Astm d3559 2003
Astm d3559 2003
e
This Document Posted By Not Affiliated Or
Public.Resource.Org, Inc., Authorized by ASTM
a California Nonprofit or by the United States
Organization. Government
Official Incorporator:
THE EXECUTIVE DIRECTOR
OFFICE OF THE FEDERAL REGISTER
WASHINGTON, D.C.
-~- -------,-,-
''1
ASTM Logo Removed
Designation: D 3559 - 03
This standard has been approved for use by agencies of the Department of Defense.
559-03
test serves to determine whether the lead content of potable 9. Interferences
water is above or below the acceptable limit. 9.1 Other metals usually do not interfere in the determina-
tion of lead by increasing or decreasing the amount of absorbed
5. Purity of Reagents radiation. The most common interference is caused by a
5.1 Reagent grade chemicals shall be used in all tests. chemical reaction in the flame that prevents conversion of the
Unless otherwise indicated, it is intended that all reagents shall lead to the atomic state.
conform to the specifications of the Committee on Analytical 9.2 High concentrations of calcium, such as those connected
Reagents of the American Chemical Society. 6 Other grades with the coal industry, will give lead concentrations higher than
may be used, provided it is first ascertained that the reagent is which actually exist. This can be corrected by using a back-
of sufficiently high purity to permit its use without lessening ground correction technique, or by the chelation-extraction
the accuracy of the determination. procedure (Test Method B).
5.2 Unless otherwise indicated, references to water shall be 9.2.1 The equipment manufacturer's instructions for use of
understood to mean reagent water conforming to Specification specific correction technique shall be followed.
D 1193, Type I. Other reagent water types may be used
provided it is first ascertained that the water is of sufficiently 10. Apparatus
high purity to permit its use without adversely affecting the
10.1 Atomic Absorption Spectrophotometer, for use at 283.3
bias and precision of the test method. Type II water was
nm.
specified at the time of round-robin testing of this test method.
NOTE I-The manufacturer's instructions shall be followed for all
6. Sampling instrumental parameters. Wavelengths other than 283.3 nm may be used if
they have been determined to be equally suitable.
6.1 Collect the samples in accordance with Practice D 1066,
Specification D 1192, and Practices D 3370, as applicable. 10.2 Lead Light Source, hollow-cathode lamps or
6.2 Samples shall be preserved with HN0 3 (sp gr 1.42) to a electrodeless-discharge lamps have been found satisfactory.
pH of 2 or less immediately at the time of collection, normally 10.3 Oxidant:
about 2 mL/L of HN0 3 . If only dissolved lead is to be 10.3.1 Air, which has been passed through a suitable filter to
determined, the sample shall be filtered through a 0.45-J.lm remove oil, water, and other foreign substances, is the usual
membrane filter before acidification. oxidant.
10.4 Fuel:
TEST METHOD A-ATOMIC ABSORPTION, 10.4.1 Acetylene-Standard, commercially available acety-
DIRECT lene is the usual fuel. Acetone, always present in acetylene
cylinders, can affect analytical results. The cylinder should be
7. Scope replaced at 50 psig (345 kPa).
7.1 This test method covers the determination of dissolved 10.4.1.1 Warning-"Purified" grade acetylene containing a
and total recoverable lead in most waters and wastewaters. special proprietary solvent rather than acetone should not be
7.2 The test method is applicable in the range from 1.0 to 10 used with poly(vinyl chloride) tubing as weakening of the
mg/L of lead. The upper limits of detectability can be increased walls can cause a potential hazardous situation.
to concentrations greater than 10 mg/L by dilution of the 10.5 Pressure-Reducing Valves-The supplies of fuel and
sample. oxidant shall be maintained at pressures somewhat higher than
the controlled operating pressure of the instrument by suitable
8. Summary of Test Method valves.
8.1 Lead is determined by atomic absorption spectropho-
tometry. Dissolved lead is determined by aspirating the filtered 11. Reagents
and preserved sample directly with no pretreatment. Total 11.1 Hydrochloric Acid (sp gr 1.19)-Concentrated hydro-
recoverable lead is determined by aspirating the sample fol- chloric acid (HCI).
