BC Biomedical NIBP 1010

The document is a user manual for the NIBP-1000 Series non-invasive blood pressure simulators from BC Biomedical. It provides warnings and cautions about proper use of the simulators for testing purposes only and not for diagnostic or treatment use on patients. It also contains an overview of the keys, screens, setup instructions and theory of operations for running tests with the simulators.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
155 views48 pages

BC Biomedical NIBP 1010

The document is a user manual for the NIBP-1000 Series non-invasive blood pressure simulators from BC Biomedical. It provides warnings and cautions about proper use of the simulators for testing purposes only and not for diagnostic or treatment use on patients. It also contains an overview of the keys, screens, setup instructions and theory of operations for running tests with the simulators.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 48

NON-INVASIVE

BLOOD PRESSURE
SIMULATORS

NIBP-1030
NIBP-1020

NIBP-1010

USER MANUAL

Downloaded from www.Manualslib.com manuals search engine


Downloaded from www.Manualslib.com manuals search engine
BC BIOMEDICAL
NIBP-1000 SERIES
TABLE OF CONTENTS

WARNINGS, CAUTIONS, NOTICES ...............................................................................ii

DESCRIPTION............................................................................................................. 1

OVERVIEW.................................................................................................................. 4

KEYS............................................................................................................................ 7

SCREENS.................................................................................................................... 8

SETUP ......................................................................................................................... 13

OPERATIONS.............................................................................................................. 16

CONNECTING PRESSURE CUFFS............................................................... 16

CONNECTING PATIENT LEADS .................................................................... 17

IBP CONNECTOR (NIBP-1030 ONLY)............................................................ 17

TEMPERATURE CONNECTOR (NIBP-1030 ONLY)....................................... 18

AUX CONNECTOR.......................................................................................... 18

THEORY OF OPERATIONS........................................................................... 19

RUNNING A TEST .......................................................................................... 21

MANUAL REVISIONS.................................................................................................. 33

WARRANTY................................................................................................................. 33

SPECIFICATIONS ....................................................................................................... 34

NOTES......................................................................................................................... 40

Downloaded from www.Manualslib.com manuals search engine


WARNING - USERS
The NIBP-1000 Series Simulators are for use by
skilled technical personnel only.

WARNING - USE
The NIBP-1000 Series Simulators are intended
for testing only and they should never be used in
diagnostics, treatment or any other capacity
where they would come in contact with a patient.

WARNING - MODIFICATIONS
The NIBP-1000 Series Simulators are intended
for use within the published specifications. Any
application beyond these specifications or any
unauthorized user modifications may result in
hazards or improper operation.

WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulators. A serious hazard may occur
if the patient is connected when
testing with the Simulators.
Do not connect any leads from the patient
directly to the Simulators or DUT.

WARNING - ADAPTER
Turn Power Off and unplug any AC Power
Adapter before cleaning the surface
of the Simulators.

WARNING - LIQUIDS
Do not submerge or spill liquids on Simulators.
Do not operate the Simulators if internal
components may have been exposed to fluid,
as the internal leakage may have caused
corrosion and be a potential hazard.

ii

Downloaded from www.Manualslib.com manuals search engine


CAUTION - SERVICE
The NIBP-1000 Series Simulators are intended to
be serviced only by authorized service
personnel. Troubleshooting and service
procedures should only be performed by
qualified technical personnel.

CAUTION - ENVIRONMENT
The NIBP-1000 Series Simulators are intended to
function between 15 and 40 °C.
Exposure to temperatures outside this range can
adversely affect the performance
of the Simulators.

CAUTION - CLEANING
Do not immerse. The Simulators should be
cleaned by wiping gently with a damp,
lint-free cloth.

CAUTION - INSPECTION
The NIBP-1000 Series Simulators should be
inspected before each use for wear and
the Simulator should be serviced
if any parts are in question.

iii

Downloaded from www.Manualslib.com manuals search engine


NOTICE – CE

The NIBP-1000 Simulators bear the mark


Based on the following testing standards:

ELECTROMAGNETIC COMPATIBILITY DIRECTIVE


EMC – Directive 89/336/EEC and 2004/108/EC as amended by
92/31/EEC, 93/68/EEC and Directive 91/263/EEC [ TTE/SES ]

EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003


“Electrical equipment for measurement, control and
laboratory use – EMC requirements”

This equipment has been type tested by an independent, accredited testing laboratory
and compliance was demonstrated to the above standard to the extent applicable.

