Fibro Scan®
Fibro Scan®
User manual
for the FibroScan 630 ®
en-GB
en-GB FIBROSCAN USER MANUAL 630
2
FIBROSCAN USER MANUAL 630 en-GB
TABLE OF CONTENTS
2. WARNINGS .................................................................................................................... 9
2.1. Peripherals not supplied by Echosens ........................................................................ 9
2.2. Peripherals supplied but not manufactured by Echosens ........................................... 9
2.3. Electrical safety ........................................................................................................... 9
2.4. Electromagnetic safety .............................................................................................. 10
2.5. Magnetic resonance safety........................................................................................ 10
2.6. Using the device ........................................................................................................ 10
2.7. Deleting measurements............................................................................................. 11
2.8. Patient safety............................................................................................................. 11
2.9. Using the B-Mode localisation module ...................................................................... 11
2.10. Switching off the unit ................................................................................................. 11
2.11. Cleaning and maintenance........................................................................................ 11
2.12. Cleaning .................................................................................................................... 11
2.13. Interpreting the result................................................................................................. 12
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
3
en-GB FIBROSCAN USER MANUAL 630
6. EXAMINATION PROCEDURE..................................................................................... 32
6.1. Selection of the probe and examination type ............................................................ 32
6.1.1. Liver examination......................................................................................... 32
6.1.2. Spleen examination ..................................................................................... 33
6.2. Performing the examination....................................................................................... 33
7. SOFTWARE INTERFACE............................................................................................ 35
7.1. Authentication page................................................................................................... 35
7.2. Home screen ............................................................................................................. 35
7.3. Using the interface keypad ........................................................................................ 36
7.4. Status bar .................................................................................................................. 36
7.5. The patient record screen.......................................................................................... 38
7.6. Acquisition screen ..................................................................................................... 40
7.6.1. Localisation phase ....................................................................................... 41
7.6.2. Acquisition screen for Liver examination ..................................................... 42
7.6.3. Acquisition screen for Spleen examination (only available with the
EXPERT version)......................................................................................... 43
7.6.4. Patient data.................................................................................................. 44
7.6.5. One-dimensional ultrasound localisation ..................................................... 44
7.6.6. B-Mode ultrasound localisation ................................................................... 45
7.6.7. Force indicator ............................................................................................. 47
7.6.8. Elastogram................................................................................................... 48
7.6.9. Valid stiffness measurement counter........................................................... 48
7.6.10.Stiffness results area ................................................................................... 49
7.6.11.CAP-LEVEL (only for CAPc)........................................................................ 50
7.6.12.CAP result zone (option).............................................................................. 50
7.6.13.Measurements displayed as image carrousel.............................................. 51
7.6.14.Displays measurements as vignettes .......................................................... 52
7.6.15.Displaying the B-Mode ultrasound localisation screenshot.......................... 52
7.6.16.Deleting measurements ............................................................................... 52
7.6.17.Adding a comment ....................................................................................... 53
7.6.18.Alert display area ......................................................................................... 54
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
4
FIBROSCAN USER MANUAL 630 en-GB
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
5
en-GB FIBROSCAN USER MANUAL 630
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
6
FIBROSCAN USER MANUAL 630 en-GB
The present User Manual has no contractual value whatsoever and under no circumstances
may Echosens be held responsible on the basis of the information contained in the present
manual.
The present User Manual details, on the one hand, all of the information required for the
implementation, use and maintenance of the FibroScan instrument and, on the other hand,
the list of information displayed.
IMPORTANT NOTE: The FibroScan 630 is available in two different versions: PRIME and
EXPERT. The EXPERT version is an extension of the PRIME version, additionally offering
spleen stiffness measurement and the B-Mode localisation module. Unless otherwise
stated, the contents of this User Manual apply to the PRIME and EXPERT variants of
the FibroScan 630 equally.
Echosens SA publishes this manual "as is'", without guarantees of any nature, whether
explicit or implicit, including, but not limited to, implicit guarantees concerning merchantability
or fitness for a particular use, for the purpose of providing simple and accurate information.
Consequently, Echosens SA cannot accept any responsibility for any incorrect interpretation
of the manual. Though all efforts have been made to offer a manual that is as accurate
as possible, the manual may nevertheless contain some technical inaccuracies and/or
typographical errors.
Echosens cannot, under any circumstances, be held responsible for any loss of profit,
loss of business, data loss, business interruption, or for any indirect, specific, accidental
or consecutive damages of any kind. In the event of damages arising from a defect
(imperfection) or error contained in the present User Manual, Echosens undertakes to send
the physician, as rapidly as possible, a hard copy or electronic document containing all
corrections made to the present manual.
This manual is updated on a regular basis. The most recent version of this manual is available
from Echosens on simple request. Should any major modifications be made to the manual,
however, Echosens undertakes to send the physician, as rapidly as possible, a new copy of
the manual in hard copy or electronic format. Note that this does not involve updating the
hardware and/or software in your possession.
The product owner must keep the present manual for as long as the product is used.
The present manual contains a chapter for troubleshooting the most commonly encountered
problems.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
7
en-GB FIBROSCAN USER MANUAL 630
Any information or modification requests pertaining to this manual should be sent to:
https://support.echosens.com.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
8
FIBROSCAN USER MANUAL 630 en-GB
2. WARNINGS
When connecting any peripheral not supplied by Echosens to the FibroScan, refer
to the manufacturer's User Manual to ensure operator and patient safety.
To avoid the risk of electric shock, this device must be connected to an earthed
power supply only.
Multi-socket adapters and extension leads must not be connected to the device.
The power cable socket is intended for use as an isolating device and must be
easily accessible to prevent connection problems.
Any parts not specified in the user manual must not be connected to the system.
Make sure that the vents are not obstructed. If they are, the electronic equipment
could overheat, causing irreversible damage.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
9
en-GB FIBROSCAN USER MANUAL 630
With the EXPERT variant fitted with the B-Mode localisation module, to avoid the
risk of electrical shock and fire hazard:
Before using the B-mode probe, inspect the probe face, housing, and cable.
Do not use the probe if the probe or its cable is damaged.
Do not use any B-mode probe that has been immersed beyond the specified
cleaning or disinfection level.
Only the Linemaster GEM SWNO control pedal is compatible with the FibroScan.
Under no circumstances should a probe not of this reference be connected to the
FibroScan.
The use of accessories not specified in the user guide may cause a non-
compliance in terms of electromagnetic compatibility (EMC).
Avoid using the FibroScan device when placed upon or near a machine that
generates electromagnetic disturbance.
With the EXPERT variant, when the B-Mode localisation probe is not being used,
secure the probe cable safely out of the way using the clip located on the right-
hand side of the device, underneath the probe holder bracket.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
10
FIBROSCAN USER MANUAL 630 en-GB
All measurements performed before the one chosen for deletion will be eliminated
from the examination after confirmation.
If second generation CAP (CAPc) is enabled on your device, the CAPc
measurements performed prior to the stiffness measurement you have selected
for deletion will also be deleted.
Never switch the unit off during an examination or whilst in configuration mode.
Never disconnect the main power supply when the device is switched on. Failure
to comply with these instructions could cause a malfunction of the machine and/or
loss of data.
These maintenance operations must not be performed by a third party other than
a technician authorised by Echosens.
