Page 1
DIGITAL R.F. PROBE INSTRUCTIONS
                                                  INDEX
                                                                                               Page
1. GENERAL DESCRIPTION ...........................................................              2
2. PRINCIPLE OF OPERATION .......................................................               2
3. SPECIFICATION ...........................................................................    3
         3.1. Digitrans probe                                                                   3
         3.2 Digitrans controller - model C2C                                                   4
4. INSTALLATION ...........................................................................     4
         4.1 Probe installation                                                                 4
         4.2 Controller installation                                                            4
         4.3 Controller terminations                                                            5
5. OPERATION OF PROBE AND CONTROLLER ........................                                   6
         5.1   The keypad                                                                       6
         5.2   Operation                                                                        6
         5.3   View or set parameters                                                           7
         5.4   Calibration options                                                              8
         5.5   Display mode                                                                     8
         5.6   Switch 4 / 20 mA                                                                 8
         5.7   Set Access code                                                                  8
         5.8   Test Watch dog                                                                   8
6. COMMISSIONING .....................................................................          9
         6.1 Power up                                                                           9
         6.2 Check probe operation                                                              9
         6.3 Loop testing                                                                      10
7. CALIBRATION ........................................................................        10
         7.1 Simple 2 point calibration                                                        10
         7.2 Automatic calibration using regression                                            11
         7.3 Calibrator                                                                        14
8. FAULT DIAGNOSIS .................................................................           16
                                  Page 2
1. GENERAL DESCRIPTION
  The DIGITRANS DIGITAL R.F. Probe consist of a probe (electrode) fitted with a
  head containing the measurement electronics which is connected to a second box
  (the controller), via a single co-axial cable.
  The controller contains a power supply, a microprocessor, keypad and display,
  A/D and D/A converters, and is also used to supply power to the probe, control
  the probe, receive signals from the probe, calculate brix from these
  measurements, and provide digital indication plus 4-20 mA output signals for
  connection into PID controllers or DCS.
  One control box can operate two probes in multiplex. In addition, for each probe
  there are two analog outputs available, and an analog input suitable for
  connecting a 2 wire temperature transmitter (one for each probe).
  R.F. readings may therefore be corrected for temperature if required.
2. PRINCIPLE OF OPERATION
  The instrument consists basically of a variable frequency oscillator built into the
  probe head which is coupled to the probe electrodes. The oscillator is switched to
  operate at two different frequencies separated by about 10 MHz and the probe is
  also switched in and out of circuit by a second switch to provide two measurement
  and two reference frequencies.
  The probe is controlled by a microprocessor controller unit which operates the
  switches in the probe head. Four output frequencies (two measurement and two
  reference), are then measured in sequence over about a 3,8 second period.
  These frequencies are designated Fa, Fb, Fc, and Fd. Fa and Fd are reference
  frequencies with the probe disconnected from the oscillator, and Fb and Fc
  measurement frequencies with the probe in circuit. The differences between the
  two measurement and reference frequencies Fb-Fa and Fc-Fd are calculated and
  scaled and designated F1 and F2. These differences are then used for calculating
  brix.
  F1 and F2 are each affected differently by massecuite resistance and capacitance
  and these signals are combined in a mathematical formula to derive an optimum
  measurement signal calibrated in brix units.
  In order to calibrate the instrument, a series of measurements of F1 and F2 are
  taken for different known brixes and brix correlated against the measurements
  using a regression technique. The constants derived from this correlation are then
  used by the microprocessor to calculate brix from F1 and F2 when in the
  measurement mode.
  For each probe there are two 4-20 mA analog outputs. These outputs (1 & 2) may
  each be used for different calibrations (eg one for brix and the other for ash), or
  for calibrating a probe for use on two different products.
                                     Page 3
   The formula used to derive probe 1 output 1 is given below:-
   Output = K3 + K4*F1 + K5*F2 + K6*F12 + K7*F23 + K8*F13 + K9*F23 + K10*T
   K3 to K10 are constants stored in memory which are inputted during calibration.
   T is temperature measured by a temperature probe (Range 0 - 100 C).
   The above formula gives the output in engineering units. The 4-20 mA output may
   be spanned to operate over any range by setting parameters K1 (zero) and K2
   (full scale)
   Parameters K1 to K10 can be entered via the keypad, but alternatively constants
   K3 to K9 may be automatically calculated and stored by the microprocessor when
   a calibration procedure is carried out.