EMISSIONS
Radiated and Line Conducted Emissions

EN 61000-3-2 Harmonic Current Emissions


EN 61000-3-3 Voltage Fluctuation and Flicker

IMMUNITY– CLASS C

EN 61000-4-2 Electrostatic Discharge


EN 61000-4-3 Radiated Electric Field Immunity
EN 61000-4-4 Electrical Fast Transients / Bursts
EN 61000-4-5 Surge Voltage
EN 61000-4-6 Conducted Disturbance
EN 61000-4-11 Voltage Dips and Short Interrupts

LOW VOLTAGE DIRECTIVE


EC – Directive 73/23/EC

EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and
laboratory use – General requirements”

This equipment has been type tested and compliance was demonstrated
to the above standard to the extent applicable.

iv

Downloaded from www.Manualslib.com manuals search engine


NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)

Center Negative
Direct Current
Per European Council Directive
2002/95/EC, do not dispose of this
product as unsorted municipal waste.

NOTICE – ABBREVIATIONS
AC Alternating Current
BP Blood Pressure
Bpm Beats per minute
brpm breaths per minute
C Celsius
° degree
DUT Device Under Test
Dias Diastolic
ECG Electrocardiogram
Euro European
HR Heart Rate
Hz hertz
k kilo- (103)
Kg Kilogram
µ micro- (10-6)
µA microampere
m milli- (10-3)
mA milliampere
mm millimeter
mmHg millimeter mercury
mV millivolt
min minutes
Neo Neonatal
NSR Normal Sinus Rhythm
Ω ohm
lbs pounds
R Respiration
Sec seconds
Syst Systolic
US United States
VDC Direct Current Voltage

Downloaded from www.Manualslib.com manuals search engine


NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE
FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS.

NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY
IS ASSUMED FOR INACCURACIES.

NOTICE – CONTACT INFORMATION


BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA

1-800-242-8428
314-638-3800

www.bcgroupintl.com
sales@bcgroupintl.com

User Manual NIBP-1000 Series Copyright © 2008


www.bcgroupintl.com Made in the USA
1/08 Rev 10

vi

Downloaded from www.Manualslib.com manuals search engine


BC GROUP
NIBP-1000 SERIES
NON-INVASIVE BLOOD PRESSURE SIMULATORS

The Model NIBP-1000 Series is a family of Microprocessor based, High Precision Non-
Invasive Blood Pressure (NIBP) Simulators. The units are small, easy to use and have
multiple features to fit many different applications. The NIBP-1020, offers ECG waveforms
that are full QRS and respiration waveforms that look real. The NIBP-1030 offers Invasive
Blood Pressure, Temperature, Arrhythmias and a Leak Rate test mode.

The graphics display not only provides multiple screens that give the pressure digitally in
mmHg, but also offers views of the plot of the overall pressure or a close-up of the BP
waveform.

The following are highlights of some of the main features:

NIBP-1010 (BASIC FEATURES, NIBP ONLY):

• LARGE GRAPHICS DISPLAY WITH CURSOR SELECTION OF OPTIONS


AND SETUP OF PARAMETERS
• FULL RANGE MANOMETER
• ADULT, NEONATAL HYPERTENSIVE AND HYPOTENSIVE MODES
• 0-500 mmHg PRESSURE RANGE
• +/- 1% OF READING PRESSURE ACCURACY
• DIGITAL PRESSURE ENVELOPE OFFSET
• OPTIONAL PEAK PRESSURE DETECT WITH SIMPLE RESET
• SpO2 READY – COMPATIBLE WITH MSP-2100 MODULE
• DIGITAL CALIBRATION – NO POTS TO TURN
• SELECTABLE DISPLAY OPTIONS AND DIGIT SIZES
• SOFTWARE ADJUSTABLE CONTRAST
• FLASH PROGRAMMABLE
• BATTERY LIFE DISPLAY (0 TO 100%)
• OPTIONAL RECHARGEABLE NiMH BATTERIES
• BATTERY ELIMINATOR
• RS232 INTERFACE

NIBP-1020 (NIBP, ECG PACE & RESP)

HAS ALL THE BASIC MODEL FEATURES PLUS:

• ECG OUTPUT WITH FULL NSR WAVEFORM


• SINUSOIDAL RESPIRATION SIMULATION
• ECG TEST WAVEFORMS
• PACE WAVEFORM
• OPTIONAL PEAK PRESSURE DETECT WITH ECG ALARM TEST
• ECG SYNCRONIZED WITH BLOOD PRESSURE
1

Downloaded from www.Manualslib.com manuals search engine


NIBP-1030 (NIBP, IBP, TEMPERATURE, ECG PACE & RESP)

HAS ALL THE 1020 MODEL FEATURES PLUS:

• BACKLIT LCD DISPLAY


• SYNCHRONIZED INVASIVE BLOOD PRESSURE OUTPUT
• SELECTABLE IBP SENSITIVITY 5/40 µV/V/mmHg
• STATIC IBP SIMULATION –10mmHg TO 400 mmHg
• LEAK RATE TEST
• ECG ARRHYTHMIA WAVEFORMS
• ECG ARRHYTHMIA SEQUENCE
• YSI 400 AND 700 SERIES TEMPERATURE SIMULATION

