A15 MS Eng
A15 MS Eng
Manual
English
Manual version Revision data Modification
02 April 20120 Added chapters: Notices and Warns,
Licences
Modification of the chapter: 3, 5.2.7, 6,
7, AII, AIII, AV, AVI
01 November 2017 Modification of the chapter 1.4, 4.2.3.4,
4.2.7 and AV
00 May 2017 Modification to add the refrigerator
model
http://www.biosystems.es
The Y15 is compliant with the low voltage directive 2014/35 / EU and the electromag-
netic compatibility directive 2014/30 / EU of the European Union.
TABLE OF CONTENTS
8
Persons for whom this manual is intended
This manual is intended for professionals belonging to the technical service who perform preventive maintenance
tasks on and repair the A15, A15C, Y15 and Y15C analyser. These professionals will have received special training
enabling them to perform the above-described tasks.
This manual describes the mechanical and electronic characteristics and service software to assist technicians in
performing maintenance and repair work. It also describes the steps for disassembling and changing the different
elements that comprise the analyser.
Symbol Description
The symbol warns of operating risks that could cause personal
injury.
WARNING
The symbol warns of a potential biological hazard.
BIOHAZARD
The symbols warns of potential damage to the system or unreliable
results.
CAUTION
The symbol warns that the information requires your attention.
NOTE
Risk of electric shock.
Explanations of the symbols used on the analyser labels and in the manual
Symbol Description
This product is compliant with European union directive
Serial number
9
Service manual
Symbol Description
Expiry date
Batch code
Catalogue number
Temperature limit
Manufacturer
Irritant
Safety precautions
Symbol Description
Preventing electric shock
To prevent the risk of electrocution. Do not remove any of the analyser
housing elements. No user intervention makes it necessary to access the
parts inside the equipment. If necessary, contact the technical assistance
service.
Preventing biological risks in handling the samples
Inappropriate handling of samples, controls and calibrators could cause
biological infection. Do not touch the samples, mixtures or waste with your
hands. Wear gloves and protective clothing when necessary.
In the event that the samples come into contact with the skin, wash
BIOHAZARD
immediately with abundant water and seek medical advice. It is advisable
to follow good laboratory practices.
10
Symbol Description
Prevention in handling reagents
Handle reagents and washing solutions with care, they contain substances
that could be corrosive.
In the event that the reagents or washing solutions come into contact with
WARNING the skin, wash immediately with abundant water and in the event of a
reaction, seek medical advice.
Consult the reagent or washing solution adaptation sheet and follow the
safety instructions.
It is advisable to follow good laboratory practices.
Preventing biological risks in handling solid waste
Take care in handling parts of the analyser that are converted to waste
such as the reactor rotor, sample tubes and reagent bottles.
Wear gloves and protective clothing when handling such waste.
BIOHAZARD Dispose of the waste in accordance with national or local legislation
for disposing of dangerous biological waste, and consult the reagent
manufacturer or distributor for more details.
Prevention of electro magnetic interferences
The analyser complies with the emission and immunity requirements
set out in standard UNE-EN 61326-2-6:2006. This equipment has been
designed and tested for class B of standard UNE-EN 55022:2000. In a
NOTE domestic environment, it may cause radio interference, in which case it
may be necessary to take measures to mitigate the interference.
Do not use the analyser near strong electro magnetic radiation sources
(such as centrifuge appliances, radio transmitters, mobile telephones), as
they could interfere with its correct operation.
Prevention at the end of the useful life of the analyser
At the end of the useful life of the analyser, the product must be disposed
of in accordance with the environmental legislation in force in each country.
If that country is a EU member state, the terms of the WEEE directive on
electrical and electronic appliances will apply. In other words, when the
appliance's useful life has ended, it is converted into waste and must be
separated from household waste for correct recycling. For this purpose,
contact the distributor for the product to be properly recycled.
Prevention when deactivating the cover sensor
The option to deactivate the covers sensor only serves to facilitate the
technical service that can carry out maintenance or repair actions. Once
the maintenance or repair is finished, the technician has to reactivate the
NOTE covers’ sensor again. The covers’ sensor are elements of security of the
analyzer.
Safety prevention for technical assistance service personnel
Wear gloves and protective clothing to manipulate the analyzer.
Before performing any manipulation, empty the analyzer of biological
material: samples, deposits, etc.
NOTE Be cautious when you have to repair analyzer points that have dangerous
voltages such as parts related to the power supplies.
Use caution when handling points that have contact with biological
material, such as tips, pipes and fittings where highly contaminated waste
flows.
Be sure to follow the instructions in the manual when you have to replace
parts or perform maintenance to ensure the safety of the service personnel
as well as the operator once the repair is complete.
11
Service manual
Symbol Description
Spare network cables
If you are going to use a network cable that is not supplied by the manufacturer,
make sure that it supports the power of the analyzer, that it is approved and
that it complies with the CE regulations.
NOTE
Operations after a repair
After a repair make sure that you have replaced all the disassembled parts
following the relevant instructions and that you have carried out the verification
tests with the service program.
NOTE
Use of consumables
Before using the consumables (cleaning solutions and system fluids) check
the expiration date. If it has expired, remove the product and use a new one
with the current expiration date.
NOTE
Operations that may compromise the safety of the equipment
Any misuse (negligence, power conditions out of tolerance, inappropriate
location or atmospheric conditions, etc.) together with internal manipulation
of the instrument by personnel not authorised by BioSystems or the use of
NOTE unoriginal consumables and spares (tubes, fuses, etc.) shall compromise
the protection assured by the analyzer.
Representations of screens shown in this manual are for illustrative purposes only. The screens do not necessarily
reflect valid data.
12
Units Definition
" Inch
˚C Degrees centigrade
A Ampere / Absorbance
GB Gigabyte
h Hour
Hz Hertz
kg Kilogram
L Liter
MB Megabyte
m Meter
min Minute
mL Milliliter
mm Millimeter
mmol Millimole
mV Millivolt
nm Nanometer
Nm Newton Meter
prep Preparation
s Second
VA Volt-ampere
V Volt
W Watt
μL Microliter
μm Micrometer
13
Service manual
The licence granted and constituted pursuant to these terms and conditions will be interpreted in accordance with
and governed by Spanish law, with the jurisdiction being the courts of the city of Barcelona, Spain. The user waives
any other applicable legislation and/or jurisdiction that is competent, if any.
The user of this licence knows and agrees that the user licence grants no rights over the use of computer program-
mes and/or applications belonging to others that may be used or which may be necessary for using or operating this
computer application, for which the user will obtain the respective legitimisation of use.
1. Intended use
• The A15/A15C analyser is used to determine analyte concentrations by in vitro biochemical, turbidimetric and
electrolyte measurements of human samples of serum, urine, plasma, cerebrospinal fluid or total blood.
The A15/A15C analyser has been optimised to operate with the BioSystems biochemistry, turbidimetric and elec-
trolyte reagent lines. Reagents not included in the A15/A15C analyser validation performed in BioSystems SA
require a complete and detailed validation by the user or the laboratory.
Users are strongly advised to validate the overall functioning of the analyser and of the reagents in the laboratory,
taking into consideration the preanalytical phase and all other relevant aspects.
The analyser is exclusively for professional use, i.e., for users who have the appropriate training and expertise
to use it. In addition to how to install the instrument, users are instructed on how to use the analyser and the
software that goes with it.
The environmental conditions for the functioning of the analyser are normal clinical analysis laboratory conditions.
These conditions are set out in the specifications chapter.
• The Y15/Y15C analyzer is used to determine analyte concentrations in wines, beverages, food and samples of
biological cultures through biochemical analysis.
The Y15/Y15C analyzer is optimized to work with the BioSystems line of FoodQuality reagents. Reagents not
included in the Y15/Y15C validation performed in BioSystems S.A., will require thorough and detailed validation
by the user or laboratory.
It is strongly recommended to validate the overall performance of the analyzer and the reagents in the laboratory
setting, taking into account the preanalytical phase and any other relevant aspect.
The analyser is exclusively for professional use, i.e., for users who have the appropriate training and expertise
to use it. In addition to how to install the instrument, users are instructed on how to use the analyser and the
software that goes with it.
The environmental conditions for the functioning of the analyser are normal analysis laboratory conditions. These
conditions are described in the specifications chapter.
14
2. Introduction
The A15 analyzer is an automatic random access analyser specially designed for performing biochemical and turbi-
dimetric clinical analyses. The instrument is controlled on-line in real time from an external dedicated PC.
A15C model has also a cooler reagents.
Y15 analyzer is an automatic random access analyzer derived from A15 is specially suitable for safe food analyses.
The Y15C has also an coller reagents.
The A15 and A15C analyzer can also be used for the measurement of stool analytes. Chapter 1.4 shows how to
modify the analyzer to determine stool analytes.
In each of the elements of the A15 analyser, BioSystems has used leading edge technology to obtain optimum ana-
lytical performance, as well as taking into account economy, robustness, easy use and maintenance. A three-axis
Cartesian operating arm prepares the reactions. Dispensing is performed by means of a pump with a ceramic piston
via a detachable thermostated needle. A washing station guarantees that the needle is kept perfectly clean through-
out the process. The reactions take place in a thermostated rotor in which absorbance readings are taken directly by
means of an integrated optical system.
This manual contains the information required for learning about, maintaining and repairing the A15 automatic analy-
zer. It should be used by the Technical Service as a learning and consultation document for the maintenance and
repair of the instrument. Chapter 2 describes the different mechanical elements that form the analyzer together with
their functionality, and chapter 3 describes the electronic system. Chapter 4 describes the Service Program. All the
adjustments and checks of the analyzer are carried out through this program, which is independent from the application
program (User Program). The separation of both programs enable it to be maintained separately and the extensions
and improvements of one do not affect the other. The user does not have the service program. The Technical Service
must install it on the user’s computer in order to carry out the service requirements. Once said tasks have been carried
out, the Technical Service must uninstall the program. Chapter 5 offers instructions for the different maintenance,
repair and cleaning operations that can be carried out by the Technical Service. The annexes contain a summary of
the technical specifications of the analyzer, the adjustment margin tables, the lists of accessories and spares, a list
of software versions and their compatibility and a software troubleshooting guide.
This is a three-axis XYZ Cartesian mechanism. The X and Y axes move the dispensing needle over the analyser
horizontally and the Z axis moves it vertically. It is operated by three step-by-step motors. In each 24-second prepara-
tion cycle, the operating arm performs the following actions: first of all, it sucks in the reagent from the corresponding
bottle. Next, the needle is washed externally in the washing station and sucks in the sample from the corresponding
tube. It is washed externally again and dispenses the sample and the reagent into the reactions rotor. Finally, it is
exhaustively washed internally and externally before proceeding with the next preparation. The arm has a system
for controlling vertical movement to detect whether or not the needle has collided into anything on descending. If a
collision occurs, as may be the case if, for example, a lid has been left on a bottle of reagent, the arm automatically
restarts, verifies the straightness of the needle and continues working issuing the corresponding alert to the user.
A vertical axis retention system prevents the needle from falling in the case of a power cut, avoiding injury from the
needle to the user or the needle being bent by an attempt to move the arm manually. The operating arm only makes
the preparations if the general cover of the analyser is closed. If the cover is raised while it is functioning, the arm
automatically aborts the task in progress and returns to its parked position to avoid injury to the user.
This system consists of a thermostated needle, supported and displaced by an operating arm and connected to a
dispensing pump. The needle is detachable to enable cleaning and replacement. The analyser has capacity level
detection to control the level of the bottles and tubes and prevent the needle from penetrating too far into the corre-
sponding liquids, thus minimising contamination. An automatic adjustment system informs the user if the needle is not
mounted or if it is too bent. The needle has a sophisticated Peltier thermostatation system, with PID control, capable
of thermostating the preparations at approximately 37º in less than 15 seconds. Dispensing is carried out by means
of a low maintenance ceramic piston pump driven by a step-by-step motor. It is capable of dispensing between 3 and
1250 ml. The exterior of the needle is kept constantly clean by a wash station included in the base. A membrane pump
transports the waste to the corresponding container.
The A15 analyser has a tray with 4 free positions for racks of reagents or samples. Each reagents rack can carry up
to 10 reagents in 20 ml or 50 ml bottles. Each samples rack can contain up to 24 tubes of samples. The samples can
be patients, calibrators or controls. The analyser can be configured to work with 13 mm or 15 mm diameter tubes of
samples with a length of up to 100 mm or with paediatric wells. Any possible configuration of racks can be mounted
from 1 rack of reagents (10 reagents) and 3 racks of samples (72 samples) to 3 racks of reagents (30 reagents) and
1 rack of samples (24 samples).
On the left of the analyser are the waste and distilled water containers. The analyser constantly controls the level of
these containers and issues the appropriate alerts if the distilled water is nearly empty or if the waste container is full.
16
The A15C/Y15C versión has a cooler. This cooler has two fixed racks. The first rack can contain up to 10 reagent
containers of 20 mL. The second one can contain 10 reagent containers of 50 mL. Each of these possitions can use
conatiners of 20 ml but not both.
The preparations are dispensed in an optical quality methacrylate reactions rotor thermostated at 37ºC. The optical
absorbance readings are taken directly on this rotor. Each reaction can be read for 10 minutes. The readings are
taken as they are programmed in each measurement procedure. The reaction wells have been designed to enable
the mixture of the sample and the reagent during the dispensing. Each rotor has 120 reaction wells. The length of the
light path is 6 mm. The minimum volume required to take the optical reading is 200 uL. The wells have a maximum
useful capacity of 800 uL. When the reactions rotor is completely full, the user must change it for one that is empty,
clean and dry. The rotor is driven by a step-by-step motor with a transmission. A Peltier system with PID control ther-
mostates the rotor at 37ºC.
