Operating Manual
                          Model : MI-2100A
                             CONTENTS
                                                                        1 / 58
                      1. BEFORE INSTALLATION
1-1. Caution / Warning Marks
            Warning
                        This mark warns the possibility to arrive death or serious injury in
                        case of wrongly used.
            Caution
                         This mark cautions the possibility to arrive serious human body
                         injury or product lose in case of wrongly used.
1-4. Inquiries
  If you have any kinds of inquiries for this model, please contact with your local agent or Head
  Office.
  Head Office : Migun ST co.,Ltd.
  Overseas Office : Migun Corporation.
  Website : http://www.miguncorp.co.kr
  Email : info@miguncorp.co.kr / sales@miguncorp.co.kr
                                                                                            2 / 58
                                      2. INTRODUCTION
       2-1. Introduction
            Thank you for your choice, this “MI-2100A” Industrial Digital Weighing Indictor..
            This “MI-2100A” model is control purpose application usage Digital Weighing Indicator, with
            powerful communication performance.
            With 4pcs control relay outputs and High Speed A/D conversion performance will lead you to
            precise weighing process.
            This “MI-2100A” Weighing Indicator is control purpose application model, and it can be used for
            most kinds of control applications.
            Please review this instruction Manual and learn more about information about “MI-2100A”.
            Enjoy your process efficiency with “MI-2100A” Weighing Indicator..
       2-2. Cautions
            1). Don’t drop on the ground or avoid serious external damage on item.
            2). Don’t install under sunshine or heavy vibrated condition.
   !        3). Don’t install place where high voltage or heavy electric noise condition.
            4). When you connect with other devices, please turn off the power of item.
Cautions
            5). Avoid from water damage.
            6). For the improvement of function or performance, we can change item specification without
                prior notice or permission.
            7). Item’s performance will be up-dated continuously base on previous version’s performance.
            8) The quantity of load cell to connect is max. 6pcs.
           2-3. Features
            1). All Modules and Option Cards are isolated to maximize accuracy and performance.
            2). External input terminal inside.
            3). By using “Photo-Coupler” on each module(Option, Analog board, In/Out), we improved
                “Impedance problem”, “Isolation ability among inputs”, “Leading power problem”, and
                “Noise covering function”.
            4). Data back-up function, when the sudden power off
            5). Polycarbonate film panel, strong against dust and water
            6). RS-232C (Com. Port1) is standard installed.
             8). Variable options(Order in advance)
                                                                                                    3 / 58
                               3. SPECIFICATION
3-1. Analog Input & A/D Conversion
 Input Sensitivity                                                     0.45㎶ / Digit
 Load Cell Excitation                                             DC 10V ( - 5V ~ + 5V )
 Max. Signal Input Voltage                                               Max.32mV
                                                                     [Zero] ±20PPM/℃
 Temperature Coefficient
                                                                     [Span] ±20PPM/℃
                                                                                            4 / 58
3-3. General Specification
                                         Trans type Power Supply, AC110/220V
 Power Supply
                                            (DIP S/W Selectable)
 Operating Temperature Range             -10℃ ~ 40℃
 Operating Humidity Range                Under 85% Rh (non-condensing)
 External Dimension                      193mm(W)x100mm(H)x140mm(L)
 Net Weight(kg)                          About 2.5kg
 Gross Weight(kg)                        About 3.0kg
And this MI-2100A indicator can’t use a BCD IN and OUT card.
                                                                               5 / 58
3-5-1. Status Lamp (ANNUNCIATORS) : Green Color Lamp is “ON”.
Low When weight value reaches to LOW set value, Lamp is turn on.
                                                                                          6 / 58
Start Weighing process, under Packer Mode, only.
Set the Free Fall value for the High set value..
Free Fall Check : No. 8 key and if press Clear key it will be
           back..
Free Fall Change : No. 8 key + Number key + Enter
Under Print installation, you can print out the “Grand-total data” of
all P/N.
Printed Data : Accumulated count and weight of All P/N.
Manual Printer
 - Key input, print output.
