XDR1
XDR1
Rakesh Pani
Introduction
X-rays were discovered by Wilhelm Roentgen who called them x - rays
because the nature at first was unknown so, x-rays are also called Roentgen
rays. X-ray diffraction in crystals was discovered by Max von Laue. The
wavelength range is 0.01 to about 10 nm.
Wilhelm Röntgen
X-rays are short wave length electromagnetic radiations produced by the deceleration
of high energy electrons or by electronic transitions of electrons in the inner orbital of
atoms
The penetrating power of x-rays depends on energy also, there are two types of x-rays.
i) Hard x-rays: which have high frequency and have more energy.
ii) soft x-rays: which have less penetrating and have low energy.
2
Max von Laue
Introduction
X-ray Diffraction (XRD) is a non-contact and non-destructive technique used to understand the
crystalline phases, different polymeric forms and the structural properties of the materials
X – ray diffraction
“ Every crystalline substance gives a pattern; the same substance always gives the same
pattern; and in a mixture of substances each produces its pattern independently of the others”
The X-ray diffraction pattern of a pure substance is, therefore, like a fingerprint of the substance.
It is based on the scattering of x-rays by crystals.
Definition
The atomic planes of a crystal cause an incident beam of X- rays to interfere with one another
as they leave the crystal. The phenomenon is called X- ray diffraction.
3
Timeline
4
Why XRD?
Measure the size, shape and internal stress of small crystalline regions
5
What is Diffraction ?
A diffracted beam may be defined as a beam composed of a large number of scattered rays
mutually reinforcing each other
Scattering
Interaction with a single
particle
Diffraction
Interaction with a
crystal
6
What is X-ray Diffraction ?
Constructive
interference
X-ray 2 occurs only
when Sir William
n λ = AB + BC Henry Bragg
AB=BC
n λ = 2AB Sin θ = AB/d
AB=d sin θ
n λ =2d sin θ
λ = 2 d hklsin θhkl
Sir Lawrence Bragg
8
Order of Diffraction
Rewrite Bragg’s law λ=2 sinθ d/n
A reflection of any order as a first order Diffraction from planes, real or fictitious,
spaced at a distance 1/n of the previous spacing.
An nth order reflection from (hkl) planes of spacing d may be considered as a first
order Diffraction from the (nh nk nl) plane of spacing d’= d/n
9
Basics of Crystallography
The atoms are arranged in a regular pattern, and there is as
smallest volume element that by repetition in three dimensions
describes the crystal. This smallest volume element is called a
unit cell.
10
Miller Indices
Crystal structures are made up of a series of planes of
atoms in which each plane is spaced with a distance d with
each other. But various atomic planes in a crystal can be
resolved with different d- spacing. For distinguishing
different planes there is a coordinate system introduced by
William Hallowes Miller called Miller indices (i.e., h, k, l).
William Hallowes
Miller
Miller indices-the reciprocals of the fractional intercepts
which the plane makes with crystallographic axes.
Reciprocals are taken to avoid ∞ in the indices.
11
Instrumentation
Production of x-rays (Source Of X-Rays)
Collimator
Monochromator
a. Filter
b. Crystal Monochromator
Detectors
a.Photographic methods
b.Counter methods
12
Production of x-rays
X-rays are produced whenever a
charged particles are accelerated.
In XRD, X-rays are generated
when high velocity electrons
impinge on a metal target.
A source of electrons – hot W
filament, a high accelerating
voltage between the cathode (W)
and the anode and a metal target,
Cu, Al, Mo, Mg.
The anode is a water-cooled
block of Cu containing desired
target metal.
13
Production of x-rays
14
Collimator
15
Monochromator
A monochromator is an
optical device that transmits
a mechanically selectable
narrow band of wavelengths
of light or other radiation
chosen from a wider range of
wavelengths available at the
input. The name is from the
Greek roots mono-, "single",
and chroma, "colour", and
the Latin suffix -ator,
denoting an agent.
16
Types Of Monochromators
Monochromatization can be broadly divided into two,
1.Interference Filters
2.Crystal Monochromator
17
Interference Filters
X-ray beam may be partly monochromatized by insertion of
a suitable filter.
