Manual Q400 - 500
Manual Q400 - 500
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
BATTERY INSTALLATION AND REMOVAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Q-400 and
UP AND RUNNING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Turning the Dosimeter On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Turning the Dosimeter Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Starting and Stopping the Logging of Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Q-500 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
SAMPLE PRINTOUTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Header Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Calibration Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setup Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Time Summary / Data Summary Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Event Summary Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Time History Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
% Time Statistical Distribution Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Exceedance Levels Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
56-253
REV. G
11/98
INTRODUCTION
The Q-400 and the Q-500 accumulate, calculate, and display noise measurement data. (See Figure 1.)
SIMPLIFIED PROGRAMMING INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
PROGRAMMING MENU STRUCTURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
CAL Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
PRINT Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
MENU ON/OFF Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
LOGGING TIME LIMITATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
ENERGY OVERLOAD . . . . . . . . . . . . . . . . . . . . . . 67
PRINCIPLES OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
General Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Microphone Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Weighting Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
ACCURACY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Accuracy of Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Microphone Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Displayed Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Microphone / Preamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Data Interface Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Power Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Questware Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Windscreens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Calibrator Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Tripods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
1
BATTERY INSTALLATION AND REMOVAL
Either dosimeter will function as a Personal Noise Dosimeter, an Environmental Monitor, or an Event Monitor. The Q400
provides Type 2 measurement accuracy, while the Q500 is a precision Type 1 instrument. The battery must be a 9 Volt alkaline type.
(Examples are: NEDA 1604A, IEC 6LF22, or IEC 6LR61)
# When used as a Personal Noise Dosimeter: # Replace the battery as follows (see Figure 2):
The dosimeter may be clipped on a belt or worn in a shirt pocket. The small microphone is simply clipped 1.) Remove the battery cover.
to the shirt or shirt collar near the ear.
2.) Lift the non-terminal end of the battery out of the battery compartment.
# When used as an Environmental Monitor or Events Monitor: 3.) Remove the battery.
The dosimeter may be either hand held or mounted to a standard camera tripod. The microphone is 4.) Notice the battery orientation drawing on the battery cover. Be sure that the battery is properly oriented
connected to the "Microphone Boom" and the unit is held similarly to a sound level meter. with respect to the battery compartment terminals.
5.) Angle the non-terminal side of the battery into the battery compartment, and press the terminal end down
# When used with a Personal Computer: into the holder.
The Quest Noise Manager Software package allows the user to analyze and print out pre-recorded data in 6.) Install the battery cover.
extreme accuracy and detail.
Note: Battery life is affected by the dosimeter measuring setup parameters. Refer to the specifications for details.
# When used with a Printer:
The dosimeter can print detailed reports of noise events. It connects to a printer by using a Quest "Parallel
Printer Interface" or "Serial Computer Interface".
The dosimeter can function as two simultaneously operating noise dosimeters. When this is done, each is programmed
with a different set-up. The dosimeter can also be programmed to perform the C - A measurement.
An internal battery powers the clock and memory. When this battery's voltage gets low, data memory and setup will not be
retained when a 9V battery is not in the dosimeter. The memory only draws current from this battery when a 9V battery is
not in the unit. Contact the factory or a Quest service center for replacement.
2 3
AC/DC ADAPTER INSTALLATION CABLE REMOVAL OR ATTACHMENT
Quest part number 056-973
Remove or attach the Cable Connector as follows:
The dosimeter can operate from one of the following power sources if desired:
# To remove the Cable Connector from the dosimeter:
# Any AC power source (9 to 18 VAC, 35 mA minimum)
(Such as the Quest Model 920 AC Power Supply) # Gently grasp and pull the Knurled Ring of the Cable Connector.
# Any DC power source (12 to 24 VDC, 35 mA minimum) # To attach the Cable Connector to the dosimeter:
(The polarity of the DC Plug does not matter.)
1.) Grasp the black Rubber Boot of the Cable Connector.
The power source must have a 3.5mm (1/8") phone plug.
2.) Gently press the Cable Connector against the dosimeter connector while slowly rotating it. When it is
Simply install the AC/DC Adapter into the dosimeter battery compartment in place of the battery as follows: properly lined up, it will stop rotating and slide into the dosimeter.
# Remove the battery from the battery compartment as follows: 3.) Insert the cable connector until a "click" is heard.
The Cable Connector is now attached.
1.) Remove the battery cover. (See Figure 2.)
2.) Lift the terminal end of the battery out of the battery compartment.
3.) Remove the battery.
2.) Snap the AC/DC Adapter into the dosimeter and fasten the cover.
3.) Plug the desired power source (see above) into the AC/DC Adapter jack.
The dosimeter is now ready for operation.
Note: Safety approvals for dosimeter use in hazardous locations are for battery operation only. Do not use the AC/DC
adapter in an environment classified as hazardous by the local electrical code or governing agency.
4 5
USING THE MICROPHONE BOOM CALIBRATION
The Microphone can be mounted on the Microphone Boom. This makes the dosimeter easier to use when hand-held or
tripod-mounted. The dosimeter should be calibrated before each use.
The Microphone Boom attaches to the back of the dosimeter as follows: To calibrate the microphone sensitivity:
1.) Connect the Microphone to the Microphone Boom in one of the following ways: ))))> "PRE-SURVEY"
Note that PRE-SURVEY Calibration can
# Clip the Microphone to the Microphone Boom. CAL Key only be done after a unit RESET.
"PRE-SURVEY" (No Run Time stored in the unit)
# Unscrew the Microphone Clip from the Microphone and screw the Microphone Boom to the
Microphone. To periodically check the Acoustic Calibration:
2.) Carefully wind the cable onto the upper portion of the Microphone Boom. ))))> "PERIODIC CHECK"
3.) Insert the Microphone Boom Pin into the center hole on the Belt Clip. CAL Key
"PERIODIC CHECK"
4.) Fasten the bottom of the Microphone Boom to the Belt Clip with the Microphone Boom Thumbscrew.
Periodically check the Acoustic Calibration of the dosimeter during and after use. Up to 8 Calibration Levels and Calibration
The exposed end of the Microphone Boom Thumbscrew is threaded so that a tripod can be connected to it. Times can be saved throughout a measurement period.
To review the stored Calibration Levels and Times:
UP AND RUNNING
))))> "REVIEW"
Turning the Dosimeter On: CAL Key
"REVIEW"
1.) ))))> " QUEST ", the Software Revision Level and "INITIALIZING" will appear. If
"CRITICAL RESET" appears, see APPENDIX, Displayed Warning Messages. See "DETAILED PROGRAMMING INSTRUCTIONS" if needed.
2.) ))))> "SETUP" menu will appear. The unit is now ready to operate.
2.) ))))> Press and hold the key. "OFF-5" will count down to "OFF-0" and the dosimeter
will turn off.
1.) ))))> Press to "RUN" (Start Logging) and press again to "PAUSE" (Stop Logging).
The dosimeter will maintain its accuracy for many months of use.
However, it is recommended that the dosimeter be returned annually to a Quest Authorized Service Station for a
recalibration.
Calibration standards, traceable to the National Institute of Standards and Technology (NIST) are maintained and used by
6 7
Quest. ABOUT THE DISPLAY
OL - Overload indicator lights when the sound level exceeds the measurement range of the instrument. While in RUN, this
indicator stays lit until the dosimeter is RESET.
SLOW or FAST - indicates the time response of the measurement being viewed.
LOBAT - when first lit, indicates 8 hours of measurement time remaining on the battery.
Display Function Structure
8 9
"PKTime" Time that Event Peak Occurred Display Operation
"MaxTime" Time that Event Max Occurred
Display Function Keys are as follows:
"MinTime" Time that Event Min Occurred
"LoggTime" Remaining Logging Time
# To alternate between Dosimeter 1 and Dosimeter 2 for LEVELS, DOSE, AVG, or TIMES:
2.) Use the UP ARROW and DOWN ARROW Keys to move the desired function so that the White Arrow (>)
points to it.
# To alternately change the size of the numbers pointed to by the White Arrow:
# Alternately press the ENTER Key.
NOTE: See "DISPLAY FUNCTION STRUCTURE" to see which functions are controlled
with each Display Function Key.
10 11
SECURITY CODES USING THE EVENT MODES
The Security Function allows you to "Lock-Out" another person's access to the dosimeter. The Event Modes (Manually-Entered or Level-Triggered) can be used to store, review, and print out up to 999 events.
# A Security Code is needed to place the dosimeter into the Secured mode.
Storing Manually Entered Events
# Only the number keys will operate when the security feature is enabled.
The dosimeter will store an Event each time that it is Run and Paused.
# The Security Code is also needed to exit the Secured mode.
Be sure that the Level-Triggered Event Mode is off (DISABLED).
# When the Security Function is disabled, all keys function normally. To disable it, see the "DETAILED PROGRAMMING INSTRUCTIONS" if needed.
"SECURE"
"SECURE ENABLE" Select to Enable Security Operate the Manually-Entered Event Mode as follows:
"CHANGE CODE" Select to Enter New Security Code 1.) Reset the dosimeter. This clears all stored data.
Note that the Security Function can be used with either Level Triggered Events or Auto-On. 2.) ))))> "EVENTS / VIEW"
See the "DETAILED PROGRAMMING INSTRUCTIONS" for more information about the various modes of operation. # "001" will be highlighted above "Cev" (Current Event).
The dosimeter arrives from the factory with the 4 digit security code set to 1111. This code is initially needed to allow
access to the dosimeter "CHANGE CODE" feature. You can then change the code to your own 4 digit security code: 3.) ))))> "RUN" or "PAUSE"
Once you have changed the security code, be sure that you don't forget it! If you do, once the unit is placed into the # Press to start and stop each Event. An Event will be stored each time that this is done. (Current
SECURE Mode ("SECURE ENABLE"), you will not be able to disable it unless you know the exact code. Event 001 will increase to 002, 003, etc. as each event is stored.)
# Accumulate as many Events as desired. The number over "Cev" (Current Event) is the last event
Memorize your code and, if you wish, record the code here.
number that has been recorded.
OPERATOR SECURITY CODE ____ ____ ____ ____
# To review the data, see "Reviewing Events" (located within this section).
# FAX Quest at 1-262-567-4047. Quest will FAX you a security code that will allow you to gain access to the Security 2.) ))))> "SETUP" menu will appear.
Function.
12 13
Storing Level-Triggered Events Reviewing Events
The dosimeter will store an Event each time that the sound level goes above the "LEVEL ON" setting and then goes below
the "LEVEL OFF" setting. Each stored Event can be reviewed as follows:
Be sure that the Level-Triggered Event Mode is enabled. 1.) ))))> "EVENTS / VIEW"
To program the dosimeter to operate in the Level-Triggered Event Mode, see "DETAILED PROGRAMMING
INSTRUCTIONS" if needed. # "XXX" (representing the Current Event Number) will be highlighted above "Cev" (Current Event).
# To review the data, see "Reviewing Events" (located within this section).
14 15
Dosimeter 1 or 2 Event Information can be reviewed for each Display Function Key: C - A WEIGHTING
Select C - A by programming the Response and Weighting for Dosimeter 1 and Dosimeter 2 as follows:
))))> "SPL" * Sound Pressure Level
"PEAK" * Absolute Unweighted Peak
"MAX" * Maximum Level "SETUP"
"MIN" * Minimum Level "RESP/WT"
"DOS1" Set both dosimeters to SLOW or
"DOS2" set both dosimeters to FAST.
They must be set identically!
))))> "DOSE" * Dose "WGHT" Weighting, Select:
"PrD.8" * 8 Hour Projected Dose "1 = A" / "2 = C"
"PrD.(Prt)" * Variable Projected Dose
"EXP" * Exposure # Also, both dosimeters (1 & 2) must have the following three parameters set identically:
# "TL" Threshold Level
# "ER" Exchange Rate
# "CL" Criterion Level
))))> "SEL" * Sound Exposure Level
"LAV" * Average Level (or LEQ) See "DETAILED PROGRAMMING INSTRUCTIONS" if needed.