Calibration mode
 - Digit setting
Whenever pressing “0”key, digit will be change 1, 2, 5, 10, and 50.
                                                               7 / 58
※ Function Keys (Combined Key functions :         key + other keys)
For combined key function, you should press “CLEAR” or “ENTER” + other key within 2 sec.
Sub-total Print
Tare Reset
                                                                                   8 / 58
3-6. Rear Panel
               ⑤        ④                       ⑥
    ①
               ①
                                                                                  9 / 58
                          4. INSTALLATION
4-1. External Dimension & Cutting Size
(External Dimension)   (unit : mm)
Cutting Size
                                            10 / 58
           4-2. Formula to plan the precise weighing system
           This “MI-2100A” weighing controller’s Max. input sensitivity is 0.45㎶ / Digit.
           And for precise weighing system, the following formula must be satisfied.
   !       Caution : “Input sensitivity” means Min. output voltage variation of weighing part to change
Cautions           1digit. So, please do not make large input voltage to make reliable weighing system.
                                                  ≤EⅹBⅹD
            Single Load cell use                                 A : Load cell capacity(kg)
                                       0.45㎶
                                                       A         B : Load cell Voltage(mV)
                                                                 D : Digit
                                                  ≤EⅹBⅹD
                                                                 E : affirmation Voltage of Load cell
            Plural Load cells use      0.45㎶                     N : Number of Load cell
                                                      AⅹN
           Example1.)
                     Number of Load cell : 1pcs
                     Load cell capacity : 500kg
                     Load cell Voltage : 2mV/V
                     Digit : 0.05kg
                     Affirmation Voltage of Load cell : 5.0V
                     Max. Capacity of Weighing System : 300kg
           Then, estimation result for this weighing system with formula,
               5000ⅹ2ⅹ0.05                            The calculated value is larger than 0.45㎶,
                                      = 1 ≥ 0.45㎶
                        500                           so this system has no problem.
           Example2.)
                     Number of Load cell : 4pcs
                     Load cell capacity : 500kg
                     Load cell Voltage : 2mV/V
                     Digit : 0.10kg
                     Affirmation Voltage of Load cell : 5.0V
                     Max. Capacity of Weighing System : 1,000kg
                                                                                                        11 / 58
                               Chapter 5. Set Up
 5-1. Calibration
Adjust weight balance between “Real weight” on the load cell(Weight Part) and “Displayed weight of
Indicator”. When you replace LOAD CELL or Indicator, you have to do Calibration process once again
                                       SET-CAL
Press          key during 5 sec                        display.
                                                                                           12 / 58
                                                                  TEST
Or turn on the Power + with pressing                                             display
                                  SET-CAL
Press                 key                           display
Remarks : Go to next step with save key / Back to previous step key
                                                            P        _1
Step 2. Multi-Calibration setting with 4 Section (                            display)
P _1 : Section number of Multi-calibration setting
2-1. If a real curve of load cell is straight line, the calibration section would be 1.
2-2. The multi calibration function is for compensate to load cell’s output by so using a multi-point at
some section when the real curve of load cell is not straight line as like under picture.
                                                                                                  13 / 58
                P        _1
2-3. Under                         display, if press a keypad No. 2, it will be change the section 2.
And, if press a keypad No. 3 or 4 or 1, it will be change the section 3 or 4 or 1.
                                      d        _1
Step 3. Digit(Division)setting (                        display)
                                          CAPA                  XXXXXX
Step 4. Max. Capacity Setting (                                                  display)
                                                                                                    14 / 58
Input Max. Capacity of Scale with No. keys.
                                   CAL _0_
Step 5. Zero Balance setting (                        display)
                                                                                              15 / 58
Input prepared Test weight value with No. keys.
                             C_UP_
Then, display will show                      and then, load prepared test weight unit on the scale.
※ Caution
For the precise Span calibration, please prepare Test weight unit, at least 10% of Max. capacity
of Scale.
                                                             XXXXXX                 C-END
Step 7. Check Span Value and Finish the Calibration(                                                 )
                     C-END
And after 3sec,                      will displayed automatically and move to weighing Mode.