   For Probe 1 output 2, the corresponding constants are K11 to K20, and for probe
   2 the constants for output 1 and 2 are K21 to K30, and K31 to K40 respectively.
3. SPECIFICATION :
3.1 Digitrans Probe
   Probe length overall          : Model T2      422 mm
                                 : Model T4      507 mm
                                 : Model T3      577 mm
   Probe length into pan         : Model T2      165 mm
                                 : Model T4      250 mm
                                 : Model T3      320 mm
   Probe diameter                : 26 mm
   Probe body material           :   316 stainless steel
   Connection into pan           :   Adaptor flange/bushing
   Probe body temperature        :   Max 120 deg.C
   Probe head temperature        :   Max 70 deg.C
   Connection                    :   N type coaxial socket
   Cable to controller           :   RG 58 C/U 50 ohm coaxial
   Power consumption             :   Typically 55 ma 16-25 V DC.
   Protection class              :   IP 65 enclosure.
                                     PCB encapsulated in epoxy for added protection.
   Measurement frequencies       : approx. 18 and 27 MHz
                                  Page 4
3.2 Digitrans Controller - Model C2C
   Power requirement             : 24 V DC 300 mA.
   Probe inputs                  : 2 off N type coaxial sockets
   Program                       : EPROM
   Parameters                    : RAM battery back up
   Display                       : 4 line alpha numeric illuminated.
   Keypad                        : 16 keys - access through password.
   Temperature                   : 50 deg. C max.
   Outputs                       : 4 x 4-20 mA into 500 ohms
   Inputs                        : 2 x 4-20 mA two wire self powered.
   Isolation                     : Power supply and 4-20- mA signals galvanically
                                    isolated from probes
   Terminations                  : Terminals in separate compartment to electronics.
   Protection class              : IP 65
4. INSTALLATION
4.1 Probe installation
   The probe is push fitted into a housing which is mounted in the bottom of a pan or
   on a pipe where liqour brixes are measured. The housing consists ot two parts, an
   adaptor flange which is welded into the vessel, and a bronze inner housing which
   bolts onto the adaptor flange. The probe is fitted into the bronze housing and is
   secured by a single locking bolt.
   A dummy plug is recommended for use when a probe is removed for cleaning or
   maintenace. Details of the probe housing and plug are given in the instruction
   manual to allow fabrication by the user. Alternatively the housing and plug may be
   ordered as optional extras.
4.2 Controller Installation
   The controller box requires a 24 volt DC supply. The negative of the supply is
   common with the 0 volts of the 4-20 mA outputs but is galvanically isolated from
   the probe circuits.
                                     Page 5
4.3 Controller Terminations
   The terminals in the termination box are designated as follows:
           Terminal                Description
              1            + 24 V supply
              2            Negative supply
              3            Screen
              4            Screen
              5            + Temp. transmitter probe 1
              6            - Temp. transmitter probe 1
              7            + Temp. transmitter probe 2
              8            - Temp. transmitter probe 2
              9            + 4-20 ma output 1 probe 1
             10            - output
             11            + 4-20 mA output 2 probe 1
             12            - output
             13            + 4-20 mA output 1 probe 2
             14            - output
             15            + 4-20 mA output 2 probe 2
             16            - output
   The coaxial fittings on the side of the terminal box are as follows:
             Right         Probe 1
             Left          Probe 2
   Screened cable must be used to connect analog outputs to the controllers or
   DCS. The screens may be connected to terminals 3, and / or 4.
   If temperature transmitters are installed, these must be galvanically isolated from
   earth. Screened cable must be used and the cable screens terminated only at the
   controller (not at the transmitter). These screens should be connected to screens
   of cables back to contol room.
   The coaxial cables connecting the probes to the controller must be of good quality
   with an impedance of 50 ohms (MIL spec. RG 58) and may be any length up to
   70 m. Although cable lengths up to 100 m may be used, it is advisable to use a
   coaxial cable with a lower loss and DC resistance (eg RG 8) in which case lengths
   greater than 100 m are possible.
   After installation of the coaxial fittings, the cables should be checked prior to
   connection. Check for continuity (resistance) and for short circuits between centre
   pin and screen. Resistance through the cable (with one end shorted) should be
   less than 10 ohms.
                                  Page 6
  The probe is supplied with 16-25 volt DC through the cable. Current drawn is
  about 55 mA. There is short circuit protection and current is limited to prevent
  damage.
