0% found this document useful (0 votes)
813 views88 pages

Datasheet Mass Intl

The document provides specifications for various mass product pressure gauges and accessories. It includes details like model numbers, measurement ranges, connection types, materials, accuracy levels, dimensions and ordering information for over 50 different pressure gauge models ranging from utility to industrial to process gauges. Technical specifications are provided for multiple models of utility pressure gauges with nominal sizes from 40mm to 63mm.

Uploaded by

Reza Pramata
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
813 views88 pages

Datasheet Mass Intl

The document provides specifications for various mass product pressure gauges and accessories. It includes details like model numbers, measurement ranges, connection types, materials, accuracy levels, dimensions and ordering information for over 50 different pressure gauge models ranging from utility to industrial to process gauges. Technical specifications are provided for multiple models of utility pressure gauges with nominal sizes from 40mm to 63mm.

Uploaded by

Reza Pramata
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 88

Mass product datasheet

Index
Model Page
PG120 Uitility gauges (Size 40, 50, 63 mm) 1
PG121 Uitility gauges (Size 100, 150, 250 mm) 3
GFB Industrial gauges crimp ring type with brass connection 5
GES Industrial gauges crimp ring type compact design 7
GFS Industrial gauges crimp ring type 9
WPS Open front process gauges 11
ZPS Stainless steel process gauges 13
SFS Solid front process gauges 15
TGS Test gauges accuracy 0.5% 17
MGS Master gauges accuracy 0.25% 19
SDS Diaphragm gauges for low pressure measurement 21
D100 Differential pressure gauges bellow type 23
D101 Differential pressure gauges diaphragm type(25 bar) 25
D104 Differential pressure gauges diaphragm type(100 bar) 27
D105 Differential pressure gauges double diaphragm type 29
D201 Differential pressure gauge magnetic piston type 31
DS20 Screwed Diaphragm seal general purpose application 33
DS21 Screwed Diaphragm seal 35
DS50 Screwed Diaphragm seal All welded construction 37
DS51 Screwed Diaphragm seal compact design 39
DS52 Sanatary seal tri-clamp type 40
DS54 Screwed Diaphragm seal for high pressure 41
DF20 Flanged Diaphragm seal 43
DF21 Flanged Diaphragm seal for exocitic material 45
DF50 Flush type flanged diaphragm seal 47
DF52 Diaphragm seals pancake type 49
BMT Bi-metallic temperature gauges 51
MTS Mercury filled temperature gauges 53
GTS Gas filled temperature gauges 55
RTH Resistant temperature detector with head 57
RTW Resistant temperature detector without head 59
TCH Thermocouple with Head 61
Mass product datasheet
TWT Screwed thermowell 63
TWF Flanged thermowell 65
EC Electric contact 67
SY1 Syphon 68
CT1,CT5 Cooling Tower 69
PD1 Pulsation dampener / Snubber 70
PL1 Pressure limiter / Gauge saver 71
GV Gauge valves / Gauge cock 72
NV1 Needle valve 73
2VM 2 valve manifold 75
3VM 3 valve manifold 77
5VM 5 valve manifold 79
Utility pressure gauges
Model PG120
Nominal size 40 mm, 50 mm, 63 mm
Accuracy: Class 2.5 (2.5%)

Features
x Rugged construction
x Protection IP54
x Overload protection 130 %
2
x Standard dial dual scale (psi, kg/cm )
x Dry

Ranges
-30 in. Hg ... 0 psi up to 0 ... 6,000 psi

Applications
Heating and ventilation
Machine and apparatus construction
Air filters and regulators
Valve positioners and compressors

Technical specification PG120


Dial size in mm 40 50 63
Construction Open front with push-in ring
Measuring principle Bourdon tube
Range in psi 15 20 30 50 60 100 150 200 250 300 400 500 600 800
1000 1500 2000 3000 4000 5000 6000
“HG / psi -30”/0 -30”/15 -30”/30 -30”/60 -30”/100 -30”/150 -30/200
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/8” NPT/BSP (M) 1/8” 1/4” NPT/BSP (M)
Connection location Lower, back
Material
Pressure connection Brass optional stainless steel 316
Tube Phosphor Bronze optional stainless steel 316
Case/ring Black finished steel case optional stainless steel case
Window Instrument glass, optional polycarbonate cap
Dial Aluminum, black markings on white background
Pointer Aluminum, black
Movement Brass optional stainless steel
Accuracy Class 2.5 (2.5%)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 65 °C
Storage temperature -40 ... 65 °C
Effect Max. 0,3 % / 10 K
Protection according IP54
Mounting Standard stem, optional panel, bracket, surface
Weight dry in gram 60 90 105

All specifications are subject to change without notice.


Page 1
IN.PG120 Rev. G 19/11/2016
General dimensions in mm

Order information
Size System Execution Process Connection Range Engineering Case Material Options
Material connection orientation units
(40) (B) Tube (D) dry IP54 (02)1/4” NPT (M) 1) (L) Lower -30”/ 01) (psi & kg/cm2) (=) Standard (DA) Dial Marking
40mm phosphor -30”/ 151) Dual scale Steel
bronze, (01)1/8” NPT (M) (C) Centre -30”/ 301) “Hg / psi & case (ON) Neutral, without
(50) socket back -30”/ 601) mmHg / black logo
brass (13)1/4” BSP (M) 1) -30”/ 1001) kg/cm2 finish
50mm (UC) U-clamp
-30”/ 1501)
(S) 316 tube, (12)1/8” BSP (M) -30”/ 2001)
(63) (YX) Stainless (FX) Front panel 1)
316 socket steel 304
63mm
0/ 151) (psi & kg/cm2) case (FW) Surface plate 1)
0/ 201) Dual scale
0/ 30 psi & kg/cm2 (PC) Polycarbonate
0/ 50 cap
0/ 60
0/ 100
(CR) Crimp ring
0/ 150
0/ 2001)
0/ 2501)
0/ 3001)
0/ 4001)
0/ 5001)
0/ 6001)
0/ 8001)
0/ 10001)
0/ 15002)
0/ 20002)
0/ 30002)
0/ 40002)
0/ 50002)
0/ 60002)
1) Not for dial size 40mm 2) not for dial size 40mm, 50mm
How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
2
63 PG120 B D 02 L 0/60 psi & kg/cm = CR

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 2
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Utility pressure gauges
Model PG121
Nominal size 100 mm, 150mm and 250mm
Accuracy: Class 2.5 (2.5%)

Features
x Rugged construction
x Protection IP54
x Overload protection 130 %
2
x Standard dial dual scale (psi, kg/cm )
x Dry

Ranges
-30 in. Hg ... 0 psi up to 0 ... 10,000 psi

Applications
Heating and ventilation
Machine and apparatus construction
Hydraulic and pneumatic systems
Boiler and refrigeration

Technical specification PG121


Dial size in mm 100 150 250
Construction Open front with push-in ring
Measuring principle Bourdon tube
Range in psi 10 15 20 30 50 60 100 150 200 250 300 400 500 600 800 1000
1500 2000 3000 4000 5000 6000 8000 10000
“Hg / psi -30”/0 -30”/15 -30”/30 -30”/60 -30”/100 -30”/150 -30/200
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/4” 3/8” 1/2” BSP / NPT (M) , M20x1.5 (M)
Connection location Lower, back
Material
Pressure connection Brass
Tube Bronze for ranges ≤ 1000 psi, SS316 for ranges > 1000psi
Case/ring Black finished steel case
Window Instrument glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black
Movement Brass
Accuracy Class 2.5 (2.5%)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 65 °C
Storage temperature -40 ... 65 °C
Effect Max. 0,3 % / 10 K
Protection according IP54
Mounting Standard stem, optional panel, surface
Weight dry in kg 0.28 0.70 1.94

All specifications are subject to change without notice.


Page 3
IN.PG121 Rev. G 19/11/201
General dimensions in mm

Order information
Size System Execution Process Connection Range Engineering Case Material Options
material connection orientation units
(100) (B) Tube (D) dry (04)1/2” NPT (M) 2) (L) Lower -30”/ 0 (psi & kg/cm2) (=) Standard (NH) Tagging wired
100mm bronze, IP54 -30”/ 15 Dual scale Steel case
socket (03)3/8” NPT (M) (C) Centre -30”/ 30 “Hg / psi & black finish (DA) Dial Marking
(150) brass3) back1) -30”/ 60 mmHg / kg/cm2
(02)1/4” NPT (M) -30”/ 100 (ON) Neutral, without
150mm
2) -30”/ 150 logo
(15)1/2” BSP (M) -30”/ 200
(250) (FX) Front panel 1)
250mm (14)3/8” BSP (M) 0/ 10 (psi & kg/cm2)
0/ 15 Dual scale (FW) Surface plate
(13)1/4” BSP (M) 0/ 20 psi & kg/cm2
0/ 30
(16)M20x1.5 (M) 2) 0/ 50
0/ 60
0/ 100
0/ 150
0/ 200
0/ 250
0/ 300
0/ 400
0/ 500
0/ 600
0/ 800
0/ 1000
0/ 1500
0/ 2000
0/ 3000
0/ 4000
0/ 5000
0/ 6000
0/ 8000
0/ 10000
1) Only for dial size 100mm 2) only for dial size 250mm 3) for ranges above 1000 psi tube material SS316
How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
2
100 PG121 B D 02 C 0/60 psi & kg/cm = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 4
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Liquid filled gauges
Model GFB
Nominal size 63mm, 100mm
Accuracy: Class 1.6 (1.6%)

Features
x Twin o-ring design for leak tight enclosure
x Protection IP65
x Usable to full scale
x Overload protection 130 %
x Measuring tubes SS316 / PB
x Minimized pulsation, Vibration and pointer flittering due to
dampening effect

Ranges
2 2
-1 ... 0 kg/cm up to 0 ... 400 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 6,000 psi

Applications
Power packs
Hydraulic machinery
Pumps
Other vibration applications

Technical specification GFB


Dial size in mm 63 100
Construction Open front cylindrical case with crimp type bezel ring
Measuring principle Bourdon tube
Range in kg/cm2 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400
-1/0 -1/0.6 -1/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/8” 1/4” BSP / NPT (M) 1/4” 3/8” 1/2” BSP / NPT (M)
Connection location Lower, back
Material
Pressure connection Brass
Tube SS 316/ Phosphorous Bronze below 6 kg/cm2, Brass
Case/bayonet ring Stainless steel 304
Window Acrylic glass
Dial Aluminum, black markings on white background
Pointer Plastic, Black Aluminum, black
Movement Brass
Accuracy Class 1.6 (1.6% F.S)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP65
Filling liquids Glycerin, silicone, others on request
Mounting Standard stem, optional panel, bracket, surface or others on request
Weight with liquid filling in kg 0.22 0.48

All specifications are subject to change without notice.


Page 5
IN.GFB Rev. F 21/11/2016
General dimensions in mm

Order information
Size Type System Executio Process Connection Range Engineering Filling Options
material n connection orientation units material
(63) GFB (B) Socket (L) IP 65 (04) 1/2” NPT (M)1) (L) Lower -1/ 02) (kg/cm2) (=) No filling (NH) Tagging wired
63 mm Brass, -1/ 0.62)
Tube (03) 3/8” NPT (M) 1)
(C) Center -1/ 1.52) (GV) Silicone (DA) Dial Marking
(100) SS316 2) back -1/ 32)
(02) 1/4” NPT (M) (GR) Glycerin (CS) Dual Scale
100 mm -1/ 52)
-1/ 9 (FX) Front panel
(01) 1/8” NPT (M) -1/ 15
-1/ 24 (FW) Surface plate
(15) 1/2” BSP(M)1)
0/ 12)
(14) 3/8”BSP(M)1) 0/ 1.62) (UF) U-clamp
0/ 2.52)
(13) 1/4” BSP(M) 0/ 42)
0/ 62)
(12) 1/8” BSP(M) 0/ 10
0/ 16
(16) M20x1.5(M)1) 0/ 25
0/ 40
0/ 60
0/ 100
0/ 160
0/ 250 psi and others
Others on request 0/ 400 on request
1) Not for 63mm 2) Tube material Phosphorous Bronze for range below 6 kg/cm2

How to order
Size Type System Execution Process Connection Range Engineering Filling/case Option
material connection orientation unit material
63 GFB B L 02 L 0/16 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 6
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Industrial Pressure gauges
Model GES
Nominal size 100 mm,
Accuracy: Class 1 (1%)

Features
x Rugged stainless steel construction
x Socket and case welded
x Protection IP65
x Overload protection 130 %
x Dry or liquid filled
x Measuring system stainless steel

Ranges
2 2
-1 ... 0 kg/cm up to 0 ... 700 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 10,000 psi

Applications
Power packs
Hydraulic machinery
Pumps
Power plants
Pipelines
Transmission stations

Technical specification GES


Dial size in mm 100
Construction Open front cylindrical case with crimp type bezel ring
Measuring principle Bourdon tube
Range in kg/cm2 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400 600
-1/0 -1/0.6 -1/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/4” 3/8” BSP / NPT (M)
Connection location Lower, back
Material
Pressure connection Stainless steel 304
Tube Stainless steel 316
Case/bayonet ring Stainless steel 304
Window Acrylic glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black
Accuracy Class 1 (1 % F.S.)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP65
Filling liquids Glycerin, silicone
Mounting Standard stem, optional panel, bracket, surface or others on request
Accessories, options Diaphragm seals, valves, dampener, gauge saver, etc

All specifications are subject to change without notice.


Page 7
IN.GES Rev. A 27/11/2016
General dimensions in mm

Order information
Size Type System Execution Process Connection Range Engineering Filling liquids Options
material connection orientation units
(100) GES (A) Socket (L) IP 65 (03) 3/8” NPT (M) (L) Lower -1/ 0 (kg/cm2) (=) No filling (NH) Tagging wired
100 mm 304, -1/ 0.6
Tube (02) 1/4” NPT (M) (B) Back -1/ 1.5 (GV) Silicone (CS) Dual Scale
316 -1/ 3
(14) 3/8”BSP(M) (GR) Glycerin (FX) Front panel
-1/ 5
-1/ 9 (FW) Surface plate
(13) 1/4” BSP(M) -1/ 15
-1/ 24 (UF) U-clamp
0/ 1
0/ 1.6
0/ 2.5
0/ 4
0/ 6
0/ 10
0/ 16
0/ 25
0/ 40
0/ 60
0/ 100
0/ 160 and other
0/ 250 equivalent
0/ 400 ranges in PSI,
0/ 600 Bar, kPa, MPa,
etc.

How to order
Size Type System Execution Process Connection Range Engineering Filling liquids Option
material connection orientation unit
100 GES A L 14 L 0/16 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 8
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Stainless steel liquid filled gauges
Model GFS
Nominal size 63mm, 100 mm,
Accuracy: Class 1.6 (1.6%) for 63 mm, Class 1 (1%) for 100 mm

Features
x Rugged stainless steel construction
x Socket and case welded
x Protection IP65
x Usable to full scale
x Overload protection 130 %
x Dry or liquid filled
x Measuring system stainless steel

Ranges
2 2
-1 ... 0 kg/cm up to 0 ... 1600 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 23,000 psi

Applications
Power packs
Hydraulic machinery
Pumps
Power plants
Pipelines
Transmission stations

Technical specification GFS


Dial size in mm 63 100
Construction Open front cylindrical case with crimp type bezel ring
Measuring principle Bourdon tube
Range in kg/cm2 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400 600 1000 1600
-1/0 -1/0.6 -1/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/8” 1/4” BSP / NPT (M) 1/4” 3/8” 1/2” BSP / NPT (M), M20x1.5(M)
Connection location Lower, back
Material
Pressure connection Stainless steel 304, 316, 316L
Tube Stainless steel 316, optional 316L
Case/bayonet ring Stainless steel 304 , optional 316L
Window Acrylic glass, optional laminated safety glass
Dial Aluminum, black markings on white background
Pointer Plastic, Black |Aluminum, black
Movement Stainless steel 304
Accuracy Class 1.6 (1.6% F.S) Class 1 (1 % F.S.)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP65
Conformity according to RL 94/9/EC
Optional, Marking CE EX II 2 GD c
appendix X for mechanical equipment
File no. XXXXXXXX at notified body 0044,
in potential hazardous areas
TUV NORD CERT (only with safety glass)
(for 63mm only)
Filling liquids Glycerin, silicone, halocarbon, others on request
Mounting Standard stem, optional panel, bracket, surface or others on request
Weight with liquid filling in kg 0.21 0.75
Accessories, options Diaphragm seals, valves, dampener, gauge saver, etc

All specifications are subject to change without notice.