"TWA" * Time Weighted Average # C - A information is only presented in the printout when the above settings are properly set. C - A information is not
"TWA (Prt)" * Projected TWA available from the dosimeter display.
The dosimeter calculates C - A (Dosimeter 2 - Dosimeter 1) for the following parameters:
))))> "RunTime" Event Run Time Lavg (dB) or Leq when Exchange Rate is 3 dB
TWA (dB) Time Weighted Average, 8 Hours
"PKTime" Time when Event Peak Occurred TWA [Prt] (dB) Projected TWA, Projection Time
"MaxTime" Time when Event Max Occurred Dose (%) Dose accumulated over Run Time
"MinTime" Time when Event Min Occurred Dose [8] (%) Projected Dose, always 8 Hour
Dose [Prt] (%) Projected Dose, Projection Time
"LoggTime" Remaining Logging Time SEL [ER] (dB) Sound Exposure Level, 1 Second
Pa2H Exposure in Pascal Squared Hours
# If desired, press to alternately enlarge or shrink the character size of the function pointed to by the
white arrow. Note that the "TIMES" information does not respond this way. It always remains When C - A is properly set up, C - A data samples will be calculated as follows:
small.
# ["C" Weighted Data] - ["A" Weighted Data] = ["C - A" Data]
NOTE: See "APPENDIX, Acoustical Definitions" for more detail about any items preceded
by "*".
16 17
USING AUTO-ON / TIMED RUN NOTES:
Program the Automatic Start/Stop function as follows: # The unit must be Off when waiting to go into the Auto-On mode.
# When the unit starts, there is a 10 second "INITIALIZATION" (warm-up) period where no data is gathered.
"AUTO-ON" If this is of concern, allow for the 10 seconds when programming the Start Time.
"ENABLE Y/N" Turn Auto-On/Timed Run On or Off # The Auto-On should be disabled after the study. Otherwise, when the unit is turned off and put away, it
"MODE DAILY" Program Daily Start/Duration may turn on again automatically. This would accumulate useless data and also wear out the battery.
"MODE WEEKDAY" Program Days of Week, Start/Duration
"MODE DATE" Program One Day for Start/Duration
MEASURING Ldn
See DETAILED PROGRAMMING INSTRUCTIONS if needed.
Ldn is a measurement where 10 dB is automatically added to all data between 2200 (10P.M.) and 0700 (7A.M.).
To Activate or De-activate the Auto-On / Timed Run Feature: Ldn operates automatically (only on DOSIMETER 1) if it is programmed into this mode.
A.) Select "Y" (Yes) or "N" (No) for "ENABLE". When Ldn is programmed Yes [Y], it automatically programs DOSIMETER 1 to the following:
To Start and Log at a specific time on selected days of the week: "SETUP"
"DOSE1"
A.) Select "MODE WEEKDAY". *"LDN" [Y] (Day/Night Levels)
B.) Enter the desired Day of the Week that the unit is to Log. (SMTWTFS) See DETAILED PROGRAMMING INSTRUCTIONS if needed.
C.) Enter the Start Time and the Duration. It will Start and Log identically for each day that is selected. NOTES:
# Ldn will be logged only when Ldn is enabled during the measurement period.
To Start and Log on one specific date:
# Ldn will be printed only if Ldn has been logged during the measurement period.
A.) Select "MODE DATE".
# The Event mode is disabled while running Ldn.
B.) Enter the desired Date that the unit is to Log. (MONTH and DAY).
C.) Enter the Start Time for that Date and the Duration. The unit will Start and Log on only that Date. NOTE: See "APPENDIX, Acoustical Definitions" for more detail about any items preceded
by "*".
B.) Proceed to "DURATION" and set the desired Timed Run length.
C.) Any time that the RUN Key is pressed, the unit will do a Timed Run for that Duration.
18 19
PRINTER USE 3.) Remove the microphone from the dosimeter and connect the printer.
Printer Connection 4.) Turn on the printer, the Interface Module, and the dosimeter.
The dosimeter requires one of the following INTERFACE Modules to connect to the printer (or computer). Selecting What to Print
3.) Insert the other cable connector (attached to the flat cable) into the Printer Port. # The EOL (End of Line) characters can be:
Preparing to Print # Both set within the printer configuration.
Only data that has been selected for logging can be printed. # Both set within the dosimeter configuration.
Use the dosimeter with a Printer as follows: # One setting within the printer configuration and one setting within the dosimeter configuration.
1.) Accumulate data with the dosimeter. # The dosimeter can be programmed to one of the following four settings:
2. Place it in the Pause Mode. LF/CR (Line Feed followed by Carriage Return)
CR/LF (Carriage Return followed by Line Feed)
LF (Line Feed only)
CR (Carriage Return only)
# To try a few lines of print, depress PRINT twice. If the printer does not work properly, change the
dosimeter setting until it works.
# If the system still does not print properly, you may need to change the printer setup. Refer to the printer
manual to aid in the printer configuration. After changing the printer setup, it is usually necessary to reset
20 21
the printer as follows: SAMPLE PRINTOUTS
# Turn the printer off for a few seconds and then back on. Most printers only read these switches The following pages describe each section of the printout.
during the turn on.
Each item may be selected from the "PRINT SELECTED" menu (below) which will affect:
If more information is needed as to how to change the dosimeter settings, see "DETAILED PROGRAMMING
INSTRUCTIONS". # What specific Section is to be printed.
# What data is to be included within Printout Sections.
# Parallel Printer Configuration
# If the printer operates in either the Serial or Parallel mode, be sure that it is set to Serial. "PARAMETERS" [Y/N] Dosimeter Setup Parameters
"SUMMARY" [Y/N] Time and Data Summary
# The printer must be RS-232C compatible.
"PEAK" [Y/N] Peak Levels
# Match the dosimeter's and the printer's baud rates. "U.L." [Y/N] Upper Limit Time
# 300 600 1200 2400 4800 9600 19200
"STATISTICS" [Y/N] % Time Statistical Distribution
# Match the dosimeter's and the printer's data formats. "LN LEVELS" [Y/N] Exceedance Levels, L01 - L100
"EVENTS" [Y/N] All Events
# Each character sent from the dosimeter consists of 10 bits:
# 1 Start bit "TIME HISTORY" Time History
# 8 data bits "LAVG" [Y/N] Average Time History
# 1 Stop Bit "MAX" [Y/N] Maximum Level Time History
# No Parity (Disabled)
"PEAK" [Y/N] Peak Level Time History
# Match the dosimeter's and the printer's Handshake options. "LC-A" [Y/N] C minus A Time History
# NONE HARDWARE XON/XOFF "TIME HISTORY" Time History
"TABULAR/GRAPHICAL" Time History Format
"SAMPLES/LINE" Enter 1 to 120
Use the UP and DOWN arrow keys to advance through the print selections. Depressing the ENTER key toggles between Y
or N.
Note that the Time History Printout (Figures 10, 11, and 12) is printed in both the "GRAPHICAL" and "TABULAR" formats.
This printout has been shortened and rearranged to provide a good example of what the various printouts will look like.
22 23
Header Section Figure 5 Calibration Section Figure 6
The Header Printout will occur only if: The Calibration Printout will occur only if:
# Header is selected. # Calibration is selected.
# "HEADER" (Selected if Yes) # "CALIBRATION" (Selected if Yes)
QUEST ELECTRONICS
Q-400 Noise Logging Dosimeter
Name _____________________________________________________________________
__________________________________________________________________________
24 25
Setup Section Figure 7 Time Summary / Data Summary Section Figure 8
The Setups for Dosimeter 1 and Dosimeter 2 Parameters are printed out as follows: Peak Level and/or Upper Limit Time will print within Dosimeter 1 or 2 Data Summaries only if one or both of the following
are selected:
# "PEAK" (Selected if Yes)
# "U.L." (Selected if Yes)
Printed data within the Summary is always computed for the Total Run Time (total of all events).
The
Tim
Auto Settings: e Time Summary:
Auto-On Disabled Mode - Daily Time 07:00:00AM Duration 00:02 H:M and Number of Events 1
Level Triggered Events Off Event Started Event Stopped
Dat
15-JAN-93 @ 10:56:38AM 15-JAN-93 @ 10:58:42AM Event 1
Dosimeter 1 Parameters: a
Range 70-140dB Weighting A Time Constant Slow Su Total Run 0:02:00 Total Pause 1:22:51
Criterion 90dB Threshold 80dB Exchange Rate 5dB mm
Prj Period 4.00H Upper Limit 115dB Ldn Off Data Summary [Dosimeter 1, A / Slow, Threshold 80dB, Exchange Rate 5dB]
ary
Peak Level 102.6dB 15-JAN-93 @ 10:56:42AM
Dosimeter 2 Parameters: are Max Level 88.0dB 15-JAN-93 @ 10:56:42AM
Range 70-140dB Weighting C Time Constant Slow prin Min Level 73.4dB 15-JAN-93 @ 10:56:42AM
Criterion 90dB Threshold 80dB Exchange Rate 5dB ted UL Time 0:00:00
Prj Period 4.00H Upper Limit 115dB Lavg 80.5dB Dose 0.11% SEL(5) 115.1dB
as
TWA 41.0dB Dose[8] 26.40% Ldn OFF
foll TWA[ 4.00] 75.5dB Dose[ 4.00] 13.20%
ow
s: Data Summary [Dosimeter 2, C / Slow, Threshold 80dB, Exchange Rate 5dB]
Peak Level 102.6dB 15-JAN-93 @ 10:56:42AM
Max Level 88.6dB 15-JAN-93 @ 10:56:42AM
Min Level 77.3dB 15-JAN-93 @ 10:56:39AM
UL Time 0:00:00
Lavg 84.5dB Dose 0.19% SEL(5) 119.0dB
TWA 44.9dB Dose[8] 45.60%
TWA[ 4.00] 79.5dB Dose[ 4.00] 22.80%
26 27
Event Summary Section Figure 9
__________________________________________________________________________
An Event Summary Printout will occur only if A and B are selected as follows:
__________________________________________________________________________
A.) Events is selected: Event Started Event Stopped
15-JAN-93 @ 10:56:38AM 15-JAN-93 @ 10:58:42AM
# "EVENTS" (Selected if Yes)
Run Time 0:02:00
B.) One or both of the following are selected: Data Summary [Dosimeter 1, A / Slow, Threshold 80dB, Exchange Rate 5dB]
Peak Level 102.6dB 15-JAN-93 @ 10:56:42AM
Max Level 88.0dB 15-JAN-93 @ 10:56:42AM
# "DOSE1" (Selected if Yes) Min Level 73.4dB 15-JAN-93 @ 10:56:42AM
# "DOSE2" (Selected if Yes) Lavg 80.5dB Dose 0.11% SEL(5) 115.1dB
TWA 41.0dB Dose[8] 26.40%
TWA[ 4.00] 75.5dB Dose[ 4.00] 13.20%
If both Dosimeters are selected, each Dosimeter will produce Time Histories based on the selection of LAVG, MAX, PEAK,
and/or LC-A.
Only data that has been logged can later be printed. See DETAILED PROGRAMMING INSTRUCTIONS if needed.
Note that the Time History Printout example (Figures 10, 11, and 12) is printed in both the "TABULAR" and "GRAPHICAL"
formats.