18
Crystal monochromators
Crystal monochromators are made up of suitable crystalline material positioned in the x-ray beam so that the
angle of reflecting planes satisfied the Bragg’s equation for the required wavelength the beam is split up into
component wavelengths crystals used in monochromators are made up of materials like NaCl, lithium
fluoride , quartz etc. Pyrolytic graphite can be used for broad band and silicon for narrow band.
19
Detectors
X-ray detectors are devices used to measure the
flux, spatial distribution, spectrum, and other
properties of X-rays.
20
Types of detectors
The x-ray intensities can be measured and recorded either by
1)Photographic methods
2)Counter methods
Both these types of methods depends upon ability of x-rays to ionize matter and differ only in the
subsequent fate of electrons produced by the ionizing process.
21
Photographic method
To record the position and intensity of x-ray beam a plane or cylindrical film is used.
The film after exposing to x-ray is developed
The blackening of the developed film is expressed in terms of density units D given by
D = log I₀/I, I₀-incident intensity
I - Transmitted intensity
D - Total energy that causes blackening of the film D is
measured by densitometer
The photographic method is mainly used in diffraction
studies since it reveals the entire diffraction pattern on a single film .
Disadvantage : time consuming and uses exposure of several hours
22
A) Geiger - Muller tube counter
Geiger tube is filled with inert gas likeargon
Central wire anode is maintained at a positive potential of 800 to 1500V .
The electron is accelerated by the potential gradient and causes the ionization of
large number of argon atoms, resulting in the production of avalanche of electrons
that are travelling towards central anode
23
B) Proportional counter
Construction is similar to Geiger tubecounter
24
C) Scintillation detector:
26
X-ray diffraction methods
There are several XRD methods which are generally used for investigating the
internal structures and crystal structures of various solid compounds.
27
The Laue method
Laue in his very first experiments used white radiation of all possible wavelengths and
allowed this radiation to fall on a stationary crystal. The crystal diffracted the X-ray beam and
produced a very beautiful pattern of spots which conformed exactly with the internal
symmetry of the crystal. Let us analyze the experiment with the aid of the Bragg equation.
The crystal was fixed in position relative to the X-ray beam, thus not only was the value for d
fixed, but the value of was also fixed.
30
Disadvantages of Laue method
Big crystals are required
Crystal orientation is determined from the position
of the spots. Each spot can be indexed,
i.e. attributed to a particular plane, using
special charts.
The Greninger chart is used for back-reflection
patterns and the Leonhardt chart for transmission
patterns.
The Laue technique can also be used to assess
Back-reflection Laue method
crystal perfection from the size and shape
31
The Bragg’s x-ray spectrometer
C - crystal
E - ionization chamber
One plate of ionization chamber is connected to the positive terminal of a H.T Battery ,
while negative terminal is connected to quadrant electrometer(measures the strength of
ionization current) 32
Working of Bragg’s x-ray spectrometer
Crystal is mounted such that ѳ=0°and ionization chamber is adjusted to receive x-rays
Crystal and ionization chamber are allowed to move in small steps
The angle through which the chamber is moved is twice the angle through which the crystal is
rotated
X-ray spectrum is obtained by plotting a graph between ionization current and the glancing angleѳ
Peaks are obtained, corresponding to Bragg’s diffraction.
Different order glancing angles are obtained with known values of λ and n and from the observed
value of ѳand d can be measured.
33
Determination of crystal structure by bragg’s law
X-Rays fall on crystal surface
The crystal is rotated and x-rays are made to reflect from various lattice planes
The intense reflections are measured by bragg’s spectrometer and the glancing
Then on applying bragg’s equation ratio of lattice spacing for various groups of planes
can be obtained.
Experimentally observed ratio’s are compared with the calculated ratio’s ,particular
Fine powder is struck on a hair with a gum ,it is suspended vertically in the axis of a cylindrical camera
When monochromatic beam is allowed to pass different possibilities may happen
1.There will be some particles out of random orientation of small crystals in the fine powder
2.Another fraction of grains will have another set of planes in the correct positions for the
reflections to occur
3. Reflections are possible in different orders for each set
36
Powder crystal method:
POWDER CRYSTALDIFFRACTION
37
Applications
• The electron density and accordingly, the position of the atoms in complex structures, such as
penicillin may be determined from a comprehensive mathematical study of the x-ray
diffraction pattern.