AVAILABLE MODELS:

NIBP1010 BASIC UNIT


NIBP1010-P + PEAK
NIBP1010-BP + BATTERY & PEAK
NIBP1010-KIT NIBP1010-BP, 5 ADAPTERS & SOFT CARRYING CASE
NIBP1020 BASIC UNIT WITH ECG
NIBP1020-PA + PEAK & ALARM
NIBP1020-BPA + BATTERY, PEAK & ALARM
NIBP1020-KIT NIBP1020-BPA, 5 ADAPTERS & SOFT CARRYING CASE
NIBP1030 BASIC UNIT WITH ECG, IBP, TEMP, PEAK & ALARM + MANY MORE
NIBP1030-B + BATTERY
NIBP1030-KIT NIBP1030-B, 5 ADAPTERS & SOFT CARRYING CASE

ACCESSORIES:

• BC20-40614 AC POWER ADAPTER (EURO)


• BC20-40613 AC POWER ADAPTER (US)

OPTIONAL ACCESSORIES:

• BC20-30111 SOFT CARRYING CASE


• BC20-40714 DINAMAP/CRITIKON ADAPTER
• BC20-40715 QUICK DISCONNECT ADAPTER
• BC20-40716 LUER ADAPTER
• BC20-40717 MARQUETTE ADAPTER
• BC20-40608 BULB ADAPTER
• BC20-40602 NIBP FITTING KIT (HP CUFF ADAPTER,
INFLATION BULB, CPC CONNECTOR, QUICK
COUPLER WITH 1/8” MNPT AND SILICONE
TUBING AND TEE)
• BC20-40605 NIBP ADAPTER KIT (11 CONNECTORS)

Downloaded from www.Manualslib.com manuals search engine


INVASIVE BLOOD PRESSURE CABLES
FOR A COMPLETE LIST CONSULT THE BC GROUP WEBSITE AT
WWW.BCGROUPINTL.COM

TEMPERATURE CABLES
BC20 - 41333 UT-1 YSI-400
BC20 - 41334 UT-2 YSI-700

Downloaded from www.Manualslib.com manuals search engine


OVERVIEW

This section looks at the layout of the NIBP-1030 and gives descriptions of the elements
that are present.

10 Snap Patient
Lead Connectors:
RA R
LA L
RL N
LL F
Male Quick V1 C1
Connect for V2 C2
LCD Graphical Pressure V3 C3
Display: Input V4 C4
Shows Test Data and V5 C5
Configuration Data V6 C6

6 Light Touch Keys for


Selecting Parameters
and Settings:
POWER for Turning Unit
On and Off 6 pin Mini-
UP and DOWN Arrows Din
for Scrolling Through connector for
Selected Options. IBP
(When No Options are
Selected, for Changing
Display Screen.) 8-Pin Mini-DIN
SELECT for Choosing Plug Connector for
Setting or Parameter to Temperature
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu 7 pin Mini-
RESET for resetting Din
peak pressure reading. connector for
RS-232 and

Power
2.1 mm Jack

Downloaded from www.Manualslib.com manuals search engine


This section looks at the layout of the NIBP-1020 and gives descriptions of the elements
that are present.

10 Snap Patient
Lead Connectors:
RA R
LA L
RL N
LL F
Male Quick V1 C1
Connect for V2 C2
LCD Graphical Pressure V3 C3
Display: Input V4 C4
Shows Test Data and V5 C5
Configuration Data V6 C6

6 Light Touch Keys for


Selecting Parameters
and Settings:
POWER for Turning Unit
On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options.
(When No Options are
Selected, for Changing
Display Screen.)
SELECT for Choosing
Setting or Parameter to 7 pin Mini-
Change Din
SETUP/RETURN for connector for
Entering and Exiting RS-232 and
Setup Menu
RESET for resetting
peak pressure reading.
Power
2.1 mm Jack

Downloaded from www.Manualslib.com manuals search engine


This section looks at the layout of the NIBP-1010 and gives descriptions of the elements
that are present.

Male Quick
LCD Graphical Connect for
Display: Pressure
Shows Test Data and Input
Configuration Data

6 Light Touch Keys for


Selecting Parameters
and Settings:
POWER for Turning Unit
On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options.
7 pin Mini-
(When No Options are
Din
Selected, for Changing
connector for
Display Screen.)
RS-232 and
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting Power
Setup Menu 2.1 mm Jack
RESET for resetting
peak pressure reading.

Downloaded from www.Manualslib.com manuals search engine


KEYS

Six tactile-touch keys are provided for system operation:

– This key turns the unit off and on. The unit will return to the screen that was

active when it was turned off.