Calculate the "Time per Line" (time between lines of print on the graph) as follows:
# The dosimeter logs information as "Time per Sample".
# (01 SEC / SAMPLE) (10 SEC / SAMPLE) (01 MIN / SAMPLE)
# To set the "Time per Sample", see PROGRAMMING MENU STRUCTURE if needed.
# The Printout is programmed as "Samples per Line":
# "SAMPLES/LINE" (1 to 120)
# Multiply the "Time per Sample" by the "Samples per Line" to get the "Time per Line".
EXAMPLE:
(1 Minute / Sample) X (10 Samples / Line) = 10 Minutes / Line
30 31
TABULAR FORMAT
10 (Sec)
22-JAN-93
Time Max
03:17:01PM 99.3dB 101.2dB 96.8dB 99.7dB 105.1dB
03:17:51PM 105.2dB 103.3dB 102.9dB 104.1dB 76.2dB
03:18:41PM 103.4dB 105.8dB 108.0dB 82.9dB 108.2dB
03:19:31PM 107.4dB 82.2dB 102.5dB 106.2dB 105.4dB
Time Max
10 (Sec)
22-JAN-93
Time Peak
03:17:01PM 121.6dB 117.9dB 120.5dB 119.1dB 123.8dB
03:17:51PM 126.3dB 126.3dB 122.3dB 130.3dB 115.7dB
03:18:41PM 127.2dB 127.9dB 128.0dB 110.2dB 126.2dB
03:19:31PM 127.2dB 120.0dB 124.0dB 128.0dB 124.6dB
Time Peak
32 33
GRAPHICAL FORMAT TABULAR FORMAT
Lavg TIME HISTORY [Dosimeter 1] Page 1 C-A Lavg TIME HISTORY [Dosimeter 2 - Dosimeter 1] Page 1
Weighting A Time Constant Slow Weighting C Time Constant Slow Weighting A Time Constant Slow
Threshold 80dB Exchange Rate 5dB Threshold 80dB Exchange Rate 5dB Threshold 80dB Exchange Rate 5dB
10 (Sec) 10 (Sec)
22-JAN-93 40 60 80 100 120 140 22-JAN-93
Time Lavg +----+----+----+----+----+----+----+----+----+----+ Time C-A Lavg
03:17:01PM 93.9dB *************************** 03:17:01PM 3.5dB 0.7dB 5.2dB 2.5dB 1.4dB
03:17:11PM 91.4dB ************************** 03:17:51PM 3.3dB 3.6dB 4.0dB 5.9dB 79.8dB
03:17:21PM 86.2dB ************************ 03:18:41PM 2.7dB 2.4dB 3.3dB 10.2dB 1.3dB
03:17:31PM 87.8dB ************************ 03:19:31PM 1.2dB 18.9dB 3.5dB 1.8dB 0.9dB
03:17:41PM 99.1dB ****************************** Time C-A Lavg
03:17:51PM 97.2dB *****************************
03:18:01PM 95.5dB **************************** C-A Max TIME HISTORY [Dosimeter 2 - Dosimeter 1] Page 1
03:18:11PM 88.8dB ************************* Weighting C Time Constant Slow Weighting A Time Constant Slow
03:18:21PM 91.4dB ************************** Threshold 80dB Exchange Rate 5dB Threshold 80dB Exchange Rate 5dB
03:18:31PM 0.0dB
Time Lavg +----+----+----+----+----+----+----+----+----+----+ 10 (Sec)
40 60 80 100 120 140 22-JAN-93
Time C-A Max
Max TIME HISTORY [Dosimeter 1] Page 1 03:17:01PM 4.0dB 0.6dB 5.0dB 0.9dB 1.3dB
Weighting A Time Constant Slow 03:17:51PM 3.9dB 4.1dB 6.0dB 5.7dB 12.8dB
Threshold 80dB Exchange Rate 5dB 03:18:41PM 2.6dB 2.4dB 2.8dB 4.3dB 1.0dB
03:19:31PM 0.9dB 7.7dB 1.2dB 0.8dB 0.4dB
10 (Sec) Time C-A Max
22-JAN-93 40 60 80 100 120 140
Time Max +----+----+----+----+----+----+----+----+----+----+ GRAPHICAL FORMAT
03:17:01PM 99.3dB ==============================
03:17:11PM 101.2dB =============================== C-A Lavg TIME HISTORY [Dosimeter 2 - Dosimeter 1] Page 1
03:17:21PM 96.8dB ============================= Weighting C Time Constant Slow Weighting A Time Constant Slow
03:17:31PM 99.7dB ============================== Threshold 80dB Exchange Rate 5dB Threshold 80dB Exchange Rate 5dB
03:17:41PM 105.1dB =================================
03:17:51PM 105.2dB ================================= 10 (Sec)
03:18:01PM 103.3dB ================================ 22-JAN-93 0 5 10 15 20 25
03:18:11PM 102.9dB ================================ Time C-A Lavg +----+----+----+----+----+----+----+----+----+----+
03:18:21PM 104.1dB ================================= 03:17:01PM 3.5dB ********
03:18:31PM 76.2dB =================== 03:17:11PM 0.7dB **
Time Max +----+----+----+----+----+----+----+----+----+----+ 03:17:21PM 5.2dB ***********
40 60 80 100 120 140 03:17:31PM 2.5dB ******
03:17:41PM 1.4dB ***
03:17:51PM 3.3dB *******
Peak TIME HISTORY Page 1 03:18:01PM 3.6dB ********
Time C-A Lavg +----+----+----+----+----+----+----+----+----+----+
0 5 10 15 20 25
Figure 11. Time History Printout (Graphical Format). Figure 12. C-A Time History (Tabular and Graphical Format).
34 35
% Time Statistical Distribution Section Figure 13 Exceedance Levels Section Figure 14
The % Time Statistical Distribution Printout will occur only if A and B are selected as follows: The Exceedance Level Printout will occur only if A and B are selected as follows:
A.) Statistics is selected: A.) LN Levels is selected:
# "STATISTICS" (Selected if Yes) # "LN LEVELS" (Selected if Yes)
B.) One or both of the following are selected: B.) One or both of the following are selected:
# "DOSE1" (Selected if Yes) # "DOSE1" (Selected if Yes)
# "DOSE2" (Selected if Yes) # "DOSE2" (Selected if Yes)
Statistics for Dosimeter 1 and Dosimeter 2 Data are printed out as follows:
The distribution is always accumulated without a threshold.
An LN Table for either Dosimeter 1 or Dosimeter 2 (or both) will be printed out as follows:
% TIME STATISTICAL DISTRIBUTION [Dosimeter 1] Page 1
Weighting A Time Constant Slow
0 20 40 60 80 100
dB Samples % Time +----+----+----+----+----+----+----+----+----+----+
73 14 0.36% *
74 66 1.71% *
75 200 5.20% ***
76 239 6.21% ****
77 254 6.60% ****
78 251 6.53% ****
79 341 8.87% *****
80 255 6.63% ****
81 380 9.88% *****
82 366 9.52% *****
83 543 14.12% ********
84 438 11.39% ******
85 201 5.23% ***
86 148 3.85% **
87 140 3.64% **
88 7 0.18% *
Figure 13. % Time Statistical Distribution Printout.
36 37
SIMPLIFIED PROGRAMMING INSTRUCTIONS
EXCEEDANCE LEVELS [Dosimeter 1]
Weighting A Time Constant Slow Programming Keys are as follows:
L01 87dB L02 87dB L03 87dB L04 86dB L05 86dB
L06 86dB L07 86dB L08 85dB L09 85dB L10 85dB
L11 85dB L12 85dB L13 84dB L14 84dB L15 84dB
L16 84dB L17 84dB L18 84dB L19 84dB L20 84dB They allow the user to:
L21 84dB L22 84dB L23 84dB L24 84dB L25 83dB # Acoustically Calibrate the dosimeter.
L26 83dB L27 83dB L28 83dB L29 83dB L30 83dB
# Select Printouts and other printer parameters.
L31 83dB L32 83dB L33 83dB L34 83dB L35 83dB # Enter menus and change settings.
L36 83dB L37 83dB L38 83dB L39 82dB L40 82dB
L41 82dB L42 82dB L43 82dB L44 82dB L45 82dB
L46 82dB L47 82dB L48 81dB L49 81dB L50 81dB Action Keys are as follows:
L51 81dB L52 81dB L53 81dB L54 81dB L55 81dB
L56 81dB L57 81dB L58 80dB L59 80dB L60 80dB
L61 80dB L62 80dB L63 80dB L64 80dB L65 79dB
L66 79dB L67 79dB L68 79dB L69 79dB L70 79dB They allow the user to:
L71 79dB L72 79dB L73 79dB L74 78dB L75 78dB # Highlight and select a function within a menu.
L76 78dB L77 78dB L78 78dB L79 78dB L80 77dB
# Enable or disable a function.
L81 77dB L82 77dB L83 77dB L84 77dB L85 77dB # Modify a function's value or setting.
L86 77dB L87 76dB L88 76dB L89 76dB L90 76dB
NOTE: See "PROGRAMMING MENU STRUCTURE" to locate various functions within the layers of
L91 76dB L92 76dB L93 75dB L94 75dB L95 75dB
L96 75dB L97 75dB L98 74dB L99 74dB L100 73dB menus.
___________________________________________________________________
140 1.) Use one of the Programming Keys (CAL, PRINT, and MENU - ON/OFF) to enter the proper menu leading
130 to the desired item to change (or check).
120 2.) Select an item and change (or check) its value as follows:
110
A.) Use either the UP ARROW and DOWN ARROW Keys or the ENTER Key (if needed) to highlight
100
an item.
90
* * B.) Use either the ENTER Key, the UP ARROW and DOWN ARROW Keys, or the NUMBER Keys to
80 * * * * * * *
* * * * * * * * * * select or change the item or value.
70 * * * * * * * * * * *
* * * * * * * * * * * 3.) Repeat Steps 2A and 2B as needed to change (or check) all settings within each menu. Change values if
60 * * * * * * * * * * * necessary.
* * * * * * * * * * *
50 * * * * * * * * * * *
* * * * * * * * * * * Exit the Programming Mode as follows:
40 * * * * * * * * * * *
L01 L10 L20 L30 L40 L50 L60 L70 L80 L90 L100 1.) Use the ENTER Key (if needed) to return the highlighting to a function.
2.) Then press the MENU - ON/OFF Key until it returns to the main menu beginning with "SETUP".
38 39
PROGRAMMING MENU STRUCTURE PROGRAMMING MENU STRUCTURE (Cont.)
PRINT Key
The menu structure of the Programming Keys is as follows:
NOTE: See "APPENDIX, Acoustical Definitions" for more detail about any items preceded W444444444444444444444444444444444444444444444444444444444444444U
by "*".
CAL Key
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
"PRINT SELECTED" Print User-Selected Data
"PRE-SURVEY" Electronic Calibration Procedure "HEADER" [Y/N] Name, Work Area, Comments, etc.
"CALIBRATION" [Y/N] Pre-survey/Periodic Checks
"DOSE1" [Y/N] Dosimeter 1 Information
W444444444444444444444444444444444444444444444444444444444444444U
"DOSE2" [Y/N] Dosimeter 2 Information
"PARAMETERS" [Y/N] Dosimeter Setup Parameters
"PERIODIC CHECK" Saves up to 8 Calibration Level Checks and the Times that they were performed. "SUMMARY" [Y/N] Time and Data Summary
*"PEAK" [Y/N] Peak Levels
*"U.L." [Y/N] Upper Limit Time
W444444444444444444444444444444444444444444444444444444444444444U *"STATISTICS" [Y/N] % Time StatisticalDistrib.
*"LN LEVELS" [Y/N] Exceedance Levels, L01 - L100
*"EVENTS" [Y/N] All Events
"REVIEW" Reviews the saved Calibration Level Checks and the Times that they were *"TIME HISTORY" Time History Selection
performed.
*"LAVG" [Y/N] Average Time History
W444444444444444444444444444444444444444444444444444444444444444U *"MAX" [Y/N] Maximum Level Time History
*"PEAK" [Y/N] Peak Level Time History
*"LC-A" [Y/N] C minus A Time History
"TABULAR" Select Tabular or Graphical
"SAMPLES/LINE" Enter 1 to 120
W444444444444444444444444444444444444444444444444444444444444444U
40 41
W444444444444444444444444444444444444444444444444444444444444444U PROGRAMMING MENU STRUCTURE (Cont.)
MENU ON/OFF Key
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
42 43
PROGRAMMING MENU STRUCTURE (Cont.)
