Moduri de Ventilatie PDF
Moduri de Ventilatie PDF
                                                                                                                                                                 Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
                                     ventilation: effects on respiratory events and
                                     implications for their identification
                                     Claudio Rabec,1 Daniel Rodenstein,2 Patrick Leger,3 Sylvie Rouault,4
                                     Christophe Perrin,5 Jésus Gonzalez-Bermejo,6 on behalf of the SomnoNIV group
1
  Service de Pneumologie et          ABSTRACT                                                      is generally applied at night, nocturnal monitoring
Réanimation Respiratoire,           Compared with invasive ventilation, non-invasive              seems the best way to assess its effects. Although
Centre Hospitalier et
                                     ventilation (NIV) has two unique characteristics: the non-    nocturnal monitoring of continuous positive
Universitaire de Dijon, Dijon,
France                               hermetic nature of the system and the fact that the           airway pressure (CPAP) has been codified in the
2
  Service de Pneumologie,            ventilator-lung assembly cannot be considered as              treatment of patients with obstructive sleep
Cliniques Universitaires Saint       a single-compartment model because of the presence of         apnoea syndrome,6 this is not the case with NIV.7 8
Luc, Université Catholique de       variable resistance represented by the upper airway.          Nocturnal monitoring of NIV is far more difficult
Louvain, Bruxelles, Belgium
3
  Service de Pneumologie,            When NIV is initiated, the ventilator settings are            and unforeseen problems arise for many reasons:
Centre Hospitalier Lyon Sud,         determined empirically based on a clinical evaluation and     (1) sleep can induce profound ventilatory changes,
Lyon France                          diurnal blood gas variations. However, NIV is                 in particular in patients with respiratory insuffi-
4
  ADEP Assistance, Paris, France     predominantly applied during sleep. Consequently, to          ciency; (2) these ventilatory changes are also exac-
5
  Service de Pneumologie,                                                                          erbated by the interactions with a preset
Centre Hospitalier de Cannes,        assess overnight patientemachine ‘agreement’ and
Cannes, France                       efficacy of ventilation, more specific and sophisticated      devicedthe ventilator can induce sleep distur-
6
  Service de Pneumologie et          monitoring is needed. The effectiveness of NIV might          bances by itself; (3) physicians caring for these
Réanimation Respiratoire,           therefore be more correctly assessed by sleep studies         patients may vary greatly in their methods of
Hôpital de la Pitié-Salpétriere   than by daytime assessment. The most available and            assessing the effects of NIV from a single blood gas
and UPMC ER10, Paris, France
                                     simple monitoring can be done from flow and pressure          measurement to full polysomnography.
Correspondence to                    curves from the mask or the ventilator circuit.                  The issue is complicated further because volume
Claudio Rabec, Service de            Examination of these tracings can give useful information     and pressure preset ventilators will respond differ-
Pneumologie et Réanimation          to evaluate if the settings chosen by the operator were       ently in certain situations and because the type of
Respiratoire, Centre Hospitalier                                                                   mask used will also have different effects on
et Universitaire de Dijon, 2 Bd
                                     the right ones for that patient. However, as NIV allows
Maréchal de Lattre de Tassigny,     a large range of ventilatory parameters and settings, it is   monitoring parameters.9 The semiology of
21079 Dijon, France;                 mandatory to have information about this to better            abnormal respiratory events during sleep is there-
claudio.rabec@chu-dijon.fr           understand patienteventilator interaction. Ventilatory        fore completely different depending on the tech-
                                     modality, mode of triggering, pressurisation slope, use or    nology used. Pressure and flow dynamics also differ
Received 14 May 2010
                                     not of positive end expiratory pressure and type of           markedly among the different type of ventilators
Accepted 10 August 2010
Published Online First               exhalation as well as ventilator performances may             available for home care.
14 October 2010                      all have physiological consequences. Leaks and upper             This article will deal with the equipment avail-
                                     airway resistance variations may, in turn, modify             able for NIV, in particular the ventilator types,
                                     these patterns. This article discusses the equipment          modes and settings. Its goal is to give the back-
                                     available for NIV, analyses the effect of different           ground necessary to understand their impact on
                                     ventilator modes and settings and of exhalation and           nocturnal monitoring of NIV.
                                     connecting circuits on ventilatory traces and gives the
                                     background necessary to understand their impact on
                                     nocturnal monitoring of NIV.                                  ISSUES OF PARTICULAR IMPORTANCE DURING
                                                                                                   NIV: LEAKS, UPPER AIRWAY RESISTANCE, TYPE
                                                                                                   OF EXHALATION PORT AND NIV VENTILATORS
                                                                                                   Compared with invasive ventilation, NIV has two
                                     Since the first studies in the early 1980s showing             unique characteristics: the non-hermetic nature of
                                     the usefulness of non-invasive ventilation (NIV) in           the system that poses the potential risk of unin-
                                     the management of some forms of respiratory                   tentional leaks and the fact that the ventilator-lung
                                     failure,1e4 the number of patients receiving this             assembly cannot be considered as a single-
                                     treatment both in the acute setting and at home is            compartment model because of the presence of
                                     continuously increasing. This is explained by                 a variable resistance represented by the upper
                                     a growing number of indications in which the                  airway (UA). Both situations may compromise the
                                     effectiveness of NIV has been proven, but also by             delivery of an effective tidal volume. As a conse-
                                     major technological advances that led to the avail-           quence, increasing the delivered volume or the
                                     ability of high-performance portable ventilators as           delivered inspiratory pressure during NIV does not
                                     well as to the development of technical support               necessarily result in increased effective ventilation
                                     infrastructure.5                                              reaching the lungs.10 When monitoring NIV using
                                        When starting NIV, it seems desirable to verify            flow or pressure tracings, these two peculiarities
                                     whether this particular form of treatment is effec-           will unavoidably influence the features of the
                                     tive and has no untoward effects. Since home NIV              respiratory traces.
