0% found this document useful (0 votes)
428 views256 pages

Definium 8000 Installation Manual: GE Healthcare

Uploaded by

Alan White
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
428 views256 pages

Definium 8000 Installation Manual: GE Healthcare

Uploaded by

Alan White
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 256

GE Healthcare

Definium 8000 Installation Manual

Direction 5137436-100
Revision 8
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 2
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

LEGAL NOTES

TRADEMARKS
All other products and their name brands are the trademarks of their respective holders.

COPYRIGHTS
All Material, Copyright © 2005 by General Electric Company, Inc. All rights reserved. The material
presented and contained herein may not be reproduced in any form or manner, without the written
permission of General Electric Company, Inc.

Legal Notes Page 3


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 4 Legal Notes


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

IMPORTANT PRECAUTIONS

LANGUAGE

• This Service Manual is available in English only.


• If a customer’s service provider requires a language other than English, it is the customer’s
responsibility to provide translation services.
• Do not attempt to service the equipment unless this service manual has been consulted and
is understood.
• Failure to heed this Warning may result in injury to the service provider, operator or patient
from electric shock, mechanical or other hazards.

• Ce Manuel de maintenance n’est disponible qu’en anglais.


• Si le technicien du client a besoin de ce manuel dans une autre langue que l’anglais, c’est au
client qu’il incombe de le faire traduire.
• Ne pas tenter d’intervention sur les équipements tant que le manuel Service n’a pas été
consulté et compris.
• Le non-respect de cet avertissement peut entraîner chez le technicien, l’opérateur ou le
patient des blessures dues à des dangers électriques, mécaniques ou autres.

• Dieses Kundendienst-Handbuch existiert nur in englischer Sprache.


• Falls ein fremder Kundendienst eine andere Sprache benötigt, ist es Aufgabe des Kunden
für eine entsprechende Übersetzung zu sorgen.
• Versuchen Sie nicht, das Gerät zu reparieren, bevor dieses Kundendienst-Handbuch zu
Rate gezogen und verstanden wurde.
• Wird diese Warnung nicht beachtet, so kann es zu Verletzungen des
Kundendiensttechnikers, des Bedieners oder des Patienten durch elektrische Schläge,
mechanische oder sonstige Gefahren kommen.

• Este Manual de Servicio sólo existe en inglés.


• Si algún proveedor de servicios ajeno a GEHC solicita un idioma que no sea el inglés, es
responsabilidad del cliente ofrecer un servicio de traducción.
• No se deberá dar servicio técnico al equipo, sin haber consultado y comprendido este
manual de servicio.
• La no observancia del presente aviso puede dar lugar a que el proveedor de servicios, el
operador o el paciente sufran lesiones provocadas por causas eléctricas, mecánicas o de
otra naturaleza.

• Este Manual de Assistência Técnica só se encontra disponível em Inglês.


• Se qualquer outro serviço de assistência técnica, quE não a GEHC, solicitar estes manuais
noutro idioma, é da responsabilidade do cliente fornecer os serviços de tradução.
• Não tente reparar o equipamento sem ter consultado e compreendido este Manual de
Assistência Técnica.
• O não cumprimento deste aviso pode por em perigo a segurança do técnico, operador ou
paciente devido a‘ choques elétricos, mecânicos ou outros.

Important Precautions Page 5


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

• Il presente manuale di manutenzione è disponibile soltanto in inglese.


• Se un addetto alla manutenzione esterno alla GEHC richiede il manuale in una lingua
diversa, il cliente è tenuto a provvedere direttamente alla traduzione.
• Si proceda alla manutenzione dell’apparecchiatura solo dopo aver consultato il presente
manuale ed averne compreso il contenuto.
• Non tenere conto della presente avvertenza potrebbe far compiere operazioni da cui derivino
lesioni all’addetto alla manutenzione, all’utilizzatore ed al paziente per folgorazione elettrica,
per urti meccanici od altri rischi.

Page 6 Important Precautions


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, write “Damage
In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (262) 312-1163 or 8*320 1163, or E-mail
@MED Claims-Traffic, immediately after damage is found. At this time be ready to supply name
of carrier, delivery date, consignee name, freight or express bill number, item damaged and extent
of damage.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared
for the equipment shall be performed by licensed electrical contractors. In addition, electrical feeds
into the Power Distribution Unit shall be performed by licensed electrical contractors. Other
connections between pieces of electrical equipment, calibrations and testing shall be performed by
qualified GE GE Healthcare personnel. The products involved (and the accompanying electrical
installations) are highly sophisticated, and special engineering competence is required. In
performing all electrical work on these products, GE will use its own specially trained field
engineers. All of GE’s electrical work on these products will comply with the requirements of the
applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to
perform electrical servicing on the equipment.

IMPORTANT...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Healthcare
Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Healthcare
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEHC iTrak (issue tracking) process to report all omissions, errors,
and defects in this publication.

Important Precautions Page 7


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 8 Important Precautions


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Revision History

Revision Date Reason for change


1 Nov. 07, 2005 Initial release of document.
2 Mar. 10, 2006 Updated X-ray tube wiring drawing. Added loctite to UIF
installation (SPR XRYge63991). Added note to check DSA
detector power supply DIP switch settings. Added wall stand and
table chiller fill procedure. Added wall stand image pasting barrier
label attachment procedure. Added table vertical brake manual
release.
3 May 26, 2006 Page 39, corrected signature text to FE and Apps, added item 21
on checklist to ’collect and process product locator cards’. Added
HHS Compliance Compatibility List to HHS chapter. Added
’Networking and PACS’ section to Electrical Installation flowchart.
Updated general installation procedures.
4 June 6, 2006 CFR 1020.30/31/32 regulatory compliance.
5 June 10, 2007 Updated Final OTS Counterpoise Adjustment procedure to call for
power removal before doing adjustment.
Added procedure to attach caution label to wall stand lateral
support arm.
Updated Electrical Installation flowchart to show correct path to the
Monitor finalization configuration procedure. PQR 13093761
Updated Wallstand Installation chapter to add note that footswitch
cable labelled J4 should be plugged into wallstand bulkhead
connector J3. SPR XRYge67329.
6 June 19, 2007 Updated HHS AEC Minimum Exposure procedure for 4-Cell Ion
Chamber.
7 Aug. 15, 2007 Chapter 15, Section 2.0: Added ’Master Control’ row to Table 15-1.
Chapter 17, Section 4.0: Updated Table 17-1 Magic PC part
numbers.
8 Aug. 22, 2007 Chapter 15, Section 2.0: Added note below Table 15-1.
Chapter 17, Section 4.0: Updated Table 17-1 Magic PC part
numbers again.

Revision History Page 9


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

List of Effected Pages

PAGES REVISION PAGES REVISION


1 through 256 8

Page 10 List of Effected Pages


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Table of Contents
Preface Publication Conventions ........................................................................ 19
Section 1.0 Safety & Hazard Information ....................................................... 19
1.1 Hazard Messages............................................................................................................ 19
1.2 Text Format of Signal Words ........................................................................................... 19
1.3 Symbols and Pictorials Used ........................................................................................... 20
Section 2.0 Publication Conventions ............................................................. 21
2.1 General Paragraph and Character Styles........................................................................ 21
2.2 Page Layout..................................................................................................................... 21
2.3 Computer Screen Output/Input Text Character Styles .................................................... 22
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys) ........................................... 22

Chapter 1 - Safety and Service............................................................................. 23


Section 1.0 Lock-Out & Tag-Out (LOTO)........................................................ 23
1.1 System Power.................................................................................................................. 23
1.2 LOTO Procedure for System Power ................................................................................ 24
Section 2.0 Applying System Power .............................................................. 25
2.1 System Preparation for Energizing Power....................................................................... 25
2.2 Energizing System........................................................................................................... 27
Section 3.0 De-Energizing System ................................................................. 29
Section 4.0 Equipment Safety - Electrostatic Discharge (ESD) .................. 29
4.1 Generating Static ............................................................................................................ 30
4.2 Personal Grounding Methods and Equipment ................................................................ 30
4.3 Grounding the Work Area ............................................................................................... 31
4.4 Recommended Materials and Equipment ....................................................................... 31

Chapter 2 System Installation Guides ................................................................. 33


Section 1.0 Mechanical Installation ................................................................ 33
Section 2.0 Electrical (Config., Calibration, Finalization) ............................. 36
Section 3.0 Installation Handoff Checklists................................................... 39
3.1 Pre-Installation Checklist ................................................................................................. 39
3.2 MI ==> FE Handoff Checklist........................................................................................... 41
3.3 FE ==> Application Handoff Checklist ............................................................................. 42
Section 4.0 Lean Carts 1 & 2 ........................................................................... 43

Chapter 3 Overhead Stationary Rails .................................................................. 45


Section 1.0 Introduction .................................................................................. 45
1.1 Summary Checklist.......................................................................................................... 45
1.2 Before You Begin............................................................................................................. 45
1.3 Special Tools and Materials............................................................................................. 46
1.4 Parts Shortages and Transportation Damage ................................................................. 46

Table of Contents Page 11


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Section 2.0 Inboard Stationary Rails .............................................................. 47
2.1 Room Orientation ............................................................................................................ 47
2.2 Inboard Stationary Rail Length Modifications.................................................................. 47
2.3 Rail Leveling Procedures (using LASER alignment tool) ................................................ 48
2.4 Rail and Shim Installation................................................................................................ 48
2.5 Longitudinal Anchor Rails ............................................................................................... 50

Chapter 4 Overhead Tube Support (OTS) ........................................................... 53


Section 1.0 Summary Checklist ..................................................................... 53
Section 2.0 Modifying the Bridge Length ...................................................... 54
Section 3.0 Bridge and Carriage .................................................................... 54
3.1 Set Up the Lifting Fixture................................................................................................. 55
3.2 Bridge Mounting Procedure ............................................................................................ 56
3.3 Lateral Drive Hardware Installation ................................................................................. 58
3.4 Bridge Stiffener Plate Installation .................................................................................... 62
3.5 Longitudinal Drive Hardware Installation......................................................................... 62
3.6 Install The Lateral Cable Drape Hardware...................................................................... 67
3.7 Carriage Cable Connections ........................................................................................... 68
3.8 Bridge Anti-Backlash Assemblies ................................................................................... 74
Section 4.0 Longitudinal and Lateral Drive Belt Assemblies ....................... 77
4.1 Longitudinal Belt Installation and Tracking...................................................................... 77
4.2 Wiring the Longitudinal Drive Assembly.......................................................................... 84
4.3 Lateral Drive Belt Alignment and Tensioning .................................................................. 84
Section 5.0 X-ray Tube Covers ........................................................................ 85
Section 6.0 Collimator ...................................................................................... 88
Section 7.0 User Interface (UIF)....................................................................... 92
Section 8.0 Longitudinal Cable Drape ............................................................ 93
Section 9.0 Initial OTS Counterpoise Balance ............................................... 94
Section 10.0 OTS Jumpers and Switches ...................................................... 94

Chapter 5 Table ..................................................................................................... 95


Section 1.0 Summary Checklist ...................................................................... 95
Section 2.0 Unpack Table ............................................................................... 95
Section 3.0 Initial Table and Bridge Alignment............................................ 100
Section 4.0 Install Table ................................................................................. 101
4.1 Drill Anchor Holes ......................................................................................................... 101
4.2 Position Table ............................................................................................................... 102
4.3 Re-check Alignment and Anchors................................................................................. 104
4.4 Adjust Pedal Stabilizing Screws.................................................................................... 104
Section 5.0 Detector and Ion Chamber......................................................... 105
Section 6.0 Chiller........................................................................................... 111
6.1 Table Chiller Mechanical Installation............................................................................. 111

Page 12 Table of Contents


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.2 Fill Table Chiller with Coolant ........................................................................................ 113
Section 7.0 Table Cables ............................................................................... 113
7.1 Routing (Under Table Entry)......................................................................................... 113
7.2 Power Connections....................................................................................................... 114
Section 8.0 Table Top .................................................................................... 115
8.1 Typical Room................................................................................................................. 115
8.2 Small Rooms (Alternate Method)................................................................................... 116

Chapter 6 Wall Stand & Extended Wall Stand .................................................. 119


Section 1.0 Summary Checklist .................................................................... 119
Section 2.0 Prerequisites............................................................................... 119
Section 3.0 Unpack Wall Stand ..................................................................... 120
Section 4.0 Install Wall Stand........................................................................ 128
Section 5.0 Detector and Ion Chamber ........................................................ 131
Section 6.0 Cables.......................................................................................... 138
Section 7.0 Wall Stand Jumpers and Switches ........................................... 140
Section 8.0 Display......................................................................................... 140
8.1 Mechanical Installation .................................................................................................. 140
8.2 Install Display Overlay ................................................................................................... 141
Section 9.0 Set Initial Wall Stand Counter Balance .................................... 141
Section 10.0 Initial Mechanical Alignment ................................................... 142
Section 11.0 Wall Stand Anchors ................................................................. 143
Section 12.0 Final OTS/Tube/WS Alignment................................................ 144
12.1 X-ray Beam to SID Travel Alignment............................................................................ 144
12.2 Detector Perpendicular to X-ray Beam Adjustment ....................................................... 145
12.3 Vertical Tracking Adjustment ......................................................................................... 146

Chapter 7 Digital Support Assembly (DSA) ...................................................... 147


Section 1.0 Summary Checklist .................................................................... 147
Section 2.0 Mechanical Installation ............................................................. 147
Section 3.0 Fill DSA Chiller with Coolant.................................................... 149

Chapter 8 Workstation ........................................................................................ 151


Section 1.0 Summary Checklist .................................................................... 151
Section 2.0 Computer .................................................................................... 152
2.1 HP xw8200 Workstation ................................................................................................ 152
2.2 HP xw8400 Workstation ................................................................................................ 152
Section 3.0 Mouse and Keyboard ................................................................. 153
Section 4.0 RCIM & Handswitch ................................................................... 154
Table of Contents Page 13
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Section 5.0 LCD Monitors .............................................................................. 154
Section 6.0 Bar Code Reader (Purchased Option) ...................................... 155

Chapter 9 Cables ................................................................................................. 157


Section 1.0 Summary Checklist .................................................................... 157
Section 2.0 Routing ........................................................................................ 157
Section 3.0 System Cables ............................................................................ 157
Section 4.0 EMC Compliance ........................................................................ 158
4.1 Ferrite Beads................................................................................................................. 158
4.2 Grounding P-Clamps..................................................................................................... 159

Chapter 10 System Cabinet................................................................................ 161


Section 1.0 Summary Checklist .................................................................... 161
Section 2.0 Procedure .................................................................................... 161

Chapter 11 System Termination Jumpers ........................................................ 169


Section 1.0 Summary Checklist .................................................................... 169
Section 2.0 CAN Bus Terminations............................................................... 169
Section 3.0 RT Bus Terminations.................................................................. 169
3.1 Wall Stand-Only Configuration..................................................................................... 169

Chapter 12 Grid Holder Installation ................................................................... 171

Chapter 13 Image Pasting Barrier (option) ....................................................... 175


Section 1.0 Summary Checklist .................................................................... 175
Section 2.0 Procedure .................................................................................... 175

Chapter 14 Flexi-DT Stretcher (option) ............................................................. 179


Section 1.0 Summary Checklist .................................................................... 179
Section 2.0 Hardware and Test Equipment .................................................. 179
Section 3.0 Floor Preparation........................................................................ 179
Section 4.0 Equipment Evaluation ................................................................ 179
Section 5.0 Flexi-DT Unpacking .................................................................... 180
5.0.1 UNPACKING PROCEDURE ............................................................................ 180
Section 6.0 Table Top Handle........................................................................ 181
Section 7.0 Pin Installation ............................................................................ 181
Section 8.0 Pin Lodging Installation ............................................................. 182

Page 14 Table of Contents


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
8.1 Drilling the Floor............................................................................................................. 182
8.2 Pin Lodging Installation Procedure ................................................................................ 182
Section 9.0 Power Cable................................................................................ 183
Section 10.0 Battery Recharging .................................................................. 183
Section 11.0 Options Installation.................................................................. 184
11.1 Handgrips ...................................................................................................................... 184
11.2 Compression Bands ...................................................................................................... 184
11.3 Lateral Cassette Holder ................................................................................................. 185
11.4 Double Pedals ............................................................................................................... 185
11.5 Remote RF Control........................................................................................................ 185
Section 12.0 Drawings ................................................................................... 186

Chapter 15 Regulatory Tests (HHS)................................................................... 187


Section 1.0 Summary Checklist .................................................................... 187
Section 2.0 HHS Compliance Compatibility List ........................................ 188
Section 3.0 Customer Information................................................................ 188
Section 4.0 Test Equipment .......................................................................... 188
Section 5.0 X-ray Generator Tests................................................................ 189
5.1 Generator Operator Indicators....................................................................................... 189
5.2 Beam Quality ................................................................................................................. 189
5.3 Technic Accuracy and Linearity..................................................................................... 190
5.4 Reproducibility Of Exposure .......................................................................................... 191
5.4.1 Overview........................................................................................................... 191
5.4.2 Procedure ......................................................................................................... 191
> Calculation for Coefficient of Variation (C.V.):............................... 192
5.5 Automatic Exposure Control (AEC) Tests ..................................................................... 193
5.5.1 AEC Minimum Exposure Time.......................................................................... 193
5.5.1.1 Overview ....................................................................................... 193
5.5.1.2 Procedure...................................................................................... 193
5.5.2 AEC Maximum mAs.......................................................................................... 195
5.5.2.1 Overview ....................................................................................... 195
5.5.2.2 Procedure (50kVp) ........................................................................ 195
5.5.2.3 Procedure (40kVp) ........................................................................ 195
Section 6.0 Collimator Tests ......................................................................... 195
6.1 Introduction .................................................................................................................... 195
6.1.1 Scope................................................................................................................ 195
6.2 General Tests ................................................................................................................ 196
6.3 SID Accuracy ................................................................................................................. 196
6.3.1 Overview........................................................................................................... 196
6.3.2 Procedure for Digital Table ............................................................................... 196
6.3.3 Procedure (Digital Wall Stand) ......................................................................... 197
6.4 Collimator Field Light Intensity....................................................................................... 197
6.4.1 Overview........................................................................................................... 197
6.4.2 Procedure ......................................................................................................... 198
6.5 Collimator Light Field to X-ray Field Alignment.............................................................. 198
6.5.1 Overview........................................................................................................... 198

Table of Contents Page 15


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.5.2 Procedure......................................................................................................... 198
6.6 Collimator Field Size Indicator Accuracy....................................................................... 200
6.6.1 Overview .......................................................................................................... 200
6.6.2 How to Measure ............................................................................................... 200
6.6.2.1 Magnification Calculation.............................................................. 200
6.6.3 Procedure......................................................................................................... 200
6.7 Digital Table Mode - Center to Center Alignment Error ................................................ 201
6.7.1 Overview .......................................................................................................... 201
6.7.2 How to Measure ............................................................................................... 201
6.7.3 Procedure......................................................................................................... 201
6.8 Digital Table Mode - Size to Size Error (PBL check) ................................................... 202
6.8.1 Overview .......................................................................................................... 202
6.8.2 How to Measure ............................................................................................... 202
6.8.2.1 Magnification Calculation.............................................................. 203
6.8.3 Procedure......................................................................................................... 203
6.9 Digital Wall Stand Mode - Center to Center Alignment Error ........................................ 204
6.9.1 Overview .......................................................................................................... 204
6.9.2 Procedure......................................................................................................... 204
6.10 Digital Wall Stand Mode - Size to Size Error (PBL check) ............................................ 205
6.10.1 Overview .......................................................................................................... 205
6.10.2 How to Measure ............................................................................................... 205
6.10.2.1 Magnification Calculation.............................................................. 206
6.10.3 Procedure......................................................................................................... 206
Section 7.0 HHS Data Sheets......................................................................... 208
7.1 Customer Information.................................................................................................... 208
7.2 Test Equipment ............................................................................................................. 208
7.3 X-ray Generator Tests................................................................................................... 209
7.3.1 Generator Operator Indicators ......................................................................... 209
7.3.2 Beam Quality.................................................................................................... 209
7.3.3 Technique Accuracy and Linearity ................................................................... 209
7.3.4 Reproducibility of Exposure ............................................................................. 210
7.3.5 AEC.................................................................................................................. 210
7.3.5.1 AEC Minimum Exposure Time...................................................... 210
7.3.5.2 AEC Maximum mAs...................................................................... 210
7.4 Collimator ...................................................................................................................... 210
7.4.1 General Tests................................................................................................... 210
7.4.2 SID Accuracy ................................................................................................... 211
7.4.2.1 SID Accuracy - Table.................................................................... 211
7.4.2.2 SID Accuracy - Wall Stand ........................................................... 211
7.4.3 Collimator Field Light Intensity ......................................................................... 211
7.4.4 Collimator Light Field to X-ray Field Accuracy Check ...................................... 211
7.4.5 Collimator Field Size Indicator Accuracy.......................................................... 211
7.4.6 Table - Center to Center Error ......................................................................... 212
7.4.7 Table - Size to Size Error ................................................................................. 212
7.4.8 Wall Stand - Center to Center Error ................................................................. 212
7.4.9 Wall Stand - Size to Size Error......................................................................... 213

Chapter 16 Final Cover Installation ................................................................... 215


Section 1.0 Summary Checklist .................................................................... 215

Page 16 Table of Contents


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Section 2.0 Table Covers............................................................................... 215
2.1 Table Top Insert Removal ............................................................................................. 216
2.2 Lower Front and Rear Cover Installation ....................................................................... 216
2.3 Lower End Cover Installation......................................................................................... 217
2.4 Upper Front and Rear Cover Installation ....................................................................... 218
2.5 Front and Rear Mid Cover Installation ........................................................................... 220
2.6 Upper End Cover Installation......................................................................................... 221
Section 3.0 OTS Final Counterpoise Adjustment and Trim Covers .......... 222
3.1 Final Counterpoise Adjustment...................................................................................... 222
3.2 Install OTS Trim Covers ................................................................................................ 223
Section 4.0 DSA Cover................................................................................... 224
Section 5.0 Wall Stand and Extended Stand Covers .................................. 226

Chapter 17 Final Tasks ....................................................................................... 241


Section 1.0 Summary Checklist .................................................................... 241
Section 2.0 Database Backup ....................................................................... 241
Section 3.0 Final Administrative Tasks........................................................ 241
Section 4.0 System Rating Plates................................................................. 242

Appendix A........................................................................................................... 247


Section 1.0 Inboard Rail Length Modifications ........................................... 247
Section 2.0 Rail Leveling Procedures (Water-Level or Transit) ................. 248
2.1 Using a Water-Level Kit ................................................................................................. 249
2.2 Using a Transit (Alternate Method)................................................................................ 252
Section 3.0 Modifying the Bridge Length..................................................... 252

Table of Contents Page 17


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 18 Table of Contents


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Preface Publication Conventions


Standardized conventions for representing information is a uniform way of communicating
information to a reader in a consistent manner. Conventions are used so that the reader can easily
recognize the actions or decisions that must be made. There are a number of character and
paragraph styles used in this publication to accomplish this task. Please become familiar with them
before proceeding forward.
It’s important that you read and understand hazard statements, and not just ignore them.

Section 1.0 Safety & Hazard Information


Proper product safety labeling allows a person to safely use or service a product. The format and
style for safety communications reflected in this publication represents the harmonization of IEC/
ISO 3864 and ANSI Z535 standards.
Within this publication, different paragraph and character styles are used to indicated potential
hazards. Paragraph prefixes, such as hazard, caution, danger and warning, are used to identify
important safety information. Text (Hazard) styles are applied to the paragraph contents that are
applicable to each specific safety statement.

1.1 Hazard Messages


Any action that will, could or potentially cause personal injury will be preceded by the safety alert
symbol and an appropriate signal word. The safety alert symbol is the triangle with an exclamation
mark within it. It’s always used next to the signal word to indicate the severity of the hazard.
Together, they are used to indicate a hazard exists.
Signal words describe the severity of possible human injures that may be encountered. The alert
symbol and signal word are placed immediately before any paragraph they affect. Safety
information includes:
1.) Signal Word - The seriousness level of the hazard.
2.) Symbol or Pictorial - The consequence of interaction with the hazard.
3.) Word Message:
a.) The nature of the hazard (i.e. the type of hazard)
b.) How to avoid the hazard.
The safety alert symbol is not used when an action can only cause equipment damage.

1.2 Text Format of Signal Words


DANGER - INDICATES AN IMMINENTLY HAZARDOUS SITUATION WHICH, IF
NOT AVOIDED, WILL RESULT IN DEATH OR SERIOUS INJURY. THIS SIGNAL
WORD IS LIMITED TO THE MOST EXTREME SITUATIONS.
WARNING - INDICATES A POTENTIALLY HAZARDOUS SITUATION WHICH, IF NOT
AVOIDED, COULD RESULT IN DEATH OR SERIOUS INJURY.
Caution - Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury. It may also be used to alert against unsafe practices.
NOTICE - Indicates information or a company policy that relates directly or indirectly to the
safety of personnel or protection of property. This signal word is associated directly with a

Preface Page 19
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
hazard or hazardous situation and is used in place of 'DANGER,' 'WARNING,' or 'CAUTION.'
It can include:
• Destruction of a disk drive
• Potential for internal mechanical damage, such as to a X-ray tube

1.3 Symbols and Pictorials Used


The following Symbols and Pictorials are be used in this publication. These graphical icons
(symbols) may be used to make you aware of specific types of hazards that could possibly cause
harm.

keep_up magnetic biohazard compressgas ppe-hearing

fragile impact corrosive heavyobject ppe-2people

static_elec heat general laser ppe-respitory

keep_dry pinch radiation poisongas ppe-loto

general explosive electrical flammable ppe-eye

torque crush/mechanical tipping Read Manual ppe-gloves

ce instuction poisonmatl entanglement instuction

Page 20 Preface
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 2.0 Publication Conventions


2.1 General Paragraph and Character Styles
Prefixes are used to highlight important non-safety related information. Paragraph prefixes (such
as Purpose, Example, Comment or Note) are used to identify important but non-safety related
information. Text styles are also applied to text within each paragraph modified by the specific
prefix.

EXAMPLES OF PREFIXES USED FOR GENERAL INFORMATION:


Purpose: Introduces and provides meaning as to the information contained within the chapter,
section or subsection (such as used at the beginning this chapter, for example).
Note: Conveys information that should be considered important to the reader.
Example: Used to make the reader aware that the paragraph(s) that follow are examples of information
possibly stated previously.
Comment: Represents “additional” information that may or may not be relevant to your situation.

2.2 Page Layout


Publication Part Number & Revision Number Publication Title

The current section and its title The current chapter and its title
are always shown in the footer of are always shown in the footer of
the left (even) page. the right (odd) page.
An exclamation point in a triangle is used Paragraphs preceeded by a symbol
to indicate important information to the user. (e.g. bullets) contain information that
Paragraphs preceeded by Alphanumeric has no specific order.
characters (e.g. numbers) contain infor-
mation that must be followed in a specific order.

Headers and footers in this publication are designed to allow you to quickly identify your location.
The document part number and revision number appears in every header on every page. Odd
numbered page footers indicate the current chapter, its title and current page number. Even page
footers show the current section and its title, as well as the current page number.

Preface Page 21
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.3 Computer Screen Output/Input Text Character Styles
Within this publication, mono-spaced character styles (fonts) are used to indicate computer text
that’s either screen input and output. Mono-spaced fonts, such courier, are used to indicated text
direction. When you type at your keyboard, you are generating computer input. Occasionally you
will see the math operator “greater-than” and “less-than” symbols used to indicate the start and
finish of variable output. When reading text generated by the computer, you are reading it as
computer generated output. In addition to direction, characters are italicized (e.g. italics) to indicate
information specific to your system or site.
Example: Fixed This paragraph’s font represents computer generated screen “fixed” output.
Output Its output is fixed from the sense that it does not vary from application
to application. It’s the most commonly used style used to indicate
filenames, paths and text that do not change from system to system. The
character style used is a fixed width such as courier.
Example: This paragraph’s font represents computer screen output that is
Variable Output “variable”. It’s used to represent output that varies from application to
application or system to system. Variable output is sometimes found placed
between greater-than and less-than operators for clarification. For
example: <variable_ouput> or <3.45.120.3>. In both cases, the < and >
operators are not part of the actual input.
Example: Fixed This paragraph’s font represents fixed input. It’s computer input that is
Input typed-in via the keyboard. Typed input that does not vary from application
to application or system to system. Fixed text the user is required to
supply as input. For example: cd /usr/3p
Example: This paragraph’s font represents computer input that can vary from
Variable Input application to application or system to system. With variable text, the
user is required to supply system dependent input or information. Variable
input sometimes is placed between greater-than and less-than operators.
For example: <variable_input>. In these cases, the (<>) operators would
be dropped prior to input. For example: ypcat hosts | grep <3.45.120.3>
would be typed into the computer as:
ypcat hosts | grep 3.45.120.3
without the greater-than and less-than operators.

2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys)
Different character styles are used to indicate actions requiring the reader to press either a hard or
soft button, switch or key. Physical hardware, such as buttons and switches, are called hard keys
because they are hard wired or mechanical in nature. A keyboard or on/off switch would be a hard
key. Software or computer generated buttons are called soft keys because they are software
generated. Software driven menu buttons are an example of such keys. Soft and hard keys are
represented differently in this publication.
Example: Hard A power switch ON/OFF or a keyboard key like ENTER is indicated by applying a character style
Keys that uses both over and under-lined bold text that is bold. This is a hard key.
Example: Soft Whereas the computer MENU button that you would click with your mouse or touch with your hand
Keys uses over and under-lined regular text. This is a soft key.

Page 22 Preface
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 1 - Safety and Service

Section 1.0 Lock-Out & Tag-Out (LOTO)


1.1 System Power

Name Of Equipment: Definium 8000 System (Workstation, System Cabinet, Table, Wall
Stand and Overhead Tube Support)

Number Of Locks: One per person, working on the system

Title(s) of Employees Authorized to Perform LOTO: Those trained in Lockout Tagout (LOTO)

Title(s) of Affected Employees and How to Notify: Hospital Personnel; notified by verbal
communication.

Location of Energy Isolating


Energy Source Yes No Magnitude of Energy
Means
Electrical X System’s Main “A1” Power 320 - 480VAC, 3 Phase
Disconnect
Pneumatic X
Hydraulic X
Gas/Water/Steam X
Chemical X
Mechanical Motion X
Gravity X
Springs X
Thermal X
Stored Energy X
Air Under Pressure X
Oil Under Pressure X
Water Under Pressure X
Gas Under Pressure X
Steam X
Other X

Type(s) of Equipment and/or Method(s) Selected to Dissipate or Isolate Stored Energy:

Allow 3 minutes for stored energy to dissipate.