MENU ON/OFF Key
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
44 45
PROGRAMMING MENU STRUCTURE (Cont.)
MENU ON/OFF Key
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
W444444444444444444444444444444444444444444444444444444444444444U
46 47
SETTING UP MEASUREMENT PARAMETERS Setup # 1
The following 2 pages provide step-by-step worksheets for programming setup parameters into the dosimeter. If you need MENU - ON/OFF Key )))> "SETUP" )))> "DOSE1"
help in programming see the "DETAILED PROGRAMMING INSTRUCTIONS" section.
"Prt" _______ hrs.
The following are some typical setups that can be programmed into the dosimeter. Although they are correct as of the time
of printing, check your regulations before taking data. Regulations change and you must be sure that your settings are "TL" _______ dB
correct. "ER" _______ dB
"UL" _______ dB
"CL" _______ dB
OSHA Hearing Conservation
"LDN" _______ Yes or No
A Weighting 80dB Threshold 5dB Exchange Rate
90dB Criterion Level LDN- No Slow Response MENU - ON/OFF Key )))> "SETUP" )))> "DOSE2"
A Weighting 80dB Threshold 4 dB Exchange Rate MENU - ON/OFF Key )))> "SETUP" )))> "RESP/WT"
84dB Criterion Level LDN- No Slow Response "DOS1" _______ SLOW or FAST
"DOS2" _______ SLOW or FAST
IEC Noise Monitoring "WGHT" _______ A / A C / C A / C
48 49
"TRIG ENABLE" _______ Yes or No Setup # 2
"LEVEL ON" _______ Any Allowable Level
MENU - ON/OFF Key )))> "SETUP" )))> "DOSE1"
"LEVEL OFF" _______ Any Allowable Level
"Prt" _______ hrs.
MENU - ON/OFF Key )))> "AUTO-ON" )))> "ENABLE Y/N" "TL" _______ dB
"ER" _______ dB
"MODE DAILY" ___:___#___ Start#Duration "UL" _______ dB
"MODE WEEKDAY" SMTWTFS#___:___#___ Day#Start#Duration "CL" _______ dB
"MODE DATE" ___ /___#___:___#___ Mo./Day#Start#Duration "LDN" _______ Yes or No
"DOSIMETERS" _______ 1 or 2
"LAVG" _______ Yes or No
"MAX" _______ Yes or No
"PEAK" _______ Yes or No
"01 SEC" _______ 01-SEC 10-SEC 01-MIN
50 51
"TRIG ENABLE" _______ Yes or No DETAILED PROGRAMMING INSTRUCTIONS
"LEVEL ON" _______ Any Allowable Level
"LEVEL OFF" _______ Any Allowable Level After becoming familiar with the dosimeter, the "PROGRAMMING MENU STRUCTURE" section should provide enough
information about programming. However, if more detailed programming instructions are desired, use this section. The
following pages provide step by step instructions.
MENU - ON/OFF Key )))> "AUTO-ON" )))> "ENABLE Y/N"
CAL Key
"PRE-SURVEY"
4.) Use the UP ARROW and DOWN ARROW Keys to set the number to match the output level listed on the
acoustic calibrator.
5.) Turn on the calibrator (1000 Hz). Listen to see that the
calibrator is producing a tone. If no tone is present,
see the calibrator manual to correct the problem.
6.) Insert the Microphone and Adapter into the Calibrator.
8.) "CALIBRATING..." will appear for a moment with "..." slowly moving left to right. This means that the
system is attempting to calibrate.
# If "BAD CALIBRATION / TO TRY AGAIN / PRESS (CAL)" appears or the SPL does not match the
entered calibrator level by +/- 0.1 dB:
# Check the calibration system.
# Repeat steps 2 through 8 to obtain a valid calibration.
9.) Exit this mode by pressing the MENU ON/OFF Key.
52 53
# For the "PEAK" level to be accurate, the "SPL" calibration level must be in the range of 100 to 140 dB. CAL Key:
"REVIEW"
# A valid "PEAK" level (at 1000Hz) is 2.0 to 4.0 dB higher than the calibration level.
# If a new "PRE-SURVEY" Calibration is performed, the old calibration is discarded along with all old Review the Initial Calibration (followed by "**") and up to 8 "PERIODIC CHECK"s of the Acoustic Calibration as follows:
"PERIODIC CHECK" Levels and Times. This is the only way to delete all previously entered "PERIODIC
1.) Press the CAL key. "PRE-SURVEY", "PERIODIC CHECK", and "REVIEW" will appear in the menu.
CHECK"s.
"PRE-SURVEY" will be highlighted.
# The Sound Level Calibrator output will be affected slightly due to altitude (atmospheric pressure). If
calibrating at high altitudes, see the calibrator manual for altitude correction values. 2.) Use the UP ARROW and DOWN ARROW keys to highlight "REVIEW".
CAL Key: 3.) Press the ENTER key. The Main Calibration is shown followed by any "PERIODIC CHECK"s that were
"PERIODIC CHECK" entered.
4.) Use the UP ARROW and DOWN ARROW keys to observe up to 8 "PERIODIC CHECK" dB values and
Up to 8 Calibration Levels and Calibration Times can be saved throughout a measurement period. times within the display.
NOTE: If you try to save more than 8 levels, the ENTER key will not appear to work. 5.) Exit this mode by pressing the MENU ON/OFF key.
Periodically check the Acoustic Calibration of the dosimeter as follows: # To delete the old "PERIODIC CHECK" levels and times, it is necessary to perform a new "PRE-SURVEY"
1.) Turn on the calibrator (1000 Hz). Listen to see if the calibrator is producing a tone. If no tone is present, calibration. The old calibration is then discarded along with all old "PERIODIC CHECK" levels and times.
see the calibrator manual to correct the problem.
PRINT Key:
2.) Insert the Microphone and Adapter into the calibrator. "PRINT ALL"
3.) Press the CAL Key. "PRE-SURVEY", "PERIODIC CHECK", and "REVIEW" will appear in the menu.
Print all available data as follows:
4.) Use the UP ARROW and DOWN ARROW Keys to highlight "PERIODIC CHECK".
1.) Press the PRINT key. "PRINT ALL" will appear highlighted.
5.) Press the ENTER Key. "CAL. CHECK" followed by "LEVEL = XX.X dB" (where XX.X is the SPL value)
will appear. 2.) Press the PRINT key. "PRINTING..." will appear and the printing will begin if all connections and printer
settings are proper.
6.) Press the ENTER Key to save the dB Level and time of check.
7.) Exit this mode by pressing the MENU ON/OFF key. See the "SAMPLE PRINTOUTS" section to see what information is contained within each printout section.
If you wish to print only certain sections of the total printout, then look at the following section, "PRINT SELECTED".
# To delete the old "PERIODIC CHECK" levels and times, it is necessary to perform a new "PRE-SURVEY"
calibration. The old calibration is then discarded along with all old "PERIODIC CHECK" levels and times.
54 55
PRINT Key: PRINT Key:
"PRINT SELECTED" "PRINTER SETUP"
1.) Press the PRINT key. Use the UP ARROW and DOWN ARROW keys to highlight "PRINTER SETUP".
Print User-Selected Data as follows:
2.) Press ENTER. The following three items will appear.
1.) Press the PRINT key. "PRINT ALL" will appear highlighted.
"BAUD" (Highlighted)
2.) Use the UP ARROW and DOWN ARROW keys to highlight "PRINT SELECTED". "EOL"
"PAGE BREAKS" ("HANDSHAKE" will not appear)
3.) Press the ENTER key. "HEADER" will appear.
# Set the Baud Rate as follows:
# To select which information to print:
("Y" will print, "N" will not print) A.) Press the ENTER key. The present Baud Rate setting will become highlighted. Use the UP
ARROW and DOWN ARROW keys to change it to one of the following:
A.) Press the UP ARROW and DOWN ARROW keys to highlight an item.
300 600 1200 2400 4800 9600 19200
See the "SAMPLE PRINTOUTS" section to see how the following selections relate to each printout.
B.) Press ENTER to again highlight "BAUD".
"HEADER" [Y/N] Name, Work Area, Comments, etc.
"CALIBRATION" [Y/N] Pre-survey/Periodic Checks C.) If desired, continue on to the next setting. Otherwise, exit this mode by pressing MENU ON/OFF.
"DOSE1" [Y/N] Dosimeter 1 Information
"DOSE2" [Y/N] Dosimeter 2 Information # Set the End of Line Characters "LF" & "CR" as follows:
"PARAMETERS" [Y/N] Dosimeter Setup Parameters
"SUMMARY" [Y/N] Time and Data Summary A.) Use the UP or DOWN ARROW keys to highlight "<EOL>".
"PEAK" [Y/N] Peak Levels
B.) Press the ENTER key to highlight the present EOL setting. Use the UP ARROW and DOWN
"U.L." [Y/N] Upper Limit Time
ARROW keys to change the "<EOL>" setting to one of the following:
"STATISTICS" [Y/N] % Time Statisticl Distribution
"LN LEVELS" [Y/N] Exceedance Levels, L01 - L100 "LF/CR" Line Feed followed by Carriage Return
"EVENTS" [Y/N] All Events "CR/LF" Carriage Return followed by Line Feed
"TIME HISTORY" Time History "LF ONLY" Line Feed only
"LAVG" [Y/N] Average Time History "CR ONLY" Carriage Return only
"MAX" [Y/N] Maximum Level Time History
"PEAK" [Y/N] Peak Level Time History C.) Press ENTER to again highlight "<EOL>". Continue on, or exit this mode by pressing MENU
"LC-A" [Y/N] C minus A Time History ON/OFF.
"TABULAR" Select Tabular or Graphical
"SAMPLES/LINE" Enter 1 to 120 # Select Page Breaks as follows:
C.) Repeat Steps A and B to select as many of the above as desired. C.) Continue, or exit by pressing MENU ON/OFF.
4.) When all selections are completed, press the PRINT key twice. "PRINTING..." will appear and the printing
# Select Handshake method as follows:
will begin if all connections and printer settings are proper.
A.) Highlight "HANDSHAKE" with the UP or DOWN ARROWS.
56 57
B.) Press ENTER to select "NONE", "HARDWARE" or "XON/XOFF". This should match the setting MENU - ON/OFF Key:
of your serial printer or computer COM port. "SETUP"
"DOSE1"
C.) Exit this mode by pressing the MENU ON/OFF key. "DOSE2"
NOTE: These settings cannot be changed if the unit has any data stored (Run Time) in it.
Setup Dosimeter #1 and/or #2 as follows:
1.) Press the MENU ON/OFF key. "OFF-5" will appear and is highlighted.
2.) Use the UP ARROW and DOWN ARROW keys to highlight "SETUP".
3.) Press the ENTER key. "DOSE1" will appear highlighted. "DOSE2" will appear but is not highlighted.
4.) Use the UP ARROW and DOWN ARROW keys to highlight either "DOSE1" or "DOSE2".
5.) Press the ENTER key. The following items will appear:
C.) Use the UP ARROW and DOWN ARROW keys to change the value.
D.) Press the ENTER key to highlight the item that just had its value changed.
A.) Use the UP ARROW and DOWN ARROW keys to highlight "LDN".
6.) When all values are properly set, press the MENU ON/OFF key twice to return to the "SETUP" menu.
58 59
MENU - ON/OFF Key: MENU - ON/OFF Key:
"SETUP" "SETUP"
"RESP/WT" "LOGGING"
NOTE: These settings cannot be changed if the unit has any data stored (Run Time) in it. Select the items to be Logged as follows:
Set up the Response and Weighting as follows: "DOSIMETERS" Select 1 or 2 Dosimeter Mode
"LAVG" [Y/N] Select to Log the Average Level
"DOS1" Response, Select "SLOW" or "FAST"
"MAX" [Y/N] Select to Log the Maximum Level
"DOS2" Response, Select "SLOW" or "FAST"
"PEAK" [Y/N] Select to Log the Peak Level
"WGHT" Weighting, Select 1 of 3:
"01 SEC" 1 Sec., 10 Sec., or 1 Min. Logging
DOS1 = WGHT / DOS2 = WGHT
1.) Press the MENU ON/OFF key. "OFF-5" will appear and is highlighted.
