Brilliance CT: 6-Slice, 10-Slice, 16-Slice, and 16 Power Configurations
Brilliance CT: 6-Slice, 10-Slice, 16-Slice, and 16 Power Configurations
Model: 4535 670 73191, 4535 670 73181, 4535 670 78851, and 4535 670 05721
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems (“Philips”) and may not be
reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal
Department. Use of this document and the information contained in it is strictly reserved for current Philips personnel and Philips customers who have a
current and valid license from Philips for use by the customer’s designated in-house service employee on equipment located at the customer’s designated site.
Use of this document by unauthorized persons is strictly prohibited. Report violation of these requirements to the Philips Legal Department at 22100 Bothell
Everett Highway, Bothell, WA 98021.
This document must be returned to Philips when the user is no longer licensed and in any event upon Philips’ first written request.
Warranty Disclaimer
PHILIPS PROVIDES THIS DOCUMENT WITHOUT WARRANTY OF ANY KIND, IMPLIED OR EXPRESSED,
INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
Limitation of Liability
PHILIPS HAS TAKEN CARE TO ENSURE THE ACCURACY OF THIS DOCUMENT. HOWEVER, PHILIPS ASSUMES
NO LIABILITY FOR ERRORS OR OMISSIONS AND RESERVES THE RIGHT TO MAKE CHANGES WITHOUT
FURTHER NOTICE TO ANY PRODUCTS HEREIN TO IMPROVE RELIABILITY, FUNCTION, OR DESIGN. PHILIPS
MAY MAKE IMPROVEMENTS OR CHANGES IN THE PRODUCT(S) OR PROGRAM(S) DESCRIBED IN THIS
DOCUMENT AT ANY TIME.
Symbol Descriptions
This document was prepared by the CT PMG Haifa Service Innovation Department.
For any additions, corrections, or suggestions, contact:
CT PMG Haifa Customer Support Department
Email: _1HFA_Support_e2@philips.com
Safety Information
To the User of This Manual
The user of this manual is directed to read and carefully review the instructions, warnings and cautions contained herein prior to
beginning installation or service activities. While you may have previously installed or serviced equipment similar to that described in this
manual, changes in design, manufacture or procedure may have occurred which significantly affect the present installation or service.
Electrical-grounding Instructions
The equipment must be grounded to an earth ground by a separate conductor. The neutral side of the line is not to be considered the
earth ground. On equipment provided with a line cord, the equipment must be connected to properly grounded, three-pin receptacle. Do
not use a three-to-two pin adapter.
carefully prior to working on the system. For your own safety and for more detailed safety information, refer only to the current version
of the documents.
Use only specified tools and auxiliary materials. After finishing any service activity, ensure that all covers of the Brilliance CT, including
TM
WARNING IN ADDITION TO THE WARNINGS LISTED ABOVE, MAKE SURE TO FOLLOW ALL SAFETY
GUIDELINES AS DESCRIBED IN THE SAFETY CHAPTER OF THIS MANUAL. FAILURE TO DO SO
CAN RESULT IN SEVERE PERSONAL INJURY.
CAUTION Always use an ESD protection wrist strap when servicing any component in the system.
Contents
Brilliance CT System Performance ......................................................................................1-1
Introduction..........................................................................................................................1-1
Scope ........................................................................................................................... 1-1
Equipment Required..................................................................................................... 1-1
System Test Phantom .........................................................................................................1-2
Brilliance CT System Configuration Geometry And Detection Overview ............................1-5
System Description and Geometry............................................................................... 1-5
Detectors and Detector Module Design........................................................................ 1-5
System Dose Efficiency................................................................................................ 1-5
Prerequisites For Performance Testing...............................................................................1-6
Verification...........................................................................................................................1-6
Performance Tests And Specifications ...............................................................................1-7
Test 1. Peak Tube Potential Check (FDA Test #303) ................................................. 1-7
Test 2. Laser Alignment Check ................................................................................... 1-9
Specification .............................................................................................................1-9
Test 3. Surview (i.e. Scan Projection Radiography) Mode: Scan Accuracy, Image Quality
and Planning Accuracy ......................................................................................... 1-10
Specification ...........................................................................................................1-11
Test 4. Impulse Response and MTF Measurements................................................. 1-12
Test 5. Axial Slice Widths .......................................................................................... 1-16
Test 6. Axial CTDI 100 Air Dose - Standard Head and Body Protocols Procedure .. 1-19
Test 7. Study Cycle Times and CT Number Stability ................................................ 1-21
Test 8. General Image Quality Check (CT Number Accuracy, Artifacts and Image Uni-
formity) .................................................................................................................. 1-23
Allowed Values.......................................................................................................1-23
Test 9. Image Noise (SD) Check............................................................................... 1-30
Test 10. CT Number Linearity Check ........................................................................ 1-33
Test 11. Dose Modulation (DOM) Check................................................................... 1-37
Test 12. Infant Mode Check ...................................................................................... 1-38
Introduction
This document provides system performance tests, test methods, and performance specifications for the Brilliance
CT 6/10/16-slice and 16 Power. Personnel using this document and performing these tests are expected to be
familiar with CT system test methods and the operation of the Brilliance CT System. A Brilliance CT System
satisfying these performance specifications is a necessary but not sufficient condition for acceptable system
performance and acceptable image quality for system shipment to customers and subsequent clinical use.