Type(s) of Equipment and/or Method(s) Used to Ensure Disconnection’s:

One Lock & Tag for Main “A1” Disconnect

Chapter 1 - - Safety and Service Page 23


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
1.2 LOTO Procedure for System Power
Name of Equipment:Definium 8000 System (Workstation, System Cabinet, Table, Wall Stand and
Overhead Tube Support)

Step Lockout/Tagout Procedure


1 Prepare for shutdown of equipment by notifying affected personnel working in the area
that lockout/tagout is being performed.
2 Bring system software down. See Section 3.0 on page 29 and return to Step 3
3 Locate main power supply (Service) disconnect feeding System Cabinet.
4 Lock Out / Tag Out the main power disconnect.
5 Using an insulated voltmeter, measure the incoming power connection to the system
cabinet at the circuit breaker.
6 If no voltage is present at the system cabinet’s circuit breaker; place the power switch on
the PDU to the OFF position and service equipment as required.
7 After service has been performed, notify affected personnel that equipment will be re-
energized and LOTO devices are being removed.
8 Assure the safety of the area and that all safety devices and guards have been replaced
9 Return system functions to a preset state.
10 Remove Lock-Out Tag-Out equipment from main disconnect.
11 Locate main power supply (Service) disconnect feeding System Cabinet and re-apply
power.
12 Perform system startup. See Energizing System on page 27.

Page 24 Section 1.0 - Lock-Out & Tag-Out (LOTO)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 2.0 Applying System Power


2.1 System Preparation for Energizing Power

POTENTIAL PERFORM LOTO BEFORE PERFORMING SERVICE OF ANY KIND. See Section 1.0 on
ELECTRICAL page 23.
SHOCK
Follow the steps below to ensure that your system powers up safely, properly and reliably.

Potential for PDU tap settings must match input voltage else there is potential for equipment damage.
Equipment
Damage
1. Open System Cabinet door and verify that the PDU’s tap settings (underneath cover at TB3)
are set correctly for the line voltage being supplied to the system. See Figure 1-1.

Figure 1-1 2293487, PDU Front Panel

2. Disengage all facility (i.e. hospital) E-Stops.

Figure 1-2 Sample Hospital E-Stops

Chapter 1 - - Safety and Service Page 25


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3. Disengage Table E-Stop Button. Turn it 1/4 turn clockwise to release button. It will pop out.
See Figure 1-3.

Figure 1-3 Table E-Stop

4. Set table chiller power switch to the “I” or “ON” position.

Figure 1-4 Table Chiller Power Switch

5. Set WS I/O board power switch to the “ON” position and the WS chiller power switch to the “I”
or “ON” position. See Figure 1-5.

Figure 1-5 WS I/O Board Power Switch & Chiller Power Supply Switch

6. On the PDU’s front panel, set circuit breakers CB1, CB2, CB3, CB4, CB5, CB6 and CB7 to the
“ON” position.

Figure 1-6 PDU Circuit Breakers

Page 26 Section 2.0 - Applying System Power


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Note: Do not use CB8 to Switch the PDU ON or OFF. CB8 in the system cabinet should not be
used to switch power “ON” and “OFF” to the PDU. If this circuit breaker is used, there is a high
probability it will trip when switched to the “ON” position. A high “in-rush” of current causes this
situation.

7.) Switch CB8 and the power switch to the “ON” position. The power switch handle points to the
“I” Position. See Figure 1-7.

Figure 1-7 PDU Circuit Breaker (CB8) & Power Switches Shown in “ON” position.

2.2 Energizing System


System power-up is controlled by both hardware and software. When power is applied, power is
switched “ON” to each of the subsystems through hardware. Software in the subsystems boot and
begin loading their operating systems and applications. Complete system startup takes
approximately 2 minutes.
1.) Prepare the system for energizing, if required. See System Preparation for Energizing Power
on page 25.
2.) Remove tag and lock from facility power. See Section 1.2 on page 24
3.) Press the green POWER ON button on the RCIM to initiate system start-up.

Figure 1-8 RCIM Power On Button

Chapter 1 - - Safety and Service Page 27


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4. The table inhibit switch should not be illuminated, after power is applied. Unless you wish to
inhibit table motion. When the pedals are inhibited, the switch will be illuminated.

Figure 1-9 Table Foot Pedals Inhibit Switch (Old and New styles respectively)

5.) System startup is complete when applications screen is displayed on workstation computer.

Page 28 Section 2.0 - Applying System Power


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 3.0 De-Energizing System


Shut down is controlled by both software and hardware. When a shutdown is initiated, application
and the operating software execution is terminated. While this takes place, signals are sent to the
PDU to shut-off system hardware. Complete shutdown takes approximately 1 minute.
1.) From the application interface, using the mouse, click on the tools icon. This selects the
Service User Interface.

2.) Now click on SHUT DOWN.

3.) Now lock out and tag out the facility power to the system, if you wish to perform service. See
Section 1.2 on page 24.

Section 4.0 Equipment Safety - Electrostatic Discharge (ESD)


A sudden discharge of static electricity from your finger or other conductor can destroy static-
sensitive devices or microcircuitry. Often the spark is neither felt nor heard, but damage occurs. An
electronic device exposed to electrostatic discharge (ESD) might not appear to be affected at all
and can work perfectly throughout a normal cycle. The device can function normally for a while, but
it has been degraded in the internal layers, reducing its life expectancy.
Networks built into many integrated circuits provide some protection, but in many cases, the
discharge contains enough power to alter device parameters or melt silicon junctions.

Chapter 1 - - Safety and Service Page 29


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.1 Generating Static
The following table shows that:
• Different activities generate different amounts of static electricity.
• Static electricity increases as humidity decreases.

Relative Humidity
Event 55% 40% 10%
Walking across carpet 7,500 V 15,000 V 35,000 V
Walking across vinyl floor 3,000 V 5,000 V 12,000 V
Motions of bench worker 400 V 800 V 6,000 V
Removing bubble pack from PCB 7,000 V 20,000 V 26,500 V
Packing PCBs in foam-lined box 5,000 V 11,000 V 21,000 V

Many electronic components are sensitive to ESD. Circuitry design and structure determine the
degree of sensitivity. The following packaging and grounding precautions are necessary to prevent
damage to electric components and accessories.
• Transport products in static-safe containers, such as tubes, bags, or boxes to avoid hand
contact.
• Protect all electrostatic parts and assemblies with conductive or approved containers or
packaging.
• Keep electrostatic sensitive parts in their containers until they arrive at static-free stations.
• Place items on a grounded surface before removing them from their container.
• When handling or touching a sensitive component or assembly, ground yourself by touching
the chassis.
• Avoid contact with pins, leads, or circuitry.
• Place reusable electrostatic-sensitive parts from assemblies in protective packaging or
conductive foam.

4.2 Personal Grounding Methods and Equipment


Use the following equipment to prevent static electricity damage to equipment:
• Wrist straps are flexible straps with a maximum of one-megohm +/- 10% resistance in the
ground cords. To provide a proper ground, wear the strap against bare skin. The ground cord
must be connected and fit snugly into the banana plug connector on the grounding mat or
workstation.
• Heel straps, toe straps, and boot straps can be used at standing workstations and are
compatible with most types of shoes or boots. On conductive floors or dissipative floor mats,
use them on both feet with a maximum of one-megohm +/- 10% resistance between the
operator and ground.
Static Shielding Protection Levels
Method Voltage
Antistatic plastic 1,500
Carbon-loaded plastic 7,500
Metallized laminate 15,000

Page 30 Section 4.0 - Equipment Safety - Electrostatic Discharge (ESD)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.3 Grounding the Work Area
To prevent static damage of components and parts in the work area:
• Cover the work surface with approved static-dissipative material. Provide a wrist strap
connected to the work surface and properly grounded tools and equipment.
• Use static-dissipative mats, foot straps, or air ionizers to give added protection.
• Handle electrostatic sensitive components, parts, and assemblies by the case or PCB
laminate.
• Handle components only at static-free work areas.
• Turn off power and input signals before inserting and removing connectors or test equipment.
• Use fixtures made of static-safe materials when fixtures must directly contact dissipative
surfaces.
• Keep work area free of non-conductive materials, such as ordinary plastic assembly aids and
Styrofoam.

4.4 Recommended Materials and Equipment


Materials and equipment that are recommended for use in preventing static electricity include:
• Antistatic tape, smocks, aprons, or sleeve protectors
• Conductive bins, foam, tabletop workstations with ground cord of one-megohm +/-10%
resistance, and other assembly or soldering aids.
• Static-dissipative table or floor mats with hard tie to ground
• ESD Field service kits
• Static awareness labels
• Wrist straps and footwear straps providing one-megohm +/-10% resistance
• Material handling packages
- Conductive plastic bags, plastic tubes and tote boxes
- Opaque shielding bags
- Transparent metallized shielding bags and shielding tubes

Chapter 1 - - Safety and Service Page 31


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 32 Section 4.0 - Equipment Safety - Electrostatic Discharge (ESD)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 2 System Installation Guides


Installation is broken into two primary tasks: mechanical and electrical. Mechanical Installation tasks are performed by
the mechanical installers (MI) and focus on the physical aspects of hardware assembly, equipment positioning, and
cabling. Electrical tasks are done by the local field engineer (FE) and are performed when system power is
applied.
The primary tasks are designed to be performed in sequence. Follow the sequence of tasks describe for each primary
task in the sections (Section 1.0 & Section 2.0) that follow.

Section 1.0 Mechanical Installation


1 Complete Pre-Installation Checklist People Required: 1 Hours to Complete: .5
Fill out the Pre-Installation Checklist to confirm completion of pre-install activities before beginning the
installation process. See 3.1 Pre-Installation Checklist on page 39.

2 Receive and Inventory System People Required: 2 Hours to Complete: 4


Inventory equipment & report discrepancies immediately
Collect Product Locator Cards and store in a secure location
Plan route and transport equipment to room
Schedule electrician to coincide with System Cabinet installation.

3 Install Rails and Overhead Tube Support (OTS) People Required: 2 Hours to Complete: 4
Select rail length, assemble rails, mount, level and secure supports, assemble and install longitudinal drive
belt anchor rails and brackets. See Chapter 3 Overhead Stationary Rails on page 45.
Assemble bridge and OTS, lateral and longitudinal drive hardware, lift bridge and mount in stationary rails,
install anti-backlash assemblies. See Bridge and Carriage, Section 3.0 on page 54.
Tension longitudinal and lateral drive belts, wire up the longitudinal drive belt assembly. See Longitudinal
and Lateral Drive Belt Assemblies, Section 4.0 on page 77.
Install X-ray tube covers. See X-ray Tube Covers, Section 5.0 on page 85.
Mount the collimator. See Collimator, Section 6.0 on page 88.
Mount the UIF. See User Interface (UIF), Section 7.0 on page 92.
Install the longitudinal drape. See Longitudinal Cable Drape, Section 8.0 on page 93.
Set up initial OTS counter balance. See Initial OTS Counterpoise Balance, Section 9.0 on page 94.

4 Install Table People Required: 2 Hours to Complete: 2


Remove table from shipping container. See Section 2.0 Unpack Table on page 95.
Position table according to site plans & drawings. See Initial Table and Bridge Alignment, Section 3.0 on
page 100.
Lay table baseplate, then position, level and secure table to floor. See Install Table, Section 4.0 on page
101.
Set up detector & attach Ion chamber. See Detector and Ion Chamber, Section 5.0 on page 105.

Chapter 2 - System Installation Guides Page 33


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
A

5 Install Wall Stand (WS) & options People Required: 3 Hours to Complete: 2
Remove wall stand from shipping container. See Unpack Wall Stand, Section 3.0 on page 120.
Position, level and secure WS. See Install Wall Stand, Section 4.0 on page 128.
Attach detector & Ion chamber. See Detector and Ion Chamber, Section 5.0 on page 131.
Attach display. See Display, Section 8.0 on page 140.
Set up initial Wall Stand counter balance. See Set Initial Wall Stand Counter Balance, Section 9.0 on page
141.
Align OTS and Wall Stand, mount Wall Stand. See Initial Mechanical Alignment, Section 10.0 on page
142, Wall Stand Anchors, Section 11.0 on page 143.
Assemble Digital Support Assembly. See Chapter 7 - Digital Support Assembly (DSA) on page 147.

6 Install Operator Console (Workstation) People Required: 2 Hours to Complete: 1


Position and connect Workstation computer, RCIM and accessories. See Chapter 8 Workstation on page
151. Including any purchased equipment (i.e.Bar Code Reader) options.

7 Pull MIS Cables People Required: 2 Hours to Complete: 8


Pull MIS cables between sub-systems. See Chapter 9 Cables on page 157.

8 Finalize Table Installation People Required: 2 Hours to Complete: 0.5


Install chiller into table base. See Chiller, Section 6.0 on page 111.
Connect table cables. See Table Cables, Section 7.0 on page 113.
Attach table top. See Table Top, Section 8.0 on page 115.

9 Install System Cabinet (Sys. Cab) People Required: 2 Hours to Complete: 2


Insure electrician is present before securing System Cabinet. Allow 1 hour for electrician to complete
power connection.
Position cabinet according to site plans & drawings, and secure. See Chapter 10 System Cabinet on page
161.

10 System Termination Jumpers People Required: 1 Hours to Complete: 0.25


Check and configure system termination jumpers. See Chapter 11 System Termination Jumpers on page
169.

11 Install Grid Holder People Required: 2 Hours to Complete: 0.5


Install grid holder onto wall. See Chapter 12 Grid Holder Installation on page 171.

12 Install Image Pasting Barrier (Optional) People Required: 1 Hours to Complete: 1.0
Install image pasting barrier in front of wall stand. See Chapter 13 Image Pasting Barrier (option) on page
175.

B
Page 34 Section 1.0 - Mechanical Installation
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
B

13 Clean Up Site People Required: 2 Hours to Complete: 1


Remove and dispose of shipping materials and tools properly.

14 Complete MI==>FE Handoff Checklist People Required: 1 Hours to Complete: 0.5


Mechanical installer to fill out MI==>FE handoff checklist and deliver it to the local FE for review at
completion of mechanical installation. See MI ==> FE Handoff Checklist, Section 3.2 on page 41.
Mechanical installer to deliver the completed Pre-Installation checklist to the local FE for review. See Pre-
Installation Checklist, Section 3.1 on page 39.

Chapter 2 - System Installation Guides Page 35


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 2.0 Electrical (Config., Calibration, Finalization)


All Class A (General) documentation for the Definium 8000 system is provided on the Definium
8000 Service Documentation CDROM, 5137441-200, shipped with the system. Any Definium 8000
documentation that is referenced in this Installation manual can be found on this CDROM or on the
Operator Manual CDROM, 5137443-200, also shipped with the system. In addition, paper copies
of the Installation manual (this document) and the Schematics manual are provided to assist in
system installation and troubleshooting.

1 Review Pre-Install & MI==>FE Handoff Checklists People Required: 1 Hours to Complete: 0.5
Local FE to review the Pre-Installation checklist with the mechanical install team.
Local FE to review the MI==>FE handoff checklist with the mechanical install team to confirm completion
of mechanical installation before continuing with the installation process.
If the customer has purchased the Auto Protocol Assist option, the Field Engineer should request a list of
the protocol codes from the HIS/RIS manager for the facility. This list will be provided to the Applications
Specialist so they can manually enter the codes.

2 Equipment and Personnel Readiness People Required: 1 Hours to Complete: 3


Apply LOTO to power. See Lock-Out & Tag-Out (LOTO), Section 1.0 on page 23.
Verify equipment is configured per the site plans and drawings. Also, see Pre-Install Manual, 5137435-
100 which can be found on the Definium 8000 Service Documentation CDROM 5137441-200, shipped
with the system.
Verify that the resistance between subsystems grounds is less than 0.1 ohms. Fix and/or report any
discrepancies.
Fill table chiller with coolant. See Fill Table Chiller with Coolant, Section 6.2 on page 113.
Fill DSA chiller with coolant. See Fill DSA Chiller with Coolant, Section 3.0 on page 149.

3 System Power Up and Minor Tasks People Required: 1 Hours to Complete: 0.5
Prepare system for Energizing. See System Preparation for Energizing Power, Section 2.1 on page 25.
Power Up system. See Energizing System, Section 2.2 on page 27.
Re-check coolant levels in table and DSA chillers; add coolant as necessary.
Install wall stand display overlay. See Install Display Overlay, Section 8.2 on page 141.

4 Configure and Calibrate System People Required: 1 Hours to Complete: 8


Finalize LCD Monitor. See System Manual 5137437-800, Repair/Parts Replacement Procedures/Work-
station/LCD Monitor(s)/Finalization, which can be found on the Definium 8000 Service Documentation
CDROM 5137441-200, shipped with the system.
Configure System parameters listed in System Configuration Chapter/ System Configuration Tool Param-
eters section of System Manual 5137437-800 which can be found on the Definium 8000 Service Docu-
mentation CDROM 5137441-200, shipped with the system.
Verify system is configured correctly per FDO for generator power (65 or 80kW) and ion chamber (3-Cell
or 4-Cell) feature.
Firmware download of ALL controllers is required for proper system operation. Perform each download
available as listed in SUIF>>Utilities>>Subsystem Software>>All Subsystems.

Page 36 Section 2.0 - Electrical (Config., Calibration, Finalization)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Start System Calibrations. Perform the positioner calibrations as identified in the System Manual
5137437-800, Calibration / When To Perform Calibrations / Section 1 Installation, which can be found on
the Definium 8000 Service Documentation CDROM 5137441-200, shipped with the system).
• Start with the "OTS Longitudinal Travel Range" calibration and stop after the "OTS Table Long/Lat"
calibration is completed. The rest of the calibrations will be completed after the final wall stand
alignment is done.
Perform final wall stand alignment. See Final OTS/Tube/WS Alignment, Section 12.0 on page 144.

Finish System Calibrations. In the order specified, perform the calibrations identified in the System Man-
ual 5137437-800, Calibration / When To Perform Calibrations / Section 1 Installation, which can be found
on the Definium 8000 Service Documentation CDROM 5137441-200, shipped with the system).
• Start with the "Wallstand Travel/Tilt Cal." calibration and continue with the rest of the calibrations as
listed in the table.
• For table detector calibration, the table top insert must be in place. Do not use the screws to mount
the insert to the table top frame as the table top insert must be removed in step 6 to install the table
covers.

5 HHS Tests People Required: 1 Hours to Complete: 8


Perform HHS Tests. See Chapter 15 Regulatory Tests (HHS) on page 187 or other equivalent procedure.
Fill out HHS data sheets and submit to HHS representative. See Section 7.0 HHS Data Sheets on page
208 or other equivalent procedure.

6 Install Subsystem Protective Covers People Required: 1 Hours to Complete: 2.5


Table covers. See Table Covers, Section 2.0 on page 215.
OTS final counterpoise adjustment and covers. See OTS Final Counterpoise Adjustment and Trim
Covers, Section 3.0 on page 222.
Wall Stand DSA cover. See DSA Cover, Section 4.0 on page 224.
Wall Stand covers and counterweights. See Wall Stand and Extended Stand Covers, Section 5.0 on page
226.

7 Networking and PACS People Required: 1 Hours to Complete: 0.5


Verify that the hospital network is accessible (i.e; “Ping” Hospital IP network Address).
Verify PACS Image Quality. Follow instructions in System Calibration Chapter/ PACS Image Display Test
section of System Manual 5137437-800 which can be found on the Definium 8000 Service Documentation
CDROM 5137441-200, shipped with the system.
HIS/RIS setup. Configure AutoProtocol Assist (Option), See Operator Manual 5137442-100 on CDROM
5137443-200 provided with the system.

Chapter 2 - System Installation Guides Page 37


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
8 Operational Checks People Required: 1 Hours to Complete: 1.0
Verify that lights and LEDs all function (OTS UIF, RCIM, Table, Wall Stand...). See Operator Manual
5137442-100 on CDROM 5137443-200 provided with the system.
Verify that safety devices function correctly (E-Stops, Inhibit switches...). See Operator Manual 5137442-
100 on CDROM 5137443-200 provided with the system.
Verify that positioning controls function correctly (Table, OTS and Tube, Wall Stand). See Operator
Manual 5137442-100 on CDROM 5137443-200 provided with the system.
Verify that all purchased options function correctly (Dual Energy, Image Pasting, Auto Positioning,
VolumeRAD, 4-Cell Ion Chamber, 130 cm Cross Table Grid, AutoProtocol Assist, etc.) See Operator
Manual 5137442-100 on CDROM 5137443-200 provided with the system.
Ion Chamber AEC Checks:
Start exam: Table and Wall Stand (Wall Stand in horizontal position)
Take 2 exposures.
- Exposure #1, unblocked.
- Exposure #2, 1 cell blocked with flat field phantom (for 4-Cell Ion Chamber, block Cell D (dotted
circle).
Verify mAs difference between exposures is significant, about a factor of 10.
Functional Performance Checks:
Start an exam with Chest PA protocol selected.
Select the Digital Wall Stand or Digital Table receptor.
Check that the FOV is set to 41x41 cm.
Take two pre-test exposures to confirm system is operational.
Take one test exposure and note down the time (in seconds) it took for the conditioned image
(processed image) to be displayed from the time the exposure tone was heard.
Verify the value for conditioned image display is within 10 seconds.

9 Final Tasks & Customer Turnover People Required: 1 Hours to Complete: 2


Back Up System Data Bases. See Section 2.0 Database Backup on page 241.
Perform general house keeping tasks. See Section 3.0 Final Administrative Tasks on page 241.
Complete & Return Product (i.e; Locator) Cards. See Section 4.0 System Rating Plates on page 242.
If your country requires the product rating plate to be in the language of that country, follow the instructions
on document #45474336. Paste the appropriate language label over the English rating plate. The rating
plate labels are: 2379361 (Spanish), 2379361-2 (German), 2379361-3 (French), 2379361-4 (Italian).
Turn system over to customer.

10 Complete FE==>Apps Handoff Checklist People Required: 1 Hours to Complete: 0.5


Field Engineer to fill out FE==>Apps handoff checklist. See FE ==> Application Handoff Checklist, Section
3.3 on page 42. FE should copy and deliver the following checklists to the Applications Specialist at the
completion of the installation:
• Pre-Installation checklist
• MI-to-FE handoff checklist
• FE-to-Apps handoff checklist

Page 38 Section 2.0 - Electrical (Config., Calibration, Finalization)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 3.0 Installation Handoff Checklists


3.1 Pre-Installation Checklist

Delivery Date: Sales Person:

Customer: FDO No.: Room #

Equipment:

Physical Requirements of Site Completed

1.) Room size adequate for intended equipment configuration?

2.) Floor and ceiling is strong enough for intended equipment and mounting
methods approved – seismic regulatory codes considered?

3.) Delivery route accommodates all intended equipment?

4.) Radiation physicist consulted?

5.) Necessary alterations made to circumvent obstructions?

6.) Modifications to room finished?

7.) Supports, platforms, suspensions, ceiling materials been provided?

8.) Support structures installed for floor, ceiling, and wall mounted equipment?

9.) Ceiling supports leveled?

10.) Has floor been modified for cable ducts?

11.) If drop-in ceiling is not used, is access panel provided (3 x 2 ft. minimum)?

12.) Electrical service in place – at the ratings specified in pre-installation


documentation?

13.) Power available to operate power tools?

14.) All non-electrical lines (air, water, oxygen, vacuum) installed?

Interconnections Completed

1.) Signal cable, power and grounding plans produced?

2.) Necessary interconnection hardware, such as junction boxes, conduit or


raceways, and fittings provided?

3.) Interconnection hardware installed?

4.) Flexible, stranded wire provided for System input power connection?

5.) System “feeder” power cables pulled and sufficient length available at
disconnect box for connections?

6.) Interconnecting cables continuity checked, and labeled?

Chapter 2 - System Installation Guides Page 39


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Interconnections Completed

7.) All high voltage cable lengths verified?

8.) Interface information available for equipment?

General Completed

1.) Ceiling, walls, and floor clear of all obstructions?

2.) Walls finished?

3.) Finished floor installed?

4.) Room lights installed?

5.) Dust-creating work completed?

6.) Old equipment within room removed?

7.) Component positions clearly marked on floor?

8.) Space available to store equipment?

9.) Lock on door, or locked room available?

10.) Room IP Addresses for DICOM and Broadband identified?

11.) Broadband connection provided for InSite connection?


OR
If Broadband connection will not be used, is dedicated inbound “dialup”
phone line provided for InSite connection?

12.) Voice phone line connection provided?

13.) Have all fire/safety inspections for occupancy been completed?

14.) Send completed checklist to the GE Healthcare installation team.

Comments:

Inspection Date(s):

Installation Project Manager Signature

Page 40 Section 3.0 - Installation Handoff Checklists


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.2 MI ==> FE Handoff Checklist
Purpose: Mechanical Installer to Field Engineer handoff checklist to confirm that all critical installation
steps have been completed. Upon completion of the mechanical installation of the system, the
Mechanical Installer must fill out this checklist and review with the local Field Engineer.
Item # Item Sign Off
1.) Power and ground cables connected and tightened:
Power cables (3-phase power and ground into System Cabinet)

Ground cables (PC, Wall Plate, Table, Wall Stand, OTS, System Cabinet buss bar)
2.) All signal cables connected and tightened.
3.) Stator cables shield P-clamps (2) installed and tightened on System Cabinet bulkhead.
4.) All mounting bolts installed and tightened (floor, wall and ceiling).
5.) Stationary rails leveled within tolerance of 1/32" (0.8 mm).
6.) Mechanical alignments done:
OTS - Wall Stand

OTS - Table

Wall Stand - Image Pasting Barrier


7.) Detector(s) installed and connected.
8.) Control room cables properly dressed.
9.) Correct language decal applied to RCIM.
10.) All system documentation (cat. no. A7998KT) turned over.
11.) All covers installed.
12.) Room cleaned.
13.) Product locator cards collected.
14.) Missing components and/or options:

15.) List of any installation drawing deviations (all deviations should be signed by customer and
reviewed by the installation specialist).
16.) Packing materials disposed of.
17.) Dollies and crates returned:
System Cabinet dolly

Detector crate(s)

OTS shipping crate and red shipping brackets

____________________________________ ____________________________________
Mechanical Installer Signature Field Engineer Signature

Chapter 2 - System Installation Guides Page 41


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.3 FE ==> Application Handoff Checklist
Purpose: Field Engineer to Application Specialist handoff checklist is used to confirm that all critical
installation steps have been completed.
Item # Item Sign Off
1.) Have all calibrations been completed?
2.) Has the operator manual CDROM contents been installed on the system?
3.) Have the sample images been installed on the system?
4.) Have purchased options been installed/enabled/configured?
5.) All appropriate labels installed:
X-ray tube focal spot
Image Pasting Barrier (purchased option)
6.) HHS testing completed?
7.) InSite checkout completed?
8.) Have functional checks been completed?
9.) Are all covers installed?
10.) Decal attached to Wall Stand display?
11.) Have all protective tape/coverings been removed?
Table footswitches
OTS UIF
12.) Are there any cosmetic issues?
13.) Are there any system issues?
14.) Are there any functional limitations?
15.) Has system documentation (paper and CDROM) been given to the customer?
Also, for VA sites, has 2nd set of documentation been given to the customer?
16.) Is the hospital scheduling system (HIS/RIS) configured? Does the modality
display MWL (Modality WorkList) responses successfully?
17.) Are printers configured and printer calibrations done on ALL printers?
18.) Are PACS configured and SMPTE and/or test images pushed to system for
successful display (be sure burn/no burn is properly set for PACS)?
19.) Has local physics inspection been completed? Some states/sites may not
allow use of system for patients until this is done.
20.) Has backup of pristine system been performed? (Patient Data Base as well as
total system backup)?
21.) Has the list of protocol codes been provided to the Applications Specialist?
22.) Have all product locator cards been collected and processed?
23.) Has Application Specialist received all checklists:
Pre-Installation
MI to FE checklist
FE to Apps checklist

____________________________________ ____________________________________
Field Engineer Signature / Date Application Specialist Signature / Date

Page 42 Section 3.0 - Installation Handoff Checklists


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 4.0 Lean Carts 1 & 2


If the Definium 8000 system has been shipped with Lean Carts 1 and 2, refer to the figures and
tables below to identify the location of the system components.

 


 




 


Figure 2-1 Lean Cart 1

Item No. Catalog Number Component Description


1 B1100PG, A8010LY, B1100LJ Computer, RCIM, Bar Code Reader, Keyboard
2 C1602LE MIS Cables
3 B1100NJ Wall Stand 100 cm Grid
4 B1100KG QAP Kit / Phantoms
5 B2055TW OTS Cover Kit
6 B1100SH Collimator
7 B1000JL Workstation Monitor
8 B1100KM Wall-mounted Grid Holder
9 B1000JL Workstation Monitor
10 B2055WP Bridge Anti-Backlash (Foot-end)
11 B2055WR Bridge Anti-Backlash (Head-end)
Table 2-1 Lean Cart 1 Contents

Chapter 2 - System Installation Guides Page 43


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

 
  

 






 

 

Figure 2-2 Lean Cart 2

Item No. Catalog Number Component Description


1 B1100LN Table Chiller (removed from box)
2 C1602KP Table Cables
3 B1000LJ Wall Stand Ion Chamber
4 B1100NJ/MW/MT Wall Stand Grid - Optional
5 C1602ME/MD Wall Stand Cables
6 B2055ZG 3 meter OTS Cable Assembly
7 B1100PJ Table Grid Frame Assembly
8 B1000KL Chiller Coolant Kit - complete (located behind #9)
9 B1100KG QAP Phantom Kit
10 B1000LL Chiller Coolant Kit - 2 bottles
11 A8091JH Emergency OFF Trip Switch
12 B1100NG Tube Cover Kit
13 B2055TY UIF Interface
14 B1100KL Wall Stand DSA Cover
15 Box containing: A7998KT, System Manuals, Loopback Kit, Lateral Cassette
B1000KE, S3918JZ, & Holder, System Software. OTS Mounting
C7040MM Hardware
16 B1100KF/LF Wall Stand Covers
17 B1100L/P & B1100RC OR Envelope Inside Door
ANY SOFTWARE OPTION
Table 2-2 Lean Cart 2 Contents

Page 44 Section 4.0 - Lean Carts 1 & 2


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 3 Overhead Stationary Rails

Section 1.0 Introduction


1.1 Summary Checklist

Modify stationary rail length (if necessary)

Install stationary rails

Level/shim stationary rails

Confirm stationary rails are parallel

Confirm stationary rails are square

Install longitudinal drive belt anchor rails

1.2 Before You Begin


Stationary rails are shipped fully assembled (i.e., wear strips, lock strips, end caps, and bumpers
assembled to the rail) to you. The mounting bolts and shims are usually shipped in the same
container as the rails. The container should be labeled as the rear rail. If you’re uncertain, the rear
rail has lock strips. Lock strips are thin metallic strip of 1/8 inch steel strips attached to the rail. See
Figure 3-1.

Stationary Rail
Lock Strip

Wear Strip
Figure 3-1 Lock Strip Location

Rails are available in various lengths ranging from 134 inches (3.4 m) to 228 inches (5.79 m).
Usually, rails require no modification to their length. If the rails require cutting, follow the procedure
described in Section 2.2; Inboard Stationary Rail Length Modifications.
Before the installation work begins, you should consult your system’s Pre-Installation manual for
specific room and layout requirements.