5. OPERATION OF PROBE AND CONTROLLER
  On power up, the microprocessor displays "Warm start" followed by "on line"
  Thereafter the pre-selected probe will be selected and the following displayed :
            PROBE 1 TEMP. ...
            OUTPUT 1 = ....
            OUTPUT 2 = ....
  This display is updated every 3,8 seconds. If there are no calibration constants, or
  the probe is disconnected, then output 1 and 2 will be zero.
  If there is no temperature transmitter connected, TEMP will also display zero.
5.1 The Keypad
  The probe is calibrated and the mode of operation set via the keypad. Keys are
  allocated as follows:
     A:     When in the run mode, pressing A allows access via a password to the
            menu used for calibration or checking. The default password is 0 and
            this may be changed by the user.
     B:     This is used to back track one digit when entering parameters, or to
            backtrack parameters when viewing data.
     ENT: This is the ENTER key and is pressed after entering data or to advance
          to the next parameter.
     F:     This is used to return to the run mode and is active only when not in the
            run mode.
     In addition there are keys for digits 0-9, negative (-) and decimal point.
5.2 Operation
  On power up, after initialising, the probe will go into the measurement mode and
  the following will be displayed:-
            PROBE 1 TEMP 0 C
            OUTPUT 1 = ....
            OUTPUT 2 = ....
  The outputs will indicate zero unless there are calibration parameters in memory.
                                    Page 7
   The temperature will display 0 unless there is a temperature transmitter
   connected.
   On pressing key A, "ENTER ACCESS CODE" will be displayed.
   Key in the acess code (default 0) followed by ENT. The following menu will be
   displayed:
             VIEW / SET VARS.            1
             CALIBRATE                   2
             DISPLAY MODE                3
             NEXT PAGE                          4
   On selecting 4 the following additional menu will be displayed:
             SWITCH 4-20 mA              1
             SET ACC. CODE               2
             TEST W. DOG                 3
             FINISHED                    4
5.3 View or set parameters
   This option allows mode of operation to be changed or calibration parameters to
   be entered or viewed.
   On selecting this option the following is displayed:
             PROBE         1
             PROBE         2
             BOTH          3
   Select the option depending on which probe is required.
   If 3 is selected the unit automatically returns to the measurement mode and both
   probes are in operation in multiplex. If 1 or 2 is selected followed by F, the unit
   also returns to the measurement mode with the selected probe in operation.
   Alternatively, after probe 1 or 2 is selected and the following appears:
             OUTPUT 1 OR 2 ?
   When the required output is selected, calibration constants K .. are displayed. The
   next parameter may be viewed by pressing ENT. If B is depressed, the previous
   parameter is displayed.
   New calibration parameters may be keyed in and entered if required.
   Pressing F returns to measurement mode.
                                   Page 8
5.4 Calibration options
   On selecting this option, a further menu allows selection of the required probe and
   output. The following new menu then appears:
             NEW CALIBRATION             1
             ADD TO CALIBRATION          2
             ENTER ANALYSIS              3
             REGRESSION                  4
   These will be used for automatic calibration and will be described later.
   If option 1 is selected, all previous calibration data is erased but not calibration
   constants (K values).
   If option 1 or 2 is selected, then the frequency differences F1 and F2 will be
   displayed. Pressing F returns to the measurement mode.
5.5 Display mode
   When this option is selected display is changed to display the four frequencies. To
   return to the normal display, this option must be again selected by pressing A etc.
5.6 Switch 4-20 mA
   When this option is selected, all outputs cycle from 4 to 20 mA maintaining each
   output for 5 seconds. This option may be used for loop testing.
   Pressing F will terminate this option.
5.7 Set access code
   When this option is selected, a new acess code may be entered. This may be any
   number of up to 8 digits.
   This code must be recorded since if forgotten access to the unit will only be
   possible by using a master code.
   Initially during calibration, it is advisable to use a simple number (eg 0) and only
   enter the final code once calibration is complete.
5.8 Test watch dog
   This option is used to check operation of the watchdog timer.
   The function of the watchdog timer is to reset the microprocessor if there is a
   failure due to a glitch in the power supply etc. This prevents the unit from "hanging
   up".
                                   Page 9
6. COMMISSIONING
6.1 Power up
  After completing installation, when power is applied, the display will light up and
  after a short while, the following will be displayed:
            PROBE 1 TEMP 0 C
            OUTPUT 1 = ....
