Workbook 6.ct.2006
Workbook 6.ct.2006
WORKBOOK 6
Radiological Council
Locked Bag 2006 P O
NEDLANDS WA 6009
WORKBOOK 6
COMPUTED TOMOGRAPHY EQUIPMENT
2006
The workbook describes standard test protocols approved by the Radiological Council for routine
compliance testing of computed tomography (CT) x-ray equipment. Variations to the recommended
test methods may be used provided they are approved by the Radiological Council. Documentation on
alternative test methods will need to be provided.
Persons who perform these tests in Western Australia must hold a licence under the Radiation Safety
Act for the purpose, or be acting under the direction and immediate personal supervision2 of a
licensed compliance tester.
Most of the tests assess compliance with the regulations to the Western Australian Radiation Safety
Act and with other Council requirements that may apply to the class of equipment under test. Where
practicable, tests are traceable to relevant Australian and International Electrotechnical Commission
(IEC) standards.
Directed essentially at radiation safety parameters, this workbook does not include all tests that might
be considered as part of a comprehensive quality assurance (QA) program. However, should the
workbook be used as a basis for a QA program, it is recommended that additional radiographic,
sensitometric and image quality tests be performed.
The Radiological Council is grateful to the personnel of Radiation Health Branch and the other
members of the compliance testing working group who contributed their time and effort to the
production of the workbooks for the compliance testing program. It is particularly grateful for the
substantial contributions of Mr Peter Henson in redrafting these workbooks. The working group
which undertook the program revision has comprised Richard Fox (Chairman); Janette Atkinson (Sir
Charles Gairdner Hospital); John Burrage & David Causer (Royal Perth Hospital); Bill Cooke
(Community Dental Services); Janelle D’Souza & Tim Finney (Coordinators, Radiation Health
Branch); Bruce Dwyer (Philips Medical Systems); Luke Ferris (MedDent); Peter Henson (Radiation
Health Branch); Gary Johnson; Brett McLoughlin & John Pereira (Health Technology Consultancy);
Max Ross (Biomedical Engineering); Roger Sandercock (Toshiba (Australia) Pty Ltd); Tony
Shackleton (Sir Charles Gairdner Hospital); Jonathon Thwaites (Medical & Scientific); Ann Watson
(ADAW Pty Ltd); Graham Wood (Siemens Pty Ltd).
Further comments during editing by Barry Cobb (Radiological Council), Hazel Upton (Secretary,
Radiological Council) and Leif Dahlskog (Radiation Health Branch) are also gratefully acknowledged.
1
The Radiological Council is the statutory regulatory authority responsible for administration of the State’s
Radiation Safety Act. For use in other jurisdictions the relevant regulatory authority would need to be
consulted where the Radiological Council is referred to in these workbooks. Organisations which wish to
reproduce all or part of the workbook for use in their own testing programs should contact the Radiological
Council for permission to do so.
2
“Immediate personal supervision” requires the licensee to be physically present and to directly observe
persons working under their direction and supervision.
CONTENTS
1. INTRODUCTION 7
1.1 SCOPE AND APPLICATION 7
1.3 DEFINITIONS 10
1.4 PRECAUTIONS 12
3. IMAGE QUALITY 27
3.1 INTRODUCTION 27
RELATED PUBLICATIONS 39
1. INTRODUCTION
1.1 SCOPE AND APPLICATION
This workbook describes protocols for the compliance testing of computed
tomographic (CT) x-ray equipment.
Persons intending to use this workbook should seek advice from the
Radiological Council on any necessary approvals required to perform the tests.
The frequency of testing for each class of x-ray equipment is prescribed by the
Radiological Council in the Program Requirements.
There are a number of CT test phantoms in use, both commercial and in-
house, some of which may not fully satisfy the requirements of this
Workbook. The CT phantom(s) required to properly perform the tests in
this Workbook must contain the following inserts —
1
All radiation measuring instruments used as part of a test must be of a suitable type and calibrated by an
organisation recognised by the Radiological Council at not more than two yearly intervals.
2
Perspex - polymethylmethacrylate (PMMA) density 1.19 g.cm-3
1.3 DEFINITIONS
Tube Voltage
For the purpose of this workbook the term kVp (kilovoltage peak) has been
adopted although for true constant potential units the term kV should be used.
Tube voltage measurements will normally be specified as kVp average, being
the average of the voltage peaks measured in the sampling period.