Page 9
IN.GFS Rev. H 21/11/2016
General dimensions in mm

Order information
Size Type System Executi Process Connection Range Engineering Filling/case Options
material on connection orientation units material
(63) GFS (A) Socket (L) IP 65 (04) 1/2” NPT (M)1) (L) Lower -1/ 0 (kg/cm2) (=) No filling (NH) Tagging wired
63 mm 304, 1)
-1/ 0.6
Tube (03) 3/8” NPT (M) (B) Back -1/ 1.5 (GV) Silicone (DA) Dial Marking
(100) 3161) -1/ 3
100 mm (02) 1/4” NPT (M) (C) Center -1/ 5 (GR) Glycerin (CS) Dual Scale
2)
back -1/ 9
(S) Socket (01) 1/8” NPT (M) -1/ 15 (GX) Halocarbon (OF) Oil free
316, -1/ 24
Tube (15) 1/2” BSP(M)1) (YW) Case (SG) Safety glass2)
0/ 1
316 material
(14) 3/8”BSP(M)1) 0/ 1.6 (FX) Front panel
316L
0/ 2.5
(1.4404)
(13) 1/4” BSP(M) 0/ 4 (FW) Surface plate
(L) Socket 0/ 6
and (12) 1/8” BSP(M) 0/ 10 (UF) U-clamp
Tube 0/ 16
316L (16) M20x1.5(M) 1)
0/ 25 (ATX) ATEX listed3)
0/ 40
0/ 60
0/ 100
0/ 160
0/ 250
and other
0/ 400 equivalent
0/ 600 ranges in PSI,
0/ 1000 Bar, kPa, MPa,
0/ 16001) etc.
1) Not for 63mm 2) For 63 mm only 3) Safety glass(SG) mandatory for Ex-proof and for 63mm only
How to order
Size Type System Execution Process Connection Range Engineering Filling/case Option
material connection orientation unit material
63 GFS S L 02 L 0/16 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91- 2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 10
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Stainless steel process gauges
Model WPS
Nominal size 63 mm, 100 mm, 115 mm, 150 mm or 250 mm
Accuracy: Class 1(1%) / Class 1.6 (1.6%) for 63mm

Features
x Rugged stainless steel construction
x Measuring system stainless steel or Monel
x Socket and case welded
x Protection IP54, IP65, IP66 and IP67
x Usable to full scale
x Overload protection 130 %
x Dry, liquid filled
x Optional contacts
x Optional external Zero Adjustment

Ranges
2 2
-1 ... 0 kg/cm up to 0 ... 2500 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 36,000 psi

Applications
Chemical & petrochemical industries, refineries
Fertilizer, thermal & nuclear power plants
OEM & machine apparatus

Technical specification WPS


Dial size in mm 63 100 115 150 250
Construction Open front cylindrical case with snap action bayonet ring with blow out disc
Measuring principle Bourdon tube
Range in kg/cm2 0.6 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400 600 1000 1600 2500
-1/0 -1/0.6 -1/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/8” 1/4” 3/8” 1/2” BSP / NPT (M) , M20x1.5 (M)
Others on request
Connection location Lower, back
Material
Pressure connection Stainless steel 316, optional 316L, Monel
Tube Stainless steel 316, optional 316L, Monel
Case/bayonet ring Stainless steel 304 , optional 316L
Window Instrument glass, optional laminated safety glass, acrylic glass
Dial Aluminum, black markings on white background, others on request
Pointer Aluminum, black optional micrometer adjustment
Movement Stainless steel 304
Accuracy Class1.6(1.6% F.S) Class 1 (1 % F.S.) Class1 (1% F.S.), optional 0.5% F.S.
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 200 °C (dry), max. 100 °C (for liquid filled)
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP54/IP65/IP66/IP67
Conformity according to RL 94/9/EC Optional, Marking CE EX II 2 GD c
appendix X for mechanical File no. XXXXXXXX at notified body
equipment in potential hazardous 0044, TUV NORD CERT
areas (for 63mm only) (only with safety glass)
Filling liquids Glycerin, silicone, halocarbon, others on request
Mounting Standard stem, optional panel, bracket, surface or others on request
Weight dry/filled in kg 0.1/0.17 0.51 / 0.86 0.53/0.88 0.88 / 1.78 2.02 /3.52
Accessories, options NACE, diaphragm seals, valves, electric contacts, dampener, gauge saver, syphon etc

All specifications are subject to change without notice.


Page 11
IN.WPS Rev. N 21/11/2016
General dimensions in mm

Order information
Size Type System Executi Process Connection Range Engineer- Filling/case Options
material on connection orientation ing units material
(63) WPS (S) Socket (D) IP54 (04)1/2” NPT (M)1) (L) Lower -1/ 0 (kg/cm2) (=) Standard (NH) Tagging wired
63mm 316, -1/ 0.6 no filling
Tube (L) IP65 (03)3/8” NPT (M)1) (B) Back -1/ 1.5 (DA) Dial Marking
(100) 316 -1/ 3 (GV) Silicone5)
(R) IP66 (02)1/4” NPT (M) (C) Centre (CS) Dual Scale
100 mm -1/ 5
(L) Socket back 2)
-1/ 9 (GR) Glycerin5) (OF) Oil free
(115) and (N) IP67 (01)1/8” NPT (M) -1/ 15
115mm Tube -1/ 24 (GX) Halocar- (MP) Micrometer
(15)1/2” BSP (M)1) bon5) pointer1)
316L
(150) (P) Socket 0/ 0.61)
150mm (14)3/8” BSP(M)1) (YW) Case (PD) Acrylic glass
and 0/ 1
material (SG) Safety glass
Tube (13)1/4” BSP(M) 0/ 1.6
(250) 316L
Monel1) 0/ 2.5
250mm (1.4404) (FX) Front panel
(12)1/8” BSP(M) 0/ 4
0/ 6 (EC) Contact (FW) Surface plate
(16)M20x1.5(M) 1) 0/ 10 type and
0/ 16 (UF) U-clamp
function
(KQ)1/2” BSPT (M)1) 0/ 25 (see (TU) Throttle screw
0/ 40 EC.1)1)3) 5)
(KR)3/8” BSPT (M)1) 0/ 60 (EZ) External Zero
(KA)1/4” BSPT (M)
0/ 100 adjustment1)6)
0/ 160
0/ 250 (AJ) Calibration 0.5 %
0/ 400 F.S.4)
0/ 600
0/ 1000 (PR) Receiver gauge 1)
0/ 16001) (ATX) ATEX listed 7)
0/ 25001) psi and
others on
request
1) Not for size 63mm 2) For 63mm only 3) For dry gauges only 4) For 150 mm, 250 mm 5) Not for 0.6 kg/cm2 6) not for 250 mm
7) Safety glass (SG) mandatory for Ex-proof and for 63mm only

How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
150 WPS S L 04 L 0/16 KG/CM2 NA NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 12
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Stainless steel process gauges
Model ZPS
Nominal size 100 mm, 115 mm, 150 mm
Accuracy: Class 1(1%)

Features
x Rugged stainless steel construction
x Socket and case welded
x Protection IP65, IP66 and IP67
x Usable to full scale
x Overload protection 130 %
x Dry, liquid filled
x Optional contacts
x Optional external Zero Adjustment

Ranges
2 2
-1 ... 0 kg/cm up to 0 ... 1000 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 15,000 psi

Applications
Chemical & petrochemical industries, refineries
Fertilizer, thermal & nuclear power plants
OEM & machine apparatus

Technical specification ZPS


Dial size in mm 100 115 150
Construction Open front cylindrical case with snap action bayonet ring with blow out disc
Measuring principle Bourdon tube
Range in kg/cm2 0.6 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400 600 1000
-1/0 -1/0.6 -1/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/8” 1/4” 3/8” 1/2” BSP / NPT (M) , M20x1.5 (M)
Others on request
Connection location Lower, back
Material
Pressure connection Stainless steel 316, optional 316L
Tube Stainless steel 316, optional 316L
Case/bayonet ring Stainless steel 304 , optional 316L
Window Toughened glass, optional Safety Glass, acrylic glass
Dial Aluminum, black markings on white background, others on request
Pointer Aluminum, black optional micrometer adjustment
Movement Stainless steel
Accuracy Class 1 (1 % F.S.)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 200 °C (dry), max. 100 °C (for liquid filled)
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP65/IP66/IP67
Filling liquids Glycerin
Mounting Standard stem, optional panel, bracket, surface or others on request
Weight dry/filled in kg 0.51 / 0.86 0.64/0.99 0.88 / 1.78
Accessories, options NACE, diaphragm seals, valves, dampener, gauge saver, syphon etc

All specifications are subject to change without notice.


Page 13
IN.ZPS Rev. F 26/11/2016
General dimensions in mm

Order information
Size Type System Execution Process Connection Range Engineering Filling/case Options
material Connection orientation units material
(100) ZPS (S) Socket (L) IP65 (04)1/2” NPT (M) (L) Lower -1/ 0 (kg/cm2) (=) Standard (NH) Tagging wired
100 mm 316, -1/ 0.6 no filling
Tube (R) IP66 (03)3/8” NPT (M) (B) Back -1/ 1.5 (DA) Dial Marking
(115) 316 -1/ 3 (GV) Silicone2)
115mm (N) IP67 (02)1/4” NPT (M) -1/ 5 (CS) Dual Scale
(L) Socket -1/ 9 (GR) Glycerin2)
(150) and (01)1/8” NPT (M) -1/ 15 (OF) Oil free
150mm Tube -1/ 24 (GX) Halocar-
316L
(15)1/2” BSP (M)
0/ 0.6 bon2) (MP) Micrometer
pointer
(14)3/8” BSP(M) 0/ 1
(YW) Case
0/ 1.6 (PD) Acrylic glass
material
(13)1/4” BSP(M) 0/ 2.5
316L
0/ 4 (SG) Safety Glass
(1.4404)
(12)1/8” BSP(M) 0/ 6
0/ 10 (FX) Front panel
(16)M20x1.5(M) 0/ 16
0/ 25 (FW) Surface plate
(KQ)1/2” BSPT (M) 0/ 40
0/ 60 (UF) U-clamp
(KR)3/8” BSPT (M) 0/ 100
0/ 160 (TU) Throttle screw
(KA)1/4” BSPT (M) 0/ 250
0/ 400 (EZ) External Zero
0/ 600 adjustment
psi and others
0/ 1000 on request
1) Specify details 2) Not for 0.6 kg/cm2

How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
100 ZPS S L 04 L 0/16 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 14
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All stainless steel safety process gauges
Model SFS
According DIN EN 837-1
Nominal size 100 mm, 115 mm, 150 mm
Accuracy: Class 1(1%)

Features
x Solid front type construction
x Rugged stainless steel construction
x Measuring system stainless steel or Monel
x Socket and case welded
x Protection IP54, IP65 and IP66
x Usable to full scale
x Overload protection 130 %
x Dry, liquid filled
x Optional external Zero Adjustment

Ranges
2
-1 ... 0 bar up to 0 ... 1400 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 20,000 psi

Applications
Chemical, petrochemical industry, petroleum refinery
Fertilizer, thermal & nuclear power generation industry
Machine and apparatus construction

Technical specification SFS


Dial size in mm 100 115 150
Construction Solid front safety pattern cylindrical case with blowout at the back
Measuring principle Bourdon tube
Range in kg/cm2 0.6 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400 600 1000 1400
-1/0 -1/0.6 -0/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/4” 3/8” 1/2” BSP / NPT (M) , M20x1.5 (M),
Others on request
Connection location Lower, back(for 150 mm only)
Material
Pressure connection Stainless steel 316, optional 316L, Monel
Tube Stainless steel 316, optional 316L, Monel
Case/bayonet ring Stainless steel 304, optional 316L
Window laminated safety glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black optional - micrometer adjustment others on request
Movement Stainless steel 304
Accuracy Class 1 (1 % F.S.), optional 0,5 % F.S.
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 200 °C (dry), max. 100 °C (for liquid filled)
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP54/IP65/IP66
Filling liquids Glycerin, silicone, halocarbon, others on request
Mounting Standard stem, optional flush, surface
Weight dry/filled in kg 0.63 / 1.02 0.76 / 1.65 1.14 / 2.08
Accessories, options NACE, diaphragm seals, valves, electric contacts, dampener, gauge saver, syphon etc

All specifications are subject to change without notice.


Page 15
IN.SFS Rev. J 26/11/2016
General dimensions in mm

Order information
Size Type System Execution Process Connection Range Engineering Filling/Case Options
material connection orientation units material
(100) SFS (S) Socket (D) IP54 (04)1/2” NPT (M) (L) Lower -1/ 0 (kg/cm2) (=) Standard (NH) Tagging wired
100 mm 316, -1/ 0.6 no filling
Tube (L) IP65 (03)3/8” NPT (M) (B) Back2) -1/ 1.5 (DA) Dial Marking
(115) 316 -1/ 3 (GV) Silicone1)
115mm (02)1/4” NPT (M) -1/ 5 (CS) Dual scale
(L) Socket -1/ 9 (GR) Glycerin1)
(150) and (15)1/2” BSP (M) -1/ 15 (OF) Oil free
150mm Tube -1/ 24 (GX) Halocar-
316L (14)3/8” BSP(M) 0/ 0.6 bon1) (MP) Micrometer
(P) Socket 0/ 1 pointer
and (13)1/4” BSP(M) 0/ 1.6 (YW) Case
Tube 0/ 2.5 material (EZ) External Zero
Monel (16)M20x1.5(M) 0/ 4 316L adjustment
0/ 6 (1.4404)
0/ 10 (FX) Front panel
0/ 16 (EC) Contact
0/ 25 type and (FW) Surface plate
0/ 40 function
0/ 60 (see (TU) Throttle screw
0/ 100 EC.1)1)
0/ 160 (AJ) Calibration 0.5 %
0/ 250
F.S.2)
0/ 400
0/ 600
(AV) Anti-Vibration3)
0/ 1000
0/ 1400
psi & others on
Others on request request Others on request
1) Not for range 0.6 kg/cm² 2) For size 150mm and range up to 0/700 kg/cm2 only
3) For pressure ranges ≥ 2.5 kg/cm2 & available with limited options, Halocarbon filling not allowed & consult engineering.

How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
100 SFS S L 04 L 0/16 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91- 2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 16
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Stainless steel test gauge open front or solid front
Model TGS
Nominal size 150 mm or 250 mm
Accuracy: 0.5%

Features
x Rugged stainless steel construction
x Protection IP65
x Usable to full scale
x Overload protection 130 %
x Dry, liquid filled
x Measuring system stainless steel all welded
x Optional external Zero Adjustment

Ranges
2 2
-1 ... 0 kg/cm up to 0 ... 1000 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 15,000 psi

Applications
Chemical & petrochemical industries
Machine and apparatus construction
Food and beverage industries
Pulp and paper industries

Technical specification TGS


Dial size in mm 150 250
Construction Open front cylindrical case with snap action bayonet ring, optional solid front for 150mm
Measuring principle Bourdon tube
2
Range in kg/cm 0.6 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400 600 1000
-1/0 -1/0.6 -1/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/4” 3/8” 1/2” BSP / NPT (M) , M20x1.5 (M)
Others on request
Connection location Lower, back
Material
Pressure connection Stainless steel 316
Tube Stainless steel 316L
Case/bayonet ring Stainless steel 304 optional 316L
Window Instrument glass, optional laminated safety glass, acrylic glass
Dial Aluminum, black markings on white background, with polished mirror band
Pointer Aluminum, black knife edge, optional - micrometer adjustment others on request
Movement Brass and Stainless steel 304
Accuracy Class 0.5 (0.5% F.S)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP65
Filling liquids Glycerin, silicone, halocarbon, others on request
Mounting Standard stem, optional panel, bracket, surface or others on request
Weight dry/filled in kg 0.87 / 1.68 1.99 / 4.09
Accessories, options Valves

All specifications are subject to change without notice.


Page 17
IN.TGS Rev. I 26/11/2016
General dimensions in mm

Order information
Size Type System Execution Process Connection Range Engineer- Filling/case ma- Options
material connection orientation ing units terial
(150) TGS (S) Socket (=) IP65 (04)1/2” NPT (M) (L) Lower -1/ 0 (kg/cm2) (=) Standard (NH) Tagging wired
150mm 316, -1/ 0.6 no filling
Tube (03)3/8” NPT (M) (B) Back -1/ 1.5 (DA) Dial Marking
(250) 316L -1/ 3 (GV) Silicone
250mm3) (02)1/4” NPT (M) -1/ 5 (CS) Dual Scale
-1/ 9 (GR) Glycerin
(15)1/2” BSP (M) -1/ 15 (OF) Oil free
-1/ 24 (GX) Halocar-
(14)3/8” BSP(M) bon (PD) Acrylic glass
0/ 0.62)
(13)1/4” BSP(M) 0/ 1
(YW) Case ma- (MP) Micrometer
0/ 1.6
terial 316L pointer
(16)M20x1.5(M) 0/ 2.5
0/ 4
(EZ) External Zero
0/ 6
0/ 10 adjustment
0/ 16
0/ 25 (SG) Safety glass
0/ 40
(FX) Front panel
0/ 60
0/ 100 (FW) Surface plate
0/ 160
0/ 250 (UF) U-clamp3)
0/ 400
0/ 600
psi and oth-
(SF) Solid front1)
0/ 1000 ers on re-
quest
1) Not for 250mm 2) Not for liquid filled 3) Not for solid front (SF)

How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
150 TGS S = 04 L 0/16 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 18
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Stainless steel master gauge open front or solid front
Model MGS
Nominal size 150 mm or 250 mm
Accuracy: Class 0.25 (DIN)

Features
x Rugged stainless steel construction
x Protection IP65
x Usable to full scale
x Easy access for calibration
x Ambient temperature compensation
x Optional external Zero Adjustment

Ranges
2 2
-1 ... 0 kg/cm up to 0 ... 1000 kg/cm
-30 in. Hg ... 0 psi up to 0 ... 15,000 psi

Applications
Test benches
Calibration laboratories
Maintenance and repair shop

Technical specification MGS


Dial size in mm 150 250
Construction Open front cylindrical case with snap action bayonet ring, optional Solid front for 150mm
Measuring principle Bourdon tube
Range in kg/cm2 0.6 1 1.6 2.5 4 6 10 16 25 40 60 100 160 250 400 600 700 1000
-1/0 -1/0.6 -1/1.5 -1/3 -1/5 -1/9 -1/15 -1/24
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/4” 3/8” 1/2” BSP / NPT (M) , M20x1.5 (M)
Others on request
Connection location Lower, back
Material
Pressure connection Stainless steel 316, optional Monel
Tube Beryllium copper, optional Monel
Case/bayonet ring Stainless steel 304, optional 316L
Window Instrument glass, optional laminated safety glass, acrylic glass
Dial Aluminum, black markings on white background, with polished mirror band
Pointer Aluminum, black knife edge, optional - micrometer adjustment others on request
Movement High precision jewel bearing with temperature compensation
Accuracy Class 0.25 (0.25% F.S)
Permissible
Ambient temperature -25 ... 60 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Automatic compensated
Protection according IP65
Mounting Standard stem, optional panel, bracket, surface or others on request
Weight dry/filled in kg 0.87 / 1.68 1.99 / 4.09
Accessories, options Valves

All specifications are subject to change without notice.


