The document discusses automated external defibrillators (AEDs), which are portable devices that treat sudden cardiac arrest by detecting abnormal heart rhythms and delivering an electric shock to restore normal rhythm. AEDs analyze the heart's rhythm and instruct laypeople, with minimal training, on applying treatment. For cardiac arrest victims, immediate CPR and defibrillation within the first few minutes provides the best chance of survival. The document outlines AED components, how they work to detect shockable rhythms and safely deliver treatment, and design considerations around ensuring operator and patient safety.

Uploaded by

Alina Barkova
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
225 views17 pages

AEDs: Essential Life-Saving Devices

The document discusses automated external defibrillators (AEDs), which are portable devices that treat sudden cardiac arrest by detecting abnormal heart rhythms and delivering an electric shock to restore normal rhythm. AEDs analyze the heart's rhythm and instruct laypeople, with minimal training, on applying treatment. For cardiac arrest victims, immediate CPR and defibrillation within the first few minutes provides the best chance of survival. The document outlines AED components, how they work to detect shockable rhythms and safely deliver treatment, and design considerations around ensuring operator and patient safety.

Uploaded by

Alina Barkova
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 17

AED

1. What is AED and why this necessary for peoples in our lives?
Automated external defibrillators (AEDs) are portable, life-saving devices
designed to treat people experiencing sudden cardiac arrest, a medical condition
in which the heart stops beating suddenly and unexpectedly.
The combination of CPR and early defibrillation is effective in saving lives when
used in the first few minutes following collapse from sudden cardiac arrest.

So AEDs are portable, life-saving devices designed to treat people experiencing sudden
cardiac arrest, a medical condition in which the heart suddenly and unexpectedly stops
beating. The AED system includes accessories, such as a battery and pad electrodes, that
are necessary for the AED to detect and interpret an electrocardiogram and deliver an
electrical shock. There are two main types of AEDs: public access and professional use.
 Public access AEDs can be found in airports, community centers, schools,
government buildings, hospitals, and other public locations. They are intended to
be used by laypeople who have received minimal training.
 Professional use AEDs are used by first responders, such as emergency medical
technicians (EMTs) and paramedics, who receive additional AED training.

Then why this platform needs:


Out-of-hospital cardiac arrest (OHCA) is a significant public health issue. In Australia
more than 30,000 OHCAs occur every year, with fewer than 10% surviving to leave
hospital (Ambulance Victoria, 2016). However, with immediate resuscitation more lives
can be saved.
In cardiac arrest the heart is no longer pumping, blood flow stops, the casualty loses
consciousness and will not be breathing normally. This person will die, unless within a
short period of time blood flow is restored. Immediate cardiopulmonary resuscitation
(CPR) will restore some blood flow.
For a short period of time, the non-pumping heart may have an abormal rhythm
(ventricular fibrillation (VF) or ventricular tachycardia (VT) which may be “shocked” by
defibrillation (a type of electric shock) back to a normal pumping rhythm.
That’s why AED is desperately needed. One of the functions of an AED is to determine
whether the casualty has a “shockable” or “non-shockable” rhythm. The time to
defibrillation is a key factor that influences survival. For every minute defibrillation is
delayed, there is approximately 10% reduction in survival if the patient is in cardiac arrest
due to VF or VT, (shockable rhythms).
2. AED System Theory and Architecture
Energy-based defibrillators can deliver energy in a variety of waveforms, broadly
characterized as monophasic, biphasic or triphasic.
- Monophasic waveform
Defibrillators with this type of waveform deliver current in one polarity and were
the first to be introduced. They can be further categorized by the rate at which the
current pulse decreases to zero. If the monophasic waveform falls to zero
gradually, the term damped sinusoidal is used. If the waveform falls
instantaneously, the term truncated exponential is used (figure 1). The damped
sinusoidal monophasic waveforms have been the mainstay of external
defibrillation for over three decades
- Biphasic waveform. This type of waveform was developed later. The delivered
current flows in a positive direction for a specified time and then reverses and
flows in a negative direction for the remaining duration of the electrical discharge
(figure 2A). With biphasic waveforms there is a lower defibrillation threshold (DFT)
that allows reductions of the energy levels administrated and may cause less
myocardial damage. The use of biphasic waveforms permits a reduction in the size
and weight of AEDs.
- Triphasic waveform. There are no human studies to support the use of multiphasic
waveforms over biphasic. Investigation in animals suggests that the benefits of
biphasic waveform could be harnessed through the use of a triphasic waveform in
which the second phase has the larger strength to lower the DFT and the third
phase the lower strength, to minimize damage