Influence of unintentional leaks                                      will have similar features whatever the type of positive pressure
Unintentional leaks are very common in NIV.11 12 Leakage may          device used for ventilation.
                                                                                                                                            Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
be absent or minimal when the patient is awake but may
worsen during sleep as a result of the loss of voluntary control
                                                                      VENTILATOR MODES AND SETTINGS: WHAT’S IN A NAME?
and decreased muscle tone. Leaks can take place at the mouth or
                                                                      When NIV was introduced there were a very limited number of
between the skin and the mask, and air could also be deposited
                                                                      modalities and types of ventilators with very few possible
in the oropharyngeal reservoir and even flow into the digestive
                                                                      settings. We now have more than 30 brands offering numerous
tract (‘internal’ leaks).13
                                                                      options for settings.22 Moreover, ventilators are not submitted
   Leaks can impair the quality of both ventilation and sleep.
                                                                      to stringent medical regulations. This leaves manufacturers free
They can largely affect ventilator triggering, pressurisation,
                                                                      to give different names to the same ventilator modalities and
volume delivered, rate of inspiratory pressuring and cycling
                                                                      settings and even to ‘create’ new modalities that frequently
function and induce sleep fragmentation.
                                                                      correspond only to small modifications of a known class. This
                                                                      explains the wide variety of existing terminology describing NIV
Influence of the upper airway                                         modalities.
During NIV a variable resistance constituted by the UA is
interposed between the ventilator and the lungs. The UA may           Influence of the ventilator mode
change its resistance to airflow, compromising the delivery of an      Theoretically, NIV could be delivered using all the modalities
effective tidal volume to the lungs. Intermittent obstruction of      used for invasive ventilation but, in real life, most ventilators
the UA is common during NIV and may be related to two                 used for NIV deliver either volume or pressure-targeted venti-
mechanisms. The first corresponds to obstructive events at the         lation. Because the place of other anecdotal modalities proposed
oropharyngeal level because of UA collapse as a result of             by some NIV devices such as synchronised intermittent
insufficient expiratory airway pressure. This mechanism may be         mandatory ventilation or other ‘hybrid’ modalities is not yet
present in patients with an unstable UA.14 Another mechanism          clear, this paper will mainly focus on the former two modalities.
corresponds to episodes of intermittent obstruction at
the glottic level reflecting cyclic glottic closure induced by
                                                                      Volume-targeted ventilation (VTV)
hyperventilation, a type of ‘ventilation resistance’ reaction.15e18
                                                                      In this modality the ventilator delivers a fixed volume during
                                                                      a given time and will generate whatever pressure is necessary to
Influence of type of exhalation device and connecting circuits        achieve this, regardless of the patient contribution to ventila-
Whereas ICU ventilators classically use a double circuit with an      tion. Pressure in the airways (Paw) is not constant and results
integrated expiratory valve, two different types of circuits can be   from the interaction between ventilator settings, compliance
used to provide NIV. The first uses a similar assembly to those        and resistance of the respiratory system and spontaneous
used in ICU devices and includes either single or double tubing       inspiratory efforts (figure 2). It should be emphasised that any
in which inspiration and expiration are separated and a true          additional inspiratory effort will not lead to changes in delivered
expiratory valve is present so that carbon dioxide (CO2)              volumes or flows but only result in a decrease in Paw. As each
rebreathing is not a significant problem (figure 1A). A second          breath is delivered with the same predetermined flow-time
type does not have a true exhalation valve and often uses             profile and as the area under the flow-time curve defines the
a single-limb circuit with a risk of rebreathing. To avoid            volume, the advantage of this modality is the strict delivery, in
rebreathing, this system includes a calibrated leak (called an        the absence of leaks, of the preset volume, whatever the values
intentional leak) either at the mask level or in the circuit (figure   of compliance (C) and resistance (R). A major disadvantage of
1B). Single circuit pressure-targeted ventilators provided with       VTV is precisely that delivery of this fixed ventilatory assistance
a calibrated leak (called bilevel ventilators) are most commonly      does not allow account to be taken of patients’ varying
used for NIV nowadays. These devices cycle between a higher           requirements. Another inconvenience is that, if there is a leak,
inspiratory positive airway pressure (IPAP) and a lower expira-       there will be no increase in flow rate to compensate for it and
tory positive airway pressure (EPAP) that can be independently        the generated pressure will be lower, so that the effectively
adjusted. With these devices, a minimum mandatory EPAP level          delivered volume will be reduced in proportion.
