Certified Weighing for Fisheries

The document summarizes information about the Marel Flowscale, an approved belt weigher for weighing fish and other materials on fishing vessels and processing plants. The Flowscale can weigh flows of up to 100 tons per hour and records weights in an alibi memory. It is calibrated and sealed by authorities to ensure accurate weighing according to legal requirements. The summary provides an overview of how the Flowscale functions, is approved, has its indicator and hardware sealed, and records weight measurements.

Uploaded by

Rodinei Ferraz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
436 views15 pages

Certified Weighing for Fisheries

The document summarizes information about the Marel Flowscale, an approved belt weigher for weighing fish and other materials on fishing vessels and processing plants. The Flowscale can weigh flows of up to 100 tons per hour and records weights in an alibi memory. It is calibrated and sealed by authorities to ensure accurate weighing according to legal requirements. The summary provides an overview of how the Flowscale functions, is approved, has its indicator and hardware sealed, and records weight measurements.

Uploaded by

Rodinei Ferraz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

MAREL FLOWSCALE

APPROVED WEIGHING FOR


SUSTAINABLE FISHERIES

1
01
Marel Flowscale
OVERVIEW
02
Approval & verification

03Flowscale Hardware
06 and Software sealing
Innova for Flowscale

05 04
Data registration & Recording of weight
protection measurements
Marel Flowscale
Design
The Marel Flowscale is a compact,
durable, high-capacity belt weigher used
to weigh a continuous flow of material
moving on a conveyor.

The scale is capable of weighing a flow


of up to 100 tons per hour.

Similar to the way water or


electricity meters function; the scale
registers the total weight of the product
flow that is moving across it and
displays the accumulated weight on an
indicator.
Flowscale Function

The Flowscale is designed and built as


W&M certified for in-weighing of pelagic or round
fish from fishing vessel.

The weight appears on the indicator display and


can be sent to Innova for data collection and/or
further data utilization.

The weight is recorded in alibi memory that


stores the weighing data.

The Marel Flowscales are used to ensure


that accurate weight registration is performed
especially when landing fish.

Furthermore to secure that it is done in


accordance to legal requirements.

4
Approval & Verification

The authorities or an accredited calibration company such


as TRESCAL or FORCE can carry out the calibration of
flowscales.
During calibration the Flowscale is sealed in order to prevent
error or misuse.

The Flowscale has EC Type Examination Certificate No. DK 0199.361 issued by Delta.

This Certificate is in accordance with the requirements for the automatic weighing instruments
in Directive 2004/22/EC of the European Parliament and Council of March
31, 2004 on Measuring Instruments (MID).

The conformity with essential requirements in Annex MI-006, chapter I & V of the
Directive 2004/22EC is met by the application of OIML R50-1:1997, section 12 & 13 OIML
D11:2004, WELMEC Guide 7:2:2011, and WELMEC Guide 8.16-4:2006

US approval NTEP CC Certificate Number: 13-019

5
Sealing of the indicator
There are two ways of sealing the indicator:
1. Using event counters CAL and CON
2. Use of external wire and plumb seal and an internal jumper switch

• The indicator is sealed by two event counters, • In the case of a remote configuration, which
CAL & CON which are affected by each is possible, the primary indicator will blink and
change in the sealed parameters. the remote adjustment will also increment the
counters.
• These event counters have the function of
sealing the adjustment parameters and • The event counter label is sealed by partially
configuration settings. covering it with an official sealing label.
• Legal authorities can use their own means of
affixing and inscribing the sealed count of the
event counter, or documenting the count for
later confirmation of the sealing status.
• If sealing of the access of the indicator is
required, a sealing label is placed across the
upper and lower part of the enclosure.
Example of event counter with official sealing label
6
Sealing of the indicator
• The second method to seal the indicator is to • During certification process, the notified
use an internal jumper switch T4 or Seal body can seal the top cover of the
Enable. indicator with a fastening wire and a
• The Seal Enable is located on the printed circuit plumb seal. This seal is placed on the
board inside the sealable indicator and will lower part of the indicator as seen on the
make it impossible to change the sealed picture below.
configuration.