17
Service manual
An optical system integrated in the rotor takes the readings directly on the reaction wells. The light source is a 10 W
halogen lamp. The detector is a silicon photodiode. The wavelength is selected by a drum with 9 positions available
for optic filters. The filters are easily changed by the user from the exterior of the analyser, without the need for disas-
sembling the filter drum. A step-by-step motor positions the drum. The optical system is capable of taking 1.25 readings
per second, with or without a filter change in between. The light beam from the lamp passes through a compensated
interferential filter to select the desired wavelength. It then passes through the rotor well and finally reaches the pho-
todiode, where the light signal is turned into an electric signal. A sophisticated analogical digital integrator-converter
system converts the electric signal into a digital value with which the analyser obtains the absorbance values. The
optical system continues to work when the general cover of the analyser is open, whereby the analyser can continue
to take readings while the user handles, for example, the sample tubes or the reagent bottles. The rotor cover must
be in placefor the optical system to work correctly.
A detector tells the analyser of the presence of the cover. The analyser aborts the readings if the user removes the
rotor cover while the optical system is taking photometric measurements. If the rotor is not covered, the analyser
informs the user so that he or she places the rotor cover when it sends samples to be analyzed.
The described elements are controlled by an electronic system based on a microprocessor. The microprocessor has
two external communication channels to connect the instrument to the computer containing the application program.
The electronic system is made up of the following independent boards:
• Microprocessor board
• Photometric system board
• Needle conditioning board
• Fluid system interconnection board
• Arm interconnection board
• Rotor interconnection board
• Power supply board
• Refrigerator control board (only A15C / Y15C models)
The application program makes it possible to control all the operations of the analyzer. From this program, the user
can monitor the state of the analyzer and the work session, program parameters, e.g. technique parameters, prepare
the work session, prepare results reports, configure different analyzer options, activate various test utilities, prepare
and maintain the instrument and carry out internal quality control processes. The purpose of this manual is not to
explain the functioning of the user program. For detailed information to this regard, please consult the User Manual
included with the analyzer.
18
2.2. Functioning of the analyser
The A15 analyser is an automatic random access analyser specially designed for performing biochemical and turbi-
dimetric clinical analyses. The analyser performs patient-by-patient analyses and enables the continual introduction
of samples. The analyser is controlled from a dedicated PC that is permanently communicated to the instrument.
The programme, installed on the computer, keeps the user constantly informed of the status of the analyser and the
progress of the analyses. As results are obtained, the computer shows them to the user immediately.
When a Work Session is begun, the analyser proposes performing the blanks, calibrators and controls programmed
for the measurement procedures it is to carry out. The user may choose between performing the blanks and the
calibrators or not. If they are not performed, the analyser uses the last available memorised data. The controls can
also be activated or not. During a session, while the analyser is working, the user can introduce new normal or urgent
samples to be analyzed. Each time a new sample is added, the analyser automatically proposes the possible new
blanks, calibrators or controls to be performed. A work session can remain open for one or more days. When a session
is closed and another new session is opened (Reset Session), the analyser again proposes performing the blanks,
calibrators and controls. It is recommended that the session is reset each working day.
The analyser determines the concentrations of the analytes based on optical absorbance measurements. To measure
the concentration of a certain analyte in a sample, the analyser uses a pipette to take a specific volume of the sample
and the corresponding reagent, quickly thermostates them in the needle itself and dispenses them into the reactions
rotor. The very dispensing speed together with the geometry of the reaction well causes the mixture to be shaken and
the chemical reaction begins. In the bireagent modes, the reaction begins when the analyser later dispenses a second
reagent in the same reaction well. The reactions can be biochemical or turbidimetric. In both cases, the reaction or
the chain of reactions produced generate substances that attenuate certain wavelengths, either by absorption or by
dispersion. Comparing the light intensity of a certain wavelength that crosses a well when there is a reaction and when
there is not a reaction can determine the concentration of the corresponding analyte. This comparison is quantified
with the physical magnitude called absorbance. In some cases, the concentration is a direct function of the absorb-
ance, and in other cases, it is a function of the variation of the absorbance over time, depending on the analysis mode.
19
Service manual
1 2 3
If you are going to use the analyzer to measure stool analytes, do not mix the different sample types in the same
analyzer.
The tip for the determination of the stool analytes is special since it ends in a sharp point to pierce
the operculum of the primary tube.
Be very careful when performing maintenance and / or repair operations that are related to the tip.
BIOHAZARD Pay attention a lot when handling or readjust the tip in not puncturing. The tip can be potentially
infectious. Wear gloves and protective clothing when handling the tip.
21
Service manual
3. Mechanical elements
• Operating arm
• X guide
• Y guide
• X carriage
• Y carriage
• Needle unit
• Dispensing system
• Thermostated probe
• Dispensing pump
• Tubes and containers
• Container level control sensors
• Racks tray with integrated washing station
• Waste pump
• Reactions rotor with integrated optical system
• Thermostated rotor and photometric system. This contains the electronic photometric system board
• Lighting system
• Electronics box.This houses the electronic boards of the microprocessor, the power supply and the front indicator
• Main cover hinges
• Base
• Housings
• Upper casing
• Front housing
• Arm casing
• Main cover
The following is a brief description of each of the mechanical elements that make up the analyzer.
This mechanism positions the dispensing needle appropriately during the preparation of the analyses. An encoder
checks the vertical movement of the needle and a spring automatically stops it from falling in the case of a power cut.
The dispensing pipe and the electrical hoses of the arm pass through the front casing
22
Reference to the Ilustration 3.1
1 – Axis control board 6 – Axis support
2 – Photodetector X position 7 – Bottom X bearing
3 – X Carriage 8 – Lower X axis
4 – X Motor
5 – Upper toothed X axis
The needle unit (5) supports the thermostated needle and can move on the Y carriage (3), which can move on the
Y axes (4). The Y axes are supported by the X carriage, which moves on the X axes (1). In this way, the needle can
be moved in the three Cartesian directions of X, Y and Z. The hose (6) houses the Teflon dispensing tube and all the
electrical hoses of the arm.
3
4
This consists of two supports (7) that hold the steel axes (1 and 2) on which the X carriage moves. The photosensor
(3) indicates the start position of the X carriage movement. The motor X (5) is moved by a rack (2). The X carriage is
supported by the second axis (2) by means of a bearing (4).
The X carriage body (1) moves along the two axes (2, 3). The upper axis (2) acts as a rack. The X motor (4) is fitted
with a pinion that moves the carriage. The X carriage also supports the interconnection PCB (7) and the Z motor (5).
To enable the movement, it uses linear bearings (8).
The body of the Y carriage (1) moves along the two axes (2) on linear bearings (9). The said axes are supported by
the X carriage. The movement is made by the Y motor (3) by the belt (4) and the return pulley (5). The start of the
movement is controlled by the tab (7) and the start photosensor (6) located on the X carriage (10). The body of the Y
carriage (1) also supports the needle unit.
23
Service manual
9 10
11
12
13
14
The Z guide (1) supports the thermostatation pipe (7) and the control PCB (8) where the heating elements are located,
together with the thermistor signal amplifier and level detection and the Z axis start photosensor. The rack (2) supports
the Z guide (1) which crosses the Y carriage (9) on two bearings. The Z motor (3) is fastened to the X carriage (10)
and is moved by a transmission axis (5) fitted with a pinion that acts on the rack. The return spring (6) acts on the
transmission axis and prevents the needle from falling in the event of a power cut: The encoder (4), which detects any
obstruction to the movement of the thermostated needle (9) is located on the same axis and on the part of the motor.
The thermostatation pipe (1) preheats the reagent during dispensing. It is fitted with two connectors at each end. The
removable needle (6) is connected to one and the Teflon dispensing pipe (3) is connected to the other, fixed by the
fastening connector (5). The PCB (2) contains the thermostatation elements, the thermistor and associated circuits.
The various thermistor and element action signals (3) pass through the hose (4).
24
15 16 17 18
19
20
21
The plastic body (1) joins the different elements that make up the pump. The transparent methacrylate fluidic cham-
ber (2) or the manifold (1&makes it possible to observe the flow of liquid through the pump. The support (4) fastens
the seal (3). The ceramic piston (5) dispenses by displacing a certain volume of liquid in the chamber. The piston is
adhered to the support (6), which moves alternatively by the rotation of the endless screw (9) fixed to the motor axle
(10). The barrier (7), joined to the piston support, obstructs the photosensor (11) when the piston reaches its start
position. The axial bearing (8) prevents any longitudinal displacement of the motor axle for greater precision in the
dispensing operation. The 3-channel electrovalve (14) makes it possible to connect the pump chamber to the distilled
water container or to the thermostated needle. The Teflon tube (13) connects the chamber to the electrovalve. It is
connected to each of these elements by the nuts (13) and (15).
25
Service manual
22
23
24
25
36
26
27
28
26
29
30
31
32
33
34
35
28
The Teflon tube (4) connects the distilled water container (1) to the electrovalve of the dispensing pump. This tube is
installed at the end of the filter container (5). It is connected to the electrovalve of the dispensing pump through the nut
(6) The Teflon pipe passes through the rubber piece (3) in the lid (2) of the container, which fastens them in position.
The PVC tube (12) connects the waste extraction membrane pump to the waste container (9). The waste container
lid (19) has a fast coupling nut (11) with automatic drip-proof closing when disconnected. All the tubes pass into the
interior of the analyzer through the rubber grommet (13).
27
Service manual
47
46
37
38
39
40
41
42
43
44
45
The analyzer has a capacitance system to control the level of the distilled water and waste containers. For this, there
is an emission plane (1) under the bottles where a signal is injected through the connector (2). The base supporting
the bottles is above this. They have 2 rods that collect the signal and indicate the presence or absence of liquid.
The plastic injection tray (1) is part of the base of the instrument. The washing station (2) is installed on the right.
The plate (3) detects the level of the dispensing needle. The PVC tube (4) connects the washing station drain to the
waste extraction pump.
28
48
49
50
51
The needle washing system has a waste extraction pump (1). This is connected to the washing station by the PVC (2).
The pump expels the waste through the pipe (4) into the waste bottle. The pipes are fastened by two safety flanges.
The dispensing system dispenses the reagents and the samples in the methacrylate rotor (1). The optical system
measures the absorbance directly on the rotor wells. The aluminium heating canal (2) surrounds the rotor and keeps it
at 37ºC. The canal is thermally insulated from the exterior by means of the moulded expanded polystyrene insulation
(3). The Peltier cells (4), with their respective radiators (5) and fans, act on the canal to control the temperature. The
sensor used to control the temperature is the probe (7). The methacrylate rotor is fastened to its centring unit (8) by
means of the screw (9). The centring unit is fixed to the heating canal through the axis (22), which is fitted on bearings
(11). The barrier obstructing the photosensor (10) when the rotor reaches its start position forms part of the centring
unit (8). The centring unit also acts as gearing. The pinion (12), fixed to the motor (13), acts through the crown (14),
which also acts as a centring unit. The separator (15) does not allow the motor temperature to reach the heating canal.
The electronic board of the photometric system (16) is housed in a cavity in the heating canal. The upper cover of this
cavity (17) supports the electronic board. The seal (18) keeps the cavity hermetically closed in the case of possible
liquid spillage. The housing of the filter drum is closed at the bottom by the cover (18). The part (19) centres the pho-
todiode with regard to the lighting system and also acts as a grill to prevent the incidence of unwanted light. The grill
(20) limits the light hitting the reactions rotor. The detector (21) tells the analyzer if the rotor cover is in position or not.
29
Service manual
The aluminium body (1) is the structure that supports all the elements of the lighting system. The lamp holder (2),
fastened to the body by means of the fastening system (4), keeps the halogen lamp (3) in position without the need
for adjustments. The filter drum (5) has 10 positions for optical filters. Position 0 must always be taken up by a cove-
red filter. The other positions can be taken up by an interferential filter (8) or by other covered filters. No position in
the drum must be left unoccupied. Each filter is fitted on a filter holder (6) and fastened to it by the nut (7). The filter
holders can be dismounted from the drum by simply pulling on them. The cover (13) allows easy access to the filter
drum. The filter drum is fastened to the axle (9). This axle can be turned by the direct action of the motor (11). Its end
is guided by the bearing (14). The photosensor (10) indicates the start position of the drum. The light from the lamp,
limited by the diaphragm (12). The light passes through the filter drum, which selects the desired wavelength, and
through the aperture(15), which adapt the form of the light beam to the geometry of the rotor wells.
1
9
2
3
12
13
11
8
22
20 19 16 17
30
3
18
13
15
10
14
21
7
31
Service manual
32 2 24 25 30 23
37
36
35
31
34
26
29
27
28
The base (2) on which all the elements of the analyser are fixed is fastened directly to the lower plastic casing. The
rack tray and washing station form part of the base. The instrument stands on 4 rubber legs . The front right leg (13)
is adjustable in height to adapt the instrument to the work surface.
The front casing (1) is fastened to the upper casing (2) and the upper casing is fastened to the lower casing (4). The
32
top cover (3) is transparent and lets users see the analyser in operation with the cover closed.
The metal cover (1) supports the mains switch (2) and the fuse holders (3), as well as the identification label (4). The
COM1 and COM2 connectors (6, 7) and the mains connector (5) are fastened to the electronics box. The cover(1)
opens on 2 hinges (7).
The two hinges enabling the raising of the main cover of the analyzer consist of an articulated steel structure (2)
operated by a hydro-pneumatic spring (1). The right-hand hinge includes a photosensor (3) to detect whether or not
the cover of the analyzer is open or closed.
5
6
7
8
9
10
33
Service manual
11
12
13
14
15 16 20 21
22
17
18
19
34
Referencias de la Ilustration 3.12, Ilustration 3.13 y Ilustration 3.14
1 – Top cover 12 – Fridge switch
2 – Upper housing 13 – Fuses
3 – Arm housing 14 – Plug
4 – Front arm protector 15 – Main connection
5 – Front housing 16 – Auxiliary connection
6 – Reaction rotor cover 17 – Cover detector plate
7 – Fridge cover 18 – Cover Detector
8 – Bottom housing 19 – Hydro-pneumatic piston
9 – Fridge housing 20 – Back plate
10 – Leg 21 – Identification label
11 – Analyzer switch 22 – Top cap support structure
The plate (12) is cooled by the peltier cells (6). These are refrigerated by the heatsinks (5) and fans (4). A temperatu-
re sensor(10) is connected to electronics module that controls the temperature. The isolators (9) in the housing (2)
avoid that the flux heat enters inside of cooler. A lid sensor (8) informs to analizer if lthe lid is seated. The lid sensor
is serial connected with the general lid of analizer.