of 4 cm H2O is needed to ensure an effective washout of CO2.20
Interestingly, a recent study showed that the exhalation port         Pressure-targeted ventilation (PTV)
position influences CO2 rebreathing with a more efficient CO2           In this modality the ventilator is set to deliver airflow by
washout when the leak is positioned within the mask.21                generating a predefined positive pressure in the airways for
   When monitoring NIV using polygraphy, the mode of exha-            a given time. Airflow is therefore adjusted in order to establish
lation used and the position of the flow sensor (in relation to the    and maintain a constant Paw. Constant analysis of the flow rate
expiratory device) both have a major influence on features of          and airway pressure determines the flow variations necessary to
respiratory traces. Typical traces of these different assemblies      maintain a flat or ‘square wave’ pressure. Flow is brisk at the
can be seen in figure 1A and B.                                        beginning of inspiration when the gradient between the circuit
                                                                      pressure and the pressure target is large. As this gradient narrows
Influence of the ventilator: ICU vs home devices                      the flow decelerates until driving pressure no longer exists and
Both ICU and home ventilators can be used to deliver NIV. The         flow ceases (figure 2).
main technical characteristic differentiating them is that, for the      Thus, for a given patient, the volume delivered is not fixed and
former, the driving pressure is supplied by compressed gas,           will depend on the interaction between the preset pressure,
whereas the latter devices incorporate their own pressure source.     patient inspiratory effort, the physical characteristics of the
Nevertheless, as the type of ventilatory support that they            respiratory system (C and R) and inspiratory time. An impor-
provide is similar, this difference has little influence on respira-   tant advantage of PTV is the ability to compensate for mild to
tory traces. Therefore, when monitoring NIV, respiratory traces       moderate leaks.
                                                                                                                                               Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
an expiratory valve and (B) intentional
leak, and typical traces obtained with
these different assemblies. Note that
when a circuit with an expiratory valve
is used, the expiratory valve may be
interposed in the circuit (single circuit)
(A) or included in the ventilator (double
circuit) (B); when a single circuit with
intentional leak is used, the leak may be
interposed in the circuit (C) or
incorporated at the mask (D). Flow
traces can be influenced by the type of
circuit used, but also by the position of
the pneumotachograph with regard to
the expiratory device. In the case of
a double circuit provided by an
expiratory valve, the expiratory slope
will reflect the expired volume only
when the flow sensor is interposed
between the mask and the exhalation
device (trace 2 in A), but not when it is
placed distally to the expiratory valve
(trace 1 in A). On the other hand, when
using a single circuit with intentional
leak, the expiratory slope does not
reflect the expired tidal volume and may
be absent (B). In this case, the position
of the pneumotachograph with regard to
the leak device does not significantly
change the expiratory slope. Modified
from Perrin et al19 with permission.
  A comparative analysis of the two modalities and of corre-        reached, the breath converts to a flow-cycled modality by
sponding flow and pressure patterns is summarised in table 1.        prolonging the inspiratory time), but most of them progressively
                                                                    adjust the pressure level during several cycles to provide a tidal
Volume-targeted pressure ventilation                                volume as close as possible to the target volume set by the
A limitation of pressure ventilation is that it cannot guarantee    clinician.
a tidal volume delivered to the patient. Volume targeting is           Whether this feature improves the effectiveness of ventilation
a feature available in some new ventilators that could allow this   is not yet clear.23 24 Different algorithms used to provide
limitation to be overcome. This hybrid modality combines            volume-targeted pressure are shown in figure 3.
features of pressure and volume ventilation. The ventilator
estimates the delivered tidal volume and adjusts its parameters     Recent ventilator modes: matching physiology and practice?
to ensure a predetermined target tidal volume. Some ventilators     Conventional ventilation delivers fixed pressure or volumes
adjust a target volume within each cycle (each breath starts as     rates. Some recent modes use a more physiological delivery of
a pressure-limited breath and if the set tidal volume is not        assistance following the spontaneous inspiratory signal coming
                                                                                                                                                                         Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
                                                                                                             Volume-targeted                    Pressure-targeted
                                                                              Pressure curve pattern
                                                                                                                                                          Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
volume and flow recording traces using
(A) an algorithm that progressively
adjusts the pressure support level over
several cycles between a preset range
of pressure to achieve a target tidal
volume (Vt) and (B) an algorithm that
adjust a target volume intracycle. Note
that, in this case, each breath starts as
a pressure-limited breath and if the set
Vt is not reached, the breath converts to
a flow-cycled mode by prolonging the
inspiratory time. In both cases the
ventilator is able to estimate the cycle to
cycle Vt effectively delivered to the
patient and to adjust the ventilatory cycle to ensure a target Vt. Press, inspiratory pressure. Reproduced with permission of A Cuvelier.
