Julabo fl300
Julabo fl300
Recirculating Coolers
FL300
FL601
English
19514818-V5.docx 04.07.19
Congratulations!
You have made an excellent choice.
JULABO thanks you for the trust you have placed in us.
This operating manual has been designed to help you gain an understanding of the principles of
operating and possibilities of our circulators. For optimum utilization of all functions, we recommend
that you thoroughly study this manual prior to beginning operation.
2
EC Conformity
The products described in the operating instructions conform to the
requirements of the following European guidelines:
Directive of the European Parliament and of the Council on the approximation of the laws of
the Member States relating to machinery.
Low voltage regulations with respect to legal harmonization of the member countries
concerning electric devices for use within certain voltage limits.
EMC guideline with respect to legal harmonization of the member countries concerning
electromagnetic compatibility.
RoHS directive on the restriction of the use of certain hazardous substances in electrical and
electronic equipment.
JULABO GmbH
Gerhard-Juchheim-Strasse 1
77960 Seelbach / Germany
3
4
5
TABLE OF CONTENTS
1. Intended use ................................................................................................................... 8
1.1. Description ........................................................................................................... 8
2. Operator responsibility – Safety instructions ................................................................... 8
2.1. Disposal...............................................................................................................12
3. Technical specifications .................................................................................................13
4. Safety notes for the user ................................................................................................20
4.1. Explanation of safety notes .................................................................................20
4.2. Explanation of other notes ...................................................................................20
4.3. Safety instructions ...............................................................................................21
5. Installation ......................................................................................................................23
5.1. Tubing .................................................................................................................26
6. Operating controls and functional elements ...................................................................27
7. Operating procedures ....................................................................................................29
7.1. Bath fluids............................................................................................................29
7.2. Power connection ................................................................................................30
7.3. Filling ...................................................................................................................31
7.4. Switching on / Start - Stop ...................................................................................32
7.5. Setting the temperatures .....................................................................................33
7.6. AUTOSTART ON / OFF ......................................................................................34
7.7. Remote control: activate – deactivate .................................................................35
8. Safety installations .........................................................................................................36
6
8.1. Excess temperature protection ........................................................................... 36
8.2. Low level protection ............................................................................................ 36
9. Troubleshooting guide / Error messages ....................................................................... 37
10. Electrical connections............................................................................................ 39
11. Remote control ...................................................................................................... 40
11.1. Setup for remote control...................................................................................... 40
11.2. Communication with a PC or a superordinated data system............................... 41
11.3. List of commands ................................................................................................ 43
11.4. Status messages................................................................................................. 44
11.5. Error messages ................................................................................................... 44
12. Cleaning / repairing the unit .................................................................................. 45
13. Adequate storing of operating manual .................................................................. 47
14. Draining ................................................................................................................. 48
15. Warranty conditions .............................................................................................. 49
7
1. Intended use
JULABO recirculating coolers have been designed for temperature application to specific fluids. The
pump connections can be used for cooling applications in an external circuit at a constant
temperature.
JULABO water baths are not suitable for direct temperature control of foods, semi-
luxury foods and tobacco, or pharmaceutical and medical products. Direct temperature
control means unprotected contact of the object with the bath medium (bath fluid).
1.1. Description
The recirculating coolers are operated via the splash-proof keypad. The
implemented microprocessor technology allows to set and to store the setpoint that
can be indicated on the LED temperature display.
The PID temperature regulation is used to withdraw heat from the bath fluid by
means of the cooling machine and to automatically regulate the required need.
Electrical connections:
PID1 1. The serial interface RS232 allows modern process technology without
additional interface.
2. Alarm output for external alarm message.
RS232
8
The operator is responsible for the qualification of the personnel operating the units.
The personnel operating the units should be regularly instructed about the dangers involved with
their job activities as well as measures to avert these dangers.
Make sure all persons tasked with operating, installing, and maintaining the unit have read and
understand the safety information and operating instructions.
When using hazardous materials or materials that could become hazardous, the unit may be
operated only by persons who are absolutely familiar with these materials and the unit. These
persons must be fully aware of possible risks.
If you have any questions concerning the operation of your unit or the information in this manual,
please contact us!