Page 19
IN.MGS Rev .I 28/11/2016
General dimensions in mm

Order information
Size Type
System Execution Process Connection Range Engineering Case material Options
material connection orientation units
(150) MGS (A) Socket (=) IP65 (04)1/2” NPT (M) (L) Lower -1/ 0 (kg/cm2) (=) Standard (NH) Tagging wired
150mm 316, -1/ 0.6
Tube (03)3/8” NPT (M) (B) Back -1/ 1.5 (YW) Case mate- (DA) Dial Marking
(250) beryllium -1/ 3 rial 316L
250mm copper (02)1/4” NPT (M) -1/ 5 (CS) Dual Scale
(M) Socket & -1/ 9
Tube (15)1/2” BSP (M) -1/ 15 (OF) Oil free
Monel -1/ 24
(14)3/8” BSP(M) (PD) Acrylic glass
0/ 0.6
(13)1/4” BSP(M) 0/ 1 (MP) Micrometer
0/ 1.6
pointer
(16)M20x1.5(M) 0/ 2.5
0/ 4
(EZ) External Zero
0/ 6
0/ 10 adjustment
0/ 16
0/ 25 (SG) Safety glass
0/ 40
(FX) Front panel
0/ 60
0/ 100 (FW) Surface plate
0/ 160
0/ 250 (UF) U-clamp2)
0/ 400
0/ 600 (SF) Solid front1)
0/ 700
0/ 1000 psi and others
on request
1) Not for 250mm 2) not for solid front (SF)
How to order
Size Type System Execution Process Connection Range Engineering Case material Option
material connection orientation unit
150 MGS A = 04 L 0/16 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 20
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm type pressure gauge
Model SDS
Nominal size 100 mm or 150 mm
Accuracy: Class 1.6 (1.6%)

Features
x Stainless steel construction
x Suitable for corrosive environments
x Protection IP54
x Usable to full scale
x Overload protection 130%

Ranges
2
-25 ... 0 mbar up to 0 ... 2,5 kg/cm
-250 ... 0 mmWC up to 0 ... 30 psi

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification SDS


Dial size [mm] 100 150
Construction Open front cylindrical case, optional with blow out disc at the back
Measuring principle Horizontal diaphragm
Range [mmwc] 250 400 600 1000 1600 2500 4000 -250/0 -400/0 -600/0 -1000/0 -1600/0
-2500/0 -4000/0
2
[kg/cm ] 0,6 1,0 1,6 2,5 -1/0 -1/0,6 -1/1,5
Overpressure limit 130 % F.S., short time.
Pressure type Gauge, vacuum and compound
Process connection 1/4” 3/8” 1/2” BSP / NPT (M), M20x1.5(M)
Direct flange as per DIN/ANSI
Others on request
Connection location Lower
Material
Process connection Stainless steel 316 optional Stainless steel 316L , PTFE lined
Diaphragm Stainless steel 316L optional PTFE protected
Case/bayonet ring Stainless steel 304 optional 316L
Window Instrument glass optional laminated safety glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black, optional micrometer adjustment
Movement Stainless steel 304
Accuracy Class 1.6 (1.6 % F.S.)
Permissible
Ambient temperature -25 ... 50 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,5 % / 10 K
Protection according IP54, Optional IP65
Mounting Standard stem or flange
Weight d 600 mbar / > 600 mbar [kg] 1.66 / 1.26 1.99 / 1.59
Accessories, options Valves, dampener, gauge saver, etc.

All specifications are subject to change without notice.


Page 21
IN.SDS Rev. J 28/11/2016
General dimensions in mm

Order information
Size Type System Process Rating Facing Range Engineer- Blow out/case Options
material connection ing units material
(100) SDS (S) Diaphragm (04) 1/2” NPT (M) (NA) (T) -250/ 0 (mmWC) (=) Standard (NH) Tagging
100 mm 316L, 2) -400/ 0 without wired
flange 316 (03) 3/8” NPT (M) -600/ 0 blowout
(150) (02) 1/4” NPT (M) 2) -1000/ 0 (CS) Dual scale
150 mm (L) Diaphragm -1600/ 0 (IP) Ingress
316L, (15) 1/2” BSP (M) -2500/ 0 protection (OF) Oil free
flange -4000/ 0 IP65
(14) 3/8” BSP(M) 2) 0/ 250
316L (MP) Micrometer
(13) 1/4” BSP(M) 2) 0/ 400
(RB) Blow out pointer
0/ 600
(SP) Diaphragm (16) M20x1.5(M) 0/ 1000 disc
316L & 0/ 1600 (SG) Safety
flange 316, ANSI Flanged (150) 150 lbs (FF) Flat 0/ 2500 (YW) Case glass
PTFE face 1) 0/ 4000 material
protected (05) 1/2” (300) 300 lbs (RF) Raised
316L (DA) Dial
(07) 3/4” (600) 600 lbs face 1) -1/ 0 (1.4404) marking
-1/ 1,5 (kg/cm2)
(10) 1” 0/ 0,6
(15) 1 ½” 0/ 1
0/ 1,6
(20) 2” 0/ 2,5
DIN Flanged (PN10)
(DN15) (PN16)
(DN25) (PN25)
(DN40)
(DN50)
psi and others
on request
1) The DIN-flange facing depends on nominal size and nominal pressure 2) Not for system material (SP)
How to order
Size Type System Process Rating Facing Range Engineering Filling/Case Option
material connection unit material
100 SDS S 04 NA T 0/400 MMWC = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 22
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Bellow differential pressure gauge
Model D100
Nominal size 150 mm
Accuracy: Class 1.6(1.6%) optional class 1(1%)

Features
x Stainless steel case and wetted parts
x Static pressure 10 times of the range
x Protection IP65
x High corrosive resistance
x Dry or liquid filled

Ranges
0 ... 2500 mmWC up to 0 ... 10 kg/cm2

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification D100


Dial size [mm] 150
Construction Open front cylindrical case with bayonet ring
Measuring principle Twin bellow
Range [mmWC] 2500 4000
2
[kg/cm ] 0,6 1 1.6 2.5 4 6 10
Max. static pressure 10 times of the range. Static pressure is also maximum pressure allowed on one side
Pressure type Differential
Process connection 1/4” 1/2” NPT (F), 1/4” 3/8” 1/2” BSP / NPT (M), M20x1.5(M), others on request
Connection location Lower, 90° lower, back
Material
Pressure connection Stainless steel 316L
Measuring bellow Stainless steel 316L
Case/bayonet ring Stainless steel 304, optional 316L
Window Laminated safety glass optional acrylic glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black, optional Micrometer Adjustment
Accuracy calibrated ascending Class 1.6 (1.6 % F.S.) optional 1.0 (1.0 % F.S.)
Permissible
Ambient temperature -20 ... 80 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection IP65
Filling liquids Glycerin, silicone, others on request
Mounting Wall mounting, optional 2" pipe mounting, others on request
Weight dry/filled [kg] 1.15 / 2.05
Accessories, options 3 or 5 way remote or direct mounted manifolds, valves

All specifications are subject to change without notice.


Page 23
IN.D100 Rev. I 03/02/2017
General dimensions in mm

Order information
Size Type
System Execution Process Connection Range Engineer- Filling/case Options
material connection orientation ing units material
(150) D100 (S) 316L (L) IP65 (25) 1/4” NPT (F) (L) Lower 0/2500 (mmWC) (=) Standard no (NH) Tagging wired
150mm process (N) 90° 0/4000 filling
connecti (50) 1/2” NPT (F) (DA) Dial marking
Lower
on and 0/ 0,6 (kg/cm2) (GV) Silicone
bellow (04)1/2” NPT (M) 0/ 1 (CS) Dual scale
(B) Back (GR) Glycerin
0/ 1,6
(03)3/8” NPT (M) 0/ 2,5 (OF) Oil free
0/ 4 (GX) Halocarbon
(02)1/4” NPT (M) (MP) Micrometer pointer
0/ 6
(YW) Case
(15)1/2” BSP (M) 0/ 10 (FX) Front panel
material
316L
(14)3/8” BSP(M) (FW) Surface plate

(13)1/4” BSP(M) (TM) 2” pipe mounting


psi and
(16)M20x1.5(M) others on (AN) Calibration 1 % F.S.
request

compound
ranges on
request
1) Specify details

How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
150 D100 S L 04 L 0/4000 MMWC = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 24
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Single diaphragm differential pressure gauge
Model D101
Max. static pressure 25 bar
Nominal size 100 mm, 115mm or 150 mm
Accuracy: Class 2.5(2.5%), optional 1.6(1.6%)

Features
x Stainless steel case and wetted parts
x Static pressure 25 bar
x Protection IP65
x Dry or liquid filled

Ranges
0 ... 600 mmWC up to 0 ... 2.5 kg/cm2

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification D101


Dial size [mm] 100 115 150
Construction Cylindrical case with bayonet ring.
Measuring principle Diaphragm, safety design, solid front not required
Range [mmWC] 600 1000 1600 2500 4000 5000
2
[kg/cm ] 0,6 1 1.6 2.5
2
Max. static pressure [kg/cm ] 25 bidirectional only
Overload capability 130 % F.S., short time
Pressure type Differential
Process connection 1/4” 1/2” NPT (F), 1/4” 3/8” 1/2” BSP / NPT (M), M20x1.5(M), others on request
Connection location Lower
Material
Pressure connection Stainless steel 316
Pressure chamber Stainless steel 316, Viton-ring
Measuring membrane Stainless steel 316L
Case/bayonet ring Stainless steel 304, optional 316L
Window Laminated safety glass optional acrylic glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black, optional Micrometer Adjustment
Movement Stainless steel 304
Accuracy calibrated ascending Class 2.5 (2.5 % FS.) Optional 1.6% FS.
Permissible
Ambient temperature -20 ... 80 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection IP65
Filling liquids Glycerin, silicone, others on request
Mounting Wall mounting, optional 2" pipe mounting, others on request
Weight dry/filled [kg] 8.1/8.4 8.2/8.6 8.4/9.0
Accessories, options 3 or 5 way remote or direct mounted manifolds

All specifications are subject to change without notice.


Page 25
IN.D101 Rev. E 21/01/2017
General dimensions in mm

Order information
Size Type System Execution Process Connection Range Engineer- Filling/Case Options
material Connection orientation ing units material
(100) D101 (S) Pressure (L) IP65 (02T) 1/4” NPT (F) (L) Lower 0/ 600 (mmWC) (=) Standard (NH) Tagging wired
100 mm chamber 0/ 1000 no filling
316 and (04T) 1/2” NPT (F) 0/ 1600 (DA) Dial Marking
(115) diaphragm 0/ 2500 (GV) Silicone
115mm 316L (04) 1/2” NPT (M) 0/ 4000 (CS) Dual Scale
0/ 5000 (GR) Glycerin
(150) (03) 3/8” NPT (M) (OF) Oil free
150mm 0/ 0,6 (kg/cm2) (GX) Halocar-
(02) 1/4” NPT (M) bon (MP) Micrometer
0/ 1
pointer
(15) 1/2” BSP (M) 0/ 1,6
(YW) Case
0/ 2,5 (PD) Acrylic glass
material
(14) 3/8” BSP(M) 316L
(FW) Wall mounting
(13) 1/4” BSP(M) bracket
(16) M20x1.5(M) (TM) 2” pipe mounting

(AL) Calibration
1.6%1)
psi and
Standard (02T) others others on
through adaptor request
1) Consult Engineering, limited to certain ranges.

How to order
Size Type System Execution Process Connection Range Engineering Filling/Case Option
material connection orientation unit material
100 D101 S L 02T L 0/600 mmWC = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 26
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm type differential pressure gauge
Model D104
Max. Static pressure 100 bar
Nominal size 100 mm or 150 mm
Accuracy: Class 1,6 optional Class 1

Features
x Stainless steel case and wetted parts
2
x Max static pressure 100 kg/cm
x High corrosive resistance
x Dry or liquid filled
x IP65

Ranges
2
0 ... 1000 mmWC up to 0 ... 25 kg/cm

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification D104


Dial size in mm 100 150
Construction Cylindrical case
Measuring principle Single diaphragms
Static Pressure 100 kg/cm2 both side
Pressure range 1000 1600 2500 4000 mmWC
0,6 1 1.6 2.5 4 6 10 16 25 kg/cm 2
Max static pressure 100 kg/cm2
Overload capability One side 40 bar (10 bar for ranges ≤ 1600 mmWC)
Pressure type Differential
Process connection 1/4” NPT (F), 1/4” NPT (M), 1/2” NPT (M), 1/4” BSP (M), 1/2” BSP (M), others on request
Connection location Lower
Material
Pressure element Stainless steel 316L (1.4404) for ranges ≤2.5 bar
Duratherm for ranges > 2.5 bar
Measuring diaphragm Stainless steel 316L
Case/bayonet ring Stainless steel 304, optional 316
Window Laminated safety glass
Dial Aluminum, black markings on white background
Pointer Micro Pointer Aluminum, black
Accuracy Class 1,6 (1,6 % F.S.), optional Class 1 (1 % F.S.)
Permissible
Zero shift ±2% at maximum static pressure
Ambient temperature -20 ... 60 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,6 % / 10 K
Protection according EN 60 529/IEC 529 IP65
Filling liquids Glycerin, silicone, others on request
Mounting Direct, optional 2" pipe mounting
Weight dry/filled in kg 2.2 dry / 2.5 filled 2.4 dry / 3.1 filled
Accessories, options 3 or 5 way remote or direct mounted manifolds, valves, NACE

All specifications are subject to change without notice.


Page 27
IN.D104 Rev. E 27/01/2017
General dimensions in mm

Order information
Size System Execution Process Connection Range Engineer- Filling/case/ Options
material connection orientation ing unit contacts
(100) (S) Pressure (L) IP65 (02) 1/4 NPT (M) (L) Lower 0/10002) (mmWC) (=) Standard (NH) Tagging wired
100 mm chamber 0/1600 no filling
316L & (04) 1/2 NPT (M) 0/2500 (CS) Dual scale
(150) Diaphragm 0/4000 (GV) Silicone
(DA) Dial marking
150 mm ≤2.5 bar (13) 1/4 BSP (M)
2
316L 0/ 0,6 (kg/cm ) (GR) Glycerin
(AN) Calibration 1 %1)
> 2.5 bar (15) 1/2 BSP (M) 0/ 1
Duratherm 0/ 1,6 (YW) Case
(FX) Front panel
(02T) 1/4 NPT (F) 0/ 2,5 material
0/ 4 316 (FW) Surface plate
0/ 6
0/ 10 (TM) 2" pipe mounting
0/ 16 psi and bracket
others on
0/ 25 request

1) Accuracy 1% is excluding Hysteresis 2) Dial arc shall be 180°

Order example
Size Type System Execution Process Connection Range Engineering Filling/case/ Option
material connection orientation unit contacts
100 D104 S L 02T L 0/2500 mmWC = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 28
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Double diaphragm differential pressure gauge
Model D105
Max. Static pressure 400 bar
Nominal size 100 mm or 150 mm
Accuracy: Class 1,6

Features
x Stainless steel case and wetted parts
2
x Max static pressure 400 kg/cm , one side load permitted
x Chamber purge and bleed connection available
x High corrosive resistance
x Dry or liquid filled
x Dual diaphragm, solid front not required
x IP65

Ranges
2
0 ... 250 mmWC up to 0 ... 25 kg/cm

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification D105


Dial size in mm 100 150
Construction Cylindrical case
Zero adjustment Optional, Externally at the top of the case.
Measuring principle Dual diaphragm safety design, solid front not required
Static Pressure 25 kg/cm2 standard for ranges 250 to 1000 mmWC
100 kg/cm2 standard for ranges 1600 mmWC to 25 bar
Optional 100, 250 & 400 kg/cm2 (refer limits below)
Pressure range 250 to 1000 mmWC 1600 mmWC to 25 bar
Max static pressure 100 kg/cm2 250 kg/cm2 & 400 kg/cm2
Overload capability Static pressure is also maximum pressure allowed on one side
Pressure type Differential
Process connection 1/4” NPT (F), 1/4” NPT (M), 1/2” NPT (M), 1/4” BSP (M), 1/2” BSP (M), others on request
Connection location Lower
Material
Pressure connection Stainless steel 316L (1.4404)
Pressure chamber Stainless steel 316L (1.4404)
Measuring diaphragm Stainless steel 316L / Stainless steel 316Ti
Case/bayonet ring Stainless steel 304, optional 316
Window Laminated safety glass
Dial Aluminum, black markings on white background
Pointer Micro Pointer Aluminum, black,
Accuracy Class 1,6 (1,6 % F.S.)
Permissible
Zero shift ±2% at maximum static pressure
Ambient temperature -20 ... 60 °C
Medium temperature Max. 100 °C
Storage temperature -40 ... 60 °C
Effect Max. 0,6 % / 10 K
Protection according EN 60 529/IEC 529 IP65
Filling liquids Glycerin, silicone, others on request
Mounting Direct, optional 2" pipe mounting
Weight small / large chamber in kg 5.1/11.0 5.7/11.6
Accessories, options 3 or 5 way remote or direct mounted manifolds, valves, NACE

All specifications are subject to change without notice.