The probe must be calibrated periodically. Beyond the period indicated on the
calibration certificate, the manufacturer no longer guarantees the performance
characteristics of the probe.
2.12. CLEANING
To prevent electric shock, switch off the device and disconnect it from the power
supply before cleaning.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
11
en-GB FIBROSCAN USER MANUAL 630
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
12
FIBROSCAN USER MANUAL 630 en-GB
3. MISCELLANEOUS INFORMATION
3.1. GUARANTEE
The terms of guarantee are stated in the Echosens terms of sale documents.
For any claims, Echosens is at the disposal of the physician and their assistants and shall, if
necessary, pass the aforementioned claim on to the competent local representative.
3.2. LIABILITY
The information displayed on the FibroScan screen is the result of complex calculations
performed by the software application built into the FibroScan. These results are then
interpreted by the physician in charge. Under no circumstances, and even if Echosens had
been notified, can Echosens be held responsible for the incorrect interpretation of these
results; Echosens' liability being limited to making the measurements, displaying them and
printing them via the FibroScan.
The data from each examination are saved on the machine's hard disk. The user is
responsible for saving the data on a regular basis. Echosens cannot under any circumstances
be held responsible for the partial or total loss of FibroScan data.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
13
en-GB FIBROSCAN USER MANUAL 630
Microsoft Excel and Windows Embedded are registered trademarks of Microsoft Corporation
in the United States and other countries.
Patents: www.echosens.com/patents
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
14
FIBROSCAN USER MANUAL 630 en-GB
The FibroScan 630 and its probes form an active, non-implantable medical device using
ultrasound.
Liver stiffness and liver CAP are indicated as an aid to diagnosis and monitoring of paediatric
patients as part of an overall assessment of liver disease.
Note: the parameters listed in Table 1, known to influence liver stiffness, shall be considered
when interpreting liver stiffness.
Table 1: parameters influencing liver stiffness.
Parameter Reference
Liver fibrosis, cirrhosis [1-15]
Acute hepatitis, inflammation, ALT flares, excessive alcohol intake [16-19, 20, 21, 22, 23]
Portal pressure [24, 25]
Extra-hepatic cholestasis [26-28]
Hepatic congestion, central venous pressure [29, 30, 31]
Amyloidosis [32-34]
Food intake [35-39]
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
15
en-GB FIBROSCAN USER MANUAL 630
References:
[2] Wong, V.W., et al., Diagnosis of fibrosis and cirrhosis using liver stiffness measurement in
nonalcoholic fatty liver disease. Hepatology, 2010. 51(2): p. 454-462.
[3] Pavlov, C.S., et al., Systematic review with meta-analysis: diagnostic accuracy of transient
elastography for staging of fibrosis in people with alcoholic liver disease. Alimentary
Pharmacology & Therapeutics, 2016. 43(5): p. 575-85.
[4] Li, Y., et al., Systematic review with meta-analysis: the diagnostic accuracy of transient
elastography for the staging of liver fibrosis in patients with chronic hepatitis B. Aliment
Pharmacol Ther, 2016. 43(4): p. 458-69.
[5] Chon, Y.E., et al., Performance of transient elastography for the staging of liver fibrosis in
patients with chronic hepatitis B: a meta-analysis. PLoS One, 2012. 7(9): p. e44930.
[6] Tsochatzis, E.A., et al., Elastography for the diagnosis of severity of fibrosis in chronic
liver disease: a meta-analysis of diagnostic accuracy. Journal of Hepatology, 2011. 54(4): p.
650-9.
[7] Sanchez-Conde, M., et al., Comparison of transient elastography and liver biopsy for the
assessment of liver fibrosis in HIV/hepatitis C virus-coinfected patients and correlation with
noninvasive serum markers. Journal of viral hepatitis, 2010. 17(4): p. 280-6.
[8] Castera, L., et al., Prospective comparison of transient elastography, Fibrotest, APRI, and
liver biopsy for the assessment of fibrosis in chronic hepatitis C. Gastroenterology, 2005.
128(2): p. 343-50.
[10] Hartl, J., et al., Transient elastography in autoimmune hepatitis: Timing determines the
impact of inflammation and fibrosis. Journal of Hepatology, 2016. 65(4): p. 769-75.
[11] Siddiqui, M.S., et al., Vibration-controlled Transient Elastography to Assess Fibrosis and
Steatosis in Patients With Nonalcoholic Fatty Liver Disease. Clinical gastroenterology and
hepatology, 2019. 17(1): p. 156-163.
[12] Afdhal, N.H., et al., Accuracy of fibroscan, compared with histology, in analysis of
liver fibrosis in patients with hepatitis B or C: a United States multicenter study. Clinical
Gastroenterology & Hepatology, 2015. 13(4): p. 772-779 e3.
[13] Nguyen-Khac, E., et al., Non-invasive diagnosis of liver fibrosis in patients with alcohol-
related liver disease by transient elastography: an individual patient data meta-analysis. The
lancet. Gastroenterology & hepatology, 2018. 3(9): p. 614-625.
[14] Eddowes, P.J., et al., Accuracy of FibroScan Controlled Attenuation Parameter and Liver
Stiffness Measurement in Assessing Steatosis and Fibrosis in Patients With Nonalcoholic
Fatty Liver Disease. Gastroenterology, 2019. 156(6): p. 1717-1730.
[15] Lee, C.K., et al., Validation of Transient Elastography Cut Points to Assess Advanced
Liver Fibrosis in Children and Young Adults: The Boston Children's Hospital Experience. J
Pediatr, 2018. 198: p. 84-89 e2.
[16] Arena, U., et al., Acute viral hepatitis increases liver stiffness values measured by
transient elastography. Hepatology, 2008. 47(2): p. 380-4.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
16
FIBROSCAN USER MANUAL 630 en-GB
[17] Coco, B., et al., Transient elastography: a new surrogate marker of liver fibrosis
influenced by major changes of transaminases. Journal of Viral Hepatitis, 2007. 14(5): p.
360-9.
[18] Sagir, A., et al., Transient elastography is unreliable for detection of cirrhosis in patients
with acute liver damage. Hepatology, 2008. 47(2): p. 592-5.
[19] Vigano, M., et al., Transient elastography assessment of the liver stiffness dynamics
during acute hepatitis B. European Journal of Gastroenterology & Hepatology, 2010. 22(2):
p. 180-4.
[20] Wong, G.L., et al., Increased liver stiffness measurement by transient elastography in
severe acute exacerbation of chronic hepatitis B. Journal of Gastroenterology & Hepatology,
2009. 24(6): p. 1002-7.
[21] Mueller, S., et al., Increased liver stiffness in alcoholic liver disease: differentiating fibrosis
from steatohepatitis. World Journal of Gastroenterology, 2010. 16(8): p. 966-72.
[22] Trabut, J.B., et al., Rapid decline of liver stiffness following alcohol withdrawal in heavy
drinkers. Alcoholism, clinical and experimental research, 2012. 36(8): p. 1407-11.
[23] Bardou-Jacquet, E., et al., Effect of alcohol consumption on liver stiffness measured by
transient elastography. World Journal of Gastroenterology, 2013. 19(4): p. 516-22.
