BC Biomedical NIBP 1010
BC Biomedical NIBP 1010
BLOOD PRESSURE
SIMULATORS
NIBP-1030
NIBP-1020
NIBP-1010
USER MANUAL
DESCRIPTION............................................................................................................. 1
OVERVIEW.................................................................................................................. 4
KEYS............................................................................................................................ 7
SCREENS.................................................................................................................... 8
SETUP ......................................................................................................................... 13
OPERATIONS.............................................................................................................. 16
AUX CONNECTOR.......................................................................................... 18
THEORY OF OPERATIONS........................................................................... 19
MANUAL REVISIONS.................................................................................................. 33
WARRANTY................................................................................................................. 33
SPECIFICATIONS ....................................................................................................... 34
NOTES......................................................................................................................... 40
WARNING - USE
The NIBP-1000 Series Simulators are intended
for testing only and they should never be used in
diagnostics, treatment or any other capacity
where they would come in contact with a patient.
WARNING - MODIFICATIONS
The NIBP-1000 Series Simulators are intended
for use within the published specifications. Any
application beyond these specifications or any
unauthorized user modifications may result in
hazards or improper operation.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulators. A serious hazard may occur
if the patient is connected when
testing with the Simulators.
Do not connect any leads from the patient
directly to the Simulators or DUT.
WARNING - ADAPTER
Turn Power Off and unplug any AC Power
Adapter before cleaning the surface
of the Simulators.
WARNING - LIQUIDS
Do not submerge or spill liquids on Simulators.
Do not operate the Simulators if internal
components may have been exposed to fluid,
as the internal leakage may have caused
corrosion and be a potential hazard.
ii
CAUTION - ENVIRONMENT
The NIBP-1000 Series Simulators are intended to
function between 15 and 40 °C.
Exposure to temperatures outside this range can
adversely affect the performance
of the Simulators.
CAUTION - CLEANING
Do not immerse. The Simulators should be
cleaned by wiping gently with a damp,
lint-free cloth.
CAUTION - INSPECTION
The NIBP-1000 Series Simulators should be
inspected before each use for wear and
the Simulator should be serviced
if any parts are in question.
iii
This equipment has been type tested by an independent, accredited testing laboratory
and compliance was demonstrated to the above standard to the extent applicable.
EMISSIONS
Radiated and Line Conducted Emissions
IMMUNITY– CLASS C
EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and
laboratory use – General requirements”
This equipment has been type tested and compliance was demonstrated
to the above standard to the extent applicable.
iv
Center Negative
Direct Current
Per European Council Directive
2002/95/EC, do not dispose of this
product as unsorted municipal waste.
NOTICE – ABBREVIATIONS
AC Alternating Current
BP Blood Pressure
Bpm Beats per minute
brpm breaths per minute
C Celsius
° degree
DUT Device Under Test
Dias Diastolic
ECG Electrocardiogram
Euro European
HR Heart Rate
Hz hertz
k kilo- (103)
Kg Kilogram
µ micro- (10-6)
µA microampere
m milli- (10-3)
mA milliampere
mm millimeter
mmHg millimeter mercury
mV millivolt
min minutes
Neo Neonatal
NSR Normal Sinus Rhythm
Ω ohm
lbs pounds
R Respiration
Sec seconds
Syst Systolic
US United States
VDC Direct Current Voltage
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY
IS ASSUMED FOR INACCURACIES.
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
vi
The Model NIBP-1000 Series is a family of Microprocessor based, High Precision Non-
Invasive Blood Pressure (NIBP) Simulators. The units are small, easy to use and have
multiple features to fit many different applications. The NIBP-1020, offers ECG waveforms
that are full QRS and respiration waveforms that look real. The NIBP-1030 offers Invasive
Blood Pressure, Temperature, Arrhythmias and a Leak Rate test mode.
The graphics display not only provides multiple screens that give the pressure digitally in
mmHg, but also offers views of the plot of the overall pressure or a close-up of the BP
waveform.
AVAILABLE MODELS:
ACCESSORIES:
OPTIONAL ACCESSORIES:
TEMPERATURE CABLES
BC20 - 41333 UT-1 YSI-400
BC20 - 41334 UT-2 YSI-700
This section looks at the layout of the NIBP-1030 and gives descriptions of the elements
that are present.