• The elucidation of structure of penicillin by xrd paved the way for the later synthesis of penicillin.
• The powder xrd pattern may be thought of as finger print of the single crystal structure, and it may
be used conduct qualitative and quantitative analysis.
• XRD can also be used to determine whether the compound is solvated or not.
• Particle size determination by applying the relation.
v= V. δθ. cos θ / 2n
It is used to assess the weathering and degradation of natural and synthetic , minerals.
38
Limitations
Homogeneous and single phase material is best for identification of an unknown
Must have access to a standard reference file of inorganic compounds (d-spacings, hkls)
For unit cell determinations, indexing of patterns for non-isometric crystal systems is
complicated
Peak overlay may occur and worsens for high angle 'reflections'
39
Conclusions
For materials including metals, minerals, plastics, pharmaceuticals and semiconductors XRD
apparatus provide highly accurate tools for non-destructive analysis.
The diffraction systems are also supported by an extensive range of application software.
• Phase analysis
• Lattice parameter determination
• Strain determination
• Texture and orientation analysis
• Order-disorder
transformation and many
more things.
40
X-RAY FLUORESECENCE
Rakesh Pani
X-ray fluorescence
Polymers
Oils
Metals, Slags
Chemicals
Ceramic
Glass
Food Products
Typical Samples for X-Ray Fluorescence
Fundamental Principles of XRF
• XRF works on methods involving interactions between electron
beams and x-rays with samples.
• Made possible by the behavior of atoms when they interact with radiation.
• When materials are excited with high-energy, short wavelength radiation
(e.g., X-rays), they can become ionized.
• If the energy of the radiation is sufficient to dislodge a tightly- held inner
electron, the atom becomes unstable and an outer electron replaces the
missing inner electron.
• When this happens, energy is released due to the decreased binding energy
of the inner electron orbital compared with an outer one.
• The emitted radiation is of lower energy than the primary incident X-rays
and is termed fluorescent radiation.
• Because the energy of the emitted photon is characteristic of a transition
between specific electron orbitals in a particular element, the resulting
fluorescent X-rays can be used to detect the abundances of elements that are
present in the sample.
Excitation of the sample
Sample
Secondary X-Rays or X-
Ray Fluorescence which is
characteristic for the
elemental composition of
the sample
Spectrometer
Principle of the excitation by X-Rays
X-Ray Photon
Layer L
Layer K
Layer N
Layer M
XRF - HOW IT WORKS
• An XRF spectrometer works because if a sample is illuminated by an
intense X-ray beam, known as the incident beam, some of the energy is
scattered, but some is also absorbed within the sample in a manner that
depends on its chemistry.
• The incident X-ray beam is typically produced from a Rh target, although
W, Mo, Cr and others can also be used, depending on the application.
• When x-ray hits sample, the sample emits x-rays along a spectrum of
wavelengths characteristic of the type of atoms present.
• If a sample has many elements present, the use of a Wavelength Dispersive
Spectrometer allows the separation of a complex emitted X-ray spectrum
into characteristic wavelengths for each element present.
• Various types of detectors used to measure intensity of emitted radiation.
• Examples of detector used the flow counter and the scintillation detector.
• Flow counters measure long wavelength(>0.15nm) x-rays typical of
elements lighter than zinc.
• The scintillation detector is commonly used to analyze shorter
wavelengths in the X-ray spectrum(K spectra of element from Nb to I; L
spectra of Th and U).
Benefits of XRF analysis
• bulk chemical analyses of major elements (Si, Ti, Al, Fe, Mn, Mg, Ca, Na, K,
P) in rock and sediment
• bulk chemical analyses of trace elements (>1 ppm; Ba, Ce, Co, Cr, Cu, Ga,
La, Nb, Ni, Rb, Sc, Sr, Rh, U, V,Y, Zr, Zn) in rock and sediment
Limitations
In theory the XRF has the ability to detect X-ray emission from virtually all
elements, depending on the wavelength and intensity of incident x-rays.
However...