Figure 1.
Figure 2.

An automated external defibrillator (AED) is a portable electronic device that


automatically diagnoses the potential life threatening cardiac arrhythmias of ventricular
fibrillation and ventricular tachycardia in a patient.
Automatic refers to the unit’s ability to autonomously analyze the patient’s condition;
to assist this, the vast majority of units have spoken prompts, while some may also have
visual displays to instruct the user. With simple audio and visual commands, AEDs are
designed to be simple to use for the layman.
Defibrillators can be external, transvenous, or implanted, depending on the type of device
used or needed. The external defibrillator could be manual or automatic by operation
method, and monophasic or biphasic waveform by energy delivery method.
Defibrillation consists of delivering a therapeutic dose of electrical energy to the
affected heart with a device called a defibrillator. This depolarizes a critical mass of the
heart muscle, terminates the dysrhythmia, and allows normal sinus rhythm to be
reestablished by the body’s natural pacemaker in the sinoatrial node of the heart. The
energy selection is decided by the AED device automatically according to the
electrocardiogram (ECG) and impedance gotten from both of the defibrillator electrodes,
then the safety processor controls the power circuit to charge the high voltage capacitor
with selected energy. After the capacitor charging is complete, the device should prompt
the user to do a shock operation, which is a high risk operation, and a double
confirmation is always needed to make sure both the operator and the patient are safe.
Before and after defibrillation, the optional multi-lead ECG monitor (3/5/10 leads) may
be used to evaluate the treatment.
The ECG in the defibrillator electrode is a simple single-lead ECG for basic ECG
measurement like R wavelet recognition, but the optional multi-lead ECG is a diagnostic
monitor level, which can detect complex issues.

3. Design consideration and Major Challenge


1) Components
The term refers to a portable and lightweight computerized device that incorporates
rhythm analysis and defibrillation systems and uses voice and/or visual prompts to guide
lay rescuers and healthcare providers to safely defibrillate victims of cardiac arrest due to
VF or pulseless VT.
Basically these devices consist of a battery, a capacitor, electrodes and an electrical
circuit designed to analyze the rhythm and send an electric shock if is needed.
- Batteries.
Essentially they are containers of chemical reactions and one of the most
important parts of the AED system. Initially lead batteries and nickel-cadmium
were used but lately non-rechargeable lithium batteries, smaller in size and with
longer duration without maintenance (up to 5 years), are rapidly replacing them.
Since extreme temperatures negatively affect the batteries, defibrillators must be
stored in controlled environments. Also it is important to dispose of the batteries
using designated containers as they contain corrosive and highly toxic substances.
- Capacitor
The electrical shock delivered to the patient is generated by high voltage circuits
from energy stored in a capacitor which can hold up to 7 kV of electricity. The
energy delivered by this system can be anywhere from 30 to 400 joules.
- Electrodes.
These are the components through which the defibrillator collects information for
rhythm analysis and delivers energy to the patient's heart. Many types of
electrodes are available including hand-held paddles, internal paddles, and self-
adhesive disposable electrodes. In general, disposable electrodes are preferred in
emergency settings because they increase the speed of shock and improve
defibrillation technique.
- Electrical circuit
AEDs are highly sophisticated, microprocessor-based devices that analyze multiple
features of the surface ECG signal including frequency, amplitude, slope and wave
morphology. It contains various filters for QRS signals, radio transmission and
other interferences, as well as for loose electrodes and poor contact. Some
devices are programmed to detect patient movement.
- Controls
The typical controls on an AED include a power button, a display screen on which
trained rescuers can check de heart rhythm and a discharge button. Defibrillators
that can be operated manually have also an energy select control and a charge
button. Certain defibrillators have special controls for internal paddles or
disposable electrodes.