Radiation Units
Exposure (C kg-1 )
The energy absorbed per unit mass in a nominated medium. Because the
amount of energy deposited from a beam of radiation will depend on the
medium irradiated, the nature of that medium should also be stated. The
gray (Gy), the unit of absorbed dose, is equal to the deposition of energy of
1 joule in a mass of 1 kg of the nominated medium.
This quantity takes into consideration the type of radiation being measured.
The unit of equivalent dose is the sievert (Sv).
where WR is the Radiation Weighting Factor and Absorbed Dose refers to the
average dose over a tissue or organ.
WR relates to the biological effects that result from exposure to different types
of radiation. For x-rays, WR = 1 and therefore an absorbed dose of 1 gray (Gy)
of x-rays gives an equivalent dose of 1 sievert (Sv).
OD = log 10 ( I 0 I t )
where I0 is the intensity of light without the film and It is the intensity of light
transmitted through the film. For example, a density of 1 means that only
10% of the light has been transmitted through the film; a density of 2 means
1% transmission, etc.
Abbreviations
1.4 PRECAUTIONS
Permissions
Before performing any of the tests in this workbook, ensure that approvals to
operate the x-ray equipment have been obtained from —
Tube warm-up is under computer control and should present no problem for
the tester. However, should an alarm, error code (or message), equipment shut
down or an abnormal waveform occur, the tester should:-
Calibration of scanner
Before any scanner image parameter tests are performed at least an air, and
preferably an air/water, calibration should be performed on the scanner to
correct for any drift which may have occurred in CT numbers. The scanner
may need to be re-calibrated should the test sequence be interrupted for any
significant time.
All image parameter tests and dose measurements must be conducted in the
routine operating mode with any beam compensation filter in position.
Errors in measurement may result if these steps are not followed correctly.
Radiation Protection
Faults found during testing should be included on the test report, even if such
faults are corrected before the completion of testing. Some faults may be
common to a particular model of x -ray equipment and failure to report them
may put other users and patients at risk.
A X-ray Control
Ø manufacturer
Ø model
Ø serial number
Ø generator type (three phase, medium frequency, etc.)
Ø generator ratings (maximum kVp, mA or mAs)
Ø exposure warning devices
Ø manufacturer
Ø model
Ø serial number
Ø inherent filtration (mm Al at a nominal kVp)
Ø added filtration
Ø filtration non-removable (without tools)
Ø marking of focal spot position
Ø manufacturer
Ø model
Ø serial number
Ø maximum rated kV
Ø focal spot sizes
D CT Scanner
Ø scanner movements (rotate/rotate, rotate only, spiral, other)
Ø reconstruction matrix sizes available
Ø type of detectors, solid or gas (name of material if available)
Purpose of Test
Equipment Required
Method
1. Ensure that any necessary x-ray tube warm up procedure has been
followed (section 1.4).
Care must be taken to ensure that the tube voltage obtained is the average of
the peak voltage throughout the sampling period (kVp average). Some
instruments provide different high voltage parameters such as kVp maximum,
kV effective, etc.
The tube voltage must satisfy the following accuracy requirements within
100 ms of initiating the exposure.
The radiation waveform is used to determine that the measurement has not
been affected by any unusual waveform characteristics. These may include
voltage spikes, inconsistent output, rising radiation waveform, etc.
The radiation output over a range of tube voltages may be used to compare
outputs from different tubes and generators within a department. For fixed
mA and time or fixed mAs exposures, a graph of the logarithm of the kVp
versus the logarithm of the output should be a straight line between, typically,
50 kVp and 100 kVp, with a correlation coefficient of r2 ≥ 0.99. The gradients
of such graphs have been found to vary between 2.0 and 2.8. A significant
deviation of the measured points from linearity on such a plot may indicate a
generator problem.
Purposes of Test
To measure the free in air radiation dose at the centre of the CT gantry and to
measure the output linearity coefficient of the scanner.
Equipment Required
Ø Small volume ion chamber with an occluding cap so that the sensitive
region of the ion chamber is less than 6 mm in length1
Ø Suitably calibrated electrometer for the small volume ion chamber
Ø Retort stand and clamps
Ø Tape measure
Method
1. Using the retort stand, locate and clamp the ion chamber with the
occluding cap at the centre of the gantry (free in air).
2. Confirm with a tape measure and the positioning lights that the chamber
is centred within the gantry opening and that the centre of the chamber is
within the slice plane.