                                                                                               16 / 58
 5-3. Simulation Calibration Mode (Without Test Weight) – Mode 2.
 - This calibration Method will be useful to make calibration more than 10ton capacity setting.
 - Guaranteed resolution will be 1/5,000 and if you need higher resolution, please make
calibration with Test weight.
                                       SET-CAL
Press          key during 5 sec                        display.
                                                           TEST
Or turn on the Power + with pressing                                  display
                                SET-CAL
Press               key                     display.
                                                                                         17 / 58
Press                key to start “Simulation Calibration Mode”.
Remarks : Go to next step with save key / Back to previous step key Step 2.
                                      CAL_ _2                 d     _1
 Step 2. Digit (Division)setting (                                           display)
                                                                                               18 / 58
                                              CAPA                 XXXXXX
Step 3. Max. Capacity of Load cell (                                                display)
                                                                     CAL _0_
Step 4. Measure/Adjustment optimal Zero balance of Scale (                             display)
                                                                                                  19 / 58
  Step 5. Input Max. Output Rate (mV/V) value of load cell
Input Max. Output Rate(mV/V) value of load cell with No. keys.
      - Under this step, input Max. Output rate(mV) of load cell.
                                                                    20 / 58
     - If you installed a few pieces of load cells, the connection will be parallel, so the rated output of
   a few load cells are as same as single load cell’s rated output.
      - The Output rate is stated on “Test report” or “Label”
- Calculated Span value will be displayed and automatically reset and move the normal
                                                                                                  21 / 58
5-4. Function Setting – Mode 1.
- Applicable model : MI – 1000/2000/3000/4000 series
To make more accuracy performance of Digital Weighing Indicator, through this Function setting.
                                                           TEST
Turn on the Power + with pressing                                      display
                              SET-CAL
Press               key                         display
                                                                                             22 / 58
Step 2. Change the F-Function No.
To change the F-function No., press          key. whenever pressing             key, Function No. will
be changed
If you want to move certain function No. directly, press function No. with keypad and press
key.
Input new set value with keypad, and press key to save new setting.
If you don’t press key, after changing the set value, the new set value will not be saved.
                                                                                              23 / 58
5-5. Function List
  Function
                                   Contents                                   Remark
    No.
F22 Weighing Finish Relay “ON” delay time setting Setting range : 0.0~9.9sec
F23 Weighing Finish Relay “ON” Duration time setting Setting range : 0.0~9.9sec
F24 Weighing Judge Relay “ON” delay time setting Setting range : 0.0~9.9sec
F25 Weighing Judge Relay “ON” Duration time setting Setting range : 0.0~9.9sec
                                                                                             24 / 58
Function
                     Contents                           Remark
  No.
                                                                    25 / 58
         5-6. Function List detailed information.
                                                 Decimal Point Setting
                        ●         0      No Decimal point
                                          st
                                  1      1 place under Zero (0.0)
            F01
                                          nd
                                  2      2 place under Zero (0.00)
                                          rd
                                  3      3 place under Zero (0.000)
                        ●         0      Normal mode
            F02
                                  1      Back up mode
         ※ Normal Mode
         When the power is off and will not be recovered within 1sec, weight memory will be out and display
         wrong weight value when power is recovered.
         Also, if there is material over 10% of max. capacity weight in the hopper, the “Un-Pass” display will be
         appear and you cannot process weighing job. Under this case, please remove material and turn on
         MI-2100A.
         If there is material less than 10% of max. capacity weight in the hopper, indicator will initialize
         with ”Pass” display, and display “0” whatever the real weight. Under this case, remove the material
         and press “Zero” key and input new “Zero” value. Then, new Zero value will be memorized and you
         can process weighing job. For “Manual initialization”, press “0” key or enter “Calibration mode”.
         ※ Weight Back up Mode
         Under this mode, MI-2100A memories “Zero” value of weighing part. So, when the power is recovered
         MI-2100A can display material weight.