[24] Carrion, J.A., et al., Transient elastography for diagnosis of advanced fibrosis and
portal hypertension in patients with hepatitis C recurrence after liver transplantation. Liver
Transplantation, 2006. 12(12): p. 1791-8.
[25] Vizzutti, F., et al., Liver stiffness measurement predicts severe portal hypertension in
patients with HCV-related cirrhosis. Hepatology, 2007. 45(5): p. 1290-7.
[26] Millonig, G., et al., Extrahepatic cholestasis increases liver stiffness (FibroScan)
irrespective of fibrosis. Hepatology, 2008. 48(5): p. 1718-23.
[27] Harata, M., et al., Liver stiffness in extrahepatic cholestasis correlates positively with
bilirubin and negatively with alanine aminotransferase. Hepatology Research, 2011. 41(5): p.
423-9.
[28] Trifan, A., et al., Increased liver stiffness in extrahepatic cholestasis caused by
choledocholithiasis. Hepatitis Monthly, 2011. 11(5): p. 372-5.
[29] Colli, A., et al., Decompensated chronic heart failure: increased liver stiffness measured
by means of transient elastography. Radiology, 2010. 257(3): p. 872-8.
[30] Lebray, P., et al., Liver stiffness is an unreliable marker of liver fibrosis in patients with
cardiac insufficiency. Hepatology, 2008. 48(6): p. 2089.
[31] Millonig, G., et al., Liver stiffness is directly influenced by central venous pressure.
Journal of Hepatology, 2010. 52(2): p. 206-10.
[32] Janssens, F., et al., Hepatic amyloidosis increases liver stiffness measured by transient
elastography. Acta Gastroenterologica Belgica, 2010. 73(1): p. 52-4.
[33] Lanzi, A., et al., Liver AL amyloidosis as a possible cause of high liver stiffness values.
European Journal of Gastroenterology & Hepatology, 2010. 22(7): p. 895-7.
[35] Mederacke, I., et al., Food intake increases liver stiffness in patients with chronic or
resolved hepatitis C virus infection. Liver International, 2009. 29(10): p. 1500-6.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
17
en-GB FIBROSCAN USER MANUAL 630
[36] Arena, U., et al., Liver stiffness is influenced by a standardized meal in patients with
chronic hepatitis C virus at different stages of fibrotic evolution. Hepatology, 2013. 58(1): p.
65-72.
[37] Berzigotti, A., et al., Effect of meal ingestion on liver stiffness in patients with cirrhosis
and portal hypertension. PLoS One, 2013. 8(3): p. e58742.
[38] Kjaergaard, M., et al., High risk of misinterpreting liver and spleen stiffness using 2D
shear-wave and transient elastography after a moderate or high calorie meal. PLoS One,
2017. 12(4): p. e0173992.
[39] Vuppalanchi, R., et al., Is Fasting Necessary for Individuals With Nonalcoholic Fatty Liver
Disease to Undergo Vibration-Controlled Transient Elastography? Am J Gastroenterol, 2019.
114(6): p. 995-997.
Liver examination
The probe and type of examination must be selected based on the recommendations
detailed in the “Examination procedure” section, page 32.
With SmartDepth option enabled, measurements should not be performed if the device
displays the “Liver too deep” warning. In this case, the operator can change the probe
and/or measurement point.
With SmartDepth option disabled, with the M+ probe: measurements should not be
performed if the PCD (probe-to-capsule distance) is more than 25 mm. In this case, the
operator can change the probe and/or measurement point.
With SmartDepth option disabled, with the XL+ probe: measurements should not be
performed if the Probe-to-capsule distance is more than 35 mm. In this case, the
operator can change the measurement point.
Spleen examination methods must not be used for the liver.
Spleen examination
The FibroScan device is indicated for use in adult patients for detecting a homogeneous
splenic parenchyma located between 25 mm and 55 mm beneath the surface of the
skin.
Liver examination methods must not be used for the spleen.
In all cases
The patient must fast for at least 3 hours before the examination.
The clinical personnel must follow normal safety procedures.
The presence of ascites between the probe and the organ to be measured may prevent
measurements from being obtained.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
18
FIBROSCAN USER MANUAL 630 en-GB
Once an optimal image is achieved, the need for prolonging the exposure cannot be justified.
4.5. CONTRAINDICATIONS
To ensure patient safety, the FibroScan device must not be used in the following situations:
On an organ other than the liver (for PRIME version).
On organs other than the spleen or liver (for EXPERT version).
On the eyes and mucosa.
On wounds.
Use of the M+ probe is not approved for patients under 14 years old.
first generation CAP is measured at the same time as the stiffness measurement, and
concerns the same explored volume. This measurement is available with the M+ and XL+
probes only.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
19
en-GB FIBROSCAN USER MANUAL 630
second generation CAP (CAPc) uses the same ultrasound acquisition sequences (emission/
reception) as those used for measuring liver stiffness. Ultrasound data is continuously
collected during the examination. The sequences retained are those that meet predefined
quality criteria. For second generation CAP, more data is therefore used than for first
generation CAP. This mechanism thus permits better CAP measurement repeatability. This
measurement is available with the S+, M+ and XL+ probes.
Prior to installation, ensure that the operating and mains voltage values match.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
20
FIBROSCAN USER MANUAL 630 en-GB
The electrical power lead provided must be connected to the FibroScan mains connector and
to an earthed socket. Correct earthing operation can only be guaranteed if the FibroScan is
connected to a socket compliant with safety standards.
When the safe use of the FibroScan is no longer possible, the device must be taken out of
operation. Steps must be taken to prevent its inadvertent use. The medical device is entrusted
to authorised technicians for inspection.
For correct and safe use and for all maintenance operations, personnel must conform to the
user precautions.
With the EXPERT variant, to protect the B-Mode probe – notably its transducer –, follow these
precautions:
excessive cables bending or twisting can cause a failure or intermittent operation;
improper cleaning or disinfecting of any system part can cause permanent damage, for
cleaning and disinfecting instructions see chapter Nettoyage, entretien et dépannage;
do not use solvents such as thinner/benzene, or abrasive cleaners on any part of the
system;
do not spill liquid on the system.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
21
en-GB FIBROSCAN USER MANUAL 630
5. EXTERNAL PRESENTATION
S+ probe
M+ probe
XL+ probe
Control pedal
Refer to the warnings in Chapter 2 concerning the use of peripherals not supplied
by Echosens.
Refer to the warnings in Chapter 2 concerning the use of peripherals supplied but
not manufactured by Echosens.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
22
FIBROSCAN USER MANUAL 630 en-GB
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
23
en-GB FIBROSCAN USER MANUAL 630
Start-up/standby button
This button is enabled only when the main switch, located at the rear of the device,
is set to "I".
The touch screen button flashes green when the device is starting up or going into standby.
It shows steady green when the device is started up and ready to use.
To switch the device on, press the button until you hear a beep. Release. This button loads
the application. After a few seconds, the home window is displayed.
The touch screen button flashes orange in quick succession in the event of device critical
error. A fast beeping sound is also emitted.
Power indicator
The status of the power indicator is defined as follows:
Off: Device in standby or not connected to power supply.
On: Device started up and ready to use.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
24
FIBROSCAN USER MANUAL 630 en-GB
View of the front panel of the FibroScan unit: A: Start-up/standby button. B: Power supply
indicator.