lowing hydrochloric-nitric acid digestion and filtration. The
NOTE 2-If the reagent blank concentration is greater than the method
same digestion procedure may be used to determine total detection limit, distill the Hel or use a spectrograde acid. Caution-When
recoverable cadmium (Test Methods D 3557), chromium (Test Hel is distilled, an azeotropic mixture is formed (approximately 6 N Hel
Methods D 1687), cobalt (Test Methods D 3558), copper (Test is formed). Therefore, whenever concentrated Hel is used in the prepa-
Methods D 1688), iron (Test Methods D 1068), manganese ration of a reagent or in the procedure, use double the volume specified if
(Test Methods D 858), nickel (Test Methods D 1886), and zinc distilled Hel is used.
(Test Methods D 1691). 11.2 Lead Solution, Stock (I mL = 1 mg lead)-Dissolve
1.5999 g of lead nitrate (Pb (N0 3 h) in a mixture of 10 mL of
HN0 3 (sp gr 1.42) and 100 mL of water. Dilute to 1 L with
6 Reagent Chemicals, American Chemical Society Specifications, American water. Warning-Lead salts are toxic. Handle with care and
Chemical Society, Washington, DC. For suggestions on the testing of reagents not avoid personal contamination.
listed by the American Chemical Society, see Analar Standards for Laborato/y
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
11.3 Lead Solution, Standard (1 mL =0.1 mg lead)-Dilute
and National Formulary, U.S. Pharmaceutical Convention, Inc. (USPC), Rockville, 100.0 mL of stock lead solution to 1 L with HN0 3 (1 + 499).
MD. Store all solutions in polyethylene bottles. Warning-Lead
03559-03
salts are toxic. Never pipet by mouth. Pipet with the end of a TABLE 1 Determination of Bias, Direct
suction device or employ other conventional means of quanti- Amount Amount
Statistically
tative measurement. Added, Found,
Sr, So,
Bias,%
Significant
mg/L mg/L (95 % Confi-
mg/L mg/L
11.4 Nitric Acid (sp gr 1.42)-Concentrated nitric acid dence Level)
(HN0 3 )· Reagent Water, Type II
1 1.01 0.08 0.04 +1.00 no
NOTE 3-If the reagent blank concentration is greater than the method 6 6.01 0.28 0.14 +0.17 no
detection limit, distill the RN03 or use a trace metal grade acid. 8 8.02 0.34 0.14 +0.25 no
Selected Water Matrices
11.5 Nitric Acid (1 + 499)-Add 1 volume of HN0 3 (sp gr
1.42) to 499 volumes of water. 1 1.00 0.00 0.06 0.00 no
6 6.11 0.25 0.16 +1.83 yes
8 7.99 0.36 0.23 -0.13 no
12. Standardization
12.1 Prepare 100 mL each of a blank and at least four
standard solutions to bracket the expected lead concentration 15. Precision and Bias 7
range to be analyzed by diluting the lead standard solution
(11.3) with HN0 3 (1 + 499). Prepare the standards each time
15.1 Fourteen operators from nine laboratories participated
in this study. One operator performed the analysis in quadru-
the test is to be performed.
plicate, twelve in triplicate and one in duplicate at each
12.2 When determining total recoverable lead, add 0.5 mL concentration level.
of HN0 3 (sp gr 1.42) to each blank and standard solution and 15.2 The bias of this test method for lead is listed in Table
proceed as directed in 13.2 through 13.4. After the digestion of 1.
the blank and standard solutions has been completed in 13.4, 15.3 These data may not apply to waters of other matrices.
return to 12.3 to complete the standardization for total recov- 15.4 This section on precision and bias conforms to Practice
erable determinations. When determining dissolved lead, pro- D 2777 - 77 which was in place at the time of collaborative
ceed with 12.3. testing. Under the allowances made in 1.4 of D 2777 - 98,
12.3 Aspirate the blank and standards and record the instru- these precision and bias data do meet existing requirements of
ment readings. Aspirate HN0 3 (1 + 499) between standards. interlaboratory studies of Committee D19 test methods.
