eCL8000 Software upgrade
This software upgrade introduction is applicable to all I05A series instruments.
Tools required for upgrading:
computer, USB flash drive, upgrade adapter (only for board
upgrade).
Upgrade step:
1 . Preparations before upgrading
Decompress the latest version of the software and put it in a USB
flash drive. Insert the USB flash drive into the USB port of the
instrument.
Note: The instrument does not have anti-virus software. Any equipment
with virus is strictly prohibited to use on the instrument.
1.1 Parameter record
Before we update the software, have to record thoses parameters (include
the parameter of 3D Arm, of 2D Arm, of the temperature compensation),
so that the reference can be recalibrated after the upgrade.
1.1.1 In the interface [Aplication], then Click [Advanced], and login.
1.1.2 After login Advanced, click [HD Adjust]
1.1.3 In the interface [HD Adjust], we have to record the information of
[MPU1] , [MPU2] and the temperature compensation data in [Other]
interface
Firstly click [MPU1], then click [Check],
and take a photo of these parameters
Then click [MP2] , in the interface MPU2, click [Check]
Take a photo to record these parameters.
Then Click [Others],
And there are 4 choose of temperature. And should click each one, and
record the temperature compensation.
1.2 Project file export
Before exporting the project file, quit the software and insert the U disk.
① About quit the software, in the upper left corner of the software and
click
Then click [Exit].
② At the back of the machine, there are several connectors of USB, and
insert the U disk .
③ Find the operating software shortcut on the desktop, right-click
after selecting it, click "Properties", "Open File Location", and find the
[items] folder in this folder.
③ Copy and save the 【items】folder into U disk or other folder.
Attention: You must exit the software when importing / exporting the
project file.
2. COP Upgrade
The installation of the lower computer program mainly update the COP
board
2.1 Open the mcuisp (the software is in the folder 【Lifotronic】on
desktop) and confirm that the serial port number is COM2, the baud rate
is 115200, and choose "Check", "Execute after programming" and
"Reinstall software before programming" check boxes, as shown below.
At the desktop, there
is a folder【Lifotronic】
Have to check:
the serial port number: COM2,
the baud rate: 115200;
choose "Verify",
"Run after ISP complete"
"Auto Reload before Program"
4.2 COP board bootloader download: Under the mcuisp download tool
software interface,at this time, open the left cover (near these switches)
press and hold the ISP button on the COP board → press and hold the
REST button→ release the REST button (the ISP button is still pressed)
→ release the ISP button, and the board enters the software upgrade
mode.
Open the left cover which
near these switches
COP board
REST
ISP
Press and hold the ISP button on the COP board →
press and hold the REST button
→ release the REST button (the ISP button is still
pressed)→ release the ISP button
Choose【COP XXXX.hex】(The document Update which was sent by our
engineer include this program ), click "Start ISP" to start the program
download, waiting for completion means that the program download is
complete.
When the software shows the below information, means the COP Update
is successful, And on the COP Board, click REST again.
3 MPU Upgrade
When upgrade the MPU, must exit the software.
3.1. Open the folder【Lifotronic】on desktop,Find【IAP downloader】
document, open the application【IAP downloader】
At the desktop, there
is a folder【Lifotronic】
At folder【Lifotronic】,
open the folder 【IAP】
Find the software【IAP
downloader】,and open it
At the top of the application, click [Setting] > [COP Setting] ;
At the interface Com-Setting, click [Traverse], and this application shows
Serial no. and Baud rate. Have to take a photo at this interface, Set the
same serial port number and Baud rate.
Serial no. and
Baudrate for
Upgrade.
Take a photo to record
these parameters
Set the same serial port
number as what shows
[ComSetting]
Set the same Baud
rate as what shows
at [ComSetting]
3.2. MPU1 Program download:Click button, choose【MPU1
XXXXXXX.Pbin 】 (The document Update which was sent by our
engineer include these) program,and then click “download”.
2
Have to check :
the serial port number: COM2,
Baud rate: 38400;
If the application shows [The current version with the existing [MPU1]
whether to download again?], Please click [YES].
The program enters the downloading state. When the download progress
shows 100%, the downloading of the program is completed.
3.3. MPU2 Program download:Click button, choose【MPU2
XXXXXXX. Pbin】 program, and then click “download”. The program
enters the downloading state. When the download progress shows 100%,
the downloading of the program is completed.
3.4. MPU3 Program download:Click button, choose
【 MPU3XXXXX.Pbin】 program, and then click “download”. The
program enters the downloading state. When the download progress
shows 100%, the downloading of the program is completed.
4 Software Upgrade
4.1 Uninstall old PC software
Click "Start" in the lower left corner of the desktop, search [Uninstall a
program]; then open "Uninstall a program" in "Control Panel", and
uninstall the computer software.
4.2 New PC software installation
① Open the folder, double click【System software.exe】(The document
Update which was sent by our engineer include this program ), start the
installation program:
② Do not change any parameter, Select “Run eCL8000 V00.XXX ”, just
click “Finish” to finish the installation, and run the software
③ After the software runs, it will enter the startup interface, and the
installation is complete.
5. After Upgrade
5.1 Item File Update (Only for version with Item File Update)
Copy all of the item files (not the whole folder) found in the [items]
folder in the Software Upgrade zip file,
Copy all the files which
included in the folder Update
then Find the eCL8000 icon on Desktop, right click it with your mouse,
then choose [Properties], then choose [Open File Location] in the window
pop up. In the folder [eCL8000], open the [items] folder, and paste and
replace the item files from the Upgrade zip file.
5.2 Delete ItemRunConfig.dat
Find the eCL8000 icon on Desktop, right click it with your mouse, then
choose [Properties], then choose [Open File Location] in the window pop
up. At the folder [eCL8000] > [data]; delete the document
[ItemRunConfig.dat]
5.3 Then delete those files changed
Open the [items] folder in the Software Upgrade zip file, take a photo to
record these updated item file.
Find the eCL8000 icon on Desktop, right click it with your mouse, then
choose [Properties], then choose [Open File Location] in the window pop
up. Open the folder [eCL8000] > [itemSetConfig]
In the [itemSetConfig] interface, delete all the .dat files corresponding to
the items listed in the photo taken.
5.4 Software version confirmation
Open the software and check whether the software versions are the same
as those in the USB flash drive in Application-About.
5.4 Parameter confirmation
After the software upgrade, re-confirm the position parameters and
temperature compensation of the instrument hardware in the system
statement. And they are in the interface [HD Adjust], check these
parameters what have recorded at the step 1.
5.5 Confirmation of other states of the instrument
Check whether all parts of the instrument are normal in the instrument
status.
5.6 Check the sample testing process
Perform at least 5 sample tests (any item), to confirm the status of the
analyzer after the upgrade.