CMC Dimension
CMC Dimension
2
Measured/Observed Readings Standard Value Avearge (xj-x) (xj-x)
mm mm mm microns microns
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 6.135 microns
Degree of Freedom, (Veff)
4
uc(y)
Veff = 4
n(ui(y) )
j=1
Vi
= 1793.317216 =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
Ue = ke x Uc = 2 x 6.135 microns
Therefore uncertainty in above measurement is = ± 12.3 microns
Feeler Gauge
Determination of Measurement Uncertainty of Feeler Gauge
Feeler Gauge Range : 1.00 mm
Reading Point = 1.00 mm
Uncer. in D. M. M. (mm): = 2.10 microns
Unit of Measurement : = microns
o
Uncertainty of Temperature Scanner = 0.39
C
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows :-
n
[S(X)]2
Standard Deviation of the Mean, S ( X) = n
= 0.0000 microns
Standard Uncertainty in Type 'A' Evaluation, UA = 0.0000 microns
Type 'B' Evalution
1. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Slip Gague
The value of Uncertainty is taken from its calibration certificate
Assuming Normal distribution =2
0.17
U1 = = 0.085 microns
2
2. Standard Uncertainty Due to the Resolution of External Micrometer
Considering half of the least count & assuming rectangular distribution
1
U2 = = 0.289 microns
2x 3
3. Standard Uncertainty Due toThermalCoefficient of UUC & Standard
Assuming rectangular distribution.
U3 = L X a X dt
3
Where L = Length = 100.000 mm
a = 20 % of a1+a2 3.24 X 10-6 / Deg C
a1 = Th. Coefficient of Expansion of DStandard = 4.7 X 10-6 / Deg C
a2 = Th. Coefficient of Expansion of UUC = 11.5 X 10-6 / Deg C
dt = Control Limit or 10% of Temp. 20 °C 1.0 Deg C
0.0648 0.037 microns
3
4. Standard Uncertainty Due to the Wringing of Slip Gauge
Considering half of the least count & assuming rectangular distribution
0
U4 = = 0.000 microns
3
5. Standard Uncertainty Due to the Temp. Variation in Master Instrument & UUC assumed as 0.5 deg C
Assuming rectangular distribution.
L X a X dt
U5 =
3
Where L = Length = 100.00 mm
a = (a1+a2)/2 8.1 X 10-6
a1 = Th. Coefficient of Expansion = 4.7 X 10-6 / O C
a2 = Th. Coefficient of Expansion = 11.5 X 10-6 / O C
d = 20% of Temp. Limit ± 1ºC 0.2 Deg C
0.162
U5 = = 0.0935 microns
3
6. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
o
The value of Uncertainty is taken from its calibration certificate i.e. = 0.39 C
L X a X dt
U6 =
2
Where L = Length = 100.0 mm
a = Th. Coefficient of Expansion of Slip Gague = 8.1 X 10-6 /o C
o
d = Uncertainty of temperature scaner = 0.39 C
0.316
U6 = = 0.1580 Deg C
2
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 0.361 microns
Degree of Freedom, (Veff)
uc(y)4
Veff =
n(ui(y)4)
j=1
Vi
= 3.0423E+44 =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
Ue = ke x Uc = 2 x 0.361 microns
0.722 microns
Therefore uncertainty in above measurement is =+ 0.7 microns
INTERNAL MICROMETER
Determination of Measurement Uncertainty of Micrometer Range upto 100 mm
INTERNAL MM Range : 100 mm Least Count: 1 micron
Accessories & Slip Gauge Set is Used For Calibration
Uncer. Of Slip Gauge 0.12 micron
Accuracy of Slip Gauge 0.10 micron
Uncer. Of Slip Gauge Accessories 2.40 micron No. of Slip 1
Unit of Measurement : = microns Reading Point : 50 mm
o
Uncertinty of Temperature Scanner = 0.39 C
Accuracy of Master Equipment (Slip Gauge) 0.1 Surfaced plate 0
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows-
n
(xj-x) (xj-x)2
Measured/Observed Readings Standard Value Avearge
microns microns
49.998 50.000 -0.5000 0.250
49.999 50.000 0.5000 0.250
49.998 50.000 -0.5000 0.250
49.999 50.000 0.5000 0.250
49.999 50.000 49.9985 0.5000 0.250
49.999 50.000 0.5000 0.250
49.998 50.000 -0.5000 0.250
49.998 50.000 -0.5000 0.250
49.998 50.000 -0.5000 0.250
49.999 50.000 0.5000 0.250
Standard Deviation = (Xj - X)2
n-1
= 0.5270 microns
[S(X)]2
Standard Deviation of the Mean, S ( X) = n
= 0.1667 microns
Standard Uncertainty in Type 'A' Evaluation, UA = 0.1667 microns
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Slip Gauge
The value of Uncertainty is taken from its calibration certificate
Assuming Normal distribution =2
0.12
U1 = = 0.060 microns
2
2. Standard Uncertainty Due toThermal Coefficient of UUC & Standard
Assuming rectangular distribution.
