MOY/SCMI/36 (Issue 4):2001
Centre for Basic, Thermal and Length Metrology
                       National Physical Laboratory
                               MOY/SCMI/36
       SPECIFICATION OF ACCURACY FOR A
            PRECISION CLINOMETER
A “Watts” Precision Clinometer fitted with a circular glass scale and an optical
micrometer graduated to read direct to 1/6th minute of arc (10 seconds of arc). A
bubble unit is supplied for fiducial setting.
Made by:      Messrs. Hilger & Watts Ltd. (Watts Division).
National Physical Laboratory
                                                       MOY/SCMI/36 (Issue 4):2001
FOREWORD
In the 1940s and 1950s, NPL was involved in drafting a special series of
Specifications of Accuracy that covered a wide range of precision measuring
apparatus. This series has been built on first hand experience gained in the design and
construction of prototype measuring equipment at NPL and in the design and
calibration of measuring equipment of British and foreign manufacture. Each
specification in the series originally conformed to a general pattern and was allocated
a permanent serial number which, in addition to its title, serves as its identity.
The MOY/SCMI (Metrology/Specification Certification Measuring Instruments)
standards are complementary to the Standards issued by the British Standards Institute
(BSI). The majority relate to measurement equipment of a proprietary kind designed
either at NPL or by British manufacturers which, in the ordinary way, would not fall
within BSI’s terms of reference. In some cases, in which the equipment is of a more
general nature, the Specification has provided a useful basis for formulating a British
Standard. The specifications are to enable manufacturers to base their inspection on
mutually agreed specifications of accuracy both in workmanship and performance.
MOY/SCMI/36 has been updated as part of a project financed by the DTI (MPU
8/61.3) concerned with Good Practice Guides and Equipment Specifications.
SCOPE
A “Watts” Precision Clinometer fitted with a circular glass scale and an optical
micrometer graduated to read direct to 1/6th minute of arc (10 seconds of arc). A
bubble unit is supplied for fiducial setting.
Made by:        Messrs. Hilger & Watts Ltd. (Watts Division).
Figure 1 A Clinometer (5 second graduations)
                                               1
                                                                             MOY/SCMI/36 (Issue 4):2001
CONTENTS
                                                                                                                   Page
1   GENERAL .............................................................................................................3
2   ALIGNMENTS ......................................................................................................3
3   BUBBLE UNIT......................................................................................................4
4   ACCURACY..........................................................................................................4
5   UNCERTAINTIES ................................................................................................5
6   REPORTING OF COMPLIANCE ........................................................................5
                                                          2
                                                MOY/SCMI/36 (Issue 4):2001
1     GENERAL
1.1   The general workmanship and finish shall be in keeping with a
      precision instrument of this class.
1.2   The instrument shall be marked with an identification number and with
      the maker’s name or trademark.
1.3   The circular glass scale and its optical micrometer scale shall focus
      simultaneously. The position of normal focus of these two scales shall
      be approximately in the centre of the eyepiece travel.
1.4   The graduation lines of both scales shall be cleanly cut and free from
      blemishes.
1.5   The readings of the exterior setting scale and those of the circular glass
      scale shall agree to within ± 0.5 degree.
1.6   The value of the scale divisions of the optical micrometer shall be
      clearly indicated.
1.7   The fine-setting device shall operate smoothly.
1.8   The steel base shall have a hardness of not less than 750 HV, when
      tested in accordance with BS EN ISO 6507-1:1998.
2     ALIGNMENTS
      NOTE: The front bearing edge shall be regarded as the datum edge for
      the clinometer.
2.1   The underside of the base shall be flat. When the instrument is placed
      on a truly flat surface it shall be free from rock to within 0.013 mm.
2.2   The front and side bearing edges of the base shall be flat to within
      0.013 mm.
2.3   The front and side bearing edges of the base shall be mutually square to
      within 0.05 mm as measured over the longer side.
2.4   The front and side bearing edges of the base shall be square to the
      underside of the base to within 0.025 mm over their depth.
                                   3
                                               MOY/SCMI/36 (Issue 4):2001
3     BUBBLE UNIT
      NOTE: This unit shall be in general accordance with the requirements
      of BS 958 with regard to:
      Scale
3.1   It is recommended that the scales on all levels be divided into 2.5 mm
      divisions (0.1 inch).
      Marking
3.2   It is recommended that the glass tube of the level be etched with the
      nominal sensitivity in terms of a base length of 10 in. (or in terms of
      mm per metre), e.g.
        1 div = 0.001 in./10 in.
       (1 div = 0.1 mm/metre)
      Accuracy of Calibration
3.3   When tested on a tilting table, readings being taken at each division at
      both ends of the bubble, the level should comply with the following
      limits of error:
      (i)     The average sensitivity shall not differ from the value marked
              on the level by more than 10 per cent
      (ii)    The maximum deviation of the value of any one division from
              the average value shall not exceed 0.2 division
4     ACCURACY
4.1   Identical fiducial settings of the bubble unit shall be made and the
      departure of any one reading from the mean of six readings shall not
      exceed one half micrometer scale division (5 seconds of arc).
4.2   The maximum error of indication between any two readings, including
      those of the optical micrometer scale, shall not exceed one-micrometer
      scale division - 10 seconds of arc.
                                  4
                                                MOY/SCMI/36 (Issue 4):2001
5     UNCERTAINTIES
5.1   It will normally be necessary to consider the uncertainty of
      measurement when ascertaining compliance (or non-compliance) with
      this specification. UKAS document M3003 ‘Uncertainty and
      confidence in measurement’ gives guidance in Appendix J.
6     REPORTING OF COMPLIANCE
6.1   Certain clauses in any specifications are necessary to support
      manufacture and assembly but may be difficult or unnecessary to check
      in subsequent checks for compliance with this specification. In certain
      cases checking a feature may require disassembly of the item, which
      may be undesirable. Although it is not essential that all clauses be
      checked on subsequent verification, it is important that those clauses
      omitted do not detract from the metrological value of the test. Where
      applicable, a performance check should always be carried out as this
      may allow indirect verification of those parameters that are not easily
      measured individually without disassembly.
6.2   When making statements of compliance or non-compliance, it is
      recommended that this specification and the relevant clauses within it
      be unambiguously identified in the calibration certificate or test report.
      Example wording for a set of angle gauges follows.
      This set of angle gauges has been examined for compliance with the
      accuracy requirements of clauses 2 and 3 of NPL Specification of
      Accuracy MOY/SCMI/18 (Issue 5), a copy of which is attached to this
      certificate.
                                  5
                                                        MOY/SCMI/36 (Issue 4):2001
For free measurement advice and information on other specifications in this series call
the NPL Help line on 020 8943 6880
E-mail: enquiry@npl.co.uk
Internet: www.npl.co.uk
National Physical Laboratory
Queens Road
Teddington
Middlesex
TW11 0LW
Further information on angle measurement can be found in ‘Measurement of Angle in
Engineering’ by J C Evans and C O Taylerson (Third Edition Revised by E W Palmer
and S P Poole).
The standard reference temperature for industrial length measurements is defined in
ISO 1:1975 Standard reference temperature for industrial length measurements.
United Kingdom Accreditation Service (UKAS) can be contacted at
21-47 High Street
Feltham
Middlesex
TW13 4UN
Tel 020 8917 8400
Internet: www.ukas.com
MOY/SCMI/36 (Issue 4):2001