Scope
This document applies to all completely assembled and fully calibrated Brilliance CT 6/10/16-slice and 16 Power
Configurations.
NOTE System Performance Data Sheets are available in pdf format on InCenter in the Brilliance CT
Performance Verification Section.
Equipment Required
System Performance Phantom p/n 4550 120 01021
System Calibration Phantom (Eye Phantom) p/n 4550 120 01031
Surview Gauge p/n 4535 671 22551
Laser Alignment Gauge p/n 4535 671 22541
COPPER WIRE
Section 1
PHYSICS LAYER
Section 2
WATER LAYER HEAD PHANTOM
45
Section 3
MULTI-PIN LAYER
BODY PHANTOM
Cross Section Details of the System Performance Phantom Composition - Multi-Pin Layer
Verification
A permanent record (as a digital file or hard copy data sheet or both) of this documents performance results shall be
recorded and stored. This record shall also be available for reference to appropriate engineering, manufacturing and
service personal via digital archive and/or hard copy.
Tube Potential Tube Potential Reading* Average of The Two Difference between the Difference
Readings KV settings and the Ave. Specification
#1 #2 Readings
90 KV +/- 3.6 kV
Specification
Automatic couch movement accuracy from Symmetry Markers to Slice Marker: plate center to plate center within +/
- 2.5 mm.
Laser Slice Marker Alignment Tolerance: within +/- 1.5 mm of center of the first axial slice.
Laser External Symmetry Markers (horizontal and vertical) Alignment tolerance: +/- 3 mm of axis of rotation or
tomographic image center.
Test 3. Surview (i.e. Scan Projection Radiography) Mode: Scan Accuracy, Image Quality and
Planning Accuracy
1 Mount the system phantom off the end of the table using the phantom
mounting bracket.
2 Use the proper test gauge, shown in Figure 1 - 5, or fabricate a stainless
steel test gauge or object according to the drawing below (a solid
copper wire may be folded to this shape in a pinch). The z dimension
length must be 150 mm +/- 1.0 mm. The two end wings should be 150 MM TEST GAUGE
approximately 20 mm long. The round diameter should be 1 +/-0.2mm.
Tape one end of this object to the top center of the body phantom with
its length extending in the air above the top center of the system head
phantom.
3 Center the System Phantom using the external symmetry laser lights.
4 Manually move the table into the gantry, so that the Nylon Body Section
of the System Phantom is just past the slice marker laser line. Image the Figure 1 - 5: Test Gauge
nylon body section but not the screw heads.
5 Perform a default PA Body Surview Scan (i.e. 30 or 50 mA and 120 KV) of the entire System Phantom (Head and
Body Phantoms). Increase the mA as needed to image the test gauge thru the body phantom but not above 50
mA. .
6 Visually examine the Surview Image for Image quality as specified below.
7 Using the software "line length measuring tool", carefully measure the length of the 150 mm long test gauge in
the Surview image. The measured length from the image must be accurate to within +/- 1 mm of the true length
as measured with an accurate scale (i.e. 150 +/- 1 mm). Result of line length measurement =
_________________ mm.
8 Carefully plan an axial sequence of 3 mm wide images using the acquisition technique defined below. Start the
plan exactly at one end of the gauge object and stopping just past the other end; as close as possible. You may
have to magnify and then re-center the Surview image in order to accurately plan the scan.
9 Acquire the above plan using the following standard body mode technique: 120KV, 250 mAs, 16 x 1.5 mm, or 10
x 15 mm, or 6 x 1.5 mm collimation mode, 3 mm slice width, contiguous mm increment, etc.
10 The wings or ends of the object should be visible in the first and the 50th image.
11 Repeat the above planning and acquisition procedure in the helical scanning mode using a helical pitch of 0.9,
thickness = 3 mm, increment = contiguous, the number of images are now determined from the helical plan.
Specification
Surview Mode Scan measurement accuracy: +/- 1mm
Image Quality: No unusual streaks or stripes; test gauge visible for planning purposes, phantom details sufficient
for planning scans.