Page 29
IN.D105 Rev. B 21/01/2017
General dimensions in mm

Order information
Size Type System Execu- Process Connection Range Engineer- Filling/case/ Options
material tion connection orientation ing unit contacts
(100) D105 (S) Pressure (L) IP65 (02) ¼” NPT (M) (L) Lower 0/2501) (mmWC) (=) Standard (NH) Tagging wired
100 mm chamber 0/400 no filling
316L (1.4404) (04) ½” NPT (M) 0/ 600 (6B) Oxygen service
(150) Diaphragm 0/1000 (GV) Silicone
(CS) Dual scale
150 mm 316L / 316 Ti (13) ¼” BSP (M) 0/16001)
0/2500 (GR) Glycerin (DA) Dial marking
(15) ½” BSP (M) 0/4000
(YW) Case (EZ) External zero
(02T) ¼” NPT (F) 0/ 0,6 (kg/cm2) material adjustment
0/ 1 316
0/ 1,6 (TM) 2" pipe mounting
0/ 2,5 bracket
0/ 4
(PN100) Static pressure
0/ 6
100 bar
0/ 10
0/ 16 (PN250) Static pressure
0/ 25 250 bar2)
psi and others (PN400) Static pressure
others on request on request 400 bar2)
1) Dial arc shall be 180° 2) only with ranges ≥1600 mmWC

Order example
Size Type System Execution Process Connection Range Engineering Filling/case/ Option
material connection orientation unit contacts
150 D105 S L 04 L 0/600 mmWC = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 30
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Differential pressure gauge
Model D201
Max. static pressure 200 bar
Nominal size 100 and 150 mm
Accuracy: 2 % Ascending

Features
x Stainless steel case and several wetted parts
x Static pressure up to 200 bar
x Protection IP65
x Smooth pointer motion
x Dry or liquid filled
x Magnetic coupling, solid front not required

Ranges
0 … 0,5 bar up to 0 ... 10 bar

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Filters, Hydraulic systems
Water treatment

Technical specification D201


Dial size in mm 100 and 150
Construction Cylindrical case
Measuring principle Magnetic coupling, solid front not required
Actuator Piston
Process fluid migration Minimal
in bar 0,5 1 1,6 2 2,5 4 6 10
Max. static pressure in bar 200
Overload capability Maximum static pressure is also maximum pressure allowed on one side
Pressure type Differential
Process connection G ¼ B female, ¼ NPT female, others on request
Connection location Lower, back, in-line
Material
Wetted parts Body material, spring, screw, magnet
Body, pressure chamber, piston Aluminum, brass or stainless steel 316 (1.4401)
Magnet Ceramic
O-Ring Buna-N, Viton
Case SS304
Window Acrylic glass, optional Safety glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black, optional red set hand or maximum pointer
Accuracy calibrated ascending 2 %, 5% for liquid filled
Permissible
Ambient temperature 0 ... 80 °C
Medium temperature Max. 80 °C
Storage temperature 0 ... 60 °C
Effect Max. 0,3 % / 10 K
Protection according IP65
Filling liquids Glycerin, Silicon on request
Mounting Direct, optional panel or 2" pipe mounting (only for inline connections), others on request
Accessories, options 3 or 5 way remote mounted manifolds, valves

All specifications are subject to change without notice.


Page 31
IN.D201 Rev. C 25/04/2017
General dimensions in mm

Order information
Size Type Body Execution Process Connection Range Engineer- Filling/case/ Options
connection orientation ing units contacts
(100) D201 (F) Body (L) IP65 (02T)1/4” NPT (F) (S) In-line 0/ 0,5 KG/CM2 (=) Standard (NH) Tagging wired
100 mm Aluminum standa 0/ 1 no filling
rd case (13T)1/4” BSP(F) (L) Lower 0/ 1,6 (CS) Dual scale
(A) Body Brass 0/ 2 (GR) Glycerin
(150) (02)1/4” NPT (M) (B) Back 0/ 2,5 (DA) Dial marking
150 mm (S) Body 316 0/ 4
(13)1/4” BSP(M) 0/ 6 (SG) Safety glass
0/ 10
(FX) Front Panel

(VD) Viton O-Rings/


diaphragm

(TM) 2" pipe mounting


bracket
(inline mounting only)

psi and others (EP) Maximum pointer


on request
Others on request

Order example
Size Type Body Execution Process Connection Range Engineering Filling/case/ Options
connection orientation unit contacts
100 D201 S L 02T S 0/1 KG/CM2 = NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral 382 729 , Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 32
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm seals threaded types
Model DS20
Threaded connections to ANSI/ASME or ISO

Features
x Metal diaphragms with Teflon gasket
x Interchangeable components
x Removable top and bottom housing
x Variety of wetted parts materials

Ranges
15 psi up to 1000 psi
2 2
1 kg/cm up to 70 kg/cm

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Offshore installations
Pulp and paper industry
Waste water treatment

Technical specification DS20


Process connection size 1/4” 3/8” 1/2” BSP / NPT , M20x1.5
Construction Bolted top and bottom housing, removable for easy maintenance
Fill port At the side with stainless steel ball and set screw
Flushing connection Optional flushing
Flushing connection size 1/4” 1/2” NPT female
Maximum pressure
2
Standard [kg/cm ] 70
Measuring principle Diaphragm welded to top housing

Instrument connections [female] 3/8” BSP, 1/2” NPT, others on request


Material (wetted parts)
Bottom housing Stainless steel 316, 316L, Polypropylene, PVC, SS 316 with Teflon protection,
Hasteloy-C 276
Diaphragm (metal type) Stainless steel 316L Optional Hasteloy-C276
Diaphragm (other types) Stainless steel 316L with Teflon protection
Material (non wetted parts)
Top housing Stainless steel 304, optional 316
Clamps, rings Steel painted black, optional stainless
Bolts and nuts Stainless steel 300 series
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
Silicone (-40/600 °F, -40/315 °C),
Halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
Vegetable oil (32/248 °F, 0/120 °C),
Others on request, also depending on selected material and configuration.
Mounting Direct or with capillary
Weight in kg 1.00
Accessories, options NACE, pressure gauges, pressure switches, transducer, capillaries

All specifications are subject to change without notice.


Page 33
IN.DS20 Rev. H 29/01/2017
General dimensions in mm

Order information
Process Type Diaphragm Bottom housing Instrument Filling fluid Options
connection material connection
(06) 1” NPT(M) DS20 (S) 316L (S) 316 (14T) 3/8” (CG) Glycerin (NH) Tagging wired
BSP (F) (direct mounted,
(05) 3/4” NPT(M) (SP) 316L with (L) 316L -18/204 °C) (OF) Oil free (only filled with
teflon (04T) 1/2” halocarbon)
(CK) Silicone
(04) 1/2” NPT(M) protection1) (SP) 316 with teflon NPT (F) (direct mounted or with
lined 1)3) (YT) Stainless steel 316 top
capillary, -40/315 °C)
(03) 3/8” NPT(M) (SJ) 316L with housing
1)2)
teflon (PP) Polypropylene (CF) Halocarbon
(02) 1/4” NPT(M) Coating1) (direct mounted or with (SE) Stainless steel clamp
1)2)
(V) PVC capillary, -56/149 °C) rings
(18) 1” BSP(M) (H) Hasteloy-
(H) Hasteloy-C276 (HA) Syltherm 800 (PU) Flushing connection 1/4”
C276
(17) 3/4” BSP(M) (direct mounted or with
capillary, -40/400 °C)
NPT (F)5)

(15) 1/2” BSP(M) (GZ) Vegetable oil (CA) Stainless steel armored
(direct mounted or with capillary 4)
(14) 3/8” BSP(M) capillary, 0/120 °C)
(CP) Stainless steel armored
(13) 1/4” BSP(M) capillary with PVC
sleeving 4)
(16) M20x1.5(M)

(06T) 1 NPT(F)

(05T) 3/4 NPT(F)

(04T) 1/2” NPT(F)


4) Max length 30m, specify length
(02T) 1/4” NPT(F) in mm
others on request EG. CP1000 = CP with 1m
1) Restricted to maximum temperature 200 °C
2) Max. Rating / temperature limit: 200 psi/14 bar at 74 °F/23 °C; 125 psi/8.6 bar at 125 °F/52 °C; 80 psi/5,5 bar at 150 °F/65 °C
3) Not for process connection (02), (13), (06T), (05T), (04T), (02T) , not available with option PU
5) Not for (SP), (PP)& (V)
How to order
Process Type Diaphragm Bottom housing Instrument Filling fluid Option
connection material connection
04 DS20 S S 14T CK CA1000

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 34
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm seals threaded types
Model DS21
Threaded connections to ANSI/ASME or ISO

Features
x Removable diaphragm seal capsule
x Interchangeable components
x Removable top and bottom housing
x Variety of wetted parts materials

Ranges
15 psi up to 1,500 psi
2 2
1 kg/cm up to 100 kg/cm

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Offshore installations
Pulp and paper industry
Waste water treatment

Technical specification DS21


Process connection size 1/4” 3/8” 1/2” BSP / NPT, M20x1.5
Construction Bolted top and bottom housing, removable for easy maintenance and
changeable capsules
Fill port At the side with stainless steel ball and set screw
Flushing connection Optional flushing
Flushing connection size 1/4” 1/2” NPT female
Maximum pressure
2
Standard [kg/cm ] 70
2
With high pressure ring [kg/cm ] 100
Measuring principle Screwed capsule
Instrument connections [female] 3/8” BSP, 1/2” NPT, others on request
Material (wetted parts)
Bottom housing Stainless steel 316, Polypropylene, PVC, Monel, Hastelloy C, SS 316 with
teflon protection,
Diaphragm (metal type) Stainless steel 316L (1.4404), stainless steel 304L (1.4306), K-Monel,
Hastelloy C, Hastelloy B, Nickel, Carpenter 20, Tantalum
Material (non wetted parts)
Top housing Stainless steel 304, optional 316
Clamps, rings Steel painted black, optional stainless
Bolts and nuts Stainless steel 300 series
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
Silicone (-40/600 °F, -40/315 °C),
Halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
Vegetable oil (32/248 °F, 0/120 °C),
Others on request, also depending on selected material and configuration.
Mounting Direct or with capillary
Weight in kg 1.32
Accessories, options NACE, pressure gauges, pressure switches, transducer, capillaries

All specifications are subject to change without notice.


Page 35
IN.DS21 Rev. G 06/12/2016
General dimensions in mm

Order information
Process Type Diaphragm Bottom housing Instrument Filling fluid Options
connection material connection
(06) 1” NPT(M) DS21 (S) 316L (S) 316 (14T) 3/8 BSP (F) (CG) Glycerin (NH) Tagging wired
(direct mounted,
(05) 3/4” NPT(M) (C) 304L (L) 316L (04T) 1/2” NPT (F) -18/204 °C) (OF) Oil free (only filled with
halocarbon)
(04) 1/2” NPT(M) (D) Carpenter (SP) 316 with teflon (CK) Silicone
(direct mounted or with (YT) Stainless steel 316 top
20 lined 1)3) capillary, -40/315 °C)
(03) 3/8” NPT(M) housing
(G) Hastelloy B (PP) Polypropylene 2) (CF) Halocarbon
(02) 1/4” NPT(M) (direct mounted or with (SE) Stainless steel clamp
2)
(H) Hastelloy C (V) PVC capillary, -56/149 °C) rings
(18) 1” BSP(M)
(N) Nickel (H) Hastelloy C (HA) Syltherm 800 (PU) Flushing connection
(17) 3/4” BSP(M) (direct mounted or with
capillary, -40/400 °C)
1/4” NPT (F)5)
(P) K-Monel (M) Monel 400
(15) 1/2” BSP(M) (GZ) Vegetable oil (CA) Stainless steel armored
(U) Tantalum (direct mounted or with capillary 4)
(14) 3/8” BSP(M) capillary, 0/120 °C)
(CP) Stainless steel armored
(13) 1/4” BSP(M) capillary with PVC
sleeving 4)
(16) M20x1.5(M)

(06T) 1 NPT(F)

(05T) 3/4 NPT(F)

(04T) 1/2” NPT(F)


4) Max length 30m, specify length
(02T) 1/4” NPT(F) in mm
others on request EG. CP1000 = CP with 1m
1) Restricted to maximum temperature 200 °C, pressure 40 bar
2) Max. Rating / temperature limit: 200 psi/14 bar at 74 °F/23 °C; 125 psi/8.6 bar at 125 °F/52 °C; 80 psi/5,5 bar at 150 °F/65 °C
3) Not for process connection (02), (13), (06T), (05T), (04T), (02T) , not available with option PU
5) Not for (SP), (PP)& (V)

How to order
Process Type Diaphragm Bottom housing Instrument Filling fluid Option
connection material connection
04 DS21 S S 14T CK CA1000

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 36
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm seals threaded types
Model DS50
Threaded connections to ANSI/ASME or ISO

Features
x Metal diaphragms with Teflon gasket
x Interchangeable components
x Removable top and bottom housing
x Variety of wetted parts materials

Ranges
15 psi up to 3,500 psi
2 2
1 kg/cm up to 250 kg/cm

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Offshore installations
Pulp and paper industry
Waste water treatment

Technical specification DS50


Process connection size 1/4” 3/8” 1/2” BSP / NPT, M20x1.5
Construction Diaphragm, top and bottom housing welded
Fill port At the side with stainless steel ball and set screw
Flushing connection Optional flushing
Flushing connection size 1/4” 1/2” NPT female
Maximum pressure
2
Standard [kg/cm ] 100
2
With high pressure ring [kg/cm ] 250
Measuring principle Diaphragm welded to top and bottom housing
Instrument connections [female] 3/8” BSP, 1/2” NPT, others on request
Material
Housing Stainless steel 316, Optional Stainless steel 316L , Hasteloy-C276
Diaphragm (metal type) Stainless steel 316L, Optional Hasteloy-C276
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
Silicone (-40/600 °F, -40/315 °C),
Halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
Vegetable oil (32/248 °F, 0/120 °C),
Others on request, also depending on selected material and configuration.
Mounting Direct or with capillary
Weight in kg 0.6
Accessories, options NACE, pressure gauges, pressure switches, transducer, capillaries

All specifications are subject to change without notice.


Page 37
IN.DS50 Rev. H 29/01/2017
General dimensions in mm

Order information
Process Type Diaphragm Housing Instrument Filling fluid Options
connection material connection
(06) 1” NPT(M) DS50 (S) 316L (S) 316 (14T) 3/8” BSP (CG) Glycerin (NH) Tagging wired
(F) (direct mounted,
(05) 3/4” NPT(M) (H) Hasteloy-C276 (L) 316L -18/204 °C) (HP) High Pressure Ring
(04T) 1/2” NPT
(04) 1/2” NPT(M) (H) Hasteloy-C276 (CK) Silicone (OF) Oil free
(F) (direct mounted or with
(only filled with halocarbon)
capillary, -40/315 °C)
(03) 3/8” NPT(M)
(CF) Halocarbon (PU) Flushing connection
(02) 1/4” NPT(M) (direct mounted or with 1/4” NPT (F)
capillary, -56/149 °C)
(18) 1” BSP(M) (CA) Stainless steel
(HA) Syltherm 800 armored capillary 1)
(17) 3/4” BSP(M) (direct mounted or with
capillary, -40/400 °C)
(CP) Stainless steel
(15) 1/2” BSP(M) (GZ) Vegetable oil armored capillary
(direct mounted or with with PVC sleeving 1)
(14) 3/8” BSP(M) capillary, 0/120 °C)

(13) 1/4” BSP(M)

(16) M20x1.5(M)

(06T) 1 NPT(F)

(05T) 3/4 NPT(F)

(04T) 1/2” NPT(F)


1) Max length 30m, specify
(02T) 1/4” NPT(F) length in mm
others on request EG. CP1000 = CP with 1m

How to order
Process Type Diaphragm Housing Instrument Filling fluid Option
connection material connection
04 DS20 S S 14T CK CA1000

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 38
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm seals threaded types
Model DS51
Threaded connections to ANSI/ASME or ISO
Features
x Metal diaphragm all welded design
x Compact design
Ranges
15 psi up to 300 psi
2 2
1 kg/cm up to 21 kg/cm
Applications
Pharmaceuticals
Waste water treatment
Machine and apparatus construction
Biotechnology and filtration industries
Technical specification DS51
Process connection size 1/4” BSP / NPT (M)
Construction Diaphragm, top and bottom housing welded
2
Pressure range in kg/cm (min || max) 1 || 21
Instrument connection [female] 1/4” BSP
Material
Housing Stainless steel 316, Optional Stainless steel 316L , Hasteloy-C276
Diaphragm (metal type) Stainless steel 316L, Optional Hasteloy-C276
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C), silicone (-40/600 °F, -40/315 °C), halocarbon (-
70/300 °F, -56/149 °C), Syltherm 800 (-40/750 °F, -40/400 °C), vegetable oil
(32/248 °F, 0/120 °C)
Mounting Direct
Weight dry in grams 220
Accessories, options Pressure gauges, transducer

General dimensions in mm

Order information
Process Type Diaphragm Housing Instrument Filling fluid Options
connection material connection
(13) 1/4” BSP(M) DS51 (S) 316L (S) 316 (13T) 1/4” BSP(F) (CG) Glycerin (OF) Oil free (only filled with
(CK) Silicone halocarbon)
(02) 1/4” NPT(M) (H) Hasteloy-C276 (L) 316L (CF) Halocarbon
(HA) Syltherm 800
(H) Hasteloy- (GZ) Vegetable oil
C276

How to order
Process Type Diaphragm Housing Instrument Filling fluid Option
connection material connection
02 DS51 S S 13T CK OF

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 39
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN.DS51 Rev. D 02/02/2017
Diaphragm seals for sanitary use
Model DS52
Standard Tri-clamp connections
Features
x SS 316L diaphragms welded to a SS316L top housing
x Compatible with Tri-Clover
x Easy to remove from process
Ranges
15 psi up to 600 psi
2 2
1 kg/cm up to 40 kg/cm
Applications
Pharmaceuticals
Dairy & Food process industry
Biotechnology and filtration industries
Breweries, distilleries, wineries and citrus juice production plants
Model DS52
Process connection size 1 1/2” Clamp 2” Clamp
2
Pressure range in kg/cm (min || max) 0/2.5 || 0/40 for 63mm, 0/1 || 0/40 for 63mm and 100mm
0/4 || 0/40 for 100mm
Instrument Connection 1/4” BSP (F)
Material, Diaphragm & Top housing Stainless steel 316L
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/300 °F, -18/150 °C), vegetable oil (32/248 °F, 0/120 °C)
Mounting Direct
Weight in grams 160 240
Accessories, options Pressure gauges, transducer

General dimensions in mm

Order information
Process Type Diaphragm and Instrument Filling fluid Options
connection top housing connection
(15) 1 1/2” Tri-clamp DS52 (S) 316L (13T) 1/4” BSP(F) (CG) Glycerin (NH) Tagging marked
(direct mounted,
(20) 2” Tri-clamp -18/150 °C) (C) Clamp
(GZ) Vegetable oil (G) Gasket
(direct mounted,
0/120 °C)
(PE) Electro polishing

others on request

How to order
Process Type Diaphragm and Instrument Filling fluid Option
connection top housing connection
15 DS52 S 13T GZ NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 40
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN.DS52 Rev. C 20/01/2017
Compact high pressure seal
Model DS54
Threaded connections to ANSI/ASME or ISO

Features
x All welded metal construction
x No gaskets or bolts
x All material stainless steel

Ranges
1500 psi up to 6000 psi
2 2
100 kg/cm up to 400 kg/cm

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Power station

Technical Specification DS54


Process connection size 1/4” 3/8” 1/2” 3/4” BSP / NPT (M), M20x1.5(M), others on request
2
Pressure range in kg/cm (min || max) 100 || 400
Material
Housing Stainless steel 316, Optional Stainless steel 316L , Hasteloy-C276
Diaphragm (metal type) Stainless steel 316L Optional Hasteloy-C276
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
Silicone (-40/600 °F, -40/315 °C),
Halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
Vegetable oil (32/248 °F, 0/120 °C)
Others on request, also depending on selected material and configuration.
Mounting Direct
Weight dry in grams 440
Accessories, options Pressure gauges, transducer

All specifications are subject to change without notice.