– In the DISPLAY MODE, these keys toggle the display through the available

main screens.

In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through

the available settings.

– On the Main screen, this key sequences through the available NIBP or ECG

simulations. On the Setup screen, there are a number of parameters that may be selected

and changed. This key sequences the cursor (Highlight) through those parameters.

– This key is used to RESET the peak pressure reading or to begin a Leak Test.

– This key toggles the unit into and out of the Setup Mode. Depressing this key

will enter the Setup screen where the configuration can be viewed and adjusted.

Depressing the key again will exit the Setup Mode and return to the previously viewed

main screen.

Downloaded from www.Manualslib.com manuals search engine


SCREENS

MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output

Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).

The available screens can be toggled using .

PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below.

Also displayed on this screen is the peak pressure and selected output waveform

The display will resemble the following:

PEAK PRESSURE
PRESSURE (Optional)

PRESSURE ENVELOPE
OFFSET
This line is only displayed
OUTPUT SELECTION if a non-zero value is
programmed into the
Respiration Systolic &/or Diastolic
Systolic/Diastolic (mean) Mean Heart Rate Shift. (See Setup for
Rate (Optional)
Adult 120/80 93 80 35 more information).
Adult 120/80 w/Pace 93 Pacemaker 70 35
Adult High 190/120 150 80 35
Adult Low 80/40 53 80 35
Neonatal 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 Only)
Leak Test (NIBP-1030 Only) - - -
Manometer - - -

Downloaded from www.Manualslib.com manuals search engine


OUTPUT WAVEFORM SCREEN – This screen shows the output waveform, as well as, the

pressure, peak pressure and output selection.

NOTE

This waveform is not intended to be


physiologically correct.

The display will resemble the following:

PRESSURE

PEAK PRESSURE
(Optional)

OUTPUT
WAVEFORM
(as generated by
the NIBP-1000
S i )
Dependant on
OUTPUT SELECTION pressure

Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 70 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 Only)
Leak Test (NIBP-1030 Only) - - -
Manometer - - -

Downloaded from www.Manualslib.com manuals search engine


PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as,

the pressure, peak pressure and output selection

PRESSURE

PEAK PRESSURE
(Optional)

PRESSURE
GRAPH
2mmHg per
pixel
OUTPUT SELECTION
Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 70 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 Only)
Leak Test (NIBP-1030 Only) - - -
Manometer - - -

10

Downloaded from www.Manualslib.com manuals search engine


ECG OUTPUT SCREEN – This screen shows the selected ECG output mode.

(NIBP-1020 & NIBP-1030 units only)

NOTE: While in this mode, the NIBP simulation does not run.

NOTE: NSR ECG output is active during NIBP testing at the rate stipulated for the selected

test.

The display will resemble the following:

LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 Only

11

Downloaded from www.Manualslib.com manuals search engine


BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage

level with Built-In Battery Option only.

NOTE: It is only an estimate of the battery life remaining.

When the level reaches 10%, the BP simulation mode will be disabled; however, the

Manometer and ECG (optional) will continue to function. Once the battery level reaches

0%, the unit will automatically turn itself off to avoid damaging the batteries.

The display will resemble the following:

Batteries charge from provided AC power adapter. The charge time is about 4 hours from

full discharge. While the batteries are charging, the display will flash “charging.”

CAUTION

USE ONLY THE AC POWER ADAPTER PROVIDED


WITH THE NIBP-1010, NIBP-1020 AND NIBP-1030.

12

Downloaded from www.Manualslib.com manuals search engine


SETUP

The Setup Mode allows the user to adjust the configuration of the simulator. The Setup

screen can be entered using the key. The parameters can be changed by using

key to highlight the line and to toggle the available options.

The Setup screen can be exited using the key.

NIBP 1010 and NIBP 1020

NIBP 1030

13

Downloaded from www.Manualslib.com manuals search engine


The following is a breakdown of the parameters available in the configuration of the unit

and their available options:

System Setup Configuration

Parameter Description Range


Adjusts the Systolic Output of the NIBP
Systolic Shift Simulation. This is not a direct mmHg adjustment. ±50.0
Adjusts the Diastolic Output of the NIBP
Diastolic Shift Simulation. This is not a direct mmHg adjustment.
±50.0

Static BP
Adjusts the Static Blood Pressure Output. -10 to 400 mmHg
(NIBP-1030 Only)