12.4 Prepare an analytical curve by plotting the absorbance
versus the concentration for each standard on linear graph 16. Quality Control
paper. Alternatively, read directly in concentration if this 16.1 In order to be certain that analytical values obtained
capability is provided with an instrument. using these test methods are valid and accurate within the
confidence limits of the test, the following QC procedures must
13. Procedure be followed when analyzing lead.
13.1 Measure 100.0 rnL of a well-mixed acidified sample 16.2 Calibration and Calibration Verification:
into a 125-mL beaker or flask. 16.2.1 Analyze at least three working standards containing
concentrations of lead that bracket the expected sample con-
NOTE 4-If only dissolved lead is to be determined, start with 13.5. centration, prior to analysis of samples, to calibrate the
instrument. The calibration correlation coefficient shall be
13.2 Add 5 rnL of HCI (sp gr 1.19) to each sample.
equal to or greater than 0.990. In addition to the initial
13.3 Heat the samples on a steam bath or hot plate in a calibration blank, a calibration blank shall be analyzed at the
well-ventilated hood until the volume has been reduced to 15 eml of the batch run to ensure contamination was not a problem
to 20 rnL, making certain that the samples do not boil. during the batch analysis.
NOTE 5-For samples having appreciable amounts of suspended matter
16.2.2 Verify instrument calibration after standardization by
or dissolved matter, the amount of reduction in volume is left to the analyzing a standard at the concentration of one of the
discretion of the analyst. calibration standards. The concentration of a mid-range stan-
dard should fall within ± 15 % of the known concentration.
13.4 Cool and filter the samples through a suitable filter
16.2.3 If calibration cannot be verified, recalibrate the
such as fine-textured, acid washed, ashless paper, into 100-rnL
instrument.
volumetric flasks. Wash the filter paper two or three times with
16.3 Initial Demonstration of Laboratory Capability:
water and adjust to volume.
16.3.1 If a laboratory has not performed the test before, or if
13.5 Aspirate each filtered and acidified sample and deter- there has been a major change in the measurement system, for
mine its absorbance or concentration at 283.3 nm. Aspirate example, new analyst, new instrument, etc., a precision and
HN0 3 (1 + 499) between samples. bias study must be performed to demonstrate laboratory
capability.
14. Calculation
14.1 Calculate the concentration of lead in each sample, in
milligrams per litre, using the calibration curve established in 7 Supporting data are available from ASTM Headquarters. Request RR:D
12.4. 19-1030.
... " ... " .. " ........ - ........ , ...... .
03559-03
in 8.1 is used for total recoverable lead. The same chelation- 22.2 When determining total recoverable lead use 125-mL
extraction procedure may be used to determine total recover- beakers or flasks, add 0.5 mL RN0 3 (sp gr 1.42) and proceed as
able cadmium (Test Methods D 3557), cobalt (Test Methods directed in 23.2 thru 23.15. When determining dissolved lead
D 3558), copper (Test Methods D 1688), iron (Test Methods use 250-mL separatory funnels and proceed as directed in 23.5
D 1068), nickel (Test Methods D 1886), and zinc (Test Meth- thru 23.15.
ods D 1691). 22.3 Construct an analytical curve by plotting the absor-
bances of standards versus micrograms of lead. Alternatively,
19. Interferences read directly in concentration if this capability is provided with
19.1 See Section 9. the instrument.
23. Procedure
20. Apparatus
23.1 Measure a volume of a well-mixed acidified sample
20.1 All apparatus described in Section 10 are required. containing less than 100 Ilg lead (100-mL maximum) into a
125-mL beaker or flask and adjust the volume to 100 mL with
21. Reagents water.
21.1 Bromphenol Blue Indicator Solution (1 glL)-Dissolve
NOTE 7-If only dissolved lead is to be determined, measure a volume
0.1 g of bromphenol blue in 100 mL of 50 % ethanol or
of filtered and acidified sample containing less than 100 Ilg of lead
2-propanol. (lOO-mL maximum) into a 250-mL separatory funnel, and start with 23.5.