Contact JULABO GmbH Tel. +49 7823 51-0 info.de@julabo.com
Gerhard-Juchheim-Strasse 1 Fax +49 7823 2491 www.julabo.com
77960 Seelbach / Germany
10
When using hazardous materials or materials that could become hazardous, the operator must affix
the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible:
1 Danger area. Attention! Observe instructions.
(operating manual, safety data sheet)
Observe the instructions in the manuals for instruments of a different make that you connect to the
circulator, particularly the corresponding safety instructions. Also observe the pin assignment of
plugs and technical specifications of the products.
11
2.1. Disposal
This unit contains the refrigerants R134a, R404A or R452A, which at this time are not considered
harmful to the ozone layer. However, over the long operating period of the unit, disposal rules may
change. Therefore, only qualified personnel should handle the disposal.
Valid in EU countries
See the current official journal of the European Union – WEEE directive.
Directive of the European Parliament and of the Council on waste electrical and
electronic equipment (WEEE).
This directive requires electrical and electronic equipment marked with a crossed-
out trash can to be disposed of separately in an environmentally friendly manner.
Contact an authorized waste management company in your country.
Disposal with household waste (unsorted waste) or similar collections of municipal
waste is not permitted!
12
3. Technical specifications
Recirculating Cooler FL300
Working temperature range °C -20 ... +40
Temperature stability °C ±0.5
Electrical connections:
Computer interface RS232
Alarm output for external alarm signal
Circulating pump:
discharge, max.at 0 bar l/min 15
pressure, max. at 0 Liters bar 0.35
13
Recirculating Cooler FL300
bar
Filling level indicator sight glass
Filling volume from ... to Liters 3 ... 4.5
Dimensions (WxLxH) cm 25x50x60
Weight kg 35
Ambient temperature range °C 5 ... 40
Return flow temperature °C 80 max.
IP class according to IEC 60 529 IP 21
14
Recirculating Cooler FL300
All measurements have been carried out at: rated voltage and frequency,
ambient temperature: 20 °C
15
Recirculating Cooler FL601
Working temperature range °C -20 ... +40
Temperature stability °C ±0.5
Temperature selection: digital
via key pad indication on LED-DISPLAY
remote control via personal computer indication on monitor
Temperature indication: LED-DISPLAY
Resolution °C 0.1
Temperature control PID 1
Temperature sensor Pt 100
Excess temperature protection 85 °C - fixed value
Low liquid level protection float switch
Cooling capacity °C +20 +10 0 -10 -20
Medium: Mixture water-glycol kW 0.6 0.5 0.4 0.33 0.2
Cooling compressor 1- stage / air cooled
Refrigerant R404A, R452A*
Electrical connections:
Computer interface RS232
Alarm output for external alarm signal
* at 230 V / 50 Hz
16
Circulating pump:
discharge, max.at 0 bar l/min 23
pressure, max. at 0 Liters bar 1.0
bar -----
Filling level indicator sight glass
Filling volume from ... to Liters 5.5 ... 8
Dimensions (WxLxH) cm 32x50x60
Weight kg 48
Ambient temperature range °C 5 ... 40
Return flow temperature °C 80 max.
IP class according to IEC 60 529 IP 21
17
Warning functions and safety installations
Excess temperature protection 85 °C - fixed value
Low liquid level protection float switch
Alarm message optical + audible (permanent)
Excess temperature - Warning function 75 °C
Overload protection for compressor and pump motor
Classification according to DIN 12876-1 class I
18
Information about the used refrigerants
The Regulation (EU) No. 517/2014 on fluorinated greenhouse gases applies to all systems which
contain fluorinated refrigerants and replaces (EC) 842/2006.
The aim of the Regulation is to protect the environment by reducing emissions of fluorinated
greenhouse gases.
Among other things it regulates the emission limits, use and recovery of these substances. It also
contains requirements for operators of systems which require / contain these substances to function.
Under Regulation 517/2014, the operator of a system of this nature has the following duties:
• The operator must ensure that the equipment is checked at regular intervals for leaks.
• These intervals depend on the CO2 equivalent of the system. This is calculated from the
refrigerant fill volume and type of refrigerant. The CO2 equivalent of your system is shown on
the model plate.
• The operator undertakes to have maintenance, repair, service, recovery and recycling work
carried out by certified personnel who have been authorized by JULABO.
• All such work must be documented. The operator must keep records and archive them for at
least five years. The records must be submitted to the relevant authority on request.