To protect the screen from any risk of damage, take care not to hang the power
cable on the top of the machine.
The software is automatically launched when the FibroScan is switched on. It is used to:
perform examinations,
manage examinations saved in archives.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
25
en-GB FIBROSCAN USER MANUAL 630
Computer connectors
Two USB 3.0 sockets: To connect an external hard drive for backups, a USB storage device,
or a USB printer.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
26
FIBROSCAN USER MANUAL 630 en-GB
View of main switch and power supply connector: A: Main switch and access to fuses. B:
Electrical power supply connector.
Main switch
This switch has two positions:
Position 0: no voltage is present in the internal circuits of the FibroScan, meaning that it
consumes no electrical energy. This is the usual position when the FibroScan is no
longer to be used (at the end of the day for example).
Position 1: the internal electrical circuits of the FibroScan are powered by the AC power
supply. This is the FibroScan's working position.
This switch assembly has a removable part. It provides access to the two FibroScan
protection fuses.
The bottom part of the switch device receives the power cable plug connection outlet.
Earthing connector
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
27
en-GB FIBROSCAN USER MANUAL 630
Auxiliary connectors
Probe connectors
Location of probe connectors: A: Probe connectors. B: Location for the connector of the
unconnected probe.
Location (B) is not operational. It protects the connector of the unconnected probe.
Before using the FibroScan device, the operator must be able to recognize the Electrostatic
discharge (ESD) symbol, and must take the appropriate precautionary measurements as
described below:
Do not touch the pins of connectors marked with the ESD warning symbol,
Before making any connections, discharge the static electricity from your body, either by
touching and remaining in contact with a grounded metal object, or using a properly
grounded antistatic bracelet or a potential equalization terminal.
Training in the meaning of the ESD symbol and ESD precautionary procedures is
strongly recommended.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
28
FIBROSCAN USER MANUAL 630 en-GB
Measurement button
As soon as this button is pressed (if sufficient force is placed on the transducer), the
electrodynamic transducer generates a shear wave (s-wave) that painlessly impacts the
patient's skin. At the same time, the ultrasound transducer performs a series of acquisitions
(emission/reception) to measure the propagation speed of this shear wave. Acquisition lasts
less than one tenth of a second.
Indicators
The indicator lights (LEDs) display a status as follows:
Lit up blue during FibroScan start-up and when standing by to launch an examination.
Flashing blue lights for the probe selected when an examination starts.
Not lit up during an examination when the operator is not applying correct force on
patient’s body.
Lit up blue during an examination when the operator is applying the correct force to the
patient's body. It is however strongly recommended that you view the force exerted by
looking at the on-screen force indicator.
Lead
This 1.5 m lead connects the probe to the FibroScan by means of a multi-pin jack with guide
pin.
The probe transducer, the probe jack, and the FibroScan connector are fragile
elements and must be handled with care.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
29
en-GB FIBROSCAN USER MANUAL 630
The probe jack has a red dot that should be aligned with the red dot on the FibroScan socket
before insertion.
The serial number marked on the connector identifies the probe uniquely.
The convex transducer array of the B-Mode ultrasound localisation probe is a “BF Type”
applied part.
Lead
This cable, no more than 3 m long, connects the probe to the FibroScan device by means of
an EX10E-280P HRS foolproof connector.
The probe transducer, the probe jack, and the connector are fragile elements and
must be handled with care.
The serial number marked on the connector identifies the probe uniquely.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
30
FIBROSCAN USER MANUAL 630 en-GB
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
31
en-GB FIBROSCAN USER MANUAL 630
6. EXAMINATION PROCEDURE
For patients younger than 18 years old, the type of examination depends on the patient's
thoracic perimeter (TP), as per the following recommendations:
If the thoracic perimeter is more than 75 cm, an M examination should be performed.
If the thoracic perimeter is between 45 and 75 cm, an S2 examination should be
performed.
If the thoracic perimeter is less than 45 cm, an S1 examination should be performed.
For adult patients, the examination should be selected based on the following
recommendations:
Start with an M examination.
If the FibroScan device recommends the XL examination, the XL+ probe should be
used. Otherwise, the M examination can be performed.
With SmartDepth option enabled: measurements should not be performed if the
device displays the “Liver too deep” warning. In this case, the operator can change the
probe and/or measurement point.
With SmartDepth option disabled, with the M+ probe: measurements should not be
performed if the PCD (probe-to-capsule distance) is more than 25 mm. In this case, the
operator can change the probe and/or measurement point.
With SmartDepth option disabled, with the XL+ probe: measurements should not be
performed if the Probe-to-capsule distance is more than 35 mm. In this case, the
operator can change the measurement point.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
32
FIBROSCAN USER MANUAL 630 en-GB
* With SmartDepth option enabled, the “Liver too deep” warning may be displayed on the
screen. If SmartDepth option is disabled, PCD should be monitored by A and TM Mode
ultrasound signals to avoid performing measurements with PCD higher than 35 mm.
Liver examination
For Liver examinations, the stages are as follows:
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
33
en-GB FIBROSCAN USER MANUAL 630
1. In a quiet environment, lie the patient down on their back with their right arm behind
their head, fully abducted.
2. Leave the patient to rest for at least 5 minutes so that they are completely relaxed.
3. Start the FibroScan examination.
4. Select the probe and type of examination appropriate for the patient (see section
“Selection of the probe and examination type”, page 32, and "Automatic probe
recommendation", page 56).
5. Position the FibroScan probe transducer on the patient’s skin, between the ribs above
the liver, along the midaxillary line.
6. Use the FibroScan probe's ultrasound localisation tools to locate an area of the liver
that meets the FibroScan examination quality criteria: Homogeneous tissue with
optimum ultrasound signal throughout the entire measurement depth (for further
details, please refer to the FibroScan training material). If necessary, use the
localisation probe in B-Mode. This phase is formalised by the Localisation phase.
7. Take a minimum of 10 valid measurements in each examination, using the same
measurement point each time. If second generation CAP (CAPc) is enabled, maintain
the probe positioned on the patient until 100 % of the CAP measurement is attained.
The final stiffness and CAP (if first generation CAP is enabled) values saved are the median
of all valid stiffness and CAP measurements performed during the examination. If second
generation CAP is enabled, the final CAP value is calculated as the mean of all CAP values
measured.
Spleen examination
For Spleen examinations, the stages are as follows:
1. In a quiet environment, lie the patient down on their back with their left arm behind
their head, fully abducted.
2. Leave the patient to rest for at least 5 minutes so that they are completely relaxed.
3. Start the FibroScan examination.
4. Locate the spleen beforehand using the B-Mode ultrasound localisation module.
5. Position the FibroScan probe transducer on the patient’s skin, between the ribs, above
the area previously located using the B-Mode localisation probe.
6. Use the FibroScan probe's ultrasound localisation tools to locate an area of the spleen
that meets the FibroScan examination quality criteria: Homogeneous tissue with
optimum ultrasound signal throughout the entire measurement depth (for further
details, please refer to the FibroScan training material).
7. Take a minimum of 10 valid measurements in each examination, using the same
measurement point each time.