L X a X dt
U2 =
3
Where L = Length = 100 mm
-6
a = 20% of ( a1 +a2) = 3.24 X 10 / Deg C
a1 = Th. Coefficient of Expansion = 4.7 X 10-6 / Deg C
a2 = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
d = Control Limit or 10% of Temp. 20 °C 1 Deg C
0.324
U2 = = 0.187 microns
3
3. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Slip Gauge Accessories
The value of Uncertainty is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
2.40
U3 = = 1.200 microns
U3 = = 1.200 microns
2
4. Standard Uncertainty Due to the Resolution of lnternal Micrometer
Considering half of the least count & assuming rectangular distribution
1
U4 = = 0.289 microns
2x 3
5. Standard Uncertainty Due to the Temp. Variation in Master Instrument & UUC
Assuming rectangular distribution.
L X a X dt
U5 =
3
Where L = Length = 100 mm
-6
a = (a1+a2)/2 8.1 X 10
a1 = Th. Coefficient of Expansion = 4.7 X 10-6 / Deg C
a2 = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
d = 20% of Temp. Limit ± 1ºC 0.2 Deg C
0.162
U5 = = 0.094 microns
3
6. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
o
The value of Uncertainty is taken from its calibration certificate i.e. = 0.39 C
Assuming rectangular distribution.
L X a X dt
U6 =
2
Where L = Length = 100 mm
a = Th. Coefficient of Expansion of Slip Gague = 4.7 X 10-6 / Deg C
o
d = Uncertainty of temperature Scanner = 0.39 C
0.183
U6 = = 0.1058 microns
1.732
2
Measured/Observed Readings Value Average (xj-x) (xj-x)
mm mm microns microns
(xj) (x)
75.0006 75 0.0000 0.000
75.0006 75 0.0000 0.000
75.0006 75 0.0000 0.000
75.0006 75 0.0000 0.000
75.0006 75 0.0000 0.000
75.0006 75 75.0006 0.0000 0.000
75.0006 75 0.0000 0.000
75.0006 75 0.0000 0.000
75.0006 75 0.0000 0.000
75.0006 75 0.0000 0.000
2
Standard Deviation , S(X) = å (Xj - X)
n-1
= 0.0000 microns
2
[S(X)]
Standard Deviation of the Mean, S ( X) = n
= 0.000 microns
Standard Uncertainty in Type 'A' Evaluation, UA = 0.000 microns
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Comp Stand
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
2.85
U1 = = 1.425 microns
2
2. Standard Uncertainty Due to 50% of Resolution of Standard Dial Gauge
Assuming Rectangular distribution
0.2
U2 = = 0.289 microns
2 3
3. Standard Uncertainty Due to Thermal Coefficien of Master Instrument & UUC .
Assuming rectangular distribution. for Master
L X a X dt
U3 =
3
Where L = Length = 75.00 mm
a = 20% of ( a1 +a2) = 0.0000032
-6 O
a1 = Th. Coefficient of Expansion = 4.7 X 10 / C
-6 O
a2 = Th. Coefficient of Expansion = 11.5 X 10 / C
d = Control Limit or 10% of Temp. 20 °C 1 Deg C
0.2430
U3 = = 0.140 microns
3
4. Standard Uncertainty Due to the Temp. Variation in Master Instrument & UUC.
Assuming rectangular distribution.