Planned axial slice alignment: +/- 1.5 mm of center of first axial slice (visible in first image and 50th image)
Planned helical slice alignment: +/- 1.5 mm of center of first helical slice (visible in first image one of the last two
images)
Std_Body_QA protocol: 120 KV, 240mAs, scan angle =360 degrees, resolution = Standard (Body), rotation time =
0.75 sec, Recon filter B, 512 x 512 matrix size, number of scans =1, etc.
4 Perform the Impulse Response and MTF measurements on selected wire images by activating the Home /
Utilities / Misc / Impulse Response program and then by selecting the above-acquired images to be analyzed
from the image archive disk. Select the central two images from each scan (Head and Body) for analysis.
5 Click on "Curs" to get the impulse response measurement cursor. Place the cursor directly on top of the circular
cross-sectional image of the wire in the selected images. Press the "calculate" button and record the results in
the table on the next page. You may have to read the required MTF values off the MTF graph.
6 Repeat the above in both the HR (High Resolution) and UHR (Ultra- High-Resolution) axial head modes. Adjust
the head protocol as follows: FOV = 100mm and 1024 matrix size as indicated in the table below and Recon
Center set to Center x (mm) = -40, Center y (mm) = +24,the rotation time = 1.5sec; collimation mode 16/10/6 x
0.75mm, slice width = 1.5mm; mAs = 400; scan mode = HR or UHR; recon filter = D for HR and E for UHR. Again
acquire one scan with zero table increment through the center of the thin copper wire in the Head section.
7 Measure and record the Impulse Response and MTF results as indicated in the Table(s) below for either 6-slice,
10-slice, 16-slice, or 16 Power.
Measured Measured
Measured I.R. Specified Specified
Rotation Specified I.R at MTF at 50% MTF at 10%
Scan # Filter FOV (mm) Resolution at 50% (mm) MTF at 50% MTF at 10%
Time (sec) 50% (mm) (lp/cm) Slice (lp/cm) Slice
Slice 1 Slice 2 (lp/cm) (lp/cm)
1 Slice 2 1 Slice 2
Standard
1 EB 100 0.75 1.43 +/- 0.1 2.8 +/- 0.5 5.5 +/- 0.5
(Head)
Standard
2 B 100 0.75 1.25 +/- 0.1 3.3 +/- 0.5 6.5 +/- 0.5
(Body)
100 (1024
3 D High (Head) 1.5 0.72 +/- 0.08 7.5 +/- 0.7 11.5 +/-1.0
matrix)
Measured Measured
Measured I.R. Specified Specified
Rotation Specified I.R at MTF at 50% MTF at 10%
Scan # Filter FOV (mm) Resolution at 50% (mm) MTF at 50% MTF at 10%
Time (sec) 50% (mm) (lp/cm) Slice (lp/cm) Slice
Slice 1 Slice 2 (lp/cm) (lp/cm)
1 Slice 2 1 Slice 2
Standard
1 EB 100 0.75 1.43 +/- 0.1 2.8 +/- 0.5 5.5 +/- 0.5
(Head)
Standard
2 B 100 0.75 1.25 +/- 0.1 3.3 +/- 0.5 6.5 +/- 0.5
(Body)
100 (1024
3 D High (Head) 1.5 0.72 +/- 0.08 7.5 +/- 0.7 11.5 +/-1.0
matrix)
Measured Measured
Measured I.R. Specified Specified
Rotation Specified I.R at MTF at 50% MTF at 10%
Scan # Filter FOV (mm) Resolution at 50% (mm) MTF at 50% MTF at 10%
Time (sec) 50% (mm) (lp/cm) Slice (lp/cm) Slice
Slice a Slice b (lp/cm) (lp/cm)
a Slice b a Slice b
Standard
1 EB 100 1 1.43 +/- 0.1 2.8 +/- 0.5 5.5 +/- 0.5
(Head)
Standard
2 B 100 0.75 1.25 +/- 0.1 3.3 +/- 0.5 6.5 +/- 0.5
(Body)
100 (1024
3 D High (Head) 1.5 0.72 +/- 0.08 7.5 +/- 0.7 11.5 +/-1.0
matrix)
4 Measure the FWHM slice width from the images of the ramps using the Slice Width program found under the
Misc menu. Once in the slice width program is selected, press the open archive icon and select one image for
measurement. Select the image where the aluminum image is approximately centered in the ramp image. Click
the ”vertical ROI” icon on the left tool bar and measure both ramps on the selected image. (If the phantom is level
and the ramps are centered in the slice, the two ramp measurements should give similar results.) Report the
average slice width value for each image on the data sheet.