asynchrony is quite common in patients during NIV.31 32                     ventilator is triggered by the patient. The pressure is maintained
Asynchronies may occur at two levels: during inspiratory trig-              as long as a minimal preset inspiratory flow is occurring. End of
gering in situations in which there is a mismatch between                   inspiratory assistance (eg, cycling from inspiration to expiration)
patient inspiratory effort and ventilator triggering (ie, ineffective       occurs when inspiratory flow reaches a predetermined
inspiratory effort, double triggering or autotriggering); or during         percentage of peak inspiratory flow. In this mode, a targeted
cycling from inspiration to expiration when ventilator cycling              inspiratory pressure, inspiratory trigger sensitivity and
does not coincide with the end of patient effort (ie, premature or          a percentage threshold of peak flow for cycling to expiration (see
delayed cycling).32                                                         below) must be selected. In some ventilators these can all be set
  Thus, the most logical approach to explain how a ventilator               by the clinician whereas in others only the inspiratory pressure
acts is to analyse the different phases of a typical positive               can be set. Because each cycle is terminated by a flow criterion
pressure ventilatory cycle (figure 4).                                       rather than by volume or time, the patient retains control of
  We will start this section by reviewing the triggering and                cycle length as well as its depth and flow profile This mode is
cycling modes which refers, respectively, to how inspiration and            also called ‘pressure support ventilation’ (PSV) (figure 5A).
expiration start and stop.
                                                                            Assist mode (A)
                                                                            In this mode the patient controls the onset of inspiration but the
How is the onset of ventilator inspiration determined?                      inspiratory length is regulated by the operator. The clinician
The patient can control the initiation (eg, triggering) and the             must select a targeted volume or pressure, an inspiratory:expi-
end of inspiration (eg, cycling into expiration) or, on the                 ratory (I:E) ratio or an inspiratory time and an inspiratory trigger
contrary, neither of them, and the ventilator controls the initi-           sensitivity (figure 5B).
ation and the end of inspiration.
                                                                            Assist-control mode (A/C)
Spontaneous mode (S)                                                        This mode operates as an assist one but also allows selecting
In this mode, only available in PTV, the patient controls the               a back-up respiratory rate (RR). If the patient’s spontaneous
beginning and end of inspiration. Inspiration starts when the               frequency is lower than the preset ventilator back-up RR, the
                                                                            system moves to control mode. Therefore, this mode allows the
                                                                            patient to trigger the ventilator but also grants a minimum
                                                                            back-up rate. In this mode the clinician must select the same
                                                                            settings as in assist mode but also add a back-up rate.
                                                                            Control mode (C)
                                                                            In the control mode there is a preset automatic cycle based on
                                                                            time. The ventilator controls the beginning and end of inspira-
                                                                            tion and thus the RR. One therefore expects the ventilator to
                                                                            capture and inhibit the patient’s respiratory centre and for the
                                                                            patient to follow the settings imposed by the ventilator. In this
                                                                            mode the clinician must select a targeted volume or pressure,
                                                                            a fixed RR and an I:E ratio or inspiratory and expiratory dura-
                                                                            tions. With this mode, the entire work of breathing is supposed
                                                                            to be performed by the ventilator. In some ventilators this mode
                                                                            is also called ‘Timed (T) mode’ and is rarely used (figure 5C).
                                                                               A particular combination of these modes is available in some
                                                                            NIV ventilators. This mode, called S/T, is basically a PSV that
                                                                            provides a back-up rate. In this particular mode, cycling from
                                                                            inspiration to expiration is flow-limited in patient-triggered
Figure 4 The ventilatory cycle as seen from a mask pressure-time            cycles and switches to time-limited when the patient’s sponta-
tracing. I to E, inspiratory to expiratory; PEEP, positive end expiratory   neous RR falls below the back-up RR, and also when the
pressure.                                                                   inspiratory time exceeds a predetermined maximal length during
                                                                                                                                                            Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
Figure 5 Different modes of inspiratory and expiratory triggering. (A) Spontaneous mode: the patient controls the beginning and end of inspiration.
Inspiration starts when the ventilator is triggered by the patient and cycling from inspiration into expiration occurs when the inspiratory flow reaches
a predetermined percentage of peak inspiratory flow. This mode is also called ‘pressure support’ and, in some ventilators, ‘spontaneous (S) mode’. (B)
Assisted mode: the patient controls the onset of inspiration but the end of inspiration is time triggered. When a backup respiratory rate (RR) is preset,
this mode is called assist/control. In this last mode, if the patient’s RR is lower than the preset ventilator backup RR, the system moves to control
mode. (C) Control mode: there is a preset automatic cycle based on time. The ventilator controls the beginning and end of inspiration and thus the RR.
In some ventilators this mode is also called ‘timed (T) mode’. Paw, pressure in the airways.
S cycles (see below). A patient-triggered cycle can be seen in                 EPAP level below the trigger threshold (when using pressure
curves of ventilation as a negative inspiratory deflection in                   triggering), both situations being able to lead to autotriggering
pressure and flow curves (see trace 2 in figures 2B and C).                      (figure 6).
                                                                                  The newer technologies (microprocessors, servo valves and
Type of trigger                                                                fast blowers) have substantially improved trigger responses.