Chapter 3 - Overhead Stationary Rails Page 45


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
1.3 Special Tools and Materials
Two people are required to install the rail system safely. A set of common tools (metric and SAE)
are required. In addition, the following items are required:
• Two, 6’ (1.8 m) step ladders.
• Steel measuring tapes, 12 feet (3.7 m) and 50 feet (15.2 m).
• Level, 2’ (61 cm) long.
• 3/4 inch (19 mm) drill bit with 1/2 inch (13 mm) shank (only if rail length needs to be modified).
• No. 29 drill bit (3.5 mm) and 8-32 inch tap (only if rail length needs to be modified).
• No. 18 drill bit (4.3 mm) for 8-32 inch clearance hole (only if rail length needs to be modified).
• No. 7 drill bit (5.1 mm) and 1/4-20 inch tap (only if rail length needs to be modified).
• Drill motors, 1/2 inch (13 mm) & 3/8 inch (9.5 mm) chuck capacity.
• Stationary Rail Leveling Methods/Tools:
- LASER kit (46-216640P1) or equivalent. Recommended leveling method.
* LASER must be capable of 1/8 inch accuracy over 30 feet.
- Transit (i.e, builder’s level). Alternate leveling method, see Appendix A - Figure 17-13 -
Checking Rail Supports with Transit (Alternate Method).
- Tripod (for use with LASER or transit levels):
* Tripod (46-194427P71) or equivalent.
* Tripod case (46-199427P72) or equivalent.
* Tripod adapter (46-216499G1) or equivalent.
- Water-level kit (GE No. 46-316872G1) with 30’ (9.14 m) of 3/8 inch (9.5 mm) I.D. plastic
tubing (ELECTRA/LEVEL by Zircon International) or equivalent capable of +/-1/8 inch
(3.2 mm) over 30 ft. (9.14 m). Alternate leveling method, see Appendix A - Appendix A - .
* Pint bottle containing Water-level liquid.
• Small brush about 1" (2.54 cm) wide for applying wear strip adhesive (only if rail length needs
to be modified).
• “Armstrong” D220 adhesive (Only if rail length needs to be modified.)
• Four, 3" (7.6 cm) C-clamps.
• Headless bolt or stud, 1/2 - 13 x 4" (10.2 cm) for wall-mounted units.
Note: Special 1/2”-13 headed bolts for mounting the stationary rails are furnished with each suspension
Only use the bridge assembly. These bolts are case-hardened “Whiz-lock” flange bolts. These bolts require no
“Whiz-lock” separate flat washer or lock washer. Bolts of this type must be used to obtain adequate clearances
bolt brand and permit unrestricted longitudinal movement of the bridge along the rails. Substitution is not
recommended.
The furnished bolts are 1-1/2 inch (3.8 cm) long (GE Part No. 59136). Thread lengths ranging from
3/4 inch to 2" (1.9 to 5.1 cm) are available from:
MacLean-Fogg Lock Nut Company
1000 Allanson Road
Mundelein, Illinois 60060
Phone (312) 566-0010

1.4 Parts Shortages and Transportation Damage


• Unpack and inspect for shortages and parts damage promptly upon receipt of equipment. See
the “Damage in Transportation” statement in the front of this direction.

Page 46 Section 1.0 - Introduction


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
• Unpack what is necessary to thoroughly inspect for possible shipping damage. Shortage
claims on small parts may be avoided if packing material, envelopes, packages etc., are
carefully examined.
• Do not discard any of packing materials until all items have been located. Refer to the Product
Delivery Instructions (PAD’s) accompanying the equipment to check that all items have been
shipped.

Section 2.0 Inboard Stationary Rails


Note: Make sure that the Unistruts that supports the Stationary Rails are parallel to each other and spaced
at a distance of 56 inches (1422 mm) center-to-center.

The inboard/outboard rails are supplied in 4" (10.2 cm) increments. Each rail has mounting holes
on 26" (66 cm) centers, with the first hole being located at 2" (5.1 cm) from end of rail. The last hole
is located either 2" (5.1 cm) or 4" (10.2 cm) from the other end of the rail with a variable space of
less than 26" (66 cm) between it and the second to the last hole. There must be a minimum of 2"
(5.1 cm) between the ends of the stationary rails and the wall.

2.1 Room Orientation


Refer to Figure 3-2 to understand the terminology regarding room orientation (head, foot, front,
rear).
REAR

HEAD-END OTS FOOT-END

X-ray Tube

UIF

(TOP VIEW)

FRONT

Figure 3-2 Room Orientation Terminology

2.2 Inboard Stationary Rail Length Modifications


If the stationary rail length(s) must be modified, refer to the Appendix A - Section 1.0 Inboard Rail
Length Modifications on page 247.

Chapter 3 - Overhead Stationary Rails Page 47


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.3 Rail Leveling Procedures (using LASER alignment tool)
Note: The LASER alignment tool is the recommended method to level the stationary rails. If a
LASER tool is not available, alternative methods to level the stationary rails can be found in the
Appendix A - Section 2.0 Rail Leveling Procedures (Water-Level or Transit) on page 248.

1.) Mount both stationary rails in place using all bolts and then tighten bolts.
2.) Determine a convenient location in the room to set up the LASER alignment tool. Set up the
LASER alignment tool tripod and place the LASER alignment tool on the tripod.
3.) Using the LASER leveling screws, level the LASER at 0 degrees and at 90 degrees. The
LASER must be level at both angle positions. Turn the LASER back and forth between 0
degrees and 90 degrees, and adjust until correct level is achieved.
4.) Attach the steel tape measure, supplied with the LASER kit, to the stationary rail at one of the
mounting bolt locations. Refer to Figure 3-3.
5.) Turn the LASER ON and rotate LASER so that the beam hits the tape measure.

Stationary Rail

Steel Tape Measure


(supplied with LASER kit)

LASER Alignment Tool


GE Part # 46-216640P1 or equivalent

Tripod Adapter

Tripod

Figure 3-3 Checking Rail Supports with LASER

6.) Take readings at each mounting bolt location and record these readings. Determine the
lowest bottom point of the stationary rails. This is the reference value. Calculate the difference
between this reference value and the other bolt locations to 1/32” (0.8 mm). Use these
difference calculations to determine height and number of shims required at each bolt hole.
Shims will be required and placed at points with calculated differences greater than 1/32”(0.8
mm).
7.) When all measurements have been recorded, turn the LASER OFF.
8.) Proceed to Section 2.3 (Rail and Shim Installation).

2.4 Rail and Shim Installation


1.) Loosen all bolts until the bolts are finger tight. At each bolt location, place the previously

Page 48 Section 2.0 - Inboard Stationary Rails


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
determined shims between the rail and the ceiling steelwork. Re-tighten all bolts.
2.) Check stationary rail parallelism:
a.) Measure the spacing between the rails. It must be within 1/8” (3.2 mm) of the specified
value. Refer to Figure 3-4. The final distance between the outside edges of the
stationary rails should be 58.36 inches +/- 1/8 in. (1482 mm +/- 3.2 mm) as shown in
Figure 3-5.
b.) Measure the distance from one end of a rail to the same point on the opposite end of the
other rail. These are the rails' diagonals. Compare both measurements. The rails are
parallel if the diagonals are equal.
3.) Re-measure stationary rail levelness at each bolt location. Adjust shims accordingly so
that all mounting bolt readings are equal within 1/32” (0.8 mm).
4.) Retighten all bolts.
5.) Using a two-foot level perpendicular to the rail at each bolt location, insure that each stationary
rail is not twisted. Adjust shims accordingly. Re-tighten bolts.
CENTERLINE OF UNISTRUT OR
EQUIVALENT FOR MOUNTING
REQUIRED CABLE DRAPE
AUXILIARY HARDWARE
SUPPORT RAIL 34-3/4 in. +/- 1/4 in.
46-177615P3 & (88.3 cm +/- 6.5 mm)
46-177615P4
56 in. +/- 1/8 in. (142 cm +/- 3 mm)
CHECK AT EACH BOLT POINT

TRIM TO
LENGTH
TIGHTEN
HARDWARE
AND RECHECK

LAT
B LEVELS OF ERA
OF S C TAB L CE
E LE NT
LIN LE E ER
ER HO LIN LIN
E NT OLT ER E
C IL B E NT
RA LC
INA OF
T UD INE ES
N
I
G LE E RL HOL
LO TAB NT LT
OF CE IL BO
RA A, B & C - ACCORDING TO THE SUBSYSTEM SUPPLIED

Figure 3-4 Inboard Stationary Rails Installation and Leveling

Chapter 3 - Overhead Stationary Rails Page 49


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 3-5 Stationary Rail Spacing, Cross-Sectional View

2.5 Longitudinal Anchor Rails


1.) Locate and open the Long Belt Drive Kit. See Figure 3-6.
2.) Remove the tensioner and anchor mounting assemblies.

Figure 3-6 Long Belt Drive Kit

3.) If present, remove metal caps from the ends of each stationary rail and discard; these are not

Page 50 Section 2.0 - Inboard Stationary Rails


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
needed for the Definium 8000 system. See Figure 3-7.

Figure 3-7 Remove and Discard End Caps

4.) Mount the anchor side assembly to the "foot" end of the stationary rails. See Figure 3-8.

Figure 3-8 Anchor Assembly at Foot End

5.) Mount the tensioner side assembly to the "head" end of the stationary rails.

Chapter 3 - Overhead Stationary Rails Page 51


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.) Install bridge bumpers.
a.) Attach the short bumpers at the head-end of the stationary rails. See Figure 3-9.

Figure 3-9 Short Bumpers on Head End

b.) Attach the long bumpers at the foot-end of the stationary rails. See Figure 3-10. The
longer bumpers are required to prevent the longitudinal drive assembly on the bridge from
hitting the longitudinal drive belt anchor assembly on the foot-end anchor rail.

Figure 3-10 Long Bumpers on Foot End

Page 52 Section 2.0 - Inboard Stationary Rails


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 4 Overhead Tube Support (OTS)

Section 1.0 Summary Checklist

Modify bridge length (if necessary)

Attach lifting fixtures to stationary rails

Place bridge on lifting fixtures

Remove stiffener plate from top of bridge

Lift and attach bridge to OTS

Install lateral drive hardware

Re-install stiffener plate to top of bridge

Install longitudinal drive hardware on bridge

Lift and install bridge into stationary rails

Install lateral cable drape channel/hooks

Connect MIS cables to OTS bulkhead

Install and adjust bridge anti-backlash assemblies

Adjust longitudinal drive belt alignment and tension

Connect longitudinal drive assembly cable

Adjust lateral drive belt alignment and tension

Install X-ray tube cover/endcaps

Install collimator, connect cables, install covers

Install user interface (UIF), connect cable

Attach auxiliary support and longitudinal cable drape


channel to unistrut

Perform initial OTS counterpoise balance


adjustment

Confirm OTS jumpers and switches (factory set)

Chapter 4 - Overhead Tube Support (OTS) Page 53


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 2.0 Modifying the Bridge Length


If the bridge length must be modified, refer to Appendix A - Section 3.0 Modifying the Bridge Length
on page 252.

Section 3.0 Bridge and Carriage


Note: Do not remove any of the red shipping brackets from the top of the OTS shipping crate until
the OTS has been moved into the exam room.

1.) Remove the plastic bag containing OTS bridge mounting hardware from the Technical
Publications box.
2.) Cut the shipping bands, remove the wood screws and remove the top wood platform (if
present).

Figure 4-1 OTS Shipping Crate

3.) Remove the top shipping bracket (red) from the top of the OTS (5 wood screws on each side
that attach the bracket to the shipping crate, 3 hex head capscrews on each side that attach
the bracket to the top of the carriage). See Figure 4-1 and Figure 4-2.
4.) Do not remove the 2 side shipping brackets (red) until the bridge has been attached to
the top of the OTS and the lifting fixtures have been attached to the bridge and
stationary rails.

Page 54 Section 2.0 - Modifying the Bridge Length


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-2 OTS Shipping Crate - Top Bracket

3.1 Set Up the Lifting Fixture


1.) If the carriage locking bolt has been removed, re-install it to lock the carriage in the bridge and
prevent lateral movement of the carriage while the bridge is being raised.
2.) Set up the lifting fixtures on the stationary rails as shown in Figure 4-3, being sure they are 37"
apart.
A 37 IN.
(940 mm) STATIONARY RAILS

CRANK
A=B

CRANK

B 37 IN.
(940 mm) 37"

CLAMP

NOTE POSITION
OF BOLTS

CRANKS SHOULD BE AWAY


FROM OTS. OPPOSITE LIFTING
FIXTURE WILL HAVE THE
CRANK ON THE OPPOSITE SIDE
OF THE BRIDGE.

BELT AND BUCKLE ENGAGEMENT-- POSITION NEAR FLOOR,


THE ONLY SAFE WAY! THEN LEVEL FIXTURE
BY ADJUSTING BELTS
IN BUCKLES

Figure 4-3 Setting up the Lifting Fixtures

3.) Loosen and remove the four (4) bolts and clamps from the tops of the lifting fixture upright
posts. See Figure 4-3.
4.) Remove the protective shipping cardboard from the bridge.
5.) Lift the bridge and place it on the lifting fixtures as shown in Figure 4-4 so that the bearings on
the bridge cross-ties are directly below the stationary rails. The end of the bridge with the
model/serial number label must be towards the rear stationary rail.

Chapter 4 - Overhead Tube Support (OTS) Page 55


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.) Re-install and tighten the four (4) bolts and clamps onto the tops of the lifting fixture upright
posts. Make sure that the ends of the clamps are inside the bridge rails.

POSITION BRACKET ON LIP


OF BRIDGE AND TIGHTEN BOLT

CRANK SHOULD BE
ON THIS SIDE

CARRIAGE LOCKING BOLT - KEEP THIS


BOLT INSTALLED UNTIL THE BRIDGE HAS
BEEN INSTALLED IN THE STATIONARY
RAILS

Figure 4-4 Positioning Bridge in Lifting Fixtures

7.) Remove the metal stiffener plate (if present) from the top of the bridge (16 bolts and lock
washers). The stiffener plate will be re-installed after the lateral drive belt has been installed.

Figure 4-5 Bridge Stiffener Plate

8.) Turn the lifting fixture cranks equally to raise the bridge until it is approximately 5 feet (1.5 m)
above the floor.
9.) Position the OTS shipping crate below the bridge so that the six mounting holes in the carriage
dolly align with the threaded holes in the top of the OTS. Refer to Figure 4-7.

3.2 Bridge Mounting Procedure


1.) Lower the bridge until it is resting on top of the OTS as shown in Figure 4-6. While lowering

Page 56 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
the bridge, re-check the alignment of the bridge and OTS and re-position the OTS shipping
crate as necessary to maintain alignment of the mounting holes.

Figure 4-6 Attaching Bridge to Top of OTS

2.) Apply Loctite 242 (Blue) to the six (6) 10 mm x 60 mm hex head capscrews. The capscrews
can be found in the large zip-lock plastic bag attached to the OTS.
3.) Turn in and tighten the six (6) hex head capscrews (and lockwashers) connecting the front and
rear of the OTS carriage to the dolly. See Figure 4-7.
4.) Leave the OTS/bridge assembly resting on the shipping crate until the bridge is raised and
installed into the stationary rails.

Note: Be extremely careful not to pinch cables between the dolly and the carriage when tightening
Don’t pinch the bolts.
cables.

Figure 4-7 Mounting OTS Carriage to Dolly

Chapter 4 - Overhead Tube Support (OTS) Page 57


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.3 Lateral Drive Hardware Installation
The following steps are taken to install the lateral drive belt mounting brackets and drive belt.

Figure 4-8 Lateral Position Sensor and Belt Anchor Bracket Locations

1.) See Figure 4-9 for assembly directions for the front (fixed) anchor. Loosely assemble the
upper belt clamp (item 1) and the lower belt clamp (item 2) using the four socket head cap-
screws (item 3).

Figure 4-9 Front Lateral Belt Anchor Assembly Directions

2.) Attach one end of the lateral drive belt to the front belt anchor assembly as shown in Figure 4-
11 and tighten the four socket head cap-screws.

Page 58 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.) Attach the front belt anchor assembly to the top of the bridge front end cap (closer to the head-
end side of the bridge). See Figure 4-10 and Figure 4-11.

Figure 4-10 Front Lateral Belt Anchor Assembly Directions

Figure 4-11 Lateral Drive Belt Anchor on Front of Bridge

4.) Carefully feed the lateral drive belt through the lateral drive assembly. See Figure 4-12.

Figure 4-12 Lateral Drive Belt Engaged

5.) See Figure 4-13 for assembly directions for the rear (tensioner) assembly. Loosely assemble
the upper belt clamp (item 1) and the lower belt clamp (item 3) using the four socket head cap-
screws (item 4).

Chapter 4 - Overhead Tube Support (OTS) Page 59


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-13 Rear Lateral Belt Tensioner Assembly

6.) Attach the belt tensioner bracket (item 1) to the top of the bridge rear end-cap (item 3) with four
bolts (item 2) per Figure 4-14.

Figure 4-14 Rear Lateral Belt Tensioner Assembly

Page 60 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.) Attach the belt to the belt tensioner assembly per Figure 4-15.

Figure 4-15 Rear Lateral Belt Tensioner Assembly

8.) Place the belt clamp assembly on the tensioner bracket per Figure 4-15.
9.) Install the tension adjustment bolt through the rear of the belt tensioner bracket, through the
locknut and into the belt clamp assembly. See Figure 4-15.
10.) The lateral drive belt tensioner adjustment will be performed after the bridge has been lifted
and installed into the stationary rails.

Chapter 4 - Overhead Tube Support (OTS) Page 61


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.4 Bridge Stiffener Plate Installation
1.) Attach the stiffener plate to the top of the bridge (16 bolts and lock washers). The small notch
in the stiffener plate should be positioned at the foot-end of the rear bridge crosstie. See
Figure 4-16 and Figure 4-17.

Figure 4-16 Attach Stiffener Plate to Bridge

Figure 4-17 Attach Stiffener Plate to Bridge

3.5 Longitudinal Drive Hardware Installation


1.) Before raising the bridge with the lifting fixtures, make sure that the carriage locking bolt is still

Page 62 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
installed as shown in Figure 4-18.

Figure 4-18 OTS Carriage Locking Bolt

2.) Completely remove all eight bolts from the front bridge cross-tie and remove the cross-tie. See
Figure 4-19.

FO
R I REA
NB R O
OA F
RD RO
REAR CROSSTIE MO OM
(WITH NYLON BEARINGS) UN
TIN
FRONT CROSSTIE G
(REMOVE TO INSTALL
BRIDGE INTO THE
STATIONARY RAILS)

FR
ON
T

Figure 4-19 OTS Bridge Installation

3.) Raise the bridge until the weight of the OTS is slightly resting on the OTS shipping crate.
4.) From both sides of the OTS, remove the four (4) bolts and four (4) wood screws attaching the
red OTS shipping brackets to the OTS assembly and wooden shipping crate. See Figure 4-20.

Chapter 4 - Overhead Tube Support (OTS) Page 63


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-20 Remove OTS Shipping Brackets

5.) Raise the bridge until it is approximately 2 inches above the OTS shipping crate.
6.) Mount the longitudinal drive assembly to the top of the bridge, over the two 1/4" studs on the
top surface of the foot-end rail. See Figure 4-21. Make sure that the longitudinal drive
assembly is mounted tightly against the side of the bridge rail, then secure the assembly to the
bridge using two lockwashers and nuts on the top and one lockwasher and bolt on the side.

Figure 4-21 Longitudinal Drive Assembly

RISK OF When installing the bridge into the stationary rails, use caution so that the plastic bearings in the
DAMAGE TO rear cross-tie do not impact against the rear stationary rail. Also, do not set the entire weight of the
PLASTIC bridge/OTS assembly on the plastic bearings; the weight must be on the steel stationary rail
BEARINGS bearings. See Figure 4-22.

Page 64 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-22 BRIDGE CROSS-TIE BEARINGS

7.) Turn the lifting fixture cranks equally to raise the bridge up to the stationary rails. Carefully
position the bridge so that the bearings of the rear cross-tie are placed into the rear stationary
rail. Note that two of these bearings are now made of a plastic material and more caution
is required to prevent damage to these bearings.
8.) Place the front cross-tie back into its original position on the bridge, with the bearings inside
the front stationary rail and secure loosely to the bridge rails with the eight screws (apply
Loctite 242 (Blue) to the screw threads) and four plates removed earlier.
If the holes in the bridge rails and cross-tie don’t line up, place a loop of strapping around the
bridge in the area of the front cross-tie. Wind it tight with a screw driver to pull the bridge rails
closer together, lining up the holes with the cross-tie. See Figure 4-23.

STATIONARY
RAIL ”

UPPER
BEARING

AV IS STRA P

FRONT CROSS-TIE LOWER BEARINGS (1 hidden) -


ONLY ON REAR CROSS-TIE
CROSS-TIE

REAR CROSS-TIE

Figure 4-23 INSTALLING OTS BRIDGE

9.) Center the front and rear cross-tie bearings on the wear strips and tighten the eight screws

Chapter 4 - Overhead Tube Support (OTS) Page 65


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
securing the front cross-tie to the bridge.
10.) Lower the bridge until the roller bearings contact the wear strips in the stationary rails.
11.) Disconnect the lifting fixtures from the bridge and lower the lifting fixtures to the floor.
12.) Adjust the upper bearings on both cross-ties for 0.02" gap between the tops of the bearings
and the upper part of the rails. The adjustment is made with an Allen wrench from the bottom
of the cross-tie. See Figure 4-23.
13.) Disconnect the lifting fixtures from the stationary rails.

Note: Be sure the bearings ride on the center of the wear strips over the complete range of OTS
Potential for longitudinal travel. If the bearings travel off of center, the stationary rails and cross-tie
mis-alignment. alignment must be checked.

14.) Back out the counterpoise locking bolt. See Figure 4-24.

Figure 4-24 Counterpoise Locking Bolt

15.) Use an Allen wrench to remove the column shipping rod from the rear of the OTS. See
Figure 4-25.

Page 66 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Use caution after removing the column shipping rod. If the OTS is not correctly counter-balanced,
the column may drift upward after being pulled down and released.

Figure 4-25 OTS Column Shipping Rod

16.) Remove the carriage locking bolt and install the plastic plug buttons into the carriage locking
bolt access holes on each side of the bridge (1 or 2 buttons each side). See Figure 4-18.

3.6 Install The Lateral Cable Drape Hardware


1.) Use 8 screws to install the lateral cable drape channel. See Figure 4-26.

GEH
Figure 4-26 Installing the Lateral Cable Drape Channel

2.) Slide two cable drape hooks into the lateral cable drape channel from the rear of the channel.
See Figure 4-27.

Chapter 4 - Overhead Tube Support (OTS) Page 67


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-27 Lateral Cable Drape Hooks

3.7 Carriage Cable Connections


1.) Connect the following cables to the OTS:
- A12J1, J2 Stator, Fan & Pressure Switch (see Figure 4-28)
- A1 Ground (see Figure 4-28)
- B1 J2 OTS CAN (see Figure 4-29)
- A7 J5 OTS Power 120VAC (see Figure 4-29)
- A12 Anode HV Cable (see Figure 4-31, Figure 4-32), apply thin layer of silicone oil
(GE Part Number: LNR8733, 1 ounce tube)
- A12 Cathode HV Cable (see Figure 4-31, Figure 4-32), apply thin layer of silicone oil
(GE Part Number: LNR8733, 1 ounce tube)
- B1 J7 Longitudinal Drive (see Figure 4-33)

Figure 4-28 Connecting Stator, Fan/Pressure & Ground Cables

Page 68 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-29 Connecting OTS CAN Cable to B1J2

Figure 4-30 Connecting OTS Power Cable

Chapter 4 - Overhead Tube Support (OTS) Page 69


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

l
Oi
one
ic
Sil

HV Receptacle

Cable from
System Cabinet

OTS A12 Bulkhead


Figure 4-31 Lubrication of HV Cable Connectors with silicone Oil

Figure 4-32 Connecting High Voltage Cables

Page 70 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-33 Connecting Longitudinal Drive Cable

2.) Check all OTS cable connections, including factory installed cables, to confirm that they are
securely fastened.
3.) Use Ty-raps to position OTS cables together to keep them from rubbing against the
counterpoise and OTS covers.
4.) Mount the lateral cable drape bracket to the OTS carriage. See Figure 4-34.

Figure 4-34 Lateral Cable Drape Bracket on Carriage

5.) Connect the lateral cable drape to the cable drape bracket, 2 cable drape hooks and bridge
endcap with cable ties. See Figure 4-35 and Figure 4-36.
- The front loop (between points 1-2) needs to be shorter than the back loop (3-4).
- The front loop should be as short as possible (recommended lengths shown below). Best
case would be that the cable loop on the X-ray tube does not hit the front loop on the
bridge at max SID (height) during tube angulations.
- See Table 4-1 below for recommendations for true cable drape length between anchor/

Chapter 4 - Overhead Tube Support (OTS) Page 71


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
hook points on 3m bridge with OTS full-forward.

Note: Tie wraps at points 2 and 3 should be tight enough so that the cable drape cannot come off the
hook.

Attachment Point Attachment Type Length

1 Anchor to bracket with tie wrap Loop 1 1000-1200 mm

2 Attach to Trolley Hook with tie wrap Loop 2 1000-1200 mm

3 Attach to Trolley Hook with tie wrap Loop 3 1300-1500 mm

4 Anchor to rear corner of bridge with tie wrap


Table 4-1 Recommended Cable Drape Lengths

Figure 4-35 Lateral Cable Drape Installed

Page 72 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-36 Cable Drape Attachment to Bracket

6.) Attach the remaing portion of the OTS cable drape to the cable drape hooks (on the auxiliary
support channel) using cables ties, evenly spaced.

Chapter 4 - Overhead Tube Support (OTS) Page 73


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.8 Bridge Anti-Backlash Assemblies
Note: The anti-backlash adjustment must be performed before the longitudinal drive belt alignment and
longitudinal drive belt tensioning.

Purpose of Anti-Backlash Assembly: The purpose of applying a specific amount of force (via the
anti-backlash assemblies) is to react against forces produced when the OTS is accelerating in a
longitudinal direction during auto-positioning or tomo sweeps. The force applied by the anti-
backlash assemblies cannot be too high or the bridge bearings may be damaged (the bearings on
the rear of the bridge are very difficult to replace). These plastic bearings were introduced so that
the bearing material does not wear on the aluminum stationary rails like steel bearings would. High
applied force will also increase moving efforts in the longitudinal direction.
1.) Install two anti-backlash assemblies to the top of the bridge rails, behind the rear stationary
rail, using two bolts, washers and lockwashers on each assembly. See Figure 4-37 and
Figure 4-38.

Figure 4-37 Anti-Backlash Assembly Locations

Page 74 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-38 Anti-Backlash Assembly

2.) On each anti-backlash assembly, adjust the tensioning bolt clockwise until the bearing just
touches the rear surface of the stationary rail (finger tight).
3.) Place a mark on the tensioning bolts so that adjustment revolutions of the bolts can be
counted.
4.) Turn the tensioning bolts 5 turns clockwise or until the bolts reach the end of travel. Over-
tightening of the tensioning bolts draws the bridge internal bearings tight to the rear suface of
the stationary rails.
5.) Run the bridge back and forth 1 - 2 meters (3 - 6 feet) several times to finish drawing the bridge
up to the rear stationary rail.
6.) Turn both of the tensioning bolts counterclockwise until the bearings no long touch the rear
surface of the stationary rail. Do not move the bridge until the adjustments in steps 7 and 8
are completed.
7.) On each anti-backlash assembly, adjust the tensioning bolt clockwise until the bearing just
touches the rear surface of the stationary rail (finger tight).
8.) On each anti-backlash assembly, adjust the tensioning bolt an additional 3 turns clockwise.
This produces the proper spring force of 50 lb-f or 225N. Make sure each bolt has equal turns
so that the bridge is perpendicular to the stationary rails. Unequal force could result in the
bridge being at an angle.
9.) Apply Loctite 242 (Blue) to the mounting screws and install the anti-backlash assembly
covers. See Figure 4-39.

Chapter 4 - Overhead Tube Support (OTS) Page 75


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-39 Anti-Backlash Assembly Covers

Page 76 Section 3.0 - Bridge and Carriage


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 4.0 Longitudinal and Lateral Drive Belt Assemblies


4.1 Longitudinal Belt Installation and Tracking
1.) Route the belt through longitudinal drive assembly; over the top idler, under the drive pulley
and back out over the top other idler. See Figure 4-40.

Figure 4-40 Longitudinal Drive Belt Installation

2.) Verify that the longitudinal drive assembly is installed flush against the rail.
a.) There should be no gap between the longitudinal drive and the bridge rail surface. See
Figure 4-41.

Figure 4-41 Bad and Good Longitudinal Drive Alignment

b.) If a gap exists, loosen the screws and push the longitudinal drive flush against the rail.
While holding the drive in position, tighten the fasteners.
3.) Attach the belt.

Chapter 4 - Overhead Tube Support (OTS) Page 77


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
a.) At the anchor (foot) end of the longitudinal rail, install the belt. See Figure 4-42.

Note: Tension Line - Use a Mono-Filament “Fishing” Line (8 pound test, 0.25 mm) or equivalent to
measure belt deflection. “It's important that the line you choose have a small mass. A high
mass causes the line to sag in the middle and your resulting tension will be wrong.”

i.) Place one end of the tensioning line into the anchor end belt clamp and position it
towards the middle on the tooth side of the belt. “We want the belt clamp to capture
both the belt and the line when the clamp is secured.”
ii.) Now place the belt between the clamp and the plate along with the tension line. “The
clamp must capture seven (minimum) teeth on the belt when properly installed.”
iii.) With the belt and tensioning line in the belt clamp, torque the belt clamp (M6 Hex
Socket) screws to 9.9 Nm (7.3 lbs-ft).

Figure 4-42 Anchor plate assembly (Foot End)

iv.) Tighten the two (M10 Hex Socket) plate screws that secure the anchor plate to the
rail.
v.) Loosen each of the plate screws that secure the anchor plate by one turn
counterclockwise. “We want the tension plate to be slightly loose so it can move
during our tracking adjustment.”
b.) At the tensioning (head) end of the longitudinal rail, install the belt. See Figure 4-43.
i.) Remove the four (M6 Hex Socket) belt clamp screws.

Figure 4-43 Tensioning Clamp

ii.) Create belt tension adjustment range; turn the tension adjustment screw
counterclockwise until the belt clamp slightly overhangs the plate by ½ of a tooth. “We
want the tensioning belt clamp as far out as possible, so you have the most tension
range. As you turn the adjustment screw, the belt clamp plate will move in (clockwise)

Page 78 Section 4.0 - Longitudinal and Lateral Drive Belt Assemblies


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
or out (counterclockwise).” See Figure 4-44.

Figure 4-44 Tensioning Belt Clamp Setup

iii.) Pull out the slack in the belt and position it parallel to the belt clamp. "We want the
belt pulled tight so that it is captured by seven teeth of the belt clamp."
iv.) Make marks (using a marker) on the belt so that it can be placed in the same position
after you cut the belt to length. See Figure 4-45.

Figure 4-45 Marking Belt Position in Clamp

a.) Place a mark on the belt where it will be cut off and still be captured by seven
clamp teeth. “The belt clamp must capture seven teeth of the belt after you cut off
the excess length.”
b.) Place a mark on the belt where it enters the belt clamp. “This mark serves as
additional reference point. It helps to ensure that you will capture the proper
amount of belt teeth in the clamp. The mark should be visible when the clamp is
secured.”

Chapter 4 - Overhead Tube Support (OTS) Page 79


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
v.) Cut the belt off at the "cut-off" mark you’ve made. See Figure 4-46.

Figure 4-46 Belt Markings

vi.) Place the cut-off end of the belt into the tensioner belt clamp and secure the belt
clamp screws by turning them clockwise. “Use the marks you made on the belt to
position the belt.” See Figure 4-47.

