English                                    Instruction SheetPanel 1
                                                                707
                                              Loop Calibrator
                                                Instruction Sheet
Introduction
The Fluke 707 Loop Calibrator (hereafter, the Calibrator) is a
compact sourcing and measuring tool. The Calibrator tests current
loops of 0-20 mA or 4-20 mA and measures dc voltage to 28 V. It
comes with a set of alligator-clip test leads, a 9 V alkaline battery,
and this Instruction Sheet.
The Calibrator is a IEC 61010, CAT I 30 V, Pollution Degree 2
instrument. A CAT I instrument is designed to protect against
transients from high-voltage, low-energy sources, like electronic
circuits or a copy machine, for example.
                        Calibrator Capabilities
         Function                  Range               Resolution
 Measure V dc                   28 V                     1 mV
 Measure mA dc
                                0 to 24 mA               1 μA
 Source mA dc
 Source loop power              24 V dc                  N/A
Battery Saver
The Calibrator automatically turns off after 30 minutes of inactivity.
To reduce this time or disable this feature:
1.   With the Calibrator OFF, press D.
     PSXX is displayed, where XX is the turn-off time in minutes.
     OFF means the power saver is disabled.
2.   Turn m to decrease or increase the turn-off time.
     To disable, turn m until the display shows OFF.
3.   The Calibrator resumes normal operation after 2 seconds.
August 2001 (English) Rev. 2, 2/08
© 2001-2008 Fluke Corporation. Product specifications are subject to
change without notice. All rights reserved.
English                                Instruction SheetPanel 2
                 W Warnings and Cautions
    To avoid electric shock, injury, or damage to the
    Calibrator:
    •    Use the Calibrator only as described in this
         Instruction Sheet or the protection provided by the
         Calibrator may be impaired.
    •    Do not use the Calibrator around explosive gas,
         vapor, or dust.
    •    Inspect the Calibrator before use. Do not use it if
         appears damaged.
    •    Check the test leads for continuity, damaged
         insulation, or exposed metal. Replace damaged
         test leads.
    •    Never apply more than 30 V between any two
         terminals, or between any terminal and earth
         ground.
    •    Use the proper terminals, mode, and range for your
         measuring or sourcing application.
    •    To prevent damage to the unit under test, put the
         Calibrator in the correct mode before connecting
         the test leads.
    •    When making connections, connect the COM test
         lead before the live lead; when disconnecting,
         disconnect the live lead before the COM lead.
    •    Never use the Calibrator with the case open.
    •    Make sure the battery door is closed before you
         use the Calibrator.
    •    Replace the battery as soon as the M (low battery)
         symbol appears to avoid false readings that can
         lead to electric shock.
    •    Remove test leads from the Calibrator before
         opening the case or battery door.
Symbols
 Symbol                            Meaning
   +       ON / OFF button.
   J       Earth ground
   W       Caution: Important information. Refer to instruction sheet
   h       Caution: Static discharge can damage parts
   T       Double insulated
   M       Battery
           Conforms to relevant Canadian Standards Association
   )
           directives. Certification # LR110460-2.
   P       Conforms to European Union requirements
   F       Direct current
           Do not dispose of this product as unsorted municipal
   ~
           waste. Go to Fluke’s web site for recycling information.
English                                   Instruction SheetPanel 3
Pushbutton Functions
  Pushbutton                              Function
     D              ON or OFF button.
                    Press A and D simultaneously to toggle
  A+D               between the mA output spans.
                    •    4 mA to 20 mA = 0 % - 100 % (default)
(Power-on Option)   •    0 mA to 20 mA = 0 % - 100 % (optional)
                    The selection is saved until it is changed.
  l+ D             Press l     + D simultaneously to turn HART
(Power-on Option)   resistor (Hr) on. Default is off.
                    Press to step through modes:
      A             •    Source mA
                    •    Simulate mA
                    •    Measure mA
                    •    Loop Power (24 V)
                    •    Measure V dc
                    Turn n to increase or decrease current output.
                    Current output can be adjusted at a resolution of
                    1 μA or 100 μA. (Default is 1 μA.)
                    •    To adjust the current in 1 μA steps, simply
                         turn the knob.
                    •    To adjust the current in 100 μA steps, press
hij                      in and turn the knob.
                    Press B to step the current up 25 % of full
     B              scale (20 mA).
                    At full scale, press B to step the current down
                    25 % of full scale.