"1 = A" / "2 = A"
2.) Use the UP ARROW and DOWN ARROW keys to highlight "SETUP".
"1 = C" / "2 = C"
"1 = A" / "2 = C" 3.) Press the ENTER key. "DOSE1" will appear and is highlighted.
1.) Press the MENU ON/OFF key. "OFF-5" will appear and is highlighted. 4.) Use the UP ARROW and DOWN ARROW keys to highlight "LOGGING".
2.) Use the UP ARROW and DOWN ARROW keys to highlight "SETUP". 5.) Press the ENTER key. "DOSIMETERS" will appear and is highlighted.
3.) Press the ENTER key. "DOSE1" will appear highlighted. # To select the number of "DOSIMETERS" to log:
4.) Use the UP ARROW and DOWN ARROW keys to highlight "RESP/WT". A.) Press the ENTER key to display either "1" or "2".
5.) Press the ENTER key. # To change Yes/No answers:
# To change "DOSE1" or "DOSE2" (Dosimeter 1 or 2): A.) Use the UP ARROW and DOWN ARROW keys to highlight either "LAVG", "MAX", or "PEAK".
A.) Use the UP ARROW and DOWN ARROW keys to highlight either "DOSE1" or "DOSE2". B.) Use the ENTER key to change Yes/No.
B.) Press the ENTER key to change "SLOW" / "FAST". # To change "01-SEC", "10-SEC", or "01-MIN":
# To change "WGHT" (Weighting): A.) Use the UP ARROW and DOWN ARROW keys to highlight the item.
A.) Use the UP ARROW and DOWN ARROW keys to highlight "WGHT". B.) Use the ENTER key to change it.
B.) Use the ENTER key to select one of the following: 6.) When all values are properly set, press the MENU ON/OFF key twice to return to the SETUP menu.
6.) When all values are properly set, press the MENU ON/OFF key twice to return to the "SETUP" menu.
60 61
MENU - ON/OFF Key: # To change the "TIME" when in the "12-HOUR" format:
"SETUP"
"CLK-SET" A.) Use the UP ARROW and DOWN ARROW keys to highlight "TIME" if it is not highlighted.
B.) Press the ENTER key. The first digit becomes highlighted. Seconds will return to 00 and will stop
NOTE: These settings cannot be changed if the unit has any data stored (Run Time) in it.
counting.
Set up the Clock as follows:
C.) Use the NUMBER keys to enter the correct Time.
"TIME" Enter with the NUMBER Keys
"DATE" Enter with the NUMBER Keys D.) Use the UP ARROW and DOWN ARROW keys to select "AM" or "PM".
"12-HOUR" Select 12 or 24 Hour Clock Format
E.) Use the ENTER key to Highlight "TIME". The clock will begin counting.
"MO.-DAY" Select MO.-DAY or DAY-MO. Format
F.) If desired, continue on to another setting. Otherwise, press the MENU ON/OFF key twice to return
1.) Press the MENU ON/OFF key. "OFF-5" is highlighted.
to the "SETUP" menu.
2.) Use the UP ARROW and DOWN ARROW keys to highlight "SETUP".
# To change the "TIME" when in the "24-HOUR" format:
3.) Press the ENTER key. "DOSE1" will appear highlighted.
A.) Use the UP ARROW and DOWN ARROW keys to highlight "TIME" if it is not highlighted.
4.) Use the UP ARROW and DOWN ARROW keys to highlight "CLK-SET".
B.) Press the ENTER key. The first digit becomes highlighted. Seconds will stop counting.
5.) Press the ENTER key. "TIME" will appear highlighted.
C.) Use the NUMBER keys to enter the correct Time.
6.) Check the settings. If necessary, correct the settings as follows:
# To select either the "12-HOUR" / "24 HOUR" or the "MO.-DAY" / "DAY-MO." format: D.) When the last digit is entered, "TIME" will appear highlighted and the clock will begin counting.
A.) Use the UP ARROW and DOWN ARROW keys to highlight the item. E.) If desired, continue on to another setting. Otherwise, press the MENU ON/OFF key twice to return
to the "SETUP" menu.
B.) Use the ENTER key to change it.
C.) If desired, continue on to another setting. Otherwise, press the MENU ON/OFF key twice to return
to the "SETUP" menu.
A.) Use the UP ARROW and DOWN ARROW keys to highlight "DATE" if it is not highlighted.
B.) Press the ENTER key. The first digit becomes highlighted.
C.) Use the NUMBER keys to enter the correct Date. After the last digit is entered, "DATE" will again
appear highlighted.
D.) If desired, continue on to another setting. Otherwise, press the MENU ON/OFF key twice to return
to the "SETUP" menu.
62 63
MENU - ON/OFF Key: # To enter the Level-Triggered Event mode:
"SETUP"
"EVENTS" A.) Press the EVENT key when ready to begin taking data.
# If "TRIG ENABLE" was set to "Y", the Level-Triggered Event mode will start when the
NOTE: These settings cannot be changed if the unit has any data stored (Run Time) in it. EVENT key is pressed. The RUN /PAUSE key will not operate in this mode. Sound
Set up Level-Triggered Events as follows: Levels will control Run and Pause.
"TRIG ENABLE" Level-Activated Event, Yes/No # If "TRIG ENABLE" was set to "N", the Level-Triggered Event mode will not be entered
"LEVEL ON" Modify, 40 to 140 dB when the EVENT key is pressed. The Manually-Triggered Event will be entered instead.
"LEVEL OFF" Modify, 40 to 140 dB
# To exit the Level-Triggered Event mode:
1.) Press the MENU - ON/OFF key until the "SETUP" Menu appears.
A.) Depress MENU - ON/OFF. The display will read "EVENT MODE EXITED".
2.) Use the UP ARROW and DOWN ARROW keys to highlight "SETUP".
3.) Press the ENTER key. A menu with "EVENTS" will appear. See "USING THE EVENT MODES" for detailed instructions on how to operate this mode.
4.) Use the UP ARROW and DOWN ARROW keys to highlight "EVENTS".
MENU - ON/OFF Key:
"OFF-5"
5.) Press the ENTER key. "TRIG ENABLE" will appear and is highlighted.
C.) Use the UP ARROW and DOWN ARROW keys to change its value.
D.) When the desired value is displayed, press the ENTER key. Either "LEVEL ON" or "LEVEL
OFF" will again appear highlighted. MENU - ON/OFF Key:
"RESET-5"
64 65
MENU - ON/OFF Key: # To enter the Security mode with or without Level Triggered Events activated:
"SECURE"
A.) Press the ENTER key. "ENTER SECURITY CODE: XXXX" will appear.
Operate the 4-Digit Security Code as follows: B.) Use the NUMBER keys to enter the 4-digit security code. When the last digit is entered, one of two
things will happen:
"SECURE ENABLE" Select to Enable Security
"CHANGE CODE" Select to Enter New Security Code # If the code is wrong, "INCORRECT CODE" will appear and the dosimeter cannot be secured. Try
again.
1.) Press the MENU - ON/OFF key until the "SETUP" Menu appears.
# If the code is correct, "PRESS RUN TO SECURE" will appear with the unit in the Pause mode.
2.) Use the UP ARROW and DOWN ARROW keys to highlight "SECURE". Secure and Run the dosimeter as follows:
3.) Depress ENTER. "SECURITY" will appear with "SECURE ENABLE" highlighted. # Depress "RUN / PAUSE". "SECURED CODE: XXXX" will appear. The dosimeter is now
Secured in the Run mode.
# To change the Security Code:
# If another key is pressed, "SECURITY DISABLED" will appear. If this happens, start over
A.) Use the UP ARROW and DOWN ARROW keys to highlight "CHANGE CODE". at Step 1.
B.) Depress ENTER. "ENTER SECURITY CODE: XXXX" will appear. # To exit the Security mode with or without Level Triggered Events activated:
A.) "SECURED CODE: XXXX" is present in the display.
C.) Use the NUMBER keys to enter the 4-Digit Security code.
B.) Use the NUMBER keys to enter the 4-digit security code. When the last digit is entered, one of two
# If the code is wrong, "INCORRECT CODE" will appear. Try again.
things will happen:
# If the code is correct, "ENTER NEW CODE: XXXX" will appear. # If the code is wrong, "INCORRECT CODE" will appear and the dosimeter cannot be removed from
the secured mode. Try again.
D.) When the code is valid and "ENTER NEW CODE: XXXX" is present, enter the new 4-Digit security
code. # When the code is correct, "SECURED / CODE: XXXX ":
E.) After enterin the code, the display asks you to "REENTER NEW CODE: XXXX" to match the first # Will disappear if Level Triggered event was not activated. Return the dosimeter display to
entry. its normal function as follows:
# If the code is not identical, "CODE MISMATCH" will appear. Try again. 1.) Press the RUN/PAUSE key.
# When the code is identical, "SECURITY CODE HAS BEEN SAVED" will appear and the 2.) Press the MENU - ON/OFF key to return to the "SETUP" Menu.
code is saved.
# Will enter the Event mode if Level Triggered Event was activated. Return the dosimeter
F.) Depress MENU - ON/OFF. The "SETUP" Menu will appear. display to its normal function as follows:
1.) Press the MENU - ON/OFF key. "EVENT MODE EXITED" will appear.
2.) Press the MENU - ON/OFF key again to return to the "SETUP" Menu.
# To enter the Security mode with Auto-On activated:
A.) Press the ENTER key. "ENTER SECURITY CODE: XXXX" will appear.
B.) Use the NUMBER keys to enter the 4-digit security code. When the last digit is entered, one of two
things will happen:
66 67
# If the code is wrong, "INCORRECT CODE" will appear and the dosimeter cannot be secured. Try MENU - ON/OFF Key:
again. "BATTERY"
# If the code is correct, "PRESS RUN TO SECURE" will appear with the unit in the Pause mode.
Secure the dosimeter in the Auto-On mode as follows: To Check the dosimeter Battery condition:
# Press the "RUN / PAUSE" key. "SECURED CODE: XXXX" will briefly appear. Then the 1.) During normal operation, press the MENU ON/OFF Key. "BATTERY" will appear in the menu.
display will go blank. The dosimeter is now Secured and will start when Auto-On tells it to.
2.) Use the UP ARROW and DOWN ARROW Keys to highlight "BATTERY".
# If any other key is pressed, "SECURITY DISABLED" will appear. If this happens, start
over at Step 1. 3.) Press the ENTER Key.
# To exit the Security mode with Auto-On activated: 4.) The display will list:
A.) Press the "MENU ON / OFF" key. "((QUEST))" will appear for a few seconds followed by "BATT = X.X VOLTS"
"GOOD FOR XX MORE HOURS OF USE."
"SECURED CODE: XXXX". You have about 8 seconds to enter the proper code to remove
Security. If you are not successful, "INCORRECT CODE" will appear and the display will go blank.
Repeat this step to try again.
# When the code is correctly entered, the display will immediately go blank. Press the "MENU ON / MENU - ON/OFF Key:
OFF" key to return to the "SETUP" Menu. "AUTO-ON"
"ENABLE Y/N"
The dosimeter has two Security Codes that can be used: See "USING AUTO-ON / TIMED RUN" for information on how to use these features.
# The Quest Security Code (only available by contacting Quest by either phone or FAX.) # Note that the Timed Run mode does not appear on the display. The duration of the Timed Run is equal to the
programmed Duration for either "DAILY", "WEEKDAY", or "DATE". Therefore, three Timed Run Durations can be
# The User Security Code. independently stored and used simply by selecting either "DAILY", "WEEKDAY", or "DATE".
See USING SECURITY CODES for more information on these codes. A location is included there so that you can record Program the Auto-On function as follows:
your code if you wish.