                                               Motion Band Range setting
                                         This is set “Steady” acceptable range of weighing part.
                                         If there is vibration on weighing part, you can set this function and
                                  0
                                         reduce the vibration effect on weighing process.
            F03          5        ∫
                                                        0 : Weak vibration
                                  9
                                                        ∫
                                                        9 : Strong Vibration
         ※ This function is compensate vibration effect on weighing part with acceptable range setting.
         During the fixed time period(F06 setting), there is smaller weight variation than Motion band range,
         due to vibration, Indicator will display “Steady”, if there is larger weight variation than Motion band
         range, due to vibration, Indicator will its weight and “Steady” condition is broken and find new “Steady”
         point.
         If there are much vibration effect, please set with large set value.
                                                                             Steady Break point
Motion Band Range
                                                                                                             Motion Band
           Zero                                                                                                Range
                                                                                                          26 / 58
          ※ In this case, if you increase “Motion Band Range”(F03-03), the “Steady Break point” will be in the
          steady range, and indicator will display “Steady” condition. – (Set value “1” means 50% of Digit)
                                                                                                                  F03-03
Motion Band Range
                                                                                                                  F03-01
                                                                                                                Motion Band
        Zero                                                                                                      Range
Display Zero
                                                                                                      time
                              about 5msec.
                                                                                                             27 / 58
Motion Band Range                                                                                          F03-01
           +0.30kg
                                                                                                           Steady
                                                                                                           Range
          Zero
-0.30kg
                            00
                                   Small set value for weak vibration        Small     set   value     more
        F06          15      ∫
                                   Large set value for strong vibration      sensitive
                            49
                                                                                                     28 / 58
Weighing Mode Setting
                                                                                      29 / 58
             ◆ Weighing Mode 1. Limit Mode 1. (F21-01 setting)
Weight
                                                                        Finish Point
                                              HIGH – Free Fall
LOW
Near Zero
 TARE(IN)
                                                                                                          Output 1
  LOW
                                                                                                          Output 2
 HIGH
                                                                  t1    t2                                Output 3
FINISH
                                                                                                          Output 4
Near Zero
                                                                                                          30 / 58
         ◆ Weighing Mode 2. Packer Mode (F21-02 setting)
Relay “ON” when “Run” input, “OFF” when the weight reaches to set value.
Weight
                                                                             Finish Point
                                                HIGH – Free Fall
LOW
Near Zero
 RUN(IN)
                                                                                                             Output 1
  LOW
                                                                                                             Output 2
 HIGH
                                                                    t1       t2                              Output 3
FINISH
                                                                                                             Output 4
Near Zero
                                                                                                           31 / 58
◆ Weighing Mode 3. Comparison Mode (F21-03 setting) - Checker Mode 1.
Weight Judge, when weight value is stable over than Empty range.
Low Steady weight < SP1 (ON) OK(Finish) SP1 ≤ Steady weight ≤ SP2
                                                                                              32 / 58
◆ Weighing Mode 4. Comparison Mode (F21-04 setting) - Checker mode 2.
                                                                                     33 / 58
          “FINISH Relay” delay time(t1) setting (Under F21- 01, 02 setting)
                         After current weight is reached to FINAL, you can set some delay
                         time of “FINISH relay ON time.
                  00                                          Steady point
F22      10        ∫
                           FINISH Relay
                                                                              ComOut3
                  99     “00” setting : At Steady point, FINISH relay output
                         “20” setting : After 2.0sec from Steady point, FINISH relay output
                         “99” setting : After 9.9sec from Steady point, FINISH relay output
                  00
                                                              Steady point
F23      10        ∫
                                                              t2
                           FINISH Relay
                                                                               ComOut3
                  99     “01” setting : FINISH relay will be “ON during 0.1sec.
                         “20” setting : FINISH relay will be “ON” during 2.0sec.
      “STEADY” Judging delay time(t3) setting (Only for F21-03, Check mode1.)
                         After current weight is reached to each set point, you can set some
                         delay time of “STEADY”.