ED-XRF
SEM-EDX
Handheld XRF
research tools for the physicists and chemists since these techniques provide structural
spectroscopy techniques enjoy the advantages over the conventional IR and Raman
spectrometer depends on the signal to noise ratio (SNR) value. The improved SNR
spectrometers. Frequency precision, good spectral subtractions and high resolution are
increasing the spectral sensitivity and thus yielding new information about the atomic
or molecular species or the functional group molecular bonds that exist in the sample.
high energy sources like lasers which allow spectrum storage and retrieval, scale
control of slit, etc. Often these are under the control of a microprocessor and
and microcells can usually be added to extend versatility or accuracy. In this chapter
Change in electric dipole moment is necessary for IR activity, whereas the Raman
criterion depends on the change in polarizability during vibration. The intensity of the
38
IR and Raman spectra mainly depends on the magnitude of the change in dipole
moment and the polarizability respectively. The IR and Raman methods are used to
study the materials existing in various states. The molecular vibrational frequencies
obtained by these methods have been used in molecular mechanics and they yield
Infrared (IR) spectroscopy is one of the most common and widely used
certain wavelength region. The absorption peaks within this region are usually sharper
when compared with absorption peaks from the ultraviolet and visible regions. In this
within a sample since different functional group absorbs different particular frequency
data bank of spectra. IR spectroscopy is very useful in the identification and structure
It can also be used for both qualitative and quantitative analysis of complex mixtures
of similar compounds.
3.3 INSTRUMENTATION
3.3.1 Introduction
The basic components of an infrared spectrophotometer are shown in Fig.3.1.
A source provides radiation over the entire range of the infrared spectrum. The
monochromator disperses the light and then selects a narrow wave number range. The
39
detector measures the energy and transforms it into an electric signal. This signal is
considered to cover the range from approximately 12500-10 cm-1 (0.8-1000 m). It is
generally subdivided into three sections: near infrared (12500-4000 cm-1), mid
infrared region (4000 - 400 cm-1) and the far infrared (400-10 cm-1). The mid infrared
covers almost all the vibrational and rotational transitions. The FT-IR spectra of most
of the samples were recorded in mid and far IR regions and are used for the present
investigation [2-7].
RECORDER AMPLIFIER
ideal infrared source gives a continuous and high radiant energy output over the entire
infrared region. The total amount of energy radiated and the spectral distribution of
this energy are dependent upon the temperature of the source. The usual sources of
infrared radiation are Nernst glower and Globar. The Nernst glower is composed of
1800ºC. The Globar is a small rod of silicon carbide usually 5 cm in length and 0.5
maximum radiation for the Globar occurs in the 5500-5000 cm-1 region. Nichrome
40
wire, carbon arc, rhodium wire and tungsten filament lamp are also used as light
filament contained in a ceramic tube is commonly used as infrared source for the
3.3.3 Monochromator
gratings. A grating or a prism disperses the radiation from the source into its spectral
elements according to the equation nλ=2dsinθ. The most common prism material is
NaCl and it has a lower frequency cut-off at about 650 cm-1. It has a good resolution
in the range 1000 to 650 cm-1 and moderate resolution throughout 4000 to 1000 cm-1.
radiation according to its wave number components (ii) it restricts the radiation falling
on the detector into a narrow wave number range, and (iii) it maintains the energy
throughout the wave number range of the instrument [9]. Some instruments use a
double monochromator. That is, the exit slit of the first monochromator serves as the
entrance slit for the second monochromator. As a result, the spectra obtained with
3.3.4 Detector
radiant energy into electrical energy. The infrared detectors may be selective or non-
selective. The selective detectors are those whose response is markedly dependent
upon the wavelength of the incident radiation. Examples of this type are photocells,
41
photographic plates, photoconductive cells and infrared phosphors. Photoconductive
cell has a rapid response and high sensitivity. These cells are made of materials such
as PbS, lead selenide or selenium. When illuminated by infrared light these cells show
an increase in conductivity. These cells are useful in the range 0.5 and 3.5 microns.
from crystals are known as pyroelectrics and take less time than other thermal
detector. Hence with these crystals, radiation can be chopped at a higher rate.
electrical signal and is amplified by the amplifiers. The amplified signal is registered
with that of a monochromator, so that, the pen moving across the chart, records the
spectroscopy. There are various methods of sample preparation to enable almost any
construct sample containers for vibrational spectrometry, because every material has
some vibrational absorption. The material that has a minimum interference in the
solid potassium bromide plate. Such plates are used in a number of ways.