12
9
10
35
Service manual
4. Electronic system
The electronics of the analyzer are made up of different boards located at different points in the analyzer and dedica-
ted to specific functions. Its different location corresponds to functionality and performance criteria for the functioning
of the analyzer.
This is the brain of the machine, containing the microprocessor (H8/3003), responsible for controlling all the elements
of the machine. The board has different data storage systems using either static RAM (U1 and U47), FLASH memory
(U10) or EPROM (U9). The slot associated with the EPROM is used to check the functionality of the board and the
recording of the MONITOR program in the production phases of the analyzer. The other two memories are associated
with the normal functioning of the analyzer. The FLASH memory holds the application itself as well as different data-
bases related to factory settings, adjustments, state of the rotor and possible extensions to the application.
The U21 device also exists on the board. This is a logical programmable device (FPGA) dedicated to the control of
motors, mapped in register memory associated with end-of-run control, electrovalves, level sensing and control of the
photometry-associated board (CIIM00027).
The motor control acts directly on the drivers corresponding to each of the analyzer’s axes (U28,U29,U30,U24,U25,U27)
to act on the motor. The driver comprises the L6228 integrated circuit. The regulation of the current of each axis can
be configured by means of a DAC that sets the current set point independently (U26).
The action on the thermostatation systems of the rotor is carried out through H-shaped bridges based on MOS te-
chnology (U45) and controlled directly from the microprocessor. The action on the needle thermostatation system is
through the Q4 transistor.
J1 Not available
36
Connector Function Pins
J6 Connection to interconnection 1 - 12 V
board 2 - GND
Rotor (home motor signals and 3 - DVALID
photometry board control signals) 4 - DCLK
5 - DOUT
6 - DXMIT
7 - RANGE2
8 - RANGE1
9 - RANGE0
10 - TEST
11 - CONV
12 - GND
13 - CLKAD
14 - GND
15 - GND
16 - V DC
17 - V DC
18 - Rotor cover
19 - GND
20 - Rotor thermistor
21 - Home motor filter drum
22 - GND
23 - Home motor rotor
24 - Front red LED
25 - Front green LED
37
Service manual
Analogical circuitry:
The waste and system liquid sensors function through U6, U5 and U4, which generate and detect the signal responsible
for detecting the waste and system liquid. These signals are sent and received through the fluid interconnection board
(connected to the CPU board by J9). The rack level detection is carried out in a similar way through U7, U8 and U2.
The signal injected to the base of the bottles goes to the fluid interconnection board through J9 and is collected after it
has been amplified by J10 (connection with the needle board). There is also a circuit for conditioning the signal of the
thermistor associated with the thermostatation of the rotor that is made up of the U1 and U2 circuits. The thermistor
is connected to the rotor interconnection board, which is connected to the CPU board.
TP1 - Waste pump control signal
TP2 - Electrovalve control signal
TP3 - Rotor thermistor signal
TP4 - RESET
TP5 - WATCHDOG
TP6 - LSO_BOT bottle detection signal
TP7 - Bottle signal
TP8 - Needle detection signal
TP9 - LSO needle detection signal
TP10 - Attenuated LSO needle detection signal
TP11 - IN1 Needle Peltier Driver
TP12 - Needle resistance driver
TP19 - ASL
TP20 - HWR_L
TP21 - LWR_L
TP22 - WE_L
TP24 - CS_FPGA_L
38
TP25 - DVALID (photometry)
TP26 - 12 Volts analogical
TP27 - IN2 Needle Peltier Driver
TP28 - EF Needle Peltier Driver
TP30 - DOUT (photometry)
TP33 - RANGE (photometry)
TP34 - CLKAD (photometry)
TP35 - Conditioned thermistor signal
TP38 - DXMIT (photometry)
TP39 - Analogical GND
TP40 - Power GND
TP41 - Power GND
TP42 - Power GND
TP43 - Digital GND
TP44 - Digital GND
TP45 - Digital GND
39
Service manual
40
4.3. Power Supply Board (CIIM00015)
This is made up of 2 different switched regulators and 1 voltage line that enable distribution of the power supply in
accordance with the requirement of each subsystem.
J1 24 V input 1 - 24V
3 - (GND)
D4 - Indicates 5V activated
D2 - Indicates 12V lamp activated
D3 - Indicates 12V analogicals activated
the level detection signal and the Z home. It receives, from the needle unit, the thermostatation elements, the ther-
mistor and the level signal detected by the needle itself.
The cables that join this board with the CIIM00026-01 board come from this needle
42
Connector Function Pins
J1 Connection motor X
J2 Connection motor Y
43
Service manual
J3 Connection motor Z
This enables communication with the exterior of the analyzer through a USB channel or a RS232 channel. It also
includes an auxiliary RS232 channel for monitoring the functions of the analyzer during its execution.
44
CN1 - USB Connector
P1 - Main RS232 connector
P2 - Auxiliary RS232 connector
45
Service manual
46
Connector Function Pins
DL1 - Peltier
DL2 - Home rotor motor
DL3 - Home filter motor
DL4 - Rotor cover
DL1 - Electrovalve
DL2 - Waste pump
DL3 - Home pump
DL4 - Instrument cover
47
Service manual
48
4.10. Refrigerator cooler board (CIIM00075)
1- 5V5
CN1 POWER SUPPLY
2- GND
1 - 5V
CN2 PUMPS- J11
2 - GND
1 - 5V
2 - GND
CN3 PUMPS- J13
3 - GND
4 - Earth
1 - Temp
CN4 FRIDGE THERMISTOR
2 - GND
1 - 5V5
2 - Peltier 1
CN5 PELTIERS CELLS
3 - 5V5
4 - Peltier 2
1 - 5V5
2 - Fan1
3 - 5V5
4 - Fan 2
CN6 PELTIER FANS
5 - 5V5
6 - Fan 3
7 - 5V5
8 - Fan 4
1 - Green
CN7 STATUS LED 2 - GND
3 - RED
49
Service manual
Adjustments loaded:
50
00000000 00000000 00000000 00000000 Hello World“
F[04]: 00000000 00000000 00000000 00000000 00000000 00000000
00000000 00000000 00000000 00000000
F[05]: 00000000 00000000 00000000 00000000 00000000 00000000 A15 MAGIC KEYS“
00000000 00000000 00000000 00000000
F[06]: 00000000 00000000 00000000 00000000 00000000 00000000 H: Help“
00000000 00000000 00000000 00000000 R: Rotor Temperature“
F[07]: 00000000 00000000 00000000 00000000 00000000 00000000 P: Probe Temperature
00000000 00000000 00000000 00000000 S: Level Scales
F[08]: 00000000 00000000 00000000 00000000 00000000 00000000 A: Last A15 Stress Results
00000000 00000000 00000000 00000000 L: Actual Sensitivity of Level Detection
F[09]: 00000000 00000000 00000000 00000000 00000000 00000000 N: Enable Level Detection Debug
00000000 00000000 00000000 00000000 K: Power Supply On
F[10]: 00000000 00000000 00000000 00000000 00000000 00000000 Buzzer Control
00000000 00000000 00000000 00000000 B: Buzzer On
b: Buzzer Off
F[01]: 000 000 000 000 000 000 000 000 000 000 Encoder
F[02]: 000 000 000 000 000 000 000 000 000 000 E: Generate Encoder Error
F[03]: 000 000 000 000 000 000 000 000 000 000 I: Enable Encoder IRQ
F[04]: 000 000 000 000 000 000 000 000 000 000 Rotor Reading
F[05]: 000 000 000 000 000 000 000 000 000 000 1: Choose Filter +
F[06]: 000 000 000 000 000 000 000 000 000 000 2: Choose Filter -
F[07]: 000 000 000 000 000 000 000 000 000 000 9: Start Rotor Readings
F[08]: 000 000 000 000 000 000 000 000 000 000 Notes: Use only in Service Mode
F[09]: 000 000 000 000 000 000 000 000 000 000 after a Base Line Test.“
F[10]: 000 000 000 000 000 000 000 000 000 000 Rotor Read
1: Choose Filter +
Zmax Reference=1130 2: Choose Filter -
- Pediatric Offset=0 User Mode Test
- 13mm Offset=0 G: Test
- 15mm Offset=0 Notes: Use only after a Worklist in
- Reagent Offset=0 Stand By.“
- Central Reagent Offset=0 This tests dumps all the preparations“
parameters received and the photometric“
A15 Mechanical History readings. Finally performs a general test“
of the analizer.“
- X axis: 0 Steps After this test press New Rotor
- Y axis: 0 Steps for continue working.“
- Z axis: 0 Steps DDC112/Photometry
- Rotor: 0 Steps D: Choose Mode
- Filter Wheel: 0 Steps - DDC112 internal test mode
- Ceramic Pump: 0 Steps - DDC112 Photometry Mode
- Washing Station Pump: 0 Cycles - Stop
- Washing Station Valve: 0 Cycles +: Integration Time +0.5ms
- Ceramic Pump Valve: 0 Cycles -: Integration Time -0.5ms
- Lamp: 0 Minutes Notes: Only works in Service Mode
This tests performs continuous“
A15 Statistics readings with the DDC112.“
Remember stop the test for “
- Biochemistry Tests: 0 continue working.“
- Turbidimetry Tests: 0
- Biochemistry Bireagent Tests: 0 Caution: Dont’t abuse of this functions while the
- Turbidimetry Bireagent Tests: 0 analizer is running.“
- Predilutions: 0
- Initial/Final Washings: 0 <>
- Washing Solution Washings: 0 Rx
- System Liquid Washings: 0 Stand by mode!
- New Rotor: 0 LC
- Bireagent Contaminations Solved: 0 TxS
Rx
Inicio modo servicio!
Tray Ref. X=675 => Distance from tray reference to tray corner X=2190 Programming FPGA XC2S50PQ208
Tray Ref. Y=10 => Distance from tray reference to tray corner Y=30 - Clearing FPGA program memory: OK!
- Programing FPGA: OK
Absolute position of tray corner X=2865 FPGA XCS50PQ208 is programmed
Absolute position of tray corner Y=-20
Generating Zmax Map:Ok
“ Initialization of level detection system
CPU settings: MDCR=c4;ABWCR=0;ASTCR=ff“ - Generating Sensitivity Map:Ok
Level Detection Mode:Normal
BioSystems A15“
51
Service manual
Pediatrico:
Rack 1 2 3 4 Probe AA
173 141 129 126 123 125 129 151 - X:X=Xcal+-1“
161 129 118 115 113 115 125 141 - Y:Y=Ycal+0“
158 128 114 111 108 111 121 138 AAEnd
156 127 113 108 107 110 118 139 Peltier Cells and Drivers Test
156 127 112 108 107 110 117 145 - Probe Peltier Driver Test: Ok! (Not implemented)
155 126 112 108 106 110 121 146 - Rotor Peltiers Driver Test: Ok!
161 128 113 110 106 112 121 151
160 132 112 111 106 113 121 151 Hardware Initializated
167 133 115 112 108 114 124 156
171 136 117 116 113 119 129 163 There are a few keys that work with the Hyperterminal,
177 145 123 122 116 124 136 172
202 160 139 135 133 138 155 190
to press some keys the analyser give information about
some element, the following keys has the function:
Tubo 13/15mm:
Rack 1 2 3 4 H: Help, help, shows the help text
162 141 126 123 119 123 132 147
149 132 115 115 111 115 121 137
R: Rotor Temperature, shows the rotor temperature
145 126 110 111 107 112 115 136 P: Probe Temperature, shows the needle temperatare
148 127 111 108 106 110 117 136 S: Level Scales, shows the scales mesures in %
148 127 112 111 105 110 118 144 A: Last A25 Stress Results
149 128 111 111 107 113 119 145
154 126 113 110 108 111 119 144
L: Actual Sensibility of Level Detection
155 128 115 111 110 113 124 145 N: Activate additional information of level detection (only
161 130 114 112 110 115 125 148 internal use)
161 133 117 112 112 116 125 154 K: Deactivate the power supply
169 140 123 121 118 122 133 161
190 155 134 133 128 134 146 176
B: Activate the buzzer
b: Deactivate the buzzer
Reactivos: E: Generate an encoder error (only internal use)
Rack 1 2 3 4 I: Activate the encoder interrupt (only internal use)
163 151 152 183
156 141 143 174
152 138 140 173 1: Increase the filter wheel position
150 134 137 172 2: Decrease the filter wheel position
154 135 138 176 9: Mesure the whole rotor, step by step
156 139 143 181
160 138 145 187
166 145 149 194 G: Once finish a work list, push the G and send to the
173 150 156 204 hyperteminal more detailed information of the work list
194 172 173 228 managemnet
Generating Pattern: 1 1 1 1 1 1 1 1 1 1 1 0 0 0 0 0 0 0 0 0
Generating Gate: 0 0 0 0 1 1 1 1 0 0 0 0 0 0 0 1 1 0 0 0
OK
Z Axis Initialization
- Axis Z in HOME
Z Axis initializated
Penc:111111111000000000| OK
Z Axis Initialization
- Axis Z in HOME
Z Axis initializated
52
4.12. Interconnection between boards
The following diagrams show the connections between the boards and the different elements that make up the analyzer.
53
Service manual
54
55
Service manual
56
57
Service manual
58
59
Service manual
60
5. Service program
The service program is used for the adjustment, checking and maintenance of the different components of the
analyzer. It is not supplied with the instrument, it is supplied to authorised technical services only. The personal
of the Technical Service must install it on the user’s computer in order to carry out the service requirements.