IBP Sens Selects the Invasive Blood Pressure Output 5 µV/V/mmHg


(NIBP-1030 Only) sensitivity. 40 µV/V/mmHg

0, 24, 30, 35, 37, 40,


Temp Selects the simulated Temperature output.
42 C
This parameter determines whether the SpO2
SpO2 Output output pulse is active. The output drives an MSP- Off/On
2100 FingerSim Module.
Determines the period of inactivity before the unit is
turned OFF. A timer is started when the unit is
turned ON and is reset each time a key is pressed.
When the timer reaches the value set in this
Auto Off Timer (Min) parameter, the power is automatically turned OFF. 0-30 Minutes
(NOTE: Setting this parameter to 0 disables the
Auto Off timer. When running from line power, the
unit does not automatically shut off. Auto Off timer
is inactive during a test.)
Contrast Adjust Sets the contrast of the display screen. 0-20
Off – Always off
1-30 sec – The elapsed time after which the backlight will
Backlight (Sec) automatically turn off. Off, 1-30 sec, ON
(NIBP-1030 Only) ON – Always ON.
The Default setting is 30 seconds.
Available only with Battery Option installed.
Displays current life of the battery. 0-100%
Battery Life At 10%, a warning screen will appear. (Read Only)
At 0%, the unit will power down automatically.
Software Displays current software program. (Read Only)

14

Downloaded from www.Manualslib.com manuals search engine


SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option

to shift test results to compensate for different methods of measuring Oscillometric NIBP by

various manufacturers and models of devices under test.

CAUTION

These adjustments must be used with caution as they will


allow the user to adjust the output results to invalid values.

These adjustments should only be used to aid in the


simplification of testing and with documented controls.

There are no absolute standards for Oscillometric NIBP readings; therefore, for a number

of reasons (including patents, technology, etc.), each manufacturer has established a

different method for evaluating the oscillometric pulses. Due to these varying methods,

precisely the same waveforms will give different results on different manufacturer’s units.

The normal technique used is to run the monitor against a fixed source like the NIBP-1000,

with the understanding that each manufacturer has a predictable error from this norm.

While this is generally the most direct method, users have asked for a method to correct

for this difference, making the monitors read the same as the test unit. The Systolic and

Diastolic Shift settings allow for just such correction.

These adjustments are indicated in a line added to the main display to inform the user of

any shift that has been programmed into the system. This is done so there is no

misunderstanding of the meaning of the results.

15

Downloaded from www.Manualslib.com manuals search engine


OPERATIONS

CONNECTING PRESSURE

The NIBP-1010, NIBP-1020 and NIBP-1030 Blood Pressure Simulators are connected

between the blood pressure cuff and the monitor (Device Under Test, DUT).

The blood pressure cuff should be disconnected from the DUT and a ‘T’ adapter inserted

between the cuff and the DUT. The NIBP simulator is then connected to the open side.

Blood Monitor
Pressure (Device Under Test)
Cuff (DUT)
NIBP-
10xx

Roll the cuff tightly on itself or around a mandrel.

16

Downloaded from www.Manualslib.com manuals search engine


CONNECTING PATIENT LEADS

For the models with ECG output, test snaps are provided along the sides and are identified

by the markings on the overlay.

IBP CONNECTOR
(NIBP-1030 ONLY)

For the models with IBP output, connect a standard BP test cable (mini-din to monitor

connector) between the NIBP-1030 and the DUT (Device Under Test). The pin-out shown

below for the mini-din is the industry standard configuration that is used on most modern

simulators. As such, there are many existing cables that may be used or you may obtain

the proper cable from BC Group.

17

Downloaded from www.Manualslib.com manuals search engine


TEMPERATURE CONNECTOR
(NIBP-1030 ONLY)

For the models with Temperature output, connect a standard Temperature test cable (mini-
din to monitor connector) between the NIBP-1030 and the DUT (Device Under Test). The
pin-out is shown below.

AUX CONNECTOR

The Aux connector is used for both serial communications and SpO2. The serial

connection is made using the BC20-41337 cable. The MSP-2100 module plugs directly

into the Aux port and provides the SpO2 function.

18

Downloaded from www.Manualslib.com manuals search engine


THEORY OF OPERATIONS

NIBP (Non-Invasive Blood Pressure) Function

A digital stepping motor and piston assembly is used to generate the output waveforms. A

differential pressure sensor is used to measure the pressure of the cuff. The pressure

sensor is read by a 16-bit differential Analog to Digital converter.

The cuff pressure is monitored and the motor/piston assembly reacts to provide the correct

response based on the selected mode.

IBP (Invasive Blood Pressure) Function

A 12-bit Digital to Analog converter is used to generate the IBP waveforms. The output

circuit is fully isolated and capable of switching between the two standard sensitivities (5

µV/V/mmHg and 40 µV/V/mmHg).

Temperature

There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of

a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe

types. There are seven different temperatures selectable for each.

19

Downloaded from www.Manualslib.com manuals search engine


ECG Function

The ECG output is performed by a 12-bit Digital to Analog converter. All waveforms for the

standard ECG and Arrhythmias are generated via data stored in the program.

