0% found this document useful (0 votes)
130 views2 pages

Iron-Cab Au400

The document provides instructions for using an iron liquicolor test for the quantitative determination of iron. It details information about the test including recommended controls, calibration procedures, control frequency, test parameters, and specific calibration details.

Uploaded by

Nghi Nguyen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
130 views2 pages

Iron-Cab Au400

The document provides instructions for using an iron liquicolor test for the quantitative determination of iron. It details information about the test including recommended controls, calibration procedures, control frequency, test parameters, and specific calibration details.

Uploaded by

Nghi Nguyen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

I liquicolor

Colorimetric test for the


quantitative determination of iron

AU 400
NOTE:
RECOMMENDATION TO APPLY THE REAGENTS ON THE ANALYZER.
THE LABORATORY HAS TO VALIDATE THE SETTINGS IN ITS OWN RESPONSIBILITY.
HUMAN WILL ACCEPT NO RESPONSIBILITY FOR THE CORRECTNESS OF THIS APPLICATION AS WE HAVE
NO CONTROL ON SOFT- AND HARDWARE MODIFICATIONS OF THE RESPECTIVE ANALYZER.

Package Size
[REF] 10229 2 x 30 ml IRON liquicolor Complete kit
10230 2 x 100 ml IRON liquicolor Complete kit
13160 4 x 5 ml AUTOCAL
Recommended control serum
13511 6 x 5 ml HumaTrol N
13512 6 x 5 ml HumaTrol P
13951 6 x 5 ml SERODOS
13151 6 x 5 ml SERODOSplus
[IVD]
R1 = [RGT], ready for use
Standard = [STD], ready for use
¼ For detailed information kindly refer to the most recent package insert.

Calibration
Calibrator: IRON Standard, included in Complete kit, or AUTOCAL

Calibration frequency is user defined. Re-calibration is recommended:


• following quality control procedures
• after reagent bottle change
• major preventive maintenance was performed on the analyser or a critical part was replaced.
Control frequency
It is recommended that 2 levels of control be assayed daily.
Setting Sheet Footnotes
# User defined
† Calibrator concentrations: see Calibration
¤ Analyser default values

SU-fe Human Gesellschaft für Biochemica und Diagnostica mbH


05-2018-1 Max-Planck-Ring 21 - D-65205 Wiesbaden - Germany
Telefon: +49 6122 9988 0 - Telefax: +49 6122 9988 100
Page 1 of 2 eMail: human@human.de – www.human.de
I liquicolor
Colorimetric test for the
quantitative determination of iron

AU 400
Specific Test Parameters
General LIH ISE Range

Test Name: Fe ∇ < > Type: Serum ∇ Operation: Yes ∇

Sample: Volume 12 µL Dilution 0 µL Pre-Dilution Rate: 1


Reagents R1 Volume 250 µL Dilution 0 µL Min OD Max OD
R2 Volume 0 µL Dilution 0 µL L 0 H 1.5
Reagent OD Limit:
Wavelength: Pri. 660 ∇ Sec. 700 ∇ First L -0.1 First H 1.5
Method: END ∇ Last L -0.1 Last H 1.5
Reaction slope: + ∇ Dynamic Range:
Measuring Point 1: First 0 Last 27 L # H #
Measuring Point 2: First Last Correlation Factor:
Linearity : % A 1 B 0
No Lag Time: ∇ On-board stability period: #

Specific Test Parameters


General LIH ISE Range

Test Name: Fe ∇ < > Type: Serum ∇

Value/Flag: # ∇ Level L: # Level H: #


Normal Ranges: Age L Age H
Sex Year Month Year Month L H
❏ 1. # ∇ # # # # # #
❏ 2. # ∇ # # # # # #
❏ 3. # ∇ # # # # # #
❏ 4. # ∇ # # # # # #
❏ 5. # ∇ # # # # # #
❏ 6. # ∇ # # # # # #
7. Normal Ranges # #
8. Out of Range 1-6 L H # #
Panic Value: # # Unit: µg/dl Decimal places: 1

Calibration Specific
General ISE

Test Name: Fe ∇ < > Type: Serum ∇

Calibration Type: AB ∇ Formula: Y = AX + B ∇ Counts: # Process: CONC ∇

Cal. No. OD CONC Factor/OD-L Factor/OD-H


Point 1: # ¤ † ¤ ¤
Point 2:
Point 3:
Point 4:
Point 5:
Point 6:
Point 7:
1-Point Cal. Point: ❏ With CONC-0 Extended Calibration: ∇
MB Type Factor: Calibration Stability Period: #

SU-fe Human Gesellschaft für Biochemica und Diagnostica mbH


05-2018-1 Max-Planck-Ring 21 - D-65205 Wiesbaden - Germany
Telefon: +49 6122 9988 0 - Telefax: +49 6122 9988 100
Page 2 of 2 eMail: human@human.de – www.human.de