21.2 Carbon Disulfide (CS 2 ).
23.2 Add 5 mL of HCl (sp gr 1.19) to each sample.
21.3 Chloroform (CRCI 3 ).
21.4 Hydrochloric Acid (sp gr 1.19)-Concentrated hydro- 23.3 Reat the samples on a steam bath or hot plate in a
chloric acid (HCl). well-ventilated hood until the volume has been reduced to 15
21.5 Hydrochloric Acid (1 + 2)-Add 1 volume of HCI (sp to 20 mL, making certain that the samples do not boil.
gr 1.19) to 2 volumes of water. NOTE 8-When analyzing brines and samples containing appreciable
21.6 Hydrochloric Acid (1 + 49)-Add 1 volume ofHCI (sp amounts of suspended matter, the amount of reduction in volume is left to
gr 1.19) to 49 volumes of water. the discretion of the analyst.
21.7 Lead Solution, Stock (1.0 mL =200 Ilg lead)- 23.4 Cool and filter the samples through a suitable filter
Dissolve 0.3198 g of lead nitrate (Pb (N0 3)2) in water such as fine-textured, acid-washed, ashless paper, into 250-mL
containing 1 mL of HN0 3 (sp gr 1.42) and dilute to 1 L with separatory funnels. Wash the filter paper two or three times
water. Warning-Lead salts are toxic. Handle with care and with water and bring to approximately a 100-mL volume.
avoid personal contamination. 23.5 Add 2 drops of bromphenol blue indicator solution and
21.8 Lead Solution, Intermediate (1.0 mL = 2.01l g lead)- mix.
Dilute 10 mL of lead stock solution and 1 mL of RN0 3 (sp gr 23.6 Adjust the pH by addition of NaOH (100 gIL) solution
1.42) to 1 L with water. Warning-Lead salts are toxic. Never until a blue color persists. Add HCI (1 + 49) by drops until the
pipet by mouth. Pipet with the end of a suction device or blue color just disappears; then add 2.5 mL of HCI (1 + 49) in
employ other convenient means of quantitative measurement. excess. The pH at this point should be 2.3.
21.9 Lead Solution, Standard (1.0 mL = 0.2 Ilg lead)-
NOTE 9-The pH adjustment in 23.6 may be made with a pH meter
Immediately before use, dilute 10.0 mL of lead intermediate instead of using an indicator.
solution and 1 mL of HN0 3 (sp gr 1.42) to 100 mL with water.
This standard is used to prepare working standards at the time 23.7 Add 10 mL of pyrrolidine dithiocarbamic acid-
of analysis. chloroform reagent and shake vigorously for 20 min
21.10 Nitric Acid (sp gr 1.42)-Concentrated nitric acid (Warning-See 21.11).
(RN0 3 )· 23.8 Plug the tip of the separatory funnel with cotton, allow
21.11 Pyrrolidine Dithiocarbamic Acid-Chloroform the phases to separate, and drain the chloroform phase into a
Reagent-Add 36 mL of pyrrolidine to 1 L of CHCI3 • Cool the 100-mL beaker.
solut~on and add 30 mL of CS 2 in small portions, swirling
23.9 Repeat the extraction with 10 mL of chloroform and
between additions. Dilute to 2 L with eHCI 3 . The reagent can drain the chloroform layer into the same beaker.
be ti~ed for several months if stored in a cool, dark place. NOTE IO-If color still remains in the CHCl 3 extract, reextract the
Warp-ing: All components of this reagent are highly toxic. aqueous phase until the chloroform layer is colorless.
Carbon disulfide is also highly flammable, prepare and use in 23.10 Place the beaker on a hot plate at low heat and
a we,ll-ventilated hood. Avoid inhalation and direct contact. evaporate just to near dryness. Remove beaker from heat and
2r'.l2 Sodium Hydroxide Solution (100 glL)-Dissolve 100 allow residual solvent to evaporate without further heating.
g of sodium hydroxide (NaOH) in water and dilute to 1 L. Warning-Perform in a well-ventilated hood.