L X a X dt
U4 =
3
Where L = Length = 75.00 mm
-6 O
a = (a1+a2)/2 8.1 X 10 / C
-6 O
a1 = Th. Coefficient of Expansion = 4.7 X 10 / C
-6 O
a2 = Th. Coefficient of Expansion = 11.5 X 10 / C
d = 20% of Temp. Limit ± 1ºC 0.2 Deg C
0.122
U4 = = 0.0702 microns
3
5. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
The value of Uncertainty is taken from its calibration certificate i.e. = 0.39 Deg C
Assuming Normal distribution.
L X a X dt
U6 =
2
Where L = Length = 75.00 mm
-6
a = Th. Coefficient of Expansion = 8.1 X 10 / Deg C
d = Uncertainty of temperature Scanner = 0.32 Deg C
0.194
U6 = = 0.0972 microns
2
6 . Standard Uncertainty Due to the uncertainty of Standard Slip Gauge set
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
0.140
U7 = = 0.068 microns
2
7. Standard Uncertainty Due to the uncertainty of Standard Dial Indicator
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
0.30
U8 = = 0.150 microns
2
8. Standard Uncertainty Due to the Accuracy of Master Equipment
Considering half of the accuracy & assuming rectangular distribution
0.12
U8 = = 0.069 micron
3
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 3.773 microns
Degree of Freedom, (Veff)
4
uc(y)
Veff = 4
n(ui(y) )
j=1
Vi
= 3.62159E+48 =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
U = k x Uc 2 x 3.773 microns
7.546 microns
Therefore uncertainty in above measurement is = ± 7.546 microns
Plain Plug Gauge
Determination of Measurement Uncertainty of Plain Plug Gauge
PP Gauge Range : 100 mm Reading points 100 mm
Range 25 75 100 50
Uncertainty in S.G. (mm): = 0.09 0.14 0.17 0.12
Uncer. in Comp Stand(micron) 2.85 microns
Unit of Measurement : = microns
Uncertinty of Temperature Scanner = 0.39 Deg C
Uncer. Of Dial Indicator = 0.3 micron
Accuracy of Slip Gauge Set 0.12 micron
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows-
n
Mean Deviation x = (å xj)¸n
j=1
2
Measured/Observed Readings Standard Value Average (xj-x) (xj-x)
mm mm microns microns
(xj) (x)
99.9864 0.000 0.0200 0.000
99.9866 0.000 0.2200 0.048
99.9864 0.000 0.0200 0.000
99.9866 0.000 0.2200 0.048
99.9866 0.000 0.2200 0.048
99.9864 0.000 99.9864 0.0200 0.000
99.9862 0.000 -0.1800 0.032
99.9860 0.000 -0.3800 0.144
99.9864 0.000 0.0200 0.000
99.9862 0.000 -0.1800 0.032
2
Standard Deviation , S(X) = å (Xj - X)
n-1
= 0.1989 microns
2
[S(X)]
Standard Deviation of the Mean, S ( X) = n
= 0.063 microns
Standard Uncertainty in Type 'A' Evaluation, UA = 0.063 microns
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Comp Stand
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
2.85
U1 = = 1.425 microns
2
2. Standard Uncertainty Due to 50% of Resolution of Standard Dial Gauge
Assuming Rectangular distribution
0.2
U2 = = 0.289 microns
2 3
3. Standard Uncertainty Due to Thermal Coefficien of Master Instrument & UUC .
Assuming rectangular distribution. for Master
L X a X dt
U3 =
3
Where L = Length = 100.00 mm
a = 20% of ( a1 +a2) = 0.0000032
-6 O
a1 = Th. Coefficient of Expansion = 4.7 X 10 / C
-6 O
a2 = Th. Coefficient of Expansion = 11.5 X 10 / C
d = Control Limit or 10% of Temp. 20 °C 1 Deg C
0.3240
U3 = = 0.187 microns
3
4. Standard Uncertainty Due to the Temp. Variation in Master Instrument & UUC.
Assuming rectangular distribution.