8. Using the same kVp, scan time and slice thickness, repeat the slice dose
readings with other mA values to cover the available range. Convert the
results to absorbed dose as in step 7.
1
The unoccluded length of the chamber should not be at the end but preferably in the central region of the
chamber. This is to avoid significant inaccuracy being introduced through uncertainty in the end wall thickness
and end field distortion region when correcting the recorded result by the active length/unoccluded length ratio.
The dose in mGy.100 mAs-1 at a specified kVp and slice width can be used for
patient dose calculations and for comparisons between scanners.
As a check on the validity of the results, radiation outputs in air at the centre
of the gantry at 120 kVp are typically in the region of 20 – 40 mGy.100 mAs-1 .
The radiation output in mGy.100 mAs-1 should be constant with mA over the
available tube current range or mAs over the available range of settings.
-
linearity coefficient = X max X min
X max + X min
where Xmax and Xmin are respectively the maximum and minimum output
values. The numerical value of the linearity coefficient must be ≤ 0.1, but the
parameter range over which it is measured depends on the type of x-ray
generator. See Appendix 2 for further information.
Purpose of Test
Equipment Required
Method
3. Confirm that movement of the table will not entangle the cable from the
ion chamber to the electrometer.
5. Add aluminium filtration between the x-ray tube and the ion chamber by
positioning the filters at the gantry aperture ensuring that the x-ray beam
is intercepted.
6. Repeat the scan of the ion chamber and record the reading.
1
also known as a scanogram, topogram or scan projection radiograph.
t b ln (2 D a D 0 ) − t a ln (2 D b D 0 )
HVL =
ln (D a D b )
The HVL measured will increase with an increase in beam filtration, tube
voltage and generator output pulse frequency.
While the HVL could be measured at any nominated kVp, 120 kVp is
recommended for consistency.
1
AS/NZS 3200.2.44:2000 Medical electrical equipment
Part 2.44: Particular requirements for safety – X-ray equipment for computed tomography.
Purpose of Test
Equipment Required
B
E
CT phantom C
A
Ion chamber
Electrometer
Method
1. Locate the complete Perspex body phantom in the centre of the gantry.
Ensure that the centre hole of the phantom is congruent with the centre of
the gantry.
2. Position the phantom so that the slice plane passes through the centre of
the phantom.
3. Place the pencil ion chamber in the central hole of the body phantom,
shown as A in Figure 1.
4. Check that the slice plane will pass through the centre of the ion
chamber.
9. Repeat at the 4 locations near the edge and within the body phantom
which are shown as B, C, D and E in Figure 1.
10. Expose the ion chamber at the centre of the gantry without the Perspex
phantom in place using a slice thickness of 8 to 10 mm. This will be
referred to as the free-in-air CTDI for the nominated slice thickness.
The CTDI of a single slice is the line integral of the total dose measured over
the length of the pencil ionisation chamber and can be expressed by the
equation —
DL
CTDI =
t
where D is the dose reading given by the electrometer, L is the active length of
the pencil chamber and t is the nominal slice thickness
The CTDI may be compared with the values quot ed in the manufacturer’s
literature. For comparative purposes the full test conditions must be stated
i.e., exposure factors used, the position of the chamber within the phantom and
the slice thickness.
For the purpose of this workbook, the “standard” CTDI is expressed as the
dose per 100 mAs with the chamber located in the centre of the phantom.
The free in air CTDI may be used in patient dose assessments, such as the CT
dose software package from the UK National Radiological Protection Board
NRPB SR250.
1
ImPACT National CT Scanner Dose Survey, 2000-2004. http://www.impactscan.org/dosesurvey.htm
Purpose of Test
To determine any areas of radiation leakage through the x-ray tube housing
and collimation, and to measure this leakage. Alternative methods are
described:-
Ø The preferred general method (A) describes the use of x-ray film or other
suitable image receptors to identify any areas of untoward leakage
radiation so that those discrete areas can be measured using an ion
chamber rate meter or an integrating dosemeter as may be appropriate to
the x-ray equipment under test.
Note: Once this test has been performed, subsequent testing should only
be necessary if there has been service or maintenance requiring
disassembly of the x -ray tube housing.
Caution: This test will require the assistance of licensed service personnel to
gain access to the x -ray tube housing. The x -ray tube must also be
stationary during the test.
Warning: Before using high voltage for this test, consult section 1.4,
Equipment Rating and Tube Voltage Range.
