0% found this document useful (0 votes)
227 views6 pages

Gcem 40e

The document provides information on the GCEM 40 Series Extractive Gas Analyser, which is a continuous emission monitoring system that can measure multiple gas species such as CO, NOx, SO2, and CH4. It utilizes infrared absorption technology and has self-calibration features. The system extracts flue gas samples via a heated probe and transfers them through a heated sample line to the analyzer cabinet for measurement and reporting of emissions data.

Uploaded by

Bình Đặng
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
227 views6 pages

Gcem 40e

The document provides information on the GCEM 40 Series Extractive Gas Analyser, which is a continuous emission monitoring system that can measure multiple gas species such as CO, NOx, SO2, and CH4. It utilizes infrared absorption technology and has self-calibration features. The system extracts flue gas samples via a heated probe and transfers them through a heated sample line to the analyzer cabinet for measurement and reporting of emissions data.

Uploaded by

Bình Đặng
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Product Data Sheet

CODEL Continuous Emission Monitoring

GCEM 40 Series Extractive Gas Analyser CO, NO, NO2, NOx, SO2, CH4, HCl, CO2 & H²O

Extractive low cost, low maintenance devices for continuous process and emissions monitoring.

Single or Multi-species infrared absorption analyser - Accurate, reliable and low maintenance technology

Extractive System - With full zero & span verification using certified gas

Gas temperature and pressure sensors - On-board normalisation to standard reference conditions

Auto routine calibration - Zero and span calibration using audit gas

Analogue and serial outputs - Export of data to SCADA, DCS and historian systems

Environmental Monitoring Solutions www.codel.co.uk


Doc i/d : 100140 Issue : A Rev : 1 Date : 5/2/2014 © 2010 CODEL International Ltd. We reserve the right to modify designs without prior notice
GCEM 40 Series Extractive Gas Analyser
The GCEM40E hot extractive multi-channel gas analyser is
CODEL’s industry-proven continuous emissions monitor for
difficult applications. Designed exclusively for use on a
wide range of applications where the flue gas temperature
is abnormally high, low or saturated.

Three decades of development, knowledge and practical


experience have been utilised to produce this advanced
technology gas analyser which gives complete flexibility of
use on process or emissions applications whilst delivering Heated Probe
superb accuracy and repeatability at a competitive price.

Many conventional extractive systems require the sampled


gas to be cleaned and dried to a very high standard prior
to analysis, invariably resulting in a high maintenance
demand. Such elaborate pre-conditioning is not required;
the GCEM40E creates ‘perfect’ duct conditions in a
temperature controlled chamber within a separate
free-standing cabinet. Process conditions are extracted
using a heated probe system which has an option of
compressed air blow-back for excessively dusty
applications. Once the sample has been drawn it is simply
cooled (or heated) then transferred along a heated sample
line, without further conditioning, to be measured using a
CODEL multi-channel analyser housed in the cabinet.

Environmental agencies demand that continuous emission


analysers have the facility to prove their performance using
known concentration audit gasses. The GCEM40E provides
the facility to automatically check and control zero
calibration point using clean, dry compressed air or Heated Sample Line
nitrogen. Where independent span checks are required,
bottled gases of known concentration can be injected
directly into the measurement chamber.

The GCEM40E analyser is capable of measuring a range of


CO, NO, NO2, NOx, SO2, CH4, HCl, N2O, CO2, H2O and O2
simultaneously and with integral temperature and pressure
sensors can compute fully normalised data directly
in mg/Nm3.

The analyser can be easily incorporated into the CODEL


SmartCEM system along with dust, flow and oxygen signals
to deliver a complete emission monitoring solution capable
of meeting today’s tough legislative demands.

GCEM40E Gas Analyser

Environmental Monitoring Solutions www.codel.co.uk


Doc i/d : 100140 Issue : A Rev : 1 Date : 5/2/2014 © 2010 CODEL International Ltd. We reserve the right to modify designs without prior notice
GCEM 40 Series Gas Analyser - System Arrangement
Sample Line

Sensor Unit

Inputs/Outputs

Power In

GCEM 40 Series Gas Analyser - Options

Embedded Touch Screen


Display

Environmental Monitoring Solutions www.codel.co.uk


Doc i/d : 100140 Issue : A Rev : 1 Date : 5/2/2014 © 2010 CODEL International Ltd. We reserve the right to modify designs without prior notice
GCEM 40 Series Extractive Gas Analyser - Overall Dimensions

450

500
1200

Note: 100mm Mounting


Brackets underneath panel

1200

Note: All dimensions are in mm

Environmental Monitoring Solutions www.codel.co.uk


Doc i/d : 100140 Issue : A Rev : 1 Date : 5/2/2014 © 2010 CODEL International Ltd. We reserve the right to modify designs without prior notice
SmartCEM Software
Detailed and customisable reports to meet Regulatory Agency requirements (including Chineese
EPA, LCPD, WID)
User configurable real-time and historical graphing facilities for data analysis
Data export to popular applications such as Excel etc
Multi-user networking capability
Real-time normalisation of data
Data storage on PC hard drive of at least 10 years
Easy installation and set-up
Intuitive operation
Will operate on Windows based operating systems
SmartCEM Emissions Monitoring Software provides the complete solution to data gathering and
analysis on CODEL-based monitoring systems. With a simple installation and set-up routine, the
program takes only minutes to load and configure and comes with a comprehensive on-board help
feature.