The final stiffness measurement value saved is the median of all of the valid stiffness
measurements performed during the examination.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
34
FIBROSCAN USER MANUAL 630 en-GB
7. SOFTWARE INTERFACE
The software loaded when the FibroScan unit is started up is used to:
perform examinations,
print the results,
manage the archives,
export the data in several formats.
If the device is connected to the hospital network, the software also makes it possible to:
import the lists of patients,
export the results of the FibroScan examination.
Before you reach the authentication page, the following messages may be displayed:
No probe connected
Probe out of calibration (see the Probe Calibration section)
Restricted user accounts (Manufacturer and Distributor) can be accessed by clicking on the
button.
The enabling of secure session opening and the user accounts are configurable.
See the Configuration section.
Beyond a certain idle time, the user is disconnected and the application returns to the login
page.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
35
en-GB FIBROSCAN USER MANUAL 630
Description of keypad:
Shift.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
36
FIBROSCAN USER MANUAL 630 en-GB
The CAPc and SmartDepth options are enabled simultaneously. These two
options together are called SmartExam.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
37
en-GB FIBROSCAN USER MANUAL 630
Complete the fields. At least the 'Name' or 'Code' field must be filled in to start the
examination.
Fields can be filled in using the barcode reader . Either by scanning the QR Code available
on the FibroScan examination report – In this case, each field completed in the report will be
copied into the corresponding field on the FibroScan interface. Or by scanning the barcode
on the patient label – In this case, only the Code field will be filled in. Barcode
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
38
FIBROSCAN USER MANUAL 630 en-GB
However, to import a patient waiting for an examination in the coming days, a manual patient
search must be performed by clicking the magnifying glass button in the Worklist
window.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
39
en-GB FIBROSCAN USER MANUAL 630
On this page, the operator's name can be changed and the patient details can be filled
in: Select indication from the list of illnesses specified in Configuration, and fill in fields on
Compliance with food intake recommendations (Fasting > 3 hrs), Weight and Size.
RAW Mode
The RAW Mode button can be used to enable / disable Raw Data saving for the scheduled
examination (only available if RAW Mode has been enabled in Configuration). In RAW Mode,
the results and conditions of all individual measurements can be saved, whether valid or not.
In addition, RAW Mode saves all of the raw data acquired via ultrasound examinations and
used for calculation of all individual measurements, whether valid or not.
Saving Raw Data can slow the application down considerably. Enable it only for
clinical studies that explicitly require it.
The Patient Code entered governs how the examination can be searched for in
the Archives. Particular care should be taken when entering this code in order to
prevent duplication of the clinical trial data concerned.
Additional fields
If the FibroScan is connected to FibroView Gateway, additional fields are available via the
examination's data entry window. Click on the button to edit their content. The names and
data entry rules for each field are defined on the Gateway.
After five minutes of idle time during an acquisition, a message tells the user, who
then has five additional minutes before the examination is automatically halted.
Beyond this time, if no action has been taken, the examination is saved and then
stopped.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
40
FIBROSCAN USER MANUAL 630 en-GB
Uncalibrated probe
If the probe is uncalibrated, the following message is displayed:
Unless examination blockage has been activated on the device, the examination may still
be performed, but with the proviso that probe performance is not guaranteed by the
manufacturer..
During this localisation phase, only stiffness measurement values are displayed. The smallest
of these values is displayed.
Once the optimum measurement zone has been identified, press the following button to
proceed to the examination:
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
41
en-GB FIBROSCAN USER MANUAL 630
CAP value (dB/m) is only displayed for a liver examination, if this option is installed.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
42
FIBROSCAN USER MANUAL 630 en-GB
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
43
en-GB FIBROSCAN USER MANUAL 630
Ultrasound signals in TM (left) and Amplitude (right) modes: A: Depth explored (mm). B:
Ultrasound signal represented in TM (Time-Motion) mode. C: Ultrasound signal represented
in A (Amplitude) mode.
As soon as the probe makes contact with the skin, i.e. when a force change is detected, the
ultrasound transducer makes ultrasound acquisitions.
The system displays two ultrasound signals used to locate a zone that satisfies the
measurement criteria:
One in Time-Movement (TM) mode, one-dimensional information that shows, in
greyscale and according to time, the different elements detected by the ultrasound
probe in the course of its scan.
And the other in A-Mode, one-dimensional information on the amplitude of the
ultrasound signal collected by the probe according to depth.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
44
FIBROSCAN USER MANUAL 630 en-GB
These two modes serve to ensure that the probe is correctly positioned to perform a
measurement on a sufficiently thick portion of the organ to be measured, visible throughout
the explored depth. They also allow the operator to ensure that the measurement will not be
disrupted by the presence of structures not of interest, such as blood vessels.
As soon as the probe comes into contact with the patient’s skin, the ultrasound signal quality
indicator shows the quality of the ultrasound signals passing through the liver by means of an
indicator light.
Indicator light: Off: The quality of the ultrasound signals is low. Green: The quality of the
ultrasound signals is high.
The greener the liver indicator, the higher the quality of the ultrasound signals, and the better
the position of the measurement point.
This tool only functions on a liver examination, with the M+ and XL+ probes.
For Liver examinations, when A and TM Modes do not allow for an area of the liver that
meets measurement criteria to be located, B-Mode, which shows a 2D image of the tissues
examined, can help the user to locate a measurement point suitable for stiffness and/or CAP
measurement.
For Spleen examinations, B-Mode, which shows a 2D image of the tissues examined, is
strongly recommended to help locate the spleen.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
45
en-GB FIBROSCAN USER MANUAL 630
As soon as the probe comes into contact with the patient’s skin, B-Mode displays a real-time,
B-Mode ultrasound signal of the area examined.
For Spleen examination, the median axis makes it possible to ensure that spleen thickness
is sufficient on the measurement line to ensure that the stiffness measurements are of the
correct quality.
Three presets are available according to the dimensions of the organ being screened and/or
depth of examination:
Standard: Central frequency = 3.0 MHz Focus = 70 mm Depth = 120 mm
Depth 1: Central frequency = 3.0 MHz Focus = 70 mm Depth = 150 mm
Depth 2: Central frequency = 2.0 MHz Focus = 70 mm Depth = 150 mm
Once you have screened a measurement area suitable for stiffness measurement, use the
following button to capture the image:
This tool must only be used as an aid to locate an area suitable for stiffness and/
or CAP measurement for the purpose of a FibroScan examination.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
46
FIBROSCAN USER MANUAL 630 en-GB
You can go back to the FibroScan examination by placing the FibroScan probe transducer
directly onto the patient’s skin. Doing so automatically disables the B-Mode ultrasound
localisation probe.
The probe contains a sensor that measures the force applied by the operator to the patient.
The force level is given by:
Software: The force indicator (green/orange/red).
The probe: blue LEDs.
Measurements may only be made when the force indicator is in the green zone.
Force too high (red). Force too low (orange). Force correct (green).
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
47
en-GB FIBROSCAN USER MANUAL 630
7.6.8. Elastogram
This image is displayed once the measurement is complete. It represents the levels of liver/
spleen deformation generated by the propagation of the shear wave as a function of time
(horizontal axis in milliseconds) and depth (vertical axis in millimetres).
The colour scale indicates the sign of the deformations (compression or expansion). Black
areas correspond to negative deformation and pale areas to positive deformation. The black
strip through the image represents deformations associated with the passage of the shear
wave, which penetrates progressively deeper into the tissues with time.