Page 41
IN.DS54 Rev. F 25/03/2017
General dimensions in mm

Order information
Process Type Diaphragm Housing Instrument Filling fluid Option
connection material connection
(05) 3/4” NPT(M) DS54 (S) 316L (S) 316 (14T) 3/8” BSP(F) (CG) Glycerin (NH) Tagging wired
(direct mounted,
(04) 1/2” NPT(M) (H)Hasteloy- (L) 316L -18/204 °C) (OF) Oil free
C276 (only filled with
(03) 3/8” NPT(M) (H) Hasteloy-C276 (CK) Silicone halocarbon)
(direct mounted or with
capillary, -40/315 °C) (PU) Flushing
(02) 1/4” NPT(M)
connection 1/4”
(GZ) Vegetable oil
(16) M20x1.5(M) (direct mounted, NPT (F)
0/120 °C)
(17) 3/4” BSP(M) (CA) Stainless steel
(CF) Halocarbon armored capillary 1)
(direct mounted or with
(15) 1/2” BSP(M) capillary, -56/149 °C)
(CP) Stainless steel
(14) 3/8” BSP(M) (HA) Syltherm 800 armored capillary
(direct mounted or with with PVC sleeving
1)
(13) 1/4” BSP(M) capillary, -40/400 °C)

1) Max length 30m, specify


length in mm
EG. CP1000 = CP with 1m

How to order
Process Type Diaphragm Housing Instrument Filling fluid Option
connection material connection
02 DS54 S S 14T CK NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 42
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm seals flanged types
Model DF20
Flanged connections to ANSI/ASME or EN

Features
x Metal diaphragms with Teflon gasket
x Removable top and bottom housing
x Variety of wetted parts materials

Pressure rating
From 150 lbs up to 1500 lbs
From PN10 up to PN100

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Offshore installations
Pulp and paper industry
Waste water treatment

Technical specification DF20


Process connection size, flanged ANSI 1/2", 3/4", 1", 1 ½", 2", 3"
DIN DN15, DN25, DN40, DN50, DN80, others on request
Rating ANSI 150 lbs, 300 lbs, 600 lbs, 900 lbs, 1500 lbs
DIN PN10, PN16, PN25, PN40, PN64, PN100, others on request
Construction Bolted top and bottom housing, removable for easy maintenance
Fill port At the side with stainless steel ball and set screw
Flushing connection Optional flushing
Flushing connection size 1/4” 1/2” NPT female
Maximum pressure
Metal bottom housing [lbs/PN] 1500/100
Other bottom housing [lbs/PN] 150/10
Measuring principle Diaphragm welded to top housing
Instrument connections [female] 3/8” BSP, 1/2” NPT, others on request
Material (wetted parts)
Bottom housing Stainless steel 316, 316L, 316 with teflon protection, polypropylene, PVC
Diaphragm (metal type) Stainless steel 316L
Diaphragm (other types) Stainless steel 316L with teflon protection
Material (non wetted parts)
Top housing Stainless steel 304, 316
Clamps, rings Steel painted black, optional stainless steel
Bolts and nuts Stainless steel 300 series
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
Silicone (-40/600 °F, -40/315 °C),
Halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
Vegetable oil (32/248 °F, 0/120 °C).
Others on request, also depending on selected material and configuration.
Mounting Direct or with capillary
Accessories, options NACE, pressure gauges, pressure switches, transducers, capillaries

All specifications are subject to change without notice.


Page 43
IN.DF 20 Rev. F 06/12/2016
General dimensions in mm

Order information
Process Type Diaphragm Bottom Instrument Rating Flange Filling fluid Options
connection housing connection facing
material
ANSI Flanged DF20 (S) 316L (S) 316 (14T) 3/8 ANSI (FF) Flat (CG) Glycerin (NH) Tagging wired
BSP(F) face 2) (direct mounted,
(05) 1/2” (SP) 316L with (L) 316L (150) 150 lbs -18/204 °C) (OF) Oil free
(only filled with
teflon (04T) 1/2” (RF) Raised
(07) 3/4” (300) 300 lbs (CK) Silicone halocarbon)
protection1) (SP) 316 with NPT(F) face 2) (direct mounted
(10) 1” teflon (600) 600 lbs or with capillary, - (YT) Stainless steel
1)
(SJ) 316L with lined (RTJ) Ring 40/315 °C) 316
(15) 1 ½” teflon (900) 900 lbs joint top housing
1) 2) (CF) Halocarbon
Coating (PP)Polypropyl type
(20) 2” ene 3) (1500) 1500 (direct mounted (SE) Stainless steel
lbs or with capillary, -
(H) Hasteloy- clamp rings
(30) 3” 3) 56/149 °C)
C276 (V) PVC
(SR4) Spacer ring with
DIN Flanged DIN (HA) Syltherm 800
(direct mounted 1/2” NPT
or with capillary, - flushing5)
(DN15) (PN10) 40/400 °C)
(SR2) Spacer ring with
(DN25) (PN16)
(GZ) Vegetable oil 1/4” NPT
(DN40) (PN25) (direct mounted flushing5)
or with capillary,
0/120 °C)
(DN50) (PN40) (CA) Stainless steel
armored capillary
(DN80) (PN64) 4)

(PN100) (CP) Stainless steel


armored capillary
with PVC sleeving
4)

4) Max length 30m, specify


length in mm
EG. CP1000 = CP with
others on request others on request 1m
1) Restricted to maximum temperature 200 °C, not available with option SR2, SR4 2) The DIN-flange facing depends on nominal size and nominal pressure
3) Only ¼ and ½ NPT, max. Rating/temp. limit: 200 psi/14 bar at 74 °F/23 °C; 125 psi/8.6 bar at 125 °F/52 °C; 80 psi/5,5 bar at 150 °F/65 °C
5) Not for (SP), (PP)& (V)

How to order
Process Type Diaphragm Bottom housing Instrument Filling fluid Option
connection capsule material material connection
04 DF20 S S 14T CK CA1000

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 44
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm seals flanged types
Model DF21
Flanged connections to ANSI/ASME or EN

Features
x Removable diaphragm seal capsule
x Metal diaphragms with Teflon gasket
x Removable top and bottom housing
x Variety of wetted parts materials

Pressure rating
From 150 lbs up to 1500 lbs
From PN6 up to PN100

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Offshore installations
Pulp and paper industry
Waste water treatment

Technical specification DF21


Process connection size, flanged ANSI 1/2", 3/4", 1", 1 ½", 2", 3"
DIN DN15, DN25, DN40, DN50, DN80, others on request
Rating ANSI 150 lbs, 300 lbs, 600 lbs, 900 lbs, 1500 lbs
DIN PN10, PN16, PN25, PN40, PN64, PN100, others on request
Construction Bolted top and bottom housing, removable for easy maintenance
Fill port At the side with stainless steel ball and set screw
Flushing connection Optional flushing
Flushing connection size 1/4” 1/2” NPT female
Maximum pressure
Metal bottom housing [lbs/PN] 1500/100
Other bottom housing [lbs/PN] 150/10
Measuring principle Screwed capsule to top housing
Instrument connections [female] 3/8” BSP, 1/2” NPT, others on request
Material (wetted parts)
Bottom housing Stainless steel 316, 316L, 316 with teflon protection, polypropylene, PVC,
Hastelloy-C, Monel
Diaphragm (metal type) Stainless steel 316L, 304L, Carpenter 20, Hastelloy B, Hastelloy C, Nickel, K-
Monel, Tantalum
Material (non wetted parts)
Top housing Stainless steel 304, 316
Clamps, rings Steel painted black, optional stainless steel
Bolts and nuts Stainless steel 300 series
Accuracy Additional 0,5 % to the specified tolerance of the mounted instrument
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
Silicone (-40/600 °F, -40/315 °C),
Halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
Vegetable oil (32/248 °F, 0/120 °C),
Others on request, also depending on selected material and configuration.
Mounting Direct or with capillary
Accessories, options NACE, pressure gauges, pressure switches, transducers, capillaries

All specifications are subject to change without notice.


IN.DF 21 Rev. G 06/12/2016
Page 45
General dimensions in mm

Order information
Process Type Diaphragm Bottom Instrument Rating Flange Filling fluid Options
connecti housing connection facing
on material
ANSI DF21 (S) 316L (S) 316 (14T) 3/8” ANSI (FF) Flat (CG) Glycerin (NH) Tagging wired
Flanged BSP(F) face 2) (direct mounted,
(C) 304L (L) 316L (150) 150 lbs -18/204 °C) (OF) Oil free
(05) 1/2” (only filled with
(04T) 1/2” (300) 300 lbs (RF) Raised
(CK) Silicone halocarbon)
(07) 3/4”
(D) Carpenter 20 (SP) 316 with NPT(F) face 2) (direct mounted
teflon (600) 600 lbs or with capillary, (YT) Stainless steel
(10) 1” (G) Hastelloy B lined1) (RTJ) Ring -40/315 °C) 316
(900) 900 lbs joint top housing
(15) 1 (H) Hastelloy C (PP)Polypropyl type 2) (CF) Halocarbon
½” ene 3) (1500) 1500 (direct mounted (SE) Stainless steel
(N) Nickel lbs or with capillary,
-56/149 °C)
clamp rings
(20) 2” (V) PVC 3)

(P) K-Monel (SR4) Spacer ring with


(30) 3” (HA) Syltherm
(H) Hastelloy 1/2” NPT
800
(U) Tantalum C (direct mounted flushing5)
DIN DIN
or with capillary,
Flanged -40/400 °C) (SR2) Spacer ring with
(M) Monel (PN10)
1/4” NPT
(DN15)
(PN16) (GZ) Vegetable flushing5)
(DN25) oil
(PN25) (direct mounted (CA) Stainless steel
(DN40) or with capillary, armored capillary
(PN40) 0/120 °C) 4)

(DN50)
(PN64) (CP) Stainless steel
(DN80) armored capillary
(PN100)
with PVC sleeving
4)

4) Max length 30m, specify


length in mm
EG. CP1000 = CP with
others on request others on request 1m
1) Restricted to maximum temperature 200 °C, not available with option SR2, SR4 2) The DIN-flange facing depends on nominal size and nominal pressure
3) Max. Rating /temp limit: 200 psi/14 bar at 74 °F/23 °C; 125 psi/8.6 bar at 125 °F/52 °C; 80 psi/5,5 bar at 150 °F/65 °C
5) Not for (SP), (PP)& (V)

How to order
Process Type Diaphragm Bottom housing Instrument Rating Flange facing Filling fluid Option
connection capsule material material connection
15 DF21 S S 14T 300 RF CK CA1000

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 46
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Flush type flanged diaphragm seal
Model DF50
Flanged connections to ANSI/ASME or EN

Features
x Welded diaphragm seal
x Variety of wetted parts material
x Filling port in the upper housing
x Suitable for pressure gauges, switches and transmitters

Size
1 ½”, 2” & 3” ASME/ANSI
DN40, DN50 & DN80

Pressure Rating
From 150 lbs up to 2500 lbs
From PN10 up to PN160

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Offshore installations
Pulp and paper industry

Technical specification DF50


Process connection size, flanged
ANSI 1 ½”, 2", 3"
DIN DN40, DN50, DN80
Back up flange Rating ANSI 150 lbs, 300 lbs, 600 lbs, 900 lbs, 1500 lbs, 2500 lbs
DIN PN10, PN16, PN25, PN40, PN64, PN100, PN160 others on request
Construction Machined flange with welded flush diaphragm
Fill port In upper housing side with stainless steel ball and set screw
Measuring principle Diaphragm welded to flange
Instrument connections [female] 3/8” BSP(F), Others on request
Sealing surfaces
ANSI/ASME B16.5 RF, optional RTJ
EN EN 1092-1 Form B1
Material (wetted parts)
Diaphragm & contact surface Stainless steel 316L, Optoinal 316L with PTFE protection, Hasteloy-C276
Material (non wetted parts)
Instrument connection Stainless steel 304
Capillary and armor Stainless steel 316 and stainless steel 304
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
silicone (-40/600 °F, -40/315 °C),
halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
vegetable oil (32/248 °F, 0/120 °C)
Others on request, also depending on selected material and configuration.
Accuracy Additional 0.5% to the specified tolerance of the mounted instrument
Mounting Direct or with capillary
Weight dry in kg Depending on flange size
Accessories, options NACE, pressure gauges, pressure switches, transducers, capillaries

All specifications are subject to change without notice.


Page 47
IN.DF50 Rev. D 28/01/2017
General dimensions in mm

Order information
Process Type Diaphragm Flange mate- Instrument Flange Flange Filling fluid Options
Size rial connection Rating Facing
ANSI DF50 (S) 316L (S) 316 (14T) 3/8 ANSI (RF) Raised (CG) Glycerin (NH) Tagging wired
Flanged BSP(F) face2) (direct mounted,
(SJ) 316L with (L) 316L (150) 150 lbs -18/204 °C) (OF) Oil free (only filled with
(15)1½” halocarbon)
Teflon (300) 300 lbs (FF) Flat 2) (CK) Silicone
(20)2” Coating1) (SJ) 316 with face (direct mounted or (CA) Stainless steel
PTFE (600) 600 lbs with capillary, armored capillary 3)
(30)3” (H) Hasteloy- coating1) -40/315 °C)
C276 (900) 900 lbs (CP) Stainless steel
(H) Hasteloy- (CF) Halocarbon armored capillary
(1500) 1500 (direct mounted or
C276
with capillary,
with PVC sleeving 3)
lbs
-56/149 °C)
(2500) 2500
lbs (HA) Syltherm 800
(direct mounted or
with capillary,
DIN DIN -40/400 °C)
Flanged
(PN10) (GZ) Vegetable oil
(DN40) (direct mounted or
(PN16) with capillary,
(DN50) 0/120 °C)
(PN25)
(DN80)
(PN40)
(PN64)
(PN100)
3) Max length 30m, specify
(PN160) length in mm
others on request others on request EG. CP1000 = CP with 1m
1) Restricted to temperature 200 °C 2) The DIN-flange facing depends on nominal size and nominal pressure

How to order
Process Type Wetted Flange Instrument Flange Flange Filling fluid Options
Size Parts material connection rating facing
20 DF50 S S 14T 150 RF CK CA1000

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 48
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Diaphragm seals pancake type
Model DF52
Pancake type to ANSI/ASME or DIN/EN

Features
x Wafer “pancake” type construction
x Flushed diaphragm
x Variety of wetted parts materials
x Suitable for slurry & high viscous service

Size
2” to 5” ASME/ANSI
DN50 to DN125

Applications
Chemical and petrochemical industry
Offshore installations
Pulp and paper industry

Technical specification DF52


Process connection size,
flanged ANSI 2", 3", 4”, 5”
DIN DN50, DN80, DN100, DN125
Back up flange Rating ANSI 150 lbs, 300 lbs, 600 lbs, 900 lbs, 1500 lbs
DIN PN10, PN16, PN25, PN40, PN64, PN100, others on request
Construction Flush welded diaphragm
Fill port At the top with stainless steel ball and set screw
Flushing connection Flushing through spacer ring
Flushing connection size 1/4” or 1/2” NPT female
Measuring principle Metallic diaphragm
Instrument connections [female] 3/8” BSP(F) through capillary
Material (wetted parts)
Diaphragm & contact surface Stainless steel 316L, 316L with PTFE protection, Gold coated,
Hastelloy-C276, Monel
Material (non wetted parts)
Counter Flange Stainless steel 304, 316, Carbon Steel
Filling liquids (temperature range) Glycerin (0/400 °F, -18/204 °C),
Silicone (-40/600 °F, -40/315 °C),
Halocarbon (-70/300 °F, -56/149 °C),
Syltherm 800 (-40/750 °F, -40/400 °C),
Vegetable oil (32/248 °F, 0/120 °C),
Others on request, also depending on selected material and configuration.
Mounting With capillary
Weight dry without capillary in kg 2”/DN50 3”/DN80 4”/DN100 5”/DN125
1.4 2.5 3.5 5.0
Accessories, options NACE, pressure gauges, pressure switches, transducers, capillaries

All specifications are subject to change without notice.


