C.
1 General
This document requires the manufacturer to compile a list of known and foreseeable hazards
associated
with the medical device in both normal and fault conditions and to consider the foreseeable
sequences
of events that can produce hazardous situations and harm. According to the definitions, a hazard
cannot
result in harm until such time as a sequence of events or other circumstances (including normal
use)
lead to a hazardous situation. At this point, the risk can be assessed by estimating both severity and
probability of occurrence of harm that could result (see Figure C.1). The probability of occurrence of
harm can be expressed as a combination of separate probabilities (P1, P2) or as a single probability
(P).
A decomposition into P1 and P2 is not mandatory
Fundamental risk concepts
                                 Depending on the complexity
                                 of the medical device, a hazard
                                 can lead to multiple
 NOTE 1                          hazardous
                                 situations, and each hazardous
                                 situation can lead to multiple
                                 harms.
                                 The probability of occurrence
 NOTE 2                          of harm (P) can be composed
                                 of separate P1 and P2 values.
 NOTE 3                          The thin arrows represent
                                 elements of risk analysis and
                                 the thick arrows depict how
                                 a hazard
                                 can lead to harm
Figure C.1 — Pictorial example of the relationship between hazard, sequence of events,
hazardous situation and harm (from ISO/IEC Guide 63:2019[2])
A good starting point for this compilation is a review of experience with the same and similar
types of medical devices. The review should take into account a manufacturer’s own experience and,
where appropriate, the experience of other manufacturers as reported in adverse event databases,
publications, scientific literature and other available sources. This type of review is particularly
useful
for the identification and listing of typical hazards and hazardous situations for a medical device and
the
associated harm that can occur. Next, this listing and aids such as the list of examples in Table C.1
can be
used to compile an initial list of hazards.
It is then possible to begin identification of some of the sequences of events that together with
hazards
could result in hazardous situations and harm. Since many hazards might never result in harm and
can
be eliminated from further consideration, it could be useful to perform this analysis by starting with
the harm that the medical device might cause and work backwards to the hazardous situations,
hazards
and initiating causes. However, although this approach is useful for the reason described, it should
be
recognised that it is not a thorough analysis. Many sequences of events will only be identified by the
systematic use of risk analysis techniques (such as those described in ISO/TR 24971 [9]). Analysis
and
identification are further complicated by the many events and circumstances that have to be taken
into
consideration such as those listed in Table C.2. Thus, more than one risk analysis technique, and
especially
complementary techniques, are often used to complete a comprehensive analysis. Table C.3
provides
examples of the relationship between hazards, sequences of events, hazardous situations, and harm.
Although compilation of the lists of hazards, hazardous situations and sequences of events should
be completed as early as possible in the design and development process to facilitate risk control, in
practice identification and compilation is an ongoing activity that continues throughout the life
cycle of
the medical device through post-production to disposal.
This annex provides a non-exhaustive list of possible hazards that can be associated with different
medical devices (Table C.1) and a list of events and circumstances (Table C.2) that can result in
hazardous situations, which can result in harm. Table C.3 provides examples in a logical progression
of
how a hazard can be transformed into a hazardous situation and produce harm by a sequence of
events
or circumstances.
Recognising how hazards progress to hazardous situations is critical for estimating the probability
of occurrence and severity of harm that could result. An objective of the process is to compile a
comprehensive set of hazardous situations. The identification of hazards and sequences of events are
stepping stones to achieve this. The lists in the tables in this annex can be used to aid in the
identification
of hazardous situations. What is called a hazard needs to be determined by the manufacturer to suit
the
particular analysis.
C.2 Examples of hazards
The list in Table C.1 can be used to assist in the identification of hazards associated with a particular
medical device, which could ultimately result in harm.
Table C.1 — Examples of hazards
 Energy hazards                    Biological and chemical hazards          Performance-related
                                                                            hazards
 Acoustic energy                   Biological agents                        Data
 — infrasound                      Bacteria                                 — access
 — sound pressure                  Fungi                                    — availability
 — ultrasonic                      Parasites                                — confidentiality
 Electric energy                   Prions                                   — transfer
 Electric fields                   Toxins                                   — integrity
 Leakage current                   Viruses                                  Delivery
 — earth leakage                   Chemical agents                          — quantity
 — enclosure leakage               Carcinogenic, mutagenic, reproduc        — rate
 Magnetic fields                   tive                                     Diagnostic information
 Static discharge                  Caustic, corrosive                       — examination result
 Voltage                           — acidic                                 — image artefacts
 Mechanical energy                 — alkaline                               — image orientation
 Kinetic energy                    — oxidants                               — image resolution
 — falling objects                 Flammable, combustible, explosive        — patient identity /
                                                                            information
 — high pressure fluid             Fumes, vapors                            Functionality
 injection
 — moving parts                    Osmotic                                  — alarm
 — vibrating parts                 Particles (including micro- and nano     — critical performance
 Potential (stored) energy         particles)                               — measurement
 — bending                         Pyrogenic
 — compression                     Solvents
 — cutting, shearing               Toxic
 — gravitational pull              — asbestos
 — suspended mass                  — heavy metals
 — tension                         — inorganic toxicants
 — torsion                         — organic toxicants
 Radiation energy                 — silica
 Ionizing radiation               Immunological agents
 — accelerated particles          Allergenic
 (alpha particles, electrons,     — antiseptic substances
 protons, neutrons)               — latex
 — gamma                          Immunosuppressive
 — x-ray                          Irritants
 Non-ionizing radiation           — cleaning residues
 — infrared                       Sensitizing
 — laser
 — microwave
 — ultraviolet
 Thermal energy
 Cryogenic effects
 Hyperthermic effects
C.3 Examples of events and circumstances
n order to identify foreseeable sequences of events, it is often useful to consider events and
circumstances that can cause them. Table C.2 provides examples of events and circumstances,
organized
into general categories. Although the list is certainly not exhaustive, it is intended to demonstrate
the
many different types of events and circumstances that need to be taken into account to identify the
foreseeable sequences of events for a medical device.