Figure 4-47 Tensioning Belt Clamp


vii.) Tighten the two (M10 Hex Socket) screws that secure the tensioner plate to the rail.
See Figure 4-48.

Figure 4-48 Tensioner Plate Mounting Screws


viii.)Turn each tensioner plate screw one revolution counterclockwise. “We want the
tensioner plate slightly loose to allow for movement during our belt tracking

Page 80 Section 4.0 - Longitudinal and Lateral Drive Belt Assemblies


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
adjustment."
4.) Align the tracking of the belt with respect to the rails. See Figure 4-49.
a.) Center the belt on the longitudinal drive pulley.
i.) Move the bridge back and forth between the head and foot ends of the stationary rails
several times. “We want the tension plate and anchor plate to align themselves to the
rails.”
ii.) Move the bridge around the stationary rails midpoint until you can find a position
where the belt tracks down the center of the longitudinal drive pulley. The belt is
centered when there is an equal gap between the belt edges and drive pulley flanges
on both sides.
b.) Measure the distance from the stationary rail to the nearest belt edge at the longitudinal
drive idler pulley and set the belt position at the head and foot ends of the rail equally.
i.) At the longitudinal drive assembly: measure and record the "tracking distance" from
the edge of the stationary rail to the nearest belt edge at the idler pulley. See
Figure 4-49.

Figure 4-49 Belt Tracking at the Longitudinal Drive

ii.) At the head end of the stationary rail: position the edge of the belt with respect to the
stationary rail at the same distance you recorded at the longitudinal drive assembly.
iii.) At the foot end of the stationary rail: position the edge of the belt with respect to the
stationary rail at the same distance you recorded at the longitudinal drive assembly.
iv.) Move the bridge back and forth between the head and foot end several times. "We
want the belt plates to align themselves again to the rail following adjustment."
v.) Measure the distance from the rail to the edge of the belt at both ends again and verify
adjustment.
c.) Secure the plates, if your measurements are equal or within specification.
i.) If the variance between the belt and rail on both ends is less than or equal to +/-
0.5 mm, then secure the plate screws on the anchor (foot) and tensioning (head)
ends of the rail.
• At the foot end, hold the anchor plate so the plate doesn’t move, while you
torque the (M10 Hex Socket) screws to 48 Nm (35.4 lbs-ft). The plate will
want to move when you tighten it because of the turning of the screw and

Chapter 4 - Overhead Tube Support (OTS) Page 81


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
washer.
• At the head end, hold the tension plate so the plate doesn’t move, while you
torque the (M10 Hex Socket) screws to 48 Nm (35.4 lbs-ft). "The plate will
want to move when you tighten it because of the turning of the screw and
washer."
ii.) If the variance between the belt and rail on either end exceeds 1 mm, repeat steps
b.) and c.) above.
d.) Finally, verify that the belt does not rub "excessively" against the metal flange of
the drive pulley at any point along its entire longitudinal travel.
“It is normal for the belt to move (walk) a little from side to side on the drive pulley. It is not
normal to have excessive contact with the drive pulley flange, which can later result in
“squeaking” noises as the bridge is moved longitudinally.”
* If the belt occasionally touches the metal flange on the drive pulley – tracking is good.
* If the belt constantly touches the metal flanges or drives hard into the drive pulley
metal flange or makes any noise during tracking – tracking is bad and should be
adjusted.
5.) Set the belt tension.

a.) Move the bridge longitudinally until it is two meters from the anchor end belt clamp, as
measured from a tangent point on top of the idler pulley to the point where the tension line
exits the anchor end belt clamp. See Figure 4-50.

Figure 4-50 Belt Tension Setup

b.) Using clamping devices (i.e. locking pliers, C-clamps, etc.), secure the bridge to the
stationary rail to prevent it from moving.
c.) Set up the tension line.
i.) Loop the tension line over the idler and down the middle of the belt.
ii.) While maintaining this position, pull the line tight. The line should make a well-defined
sound (pitch) if pulled and released.
iii.) Secure the line to prevent loss of tension.
iv.) Place a mark on the belt that is one meter from where the tension line exits the anchor
end belt clamp.

Page 82 Section 4.0 - Longitudinal and Lateral Drive Belt Assemblies


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
v.) Using a ruler, measure the distance from the tension line to the belt teeth immediately
above the 1-meter marking.
vi.) The distance (d) between the tension line and the belt at this point should be 6
millimeters (+/- 1 millimeter). Set the tension at the tensioner (head) end of the
stationary rail to achieve this.
a.) Turn the tension screw (located on the head end tension assembly) until 5 or 6
millimeters of deflection is achieved.
• Clockwise to increase tension.
• Counterclockwise to decrease tension.
b.) Torque the belt clamp (M6 Hex Socket) screws to 9.9 Nm (7.3 lbs-ft) each.
vii.) Remove any clamps used to hold the bridge stationary.
viii.)Do not remove the tension line from the anchor clamp. Store the excess line and
secure it to the top of the rail out of sight for future use.
6.) Install tensioner assembly cover plate.
a.) Apply a small amount of non-permanent thread locker, such as Loctite 242 (Blue), to
screw (M6 Hex Socket) threads.
b.) Torque screws 9.9 Nm (7.3 lbs-ft).

Chapter 4 - Overhead Tube Support (OTS) Page 83


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.2 Wiring the Longitudinal Drive Assembly
1.) Route the longitudinal drive cable through the cable trough (located on foot-end bridge rail) and
connect it to the longitudinal drive assembly per Figure 4-51 and Figure 4-52.

Figure 4-51 Routing the Longitudinal Assembly Cable

2.) Apply Loctite 242 (Blue) to the mounting screws and attach the cover to the longitudinal drive
assembly. See Figure 4-52.

Figure 4-52 Connecting the Longitudinal Assembly Drive Cable and Cover

4.3 Lateral Drive Belt Alignment and Tensioning


1.) Adjust front and rear brackets for proper side-to-side alignment of the lateral drive belt as

Page 84 Section 4.0 - Longitudinal and Lateral Drive Belt Assemblies


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
follows:
a.) Position the OTS in the center of the lateral range of travel.
b.) Measure the distance from the head-end bridge rail to the lateral drive belt where it enters
the lateral drive assembly on top of the OTS.
c.) Adjust the front bracket longitudinally so that the distance from the head-end bridge rail
to the lateral drive belt (where it enters the front bracket) is equal to the distance obtained
in step b above.
d.) Tighten the capscrews when adjustment is completed.
e.) Adjust the rear bracket longitudinally so that the distance from the head-end bridge rail to
the lateral drive belt (where it enters the rear bracket) is equal to the distance obtained in
step b above.
f.) Tighten the capscrews when adjustment is completed.
2.) Position the OTS to the front of the bridge.
3.) Loosen the locknut on the rear lateral belt anchor assembly. See Figure 4-53.
4.) Adjust the belt tensioning bolt (clockwise) just to the point where you can no longer see any
change in the lateral belt sag, then tighten the tensioning bolt an additional 1 1/4 turns
clockwise.
5.) Tighten the locknut.

Figure 4-53 Rear Lateral Belt Anchor Assembly

Section 5.0 X-ray Tube Covers


1.) Place the tube/fan cover in position on the X-ray tube and connect the tube select lamp wires

Chapter 4 - Overhead Tube Support (OTS) Page 85


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
to terminals 4 and 5 on the X-ray tube terminal strip. See Figure 4-55 and Figure 4-55.

Figure 4-54 Tube Select Lamp Wiring

2.) Use 3.5mm (9/64 inch) Allen wrench to attach the tube/fan cover and cable entrance cover to
the tube unit. See Figure 4-55 and Figure 4-56.

Note: For detailed X-ray tube stator cable connections, refer to the Definium 8000 System Schematics
and Drawings manual, 5137436-100, MIS Chart 020013.

Figure 4-55 Install Tube/Fan Cover and Cable Entrance Cover

Page 86 Section 5.0 - X-ray Tube Covers


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-56 Tube Cables Routing

3.) Use 4mm Allen wrench to attach fan grill. See Figure 4-57.

Figure 4-57 Install Fan Grill

4.) Use 3.5mm (9/64 inch) Allen wrench to attach plastic endcaps to anode and cathode end of X-
ray tube using 6 capscrews. See Figure 4-58. Install each endcap so that the GE logo is
correctly oriented when the X-ray tube is in the standard (non-rotated) position.

Figure 4-58 Install X-ray Tube Endcaps

Chapter 4 - Overhead Tube Support (OTS) Page 87


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 6.0 Collimator


1.) Carefully rotate the tube assembly 180 degrees (pointing up). See Figure 4-59.

Figure 4-59 Tube Pointing Up With Collimator Removed

2.) Apply the supplied grease (shipped in collimator box) to the top and bottom of the collimator
mounting flange. See Figure 4-59.

Figure 4-60 Tube Pointing Up With Collimator Removed

3.) Remove the two collimator blade shipping screws and remove the screw that holds the detent

Page 88 Section 6.0 - Collimator


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
lever away. See Figure 4-61.

Figure 4-61 Collimator Blade Shipping Screws

4.) Remove the blade protector by backing off the three collimator mounting screws. See Figure 4-
61.
5.) Make sure that the collimator blades are fully closed before installing the collimator onto the X-
ray tube.
6.) Being careful not to damage the collimator fingers, position the collimator on the X-ray tube.
See Figure 4-62. Be sure the detent lever is not preventing the collimator from seating fully.

Figure 4-62 Collimator Mounting


Chapter 4 - Overhead Tube Support (OTS) Page 89
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.) Tighten the three collimator mounting screws until they are snug and then back out 1/4-1/2
turn. Verify that the collimator is still able to pivot without too much effort when the tube is
rotated and pointing down. Verify that when the collimator is not in a detent location, a
minimum force of 4 lbs. is required to rotate the collimator.
8.) Remove the collimator cable trim cover. Remove the 2 screws to open up the strain relief. See
Figure 4-64

Figure 4-63 Collimator Cable Trim Cover

9.) Attach the collimator cable to the collimator connector. Clamp the cable in the metal strain
relief and secure with 2 screws. Make sure the metal strain relief contacts the cable metal
braid. See Figure 4-64.
10.) Attach the detent switch pigtail (orange and white wires) to the detent switch connector. See
Figure 4-64.
11.) Arrange the collimator cable covering around the strain relief with the end of the covering
positioned at the left side of the strain relief as shown in Figure 4-64. Run a Ty-rap through
the seam at the end of the cable covering and pull it tight around the collimator cables. Trim
any excess Ty-rap.

Page 90 Section 6.0 - Collimator


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 4-64 Connect Collimator Cable

12.) Re-install the trim cover over the collimator cable connector. See Figure 4-65. The end of the
cable covering should be tucked inside the collimator cover. The detent switch wires (orange
and white) will exit the cover from the right side.

Figure 4-65 Install Collimator Cover

Chapter 4 - Overhead Tube Support (OTS) Page 91


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 7.0 User Interface (UIF)


1.) Apply Loctite 242 (Blue) to the two (2) socket head cap screws.
2.) Secure the UIF to the UIF mounting bracket with the two cap screws. See Figure 4-66

Figure 4-66 UIF Installation

3.) Connect the cable to the UIF. See Figure 4-67.

Figure 4-67 UIF Cable Connection

Page 92 Section 7.0 - User Interface (UIF)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 8.0 Longitudinal Cable Drape


1.) Attach the auxiliary support to the Unistrut. The auxiliary support is pre-drilled for perpendicular
Unistrut on 26", and 28", centers or it can be attached to an additional parallel Unistrut as
shown in Figure 4-68.
Keep the auxiliary rail at least 3 inches (8 cm) from the wall to allow for cable clamps to clear
the wall.
2.) Attach the longitudinal cable drape rail to the bottom of the auxiliary support.
3.) Install 2 to 4 cable drape hooks in the rail.
4.) Install 2 to 4 clamps around the cable assembly, evenly spaced across the length of the
longitudinal cable assembly. These clamps are the same as the one used at the middle of the
lateral cable drape.
5.) Attach the cable clamps to the cable drape hooks.

PARALLEL
UNISTRUT

AUXILIARY
SUPPORT:
46−177615P3
&
46−177615P4

CABLE DRAPE
RAIL

Figure 4-68 Longitudinal Cable Drape Installation

Chapter 4 - Overhead Tube Support (OTS) Page 93


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 9.0 Initial OTS Counterpoise Balance


1.) Position the vertical column so that it is in the middle of the vertical range of travel.
2.) If the vertical column tends to drift upward or downward when released, adjust the
counterpoise tensioning bolt to eliminate drift (clockwise rotation results in more lift).
Throughout the range of vertical column travel, the tube/collimator shoud not drift
downward. See Figure 4-69.

Figure 4-69 OTS Counterpoise Tensioning Bolt

Section 10.0 OTS Jumpers and Switches


OTS jumpers and switches are factory configured. If desired, the jumper and switch positions can
be confirmed by referring to the Definium 8000 System Schematics and Drawings manual
(5137438-100).

Page 94 Section 9.0 - Initial OTS Counterpoise Balance


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 5 Table

Section 1.0 Summary Checklist

Unpack table

Perform initial table-to-bridge alignment

Drill table anchor holes

Level table, re-check alignment to bridge, mount to


floor

Adjust table pedal stabilizing screws

Install detector and ion chamber in table

Install chiller in table

Fill table chiller with coolant (FE task)

Route and connect table cables and hoses

Install table top

Section 2.0 Unpack Table


This section will describe the basic steps required to un-crate the table.
1.) Remove the top cover and four top cross members from the table shipping crate. See Figure 5-
1.

Figure 5-1 TABLE CRATE WITH TOP REMOVED

Chapter 5 - Table Page 95


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.) With the help of another person, remove the table base-plate and table top and store them in
a safe place. See Figure 5-2.

Figure 5-2 TABLE TOP REMOVED

3.) Remove any additional cross members from the inside of the crate. See Figure 5-3

Figure 5-3 CRATE CROSS MEMBERS

4.) Remove the 4 sides of the shipping crate. See Figure 5-4.

Page 96 Section 2.0 - Unpack Table


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 5-4 REMOVE CRATE SIDES

5.) Remove the wooden ramp from the shipping crate.

Figure 5-5 REMOVE WOODEN RAMP

6.) Remove the protective covering. See Figure 5-6.

Figure 5-6 REMOVING PROTECTIVE COVERING

7.) Remove all braces used to secure the table base. See Figure 5-7.

Chapter 5 - Table Page 97


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 5-7 REMOVING BRACES

8.) Release the vertical brake by moving and holding the manual release lever. See Figure 5-8.
9.) Use a hex socket wrench S19 to turn the hex-head bolt at the end of the vertical drive motor.
See Figure 5-8. Clockwise rotation of the motor will lift the table base above the surface of the
shipping crate. Continue turning the motor until the table base is about 1-2 inches (25 - 50 mm)
above the shipping crate. Once you feel increased resistance with turning the motor, then you
have reached the limit of travel; do not continue turning the motor or you may damage the
motor assembly.

Figure 5-8 TABLE DOLLY

10.) Position the wooden ramp at the end of the shipping crate. See Figure 5-9. Carefully guide the

Page 98 Section 2.0 - Unpack Table


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
table dolly down the wooden ramp until the table has reached the floor.

Figure 5-9 WOODEN RAMP PLACEMENT

11.) Move table into room and begin the alignment process. The table dolly casters may be rotated
for better clearance through doorways.

Chapter 5 - Table Page 99


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 3.0 Initial Table and Bridge Alignment


Note: Install the table baseplate (See Figure 5-13) BEFORE any table cables are pulled.

The OTS bridge must be installed and able to move longitudinally in the stationary rails.
1.) Move the table to the location specified on the plan drawings. Leave the table on the dollies
but lower the table base all the way down to the floor.
2.) Move the detector housing to the foot (right) end of the table.
3.) Hang a plumb bob from the outside of the bridge rail (not carriage) and position it just above
the detector barrier. See Figure 5-10.

Figure 5-10 PLUMB BOB ATTACHED TO BRIDGE

4.) Place a piece of masking tape on the detector barrier and directly beneath the plumb bob. See
Figure 5-11.

Figure 5-11 POSITION PLUMB BOB OVER TAPE

Page 100 Section 3.0 - Initial Table and Bridge Alignment


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
5.) Mark the position of the plumb bob on the tape. See Figure 5-12.

Figure 5-12 MARK PLUMB BOB LOCATION

6.) Move the detector and bridge longitudinally to the head end. Position the bridge so that the
plumb bob is over the mark made when at the foot end.
7.) Carefully move the head end of the table laterally to position it beneath the plumb bop. Try not
to move the foot end of the table.
8.) Re-check the alignment at the foot-end and head-end and adjust the table position until the
plumb bob is on the mark at one end of travel and within +/- 2 mm (0.08 in) of the mark at
the other end of travel.
9.) The table should now be positioned close to the final mounting location. Re-check the table
position to verify it is at the location specified by the site drawings. If necessary re-position and
repeat the initial alignment.

Section 4.0 Install Table


4.1 Drill Anchor Holes
1.) Use a center punch or marking pen to mark the location of the 4 mounting holes in the corners
of the table base.
2.) Raise the table and move it away from the mounting location.
3.) Place the table base-plate on the floor and align it with the markings for the anchor holes. Use
the table base-plate to confirm the proper locations for the anchor holes and remark the hole
locations if necessary. Move the table base-plate out of the way before drilling the anchor
holes. See Figure 5-13.

Figure 5-13 TABLE BASEPLATE

Chapter 5 - Table Page 101


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.) Use a masonry bit and impact drill to drill four (4) 12mm holes for the anchor bolts, M8 X 190
mm (2400546, supplied).

Note: Anchors require a minimum engagement of 80 mm (3.15 in.) into the concrete. If the floor
thickness is less than 4” it is recommended that the unit should be secured using a through-
bolt method with a reinforcement plate on the back side.

5.) Remove all construction debris from table mounting location.

4.2 Position Table


1.) Place table base-plate on floor and center it around the anchor holes in the floor.
2.) Reposition the table over the table base-plate and anchor holes.
3.) Lower the table onto the table base-plate and remove the lifting fixtures.
4.) Install the 4 mounting bolts through the centers of the four table base adjusters and into the
anchor holes in the floor. See Figure 5-14. Do not tighten the mounting bolts until the table
has been leveled.

Figure 5-14 TABLE MOUNTING BOLTS

Page 102 Section 4.0 - Install Table


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
5.) Check the levelness of the table (use a 4-foot level across the table top bearings) in both
lateral and longitudinal directions. See Figure 5-15 and Figure 5-16.

Figure 5-15 TABLE BASE LEVELING IN LATERAL DIRECTION

Figure 5-16 TABLE BASE LEVELING IN LONGITUDINAL DIRECTION

6.) Adjust the table base leveling adjusters (See Figure 5-17) until the table is level in both lateral
and longitudinal directions.

Chapter 5 - Table Page 103


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.) Tighten the table base leveling adjuster locknuts. See Figure 5-17.

Figure 5-17 TABLE BASE LEVELING ADJUSTERS

4.3 Re-check Alignment and Anchors


1.) Repeat the table to bridge alignment procedure (Initial Table and Bridge Alignment, Section
3.0 on page 100). The clearance between the anchors and the mounting holes should allow
for minor alignment adjustments.
2.) Repeat the table leveling procedure until the table is level.
3.) Repeat the table to bridge alignment procedure and the table leveling procedure until both are
within the required specifications.
4.) Torque all 4 anchors to 25 N-m (220 in-lbs).

4.4 Adjust Pedal Stabilizing Screws


1.) On the front pedal assembly, remove the 3 flathead screws to access the pedal stabilizing
Allen screws. See Figure 5-18.
2.) Adjust the Allen stabilizing screws until they just touch the table baseplate.
3.) Re-install the 3 flathead screws.
4.) Repeat the above steps for the rear pedal assembly.

Figure 5-18 PEDAL STABILIZING SCREWS

Page 104 Section 4.0 - Install Table


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 5.0 Detector and Ion Chamber


1.) Remove the detector shipping bracket. See Figure 5-19.

Figure 5-19 DETECTOR SHIPPING BRACKET

2.) Remove the mylar detector barrier. See Figure 5-20.


a.) Remove four screws and the front trim strip.
b.) Remove three screws (each) and remove the left and right trim strips.
c.) Remove six screws and the rear trim strip.
d.) Remove the mylar detector barrier.

Figure 5-20 DETECTOR ASSEMBLY

Chapter 5 - Table Page 105


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.) Remove the left end cover from the detector housing. See Figure 5-21.

Figure 5-21 REMOVE DETECTOR END COVER

4.) Secure the ion chamber pre-amp with a nut at the foot end and a screw at the head end. Then
connect the ion chamber cable to the pre-amp. See Figure 5-22.

Figure 5-22 PRE-AMP IN DETECTOR HOUSING

Page 106 Section 5.0 - Detector and Ion Chamber


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
5.) Move the detector assembly until the pre-amp head end mounting screw is over the access
hole in the table frame. Install the screw from under the table frame. See Figure 5-23.

Figure 5-23 PRE-AMP IN DETECTOR HOUSING

6.) Route cables and water hoses out of the access opening on the bottom of the housing. See
Figure 5-22.

Potential for Equipment Damage: The detector is VERY fragile. Use extreme caution when
installing the detector in the detector housing. Use the cable handles provided. Do NOT
handle the detector by the hoses or cables.

Potential for Potential for Finger Pinch Injury when sliding detector into the housing.
Injury.
7.) Remove the 4 detector centering capscrews (2 on front and 2 on rear of detector housing). See
Figure 5-27.
8.) Slide the detector into the detector housing while guiding the detector cables and hoses over
the lower rail. See Figure 5-24.

Chapter 5 - Table Page 107


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
9.) If present, remove the two handles from the detector. Store the handles in a convenient place
in the event the detector may need to be removed at some later date. Be sure to re-install the
handle screws in the detector housing.

Figure 5-24 DETECTOR MOUNTED DETECTOR HOUSING

10.) Through the access opening in the bottom of the detector housing, connect the following
cables and hoses in this order: See Figure 5-25. The excess lengths of the cables/hoses
should be carefully pushed back up through the access opening.
- Ethernet cable
- Detector power cable
- 2 water hoses (not direction sensitive)
- Ground cable (attached to side of detector housing)

Figure 5-25 ION CHAMBER & GROUND CONNECTION

11.) Install the ion chamber using two corner spacers (not reversible) and two side spacers
(grooved spacer in left front position). See Figure 5-25.

Page 108 Section 5.0 - Detector and Ion Chamber


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
12.) Position the pre-amp to ion chamber wires in the grooves in the left front spacer. See Figure 5-
26.

Figure 5-26 GROOVED SPACER

13.) Install the left end cover on the detector housing without pinching the wires. See Figure 5-21.

Potential for Equipment Damage: In the next step, you must use the 4 detector
centering capscrews that were originally removed from the detector housing. Do not
substitute longer capscrews or the detector will be damaged.

14.) Install and tighten the 4 detector centering capscrews (2 on front and 2 on rear of detector
housing). See Figure 5-27.

Figure 5-27 DETECTOR CENTERING SCREWS

Chapter 5 - Table Page 109


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
15.) Position the mylar barrier on the top of the detector housing and install with the 16 screws and
4 trim strips. See Figure 5-28.

Figure 5-28 MYLAR DETECTOR BARRIER AND TRIM STRIPS

16.) Install the Grid/IQST assemblies as shown in Figure 5-29.

Figure 5-29 Grid Detection Optical Switches

17.) The "Grid In" and "IQST In" optical switches should be adjusted to the specifications shown in
Figure 5-30. Brackets 2270512 and 2270512-2 have vertical slots for adjustment.

Page 110 Section 5.0 - Detector and Ion Chamber


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
18.) Adjust switch to allow the vane on the GRID/IQST frame to breach the switch gap with
sufficient clearance as not to have interference between the vane and the switch.

Figure 5-30 Grid Detection Optical Switch Adjustment Specifications

Section 6.0 Chiller


6.1 Table Chiller Mechanical Installation
Note: This section is to be performed by the mechanical installer.

1.) Place chiller at head end of table base and connect the two water lines to the rear of the chiller
unit. See Figure 5-31.

Figure 5-31 WATER LINE CONNECTIONS

Chapter 5 - Table Page 111


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.) Install chiller in table base as shown in Figure 5-32.

Figure 5-32 TABLE CHILLER CABLE CONNECTIONS

3.) Connect MIS 020004 to J1 on chiller. See Figure 5-32.


4.) Connect internal power cable to the chiller. See Figure 5-32.
5.) Connect table ground wires to brass grounding stud. See Figure 5-32 and Figure 5-33. The
ground connections should be made exactly as shown in Figure 5-33.

Brass Ground Stud

Brass Nut
Brass Lock-Washer
MIS GND
Cable

Table GND
Cables
Brass Washer

Table Base
Paint Mask Area

Figure 5-33 TABLE GROUND CONNECTIONS

Page 112 Section 6.0 - Chiller


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.2 Fill Table Chiller with Coolant
Note: This section is to be performed by the local field engineer.

Always use a funnel when adding coolant to the chiller. The coolant provided for filling the chiller is
water protected by a fungicide. In its undiluted form this fungicide is toxic, but in the solution
provided it is very dilute and is not dangerous to health when handled correctly. However, when
handling the solution you should treat it with care and avoid unnecessary contact (wear goggles).
Waste solution in the concentration used will not cause pollution if disposed in a non-sensitive
environment. See below for First Aid advice.

Note: 1.) In case of contact with chiller coolant:


- Immediately flush eyes or skin with plenty of water for at least 15 minutes.
- In case of eye contact, call a physician immediately.
- If swallowed, drink one to two glasses of water or milk. Avoid alcohol. Call a physician
immediately.
- Remove and wash contaminated clothing before reuse.
2.) When filling or topping up the unit, take great care to keep water away from the vents, the ON/
OFF switch and the power connection. Do not immerse the unit in any liquid.
3.) The fungicide used in the coolant is toxic. Dispose of old surplus coolant with care.

1.) Remove the reservoir cap located on the top of the detector chiller.
2.) Place paper towels on chiller to prevent coolant spills from entering chiller.
3.) Using a funnel, pour coolant liquid into the reservoir until the coolant level is about 1” (2.5 cm)
from the top of the reservoir.
4.) After system powerup, run the chiller for 5 minutes then recheck the level and add additional
coolant if necessary.
5.) Reinstall the cap on the reservoir and tighten it securely.

Section 7.0 Table Cables


Note: This section should be performed by a professional mechanical installer. The MIS cable
connections between the Table and the System Cabinet cannot be made until the system MIS
cables have been pulled.

7.1 Routing (Under Table Entry)


Two access holes are provided in the XR/d table for cable entry. The access at the rear of the head
end is for surface raceway and the access in the bottom is for under table channeling.
If the cables to the table are to be run under the table and up through the access provided in the
bottom of the table they should be routed through the table as shown in Figure 5-34 and Figure 5-
35.

Chapter 5 - Table Page 113


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 5-34 TABLE CABLE ROUTING

Figure 5-35 TABLE CABLE ROUTING

7.2 Power Connections


The power cable (MIS 020005) is connected to J1.

Page 114 Section 7.0 - Table Cables


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 5-36 Table Power Connection J1

Section 8.0 Table Top


Note: This section should be performed by mechanical installer.

The table top is installed by sliding the top onto the table bearing rails from either the head end or
the foot end of the table. This requires a minimum of 2200 mm of clearance from the head or foot
edge of the table frame. If there is enough clearance, install the table top according to the method
described in Section 8.1 (page 115). If there is not enough clearance, install the table top using the
alternate method described in Section 8.2 (page 116).

8.1 Typical Room


1.) If installed, remove the rubber bumper from one end of the table top.

Potential for Equipment Damage: Avoid damage to the Table Top Collision Sensor
Switches during table top installation.

2.) Carefully position the table top on the edge of the table frame. Do NOT begin to slide the top
onto the bearing rails.

Potential for Potential for Finger Pinch Injury when installing table top. Use a tool to depress
Injury. switches and magnetic locks.

3.) Depress the two magnetic locks (rotate to allow up/down movement) and the actuators of the
two table top collision sensor switches. See Figure 5-37.

Chapter 5 - Table Page 115


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 5-37 TABLE TOP INSTALLATION

4.) Slowly slide the table top over the locks and switches being careful not to damage the switch
actuators or pinch your fingers.
5.) Slide the top towards the opposite end of the table, stopping before the magnetic locks and
collision sensor switches are encountered.
6.) Depress the two magnetic locks (rotate to allow up/down movement) and the actuators of the
two table top collision sensor switches.
7.) Slowly slide the table top over the locks and switches being careful not to damage the switch
actuators or pinch your fingers.
8.) Install the rubber bumper to prevent the top from accidentally moving beyond the end bearings.
See Figure 5-38.

Figure 5-38 TABLE TOP BUMPER INSTALLATION (Bottom View)

8.2 Small Rooms (Alternate Method)


1.) Remove the front and rear bearing rails from the table top carriage. See Figure 5-39 and

Page 116 Section 8.0 - Table Top


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Figure 5-40. The front rail has horizontal rollers for lateral alignment.

Figure 5-39 BEARING RAILS

Figure 5-40 BEARING RAIL MOUNTING BOLTS

2.) Insert the removed bearing rails into the rail slots on either side of the table top. Keep the top
side of rail towards the top of table top and large vertical bearing to the outside of table.
3.) Carefully pick up the table top while keeping the bearing rails from sliding down the rail slots
and pinching your fingers.
4.) Position the top over the table top carriage with the front bearing rail (with horizontal rollers) to
the front of the table.
5.) Lower top onto carriage, while aligning bearing rails with the table top carriage.
6.) Slide the top towards either end and expose the bearing rail mounting holes. Keep extended
end supported.
7.) Insert and lightly tighten the (4) mounting bolts (2/per rail). See Figure 5-40.
8.) Slide the table top towards the magnetic locks and collision sensor switches.
9.) Depress the two magnetic locks (rotate to allow up/down movement) and the actuators of the
two table top collision sensor switches. See Figure 5-41.

Chapter 5 - Table Page 117


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 5-41 TABLE TOP INSTALLATION

10.) Slide the table top towards the opposite end to expose the other mounting holes.
11.) Install the other mounting bolts and align the end and outer edge of the rail with the carriage.
See Figure 5-41.
12.) Tighten the four exposed mounting bolts.
13.) Carefully slide the table top over the magnetic locks and collision sensor switches to expose
the first mounting bolts.
14.) Depress the two magnetic locks (rotate to allow up/down movement) and actuators of the (2)
table top collision sensor switches. See Figure 5-41.
15.) Repeat the alignment process for the first end and tighten the four exposed mounting bolts.
16.) Install the rubber bumper to prevent the top from accidentally moving beyond the end bearings.
See Figure 5-42.

Figure 5-42 TABLE TOP BUMPER INSTALLATION (Bottom View)

Page 118 Section 8.0 - Table Top


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 6 Wall Stand & Extended Wall Stand

Section 1.0 Summary Checklist

Unpack wall stand

Position wall stand in room and level wall stand

Drill rear mounting hole in floor and mount wall stand to floor

Install wall stand detector and ion chamber

Attach cables/hoses to detector, ion chamber and wall stand base

Check wall stand jumpers and switches (factory set)

Attach display

Install display overlay (FE task)

Set initial wall stand counterbalance

Perform initial mechanical alignment between wall stand and OTS

Drill remaining mounting holes in floor and install remaining anchors

Perform final wall stand to OTS alignment with mirror (FE task)

Section 2.0 Prerequisites


This section assumes that the table has been mechanically installed and is aligned longitudinally to
the OTS, and the collimator is level.