Page 49
IN.DF52 Rev. D 07/12/2016
General dimensions in mm

Order information
Process Type Wetted Parts Instrument Backup Flange Backup Flange Filling fluid Options
Size connection Rating Material
ANSI DF52 (S) 316 (14T) 3/8” ANSI (NA) No backup (CG) Glycerin (NH) Tagging wired
Flanged BSP(F) flange (direct mounted,
(L) 316L (NA) No backup -18/204 °C) (OF) Oil free (only filled with
(20) 2” flange (C) SS 304 halocarbon)
(SJ) 316 with (CK) Silicone
(30) 3” (150) 150 lbs (S) SS 316 (direct mounted or (SR4) Spacer ring with
teflon with capillary, - 1/2” NPT flushing3)
(40) 4” coating1) (300) 300 lbs (CS) Carbon steel 40/315 °C)
(SR2) Spacer ring with
(50) 5” (SG) 316 with (600) 600 lbs (CF) Halocarbon 1/4” NPT flushing3)
GOLD (direct mounted or
(900) 900 lbs with capillary, - (CA) Stainless steel
coating 56/149 °C)
armored capillary 2)
(1500) 1500 lbs
(H) Hastelloy - (HA) Syltherm 800 (CP) Stainless steel
DIN C276 DIN (direct mounted or
with capillary, - armored capillary
Flanged 40/400 °C) with PVC sleeving
(M) Monel (NA) No backup 2)
(DN50) flange (GZ) Vegetable oil
(direct mounted or
(DN80) (PN10) with capillary, 0/120
°C)
(DN100) (PN16)
(DN125) (PN25)
(PN40)
(PN64)
(PN100)
2) Max length 30m, specify
length in mm
others on request others on request others on request EG. CP1000 = CP with 1m
1) Restricted to temperature 200 °C 3) Not for (SJ) & (SG)

How to order
Process Type Wetted Parts Instrument Backup flange Backup flange Filling fluid Options
Size connection rating material
30 DF52 S 14T 300 CS CK CA1000

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 50
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All stainless steel bimetal thermometer
Model BMT
Nominal size 63 mm, 100 mm, 115 mm or 150 mm
Accuracy: Class 1

Features
x Rugged stainless steel construction
x Fast response
x Protection IP65
x External zero adjustment
x Dry or liquid filled
x Silicone damped coil

Ranges
-50 ... 50 °C up to 0 ... 600 °C
-80 ... 120 °F up to 30 ... 1100 °F

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification BMT


Dial size 63mm 100 mm 115 mm 150 mm
Construction Cylindrical case with bayonet ring
Measuring principle Bimetallic coil
Range in °C 0/50 0/60 0/80 0/100 0/120 0/150 0/160 0/200 0/250 0/30 0/400 0/500 0/600
-10/50 -10/110 -20/40 -20/60 -20/180 -30/50 -30/70 -40/40 -40/60 -40/160
Ranges in °F and dual scales on request
Temperature limits for ranges ≤500°C >500 °C
Max. process over temperature 130 % of span 120 % of span
Stem
Diameter 6 mm, 8 mm, 10 mm, 12 mm
Length 63 ... 1000 mm, minimum length depends on range, (refer chart)
Process connection Plain
1/2” 3/4” 1” BSP / NPT (M) fixed or adjustable
Connection location Back / Every angle Lower / Back / Every angle (rotatable)
(rotatable)
Material
Connection Stainless steel 316
Stem Stainless steel 316L
Case/ring Stainless steel 304, optional 316L
Window Instrument glass, optional laminated safety glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black
Accuracy Class 1
Permissible
Ambient temperature -25 ... 60 °C
Storage temperature -40 ... 60 °C
Protection IP65
Filling liquids Silicone, not for lower connection, maximum range up to 300 °C
Accessories, options Thermowells, dual scale

All specifications are subject to change without notice.


Page 51
IN.BMT Rev. M 19/05/2017
General dimensions in mm

Order information
Size Stem Filling Process connection Connection Stem Range Engineer- Options
diameter orientation length ing unit
(63) (06) 6 mm (=) Standard (40) Plain (75) Adjustable (L) Lower2) 63 to 0 / 50 °C (NH) Tagging
63mm (no filling) (60) Fixed 1/2 union 1 1000 mm 0 / 60 wired
(08) 8 mm BSP (M) (R) Back (for minimum 0 / 80
(100) NPT (M) length refer (DA) Dial mark-
(L) Liquid filled 0 / 100
(61) Fixed 1/2 (76) Adjustable (E) Every above table)
100 mm (10) 10 mm with silicone 0 / 120 ing
union 1/2 angle
max. up to BSP (M) 0 / 150
(115) NPT (F) (CS) Dual scale
(12) 12 mm 300 °C (70) Adjustable 0 / 160
115 mm (not for lower (77) Adjustable 0 / 200
connection, ambi- union 1/2 (SG) Safety
ent temperature NPT (M) union 1/2 0 / 250
(150) BSP (F) 0 / 300 glass
maximum 65 °C)
150mm (71) Adjustable 0 / 400
(90) Adjustable (YW) 316L hous-
union 1/2 0 / 500
BSP (M) union ing
M20X1.5 0 / 550
(72) Adjustable (110) Adjustable 0 / 600 (EZ) External
union 3/4 union 1/4 zero ad-
NPT (M) NPT (M) justment1)
-10/ 50
(73) Adjustable (112)Adjustable -10/ 110
union 3/4 union 3/8
BSP (M) -20/ 40
BSP (M) -20/ 60
(113) Adjustable
(74) Adjustable union 1/4 -20/ 180
union 1 BSP (M) -30/ 50
NPT (M) -30/ 70
-40/ 40
-40/ 60
-40/ 160
Others on
request
1) For size 100mm, 115 mm and 150mm only, and not for connection orientation Lower. 2) Not for size 63mm

How to order
Size Type Stem Filling Process Connection Stem Range Engineering Option
diameter connection orientation length Unit
100 BMT 12 = 70 E 150 0/200 C NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 52
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All stainless steel mercury actuated thermometer
Model MTS
Nominal size 100 mm or 150 mm
Accuracy: Class 1 (1%)

Features
x Rugged stainless steel construction
x Fast response
x Protection IP65
x High repeatability and small hysteresis
x Dry or liquid filled
x Rigid stem or bulb with capillary

Ranges
-38 ... 50 °C up to 0 ... 600 °C
-35 ... 120 °F up to 30 ... 1100 °F

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification Rigid stem With capillary


Dial size in mm 100 150 100 150
Construction Cylindrical case with bayonet ring
Measuring principle Mercury actuated
Range in °C 0/60 0/80 0/100 0/120 0/160 0/200 0/250 0/300 0/400 0/500 0/600
-10/50 -10/110 -20/40 -20/60 -20/100 -20/12 -20/180 -30/50 -30/70 -30/170
-38/50
Over temperature limit Ranges in °F and dual scales on request
Max. 130 % F.S.
Stem or bulb diameter 6 mm, 8 mm, 10 mm, 12mm
Lengths 80…4000 mm, other length on request, min. length depends on bulb diameter (refer chart)
Capillary length N.A. Max. 30 m
Process connection Plain
1/2” 3/4” 1” BSP / NPT (M), M20x1.5, M24x1.5, M27x2
Fixed, adjustable or with swivel nut
Connection location Back, lower, Every angle
Material
Connection Stainless steel 316
Stem Stainless steel 316L
Case/ring Stainless steel 304 , optional 316L
Capillary Stainless steel 316
Armor Flexible, SS 304, optional PVC covered
Window Instrument glass, optional laminated safety glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black, optional micrometer pointer others on request
Movement Stainless steel 304
Accuracy Class 1
Protection IP65
Filling liquids Silicone
Accessories, options Thermowells, dual scales, magnet spring- or inductive contacts, calibration certification

All specifications are subject to change without notice.


IN.MTS Rev. L 19/05/2017
Page 53
General dimensions in mm

Order information
Size Bulb Length of Length of Process Connection Connection Range Engineer- Options
diameter bulb/stem capillary connection size orientation ing units
(100) (06) 6 mm 115 to (R) Rigid (T) Turning nut (50) 1/2” BSP (L) Lower 0/ 60 °C (NH) Tagging wired
100 mm 4000 mm2) stem 0/ 80
(08) 8 mm (CS2) Adjustable (75) 3/4” BSP (B) Back 0/ 100 (DA) Dial marking
(150) (A) Capillary union 0/ 120 (CS) Dual scale
150mm (10) 10 mm SS armored1) connection (10) 1” BSP (E1) Every 0/ 160
sliding on angle 0/ 200 (SG) Safety glass
(12) 12 mm (P) Capillary capillary (50N) 1/2” NPT 0/ 250
0/ 300 (MP) Micro pointer
SS armored
with PVC (CS3) Adjustable (75N) 3/4” NPT 0/ 400 (FX) Front panel
covering1) union 0/ 500
connection (10N) 1” NPT 0/ 600 (FW) Surface plate
sliding on
(UF) U-clamp
stem (M20) M20x1.5 -10/ 50
-10/ 110 (TM) 2" pipe
(M32) M32x2 -20/ 40 mounting
-20/ 60 bracket
-20/ 100
-20/ 120 (EC) Contact type
-20/ 180 and function
-30/ 70 (see EC.1)
-30/ 50
(GV) Silicon (max.
-30/ 70
up to 300˚C)
-30/ 170
-38/ 50 (YW) 316L housing
1) Max length
30m, specify
length in m °F and dual
others on EG. A1.5 =A scale on other options on request
request with 1.5m others on request others on request request
2) Minimum stem length depends on stem diameter refer table “T1”

How to order
Size Type Bulb Length of Length of Process Connection Connection Range Engineering Options
diameter bulb/stem capillary connection size orientation unit
100 MTS 08 125 R CS2 50N L 0/100 C FW

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91- 2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 54
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All stainless steel gas actuated thermometer
Model GTS
Nominal size 100 mm, 115mm or 150 mm
Accuracy: Class 1 (1%)

Features
x Rugged stainless steel construction
x Fast response
x Protection IP65
x High repeatability and small hysteresis
x Dry or liquid filled
x Rigid stem or bulb with capillary

Ranges
-50 ... 50 °C up to 0 ... 600 °C
-60 ... 120 °F up to 30 ... 1100 °F

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification Rigid stem With capillary


Dial size in mm 100 115 150 100 115 150
Construction Cylindrical case with bayonet ring
Measuring principle Inert gas actuated
Range in °C 0/60 0/80 0/100 0/120 0/160 0/200 0/250 0/300 0/400 0/500 0/600
-10/50 -10/110 -20/40 -20/100 -20/120 -20/180 -30/50 -30/70 -30/170 -40/40
-40/60 -40/160 -50/50
Ranges in °F and dual scales on request
Over temperature limit Max. 130 % F.S. Subjected to Max. 800°C Over temperature
Stem or bulb diameter 6 mm, 8 mm,10 mm, 12mm
Lengths 100 ... 4000 mm, other length on request, min. length depends on bulb diameter
Capillary length N.A. Max. 30 m
Process connection Plain
1/4” 3/4” 1” BSP / NPT (M), M20x1.5, M24x1.5, M27x2
Fixed, adjustable or with swivel nut
Connection location Back, lower, knee joint bendable (180°), knee joint bendable (360°)
Material
Connection Stainless steel 316
Stem Stainless steel 316L
Case/ring Stainless steel 304 , optional 316L
Capillary Stainless steel 316
Armor Flexible, SS 304, optional PVC covered
Window Instrument glass, optional laminated safety glass
Dial Aluminum, black markings on white background
Pointer Aluminum, black, optional micrometer pointer others on request
Movement Stainless steel 304
Accuracy Class 1
Protection IP65
Filling liquids Silicone
Weight dry/filled in kg Depend on dial size length of stem and capillary
Accessories, options Thermowells, dual scales, magnet spring- or inductive contacts, calibration certification

All specifications are subject to change without notice.


Page 55
IN.GTS Rev. I 19/05/2017
General dimensions in mm

Order information
Size Bulb Length of Length of Process Connection Connection Range Engineer Options
diameter bulb/stem capillary connection size orientation ing units
(100) (06) 6 mm 115 to (R) Rigid (T) Turning (50) 1/2” BSP (L) Lower 0/ 60 °C (NH) Tagging wired
100 mm 4000 mm2) stem nut 0/ 80
(08) 8 mm (75) 3/4” BSP (B) Back 0/ 100 (DA) Dial marking
(115) (A) Capillary (CS2) Adjustable 0/ 120 (CS) Dual scale
115mm (10) 10 mm SS armored1) union (10) 1” BSP (E1) Every 0/ 160
connection angle 0/ 200 (SG) Safety glass
(150) (12) 12 mm sliding on (50N) 1/2” NPT 0/ 250
150mm (P) Capillary (MP) Micro pointer
SS armored capillary 0/ 300
with PVC (75N) 3/4” NPT 0/ 400 (FX) Front panel
covering1) (CS3) Adjustable 0/ 500
union (10N) 1” NPT 0/ 600 (FW) Surface plate
connection
(UF) U-clamp
sliding on (M20) M20x1.5 -10/ 50
stem -10/ 110 (TM) 2" pipe
(M32) M32x2 -20/ 40 mounting
-20/ 100 bracket
-20/ 120
-20/ 180 (EC) Contact type
-30/ 50 and function
-30/ 70 (see EC.1)
-30/ 170
(GR) Glycerin (max.
-40/ 40
up to 100˚C)
-40/ 60
-40/ 160 (GV) Silicon (max.
-50/ 50 up to 300˚C)
(YW) 316L housing
1) Max length
30m, specify
length in m °F and dual
EG. A1.5 =A scale on other options on request
with 1.5m others on request others on request request
2) Minimum stem length depends on stem diameter refer above table

How to order
Size Type Bulb Length of Length of Process Connection Connection Range Engineering Options
diameter bulb/stem capillary connection size orientation unit
100 GTS 08 125 R CS3 50N L 0/100 °C FW

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 56
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Resistant temperature detector with head
Model RTH
General purpose and Ex-proof

Features
x High stability sensor
x Mineral insulated for flexibility and durability
x General purpose with spring loaded sensor
x Available with weather proof and explosion proof head
x Custom designs for specific applications

Ranges
-200 up to 600 °C
-325 up to 1100 °F

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Pharmaceuticals, Dairy & Food process industry
Pulp and paper industry

Technical specification RTH


Head types Compact, Weatherproof and Ex-proof heads.
Element Pt100, Pt500, Pt1000
Element type Simplex or duplex
Measuring principle
Range From -200 to 600 oC
Sheath diameter 6mm, 8mm, others on request
Sheath length Rigid up to 1000mm,flexible with MI cable 6X5 upto 20000mm
Process connection 1/4”, 1/2”, 3/4” NPT/BSP, adjustable, Nipple or Nipple union nipple connections
Material
Pressure connection Stainless steel 316, optional SS321
Sheath Stainless steel 316, optional SS 321
Housing Aluminum, epoxy coated
Electrical specification
Wiring configuration 2 wire, 3 wire and 4 wire
Lead wire 26-gauge standard copper wire, silver plated
Terminal Ceramic block with nickel plated brass terminal, optional spring loaded terminal
Ex-proof Zone 1 & 2 IIA / IIB / IIC T6 Enclosure Ex-d IEC 60079-1:2007
Insulation 100 X 106 ohms minimum insulation when measured at 500 VDC
Electrical connection 3/4” ET(F) for compact head, 1/2” NPT(F) for WP & EX, others on request.
General specification
Accuracy Class “A” and Class “B” as per DIN EN 60751, optional 1/3rd & 1/10th DIN
Stability ±0.04 maximum ice-point resistance shift following 1,000 hrs at maximum specified temperature
Protection IEC 60529 IP65, IP66 for Ex-Proof
Accessories, options Thermowell, Head mounted transmitters, etc.

All specifications are subject to change without notice.


