0% found this document useful (0 votes)
153 views2 pages

Wika Model TI 30

This document provides information on the TI.30 bimetal thermometer model, including its applications, features, and specifications. The TI.30 is an industrial thermometer designed for use with fluid media that do not corrode stainless steel. It has a stainless steel casing and wetted parts, as well as a back connection for external reset. Standard models measure between -100°F to 1000°F and have accuracies within 1% of the full scale value. The thermometer also offers optional extras such as thermowells, special scales, and calibration certification.

Uploaded by

Michael Sheehan
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
153 views2 pages

Wika Model TI 30

This document provides information on the TI.30 bimetal thermometer model, including its applications, features, and specifications. The TI.30 is an industrial thermometer designed for use with fluid media that do not corrode stainless steel. It has a stainless steel casing and wetted parts, as well as a back connection for external reset. Standard models measure between -100°F to 1000°F and have accuracies within 1% of the full scale value. The thermometer also offers optional extras such as thermowells, special scales, and calibration certification.

Uploaded by

Michael Sheehan
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Mechanical Temperature Measurement

Bimetal Thermometer Model TI.30, Stainless Steel Case & Wetted Parts
Datasheet TI.30

Applications
Suitable for fluid medium which does not corrode 304 stainless steel

Special features
Industrial design Stainless steel case and wetted parts Back connection with external reset

Standard version
Application Industrial type design for fluid medium which does not corrode 304 stainless steel. Sizes 3 (76.2 mm)

Pressure Gauge TI.30

Stem Material: 304 stainless steel Diameter: (6.35 mm) Length: 2 to 72 (63.5 mm to 1,828.8 mm) Case Material: 304 stainless steel Hermetically sealed per ASME B40.3 standard Ingress protection IP 65 External reset slotted hex head on back of case Dial White aluminum, dished, with black markings Dampening Inert gel to minimize pointer oscillation Standard Scales Single: Fahrenheit or Celsius Dual: Fahrenheit (outer) and Celsius (inner) Window Gasket Neoprene Silicone (-100 oF and over 550 oF) Window Flat instrument glass Weight 7 oz. (3 dial); Add 1 oz for every 2 of stem length Movement Viscous inert gel to enhance pointer operation
Page 1 of 2
R

Type TI.30

Accuracy + 1.0% full scale value (ASME B40.3) Ranges -100 oF to 1000 oF (and equivalent Celsius) Working Range Steady: full scale value 110% of full scale value Short time: Over Range Temporary over or under range tolerance of 50% of scale up to 500 oF. (260C). For ranges above 500F, maximum over range is 800F; continous. 1000F intermittent. Connection Material: 304 stainless steel Center back mount (CBM), 1/2 NPT Measuring Element Bi-metal helix Pointer Black aluminum
Datasheet TI.30 8/2009

Optional Extras
Thermowells Silicone fill Dampened Movement Special scales and dial markings Acrylic and safety glass windows Calibration certification traceable to NIST Min/max pointer DIN standards

STANDARD RANGES Dual Scale F & C Fahrenheit Single Scale F Outer, C Inner -100/150 F -100/150 F & -70/70 C -40/120 F -40/120 F & -40/50 C 0/140 F 0/140 F & -20/60 C 0/200 F 0/200 F & -15/90 C 0/250 F & -20/120 C 0/250 F 20/240 F 20/240 F & -5/115 C 25/125 F 25/125 F & -5/50 C 50/300 F 50/300 F & 10/150 C 50/400 F 50/400 F & 10/200 C 50/550 F 50/500 F & 10/260 C 150/750 F 150/750 F & 65/400 C 200/1000 F1 200/1000 F & 100/540 C1
1

Celsius Single Scale -50/50 C -20/120 C 0/50 C1 0/100 C 0/150 C 0/200 C 0/250 C 0/300 C 0/450 C1 100/550 C1

Not recommended for continous service over 800F (425C)

Dimensions
Standard versions

Stem Length 2 (63.5 mm) 4 (101.6 mm) 6 (152.4 mm) 9 (228.6 mm) 12 (304.8 mm) 15 (381.0 mm) 18 (457.2 mm) 24 (609.6 mm)

WIKA Type 30

DIAL SIZE 3 (76.2 mm)

A 3-1/4 (82.6 mm)

B 15/16 (23.8 mm)

S (Stem Length) As Specified

Note: Thermowells for temperature instruments are recommended for all process systems where pressure, velocity, or viscous, abrasive and corrosive materials are present individually or in combination. A properly selected thermowell protects the temperature instrument from possible damage resulting from these process variables. Furthermore, a thermowell permits removal of the temperature instrument for replacement, repair or testing without effecting the process media or the system.

Ordering information State computer part number (if available) /type number/size/range/connection size and locations/options required. WIKA reserves the right to make changes without prior notice.

Page 2 of 2

WIKA Datasheet TI.30 8/2009


R

WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-WIKA-USA /770-513-8200 (in GA) Fax 770-338-5118 info@wika.com www.wika.com