Saep 36
Saep 36
                                                       Contents
                                                  1     Scope ................................................................ 2
                                                  2     Conflicts and Deviations .................................... 2
                                                  3     Applicable Documents ....................................... 2
                                                  4     Definitions and Acronyms .................................. 3
                                                  5     General .............................................................. 5
                                                  6     Responsibilities.................................................. 7
                                                  Revision Summary................................................. 10
1 Scope
                   This procedure establishes the instructions and responsibilities for the calibration of the
                   following Proving Systems used in Saudi Aramco Royalty/Custody Metering Systems:
                   (1) Displacement Prover
                   (2) Master Meter
                   (3) Atmospheric Portable Tank Prover
                   2.1      Any conflicts between this procedure and other applicable Saudi Aramco
                            Engineering Standards (SAES’s), Materials System Specifications (SAMSS’s),
                            Saudi Aramco Standard Drawings (SASD’s), or industry standards, codes, and
                            forms shall be resolved in writing by the Manager, Process & Control Systems
                            Department of Saudi Aramco, Dhahran.
                   2.2      Direct all requests to deviate from this procedure in writing in accordance with
                            SAEP-302 to the Manager, Process & Control Systems Department of Saudi
                            Aramco, and Dhahran.
                   2.3      Direct all requests for interpretation of this procedure in writing to the chairman,
                            Custody Measurement Standards Committee for resolution. The Chairman,
                            Custody Measurement Standards Committee shall be solely responsible for
                            determining whether a proposed request meets the requirements of this
                            procedure.
3 Applicable Documents
                   The procedures covered by this document shall comply with the latest edition of the
                   references listed below, unless otherwise noted:
4.1 Definitions
Master Meter: A meter used as a reference for the proving of another meter.
                            Proving System: A device used to generate a meter factor. Meters are proved
                            by passing a known volume of fluid through them at a known rate and
                            comparing this volume with a reference volume.
                   4.2      Abbreviations
                            API            American Petroleum Institute
                            BPV            Base Prover Volume
                            CMU            Custody Measurement Unit of Process & Control Systems
                                           Department
                            MEIM           Ministry of Energy, Industry, and Mineral Resources
                            MPMS           Manual of Petroleum Measurement Standards
                            MM             Master Meter
                                                     Saudi Aramco: Company General Use
5 General
                            Depending upon the facility and the application, either the U.S. Customary
                            (USC) or metric (SI) System of units shall be used.
                            5.3.5     The calibration of the displacement Prover using Master Meter method
                                      shall be performed in accordance with API MPMS 4.9.3 and API MPMS
                                      12.2.5.
5.4.6 Exceeding the calibration frequency as per Section 5.5 of this document.
                            5.5.1     All new Proving Systems must undergo a calibration at field site prior to
                                      being put in service. The initial calibration for new Proving System shall
                                      be by Waterdraw.
                            5.5.2     The maximum period shall not exceed the validity period depending on
                                      the calibration method used as indicated in the table 1 below.
                            5.5.3     For all types of Proving System, the deviation between the previous BPV
                                      to the new BPV shall not exceed 0.05% and subsequent calibration
                                      frequency as per API MPMS 4.9.1, Appendix B.
                            5.6.1     The calibration procedure shall at the minimum meet the requirement of
                                      API MPMS 4.9.1, Appendix A.
6 Responsibilities
                            6.1.2     Initiate request to the TPIA for calibrating the new Proving System at the
                                      site in accordance with SAEP-21 or SAEP-50, as applicable.
                            6.1.4     Provide Proving System data (type, capacity, material, pipe diameter,
                                      wall thickness, etc.) to TPIA.
                            6.1.5     Inform the Proponent to witness the calibration process for new Proving
                                      System.
                            6.1.6     Submit the new calibration certificate with Proving System Base Volume
                                      Calculation to Proponent for review and to CMU, in case of Royalty
                                      Measurement to obtain MEIM approval.
                            6.1.7     Provide paper and electronic copy of the final calibration certificate to
                                      Proponent & CMU in case of Royalty Measurement.
                            6.1.8     Upon CMU notification, inform the proponent of MEIM approval of the
                                      Base Proving System Volume for Proving System used in Royalty
                                      Measurement.
                            6.2.1     Initiate requests to TPIA for Proving service. This request must be
                                      initiated as early as possible to allow sufficient time for preparation
                                      before the calibration due time.
                            6.2.4     Provide Proving System data (material, meter capacity, pipe diameter,
                                      wall thickness, etc.) to TPIA.