10 Snap Patient
Lead Connectors:
RA R
LA L
RL N
LL F
Male Quick V1 C1
Connect for V2 C2
LCD Graphical Pressure V3 C3
Display: Input V4 C4
Shows Test Data and V5 C5
Configuration Data V6 C6
Power
2.1 mm Jack
10 Snap Patient
Lead Connectors:
RA R
LA L
RL N
LL F
Male Quick V1 C1
Connect for V2 C2
LCD Graphical Pressure V3 C3
Display: Input V4 C4
Shows Test Data and V5 C5
Configuration Data V6 C6
Male Quick
LCD Graphical Connect for
Display: Pressure
Shows Test Data and Input
Configuration Data
– This key turns the unit off and on. The unit will return to the screen that was
– In the DISPLAY MODE, these keys toggle the display through the available
main screens.
In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through
– On the Main screen, this key sequences through the available NIBP or ECG
simulations. On the Setup screen, there are a number of parameters that may be selected
and changed. This key sequences the cursor (Highlight) through those parameters.
– This key is used to RESET the peak pressure reading or to begin a Leak Test.
– This key toggles the unit into and out of the Setup Mode. Depressing this key
will enter the Setup screen where the configuration can be viewed and adjusted.
Depressing the key again will exit the Setup Mode and return to the previously viewed
main screen.
MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output
Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).
PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below.
Also displayed on this screen is the peak pressure and selected output waveform
PEAK PRESSURE
PRESSURE (Optional)
PRESSURE ENVELOPE
OFFSET
This line is only displayed
OUTPUT SELECTION if a non-zero value is
programmed into the
Respiration Systolic &/or Diastolic
Systolic/Diastolic (mean) Mean Heart Rate Shift. (See Setup for
Rate (Optional)
Adult 120/80 93 80 35 more information).
Adult 120/80 w/Pace 93 Pacemaker 70 35
Adult High 190/120 150 80 35
Adult Low 80/40 53 80 35
Neonatal 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 Only)
Leak Test (NIBP-1030 Only) - - -
Manometer - - -
NOTE
PRESSURE
PEAK PRESSURE
(Optional)
OUTPUT
WAVEFORM
(as generated by
the NIBP-1000
S i )
Dependant on
OUTPUT SELECTION pressure
Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 70 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 Only)
Leak Test (NIBP-1030 Only) - - -
Manometer - - -
PRESSURE
PEAK PRESSURE
(Optional)
PRESSURE
GRAPH
2mmHg per
pixel
OUTPUT SELECTION
Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 70 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 Only)
Leak Test (NIBP-1030 Only) - - -
Manometer - - -
10
NOTE: While in this mode, the NIBP simulation does not run.
NOTE: NSR ECG output is active during NIBP testing at the rate stipulated for the selected
test.
LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 Only
11
When the level reaches 10%, the BP simulation mode will be disabled; however, the
Manometer and ECG (optional) will continue to function. Once the battery level reaches
0%, the unit will automatically turn itself off to avoid damaging the batteries.
Batteries charge from provided AC power adapter. The charge time is about 4 hours from
full discharge. While the batteries are charging, the display will flash “charging.”
CAUTION
12
The Setup Mode allows the user to adjust the configuration of the simulator. The Setup
screen can be entered using the key. The parameters can be changed by using
NIBP 1030
13
Static BP
Adjusts the Static Blood Pressure Output. -10 to 400 mmHg
(NIBP-1030 Only)
14
to shift test results to compensate for different methods of measuring Oscillometric NIBP by
CAUTION
There are no absolute standards for Oscillometric NIBP readings; therefore, for a number
different method for evaluating the oscillometric pulses. Due to these varying methods,
precisely the same waveforms will give different results on different manufacturer’s units.
The normal technique used is to run the monitor against a fixed source like the NIBP-1000,
with the understanding that each manufacturer has a predictable error from this norm.
While this is generally the most direct method, users have asked for a method to correct
for this difference, making the monitors read the same as the test unit. The Systolic and
These adjustments are indicated in a line added to the main display to inform the user of
any shift that has been programmed into the system. This is done so there is no
15
CONNECTING PRESSURE
The NIBP-1010, NIBP-1020 and NIBP-1030 Blood Pressure Simulators are connected
between the blood pressure cuff and the monitor (Device Under Test, DUT).