Once the tasks have finalised, the program must be uninstalled. To install the program, follow the instructions
on the installation CD ROM called Service. The original password for using this program is A15. The password
can be changed from the service program itself. If the service personnel forget the password, the original pas-
sword can be reinstalled by deleting the hidden file code.A15 from the application directory and relaunching
the program. Once the password has been introduced, the analyzer serial number is given and the name of
the operator is requested (by default Operator1). Press the Accept button and the main program appears. The
different functions of the service program are classified in the following categories:
• Adjustments: These make it possible to make different parameter adjustments required for the correct
functioning of the analyzer.
• Tests: Tests for checking the functionality of the analyzer.
• Utilities: Different technical utilities, such as, for example, washing or priming the dispensing system
or changing an optical filter.
• Registers:This enables the management of past adjustments, tests, incidences, repairs and maintenance
of the instrument.
• Monitor:These enable the low level communication with the analyzer to load new versions of the program
in the flash memory of the analyzer (firmware) or to consult the internal parameters of the instrument.
An emergency stop button (STOP on a red background) will be accessible at all times, and when pressed, it
switches off the analyzer and closes the application quickly.
61
Service manual
62
In order to get a printed copy of this initialization report, you should press the Print button.
NOTE: If an error has been reported and the technician continues working with the service program, he must
consider that there is a hardware element that is not working properly.
5.2. Adjustments
These make it possible to make different parameter adjustments required for the correct functioning of the analy-
zer. All the values to be adjusted have certain limited ranges, indicated by the service program. These values are
also given in an appendix at the back of this manual. If, after varying any of the parameters within its permitted
range, the analyzer is not tuned up, it indicates that the corresponding system is broken and in need of repair.
63
Service manual
This screen adjusts the needle thermostatation in such a way that the dispensing temperature of the reactions
is as close as possible to 37ºC. To make this adjustment, the analyzer must be initialised. The liquid to be dis-
pensed is taken from the system liquid container or from the bottle of reagent selected by the technician. The
technician must measure the temperature of the dispensed liquid with a thermometer calibrated at 37ºC. The
program shows the control set point temperature, which is the parameter that must be adjusted for the dispen-
sing temperature to be correct. This parameter must be different from 37ºC. When the technician so indicates,
the analyzer dispenses thermostated distilled water on a certain position in the racks tray shown on the screen.
The technician must measure the temperature of the water with the calibrated thermometer and introduce the
temperature on the screen. The analyzer automatically modifies the set point temperature in accord with the
temperature measured with the thermometer for the dispensing temperature to be 37ºC. The technician can
modify this set point temperature proposed by the program. On pressing Adjust, the analyzer thermostates
the needle with the new set point and, when the technician so requests, performs new dispensing operations.
Each time the set point temperature is modified, wait 1 minute before performing new dispensing operations
for the needle temperature to become stabilised. The technician must repeat this process until the dispensing
temperature is as near as possible to 37ºC. Pressing the Store button, the analyzer stores the current value
of the adjusted set point temperature. Pressing the Cancel button keeps the last stored value and the current
value is not stored. Pressing the Restore button restores the initial screen input value.
This screen makes it possible to adjust the thermostation system of the rotor in such a way that the reactions
temperature is 37ºC. To make this adjustment, place a well rotor in position and ensure that the analyzer has
been initialised. The rotor can be automatically filled with distilled water by pressing the corresponding button.
Once filled, the technician must wait a few minutes for the rotor to be thermostated. The temperature in the
rotor wells must be measured with a temperature calibrated at 37ºC through the dispensing hole of the rotor
cover. A button makes it possible to turn the rotor in increases of 15 wells to change the well on which the
measurement is being taken. The program shows the control set point temperature, which is the parameter that
must be adjusted for the temperature of the rotor to be correct. This parameter must be other than 37ºC. The
technician must measure the temperature of the water with the calibrated thermometer in the wells and enter
the temperature on the screen. The analyzer automatically modifies the set point temperature in accord with
the temperature measured with the thermometer for the rotor reactions temperature to be 37ºC. The technician
can modify this set point temperature proposed by the program. On pressing Adjust, the analyzer thermostates
the rotor with the new set point. Each time the set point temperature is modified, wait 5 minutes before perfor-
ming new dispensing operations for the rotor temperature to become stabilised. The technician must repeat
this process until the rotor temperature is as near as possible to 37ºC. Pressing the Store button, the analyzer
64
stores the current value of the adjusted set point temperature. Pressing the Cancel button keeps the last stored
value and the current value is not stored. Pressing the Restore button restores the initial screen input value.
This screen makes it possible to adjust the horizontal positioning (X, Y) of the arm. The arm housing must be
removed to see the position of the needle. Before making the adjustments, visually check the verticality of the
needle. If necessary, carefully straighten it up ensuring you do not damage it. On the screen, select the point
at which you wish to adjust the horizontal positioning. On pressing the Adjust button the arm initialises and
positions itself over said point. The technician has buttons to move the arm step by step over the horizontal
plain (X, Y) and vertically (Z). The arm can also be moved introducing a certain number of absolute movement
steps. These absolute movements of the arm must be made with the needle at its highest position so as not to
damage it (coordinate 0). The technician must lower the needle to the adjustment point and adjust its horizontal
position. When the position is satisfactory, save the current coordinates (X, Y) by pressing the Store button.
Pressing the Cancel button keeps the last adjustment values stored. Pressing the Restore button restores the
initial screen input values. At all times, the screen shows the current coordinates of the arm for the selected
point, the last coordinates stored and the initial screen input coordinates, as additional information for the te-
chnician. The technician may repeat the procedure to adjust the positioning of the arm at the different possible
adjustment points. These points are as follows:
Note: the pediatric rack is a 15 diameter rack with pediatric adapters and wells
This process is used to adjust each rack individually in the three coordinates, X, Y and Z as much for pediatric
65
Service manual
as for reagent racks. For that, each rack of the tray will be adjusted to positions 1 and 12 for pediatric racks
and 1 and 10 for the reagent ones.
There are two tools to carry out the adjustment process: one to adjust the reagent rack and the other for the
pediatric rack.
In order to carry out this XYZ rack adjustment, you may use the screen or keyboard buttons. Movements using
keyboard:
• X axis movements: right and left cursor button
• Y axis movements: up and down cursor button
• Z axis movements: page up and page down button
66
Adjustment process
1. Select the type of adjustment which you wish to carry out: adjustment of the tray (XYZ) of the pediatric
diameter rack or of the reagent rack.
2. Press the button Adjust.
3. Deactivate the option Adjust with the tool
4. Select the number of rack to start the adjustment process. By default, it starts with number 1.
5. Place the rack in the selected position, place the well or bottle in well number 1 of the rack and press Start.
6. Probe will automatically descend to a distance over the well/bottle, in order to avoid colliding with it. Then,
probe has to be put into the well/bottle just a little. In order to do so:
- If probe is not centered in XY and out of the well or bottle opening: move the probe just by XY movements
(screen or keyboard) before carrying out Z movements to get the probe down.
- Then, get the probe slightly down with Z movements (screen or keyboard) just to make the adjustment
of the center easier.
From this position of approach, you may start the adjustment of values:
7. Adjust the well/bottle center: to do so, move the probe by X and Y movements (screen or keyboard)
8. Adjust the bottle or well bottom: to do so, move the probe by Z movements (screen or keyboard) until it
reaches the bottom. To check it: move the bottle/well up and down.
9. To finish the XYZ adjustment, press Accept.
10. The new adjusted XYZ values will appear in the upper table. (Nevertheless, they are not saved in the
instrument yet).
11. The arm will automatically move to position 10 or 12, depending on the type of rack selected. Repeat steps
6 to 9.
12. Once the adjustments in position 10 or 12 are carried out, the arm is parked so the rack can be moved.
13. Place the rack in the following position of the tray, press Start. Repeat steps 6 to 9 to carry out the ad-
justments in the every position of the rack tray.
14. Once the adjustments are finished, adjustments have to be kept in the instrument, so press Save.
67
Service manual
The adjust is performed in the same way of an instrument without reagent cooler. Keep in mind that the adjust
for sample racks (pediatics or tubes ) only will perform in the 3 and 4 positions:
The reagent cooler will only fit in positions 1 and 2 and adjusts similarly as a reagent rack. To do this select
XYZ Tray-fridge.
When the adjustment of the tray of pediatric racks is selected, it appears another adjustment: the Z relation
between pediatric and tube.
68
In order to carry out this adjustment, follow the following steps:
1. Place a diameter 15 rack in position 2 of the tray, with a tube in rack position 1.
2. Insert a value in the box of Z pediatric-tube relation. This value shows the separation steps between a
pediatric well and a primary tube.
3. Press Start.
4. Check that the probe has not collided to the bottom of the tube.
5. Move the probe by Z movements (screen or keyboard) until it reaches the well bottom.
6. Once the adjustments are finished, adjustments have to be kept in the instrument, so press Save.
When you want to use the stool tubes, it is necessary to carry out the following adjustment, once the tip model
has been changed:
1. Place in the position 2 of tray a 15 diameter rack with a stool tube in position 1.
2. Enter a value in the Z pediatric-stool tube relation box. This value indicates the separation steps between
a pediatric well and the tube. Enter a value so that the tip touches the inner liquid of the stool tube.
This ratio of Z-Pediatric to Z-stool tube is negative. Unlike the Z-Pediatric ratio with 15mm/13mm tube
which is positive
3. Press Save.
NOTE: If the instrument comes from a previous Firmware and has been updated by the technical service to
the version that incorporates the stool tubes (v6.16 or higher), the ratio Z pediatric-stool tube will be 0. This
means that in the positioning to make the adjustment the tip will lower to the position of Z for pediatric well. It
is recommended not to use a stool tube in that first case since the tip can get too low and collide. Using a tube
or nothing in that position, perform a pre-adjustment leaving the offset at -500.
Proceed then to make the adjustment in the indicated way.
This screen enables the adjustment of the positioning of the rotor with regard to the dispensing point and the
optical system. One or the other is selected by means of two different tabs.
69
Service manual
Put the tool AC15920 in the first well of the rotor. Ensure that the component is positioned until the bottom of
the well. The point of dispensation of the rotor is indicated with the hole of the tool.
The analyzer initialises the rotor and positions the first rotor well at the currently programd dispensing position.
The technician has buttons to move the rotor step by step to adjust, if necessary, this position and buttons for
finer adjustment of the X coordinate over the dispensing point. At all times, the screen shows the current dis-
pensing coordinate on the first well and of the X axis position, the last coordinate stored and the initial screen
input coordinate, as additional information for the technician. When this is satisfactory, the current coordinate
of the dispensing point of the first well can be stored by pressing the Store button. Pressing the Cancel button
keeps the last stored value and the current value is not stored. Pressing the Restore button restores the initial
screen input value.
70
5.2.4.2. Centering of the rotor with regard to the optical system
This adjustment is necessary only if the Rotor Centering Adjustment has been carried out with regard to the
dispensing point (4.1.4.1.). This adjustment must be made with the rotor cover in position. The analyzer initia-
lises the rotor and fills the first 3 wells of the rotor with distilled water. Next, step-by-step optical readings are
made through these wells at the wavelength selected by the technician. Once the readings have ended, the
program shows a graph of the light intensity measured on the rotor steps. On this graph, the program indica-
tes at which points the optical readings are made on each of the 3 wells when the analysis is made, with the
coordinate of the reading point of the first well currently programd in the analyzer. If necessary, the technician
can move the reading points over the graph jointly using two buttons. The optimum reading point is that which
globally maximises the light intensity for the three wells. At all times, the screen shows the current coordinate
of the reading in the first well and the last coordinate stored, as additional information for the technician. When
the position is satisfactory, the current coordinate of the reading point of the first well can be stored by pressing
the Store button. Pressing the Cancel button keeps the last stored value and the current value is not stored.
This adjustment must be made with the rotor cover in position. The analyzer initialises the rotor and the filter
wheel and fills the first rotor well with distilled water. Next, it takes optical readings through this well, turning the
filter wheel step by step, with a certain integration time as indicated by the technician (the concept of integration
time is explained in the section on photometric adjustments). Once the readings have ended, the program shows
a graph of the light intensity measured on the steps of the filter wheel. On this graph, the program indicates at
which points each of the filters is positioned when optical readings are taken when the analysis is carried out,
with the coordinate of the positioning of the filter 0 currently programd in the analyzer. If necessary, the technician
can move the reading points over the graph jointly using two buttons. The optimum reading point is that which
globally maximises the light intensity for all the filters. At all times, the screen shows the current coordinate
of the filter 0 and the last coordinate stored, as additional information for the technician. When the position is
satisfactory, the current coordinate of the positioning of the filter 0 can be stored by pressing the Store button.
Pressing the Cancel button keeps the last stored value and the current value is not stored.
71
Service manual
This screen makes it possible to set the level control scales with the empty waste and distilled water containers
(0% capacity) and when they are full (100% capacity). The maximum capacity of the containers is approximately
3L. The technician must choose whether he wishes to set the distilled water or waste container scales, with the
corresponding container full or empty. According to the requested adjustment, the corresponding container, full
or empty, must be placed in position and the Adjust button pressed. Based on the settings made, the analyzer
automatically adjusts the scales. On pressing the Store button, the analyzer saves the new values of the adjus-
ted parameters. Pressing the Cancel button keeps the last stored values and the current values are not stored.
72
5.2.7. Adjustment of the level detection sensitivity
This screen allows fitting the sensitivity of the capacity level detection system of the probe. In order to make
the adjustment, first of all you have to select the typs of racks: metal filled racks (grey color) or plastic racks
(black color).
Then select the type: pediatric, diameter 13 / 15mm, reagent or stool collector
In the rack rack position 2 (starting from the left) the selected rack must be placed with the following configu-
ration depending on the type of rack:
• Paediatric Rack: place 4 paediatric glasses in positions 6, 7, 18 and 19 with 150 uL of liquid system.