42
(a) Solids
Solids are sampled in a wide variety of ways. If the sample is soluble, it may be
dissolved and handled as for a liquid. Solid samples for which no solvent is suitable
can be prepared for analysis by incorporating them into a pressed pellet of alkali
halide, usually potassium bromide. The sample is mixed with a weighed amount of
tones in a die, to produce a highly transparent plate or disc which can be inserted into
the spectrophotometer. The use of KBr eliminates the problems of additional bands
due to mulling agent. KBr does not absorb infrared light in the region 2.5–15 m and a
complete spectrum of the sample is obtained. Solid samples have also been examined
surface of a salt plate. Another method, called mulling has also been developed, in
which the powdered sample is mixed to form a paste with little heavy paraffin oil. The
mull is sandwiched between salt plates for measurement. Mulls are formed by
(b) Liquids
The spectra of a pure liquid can be measured as very thin films squeezed
between two alkali halide windows of a demountable cell. This technique can produce
a film of thickness 0.01 mm or less. The cells are then taken apart and cleaned. This
method is most useful for qualitative work only because the sample thickness cannot
be controlled. For quantitative work sealed liquid cells of fixed path length in the
range 0.01 to 0.1mm are used. Liquid cells consist of two alkali halide windows
43
(c) Gases
The vapour is introduced into a special cell, usually about 10cm long that can
be placed directly in the path of one of the infrared beams. The end walls of the cell
are usually made of sodium chloride, which is transparent to infrared. Most organic
compounds have too low a vapour pressure for this phase to be useful. The low
frequency vibrational changes in the gaseous phase often split the high frequency
(d) Solvents
available and the spectra of the sample dissolved in carbon tetrachloride and carbon
solvent and is frequently used because it shows absorptions though it has less
uses an interferometer and (b) a dedicated computer which stores data performs
computations on data and plots the spectra. A schematic diagram of the essential
44
Fig.3.2. It consists of two perpendicular mirrors; one of which is a stationary mirror
and the other a movable mirror which can be displaced perpendicularly to the fixed
mirror at a constant velocity. Between these two mirrors the beam splitter is set at 450
from the initial position of the movable mirror. A parallel beam of radiation from an
infrared source is passed to the mirrors through the beam splitter. The beam splitter
reflects about half of the beam to the fixed mirror which reflects it back to the beam
splitter and transmits the other half to the movable mirror which reflects it back to the
beam splitter. The returning beams are again split and mixed about half going back to
source and half passing through the sample compartment. The composition of the
beam splitter depends on the spectral region of interest. For example, in the mid-
infrared region (4000-400 cm-1), a beam splitter of germanium coated on KBr plate
(substrate) is often used. Germanium reflects the radiation while KBr transmits most
of the desirable radiation. In the far infrared region, germanium coated on CsI
(650–10 cm-1) is used as beam splitter. A thin film of the beam splitter material is
The return beams from both the mirrors along the same path length as their
incident path are recombined into a single beam at the beam splitter. The path length
of one of the return beams is changed in order to create phase difference to cause an
interference pattern. The recombined radiation is then directed through the sample and
focused on to the detector. The detector measures the amount of energy at discrete
The movable mirror can be moved in a range of say ±5 cm. The mirror
velocities from 0.05 to 5 cm s-1 are used. Interferometer instruments need detectors
with response times short enough to detect and transmit rapid changes to the recorder.
45
The detector used in conjunction with rapid scanning interferometers in the mid-
pyroelectric bolometer. It has a high response time. Other most common detectors
used such as thermocouples, bolometers and Golay detectors have short response
any infrared region possible by simply changing the beam splitter and the detector [2].
The main advantages of FT spectroscopy are the greater ease and speed of
measurement. The entire spectrum can be recorded within few seconds using
led to higher resolution, total wavelength coverage, higher accuracy in frequency and
samples. In FT method, all the source energy passes through the instrument and the
resolving power is constant over the entire spectrum. The signal to noise ratio is also
improved [15]. The smoothening of peaks and the vertical and horizontal expansion
The Raman spectroscopy is made sophisticated with the advent of gas lasers
and computers. The advantages of lasers are their high intensity, high
monochromaticity, narrow band width, high resolution and coherence. From the time
compounds have been effectively used for the determination of molecular structure
and also for the quick identification of the presence of the characteristic group
46
Fig.3.2 Simplified diagram of Fourier transform infrared spectrometer.