            OUTPUT 2 = ....
  If the display does not light up, check for 24 volts across terminals 1 and 2 of the
  controller terminations. Terminal 1 should be + 24 volts. If polarity is incorrect,
  then it will be necessary to replace the fuse in the controller after polarity is
  corrected. This is done by removing the controller cover. This should be replaced
  by a 0,5 amp fuse.
  If the display lights up but is blank, check supply voltage.
  After sucessful power up, continue as follows:
6.2 Check probe operation
     i)     Press A, enter the acess code (default 0) , and select CALIBRATE.
            The display will prompt :-
            PROBE 1 OR 2 ?
     ii)    Press 1 to select probe 1. The following will now be displayed:-
            OUTPUT 1 OR 2 ?
     iii)   The display will now give the following options:
            NEW CALIBRATION             1
            ADD TO CALIBRATION 2
            ENTER ANALYSIS              3
            REGRESSION                  4
     iv)    Press 1. The display will now give:-
            SAMPLE 1
            F1 = ......
            F2 = ......
  Readings of F1 and F2 (between 1 and 20) will be shown if the probe is operating
  correctly. If no readings are obtained, check connections to probe.
                                     Page 10
   If probe 1 is operating correctly, press F to return to the measurement mode and ,
   if there are two probes installed, repeat selecting probe 2.
6.3 Loop testing
   Press A, enter the acess code (default 0), and select NEXT PAGE ( Key 4) from
   menu, then select SWITCH 4/20 mA (Key 1).
   All outputs will switch alternatively between 4 and 20 mA. The signals can then
   be tested at the control room or controller. When complete select FINISHED
   (Key 4).
7. CALIBRATION
   Two calibration procedures are available, a simple 2 point calibration based on F1
   or F2 measurements which will provide a useful measurement signal (output
   0-100%) without requiring laboratory analyses, and a more accurate regression
   calibration requiring laboratory analyses which will provide optimum combination
   of F1 and F2 signals, and provide a linearised readout in engineering units.
7.1 Simple 2 point calibration
   The F1 or F2 signal will normally correlate to brix. Experience has shown that in
   most applications the F2 signal provides a better correlation to brix than F1.
   The method for setting up the probe is as follows:
   Go into the calibration mode, select the required probe (PROBE1), select
   “OUTPUT 1”, and then select “NEW CALIBRATION”.
   F1 and F2 signals will then be displayed.
   Reduce brix or solids content in the process to the lowest value required and
   record F1 and F2 values. These values are designated F1L and F2L.
   Increase brix or soloids content to the highest value required, and then record F1
   and F2 values. These are designated F1H and F2H.
   If the F2 signal is to be used, calculate the following :
      K3 =    100 F2L / (F2L-F2H)
      K5 = - 100 / (F2L-F2H)
   These parameters are entered into the controller as follows:
   Press A and select 1 (VIEW / SET VARS). The following will then be displayed:-
             PROBE          1
             PROBE          2
                                     Page 11
             BOTH           3
   Press 1. The following is displayed :
             OUTPUT 1 OR 2 ?
   Press 1. Constant K1 = O is now displayed.
   Press E. Constant K2 = 100 is now displayed.
   These constants represent zero and span for the 4-20 mA output signal.
   Press E and K3 is then displayed. Enter the value for K3 as calculated above and
   then press ENT. K4 is then displayed. This should be zero. Press ENT again abd
   K5 is displayed. Enter the value for K5 calculated above and press ENT. Check
   that contants K6 to K10 are zero and then press F. The probe is now calibrated
   and will provide a readout of 0 -100 corresponding to 4-20 mA output.
   Alternatively if the F1 signal is to be used, calculate the following :
             K3 =    100 F1L / (F1L-F1H)
             K4 = - 100 / (F1L-F1H)
   These values are then entered as described above.
   Note the probe output is determined by:
   OUTPUT = K3 + K4*F1 + K5*F2 + K6*F12 + K7*F22 + K8*F13 + K9*F23 + K10*T
7.2 Automatic calibration using regression
   The probe can be calibrated against brix or solids content measured in the
   laboratory of a number of samples taken from the process. It is important that
   these samples cover as wide range of brix values as possible and are not all
   around the same brix value.
   The procedure is as follows:
7.2.1 Taking samples
   Go into the calibration mode as described before.
   After a short while the following is displayed:
             CALIB. PROBE 1 OR 2 ?