"MODE DAILY" Program Daily Start/Duration
"MODE WEEKDAY" Program Days of Week for Start/Duration
"MODE DATE" Program One Day for Start/Duration
2.) Use the UP ARROW and DOWN ARROW keys to highlight "AUTO-ON".
3.) Depress ENTER. "ENABLE" will appear highlighted, followed by the present mode ("DAILY",
"WEEKDAY", or "DATE") that the dosimeter is in.
68 69
# When Auto-On / Timed Run is enabled, be sure that the proper selection ("DAILY", "WEEKDAY", I.) Press the MENU ON/OFF key twice to return the dosimeter to the "SETUP" menu.
or "DATE") appears under "ENABLE" in the display before leaving the menu. Auto-On will then
automatically operate in this mode. Timed Run will operate for the Duration that is associated with # Turn the dosimeter Off. If Auto-On is enabled, the unit will automatically Start and Run per the
"DAILY", "WEEKDAY", or "DATE". "DAILY" programmed setup.
# Each of the following modes ("DAILY", "WEEKDAY", or "DATE") can be set and stored # To program "WEEKDAY":
independently. Therefore, each mode can have a different Start Time and duration programmed
into memory. A.) Use the UP or DOWN ARROW keys to highlight "MODE".
# To program "DAILY": B.) Depress ENTER to move the highlight to the right. "DAILY", "WEEKDAY", or "DATE" will be
highlighted.
A.) Use the UP ARROW and DOWN ARROW keys to highlight "MODE".
C.) Press the UP ARROW and DOWN ARROW keys until "WEEKDAY" appears.
B.) Depress ENTER to move the highlighting to the right. Either "DAILY", "WEEKDAY", or "DATE"
will be highlighted. D.) Press the ENTER key to select it. "MODE WEEKDAY" will appear. "DAYS" will be highlighted
followed by seven character positions:
C.) Press the UP ARROW or DOWN ARROW keys until "DAILY" appears.
# SMTWTFS = Sun Mon Tues Wed Thurs Fri Sat
D.) Depress ENTER. "MODE DAILY" will appear. "TIME" will be highlighted.
# Each Letter ("SMTWTFS") represents a day that the dosimeter will place itself into the
"XX:YY:ZZ AA" represents the Start Time. Run mode.
XX = Hours
YY = Minutes # If a Dash ("-") appears, this represents a day that the dosimeter will remain in the Pause
ZZ = Seconds mode.
AA = AM or PM (only appears in 12 Hour Mode)
# Enter each Run Day as follows:
E.) Depress ENTER. The first digit (X) will appear highlighted.
1.) Press the ENTER key. A character position will appear highlighted.
F.) Set the Start Time in either the 12 or 24 hour mode as follows:
2.) Use the UP or DOWN ARROW keys to select either a "-" or a day that the dosimeter will
# 12-Hour mode: Run.
1). Use the NUMBER keys to enter Hours, Minutes, and Seconds. Use the UP ARROW and 3.) Repeat steps 1 and 2 until all Run Days are entered.
DOWN ARROW keys to change between "AM" and "PM".
4.) Depress ENTER. "DAYS" will be highlighted.
2.) Depress ENTER. "DURATION" will appear.
E.) Use the UP ARROW and DOWN ARROW key to highlight "TIME". Enter the Start Time as
# 24-Hour mode: follows:
1). Use the NUMBER keys to enter Hours, Minutes, and Seconds. "XX:YY:ZZ AA" represents the Start Time.
XX = Hours
2.) When the last digit of the minutes is entered, "DURATION" will appear. YY = Minutes
ZZ = Seconds
G.) Use the NUMBER keys to enter "DURATION" as follows: AA = AM or PM (only appears in 12 Hour Mode)
F.) Press the ENTER Key. The first digit (X) will appear highlighted.
2 digits for "HOURS:" (0 to 99)
2 digits for "MIN:" (0 to 59) G.) Set the Start Time in either the 12 or 24 hour mode as follows.
H.) When the last digit is entered, "MODE DAILY" will appear. # 12-Hour mode:
70 71
1). Use the NUMBER Keys to enter Hours, Minutes, and Seconds. Use the UP ARROW and
DOWN ARROW Keys to change between "AM" and "PM". E.) Press the ENTER Key. The first digit (X) will appear highlighted.
2.) Press the ENTER Key. "DURATION" will appear. F.) Use the NUMBER Keys to enter all 4 digits. When the last digit is entered, "DATE" will again
become highlighted.
# 24-Hour mode:
G.) Use the UP ARROW and DOWN ARROW Key to highlight "TIME".
1). Use the NUMBER Keys to enter Hours, Minutes, and Seconds.
"XX:YY:ZZ AA" represents the Start Time.
2.) When the last digit of the minutes is entered, "DURATION" will appear. XX = Hours
YY = Minutes
H.) Use the NUMBER Keys to enter "DURATION" as follows: ZZ = Seconds
AA = AM or PM (only appears in 12 Hour Mode)
2 digits for "HOURS:" (0 to 99) H.) Press the ENTER Key. The first digit (X) will appear highlighted.
2 digits for "MIN:" (0 to 59) I.) Set the Start Time in the 12 or 24 hour mode as follows.
I.) When the last digit is entered, "MODE WEEKDAY" will appear. # 12-Hour mode:
J.) Press the MENU ON/OFF Key twice to return the dosimeter to the "SETUP" menu. 1). Use the NUMBER Keys to enter Hours, Minutes, and Seconds. Use the UP ARROW and
DOWN ARROW Keys to change between "AM" and "PM".
# Turn the dosimeter Off. If Auto-On is enabled, the unit will automatically Start and Run per the
"WEEKDAY" programmed setup. 2.) Press the ENTER Key. "DURATION" will appear.
# 24-Hour mode:
# To program "DATE":
1). Use the NUMBER Keys to enter Hours, Minutes, and Seconds.
A.) Use the UP ARROW and DOWN ARROW Keys to highlight "MODE".
2.) When the last digit of the minutes is entered, "DURATION" will appear.
B.) Press the ENTER Key to move the highlighting to the right. Either "DAILY", "WEEKDAY", or
"DATE" will be highlighted. J.) Use the NUMBER Keys to enter the "DURATION" as follows:
C.) Press the UP ARROW and DOWN ARROW Keys until "DATE" appears. # 2 digits for "HOURS:" (0 to 99)
D.) Press the ENTER Key to select it. "MODE DATE" will appear. "DATE" will be highlighted. # 2 digits for "MIN:" (0 to 59)
"XX-YY M-D"(Month-Day) or "YY-XX D-M"(Day- Month) K.) When the last digit is entered, "MODE DATE" will appear.
XX = Month
YY = Day L.) Press the MENU ON/OFF Key twice to return the dosimeter to the "SETUP" menu.
# Turn the dosimeter Off. If Auto-On is enabled, the unit will automatically Start and Run per the
"DATE" programmed setup.
72 73
SPECIFICATIONS Battery Test: Appears in the display. Use the keypad to request the Battery Voltage and Remaining Time on the
battery.
Standards: Q-400: type 2 Q-500: type 1
ANSI S1.25 - 1991, ANSI S1.4 - 1983 Memory and Clock/Calendar Battery Backup: Lithium cell, 1 to
IEC 651 - 1979, IEC 804 - 1985 2 year life not including time that the 9 volt
battery is installed. For long life, keep a
Measuring Range: 40 - 140dB with both dosimeter 1 and 2 set to either A or C weighting. Maximum level is for good 9 volt battery installed.
sinusoidal signals. A signal with a 10dB crest factor will be measured accurately if its RMS level is (Replacement only by Authorized Service Center)
10dB below the maximum level.
Note: If one dosimeter is set to A weighting and the other to C weighting, the measuring Temperature: -10o to +50oC operating; -20o to +60oC storage (battery removed)
range is 70 - 140dB.
Humidity: 0 to 95% non-condensing
Detector: True RMS, 63dB Pulse Range
Magnetic Field: Negligible below 50 Oersteds at 50 to 60 Hz.
Display: Multi-line LCD with Annunciators for SLOW, FAST, C, A, LOBAT, RUN and PAUSE Effects
Tested for RF succeptibility with <1dB error at field strengths to 10 V/m over the frequency
Data Output: Uses Interface Modules to condition the data. range of 10 MHz to 500 MHz.
Selectable Baud:
# 300 600 1200 2400 4800 9600 19200. Size: 5.5 x 2.8 x 1.4 inches (140 x 70 x 40 mm)
Connector: Uses the microphone connector.
Weight: 15.5 oz. (440 grams)
Microphone: Q-400 8mm Shoulder-Mount, Type 2.
One piece system: Construction: Cast aluminum housing with tamper-, water-, and dust-resistant
# Cable, Connector, and Microphone security cover.
Integration Time: Signal dependent: With a 3dB Exchange (doubling) rate and a constant input level of 140dB the *# Sound Exposure Level [SEL]
integration time would be 62 hours, 24 minutes. As the signal level decreases and the exchange *# Average Sound Level [LAV]. This is also called Equivalent Level [LEQ] when using 3dB Exchange Rate.
rate increases, the integration time increases. *# Time Weighted Average [TWA]
*# Projected Time Weighted Average [TWA (Prt)]
*# Run Time [RunTime], Pause Time [PSTime], Upper Limit Time [ULTime], Peak Time [PKTime], Maximum
Time [MaxTime], Minimum Time [MinTime], and Remaining Logging Time [LoggTime].
*# Up to 999 different events.
74 75
# 1-second, 10-second or 1-minute Time History. LOGGING TIME LIMITATIONS
Additional dosimeter "SETUP" features that are user-programmable from the keyboard: The following control the maximum Logging Time:
Security Code Feature [SECURE]. Locks out Keyboard. # The Logging Sample Rate (Time per Sample)
Auto-On; [DAILY], [WEEKDAY], or [DATE]. # The Total Number of Items (above) to log.
Timed Run; Allows the operator to manually place the unit into the Run mode for one of three user- The following chart shows logging time versus logging rate and number of items logged.
programmed durations. (Duration time is programmed into the Auto-On feature -- DAILY, WEEKDAY, or
DATE.)
Total APPROXIMATE LOGGING TIME
Selection of Dosimeter 1 or Dosimeter 2 data. Items
Logged 1 Sec. Log 10 Sec. Log 1 Min. Log
Projection Time [Prt], 1 HR to 16 HR (1/4 HR increments). Sample Rate Sample Rate Sample Rate
Exchange Rates [ER] of 3, 4, 5, or 6 dB. 1 17.6 Hours 176 Hours 1056 Hours
2 8.8 Hours 88 Hours 528 Hours
Threshold Level setting [TL] from 40 to 140 dB.
3 5.8 Hours 58 Hours 352 Hours
Criterion Level setting [CL] from 40 to 140 dB.
4 4.4 Hours 44 Hours 264 Hours
Upper Limit setting [UL] from 40 to 140 dB. 5 3.5 Hours 35 Hours 210 Hours
SLOW or FAST Response, A or C Weighting. ENERGY OVERLOAD
Selection of either 12- or 24-hour Real Time clock.
The amount of acoustic energy that can be measured between instrument resets is limited. If this energy is exceeded, math
PRINT Key Menu allows the user to print specific sections of the Printout as desired. overflow errors will result. The rate at which energy is accumulated depends upon the SPL and the exchange rate and the
value of SEL is a good indication of this. The limits per exchange rate are as follows:
Maximum SEL Exchange rate
193.5 3dB
244 4dB
295 5dB
347 6dB
For example the 3dB limit would be reached after 62 hours, 24 minutes with a constant level of 140dB. The time would
double for every 3dB decrease in level, so for practical situations encountered in the field the limitations should not be a
problem.
76 77
PRINCIPLES OF OPERATION Microphone Characteristics
The Q-400 Noise Dosimeter uses an 8 mm omnidirectional ceramic microphone. It is buffered by a high impedance FET
General Characteristics input stage. (See Figure 16.)