                            6.2.5     Ensure all equipment used for calibration process have been currently
                                      certified to traceable national or internationally recognized metrology
                                      institution
                            6.2.6     The proponent shall ensure that Saudi Aramco operation safety be
                                      adhered to through the calibration process.
                            6.2.7     Ensure that all the necessary preparation of the site and minimum but not
                                      limited Proving System verification as follows:
                                          Prover cleaning.
                                          4-way valve is in good working condition.
                                          The displacer shall be removed from the Proving System and
                                           examined before Proving.
                                          Piston Prover displacer shall be subjected to leak test as prescribed
                                           by API MPMS 4.9.2.
                                          The seals shall be inspected and replaced if there is any sign of
                                           mechanical damage or of softening by chemical action.
                                          Spheres shall also be inspected for roundness and proper inflation as
                                           per API MPMS 4.8.
                            6.2.8     Witness the field calibration process as per Section 5.6 of this document
                                      and ensure that the TPIA is performing the Proving System calibration in
                                      accordance API MPMS 4.8, API MPMS 4.9.1, API MPMS 4.9.2 and
                                      API MPMS 4.9.3 when applicable and sign on the raw calibration data.
                            6.2.9     Consult with CMU if a technical inquiry about the calibration and/or
                                      documentation process cannot be resolved in-house,
                            6.2.10 Ensure that the TPIA is performing the Proving System calibration
                                   calculation in accordance API MPMS 12.2.4 and API MPMS 12.2.5
                                   when applicable.
                            6.2.11 Review the draft calibration certificate. The Prover calibraiton certificate
                                   shall contain the details as per Attachment-II. If the Proving System
                                   results is not within the acceptance criterion, the operating organization
                                   and Third Party Inspection Agencies (TPIA) shall inspect the Proving
                                   System, rectify the problem, and recalibrate the system.
                            6.2.13 Upon CMU notification, inform the proponent of MEIM approval of the
                                   Base Proving System Volume for Proving System used in Royalty
                                   Measurement.
                            6.2.14 Update Flow Computer with the new approved Proving System Base
                                   Volume.
                                                     Saudi Aramco: Company General Use
6.2.15 Inform SAPMT of the completion of the new Proving System calibration.
                   6.3      Custody Measurement Unit (CMU), Process & Control Systems Department
                            (P&CSD)
6.3.1 If requested, review the qualifications of the TPIA per Attachment III.
                            6.3.2     For Royalty Proving System, inform the Ministry of Energy, Industry,
                                      and Mines (MEIM) of the calibration date.
                            6.3.5     Upon request from Proponent, review the Proving System calibration
                                      draft calculation & certificate and make sure they comply with all the
                                      requirements as per Section 5.6 of this document.
                            6.3.7     For Proving System used in Royalty Measurement, inform the Proponent
                                      or SAPMT of MEIM approval of the Base Proving System Volume.
          Revision Summary
          2 January 2012         New Saudi Aramco Engineering Procedure.
          12 December 2013       Editorial revision to change the primary contact.
          11 January 2017        Major revision using revised API standards.
          5 May 2019             Editorial revision as part of content confirmation assessment
                   The following matrix summarizes the general sequence of activities and corresponding
                   responsible organizations for calibrating Proving System. Detailed requirements for
                   each organization are specified in Section 6 of this document.
                              Review the qualifications of the Third Party Inspection Agencies and make
                    1                                                                                             CMU
                              sure they meet all of the requirements as per Attachment III.
                                                                                                                  SAPMT /
                    3         Request a list of Approved TPIA from CMU.
                                                                                                                  Proponent
4 Upon request from the Proponent or SAPMT, submit list of Approved TPIA. CMU
                              Initiate request to the approved TPIA for calibrating the new Proving System
                    5                                                                                             SAPMT(1)
                              at the site in accordance with SAEP-21 or SAEP-50, as applicable.
                              Initiate requests to TPIA for Proving service. This request must be initiated as
                    6         early as possible to allow sufficient time for preparation before the calibration   Proponent
                              due time.
7 Select the method of calibration according to Section 5.5 of this document. Proponent
                               For Royalty Proving System, inform the Ministry Energy, Industry, and Mines
                    10                                                                                            CMU(2)
                               (MEIM) of the calibration date.
                              Provide Proving System data (material, pipe diameter, wall thickness, etc.)         SAPMT /
                    11
                              and station flow rate to TPIA.                                                      Proponent
                              Inform the Proponent to witness the calibration process for new Proving
                    12                                                                                            SAPMT
                              System.