Test 6. Axial CTDI 100 Air Dose - Standard Head and Body Protocols Procedure
1 Collect and assemble the radiation meter (Radcal model 9015 or equivalent) with the CT pencil probe ion
chamber.
2 Align the pencil probe along the axis of rotation and centered in the x-ray beam. The probe should extend off the
end of the table so that the radiation can be measured in air without table attenuation or other scatter. You may
need to develop a fixture for holding the probe in this position.
3 Using the following head technique with fused slices, (i.e. Std. Res., 120kV, 100mAs, 0.75 sec rot time, 16 x 1.5
mm collimation mode, 6mm slice width, 0 increment, 4 slices). Make 5 exposures one at a time with zero table
increment, and record the 5 readings on the data sheet.
4 Repeat these 5 exposures at 90 KV and record the readings on the data sheet.
5 Repeat this measurement for the body scan technique of 140 kV, 100mAs, 0.75 sec rot time, 16 x 1.5 mm
collimation mode, 6 mm slice width, 0 increment, 4 slices).
6 Repeat single exposures at the slice widths indicated in rows 6,7, and 8.
7 Calculate the average values and the maximum deviations for the three different techniques and record them in
the data sheet.
Table 1 - 7: Axial CTDI 100 Air Dose - Standard Head and Body Protocols Procedure Data Sheet
Head Mode Air Radiation Values (Mr.) Body Mode Value (Mr.)
Exposure Number
120 KV Spec 90 KV Spec 140 KV Spec
2
3
8 100 48 142
(4 x 0.75 mm collimation mode) ±20% ±20% ±15%
* For helical evolving, time only the final full matrix recon time for the 40 images (do not include the Evolving preview
image recon time).
Test 8. General Image Quality Check (CT Number Accuracy, Artifacts and Image Uniformity)
1 Before performing this test, heat the x-ray tube to 30% anode heat. From Home, select Tube Conditioning >
Short TC. Select Utilities > Misc. > Show Tube Heat to display tube heat.
2 Click Home > Utilities > Diagnostic > Service Parameters. Verify Phantom Calibration = Yes, Image Balance=
Yes.
3 From the Home menu, go to Air Calibration and select and perform All Calibrations.
4 Use the Physics System phantom and perform one scan for each mode according to the tables below for 6-slice,
10-slice, 16-slice, or 16 Power. If a test fails, repeat the test for a second try at passing the test.
5 Look for possible rings or other artifacts according to the next table.
6 Check p in the appropriate column of the table(s) below for 6-slice, 10-slice, 16-slice, or 16 Power if the CT
number for all slices is within the allowed tolerance.
NOTE CT number mean value is measured by a 7,000+/-500 mm2 area round ROI at the center of the Head
images and a 13,000+/-500 mm2 area ROI on the Body images. For Head pins scans, use a central ROI of
1,500 +/- 500 mm2 area to measure the water ct number.
7 For Head only-Check the image CT number uniformity for scans with filter UB by using a 1,000+/-100 mm2 area
ROI in the center and a sliding 1000+/-100 mm2 area ROI to be moved around, but not closer than, 1 cm from the
"bone" or phantom edge. Uniformity within an image should be within +/- 2 CT numbers of central value for the
indicated Head scan below.
Allowed Values
• No rings or bands are allowed on the images. Use a window width of 100 for Head and 150 for Body with the
window centered on the CT number of the material.
• Allowed CT number means that the reference CT value for water (Head protocol) should be zero. Reference CT
Value for Body Aculon (defined as X) should be read from System Phantom CT Calibration Values on System
Phantom CT Calibration Values for each voltage respectively.
• CT number uniformity tolerance is Y +/- 2 CT for Head scans mode with filter UB, where Y is the mean CT
number measured by center ROI in each case.
NOTE If the CT number is outside the limit, note and record the value and the slice number. Repeat test. It
test fails a second time, perform HCOR calibration. From the Home menu, select Calibration >
Selective Calibration > HCOR Calibration. Then repeat General Image Quality Check test.
NOTE All measurements of CT numbers and STD (noise) should be done on 4 on 1 viewer maximum size
window.