23.11 Hold the beaker at a 45° angle and slowly add
22. Standardization dropwise 2 mL of HN0 3 (sp gr 1.42), rotating the beaker to
22.1 Prepare a blank and sufficient standards from 0.0 to effect thorough contact of the acid with the residue.
1000 IlglL lead from the lead standard solution (21.9) by 23.11.1 If acid is added to the beaker in a vertical position,
making appropriate dilutions in water. Prepare standards im- a violent reaction will occur accompanied by high heat and
mediately prior to use. spattering.
03559-03
35.2 Place exactly 10.0 mL of a well-mixed sample con- TABLE 3 Determination of Precision and Bias for Lead by
Differential Pulse Anodic Stripping Voltammetry
taining less than 100 f,tglL of lead into the cell.
Amount Amount Statistically
NOTE16-Concentrations greater than 100 /-lg/L of lead may be ST. So. %
Added. Found. Significant
11 giL Ilg/L Bias
95 % Level
determined by dilution. j.lg/L j.lg/L
35.3 Add 2.0 mL of redistilled HN0 3 to each sample. Reagent Water Type VI
35.4 Evaporate the samples without boiling on a hot plate or 20 21.6 4.00 2.95 + 8.00 No
40 38.4 5.70 3.71 -4.00 No
steam bath until the sample just reaches dryness (do not "bake" 80 78.5 10.21 9.89 -1.88 No
as this may cause losses due to volatilization). Steps 35.3 and Water of Choice
35.4 may be repeated if necessary for samples containing large
20 22.4 5.33 4.65 + 12.0 No
amounts of organic matter. 40 40.3 6.59 6.91 + 0.75 No
35.5 Cool, add 5.0 mL of citrate buffer, and 100 f,tL of 80 78.8 10.27 11.10 - 1.50 No
hydroxylamine solution. Warm the solution 15 min to reduce
the ferric iron and to effect dissolution of the metals in the
buffer.
35.6 Bring to volume of 10 to 12 mL with citrate buffer (pH (4)
3.0). The exact volume need not be known because the
standard additions method will be used to quantitate. where:
35.6.1 To overcome selenium at levels up to 1000 f,tglL, add i1 = stripping peak height for the sample,
1 mL of ascorbic acid. i2 = stripping height for the sample plus standard,
35.7 Deaerate for 10 min at 105 mm 3 /min with an oxygen- v = volume of standard taken for spiking,
free stream of nitrogen. V = volume of sample before digestion,
35.8 After deaeration is complete, extrude with the hanging Cs = concentration of standard used in spike, mglL, and
mercury drop electrode a mercury droplet whose area is 1.5 to CLI = concentration of the unknown in the sample, mglL.
3 mm2, as determined in Annex A2. Turn on the magnetic 36.2 The following is a sample calculation using this
stirrer and adjust the stirring rate so that the solution beneath equation:
the mercury droplet is well stirred but there is no visible i l = 0.459 /-lA (5)
movement of the mercury droplet. The stirrer is turned on 15 s i2 = 1.24 /-lA
prior to deposition to assure uniform rotational speed. v = 0.02mL
35.9 Connect the cell. Deposit at - 0.80 V versus SCE for
V= 1O.0mL
exactly 2 min, switch off stirrer, and wait exactly 30 s before
beginning the scan. The quiescent period between deposition Cs = 10.0 mg/L
and scan allows convection to cease.
35.9.1 Appendix X4 gives typical stripping curve shapes, (0.459)(0.02)(10)
peak potential, and sensitivities (in f,tA/5 f,tglL) for lead CII = (1.24)(0.02) + (1.24 _ 0.459)(10) = 0.01172 mg/L
deposited into a mercury droplet with a 2.9-mm2 area for 2 min
with stirring plus 30 s without stirring. 37. Precision and Bias 13
35.10 The following typical settings are for polarographic 37.1 Eight operators from seven laboratories participated in
instrumentation capable of performing differential pulse this study by determining three replicates at each concentration
work: 8 electrode, hanging mercury drop electrode (area 1.5 to level.