The blood pressure cuff should be disconnected from the DUT and a ‘T’ adapter inserted
between the cuff and the DUT. The NIBP simulator is then connected to the open side.
Blood Monitor
Pressure (Device Under Test)
Cuff (DUT)
NIBP-
10xx
16
For the models with ECG output, test snaps are provided along the sides and are identified
IBP CONNECTOR
(NIBP-1030 ONLY)
For the models with IBP output, connect a standard BP test cable (mini-din to monitor
connector) between the NIBP-1030 and the DUT (Device Under Test). The pin-out shown
below for the mini-din is the industry standard configuration that is used on most modern
simulators. As such, there are many existing cables that may be used or you may obtain
17
For the models with Temperature output, connect a standard Temperature test cable (mini-
din to monitor connector) between the NIBP-1030 and the DUT (Device Under Test). The
pin-out is shown below.
AUX CONNECTOR
The Aux connector is used for both serial communications and SpO2. The serial
connection is made using the BC20-41337 cable. The MSP-2100 module plugs directly
18
A digital stepping motor and piston assembly is used to generate the output waveforms. A
differential pressure sensor is used to measure the pressure of the cuff. The pressure
The cuff pressure is monitored and the motor/piston assembly reacts to provide the correct
A 12-bit Digital to Analog converter is used to generate the IBP waveforms. The output
circuit is fully isolated and capable of switching between the two standard sensitivities (5
Temperature
There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of
a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe
19
The ECG output is performed by a 12-bit Digital to Analog converter. All waveforms for the
standard ECG and Arrhythmias are generated via data stored in the program.
Respiration Function
A 12-bit Digital to Analog converter is used to generate the respiration waveform. The
respiration output is present on the LA ECG lead only. Refer to Monitor manual to
determine which lead the Monitor uses for respiration. It may be necessary to reverse the
20
Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This
section will walk through each of the tests and their basic operation.
The Main Tests are accessible with a single key. The key will scroll through the
Adult 120/80
Adult 120/80 w/Pace
Adult High 190/120
Adult Low 80/40
Neonatal 70/40
Alarm Test
Arrhythmia Seq
Leak Test
Manometer
NIBP:
The first 5 test modes deal with various NIBP setups. To run an NIBP simulation, the cuff
and monitor are connected to the pressure input. Then the measurement is initiated by the
monitor and the NIBP-1000 unit will output the proper waveform based on the cuff pressure
The NIBP output mode can be changed by pressing the key. Once the desired
operating mode is selected, the output will automatically begin when the correct pressure is
detected.
21
NOTE: The screens for the first test will have the individual components labeled. The
Adult 120/80
(Typical Screens)
PEAK PRESSURE
PRESSURE
(Optional)
PRESSURE ENVELOPE
OFFSET
NIBP TEST
This line is only displayed
Type Adult 120/80 if a non-zero value is
Heart Rate 80 programmed into the
Respiration Rate 35 Systolic &/or Diastolic
(Optional) Shift. (See Setup for
more information).
(Pressure Only Screen)
22
23
Neonatal 70/40
(Typical Screens)
24
Use the key to cycle through the Test Modes until Alarm Test appears. The
purpose of this function is to help test the alarms of the monitor under test. The ECG
output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR
with 80 brpm respiration. The time interval between alternations is 30 seconds. A count
down timer is displayed to indicate the time still remaining for the current output.
PRESSURE ENVELOPE
OFFSET
This line is only displayed
ALARM TEST if a non-zero value is
programmed into the
Heart Rate 47/160 Systolic &/or Diastolic
Respiration Rate 0/80 Shift. (See Setup for
more information).
(Pressure Only Screen)
25
Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.
The purpose of this function is to provide a timed sequence of some of the more common
Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed
below. A count down timer is displayed to indicate the time still remaining in each step.
26
Use the key to cycle through the Test Modes until Leak Test appears. The
purpose of this function is to provide a standard pressure leak test. The pressure input is
connected to the system that is to be monitored. Pressure is then applied to the system.