   Press 1 or 2 to select the probe to be calibrated and the next prompt is :
             OUTPUT 1 OR 2.
                                Page 12
Each probe has two analog outputs. Normally only one needs to be used. Press
1 or 2 to select the output to be calibrated and then the following options will be
displayed:
          NEW CALIBRATION            1
          ADD TO CALIB.              2
          ENTER ANALYSES             3
          REGRESSION                 4
Option 1 is for a complete new calibration where a new set of samples are to be
taken. Pressing 1 erases all previous calibration data from memory.
Option 2 allows additional calibration data to be added to the data bank.
Assuming a new calibration is to be done, enter "1".
The following is then displayed:
          SAMPLE 1
          FI = ...
          F2 = ...
F1 and F2 are the two frequency differences. These frequency differences are
used in the multi- linear regression.
Sample 1 is taken from the pan and at the same time the E key pressed. This
then stores F1 and F2 against sample 1.
          SAMPLE 2 is then displayed.
The pan conditions are changed to give a different brix and this will give different
values of F1 and F2. Take a second sample and press "E".
          SAMPLE 3 is then displayed.
By pressing "B" one can go back and resample if a mistake has been made.
Pressing "A" moves forward without changing the previous F1 and F2 values
stored.
Between samples, it may be desirable to go back on line so that the pan can be
controlled and the set point changed to vary brix. This is accomplished by
pressing F.
When returning to calibrate to take a sample, select "ADD TO CALIB." which will
allow new data to be added to the stored data.
At least 8 samples (up to 20) must be taken to calibrate a probe. The
microprocessor will not carry out an automatic calibration unless there are 8 or
more samples entered.
                                      Page 13
7.2.2 Entering of laboratory data
   Send the samples to the laboratory for analysis and when the results are
   obtained, go into CALIBRATE mode, select PROBE NO., and OUTPUT as
   before, then select " ADD ANALYSES" by pressing 3.
   The following is then displayed:
             SAMPLE 1 0
             F1 = ...
             F2 = ...
   Key in the laboratory brix for sample 1 and then press "E".
   Pressing "B" will backtrack if an error is made.
   The data will be stored and SAMPLE 2 ... then displayed.
   When all analyses have been keyed in, press "B" to backtrack if the data is to be
   checked, otherwise press "F" to go back on line.
   It is desirable to write down the stored values of F1 and F2 and the analysis for
   each sample for future reference.
   If a set of data (for example sample 5) is wrong, it can either be overwritten by
   going to add to calibration and taking a new sample and then keying in the
   analysis, or in the ADD analysis mode, the data may be erased by pressing A. In
   this case, sample 6 becomes sample 5 etc.
7.2.3 Regression
   The microprocessor can now be used to automatically calibrate the probe in brix
   units as follows:-
   Go into the calibration mode, select probe and output as before and then select
   "REGRESSION" by pressing 4.
   The microprocessor then checks the data for completion and may come up with
   the following messages:
                ERROR IN DATA
             or INSUFFICIENT DATA
   There must be at least 8 sets of data to be accepted.
   If ERROR IN DATA is displayed press "F" to return on line and then go to ADD
   ANALYSES. Check the data by pressing "E" to go forward, or "B" to backtrack.
   Correct the error. There must be an analysis for each set of frequencies and vice
   versa.
                                     Page 14
   If the data was acceptable, the microprocessor then does two separate
   regressions between F1 and F2, and the analyses.
   During calculation the following messages are displayed:-
             CALCULATING
             REGR. No 1
             MATRIX COMPLETE
             REGR. No 2
             MATRIX COMPLETE
   When calculation is complete, the standard errors of the two regressions are
   displayed and the regression chosen (one with the lowest standard error)
   The standard error is a good indication of the accuracy that may be expected in
   that application.
   If the results are acceptable, press "E" and the calculated regression constants
   (for example K3 to K9 for probe 1, output 1) go into memory and are then used to
   calculate brix in the measurement mode. The probe then goes back on line.
   If F is pressed, the results are ignored and the probe goes back on line.
   The various regression constants are as follows:
             Probe 1 output 1   ...........   K3 to K9
             Probe 1 output 2   ...........   K13 to K19
             Probe 2 output 1   ...........   K23 to K29
             Probe 2 output 2   ...........   K33 to K39
   On completion of calibration, the K constants should be recorded for future
   reference. These constants are applicable for the probe calibrated (record probe
   serial no.) and may be entered into another microprocessor to exactly duplicate
   the calibration. Note that this calibration is only valid for the probe used during the
   calibration.