3 mm2); initial potential, - 0.80 V versus SeE; scan rate, 5 37.2 The precision and bias of this test method for lead is
mV/s; scan direction, "+"; modulation amplitude, 25 mY; listed in Table 3.
current range, 1 to 20 f,tA; drop time, 0.5 s; display direc- 37.3 These data may not apply to waters of other matrices.
tion,"-"; low pass filter, off; mode, differential pulse; deposi- 37.4 This section on precision and bias conforms to Practice
tion time, 2 min with stirring plus 30 s quiescent; scan range, D 2777 - 77 which was in place at the time of collaborative
stop - 0.20 V. testing. Under the allowances made in 1.4 of D 2777 - 98,
35.10.1 The linearity of this test method has been tested up these precision and bias data do meet existing requirements of
to currents of 20 f,tA. If the sample gives stripping peaks with interlaboratory studies of Committee D19 test methods.
currents larger than 20 f,tgA, one may decrease the deposition
time (see Appendix Xl), although this technique has not been 38. Quality Control
interlaboratory tested. The recommended procedure is to dilute 38.1 In order to be certain that analytical values obtained
the sample and proceed as directed in 35.2 through 35.10. using these test methods are valid and accurate within the
35.11 To obtain a blank, place exactly 10.0 mL of Type IV confidence limits of the test, the following QC procedures must
water into the cell and proceed as directed in 35.2 through be followed when analyzing lead.
35.10. 38.2 Calibration and Calibration Verification:
36. Calculation
36.1 Calculate the concentration of lead determined by the 13 Supporting data are available from ASTM Headquarters. Request RR: D19-
ANNEXES
(Mandatory Information)
ALl Leach the voltammetric cells in concentrated HN0 3 A1.2 After the leaching period, rinse the glassware with
for at least 24 h prior to use. During the leaching period for the reagent water and place in an oven to dry. Exclude the oven
cells, fill other glassware with the HN0 3 (2 + 3), cover by a drying step for the platinum wire and outgassing tube. Clean
sheet of plastic film 16 to prevent contamination by trace metals the reference electrode salt bridge tube initially by soaking for
in atmospheric particles, and soak for at least 24 h. Clean all 24 h with HNOi2 + 3) but thereafter keep it immersed in a
glassware that may contact the sample solution. This includes small amount of the purified buffer solution. Again, use plastic
the voltammetric cells, digestion beakers, stirring bars, plati- film 16 to cover any areas that might tend to accumulate dust.
num wire, and outgassing tubes. For very low-level determi-
nations leach instead in aqua regia (1 + 1) for 1 h prior to use. A1.3 Remember when pelforming trace analyses that any
solution or any equipment that is left open to the air can
become contaminated by the trace metals from atmospheric
16 Parafilm available from Fisher Scientific Co., Fairlawn, NJ, has been found particles. Care should be taken to prevent this from happening
satisfactory. by liberal use of plastic film. 16
A2.4 Obtain the weight of the beaker plus the mercury A2.6.3 Area of Hg drop = 4 'IT i
(3WH 4 'TI'PHg) 2/3
(WT )·
A2.7 Sample Calculation:
A2.5 Discard the mercury and obtain the weight of the WHg = 0.006228 for 1 Hg drop
beaker (WB ). (3)(0.006228 g)
Area of 1 Hg drop = 4 'IT ( 3) 2/3
A2.6 Calculate the mercury drop area (assuming a spherical 4'IT 0.0135438 g/mm
drop) as follows: = 12.56636 (1.0978 X 10 -I) 2/3
A2.6.1 = 2.881 mm 2.
WHg = (WT - WB)/l0 = weight of a single mercury drop.
A2.6.2 Obtain the density of mercury at room temperature, A2.8 Tabulated here are typical surface areas for each small
pHg, from the following table 17. If the room temperature used vertical division on a manually operated hanging mercury drop
is not listed here, find the density at the correct temperature electrode. 9
from a suitable reference source. Surface Area, mm 2 Reading, Small Vertical Division
Temperature,O C pHg, g/mL17 1.42 2
1.86 3
2.23 4
2.60 5
17 Reproduction from Handbook of ChemistJy and Physics, 44th Edition, The 2.92 6
Chemical Rubber Publishing Co., Cleveland, OH, 1963, p. 2199. 3.23 7
APPENDIXES
(N onmandatory Information)
XU The sensitivity of DPASV is dependent upon a time. The use of deposition times as long as 30 min has been
number of factors and thus can be varied if so desired. The reported in the literature when detection limits below 0.1 /-tglL
experimental conditions chosen for this work are those which (ppb) were required. When sensitivity such as this is required,
are best suited for the concentration range covered by the the additional time required is well spent.