Equipment Required
Methods
1. Set the scanner exposure parameters for a scout scan that provides the
highest tube voltage and the lowest tube current. Record these values.
2. Open the covers of the scanner gantry and fix the 3 mm lead section
across the collimator of the x-ray tube so that the primary x-ray beam
will be fully intercepted.
4. Develop the film and visually check for any areas of significant 1
radiation leakage.
B - Dosemeter method
1. Set the scanner exposure parameters for a scout scan that provides the
highest tube voltage and the lowest tube current. Record these values.
2. Open the covers of the scanner gantry and fix the 3 mm lead section
across the collimator of the x-ray tube so that the primary x-ray beam
will be fully intercepted.
3. Set the ionisation chamber, in integrate mode, at 100 cm from the tube
focal spot in an accessible direction. With the tube stationary, make an
exposure to give a suitable detector reading. Repeat the exposure with
the dosemeter in other accessible directions around the tube housing.
4. Select the maximum value of dose obtained and use it to calculate the
leakage radiation value.
1
"Significant" is difficult to define and depends to some extent on judgement and experience. Any areas of
increased film density could be considered for measurement.
3. IMAGE QUALITY
3.1 INTRODUCTION
Mean CT number, uniformity and noise are important measures of the
imaging performance of a CT scanner. The owner should ensure that these
parameters are tested at regular intervals not exceeding one month.
Purpose of test
Equipment Required
Method
1. Position and secure the CT water phantom in the centre of the gantry.
2. Scan the phantom through the central region using exposure factors of
120 kVp, 300 mAs, 8 mm slice width for both head and body sized
phantoms. If these values of kVp, mAs and slice width are not available,
use the nearest practicable values. Record these values.
3. Select regions of interest (ROIs) at the centre of the phantom image and
at least four other sites approximately two-thirds of the phantom’s radius
towards the phantom’s edge. The ROIs should be circles of 20 mm
diameter.
♦ mean CT numbers for water at the central and outer ROI positions;
4. Print a hard copy of the image(s), recording the scan factors and other
data.
The mean CT number of water in the central ROI should not deviate from zero
by more than ± 4 CT numbers.
The uniformity of the CT numbers in the image should be such that the mean
CT number of the central ROI is within ± 2 CT numbers of the mean CT
number of each outer ROI.
To make allowance for varying scan factors being used for the test, standard
scan factors of 120 kVp, 300 mAs, 8 mm slice width may be used to assess
values of noise variations between the different ROIs. To convert the noise
value, σm, of the measured set (m) of scan parameters to the estimated noise
value σs for the standard set, use the formula —
The noise values of the outer ROI’s should not deviate by more than 2 CT at
the standard factors i.e.-
maximum noise – minimum noise ≤ 2 CT.
Purpose of Test
Equipment Required
Method
1. Position and secure the CT phantom so that a slice will pass through the
central area of the linearity insert.
2. Using factors of 120 kVp, 300 mAs and 8 mm slice width, or as near
these factors as practicable, scan the phantom through the central region.
4. Print a hard copy of the image(s) recording the scan factors and other
data.
Purpose of test
Equipment Required
Method
3. Scan the phantom and examine the image using the values of the
parameters shown in Table 3.
5. Print hard copies of all images taken, recording the scan factors and other
data. Record the values of any scan factor used which does not appear
on the films.
For hole or line-pair test inserts use a wide window to visually assess the
scanner resolution.
If the test insert consists of rows of holes, record the diameter of the holes
from the row in which the smallest holes are completely resolved from each
other.
If a line-pair test object is used, record the highest frequency which can be
completely resolved in line pairs.mm-1 .
In the case of MTF calculations, if the test insert is, for example, a bead,
ensure that the row of pixels from which the bead profile is taken runs through
the centre of the image in either, or both of, the horizontal and vertical
directions. The profile must also run fully across the bead from one side to the
other so that the CT numbers start and end in those of the background
material. Avoid including any noise spikes present in the background as these
will influence the high frequency end of the MTF calculated from the bead
point spread function.
Take the MTF cut -off value for assessing resolution as 0.02 unless the MTF
curve is approaching the spatial frequency axis extremely slowly in this
region, in which case use 0.05.
Purpose of test
To compare the nominal slice width with the full width half maximum
(FWHM) value of the slice sensitivity profile for axial, helical and multiple
slice scans.