Respiration Function

A 12-bit Digital to Analog converter is used to generate the respiration waveform. The

respiration output is present on the LA ECG lead only. Refer to Monitor manual to

determine which lead the Monitor uses for respiration. It may be necessary to reverse the

LA and LL leads for respiration to be detected.

20

Downloaded from www.Manualslib.com manuals search engine


RUNNING A TEST

Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This

section will walk through each of the tests and their basic operation.

The Main Tests are accessible with a single key. The key will scroll through the

following tests in a continuous loop:

Adult 120/80
Adult 120/80 w/Pace
Adult High 190/120
Adult Low 80/40
Neonatal 70/40
Alarm Test
Arrhythmia Seq
Leak Test
Manometer

BASIC TEST MODES

NIBP:

The first 5 test modes deal with various NIBP setups. To run an NIBP simulation, the cuff

and monitor are connected to the pressure input. Then the measurement is initiated by the

monitor and the NIBP-1000 unit will output the proper waveform based on the cuff pressure

provided by the monitor and the selected simulation.

The NIBP output mode can be changed by pressing the key. Once the desired

operating mode is selected, the output will automatically begin when the correct pressure is

detected.

21

Downloaded from www.Manualslib.com manuals search engine


There are 5 selectable NIBP Basic Test Modes; Adult, Adult w/pace, Adult High, Adult Low

and Neonatal. The displays will resemble the following examples:

NOTE: The screens for the first test will have the individual components labeled. The

component labels for subsequent tests are the same.

Adult 120/80
(Typical Screens)
PEAK PRESSURE
PRESSURE
(Optional)

PRESSURE ENVELOPE
OFFSET
NIBP TEST
This line is only displayed
Type Adult 120/80 if a non-zero value is
Heart Rate 80 programmed into the
Respiration Rate 35 Systolic &/or Diastolic
(Optional) Shift. (See Setup for
more information).
(Pressure Only Screen)

PRESSURE PEAK PRESSURE


(Optional)

NIBP TEST OUTPUT


WAVEFORM
Type Adult 120/80 (as generated by
Heart Rate 80 the NIBP-1000
Respiration Rate 35 S i )
(Optional) Dependant on
pressure
(Output Waveform Screen)

PRESSURE PEAK PRESSURE


(Optional)

NIBP TEST PRESSURE


GRAPH
Type Adult 120/80
Heart Rate 80 2mmHg per
Respiration Rate 35 pixel
(Optional)
(Pressure Graph Screen)

22

Downloaded from www.Manualslib.com manuals search engine


Adult 120/80 with Pace
(Typical Screens)

Adult High 120/80


(Typical Screens)

23

Downloaded from www.Manualslib.com manuals search engine


Adult Low 80/40
(Typical Screens)

Neonatal 70/40
(Typical Screens)

24

Downloaded from www.Manualslib.com manuals search engine


ALARM TEST:

Use the key to cycle through the Test Modes until Alarm Test appears. The

purpose of this function is to help test the alarms of the monitor under test. The ECG

output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR

with 80 brpm respiration. The time interval between alternations is 30 seconds. A count

down timer is displayed to indicate the time still remaining for the current output.

NOTE: While in this mode, the NIBP simulation output is disabled.

The displays will resemble the following examples: (Typical Screens)

PRESSURE PEAK PRESSURE


TIMER
(Information not relevant to this (Information not relevant to
test, NIBP is disabled.) this test, NIBP is disabled.)

PRESSURE ENVELOPE
OFFSET
This line is only displayed
ALARM TEST if a non-zero value is
programmed into the
Heart Rate 47/160 Systolic &/or Diastolic
Respiration Rate 0/80 Shift. (See Setup for
more information).
(Pressure Only Screen)

(Output Waveform Screen) (Pressure Graph Screen)

25

Downloaded from www.Manualslib.com manuals search engine


ARRHYTHMIA SEQUENCE: (NIBP-1030 Only)

Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.

The purpose of this function is to provide a timed sequence of some of the more common

Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed

below. A count down timer is displayed to indicate the time still remaining in each step.

NOTE: While in this mode, the NIBP simulation output is disabled.

Step Display Total Time ECG Output


1 Single PVCs 18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC
2 Pair of PVCs 9 sec 10 NSR, Pair of PVCs
3 Single PVCs 18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC
4 Run of 5 PVCs 12 sec 11 NSR, Run of 5 PVCs
5 Vtach 30 sec 15 NSR, VTach
6 Vfib 30 sec 15 NSR, VFib
7 Asystole 30 sec 15 NSR, Asystole

The displays will resemble the following examples: (Typical Screens)

PRESSURE PEAK PRESSURE


TIMER
(Information not relevant to this (Information not relevant to
test, NIBP is disabled.) this test, NIBP is disabled.)