                              Ensure all equipment used for calibration process have been currently certified
                    13                                                                                            Proponent
                              to traceable national or internationally recognized metrology institution.
                              The proponent shall ensure that Saudi Aramco operation safety be adhered to
                    14                                                                                            Proponent
                              through the calibration process.
                               Ensure that all the necessary preparation of the site and minimum but not
                               limited Proving System verification as follows:
 Prover cleaning.
                                  The displacer shall be removed from the Proving System and examined
                                   before Proving.
                    15                                                                                           Proponent
                                  Piston Prover displacer shall be subjected to a leak test as prescribed by
                                   API MPMS 4.9.2.
                              Witness the field calibration process as per Section 5.6 of this document and
                                                                                                                 Proponent /
                    16        ensure that the TPIA is performing the Proving System calibration as per
                                                                                                                 MEIM(2)
                              Section 5.6 of this document and sign on the raw data of the calibration results
                              Ensure that the TPIA is performing the Proving System calibration calculation
                    17                                                                                           Proponent
                              in accordance API MPMS 12.2.4 and API MPMS 12.2.5 when applicable
                              Upon request from Proponent, review the Proving System calibration draft
                    20        calculation & certificate and make sure they comply with all the requirements      CMU
                              as per Section 5.7 of this document.
                              Submit the new calibration certificate with Proving System Base Volume
                    21        Calculation to Proponent for review and to CMU, in case of Royalty                 SAPMT(2)
                              Measurement, for review and to obtain MEIM approval.
                    25        In case of calibrating Royalty Proving Systems, obtain the approval of the         CMU(2)
                                                      Saudi Aramco: Company General Use
                               Upon CMU notification, inform the proponent of MEIM approval of the Base
                    26                                                                                              SAPMT (2)
                               Proving System Volume for Proving System used in Royalty Measurement.
                               Inform SAPMT of the completion of the Proving System calibration for new
                    27                                                                                               Proponent
                               Prover.
                               For Royalty Proving System, inform the Proponent or SAPMT of MEIM approval
                    28                                                                                               CMU(2)
                               of the Base Proving System Volume.
29 Update Flow Computer with the new approved Proving System Base Volume. Proponent
Notes:
                   1)    SAPMT initiates calibration request for new Proving System. The Proponent initiates calibration Request for
                         Proving System in-service as per the frequency requirements set forth in Section 5 above. For BI-1900, the
                         Proponent will assume the responsibilities of SAPMT.
2) MEIM involvement is only required for calibration of Royalty Measurement Proving System.
          The final Proving System certificate shall contain the following minimum details as per
          API MPMS 4.8. Units of calibration shall be as per Section 5.2.
1) Prover Identification
                 d)     The name and address of the calibration authority or company which carried out the
                        calibration.
2) Product Information
                 The product name and density of the liquid flowing in the Prover when in service shall be
                 used in the computation of the Prover Base Volume.
3) Operational Details
                 a)     The standard temperature (60ºF for U.S. Customary, or 15ºC for metric tables) for
                        which the Base Volume has been calculated.
d) Method of Calibration
g) The Third Party Inspection Agency shall sign each page of the certificate.
5) Prover Dimensions
e) Prover previous Base Volume to new Prover Base volume to meet 0.05% deviation
                 f)     Type of sphere, sphere size, color & type of fluid used to fill sphere in a case of
                        conventional pipe Prover
The Third Party Inspection Agency shall meet all of the following requirements:
          a) The agency has provided similar services for a minimum of 3 years. It shall submit a list of
             companies for which it has provided similar calibration services over the preceding 3 years.
          b) The agency's personnel performing the field measurements and calculations must have a
             minimum of one year experience with the application of the Prover calibration standards.
             Experience shall be documented by submittal of resume and verifiable work histories.
          c) The agency shall have written procedures which meet the requirements of API MPMS 4.2:
             It shall submit the written procedures for review.
          d) The agency shall have equipment as specified in API MPMS 4.2 & API MPMS 4.8. It shall
             submit a list of the equipment it intends to use along with corresponding valid calibration
             certificates.
          f) The agency shall demonstrate that it is capable of performing the Prover calibration
             calculations as per API MPMS 12.2, API MPMS 13.1 & API MPMS 13.2. It shall provide
             a minimum of two samples of previous calibration results, including field data and all
             calculations.
          g) The agency shall provide financial and insurance documentations as deemed necessary by
             the Contracting Unit to ensure he has adequate liability coverage for damages done to Saudi
             Aramco facilities.