Page Page 1 of 2
Format No.: QC047/F02 Version No.: 00 Effective Date:
HPLC CALIBRATION CERTIFICATE
Instrument Name
Instrument Model No.
Instrument ID
Analyst
Date of Calibration
Next Calibration due date
S. Description of the
Acceptance Criteria Observation Remarks
No. calibration parameter
Calibration of flow rate
1.
Port A B C D ---
0.49 mL/min to 0.51
0.5 mL/ min
mL/min.
0.98 mL/min to 1.02
1.0 mL/min
mL/min.
1.96 mL/min to 2.04
2.0 mL/min
mL/min.
Calibration of GPV
2.
A/B/C A/B/D ---
The relative height of
19.0 to 21.0 %
peak
The measure value
Injection volume should be between ± 1
3.
accuracy µl of the injection
volume 50 µl.
4. Injector linearity
Injection volume (µL) 5 10 20 50 100 ---
%RSD of area NMT 1.0
%RSD of RT NMT 0.5
Correlation
coefficient NLT 0.999
S. Description of the Acceptance Criteria Observation Remarks
Format No. QA001/F02
Page Page 2 of 2
Format No.: QC047/F02 Version No.: 00 Effective Date:
No. calibration parameter
The auto sampler carry
Auto sampler over should not be more
5.
carryover than 0.1%.
Calibration of Column Heater Temperature
6.
Temperature
The observed
30° C
temperature should be
between ± 2°C of the set
70° C temperature
Calibration of Sample tray chiller temperature
7.
Temperature
The observed
5° C
temperature should be
between ± 3°C of the set
15° C temperature
8. Calibration of Refractive Index Detector
Detector Linearity Correlation coefficient
NLT 0.99
9. Calibration of Photo Diode Array Detector (PDA):
The area should be
minimum at 244 ± 2 nm
9.1 Wavelength accuracy
and maximum at 205±
2nm, 272 ± 2 nm.
Detector Linearity Correlation coefficient
9.2
NLT 0.99
Calibration complies/does not complies
Calibrated By Checked By Approved By
Format No. QA001/F02