Table C.2 — Examples of events and circumstances
 General category      Events and circumstances
                   Inadequate specification of:
                   — design parameters
                   — operating parameters
                   — performance requirements
                   — in-service requirements (e.g. maintenance, reprocessing)
                   — end of life
Requirements
                   Insufficient control of:
                   — manufacturing processes
                   — changes to manufacturing processes
                   — materials
Manufacturing      — materials compatibility information
processes          — subcontractors
                   Inadequate packaging
Transport and
                   Contamination or deterioration
storage
                   Inappropriate environmental conditions
                   Physical factors (e.g. heat, pressure, time)
                   Chemical factors (e.g. corrosion, degradation, contamination)
Environmental
                   Electromagnetic fields (e.g. susceptibility to electromagnetic disturbance)
factors
                   Inadequate supply of power
                   Inadequate supply of coolant
Cleaning,          Lack of validated procedures
disinfection and   Inadequate specification of requirements
sterilization      Inadequate performance of cleaning, disinfection or sterilization
Disposal and       No or inadequate information provided
scrapping          Use error
                   Biodegradation
                   Biocompatibility
Formulation        No information or inadequate specification provided
                   Incorrect formulations
                   Use error
                     Confusing or missing instructions for use
                     Complex or confusing control system
                     Ambiguous or unclear state of the medical device
                     Ambiguous or unclear presentation of settings, measurements or other
                     information
                     Misrepresentation of results
                     Insufficient visibility, audibility or tactility
                     Poor mapping of controls to actions, or of displayed information to actual
                     state
 sability
                     Controversial modes or mapping as compared to existing equipment
                     Use by unskilled or untrained personnel
                     Insufficient warning of side effects
                     Inadequate warning of hazards associated with re-use of single-use medical
                     devices
                     Incorrect measurement and other metrological aspects
                     Incompatibility with consumables, accessories, other medical devices
                     Incorrect patient identification
                     Slips, lapses and mistakes
                     Loss of electrical or mechanical integrity
                     Deterioration in performance (e.g. gradual occlusion of fluid or gas path,
                     change
 Functionality       in resistance to flow, electrical conductivity) as result of ageing, wear and
                     re
                     peated use
                     Failure of a component due to ageing, wear or fatigue
                     Unsecured data ports that are externally accessible (e.g. network, serial or
                     USB ports)
 Security            Data without encryption
                     Software vulnerabilities that can be exploited
                     Software updates without authenticity confirmation
C.4 Examples of relationships between hazards, foreseeable sequences of events,
hazardous situations and the harm that can occur
Table C.3 illustrates the relationship between hazards, foreseeable sequences of events, hazardous
situations and harm for some simplified examples. Remember that one hazard can result in more
than
one harm and that more than one sequence of events can give rise to a hazardous situation.
The decision on what constitutes a hazardous situation needs to be made to suit the particular
analysis
being carried out. In some circumstances it can be useful to describe a cover being left off a high
voltage
terminal as a hazardous situation, in other circumstances the hazardous situation can be more
usefully
described as when a person is in contact with the high voltage terminal.
 Hazard                       Foreseeable sequence of              Hazardous situation       Harm
                              events
 Electromagnetic              (1) Electrode cable                  Line voltage appears on Serious
 energy                       unintentionally                      electrodes              burns
 (high voltage)               plugged into power line                                      Heart
                              receptacle                                                   fibrillation
 Chemical (vol                (1) Incomplete removal of            Development of gas        Infarct
 atile solvent,               volatile                             embo                      Brain
 embolus)                     solvent used in manufacturing        lism (bubbles in the      damage
                              (2) Solvent residue converts to      blood
                              gas at                               stream) during dialysis
                              body temperature
 Biological (mi               (1) Inadequate instructions          Bacteria released into    Bacterial
 crobial contam               provided                             airway of patient         infection
 ination)                     for decontaminating re-used          during
                              anaesthesia tubing                   anaesthesia
                              (2) Contaminated tubing used
                              during
                              anaesthesia
 Functionality                (1) Electrostatically charged        Failure to deliver        Minor
 (no delivery)                patient                              insulin                   organ
                              touches infusion pump                to patient with           damage
                              (2) Electrostatic discharge (ESD)    elevated                  Decreased
                              causes                               blood glucose level, no   conscious
                              pump and pump alarms to fail         warning given             ness
 Functionality                (1) Implantable defibrillator        Defibrillator cannot      Death
 (no output)                  battery                              deliver shock when an
                              reaches the end of its useful life   arrhythmia occurs
                              (2) Inappropriately long interval
                              between clinical follow-up visits
 Measurement                  (1) Measurement error                Incorrect information     Progression
 (incorrect in                (2) No detection by user             reported to clinician,    of disease
 formation)                                                        lead                      Serious
                                                                   ing to misdiagnosis       injury
                                                                   and/
                                                                   or lack of proper
                                                                   therapy