Chapter 6 - Wall Stand & Extended Wall Stand Page 119


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 3.0 Unpack Wall Stand


1.) Remove the top cover of the shipping container. See Figure 6-1.

Figure 6-1 TOP COVER

2.) Remove the 2 top cross braces. See Figure 6-2.

Figure 6-2 TOP CROSS BRACES

3.) Unpack and store the covers.

Page 120 Section 3.0 - Unpack Wall Stand


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.) Turn caster bolt clockwise to allow the removal of cross brace. See Figure 6-3. It may be
necessary to back off the lock nut.

Figure 6-3 LOWER CASTERS

5.) Remove the 3 lower cross braces. See Figure 6-4.

Figure 6-4 LOWER CROSS BRACES

Chapter 6 - Wall Stand & Extended Wall Stand Page 121


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.) Remove the two end panels of the shipping container. The end panel with ramps (at base end
of stand) will be used to move the stand off the skid base. See Figure 6-5.

Figure 6-5 PANEL WITH RAMPS

7.) Position ramp as shown in Figure 6-6. Use the cross braces removed earlier to temporarily
raise the end of the ramp.

Figure 6-6 PANEL WITH RAMPS

8.) Remove any other parts from the shipping container:


- Counterweights
- Lateral Bar
- Hand Grips
9.) Adjust caster bolts to raise the casters until they are above the bottom of the base-plate. This
will prevent the stand from rolling when it is raised.

Page 122 Section 3.0 - Unpack Wall Stand


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
10.) Depending on the wall stand that is to be installed, follow either step a) or b).
a.) Wall Stand (Regular Arm Length): Without removing the sides of the shipping
container, carefully raise the unit until it is resting on its base-plate. This a 3 person
operation (1 person to hold the base and 2 people to lift the wall stand to the vertical
position). See Figure 6-7.

Figure 6-7 RAISING STAND

b.) Extended Wall Stand: Remove the sides of the shipping container and carefully raise
the unit until it is resting on its base-plate. This is a 4 person operation (1 person to hold
the base and 3 people to lift the wall stand to the vertical position). See Figure 6-8 for a
display of the crate without sides, and Figure 6-7 for how to raise the wall stand.

Figure 6-8 UNCRATING EXTENDED WALLSTAND

11.) Lower the casters until the baseplate is 1-1/2 inches (38 mm) above the floor.
Chapter 6 - Wall Stand & Extended Wall Stand Page 123
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
12.) Roll the wall stand back into the shipping container and remove the braces used to level the
ramp. See Figure 6-9

Figure 6-9 RAMP LOWERED

13.) Complete this step if you need to tilt the detector carriage to move the wall stand from the
loading dock into the exam room:
a.) Remove the bolts and shipping bracket from underneath the detector housing. See
Figure 6-10.

Figure 6-10 REMOVE DETECTOR SHIPPING BRACKET

Page 124 Section 3.0 - Unpack Wall Stand


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
b.) Hold the detector housing securely and manually release the detector pivot lock by
carefully squeezing both lock plates together. See Figure 6-11.

Figure 6-11 SQUEEZE LOCK PLATES TO RELEASE TILT

c.) With the tilt lock released, tilt the detector to the horizontal position. You may now move
the wall stand through the site to the exam room.

Potential for Potential for Equipment Damage: The Ethernet Cable is extremely fragile. Avoid stepping on
Equipment the cable.
Damage:
14.) The ethernet cable comes already attached to the wall stand. Please take care not to damage
the ethernet cable while moving the wall stand into the room.

Potential for Potential for Equipment Damage: The tilt wheel assembly can damage the cables that exit
Equipment the receptor or tube stand if the tilt wheels are allowed to pivot all the way down. See
Damage: Figure 6-12.

Figure 6-12 CABLE DAMAGE AVOIDANCE

Chapter 6 - Wall Stand & Extended Wall Stand Page 125


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
15.) Pull the tilt wheel assembly away from the vertical column and connect it to the support arm
using bolt supplied on the support arm. See Figure 6-13 and Figure 6-14.

Figure 6-13 TILT WHEEL ASSEMBLY INSTALLATION

Figure 6-14 TILT WHEEL ASSEMBLY INSTALLATION

16.) Turn the casters as shown in Figure 6-14 and tilt the stand backwards until it rests on the tilt
wheels.

Page 126 Section 3.0 - Unpack Wall Stand


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
17.) Roll unit down the ramp and onto the floor. See Figure 6-15.

Figure 6-15 MOVING STAND DOWN RAMP

Note: To make movement of the receptor and tube stands easier during transport to the room or
during alignment, use the support arm of the tilt wheel assembly for added leverage.
18.) The unit may now be moved through areas with height limitations. This arrangement restricts
the freedom of movement and it is recommended that the unit be moved in the vertical position
whenever possible.
19.) After the unit is moved to the site room, remove the tilt wheels and store. See Figure 6-16.
a.) Remove the bolt that connects the support arm to the tilt wheels.
b.) Remove the large bolt that secures the tilt wheels to the baseplate.

Figure 6-16 TILT WHEELS REMOVAL


c.) Do NOT remove the 2 lower brackets that are attached to the baseplate. They will be used
to anchor cables. See Figure 6-16.
d.) Do NOT remove the support arm at this time. It will give added leverage to help position
the wall stand during the alignment process.

Chapter 6 - Wall Stand & Extended Wall Stand Page 127


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 4.0 Install Wall Stand


1.) From room layout drawings, locate wall stand mounting location.

Potential for Potential for Injury if Equipment is not fastened properly: Anchors must be a minimum
Injury. of 150mm from any concrete edge including ducts and cracks. In addition, the general
condition of the concrete in the immediate mounting area should be inspected to
ensure that anchors will be set in good quality concrete.

2.) Move wall stand into position and slowly lower wall stand until baseplate contacts the floor and
the front leveling bolt is directly over the intersection of the lines locating the first hole and rear
leveling bolt is bisected by the longitudinal location line. (Leveling bolts must not extend
beyond the bottom of baseplate.) The support arm can be used for added leverage. See
Figure 6-17.

Figure 6-17 SUPPORT ARM USED FOR POSITIONING

3.) Turn wall stand leveling bolts until they contact the mounting sub-plate.
4.) Remove the 4 caster assemblies from the baseplate.
5.) Completely loosen (but do not remove) the 3 nuts that secure the sub-plate to the baseplate.
See Figure 6-18. This will allow the sub-plate to float away from the baseplate during leveling.

Figure 6-18 SUB-PLATE BOLT

Page 128 Section 4.0 - Install Wall Stand


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.) Attach the supplied plumb line by screwing the bushing into the hole in the bottom of the
horizontal arm. The plumb bob should rest just above the top of the baseplate. See Figure 6-
19, Figure 6-21 and Figure 6-23. Adjust the plumb bob height by tying a knot in the line or
taping it to the carriage.

Figure 6-19 PLUMB LINE ATTACHMENT

Note: If a leveling bolt is turned to raise one side of the stand, the two leveling bolts which affect the
other plane must be adjusted equally until they touch the sub-plate, since they will have been
raised off the sub-plate. Like wise, if the leveling bolt is turned to lower one side of the stand,
the two leveling bolts which affect the other plane must be adjusted upward because the stand
will be pivoting on them.
7.) Remove the support arm by removing the two brackets that mount the support arm to the
vertical column of both stands.
If desired, put the four mounting bolts back in the vertical column. However, be sure to
verify that the bolts do not interfere with the counterweight vertical motion. See
Figure 6-20.

Figure 6-20 SUPPORT ARM REMOVAL

Chapter 6 - Wall Stand & Extended Wall Stand Page 129


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 6-21 PLUMB LINE

8.) Adjust leveling bolts until the wall stand is vertical with the plumb bob directly over the center
punch mark on the top of the baseplate. See Figure 6-22 and Figure 6-23.

Figure 6-22 LEVELING BOLTS

Figure 6-23 LEVELING MARK

Page 130 Section 4.0 - Install Wall Stand


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
9.) Tighten the locking ring on the leveling bolts. See Figure 6-24.

Figure 6-24 LOCKING NUT

10.) Using the leveling bolt as a drill guide, drill one 13 mm (0.5") diameter hole 152 mm (6.0") deep
(4.0" into floor + 2.0" to top of leveling bolt) in the rear position only.
11.) Clean hole of all dust and chips. Make sure that you have adequate depth (no dust).
12.) Insert supplied 1/2 inch x 5-3/4 inches Rawl power-bolt anchor.
13.) Drive the anchor into the hole.
14.) Torque to 68 N-m (600 in-lbs) and then back off several turns to allow the wall stand to be
rotated around the rear anchor.

Section 5.0 Detector and Ion Chamber


1.) Position the carriage to a convenient height for mounting the detector housing.
2.) If present, remove the shipping bracket from the back side of the detector housing. See
Figure 6-10.
3.) On the back of the receptor housing remove the curved support plate that houses the support
pivot for the lateral patient support bar. See Figure 6-25.

Figure 6-25 SUPPORT PLATE

Chapter 6 - Wall Stand & Extended Wall Stand Page 131


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.) Remove the top cover from the detector housing by removing 4 screws on the top, 3 screws
on the back and 7 screws along the top front. See Figure 6-26.

Figure 6-26 TOP COVER REMOVAL

Potential for Potential for Injury from moving parts: Releasing the detector housing pivot lock before the
Injury. detector housing is fully loaded will allow the housing to pivot. Use caution to prevent
injuries.

5.) Hold the detector housing securely and manually release the detector pivot lock by carefully
squeezing both lock plates together. See Figure 6-27.

Figure 6-27 PIVOT LOCK RELEASE

6.) With the tilt lock released, tilt the detector to the horizontal position.

Note: Be careful not to drop the cover mounting screws inside the detector housing.

Page 132 Section 5.0 - Detector and Ion Chamber


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.) Remove the seven screws that secure the mylar barrier/grid holder to the detector cover. See
Figure 6-28.

Figure 6-28 DETECTOR HOUSING COVER REMOVAL

8.) Carefully remove the top cover. The rubber gasket will make the top stick to the housing frame.
Do not damage the gasket.
9.) With the tilt lock released, tilt the detector to the vertical position.
10.) Pull the hoses and cables out of the access hole. See Figure 6-29.

Figure 6-29 PULL HOSE OUT

Chapter 6 - Wall Stand & Extended Wall Stand Page 133


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
11.) Connect the ion chamber cable to the pre-amp. See Figure 6-30.

Figure 6-30 ION CHAMBER CABLE

12.) Position the ion chamber pre-amp over the flat head screw and secure with screw through the
back of the detector housing. See Figure 6-31 and Figure 6-32. Only the pre-amp is installed
at this point; the ion chamber should be placed on the arm housing and will be installed after
the detector has been installed.

Figure 6-31 ION CHAMBER PRE-AMP INSTALLATION

Page 134 Section 5.0 - Detector and Ion Chamber


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 6-32 ION CHAMBER PRE-AMP INSTALLATION

Potential for Potential for Equipment Damage: The detector is VERY fragile. Use extreme caution when
Equipment installing the detector in the detector housing. Use the cable handles provided. Do NOT
Damage handle the detector by the hoses or cables.

13.) Carefully remove the detector from the shipping containers and inspect the external surfaces
of the detector for any foreign objects. Clean the detector, if necessary, to remove any foreign
objects.
14.) Slide detector assembly into receptor housing as shown in Figure 6-33. Be careful not to pinch
or kink cables and hoses. Do not remove the socket head screws; they are used as guides and
fit in the grooves in the side channels.

Figure 6-33 DETECTOR INSTALLATION

Chapter 6 - Wall Stand & Extended Wall Stand Page 135


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
15.) Lower the detector all the way into the detector housing being careful not to allow the cables
or water hoses to be trapped under the detector. When the detector is fully seated in the
detector housing, there is approximately 10 mm between the top of the detector and the top of
the detector housing. See Figure 6-34.

Figure 6-34 DETECTOR INSTALLATION

16.) The two detector handles do not need to be removed from the top of the detector. Loosen the
four detector handle screws, rotate the handles as shown in Figure 6-35, then tighten the
screws. If you choose to remove the handles, re-install the screws in the detector housing.

Figure 6-35 DETECTOR HANDLE RE-POSITIONING

17.) Install the ion chamber into the detector housing, in front of the detector. See Figure 6-36.

Page 136 Section 5.0 - Detector and Ion Chamber


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 6-36 ION CHAMBER INSTALLATION (3-CELL ION CHAMBER SHOWN)

18.) For 4-Cell ion chamber patient barrier, install as shown in Figure 6-37

Figure 6-37 PATIENT BARRIER INSTALLATION

Chapter 6 - Wall Stand & Extended Wall Stand Page 137


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 6.0 Cables


1.) Make the following connections to the wall stand detector. See Figure 6-38.
- Ethernet cable
- Water lines
- Detector power cable

Figure 6-38 DETECTOR CONNECTIONS

2.) Connect the ground cable. See Figure 6-39.

Figure 6-39 DETECTOR GROUND CONNECTION

Page 138 Section 6.0 - Cables


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.) Install the cover on the detector housing.

Note: Tighten the 4 screws on the top of the cover and the 4 screws on the rear of the cover. Place
the detector horizontal to install the seven screws along the top front edge.

Figure 6-40 TOP COVER INSTALLATION

4.) Other cable connections (see Figure 6-41):


- J1 - Wall Stand Power Cord MIS 11756A
- J2 - Detector Power MIS 020020
- J3 - Foot Switch Note: Foot Switch cable is labelled as "J4". Please
connect to J3 plug.
- J4 - Ion Chamber MIS 020021
- J7 - Wall Stand CAN MIS 020018
- J8 - DSA CAN MIS 020019

Figure 6-41 Wall Stand Connections

Chapter 6 - Wall Stand & Extended Wall Stand Page 139


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 7.0 Wall Stand Jumpers and Switches


Wall Stand jumpers and switches are factory configured. If desired, the jumper and switch positions
can be confirmed by referring to the Definium 8000 Renewal Parts manual (5137439-100),
Disassembly and Reassembly section.

Section 8.0 Display


8.1 Mechanical Installation
Note: This section is to be performed by the mechanical installer.

The wall stand display assembly must be attached to the top of the bracket that holds the remote
control arm.
1.) Position the display as shown in Figure 6-42.

Figure 6-42 DISPLAY INSTALLATION

2.) Secure display with two screws (screws are shipped in the display overlay bag). The other two
countersunk holes are used when the display and remote control arm are mounted on the right
side of the wall stand.

Figure 6-43 DISPLAY INSTALLATION

Page 140 Section 7.0 - Wall Stand Jumpers and Switches


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
8.2 Install Display Overlay
Note: This step is to be performed by the local field engineer.

1.) The display overlay should be attached to the display after power has been applied to the
system. The overlay may block the display LEDs if it is not aligned correctly. After power has
been applied to the system, place the overlay in position on the display to determine the best
position. Remove the adhesive backing and attach the overlay to the display.

Section 9.0 Set Initial Wall Stand Counter Balance


1.) Position the receptor carriage to gain access to the counterweight box.

Figure 6-44 COUNTERWEIGHT INSTALLATION

2.) Install counterweights until the counterweight box is 75% full. Final counterweighting will be
done later after all covers are installed.

Chapter 6 - Wall Stand & Extended Wall Stand Page 141


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 10.0 Initial Mechanical Alignment


1.) Attach a plumb bob to the OTS.
2.) Place the receptor housing in the horizontal position and position a 24" carpenter’s square
as shown in Figure 6-45.

Figure 6-45 POSITIONING SQUARE

3.) If the wall stand is mounted off the head or foot end of the table:
- Clamp the OTS carriage to prevent it from traveling laterally across the detector.
4.) If the wall stand is mounted off the front or back wall:
- Clamp the OTS bridge to prevent it from traveling longitudinally across the detector.
5.) Move the OTS to make the plumb bob travel along the edge of the T-square. (Longitudinally
with respect to the table if the wall stand is mounted off the head or foot end of the table and
laterally with respect to the table if the wall stand is mounted off the front or back wall). See
Figure 6-46.

Figure 6-46 OTS ALIGNMENT USING PLUMB BOB

6.) Pivot the wall stand about the rear mounting hole as necessary to get the OTS plumb bob to
travel directly over the edge of the T-square.
7.) Verify the wall stand is still vertical using the plumb bob.

Page 142 Section 10.0 - Initial Mechanical Alignment


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 11.0 Wall Stand Anchors


1.) Using the leveling bolts as a drill guide, drill three 13 mm (0.5") diameter holes 152 mm (6.0")
deep (4.0" into floor + 2.0" to top of leveling bolt) with masonry drill.

Note: Anchors require a minimum engagement of 64mm (2 1/2 in.) into the concrete. If
reinforcement are encountered, alternate mounting locations are provided. Remove the
leveling assembly from the original location, move it to one of the alternate locations, check
leveling, and re-drill the mounting hole. See Figure 6-47.
If the floor thickness is less than 4" it is recommended that the unit should be secured using a
through bolt method with a reinforcement plate on the back side.

Figure 6-47 ALTERNATE MOUNTING HOLES

2.) Clean holes of all dust and chips. Make sure that you have adequate depth (no dust).
3.) Insert supplied 1/2 inch x 5-3/4 inches Rawl anchors.
4.) Drive the anchors into the hole.

Note: Check and adjust plumb alignments as anchors are tightened. This is very important if the
surface is a material that can be compressed.

5.) Torque all 4 anchors to 27 N-m (240 in-lbs), then to 41 N-m (360 in-lbs), then to 54 N-m (480
in-lbs) and finally to 68 N-m (600 in-lbs).

Note: If an anchor is loosened after it has once been torqued to its 68 N-m (600 in-lbs) it should only
be re-torqued to 41 N-m (360 in-lbs).

Chapter 6 - Wall Stand & Extended Wall Stand Page 143


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 12.0 Final OTS/Tube/WS Alignment


The following procedures must be done before performing positioner calibration of the wall stand.

Note: This chapter is to be performed by the local field engineer.

12.1 X-ray Beam to SID Travel Alignment


The X-ray beam must be parallel to the axis used to vary the wall stand SID to assure proper
alignment at all SIDs.
1.) Set the the keyswitch on the rear of the UIF to ’override’ position (yellow and green LEDs flash
on the left side of the UIF).
2.) Rotate the X-ray tube and the OTS column until the collimator faces the wall stand. It is
important to position the X-ray tube and column in the same manner as it will be used at the
site.
3.) Using a level, verify that the patient barrier is in the vertical position (0 degrees). See Figure 6-
48.

Figure 6-48 Patient Barrier Vertical (0 degrees) and Horizontal (90 degrees)

4.) Using a level, verify that the rails of the collimator are vertical.
5.) Position the OTS for a SID of 100 cm (if using tape measure, set to 95 cm).
6.) With the collimator crosshair shining on the center of the patient barrier (at 0 degree position),,
move the OTS to vary the SID while preventing the OTS from moving horizontally (left to right)
across the barrier.
7.) Verify that the collimator crosshair does not shift more than 3 mm horizontally as the SID is
varied from 100 cm to 180 cm (if using tape measure, vary from 95 cm to 175 cm).
8.) If adjustment is necessary, perform the OTS Column Rotation calibration. If the alignment is
still off, recheck Wall Stand to Stationary Rail alignment in Section 10.0 Initial Mechanical
Alignment on page 142.
9.) Position the OTS for a SID of 100 cm (if using tape measure, set to 95 cm).
10.) With the laser line shining on the center of the patient barrier (at 90 degree position),, move

Page 144 Section 12.0 - Final OTS/Tube/WS Alignment


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
the OTS to vary the SID while preventing the OTS from moving vertically (up or down) across
the barrier.
11.) Verify that the laser line does not shift more than 3 mm vertically as the SID is varied from 100
cm to 180 cm (if using tape measure, vary from 95 cm to 175 cm).
12.) If adjustment is necessary, perform the X-ray Tube Angulation calibration. Another possible
cause of excess vertical shift can be improperly leveled stationary rails.

12.2 Detector Perpendicular to X-ray Beam Adjustment


1.) Move the wall stand detector until it is approximately in the middle of its vertical travel.
2.) Rotate the collimator until the laser light projects a vertical line on the wall stand.
3.) Attach a mirror to the face of the patient barrier. See Figure 6-49.

Figure 6-49 Mirror on Detector Patient Barrier

4.) Turn on the laser and position the OTS so the laser line hits the mirror.

Potential for damage to eyesight. Do not look directly into source or reflected laser beam.

5.) Observe the reflected laser beam line at the collimator. Use a piece of paper to see the line.
See Figure 6-50.

Chapter 6 - Wall Stand & Extended Wall Stand Page 145


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 6-50 Reflected Line Within Limits

6.) Rotate the wall stand until the reflected line is within the edges of the opening of the laser
source. This assures that the detector is perpendicular to the X-ray beam.
7.) Move the OTS to vary the SID from 100 cm to 180 cm. The reflected line must stay within the
edges of the opening of the laser source.
8.) Remove mirror from wall stand.

12.3 Vertical Tracking Adjustment


1.) Set the the keyswitch on the rear of the UIF to ’normal’ position (yellow and green LEDs no
longer flashing on the left side of the UIF).
1.) Position the OTS for a SID of 180 cm.
2.) Lower the detector to its minimum height and vertically position the OTS to maintain alignment.
3.) Center the laser light on the patient barrier.
4.) Raise the detector to its maximum height and vertically position the OTS to maintain alignment.
5.) Note the location of the laser light on the patient barrier.
6.) If the light shifts laterally (left to right) as the vertical height is varied, it will be necessary to
adjust the lateral tilt of the wall stand so that it tracks the offset of the OTS as the vertical
position of the OTS is varied. When adjusting the lateral tilt of the wall stand, do not allow the
front-to-rear tilt of the wall stand to change from vertical.

Page 146 Section 12.0 - Final OTS/Tube/WS Alignment


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 7 Digital Support Assembly (DSA)

Section 1.0 Summary Checklist

Attach cables/hoses to DSA

Fill chiller with coolant (FE task)

Section 2.0 Mechanical Installation


Note: This section is to be performed by the mechanical installer. The MIS cable connections between
the DSA and the System Cabinet cannot be made until the system MIS cables have been pulled.

1.) Position the DSA, (chiller and power supply assembly) less cover, on the desired side of the
wall stand.
2.) Confirm that the detector power supply (located on top of the DSA) has the correct DIP switch
settings for the wall stand. Refer for Direction 5137438-100 (Definium 8000 Schematics &
Drawings), Chapter 2 (System), Section 3 (Jumpers, Switches, LEDs).
3.) Remove the top half of the DSA cable clamp. See Figure 7-3.
4.) Feed ground wire through the cable entrance opening and attach to ground stud.
5.) Feed cables with connectors through the cable opening starting with the largest connector.
6.) Connect to chiller and power supply per the MIS map. See Figure 7-1 and Figure 7-2 for
connector identification.

Figure 7-1 DSA CONNECTIONS

Chapter 7 - Digital Support Assembly (DSA) Page 147


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 7-2 DSA CHILLER CONNECTIONS

7.) Feed the water hoses (2) through the cable opening and connect per MIS map and Figure 7-2.
8.) Insert the hoses and cables going to the DSA into the slit duct hose.
9.) Position the duct hose so the slit is hidden (towards the floor).
10.) Secure duct hose to DSA by inserting hose under the lips on the hose/cable clamp casting on
the DSA, then install the top half of the cable clamp and secure with 2 screws
11.) Route hoses and cables so that they do not extend beyond the edge of the DSA baseplate.

Potential for Potential for Equipment Damage: Arrange the hoses so they are not pinched when
Equipment cable clamp is secured. See Figure 7-3 for suggested hose orientation.
Damage

Figure 7-3 DSA CABLE INSTALLATION

Page 148 Section 2.0 - Mechanical Installation


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 3.0 Fill DSA Chiller with Coolant


Note: This section is to be performed by the local field engineer.

Always use a funnel when adding coolant to the chiller. The coolant provided for filling the chiller is
water protected by a fungicide. In its undiluted form this fungicide is toxic, but in the solution
provided it is very dilute and is not dangerous to health when handled correctly. However, when
handling the solution you should treat it with care and avoid unnecessary contact (wear goggles).
Waste solution in the concentration used will not cause pollution if disposed in a non-sensitive
environment. See below for First Aid advice.

Note: • First Aid in case of contact with chiller coolant:


- Immediately flush eyes or skin with plenty of water for at least 15 minutes.
- In case of eye contact, call a physician immediately.
- If swallowed, drink one to two glasses of water or milk. Avoid alcohol. Call a physician
immediately.
- Remove and wash contaminated clothing before reuse.
• When filling or topping up the unit, take great care to keep water away from the vents, the ON/
OFF switch and the power connection. Do not immerse the unit in any liquid.
• The fungicide used in the coolant is toxic. Dispose of old surplus coolant with care.

1.) Remove the reservoir cap located on the top of the detector chiller. See Figure 7-4.

Figure 7-4 REMOVE CHILLER CAP


2.) Using a funnel, pour coolant liquid into the reservoir until the coolant level is about 1” (2.5 cm)
from the top of the reservoir. See Figure 7-5.

Figure 7-5 FILL CHILLER WITH COOLANT


3.) After system powerup, run the chiller for 5 minutes then recheck the level and add additional
coolant if necessary.
4.) Reinstall the cap on the reservoir and tighten it securely.

Chapter 7 - Digital Support Assembly (DSA) Page 149


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 150 Section 3.0 - Fill DSA Chiller with Coolant


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 8 Workstation
Note: System cables cannot be connected to the computer and RCIM until they have been pulled.

Section 1.0 Summary Checklist

Place computer on work desk, connect cables

Place mouse and keyboard on work desk, connect cables

Place RCIM & handswitch on work desk, connect cables

Place LCD monitors on work desk, connect cables

Place Bar Code Reader (Purchased Option) on work desk, connect cables

Chapter 8 - Workstation Page 151


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 2.0 Computer


2.1 HP xw8200 Workstation
1.) Placement
2.) Connections:

Power

Keyboard

4 Key 3
4 Key 3

6
6

2
2

1
1

5
5
Do NOT use
USB: lower USB ports!
-Mouse
-Barcode Rdr Hospital
Network

Ground
J1 - System J2 - JEDI
CAN CAN
Video
Serial B- Serial A-
DSA Chiller Table
Chiller
J4 -
Wallstand
Detector
J5 -
Table
Detector

Figure 8-1 MAGIC PC CONNECTIONS (HP 8200)

a.) Ethernet:
* External (Hospital Network)
* Table Detector MIS 020003
* Wall Stand Detector) MIS 020002
b.) CAN Bus:
* Jedi (System I/O Board) MIS 11767A
* System (CAN Bulkhead) MIS 11768A
c.) Serial:
* DSA Chiller MIS 020011
* Table Chiller MIS 020010
d.) Power & Ground MIS 11770A and 11776A (Connect to wall plate only)

2.2 HP xw8400 Workstation


1.) Placement
2.) Connections:

Page 152 Section 2.0 - Computer


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Power

Keyboard
(English
only)

4
4

6
6

2
2

Key 3
Key 3

1
1

5
5
USB: Do NOT use
-Keyboard* lower USB ports!
-Mouse
-Barcode Rdr
Hospital
Network
Serial B- Ground
DSA Serial A-
Chiller Table
Video 2 Chiller
Video 1
J4 -
Wallstand J5 -
Detector Table
Detector
J1 -
System J2 - JEDI
CAN CAN

*USB Keyboard (non-English)

Figure 8-2 MAGIC PC CONNECTIONS (HP 8400)

a.) Ethernet:
* External (Hospital Network)
* Table Detector MIS 020003
* Wall Stand Detector) MIS 020002
b.) CAN Bus:
* Jedi (System I/O Board) MIS 11767A
* System (CAN Bulkhead) MIS 11768A
c.) Serial:
* DSA Chiller MIS 020011
* Table Chiller MIS 020010
d.) Power & Ground MIS 11770A and 11776A (Connect to wall plate only)

Section 3.0 Mouse and Keyboard


1.) Placement
2.) Connections
- Attach mouse to computer USB connector.
Chapter 8 - Workstation Page 153
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
- Attach keyboard to computer USB (non-English) or PS2 (English-only) connector .

Section 4.0 RCIM & Handswitch


1.) Placement
2.) Connections
- Digital I/O for RCIM, to System Cabinet
3.) RCIM Overlays for International Use
a.) Remove the existing overlay from the RCIM.
b.) Locate the appropriate language overlay (shipped with the RCIM).
c.) Attach the overlay to the RCIM.
d.) Verify that all RCIM buttons operate without interference.

Section 5.0 LCD Monitors


1.) Placement
2.) Connections:
a.) Power & Ground (Connect to wall plate only)
b.) Video (DVI)
i.) With the HP xw8200 Workstation:
> Plug the video Y-cable into the rear of the workstation PC. See Figure 8-1 and
Figure 8-3.

Figure 8-3 VIDEO Y-CABLE

> Connect DVI video cables to INPUT 1 on the left and right LCD monitors. See
Figure 8-4.
> Connect the left LCD monitor video cable to output "1" of the video Y-cable.
> Connect the right LCD monitor video cable to output "2" of the video Y-cable.
ii.) With the HP xw8400 Workstation:
> Connect video cables to D-SUB on the left and right LCD monitors. See Figure 8-
4.
> Connect the left LCD monitor video cable to DVI video "2" on the Magic PC. See
Figure 8-2.
> Connect the right LCD monitor video cable to DVI video "1" on the Magic PC. See
Figure 8-2.

Page 154 Section 4.0 - RCIM & Handswitch


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
DVI-D D-SUB DVI-D D-SUB
REVIEW MONITOR ACQUISITION MONITOR
(RIGHT) (LEFT)

FOR HP xw8200, CONNECT FOR HP xw8200, CONNECT


VIDEO CABLE FROM VIDEO CABLE FROM
OUTPUT “2” ON VIDEO OUTPUT “1” ON VIDEO
Y-CABLE TO INPUT “DVI-D” Y-CABLE TO INPUT “DVI-D”
ON THE MONITOR. ON THE MONITOR.

FOR HP xw8400, CONNECT FOR HP xw8400, CONNECT


VIDEO CABLE FROM VIDEO CABLE FROM
OUTPUT VIDEO “1” AT MAGIC OUTPUT VIDEO “2” AT MAGIC
PC TO D-SUB CONNECTOR PC TO D-SUB CONNECTOR
ON THE MONITOR. ON THE MONITOR.

DVI-D DVI-I D-SUB

Figure 8-4 LCD MONITOR VIDEO CABLE CONNECTIONS

Section 6.0 Bar Code Reader (Purchased Option)


1.) Placement
2.) Connections: Attach bar code reader to computer USB connector.