2) Consider and implement functions


Safety is the first priority in AED design. Any operation must ensure the safety of both
operator and patient, so some redundant designs are necessary.
- Both the safety and the operation processor need to check each other to ensure the
right decision.
- Discharge the charged capacitor if it times out.
- Double confirmation is required for energy delivery.
- An audio prompt is helpful.
- Disable the energy delivery if the target impedance is not in range of the human
body.
Isolation is critical for matters internal and external.
- The device must ensure enough insulation between the internal high voltage and the
device surface/port.
- The device must provide an insulation mechanism between the internal high voltage
and low voltage part. As you know, the defibrillation works in high voltage mode
while the signal processor works in low voltage mode. Therefore, a path switch
based on the relay can be used.
Fast response is critical. The AED is a device for life saving, so the faster the response
of the device, the greater possibility of life saving.
- Fast boot-up for operation.
- Fast response to an external signal like the external patient monitor trigger out.
- Real-time R wavelet recognition for exact time to delivery energy.
- Real-time energy control for the shock procedure; it is IP related to a different
energy delivery waveform.
- Fast charge and energy delivery to save time.
Reliability is critical. The AED can be used in many fields: in the hospital and out of
the hospital, in high vibration conditions like ambulances and helicopters, and in
outdoor applications like in sunshine and rainy weather. So the AED may need
antivibration, waterproofing, and so on for complex conditions.
- Wide operation temperature range.
- Lower performance drift over temperature range, like bias current and noise.
- The power circuit should work well for large current surges.
Interaction is helpful to operate the AED easily.
- An audio prompt can be used to indicate how to do the next step.
- An audio recorder can be used to record the rescue procedure for evidence.
- Connectivity is necessary for the modern AED.
Wireless connectivity like ISM and Wi-Fi is helpful for the device in the hospital;
GPRS/3G is helpful for the device out of the hospital. This means cellular module.
- And have to consider use some peripherals like LAN, UART, and a memory card are
used for the electrical medical system.
4. Prototype Diagram
Below figure3 is show general AED operate order.

Figure3. Diagram
Figure4. General AED
5. Desired Design
As shown in Fig. 1, the AED is divided into two main electric boards powered by a
battery: a High-voltage board (HV-B) and a control board (C-B).
The HV-B contains the circuitry necessary to perform defibrillation:
- capacitor, used to store the energy to be released to the patient;
- charging circuit, which rapidly charges the capacitor;
- H-Bridge circuit, required to perform biphasic defibrillation;
- internal discharge circuit, used to dump the unused residual energy; and
- two selectors, used to isolate the patient from the capacitor, and to route the ECG
signal to the C-B.
The C-B represents the device’s brain. It contains a Programmable System on Chip
(Cypress PSoC), which amongst other functions- integrates the analogical front-end used
to acquire the ECG and impedance measurements from the patient.
The PSoC also contains an ARM Cortex-M3 CPU, used to analyze the signals and asses the
defibrillation needs.
According to MDD 93/42, AEDs are classified as Class IIb devices because they are
‘‘therapeutic devices intended to administer energy in a potentially hazardous way”.
The design has been done ensuring compliance with the standards of the IEC 60601
family.
Comparing the standards above with the commercially available AEDs, we can outline the
following technical specifications:
- Nominal voltage: 1700 V;
- Defibrillation energy: 200 J;
- Charging time: 12 s, defined as the maximum time from activation of the rhythm
detector to the defibrillator being ready
for discharge at maximum energy;
- Patient leakage currents (not during defibrillation): 50 uA;
- SCA recognition algorithms: Sensitivity >95%, specificity >95%.