47
In order that for a molecular vibration to be Raman active there must be a
spectra over infrared lies in the fact that water does not cause interference; indeed,
Raman spectra can be obtained from aqueous solutions. In addition, glass or quartz
cells can be employed, thus avoiding the inconvenience of working with sodium
conventional dispersive Raman spectroscopy is its ability to render spectra that are
3.8 INSTRUMENTATION
3.8.1 Introduction
In Raman spectrometer the sample is irradiated with monochromatic light and
the scattered light is observed at right angles to the incident radiation. Raman
3.8.2 Source
The sources used in modern Raman spectroscopy are nearly always lasers
intensity to be measured with a reasonable signal to-noise ratio. Five of the most
common lasers along with their wavelength (nm) used for Raman spectroscopy are
Argon ion (488 or 514.5 nm), Krypton ion (530.9 or 647.1 nm), Helium / Neon (632.8
nm), Diode laser (782 or 830 nm) and Nd: YAG (1064 nm). Because the intensity of
Raman scattering varies as the fourth power of the frequency, argon and krypton ion
48
48
sources that emit in the blue and green region of the spectrum have an advantage over
the other sources. Diode and Nd:YAG laser which emit near-infrared radiation are
used as powerful excitation sources. Near-infrared sources have two major advantages
over shorter wavelength lasers. The first is that they can be operated at much higher
power (upto 50W) without causing photodecomposition of the sample. The second is
that they are not energetic enough to populate a significant number of fluorescence
fluorescence is generally much less intense or non-existent with these lasers. The
two lines of the diode array laser at 782 and 830 nm also markedly reduce
The laser radiation is directed to the sample by means of a lens and a parabolic
mirror and the scattered light from the sample is collected and passed to a beam
splitter and to the moving mirrors. It is then passed through a series of dielectric filters
3.8.3 Detectors
Basically there are two different ways to detect and record Raman lines. The easiest
way is to gather the scattered light emerging through a glass window at the end of the
Raman sample tube. It is passed through a prism or grating and then focused on a
photographic plate. The plate is then developed and both the line frequencies and
have photo multiplier tubes are direct measurements and facilitate automatic scanning
49
photo multiplier type detector. This detector employs an amplifier and a recorder. It
than for infrared spectroscopy because glass can be used for windows, lenses and
other optical components instead of the more fragile and atmospherically less stable
crystalline halides. In addition, the laser source is easily focused on a small sample
area and the emitted radiation efficiently focused on slit. Consequently, very small
about 0.3 ml of a liquid may be required. The sample could be taken in glass or silica
containers or capillaries. The spectra can be measured directly from the reaction
vessel. Water is a good solvent for recording the Raman spectra. Water absorbs
strongly in the infrared but it gives a poor Raman scattering. Raman spectroscopy is
can be recorded. No medium such as null, KBr or solvent is needed. A few milligrams
of the solid samples are required. Solid can be packed into a capillary tube as a
powder. The crystal can be mounted in a goniometer on a glass or silica fibre. The
spectra can be measured for different orientation of the crystal. For a single crystal,
the Raman spectrum varies depending on the direction of the crystal axis, when
50
polarized light is used as incident radiation. Raman spectra of adsorbed species can be
and hence may require cells of larger path length. The gas may be filled in a glass or
silica tube of 1 to 2 cm diameter. If the resolving power of the instrument is good and
if the molecule has sufficiently low moment of inertia, the rotational fine structure
may be observed on either side of the Rayleigh line. Generally a broad band contour
may be observed. The main advantage of Raman spectroscopy is that it may be used
for a wide variety of sizes and forms of the sample. Samples in gas, liquid and solid
states can be examined easily. In this study, the spectral data are recorded on
infrared laser, the collection of the scattered light and its analysis using appropriately
transform Raman spectra are almost exclusively obtained with neodymium yttrium
aluminum garnet (Nd: YAG) lasers. The laser radiation is filtered to perfect its
monochromaticity and is then focused on to the sample. Light reflected and scattered
off the sample in a direction the reverse of illumination the so-called back scattering
arrangement is then filtered to remove the Rayleigh scattered light and allow only the
Raman scattering to pass. The noise associated with the intense Rayleigh scattering is
distributed over the entire spectrum in the Fourier transformation step and it seriously
scattering by means of a laser line rejection filter must be achieved before good
51
quality FT-Raman spectra can be obtained. Then the scattered light passes through a
component capable of operating from 1nm and through the mid infrared or it may be a
limited range device. The interferogram is then collected and detected on a near
infrared detector. The germanium photo resistor operating at the liquid nitrogen
temperature or indium doped gallium arsenide photo detector which operates at room
spectra can be interpreted in terms of unit cells which may contain a few elements of
different polymer chains. Since peak overlap is low in Raman spectroscopy, mixtures
spectra have been carried out by following empirical correlation of group frequencies,
infrared and Raman selection rules, the magnitude and relative intensities of the
spectra [17, 18]. In the case of polyatomic molecules, the interpretation of spectra is
52
Fig.3.3 Basic diagram of an FT-Raman spectrometer.