                                                              Steady point
                  00
F24 10 ∫ t3
                                                                               Com-Out3
                  99
                         “00” setting : At Steady point, FINISH relay output
                         “20” setting : During 2.0sec, hold “Steady” relay
                         “99” setting : During 9.9sec, hold “Steady” relay
      “STEADY” Judging “ON” time(t4) setting (Only for F21-03, Check mode1.)
                         After current weight is reached to each set point, you can set some
                         delay time of “STEADY” or after output of each set point relay
                         .
                  00
                                           Steady point
F25      10        ∫
                                                 t4
                                                                        Com-Out3
                  99
                         “00” setting : During or after the weight reaches to “Empty Range”.
                         “20” setting : During or after 2.0sec, Relay will be on.
                         “99” setting : During or after 9.9sec, Relay will be on.
                                                                                          34 / 58
Communication setting
0 600bps
1 1,200bps
2 2,400bps
3 4,800bps
F31 ● 4 9,600bps
5 19,200bps
6 38,400bps
7 57,600bps
8 115,200bps
                             Finish Mode : Single time data transfer, when the weight is finish
  F32                1
                              - When Finish Relay output, Data will be output.
2 Print Mode : Single time data transfer, when print key input
● 0 Simplex Mode
ID No. setting
            ●        0       Format 1.
  F35
                     1       Format 2. (Format 1 + time)
                                                                                             35 / 58
                         Data Transference count setting – Port 1(Standard)
0 About 40times/sec
1 About 30times/sec
2 About 20times/sec
4 About 10times/sec
5 About 5times/sec
6 About 3times/sec
F40 1 g
2 ton
● 0 200times/sec
F41 5 20times/sec
9 1time/sec
                                Continuous Print
              ●           0
                                Serial No. and Weight will be printed continuously.
   F43
                                Single Print
                          1
                                Date, Time, S/N, ID No. Weighing Data will be print
                                                                                           36 / 58
                    SUB/GRAND Total Data Delete selection
                0
F45   4         ∫      Whenever set value increased, 1line will be added.
                9
                0
F46   1         ∫      Whenever set value increased, 1line will be added.
                9
      ●         0      KOREAN
F47
                1      ENGLISH
               00     00 : No Delay time
F48   00        ∫
               99     99 : 9.9sec later, print output
      ●         0      Manual
F49
                1      Automatically
                                                                             37 / 58
Other Setting
                                        EMPTY Range setting
                                You can set “EMPTY” Range.
                                Within set range, indicator will not display current weight and just
                                display “Zero”.
                X.X.X.X.X.X.
   F80
                (0.0.0.0.1.0)   “0.000” setting : When Net Zero, “Zero” status lamp and Near Zero
                                                  relay will be output.
                                “0.190” setting : Within 190, “Zero” Status lamp and Near Zero relay
                                                  will be output.
   F89            XXXXXX         ※ If you have difficulty to process Calibration again, the best way to
                                 matching the net weight and display weight is doing Calibration
                                 process once again.
F90 Check Current DATE data or you can Change to new date
F91 Check Current TIME data or you can Change to new TIME
                                 At this function, you can check the Calculated Span value for multi
   F96            XXXXXX
                                 calibration 1.
                                 At this function, you can check the Calculated Span value for multi
   F97            XXXXXX
                                 calibration 2.
                                 At this function, you can check the Calculated Span value for multi
   F98            XXXXXX
                                 calibration 3.
                                                                                               38 / 58
                              Chapter 6. Interface
1. Rs-232C (Standard Installed)
RS-232C Serial Interface is sensitive/weak for electric Noise.
So, please isolate with AC power cable and use shield cable to reduce the electric noise effect.