samples which were analyzed. Experimental settings that can
be varied to improve the sensitivity include: hanging mercury X1.4 Another procedure to improve the sensitivity is to use
drop electrode size, deposition time, modulation amplitude, larger pulse modulation amplitudes. For typical differential
instrument gain, and stirring rate. pulse polarographic instrumentation, 8 the pulse modulation
amplitude may be increased to 50 mV with no significant loss
X1.2 The size of the mercury drop can be decreased to of resolution.
increase the sensitivity of this method. The recommended
mercury droplet size is six divisions (see Annex A2), but X1.5 One may also change the gain on the instrument to
mercury droplets of eight divisions can be used. 9 Droplets improve the sensitivity. The highest gain that can be used in
larger than this are not practical because they are very easily this experiment gives a current of 0.1 /-tNin. This gain can be
dislodged from the capillary. increased by a factor of 2 to 5, and sensitivity is increased by
a corresponding amount. When using the higher gain on the
Xl.3 The sensitivity of DPASV is variable over a wide instrument, it should be noted that the current at the beginning
range by increasing the deposition time. A deposition time of 2 of the deposition may be well above the limiting value and the
min is chosen for the concentration range investigated because instrument overload light will be on. This does not mean the
this time gives adequate sensitivity in a reasonable length of instrument is malfunctioning and the experiment can be
X2.l Remove oxygen from nitrogen by anyone of a variety technique because of the critical dependence of the scrubbing
of techniques. It is recommended that the nitrogen be scrubbed efficiency on the acid concentration. Note that several com-
with 0.1 M chromous chloride in 2.4 M Hel containing mercial systems are available for removing oxygen at room
amalgamated zinc 18 with a 0.8 to 3.2-mm pore size 19 or be temperature 20 or at high temperatures. 21
scrubbed with vanadous chloride, 18 which is a less suitable
20 One suitable system is available from Applied Science Laboratories, State
College, PA. It removes oxygen at room temperature with a Dow gas purifier
18 Meites, L., Polarographic Techniques, 2nd edition, Interscience Publishers, preceded by a Hydro-Purge Unit.
New York, NY, 1967, pp. 89-90. 21 Hewlett-Packard, Avondale, PA, Model 19046A gas purifier uses a furnace at
19 Amalgamated zinc with a pore size of 0.8 to 3.2 mm for a Jones reductor 475°C that is packed with copper, and SupeIco (Bellefonte, PAl Model 02-2315 gas
(Fisher Scientific Co., Fairlawn, NJ) has been found satisfactory. purifier uses a furnace at 600°C containing a special catalytic converter.
X3. CHEMICAL MATRIX MODIFIERS FOR LEAD DETERMINATIONS USING GRAPHITE FURNACE
SPECTROPHOTOMETRY
X3.l The determination of lead using graphite furnace used. Table X3.1 is a summary of the most commonly used
spectrophotometry may require the use of a chemical matrix modifiers.
modifier. Numerous chemical modifiers have been reportedly
X4. VOLTAMMOGRAM
X4.l The voltammogram shown in Fig. X4.1 gives typical droplet with a 2.9-mm2 area for 2 min with stirring plus 30 s
stripping curve shapes, peak potentials, and sensitivities (in [.LA without stirring.
per 5 [.LglL) for cadmium and lead deposited into a mercury
A = BLANK CURVE
B = SAMPLE CURVE
i
SIGNAL
o..lo.lj.lA/5j.1g/L Cd /0050~A/5~9/L Pb
SUMMARY OF CHANGES
This section identifies the location of selected changes to these test methods that have been incorporated since
the last issue. For the convenience of the user, Committee D19 has highlighted those changes that may impact
the use of these test methods. This section may also include descriptions of the changes or reasons for the
changes, or both.
(1) The QC sections were added to the test method. (2) Sections 15.4,25.4,37.4, and 47.3 were added.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed evel}' five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).