When the number of valid stiffness measurements is equal to 1, the IQR and the
IQR/median ratio are undefined; they are therefore not displayed.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
48
FIBROSCAN USER MANUAL 630 en-GB
Median
Stiffness is expressed in kilopascal (kPa). This value is the median of all valid measurements
performed during the examination.
If the repeat measurement is invalid, the median is not re-computed. To obtain a reliable and
representative liver or spleen stiffness measurement, at least 10 valid measurements must
be performed.
IQR/Median
This value, expressed as a percentage, is the ratio of the IQR to the median stiffness. It is
only displayed for Liver examinations, and is recalculated after each new valid measurement.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
49
en-GB FIBROSCAN USER MANUAL 630
CAP-LEVEL indicates whether a sufficient amount of data has been acquired to provide a
CAP result (CAPc only). This amount is expressed in %. A CAP result is displayed when
CAP-LEVEL reaches 50 %. The examination should be continued until CAP-LEVEL
reaches 100 %. Below this level, insufficient data has been acquired and the examination will
only be saved if the operator confirms that they wish to save it.
The CAP is only displayed on a liver examination, with the M+ and XL+ probes.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
50
FIBROSCAN USER MANUAL 630 en-GB
Median
The CAP (Controlled Attenuation Parameter) is expressed in decibels per metre (dB/m). This
value is the median of all valid measurements performed during the examination.
The CAPc is displayed on a Liver examination, with the M+, S+ and XL+ probes.
Mean
CAPc is continuously calculated once a valid measurement has been made. The result is
displayed once the CAP-LEVEL value has reached 50 %. The examination will be saved by
default once the CAP-LEVEL value reaches 100 %. If it is below this value, the operator will
be asked if they wish to save the examination.
The CAP-LEVEL increases when the ultrasound signal quality is sufficient in terms of
homogeneity throughout the entire measurement depth. To assess the homogeneity of the
area being examined, ultrasound localisation tools should be used in A and TM Modes, along
with the ultrasound signal quality indicator: if not shown in steady green, the CAP-LEVEL
value will take longer to reach 100 %.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
51
en-GB FIBROSCAN USER MANUAL 630
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
52
FIBROSCAN USER MANUAL 630 en-GB
Comments can be added and measurement conditions can be entered during the
examination:
If the option is enabled in Configuration, the conditions under which the examination was
conducted can be filled in immediately after finishing the examination. The following input
fields are offered:
patient position,
Choice of optimal measurement position,
Thick subcutaneous tissues.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
53
en-GB FIBROSCAN USER MANUAL 630
Electromagnetic disturbances.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
54
FIBROSCAN USER MANUAL 630 en-GB
Overheat.
Repeat measurement.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
55
en-GB FIBROSCAN USER MANUAL 630
The choice of examination type is governed by the selection of M/XL or S1/S2 when
examination data is entered.
For example, if an M+ probe and an S+ probe are connected to the device, S1/S2
examinations will not be available if M/XL was selected while entering the examination data.
If only an XL+ probe is connected to the device, only XL examinations will be available.
If an M+ probe and an XL+ probe are connected to the device, the examination type will be M
by default.
This tool only functions for Liver examinations, with the M+ and XL+ probes.
The result of this feature is displayed at the top right-hand side of the screen and may be one
of the following three cases:
1. "M": The tool measures a PCD that justifies the use of the M+ probe. The M
examination icon flashes.
2. "XL": The tool measures a PCD that justifies the use of the XL+ probe. The XL
examination icon flashes.
3. “Liver too deep” warning only with SmartDepth enabled: The tool detects a PCD that is
too large. In this case, the operator can change the probe and/or measurement point.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
56
FIBROSCAN USER MANUAL 630 en-GB
With SmartDepth enabled, the warning that the liver is located too deeply beneath the skin is
shown with the symbol:
You are strongly advised to use the recommended tool, to guarantee reliable
results. The decision whether or not to apply this recommendation, however, rests
with the user.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
57
en-GB FIBROSCAN USER MANUAL 630
Opens the tool for exporting the patient file and examination report.
Blocks/Unblocks session.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
58
FIBROSCAN USER MANUAL 630 en-GB
Add a comment.
To display the patient file archives, press the button in the home screen.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
59
en-GB FIBROSCAN USER MANUAL 630
Using the keypad to select files: [CTRL]+click to select non-consecutive files, [SHIFT]+click
the first and last to select a series of files, and [CTRL]+[A] to select all files.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
60
FIBROSCAN USER MANUAL 630 en-GB
The Excel file has two data sheets (Data and Parameters) and may contain all or part of the
patient database.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
61
en-GB FIBROSCAN USER MANUAL 630
Opens the tool for exporting the patient file and examination report.
A single click lists the connections configured on the device. If at least one of these
connections is marked with a green symbol (active connection), a [Retry] button is displayed.
Clicking on this button displays the list of all examinations queued for export for each active
connection.
The [Send all again] button lets you attempt the export again.
If the export is successful, the following symbol is displayed next to each exported
examination. If it is not, the following symbol is displayed.
A connection with a problem is indicated by a red symbol. You should restore this
connection before attempting to retry export.
You may need to click on the button to refresh the status of connections.
Lastly, the [Clear list] button lets you clear the list of queued exports.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
62
FIBROSCAN USER MANUAL 630 en-GB
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
63
en-GB FIBROSCAN USER MANUAL 630
In the event of malfunction, only the staff of Echosens or its local representative is authorised
to work on the FibroScan and its accessories. Any work performed by an unqualified person
will terminate the guarantee.
After opening the package, the device and probes must be cleaned and disinfected
as per the recommendations below.
Apply the following recommendations to clean or disinfect the device, probes, and
accessories.
Failure to observe these recommendations may result in damage to the device and the
probes, which will then no longer be covered by the guarantee.
Recommendations
Always wear eye protection and gloves to prevent injury.
Do not use disinfectant product beyond its expiry date.
Carefully follow the instructions for use that come with the cleaning or disinfectant
product.
Ensure appropriate concentration of cleaning or disinfection product used and time left
in contact with the surface of the device, probes or accessories.
Carefully read the recommendations from the Association for Professionals in Infection
Control and Epidemiology (APIC) and the Food and Drug Administration (FDA), if
applicable in the country. We also recommend careful reading of the “Guideline for
Disinfection and Sterilization in Healthcare Facilities (CDC)”.
Whether surfaces are glass, metal or plastic, they must be cleaned strictly in the order
described below:
1. clean using a soft cloth soaked in the recommended cleaning product,
2. wipe down with a soft cloth dampened with water,
3. dry carefully with a clean, soft and absorbent cloth.
Precautions
Do not spray any cleaning or disinfectant product directly on the device. Leaks may damage
the system, which would then no longer be covered by the guarantee.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
64
FIBROSCAN USER MANUAL 630 en-GB
Do not use paper towels or abrasive products for cleaning, as they may scratch the surface
of the screen.
Surfaces of the probe must be cleaned and disinfected strictly in the order described below:
1. Point the probe and transducer downwards.
If any damage is observed, stop using the probe and contact Echosens or its local
representative: https://support.echosens.com.
Precautions
Do not submerge or soak the probe.