The Leak Test is initiated by pressing the key. The unit traces and displays the
pressure drop and the time since the test was initiated. It also, calculates and displays the
PEAK PRESSURE
PRESSURE
(Optional)
LEAK TEST
Drop in mmHg
Time in minutes:seconds
Leak Rate in mmHg/minute
27
Use the key to cycle through the Test Modes until Manometer appears. The
purpose of this function is to allow the unit to be used strictly as a Manometer with a full
NOTE: The Graphic Screens are also available in this mode for maximum flexibility
PEAK PRESSURE
PRESSURE
(Optional)
PRESSURE ENVELOPE
OFFSET
This line is only displayed
if a non-zero value is
programmed into the
Systolic &/or Diastolic
Shift. (See Setup for
more information).
(Pressure Only Screen)
28
PEAK: (Optional)
The optional Peak test feature is available whenever the display is in a pressure mode. It
will continuously monitor the pressure input and display and hold the maximum value. This
29
The ECG output is selected by pressing until the ECG Waveforms menu is
displayed. The key is used to choose the desired ECG output. There are 16 fixed
waveforms available.
NOTE: While in this mode, the NIBP simulation is disabled, but the IBP simulation is active
(NIBP-1030 only).
LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 Only
30
The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG
Waveform mode.
the SETUP Mode before using IBP. (See Setup section for more information.)
A static Blood Pressure output may be selected in the SETUP mode. This fixed value will
the SETUP Mode before using IBP. (See Setup section for more information.)
31
The temperature setting can be selected in the SETUP mode. This fixed value will remain
on the Temperature output until changed. The output will simulate both YSI 400 and YSI
700 Temperature probes.
32
LIMITED WARRANTY
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
33
NIBP-1010
Blood Pressure
RANGE 0-500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm
ACCURACY +/- 1%
Physical
DISPLAYS LCD Graphical 128 X 64 Pixels
7 x 5 x 4 Inches
ENCLOSURE (177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
WEIGHT
(< 1.36 Kg)
FACE PLATE Lexan, Back printed
OPERATING RANGE 15 to 40 C
STORAGE RANGE -20 to 65 C
Electrical
12VDC, 500mA
BATTERY ELIMINATOR
34
Blood Pressure
RANGE 0-500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm (synchronized to ECG)
ACCURACY +/- 1%
ECG NSR
RATE 30,60,120,240 BPM
ACCURACY +/- 1%
AMPLITUDE 2.75 mV
ACCURACY +/- 2% @ Lead II
ECG Performance
SINE WAVE 10,60,100 Hz
SQUARE WAVE 0.125, 2.000 Hz
TRIANGLE WAVE 2.000 Hz
RATE ACCURACY +/- 1%
AMPLITUDE 2.75 mV
AMPLITUDE
+/- 2% @ Lead II
ACCURACY
Pacemaker Waveforms
AMPLITUDE 3 mV
ACCURACY +/- 10%
WIDTH 6 ms
ACCURACY +/- 5%
35
Physical
LCD Graphical 128 X 64 Pixels
DISPLAYS
White LED Backlight
7 x 5 x 4 Inches
ENCLOSURE (177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
WEIGHT
(< 1.36 Kg)
FACE PLATE Lexan, Back printed
OPERATING RANGE 15 to 40 C
STORAGE RANGE -20 to 65 C
Electrical
12VDC, 500mA
AC POWER ADAPTER
36
Blood Pressure
RANGE +/- 500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm (synched to ECG)
ACCURACY +/- 1%
ECG NSR
RATE 30,60,120,240 BPM
ACCURACY +/- 1%
AMPLITUDE 2.75 mV
ACCURACY +/- 2% @ Lead II
37
Pacemaker Waveforms
AMPLITUDE 3 mV
ACCURACY +/- 10%
WIDTH 6 ms
ACCURACY +/- 5%
Respiration
RATE ACCURACY +/- 1%
IMPEDANCE DELTA 3.0 ohms
ACCURACY +/- 10%
BASELINE 1000 ohms
ACCURACY +/- 5%
Temperature
0,24,30,37,40 C (32.0,
Temperature
75.2,86.0,98.6,104.0 F)
Accuracy +/- 0.1 C
Type YSI Series 400 and 700
38
Electrical
12VDC, 500mA
AC POWER ADAPTER
P:\Manuals\BCGroup\…\NIBP1000_UM_Rev10.doc