1-1. Connection
                                                                                               39 / 58
1-3. Data Protocol (Data Format 1. – Total 18byte)
, , K g CR LF
, , , K g CR LF
 ▶ Header 2
    - NT : Net Weight (Without TARE Weight)
    - GS : Gross Weight (With TARE Weight)
 ▶ ID No. : Function 34 setting (Default No is 1)
 ▶ Lamp : Status Lamp Condition
   bit7         bit6         bit5           bit4         bit3        bit2             bit1           bit0
    1          Stable         1             Hold         Print      Gross            TARE            Zero
 ▶ DATA(8) Symbol(1) , Decimal Point(1) , Weight (6) = total 8BYTE, like +000.190
    - 2B(H): "+"PLUS
    - 2D(H): "-"MINUS
    - 2O(H): " "SPACE
    - 2E(H): "."Decimal point
   ▶ UNIT
    - Kg , g
                                                                                                     40 / 58
2. Current Loop Interface (Standard installed)
“Current Loop” Interface is stronger for Electric Noise than “RS-232C” interface.
So, it can be used for long distance communication.(About 100m long distance).
2-1. Connection
                                                                                              41 / 58
3. Rs-422 Serial Interface (Option)
RS-422/485 serial interface is more stable for electric noise effect compare with other communication
method, using electric current difference.
But, install isolated place from Power cable or other electric cables and wires, and please use shielded
cable for better performance.
Recommendable communication distance is about 1.2km.
3-1. Connection
3-3. Data Protocol (Data Format 1. – Total 18byte) - As same as “Rs-232c Serial Interface
, , K g CR LF
3-4. Data Protocol (Format 2 – Total 22byte) - As same as “Rs-232c Serial Interface
, , , K g CR LF
                                                                                                        42 / 58
▶ COMMAND MODE – MI-2000A
PCIndicator            (ASCII)
Format                                           (HEX)
                                                                                      Time Data
Response                           (ASCII)
from                                                                        (HEX)
Indicator
PCIndicator            (ASCII)
Format                                          (HEX)
                                                                                      Serial No.
Response                           (ASCII)
from                                                                        (HEX)
Indicator
PCIndicator            (ASCII)
Format                                           (HEX)
                                                                                      Code No.
Response                           (ASCII)
from                                                                        (HEX)
Indicator
PCIndicator            (ASCII)
Format                                          (HEX)
                                                                                          Part No.
Response                     (ASCII)
from                                                          (HEX)
Indicator
PCIndicator            (ASCII)
Format                                           (HEX)
                                                                                      Key TARE
Response                           (ASCII)                                           weight value
from                                                                        (HEX)
Indicator
PCIndicator            (ASCII)
Format                                           (HEX)
                                                                                      Current
Response                                        (ASCII)                              Weight value
from
Indicator
                                                                  (HEX)
                                                                                             43 / 58
Remark          STX(1) ID(2) Command(4) Status1(2) Status2(2) Symbol(1)
                Weight (Include decimal point)(7) Unit(2) ACK(1) ETX(1)
                = Total 23 BYTE
PCIndicator              (ASCII)
Format                                          (HEX)
                                                                                      SP1(Low)
Response                              (ASCII)                                          DATA
from                                                                      (HEX)
Indicator
PCIndicator              (ASCII)
Format                                          (HEX)
                                                                                     SP2(High)
Response                              (ASCII)                                          DATA
from                                                                      (HEX)
Indicator
PCIndicator              (ASCII)
Format                                          (HEX)
                                                                                         Free Fall
Response                              (ASCII)                                             DATA
from                                                                      (HEX)
Indicator
TxD
PCIndicator               (ASCII)
Format                                          (HEX)
                                                                                   TARE input
Response                   (ASCII)
from                                                (HEX)
Indicator
PCIndicator               (ASCII)
Format                                          (HEX)
                                                                                  TARE RESET
Response                   (ASCII)
from                                                (HEX)
Indicator
PCIndicator               (ASCII)
Format                                          (HEX)