As described above, in the A and A/C modes the ventilator has                  Moreover, adjustable inspiratory trigger is an option presently
to recognise the patient’s inspiratory effort. This is called trig-            available in most home ventilators. Some of these also propose
gering function. Classically, NIV devices have two types of                    automated complex trigger algorithms (‘flow waveform method
triggers. The first, called ‘pressure-based’ trigger, present in old            of triggering’) in which the flow-time waveform is used to
ventilators, is based on a drop in proximal airway pressure and                trigger the ventilator. The respective advantages of these
requires a closed circuit. The second, called ‘flow-based’ trigger,             sophisticated trigger systems have not been assessed in rigorous
present in almost all recent ventilators, is based on detection of             studies. It must be emphasised that some adjustable trigger
an inspiratory flow in the presence of continuous flow washing                   devices are graduated by arbitrary units which makes them
out the circuit during expiration. Asynchrony during inspiratory               difficult to use.
triggering is quite common during sleep in patients during NIV
and may compromise ventilatory efficacy and sleep quality. It is                Pressurisation rate
mainly influenced by the delay duration (that can vary between                  As correct pressurisation is essential to decrease inspiratory
different ventilators33), the trigger sensitivity and the amount of            effort and improve synchronisation, during this phase inspira-
inspiratory effort (which depends itself on respiratory drive and              tory flow should be sufficient to match inspiratory demand.36
muscle strength).31                                                            Circumstances influencing the rate of pressurisation are the level
   Ventilators that use flow triggering generally have the shorter              of ventilatory support, the amount of time required to reach the
trigger delays.34 35 Leaks may greatly affect trigger function,                pressure target (pressurisation slope, also called ‘rise time’), the
either by preventing the detection of patient inspiratory effort               compliance and resistance of the respiratory system and the
(leading to ineffective inspiratory effort) or by mimicking an                 patient inspiratory effort. Studies comparing different ventila-
‘inspiratory flow’ (when using flow triggering) or dragging the                  tors also emphasise the influence of the type of device on
                                                                                                                                                            Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
occurrence of at least three consecutive
pressurisations at a ventilator frequency
of >40/min not synchronised with
patient respiration.43
pressurisation, in particular in situations of high inspiratory               emphasised that, even if recent ventilators have a good capa-
demand.33                                                                     bility to compensate mild to moderate leaks, greater leaks may
  A faster rise time has been shown to unload respiratory                     compromise the ability of the device to attain a desired level of
muscles more completely.36 As the slope becomes flatter, the                   inspiratory pressure.
machine delivers lower flow rates and the patient’s work of
breathing increases.36 On the other hand, it must be emphasised               Cycling from inspiration to expiration
that, if slow pressurisation could increase inspiratory work, an              Switching from inspiration to expiration can be time-cycled or
excessive peak flow could also have adverse effects as it may                  flow-cycled. In time-cycled modes, ventilators use time criteria
increase the sensation of dyspnoea,37 induce double triggering32              chosen by the clinician (figure 4B). In flow-cycled modes, cycling
and lead to high peak mask pressure, favouring leaks.                         occurs as inspiratory flow decreases to a predetermined
                                                                              percentage of the peak inspiratory flow which is supposed to
Ability to sustain the inspiratory plateau                                    indicate the end of inspiratory effort (figure 4A).
The inspiratory pressure level is one of the main determinants of                The criteria used to end inspiration may have a clinically
the efficaciousness of NIV. Determination of the optimal level is              relevant impact on the quality of ventilation. Ideally, cycling
the result of balancing two opposing aims: the desire to provide              should coincide with the end of patient effort. However,
effective minute ventilation and the need to minimise leaks and               synchronisation between the end of neural inspiration and
discomfort caused by excessive inspiratory pressure. It must be               ventilator cycling into expiration is mainly determined by
Figure 7 Impact of leaks on inspiratory to expiratory (I to E) cycling during pressure support ventilation (S in bilevel devices) mode (A) without leaks
and (B) with leaks. Note that during leaks the flow increases to compensate for them, prolonging the inspiratory time (dashed lines) and switching to
time-limited if maximal inspiratory time (Timax) is available. Interestingly, Timax is adjustable in most recent non-invasive ventilation devices. Paw,
pressure in the airways.
respiratory mechanics, moving from a premature cycling in             and back, the answer to three basic questions may be used as
restrictive diseases to a late cycling in obstructive ones.33 38      a guide to simplify terminology:
                                                                                                                                                                         Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
Moreover, when flow cycling is used, leaks may also delay              1. What is the controlled variable (pressure or volume)?
switching into expiration because, in an attempt to maintain          2. What causes the start of ‘ventilator inspiration’? It can be
pressure, the flow rate is maintained above the level at which            either the patient (triggering) or the machine.
cycling into expiration occurs (figure 7). Both these conditions       3. What determines the end of inspiration? It is either based on
may lead to patient-ventilator expiratory asynchrony.38                  time (time-cycled) or determined by the patient (flow-cycled).40
   In older ventilators cycling into expiration was fixed at 25% of      Table 2 summarises the technical characteristics of NIV
peak flow, but more recent ventilators offer adjustable values.        ventilators and their influence on monitoring tracings.