ARRHYTHMIA PRESSURE ENVELOPE


OFFSET
Single PVCs
Pair of PVCs This line is only displayed
Single PVCs if a non-zero value is
Run of 5 PVCs programmed into the
Vtach Systolic &/or Diastolic
Vfib Shift. (See Setup for
Asystole more information).
(Pressure Only Screen)

(Output Waveform Screen) (Pressure Graph Screen)

26

Downloaded from www.Manualslib.com manuals search engine


LEAK TEST: (NIBP-1030 Only)

Use the key to cycle through the Test Modes until Leak Test appears. The

purpose of this function is to provide a standard pressure leak test. The pressure input is

connected to the system that is to be monitored. Pressure is then applied to the system.

The Leak Test is initiated by pressing the key. The unit traces and displays the

pressure drop and the time since the test was initiated. It also, calculates and displays the

leak rate in mmHg per minute.

NOTE: While in this mode, the BP simulation outputs are disabled.

The displays will resemble the following examples: (Typical Screens)

PEAK PRESSURE
PRESSURE
(Optional)

LEAK TEST
Drop in mmHg
Time in minutes:seconds
Leak Rate in mmHg/minute

(Pressure Only Screen)

(Output Waveform Screen) (Pressure Graph Screen)

27

Downloaded from www.Manualslib.com manuals search engine


MANOMETER:

Use the key to cycle through the Test Modes until Manometer appears. The

purpose of this function is to allow the unit to be used strictly as a Manometer with a full

scale range of +/- 500 mmHg.

NOTE: The Graphic Screens are also available in this mode for maximum flexibility

The displays will resemble the following examples: (Typical Screens)

PEAK PRESSURE
PRESSURE
(Optional)

PRESSURE ENVELOPE
OFFSET
This line is only displayed
if a non-zero value is
programmed into the
Systolic &/or Diastolic
Shift. (See Setup for
more information).
(Pressure Only Screen)

(Output Waveform Screen) (Pressure Graph Screen)

28

Downloaded from www.Manualslib.com manuals search engine


SUB TEST MODES

PEAK: (Optional)

The optional Peak test feature is available whenever the display is in a pressure mode. It

will continuously monitor the pressure input and display and hold the maximum value. This

value may be reset at any time by pressing the key.

The displays will resemble the following examples: (Typical Screens)

29

Downloaded from www.Manualslib.com manuals search engine


ECG PERFORMANCE TEST: (NIBP-1020 & NIBP-1030)

The ECG output is selected by pressing until the ECG Waveforms menu is

displayed. The key is used to choose the desired ECG output. There are 16 fixed

waveforms available.

NOTE: While in this mode, the NIBP simulation is disabled, but the IBP simulation is active

(NIBP-1030 only).

The display will resemble the following example:

LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 Only

30

Downloaded from www.Manualslib.com manuals search engine


DYNAMIC IBP: (NIBP-1030 Only)

The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG

Waveform mode.

NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in

the SETUP Mode before using IBP. (See Setup section for more information.)

STATIC IBP: (NIBP-1030 Only)

A static Blood Pressure output may be selected in the SETUP mode. This fixed value will

remain on the IBP output until changed.

NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in

the SETUP Mode before using IBP. (See Setup section for more information.)

31

Downloaded from www.Manualslib.com manuals search engine


TEMPERATURE SIMULATION: (NIBP-1030 Only)

The temperature setting can be selected in the SETUP mode. This fixed value will remain
on the Temperature output until changed. The output will simulate both YSI 400 and YSI
700 Temperature probes.

NOTE: Both outputs are available at the output connector simultaneously.

32

Downloaded from www.Manualslib.com manuals search engine


MANUAL REVISIONS

Revision # Program # Revisions Made

Rev 01 DT7355CA Origination


Rev 02 DT7355CA Specification Information Updated
Rev 03 DT7355CA Miscellaneous Edits
Rev 04 DT7355CF Add Systolic/Diastolic adjustments
Rev 05 DT7355CG Output Selection edited
Rev 06 DT7355CI Miscellaneous Edits
Rev 07 DT7355CI Warnings, Cautions, Notices Updated
Rev 08 DT7359CA NIBP-1030, IBP, ECG Arrhythmia Waveforms Added
Rev 09 DT7359CA Temperature Added
Rev 10 DT7359CA Temp Connector Updated, Static 120 Added

LIMITED WARRANTY

WARRANTY: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE


FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.

EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR


IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE.

BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL


DAMAGES.

NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR


ASSUME ANY LIABILITY.

REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.