Table 1 - 8: Head QA Head Protocol (Axial), FOV 250 mm, 512 matrix, 360 degree scan angle, System Head Phantom Aligned
and Centered, Window Center 0 (zero), Window Width = 100 (6-slice)
1 Head Water 120kV, 4x6 (SW 6mm), STD, 0.75s, 300 mAs, EB 0 +/- 4
2 Head Water 120kV, 4x4.5 (SW 4.5mm), STD, 1.5s, 440 mAs, UC 0 +/- 4
0 +/- 4
120kV, 6x3 (SW 6mm), STD, 1.0s, 330 mAs, UB 420
3 Head Water Uniformity Within
degree scan angle
Images (Y +/- 2)
5 Head Pins 120kV, 6x0.75 (SW 0.75mm), High, 1.0s, 300 mAs, C 0 +/- 10
6 Head Water 90kV, 6x3 (SW 6mm), STD, 1.0s, 250 mAs, EB 0 +/- 4
7 Head Water 140kV, 6x3 (SW 6mm), STD, 0.75s, 300 mAs, UC 0 +/- 7
Copy the Reference CT Values for Body Aculon (defined as X) from the System Phantom Label of the System
Phantom used for this test in the table below. If these reference values are not available for this system or out of
date, then re-measure the X values using the CT Linearity Check protocol (Test 10) and record below.
KV X
Body Aculon 140 kV
Body Aculon 120 kV
Body Aculon 90 kV
Table 1 - 9: QA Body Protocol (Axial), FOV 350 mm, 512 matrix, System Body Phantom Aligned and Centered,
Window Center = X. Window Width =150 (6-slice)
CHECK p IF CT FOR ALL
SCAN PARAMETERS KILOVOLT, SW, ALLOWED CT
SCAN # PHANTOM SLICES ARE OK OR ENTER
RESOLUTION, ROT.TIME, mAs, FILTER NO.
THE NUMBER
1 Body 120kV, 4x6 (SW 6mm), STD, 0.75s, 300 mAs, B X +/- 4
2 Body 120kV, 6x0.75 (SW 0.75mm), STD, 1.0s, 325 mAs, B X +/- 6
3 Body 90kV, 6x1.5 (SW 1.5mm), STD, 1.0s, 350 mAs, B X +/- 6
120kV, 6x0.75 (SW 1.5mm), U-fast, 0.42s, 165 mAs,
4 Body X +/- 6
B
5 Body 120kV, 6x1.5 (SW 1.5mm), STD, 0.5s, 200 mAs, B X +/- 6
6 Body 120kV, 6x0.75 (SW 0.75mm), High, 2.0s, 350 mAs, B X +/- 8
7 Body 140kV, 6x1.5 (SW 1.5mm), High, 1.0s, 250 mAs, B X +/- 8
120kV, 6x0.75 (SW 0.75mm), UF detail, 0.42s, 135
8 Body X +/- 8
mAs, B, (FOV = 250 mm)
Helical Abdomen Scan, 120kV, 6x3 (SW 5mm), STD, Evaluate two images
9 Body X +/- 8
0.5s, 200 mAs, pitch=0.9, Filter=B, SP Filter=yes centered in the body section
SURVIEW SCAN, 500 MM LENGTH
120kV, 50 mA, Scan angle 180 deg -- Check OK if no
Head and vertical and horizontal lines appear on the image and
10
Body the phantom details are clearly visible for planning
purposes.
Table 1 - 10: Head QA Head Protocol (Axial), FOV 250 mm, 512 matrix, 360 degree scan angle, System Head Phantom
Aligned and Centered, Window Center 0 (zero), Window Width = 100 (10-slice)
1 Head Water 120kV, 8x3 (SW 6mm), STD, 0.75s, 300 mAs, EB 0 +/- 4
2 Head Water 120kV, 4x4.5 (SW 6mm), STD, 1.5s, 440 mAs, UC 0 +/- 4
0 +/- 4
120kV, 4x3 (SW 3mm), STD, 1.0s, 330 mAs, UB 420
3 Head Water Uniformity Within
degree scan angle
Images (Y +/- 2)
4 Head Pins 120kV, 10x0.75 (SW 0.75mm), U-high, 1.5s, 400 mAs, D 0 +/- 10
5 Head Pins 120kV, 10x0.75 (SW 0.75mm), High, 1.0s, 300 mAs, C 0 +/- 10
6 Head Water 90kV, 8x3 (SW 6mm), STD, 1.0s, 250 mAs, EB 0 +/- 4
7 Head Water 140kV, 8x3 (SW 6mm), STD, 0.75s, 300 mAs, UC 0 +/- 4
Copy the Reference CT Values for Body Aculon (defined as X) from the System Phantom Label of the System
Phantom used for this test in the table below. If these reference values are not available for this system or out of
date, then re-measure the X values using the CT Linearity Check protocol (Test 10) and record below.
KV X
Body Aculon 140 kV
Body Aculon 120 kV
Body Aculon 90 kV
Table 1 - 11: QA Body Protocol (Axial), FOV 350 mm, 512 matrix, System Body Phantom Aligned and Centered,
Window Center = X. Window Width =150 (10-slice)
CHECK p IF CT FOR ALL
SCAN PARAMETERS KILOVOLT, SW, ALLOWED CT
SCAN # PHANTOM SLICES ARE OK OR ENTER
RESOLUTION, ROT.TIME, mAs, FILTER NO.