                    Press B + C simultaneously to enter the
 B+ C               Auto Ramp mode and select a ramp form.
                    A continuously applied or controlled mA ramping
                    signal is produced in one of three ramp forms.
                    e (slow), g (fast), or f (step) identifies the
                    selected ramp form.
                    Press C to start the SpanCheck™ at 0 % of
     C              selected current span, i.e., 0 mA for 0-20 mA span
                    or 4 mA for 4-20 mA span.
                    d is displayed.
                    Press again for 100 % of selected current span.
English                                                 Instruction SheetPanel 4
Using the mA Sourcing (Output) Modes
The Calibrator outputs current for calibrating and testing 0 to 20 mA
and 4 to 20 mA current loops and instruments.
In SOURCE mode, the Calibrator supplies the current.
In SIMULATE mode, the Calibrator simulates a 2-wire transmitter in
an externally-powered current loop.
Changing the mA Output Span
The Calibrator has two mA output spans:
•   4 mA to 20 mA (0 % to 100 %) [default]
•   0 mA to 20 mA (0 % to 100 %) [optional]
To change the output span, turn the Calibrator off. Press A + D
simultaneously. The selected setting is saved until it is changed
again.
Sourcing mA
Use SOURCE mode to supply current to a passive circuit.
A path must exist for current to flow between the + and COM
terminals. Otherwise the display indicates an overload (OL) when
you set an output value.
                                     707    LOOP CALIBRATOR
                                                              Press simultaneously
                                                                 once for slow ramp
                                                                 twice for fast ramp
                                                                 three times for
     Press until SOURCE                                          stepped ramp.
     mA is displayed.
                                                              Turn to step up/down
                                                              1 A. Press in and
     Press to step current                                    turn to step up/down
     up/down in 25 %                  CAT
                                                              in 100 A increments.
     steps of full scale.      COM
                                      30V
                                                                                ADA04F.EPS
English                                   Instruction SheetPanel 5
Contacting Fluke
To contact Fluke for product information, operating assistance,
service, or to get the location of the nearest Fluke distributor or
service center, call:
   1-888-44-FLUKE (1-888-443-5853) in U.S.A
   1-800-36-FLUKE in Canada
   +31-402-675-200 in Europe
   +81-3-3434-0181 Japan
   +65-738-5655 Singapore
   +1-425-446-5500 from other countries
Or visit Fluke's web site at: www.fluke.com.
Register your Calibrator at: http://register.fluke.com.
Address correspondence to:
   Fluke Corporation               Fluke Europe B.V.
   P.O. Box 9090,                  P.O. Box 1186,
   Everett, WA 98206-9090          5602 BD Eindhoven
   U.S.A.                          The Netherlands
             Limited Warranty & Limitation Of Liability
  This Fluke product will be free from defects in material and work-
  manship for 3 years from the date of purchase. This warranty
  does not cover fuses, disposable batteries, or damage from
  accident, neglect, misuse, alteration, contamination, or abnormal
  conditions of operation or handling. Resellers are not authorized
  to extend any other warranty on Fluke’s behalf. To obtain service
  during the warranty period, contact your nearest Fluke authorized
  service center to obtain return authorization information, then
  send the product to that Service Center with a description of the
  problem.
  THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
  WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
  PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT
  LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
  CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM
  ANY CAUSE OR THEORY. Since some states or countries do
  not allow the exclusion or limitation of an implied warranty or of
  incidental or consequential damages, this limitation of liability
  may not apply to you.
                                                                      11/99
English                                                 Instruction SheetPanel 6
Simulating a Transmitter
When simulating the operation of a transmitter, the Calibrator
regulates the loop current to a known value selected by you.
A 12 V to 28 V loop supply must be available. Insert the test leads as
shown below.
                                     707    LOOP CALIBRATOR
                                                              Press simultaneously
                                                                 once for slow ramp
                                                                 twice for fast ramp
                                                                 three times for
    Press until                                                  stepped ramp.
    SIMULATE mA is
    displayed.
                                                              Turn to step
                                                              up/down 0.1 A.
           Press                                              Press in and rotate
    to step current                                           to step up/down in
                                      CAT
    up or down 25 %            COM
                                      30V
                                               +              100 A increments.
    of full scale.
                            30 V dc                Power
                             max.                  supply
                                                                              ADA05F.EPS
English                                              Instruction SheetPanel 7
Auto Ramping the mA Output
Auto ramping allows you to continuously apply a varying current from
the Calibrator to a passive (sourcing) or active (simulate) loop. Your
hands remain free to test the transmitter’s response.