Indicator wire and plumb seal


Seal Enable, default set to Open or No (Off)

7
Hardware sealing
To secure the Flowscale hardware the height of the conveyor is sealed, this prevents changes to the settings
that could affect weighing results.
This seal is placed under the weighing belt, see item marked 22 in below pictures.

Conveyor center height adjustment bolts & sealing

If this seal is broken, the Flowscale needs to be tested and re-verified.


The scale will not function in line with standards or regulations if neither of the two sealing methods are used.
8
Recording of weight measurements
• The Marel Flowscale has only one approved indication of weight which is called a Master Totalizer or
General Totalizer. On the “Home” display the Master Totalizer is presented in large numerical along with
the weighing unit (kg/lb). The weighing division is 1 kg (2lb).

Master totalizer

• The Master Totalizer will keep incrementing until one million tons and is non-resettable, this is done to
prevent tampering.
• When the Master totalizer reaches the limit it will start counting from zero again and starts totalizing
weight from that value. If the Master Totalizer would be removed from the firmware menu, then this would
increment the CON audit trail counter.
• The approved way of recording weight runs is to write down the value of the Master Totalizer at the start
and the end of each run.
9
Data registration & protection
• The Flowscale has a built-in weight-recording device, which
is operated by pressing the “RECORD” key on the M2200
indicator.
Accessing weight records
• A Lua script triggers the recording function of the Flowscale
weighing indicator firmware. If the “RECORD” key is pressed The weight records are
a weight record is stored in the alibi memory of the scale. numbered and can be looked up from
When this key is pressed the Lua script sends a request to the memory by keying in the record
number in the firmware menu.
the firmware of the indicator and asks for the value to record,
or to transmit over Ethernet to an external computer.
• The M2200-B03 firmware returns the data, but it also stores
this data in a persistent memory, generally called the alibi
memory. The alibi memory then becomes the recording The last record number is marked with
element of the belt weigher and captures the weight the value “R” in the second line above
information from the Master totalizer. the Start/Stop key on the indicator.

• Each record is protected by a 16 bit checksum and the


weight unit and the division recorded is the same as for the
master weight totalizer.

10
Registration of data
A weight can only be recorded when the Flowscale runs “Idle” or if it is “Stop”.
If the status is “Flow” the “RECORD” key will be inactive.

Alibi number is the record number


Example of weight
registration in alibi
Status message could indicate if the amount weighed was less than
minimum weight

Date, time and quantity

Amount of weight quantity that was weighed at a flow rate above


the maximum flow rate (Qmax) or below the minimum flow rate (Qmin)

About one million records can be stored in the alibi memory until the
Example of weight registration in alibi oldest records are written over by the latest ones

The Flowscale can be connected to a computer system and a


remote computer can be used to initiate the start of recording

11
Innova for Flowscale
Innova provides full production control of the machines, processes and procedures in the plant; from incoming raw
material, through processing and on to final goods inventory and dispatch

 Innova Food Processing Software supports


the receiving of raw material, or any other
flow of material within processing facilities
that operate the Flowscale.

 With Innova, processors ensure consistent


and reliable monitoring and reporting of the
flow of material through the Flowscale.

 Using Innova can provide additional


information with the Yield control report that
shows the input product detail by purchase
order.

12
Data registration with Innova
• When Innova receives a record from the
flowscale it creates a pack registration with the
alibi number attached as well as a product
transaction.
• Innova can also receive flow status messages
from the flowscale and record them to a second
path for informational purposes, but that is a
configurable behavior.
• If for some reason a batch is not recorded in
Innova, it is possible to add the information
manually into the Innova system by reviewing
the recent alibi number that was recorded. It is
best to access the flowscale’s web page via a
web browser.
• Regardless of what is registered in Innova the
information there can be crosschecked by
comparing it to the information registered in the
alibi memory.
13
Flowscale Video

Link to video 14
Further information available on the website and brochure