This may allow tailoring settings to the patient’s underlying
condition. For instance, Tassaux et al demonstrated in patients       IMPACT OF VENTILATOR MODE, SETTINGS AND FUNCTION ON
with chronic obstructive pulmonary disease during invasive            MONITORING DURING NIV
ventilation that increasing cycling from inspiration to expiration    Using NIV, being connected to a ventilator is not synonymous to
from 10% to 70% of peak flow (this means shortening inspira-           being ventilated. It is therefore essential to assess NIV efficacy. A
tion to allow a greater expiratory time) was associated with          first step is to perform a clinical assessment with the patient
a marked reduction in delayed cycling and intrinsic PEEP.39           awake. However, as NIV is generally applied for several hours
   Finally, additional mechanisms proposed by some ventilators        during sleep, more specific and sophisticated monitoring is
control the end of inspiration to prevent undesirable prolonga-       needed to assess overnight patient-machine ‘agreement’ and
tion of inspiratory time. Sudden increases in pressure (that can      efficacy of ventilation.
be assumed as secondary to an active expiratory effort) produce       Table 2 Summary of technical characteristics of NIV ventilators and
early cycling to expiration in almost all the devices. Another        their influence on traces semiology
mechanism is to limit maximal inspiratory time. This maximal                                                           Normal semiological aspect
inspiratory time acts as a safety feature to prevent unsuitable
lengthening of inspiratory duration and is highly variable in         Ventilatory modality
different ventilators. Moreover, it may be adjustable for some          Volume target                                  Squared flow
                                                                                                                       Increasing pressure
devices (figure 7).                                                      Pressure target                                Squared pressure
                                                                                                                       Decreasing flow
PEEP level                                                              Hybrids                                        Variable (depends on modality of volume
                                                                                                                       targeting)
In patients with obstructive disease, intrinsic positive end expi-
ratory pressure (iPEEP) can reduce the effective trigger threshold
                                                                      Inspiratory trigger function
leading to a significant delay between the onset of patient
                                                                         Spontaneous                                   Negative deflection in pressure and flow
inspiratory effort and ventilator triggering or even to ineffective                                                    curves
efforts. In these cases, providing an external PEEP may counter-        Assisted                                       Negative deflection in pressure and flow
balance iPEEP, improving patient-ventilator synchrony. PEEP is                                                         curves
an above atmospheric (positive) pressure applied during expira-         Controlled                                     e
tion. It is called EPAP in some ventilators. If PEEP is used, the       Assisted/controlled                            Negative deflection in pressure and flow
                                                                                                                       curves (only in a patient-triggered cycle)
pressure will never come back to zero. Indeed, IPAP results from
                                                                      Pressurisation
pressure support plus PEEP. As a result and with regard to the
                                                                        Rise time
ventilator category (ICU or home ventilator), the PEEP setting
                                                                           Fixed                                       Inspiratory slope until reaching the
may interfere with either pressure support or IPAP levels. In fact,                                                    pressure target
                                                                           Adjustable
ICU ventilators propose PEEP and pressure support settings, but
PEEP and IPAP settings are usually associated on home ventila-        Ability to sustain inspiratory plateau           Maintain constant pressure and/or flow
tors. Thus, the PEEP setting increases the IPAP level on ICU                                                           level
ventilators and decreases the pressure support level on home
ventilators. Providing an external PEEP during NIV has many           I to E cycling
other theoretical advantages: flushing CO2 from the deadspace,            Time cycled                                   Fixed inspiratory time
preventing rebreathing within the mask, preserving the airway            Flow cycled
patency in patients with unstable UA during sleep and recruiting            Fixed                                      Variable inspiratory time
alveoli. Unnecessary increases in PEEP levels must be avoided               Adjustable (% of peak flow)
because inspiratory pressure must be increased in parallel if               ‘Intelligent’ algorithms
inspiratory assistance is to be maintained, and this can lead to
intolerance and favour leaks. Leaks, if important, may make it        PEEP level                                       Level of expiratory pressure
impossible to maintain the set EPAP level.
                                                                      Type of exhalation
                                                                        Single or double limb with expiratory          Expiratory slope reflects expired Vt*
NIV MODALITIES AND SETTINGS: TIME TO DEFINE                             valve
                                                                        Single limb with calibrated leak               Expiratory slope does not reflect expired
A STANDARDISED NOMENCLATURE?                                                                                           Vt and may be absent*
As underlined above, there is a wide variety of existing termi-                                                        Any level of positive expiratory pressure
nologies describing NIV modalities without a common nomen-                                                             is recognisedy
clature and with a lot of confusion. It even happens that the          *Visual recognition of expiratory slope will depend on the position of the pneumotachograph
same acronyms correspond to different modalities and that              with regard to the expiratory device. In this way the expiratory slope will reflect the expired
identical modalities are called differently.                           volume only when the flow sensor is interposed between the mask and the exhalation
                                                                       device.
   As ventilators can be categorised by the way that they deliver      yAs a mandatory PEEP level is needed to impede rebreathing.
gas flow and by how they shift from inspiration to expiration           NIV, non-invasive ventilation; PEEP, positive end expiration pressure; Vt, tidal volume.
   The most available and simple monitoring can be done from                                      18.   Delguste P, Aubert-Tulkens G, Rodenstein DO. Upper airway obstruction during
flow and pressure curves. Examination of these tracings can give                                         nasal intermittent positive-pressure hyperventilation in sleep. Lancet
                                                                                                                                                                                                  Thorax: first published as 10.1136/thx.2010.142661 on 14 October 2010. Downloaded from http://thorax.bmj.com/ on May 23, 2020 by guest. Protected by copyright.