The Q-400 and Q-500 dosimeters use low power state-of-the-art circuitry. Each dosimeter is very stable and reliable over a
wide range of environmental conditions.
The low power circuitry gives each dosimeter a long battery life. When the 9 volt battery is changed, all dosimeter
information is retained due to an internal lithium battery. The lithium battery lasts for many years before needing
replacement. (See SPECIFICATIONS)
A rubberized Key Pad is used for all data entry and settings. Menus appear on the display and the keys are used to enter
changes.
A block diagram of the Q-400 and Q-500, is shown in Figure 15.
The Q-500 Noise Dosimeter uses a 1/2-inch free field prepolarized condenser (electret) microphone. It is buffered by a high
impedance FET preamp. (See Figure 17.)
78 79
Weighting Characteristics TYPICAL OPERATING PROCEDURE
The Q-400 and Q-500 both have "A" and "C" weighting characteristics as shown in Figure 18. For most industrial and
community noise measurement requirements, the "A" weighting should be used. The "A" weighting has a response similar As a Personal Noise Dosimeter
to the human ear. The "C" weighting is used for measuring noise reduction in hearing protectors and other scientific
purposes. 1.) Check all dosimeter setups that are relevent to the measurement. Change where needed. See DETAILED
OPERATING INSTRUCTIONS if necessary.
Both the Q-400 and Q-500 can be programmed to perform the "C - A" function. (See USING C - A)
2.) Reset the dosimeter.
4.) Attach the microphone to the operator's shirt as shown in Figure 19. Keep it high on the shoulder and away
from the neck as far as practical.
80 81
5.) Press the RUN/PAUSE Key to place the dosimeter into the RUN mode. As an Environmental Monitor
6.) Install the Security Cover and attach the unit to the operator's belt or pocket. With the unit in the Run mode, it will integrate all noise.
If installing or removing the Security Cover, do it very gently!
7.) At the end of the measurement period, remove the unit from the operator's belt or pocket.
1.) Check all dosimeter setups that are relevent to the measurement. Change where needed. See DETAILED
8.) Remove the security cover and press the RUN/PAUSE Key. OPERATING INSTRUCTIONS if necessary.
2.) Perform a Pre-Survey Calibration of the dosimeter.
9.) Remove the dosimeter and microphone from the operator.
CAL Key
10.) Evaluate the desired data by the following method(s): "PRE-SURVEY"
# Use the Display Function Keys to display the data.
3.) Connect the Microphone Boom to the dosimeter.
# Use a printer (or computer) to list a printout.
4.) Wrap the cable neatly around the Boom and connect the microphone to the top of the Boom.
# Transfer the data to a computer for further analysis.
5.) Attach the unit to a tripod if desired.
11.) Turn the dosimeter off.
6.) For the best accuracy, point the microphone as follows:
MENU ON/OFF Key
"OFF-5" # Q-400 Upward forming approximately a 70 degree angle with the noise source.
# Q-500 Directly at the noise source.
7.) Reset the dosimeter.
8.) Press the RUN/PAUSE Key to place the dosimeter into the RUN mode.
11.) Perform a Periodic Check of the dosimeter at the end of the measurement period if necessary.
CAL Key
"PERIODIC CHECK"
82 83
As a Survey Event Monitor ACCURACY
2.) Reset the dosimeter. A few items related to accuracy are as follows:
3.) Perform a Pre-Survey Calibration of the dosimeter. The effect of the "Noise Floor" on low level readings can cause inaccurate data.
CAL Key In a "Perfectly Quiet" room, the "Noise Floor" produced by the microphone is approximately:
"PRE-SURVEY"
# 35 dB on "A" Weighting.
4.) Connect the Microphone Boom to the dosimeter. # 45 dB on "C" Weighting.
5.) Wrap the cable neatly around the Boom and connect the microphone to the top of the Boom. Measurements must always be at least 5 dB above the "Noise Floor". to be valid. Therefore, the lowest possible
measurements of the dosimeter are approximately:
6.) For the best accuracy, point the microphone as follows:
# 40 dB on "A" Weighting (Lowest Valid Reading).
# Q-400 Upward forming approximately a 70 degree angle with the noise source.
# 50 dB on "C" Weighting (Lowest Valid Reading).
# Q-500 Directly at the noise source.
7.) Press the EVENT Key to monitor the event number. Accuracy of Readings
(Cev = Current Event Number)
There is a tendency to overestimate the accuracy of digital readings.
8.) Use the RUN/PAUSE Key to Run and then Pause the dosimeter at each event location as needed.
The values (such as Leq) are computed to a precision of 0.1 dB. However, the absolute accuracy of the reading is not 0.1
1st Location: "Cev" will be "1". dB, but is accurate to the stated overall accuracy of the instrument.
2nd Location: "Cev" will be "2". The 0.1 dB resolution is useful in determining the minimum sample time required to get an accurate short term
etc. measurement.
Be sure that a representative sample of noise in each location has been accumulated. # If, for example, the Leq is increasing 0.3 dB every second, then a longer sample time is required.
See USING THE EVENT MODES for details on how to use Events. # If the Leq is remaining stable from second to second within a few tenths of a dB, then the sample time is
long enough.
84 85
Microphone Positioning Microphone Windscreen
The dosimeter measures sound most accurately without the presence of sound reflecting or absorbing objects. It is recommended that a windscreen be used at all times. The Q-400 uses the WS-5 Windscreen and the Q-500 uses the
WS-7 Windscreen. Using a windscreen will improve the accuracy by minimizing the effect of the following:
Any near-by object or surface (including the operator) will act as a reflector or absorber of sound.
# Wind blowing across the microphone can produce higher than normal sound level readings.
To minimize these errors, use the dosimeter as follows:
# The windscreen blocks wind from direct contact with the microphone, producing less unwanted
# When taking Hand-Held measurements: noise.
# Connect the Microphone Boom. (See USING THE MICROPHONE BOOM.) # Clothing brushing against the microphone can produce higher than normal sound level readings.
# Hold the dosimeter with your arm extended and point the microphone as follows: # The windscreen helps position the microphone so that clothing cannot brush against it as easily.
# Q-400 Upward forming approximately a 70 degree angle with the noise source. # Dirt entering the microphone, in time, can damage the microphone, possibly changing the microphone's
frequency characteristics.
# Q-500 Directly at the noise source.
# The windscreen catches and collects this dirt before it enters the microphone.
# When taking Tripod-Held measurements:
Simply insert the microphone into the windscreen and gently pull it over the Velcro strip.
# Connect the Microphone Boom. (See USING THE MICROPHONE BOOM.)
TROUBLESHOOTING
# Mount the dosimeter to the tripod mount (located on the belt clip) with the microphone pointed as follows:
# Q-400 Upward forming approximately a 70 degree angle with the noise source. Blank display when the ON/OFF Key is pressed:
# Q-500 Directly at the noise source. # Replace the battery with a known fresh battery.
# When taking Personal Noise Dosimeter Measurements: Unit does not calibrate:
# Connect the Microphone to the shirt collar or shoulder. (See Figure 19.) Keep it high on the shoulder and # Perform a Battery Check. Battery must check OK.
away from the neck if possible.
# Check the Calibrator; Listen for a tone from its output.
# If the noise seems to be coming from one location, place the microphone near the ear facing that location.
# Reset the dosimeter and try again.
# Install a Windscreen on the microphone. It helps to do the following:
# Try a different microphone if you have another dosimeter. If this works, the microphone needs replacing.
# holds the microphone in an upright position.
# Return for service.
# keeps the microphone from brushing against clothing. This can produce higher than normal noise
levels into the microphone.
Unit is erratic:
# Try a different microphone if you have another unit. (The microphone or preamp could be intermittant.)
# Return for service.
86 87
Displayed Warning Messages Displayed Warning Messages, continued
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444U
If the MENU ON/OFF Key is pressed while in the RUN mode, the display will read: If the dosimeter reaches the end of its logging memory, the following message will appear on the display:
"KEY UNAVAILABLE, UNIT MUST BE IN THE (PAUSE) MODE" "RUN STOPPED / OUT OF MEMORY"
If this appears, press the PAUSE Key to enter the Pause mode. The dosimeter is out of memory and it is automatically placed into the Pause mode.
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444U
When the PRINT Key is pressed and the dosimeter is finished sending data to the printer, the display will read: On power up, the display normally reads "Q400 REV. x.xx"
"INITIALIZING"
"PRINTOUT COMPLETED OR TERMINATED"
If "INITIALIZING" is replaced by one of the following statements, setup data has been lost, probably due to a weak internal
If this appears and the unit did not print properly, recheck the setup and try to print again. memory backup battery. The battery should be replaced by an authorized Quest service center.
"DEFAULT 1" indicates the peak calibration values have been lost,
W444444444444444444444444444444444444444444444444444444444444444U resulting in inaccurate peak readings.
If the dosimeter has accumulated some Run Time, the unit must be reset before setup parameters can be changed. "DEFAULT 2" indicates that the serial number has been lost. The serial number may be re-entered in memory by sending
Otherwise, the total stored data could be accumulated with different setups, thus causing bad data.
the command #SAQxxxxxxxx (where xxxxxxxx is the remainder of the unit's serial number) via the serial data interface
The following message will appear if you attempt to change the setup during data accumulation: module. A communications package such as Procomm or Windows Terminal may be used.
"THIS SELECTION MAY NOT BE MADE UNTIL UNIT IS RESET" "DEFAULT 4" indicates that the dosimeter setup information has been lost. Setup has reverted back to factory default
settings. Check setup parameters before performing a study.
If this appears, do a RESET of the dosimeter before changing setup parameters.
"DEFAULT 3" indicates that both 1 and 2 have occurred.
W444444444444444444444444444444444444444444444444444444444444444U
88 89
ACCESSORIES QuestSuite for Windows Software
Provides data downloading and dosimeter setup capability using a personal computer. Graphing
Microphone / Preamp and data editing functions are provided for doing noise exposure predictions. Auto searches for
levels or database parameters. Compatible with Windows 3.1, Windows 95 or Windows NT.
56-963 Q-400, 8mm Shoulder-Mount Dosimeter Microphone, Type 2. One piece system including Cable,
Connector, and Microphone Windscreens
56-826 Q-500, 1/2-inch Shoulder-Mount Dosimeter Microphone Preamp, Type 1. Consists of a Cable, Connector, 58-452 WS-5 Windscreen, (8mm I.D.), package of 10.
and Preamp. Requires the QE-4146 Precision Microphone. (See next item.) # Use with the Q-400 Microphone.
59-491 QE-4146 Precision Microphone, 1/2-inch free-field prepolarized condenser (electret). Q-500 only. 59-344 WS-7 Windscreen, (1/2 inch I.D.), package of 3.
# Use with the Q-500 Microphone.
58-852 Earloops to hold microphone at the ear, package of 10. (For use with the Q-400 8mm Shoulder-Mount
Dosimeter Microphone only.) Calibrator Adapters
56-830 Clothing Clips, package of 5, for microphone cable. (For use with the Q-400 or Q-500.) 56-989 Calibrator Adapter, 8 mm to 1 inch coupler.
# Use with the Q-400 Microphone.
Data Interface Modules # Fits QC-10 and QC-20.
56-957 Parallel Printer Interface 58-839 Calibrator Adapter, 8 mm to 1 1/8 inch coupler.
# Centronics Compatable # Use with the Q-400 Microphone.
# Fits CA-12, CA-15, CA-22 and CA-32.
56-956 Serial Communications Interface
# 25-pin, RS-232 Female 56-990 Calibrator Adapter, 1/2 inch to 1 inch coupler.
# Powered by either: # Use with the Q-500 Microphone.
# One 9 Volt Battery, included. # Fits QC-10 and QC-20.
# MODEL 920 AC Power Supply, not included.
58-928 Calibrator Adapter, 1/2 inch to 1 1/8 inch coupler. # Use with the Q-500 Microphone.