Real-time Displays
Using a serial link to communicate with the sensors, SmartCEM updates
the live data screen every 30 seconds to keep the user aware of the
current situation on plant, while alarms warn the operator of high emis-
sion levels.
Data is displayed in numerous formats - ppm, %, mg/Nm3,
mg/m3, kg/hr assuming flow measurement is available etc. - and the
change of format is instantaneous.

Graphs
A key feature of SmartCEM is the provision of real-time and historical
graphs providing detailed analysis of emissions data. One useful feature
enables the user to highlight a period of time on the graph and obtain
the average emissions. Data is easy to read and is configurable to
individual operator requirements.

GSM Unit
The GSM unit allows CODEL engineers direct access to any analyser in our range giving them full
online remote diagnostic information for technical support.*

This service can be offered for a nominal monthly charge

* Network coverage can limit the use of this service

Environmental Monitoring Solutions www.codel.co.uk


Doc i/d : 100140 Issue : A Rev : 1 Date : 5/2/2014 © 2010 CODEL International Ltd. We reserve the right to modify designs without prior notice
GCEM 40 Series Gas Analyser - Technical Specification
Sensor Unit
Gas Species Options Single or multi-gas measurements available:
Maximum 3 gases from: CO, NO, NO2, NOx, SO2, CH4 plus gas temperature
Plus optional : CO2, H2O, pressure for data normalisation
Measuring units ppm, mg/Nm³, mg/m³ , %
Response Time Less than 200 Seconds (T90)- MCERTS Specification
Gas Temperature Below dewpoint to 1300°C, See probe options below
Calibration Automatic and manual zero/span verification
Gas Species CO, NO, NO2, NOx, SO2, HCl, CH4, N2O CO2, H2O
Max Measuring Range 0 - 6000 ppm, higher ranges available on request 0 - 25%
Accuracy +/- 2ppm or 2% of span 0.5% or 2% of span
Resolution +/- 1ppm 0.1%
Zero & span drift +/- 2ppm or 2% of span per month 0.5% or 2% of span
Linearity +/- 2% of span 2% of span
Repeatability +/- 5ppm or 1% of span 0.3% or 1% of span
Ambient Temperature -20°C to +50°C
Construction Corrosion resistant epoxy coated aluminium housing sealed to IP66

Heated Sample Line


Single Core PTFE tube, non interchangable, self regulating
Probe
Standard Probe 0 - 600°C Stainless Steel
High Temp Probe 0 - 900°C Hasteloy, 0 - 1100°C Inconol, 0 - 1300°C Kanthol
High Dust Loads 0 - 600°C Stainless Steel
Compliances
EMC 89/336/EEC directive compliant
Low Voltage 73/23/EEC directive compliant

Analyser Cabinet
Analogue outputs 4-20mA current outputs for each gas channel supplied, isolated, 500Ω load max, fully
configurable from software.
Logic Outputs up to 8 x volt-free SPCO contacts, 50V, 1A max, configurable as alarm contacts
1 x volt-free SPCO contact, 50V, 1A max, for data valid signal
Inputs 1 x 4-20mA for oxygen as standard (upto 8 optional)
Serial Data RS232 / RS485 (modbus protocol)
Construction Mild steel construction powder coated to IP66 or 55 (Double door)
Ambient Temperature -20°C to +50°C
Power Supply 180 - 264VAC by switch
Air Dryer For clean, dry, oil-free air
Services
Power Mains 230 VAC, single phase, 50/60hz
Air Requirement Clean and dry compressed air @ 4bar
Optional Items
Central Data Controller For connection to additional SmartCEM systems
Plug-in PCB 4 - 20mA Input PCB
Plug-in PCB Contact input PCB
Plug-in PCB RS485 PCB for Modbus Output
Software CODEL SmartCEM Software - See SmartCEM software data sheet
Distributor
CODEL International Ltd Tel : +44 (0)1629 814351 Bechstein-Messtechnik GmbH
Station Building Fax : +44 (0)8700 566307 Zum Stollen 4
Station Road Web : www.codel.co.uk 51674 Wiehl
Germany
Bakewell Email : sales@codel.co.uk Tel: +49 (0) 2262 752 797
Derbyshire Fax: +49 (0) 2262 752 801
DE451GE info@bechstein-messtechnik.de
www.bechstein-messtechnik.de
United Kingdom

You might also like