Chapter 8 - Workstation Page 155


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 156 Section 6.0 - Bar Code Reader (Purchased Option)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 9 Cables

Section 1.0 Summary Checklist

Pull all MIS cables between sub-systems

Connect MIS cables to (installed) sub-systems

Section 2.0 Routing


• System MIS cables should be pulled to/from the System Cabinet before the System Cabinet
is moved into the final mounting position.
• Some system cables are pre-connected to a sub-system and should not be disconnected in
order to pull the cable(s) through the conduits or wall ducts. Therefore, the following cables
should be pulled in the directions listed below:
- Cables between the Wall Stand and the System Cabinet should be pulled from the Wall
Stand to the System Cabinet.
- Cables between the Table and the System Cabinet should be pulled from the Table to
the System Cabinet.
- Cables between the Wall Stand and the Magic PC should be pulled from the Wall Stand
(through the System Cabinet) to the Magic PC.
• If excess cables must be coiled and stored, they must be coiled in a “figure 8" configuration,
NOT a single coil stack.
• Power cables must be kept separate from signal cables as much as possible. In wall ducts or
raceway, power cables should be routed in a separate channel from signal cables. Inside the
System Cabinet, power cables should not be bundled together with signal cables.

Section 3.0 System Cables


Note: Refer to the following document for details on system cabling: 5137438-100, Definium 8000 -
System Schematics and Drawings (this document contains the MIS Map).

1.) Perform routing and connection of system, power and ground cables.
2.) All MIS cables connecting to the Magic PC will pass through the wallbox (installed by electrical
contractor) and wall plate (supplied by GE with set of MIS cables, installed on the wallbox. See
Figure 9-1.

Chapter 9 - Cables Page 157


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 9-1 Magic PC Wall Plate

Section 4.0 EMC Compliance


4.1 Ferrite Beads
If Ferrite Beads are removed from cables in order to ease cable pulling, the beads must be re-
installed exactly as they were prior to removing them. See Figure 9-2.

Figure 9-2 Ferrite Bead

Page 158 Section 4.0 - EMC Compliance


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.2 Grounding P-Clamps
1.) For EMC compliance, several cables require additional grounding through the use of P-clamps
connected to the cable shields. It is important that these P-clamps connections are made
properly in order to provide the necessary electrical ground. When more than one P-clamp
must be attached at the same mounting location, refer to Figure 9-3 and Figure 9-4 for correct
orientation of the P-clamps.

Figure 9-3 Correct Installation of Multiple P-Clamps

Figure 9-4 Incorrect Installation of Multiple P-Clamps

Chapter 9 - Cables Page 159


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 160 Section 4.0 - EMC Compliance


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 10 System Cabinet

Section 1.0 Summary Checklist

Position cabinet near mounting position

Pull MIS cables into cabinet

Mount cabinet to wall/floor

Connect MIS cables to cabinet

Connect P-clamps to cabinet cables

Connect cabinet to wiring for door interlock and room warning light

Section 2.0 Procedure


Items required:
Standard Service Tools

Parts:
• Wall mounting bracket (supplied with cabinet).
• Customer or contractor supplied wall and floor anchoring devices that meet local requirements
(seismic, etc.).
• Customer or contractor supplied wiring/cable conduits, fittings, etc.

Initial Conditions:
Before installation work begins, the site should be checked to verify the correctness of junction
box sizes, their locations, the number and sizes of interconnecting wiring and proper power
availability in the room. In addition, cabinet mounting methods should already be determined
prior to cabinet installation (customer supplied attaching devices). Refer to Direction 5137435-
100, Definium 8000 Pre-Installation.

Procedure:
1.) Unbolt the wall mount bracket from the top of the System Cabinet. If the cabinet will be
mounted flush to the wall, then the wall mount bracket should be bolted flush to the rear of the
cabinet as shown in Figure 10-1. If the cabinet cannot be mounted flush to the wall due to an
obstruction (such as wall-mounted duct-work or raceway), then the wall mount bracket should
be bolted to the cabinet in the extended position by using the other set of mounting holes.
2.) The System Cabinet will be shipped attached to shipping dollies. Position the cabinet (while
on the dollies) according to your room plan.
3.) Route the system cabinet MIS cables through the cable access opening (rear, bottom or side)
of the cabinet. Lower the cabinet until contact is made with the floor.

Chapter 10 - System Cabinet Page 161


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.) Remove the front dolly retaining cables and open the front cabinet door.
5.) Using the base of the cabinet as a template, mark the front mounting holes on the floor.
6.) Re-install the front dolly retaining cables, raise the dolly from the floor, and move the cabinet
from the mounting area. Use care not to damage the MIS cables.
7.) Drill mounting holes to fit the required cabinet anchors (3/8" x 2").
8.) Re-position the cabinet (on the dollies) over the mounting holes and lower the cabinet to make
contact with the floor. Remove the dollies from the cabinet.
9.) Install locally approved floor anchoring devices through the cabinet base into the floor.
10.) Attach the cabinet to the wall with the supplied bracket using locally approved wall anchoring
devices.

Figure 10-1 System Cabinet Mounting Plate (Flush-Mount Position)

11.) Insert the plastic push-in plugs into the threaded lifting bolt holes on the top of the system
cabinet. See Figure 10-2. The plugs can be found inside the system cabinet, either attached
to the side of the cabinet or inside a plastic bag.

Figure 10-2 System Cabinet Plugs

12.) Connect cables to System Cabinet as shown in Figure 10-3 through Figure 10-7. Refer to the
system MIS Map (see System Schematics and Drawings manual, 5137438-100) for details.
- Bulkhead and ground connections. See Figure 10-3.
- P-clamp connections on bulkhead. See Figure 10-4.
- High voltage cables. See Figure 10-5. Check the HV cable wells to verify that the proper
level of transformer oil is present and fill to a depth 10 mm (0.375 in.) as necessary.
- PDU room interface connections. See Figure 10-6, Figure 10-7 and Table 10-1.
- PDU power and ground connections. See Figure 10-6.

Page 162 Section 2.0 - Procedure


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 10-3 System Cabinet Cables - Bulkhead and Ground Connections

13.) After connecting the stator cable (MIS 020013) to the System Cabinet bulkhead (SKL1-A25),
install the 2 P-clamps as shown in Figure 10-4. The P-clamps are shipped attached to the hex
standoff on the A25 bulkhead .

Figure 10-4 P-Clamp Location on System Cabinet Bulkhead

Chapter 10 - System Cabinet Page 163


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 10-5 System Cabinet Cables - High Voltage Cable Connections

Figure 10-6 System Cabinet Cables - PDU Connections

14.) Make connections to the PDU for the door interlock and X-ray On warning light circuits. These
room interface connections are provided by the TB2 terminal block. See Table 10-1,
Figure 10-6 and Figure 10-7.

Risk of minor The power for the X-ray On warning light circuit is provided by the customer. Therefore,
injury from low these wires may already have 24VAC present during installation. Use caution that you do
voltage not touch the conductors or allow them to come in contact with any system cabinet
components or ground.

Page 164 Section 2.0 - Procedure


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Position Signal Name


1 Door Interlock
2 24VDC
3 Room Light Switch (not used)
4 Room Light Power - provided by the hospital room (not used)
5 X-ray ON Light Switch
6 X-ray ON Light Power (24VAC provided by the hospital room)
7 24VDC (not used)
8 Emergency Power Off (not used)
Table 10-1 Room Interface Signals on TB2 Terminal Block

Chapter 10 - System Cabinet Page 165


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Connect the Customer
supplied breaker box to the
System Emergency Off (SEO) button
using Customer supplied wires.
3 Phase & GND Use either normally "closed" or normally
"open" contacts, depending on the box.

SEO, 600 V max AC/DC 10 amp

Feeder wire and grounding cable supplied Normally


closed
by the customer. Wires are to be provided Customer supplied
by customer with inlet to SKL with 79 in. breaker box Normally
(2m) for internal cabinet routing. open

Emergency Shutoff Kit - obtain


from GE (see Note 1 below)

3 Phase & GND

Normally “open” relay


contact in Cabinet
SKL
(System Cabinet)
DLK
Customer provided and Wire supplied with PDU
installed door switch Emergency Shutoff Kit TB2
(See Note 1 below) 1 2 3 4 5 6 7 8

24 VDC

Warning Light
XR On Light Supply
220/120
XRL XR On Light Switch 24 VAC
to 24 VAC
Max. Current Interface
10A, 20A Peak
Note 1: The Emergency Shutoff Kit (GE Part # A8091JH) is
220/120VAC
shipped with the system. The kit includes emergency shutoff
switch/nameplate and 60m spool of shielded interconnect cable
used for hookup of door interlock switch to System Cabinet PDU.

Figure 10-7 PDU Room Interface Wiring

Page 166 Section 2.0 - Procedure


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
15.) Remove the bubble-wrap from around the 2 ferrite beads. One is located on the back of the
swingout panel (Figure 10-8) and the other is located above the JEDI control assembly
(Figure 10-9).

Figure 10-8 Ferrite Bead on Rear of Swingout Panel

Figure 10-9 Ferrite Bead on JEDI Control Assembly

Chapter 10 - System Cabinet Page 167


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 168 Section 2.0 - Procedure


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 11 System Termination Jumpers

Section 1.0 Summary Checklist

Check/set CAN bus and RT bus terminations

Section 2.0 CAN Bus Terminations


In the previous product (XR/d - 2X Platform), a CAN loopback jumper was required on the System
Cabinet A25 bulkhead for any CAN device that was not present. Also, for a Wall Stand-only
configuration (XR/d - 2X Platform), a CAN loopback jumper was required on the Wall Stand I/O
board. On the Definium 8000 system, these loopback jumpers are NOT
required. Do not install any CAN loopback jumpers on the System Cabinet
A25 bulkhead or Wall Stand I/O board. Remove any existing CAN loopback
jumpers from the System Cabinet A25 bulkhead.

Section 3.0 RT Bus Terminations


3.1 Wall Stand-Only Configuration
System Cabinet: E-stop Loopback jumper installed on A25 J110 (TABLE RT BUS).

Chapter 11 - System Termination Jumpers Page 169


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 170 Section 3.0 - RT Bus Terminations


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 12 Grid Holder Installation


The grid holder will hold three grids (one from the table and two from the wall stand, the IQST
phantom holder, the wall stand patient support bar, and the flat field test phantom. See Figure 12-1.

Figure 12-1 GRID HOLDER

The following additional items will be needed to install the grid holder:
• Anchors or other mounting screws
• Hand drill

Note: Observe the following:


The mounting panel is designed with numerous mounting holes to allow mounting the grid
holder assembly to two studs at almost any spacing.
The suggested height from the floor to the top of the grid holder assembly is approximately 30
inches (76 cm), depending upon hospital personnel. This means the top of the mounting panel
should be 28 inches (71 cm) from the floor.

1.) Locate the wall studs in the area where the grid holder will be mounted.

Chapter 12 - Grid Holder Installation Page 171


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.) Using the mounting panel as a template, level the panel at the desired mounting height and
mark on the wall where the mounting holes align with the wall studs or where the anchoring
devices are to be installed. See Figure 12-2.

Figure 12-2 MOUNTING PANEL

3.) Drill holes where mounting screws or anchors are to be installed.


4.) Install anchors if used.
5.) Secure mounting panel to the wall. See Figure 12-2.
6.) Install the grid holder assembly over the top lip on the mounting panel. As the assembly is
lowered into position, make sure the stud on the bottom of the panel protrudes through the tab
on the bottom of the grid holder assembly. See Figure 12-3.

Figure 12-3 MOUNTING STUD

7.) Secure the grid holder assembly to the mounting panel with a 6 mm nut and washer provided.

Page 172 Section 3.0 - RT Bus Terminations


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
8.) Install the unused table and wall stand grids, the IQST phantom holder, patient support bar and
flat field phantom as shown in Figure 12-1 and Figure 12-4.

Figure 12-4 GRID AND IQST PHANTOM IN GRID HOLDER

Chapter 12 - Grid Holder Installation Page 173


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 174 Section 3.0 - RT Bus Terminations


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 13 Image Pasting Barrier (option)

Section 1.0 Summary Checklist

Align template to wall stand base

Mark sets of image pasting barrier mounting holes (standard or extended wall
stand) on floor

Drill holes and mount receptor brackets

Attach measurement labels to handles and base

Section 2.0 Procedure

DO NOT DRILL THE IMAGE PASTING BARRIER MOUNTING HOLES MUST NOT BE DRILLED UNTIL THE
HOLES UNTIL FINAL WALLSTAND ALIGNMENT HAS BEEN COMPLETED BY THE LOCAL FIELD
WALL STAND ENGINEER.
ALIGNMENT IS
DONE
1.) Align the U-side of the metal template (Figure 13-1) to the wall stand floor plate.

PLACE AGAINST FRONT EDGE


OF WALLSTAND BASE

STANDARD ARM
MOUNTING HOLES

EXENDED ARM
MOUNTING HOLES

Figure 13-1 Image Pasting Barrier Drilling Template

2.) The template should touch the front and sides of the wall stand floor plate. Make sure the
template is not cocked before you start drilling holes in the floor.
3.) The template has two sets of mounting holes:
- one set of holes for a regular wall stand

Chapter 13 - Image Pasting Barrier (option) Page 175


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
- one set of holes for an extended wall stand
Use the appropriate set of mounting holes for the wall stand you have.
4.) With the template in place, use a 6 mm drill bit to drill the pilot hole for the center hole and the
three 6 mm holes surrounding the pilot hole. Repeat for the other side.
5.) Remove the template. Use the pilot hole in the center to drill a 1” diameter hole, 2” deep.
Repeat for the other side.
6.) Insert the 6 plastic anchors into the 6 mm holes.
7.) Center the adjuster plates (Figure 13-2) over the 6 mm mounting holes.
8.) On the receptor bracket, back out the 3 adjustment set screws until the end of each set screw
is even with the bottom surface of the receptor bracket.
9.) Insert the metal pin on the receptor brackets (Figure 13-2) through the adjuster plates and into
the 1” holes and align with the three 6 mm mounting holes.

Adjuster Plate

Figure 13-2 Receptor Bracket & Plug and Adjuster Plate

10.) Install the 3 screws through each bracket and into the mounting holes, but do not completely
tighten.
11.) Wheel the barrier (Figure 13-3) over to the receptor brackets and and confirm that the barrier
locking pins align with the receptor brackets. If the alignment is acceptable, tighten the screws
you have access to. Remove the barrier and finish tightening the rest of the screws. The
receptor bracket plugs (Figure 13-2) should be placed into the receptor bracket to prevent
debris from falling into the holes when the Image Pasting Barrier is not in use.

Figure 13-3 Image Pasting Barrier

12.) If the barrier locking pins do not align with the receptor brackets:
a.) Remove the barrier.
b.) Unscrew 1 set of receptor screws, leaving the receptor bracket in the hole.

Page 176 Section 2.0 - Procedure


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
c.) Bring the barrier back over the receptor brackets and engage the barrier pins.
d.) Measure the distance between the barrier and the wall stand on both sides. Align the
barrier so that both measurements are equal.
e.) Rotate the receptor 60 degrees and mark the locations for three new 6 mm holes.
f.) Drill three new holes and mount the receptor bracket into place.
13.) There are 3 leveling set screws on each receptor bracket. You will only level the receptor
brackets if the floor is so severely out of level that the barrier pins cannot be inserted into the
receptor brackets.
14.) Once the receptor brackets are secured, the barrier may be rolled into position and the pins
locked into the receptor brackets. Press on both of the side foot pedals to lock the barrier into
place.
15.) Install the image pasting barrier measurement labels as follows:

Note: • The labels give a measurement of the distance between the center of the patient and the mylar
surface. As such, the labels should be attached so that the lower numbers are closer to the
mylar surface.
• Before placing the labels, clean the sides of the lateral arm bars and the recessed area in the
footstand with Windex, Formula 409, or soap and water. Dry thoroughly.

a.) Install the footstand labels:


* The footstand labels are the wider labels.
* Based on the customer’s preference, select the metric or english measurement
labels.
* Place the labels in the recessed areas on the footstand with the lower numbered
ends closest to the mylar surface. See Figure 13-4.

Figure 13-4 Footstand Labels

Chapter 13 - Image Pasting Barrier (option) Page 177


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
b.) Install the lateral arm labels (these labels may already be factory-installed):
* The lateral arm labels are the narrower labels.
* These should go on the sides of the lateral arms when the arms are rotated all the
way down as shown in Figure 13-5.
* To gauge the position along the arm for the label, the 45 cm mark should be 25 mm
or 1 inch from the end of the arm as shown in Figure 13-6.
* If not using the metric mark, the 18-inch mark should be ¾ of an inch from the end of
the arm.

Image
Pasting
Barrier

Lateral
Arm

Label Label
Location Location

Figure 13-5 Lateral Arm Labels

Figure 13-6 Lateral Arm Label Placement

Page 178 Section 2.0 - Procedure


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 14 Flexi-DT Stretcher (option)

Section 1.0 Summary Checklist

Unpack stretcher

Install handles

Measure and mark lodging holes on floor

Drill holes and mount lodgings in floor

Section 2.0 Hardware and Test Equipment


The following list summarizes the tools and test equipment needed to install the digital table Flexi-
DT:
• Wrench set
• Electric hammer drill model TE76 or similar and a 69 mm drill bit
• Digital multimeter
• 4 ft level (or two standard levels)

Section 3.0 Floor Preparation


The Flexi-DT has a mechanism named PINS, to fix the table into the floor. These PINS go into the
pin locations, or lodgings.
The lodgings must be installed in the floor, once the table has been unpacked. See Section 8.0 Pin
Lodging Installation on page 182.

Note: • The floor where the table is going to be used or displaced along must be rigid enough to
support table weight. The recommended floor material is wood or ceramic.
• Floor requirements: the floor where the table is to be installed must be flat and with a minimum
thickness of 150 mm to enable lodgings installation.
• Before installing the Flexi-DT, the room must comply with all data specified in PRE-
INSTALLATION MANUAL.

Section 4.0 Equipment Evaluation


Preliminary Checks:
• Do not discard any packing material such as envelopes, boxes, bags until all parts are
compared to the packing list.

Chapter 14 - Flexi-DT Stretcher (option) Page 179


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
• Check the general condition and external appearance of all parts for possible damages or
missing items. Notify the distribution center immediately of any damage or missing parts.

Section 5.0 Flexi-DT Unpacking


The digital table Flexi-DT is shipped in 1 crate to facilitate transport and installation. See
specification chapter for crate dimensions.

5.0.1 UNPACKING PROCEDURE


1.) Move the digital table, in the shipping container, to the installation area. Leave the container in
the horizontal position.
2.) Remove the wood panels covering the equipment that are clamped to the wooden pallet.
3.) The two lateral packing wood panels are designed to be used as a ramp. Remove thse covers
and put them on the floor. See Figure 14-2.
4.) Remove the handles, cable, and documentation below table.
5.) Loosen the screws shown in Figure 14-1. Three screws at each side of the table.
6.) Roll off the table carefully using the ramps. The brakes shown in Figure 14-2 must be in the
horizontal position to allow wheel movement.
7.) Move the table to the radiology room.
8.) Set the table to the floor or keep it mobile.
9.) Free the tabletop by removing the red shipping bracket. See Figure 14-3.

Figure 14-1 Table Shipping Crate

Figure 14-2 Table Shipping Crate Ramps

Page 180 Section 5.0 - Flexi-DT Unpacking


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 14-3 Shipping Bracket

Section 6.0 Table Top Handle


The Flexi-DT handles come disassembled. To assemble, follow the procedure shown in the Flexi-
DT Service Manual SM-S0015068 (located on the Definium 8000 Service Documentation CDROM
5137441-200), Disassembly/Reassembly chapter. See Figure 14-4.

Figure 14-4 Table Top Handle

Section 7.0 Pin Installation


Some Flexi-DT models come with the pin disassembled. To assemble, follow the procedure shown
in the Flexi-DT Service Manual SM-S0015068 (located on the Definium 8000 Service
Documentation CDROM 5137441-200), Disassembly/Reassembly chapter.

Chapter 14 - Flexi-DT Stretcher (option) Page 181


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 8.0 Pin Lodging Installation

Figure 14-5 Table Pin Lodging

8.1 Drilling the Floor


Two 69 mm diameter holes for pin lodgings must be done in the floor. The tool recommended for
these holes is a drill machine with a 69 mm drill bit.
The holes should be drilled when the Wall Stand is already in place. See Section 12.0 Drawings on
page 186.
See Section 12.0 Drawings on page 186 for the distance between pins and lodging dimensions.

8.2 Pin Lodging Installation Procedure


Note: To install lodgings correctly, it is necessary to mark the screw holes with the table fixed, to assure
that the table pins fits exactly.
1.) Place the lodging into the floor holes.
2.) Fix the two pin into the lodging by pressing the lever.
3.) Mark the screw drill holes.
4.) Unfix the pin and the lodgings.
5.) Proceed to make the drill holes.
6.) Fix the lodgings with the screws (three screws for each lodging).

Page 182 Section 8.0 - Pin Lodging Installation


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 9.0 Power Cable


The cable shown in Figure 14-6 (S0009262) is provided with the Flexi-DT. It has 6 meters of length.
The table connector is a standard IEC plug and the final is a 3-terminal strip. Service personnel
must place the desired connector.
Wall plug location:

Figure 14-6 Table Power Cable

Note: The table plug is located on the right side of table. It is customer responsibility to provide the wall
plug at the adequate distance for connecting the power cable.

Section 10.0 Battery Recharging


The batteries must be charged for the table operation. Connect the power cable S0009262 to
charge the batteries before use.
The maximum recharging time for the batteries is 12 hours.

Note: The equipment can be used while the batteries are being recharged. The indicator of charging is
located on the front of the control console.

The table is powered by gel batteries, but it can be powered by means of the power cable supplied.
Check the battery level before use and recharge them if necessary.
When the integrity of the cable is in doubt, the equipment will be actuated from its internal batteries.

Chapter 14 - Flexi-DT Stretcher (option) Page 183


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 11.0 Options Installation


11.1 Handgrips
The handgrips must be placed one at each side of the tabletop, as shown in Figure 14-7.
Use the knob to tighten and fix, or loosen and move the handgrip.

Figure 14-7 Table Handgrip

11.2 Compression Bands


Place the compression as shown in Figure 14-8 and Figure 14-9.

Figure 14-8 Table Compression Band

Figure 14-9 Table Compression Band

Page 184 Section 11.0 - Options Installation


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
11.3 Lateral Cassette Holder
This option will be placed by the user. See Figure 14-10.

Figure 14-10 Table Cassette Holder

11.4 Double Pedals


This option is an additional row of pedals to be placed on the back side of the table, in order to
control the movements from that side. This is a field upgradable option.
For part number see the Flexi-DT Service Manual SM-S0015068 (located on the Definium 8000
Service Documentation CDROM 5137441-200), Equipment Description chapter.
See Procedure in the Flexi-DT Service Manual SM-S0015068 (located on the Definium 8000
Service Documentation CDROM 5137441-200), Disassembly/Reassembly chapter.

11.5 Remote RF Control


This option is a upgradable option to control the table movements from a distance.
For part number see the Flexi-DT Service Manual SM-S0015068 (located on the Definium 8000
Service Documentation CDROM 5137441-200), Product Identification section in the Equipment
Description chapter.
See Procedure in the Flexi-DT Service Manual SM-S0015068 (located on the Definium 8000
Service Documentation CDROM 5137441-200), Disassembly/Reassembly chapter.

Chapter 14 - Flexi-DT Stretcher (option) Page 185


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 12.0 Drawings

Figure 14-11 Table Pin Lodging Hole Locations

Figure 14-12 Table Pin Lodging Specifications

Page 186 Section 12.0 - Drawings


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 15 Regulatory Tests (HHS)


Note: This chapter is to be performed by the local field engineer.

Section 1.0 Summary Checklist

Perform X-ray generator tests:

• Generator operator indicators

• Beam quality

• Technic accuracy and linearity

• Reproducibility of exposure

• Automatic Exposure Control (AEC)

Perform Collimator Tests:

• General Tests

• SID Accuracy

• Collimator Field Light Intensity

• Collimator Light Field to X-ray Field Alignment

• Collimator Field Size Indicator Accuracy

• Digital Table Mode - Center to Center Alignment Error

• Digital Table Mode - Size to Size Error (PBL check)

• Digital Wall Stand Mode - Center to Center Alignment Error

• Digital Wall Stand Mode - Size to Size Error (PBL check)

Chapter 15 - Regulatory Tests (HHS) Page 187


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 2.0 HHS Compliance Compatibility List


Product Category Product Description Model Number
• HP Workstation xw8200
Master X-ray Control - Console Magic PC
• HP Workstation xw8400
Generator JEDI Generator Jedi 80 RD 1T
X-ray Tube Housing (Casing) MX-100 46-155400G46
Beam Limiting Device (Collimator) AL01C Auto Collimator 2266999
41cm Digital Detector Digital Rad. Detector 5134673
Table Table 5131070
Wall Stand Wall Stand 5141608
Extended Wall Stand Extended Wall Stand 5141608-2
• E6401J - non-elevating (Americas-only)
Table - Extended Wall Stand Stretcher
Stretcher • E6401K - elevating (Americas-only)
(optional)
• E10811VB - elevating (Europe-only)
Table 15-1 HHS Compliance Compatibility List

Note: The Magic PC part number 5261130 Manufactured On or After August 17, 2007 is the Master
X-ray Control - Console. For all previous system configurations, the Jedi Generator is the Master
Control.

Section 3.0 Customer Information


Fill in the Customer Information data sheet: Customer Information on page 208.

Section 4.0 Test Equipment


Fill in the Test Equipment data sheet: Test Equipment on page 208.

Page 188 Section 2.0 - HHS Compliance Compatibility List


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 5.0 X-ray Generator Tests


5.1 Generator Operator Indicators
Fill in the Generator Operator Indicators data sheet: Generator Operator Indicators on page 209.

5.2 Beam Quality


1.) Position radiation probe as shown in Figure 15-1, "Good Geometry".

Figure 15-1 Half Value Layer Probe Positioning - "Good Geometry"

2.) The optimum position for the standard absorber is midway between the focal spot and the
radiation probe. Other positions will introduce reading errors. For example, if the absorber is
placed close to the probe, then the probe will receive excess scatter and the absorber will
appear to have absorbed less.
3.) Set technic factors as follows:
- 80 kVp, 50 to 100 mA, 1/2 to 1 second
- mA and time should be adjusted so that the reading in step 6 contains three significant
digits.
4.) Remove or dial out all filters that are removable by the operator and collimate slightly larger
than the active volume of the probe using the light field.
5.) Make an exposure and measure exposure with no added filters. Record the reading. This
represents 100% transmission level.
Exposure (without absorber) ___________

Chapter 15 - Regulatory Tests (HHS) Page 189


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.) Place standard absorber 46-173632G1 (or equivalent) in beam. Absorber must be on table
top or in collimator accessory rails.
a.) For systems manufactured after June 10th, 2006, add 0.5 additional Aluminum filtration
using filters from HVL Attenuator part # 46-194427P274. Total aluminum filtration = 3.0
mm.
b.) Attach additional aluminum filtration to probe side of the standard absorber using masking
tape. Ensure tape does not encroach upon usable beam.
c.) Reference the OTS Rating Plate to determine manufactured date. Note: The rating plate
shows month and year only. For systems manufactured during June, 2006 you must
verify the manufactured date in the GE GIB database. Contact your local Service
Representative for assistance.
7.) Make an exposure and measure exposure with standard absorber. Record the reading.
Exposure (with absorber) ___________
8.) Determine the ratio: Ratio = Exposure (with absorber) / Exposure (without absorber)
9.) Fill in the Beam Quality data sheet: Beam Quality on page 209.
If Ratio Is: Corrective Action:
Before June 10th, 2006 0.5 or greater No Change. Test Passes
Requires 2.5 mm Standard Less than 0.5 • Verify kVp Accuracy.
Absorber Part # 46-173632G1 • Verify test setup is correct. This is the
most common error.
• Verify test equipment is functioning
properly and within calibration.
• Issue resides at system level of
Generator, HV Cables, Tube, and
Collimator interactions. Detailed analysis
must be performed to identify failing
component.
On or After June 10th, 2006 0.5 or greater No Change. Test Passes
Requires 2.5 mm Standard Less than 0.5 • Verify kVp Accuracy.
Absorber Part # 46-173632G1 and • Verify test setup is correct. This is the
0.5 mm of additional Aluminum most common error.
filtration using filters from HVL
Attenuator part # 46-194427P274. • Verify test equipment is functioning
Total aluminum filtration = 3.0 mm properly and within calibration.
• Issue resides at system level of
Generator, HV Cables, Tube, and
Collimator interactions. Detailed analysis
must be performed to identify failing
component.
Reference the OTS Rating Plate to determine manufactured date.
Note: The rating plate shows month and year only. For systems manufactured during June 2006 you must
verify the manufactured date in the GE GIB database. Contact your local Service Representative for
assistance.

5.3 Technic Accuracy and Linearity


1.) Turn off system power.
2.) Install an invasive HV divider (i.e., Dynalyzer or equivalent) in series with the X-ray tube.
3.) Turn on system power and boot up system. Wait 1 hour and 15 minutes before making
measurements.

Page 190 Section 5.0 - X-ray Generator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.) Perform X-ray exposures as indicated in the table below, selecting a time that results in a 20
mAs exposure. Record the results in the table below.

kVp and mA/mAs Accuracy and Linearity


Focus Selected Selected Measured Measured mA/mAs kVp
S/L mA/mAs kVp ma/mAs kVp Error Error
S 10 60
S 200 60
S 10 80
S 200 80
S 10 130
S 200 130
L 250 60
L 500 70
L 250 80
L 800 80
L 250 130
L 500 130

kVp Accuracy ±3% for long exposures (>=500 ms) ±(5%+2 kV) for short exposures
Specification
mA Accuracy ±(5%+1 mA) for long exposures (>=500 ms) ±20% for short exposures
Specification

5.) Verify that the errors do not exceed the specifications given in the table above.
6.) Fill in the Technic Accuracy and Linearity data sheet: Technique Accuracy and Linearity on
page 209.

5.4 Reproducibility Of Exposure


5.4.1 Overview
• For any specific combination of selected technique factors, the estimated coefficient of
variation (CV) of radiation exposures must be no greater than 0.05. HHS rejection limit is
0.045.
• Test Equipment: Radiation meter with integrate mode, 1-1/2” (3.8 cm) thickness of Aluminum.
5.4.2 Procedure
1.) Select Digital mode.
2.) Align the OTS with the detector, Lateral Detent ON.
3.) SID = 100 cm.
4.) For Digital table place 1-1/2” (3.8 cm) aluminum on the table, center in the collimator light.
For Digital wall stand, rotate the detector housing to 90 degrees and place the aluminum on
the detector barrier (top surface), center in the collimator light.
5.) Center radiation probe 12 in. (30 cm) above the table top or detector barrier (wall stand).
6.) Collimate to a field size of 5 in. (13 cm) square.
7.) Select 80 kVp, AEC mode and mA such that the exposure will be at least 0.1 second long.
8.) Measure the exposure in mR for each of the 10 exposures. After each exposure, recycle the
technique (switch to other values and then return them to the original values). Also vary the
time between exposures to record any filament heating variations.

Chapter 15 - Regulatory Tests (HHS) Page 191


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Test Results:
Exposure
kVp mA Time mR
Number
1
2
3
4
5
6
7
8
9
10
Coefficient of Variation = ______________ (see formula at the end of this section)
9.) Verify that the coefficient of variation is < 0.045 for AEC exposures.
10.) Fill in the Reproducibility of Exposure - Table data sheet: Reproducibility of Exposure on page
210.
11.) Repeat the above procedure for the wall stand detector. Record the results in the table below.
Test Results
Exposure
kVp mA Time mR
Number
1
2
3
4
5
6
7
8
9
10

Coefficient of Variation = ______________ (see formula at the end of this section)


12.) Verify that the coefficient of variation is < 0.045 for AEC exposures.
13.) Fill in the Reproducibility of Exposure - Wall Stand data sheet: Reproducibility of Exposure on
page 210.

