium Hypochlorite) or 70% isopropyl alcohol may be used as a
disinfectant. Take special care not to spill any liquid into the read well.
3.2 Maintenance
Absolute calibration can be verified using a specifically designed product called DRI-DYE®
Check Strips, available from your instrument supplier.
The best way to assure quality instrument performance is to include a sufficient number of
controls in each assay to cover the entire operational range.
In vertical photometry, the fill volume and degree of meniscus determine the pathlength; and
absorbance is proportional to pathlength. Since lab test results are typically based upon
standards rather than upon absolute absorbances, the linearity of the instrument is the more
critical indicator of instrument performance.
Alternately, the degree of linearity can be checked by reading a 1:2 serial dilution of a
material having a peak absorbance at or near one of the wavelengths, and observing a 1:2
ratio in the absorbance readings.
If you prepare dilutions to verify linearity, uniform pipetting (pathlength) and minimal transfer
error are important, since the tolerance limits which you will establish must take into
consideration the sources of error that are not due to the instrument.
For example, if the 1/4 dilution reads .520 A, then you can expect your 1/2 dilution sample to
read twice as much or 1.04 A ± [(1% of 1.04) +0.01 A], which is ± .02.
A good working instrument would then give a result between 1.02 A and 1.06 A. You will
know that your filters are deteriorating when darkly colored solutions consistently read
lower than expected.
Since the procedures for verifying the calibration and linearity of a vertical photometer are
quite cumbersome and not conducive to stringent criteria, the best way to assure quality
instrument performance is with DRI-DYE® Check Strips.
3.2.2 Storage
The instrument may be stored under the following recommended environmental conditions:
• Temperature: 10 to 50°C
Flags are displayed, for example, when an invalid range is entered. They are intended to help the
user correct the problem. Some error messages appear as dialog boxes that require action in
order to clear; others appear on the display screen and appear on the printout.
Message/Flag Explanation
>10**6” Appears in Concentration Column when Concentration is
greater than 10**6
“******” 1. A sample has exceeded the absorbance of the reference
well in a log-logit test. Curve Axes is Y=Logit(Abs),
X=Log(Conc) and Absorbance >reference(zero) absorbance
value.
2. Will print if a curve with a log axis tries to log a zero, a
negative number, or an invalid curve has been read.
3. Prints in Concentration column whenever an error occurs
while calculating the concentration.
“>3” Prints in ABS column if a read absorbance is greater than
the maximum readable absorbance for the device.
“%Absorbance Mode This message will display if the user is in %Absorbance
should not be used with Mode and a standard is read which has a negative
negative absorbances!” absorbance.
“Assay Cannot Be User has tried to edit an OEM preprogrammed test that has
Edited!” flag set to prevent it from being edited.
“Cannot match test Pop up message that appears when running a test; occurs if
wavelength with filters the device cannot find filters for the wavelengths set in the
currently installed on test definition.
device!”
“Do not select the same Appears on Page 1 of Test Definition when the same
filter for primary and wavelengths are entered for the Primary and the Differential
differential” filters.
Message/Flag Explanation
Message/Flag Explanation
“Pos must be greater than or This message appears during test definition when the
equal to Neg!” user is defining positive/negative interpretations. A
value was entered for both the positive and negative,
but the positive value is less than the negative value
entered.
“Range Error” This indicates there is a limit to the range of numbers
that can be entered
“>Ref” Prints when a sample exceeds the absorbance of the
reference well in a log-logit test.. This is usually due to
the calibrator concentrations being entered incorrectly
or the calibrators being pipetted into the strip
incorrectly. The first, or "Zero" calibrator must have the
highest absorbance (darker) and the remaining
calibrators must be of descending absorbance and
increasing concentration.
Test Data Corrupted! Test (assay) Data (including calibration data) does not
match what was stored.
OR
OR
Test Definition Corrupted!
Test Definition does not match the data stored.
Turn the instrument off, then back on and read a plate to see
the message has been eliminated
OR
FTDI initialization error This message will be seen if the FTDI startup
initialization fails, tries again, but then fails a second
=========================== time.
Hardware Init Error (#10) =======================================
5. References
1. Engineering data supplied by Awareness Technology, Inc. Palm City, Florida (1987 to
current date)
2. Data on DRI-DYE® Check Strips, provided by Awareness Technology, Inc. Palm City,
Florida (1989 to current date)
The charts provided with these kits are designed to replace the normal calculations with easy
to visualize, built-in acceptance ranges. Kits are available for the following wavelengths:
®
o DRI-DYE Check Strips-405 - for 405nm
®
o DRI-DYE Check Strips-450 - for 450nm
®
o DRI-DYE Check Strips-492 - for 492nm
Factory Information:
Mailing Address:
Important: When contacting us, please have the Model and Serial Number of the Stat Fax® 4200 in
question. Have a description of the problem with as much detail as possible. Save any relevant jobs
or logs and send or e-mail us the information.
Model:__________________
Serial Number:_____________
Install the program by double clicking on the setup.exe file found in the SF_Capture folder. The
installation process is the same as other software – follow the prompts until it is done.
Power on the instrument, connect the serial port to USB cable from the reader to the PC. Verify
that External Export has been switched to ON from the instrument's Settings menu
NOTE: An installation of drivers will need to be done the first time SF_Capture is installed
on a PC.
Once the cable connection is made between the reader and the PC, Windows® will pop up a
message box stating that the VNC1-A As Slave is not installed and that drivers will need to be
found.
Correct response is to let Windows® look for the appropriate driver this one time.
First, it will search for and install USB Serial Converter driver(s).
The cables can be connected and the instrument's power up can be done in any sequence. If the
instrument is OFF, but SF_Capture is open and the cable is connected to the instrument, select
ReConnect.
Select Settings. This will open a dialog box where you select the instrument type and the COM
port connected to on the PC side
Data is exported to Excel, to Notepad and to a Log file each time QUIT/Test Ended occurs.
The default location for these files is C:\4200. This can be changed from the default by using the
Browse option for the Device Root Folder.
1. Erase/delete anything that is on the USB pen drive (or any USB drive used).
2. Unzip the attached file and install the folder romimage.42x on the USB drive.
3. Turn off the reader and insert the USB drive with only the one romimage.42x folder
on it into the port.
4. Place a finger on the display and then power up the reader.
5. After the instrument begins to beep you can remove your finger from the screen.
6. The update process will take about 3-5 minutes, depending upon the instrument’s
bootloader version. To determine the version installed, go to Utilities – Show Version
Details.
.
Selection of each of the eight squares should produce numbers on the display. If an area outside
of the acceptable range is pressed with the stylus or mouse, a sound will be produced. Each of
the pairs (top, left, bottom and right) must have a difference between them of not more than 3
points, such as on the left, 17 and 20 are good, whereas 17 and 24 are not. The SAVE button will
not appear until acceptable points have been selected. It is necessary to select SAVE in order to
store the new values.