                                                                                   ZERO input
Response                    (ASCII)
from                                                (HEX)
Indicator
PCIndicator              (ASCII)
                                                                                    Print input
Format                                          (HEX)
                                                                                            44 / 58
Response                   (ASCII)
from                                                 (HEX)
Indicator
PCIndicator              (ASCII)
Format                                           ( HEX)
                                                                          Sub-Total Print
Response                   (ASCII)
from                                                 (HEX)
Indicator
PCIndicator              (ASCII)
Format                                           (HEX)
                                                                           Grand-Total
Response                   (ASCII)                                            Print
from                                                 (HEX)
Indicator
PCIndicator                          (ASCII)
  Format                                                          (HEX)
PCIndicator                          (ASCII)
  Format                                                          (HEX)
PCIndicator                 (ASCII)
  Format                                                  (HEX)
                                                                                  45 / 58
PCIndicator                         (ASCII)
  Format                                                         (HEX)
PCIndicator              (ASCII)
Format                                           (HEX)                    Start(Run) Input
Response                   (ASCII)                                           (F21 – 02)
from                                                                      (PACK MODE)
                                                     (HEX)
Indicator
PCIndicator              (ASCII)
Format                                           (HEX)                      STOP Input
Response                   (ASCII)
                                                                            (F21 – 02)
from                                                                      (PACK MODE)
                                                     (HEX)
Indicator
PCIndicator                         (ASCII)
  Format                                                         (HEX)
PCIndicator                         (ASCII)
  Format                                                         (HEX)
PCIndicator                           (ASCII)
  Format                                                          (HEX)
                                                                                   46 / 58
4. Analogue Output (0~10V / Option)
This Option card converts weight value to Analog Voltage output(0~10V) and transfers to external
devices(Recorder, P.L.C), controlled by voltage output.
 4-1. Specification
   - Output Valtage : 0~10V DC output
   - Accuracy : More than 1/1,000
※As we convert Digital signal(1/30,000 accuracy) to Analogue, so the accuracy will be lower
than Digital signal
                                                                              1 : HI(+), 5 : (-)
 ※ This Voltage output is proportioned on weight calibration and outputs 0~10V.
  4-3. Adjustment
  This output is adjusted as when the weight is “Zero”, output is 0V and When the weight is “Full
capacity”, output is 10V.
 If you need additional adjustment, please adjust with “VR1(Zero)”, “VR2(Span) on the Analog Output
PCB.
※ Remark
This Analog option card converts Displayed weight value(Micro-process data) to analog value on D/A
Converter(Digital to Analog converter)
This D/A Converter has Max. 1/4,000 accuracy, so this output is not suitable for high accuracy
application, like more than 1/3,000.
For 0~5VDC or 1~5VDC analog output, please inform when you inquiry.
                                                                                                   47 / 58
5. Analogue Output (4~20mA / Option)
This Option card converts weight value to Analog Voltage output(4~20mA) and transfers to external
devices(Recorder, P.L.C), controlled by voltage output.
 5-1. Specification
   - Output Voltage : 4~20mA output (Max.2~22mA)
   - Accuracy : More than 1/1,000
   - Temperature Coefficient : 0.01%/℃
   - Max. Loading Impedance : Max. 500Ω
※As we convert Digital signal(1/30,000 accuracy) to Analogue, so the accuracy will be lower
than Digital signal
                                                                              1 : HI(+), 5 : (-)
※ “LO” terminal is not a “GND”, so this “LO” terminal do not be connected with other “GND” terminal
on other devices.
※ This output is proportioned on weight calibration and outputs 4~20mA.
※ Remark
This Analog option card converts Displayed weight value(Micro-process data) to analog value on D/A
Converter(Digital to Analog converter)
This D/A Converter has Max. 1/4,000 accuracy, so this output is not suitable for high accuracy
application, like more than 1/3,000.
                                                                                                   48 / 58
6. BCD Input (Option)
This “BCD interface” option card can be applied on PLC (Programmable Logic Controller), or Score Board
applications.
Each Input circuit is isolated with “Photo-Coupler”, from external devices electrically.
13 5 4 3 2 1
25 17 16 15 14
80 20 8 2
This Option card can be used for changing Part No. setting from external devices.
                                                                                                     49 / 58
 7. BCD Output (Option)
This “BCD interface” option card can be applied on PLC (Programmable Logic Controller), or Score
Board applications.
Each Input circuit is isolated with “Photo-Coupler”, from external devices electrically.