• Rack of 13/15mm sample: place 4 tubes in positions 6, 7, 18 and 19 with 1000 uL of liquid system.
• Rack of reagents: place 2 bottles of 20mL in positions 5 and 6 with 2mL of liquid system.
It is possible to see a graphic with the position of the tubes and racks in the screen photo. When pressing the
Adjust button, the arm takes some sensitivity readings automatically until detecting the water in each one of
the tubes. Once this operation is finished, the arm is parked in its original position and it shows the sensitivity
results of each one of the tubes. Move rack to position 5 and repeat the adjustment. Once these second rea-
dings are finished, the program calculates the average of all sensitivities; the result is the average sensitivity.
The technician should notice that the sensitivity values of each tube have to be similar; otherwise, he should
repeat the whole measurement.
Repeat this adjustment per each rack type.
The sensitivity value can also be introduced manually for each rack in the corresponding box.
Pressing the button Save, the analyser saves the new adjusted sensitivity value. Pressing the button Close,
the old value stays. Pressing the button Restore, the initial value of entry to the screen is restored. A manual
sensitivity value can be inserted in the corresponding box.
In instruments with reagent cooler the adjust of pediatrics cups will made in the position 3, as shown in the
figure. To adjust the level sensibility of reagent racks outside the cooler proceed in the same way.
73
Service manual
To adjust the level sensitivity of the reagent racks of cooler place 2 containers of 20 mL with 2 mL of system
liquid in the positions 5 and 6.
With the new tip to determine stool analytes readjust the sensitivity of the level detection.
• Stool collector rack: manually enter the sensitivity value to 255
74
5.2.10. Select the model of the instrumento
To select the model of the instrument go to the main menu option Adjusts/Enable fridge. A drop down menu will
shown. Select the instrument model:
• Analizer with or without fridge.
5.3. Tests
Various tests make it possible to check that the different components of the analyzer function correctly.
Through these tests, the technician can check the correct functioning of all the analyzer motors step by step.
The screen makes it possible to choose the motor to be tested and the test that is to be carried out. The analyzer
uses the following motors step by step:
• X axis of the operating arm.
75
Service manual
This test displaces any of the mobile components to the desired point along its range of functioning, introducing
the corresponding absolute coordinate or moving it step by step. The speed and acceleration of the movement
are those used in the normal functioning of the analyzer.
76
5.3.1.3. Loss step test
This test makes it possible to check if a motor misses steps when performing a certain sequence of movements.
The test can be carried out with the speed and acceleration used in the normal functioning of the analyzer or
with these magnitudes increased by 10% to check the functioning safety margin.
This test makes it possible for a certain sequence of movements to be performed continually. The technician
can program the duration of the test, which can be cancelled at any moment. Depending on the motor selected,
there is a minimum stress mode time (but in no case is it higher than 50 seconds).
77
Service manual
The Z axis of the operating arm has an encoder to detect if there have been missed steps as a result of a co-
llision with the needle. In the case of a power failure, a mechanical system automatically raises the needle. On
selecting the corresponding options, the analyzer checks the functioning of each of these devices.
This test checks that the needle does not collide with the bottles on the rack tray. Select the rack type (reagent,
paediatric, 30 mm or 15 mm, stool tube), the position of the rack on the tray and the position of the bottle or
well on the rack.
Press the Start button to move to the selected position and check if the needle collides with the bottle or well
or if there is space between the needle and the bottle.
Repeat the process in the positions required by the user.
78
5.3.2. Diaphragm pumps and electrovalves test
The analyzer uses a 3-way electrovalve to manage the dispensing operations. The washing system of the needle
uses a 2-way electrovalve and two diaphragm pumps. The screen makes it possible to choose the device to be
tested and the test that is to be carried out. The devices that can be tested independently are:
• 3-channel electrovalve of the dispensing pump.
• 2-channel electrovalve of the washing system.
• Washing system diaphragm pumps
To carry out these tests, the dispensing system should be primed. The following is a description of the different
tests that can be performed.
This test makes it possible for a certain sequence of device switching to be performed continually. The technician
can program the duration of the test, which can be cancelled at any moment.
This test makes it possible to check the functioning of the needle self-centering system. During its initialisation,
the analyzer uses this system to check the presence of the needle and its verticality and automatically correct
small deviations. The test consists of simply running this process. The technician can remove the housing of
the arm to observe the test. On the finalisation of the test, the program shows the deviation (x, y) found in the
motor steps.
79
Service manual
This test checks the functioning of the system for detecting the capacitive needle in bottles of reagent, pediatric
well, sample tubes and stool tubes.
This test checks the functioning of the system for detecting the capacity of the needle in reagent bottles and
sample wells. The test can be performed in any position on the tray.
First select the rack type, then the position of the rack on the tray and, finally, the position of the bottle/well on
the rack. Press the Test button and the program will move the arm to the indicated position and check whether
or not liquid is detected, depending on whether the bottle is full or empty.
Repeat the test as many times as the user considers necessary.
80
5.3.5. Needle thermostatation system test
This screen makes it possible to check that the dispensing temperature of the reactions is around 37ºC. To make
this adjustment, the analyzer must be initialised. The technician must measure the temperature of the dispen-
sed liquid with a thermometer calibrated at 37ºC. The program shows the set point temperature of the current
control. This parameter must be different from 37ºC. When the technician so indicates, the analyzer dispenses
thermostated distilled water on a certain position in the racks tray shown on the screen. The technician must
measure the temperature of the water with the calibrated thermometer and introduce the temperature on the
screen. The program indicates if the temperature measured is within the tolerated error margins and stores this
value for the test result reports. The liquid to be dispensed is taken from the system liquid container or from the
bottle of reagent selected by the technician.
81
Service manual
This screen makes it possible to check that the temperature of the rotor reactions is 37ºC. To make this test, the
analyzer must be initialised. The methacrylate rotor can be automatically filled with distilled water by pressing
the corresponding button. Once filled, the technician must wait a few minutes for the rotor to be thermostated.
The temperature in the rotor wells must be measured with a temperature calibrated at 37ºC through the dis-
pensing hole of the rotor cover. A button makes it possible to turn the rotor in increases of 15 wells to change
the well on which the measurement is being taken. The program shows the set point temperature of the current
control. This parameter must be other than 37ºC. The technician must measure the temperature of the water
with the calibrated thermometer in the wells and enter the temperature on the screen. The program indicates if
the temperature measured is within the tolerated error margins and stores this value for the test result reports.
This screen contains a set of tests to check the functioning of the optical system. The tests are classified under
different tabs. First of all, the base line and darkness count tests must be made in order to be able to carry out
the remaining tests. To perform these tests, the analyzer must be initialised.
The optical system has a photodiode that generates an electrical current proportionate to the light intensity on
it. time. An AD converter converts the accumulated load into a digital value called count number, between 0
and 1048576. During normal functioning, the analyzer automatically adjusts the integration time for each filter
when the analysis begins and after initialisation. When the first photometry test is performed, the integration
times are also automatically adjusted. These times are adjusted in such a way that the count number of the
base line for each wavelength is as near as possible to 950000. In this way, the dynamic range of the detection
system is adapted to the light intensity present at each wavelength. The filter wheel has 10 positions. Position
0 must always contain a covered filter so that the analyzer can perform the darkness adjustment. Positions 1
to 9 can be used for optical filters.
When this test is run for the first time, the analyzer fills the first 3 rotor wells with distilled water. The analyzer
automatically adjusts the integration times and makes a base line with each of the available filters in each of
the 3 wells. The program shows the current integration times for each of the filters and the average for the 3
wells of the count numbers obtained with each filter. The screen shows the corresponding alarms in the case
of anomaly. It is also possible to access a screen where it is possible to manually vary the integration times
82
to check their effect on the count numbers. And another screen where it is possible to assign calculated in-
tegration times as reference integration times for each filter. This screen is recommended when a filter or the
lamp is physically changed. After performing the test, the analyzer continues to take optical readings using the
automatically adjusted integration times.
83
Service manual
The program shows the current integration times for each filter. On running the test, the analyzer positions the
covered filter and measures the darkness counts with each of the integration times. Each time an optical reading
is taken, the analyzer subtracts these darkness counts from the count numbers measured to obtain the light
intensity. The program shows the values obtained and issues the corresponding alarms in case of anomaly.
The values should be around 4100 - 4300. All the count numbers shown by the tests given as follows have the
darkness counts subtracted.
To perform this and the following tests, the base line and darkness count test must have first of all been perfor-
med. This test takes absorbance readings during 1 minute with the filter wheel in fixed position. The technician
can choose the rotor well on which he wishes to take the readings and fill it with the liquid he desires. He can
choose which wavelength he wishes to use. The test can also be performed with the filter covered. When the
readings end, the screen graphically displays the count numbers obtained and the absorbances with regard to
the corresponding base lines. The program also shows the averages and/or standard deviations of the count
84
numbers and the absorbances.
5.3.7.4. Stability
This test takes absorbance readings during 30 minute with the filter wheel in fixed position. The technician
can choose the rotor well on which he wishes to take the readings and fill it with the liquid he desires. He can
choose which wavelength he wishes to use. The test can also be performed with the filter covered. The test can
be cancelled at any time. When the readings end, the screen graphically displays the count numbers obtained
and the absorbances with regard to the corresponding base lines. The program also shows the averages and/
or standard deviations of the count numbers and the absorbances.
This test takes absorbance readings during 10 minute moving the filter wheel randomly. The technician can
choose the rotor well on which he wishes to take the readings and fill it with the liquid he desires. The test can
be cancelled at any time. When the readings end, the screen graphically displays the count numbers obtained
and the absorbances for each filter with regard to the corresponding base lines. The program also shows the
85
Service manual
averages and/or standard deviations of the count numbers and the absorbances for each filter.
This test enables individual absorbance readings. The technician can choose the rotor well on which he wishes
to take the readings and fill it with the liquid he desires. He can choose which wavelength he wishes to use.
The screen shows the count number obtained, the absorbance with regard to the corresponding base line, the
value of the base line.
The user can use this test to check the optical status of a reactions rotor. He or she can choose the optical filter
with which the test is to be performed. The technician must place the rotor in the analyzer and press the Test
button. If the Automatic Fill option has been chosen, the analyzer fills the 120 rotor wells with distilled water
and then makes a base line on each well with the chosen filter. The analyzer graphically displays the absorban-
ces related to the average of all the wells and tells the technician the state of the rotor (optimal, adequate or
unusable). After the test, the user must remove the rotor of the analyzer, empty it and dry it completely before
using it for analyses.
86
5.3.8. Level control scales test
This screen makes it possible to check the functioning of the level control scales of the waste and distilled water
containers. The technician must select which scales he wishes to check and place a certain amount of liquid
in the corresponding container. On pressing the Test button, the screen shows the level of liquid measured by
the analyzer(only 0 % and 100%).
This test makes it possible to check the functioning of the different detectors incorporated in the analyzer.
- Open detector of the general cover of the analyzer.
Rotor cover presence detector.
The technician can manipulate the corresponding components, for example, open and close the cover of the
analyzer and the screen shows the state of the detectors in each case.
87
Service manual
On pressing the Test button, the computer attempts to establish communication with the analyzer. The program
tells the technician if it has been possible or not.
The technician can select Automatic Configuration or Manual Configuration. In the case of the latter, he can
define the Port and the Speed.
This test makes it possible to continually reproduce work cycles of the analyzer similar to those made during
the preparation and reading of reactions in a normal working routine, but dispensing at the washing station
instead of the rotor. It is necessary for the dispensing system to be primed so that the piston does not function
dry. All the racks must be removed from the racks tray. This test can be made without the covers and housing
88
of the analyzer. The technician can program the number of cycles he wishes (1 cycle = 15 seconds). The test
can be cancelled at any time.
Once the test has been launched, the screen provides regular information about the current status of the pro-
cess. If an error occurs during the process, the test ends and the screen displays a message indicating the
element causing the error.
Partial stressing of the elements of the analyser is possible. The following elements can be stressed partially:
• X axis
• Y axis
• Z axis
• Reactions rotor
• Filter wheel
• Dispensation pump
• Membrane pumps
• 2-way electrvalve
• 3-way electriovalve
This option is used with the Photometry tool (AC15222). It is used for automate the reading process of the tool.
To use the photometry tool folow the next stetp:
1. Switch on the analyzer.
2. Run the service programme. (The older version to use the tool is the 4.1)
3. Left the analyzer switch on initialized with the service programme for 20 minutes, to preheat the lamp.
4. Select the test menu and Photometry tool option.
5. Press the Load Parameters button.
6. Insert the CD-ROM and select the file ReferenciaUtilFotometria.bin. Push Accept button.
7. Insert the tool (1) in the place of the rotor.
8. Press the Read ABS button.
9. Press the Report button to print the results report.
89
Service manual
5.4. Utilities
The program contains various technical utilities. These utilities are also accessible from the user program.
On clicking on the Disassemble Needle button, the operating arm positions itself over the rack tray. The program
alerts the technician to remove any object positioned under the arm. On clicking OK, the needle descends
and the technician can remove it to work with it or change it. To remove the needle, unscrew it by holding the
top fitting. If, while handling the needle, the carriage rises due to the pressure made by the technician, press
the Lower Needle button for the needle to descend once again. Once the needle has been reassembled on
the analyzer, press the Park button for the needle to rise. It performs the self-centering test and the arm finally
returns to its parked position. These operations must be done with utmost care since they are carried out with
the analyzer cover open and the needle may be contaminated. Laboratory gloves must always be used.
90
5.4.2. Fluid system supply
On pressing the Test button, the analyzer fills the conduits of the dispensing system and the washing station
with distilled water. To perform this operation, the operating arm is moved to the washing station. The technician
can choose whether he wishes to prime the dispensing system, the washing system or both.