Page 57
IN.RTH Rev. E 28/12/2016
General dimensions in mm

Order information
Head Type Ele- Element Accuracy Sheath Sheath Sheath Process Connection Electrical connec- Options
ment type class material diameter length connection Type tion
th
(E) Compact Pt100 (S2) Simplex (1) 1/10 Din (S) SS316 (D6) 55 to (=) None (=) None 3) (NH) Tagging wired
(=) Standard
weatherproof 2 wire 6 mm 1000
head Pt500 (3) 1/3rd Din (SA) SS321 mm (02) (F) Fixed Hex (JM) M20X1.5 mm (ST) Spring loaded
female 4)
(S3) Simplex (D8) 1/4” NPT(M) nut terminal
(W) Weather Pt1000 3 wire (A) Class ‘A’ 8 mm (B1) Ni plated
2)
proof -70/300C (04) (NS) Nipple brass double (HT) Head mounted
(S4) Simplex (D10) 1/2” NPT(M) compression transmitter
(B) Ex proof 4 wire (B) Class ‘B’ 10 mm (US) Nipple cable gland7)
enclosure -70/500C (05) union (DE) Double entry for
IIA/IIB certi- (D2) Duplex 3/4” NPT(M) 2) (C1) Ni plated 5)
Nipple head
fied 2 wire (EA) Extend- brass Ex
ed tempera- (13) proof(IIA/IIB)
(D) adjustable (CM) MI cable
(C) Ex proof (D3) Duplex ture Class A 1/4” BSP(M) cable gland 7) 1)
connection extention
enclosure 3 wire -200/600C sliding on (D1) Ni plated brass
IIC certified (15) stem Ex proof(IIC) (CA) Stainless steel
1/2” BSP(M) cable gland7) armored PTFE
(F1) SS double insulated
(17) compression extension
1)6)
3/4” BSP(M) cable gland7) cable

(16) (G1) SS Ex proof (TTS) Teflon- Teflon,


M20x1.5(M) (IIA/IIB) cable SS braided ex-
7)
gland tension cable
(H1)SS Ex proof Ex
proof(IIC) cable
gland 7)
(I1) Plastic cable
7)
gland PG13.5

1) Specify length in meter EG. CM1.5 = CM with 1.5m, Max length 20m, only for sheath diameter 6mm
2) Only with connection 1/2” NPT/BSP 3) 3/4” ET for head (E) & 1/2” NPT for others 4) only with weather proof head 5) not for Head type (E)
6) Max temperature limit for extension cable is 250°C 7) For two cable gland, specify the codes as B2, C2… applicable only for (DE)

How to order
Head Type Model Element Element Accuracy Sheath Sheath Sheath Process Connection Electrical Option
type Class Material Diameter Length Connection Type Connection

C RTH Pt100 S3 A S D6 100 04 D = ST

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 58
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Resistant temperature detector without head
Model RTW
General purpose with or without extension cable

Features
x High stability sensor
x Mineral insulated for flexibility and durability
x Custom designs for specific applications

Ranges
-200 up to 600 °C
-325 up to 1100 °F

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Pharmaceuticals, Dairy & Food process industry
Pulp and paper industry
Storage Tanks
Power generation
Reactors

Technical specification RTW


Element Pt100, Pt500, Pt1000
Element type Simplex or duplex
Measuring principle
Range From -200 to 600 oC
Sheath diameter 4mm, 6mm, 8mm, others on request
Sheath length Rigid up to 1000mm,flexible with MI cable 6X5 upto 20000mm
Process connection 1/4”, 1/2”, 3/4” NPT/BSP, Fix, adjustable
Material
Pressure connection Stainless steel 316, optional SS321
Sheath Stainless steel 316, optional SS 321, Inconel 600
Housing Aluminum, epoxy coated
Electrical specification
Wiring configuration 2 wire, 3 wire and 4 wire
Lead wire 26-gauge standard copper wire, silver plated
Insulation 100 X 106 ohms minimum insulation when measured at 500 VDC
General specification
Accuracy Class “A” and Class “B” as per DIN EN 60751, optional 1/3rd & 1/10th DIN
Stability ±0.04 maximum ice-point resistance shift following 1,000 hrs at maximum specified temperature

All specifications are subject to change without notice.


Page 59
IN.RTW Rev. A 04/01/2017
General dimensions in mm

Order information
Model Element Element Accuracy Sheath Sheath Sheath Process Connection Extension Cable Option
type class material diameter length connection Type
th
RTW Pt100 (S2) Simplex (1) 1/10 Din (S) SS316 (D4) 4mm 55 to 1000 (=) None (=) None (=) None (NH) Tagging wired
2 wire mm
rd
Pt500 (3) 1/3 Din (SA) SS321 (D6) 6mm (02)1/4” NPT(M) (F) Fixed Hex (CM) MI cable exten-
(S3) Simplex nut sion
Pt1000 3 wire (A) Class ‘A’ (I6) Inconel (D8) 8mm (04)1/2” NPT(M)
-70/300C 600 (D) Adjustable (CA) Stainless steel
(S4) Simplex (D10) 10mm (05)3/4” NPT(M) connection armored PTFE
4 wire (B) Class ‘B’ sliding on insulated exten-
-70/500C (13)1/4” BSP(M) stem sion cable
(D2) Duplex
2 wire (EA) Ex- (15)1/2” BSP(M) (TT) Teflon- Teflon,
tended extension cable
(D3) Duplex temperature (17)3/4” BSP(M)
3 wire Class A (TTS) Teflon- Teflon,
-200/600C (16)M20x1.5(M) SS braided ex-
tension cable

(FF) Fiber- Fiber, ex-


tension cable

Others on request

1) Specify length of extension cable in meter EG. CM1.5 = CM with 1.5m, Max length 20m, only for sheath diameter 6mm

How to order
Model Element Element Accuracy Sheath Sheath Sheath Process Connection Extension Ca- Option
type Class Material Diameter Length Connection Type ble

RTW PT100 S3 A S D6 100 04 D TT NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +912764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 60
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Thermocouple with Head
Model TCH
General purpose and Ex-proof

Features
x High quality TC cable
x Mineral insulated for flexibility and durability
x General purpose with spring loaded sensor
x Available with weather proof and explosion proof head
x Custom designs for specific applications

Ranges
-200 up to 1200 °C
-325 up to 2200 °F

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Pharmaceuticals, Dairy & Food process industry
Pulp and paper industry

Technical specification TCH


Head types Compact, Weatherproof and Ex-proof heads.
Element J, K, T, E, N, others on request
Element Type Simples, Duplex
Junction Grounded, Ungrounded
Measuring principle
Range J type (-40 to 750°C, -40 to 1380°F), K type (-200 to 1200°C, -325 to 2200°F) ,
T type (-200 to 350°C, -325 to 660°F), E type (-200 to 900 °C, -325 to 1650 °F),
N type (-200 to 1200 °C, -325 to 2200 °F), Others on Request
Sheath diameter 3mm, 4.5mm, 6mm, 8mm
Sheath length up to 20000mm
Process connection 1/4”, 1/2”, 3/4” NPT/BSP, adjustable, Nipple or Nipple union nipple connections
Material
Pressure connection Stainless steel 316, optional SS321
Sheath Stainless steel 316, optional SS 321
Housing Aluminum, epoxy coated
Electrical specification
Terminal Ceramic block with nickel plated brass terminal optional spring loaded terminal
Ex-proof Zone 1 & 2 IIC T6 Enclosure Ex-d IEC 60079-1:2007
Insulation 100 X 106 ohms minimum insulation when measured at 500 VDC
Electrical connection 3/4” ET(F) for compact head, 1/2” NPT for WP & EX, others on request.
General specification
Accuracy (greater of) J K T E N
> 0 °C ±2.2°C or ±0.75% ±2.2°C or ±0.75% ±1.0°C or ±0.75% ±1.7°C or ±0.5% ±2.2°C or ±0.75%
< 0 °C ±2.2°C or ±2.0% ±1.0°C or ±1.5% ±1.7°C or ±1.0% ±2.2°C or ±2.0%
Protection IEC 60529 IP65, IP66 for Ex-Proof
Accessories, options Thermowell, Head mounted transmitters, etc.

All specifications are subject to change without notice.


Page 61
IN.TCH Rev. E 28/12/2016
General dimensions in mm

Order information
Head Type Element Element Junction Sheath Sheath Sheath Process Connection Electrical connec- Options
Type material diameter length connection Type tion
(E) Compact (J) Iron - (S) Simplex (G) Grounded (S) SS 316 (D3) 55 to (=) None (=) None 3) (NH) Tagging
1) (=) Standard
weatherproof Constantan 3mm 20000 mm wired
head (D) Duplex (U) Un- (SA) SS (02) (F) Fixed Hex nut (JM) M20X1.5 mm
(K) Chromel grounded 321 (D4.5) 1/4” NPT(M) female (ST) Spring
2)
(W) Weather - Alumel 4.5mm (NS) Nipple (B1) Ni plated loaded
4)
proof (04) brass double terminal
(T) Copper - (D6) 1/2” NPT(M) (US) Nipple union compression
(B) Ex proof Constantan 2)
6mm Nipple cable gland6) (HT) Head
enclosure (05) mounted
IIA/IIB certi- (E) Chromel 3/4” NPT(M) (D) adjustable (C1) Ni plated
(D8) transmitter
fied - Constantan brass Ex
8mm connection
(13) proof(IIA/IIB)
sliding on 6) (DE) Double entry
(C) Ex proof en- (N) Nicrosil - 1/4” BSP(M) cable gland 5)
stem for head
closure IIC Nisil (D1) Ni plated brass
certified (15) Ex proof(IIC)
1/2” BSP(M) cable gland6)
(F1) SS double
(17) compression
3/4” BSP(M) cable gland
6)

(16) (G1) SS Ex proof


M20x1.5(M) (IIA/IIB) cable
6)
gland
(H1)SS Ex proof Ex
proof(IIC) cable
gland6)
Others on
request (I1) Plastic cable
6)
gland PG13.5

1) Element type simplex only and max length 5000mm 2) Only with connection 1/2” NPT/BSP 3) 3/4” ET for head (E) & 1/2” NPT for others 4) Only with weather proof head
5) not for Head type (E) 6) for two cable gland, specify the codes as B2, C2… applicable only for (DE)

How to order
Head Type Model Element Element Junction Sheath Sheath Sheath Process Connection Electrical Option
Type Material Diameter Length Connection Type Connection

E TCH K S G S D6 100 04 D = ST

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 62
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Screwed Thermowells
Model TWT
Features
x Solid bore construction
x Materials according NACE MR 01-75/ISO 15156
x Selection of materials
x Standard and customer specifications
x ISO 228-1 and ANSI/ASME B1.20.1 connections

Ratings
from 620 bar at 20 °C
to 175 bar at 650 °C

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification S R H
Solid bore design

Construction shank Straight Stepped Tapered


Bore diameter in mm 6.6 7.0 9,8 10.5 12.5, others on request
Depending on process connection (½", ¾" or 1") and bore diameter (0,260"/6,6 mm) for
standard
Pressure rating
thermowell material stainless steel 316L (1.4404)
(These data don’t apply to other materials)
at process connection in inch ¾, 1 ½ ¾, 1
°F °C PSI BAR PSI BAR PSI BAR PSI BAR
70 21 6500 448 6000 413 6000 413 9000 620
200 93 5500 379 5200 358 5200 358 7700 531
400 204 5000 345 4500 310 4500 310 6700 462
600 316 4000 275 4000 275 4000 275 5800 400
800 427 3800 262 3500 241 3500 241 5300 365
1000 538 3500 241 3300 227 3300 227 5000 345
1200 649 2500 172 2500 172 2500 172 3700 255
Bore 0,385 (9,8 mm) 40 % less
Process connection ½” BSP (M), ¾” BSP (M), 1” BSP(M)
½” NPT (M), ¾” NPT (M), 1” NPT (M), 1¼” NPT (M)
Others on request
Instrument connection ½” NPT (F) or ½” BSP (F)
Material Stainless steel 304 (1.4301), stainless steel 304L (1.4306), stainless steel 321 (1.4541)
Stainless steel 316 (1.4401), stainless steel 316L (1.4404), stainless steel 316Ti (1.4571)
Hastelloy B, Hastelloy C, Monel, Duplex, Nickel, Inconel
Other materials or coatings on request
Certifications & tests Material certificate 3.1.B to EN 10 204, hydrostatic test, dye penetration test, X-ray, hardness
test,
NACE compliance certificate
Accessories, options Thermometers, RTD, thermocouples

All specifications are subject to change without notice.


Page 63
IN.TWT Rev. G 01/09/2015
General dimensions in mm

Order information
Process U-length Instrument Extension Shank type Bore Material Options
connection in mm connection length
(07) 1¼” NPT(M) (U…) (04T) 1/2” (=) Standard, (S) Straight (6.6) 6.6mm (B) Carbon steel (NH) Tagging stamped
Specify NPT(F) 45mm
(06) 1” NPT(M) length in (R) Stepped (7.0) 7.0mm (C) 304 (1.4301) (CC) Cap & chain
mm2) (15T) 1/2” (EX) Specify stainless steel
(05) 3/4” NPT(M) BSP(F) extension (H) Tapered (8.5) 8.5mm (CL) 304L
E.g. (250) for length in (1.4306) (BP) Buffed and
(04) 1/2” u length (9.8) 9.8mm
250mm mm polished
NPT(M)1) E.g. (E100) for (S) 316 (1.4401)
100mm extension (10.5) 10.5mm3)
(18) 1” BSP(M) length (SL) 316L
(12.5) 12.5mm3) (1.4404) Certification
(17) 3/4” BSP(M) (CDWFC) Wake frequency
(SH) 321 (1.4541) ASME PTC 19.3
(15) 1/2”
BSP(M)1) (ST) 316Ti (CDRH) Rockwell
(1.4571) hardness test
(33) M33 x 2(M)
(M) Monel (X-RAY) X-ray photo
(H) Hastelloy C (CDW9) 100 bar 1 min.
(G) Hastelloy B (CD3) 3.1 Material
certificate MR-01-75
(N) Nickel
(CD5) NACE MR 0175 /ISO
(DU) Duplex 15156-2003 MR-01-75
4)
(IN) Inconel
(CDMA) NACE MR 0175 /ISO
15156-2005 5)
others on request

others on request
1) Max bore diameter 9.8mm and for bore 7.0, 8.5 & 9.8 only shank type (S) straight is available
2) Max length for solid bar stock is 700mm for stainless steel and 600mm for other material, lengths above as fabricated bar stock
3) Not for shank type R (stepped) 4) only for material (H) and (DU) 5) only for material (SL)

How to order
Process Type U-length Instrument Extension Shank type Bore diameter Material Options
Connection connection length
05 TWT 250 04T = H 8.5 SL NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral 382 729 , Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91- 2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 64
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Flanged Thermowells
Model TWF
According ANSI/ASME or DIN 43772

Features
x Solid bore construction
x Materials according ISO 15156/NACE MR 01-75
x Selection of materials
x Standard and customer specifications
x DIN and ANSI/ASME process connections

Ratings
DIN up to PN100
ANSI/ASME up to 1500 lbs

Applications
Chemical and petrochemical industry
Machine and apparatus construction
Food and beverage industry
Pulp and paper industry

Technical specification S R H
Solid bore design

Shank solid bore, flange welded on shank standard. Optional one piece design, built up design
or to customer specification
Construction shank Straight Stepped Tapered

Bore diameter in mm 6.6 7.0 9,8 10.5 12.5, others on request


Pressure rating Depending on flange rating
Process connection
DIN
Size DN25, DN40, DN50, DN80, DN100, others on request
Pressure class PN10, PN16, PN25, PN40, PN64, PN100, others on request
ANSI/ASME
Size [inch] 1, 1½, 2, 3, 4, others on request
Pressure class [lbs] 150, 300, 600, 900, 1500
Facing Raised face, Flat face, Ring joint type)
Other forms, connections or standards are available on request
Instrument connection ½” NPT (F) or ½” BSP (F)
Material Carbon steel, Stainless steel 304 (1.4301), 304L (1.4306), 316 (1.4401), 316L (1.4404), 316Ti
(1.4571), 321 (1.4541)
Monel, Hastelloy B, Hastelloy C, Inconel, Tantalum, Duplex, Nickel
Other materials or coatings on request
Certifications & tests Material certificate DIN EN 10 204 3.1, Hydrostatic test, dye penetration test, X-ray, Hardness
test,
ISO 15156/NACE certificate, stress calculation according to ASME PTC 19.3
Accessories, options Thermometers, RTD, thermocouples

All specifications are subject to change without notice.


Page 65
IN.TWF Rev. G 19/05/2017
General dimensions in mm

Order information
Process U- Instrument Extension Type Bore Material Flange Pressu Options
connecti length connection Length facing re
on in mm rating
ANSI (U…) (04T) 1/2” (=) Standard, (H) Standard (6.6) 6.6mm (B) Carbon ANSI ANSI (NH) Tagging stamped
Flanged Specify NPT(F) 57mm with steel Flange Flange
length in Welded (7.0) 7.0mm (CC) Cap & chain
(10) 1” mm2) (15T) 1/2” (EX) Specify Flange (C) 304 (RF) 150 stainless steel
BSP(F) extension Tapered (8.5) 8.5mm (1.4301) Raised 300 (BP) Buffed and polished
(15) 1 ½” E.g. (250) length in Shank face 600
for u length (9.8) 9.8mm (CL) 304L (FL.) Specify flange as
(20) 2” 250mm mm (1.4306) (FF)Flat 900 per material code
(S) Welded 1500
(30) 3” E.g. (E100) for (10.5) 10.5mm1) face 2) (e.g. FLM flange material
Flange (S) 316 Monel)
100mm extension
length Straight (12.5)12.5mm1) (1.4401) (RTJ) Certification
Shank Ring joint
(SL) 316L (CDWFC) Wake frequency
(R) Welded (1.4404) type ASME PTC 19.3
Flange (CDW2) Dye penetration
(SH) 321
Stepped test
(1.4541)
Shank (CDRH) Rockwell
(only bore (ST) 316Ti hardness test
DIN DIN DIN
0,260 (1.4571)
Flanged (6,6mm)) flange flange (BC) Bore
(M) Monel Concentricity
(DN25) the flange PN10 test
(H) Hastelloy facing depends
(DN40) on the process PN16 (X-R1) X-ray for weld
C connection PN25 joint
(DN50) (nominal size PN40
(N) Nickel and nominal (X-R2) X-ray for bore
pressure) PN64
(DN80) PN100 concentricity
(DU) Duplex
(DN100) (CDW9) 100 bar 1 min.
(IN) Inconel
(CD3) 3.1 Material
flange material is certificate
the same as the
shank material (CD5) NACE
other construction unless specified
such as: Solid MR 01-75 /
with code (FL.),
machined, built up see options
ISO 15156-2003 3)
and to customer
others on specification on (CDMA) NACE
request request MR 0175 /
ISO 15156-2005 4)
1) Not for shank type R (stepped) 2) only for size < 3” 3) only for material (H) and (DU) 4) only for material (SL)
How to order
Process Type U-length Instrument Extension Shank type Bore Material Flange Pressure Options
connection connection length diameter facing rating
10 TWF 250 04T = H 9.8 SL RF 150 NH

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral 382 729 , Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 66
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
Order information
Magnetic spring contacts Electrical contact devices
Contact Switch function Typical diagram
code at increasing For pressure and temperature gauges
process For dial size 100 and 150 mm
Single contact
EC1000 Contact closes Switch rating : DC max 30W, AC max 50VA
(NO) Dead Band : ± 2 to 4% FS
Adjustable range : Over full scale