33

Downloaded from www.Manualslib.com manuals search engine


SPECIFICATIONS

NIBP-1010

Blood Pressure
RANGE 0-500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm
ACCURACY +/- 1%

Physical
DISPLAYS LCD Graphical 128 X 64 Pixels
7 x 5 x 4 Inches
ENCLOSURE (177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
WEIGHT
(< 1.36 Kg)
FACE PLATE Lexan, Back printed
OPERATING RANGE 15 to 40 C
STORAGE RANGE -20 to 65 C

Electrical
12VDC, 500mA
BATTERY ELIMINATOR

BATTERY 6 AA NiMH Rechargeable


(OPTIONAL) (Not user serviceable)
BATTERY
500 Test Cycles between charges
RUNTIME
BATTERY
1 Year from full charge
STORAGE LIFE

34

Downloaded from www.Manualslib.com manuals search engine


NIBP-1020

Blood Pressure
RANGE 0-500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm (synchronized to ECG)
ACCURACY +/- 1%

ECG NSR
RATE 30,60,120,240 BPM
ACCURACY +/- 1%
AMPLITUDE 2.75 mV
ACCURACY +/- 2% @ Lead II

ECG Performance
SINE WAVE 10,60,100 Hz
SQUARE WAVE 0.125, 2.000 Hz
TRIANGLE WAVE 2.000 Hz
RATE ACCURACY +/- 1%
AMPLITUDE 2.75 mV
AMPLITUDE
+/- 2% @ Lead II
ACCURACY

Pacemaker Waveforms
AMPLITUDE 3 mV
ACCURACY +/- 10%
WIDTH 6 ms
ACCURACY +/- 5%

35

Downloaded from www.Manualslib.com manuals search engine


Respiration
RATE ACCURACY +/- 1%
IMPEDANCE DELTA 3.0 ohms
ACCURACY +/- 10%
BASELINE 1000 ohms
ACCURACY +/- 5%

Physical
LCD Graphical 128 X 64 Pixels
DISPLAYS
White LED Backlight
7 x 5 x 4 Inches
ENCLOSURE (177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
WEIGHT
(< 1.36 Kg)
FACE PLATE Lexan, Back printed
OPERATING RANGE 15 to 40 C
STORAGE RANGE -20 to 65 C

Electrical
12VDC, 500mA
AC POWER ADAPTER

BATTERY 6 AA NiMH Rechargeable


(OPTIONAL) (Not user serviceable)
BATTERY
500 Test Cycles between charges
RUNTIME
BATTERY
1 Year from full charge
STORAGE LIFE

36

Downloaded from www.Manualslib.com manuals search engine


NIBP-1030

Blood Pressure
RANGE +/- 500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm (synched to ECG)
ACCURACY +/- 1%

Invasive Blood Pressure


-10,-5,0,20,40,50,60,80,100,120,150,
STATIC PRESSURE
160,200,240,250,300,320,400 mmHg
+/- (1% Full Range + 1 mmHg) or
ACCURACY +/- (2% Setting + 2 mmHg)
whichever is better
IMPEDANCE 300 ohms
ACCURACY +/- 10%
EXCITATION RANGE 2 to 16 V RMS
EXCITATION
DC to 5 KHz
FREQUENCY
SENSITIVITY 5 or 40 µV/V/mmHg

ECG NSR
RATE 30,60,120,240 BPM
ACCURACY +/- 1%
AMPLITUDE 2.75 mV
ACCURACY +/- 2% @ Lead II

37

Downloaded from www.Manualslib.com manuals search engine


ECG Performance
SINE WAVE 10,60,100 Hz
SQUARE WAVE 0.125, 2.000 Hz
TRIANGLE WAVE 2.000 Hz
RATE ACCURACY +/- 1%
AMPLITUDE 2.75 mV
AMPLITUDE
+/- 2% @ Lead II
ACCURACY

Pacemaker Waveforms
AMPLITUDE 3 mV
ACCURACY +/- 10%
WIDTH 6 ms
ACCURACY +/- 5%

Respiration
RATE ACCURACY +/- 1%
IMPEDANCE DELTA 3.0 ohms
ACCURACY +/- 10%
BASELINE 1000 ohms
ACCURACY +/- 5%

Temperature
0,24,30,37,40 C (32.0,
Temperature
75.2,86.0,98.6,104.0 F)
Accuracy +/- 0.1 C
Type YSI Series 400 and 700

38

Downloaded from www.Manualslib.com manuals search engine


Physical
DISPLAYS LCD Graphical 128 X 64 Pixels
7 x 5 x 4 Inches
ENCLOSURE (177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
WEIGHT
(< 1.36 Kg)
FACE PLATE Lexan, Back printed
OPERATING RANGE 15 to 40 C
STORAGE RANGE -20 to 65 C

Electrical
12VDC, 500mA
AC POWER ADAPTER

BATTERY 6 AA NiMH Rechargeable


(OPTIONAL) (Not user serviceable)
BATTERY
500 Test Cycles between charges
RUNTIME
BATTERY
1 Year from full charge
STORAGE LIFE

P:\Manuals\BCGroup\…\NIBP1000_UM_Rev10.doc

39