THE NUMBER
1 Body 120kV, 8x3 (SW 6mm), STD, 0.75s, 300 mAs, B X +/- 4
2 Body 120kV, 10x0.75 (SW 0.75mm), STD, 1.0s, 325 mAs, B X +/- 6
3 Body 90kV, 10x1.5 (SW 1.5mm), STD, 1.0s, 350 mAs, B X +/- 6
120kV, 10x0.75 (SW 1.5mm), U-fast, 0.42s, 165 mAs,
4 Body X +/- 6
B
5 Body 120kV, 10x1.5 (SW 1.5mm), STD, 0.5s, 200 mAs, B X +/- 6
6 Body 120kV, 10x0.75 (SW 0.75mm), High, 2.0s, 350 mAs, B X +/- 8
7 Body 140kV, 10x1.5 (SW 1.5mm), High, 1.0s, 250 mAs, B X +/- 8
120kV, 10x0.75 (SW 0.75mm), UF detail, 0.42s, 135
8 Body X +/- 8
mAs, B, (FOV = 250 mm)
Helical Abdomen Scan, 120kV, 10x1.5 (SW 5mm),
Evaluate two images
9 Body STD, 0.5s, 200 mAs, pitch=0.9, Filter=B, SP X +/- 8
centered in the body section
Filter=yes
SURVIEW SCAN, 500 MM LENGTH
120kV, 50 mA, Scan angle 180 deg -- Check OK if no
Head and vertical and horizontal lines appear on the image and
10
Body the phantom details are clearly visible for planning
purposes.
Table 1 - 12: Head QA Head Protocol (Axial), FOV 250 mm, 512 matrix, 360 degree scan angle, System Head Phantom
Aligned and Centered, Window Center 0 (zero), Window Width = 100 (16-slice, 16 Power)
1 Head Water 120kV, 16x1.5 (SW 6mm), STD, 0.75s, 300 mAs, EB 0 +/- 4
2 Head Water 120kV, 16x1.5 (SW 6mm), STD, 1.5s, 440 mAs, UC 0 +/- 4
0 +/- 4
120kV, 16x1.5 (SW 6mm), STD, 1.0s, 330 mAs, UB 420
3 Head Water Uniformity Within
degree scan angle
Images (Y +/- 2)
4 Head Pins 120kV, 16x0.75 (SW 0.75mm), U-high, 1.5s, 400 mAs, D 0 +/- 10
5 Head Pins 120kV, 16x0.75 (SW 0.75mm), High, 1.0s, 300 mAs, C 0 +/- 10
6 Head Water 90kV, 16x1.5 (SW 6mm), STD, 1.0s, 250 mAs, EB 0 +/- 4
7 Head Water 140kV, 16x1.5 (SW 6mm), STD, 0.75s, 300 mAs, UC 0 +/- 4
Copy the Reference CT Values for Body Aculon (defined as X) from the System Phantom Label of the System
Phantom used for this test in the table below. If these reference values are not available for this system or out of
date, then re-measure the X values using the CT Linearity Check protocol (Test 10) and record below.
KV X
Body Aculon 140 kV
Body Aculon 120 kV
Body Aculon 90 kV
Table 1 - 13: QA Body Protocol (Axial), FOV 350 mm, 512 matrix, System Body Phantom Aligned and Centered,
Window Center = X. Window Width =150 (16-slice, 16 Power)
CHECK p IF CT FOR ALL
SCAN PARAMETERS KILOVOLT, SW, ALLOWED CT
SCAN # PHANTOM SLICES ARE OK OR ENTER
RESOLUTION, ROT.TIME, mAs, FILTER NO.
THE NUMBER
1 Body 120kV, 16x1.5 (SW 6mm), STD, 0.75s, 300 mAs, B X +/- 4
2 Body 120kV, 16x0.75 (SW 0.75mm), STD, 1.0s, 325 mAs, B X +/- 6
3 Body 90kV, 16x1.5 (SW 1.5mm), STD, 1.0s, 350 mAs, B X +/- 6
120kV, 16x0.75 (SW 1.5mm), U-fast, 0.42s, 205 mAs,
4 Body X +/- 6
B
5 Body 120kV, 16x1.5 (SW 1.5mm), STD, 0.5s, 245 mAs, B X +/- 6
6 Body 120kV, 16x0.75 (SW 0.75mm), High, 2.0s, 350 mAs, B X +/- 8
7 Body 140kV, 16x1.5 (SW 1.5mm), High, 1.0s, 250 mAs, B X +/- 8
120kV, 16x0.75 (SW 0.75mm), UF detail, 0.42s, 135
8 Body X +/- 8
mAs, B, (FOV = 250 mm)
Helical Abdomen Scan, 120kV, 16x1.5 (SW 5mm),
Evaluate two images
9 Body STD, 0.5s, 400 mAs, pitch=0.9, Filter=B, SP X +/- 8
centered in the body section
Filter=yes
SURVIEW SCAN, 500 MM LENGTH
120kV, 50 mA, Scan angle 180 deg -- Check OK if no
Head and vertical and horizontal lines appear on the image and
10
Body the phantom details are clearly visible for planning
purposes.