Apply the cleaning product and disinfectant solution to the soft cloth, not directly on the
surface to be cleaned.
The probe must be cleaned and disinfected between patients. Prior cleaning is necessary in
order to ensure effective disinfection.
Do not use surgical brushes or abrasive products to clean the probe, as they may damage it.
Even the use of soft brushes could damage the probe.
Take care not to introduce any cleaning product or disinfectant solution into the probe
connector.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
65
en-GB FIBROSCAN USER MANUAL 630
Users accept sole and full responsibility should they use any disinfectant product
not featured in list of products recommended by Echosens.
If you are unsure, contact Echosens or its local representative:
https://support.echosens.com.
9.1.5. Cleaning and disinfectant products that are prohibited for use
on the FibroScan probe
Any product not specifically intended for cleaning or disinfecting medical devices must not
be used.
The products listed below must not be used on the device, probes or accessories:
Sodium hydroxide and/or sodium hypochlorite disinfectant products
Hydrogen peroxide disinfectant products
Household products, particularly scouring powders or alkaline detergents (pH > 9)
containing bleach, caustic soda, potash or ammonia
Pure or dilute acid solutions of whatever type, including household vinegar
Hydrocarbon solvents: alkanes, alkenes, benzene, toluene, xylene, gasoline
Oxygenated solvents: ethanol, methanol (methylated spirits), acetone, MIBK, acetic
acid, butyl acetate, ethyl acetate (nail varnish remover), ether, glycol ethers, DMF,
DMSO, HMPT
Halogenated solvents: perchloroethylene, trichlorethylene, dichloromethane, chloroform,
tetrachloromethane
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
66
FIBROSCAN USER MANUAL 630 en-GB
Take care to prevent any solution from entering the probe connector. Protect the connector
with a plastic bag to prevent it from coming into contact with liquids.
Do not use alcohol or alcohol-based solutions. After wiping the germicide off the probe, rinse
it thoroughly with sterile distilled water to remove any possible residues. Do not use isopropyl
alcohol to wipe the strain relief, the joint between the probe cable and housing, or the cable.
Isopropyl alcohol can damage these components of the transducer.
The probe must therefore be calibrated periodically. Beyond the period indicated
on the calibration certificate, the manufacturer no longer guarantees the
performance characteristics of the probe.
The B-Mode ultrasound localisation probe does not require regular calibration.
When an examination is opened, a window displays the expiry of the calibration of your probe.
When this is displayed, contact Echosens or its local representative to arrange calibration:
https://support.echosens.com
During the examination, the message "Calibrate the probe" is displayed in the message zone.
At the end of an examination, the message "Probe out of calibration" is displayed on the
printed examination report.
9.3. TROUBLESHOOTING
Events Solutions
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
67
en-GB FIBROSCAN USER MANUAL 630
Events Solutions
The start-up/standby button is Check that the instrument is connected to a correctly powered
inoperative. When pressed, AC supply socket (test another electrical device on this same
the instrument will not turn on. socket) and that the main switch is in the I position.
Have the main switch's fuses checked by the maintenance
department.
The start-up/standby button is Put device into standby and then reboot.
on, but the software is not
booting.
In the event of a failure or malfunction, please contact Echosens or its local representative:
https://support.echosens.com.
If patient leakage current tests are carried out, take the utmost care not to immerse any
part of the probe other than the transducer.
After drying, return the probe to its resting position with the transducer facing downwards.
If any damage is observed, stop using the probe and contact Echosens or its local
representative: https://support.echosens.com.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
68
FIBROSCAN USER MANUAL 630 en-GB
The available passwords and identifiers are in ascending order of the features to which they
give access:
Operator Level
User Name: Operator
Password: U S E R 1
Password: F S C O N F 1
In general, click [Apply] to confirm the input and save the new data.
For an Operator
Language
Select the interface language and examination report from the list.
Report URL
Gives the address of the examination report accessible by QR Code, according to
geographical area.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
69
en-GB FIBROSCAN USER MANUAL 630
Logo
Press [Modify]. If an image file named logo_institution.png is present in the root of the
USB storage device connected to the machine, the logo is automatically modified. The logo
appears on the examination report.
For an Operator
Auto-print count
Enter a digit corresponding to the number of reports to automatically print at the end of each
examination.
Anonymised auto-print
Activates anonymisation when printing off an examination report.
Printing format
A4 / Letter Printing format of the examination report. The Letter format matches
the American US Letter paper format.
Printer available
Enables the manual print button displayed during examination review.
Printers
[Add printer] Lets you add a printer. You can add by Port, by IP address or
by selecting a shared printer.
[Delete printer] Lets you delete a printer. Follow the on-screen instructions.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
70
FIBROSCAN USER MANUAL 630 en-GB
For an Operator
Import
Imports examination files in FIBX/.FIBX2 format from a USB storage device.
The files are stored in the root directory of the storage device.
Examination files created with a software version prior to 2.0 cannot be imported
into version 4.0.
Export
Launches manual and automatic exports of examination files (FIBX2 format) and examination
reports (PDF format) to a USB storage device upon completion of each examination. The files
are stored in the root directory of the storage device and can be anonymised if necessary.
For an Operator
By selecting their own account, the user can change their password by clicking the
button.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
71
en-GB FIBROSCAN USER MANUAL 630
user]
[Export] Exports the list of user accounts on the device to a USB storage device.
Network
This tab is used to specify the network parameters of the device. This includes: The Name of
the device stated on network (host name), IP address, sub-network mask, gateway, primary
and secondary DNS. A host name, IP address and sub-network mask must be entered.
The IP address of the FibroScan can be configured statically or dynamically via the DHCP
protocol. To configure the IP address manually, click [Manual] and complete the fields.
Directory
Directory sharing consists of making the contents of a directory available via the network. This
tab lets you define a shared directory on the network. To do so, simply specify the absolute
access path of the directory (\\computer\name_of_share), as well as the associated login and
password.
Once directory sharing has been correctly configured, according to the option chosen
(Examination file and/or Examination result), examination results are automatically exported
to the shared directory at the end of each examination. Manual export from Archives is also
The [Check] button lets you test the connection with the shared directory.
DICOM
This tab is used to specify the DICOM connections of the device.
A key is required to enable the DICOM connectivity (to obtain this activation key,
contact ECHOSENS or its local representative).
The hospital IT department will provide user support for the configuration of these
parameters. If necessary, refer to the FibroScan Device DICOM Conformance Statement.
Worklist Specify the server through which the device carries out patient searches.
To filter the list of available patients on the DICOM server, you can query
by element (DICOM tag 0008,0060). It is filtered by AE Title by default.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
72
FIBROSCAN USER MANUAL 630 en-GB
Storage Specify the server through which the device exports the examination
reports as a JPG and/or a PDF document.
The parameters Name, IP address, Port and AE Title must be entered for each
of the DICOM servers defined. These parameters are generally available from the
network administrator.
The AE Title is a unique identifier of the device, and of each of the servers defined.
Please note that it is case-sensitive.
The [Check] button lets you test the connection with the selected server.
FibroView
This tab is used to activate the connection with FibroView.
The network address of the FibroView server must be entered, starting with http://. The
[Check] button lets you test the connection with FibroView.
