0% found this document useful (1 vote)
886 views11 pages

1 Nabl

The document is a calibration certificate for two items: a digital vernier caliper and a vernier caliper. It summarizes the calibration results, which found the measurements to be within the expanded uncertainty range when compared to the reference standards. The calibrations were performed by a calibration engineer and approved by the technical head.

Uploaded by

Neha Sharma
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
886 views11 pages

1 Nabl

The document is a calibration certificate for two items: a digital vernier caliper and a vernier caliper. It summarizes the calibration results, which found the measurements to be within the expanded uncertainty range when compared to the reference standards. The calibrations were performed by a calibration engineer and approved by the technical head.

Uploaded by

Neha Sharma
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 11

Page: 01 of 02

Calibration Certificate / ULR No.: CC202221000001412F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

CALIBRATED FOR: M/S PAPER TESTING LAB


IIT Roorkee, Saharanpur Campus

ITEM’S DESCRIPTION: Name : Digital Vernier Caliper


Make / Model No. : ---------------
Serial No. : ---------------
Identification No. : IITR/SRE/PTL/16
Range : 0 to 150 mm
Least Count : 0.01 mm
Location : Paper Testing Laboratory
Physical Condition : OK Zero Error: 00

Temperature : 20.0 °C ± 2.0 °C


Environmental Condition
Relative Humidity : 50 % ± 20 %

STANDARD REFERENCE DETAILS (Traceable to National Standard)


Uncertainty
Make / Calibration
Name of Standard I.D. No. Certificate No. Traceable (at 95% C.L.
Model No. Due Date
approx.)
Tanson As per
Slip Gauges Set Aditya / DCL/STD/
TI/G/SGS/017/20 Instrument, 07-07-2022 Certificate
“0” Grade M88 SG-01
Faridabad (k=2)
Tanson
DCL/STD/ ± 1.47 µm
Caliper Checker Aditya TI/G/CC/002/20 Instrument, 21-07-2022
CC-01 (k=2)
Faridabad
Equipments Used : Slip Gauge Accessories
Principle / Methodology of Calibration: As per Calibration Procedure No. DCL/CP/M-02
Reference Standard : IS-3651

CALIBRATION RESULTS (IN-LAB)

Reference Standard Average Measured U.U.C. Value (mm) Expanded


Value (mm) Uncertainty^
External Measurement Internal Measurement Depth Measurement
2.00 2.00 ------ ------ ± 13.1 µm
5.00 5.00 ------ ------ ± 13.1 µm
10.00 10.00 10.00 10.00 ± 13.1 µm
20.00 20.00 20.00 20.00 ± 13.1 µm
50.00 50.00 49.99 50.00 ± 13.1 µm
100.00 100.00 99.99 100.00 ± 13.1 µm
150.00 149.99 149.98 150.00 ± 13.1 µm

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 02 of 02

Calibration Certificate / ULR No.: CC202221000001412F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

Remarks:
1- Expanded Uncertainty^ of measurement evaluated at a confidence Level of approximately 95% with Coverage factor k=2 for a
normal probability distribution.
2- U.U.C. defined above as Unit Under Calibration.
*End of Report*

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 01 of 02

Calibration Certificate / ULR No.: CC202221000001413F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

CALIBRATED FOR: M/S PAPER TESTING LAB


IIT Roorkee, Saharanpur Campus

ITEM’S DESCRIPTION: Name : Vernier Caliper


Make / Model No. : PRECISE
Serial No. : ---------------
Identification No. : IITR/SRE/PTL/15
Range : 0 to 300 mm
Least Count : 0.02 mm
Location : Paper Testing Laboratory
Physical Condition : OK Zero Error: 00

Temperature : 20.0 °C ± 2.0 °C


Environmental Condition
Relative Humidity : 50 % ± 20 %

STANDARD REFERENCE DETAILS (Traceable to National Standard)


Uncertainty
Make / Calibration
Name of Standard I.D. No. Certificate No. Traceable (at 95% C.L.
Model No. Due Date
approx.)
Tanson As per
Slip Gauges Set Aditya / DCL/STD/
TI/G/SGS/017/20 Instrument, 07-07-2022 Certificate
“0” Grade M88 SG-01
Faridabad (k=2)
Tanson
DCL/STD/ ± 1.47 µm
Caliper Checker Aditya TI/G/CC/002/20 Instrument, 21-07-2022
CC-01 (k=2)
Faridabad
Equipments Used : Slip Gauge Accessories
Principle / Methodology of Calibration: As per Calibration Procedure No. DCL/CP/M-02
Reference Standard : IS-3651

CALIBRATION RESULTS (IN-LAB)

Reference Standard Average Measured U.U.C. Value (mm) Expanded


Value (mm) Uncertainty^
External Measurement Internal Measurement Depth Measurement
2.00 2.00 ------ ------ ± 15 µm
5.00 5.00 ------ ------ ± 15 µm
10.00 10.00 10.00 10.00 ± 15 µm
20.00 20.00 20.00 20.00 ± 15 µm
50.00 50.00 50.00 50.00 ± 15 µm
100.00 100.00 100.00 100.00 ± 15 µm
200.00 200.00 200.00 200.00 ± 15 µm
300.00 300.00 300.00 300.00 ± 15 µm