Equipment required
Method
Axial scanner
3. Select a slice width and scan the phantom using factors of 120 kVp,
300 mAs, or as close as practicable, and the number of projections, scan
field diameter and reconstruction filter normally used for a head scan.
4. Analyse the image of the ramp using the profile software of the scanner
and determine the FWHM value of the profile.
The image width can then found by setting the cursor on each column
and using the scanner software to give the distance interval, or by
multiplying the pixel size by the number of pixels between these two
columns. The result must be multiplied by the tangent of the ramp angle
with the scanner plane to give the slice thickness.
e.g. for 45° the tangent = 1; for 26.6° the tangent = 0.5
5. Print a hard copy of the image containing the profile data obtained.
Helical scanner
Perform steps 1 - 6 above as for the axial scan method with zero table
movement.
2. Analyse the image of the ramp for each of the multiple slice profiles
obtained and determine the FWHM value of at least one profile at the
centre of the beam and one at the edge.
3. Print a hard copy of the image containing the profile data obtained.
The slice width is the FWHM value of the slice sensitivity profile. The service
contract for the scanner should involve collimator tests to ensure that only the
minimum volume of tissue necessary is exposed to x -rays in obtaining the
image.
Ramp inserts must be set up at the correct angle to the slice plane and also not
rotated with respect to the horizontal plane. The double opposed ramp
technique allows self cancellation of small angular errors in alignment of the
test insert with respect to the slice plane, but care should always be taken to
achieve accuracy in set -up to within at least 5°. Single-ramp techniques,
however, have no means of correcting for an introduced error in set -up angle.
If, for example, a ramp using a factor of 0.5 is misaligned by ± 3°, errors in
calculated slice width of –12% and +15% respectively will result. For a ramp
of factor 1, the respective errors will be -10% and +11%.
The calculated slice width should be within ± 0.5 mm of the nominal value for
all slice widths. If this criterion cannot be met but the result is within the
manufacturer’s specifications, include a copy of the relevant specification item
together with the completed report form.
For multiple slice scanners, differences in profile FWHM values may indicate
misalignment in the target-collimator-detector system.
Purpose of Test
Equipment Required
Method
1. Position the CT phantom so that a slice will pass through the central area
of the slice thickness insert.
2. Scan the phantom through the central region using 120 kVp, 300 mAs, or
as near as practicable, and the minimum slice thickness.
3. From the CT control console move the phantom into the scanner the
width of the slice thickness selected.
6. Repeat the table movement four more times and find the average value of
the five table movements.
7. Remove the CT phantom and load the table with the simulated patient.
Fix the ruler to the table support and the marker to the table. Set the
marker to the required zero position and note the position indicator
reading.
8. Move the table, either manually or under computer control, to its extreme
position in one direction and return it to its zero point. Note any
difference in the position indicator reading. Drive the table to its extreme
position in the opposite direction and note any position discrepancy.
9. Repeat the procedure at least three times and find the mean of the values
of position discrepancy.
There should also be no gaps between the images if the table is indexing
properly.
Purpose of Test
To assess the accuracy of the slice positioning light with respect to the x-ray
beam.
Equipment Required
Ø CT phantom
Ø Short lengths of ≤ 1 mm diameter wire or solder
Ø Adhesive tape
Method
1. Position the CT phantom at the centre of the gantry and turn on the
positioning light.
3. Scan the phantom with a single slice using the minimum slice width
available.
The alignment is acceptable if the full lengths of the marker segments are
visible with high contrast in the image. If the markers are not seen at all in
the image, the alignment is not acceptable.
RELATED PUBLICATIONS
1. Radiation Safety Act 1975 (Western Australia), and Amendments
Radiation Safety (General) Regulations 1983 (Western Australia), and Amendments,
Schedule IX
7. Poletti J. Radiation Dose and Image Qua lity of CT Scanners: Summary of NRL
Surveys 1980 - 1987. Report NRL 1988/1, 1988. National Radiation Laboratory,
Christchurch, New Zealand
APPENDIX 1
Blank page
Measurement Errors
Two general cases of error are recognised; namely the error in the measurement of a quantity
such as dose rate and the combination of the required limit on error in kVp with the error of
the measuring instrument. It has been acknowledged in the workbook that allowance needs to
be made for the error in measuring instruments under defined conditions.
This is a case in which the required limit is given as an error in a quantity rather than the
quantity itself. In this case the following approach has been taken.