The blue blocks on the left side constitute the power supply module. The module
comprises the battery and the voltage
regulators included in the control board (C-B), and the high-voltage board (HV-B). The
yellow blocks represent the C-B, which is composed of a PSoC and
some input-output components such as switches and LED diodes. Lastly, the green blocks
represent the HV-B. The HV-B includes a charging circuit, a
condenser, an H-Bridge circuit, two relays used as selectors, and an internal discharge
circuit composed of a power resistor. (For interpretation of the
references to colour in this figure legend, the reader is referred to the web version of this
article.)
Figure5. Diagram

1) Charging circuit
From an engineering point of view, defibrillation can be approximated as a capacitive
discharge on a pure resistive load.
Defibrillation can be schematized with an RC circuit, where the patient is represented as a
resistor, while a capacitor represents the defibrillator. For this reason, the most critical
aspect of an AED is the charging time of the capacitor, because more than one charge-
discharge cycle is usually needed to save a life.
An AED needs a charging circuit with the following characteristics:
- ability of producing a high-voltage output from a low-voltage input;
- high efficiency;
- fast charging speed;
- compact dimensions and low weight;
- low cost;
- compatibility with pure capacitive loads; and
- safety and robustness.
The final charging circuit was designed as a self-oscillating flyback converter, also
known as Ringing Choke Converter (RCC). The RCC is derived from the flyback converter.
The flyback converter works using a PWM signal to operate its switch. When this signal is
high, the switch is closed and a current flows through the primary coil of the transformer.
The flyback diode on the secondary side of the circuit prevents the current from
flowing, therefore the transformer accumulates energy. When the PWM signal is low, the
switch is open and the transformer releases the energy it had accumulated during the
‘‘on” time in the secondary coil. In a traditional flyback configuration, the PWM signal has
a fixed frequency while its duty cycle may vary according to the load. The disadvantage of
this configuration is that when the duty cycle is low (meaning a short ‘‘on” time), there
will be an excess of dead time, during which the circuit has no current flowing in it.
On the other hand, if the duty cycle is high (meaning a long ‘‘on” time), the
transformer might not have enough time to release all its energy onto the capacitor.
These two operating modes are commonly known as discontinuous conduction mode and
continuous conduction mode respectively.
The purpose of the RCC is to avoid working in both of these two conduction modes. In
fact, the RCC is obtained by adding a third coil to the flyback transformer, which will be
used to automatically generate a PWM signal. The peculiarity of this approach is that only
a few components are sufficient to bring the flyback circuit to work at the exact point of
transition between discontinuous and continuous conduction modes. In this condition, as
soon as the transformer releases all the energy, the switch is closed again avoiding dead
times and energy residuals in the transformer.
Figure 6. Flyback Converter

Figure 6.

2) Control Board
The Cypress PSoC 5LP is at heart of the C-B. The PSoC is a one-chip solution integrating
analog front-end, digital logic and user interface integrated circuits, with an ARM Cortex-
M3 CPU. Its internal circuits can be re-arranged using the Cypress software, in order to
obtain various circuit configurations.
In the AED, the PSoC blocks are configured to perform various operations. The general I/O
blocks are reported follow image.
These are necessary to communicate with the operator and the High-voltage board and
are hereinafter explained:
- The first is a redundant control for the H-Bridge circuit. The block on the left is a
software-set register, updated every time the firmware wants to control the H-
Bridge. The output is evaluated with logic blocks in order to provide a degree of
redundancy. The right block represents the connection to physical pins.
- These are the physical pins used to control the inner, and outer selectors.
- ”Charge_En” is the pin used to enable the charging circuit.
- The first block on the right represents the physical pin to be connected at the push-
button that the operator should press to release the defibrillation. Due to its
criticality, a de-bouncer is used, which directly calls an interrupt (here represented as
lightening).
- This contains the pins connected to the LED diodes. These, as the resistors, are not
inside the PSoC, but on the control board.
- This block is required to implement USB UART protocols.
- This block is used to evaluate the capacitor voltage. The label ”Vsense” represents
the pin attached to the charging circuit buffer output, presented. This pin is
connected to two comparators. When it reaches a value of 2.5 V, the first
comparator calls an interrupt to communicate to the processor that the capacitor is
ready. On the other hand, when its value falls below 256 mV, the second comparator
calls another interrupt, used to communicate to the processor that the capacitor’s
voltage is below 10% of its nominal value V0.
- Finally, the last block on the left is the DAC that controls the speaker. The speaker is
used to warn bystanders that a defibrillation is about to be released. As for (e), the
speaker and the passive elements are outside the PSoC.
Figure 7.