19
4. Safety notes for the user
4.1. Explanation of safety notes
In addition to the safety warnings listed above, warnings are posted throughout the
manual. These warnings are designated by an exclamation mark inside an equilateral
triangle. “Warning of a dangerous situation (Attention! Please follow the
documentation).”
The danger is classified using a signal word.
Read and follow these important instructions.
Warning:
Describes a possibly highly dangerous situation. If these instructions are not followed,
serious injury and danger to life could result.
Caution:
Describes a possibly dangerous situation. If this is not avoided, slight or minor injuries
could result. A warning of possible property damage may also be contained in the text.
Notice:
Describes a possibly harmful situation. If this is not avoided, the product or anything in
its surroundings can be damaged.
Important!
Indicates usage tips and other useful information.
20
4.3. Safety instructions
Follow the safety recommendations to prevent damage to persons or property.
Further, the valid safety instructions for working places must be followed.
• Only connect the unit to a power socket with earthing contact (PE – protective
earth)!
• The power supply plug serves as a safe disconnecting device from the line and
must always be easily accessible.
• Place the instrument on an even surface on a pad made of non-inflammable
material.
• Do not stay in the area below the unit.
• Make sure you read and understand all instructions and safety precautions listed
in this manual before installing or operating your unit.
• Never operate the unit without bath fluid in the bath.
• Do not drain the bath fluid while it is hot or cold!
Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment for example).
• Use suitable connecting tubing.
• Make sure that the tubing is securely attached.
• Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
• Regularly check the tubing for material defects (e.g., for cracks).
• Never operate damaged or leaking equipment.
• Always turn off the unit and disconnect the mains cable from the power source
before performing any service or maintenance procedures, or before moving the
unit.
21
• Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
• Always empty the bath before moving the unit.
• Transport the unit with care.
• Sudden jolts or drops may cause damage in the interior of the unit.
• Observe all warning labels.
• Never remove warning labels.
• Never operate equipment with damaged mains power cables.
• Repairs are to be carried out only by qualified service personnel.
• Some parts of the bath tank and the pump connections may become extremely
cold during continuous operation. Therefore, exercise particular caution when
touching these parts.
22
5. Installation
• Place the unit on an even surface on a base made of
nonflammable material.
• Cooling machine, pump motor and electronics produce intrinsic
heat that is dissipated via the venting openings ! Never cover
these openings!
• Keep at least 20 cm of open space on the front and rear venting
grids.
• The place of installation should be large enough and provide
sufficient air ventilation to ensure the room does not warm up
excessively because of the heat the instrument rejects to the
environment.
For a fault (leakage) in the refrigeration system, the standard EN
378 prescribes a certain room space to be available for each kg
of refrigerant.
The necessary amount of refrigerant is specified on the type
plate.
> For 0.25 kg of refrigerant R134a, 1 m3 of space is required.
> For 0.52 kg of refrigerant R404A, 1 m3 of space is required.
> For 0.423 kg of refrigerant R452A, 1 m3 of space is required.
•
23
Example: FL300 • Connect the tubings for cooling the external system to
the pump connectors M16x1 for feed and return (12) on
the rear of the recirculating cooler.
15
A
M
P
15
A
M
P
- Feed – Return
Caution:
Securely attach all tubing to prevent slipping.
24
Notice: Flood hazard!.
In case the system to be cooled is located at a higher level than the recirculating
cooler, take note of bath liquid flowing back when the unit is switched off.
Return flow safety device
Should the filling volume of the bath tank not be sufficient, prevent the liquid from
flowing back by using shut-off valves..
The following questions shall help to recognize possible dangers and to reduce the
risks to a minimum.
• Are all tubes and electrical cables connected and installed?
Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.
• What to do when a dangerous substance was spilled on or in the unit?
Before starting to work, obtain information concerning the substance and
determine the method of decontamination.
25
5.1. Tubing
Caution:
• Employ suitable connecting tubing.
• Make sure that the tubing is securely attached.
• Avoid sharp bends in the tubing, and maintain a sufficient distance from surrounding
walls.
• Regularly check the tubing for material defects (e.g. for cracks).
• Preventive maintenance: Replace the tubing from time to time.