The required kVp accuracy stated in the first edition of the workbook is the lower of ± 5.0%
or ± 5 kVp. This requirement still stands but an allowance for the error introduced into the
measurement by the measuring instrument used has now been incorporated into the
assessment by using the root mean square method of combining independent errors of x and y
to give a total probable measured error, where
< ± 5.5%,
≥ ± 5.5% and ≤ ± 6.0%,
> ± 6.0%.
• the kVp is borderline and attention may be required to the unit if the measured
error is equal to or greater than ± 5.5% and less than or equal to ± 6.0%. In this
case the qualified expert must exercise judgment on whether or not the unit meets
compliance requirements (e.g. in considering the number of kVp settings
involved).
At tube voltages in excess of 100 kVp, the basic requirement remains at ± 5.0 kVp.
For the same assumed measurement instrument error of 3.0% the three assessment
bands are then taken as
• the kVp passes if the measured error is less than ± 5.5 kVp
• the kVp is borderline and attention may be required to the unit if the measured
error is equal to or greater than ± 5.5 kVp and less than or equal to ± 6.0 kVp. In
this case the qualified expert must exercise judgment on whether or not the unit
meets compliance requirements (e.g. in considering the number of kVp settings
involved).
• the kVp fails if the measured error is greater than ± 6.0 kVp.
Summary
APPENDIX 2
Blank page
X1 − X 2
LC =
X1 + X 2
where X1 and X2 are the radiation outputs per mAs at two different mA or time settings. The
coefficient is a function of either tube current or exposure time in machines in which these
can be selected and one maintained constant, and on the mAs in machines in which they
cannot. The linearity coefficient value is therefore a function of the errors in these quantities.
There are three cases to consider:-
a) For constant mA units, eg dental units, the linearity coefficient depends only on the
timer error, assuming the coefficient of variation of the current reproducibility is small
by comparison. The allowed timer error is subject to recommendation by Standards
Australia. It can be easily shown from the above equation that if the allowed error in
the timer is ± p%, the value of the linearity coefficient is p/100. This simple
relationship does not hold if the error is stated as a fixed parameter value e.g.
± 1 pulse. In the timer range where the requirement is ± 10%, this gives the generally
accepted value of 0.1 for the linearity coefficient.
For this type of machine, if the mA were truly constant, the linearity coefficient test
would be redundant in those cases where timer linearity is shown by direct
measurement of timer accuracy. There are some cases, however, where a direct
comparison with a known time setting may not be possible, eg in object programmed
dental units. The linearity test is necessary in this case and it is required for all others
because it can also give information on the constancy of the tube current across the
timer range.
b) For units in which the mA can be varied and the time held constant for the linearity
test, the coefficient is a function of the mA error. The use of 0.1 for the linearity
coefficient implies acceptance of a ± 10% error in mA, and practical results of tests
imply that errors less than this can be routinely achieved in variable current
radiographic units. The test can also be applied to fluoroscopic units by determining
the ratios of output dose rate to the set mA over a range of tube currents at a given
kVp.
c) The use of 0.1 for the coefficient implies a combined error in the mA and time in mAs
machines of ± 10%. Again, practical results of the linearity coefficient test indicate
that errors of less than ± 10% can be routinely achieved in these units.
In practice, the recommendations for timer accuracy can be used as a model to set the linearity
coefficient for all types of x-ray units. If an accuracy of ± 10% is used in all cases for the
higher range of the parameter, the linearity coefficient will be uniformly 0.1.
However, if a fixed value of the error parameter is recommended, that value becomes a
varying percentage of the parameter setting. This results in the linearity coefficient varying
with the setting. If variable linearity coefficients were tabulated in the relevant working
ranges for each of the foregoing three types of unit, the workbooks, and their application,
would become unnecessarily complicated. The use of the linearity coefficient is therefore
discontinued in these potentially diffic ult, low parameter range and patient dose regions.
Summary
Fixed mA -
Half-wave rectified 0.1 at ≥ 200 ms No test < 200 ms
Full-wave rectified 0.1 at ≥ 100 ms No test < 100 ms
Other generators 0.1 at ≥ 100 ms No test < 100 ms
Variable mA –
tested at constant time in 0.1 At all available mA settings
radiographic mode
Variable mA –
tested as dose rate in 0.1 at ≥ 1 mA No test < 1 mA
fluoroscopic mode
mAs units – At all available mAs
0.1
(ex capacitor discharge) settings
Capacitor discharge units No test No test