CERTIFICATE OF CALIBRATION
QC Internatio nal Record Information
16 ave A 3-94 zona 15
guatemala, 01015 Record No: 464-2
50252087273 Internal ID: 464
Serial No: 201807180213
Asset No: 201807180213
PO: 18832 & 18833
SO: P1922
QC Certificate number: Q-12764
Gage Information
Gage Status - As Found: Active Range or Size: 0-6700K
Type: Visual Inspection Machine Accuracy : 1
Model : SR-780 Reference Standard: No
Manufacturer : TABONY Measurement Types : Light temperature
Units Of Measure: Kelvin Degree (K); Foot Candle
(Fc); Lux (lx) Interval : Scheduled (1 years, 0 months, 0 days)
Calibration Information
Condition Received: Used Measurement Uncertainty : N/A
Calibration Procedure: NFC counter verification Temperature: 33.8- 33.8 °C
Humidity: 39 - 37 %
Gage Assignment
Country : El Salvador Company : DECOTEX INTERNATIONAL LTDA. DE C.V.
Assignee: Esmeralda de Argueta
Calibration performed under the following environmental conditions :
Nominal | Minimum | Reference
Test No Attribute Maximum Standard ID As Found As Left Result
1 Speed 75 ± 3 RPM 62 75 RPM 75 RPM Pass
Top Light
2 Temperature 6500K - 6300K - 6700K 51 6498 K 6498 K Pass
Left
1500 lux - 1075 lux - 2500
3 Lux 51 1643 lux 1643 lux Pass
lux
© 2023 GageList | All rights reserved | www.GageList.com | 281-816-7130 | Page 1 of 2
CERTIFICATE OF CALIBRATION
Top Light
4 Temperature 6500K - 6300K - 6700K 51 6536 K 6536 K Pass
Center
1500 lux - 1075 lux - 2500
5 Lux 51 1693 lux 1693 lux Pass
lux
Top Light
6 Temperature 6500K - 6300K - 6700K 51 6300 K 6300 K Pass
Right
7 Lux 1500 lux - 1075 lux - 2500 51 2225 lux 2225 lux Pass
lux
This Gage is within Specifications: (As Found) Pass - (As Left (overall pass/fail)) Pass
Gage is Fit for Intended Purpose: (As Left) Yes
Calibration Actions Adjustments Required: No
Repairs Required: No
Action Required: No
Statement : This instrument has been inspected and calibrated in accordance with manufacturer specifications
and all standards used are traceable to the National Institute of Standards and Technology (NIST). Instruments
are calibrated in Compliance with ISO/IEC 17025:2017. The collective uncertainty of the measurement standards
shall not exceed 25% of the acceptable tolerance allowed for the individual characteristics measured, the
uncertainty of the measurement standards for this calibration are based upon 95% confidence limits.
According to manufacturing standards and specifications all calibrations should be performed in the following
laboratory conditions: 22 ± 7 °C / 50 ± 30 %. The Calibration records are on file for review at QC Internacional,
upon request.
Review and Approval
Calibrated By : Ricardo Cardona Calibration Date: 02/12/2023
Next Calibration Due: 02/12/2024 Gage Status - As Left : Active
Calibration Standards Used:
Next
Internal
Manufacturer ID Serial No Model Type Calibration
Due
Sekonic 51 356 C-700 Spectroradiometer 23/08/2023
Elmed 62 09100644 HELIO-STROB micro Digital Tachometer 31/05/2024
© 2023 GageList | All rights reserved | www.GageList.com | 281-816-7130 | Page 2 of 2