                                                                                                        1991;338:1295e7.
useful information to evaluate whether the settings chosen by                                     19.   Perrin C, Jullien V, Lemoigne F. [Practical and technical aspects of noninvasive
the operator are the right ones for that patient. However, as NIV                                       ventilation]. Rev Mal Respir 2004;21:556e66.
allows a large range of ventilatory parameters and settings, it is                                20.   Lofaso F, Brochard L, Touchard D, et al. Evaluation of carbon dioxide rebreathing
                                                                                                        during pressure support ventilation with airway management system (BiPAP)
mandatory to have accurate information about these issues to                                            devices. Chest 1995;108:772e8.
better understand the interplay between the patient and the                                       21.   Schettino GP, Chatmongkolchart S, Hess DR, et al. Position of exhalation port and
ventilator. Ventilatory modality, mode of triggering, pressurisa-                                       mask design affect CO2 rebreathing during noninvasive positive pressure ventilation.
                                                                                                        Crit Care Med 2003;31:2178e82.
tion slope, use or not of PEEP and type of exhalation as well as                                  22.   Gonzalez-Bermejo J, Laplanche V, Husseini FE, et al. Evaluation of the
ventilator performances may all have semiological consequences                                          user-friendliness of 11 home mechanical ventilators. Eur Respir J
on ventilatory traces. Leaks and variations in UA resistance may,                                       2006;27:1236e43.
in turn, modify these patterns.                                                                   23.   Janssens JP, Metzger M, Sforza E. Impact of volume targeting on efficacy of
                                                                                                        bi-level non-invasive ventilation and sleep in obesity-hypoventilation. Respir Med
   Understanding the influence of non-invasive ventilator func-                                          2009;103:165e72.
tions on the appearance of abnormal respiratory events and an                                     24.   Storre JH, Seuthe B, Fiechter R, et al. Average volume-assured pressure
appropriate mastery of the available tools for monitoring NIV                                           support in obesity hypoventilation: a randomized crossover trial. Chest
                                                                                                        2006;130:815e21.
during sleep are prerequisites to a better understanding of how                                   25.   Ambrosino N, Rossi A. Proportional assist ventilation (PAV): a significant advance or
patients and ventilators interact. These issues are analysed in                                         a futile struggle between logic and practice? Thorax 2002;57:272e6.
two other articles in this review series.41 42                                                    26.   Ranieri VM, Giuliani R, Mascia L, et al. Patient-ventilator interaction during acute
                                                                                                        hypercapnia: pressure-support vs. proportional-assist ventilation. J Appl Physiol
                                                                                                        1996;81:426e36.
Acknowledgements The authors thank ADEP Assistance (Paris, France) for the                        27.   Giannouli E, Webster K, Roberts D, et al. Response of ventilator-dependent patients
organisation of the workshops of the SomnoNIV Group.                                                    to different levels of pressure support and proportional assist. Am J Respir Crit Care
Competing interests None.                                                                               Med 1999;159:1716e25.
                                                                                                  28.   Ambrosino N, Vitacca M, Polese G, et al. Short-term effects of nasal proportional
Provenance and peer review Commissioned; externally peer reviewed.                                      assist ventilation in patients with chronic hypercapnic respiratory insufficiency. Eur
                                                                                                        Respir J 1997;10:2829e34.
                                                                                                  29.   Polese G, Vitacca M, Bianchi L, et al. Nasal proportional assist ventilation unloads
REFERENCES                                                                                              the inspiratory muscles of stable patients with hypercapnia due to COPD. Eur Respir J
 1.   Leger P, Bedicam J, Cornette A, et al. Nasal intermittent positive pressure                       2000;16:491e8.
      ventilation. Long term follow up in patients with severe chronic respiratory failure.       30.   Sinderby C, Beck J. Proportional assist ventilation and neurally adjusted ventilatory
      Chest 1994;105:100e5.                                                                             assistebetter approaches to patient ventilator synchrony? Clin Chest Med
 2.   Leger P, Jennequin J, Gerard M, et al. [Nocturnal mechanical ventilation in                       2008;29:329e42, vii.
      intermittent positive pressure at home by nasal route in chronic restrictive respiratory    31.   Fanfulla F, Taurino AE, Lupo NDA, et al. Effect of sleep on patient/ventilator
      insufficiency. An effective substitute for tracheotomy]. Presse Med 1988;17:874.                  asynchrony in patients undergoing chronic non-invasive mechanical ventilation. Respir
 3.   Rideau Y, Gatin G, Bach J, et al. Prolongation of life in Duchenne’s muscular                     Med 2007;101:1702e7.
      dystrophy. Acta Neurol (Napoli) 1983;5:118e24.                                              32.   Thille AW, Rodriguez P, Cabello B, et al. Patient-ventilator asynchrony during
 4.   Sullivan CE, Issa FG, Berthon-Jones M, et al. Reversal of obstructive sleep apnoea                assisted mechanical ventilation. Intensive Care Med 2006;32:1515e22.