EC2000 Contact opens


(NC)

Dual contacts Measuring type Pressure Diff. pressure Temperature


EC1100 Contact 1 closes Measuring principle Bourdon Diaphragm Diaphragm Gas actuated
(NOxNO) Contact 2 closes tube
Minimum range in [kg/cm2] [mmWC] [mmWC] [°C]
For dial size 100&150 mm
EC2200 Contact 1 opens and 1 magnetic contact 1,0 1000 1000 All ranges are
(NCxNC) Contact 2 opens 2 magnetic contacts 1,6 1000 1000 possible
3 magnetic contacts* 4,0 2500 2500
* For triple magnetic contacts consult factory
EC1200 Contact 1 closes How to order
(NOxNC) Contact 2 opens
`

EC2100 Contact 1 opens


(NCxNO) Contact 2 closes

Triple contacts
EC1110 Contact 1 closes
(NOxNOxNO) Contact 2 closes
Contact 3 closes

EC2220 Contact 1 opens


(NCxNCxNC) Contact 2 opens
Switching capacity graphs for contact devices
Contact 3 opens

240
EC1220 Contact 1 closes
(NOxNCxNC) Contact 2 opens Magnetic spring contact
220
Contact 3 opens Max. Load: 30 W-50 VA-1 A
200
Max. switching voltage: 250 V
EC2110 Contact 1 opens
(NCxNOxNO) Contact 2 closes
180
Contact 3 closes
160
EC1210 Contact 1 closes
Switching voltage in V

(NOxNCxNO) Contact 2 opens


140
Contact 3 closes
120
EC2120 Contact 1 opens
(NCxNOxNC) Contact 2 closes 100
Contact 3 opens
80
EC1120 Contact 1 closes
(NOxNOxNC) Contact 2 closes 60
Contact 3 opens 50 VA (AC)
40
EC2210 Contact 1 opens
(NCxNCxNO) Contact 2 opens
30 W (DC)
20
Contact 3 closes
0
0 100 200 300 400 500 600 700 800 900 1000
Switching current in mA

Contact type
EC1200 Add contact code to the selected gauge coding

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral 382 729 , Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 67
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN.EC Rev. B 01/09/2015
Syphon
Model SY1
Features
x Similar to DIN 16 282/ Industrial standards
x Male thread end for process side and weld end on process side
x Protect pressure measuring instrument form direct heating.
x U-Type for horizontal tapping
x Q-Type for vertical tapping
Applications
To protect pressure measuring instrument excessive heating
For liquids, gas or steam
Mounted on Shut-off valve or stop cock
Reduce the effect of rapid pressure surges

Permissible Pressure and Temperature Rating


100 kg/cm2 @ 100°C
95 kg/cm2 @ 150°C
85 kg/cm2 @ 250°C
80 kg/cm2 @ 300°C
70 kg/cm2 @ 350°C
63 kg/cm2 @ 400°C

General dimensions in mm

U-type Q-type

Order information
Type Connection Form1) System material Design/schedule Options

SY1 (04)1/2” NPT (M/F) Male X female (C) SS 304 (U40) U type SCH 40 (NH) Tagging Wired
(15)1/2” BSP (M/M)Male X Male (S) SS 316 (Q40) Q type SCH 40 (CDIB) IBR Certified
(16)M20x1.5 (F/F) Female X female (L) SS 316L (Q80) Q type SCH 80 (CD3) Material certificate
(F/B) Female X Blank for weld (CS) Carbon steel for wetted parts
(M/B) Male X Blank for weld ASTM A106
Others on request
1) For Form, First character indicates Process connection and Second character indicates Instrument connection

How to order
Type Connection Form System Design / Options
material schedule

SY1 04 M/F S U40 CDIB

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 68
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN.SY1 Rev. B 28/01/2017
Cooling Tower
Model CT1/CT5
Features
x Protect pressure instruments from high temperature
x Used where the service temperature exceeds the
rating of the instrument
x Stainless steel construction which is compatible
with many process media
x Perforated and finned design

Applications
Process plant, Power plant
Chemical and Petrochemical Industries
Off shore, on shore refineries
Clean air, gas and non-crystallizing media

Technical specification CT1 CT5


Operating principle Ambient air convection
Type Perforated design Finned design
Connection 1/4” 3/8” 1/2” BSP / NPT (M) , M20x1.5 (M)
Body Stainless steel 316L
Tube Stainless steel 316
Permissible pressure ratting 100 kg/cm2
Permissible temperature ratting 500°C
Weight in kg 0.4 0.55
Mounted below Pressure gauge, Pressure switch, Pressure Transmitter/Transducer

General dimensions in mm

CT1 CT5
Order information
Type Connection Form1) Options

CT1 (02)1/4” NPT (13)1/4” BSP (M/F) Male X female (NH) Tagging Stamped
CT5 (03)3/8” NPT (14)3/8” BSP (OF) Oil free
(04)1/2” NPT (15)1/2” BSP (CD3) Material certificate for wetted parts
(16)M20x1.5 (CD5) NACE MR 0175/ ISO15156 -2005
Others on request
1) For Form, First character indicates Process connection and Second character indicates Instrument connection

How to order
Type Connection Form Option

CT1 04 M/F CD3

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 69
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN.CT1/CT5 Rev. A 07/02/2017
Pulsation dampener / Snubber
Model PD1
Features
x Protects pressure instrument from line surges
x Reduce pulsation of process fluid
x Stainless steel body optional monel
x Improve reading and accuracy of instruments
x Horizontal and verticle installation

Applications
Reciprocating pumps and compressor
Hydraulic Machines and fluid power system
Process plant, Power plant
Chemical and Petrochemical Industries
Off shore and on shore refineries
Clean air, gas and non-crystallizing media

Permissible Pressure rating:


6000psi @ 100°C

General dimensions in mm

Order information
Type Connection Form1) System Options
material
PD1 (02)1/4” NPT (M/F) Male X female (C) SS 304 (NH) Tagging Stamped
(04)1/2” NPT (S) SS 316 (OF) Oil free
(13)1/4” BSP (L) SS 316L (CD3) Material certificate for
wetted parts
(14)3/8” BSP (P) Monel (10K) Pressure ratting
(15)1/2” BSP 10000PSI
(16)M20x1.5
Others on request
1) For Form, First character indicates Process connection and Second character indicates Instrument connection

How to order
Type Connection Form System Option
material
PD1 04 M/F S CD3

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 70
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN.PD1 Rev. B 28/01/2017
Pressure limiter / Gauge saver
Model PL1
Features
x Metal Compact design
x Various different adjustable ranges
x Stainless steel body optional monel
x Protect Instrument form over pressure
x Limits the excessive pressure and sudden spike
x Blocks higher pressure exceeding set value until pressure
x come to its normal value
x Isolate low pressure indicator from a system that requires to read
alternatively low & high pressure.

Applications
Machine and apparatus construction
Process plant, Power plant
Chemical and Petrochemical Industries
Off shore, on shore refineries and Mining

Permissible Pressure rating:


400 kg/cm2 @ 100°C

General dimensions in mm

Order information
Type Connection Form1) System material Adjustable range Options
(kg/cm2)
PL1 (04)1/2” NPT M/F Male X female (S) 316(1.4401) (A) 1.0 to 2.0 (NH) Tagging Stamped
(15)1/2” BSP (L) 316L (1.4404) (B) 2.0 to 4.0 (OF) Oil free
(16)M20x1.5 (P) Monel (C) 4.0 to 10.0 (CD3) Material certificate for wetted parts
(D) 10.0 to 40.0 (CD5) NACE MR 0175 /ISO 15156 -2005 2)
(E) 40.0 to 160.0
(F) 160.0 to 250.0
Others on request (G) 250.0 to 400.0
1) For Form, First character indicates Process connection and Second character indicates Instrument connection 2) Not for material (S), for material (L) restricted
up to 60oC and for material (P) unrestricted.

How to order
Type Connection Form System Range Options
material

PL1 04 M/F S A CD3

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 71
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN.PL1 Rev. B 28/01/2017
Gauge valves / Gauge cock
Model GV1/GV2/GV3
Features
x As per industrial standards
x Stainless steel body
x Provision of optional vent hole for drain
x Provide complete isolation of the media.
x Perfect alignment for positive air tight shut-off
x Designed for mounting below test pressure instrument

Applications
Isolation of instrument from service
Liquid and vapour service
Process industry, Power plant
Chemical and Petrochemical Industries
Off shore and on shore processes
Oil and gas refineries

Permissible Pressure rating:


70 kg/cm2 @ 100°C

General dimensions in mm

Order information
Type Connection Form1) System Options
material
GV1 Gauge isolation valve (04) 1/2” NPT (M/F) Male X female (C) SS 304 (NH) Tagging Stamped
GV2 Two way gauge cock (14) 3/8” BSP (F/F) Female X female (S) SS 316 (OF) Oil free
(CD3) Material certificate
GV3 Three way gauge cock (15) 1/2” BSP (L) SS 316L
for wetted parts
Others on request
1) For Form, First character indicates Process connection and Second character indicates Instrument connection

How to order
Type Connection Form System Options
material

GV1 04 M/F S CD3

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 72
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
All specifications are subject to change without notice. IN. GV1/GV2/GV3 Rev. B 28/01/2017
Needle valve
Model NV1
Features
x As per industrial standards
x Provide complete isolation or throttling of the media.
x perfect alignment for positive air tight shut-off
x optional vent plug for draining purpose
x Designed for mounting below Pressure
measuring instruments
x Stainless steel construction and optional
exotic material

Applications
Isolation of instrument from service
High pressure lines
Liquid and vapour service
Process industry, Power plant
Chemical and Petrochemical Industries
Off shore and on shore processes
Oil and gas refineries

Technical specification 2VM


Body Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Stem Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Valve assembly Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Bonnet Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Tip Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Packing PTFE up to 200 °C (grafoil up to 550 °C)
T-handle Stainless steel 304
Permissible pressure ratting 6000 PSI @ 240°C
Drain Port(Optional) M8 (F)
Vent plug Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276

Order information
Type Connection Form1) System material Options

NV1 (02) 1/4” NPT (M/F) Male X female (S) 316(1.4401) (NH) Tagging Stamped
(03) 3/8” NPT (F/F) Female X female (L) 316L (1.4404) (VP) Vent Plug
(OF) Oil free
(04) 1/2” NPT (M/M) Male X Male (P) Monel
(CD3) Material certificate for
(13) 1/4” BSP (H) Hastelloy-C276 wetted parts
(14) 3/8” BSP (CD5) NACE MR 0175/
(15) 1/2” BSP ISO15156 -2005
(16) M20x1.5
Others on request
1) For Form, First character indicates Process connection and Second character indicates Instrument connection

How to order
Type Connection Form System Options
material

NV1 04 M/F S CD3

All specifications are subject to change without notice.


Page 73
IN. NV1 Rev. B 28/01/2017
General dimensions in mm

Optional Vent or Drain design

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 74
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
2 valve manifold
Model 2VM

Features
x Stainless steel construction Optional Monel & Hastelloy-C276
x Isolates and bleeds process from Measuring Instrument assembly
x Provision of dust caps that prevent dust Ingress
x Supplied with wetted parts according to NACE MR0175
x The controlled vent provides the option of draining
to process fluid to colletor or atmospher
x PTFE packing optional grafoil

Applications RM-type
Chemical & petrochemical industries& refineries
Off shore and on shore fields
Storage tanks of fluids/gas
Thermal & nuclear power generation units
Waste treatment plant
OEM & machine apparatus
Seawater desalination plants

R-type

Technical specification 2VM


Body Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Stem Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Valve assembly Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Packing PTFE up to 200 °C (grafoil up to 550 °C)
T-handle Stainless steel 304
Permissible pressure ratting 6000 PSI @ 100°C
Drain Port ¼” NPT(F)
Vent plug Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276

Order information
Type Mounting Connection Form System material Options
Inlet X Outlet

2VM (RM) Remote mount (02) 1/4” NPT (M/F) Male X female1) (S) 316(1.4401) (NH) Tagging Stamped
(R) Pipe to pipe (04) 1/2” NPT (F/F) Female X female (OF) Oil free
(L) 316L (1.4404)
(03) 3/8” NPT (CD3) Material certificate
(13) 1/4” BSP (P) Monel for wetted parts
(T) Pipe to flange (14) 3/8” BSP (F/FL) Female X flange (10K) Pressure ratting
(H) Hasteloy-C276 10000PSI2)
(15) 1/2” BSP
(G) Grafoil packing
(H) Flange to flange (=) Flange (FL/FL) Flange X flange (CD5) NACE MR 0175/
Others on request ISO15156 -2005
1) Only for (RM) type 2) Detailed drawing shall be provided on request

How to order
Type Mounting Connection Form System Options
material

2VM R 04 F/F L CD5

All specifications are subject to change without notice.


Page 75
IN. 2VM Rev. B 28/01/2017
2 valve manifold

RM-type

R-type

T-type

H-type

Pressure Rating: 6000psi @ 100°C


Dimension shall be differ for option 10K
Precision Mass Products Pvt. Ltd.
Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91- 2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 76
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
3 valve manifold
Model 3VM
Features
x Stainless steel construction Monel & Hastelloy-C276
x Isolates each port and equalizes pressure
x Provision of dust caps that prevent dust Ingress
x Supplied with wetted parts according to NACE MR0175
x The controlled vent provides the option of draining
to process fluid to colletor or atmospher
x PTFE packing optional grafoil

Applications
Chemical & petrochemical industries, refineries
Off shore and on shore fields
Storage tanks of fluids/gas
Thermal & nuclear power generation units
Waste treatment plant
OEM & machine apparatus
Seawater desalination plants
Chemical injecttion skids

Technical specification 3VM


Body Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Stem Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Valve assembly Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Packing PTFE up to 200 °C (grafoil up to 550 °C)
T-handle Stainless steel 304
Permissible pressure ratting 6000 PSI @ 100°C
Drain Port ¼” NPT(F)
Vent plug Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276

Order information
Type Mounting Connection Form System material Options
Inlet X Outlet

3VM (R) Pipe to pipe (02) 1/4” NPT (F/F) Female X female (S) 316(1.4401) (NH) Tagging Stamped
(04) 1/2” NPT (OF) Oil free
(L) 316L (1.4404)
(03) 3/8” NPT (CD3) Material certificate
(13) 1/4” BSP (P) Monel for wetted parts
(T) Pipe to flange (F/FL) Female X flange (10K) Pressure ratting
(14) 3/8” BSP (H) Hasteloy-C276 10000PSI1)
(15) 1/2” BSP
(G) Grafoil packing
(H) Flange to flange (=) Flange (FL/FL) Flange X flange (CD5) NACE MR 0175/
Others on request ISO15156 -2005
1) Detailed drawing shall be provided on request

How to order
Type Mounting Connection Form System Options
material

3VM R 04 F/F L CD5

All specifications are subject to change without notice.


IN. 3VM Rev. B 28/01/2017 Page 77
3 valve manifold

R-Type

T-type

H-type

Pressure Rating: 6000psi @ 100°C


Dimension shall be differ for option 10K

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 78
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
5 valve manifold
Model 5VM
Features
x Stainless steel construction Optional Monel & Hastelloy-C276
x Isolates and vent the process and equalizes pressure
x Provision of dust caps that prevent dust Ingress
x Supplied with wetted parts according to NACE MR0175
x The controlled vent provides the option of draining
to process fluid to colletor or atmospher
x PTFE packing optional grafoil

Applications
Chemical & petrochemical industries, refineries
Off shore and on shore fields
Storage tanks of fluids/gas
Thermal & nuclear power generation units
Waste treatment plant
OEM & machine apparatus
Seawater desalination plants
Chemical injecttion skids

Technical specification 5VM


Body Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Stem Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Valve assembly Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276
Packing PTFE up to 200 °C (grafoil up to 550 °C)
T-handle Stainless steel 304
Permissible pressure ratting 6000 PSI @ 100°C
Drain Port ¼” NPT(F)
Vent plug Stainless steel 316 optional Stainless steel 316L, Monel & Hasteloy-C276

Order information
Type Mounting Connection Form System material Options
Inlet X Outlet

5VM (R) Pipe to pipe (02) 1/4” NPT (F/F) Female X female (S) 316(1.4401) (NH) Tagging Stamped
(04) 1/2” NPT (OF) Oil free
(L) 316L (1.4404)
(03) 3/8” NPT (CD3) Material certificate
(13) 1/4” BSP (P) Monel for wetted parts
(T) Pipe to flange (14) 3/8” BSP (F/FL) Female X flange (10K) Pressure ratting
(H) Hasteloy-C246 10000PSI2)
(15) 1/2” BSP
(G) Grafoil packing
(H) Flange to flange (=) Flange (FL/FL) Flange X flange (CD5) NACE MR 0175/
Others on request ISO15156 -2005
1) Detailed drawing shall be provided on request

How to order
Type Mounting Process Form System Options
Connection material

5VM R 04 F/F L CD5

All specifications are subject to change without notice.


Page 79
IN. 5VM Rev. B 28/01/2017
5 valve manifold

R-type

T-type

H-type

Pressure Rating: 6000psi @ 100°C


Dimension shall be differ for option 10K

Precision Mass Products Pvt. Ltd.


Plot No.: 2306, G.I.D.C. Chhatral, Ta: Kalol, Dist.: Gandhinagar, • North Zone : sales.delhi@precisionmass.com
Chhatral -382729, Gujarat, India. • East zone : sales.kolkata@precisionmass.com
Telephone: +91-2764 233681/82/83 Fax: +91-2764-233440 • West zone : sales.mumbai@precisionmass.com Page 80
Email: sales@precisionmass.com www.precisionmass.com • South zone : sales.chennai@precisionmass.com
NOTES
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________
___________________________________________________________

You might also like