On pressing the Wash button, the analyzer washes the dispensing system internally and externally. To perform
this operation, the operating arm is moved to the washing station. The technician can choose between performing
the wash with distilled water or wash solution. In the case of the latter, the analyzer asks the technician to place
a bottle of wash solution in stead of the distilled water container or to fill the latter with wash solution. Once the
wash has been performed, the analyzer asks for the distilled water container to be put back in position. Finally,
the analyzer primes the system with distilled water.
91
Service manual
When entering the screen, it is possible to choose between: Changing or checking the lamp. When a new
lamp is installed, this utility must be used to notify the analyzer that the lamp has been changed and optimize
the luminosity of the photometric system. The lamp must be changed with the analyzer in sleeping mode. If
the analyzer is on standby mode, the program shuts it down automatically. The lamp must never be touched
with fingers. Once the new lamp has been installed and the covers of the optic and rotor put back, access the
change lamp utility and press the Test button. The program starts up the analyzer, checks the light intensity of
the optical system, shuts down the analyzer and then requests the technician to remove the lamp holder again
and replace it again turning it 180º on the axis of the lamp. If the temperature of the lamp holder is high, wait until
it cools down or use pincers to hold it. The program starts up the analyzer again, measures the light intensity
of the optical system again, compares the light intensity in both possible positions and chooses the greatest
luminosity. If it is the current position, it tells the technician that the test is complete. If the best position were
the previous one, the program shuts down the analyzer and asks the technician to remove the lamp holder and
replace it, turning it 180º on the axis of the lamp, returning the lamp to its initial position. If the option selected
at the beginning was to Check the Lamp, the process is the same but without shutting down the analyzer at
the beginning.
92
5.4.5. Configuration of the filter wheel
This screen enables the modification of the analyzer filter wheel. The wheel has 10 positions. Position 0 must
always contain a covered filter so that the analyzer can perform the darkness adjustment. Positions 1 to 9 can
be used for optical filters. All the positions of the wheel must be occupied for it to work correctly. The positions
that do not contain an optical filter must be occupied by a covered filter. The analyzer includes as standard 8
optical filters in positions 1 to 8 and two covered filters in positions 0 to 9. If one of the filters is to be changed,
select the desired position of the wheel and press the Change Filter button. The analyzer automatically posi-
tions the filter wheel appropriately so that the technician can change the filter through the window of the optical
system. Next, if it is different, introduce the wavelength of the new filter that has been installed. If the filter is
covered, introduce value 0. On closing the screen, the analyzer asks if the filters have actually been physically
changed and a series of warnings are given to the technician telling him he must bear in mind whether or not
he has changed a filter.
93
Service manual
On pressing the Start button, the analyzer activates some of its mobile components, imitating functioning during
a work routine. The activated mechanical components are the operating arm, the reactions rotor and the filter
wheel. On pressing the Cancel button, the analyzer finishes the current cycle and returns to its rest position.
From this screen, it is possible to read the current adjustments that the analyser is using by pressing the button
Read Adjustments.
It is allowed to save these adjustments in a file. The technician selects the name and location of this file.
Also from this same screen and with the button Load Adjustments, the technician is allowed to select an ad-
justment file and to load it in the analyzer. Once the adjustment loading is made, the analyser turns off and
the application is closed. When reinitiating the application, the new loaded adjustments will be already active.
From the firmware version 2.80, the programme counts the number of cycles of each element and the task of
the analyser. From this menu, it is possible to read the cycles completed by the analyser. The screen displays
the said cycles with the corresponding units.
The programme automatically saves a copy of the adjustments and cycles read in a file. This file is located in
the following folder:
When a physical element of the analyser has to be changed, e.g. the Z axis belt, the counter must be reset
to zero for it to correspond to the number of cycles actually stored in the analyser. To perform this operation,
select the box of the element that is to be initialised and enter the number of cycles in the enabled box. Then
press the Load cycles button.
Using the Load adjustments button, this screen also enables the technician to select an adjustments file and load
it in the analyser. When the adjustments are loaded, the cycles are also loaded. Perform this operation when
a CPU board has to be changed. This avoids having to completely readjust the analyser; only the following
sections will have to be readjusted:
• Scales
94
• Level detection sensitivity
• Needle thermostatation
• Rotor thermostatation
In this screen the type of rotor is introduced. Each rotor comes labelled with a letter in its top part. Select in
this screen the type of rotor to use. For rotors marked with A letter, only select the letter. For the rotors marked
with other letters, select OTHERS and then introduce the light path that will come it within the box of rotors or
of the distributor.
95
Service manual
5.5. Register
This enables the management of past adjustments, tests, incidences, repairs and maintenance of the instrument.
The technician can enter the analyzer serial number so that it appears on printed service reports. If an ente-
red serial number is changed, the service records are reinitiated. In this case, the technician can store all the
previous data in a file.
The technician can enter his name so that it appears on the printed service reports.
96
5.5.2. Service Reports
The program can display and print various service reports. The printed reports contain the analyzer serial number
and the name of the current technician.
Reports are stored organised by: Adjustments, Tests, Utilities, Monitor and Summary of actions and tasks
carried out.
In all cases, it is possible to select the actions carried out within a range of dates chosen by the technician.
The technician can enter short descriptions of the incidences that may happen in the analyzer and the repairs
and maintenance operations that may be performed to the instrument in the Observations box.
This makes it possible to choose the language used in the service program.
97
Service manual
5.5.4. Users
From the Users menu, it is possible to create, delete and change users. The Change password option is for
each user to change his own password.
5.6. Monitor
These enable the low level communication with the analyzer to load new versions of the program in the flash
memory of the analyzer (firmware) or to restore the default adjustment parameters.
The firmware of the analyzer resides in a permanent flash memory. The change of this program can be made
through the computer without the need for changing the memory chip. Once the program has changed, the
analyzer is restarted with the new version of the program. While the copy process is being performed, the screen
indicates the percentage completed. To load the new version, press the Start button, previously indicating where
the program is located using the Open button. First of all, the current content of the flash memory is deleted
and then the new program is loaded. This operation may take several minutes.
There is also the option to Restore Default Adjustments, selecting the option and pressing Start.
If the technician wants the analyzer to enter monitor mode (e.g. because the analyzer does not respond because
the firmware was incorrectly loaded, he may do so by shutting down the analyzer, pressing the Force Monitor
button and then rebooting.
Once the new programme has been loaded or the default adjustments have been restored, exit the monitor by
pressing the Close button.
98
Each section explains how to manage and create different levels of access to the user program of the analyser.
When the program is installed for the first time, there is not a created user and access to the program is complete.
To activate the option of level of access to the analyser, the first time you should enter as administrator, whose
values are:
with this screen, the application with the operation by passwords is configured.
The first time that the program is activated, it forces the user to change the initial password.
It is possible to create three types of user with different access levels:
• Operator, is the user with a lower level of access to the application. He can only do working sessions, reports
of current and historical results, and validate quality control results. In the screens of programming of techniques
and contaminations, he can look up programming values, but he can not modify any parameter. He can not
delete results or alarms. This user has total access to the rack and profile programming and to the analyser’s
configuration (except for changes of filters). He can change his own password.
• Supervisor, is the user with a medium access level. This user has got the same privileges as the operator
user’s and, in addition, he has got permissions to modify the programming of techniques in the calibration
parameters and the control values. He can create a restricted number of new techniques, that is defined at
the moment of creating such user and that it is a default setting of 5. He can also modify the programming of
contaminations and change the analyser’s filters. He can change his own password.
• Administrator, is the user with total access to the analyser’s functions. He can create new users -as much at
supervisor as at operator level-, eliminate or modify users. When creating supervisor users, he has to indicate
the maximum number of new techniques that can create. He can activate or deactivate Work Without Passwords
(option within the Configuration menu). He can also activate/deactivate the working without cover detection
(this option is useful for the technical service to make verifications without needing to let the cover down). This
option activates solely when the passwords are active.
99
Service manual
When users are created, the access is limited to different parts of the program. When starting the program, an
identification of the user is requested, by the user name and a password, and then the program will automatically
restrict the different parts of the program depending on the access level permitted.
Whenever you want, you can change the user by means of the option Change of user from the User menu.
It is also allowed to eliminate users already created. Each user is capable of changing his password. All these
options can be reached from the user menu.
In order to access the consumption of reagents, it is first necessary to configure the program with the option
of working with passwords. The administrator user is the only one that can access this menu -this option is
deactivated for any other user.
In order to generate a list of the consumption of reagents, the administrator has to introduce the dates between
which he wants to know the consumption. For this, it appears a screen like this:
Such option creates two files of results, one in text format .txt and the another one in excel format .xls. These
files will be located at directory within the application directory, it will usually be:
100
Location of the folder up to version v5.4
c:\Program Files\A15\Reagents
Test Blank Prep. Calibrator Pre Control Prep. Patient Prep. Total Prep. Vol. R1 (uL) Vol. R2
(uL)
glucose 1 0 0 5 6 1332 0
alt 1 0 0 3 4 888 0
bilirrubin 9 0 0 23 32 7104 3552
Reagents
101
Service manual
To be able to access the folder of c:\ProgramData, you have to change the hiding folder option of the operating
system, for this follow the following steps, See Figure 5.1:
• Open windows explorer
• Select the View menu option
• Select the option of hidden files
102
6. Spare parts explanation
Disassembly:
1. Remove the side screws that hold the housing (1).
2. Pull the upper housing
3. Pull down the bottom housing, be careful with the tip not to bend it
Mounting:
1. Place the bottom housing
2. Place the 4 screws (1) that hold it
3. Place the top housing
4. Place the 4 screws (1) that hold it
Disassembly:
1. Remove the 2 cable glands that guide the cables (2)
2. Remove the 4 screws that hold the housing (3).
3. Remove the front cover by pulling up and forward
Mounting:
1. Attach the front cover
2. Install the 4 screws (3) that hold the front cover
103
Manual de servicio
3. Place the cable glands (2) that guide the cables in their housing
Disassembly:
1. Remove the 2 screws that hold the cover (4).
2. Extract the cover by pulling up
Mounting:
1. Place the cover
2. Place the 2 screws (4) that hold the cover
104
6.1.4. AC17318 - Refrigerator housing
Disassembly:
1. Remove the refrigerator lid and all reagent bottles
2. Remove the 4 screws that hold the housing
3. Take out the housing a little
4. Disconnect the 2 cables from the cover sensor and the status led
5. Removing completely the refrigerator housing
Mounting:
1. Install the refrigerator housing
2. Connect the two cables of the cover sensor and the status led
3. Put the 4 screws that hold the refrigerator housing
Disassembly:
1. Extract the main cover. See chapter 6.1.3
2. Remove the front cover. See chapter 6.1.2
3. Remove the refrigerator housing. See chapter 6.1.4
4. Remove the rotor cover and manually move the arm into a centered position.
5. Remove the 6 front screws (6) and the 5 rear screws (7) that hold the housing.
6. Remove the casing by lifting it
Mounting:
1. Put on the new top case
105
Manual de servicio
Disassembly:
1. Remove the upper case. See chapter 6.1.5
2. Remove the 2 screws (9) and (10) and the nut (8)
3. Extract the piston
Mounting:
1. Install the new piston
2. Follow the steps in reverse order
3. Place the main cover and check that it slides smoothly and does not fall when it is in the upper position
106
8
10
Disassembly:
1. Remove the upper case. See chapter 6.1.5
2. Remove the interconnect board. See chapter 6.5.1
3. Remove the 4 screws (1) that hold the motor (2)
4. Remove X motor
Mounting:
1. Put the new X motor
2. Place the 4 screws (1) to hold the motor loosely
3. Tighten the motor pinion with the linear rack and then tighten the 4 screws (1)
4. Place the interconnect board and housing
5. Carry out some stress cycles with the service program to verify that the assembly has been carried out correctly
107
Manual de servicio
Engine disassembly:
1. Remove the upper case. See chapter 6.1.5
2. Disconnect Y motor cable
3. Remove the 4 screws (5) with their washers that hold the motor (3)
4. Remove the Y motor from hoding plate (4)
Belt removal:
1. Remove the belt from the counter pulley
Belt mount:
1. Put on the new belt, placing it first on the counter pulley
2. Follow the Y motor mounting instructions
Y motor mount:
1. Place the new Y motor (3) with the holding plate (4) and the belt on the pulley
2. Place the 4 screws (5) with the washers to hold the motor loosely
3. Tighten the belt by displacing the motor and then tighten the screws (5)
4. Place the cover
5. Carry out some stress cycles with the service program to verify that the assembly has been carried out correctly
108
3
Encoder removal:
1. Remove the upper case. See chapter 6.1.5
2. Loosen the 4 grub screw (8) that hold the encoder (9)
3. Pull out the shaft (10) to release the encoder
Encoder removal:
1. Remove the encoder first
2. Remove the 4 screws (6) that hold the motor (7)
3. Disconnect Z motor cable
4. Remove the 4 screws (8) that hold the encoder (9)
5. Extract the motor to the side (7)
Motor mounting:
1. Fit the new Z motor
2. Place the 4 screws (8) to hold the motor (7)
3. Place the encoder (9)
Encoder mount:
1. Place the encoder on the motor (7) and the shaft (10)
2. Adjust the encoder to be centered with the photodetector
3. Tighten the 4 grub screw (8)
4. Carry out some stress cycles with the service program to verify that the assembly has been carried out correctly
109
Manual de servicio
10
Disassembly:
1. Remove the upper case. See chapter 6.1.5
2. Remove the Y-belt. See chapter 6.2.3
3. Remove the 2 screws (12) that hold the Y axis
4. Remove the 2 screws (11). Be careful and unscrew strongly when loosening the screws because they have
threadlocker.
5. Remove the spring assembly.
Mounting:
1. Place the spring assembly
2. Fit the 2 screws (11) without tightening them. Before tightening them, check that the 2 axes are perfectly aligned.
3. Tighten the motor pinion with the linear rack and then tighten the 4 screws (1)
4. Place the belt and tighten it. See chapter 6.2.3
5. Put the 2 screws (12) loosely.
6. Tighten the spring:
a. Keep the probe set in the upper position
b. Using a wrench, turn the pulley counterclockwise. Make 3-4 laps
c. Tighten the screws (12), matching with the flat part of the shaft.