                                                                                        50 / 58
8. Serial Printer Interface (Standard).
This interface can be connected all kinds of serial interface installed printer devices.
But, programmed print format is specialized with our serial printer only.
So, if you use different model, the format can be changed or not printed.
                                                                                               51 / 58
9. Serial Print Format
Single                   Continuous
Print                    Print Format
Format
                         Sub-Total
                         Print Format
                         Grand Total
                         Print Format
                                        52 / 58
                Chapter 7. Error and Treatment
1. TEST Mode
※If you installed Analogue Option card, you can test Analogue output test with “TEST 2”
mode. (Please check detailed information)
                                                                                            53 / 58
Turn on the power with pressing             key.
                            TEST
Then, display will show                  , then press No. key and move to the certain TEST mode.
                            TEST
Then, display will show
                      SET-CAL                                                  TEST
※ Under TEST 3.                      mode, press              , key to exit                 .
                                                                                         54 / 58
2. Error and Treatment
   2-1. Load Cell Installation
           Error                   Cause                    Treatment                 Remark
                                                                                 1).Input Resistance
                            1). Load cell broken                                 of “EX+” and “EX-
                                                      1).            Measure
                            2).     Load      cell                               “      is     about
                                                      input/output resistance
                            isolation                                            350Ω~450Ω.
                                                      of Load cell.
                            resistance error
                            3). Weighing part                                    2).         Output
      Weight Value is                                 2). Measure Load cell
                            touches         other                                Resistance of “EX-
        unstable                                      isolation resistance
                            devices or some                                      “ and “EX+” is
                            weight is on the                                     about 350Ω.
                                                      3) Check attach point
                            weighing part
                                                      with other devices.
                            4). Summing Board                                    3).          Isolate
                            Error                                                Resistance is more
                                                                                 than 100Ω
    “OL” or “UL” display   1). Load cell broken       1). Load cell Check
                           or           Indicator     2). Load cell connection
                           connection Error           Check
                           2). Loading over than      3).    Remove       over
                           Max. Capacity              loaded weight
                                                                                                55 / 58
2-2. Calibration Process
   Error              Cause                                  Treatment
            When Max. capacity/digit       Re-input the Max. Capacity, less than 30,000
   Err 01
            value is over 30,000           (Max. Capacity / Digit)
            1. Amp. Gain is too big        Check standard weight’s weight with set value.
            2. Sig+ and Sig- wire          If there is difference between set value and real
   Err 06
               connection error            weight, please re-input the value
            3. Test weight is not loaded   (set value is too small)
            1. Amp. Gain is too small      Check standard weight’s weight with set value.
            2. Sig+ and Sig- wire          If there is difference between set value and real
   Err 07
               connection error            weight, please re-input the value
            3. Test weight is not loaded   (set value is too big)
                                                                                       56 / 58
                                    WARRANTEE CETIFICATION
WARRANTEE CLAUSE
1. The Warrantee period, we can guarantee, is one(1) year from your purchasing date
2. Warrantee Exception Clause
      - Warrantee period is expired.
      - Any kinds of Mal-function or defection caused by Modification or Repair without Migun ST’s
       permission.
      - Any kinds of Mal-function, Defection, or External damage, caused by operator
      - Any kinds of Mal-function, Defection, caused by using spare part from Non-Authorized
       Distributor or Agent.
      - Any kinds of Mal-function, Defection, caused by not following Warnings or Cautions
       mentioned on this manual.
      - Any kinds of Mal-function, Defection caused by “Force Majeur”, like Fire, Flood.
      - Without presentation of this “Warrantee Certification”.
3. Other
      - Any kinds of “Warrantee Certification” without authorized Stamp is out of validity
.
Overseas Office
MIGUN CORPORATION
1315, Nasung Plaza Building, Gasan Dong,
Geumcheon Gu, Seoul, KOREA.                           AUTHORIZED
Email : info@miguncorp.co.kr /                          STAMP
sales@miguncorp.co.kr
http://www.miguncorp.co.kr
57 / 58