1 Head Pins FOV 250, 120 kV, 6x3 (SW 3mm), STD, 0.75s, 200 mAs, EB
2 Body FOV 350, 140 kV, 6x3 (SW 3mm), STD, 0.75s, 200 mAs, B
3 Body FOV 350, 120 kV, 6x3 (SW 3mm), STD, 0.75s, 200 mAs, B
4 Body FOV 350, 90 kV, 6x3 (SW 3mm), STD, 0.75s, 200 mAs, B
1 Head Pins FOV 250, 120 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, EB
2 Body FOV 350, 140 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, B
3 Body FOV 350, 120 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, B
4 Body FOV 350, 90 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, B
1 Head Pins FOV 250, 120 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, EB
2 Body FOV 350, 140 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, B
3 Body FOV 350, 120 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, B
4 Body FOV 350, 90 kV, 4x3 (SW 3mm), STD, 0.75s, 200 mAs, B
Nylon (Aculon)
Lexan Nylon (Aculon) Polyethylene
Body 90 kV 85 ± 15 N/A
Teflon
% Standard Deviation Difference = 100 x (S.D. DOM On - S.D. DOM Off) / (S.D. DOM Off)
1 Install the Infant head and Body phantom off the end of the table. This phantom has as 11 cm diameter head
water section and a 15 cm diameter body water section.
2 Using external lasers, align the phantom with the axis of rotation.
3 Scan the smaller head section of this phantom using an axial infant brain protocol at 120 kVp. Make sure that the
infant mode is selected for this scan. (Infant radio button checked when entering patient name).
4 Using a circular central ROI with a diameter of half of the water image diameter, measure the CT number of a
central slice of water and record the number in the data table below.
5 Scan the larger body section of this phantom using a helical infant body protocol (2-10 kg abdomen protocol).
Make sure that the infant mode is selected for this scan. (Infant radio button checked when entering patient
name).
6 Again, make sure that the infant mode is selected for this scan. (Infant radio button checked when entering
patient name).
7 Check that both the infant head and infant body images (away from the end caps and transitions) appear uniform
and free of ring and/or shading artifacts.
The Constancy test utility enables you to run the Acceptance and Constancy tests. These tests may be required by local
authorities or by the site’s internal guidelines. The tests are performed by an authorized Philips Service representative.
The Acceptance test can be performed when the CT scanner is installed or run for the first time, after a significant upgrade, or
when the configuration is changed. Acceptance Test results are compared to factory default tolerances. The Acceptance test
sets the baseline.
The Constancy test is performed to ensure that the image quality of the CT scanner remains at the highest standards at all
times. It can be performed periodically at the discretion of the local authorities. Constancy Test results are compared to the
baseline generated by the Acceptance test.
The Constancy Test Automation tool supports automated stages. The tool supports two modes of operation: Baseline
measurement and Periodic measurement. These are the items checked by the Constancy Test Utility:
• Mean CT numbers of Head and Body protocols.
• CT Uniformity numbers of Head and Body protocols.
• Image Noise - Standard Deviation of Head and Body protocols.
• Spatial resolution as measured by the MTF at 10% and 50% in lp/cm.
• Slice Thickness measured at FWHM (in mm).
NOTE The CT scanner should be nominally calibrated by an authorized Philips service staff member. It is
recommended to perform the Air calibration immediately before running the Acceptance or Constancy tests.
Getting Started
Perform a visual check of the system phantom to make sure it is in proper operating condition before running the Constancy
Test Utility. Make sure there isn’t an air bubble in the water pin of the body section or in the head section. If an air bubble is
present, open the screw on the cover of the water pin and inject some water. You must use the Brilliance CT system phantom
(p/n 455012001021).
Setup
Before you can conduct the Acceptance or Constancy tests, you must complete a special log-in procedure. Follow these steps
to log-in.
1 Log onto the system using this user name: auto_test.
2 Type in the password: constancy.
3 On the Constancy Test Utility, click the Start Scanner Application button.
4 Once the Scanner Application is loaded, then click the Run Constancy Test button on the Constancy Test Utility.
5 Click OK on the opening dialog box that appears.
6 The configuration dialog box opens if system information has not been entered.
10 The system conducts an automatic application initialization procedure. When it is complete, the patient data window
displays:
NOTE This window may be minimized to the Windows Taskbar.