53
3.13 ULTRAVIOLET SPECTROSCOPY
the transitions between the electronic energy levels. It helps to identify the functional
groups and nature of the bond in the organic compounds. It also provides the
3.13.1 Instrumentation
radiation coming from the monochromator is split into beams with the help of a beam
splitter. These are passed simultaneously through the reference and the sample cell.
The transmitted radiations are detected by the detectors and the difference in the
signal at all the wavelengths is suitably amplified and sent for the output. UV-visible
hydrogen or deuterium lamps, the xenon discharge lamps and mercury arcs. The
sources should provide stable output over the entire UV- visible range (190 nm to 780
nm). For measurements in the UV region, electric discharge sources like hydrogen or
a deuterium lamp are used. In these, the excitation of the gaseous molecules is
brought about by the passage of electrons through the gas at low pressures. A
wavelength region of 160-375 nm. The radiant power of the hydrogen lamp is low
filament lamp as the radiation source. This consists of a thin, coiled tungsten wire that
54
350-2200 nm. As the output depends on the voltage, the tungsten lamp is energized by
3.13.3 Monochromators
The monochromator is used to disperse the radiation according to the
dispersing element and an exit slit. The entrance slit sharply defines the incoming
heterochromatic radiation into its component wavelengths, whereas exit slit allows the
nominal wavelength together with a band of wavelengths on either side of it. The
position of the dispersing element is always adjusted by rotating it to vary the nominal
wavelength passing through the exit slit. The dispersing element may be a prism or
grating. Quartz and fused silica prisms which are transparent throughout the entire
3.13.4 Detectors
The detectors are used to convert a light signal into an electric signal which
can be suitably measured and transformed into an output. The detectors used in most
of the instruments generate a signal, which is linear in transmittance i.e. they respond
linearly to the radiant power falling on them. The transmittance values can be changed
the signal read out device. Phototube, photomultiplier tube and diode array detectors
The UV-vis absorption spectra are usually determined either in vapour phase
called a cuvette which is transparent to the wavelength of light passing through it.
55
Fig.3.4 UV-visible double beam spectrophotometer.
56
Most of the spectrophotometers employ quartz cuvettes for both visible and UV
transparent in the UV region. Hexane, ethanol and methanol are commonly employed
taken in a suitable cuvette and the spectrum is run in the desired range of the
spectrum for the solvent alone in the same or identical cuvette and subtracting it from
the spectrum of the solution. This gives the spectrum only due to the absorption
species under investigation. In double beam spectrometers, the sample and the
57
REFERENCES
1. P.S. Sindhu, Molecular Spectroscopy, Mc Graw Hill, New Delhi, 1985.
Edition, 1995.
4. H.H. Willard, L.L. Merritt, J.A. Dean, F.A. Settle, Instrumental Methods of
8. N.B. Colthup, L.H. Daly, S.E. Wiberly, Introduction to Infrared and Raman
10. R.A. Nyquist, C.L. Pulzig, M.A. Leugers, The Hand Book of Infrared and
58
14. P.R. Griffiths, J.A. de Haseth, Fourier Transform Infrared Spectroscopy,
59