Printer # Fits CA-12, CA-15, CA-22 and CA-32.
56-022 80 Column Parallel Printer, 110 volt only. Tripods
90 91
APPENDIX W444444444444444444444444444444444444444444444444444444444444444U
EXP: Units: Pascal Squared Hours (Pa2H)
Definitions Exposure
It is a method of measuring dosage. Pa2H is a linear unit rather than a percentage.
All definitions are in reference to the way that they are used with either the Q-400 or Q-500 dosimeter.
1 Pascal is equal to 94 dB. 94 dB for 1 Hour equals 1 Pa2H.
W444444444444444444444444444444444444444444444444444444444444444U
BAUD: Units: Bits per Second Examples of the 94 dB/time relationship:
Baud Rate
The rate of data transfer between the dosimeter and a printer or computer in the serial output mode. # 94 dB for 1 Hour = 1.00 Pa2H
# 94 dB for 8 Hours = 8.00 Pa2H
W444444444444444444444444444444444444444444444444444444444444444U
CL: Units: decibels (dB) The accumulation of Pa2H will double (or halve) for every 3 dB of change from 94 dB as follows:
Criterion Level
It is the constant sound level that, if applied for 8 hours, would accumulate a DOSE of 100%. (Used in Dose # 97 dB for 1 Hour = 2.00 Pa2H
measurements.) # 94 dB for 1 Hour = 1.00 Pa2H
# 91 dB for 1 Hour = .50 Pa2H
W444444444444444444444444444444444444444444444444444444444444444U # 88 dB for 1 Hours = .25 Pa2H
DOSE: Units: Percent (%) # 85 dB for 1 Hours = .125 Pa2H
Dose
1 Pa2H is typically the maximum allowable Exposure.
A percentage of the maximum allowable daily noise dose. This as a computation that is based on the following
variables: Criterion Level (CL), Lower Threshold (LT), and Exchange Rate (ER). # 85 dB for 8 Hours = 1 Pa2H
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444UFAST:
EOL: Units: LF and CR
End of Line Character Units: Time (milliseconds)
Fast Response
These are printer instructions that can be placed at the end of each line of type in the printout. The dosimeter can A Time Constant of 125 milliseconds. When Fast is used, a fluctuating noise into the dosimeter will cause SPL to
be programmed to send Line Feed (LF) and Carriage Return (CR) instructions. closely track the fluctuation.
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444U
ER: Units: decibels (dB) LAVG: Units: decibels (dB)
Exchange Rate Average Level
It is the number of dB that a sound must change to either halve or double the rate of dose accumulation. It is the average sound level for the measurement period based on either a 4, 5, or 6 dB Exchange Rate (ER). If
(3, 4, 5, or 6 dB exchange rates are common.) the Exchange Rate (ER) is 3 dB, then LAVG becomes LEQ.
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444U
EVENTS: Units: Each occurance LC - A: Units: decibels (dB)
Events C minus A
Each time that the dosimeter is in the Run mode and then Pauses creates an event. The dosimeter calculates C - A (Dosimeter 2 - Dosimeter 1) for the following parameters:
W444444444444444444444444444444444444444444444444444444444444444U # Lavg (dB) (or Leq when Exchange Rate is 3 dB)
# TWA (dB) (Time Weighted Average, 8 Hours)
# Ta [Vt] (dB) (Projected TWA, Variable Time)
# Dose (%) (or Leq when Exchange Rate is 3 dB)
# Dose [8] (%) (Projected Dose, always 8 Hour)
# Dose [Vt] (%) (Projected Dose, Variable Time)
# SEL [ER] (dB) (Sound Exposure Level, 1 Second)
# Pa2H (Exposure in Pascal Squared Hours)
# Can be calculated with 3, 4, 5, or 6 dB Exchange Rates.
W444444444444444444444444444444444444444444444444444444444444444U
92 93
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444U
LEQ: Units: decibels (dB) SEL: Units: decibels (dB)
Equivalent Continuous Sound Level Sound Exposure Level
It is the average sound level for the measurement period based on a 3 dB Exchange Rate (ER). It is the constant sound level which, if lasting for one second, would deliver the same amount of acoustical energy
as that delivered over the entire measurement period.
If the Exchange Rate (ER) is 4, 5, or 6 dB, then LEQ becomes LAVG.
Technically speaking, it is usually measured with a 3 dB Exchange Rate. However, the dosimeter will also allow
W444444444444444444444444444444444444444444444444444444444444444U SEL to be measured with 4, 5, or 6 dB Exchange Rates. On a printout, the exchange rate is shown in parenthesis.
LDN: Units: decibels (dB)
Day/Night Sound Level W444444444444444444444444444444444444444444444444444444444444444U
It is the average sound pressure during a 24 hour calendar day. However, all sound pressures between the hours SLOW: Units: Time (seconds)
of 10 pm and 7 am (2200 to 0700 hours) are increased by 10 dB before being averaged. Unless otherwise stated, Slow Response
A-Weighting should be used and is generally assumed. A Time Constant of 1 second. When Slow is used, a fluctuating noise into the dosimeter will cause SPL to operate
in a much slowed-down fashion.
The dosimeter will only measure LDN when programmed to do so. It automatically will program itself to the 3 dB
Exchange Rate since LDN is not defined for other Exchange Rates. W444444444444444444444444444444444444444444444444444444444444444U
SPL: Units: decibels (dB)
W444444444444444444444444444444444444444444444444444444444444444U Sound Pressure Level
LN: Units: decibels (dB) / % of Run Time It is the sound pressure, referred to 20 uPa (0.00002 N/m2). The word "Level" indicates that the sound pressure is
Exceedance Level a certain level above the reference level. The SPL is displayed each second as the maximum value (Slow or Fast
Each Exceedance Level shows the level that was exceeded for the percentage of total Run Time. Response) for the previous 1 second period.
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444U
MAX: Units: decibels (dB) STATISTICS:
Maximum Level % Time Statistical Distribution
The highest sound pressure level that occurs during a given time period. For a given run time, the percentage of time that a sound level occured at a specific dB level.
W444444444444444444444444444444444444444444444444444444444444444U W444444444444444444444444444444444444444444444444444444444444444U
MIN: Units: decibels (dB) TIME HISTORY: Units: Listed Form (Min. or Sec.)
Minimum Level Time History
The lowest sound pressure level that occurs during a given time period. A Printout list (and Graph) showing how levels were accumulated over time. A printout can be made for each of the
following: Lavg, Lmax, Peak, and LC-A.
W444444444444444444444444444444444444444444444444444444444444444U
PEAK: Units: decibels (dB) C - A is used mainly when determining earmuff noise reduction effectiveness.
Absolute Unweighted Peak
It is the highest instantaneous sound pressure that occurs during a given time period. W444444444444444444444444444444444444444444444444444444444444444U
TL: Units: decibels (dB)
W444444444444444444444444444444444444444444444444444444444444444U Threshold Level
PrD.(Prt): Units: Percent (%) It is a preset dB level below which sound is not accumulated or averaged into LAVG, LEQ, or Dose.
Projected Dose
It is computed by measuring dose for some time period and extrapolating it to a different time period. W444444444444444444444444444444444444444444444444444444444444444U
(Example: 50% Dose / 4 hrs = 75% Projected Dose / 6 hrs)
W444444444444444444444444444444444444444444444444444444444444444U
94 95
W444444444444444444444444444444444444444444444444444444444444444U Acoustical Formulas
TWA (Prt): Units: decibels (dB)
Projected Time Weighted Average Both the Q-400 and Q-500 use the following formulas to calculate the accumulated data:
It is used to determine the TWA when the operator wishes to use a measurement time that is different from the
worker's exposure time.
Note that a short term sample can only be used if the average noise in the work area is relatively constant.
For example, a worker is only working a 4 hour day and you wish to determine the worker's TWA by making a 5
minute measurement, do as follows:
1.) Program the Projection Time (Prt) to 4 hours. Note that the preset time period can be set from 1 to 16
Hours in 1/4 Hour steps.
2.) Take a 5 minute sample. This average level is then assumed to be constant over the Projection Time (Prt).
3.) Read out the Projected Time Weighted Average [TWA (Prt)].
W444444444444444444444444444444444444444444444444444444444444444U
TWA: Units: decibels (dB)
Time Weighted Average
It is the sound level that is accumulated for any time period but with its average level computed over an 8 hour time
period.
# If the time period is less than 8 hours, the Time Weighted Average will always be less than the Average
Sound Level (LAVG).
# If the time period is more than 8 hours, the Time Weighted Average will always be more than the Average
Sound Level (LAVG).
It is usually measured with A Weighting, Slow Response, and a 5 dB Exchange Rate. However, the dosimeter will
allow either A or C-Weighting, either Slow or Fast Response, and either a 3, 4, 5, or 6 dB Exchange Rate.
W444444444444444444444444444444444444444444444444444444444444444U
UL: Units: Minutes / Seconds
Upper Limit Time
It is the total time that the sound level exceeds a preset level.
W444444444444444444444444444444444444444444444444444444444444444U
96 97
Where:
LS = Sound Level in dB with the selected Time Constant (Slow or Fast). Its value is entered only when the
Sound Level is greater than the Threshold Level. It is entered as minus infinity if the level is less than the
Threshold Level.
LD = Same as LS except 10 dB is added to the Sound Level between 2200 (10 P.M.) and 0700 (7 A.M.) if the
dosimeter is operating in the Ldn mode.
DOSE CL LAVG
PrD (Projected Dose) ER LEQ
2
TWA Pa H (Exposure) SEL
TWA (Projected TWA) LDN % TIME STAT DIST
See APPENDIX, Acoustical Definitions.
98 99
Microphone Input / Data Output Connector Parallel Printer Interface
100 101
Serial Communications Interface QUEST SERVICE AND WARRANTY POLICY
The Serial Communications Interface is compatible with most Computer COM Ports and most Serial Printers. It converts Service Policy
data from the Q-400 and Q-500 into serial information. The Quest product you have purchased is one of the finest acoustic instruments available. It is backed by our full one year
warranty which seeks complete customer satisfaction. This is your assurance that you can expect prompt courteous service
for your equipment from the entire Quest service organization.
Should your Quest equipment need to be returned for repair or recalibration, please contact the Service Department at
1(800)245-0779 (USA) or Fax (262)567-4047 for a Return Authorization Number. The RA number is valid for 30 days, and
must be shown on the shipping label and purchase order/cover letter. If you are unable to return instruments in that time
call for a new RA number. Send it prepaid and properly packed in the original shipping carton directly to Quest
Technologies, 1060 Corporate Center Dr., Oconomowoc, WI 53066 U.S.A.
Repair or replacement work done under warranty will be performed free of charge, and the instrument will be returned to
you prepaid. Your copy or a photocopy of the Quest Registration Card will serve as proof of warranty should the factory
require this information.
If for any reason you should find it necessary to contact the factory regarding service or shipping damage, please direct
your calls or letters to the attention of the Service Manager, Quest Technologies, (262) 567-9157 or (800) 245-0779. Office
hours are from 8 AM to 5 PM (Central Standard Time) Monday through Friday.
Warranty
Quest Technologies warrants our instruments to be free from defects in materials and workmanship for one year under
normal conditions of use and service. For U.S.A. customers we will replace or repair (our option) defective instruments at no
charge, excluding batteries, abuse, misuse, alterations, physical damage, or instruments previously repaired by other than
Quest Technologies. Microphones, sensors, printers and chart recorders may have shorter warranty periods. This
warranty states our total obligation in place of any other warranties expressed or implied. Our warranty does not include
any liability or obligation directly resulting from any defective instrument or product or any associated damages, injuries, or
property loss, including loss of use or measurement data.
For warranty outside the U.S.A., a minimum of one year warranty applies to the same limitation and exceptions as above
with service provided or arranged through the authorized Quest sales agent or our Quest European Service Laboratory.
Foreign purchasers should contact the local Quest sales agent for details.
Figure 22. Serial Communications Interface.
102 103