3) Firmware
AED was programmed with custom firmware in C programming language. The firmware
acts as a finite-state machine, in which each state is used to enable or disable specific
circuits depending on the state itself. The five different states are presented in Fig. 8, and
their operations are:
- Measurement mode is the starting state. In measurement mode, the device has not
yet diagnosed SCA in the patient. Therefore, the operations are limited to
continuously acquiring ECG and impedance signals from the patient.
- Charging mode state is reached when AED successfully diagnoses SCA for the first
time. As a result, the charging circuit is enabled, while the patient is still monitored
for SCA.
- Discharge enabled mode is entered when the patient is both suffering from SCA and
the capacitor is ready. In this mode the defibrillator is armed and ready to deliver the
shock when the operator will press the ‘‘defibrillate” button.
Figure 9. Finite-state machine diagram
- Internal discharge mode represents an emergency stop for AED. Whenever
something is not working properly and the capacitor is charged (even partially), AED
will dump the defibrillation energy to the internal discharge circuit, which is a power
resistor.
- Lead-Off mode is an idle state in which AED just waits for a patient to be connected.
Thereafter in this state, AED will only perform impedance acquisitions. When it
recognizes a patient, lead-off mode is switched to Measurement mode.

The firmware also includes five different algorithms used in combination to analyze
the ECG signals and assert whether the patient is suffering from SCA. These algorithms
are: Threshold Crossing Interval (TCI), VF filter, Threshold Crossing Sample Count
(TCSC), Phase Shift Reconstruction (PSR), and Hilbert Transform Algorithm (HTA).
Each ECG segment obtained by the acquisition chain is evaluated with all the
algorithms and the decision is majority based. If at least two of the last three ECG
segments are SCA positive, then AED diagnoses the heartbeat as pathologic.

6. References
[1] Newman, Mary. “Sudden Cardiac Arrest: A Healthcare Crisis”.
http://www.sca-aware.org/about-sca.
[2] American Heart Association. “ECC Guidelines – Part 4: The Automated External
Defibrillator”. American Heart Association Journals.
ISSN: 1524-4539. http://circ.ahajournals.org/content/102/suppl_1/I-60.full
[3] David Arney, Raoul Jetley, Paul Jones, Insup Lee, and Oleg Sokolsky. 2007. Formal
Methods Based Development of a PCA Infusion Pump Reference Model: Generic Infusion
Pump (GIP) Project.
In Proceedings of the 2007 Joint Workshop on High Confidence Medical Devices,
Software, and Systems and Medical Device Plug-and-Play Interoperability (HCMDSS-
MDPNP '07). IEEE Computer Society,
Washington, DC, USA, 23-33.
[4] Amann, A., Tratnig, R., & Unterkofler, K. (2007). Detecting ventricular fibrillation by
time-delay methods. Biomedical Engineering, IEEE Transactions on, 54(1), 174-177.
[5] Kewei, H., Jie, L., Ningjun, F., Yuebin, L., Xiaolin, H., & Luo, W. (2009, July). Modeling
analysis and simulation of high-voltage flyback DC-DC converter. InIndustrial Electronics,
2009. ISIE 2009. IEEE
International Symposium on(pp. 813-818). IEEE.
[10] Edmar C. Gurjao, A Model for Architecture Centric Development of Automated
External Defibrillators