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 02 of 02

Calibration Certificate / ULR No.: CC202221000001413F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

Remarks:
1- Expanded Uncertainty^ of measurement evaluated at a confidence Level of approximately 95% with Coverage factor k=2 for a
normal probability distribution.
2- U.U.C. defined above as Unit Under Calibration.
*End of Report*

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 01 of 01

Calibration Certificate / ULR No.: CC202221000001414F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

CALIBRATED FOR: M/S PAPER TESTING LAB


IIT Roorkee, Saharanpur Campus

ITEM’S DESCRIPTION: Name : Steel Scale


Make / Model No. : KRISTEEL
Serial No. : ---------------
Identification No. : IITR/SRE/PTL/18
Range : 0 to 600 mm
Least Count : 0.5 mm (≤ 100 mm) & 1 mm (> 100 mm)
Location : Paper Testing Laboratory
Physical Condition : OK Zero Error: ------

Temperature : 20.0 °C ± 2.0 °C


Environmental Condition
Relative Humidity : 50 % ± 20 %

STANDARD REFERENCE DETAILS (Traceable to National Standard)


Uncertainty
Make / Calibration
Name of Standard I.D. No. Certificate No. Traceable (At 95% C.L.
Model No. Due Date
approx.)
Khushi
As per cal.
Tape & Scale DCL/STD/ CC237920 Calibration
Detron 27-11-2021 certificate
Calibration System STCS-01 000004602F Laboratory,
(k=2)
Faridabad
Equipments Used : Nil
Principle / Methodology of Calibration: As per Calibration Procedure No.: DCL/CP/M-18
Reference Standard : IS-1481

CALIBRATION RESULTS (IN-LAB)


Average U.U.C. Value Average Reference Standard Value Expanded Uncertainty^
(mm) (mm) (µm)
10 10.005 ± 290.9
50 50.016 ± 290.9
100 100.026 ± 290.9
200 200.049 ± 291.2
300 300.073 ± 291.2
450 450.108 ± 291.2
600 600.143 ± 291.2
Remarks:
1- Expanded Uncertainty^ of measurement evaluated at a confidence Level of approximately 95% with Coverage factor k=2 for a
normal probability distribution.
2- U.U.C. defined above as Unit Under Calibration.
*End of Report*

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 01 of 02

Calibration Certificate / ULR No.: CC202221000001415F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

CALIBRATED FOR: M/S PAPER TESTING LAB


IIT Roorkee, Saharanpur Campus

ITEM’S DESCRIPTION: Name : Digimatic Micrometer


Make / Model No. : PRECISE
Serial No. : ---------------
Identification No. : IITR/SRE/PTL/24
Range : 0 to 25 mm
Least Count : 0.001 mm
Location : Paper Testing Laboratory
Physical Condition : OK Zero Error: 00

Temperature : 20.0 °C ± 2.0 °C


Environmental Condition
Relative Humidity : 50 % ± 20 %

STANDARD REFERENCE DETAILS (Traceable to National Standard)


Uncertainty
Make / Calibration
Name of Standard I.D. No. Certificate No. Traceable (at 95% C.L.
Model No. Due Date
approx.)
Tanson As per
Slip Gauges Set Aditya / DCL/STD/
TI/G/SGS/017/20 Instrument, 07-07-2022 Certificate
“0” Grade M88 SG-01
Faridabad (k=2)
Equipments Used : Nil
Principle / Methodology of Calibration: As per Calibration Procedure No. DCL/CP/M-03
Reference Standard : IS-2967

CALIBRATION RESULTS (IN-LAB)


Reference Standard Value Average Measured U.U.C. Value Expanded Uncertainty^
(mm) (mm) (µm)
2.500 2.500 ± 1.8
5.100 5.100 ± 1.8
7.700 7.700 ± 1.8
10.300 10.300 ± 1.8
12.900 12.900 ± 1.8
15.000 15.001 ± 1.8
17.600 17.601 ± 1.8
20.200 20.201 ± 1.8
22.800 22.802 ± 1.8
25.000 25.002 ± 1.8

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 02 of 02

Calibration Certificate / ULR No.: CC202221000001415F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

Remarks:
1- Expanded Uncertainty^ of measurement evaluated at a confidence Level of approximately 95% with Coverage factor k=2 for a
normal probability distribution.
2- U.U.C. defined above as Unit Under Calibration.
*End of Report*

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 01 of 02

Calibration Certificate / ULR No.: CC202221000001416F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

CALIBRATED FOR: M/S PAPER TESTING LAB


IIT Roorkee, Saharanpur Campus

ITEM’S DESCRIPTION: Name : Vernier Height gauge


Make / Model No. : PRECISE
Serial No. : ---------------
Identification No. : IITR/SRE/PTL/21
Range : 0 to 300 mm
Least Count : 0.02 mm
Location : Paper Testing Laboratory
Physical Condition : OK Zero Error: ----