110
11
12
Cable disassembly:
1. Remove the upper case. See chapter 6.1.5
2. Unscrew the clamping fitting (14)
3. Cut all the ties that hold the tube and cables
4. Disconnect the Teflon tube at both ends (tip and electrovalve) (13)
5. Disconnect the connector that goes to the CPU board and the grounding
6. Remove the cables from the arm
111
Manual de servicio
13
14
Disassembly:
1. Remove the arm housing. See chapter 6.1.1
2. Remove the tip
3. Extract the tube from the top of the board
4. Remove the 3 screws (15), also remove the nuts on the other side and the protection plate
5. Remove the tip plate
112
15
16
Disassembly:
1. Unscrew the 4 screws (1) and separate the manifold (2) from the rest of the pump.
2. Carefully stretch the manifold to remove the piston from the seals that comes with the manifold.
3. Turn over the manifold and unscrew the 4 screws (3) to separate the parts holding the chamber (4), the O-ring
(5), the bubble suppressor (6) and the adapter seal with the seal (7).
Mounting:
1. Replace the manifold.
2. Orient each part correctly (9) according to Illustration 6.14
3. Rotate the manifold holding the pieces (14).
4. Verify that the O-ring (14) is clean. Put it in your accommodation. Check that it has been inserted correctly and
does not come out of the guides (15). See Illustration 6.15
5. Check that the bubble suppressor (16) is free of burrs and dirt. Place it on top of the O-ring. Once assembled,
verify that the O-ring is not visible. See Illustration 6.15
6. Orient the seal and seal adapter (17) correctly. The flat face of the adapter must coincide with the window of the
manifold (18). See Illustration 6.16
7. Screw in the 4 screws (19) with a tightening torque of 30 Ncm and crosswise. See Illustration 6.16
8. Lubricate piston and seal with WS (Washing Solution)
9. Orient the manifold correctly with the pump. The window of the manifold (21) has to coincide with the guide screw
(22) of the pump. See Illustration 6.16
10. Insert the piston into the seal
11. Tighten the manifold against the pump, position the 4 screws (20) and cross tighten with a torque of 60 Ncm.
12. Mount the pump on the instrument. Connect the motor and home detector cable. Connect the tubes
113
Manual de servicio
13. Check the operation of the pump with the service program.
4
1
5
6
7
3
114
8
10
Ilustración 6.15
11
13
12
Ilustración 6.16
115
Manual de servicio
14
15
16
Ilustración 6.17
Tools:
Disassembly:
1. Follow the disassembly steps according to chapter 6.3.1
2. Detach the seal (17) from the seal adapter (11).
Mounting:
1. Follow the mounting steps according to chapter 6.3.1
2. Before performing point 7, replace the seal (17) with a new one. Dampen the seal adapter and seal with WS
(Washing Solution) to facilitate insertion of the seal into its housing. Use a flat piece to press the seal.
3. Verify the seal is properly leveled with the seal adapter.
4. Continue the assembly steps according to chapter 6.3.1
116
17
Ilustración 6.18
Tools: 2 and 3 Allen wrench and 7 open end wrench, ceramic pump grease
Disassembly:
1. Follow the disassembly steps according to chapter 6.3.1
2. Remove the home detector.
3. Remove the 4 screws with nuts that secure the pump body (18) with the motor.
4. Pull out the pump body (18).
Mounting:
1. Replace the defective part.
2. Unscrew the piston support from the spindle to grease the thread.
3. Grease the guide of the piston support. Put enough grease in the hole in the bracket where the spindle enters.
4. Screw in the piston support with the spindle. Wipe off excess grease with a cloth.
5. Place the pump chamber. Align the groove in the bracket (20) with the body anti-rotation system (19).
6. Install the screws and nuts to secure the body with the motor.
7. Place the home detector.
8. Follow the manifold mounting steps according to chapter 6.3.1
117
Manual de servicio
18
19
20
21
Ilustración 6.16. Lubricating the spindle and mounting the pump body
Disassembly:
1. Remove the upper case. See 6.1.5
2. Remove the connection tubes from the pump and the connector
3. Remove the 3 screws (20) that hold the pump
4. Remove the pump (18)
Mounting:
1. Replace the pump with a new one
2. If the analyzer does not have the support plate (19), put it in place with the screws
3. Install the 3 pump retaining screws (20)
4. Connect the tubes to the pump
5. Verify the operation of the washing station with the service program
118
18
19
20
Disassembly:
1. Remove the upper case. See 6.1.5
2. Remove the connection pipes from the manifold (21)
Mounting:
1. Replace the tube with a new one
2. Put it in its position
3. Verify operation with service program
21
22
23
119
Manual de servicio
Disassembly:
1. Remove the upper case. See 6.1.5
2. Disconnect the connection cables to the rotor and remove the lamp holder
3. Remove the 3 screws (1) and (2) that hold the assembly to the base
Mounting:
1. Follow the reverse order to disassembly
Disassembly:
1. Remove the upper case. See 6.1.5
2. Disconnect the cables from the temperature probe
3. Remove the thermal insulation from the temperature sensor
4. Remove the temperature sensor (3)
Mounting:
1. Replace temperature sensor
2. Remove the thermal silicone from the housing and put new thermal silicone on the end of the new sensor
3. Fit the new temperature sensor
4. Perform rotor temperature adjustment
120
3
Disassembly:
1. Remove the upper case. See chapter 6.1.5
2. Extract the rotor assembly. See chapter 6.4.1
3. Loosen the screw (5) that holds the rotor insulator
4. Remove the grub screw (7) that holds the cover detector
5. Remove the cover detector by stretching from the bottom or with the help of a screwdriver press in the direction
of the arrow (6)
Mounting:
1. Follow the reverse order to disassembly
Fan disassembly:
1. Remove the upper case. See 6.1.5
121
Manual de servicio
Peltier disassembly:
1. Extract the 4 fans
2. Remove the 4 radiators (9)
3. Extract the 4 peltiers. The peltiers are all connected with a single wiring
Peltier mount
1. Clean the silicone residues from the heating and the radiators. Use alcohol.
2. Place a thin layer of thermal silicone on both sides of the peltier (10)
3. Be careful not to put too much silicone so as not to make a thermal bridge with the excess
4. Orient the 4 peltiers as indicated in Illustration 6.26
5. Place each peltier with its radiator (9) with its screws
6. Fit the fans (8) with their screws
7. Make rotor position and temperature adjustments
10
122
11
Disassembly:
1. Remove the upper case. See 6.1.5
2. Extract the rotor assembly. See chapter 6.4.1
3. Remove the 4 screws (12) that hold the upper rotor plate
4. Remove the 2 nuts (13) that hold the motor
5. Separate the motor assembly (14) from the heating channel
6. Remove the 4 screws (15) to separate the motor (16) from the spacer
Mounting
1. Assemble in reverse order of disassembly
2. Before tightening the nuts (13), adjust the separations between the gear and the pinion.
3. Tighten the nuts
4. Check the movement of the gear with the pinion. To verify movement, slowly rotate one full turn of the crown and
verify that the crown is never blocked with the pinion. Also check that the crown is not too loose with respect to
the pinion.
123
Manual de servicio
12
13
124
14
15
16
Disassembly:
1. Remove the upper case. See 6.1.5
2. Extract the rotor assembly. See chapter 6.4.1
3. Rotate the rotor assembly
4. Remove the 4 screws (17) that hold the protection plate
5. Remove the 2 studs (18) that hold the filter drum with the shaft
6. Remove the 4 screws (19) that hold the motor (21)
125
Manual de servicio
Mounting
1. Place the motor (21) in its location together with the filter drum (20)
2. Match the flat part of the motor shaft with the grub screw position (18). Put threadblocker on
the grub screw thread.
3. Install the screws (19) and tighten them
4. Check that the filter drum rotates freely
5. Put on the protection cap
6. Mount the rotor assembly on the analyzer
7. Make rotor adjustments
17
Ilustración 6.30
18
Ilustración 6.31
126
19
Ilustración 6.32
20
21
Ilustración 6.33
Disassembly:
1. Remove the upper case. See chapter 6.1.5
2. Disconnect the cables from the board
3. Remove the 4 screws (1)
4. Remove the protection plate and the connection board
Mounting:
1. Place the new board
127
Manual de servicio
Disassembly:
1. Fold down the rear cover of the electronics.
2. Disconnect all the cables from the board.
3. Remove the screws (2) or (3) depending on the board to be removed
Mounting:
1. When reinstalling the board, carefully reconnect all connectors.
128
2
Disassembly:
1. Fold down the rear cover of the electronics.
2. Remove the input and output connectors by loosening the screws on the sources (4).
3. Cut the ties (7)
4. Remove the 2 screws (5) that hold the refrigerator power supply.
5. Remove the 5 screws (6) that hold the main power supply
6. Extract the power supply
Mounting:
1. Follow the reverse order to disassembly
129
Manual de servicio
5
4
130
2
1
Disassembly:
1. Fold down the rear cover of the electronics.
2. Disconnect the cables from the communication board
3. Remove the 6 nuts (1)
4. Remove the communication board
Mounting:
1. Install a new board
2. Perform the reverse disassembly process
Ilustración 6.39
Disassembly:
1. Remove the screws (1) from the support cover (2) of the photometric system.
2. Remove the screws (3) from the support cover (4) of the photometric plate.
3. Slightly move the support cover (4) towards the center of the rotor and remove it from its housing.
4. Disconnect the flat band from the photometric system plate.
5. Remove the screws (5) to separate the support cover from the photometric plate.
Mounting:
1. Reinstall the board, carefully reconnect all connectors.
2. Perform the reverse disassembly process
Ilustración 6.40
Ilustración 6.41
132
5
Ilustración 6.42
Disassembly:
1. Remove all cables from the board
2. Remove the 4 screws (7) that hold the interconnection board (6)
Mounting:
1. Install the new board, carefully reconnect all connectors.
2. Perform the reverse order process
Ilustración 6.43
133
Manual de servicio
To replace the status indicator led or the lid sensor it is necessary to replace the fridge body.
Disassembly:
1. Remove the 4 screws (1)
2. Body Pull (2)
3. Disconnect the connectors (3)
Mounting:
1. Proceed in reverse order to replace the fridge body
1 2 3
Ilustración 6.44
Peltier disassembly:
1. Remove the body from the fridge as explained in chapter 6.6.1
2. Open the electronic cover.
3. Remove the ties (1) and the cover (2) to free the cables that go to the reagent refrigerator. Disconnect all the
connectors that go to the refrigerator board.
4. Close the electronic cover again.
5. Remove the front cover.
6. Remove the ties (3).
7. Remove the general cover of the equipment. See chapter 6.1.5
8. Flip the instrument over so that it rests on its back.
9. Remove the protective plate (4) by removing the screws (5).
10. Remove the 4 outer screws (6) and remove the assembly.
11. Once removed from its housing, remove the ties (7)
12. Unplug the fan connectors (8)
13. Remove the temperature sensor (9) by removing the screw (10)
14. Remove the level detection signal cable (11) by removing the screw (12). Completely remove the fridge together
with the cable from the peltier cells.
15. Once the fridge is completely removed, remove the caps and screws (13), in order to remove the radiators.
134
16. Remove the peltiers (14). Attention remember the polarity of the same and then reassemble the new ones with
the same polarity. Look at the color of the cables.
Fan disassembly:
1. Follow the previous steps up to point 14
2. Remove the faulty fan
Mounting:
1. Clean the surfaces of the radiators (15) and the cold plate (16) of traces of thermal conductor. To do this, first use
a cutter taking care not to scratch the surfaces. Finish cleaning these surfaces with alcohol.
2. Place the thermal conductors (16) on both sides of the peltiers. Remove the paper that protects one of the faces.
Apply heat to this face. It is enough applying a little breath on it. Paste the conductor in the peltier.
3. Place the peltiers in their housing, respecting the polarity. Mount them in the same position that the replaced ones
were. Pay attention to pass the cable through the housing for this purpose (17).
4. Replace the radiators. Tighten the screws evenly in the order shown. Do not overtighten the screws to avoid
damaging the peltiers.
5. Reassemble the refrigerator in its housing. Follow the steps above in reverse order, placing new ties where they
were removed.
1
2
Ilustración 6.45
Ilustración 6.17.
135
Manual de servicio
Ilustración 6.46
7
8
10 12
6 11
Ilustración 6.47
14
13
Ilustración 6.48
136
16
15
Ilustración 6.49
17
4 2
1 3
Ilustración 6.50
Disassembly:
1. Proceed as in chapter 6.6.2 above up to point 13
Mounting:
1. Reassemble the temperature sensor having previously placed a certain amount of thermal silicone.
2. Reassemble the entire assembly following the steps performed in reverse order.
137
Manual de servicio
7. Preventive maintenance
In this chapter, firstly it describes step by step the different operations necessary for both preventive maintenance
and repair of the analyzer. Then, it gives some basic recommendations for preventive maintenance of the instrument.
Finally, a series of indications for its care and cleaning are exposed.
Annual maintenance
1 Clean electronic boards and fans from dust
2 Check the fluidic connections from the pump to the probe and the connections of the washing
station, Check the tightness of the fittings and that none of them leaks liquid.
3 Check the seal of the dosing pump. Verify that there are no leaks and that no bubbles are produced.
4 Clean the filters in the distilled water container
5 Clean the washing station
6 Disassemble and clean the probe. Check your status.
7 Check the arm (Check the wear of the belts and their tension)
8 Check probe shock detection and protection system
9 Check the condition of the heating channel
138
Annual maintenance
10 Check belt and reaction rotor tension
11 Clean filters, slit and photodiode
12 Check placement and detector of general cover and reaction cover
13 Check that the general cover is properly supported and does not fall.
14 Carry out the tests with the service program
After carrying out any maintenance, make sure that all the elements that have been replaced or dis-
mounted have been correctly assembled.