Calculation for Coefficient of Variation (C.V.):


Formula: C.V. = s / X(avg.) where s = sample standard deviation (n-1)
Sample standard deviation (s) can be calculated on a calculator with statistics functions (such
as the TI36Y) as follows:
* Clear the stat register, press (CSR).
* Enter the first reading and then press (Σ+). Repeat this step for all readings.
* Press (mean) to get X(avg.).
* Press sample standard deviation key (n-1).
* Enter the results in the formula above.

Page 192 Section 5.0 - X-ray Generator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
5.5 Automatic Exposure Control (AEC) Tests

5.5.1 AEC Minimum Exposure Time

5.5.1.1 Overview
At 50 kVp or more, minimum exposures must be less than 16 msec.or less than the time to deliver
5 mAs, whichever time is greater.

5.5.1.2 Procedure
1.) Start a Walls Stand digital exam.
2.) Select 50 kVp, AEC (area 1 selected) and maximum mA. 100 msec or 10mAs.
3.) Align the OTS to the detector and collimator is in PBL mode; green LED illuminated.
4.) Lateral Detent ON.
5.) Set the SID to 40 inches.
6.) Remove all phantoms.
7.) Place Wall Stand detector in the vertical position.
8.) Acquire an AEC exposure.
9.) Record the exposure time and repeat the process for each of the three cells (Figure 15-2):
Ion Chamber Cell A (Left) _____________ msec
Ion Chamber Cell C (Right) _____________ msec
Ion Chamber Cell B (Center)_____________ msec

A C

D E

Figure 15-2 Ion Chamber Cells Used in Vertical Position

10.) Verify that the exposure time is less than 16 msec, or less than 5 mAs.

Chapter 15 - Regulatory Tests (HHS) Page 193


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
11.) Repeat the above procedure with the Wall Stand detector in the horizontal position.
Record the exposure time and repeat the process for each of the three cells (Figure 15-3):
Ion Chamber Cell D (Left) _____________ msec
Ion Chamber Cell A (Right) _____________ msec
Ion Chamber Cell B (Center)_____________ msec

C
A

B
D

Figure 15-3 Ion Chamber Cells Used in Horizontal Position

12.) Verify that the exposure time is less than 16 msec, or less than 5 mAs.
13.) Fill in the AEC Minimum Exposure Time data sheet: AEC Minimum Exposure Time on page
210.
14.) Start a Table digital exam.
15.) Select 50 kVp, AEC (area 1 selected) and maximum mA. 100 msec or 10mAs.
16.) Align the OTS to the detector and collimator is in PBL mode; green LED illuminated.
17.) Lateral Detent ON.
18.) Set the SID to 40 inches.
19.) Remove all phantoms.
20.) Acquire an AEC exposure.
21.) Record the exposure time and repeat the process for each of the three cells (Figure 15-4):
Ion Chamber (Left) _____________ msec
Ion Chamber (Right) _____________ msec
Ion Chamber (Center) _____________ msec

Left Right

Center

Figure 15-4 Ion Chamber Cells Used in Table

22.) Verify that the exposure time is less than 16 msec, or less than 5 mAs.
23.) Fill in the AEC Minimum Exposure Time data sheet: AEC Minimum Exposure Time on page
210

Page 194 Section 5.0 - X-ray Generator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
5.5.2 AEC Maximum mAs

5.5.2.1 Overview
• At 50kV or more, exposures shall not exceed 600mAs per exposure.
• Below 50 kV exposures shall not exceed 2,000 mAs.

5.5.2.2 Procedure (50kVp)


1.) Select Digital mode.
2.) Select 50 kVp, AEC (all three areas selected) and an mA that selects 600 mAs.
3.) Position the X-ray tube away from the Ion chamber selected, close collimator blades and select
0.3 mm CU filter.
4.) Perform an exposure.
5.) Record the selected mAs and actual mAs. After exposure, mAs will be displayed on the
monitor screen.
Selected mAs ________________
Measured mAs _______________
6.) Verify that the AEC exposure is less than 600 mAs.
7.) Verify that after the exposure is completed, a backup mAs popup window is displayed. Click
the OK button. Verify popup window is removed and exposures are allowed.
8.) Fill in the AEC Maximum mAs - 600 mAs data sheet: AEC Maximum mAs on page 210.

5.5.2.3 Procedure (40kVp)


1.) Select Digital mode.
2.) Select 40 kVp, AEC (all three areas selected) and an mA that selects 2000mAs.
3.) Position the X-ray tube away from the Ion chamber selected, close collimator blades and select
0.3 mm CU filter.
4.) Perform an exposure.
5.) Record the selected mAs and actual mAs. After exposure, mAs will be displayed on the
monitor screen.
Selected mAs ________________
Measured mAs _______________
6.) Verify that the AEC exposure is less than 2000 mAs.
7.) Verify that after the exposure is completed, a backup mAs popup window is displayed. Click
the OK button. Verify popup window is removed and exposures are allowed.
8.) Fill in the AEC Maximum mAs - 2000 mAs data sheet: AEC Maximum mAs on page 210.

Section 6.0 Collimator Tests


6.1 Introduction

6.1.1 Scope
System level test procedures for the Regulatory requirements of the Digital Radiographic Imaging
System.

Chapter 15 - Regulatory Tests (HHS) Page 195


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.2 General Tests
Fill in the General Tests data sheet: General Tests on page 210.

6.3 SID Accuracy

6.3.1 Overview
The difference between the indicated SID and the actual SID may not exceed 1.8% of the indi-
cated SID.
Measurements will be made between the focal spot locator decal (on the X-ray tube) and the table
top for Digital Table.
Measurements will be made between the focal spot locator decal (on the X-ray tube) and the
patient barrier for Digital Wall Stand.

6.3.2 Procedure for Digital Table


1.) Prepare the Magic PC for an exam.
2.) Select Digital Table mode.
3.) Align the X-ray tube to the detector.
4.) Set the Tube Angle to 0 degrees.
5.) Lateral center detent.
6.) Following Table 15-2 below, set the OTS/Table to the SID position.
7.) Measure the distance from the focal spot label to the table top. Enter the value into the
"Measured" column in Table 15-2. Note: Add 6.6 cm for distance from table top to detector.

8.) Determine the error as follows:


Accuracy = ((Display value-Measured value) / Measured value) × 100
( Display Value – MeasuredValue )
--------------------------------------------------------------------------------------- × 100
Measured Value
SID Displayed Measured Error % of SID Error
102 cm (40 in)
120 cm (47 in)
152 cm (60 in)
Table 15-2 SID Test Data - Table

9.) Verify that the SID error is < or = 1.8% SID.


10.) Fill in the SID Accuracy - Table data sheet: SID Accuracy - Table on page 211.

Page 196 Section 6.0 - Collimator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.3.3 Procedure (Digital Wall Stand)
1.) Prepare the Magic PC for an exam.
2.) Select Digital Wall Stand mode.
3.) Align the X-ray tube to the detector.
4.) Set the Tube Angle to 90 degrees.
5.) Lateral center detent.
6.) Following Table 15-3 below, set the OTS/Wall Stand to the SID position.

Wall SID Displayed Measured Error % of SID Error


Stand
Horizontal 102 cm (40 in)
Horizontal 120 cm (47 in)
Vertical 102 cm (40 in)
Vertical 152 cm (60 in)
Vertical 183 cm (72 in)
Table 15-3 SID Test Data - Wall Stand

7.) Measure the distance from the focal spot label to the patient barrier. Enter the value into the
"Measured" column in Table 15-3.

Note: Add 4.9 cm for distance from the patient barrier to the detector.

Focal
Spot

Patient
Barrier
49 mm

Detector
Determine the error as follows:
Accuracy = ((Display value - Measured value) / Measured value) ×100
( Display Value – MeasuredValue )
--------------------------------------------------------------------------------------- × 100
Measured Value
Verify that the SID error is < or = 1.8% SID.
8.) Fill in the SID Accuracy - Wall Stand data sheet: SID Accuracy - Wall Stand on page 211.

6.4 Collimator Field Light Intensity

6.4.1 Overview
Tools Required: Light meter, tape measure.
The average illumination at a distance of 100 cm (39.37”) from the focal spot shall be no less than
15 foot candles (160 lux).
Verify that the light meter used for this test has an accuracy of 5% or better.

Chapter 15 - Regulatory Tests (HHS) Page 197


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.4.2 Procedure
1.) Place the light meter on a horizontal surface (table top) with sensor directed toward collimator
light source.
2.) Adjust the distance from the X-ray tube focal spot to light meter sensor to 100 cm (39.37”).
3.) Turn on the collimator lamp and adjust the field size to 10” x 10” (25.4 x 25.4 cm).
4.) With the collimator lamp OFF, measure the ambient light level. If possible, room lights should
be turned off during testing.
5.) Record ambient light level measurement______________
6.) Turn collimator Lamp ON.
7.) Centering the light meter in each of the 4 quadrants, measure light level and record data.
Quadrant 1 _____________
Quadrant 2 _____________
Quadrant 3 _____________
Quadrant 4 _____________

Q uadrant 1 Q uadrant 2

Q uadrant 3 Q uadrant 4

8.) Verify that the average illumination at 100 cm from the focal spot is no less than 15 foot-
candles (160 lux) above ambient brightness.
9.) Fill in the Collimator Field Light Intensity data sheet: Collimator Field Light Intensity on page
211.

6.5 Collimator Light Field to X-ray Field Alignment

6.5.1 Overview
Tools Required: Light to X-ray test pattern (part number 46-303843P1)
The total misalignment of the X-ray and light fields in either the lateral or longitudinal directions
may not exceed 1.8%.

6.5.2 Procedure
1.) Prepare the Magic PC for an exam.
2.) Select Digital Table or Digital Wall Stand mode.
3.) Align the X-ray tube to the detector.
4.) Place the light to X-ray test pattern on the table (or tape to the patient barrier on the wall stand).
5.) Collimator in override mode.
6.) Adjust the +/- buttons on the collimator to obtain a SID of 100 cm.

Page 198 Section 6.0 - Collimator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.) Using the collimator dials, adjust the light field to the 10 inch outline square on the test pattern.
Center pattern as necessary.
8.) Select 60 kVp, 200ma, small focal spot, and AEC mode.
9.) Perform a small focal spot exposure.
10.) Select the large focal spot.
11.) Select 60 kVp, 200ma and AEC mode.
12.) Perform an exposure.
13.) Select the RAW image for small focal spot; Using the marks on the test pattern, record the
differences between the X-ray image and the test pattern. Enter the values below for N, S, E
and W.
Small Spot
N S E W
Light Field 5 in. 5 in. 5 in. 5 in.
X-ray Field _______ _______ _______ _______

N+S error = ________


E+W error = ________
14.) Determine the percentage misalignment: (misalignment divided by SID) x 100
( N+S error / SID ) x 100 = __________
( E+W error / SID) x 100 = __________
15.) Verify that the misalignment in both calculations is less than 1.8 percent of SID.
16.) Fill in the Collimator Light Field to X-ray Field - Small data sheet: Collimator Light Field to X-
ray Field Accuracy Check on page 211
17.) Select the Large spot image. Using the marks on the test pattern, record the differences
between the X-ray image and the test pattern. Enter the values below for N, S, E and W.
Large spot
N S E W
Light Field 5 in. 5 in. 5 in. 5 in.
X-ray Field _______ _______ _______ _______
N+S error = ________
E+W error = ________
18.) Determine the percentage misalignment: (misalignment divided by SID) x 100
( N+S error / SID ) x 100 = __________
( E+W error / SID) x 100 = __________
19.) Verify that the misalignment in both calculations is less than 1.8 percent of SID.

Chapter 15 - Regulatory Tests (HHS) Page 199


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
20.) Fill in the Collimator Light Field to X-ray Field - Large data sheet: Collimator Light Field to X-
ray Field Accuracy Check on page 211

6.6 Collimator Field Size Indicator Accuracy

6.6.1 Overview
• The difference between the indicated field size and the actual field size shall not exceed 1.7%
SID in either direction.
• To determine Image Size and Centering, exposures will be completed in Application mode.
• All measurements shall use the RAW Image.

6.6.2 How to Measure


• Select the Measurement tools Icon, select Line. Using the Line measurement tool measure the
size of the X-ray image.
• Use the Line tool to measure the Image at the table top plane. Magnification calculation needs
to be added to measure values.
• Record the measure values.
• Line values are displayed at the bottom right corner of the monitor.
• Measure the distance between front blade and rear blade of the Image = d1.
• Measure the distance between left blade and right blade of the Image = d2.

6.6.2.1 Magnification Calculation

100cm
------------------ = 1.07 Magnification Factor for Table
93.4cm
100cm
------------------ = 1.06 Magnification Factor for Wall Stand
94.1cm

6.6.3 Procedure
1.) Select Digital Mode.
2.) Align the X-ray tube to the detector.
3.) Lateral detent position.
4.) SID =100 cm.
5.) Set X-ray tube angle = 0 deg.
6.) Manually adjust collimator knobs to set collimator FOV to 24 x 24 cm.

Page 200 Section 6.0 - Collimator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.) Take an 80kVp, 50ma, AEC exposure.
8.) Select RAW Image.
9.) Measure actual image height and width. Multiply the dimensions of the image by the
magnification factor to determine the size of the X-ray field at the detector plane.
Height of image ________x Magnification factor = _________ actual size (d1)
Width of image ________ x Magnification factor = _________ actual size (d2)
10.) Calculate indicator to actual size error.
11.) Record the measure values.
(24 cm- actual size d1) (24 cm- actual size d2)
------------------------------------------------------- × 100 and ------------------------------------------------------- × 100
SID SID
where, (d1) = height of image % error ____________________
(d2) = width of image % error_____________________

12.) Verify that the difference between the indicated field size and the actual field size does not
exceed 1.7% SID in either direction.
13.) Fill in the Collimator Field Size Indicator Accuracy data sheet: Collimator Field Size Indicator
Accuracy on page 211.

6.7 Digital Table Mode - Center to Center Alignment Error

6.7.1 Overview
• Misalignment of the center of the X-ray field and the center of the receptor shall not exceed
1.8%.
• To determine the alignment error, exposures will be completed in Application mode.
• All measurements shall use the RAW Image.

6.7.2 How to Measure


• Select the Measurement tools Icon, select Line.
• Using the Line measurement tool, draw diagonal lines on the image to identifying both the
center of the X-ray field and the center of the active area.
• Using the Line measurement tool, Measure the separation of the two centers.
• Record the values in the "Measured" column in Table 15-4 below.
• Line values are displayed at the bottom right corner of the monitor.

6.7.3 Procedure
1.) Select Digital Table mode.
2.) Place the OTS at midrange of longitudinal travel.
3.) Lateral center detent.
4.) Position the OTS to mid-vertical height, position the Table to mid-vertical height.
5.) Position OTS to obtain SID =100 cm.
6.) Set X-ray tube angle = 0 deg; set collimator rotation = 0 deg.
7.) Adjust collimator FOV = 30 x 24 cm.
8.) Verify that the collimator is in PBL mode; green LED illuminated.
9.) Verify that the tracking positioning is completed; detent lights are ON.
10.) Acquire a 80kVp, AEC exposure, small focal spot.

Chapter 15 - Regulatory Tests (HHS) Page 201


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
11.) Select RAW image, using the measurement tool record the separation of the two centers.
12.) Determine the error as follows:
Center to Center Error in centimeters (cm)
Error % = ------------------------------------------------------------------------------------------------------ × 100
(Current SID Value)
Table Height OTS FOV Center to Center Error % of SID Error
Head-end 102 cm 30 x 24 cm
Minimum
Foot-end 102 cm 30 x 24 cm
Minimum
Head-end Max. SID 24 x 30 cm
Maximum
Foot-end Max. SID 24 x 30 cm
Maximum
Table 15-4 Center to Center Error - Table

13.) Verify center to center error is < or = 1.8% SID.


14.) Fill in the Table Center to Center Error data sheet: Table - Center to Center Error on page 212.

6.8 Digital Table Mode - Size to Size Error (PBL check)

6.8.1 Overview
• The difference between the X-ray field size and the nominal size of the image shall not 2.8%
in one direction and 3.8% for both directions combined.
• Maximum longitudinal sizing error in the receptor plane < 2.8% SID.
• Maximum transverse sizing error in the receptor plane < 2.8% SID.
• Maximum sum of longitudinal plus transverse sizing errors in the receptor plane < 3.8% SID.

6.8.2 How to Measure


• To determine the size to size error, exposures will be completed in Application mode.
• All measurements shall use the RAW image.
• Select the Measurement tools Icon, select Line.
• Using the Line measurement tool, measure the size of the X-ray image.
• Line tool measure the Image at the tabletop plane, magnification calculation needs to be added
to measure values.
• Record the measure values.
• Line values are displayed at the bottom right corner of the monitor.
• Measure the distance between front blade and rear blade of the Image = d1.
• Measure the distance between left blade and right blade of the Image = d2.

Page 202 Section 6.0 - Collimator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.8.2.1 Magnification Calculation

100cm
------------------ = 1.07 Magnification Factor
93.4cm
100cm
------------------ = 1.06 Magnification Factor
94.1cm

6.8.3 Procedure

1.) Select Digital Table Mode.


2.) Place the OTS at mid-range of longitudinal travel, lateral detent position.
3.) Set the SID =100 cm.
4.) Set the X-ray tube angle = 0 deg.
5.) Set the collimator FOV = Full FOV (on OTS user interface).
6.) Check that PBL is indicated on collimator face (green LED is on).
7.) Rotate the collimator 45 degrees. Set 41 x 41 cm FOV. Defeat the collimator rotation switch.
8.) Take a 60kVp, 50ma, AEC exposure.
9.) Select RAW image.
10.) Measure actual image height and width. Multiply the dimensions of the image by the
magnification ratio to determine the size of the X-ray at the detector plane.

Height of image ________ x Magnification factor = _________ actual size (d1)


Width of image ________ x Magnification factor = _________ actual size (d2)
Calculate indicator to actual size error.

11.) Record the measure values.

Chapter 15 - Regulatory Tests (HHS) Page 203


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
(41 cm- actual size d1) (41 cm- actual size d2)
------------------------------------------------------- × 100 and ------------------------------------------------------- × 100
SID SID
(d1) = height of image % error ____________________
(d2) = width of image % error _____________________

Height of Width of Height


Width % Total
Image x Image x Height Width % of
FOV SID of SID Error %
Mag. Mag. Error Error SID
Error of SID
Factor Factor Error
41 x 41 cm 102 cm
24 x 24 cm 102 cm
18 x 18 cm 102 cm
Table 15-5 Size to Size Error - Table

12.) Verify that both the height and width size to size error are < or = 2.8%.
13.) Verify that the sum of both is < or = 3.8% SID.
14.) Fill in the Table Size to Size Error data sheet: Table - Size to Size Error on page 212.

6.9 Digital Wall Stand Mode - Center to Center Alignment Error

6.9.1 Overview
• Misalignment of the center of the X-ray field and the center of the receptor shall not exceed
1.8%.
• To determine center to center error, exposures will be completed in Application mode.
• All measurements shall use the RAW image.

6.9.2 Procedure
1.) Select the Measurement tools Icon, select Line.
2.) Using the Line measurement tool, draw diagonal lines on the image to identifying both the
center of the X-ray field and the center of the active area.
3.) Measure the separation of the two centers, using the Line measurement tool.
4.) Record the center-to-center error values in Table 15-6 below.
5.) Line values are displayed at the bottom right corner of the monitor.

6.) Determine the error as follows:

Center to Center Error in centimeters (cm)


Error % = ------------------------------------------------------------------------------------------------------ × 100
Current SID Value

Page 204 Section 6.0 - Collimator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Wall Stand SID/OTS FOV Center to Center Error % of SID Error


Horizontal Min. 102 cm (40 in) 30 x 24 cm
Height
Horizontal 102 cm (40 in) 30 x 24 cm
Max. Height
Vertical Mid. 102 cm (40 in) 30 x 24 cm
Height
Vertical Mid. 183 cm (72 in) 30 x 24 cm
Height
Vertical Min. 183 cm (72 in) 30 x 24 cm
Height
Vertical Max. 183 cm (72 in) 30 x 24 cm
Height
Table 15-6 Center to Center Error - Wall Stand

7.) Verify that the center to center error is < or = 1.8% SID.
8.) Fill in the Wall Stand Center to Center Error data sheet: Wall Stand - Center to Center Error
on page 212.

6.10 Digital Wall Stand Mode - Size to Size Error (PBL check)

6.10.1 Overview
• The difference between the X-ray field size and the nominal size of the image shall not exceed
2.8% in one direction and 3.8% for both directions combined.
• Maximum longitudinal sizing error in the receptor plane < 2.8% SID.
• Maximum transverse sizing error in the receptor plane < 2.8% SID.
• Maximum sum of longitudinal plus transverse sizing errors in the receptor plane < 3.8% SID.
• To determine size to size errors, exposures will be completed in Application mode.
• All measurements shall use the RAW image.

6.10.2 How to Measure


• Select the Measurement tools Icon, select Line.
• Using the Line measurement tool, measure the size of the X-ray image.
• Using the Line tool, measure the Image at the barrier plane. Magnification calculation needs
to be added to measured values.

Chapter 15 - Regulatory Tests (HHS) Page 205


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
6.10.2.1 Magnification Calculation

100cm
------------------ = 1.07 Magnification Factor for Table
93.4cm
100cm
------------------ = 1.06 Magnification Factor for Wall Stand
94.1cm
• Record the measured values in Table 15-7 below.
• Line values are displayed at the bottom right corner of the monitor.
• Measure the distance between front blade and rear blade of the image = d1.
• Measure the distance between left blade and right blade of the image = d2.

6.10.3 Procedure
1.) Rotate the collimator 45 degrees. Set 41 x 41 cm FOV. Defeat the collimator rotation switch
should be disabled.
2.) Take a 60kVp, 50ma, AEC exposure.
3.) Select RAW image.
4.) Measure actual image height and width. Multiply the dimensions of the image by the
magnification ratio to determine the size of the X-ray at the detector plane.
Height of image ________ x Magnification factor = _________ actual size (d1)
Width of image ________ x Magnification factor = _________ actual size (d2)

5.) Calculate indicator to actual size error.


6.) Record the measure values.

(41 cm- actual size d1) (41 cm- actual size d2)
------------------------------------------------------- × 100 and ------------------------------------------------------- × 100
SID SID
(d1) = height of image % error ____________________
(d2) = width of image % error _____________________

Page 206 Section 6.0 - Collimator Tests


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Height of Width of
Height Width Height % of Width % of Total Error
FOV SID Wall Stand Image x Image x
Error Error SID Error SID Error % of SID
Mag. Factor Mag. Factor
41 x 41 cm 102 cm Horizontal,
(40 in) Mid. Height
24 x 24 cm 102 cm Horizontal,
(40 in) Mid. Height
18 x 18 cm 102 cm Horizontal,
(40 in) Mid. Height
41 x 41 cm 102 cm Vertical, Mid.
(40 in) Height
24 x 24 cm 102 cm Vertical, Mid.
(40 in) Height
18 x 18 cm 102 cm Vertical, Mid.
(40 in) Height
41 x 41 cm 183 cm Vertical, Mid.
(72 in) Height
24 x 24 cm 183 cm Vertical, Mid.
(72 in) Height
18 x 18 cm 183 cm Vertical, Mid.
(72 in) Height

Table 15-7 Size to Size Error - Table

7.) Verify that both the height and width size to size error are < or = 2.8%.
8.) Verify that the sum of both is < or = 3.8% SID.
9.) Fill in the Wall Stand Size to Size Error data sheet: Wall Stand - Size to Size Error on page 213.

Chapter 15 - Regulatory Tests (HHS) Page 207


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 7.0 HHS Data Sheets


7.1 Customer Information

GE Service Personnel: Be sure that you return FDA2579 to Environmental EHS. Failure to do
so, except in case of a tube change, is a violation of FDA regulations and GEHC policy.

Customer Name:

Street:

City, State, ZIP:

System ID:

FDO#:

Room Name:

Field Engineer: Test Date:

Factory Engineer: Test Date:

7.2 Test Equipment

Cal. Due Date


Item Manufacturer Model No. Serial No.
mm/dd/yyyy

Radiation Meter MDH 2025

Non-invasive kV Keithley 35080A


Meter

Light Meter Digaphot

DVM

HHS Test Kit GEMS 46-303879G1 or N/A N/A


46-315694G1

mA Meter

mAs Meter

Oscilloscope

HV Divider

Page 208 Section 7.0 - HHS Data Sheets


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.3 X-ray Generator Tests

7.3.1 Generator Operator Indicators

GENERATOR OPERATOR INDICATORS Yes No N/A

Tube unit indicator lights

AEC reset

Tones and buzzers

Warning label present

Tech. factors indicated prior to exposure

7.3.2 Beam Quality


R

BEAM QUALITY (HALF VALUE LAYER)

Radiation Reading Aluminum Filtration Value Manufacture Date


One of the following:
With Without
Standard Standard Ratio (with/without) 2.5, or 3.0 Aluminum
Absorber Absorber

Acceptance Criteria Before June 10th, 2006:


- Ratio greater than or equal to 0.50 @ 2.5 mm Aluminum
Acceptance Criteria On or After June 10th, 2006:
- Ratio greater than or equal to 0.50 @ 2.9 mm Aluminum Minimum

7.3.3 Technique Accuracy and Linearity

kVp and mA/mAs Accuracy and Linearity


Focus Selected Selected Measured Measured mA/mAs kVp
S/L mA/mAs kVp ma/mAs kVp Error Error

S 10 60

S 200 60

S 10 80

S 200 80

S 10 130

S 200 130

L 250 60

L 500 70

L 250 80

L 800 80

L 250 130

Chapter 15 - Regulatory Tests (HHS) Page 209


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

L 500 130

7.3.4 Reproducibility of Exposure

REPRODUCIBILITY OF EXPOSURE

Exposure - mR
kVp mA C.V.
1 2 3 4 5 6 7 8 9 10

Table 80

Wall 80
Stand

7.3.5 AEC

7.3.5.1 AEC Minimum Exposure Time

AEC MINIMUM EXPOSURE TIME


Minimum exposure time requirement met for each sensing area of: Yes No N/A

Wall Stand Detector Vertical

Wall Stand Detector Horizontal

Table Detector

7.3.5.2 AEC Maximum mAs

AEC MAXIMUM mAs Yes No

Maximum mAs requirements met @ < 600 mAs?

Maximum mAs requirements met @ < 2000 mAs?

7.4 Collimator

7.4.1 General Tests

AUTOMATIC COLLIMATORS Yes No

Does the collimator respond to all changes in 5 seconds or less? (Y/N)

When going from Maximum SID towards a lower SID, the collimator
switches to manual mode at a SID of: (< 35.5in or 90.17cm)

When going from a low SID to a higher SID, the collimator switches into
automatic mode at a SID of: (< 35.5in or 90.17cm)

Page 210 Section 7.0 - HHS Data Sheets


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.4.2 SID Accuracy

7.4.2.1 SID Accuracy - Table

SID Displayed Measured Error % of SID Error

102 cm (40 in)

120 cm (47 in)

152 cm (60 in)

7.4.2.2 SID Accuracy - Wall Stand

Wall SID Displayed Measured Error % of SID Error


Stand

Horizontal 102 cm
(40 in)

Horizontal 120 cm
(47 in)

Vertical 102 cm
(40 in)

Vertical 152 cm
(60 in)

Vertical 183 cm
(72 in)

7.4.3 Collimator Field Light Intensity

COLLIMATOR FIELD LIGHT INTENSITY Yes No

The average illumination at a distance of 100 cm (39.37”) from the focal spot
is no less than 15 foot candles (160 lux).)

7.4.4 Collimator Light Field to X-ray Field Accuracy Check


SID Focal Spot Error Lateral (cm) Error Long (cm) % of SID Lateral % of SID Long

100 cm Small

100 cm Large

7.4.5 Collimator Field Size Indicator Accuracy

SID Focal Spot Error Height (cm) Error Width (cm) % of SID Height % of SID Width

100 cm Small

Chapter 15 - Regulatory Tests (HHS) Page 211


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.4.6 Table - Center to Center Error

Table Height OTS FOV Center to Center Error % of SID Error

Head-end 102 cm 30 x 24 cm
Minimum

Foot-end 102 cm 30 x 24 cm
Minimum

Head-end Max. SID 24 x 30 cm


Maximum

Foot-end Max. SID 24 x 30 cm


Maximum

7.4.7 Table - Size to Size Error

Height of Width of Height Width


Image x Image x % of % of Total
Mag. Mag. Height Width SID SID Error %
FOV SID Factor Factor Error Error Error Error of SID

41 x 41 cm 102 cm

24 x 24 cm 102 cm

18 x 18 cm 102 cm

7.4.8 Wall Stand - Center to Center Error

Wall Stand SID/OTS FOV Center to Center Error % of SID Error

Horizontal 102 cm (40 in) 30 x 24 cm


Min. Height

Horizontal 102 cm (40 in) 30 x 24 cm


Max. Height

Vertical Mid. 102 cm (40 in) 30 x 24 cm


Height

Vertical Mid. 183 cm (72 in) 30 x 24 cm


Height

Vertical Min. 183 cm (72 in) 30 x 24 cm


Height

Vertical Max. 183 cm (72 in) 30 x 24 cm


Height

Page 212 Section 7.0 - HHS Data Sheets


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.4.9 Wall Stand - Size to Size Error

Height of Width of Height Width


Image x Image x % of % of Total
Wall Mag. Mag. Height Width SID SID Error %
FOV SID Stand Factor Factor Error Error Error Error of SID

41 x 41 cm 102 cm (40 in) Horizontal,


Mid. Height

24 x 24 cm 102 cm (40 in) Horizontal,


Mid. Height

18 x 18 cm 102 cm (40 in) Horizontal,


Mid. Height

41 x 41 cm 102 cm (40 in) Vertical,


Mid. Height

24 x 24 cm 102 cm (40 in) Vertical,


Mid. Height

18 x 18 cm 102 cm (40 in) Vertical,


Mid. Height

41 x 41 cm 183 cm (72 in) Vertical,


Mid. Height

24 x 24 cm 183 cm (72 in) Vertical,


Mid. Height

18 x 18 cm 183 cm (72 in) Vertical,


Mid. Height

Chapter 15 - Regulatory Tests (HHS) Page 213


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 214 Section 7.0 - HHS Data Sheets


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 16 Final Cover Installation


Note: This chapter is to be performed by the local field engineer.

Section 1.0 Summary Checklist

Install table covers

Perform final OTS counterpoise adjustment

Install OTS trim covers and perform final wall stand counterbalancing

Install DSA cover

Install wall stand covers

Section 2.0 Table Covers


The installation of the table covers requires the table top insert be removed to gain access to the
cover mounting screws.

Chapter 16 - Final Cover Installation Page 215


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.1 Table Top Insert Removal
1.) On the underside of the table top, remove the four M4 hex socket head screws (two on each
end) that secure the table top laminate insert to the table top frame. See Figure 16-1.