L X a X dt
U4 =
3
Where L = Length = 100.00 mm
-6 O
a = (a1+a2)/2 8.1 X 10 / C
-6 O
a1 = Th. Coefficient of Expansion = 4.7 X 10 / C
-6 O
a2 = Th. Coefficient of Expansion = 11.5 X 10 / C
d = 20% of Temp. Limit ± 1ºC 0.2 Deg C
0.162
U4 = = 0.0935 microns
3
5. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
The value of Uncertainty is taken from its calibration certificate i.e. = 0.39 Deg C
Assuming Normal distribution.
L X a X dt
U6 =
2
Where L = Length = 100.00 mm
-6
a = Th. Coefficient of Expansion = 8.1 X 10 / Deg C
d = Uncertainty of temperature Scanner = 0.32 Deg C
0.259
U6 = = 0.1296 microns
2
6 . Standard Uncertainty Due to the uncertainty of Standard Slip Gauge set
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
0.136
U7 = = 0.068 microns
2
7. Standard Uncertainty Due to the uncertainty of Standard Dial Indicator
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
0.30
U8 = = 0.150 microns
2
8. Standard Uncertainty Due to the Accuracy of Master Equipment slip gauge
Considering half of the accuracy & assuming rectangular distribution
0.12
U8 = = 0.069 micron
3
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 3.777 microns
Degree of Freedom, (Veff)
4
uc(y)
Veff = 4
n(ui(y) )
j=1
Vi
= 117040446 =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
U = k x Uc 2 x 3.777 microns
7.554 microns
Therefore uncertainty in above measurement is = ± 7.554 microns
0.851
Snap Gauge
Determination of Measurement Uncertainity of Snap Gauge
Snap Gauge Range : 100 mm Reading Point : 100 mm
Uncertainity in S.G. (mm): = 20 75 100
Uncertainty due to Wringing of S.G. 0.09 0.14 0.17
Unit of Measurement : = microns
Uncertinty of Temperature Scaner = 0.39 Deg C
Accuracy of master equipment 0.12 Micron
Type 'A' Evalution
Five Readings are taken and the deviation from the nominal value is as follows-
n
Mean Devitation x = (å xj)¸n
j=1
= 0.000 microns
Standard Uncertainity in Type 'A' Evalutation, UA = 0.000 microns
Type 'B' Evalution
1. Standard Uncertainity Due to the uncertainty of Standard Equipment used i.e. Slip Gague
The value of uncertainity is taken from its calibration certificate (Up to 20 mm)
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
0.17
U1 = = 0.085 microns
2
2. Standard Uncertainity Due to Thermal Coefficient between UUC & Standard.
Assuming rectangular distribution.
L X a X dt
U2 =
3
Where L = Length = 100.0 mm
a = Th. Coefficient of Expansion of Slip Gague = 4.7X 10-6 / Deg C
dt = Temperature Variation or Temp. Difference = 1 Deg C
0.470
U2 = = 0.271 microns
3
3. Standard Uncertainity Due to the Wringing of Slip Gagues
Assuming Rectangular distribution
-0.20
U3 = = -0.115 microns
3
4. Standard Uncertainity Due to the Temp. Variation in Master Instrument & UUC assumed as 0.5 deg C
Assuming rectangular distribution.
L X a X dt
U4 =
3
Where L = Length = 100.0 mm
a = Th. Coefficient of Expansion of Slip Gague = 6.8 X 10-6 / O C
dt = Temperature Variation or Temp. Difference = 1 Deg C
0.470
U4 = = 0.271 microns
3
5. Standard Uncertainity Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
The value of uncertainity is taken from its calibration certificate i.e. = 0.39 Deg C
Assuming Normal distribution.