Automatic addition of patients to the Worklist from FibroView is not compatible with
automatic addition from a DICOM server. Enabling one disables the other.
For an Operator
Clinical Research
Lets you view the studies created for the Clinical Research Mode.
Indications
Lets you fill in, edit, and reset the list of indications.
A key is required to enable the CAP (to obtain this activation key, contact
ECHOSENS or its local representative).
Options
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
73
en-GB FIBROSCAN USER MANUAL 630
Calibration
Lets you enter the name and details of the person at Echosens, or the local representative,
to contact when a probe is due for calibration.
Options
Clinical Research
Lets you enable the Clinical Research Mode and then create, edit, delete, and conceal study
codes.
For an Operator
DICOM log Exports the log files for DICOM connectivity if a DICOM connection
is active.
From all user account types, export log files by clicking on the button [Export log files], having
first connected a USB storage device to the device. Log files are shown as compressed files
copied to the root directory.
The USB storage device may not be recognised immediately after insertion. In that
case, press [Export log files] again if an error message is displayed.
Information
This tab provides information on the device: Serial numbers (device and connected probes),
software version number, firmware versions (CPU, FPGA, eSW, Examination Engine),
remaining disk space and total disk space, device use time.
Logs
The log file tracks system activity and gives the operator a history of the events that occurred
during use of the FibroScan software. This tab lets you view and export the log files.
General
Delay before Period of idle time after which the application logs out. This
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
74
FIBROSCAN USER MANUAL 630 en-GB
Barcode
Service
This tab lets you access the options required for device maintenance and updates.
[Clear Clears the database. Note that the examination files remain
Database] physically present on the hard drive.
General
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
75
en-GB FIBROSCAN USER MANUAL 630
Anti-malware
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
76
FIBROSCAN USER MANUAL 630 en-GB
USB connector
Connection 5 V
Manufacturer
Year of manufacture
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
77
en-GB FIBROSCAN USER MANUAL 630
Medical Device
Alternating current
Start-up/standby button.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
78
FIBROSCAN USER MANUAL 630 en-GB
Manufacturing date
11.2. NOTE
The serial number and Data Matrix marked on the device identify the FibroScan uniquely.
The Serial Number and Data Matrix shown on the B-Mode ultrasound localisation probe are
the unique identifiers of this probe.
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
79
en-GB FIBROSCAN USER MANUAL 630
IP Code IP20: The instrument without probe is not protected against liquids.
Software library The software uses the Qt library licensed under LGPL.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
80
FIBROSCAN USER MANUAL 630 en-GB
Stiffness E* (kPa)
Zone 1 1 − 37 2 − 38 3 − 22 2 − 22 2
3.0 2 − 37 1 − 33 0 − 35 2 − 29 3
Zone 2 1 − 31 2 − 32 2 − 32 1 − 32 1
10.0 2 − 30 2 − 32 2 − 31 1 − 30 1
Zone 3 1 − 30 1 − 28 1 − 29 1 − 29 1
16.9 2 − 29 2 − 27 3 − 26 1 − 29 1
Zone 4 1 − 19 1 − 12 1 − 15 3 − 15 3
25.0 2 − 17 1 − 10 2 − 20 1 − 21 3
Stiffness E* (kPa)
M+
Zone 1 1 − 41 1
9.6 2 − 40 1
Zone 2 1 − 43 1
14.6 2 − 44 0
Zone 3 1 − 39 1
21.7 2 − 38 0
Zone 4 1 − 31 1
50.9 2 − 31 1
CAP (dB/ Bias** Preci- Bias** Preci- Bias** Preci- Bias** Preci-
m) (in %) sion** (in %) sion** (in %) sion** (in %) sion**
(in %) (in %) (in %) (in %)
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
81
en-GB FIBROSCAN USER MANUAL 630
CAP (dB/ Bias** Preci- Bias** Preci- Bias** Preci- Bias** Preci-
m) (in %) sion** (in %) sion** (in %) sion** (in %) sion**
(in %) (in %) (in %) (in %)
Only second generation CAP (CAPc) is available with the S probe, and only in
certain geographical locations.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
82
FIBROSCAN USER MANUAL 630 en-GB
Model ES-C5-2R60S-3
IP Code IPX7
Radius of Curvature 60 mm
Laser device class Class 1: Safe-use laser devices, including when looking directly at the beam for
(IEC 60825-1 Ed. 3) prolonged periods, even when exposure takes place during use of telescopic optical
devices. Looking at laser beams produced by Class 1 laser devices that emit visible
radiant energy may cause visual dazzling, particularly in low lighting conditions.
IP Code IP68
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
83
en-GB FIBROSCAN USER MANUAL 630
12.6. CONSUMABLES
Not applicable.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
84
FIBROSCAN USER MANUAL 630 en-GB
The FibroScan 630 is designed for use in the electromagnetic environment defined in the
following chapter. The customer or the user of the FibroScan 630 must ensure that it is used
in that type of environment.
The use of cables and/or accessories not specified in the User Manual may damage the
device's electromagnetic performance.
13.1. CLASSIFICATION
Group 1 The FibroScan 630 uses RF energy for its internal functions only. Consequently,
its RF emissions are very low and unlikely to cause any interference with nearby
electronic equipment.
Class A The FibroScan 630 may be used on all premises other than domestic premises
and those directly connected to the public low voltage energy grid used to supply
domestic buildings.
NOTE: The emissions characteristics of this device permit its use in industrial zones and
hospital environments (Class A as defined in CISPR 11). When used in residential
environments (for which Class B as defined in CISPR 11 is normally required), this device
may interfere with radio frequency communications services. The user may need to take
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
85
en-GB FIBROSCAN USER MANUAL 630
13.2. IMMUNITY
The FibroScan device 630 is intended for use in professional healthcare environments.
Voltage drops, short interruptions and 0 % UT 1, over 0.5 cycles for 0°, 45°, 90°, 135°, 180°,
supply inlet voltage variation
IEC 61000-4-11 225°, 270° and 315° phase shifts.
0 % UT, over 1 cycle
70 % UT, over 25 cycles at 50 Hz and 30 cycles at
60 Hz, for a 0° phase shift
NOTE: In case of any disturbance in the electromagnetic environment of the FibroScan 630, a
message is displayed (see the Message Area chapter) and no measurements can be carried
out.
13.3. EMISSIONS
The FibroScan device 630 is intended for use in professional healthcare environments.
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
86
FIBROSCAN USER MANUAL 630 en-GB
73 dB μV quasi-peak
from 0.5 MHz to 5 MHz
73 dB μV quasi-peak
from 5 MHz to 30 MHz
47 μV/m quasi-peak
from 230 MHz to 1 GHz
Test Compliance
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
87
en-GB FIBROSCAN USER MANUAL 630
14. INDEX
11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED E303M007.10
88
FIBROSCAN USER MANUAL 630 en-GB
E303M007.10 11/2020 - ECHOSENS TM AND FIBROSCAN® ARE REGISTERED TRADEMARKS © COPYRIGHT ECHOSENS ALL RIGHTS RESERVED
89
blank
Echosens
6 rue Ferrus
75014 PARIS
France
Tel: +33 (0)1 44 82 78 50
Fax: +33 (0)1 44 82 78 60
Website: www.echosens.com
Email: info@echosens.com