Figure 16-1 UNDERSIDE OF TABLE TOP

2.) Raise the insert out of the metal frame. If the insert will not come out of the frame it may be
necessary to loosen the rubber bumper on one end of the table top. See Figure 16-1.

2.2 Lower Front and Rear Cover Installation


1.) If the hole in the tabs on the lower corners of the front and rear covers is large enough to allow
the head of the mounting screw to pass through it, start the four mounting screws a few turns.
Figure 16-2.

Figure 16-2 FRONT AND REAR LOWER COVER MOUNTING TABS

Page 216 Section 2.0 - Table Covers


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.) Slip mounting tabs on front and rear cover over screw heads (or hold cover in position and start
screw). See Figure 16-2. Tighten all four cover mounting screws installed in step 1. See
Figure 16-3.

Figure 16-3 LOWER FRONT COVER

2.3 Lower End Cover Installation


1.) Install the four lower end cover mounting screws a few turns. See Figure 16-4.

Figure 16-4 LOWER END COVER MOUNTING SCREWS

2.) Position the lower end cover as shown in Figure 16-5 by slipping it under the front cover and
over the partially installed mounting screws.

Figure 16-5 LOWER END COVER INSTALLATION

Chapter 16 - Final Cover Installation Page 217


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.) Rotate cover until it is vertical and secure by installing two upper mounting screws and
tightening the two lower mounting screws that were only started earlier. See Figure 16-6.

Figure 16-6 LOWER END COVER INSTALLATION

4.) Verify that the rear lower cover is not touching the table elevating mechanism. See Figure 16-
7.

Figure 16-7 COVER CLEARANCE CHECK

2.4 Upper Front and Rear Cover Installation


Note: If above grade floor mounted cable trough is being used, it is possible that the upper covers
could contact the cable trough when the table is driven to its lowest possible height. To avoid
any contact, the lower limit and position switches must be re-adjusted. Refer to Direction
5137439-100, Definium 8000 Renewal Parts and Disassembly/Reassembly, to identify the
lower safety and limit switches. This will mean that the table can not be driven to the lower
specification height and the customer should be notified of this change.

Page 218 Section 2.0 - Table Covers


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
1.) Install the upper front and rear covers with one screw in each upper corner. The screws do not
need to be tightened since the mid-covers will also use these screws to hold them in place.
See Figure 16-8.
Note: The upper front cover has one long recess cut out to allow the front detector assembly bearing
to roll on the bearing strip. The upper rear cover has three small recesses cut out to allow for
the detector bearing rod supports.

Figure 16-8 UPPER COVER INSTALLATION

2.) Adjust cover in the left to right direction to equalize the gap between the upper and lower
covers. See Figure 16-9.

Figure 16-9 COVER GAP ADJUSTMENT

Chapter 16 - Final Cover Installation Page 219


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.5 Front and Rear Mid Cover Installation
Note: The front mid cover has a long cut out and the rear mid cover is solid.
1.) Position the mid covers as shown in Figure 16-10. The detector assembly needs to be raised
to allow the detector handle to fit through the opening in the front mid cover.
Note: The table top must be moved to access these screws when the mid cover is on.
2.) Secure mid covers with the same screws that were used to hold the upper front and rear
covers. Figure 16-10.
Verify that the top of the detector handle does not rub on the opening in the mid-cover. If
necessary, shim the cover to prevent rubbing.

Figure 16-10 MID COVER INSTALLATION

3.) Adjust the front and rear mid covers to prevent rubbing on the detector housing. See Figure 16-
11.

Figure 16-11 MID COVER GAP ADJUSTMENT

Page 220 Section 2.0 - Table Covers


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.6 Upper End Cover Installation
1.) Position the upper end covers as shown in Figure 16-12.

Figure 16-12 UPPER END COVER INSTALLATION

2.) Secure end covers with four screws, two in the upper corners, see Figure 16-12 and Figure 16-
13 and two under the middle lip, see Figure 16-14.

Figure 16-13 UPPER END COVER MOUNTING SCREWS

Chapter 16 - Final Cover Installation Page 221


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 16-14 UPPER END COVER MOUNTING SCREWS

3.) Re-install the table top insert.

Section 3.0 OTS Final Counterpoise Adjustment and Trim


Covers
3.1 Final Counterpoise Adjustment
1.) Shut down the system, then turn off system power.
2.) Position the vertical column so that it is in the middle of the vertical range of travel.
3.) If the vertical column tends to drift upward or downward when released, adjust the
counterpoise tensioning bolt to eliminate drift (clockwise rotation results in more lift). See
Figure 16-15. Throughout the range of vertical column travel, the tube/collimator shoud
not drift downward.

Figure 16-15 OTS Counterpoise Tensioning Bolt

4.) Move the vertical column through the complete range of vertical travel. Check for drift at the
top and bottom limits of travel and adjust the counterpoise tensioning bolt to equalize the
amount of drift.

Page 222 Section 3.0 - OTS Final Counterpoise Adjustment and Trim Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
3.2 Install OTS Trim Covers
1.) Install the carriage shipping bolt through the bridge rail and into the carriage to prevent the
OTS carriage from moving during installation of the covers. See Figure 16-16.

Figure 16-16 OTS Carriage Shipping Bolt

Note: Refer to Figure 16-17 for OTS trim cover component identification and locations.

2.) Attach the cover pin receivers to the the sides of the OTS casting. The wider pin receiver is
mounted to the left side (head-end) of the casting. The smaller pin receiver is mounted on the
right side (foot-end) of the casting as shown in Figure 16-17.

Note: When installing or adjusting the brackets on the OTS front and rear covers, place the covers on the
plastic shipping bags to prevent any cosmetic damage to the covers.

3.) Attach the front mounting brackets to the left and right sides of the front cover. Leave the front
mounting bracket screws slightly loose so that they can be adjusted when the cover is
installed.
4.) Attach the baffle covers to the largest OTS column. Slide the baffle up until it touches the
bottom of the cable bracket at the rear of the column, then tighten the four screws.
5.) Position the ladder in front of the OTS. Lift the front cover into position. Rest the left front
mounting bracket on the TSL (Tube Support Logic) assembly. Rest the right front mounting
bracket on the cable drape bundle. It may be necessary to rotate the left and right front
mounting brackets to obtain clearance between the cover and the TSL and/or cable drape
bundle.
6.) When the front cover is properly positioned, attach the front cover brackets to the OTS casting
with four screws and washers. Do not completely tighten the screws; just get them started in
the threaded holes. The bottom of the front cover should be underneath the baffle.
7.) On the left side (head-end), slide the front cover pin into the cover pin receiver. Repeat on the
right side (foot-end). The cover can be flexed in order to slide the pins into the receivers.
8.) Tighten the four front mounting bracket screws to the OTS casting and tighten the two screws
that attach the front mounting brackets to the OTS front cover.
9.) Position the ladder at the rear of the OTS. Lift the rear cover into position and slide the rear
cover pins into the cover pin receivers. Attach the rear cover bracket to the OTS casting with
two screws. If the mounting bracket holes do not line up with the threaded holes in the casting,
the rear mounting bracket can be loosened from the cover adjusted as necessary. The bottom
of the rear cover should be underneath the baffle and the front of the rear cover should engage
with the lip at the rear of the front cover.
Chapter 16 - Final Cover Installation Page 223
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Cover Pin Receiver


Brackets, screws (4) &
12mm O.D. washers (4)
Rear
Cover Mounting
Bracket Screws (2)

Rear
Front Cover Mounting
Mounting Bracket
Bracket
screws (4) &
18mm O.D. washers (4) Rear
Cover
Front Mounting Pins
Brackets Front
Cover
Pins

Rear Cover

Baffle

Front Cover
Figure 16-17 OTS COVERS

Section 4.0 DSA Cover


1.) Remove the cover mounting screws located in the grills on both ends of the DSA.
2.) Slide cover down over the top of the DSA frame. Do NOT pinch cables or hoses. The fan grill
and cable entrance grill must be inside the cover. See Figure 16-18.
3.) Position guide blocks on the inside of cover over the guide pins on the baseplate. See
Figure 16-18. Allow fingers to feel the pins and hole in guide block.

Page 224 Section 4.0 - DSA Cover


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 16-18 DSA COVER POSITIONING

4.) Secure the cover to the frame with the screws removed earlier. See Figure 16-19.

Figure 16-19 DSA INSTALLED

Chapter 16 - Final Cover Installation Page 225


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 5.0 Wall Stand and Extended Stand Covers


1.) Install the 4 hole protectors on the vertical column. See Figure 16-20 and Figure 16-21.

Figure 16-20 HOLE PROTECTOR INSTALLATION

Figure 16-21 HOLE PROTECTOR INSTALLATION

Page 226 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.) Install a patient handle on both sides of the detector housing. See Figure 16-22.

Figure 16-22 PATIENT HANDLE INSTALLATION

3.) Position the front cover as shown in Figure 16-23.

Figure 16-23 FRONT COVER INSTALLATION

Chapter 16 - Final Cover Installation Page 227


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
4.) Secure the front cover with two screws in each side and three screws from the back side of the
top rail. Be careful not to drop the screw into the slot for the grid. See Figure 16-24 and
Figure 16-25.

Figure 16-24 FRONT COVER INSTALLATION

Figure 16-25 FRONT COVER INSTALLATION

5.) Position the back plate as shown in Figure 16-26 and secure with three screws.

Page 228 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 16-26 BACK PLATE INSTALLATION

6.) Position the lateral patient support bar anchor plate, with pivot attached, as shown in
Figure 16-27 and secure with six bolts.

Figure 16-27 LAT BAR SUPPORT INSTALLATION

Chapter 16 - Final Cover Installation Page 229


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
7.) With the lateral support pivot centered laterally, position the top rear cover loosely as shown
in Figure 16-28 and connect the wires to the mating connector in the lower right corner.

Figure 16-28 TOP REAR COVER INSTALLATION

8.) Position the bottom rear cover as shown in Figure 16-29 in preparation for installation and
connect the wires to the mating connector in the lower right corner.

Figure 16-29 BOTTOM REAR COVER INSTALLATION

9.) Before swinging the bottom rear cover into position, engage the bracket on the bottom of the
cover in the mating bracket on the bottom of the front cover. See Figure 16-30.

Figure 16-30 BOTTOM REAR TO FRONT COVER BRACKET

Page 230 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
10.) Swing the bottom rear cover up and engage the side alignment brackets in the slot on the front
cover brackets. See Figure 16-31.

Figure 16-31 BOTTOM REAR COVER ALIGNMENT BRACKET

11.) Secure the bottom rear cover with two screws. See Figure 16-32.

Figure 16-32 BOTTOM REAR COVER INSTALLATION

Chapter 16 - Final Cover Installation Page 231


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
12.) Swing the top rear cover down and engage the side alignment brackets in the slot on the front
cover brackets. See Figure 16-33.

Figure 16-33 TOP REAR COVER INSTALLATION

13.) Secure the top rear cover with two long screws (and washers) from the back and two short
screws from the top. See Figure 16-34 and Figure 16-35.

Figure 16-34 TOP REAR COVER INSTALLATION

Page 232 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 16-35 TOP REAR COVER INSTALLATION

14.) Secure the support pivot access panel with two screws. See Figure 16-36.

Figure 16-36 SUPPORT ACCESS PANEL INSTALLATION

15.) Install the grid. See Figure 16-37. Grid must be installed before final counterbalancing is
performed.

Figure 16-37 GRID INSTALLATION

Chapter 16 - Final Cover Installation Page 233


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
16.) Position the upper horizontal arm cover and secure with two screws. See Figure 16-38

Figure 16-38 UPPER ARM COVER INSTALLATION

17.) Position the front horizontal arm cover and secure with two screws. See Figure 16-39.

Figure 16-39 FRONT ARM COVER INSTALLATION

18.) Position the lower horizontal arm cover and secure with four screws. See Figure 16-40.

Page 234 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Figure 16-40 LOWER ARM COVER INSTALLATION

19.) Position the back collar against the rear of the receptor carriage and snap over the ball studs.
See Figure 16-41. The back collar must be installed prior to final counterbalancing of the wall
stand.

Figure 16-41 BACK COLLAR COVER

Note: Before proceeding with the final counterbalancing of the wall stand, be sure that all wall stand
components that will affect counterbalancing have been installed. These items include: all detector
assembly covers, all wall stand arm covers, back collar cover and grid. Do not install the lateral
patient support bar.

Chapter 16 - Final Cover Installation Page 235


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
20.) Release the vertical lock (clutch) and add counterweights until the vertical moving effort is the
same in either direction. See Figure 16-42.

Figure 16-42 COUNTERWEIGHT INSTALLATION

21.) Install the counterweight restraint. See Figure 16-43.

Figure 16-43 COUNTERWEIGHT RESTRAINT INSTALLATION

Page 236 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
22.) Remove the 4 screws and top cover. See Figure 16-44.

Figure 16-44 REMOVE TOP COVER

23.) Raise the receptor carriage and position the lower front trim cover against the front of the stand
and snap over the ball studs near the top of the cover. See Figure 16-45.

Figure 16-45 LOWER FRONT COVER INSTALLATION

Chapter 16 - Final Cover Installation Page 237


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
24.) Lower the receptor carriage and position the upper front trim cover against the front of the
stand and snap over the ball studs near the top and bottom of the cover. See Figure 16-46.

Figure 16-46 UPPER FRONT COVER INSTALLATION

25.) Remove the back collar from the rear of the receptor carriage so that the rear trim covers can
be installed. See Figure 16-41.
26.) Position the back trim cover against the rear of the stand and snap over the ball studs near the
top of the unit. See Figure 16-47.

Figure 16-47 BACK COVER INSTALLATION

Page 238 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
27.) Align the mounting holes for both the back and lower front covers with the mounting holes in
the baseplate. Secure the covers with 1 screw, bushing and trim cap on each side. See
Figure 16-48.

Figure 16-48 COVER SCREWS

28.) Position top cover and secure top cover with 4 screws and trim bushings. See Figure 16-49.

Figure 16-49 TOP COVER INSTALLATION

Chapter 16 - Final Cover Installation Page 239


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
29.) Position the back collar against the rear of the receptor carriage and snap over the ball studs.
See Figure 16-41.
30.) Clean the lateral support arm with isopropyl alcohol (or equivalent) in the area where the yellow
Caution label is to be attached, then attach the label (in the appropriate language) to the arm.
See Figure 16-50. The Caution labels are shipped in the Wall Stand counterweight box.

Figure 16-50 LATERAL SUPPORT ARM CAUTION LABEL

31.) Wall Stand installation is complete.

Page 240 Section 5.0 - Wall Stand and Extended Stand Covers
GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Chapter 17 Final Tasks


Note: This chapter is to be performed by the local field engineer.

Section 1.0 Summary Checklist

Backup database

Clean up room

Collect/send product locator cards

Section 2.0 Database Backup


Perform complete database backup. See Software Chapter, System Data, Backup and Restore
Section in System Manual 5137437-800 which can be found on the Definium 8000 Service
Documentation CDROM 5137441-200, shipped with the system.

Section 3.0 Final Administrative Tasks


1.) Ready system for normal customer operation.
2.) Remove trash, remove installation equipment and tools, clean up room and store service
manuals.
3.) Fill out Product Locator Cards and return them. See Section 4.0 on page 242.
4.) Record installation time in Service database.

Chapter 17 - Final Tasks Page 241


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Section 4.0 System Rating Plates

3 1

Figure 17-1 SYSTEM RATING PLATE LOCATIONS

Page 242 Section 4.0 - System Rating Plates


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

 -/$%, .5-"%2 490%


 
$%3)'.!4)/. *%$) 2$ 4
-!.5&!#452%$ 8888888 8
,/#!4)/. "!.'!,/2% ).$)!
3. 8888888
6/,43   6!#
!-0 -/-%.4!29  !
!-0,/.' 4%2-  !
K6! +6!
0(!3%  (Z 0(   (Z

3YSTEM -/$%, .5-"%2 490%




#ABINET -!.5&!#452%$
,/#!4)/.
3.
8888888 8
"!.'!,/2% ).$)!
8888888


 -!.5&!#452%$ &/2 -/$%, ./ ).8  '%-3
'%.%2!, %,%#42)# #/-0!.9 3%2)!, ./ 88888888888
-),7!5+%% 7)3#/.3). -!.5&!#452%$ 88888888888 8
"9 $%3#2)04)/. 4(5.$%2 )/. #(!-"%2

4ABLE !$6!.#%$ ).3425-%.4 $%6%,/0-%.4


-%,2/3% 0!24 ), 53!
).#
  6$#  !

-!.5&!#452%$ &/2
'%.%2!, %,%#42)# #/-0!.9
-),7!5+%% 7)3#/.3).
"9 0+)
 
-!$% ). 53!
-/$%, 
3. 88888888
-!.5&!#452%$ 88888888 8
$%3# 4(5.$%2 $%4%#4/2
6^ (Z !

Figure 17-2 SYSTEM RATING PLATE LOCATIONS

Chapter 17 - Final Tasks Page 243


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100



-!.5&!#452%$ &/2 ).8  '%-3 -/$%, ./

'%.%2!, %,%#42)# #/-0!.9 88888888888 3%2)!, ./

-),7!5+%% 7)3#/.3). -!.5&!#452%$ 88888888888 8


"9 $%3#2)04)/. 4(5.$%2 )/. #(!-"%2
!$6!.#%$ ).3425-%.4 $%6%,/0-%.4 ).#
-%,2/3% 0!24 ), 53!
  6$#  !


-!.5&!#452%$ &/2
'%.%2!, %,%#42)# #/-0!.9
-),7!5+%% 7)3#/.3).
0+)
"9
 
-!$% ). 53!
-/$%, 
3. 88888888
-!.5&!#452%$ 88888888 8
4(5.$%2 $%4%#4/2
7ALL 3TAND
$%3#

6^ (Z !

'%.%2!, %,%#42)# #/-0!.9


 -),7!5+%% 7)3#/.3). -!$% ). 53!
#!4 ./ .!
-/$%, 
3. XXXXXXX7+X
-!.5&!#452%$ XXXXXXX X
$%3# 4(5.$%2 $%4%#4/2 3500/24
-!$% ). '%-3 !- !33%-",%$ ). !0/$!#! -%8)#/

$3! 8 2 !9 % 1 5 ) 0 - % . 4 # , ! 3 3 ) & ) % $ " 9 5 . $ % 27 2 ) 4 % 2 3 , ! " / 2 !4/ 2 ) % 3 ) . #


! 3 4/ % , % # 4 2 ) # ! , 3 ( / # + & ) 2 % ! . $ - % # ( ! . ) # ! , ( ! : ! 2 $ 3 / . ,9    ,

Figure 17-3 SYSTEM RATING PLATE LOCATIONS

Page 244 Section 4.0 - System Rating Plates


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
Item System Component Model Number Location
1 Bridge 5127305 Top of rear endcap
2 Overhead Tube Suspension (OTS) 5135678 Rear of largest column.
3 X-ray Tube Casing MX100 09PS 46-155400G46 Rear of X-ray tube.
X-ray Tube Insert .6-1.25 12.5 degree 2336058
MX100
4 Collimator 2266999 Rear of collimator
5 Workstation PC • 5117866-27 (xw8200) Top front of PC
• 5261130 (xw8200)
• 5183547-28 (xw8400)
• 5261133(xw8400)
6 RCIM 2383880 Bottom of RCIM
7 System Cabinet 5128145 • Top of unit, on right side towards
front.
Jedi 80 Rad 1T 2374870 • Top of unit, on right side towards
front. Also inside system cabinet,
on left side of Jedi Control
Assembly
Jedi HV Tank 2186730 • Inside system cabinet, on front of
Jedi HV Tank
8 • Jedi 80 Rad 1T 2374870 • Top of unit, on right side towards
front. Also inside system cabinet,
on left side of Jedi Control
Assembly
• Jedi HV Tank 2186730 • Inside system cabinet, on front of
Jedi HV Tank
9 Table 5131070 Left side, upper corner towards rear of
table. Also on rear center of table, but
concealed by cover.
10a Table Ion Chamber 5143310 Inside table detector assembly.
10b Table Digital Detector 5134673 Inside table detector assembly.
10c Table Grid 46-286129P51
11 Wall Stand - standard arm 5141608 Rear top of vertical column cover.
Wall Stand - extended arm 5141608-2
12a Wall Stand Ion Chamber 5143310 Inside wall stand detector assembly.
12b Wall Stand Digital Detector 5134673 Inside wall stand detector assembly.
12c Wall Stand Grids: 100 cm 46-286129P50
100-180 cm 46-286129P56
180 cm 46-286129P48
13 Detector Support Assembly (DSA) 5137888 Under cover, on base next to cable
clamp.
QAP Phantom Carrier 2320105
QAP Phantom 2220170
Table 1 - Definium 8000 Rating Plates

Chapter 17 - Final Tasks Page 245


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 246 Section 4.0 - System Rating Plates


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Appendix A

Section 1.0 Inboard Rail Length Modifications


If rail mounting bolt holes do not match the holes in the overhead structure (rail supports), Unistrut
part number P1000T or Halfen equivalent could be installed to provide correct stationary rail-to-rail
support, mounting hole alignment.

PROCEDURE
1.) Determine new rail length and scribe the rails with a scribe.
2.) If present, remove the end cap and bumper from both ends of the stationary rail. These parts
can be discarded, as they are not used on the Definium 8000 system. Refer to Figure 17-4.

1/40--20 x 1/2 (12.7 mm)


HHC SCREW

EN D CAP
10 --- 32 SCR EW
rubber bumper

Figure 17-4 End Cap and Bumper Mounting

3.) Starting with rear rail, shorten wear strip as follows. Refer to Figure 17-5.
a.) Score the hardened wear strip with a file, 1/4” (6.35 mm) shorter than the rail is to be cut.
Pry the end of the wear strip loose and break off excess wear strip.
Do not confuse this procedure with the procedure for modifying the bridge rails that are
responsible for lateral XT movement.
b.) Clean wear strip mounting surface with alcohol, and re-bond wear strip using Armstrong
D220 cement.
HARDENED
WEARSTRIP NEW RAIL END

scored
wear
strip

1/4"
(6.35 mm)
STATIONARY RAIL

Figure 17-5 Wear Strip Modification

4.) Cut lock strip (for rear rail only) as follows. Refer to Figure 17-6.

Chapter 17 - Final Tasks Page 247


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
1" (2.54 cm)

newlockstrip
stationary rail clear ance hol e
8--32 screws 8"
(203 mm)

NEW END OF RAIL NEW END OF


LOCKSTRIP

Figure 17-6 Lock Strip Modification

a.) Remove 8-32 screws holding lock strip on rail end to be shortened. Refer to Figure 17-6.
b.) Pry lock strip loose.
c.) Scribe mark the lock strip 8" (20.3 cm) shorter than rail is to be cut, and cut lock strip.
d.) Drill and countersink a No. 18 clearance hole (4.3 mm) for an 8-32 flat head screw, 1"
(2.54 cm) from cut end of lock strip.
e.) Using hole in lock strip as a guide, drill a No. 29 hole (3.5 mm) in rail and tap hole with 8-
32 tap. Refer to Figure 17-7.
f.) Clean out drill chips, and clean rail surface with alcohol.
g.) Using Armstrong D220, bond lock strip to rail.
h.) Reinstall 8-32 screws holding lock strip.
5.) Cut rail to proper length, ensuring end is square.
6.) Step-drill a 3/4” (1.9 cm) hole in the top (a) or the side (b) of the rail, depending on mounting
method of inboard or outboard rail.
(2) 3/4"
(19.05mm)
HOLES
(a)

4” MAX
(1016 mm)
1--1/16"
(2.7 cm)
1- 3/16”
(30.2 mm)

(b)
stationary
rail

Figure 17-7 Stationary Rail Modifications

7.) Modify front rail as outlined in steps 1, 2, 3, 5, 6 and 7.

Section 2.0 Rail Leveling Procedures (Water-Level or Transit)


Note: Bubble levels are NOT accurate enough to obtain a 1/32" (0.8 mm) tolerance.

Page 248 Section 2.0 - Rail Leveling Procedures (Water-Level or Transit)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
2.1 Using a Water-Level Kit
1.) Slide open the battery door on the back of the unit and connect a 9 volt alkaline battery.
Use an alkaline battery only. The use of a non-alkaline or heavy-duty battery will cause unit to
malfunction.
2.) Attach a 30’ (9.14 m) clear plastic 3/8” (9.5 mm) I.D. tube to the connector on the bottom of the
water-level case.
To meet any room installation requirement, greater than 30’ (9.14 m) of tubing can be used
with the water-level unit. The greater the distance from the water-level case (i.e., longer
lengths of tube), the more time that must be allowed for the water line to damp (i.e. reach
equilibrium).
3.) Prepare 3 cups of water solution containing 1/8 teaspoon of salt and enough food color to
enable the water level to be seen more easily.
4.) Remove the plug from the short tube on the water-level case side.
5.) Fill the long plastic tube with water solution. Refer to Figure 17-8.
a.) Place the free end of the long plastic tube in the water solution container and elevate the
container.
b.) Place the water-level case below the level of the water solution container.
c.) Siphon water solution into the long plastic tube by sucking on the short tube on the water-
level case side.
d.) Fill the long plastic tube until the water solution level is slightly below the water-level case.
6.) Insert plugs into both the short and long free tube ends to prevent loss of water.

short tube used to start water siphon


elevated
water
solution
water--level
case

long plastic tube

Figure 17-8 Filling Water-level with Liquid

7.) Mount both stationary rails in place using all bolts and then tighten the bolts. There must be a
minimum of 2" (5.1 cm) between the ends of the stationary rails and the wall.
8.) Tape a six (6) foot or a two (2) meter flexible steel scale to a support rod to stiffen the scale or
tape the zero scale end of a six (6) foot or a two (2) meter flexible steel scale to a stationary
rail. Refer to Figure 17-9.

Chapter 17 - Final Tasks Page 249


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

Stationary Rail

Lock Strip
Zero Scale End

Flexible Scale
Without Rod
Wear Strip

Figure 17-9 Stationary Rail (End View of Rail)

9.) If using the steel scale with a support rod, hold the tape scale and rod assembly up against the
bottom of the stationary rail surface.
10.) Mount the water-level case in an upright position with the reference line on the side of the case
at some convenient position above the lower end of the tape scale. To insure that the
instrument remains plumb, secure the top and both sides of the water-level case in position.
Refer to Figure 17-10.
water--level
plug case

short
plastic
tube case
leveling
mark

long plastic
tube water line

Figure 17-10 Mount Water-level Case

11.) Lay out the plastic tube and remove any knots or kinks. Check to make sure that there are no
air bubbles in the tube. Air bubbles will cause inaccurate readings.
To remove air bubbles, remove the plugs from the long tube free end and from the short tube
end on the instrument case. Hold the long tube below the water-level case (at about a 45
degree angle), until a small amount of water flows from the long free tube end. All bubbles
should be gone.

Page 250 Section 2.0 - Rail Leveling Procedures (Water-Level or Transit)


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
12.) With both plugs removed, bring the free end of the long tube alongside the same tube at the
mounted water-level case. Refer to Figure 17-11. Verify that the water line in the long plastic
tube settles at an equal level at both tubes ends. If the level is not equal, check for air bubbles
in the tube and remove.
water--level case

plug removed from tube


end

plug

long level
plastic water
tube line

Figure 17-11 Verify Water Lines are Level

13.) To measure the number of shims required, hold the tape scale and rod assembly or tape the
flexible scale up against the stationary rail bottom at each bolt location.
14.) Turn ON the water-level switch.
15.) With both plugs still removed, slowly raise the long tube free end along with the scale and rod
assembly up against the ceiling steelwork at any mounting location. Adjust the water line (up
or down) at the long tube free end until the buzzer first sounds. Refer to Figure 17-12.
TO ST ATIO NA RY RA IL

scale and
rod assy long tub
free end

plug

water
line

Figure 17-12 Water Line Adjustment

16.) Take readings at each mounting bolt location and record these readings. This allows you to
determine the lowest bottom point of the stationary rails. This is the reference value. Calculate
the difference between this reference value and the other bolt locations greater than 1/32” (0.8
mm). Use these difference calculations to determine height and number of shims required at
each bolt hole location.
Shims will be required and placed at all points with a calculated difference of greater than 1/
32” (0.8 mm).
17.) When all measurements have been recorded, turn OFF the water-level switch and replace the
plugs at both the long tube free end and at the short tube on the water-level case side.

Chapter 17 - Final Tasks Page 251


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
18.) Proceed to Section 2.3 (Rail and Shim Installation).

2.2 Using a Transit (Alternate Method)


1.) Mount both stationary rails in place using all bolts and then tighten bolts.
2.) Set up a transit (builder's level) in the center of the area encompassed by the rails. Refer to
Figure 17-13.
Level the transit according to the instructions supplied with the instrument.
3.) Tape a six (6) foot or a two (2) meter flexible steel scale to a support rod to stiffen the scale or
tape the zero scale end of a six (6) foot or a two (2) meter flexible steel scale to a stationary
rail. Refer to Figure 17-9.
4.) To measure the number of shims required, move the scale and hold up to (or tape to) the
stationary rail at each bolt location. Refer to Figure 17-13.
5.) View the scale through the transit. Recheck transit level at each bolt location and re-level if
necessary.
6.) Take readings at each mounting bolt location and record these readings. This allows you to
determine the lowest bottom point of the stationary rails. This is the reference value. Calculate
the difference between this reference value and the other bolt locations greater than 1/32” (0.8
mm). Use these difference calculations to determine height and number of shims required at
each bolt hole location.
Shims will be required and placed at all points with calculated differences greater than 1/32”
(0.8 mm).
7.) Proceed to Section 2.3 (Rail and Shim Installation).

CEILING
steelwork

STEEL SCALE
and support rod
STATIONARY
RAIL

centrally
located
transit
leveling
line of sight

.
Figure 17-13 Checking Rail Supports with Transit (Alternate Method)

Section 3.0 Modifying the Bridge Length


Note: Modifying bridge length is NOT recommended by GEHC Engineering.
If modification of the bridge is required, YOU MUST CHECK WITH YOUR INSTALLATION
SPECIALIST. Confirm that the length you desire will not limit your ability to purchase future
product options or upgrades (i.e. VolumeRAD).

Page 252 Section 3.0 - Modifying the Bridge Length


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100
If necessary, the OTS bridge may be cut off on either/or both ends using the following procedure:
1.) Remove end cap.
2.) Pry hardened wear strip loose, score with file, and break off 3" shorter than rails are to be cut.
Do not confuse this procedure with the procedure for modifying the stationary rails.
3.) Cut rails square and of equal length.
4.) Chamfer end of rails so that end cap will fit.
5.) Re-bond wear strip.
6.) Place end cap on rails and use as template to drill (3) new holes in each rail.
7.) Replace end cap and mount with original hardware.

HARDENED
WEARSTRIP

BRIDGE RAIL

Figure 17-14 OTS Bridge Length Modification

Chapter 17 - Final Tasks Page 253


GE HEALTHCARE DEFINIUM 8000 INSTALLATION MANUAL
REVISION 8 DIRECTION 5137436-100

This page left intentionally blank

Page 254 Section 3.0 - Modifying the Bridge Length


AY
XR
GE HEALTHCARE
© 2005, General Electric Company.
GE Medical Systems, a General Electric Company, going to market as GE Healthcare.
3000 N. Grandview Boulevard
Waukesha, Wisconsin 53188, USA
www.gehealthcare.com
256

You might also like