      by continuous positive airway pressure applied through the nares. Lancet                    33.   Battisti A, Tassaux D, Janssens JP, et al. Performance characteristics of 10 home
      1981;1:862e5.                                                                                     mechanical ventilators in pressure-support mode: a comparative bench study. Chest
 5.   Farre R, Lloyd-Owen SJ, Ambrosino N, et al. Quality control of equipment in home                  2005;127:1784e92.
      mechanical ventilation: a European survey. Eur Respir J 2005;26:86e94.                      34.   Aslanian P, El Atrous S, Isabey D, et al. Effects of flow triggering on breathing
 6.   Kushida CA, Littner MR, Hirshkowitz M, et al. Practice parameters for the use of                  effort during partial ventilatory support. Am J Respir Crit Care Med
      continuous and bilevel positive airway pressure devices to treat adult patients with              1998;157:135e43.
      sleep-related breathing disorders. Sleep 2006;29:375e80.                                    35.   Nava S, Ambrosino N, Bruschi C, et al. Physiological effects of flow and pressure
 7.   Gonzalez MM, Parreira VF, Rodenstein DO. Non-invasive ventilation and sleep.                      triggering during non-invasive mechanical ventilation in patients with chronic
      Sleep Med Rev 2002;6:29e44.                                                                       obstructive pulmonary disease. Thorax 1997;52:249e54.
 8.   Elliott M. Non-invasive ventilation during sleep: time to define new tools in the           36.   Bonmarchand G, Chevron V, Menard JF, et al. Effects of pressure ramp slope
      systematic evaluation of the technique. Thorax 2011;66:82e4.                                      values on the work of breathing during pressure support ventilation in restrictive
 9.   Borel JC, Sabil A, Janssens JP, et al. Intentional leaks in industrial masks have                 patients. Crit Care Med 1999;27:715e22.
      a significant impact on efficacy of bilevel noninvasive ventilation: a bench test study.    37.   Manning HL, Molinary EJ, Leiter JC. Effect of inspiratory flow rate on
      Chest 2009;135:669e77.                                                                            respiratory sensation and pattern of breathing. Am J Respir Crit Care Med
10.   Parreira VF, Delguste P, Jounieaux V, et al. Glottic aperture and effective minute                1995;151:751e7.
      ventilation during nasal two-level positive pressure ventilation in spontaneous mode.       38.   Calderini E, Confalonieri M, Puccio PG, et al. Patient-ventilator asynchrony during
      Am J Respir Crit Care Med 1996;154:1857e63.                                                       noninvasive ventilation: the role of expiratory trigger. Intensive Care Med
11.   Bach J, Robert D, Leger P, et al. Sleep fragmentation in kiphoscoliotic individuals               1999;25:662e7.
      with alveolar hypoventilation treated by NIPPV. Chest 1995;107:1552e8.                      39.   Tassaux D, Gainnier M, Battisti A, et al. Impact of expiratory trigger setting on
12.   Rabec C, Georges M, Kabeya NK, et al. Evaluating noninvasive ventilation using                    delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med
      a monitoring system coupled to a ventilator: a bench-to-bedside study. Eur Respir J               2005;172:1283e9.
      2009;34:902e13.                                                                             40.   Chopin C, Chambrin M. An attempt to classify the current positive airway pressure
13.   Rabec CA, Reybet-Degat O, Bonniaud P, et al. [Leak monitoring in noninvasive                      modes of mechanical ventilation. Réanimation Urgences 1998;7:87e99.
      ventilation]. Arch Bronconeumol 2004;40:508e17.                                             41.   Gonzalez-Bermejo J, Perrin C, Janssens JP, et al. Proposal for a systematic
14.   Pepin JL, Chouri-Pontarollo N, Orliaguet O, et al. [Alveolar hypoventilation during               analysis of polygraphy or polysomnography for identifying and scoring abnormal
      sleep and domiciliary assisted ventilation]. Rev Mal Respir 2005;22:113e25.                       events occurring during non invasive ventilation. Thorax Published Online First: 22
15.   Jounieaux V, Aubert G, Dury M, et al. Effects of nasal positive-pressure                          October 2010. doi:10.1136/thx.2010.142653.
      hyperventilation on the glottis in normal awake subjects. J Appl Physiol                    42.   Janssens JP, Borel JC, Pepin JL. Nocturnal monitoring of home non-invasive
      1995;79:176e85.                                                                                   ventilation: contribution of simple tools such as pulse-oximetry, capnography, built-in
16.   Jounieaux V, Aubert G, Dury M, et al. Effects of nasal positive-pressure hyperventilation         ventilator software and autonomic markers of sleep fragmentation. Thorax Published
      on the glottis in normal sleeping subjects. J Appl Physiol 1995;79:186e93.                        Online First: 22 October 2010. doi:10.1136/thx.2010.139782.
17.   Parreira VF, Jounieaux V, Aubert G, et al. Nasal two-level positive-pressure                43.   Guo YF, Sforza E, Janssens JP. Respiratory patterns during sleep in obesity-
      ventilation in normal subjects. Effects of the glottis and ventilation. Am J Respir Crit          hypoventilation patients treated with nocturnal pressure support: a preliminary report.
      Care Med 1996;153:1616e23.                                                                        Chest 2007;131:1090e9.