WARNING
1. Never use detergents or abrasive products for cleaning the surface of the analyzer. Use only a damp cloth with
water and pH-neutral soap.
2. If a reagent or corrosive product spills or splashes onto the apparatus, clean it with a damp cloth and soap imme-
diately. If necessary, protect your hands with appropriate laboratory gloves.
3. All the components of the analyzer have drainage conduits leading to the exterior to enable the elimination of any
liquid spilled and to prevent the apparatus from flooding. If the spillage is significant, the liquid spilled onto the
table through the drainage conduits and the analyzer must be adequately cleaned.
4. When not in use, close the main cover of the analyzer to protect it from dust.
The components of the optical system must be cleaned periodically in order to keep them free from dust and dirt.
These components are the lamp, the lenses, the filters and the photodiode. The recommended necessary material
is as follows:
• Special paper for cleaning optical components (non-abrasive paper which does not leave solid residue).
• Ether and alcohol solution at 50%
• Cotton wool buds.
• Small bellows.
All the optical components must be handled in an area of maximum cleanliness and with great care given that they are
fragile and delicate. To remove them and refit them, the corresponding instructions given in the Maintenance chapter
must be carefully followed. Avoid touching the useful area of these components with fingers. The lenses, filters and the
photodiode must be held by the sides. Do not touch the lamp bulb. To handle the lamp, use the wrapping, cutting it at
the terminal ends and squeezing them until they come out. To clean the optical components, bear in mind the following
1. Remove the dust on the surface of the component with the bellows. This will avoid scratching by particles when
cleaning with paper.
2. Carefully clean the surface of the component with the cleaning paper.
3. If the dirt is persistent or greasy, clean the component with the paper moistened with the alcohol and ether solu-
tion. Then dry with dry paper. To clean the filters or the window of the photodiode, use the cotton wool buds with
the paper in difficult and delicate areas.
4. Finally, it is recommended that you use the bellows once again, thus removing any remains of paper or cotton.
139
Manual de servicio
The analyser automatically performs conditioning of the fluid system at warm-up and shut-down. This conditioning
consists of a 3 primings of the dosing system, the first with air, the second with washing solution and the third and
last with system liquid. To perform this conditioning, the bottles of system liquid (marked with a blue dot) and washing
solution (marked with a green dot) must be full, and the waste bottle (marked with a red dot) must be empty. The final
conditioning consists of performing a priming with washing solution.
With initial wash, the system is ready to work in optimum conditions for the whole workday, offering its best performan-
ce. With the final wash, the analyser cleans the tip at the end of the day’s work, thus keeping it in perfect condition
for the next workdays. The user can also perform conditionings of the dosing system when desired by pressing the
conditioning button in the vertical button bar on the monitor screen while the analyser is in standby mode. It is also
recommended to clean and check the status of the filters of the system liquid container at least once every 3 months.
If the needle is obstructed by solid residue, it must be removed and cleaned using the metal cleaning rod supplied
with the analyzer. For this, the Remove the dispensing needle utility should be used. It is also recommendable to
clean the outside surface of the needle with a piece of cotton or a soft cloth dampened with alcohol. The needle must
be replaced if it is noticeably deteriorated.
It is important for the interior of the instrument to be free from dust at all times in order to preserve the correct functio-
ning of the different components. For this, remove the front housing of the analyzer and carefully clean the dust inside
the instrument.
To clean the refrigerator, disassemble the refrigerator body as explained in section 6.1.4 Clean the cold plate with a
damp cloth with water and neutral soap. Clean the body of the fridge in the same way.
This is an standard process for verifying the correct operation of the instrument after doing an intervention.
This process can be applied to the following operations:
• Instrument reception (in the office, before installation)
• Maintenance
• After a repair
The goal is to standarize the verification process and the acceptance criteria.
The duration time of the process is: 1,5 hours
Use Biosystems reagent and controls or equivalents codes, use always new fresh reconstituted calibrators and controls.
140
Model A15 / A25
Code Description
18010 / 18042 Bovine / Human Control Serum II 5mL
100-1000 µm Pipette (Calibrated)
Disposable pipette tips
Distilled water Type II
141
Manual de servicio
15. You can use same Distilled water bottle from the customer
16. Look for empty spaces in the rotor. Try not to remove the customer’s reagents form the rotor
17. Fill the cups and bottles with the following liquids, you can fill the cups with more volume, but in that case, you
will not check the correct adjustment of the sample for the dead volume.
18. Position the reagents and samples
Place the bottles and the samples in the positions where you positioned them in the software
19. Prepare the washing solution
20. Run the session
21. Export the results
22. When it finishes, go to the test result screen, press the option TEST. Press print current test button
23. Export the current test results in .CSV format
24. Copy the content of the file to the “Data” sheet of the excel file
25. Enter the rest of the data in the “instruction and batch” sheet of the excel file
26. In the “summary” and in the “verification results” you will get all the results and the final result.
27. Save and print the excel file with the serial number as a name. Discard all the rest of the samples and calibration.
142
The following table shows the acceptance criteria of the verification:
Values of the coeficient of variation
Model: A15/A25
GLUCOSE AST PHOSPHORUS
SCI SCII SCI SCII SCII
Tolerance (CV%) 3.00 3.00 5.00 3.00 3.00
Model: Y15/Y25
Glucose-Fructose Acetico Acid Glycerol PAN
N2 N4 N2 N4 N2 N4 N4
Tolerance (CV%) 3.00 3.00 5.00 3.00 3.00 3.00 3.00
Also all the concentration results must be between the range of the control.
143
Manual de servicio
Analyzer Model:
Serial number:
Instalation date:
144
Preventive maintenance
record
Laboratory name:
Analyzer Model:
Serial number:
Instalation date:
7 Check the arm (Check the wear of the belts and their tension)
8 Check probe shock detection and protection system
9 Check the condition of the heating channel
10 Check the separation between gear and pinion of the rotor
11 Clean filters, slit and photodiode
12 Check placement and detector of general cover and reaction cover
13 Check that the general cover is properly supported and does not
14 Carry out the tests with the service
145
Service manual
GENERAL SPECIFICATIONS
Automatic random access analyzer
Preparation cycle time 24 s (up to 150 prep/h)
Warm-up time 25 mins
Reading time for each preparation Every 24 s, up to 10 mins
Possible configurations
Sample racks Reagent racks Number of samples Number of reagents
1 3(1)* 24 30
2 2(0)* 48 20
3** 1 72 10
REAGENT COOLER*
Cooling capacity 10 ºC under ambient temperature (Ambient
temperature of de 25 ºC)
Power supply independent of analizer. Dedicated
power source and switch.
Open lid sensor and indicator power led.
*A15C(Reagent cooler version)
146
Actuator Resistive elements
Control PID
Dispensation temperature 37 °C
Repeatability ± 0.5 °C
DISPENSING PUMP
Ceramic piston with PTFE-graphite seal
Piston diameter 8 mm
Dispensing volume 3 µL - 1250 µL
Programmable reagent volume 10 µL -440 µL
Programmable sample volume 3 µL - 40 µL
OPTICAL SYSTEM
Halogen lamp 6 V, 10 W
Wavelength selection with compensated
interferential filters
Detection system with silicon photodiode and 20-
bit AD integrator-converter
Measurement range from -0.05 A to 3.0 A
Reading speed 1.25 readings/s
Maximum number of filters 9
Base configuration of the filter wheel A15 340, 405, 505, 535, 560, 600, 635, 670 nm
Base configuration of the filter wheel Y15 340, 405, 420, 480, 520, 560, 600, 620, 635 nm
Wavelength precision ± 2 nm
147
Service manual
Bandwidth 10 ± 2 nm
Digital resolution < 0.0001 A
Base line stability max. 0.004 A in 30 min, at 505 nm
POWER REQUIREMENTS
Input voltage 125 Vac - 230 Vac, ±10%, 50/60 Hz
Power 150 VA (200VA)*
Electrical installation category (overvoltage II
category)
The power point must be officially approved, earthed and the cable must have a minimum cross-
section of 1.5 mm2.
*A15C(Reagent cooler version)
ATMOSPHERIC CONDITIONS
Interior use
Height < 2500 m
Temperature 10°C - 35°C
Relative humidity < 75%
Contamination level 2
The manufacturer reserves the right to modify any technical specification without prior notice.
148
COMPLIANCE WITH DIRECTIVES
Y15/Y15C model:
Low voltage directive 2014/35/EU
Electromagnetic compatibility directive 2014/30/EU
A15/A15C/Y15/Y15C:
Directive on restriction to the use of certain 2011/65/EU
dangerous substances in electrical and electronic
equipment.
149
Service manual
151
Service manual
Accessories list
Code Representation Description
AC13188 User Program CD
152
Accessories list
Code Representation Description
BO11493 50 mL bottle with stopper (10 units)
153
Service manual
Accessories list
Code Representation Description
FI13812 520nm filter assembly
154
Accessories list
Code Representation Description
FI11488 System Liquid Container Filters (2)
155
Service manual
Accessories list
Code Representation Description
AC13199 Height adjustable leg
Spare parts
Code Represntations Description
CA13393 Base housing
AC17356
156
Spare parts
Code Represntations Description
CA13379 Probe bottom housing
AC17350
157
Service manual
Spare parts
Code Represntations Description
IN11557 Main switch
158
Spare parts
Code Represntations Description
FO11562 photodetectors (5)
AC17316 Manifold
159
Service manual
Spare parts
Code Represntations Description
AC17317 Manifold tube
160
Spare parts
Code Represntations Description
MO13363 Ceramic pump motor
AC16594
ZO13374 Lampholder
161
Service manual
Spare parts
Code Represntations Description
AC17327 Rotor centering
162
List of necessary tools
Code Representation Description
Metric Allen wrench set.
Heat conductive silicone grease or the like
Welder
Screwdrivers or two 3mm diameter allen wrenches
AC13694 Torx wrench set
AC15212 Loctites set: 243, 641, 415
AC13935 Fluke thermometer
AC13079 ELESA NT1 mechanical grease (only for ceramic
pump)
AC11860 Temperature adjustment tool
163
Service manual
Compatibilities table
Usua. A15 4.7 4.8.3 4.9.1 4.9.3 5.2 5.4
firmware
3.70 x x x x
4.12 x x x x
5.10 x x x x
6.02 x x
6.16 x x
6.19 x x
166
fixed time
Be able to perform absorbance techniques
167
Service manual
This chapter explains the instructions for the refurbish and control of the instruments traded as rental or loan, or
second hand.
Terms
• Life: Analyzers have the functionality for life as long as they keep maintenance conditions specified by the ma-
nufacturer and the use of original spare parts. Note that the manufacturer is obliged to provide spare parts up to
5 years after discontinuing the analyzer.
• Number of times can be refurbished: You can refurbish as many times as necessary, provided the overhaul
following the instructions described in this section and the use of original spare parts.
• Accessories used in refurbished devices: Any accessories used in refurbished analyzer are provided by the
manufacturer. Any fungible remaining from the last use such as rotors, extra reagent bottles, cleaning solution,
fluid system, and so on, will discarded.
• Staff: The refurbishment will be done by personnel authorized by Biosystems. It will normally be the distributor
service staff, or someone has received the appropriate training.
Instructions
• Users carrying out disinfection processes must wear gloves and protective apparel. All consumable materials
(reaction rotor, reagent bottles, sample tubes) will be treated as potentially infectious waste.
• Check whether there are any reaction rotors in their compartments or reagent bottles in the refrigerator. If there
are, dispose of them.
• Empty the high contamination waste tank and washing solution tank.
• Wash the inside of the equipment using a damp cloth and neutral soap. If there are splashes, wash them with
alcohol.
• Follow the applicable national guidelines for disposing of waste that is considered potentially infectious.
Components to change
• Change the tip. (AC11500) Use the utility change the tip of the service program. Discard the tip. Treat as potentially
infectious. To manipulate the tip wear gloves to protect the user.
• Changing the inner tubes of Teflon (AC13365 and TU13366)
• Changing the lamp and he lamp socket. (LA13195 and ZO13374) Use the utility lamp replacement of the service
program.
• Changing the pump seal ceramic (AC13361)
• Update firmware and software
Carry out the following checks, consult the service manual (chapter 3) to carry out the verifications. If a verification
does not fit within the values, adjust again
168
• Stability at 505 nm
• Repeatability by moving filters
To carry out the functional verification of the instrument, follow the detailed instructions in chapter 7.2.6 and fill in the
excel sheet with the verification data and then print it out for registration.
Ask the technical service department for the Excel file.
Ending
• Place an indelible label to the refurbished analyzer. The label contains the following information (check with the
national legislation of each country):
1. The name and address of the company responsible for implementing and refurbishing
2. The date of last revision
3. The number of refurbish performed
4. Indication that it is a renewed equipment
5. Lifetime after refurbishing.
• Wear gloves and protection for the user who will perform the disinfection process. Treat disposables (rotors,
reagent bottles, sample tubes) as potentially infectious.
• Check if there is any reaction rotor in the housing and reagent bottles. If so, discard them.
• Clean the outside of the computer with a damp cloth and mild soap. In case of splash, locally wipe with alcohol.
• In case of changing or disposing of the tip, use the tip change utility of the service program. Treat it as potentially
infectious.
• Follow the current national regulations for disposal for potentially infectious waste.
169
RECORD OF REFURBISH PROCESS
1. Spare change
TIp change Ok / No Ok
Prepared by:
Teflon tubing change Ok / No Ok
Lamp change Ok / No Ok
Pump seal change Ok / No Ok
Update firmware and
Ok / No Ok
software
FUNCTIONAL VERIFICATION
Verification according to
Ok / No Ok
section 7.2.6
Print and save the results
Ok / No Ok
of the Excel sheet
TERMINATION
Finish
Place label of refurbished Ok / No Ok
equipment
Add full box of accesories Ok / No Ok
Check packaging Ok / No Ok