Press B + C simultaneously to enter the Auto Ramp mode and
step to a ramp type.
The Calibrator applies or controls a continuously repeating mA signal
over a 0-20 mA or a 4-20 mA span in one of three ramp types:
Slow (e)         0 % to 100 % to 0 % smooth ramp over 40 sec.
Fast (g)         0 % to 100 % to 0 % smooth ramp over 15 sec.
Step (f)         0 % to 100 % to 0 % stair-step ramp in 25 % steps,
                 pausing 5 sec at each step.
To exit, press any pushbutton or turn the Calibrator off.
Using the SpanCheck Function
The SpanCheck™ function checks the zero and span points of a
transmitter in either SOURCE or SIMULATE mode.
To select SpanCheck, press C.
To exit, press any pushbutton or turn the knob.
            or
                                           707    LOOP CALIBRATOR
      Put in SOURCE mA
      or SIMULATE mA.
      Press to start
      SpanCheck at 0%.
      Press again for 100%.                 CAT
                                            30V
                                     COM             +
                                                              Device under
                                                                  test
                                30 V dc max.
                                                                             ADA02F.EPS
English                                       Instruction SheetPanel 8
Measuring dc mA
                               W Caution
     To prevent damage to the unit under test, ensure that the
     Calibrator is in the correct mode before connecting the
     test leads.
To measure dc mA:
1.    Press A to step to MEASURE mode.
      MEASURE mA is displayed.
2.    Touch test lead probes to the circuit across the load or power
      source as shown below.
                                            707    LOOP CALIBRATOR
       Press until MEASURE
       mA is displayed.
                                             CAT
                                             30V
                                      COM             +
                     30 V dc max.
                                                                     ADA03F.EPS
English                                              Instruction SheetPanel 9
Measuring dc mA with Loop Power
                               W Caution
     To prevent damage to the unit under test, ensure that the
     Calibrator is in the correct mode before connecting the
     test leads.
Loop power provides +24 V to power a transmitter and to read loop
current simultaneously.
To measure dc mA with Loop Power:
1.    Press A to step to Loop Power mode.
      MEASURE mA and Loop Power are displayed.
2.    Touch test lead probes to the circuit across the load or power
      source as shown below.
To exit Loop Power, change measurement mode.
                                           707    LOOP CALIBRATOR
         Press until Loop
         Power is displayed.
                                            CAT
                                                                          +24 V output
                                            30V
                                     COM             +
                                                                     Two-wire
                                                                    transmitter
                               30 V dc max.
                                                                                         ADA06F.EPS
English                                Instruction SheetPanel 10
Measuring dc Volts
                               W Caution
     To prevent damage to the unit under test, ensure that the
     Calibrator is in the correct mode before connecting the
     test leads.
To measure dc Volts:
1.    Press A to step to MEASURE mode.
      MEASURE V is displayed.
2.    Touch test lead probes across the load or power source.
                                            707    LOOP CALIBRATOR
          Press until MEASURE
          V dc is displayed.
                                             CAT
                                             30V
                                      COM             +
                      30 V dc max.
                                                                     ADA01F.EPS
English                              Instruction SheetPanel 11
Maintenance
                           W Warning
    To avoid electric shock, personal injury, or damage to the
    Calibrator:
    •    Do not service this product other than as described in
         this Instruction Sheet unless you are a qualified
         technician and have the required equipment and
         service information.
    •    Remove any input signals prior to removing test leads
         and opening case.
    •    When servicing the Calibrator, use only specified
         replacement parts.
    •    Do not allow water to get in the case.
For maintenance procedures not described in this Instruction Sheet,
contact a Fluke Service Center.
English                                 Instruction SheetPanel 12
In Case of Difficulty
•    Make sure you are using the Calibrator as described in this
     Instruction Sheet.
•    Check the battery and test leads. Replace as needed.
Contact a Fluke Service Center if the Calibrator needs repair or does
not seem to be operating properly.
If the Calibrator is under warranty, refer to the warranty statement for
warranty terms, conditions, and product return information.
If the warranty has lapsed, the Calibrator will be repaired and
returned for a fixed fee.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do not
use abrasives or solvents.
Calibration
Calibrate the Calibrator once a year to ensure that it performs
according to its specifications.
English                                  Instruction SheetPanel 13
Replacing the Battery
                              W Warning
     To avoid false readings, which could lead to electric shock
     or injury, replace the battery as soon as M (low battery
     indicator) appears on the display.