7.3 Calibrator
   A calibrator consisting of a number of resistors and capacitors which may be
   switched in parallel across the probe electrodes to simulate product impedance is
   provided. There are 4 various types of calibrator available, the CA1 which is used
   for Raw Sugar massecuites, the CA2 calibrator used for Refinery White boilings,
   the CA3 which is used for Syrup and Remelt, and the CAX universal calibrator
   which covers all the ranges and can be used on all products.
   The CAX calibrator has four binary switches. The first two set the resistance
   value, and the second two the capacitance value. Each switch has 15 positions ie
                                    Page 15
   0, 1, 2, 3, 4, 5. 6, 7, 8, 9, 10, A, B, C, D, E, and F. As the swich is increased,
   resistance or capacitance increases.
   The first switch on the left gives high resistance values, going from 680 ohms for
   position 1, reducing down to 42,8 ohms for position F. The second switch gives
   resistance values from 39 ohms down to 3,6 ohms.
   The third switch sets capacitance from 5,6 pF to 48,5 pF. The fourth switch is only
   used from position 1 to 7 and provides capacitances from 100 pF to 680 pF.
   By switching in various combinations, a large range of different resistance and
   capacitance values can be simulated.
7.3.1 Using the calibrator
   Once a probe calibration has been carried out using the regression, record the
   values of F1 F2 and brix used in the regression. (These can be read in the
   calibrate mode by going to " ADD ANALYSES ").
   Plot F1 vs F2 on a graph.
   This will give roughly a curve but with scatter.
   Remove the probe from the pan, fit the calibrator, then go into the calibrate mode.
   Go to "ADD TO CALIB." and F1 and F2 will be displayed
   Now we require 10 calibration points with F1 and F2 values similar to those
   obtained during calibration. These must be scattered around the calibration data
   on the graph to cover a similar range of F1 and F2.
   Adjust the switch positions to give a calibration point as above.
   Note the switch position for this calibration point and mark F1 and F2
   corresponding to this calibration point on the graph.
   Adjust the switch positions to give another calibration point and repeat until 10
   points have been obtained.
   Now go back into the measurement mode and set the switch positions to each of
   the calibration points chosen.
   Write down the switch positions and the brix reading for each of the 10 calibrator
   settings, for example using a CA1 calibrator, one of these may be as follows:
      1   2 3 4 5 6 7 8 9 10                          Brix
       0 1 0 1 1 0 1 1 0 0                            90.3
   Alternatively, if the wide range calibrator type CAX is used, the readings may be
   as follows:
      1 2 3 4                                         Brix
      5 A 3 1                                         90.3
                                   Page 16
   These 10 settings and corresponding brixes can now be used for recalibration of
   any other probe using the regression procedure.
   Note that this calibration is only valid in the pan in which original calibration is
   carried out. If the probe is used in a different pan having a different physical
   configuration (eg calandria closer to probe) then calibration may be different.
   The calibration curve may be elevated or suppressed by adjusting parameter K3
   (for probe 1 output 1) etc.
8. FAULT DIAGNOSIS
   1. No power-     Check for power at terminals 1 and 2 and if there is power check
                    polarity. Wrong polarity will blow a fuse on the main controller
                    PCB. To replace undo the screws on the main controller panel
                    and carefully remove the front panel. Check and replace with 0,5
                    amp fuse if blown.
   2. Unit working but no reading or erratic readings from probe.
                    Go into calibration mode "ADD TO CALIB".
                    Check if FI and F2 displays a reading. If no response, unplug
                    the probe cable from the controller and measure volts across the
                    plug. The voltage should vary from 16 to 25 volts.
                    If the voltage is present, check the probe and cable.
   4. Unit working but keyboard faulty.
                    Check cable from keyboard to plug in board.
   5. Display faulty.
                    Check cable from display to plug in board.
   6. Loss of parameters on power down.
                    This indicates failure of the RAM battery back up and may occur
                    after 5 to 10 years of operation. The battery is soldered onto the
                     mother board and may be checked with a voltmeter. The
                    voltage should be greater than 3 volts ( new battery 3,6V ) This
                    may be unsoldered from the top of the board if faulty and
                    replaced with a new unit by a qualified technician.
Revision 6   3- 7- 03