Acceptance test
Use this procedure to run an Acceptance test.
1 On the Constancy Test Utility screen, click the Run Acceptance Test button. Click OK to continue:
NOTE If you have never run an Acceptance Test before, or if the last test was completed, skip to step 2.
If the previous run was not completed (not all stages were passed), a
confirmation message appears. Follow Step a below.
a Select one of the following:
• Yes - to complete the last run (failed stages will run).
• No - to run tests from the beginning (all stages will run).
• Cancel - to cancel the test.
3 You will get the message at right. Click the OK button to continue, or
the Cancel button to stop the process.
4 This message below will display. Follow the instructions on the screen.
CAUTION It is extremely important to make sure the laser markers are properly aligned with the
marks on the phantom. If they are not, the Acceptance Test will fail.
c Click OK to continue. A series of tests run at this time. Do not interrupt the process until the results box appears. You can
also click Cancel to stop the process.
CAUTION Do not use the CT Box to attempt to control the sequence. The tests run automatically.
CAUTION It is not recommended that you interrupt the test procedure however, you can do so
from the following dialog box that appears on the background of your terminal.
d After the first test, the utility will prompt you to adjust the height of the couch. Follow the instructions on the screen.
e The system displays several images during the process.
5 After the run ends, a baseline is automatically saved. You receive the following message
6 Click OK.
NOTE A baseline is saved only if all stages passed and the measurements are within the required ranges.
8 Click OK after you fix the cause of the failure, and retry the tests.
9 Reports are generated automatically and opened in Microsoft Explorer. The following is an example of a test report
.
Constancy test
Use this procedure to run a Constancy test.
1 On the Constancy Test Utility screen, click the Run Constancy Test button:
NOTE If you have never run an Constancy Test before, or if the last test was completed, skip to step 2.
If the previous run was not completed (not all stages were passed), a
confirmation message appears. Follow Step a below.
a Select one of the following:
• Yes - to complete the last run (failed stages will run).
• No - to run tests from the beginning (all stages will run).
• Cancel - to cancel the test.
2 If you choose Yes, the Scenario setup dialog box will appear.
• Fill in your name and the reason for performing the test.
• Click OK.
3 You will get the message at right. Click the OK button to continue, or
the Cancel button to stop the process.
4 This message below will display. Follow the instructions on the screen.
CAUTION It is extremely important to make sure the laser markers are properly aligned with the
marks on the phantom. If they are not, the Constancy Test will fail.
c Click OK to continue. A series of tests run at this time. Do not interrupt the process until the results box appears. You can
also click Cancel to stop the process.
CAUTION Do not use the CT Box to attempt to control the sequence. The tests run automatically.
CAUTION It is not recommended that you interrupt the test procedure however, you can do so
from the following dialog box that appears on the background of your terminal.
d After the first test, the utility will prompt you to adjust the height of the couch. Follow the instructions on the screen.
e The system displays several images during the process.
5 After the run ends, a baseline is automatically saved. You receive the following message
6 Click OK.
NOTE A baseline is saved only if all stages passed and the measurements are within the required ranges.
8 Click OK after you fix the cause of the failure, and retry the tests.
The Status/Progress Indicator is located at the bottom of the screen and displays messages regarding the actions the utility
performs (such as generating report or idle).
Creating a baseline
Use this procedure to create a baseline from the Constancy Test results.
1 In the Re-define Acceptance/Constancy Baselines box, click Add Constancy
Baseline.
2 Select the timestamp of the test that was just completed.
3 Click OK to close the dialog box.
Removing a Baseline
1 In the Re-define Acceptance/Constancy Baselines box, select the baseline you want to remove from the list.
2 Click Remove Baseline. This message displays.
3 Select an option:
• Click Yes to the remove the baseline.
• Click No to keep the baseline.
Viewing a baseline
In the Re-define Acceptance/Constancy Baselines box, click View Baseline. This
opens a window with the actual tolerances associated with the selected baseline.
Showing reports
To Show Reports, click the Show Report button in the Constancy Test
utility window.
A dialog box displays which consists of two lists with Acceptance and
Constancy test histories.
1 Select the report you want to view.
2 Click Show Report(s). You receive the report in Microsoft Explorer.
Exporting reports
1 Save a report by clicking Export Reports. You receive the following dialog box which
allows you to select where you want to save the report.
2 Select the location where you want to save the file.
3 Click OK.
You can also reach this dialog box by selecting File > Export Report.
1 Select the desired report(s).
2 Click the Export Report button.