     Use only a single 9 V battery, properly installed, to power
     the Calibrator.
The Calibrator uses a single 9 V, alkaline battery (ANSI/NEDA
1604A or IEC 6LR61). To replace the battery:
1.   Press D to turn the Calibrator OFF.
2.   Remove the test leads from the terminals.
3.   Remove the holster.
4.   Lift off the battery door on the back of the Calibrator as shown.
5.   Remove the battery.
6.   Insert the replacement battery and restore the battery door.
     Make sure it is securely in place.
7.   Return the Calibrator to its holster.
                                                                ADA07F.EPS
HART Mode
To enable or disable the Calibrator's HART mode (Highway-
Addressable Remote Transducer), see the Power-on Option in the
Pushbutton Functions section. Default is HART resistor off.
English                                    Instruction SheetPanel 14
Replaceable Parts
.
.
.
                       5
          9                                                 6
                       7
                                                            8
                             10
                                              12
                                    11
                                                                   ADA10F.EPS
    Item                     Description                Part No.      Qty.
    1.        Rotary knob                               1618022       1
    2.        Case top                                  1618355       1
    3.        Keypad                                    1612222       1
    4.        Case bottom                               1618005       1
    5.        Non-skid foot                             885884        1
    6.        Battery. 9 V, Alkaline, ANSI / NEDA       614487        1
              1604A or IEC 6LR61
    7.        Case screws                               665098        4
    8.        Battery door                              665106        1
    9.        CD ROM                                    2088974       1
    10.       Test lead set                             TL75          1
    11.       Alligator clips                           AC72          1
    12.       Holster, yellow                           C10           1
English                                 Instruction SheetPanel 15
Accuracy Specifications
Accuracy is specified for 1 year after calibration at operating
temperatures of 18 °C to +28 °C and is given as:
                ±( [ % of reading ]+[ counts ] )
MEASURE V dc
   Range: +28 V (+30 V max)
   Resolution: 1 mV
   Input Impedance: 1 MΩ
   Accuracy: ±(0.015 % of reading+2 counts)
MEASURE mA dc
     Range: 20 mA (24 mA max)
     Resolution: 1 μA
     Accuracy: ±(0.015 % of reading+2 counts)
SOURCE / SIMULATE mA dc
   Range: 0 mA to 20 mA (24 mA max)
   Resolution: 1 μA
   Accuracy: ±(0.015 % of reading+2 counts)
  Source mode:
  Compliance: To 1200 Ω at 20 mA
                To 950 Ω at 20 mA in HARTTM Mode
  Simulate mode:
   External loop voltage requirement: 24 V nominal, 30 V maximum,
   12 V minimum
Loop Power
   ≥24 V
Percent display
   -25 % to 125 %
Input / Output Protection
   Fuseless protection
English                              Instruction SheetPanel 16
General Specifications
Maximum voltage between any terminal and earth ground or
between any two terminals:
   30 V
Storage temperature:
   -40 °C to 60 °C
Operating temperature:
   -10 °C to 55 °C
Operating altitude:
   3000 meters maximum
Temperature coefficient:
   ±0.005 % of range per °C for temperatures of -10 °C to 18 °C and
   28 °C to 55 °C
Relative humidity:
   95 % up to 30 °C;
   75 % up to 40 °C
   45 % up to 50 °C
   and 35 % up to 55 °C
Vibration:
   Random 2 g, 5 to 500 Hz
Shock:
   1 meter drop test
Safety Compliance:
   Complies with IEC 61010-1-95 CAT I, 30 V; CSA C22.2 No. 1010-
   992 NRTL; ANSI/ISA S82.02.01-1994.
CE:
   Complies with EN61010-1 and EN61326
Power requirements:
   Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Battery life (typical):
   SOURCE mode: 18 hours; 12 mA into 500 Ω;
   MEASURE / SIMULATE mode: 50 hours
Size:
   69.85 mm (W) x 142.87 mm (L) x 50.80 mm (H)
   [ 2.75 in (W) x 5.625 in (L) x 2.00 in (H) ]
   With holster and Flex-Stand:
   76.20 mm (W) x 158.75 mm (L) x 54.61mm (H)
   [ 3.00 in (W) x 6.25 in (L) x 2.15 in (H) ]
Weight:
   224 g (8 oz); With holster and Stand: 349 g (12.3 oz)
Protection Class:
   Pollution Degree II