L X a X dt
U5 =
2
Where L = Length = 100.0 mm
a = Th. Coefficient of Expansion of Slip Gague = 4.7X 10-6 / Deg C
dt = Uncertainty of temperature scaner = 0.39 Deg C
0.183
U5 = = 0.0917 microns
2
0.12
U6 = = 0.069 microns
3
Combined Uncertainity:
Uc = (UA)2+(U1)2+(U2)2+(U3)2……………………………………………
Uc = 0.425 microns
Degree of Freedom, (Veff)
uc(y)4
Veff =
ån(ui(y)4)
j=1
Vi
= #DIV/0! = ¥
Expanded Uncertainity of Overall Uncertainity or Uncertainity of Measurement :
From thestudent's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
U = k x Uc = 2 x 0.425 microns
0.851 microns
Therefore uncertainty in above measurement is =+ 0.851 microns
Dial Thickness gauge
Determination of Measurement Uncertainty
Dial Gauge Range: 50 mm Least Count: 10 microns
Set of Slip Gauge used for calibration
Size (mm) :- 50 Reading Pionts : 50 mm
Uncertainty in measurement (mm): = 0.12
Unit of Measurement : = microns
Uncertainty of Temperature Scanner = 0.39 Deg C
Accuracy of master Equip. 0.1
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows-
n
Mean Deviation x = ( xj)n
j=1
2
Measured/Observed Readings Standard Value Average (xj-x) (xj-x)
mm mm mm micron microns
(xj) (x)
49.990 0.0000 0.0000 0.000
49.990 0.0000 0.0000 0.000
49.990 0.0000 0.0000 0.000
49.990 0.0000 0.0000 0.000
49.990 0.0000 0.0000 0.000
49.990 0.0000 49.990 0.0000 0.000
49.990 0.0000 0.0000 0.000
49.990 0.0000 0.0000 0.000
49.990 0.0000 0.0000 0.000
49.990 0.0000 0.0000 0.000
2
Standard Deviation , S(X) = (Xj - X)
n-1
= 0.0000 micron
2
[S(X)]
Standard Deviation of the Mean, S ( X) = n
= 0.000 microns
Standard Uncertainty in Type 'A' Evaluation, UA = 0.000 microns
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Set of Slip Gauge
The value of Uncertainty is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
0.12
U1 = = 0.059 microns
2
2. Standard Uncertainty Due to the Resolution of dial thickness gauge
Considering half of the least count & assuming reactangular distribution
10
U2 = = 2.8868 microns
2x 3
3. Standard Uncertainty Due to Thermal Coefficient bitween Master Instrument & UUC
Assuming rectangular distribution.
L X a X dt
U3 =
3
Where L = Length = 50 mm
-6
a = 20% of ( a1 +a2) = 4.60 X 10 / Deg C
a1 = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
a2 = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
d = Control Limit or 10% of Temp. 20 °C 1 Deg C
0.230
U3 = = 0.133 microns
3
4. Standard Uncertainty Due to the Temp. Variation in Master Instrument & UUC
Assuming rectangular distribution.
L X a X dt
U4 =
3
Where L = Length = 50 mm
-6
a = (a1+a2)/2 11.5 X 10
a1 = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
a2 = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
d = 20% of Temp. Limit ± 1ºC 0.2 Deg C
0.115
U4 = = 0.066 Deg C
3
5. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
Assuming Normal distribution.
The value of Uncertainty is taken from its calibration certificate i.e. = 0.39 Deg C
L X a X dt
U7 =
2
Where L = Length = 50 mm
a = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
d = Uncertainty of temperature scanner = 0.39 Deg C
0.224
U7 = = 0.112 Deg C
2
0.10
U6 = = 0.058 microns
3
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 2.894 microns
Degree of Freedom, (Veff)
4
uc(y)
Veff = 4
n(ui(y) )
j=1
Vi
=#DIV/0! =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =. 2
U = k x Uc = 2 x 2.894 microns
5.788 microns
Therefore uncertainty in above measurement is = ± 5.8 microns
comparator stand
Determination of Measurement Uncertainity of comparator stand
1.3499 microns
[S(X)]2
Stabdard Devitation of the Mean, S ( X) = n
= 0.4269 microns
3. Standard Uncertainity Due to the uncertainty of Standard Equipment used i.e. Puppy Dial
The value of uncertainity is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab = 2
1.51
U3 = = 0.755 microns
2
4. Standard Uncertainity Due to the uncertainty of Standard Equipment used i.e. Surface Plate
The value of uncertainity is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab 2
0.00
U4 = = 0.000 microns
2
5. Standard Uncertainity Due toThermal Coefficient between UUC & Standard
Assuming rectangular distribution.