Temperature : 20.0 °C ± 2.0 °C


Environmental Condition
Relative Humidity : 50 % ± 20 %

STANDARD REFERENCE DETAILS (Traceable to National Standard)


Uncertainty
Make / Calibration
Name of Standard I.D. No. Certificate No. Traceable (at 95% C.L.
Model No. Due Date
approx.)
Tanson As per
Slip Gauges Set Aditya / DCL/STD/
TI/G/SGS/017/20 Instrument, 07-07-2022 Certificate
“0” Grade M88 SG-01
Faridabad (k=2)
Tanson
DCL/STD/ ± 1.47 µm
Caliper Checker Aditya TI/G/CC/002/20 Instrument, 21-07-2022
CC-01 (k=2)
Faridabad
Equipments Used : Dial Test Indicator

Principle / Methodology of Calibration: As per Calibration Procedure No. DCL/CP/M-11

Reference Standard : IS-2921

CALIBRATION RESULTS (IN-LAB)


Reference Standard Value Average Measured U.U.C. Value
Expanded Uncertainty^
(mm) (mm)
5.000 5.00 ± 15.0 µm
10.000 10.00 ± 15.0 µm
20.000 20.00 ± 15.0 µm
50.000 50.00 ± 15.0 µm
100.000 100.00 ± 15.0 µm
200.000 200.00 ± 15.0 µm
300.000 299.98 ± 15.0 µm

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 02 of 02

Calibration Certificate / ULR No.: CC202221000001416F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

Remarks:
1- Expanded Uncertainty^ of measurement evaluated at a confidence Level of approximately 95% with Coverage factor k=2 for a
normal probability distribution.
2- U.U.C. defined above as Unit Under Calibration.
*End of Report*

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 01 of 02

Calibration Certificate / ULR No.: CC202221000001417F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

CALIBRATED FOR: M/S PAPER TESTING LAB


IIT Roorkee, Saharanpur Campus

ITEM’S DESCRIPTION: Name : Digital Thermo-Hygrometer


Make / Model No. : MEXTECH / TM-2
Serial No. : ---------------
Identification No. : IITR/SRE/PTL/14
Range : 0 °C to 50 °C & 20 %RH to 90 %RH
Least Count : 0.1 °C & 1 %RH
Location : Paper Testing Laboratory
Physical Condition : OK Zero Error: ------

Temperature : 25.0 °C ± 2.5 °C


Environmental Condition
Relative Humidity : 50 % ± 20 %

STANDARD REFERENCE DETAILS (Traceable to National Standard)

Uncertainty
Make / Calibration
Name of Standard I.D. No. Certificate No. Traceable (At 95% C.L.
Model No. Due Date
approx.)
± 0.32 °C,
Dig. Thermometer
Yudian / DCL/STD/ Belz, ± 0.20 °C &
with RTD (Pt-100) 30030397 14-10-2021
AI5600 TM-03 Faridabad ± 0.80 °C
4 wire sensor
(k=2)
Dig. Humidity
Polltech / DCL/STD/ Belz, ± 0.8 %RH
Indicator with RH 30030394 14-10-2021
PSI-RHT1 THM-04 Faridabad (k=2)
Sensor
Equipments Used : Humidity Chamber
Principle / Methodology of Calibration: As per Calibration Procedure No.: DCL/CP/TH-07

CALIBRATION RESULTS (IN-LAB)


Average U.U.C. Value Average Standard Reference Value Expanded Uncertainty^
(°C) (°C ) (°C)
10.0 10.032 ± 0.30
20.0 20.045 ± 0.30
30.0 30.067 ± 0.30
40.0 40.096 ± 0.30
50.0 50.115 ± 0.30

Calibrated by: Nayan Kumar Approved by: B.G. Sharma


(Sr. Calibration Engineer) (Technical Head)
Page: 02 of 02

Calibration Certificate / ULR No.: CC202221000001417F Calibration Request No.: DCL/2021/N/5167


Certificate issued on: 05-02-2021 Date of Receipt: 02-02-2021
Suggested date for the next calibration: 03-02-2022 Date of Calibration: 03-02-2021

CALIBRATION RESULTS (IN-LAB)


Average U.U.C. Value Average Standard Reference Value Expanded Uncertainty^
(%RH) (%RH) (%RH)
30 30.21 ± 1.64
50 50.35 ± 1.64
60 60.57 ± 1.64
70 70.85 ± 1.64
90 90.89 ± 1.64

Remarks:
1- Expanded Uncertainty^ of measurement evaluated at a confidence Level of approximately 95% with Coverage factor k=2 for a
normal probability distribution.
2- U.U.C. defined above as Unit Under Calibration.
3- The temperature measurements have been observed at 50 %RH (approx.) and humidity measurements have been observed
at 25 °C (approx.).
*End of Report*