L X a X dt
U5 =
3
Where L = Length = 250.000 mm
a = Th. Coefficient of Expansion = 11.5 X 10-6 / Deg C
dt = Control Limit or 10% of Temp. 20 °C 1.0 Deg C
2.875 1.660
microns
3
7. Standard Uncertainity Due to the Temp. Variation in Master Instrument & UUC assumed as 0.5 deg C
Assuming rectangular distribution.
L X a X dt
U7 =
3
Where L = Length = 250 mm
a = Th. Coefficient of Expansion = 11.5 X 10-6 / o C
o
d = 20% of Temp. Limit ± 1ºC 0.2 C
0.575
U7 = = 0.3320 microns
3
8. Standard Uncertainity Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
The value of uncertainity is taken from its calibration certificate
L X a X dt
U8 =
2
Where L = Length = 250 mm
a = Th. Coefficient of Expansion o = 11.5 X 10-6 / o C
o
d = Uncertainty of temperature scaner = 0.39 C
1.121
U8 = = 0.6474 microns
2
12. Standard Uncen. due to Accuracy Master Equipment. LTDG
Considering half of the accuracy & assuming rectangular distribution
3.00
U12 = = 1.732 microns
3
5.00
U13 = = 2.887 microns
3
Combined Uncertainity:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 3.921 microns
Ue = ke x Uc = 2 x 3.921 microns
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 4.731 microns
Degree of Freedom, (Veff)
uc(y)4
Veff =
n(ui(y)4)
j=1
Vi
= 2.40412E+57 =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
Ue = ke x Uc = 2 x 4.731 microns
Therefore uncertainty in above measurement is = ± 9.46 microns
Dial Gauge
Determination of Measurement Uncertainty
Range : 50 mm Least Count: 10 microns
Size of slip 50.0
Uncer. Of S.G.(micron) 0.12
Uncertainty Of Comparator stand(micron) 2.85
Unit of Measurement : = microns Reading Point : 50 mm
o
Uncertainty of Temperature Scanner = 0.39 C
Accuracy Slip Gauge 0.1 Comparator 6 microns
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows-
n
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 4.733 microns
Degree of Freedom, (Veff)
uc(y)4
Veff =
n(ui(y)4)
j=1
Vi
= #DIV/0! =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
Ue = ke x Uc = 2 x 4.733 microns
Therefore uncertainty in above measurement is = ± 9.47 microns
Dial Gauge
Determination of Measurement Uncertainty
Range : 25 mm Least Count: 10 microns
Size of slip 20.0 5
Uncer. Of S.G.(micron) 0.09 0.09
Uncertainty Of Comparator stand(micron) 2.85
Unit of Measurement : = microns Reading Point : 25 mm
o
Uncertainty of Temperature Scanner = 0.39 C
Accuracy Slip Gauge 0.1 Comparator 6 microns
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows-
n
Combined Uncertainty:
2 2 2 2……………………………………………
Uc = (UA) +(U1) +(U2) +(U3)
Uc = 4.732 microns
Degree of Freedom, (Veff)
uc(y)4
Veff =
n(ui(y)4)
j=1
Vi
= #DIV/0! =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
Ue = ke x Uc = 2 x 4.732 microns
Therefore uncertainty in above measurement is = ± 9.46 microns
Measuring Pin
Determination of Measurement Uncertainty
Range : 20 mm Least Count: microns
Size of slip 20.0 0
Uncer. Of S.G.(micron) 0.09 0
Uncertainty Of Comparator stand(micron) 2.85
Unit of Measurement : = microns Reading Point : 20 mm
o
Uncertainty of Temperature Scanner = 0.39 C
Accuracy Slip Gauge 0.1 Comparator 6 microns
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows-
n
Combined Uncertainty:
Uc = (UA)2+(U1)2+(U2)2+(U3)2……………………………………………
Uc = 3.759 microns
Degree of Freedom, (Veff)
uc(y)4
Veff =
n(ui(y)4)
j=1
Vi
= 1345169092 =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
Ue = ke x Uc = 2 x 3.759 microns
Therefore uncertainty in above measurement is = ± 7.52 microns
ultra sonic gauge
Determination of Measurement Uncertainty of ultra sonic gauge Up to 150 mm
Range : 100 mm Least Count: 10 microns
Size of Caliper 100
Uncer. Of S.G.(micron) 0.17 0
Uncertainty Of surface plate(micron) 4.6
Unit of Measurement : = microns Reading Point : 100 mm
o
Uncertainty of Temperature Scanner = 0.39 C
Accuracy Slip Gauge 0.12 Surface Plate 5 microns
Type 'A' Evaluation
Five Readings are taken and the deviation from the nominal value is as follows-
n
2
Measured/Observed Readings Standard Value Avearge (xj-x) (xj-x)
mm mm mm microns microns
[S(X)]2
Standard Deviation of the Mean, S ( X) = n
= 1.6667 microns
Standard Uncertainty in Type 'A' Evaluation, UA = 1.6667 microns
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. slip gauge
The value of Uncertainty is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
0.17
U1 = = 0.085 microns
2
2. Standard Uncertainty Due to the wringing of Standard Equipment used i.e. slip gauge/Slips
The value of Uncertainty is taken from its calibration certificate
Assuming Rectangular distribution
0
U2 = = 0.000 microns
1.732
3.Standard Uncertainty Due toThermal Coefficient between UUC & Standard
Assuming rectangular distribution.
U3 = L X a X dt
3
Where L = Length = 100.000 mm
a = 20 % of a1+a2 3.24 X 10-6 / Deg C
a1 = Th. Coefficient of Expansion of DStandard = 4.7 X 10-6 / Deg C
a2 = Th. Coefficient of Expansion of UUC = 11.5 X 10-6 / Deg C
dt = Control Limit or 10% of Temp. 20 °C 1.0 Deg C
0.320 0.185 microns
3
4.Standard Uncertainty Due to the Resolution of ultra sonic gauge
Considering half of the least count & assuming rectangular distribution
10
U4 = = 2.887 microns
2 x 1.732
5.Standard Uncertainty Due to the Temp. Variation in Master Instrument & UUC
Assuming rectangular distribution.
L X a X dt
5 =
3
Where L = Length = 100.00 mm
-6
a = (a1+a2)/2 11.5X 10
-6 O
a1 = Th. Coefficient of Expansion = 11.5X 10 / C
-6 O
a2 = Th. Coefficient of Expansion = 11.5 X 10 / C
d = 20% of Temp. Limit ± 1ºC 0.2 Deg C
0.230
U5 = = 0.1328 microns
3
6.Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Temp.Scanner
The value of Uncertainty is taken from its calibration certificate i.e. =0.35
L X a X dt
U6 =
2
Where L = Length = 100 mm
-6 o
a = Th. Coefficient of Expansion of slip gauge = 11.5 X 10 / C
o
d = Uncertainty of temperature scanner = 0.39 C
0.449
U6 = = 0.2243 microns
2
7.Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Surface Plate
The value of Uncertainty is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab =2
4.60
U7 = = 2.300 microns
2
Combined Uncertainty:
Uc = (UA)2+(U1)2+(U2)2+(U3)2……………………………………………
Uc = 4.985 microns
Degree of Freedom, (Veff)
4
uc(y)
Veff = 4
n(ui(y) )
j=1
Vi
= 720.2495681 =
Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :
From the student's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k =2.
Ue = ke x Uc = 2 x 4.985 microns
Therefore uncertainty in above measurement is = ± 10.0 microns