Recommended tubing:
Order No. Suitable for
8930008 1 m CR -tubing 8 mm inner dia. (-20 ... +120 °C)
® FL300
8930308 1 m Reinforced tubing 8 mm inner dia. (-40 ... +120 °C) FL601
8930312 1 m Reinforced tubing 12 mm inner dia. (-40 ... +120 °C) FL601
Tubing insulation
8930410 1 m Insulation, 14 mm inner dia.. CR®-tubing 8 mm inner dia
8930412 1 m Insulation, 18 mm inner dia. Reinforced tubing 8 mm inner dia.
8930413 1 m Insulation, 23 mm inner dia. Reinforced tubing 12 mm inner dia.
Tube clamps
8970480 2 Tube clamps, size 1 CR®-tubing 8 mm inner dia
8970481 2 Tube clamps, size 2 Reinforced tubing 8 mm inner dia.
8970482 2 Tube clamps, size 3 Reinforced tubing 12 mm inner dia.
26
6. Operating controls and functional elements
Example: FL300 Front view Top view Rear view
Push
10
14b
15 15
A A
11
M M
P P
15
Push 12
13
27
1 Mains power switch, spash-water protected
on
off
2.0 Keypad, spash-water protected
2.1 Edit keys (set point increase or decrease)
2.2 Enter key Store set point value / parameter
3.0 Indication
°C
3.1 LED temperature display
12
Pump connector M16x1: - Feed
Pump connector M16x1: – Return
28
13 Overflow connector
7. Operating procedures
7.1. Bath fluids
Caution:
No liability for use of other bath liquids!
Please contact JULABO before using other than recommended bath fluids. JULABO
takes no responsibility for damages caused by the selection of an unsuitable bath fluid
Do not use alcohols.
Water:
The quality of water depends on local conditions.
• Due to the high concentration of lime, hard water is not suitable for temperature
control because it leads to calcification in the bath.
• Ferrous water can cause corrosion - even on stainless steel.
• Chloric water can cause pitting corrosion.
• Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.
Water: - No liablity for use with water.
Danger of freezing at working temperatures <5 °C.
29
Recommended bath fluids:
Bath fluids Temperature range
soft/decalcified water 5 °C to 80 °C
Make sure that the line voltage and frequency match the supply voltage specified on
the type plate.
30
7.3. Filling
Take care that no liquid enters the interior of the circulating
Top view cooler.
Connect the tubing from the external system to the pump
connectors and check for leaks
Push Check to make sure that the drain tap (8) is closed.
31
7.4. Switching on / Start - Stop
Switching on:
The recirculating cooler is turned on and off with the mains switch.
The unit performs a self-test. All segments of the 4-digit LED temperature
DISPLAY and all indicator lights will illuminate
(as illustrated on the left).
Then the software version and the type of unit is indicated.
Examples: (v 1.02) (FL300)
The display "OFF" indicates the unit is ready to operate (standby mode).
32
7.5. Setting the temperatures
Factory setting: Setting can be carried out in the start/stop condition.
25 °C
1. Press one of the keys for a short moment.
The setpoint value instead of the actual value is indicated on the display for
about 8 seconds.
The value can now be changed.
2. Change value:
Press to set a higher value.
Press to set a lower value.
Keep the keys depressed for the value to change fast.
3. Press enter to store the value.
33
7.6. AUTOSTART ON / OFF
The recirculating cooler has been configured and supplied by JULABO
according to N.A.M.U.R. recommendations. This means for the start mode,
that the unit must enter a safe operating state after a power failure (non-
automatic start mode). This safe operating state is indicated by „OFF“ on the
LED temperature display. A complete shutdown of the main functional
elements such as compressor and circulating pump is effected
simultaneously.
Should such a safety standard not be required, the AUTOSTART function
(automatic start mode) may be activated, thus allowing the start of the
circulator directly by pressing the mains power switch or using a timer.
Keep depressed enter and
turn on the unit with the mains power switch.
For a short while the LED DISPLAY indicates the effective start mode:
34
7.7. Remote control: activate – deactivate
The recirculating cooler is to be prepared for remote control by a
personal computer via the serial interface RS232.
Set the interface item from >IOFF< to >ION<.
35
8. Safety installations
8.1. Excess temperature protection
This safety installation is independent of the control circuit.
When the temperature of the bath fluid has reached the safety temperature
+ The alarm is indicated by optical and audible signals (continuous tone) and on
the LED-DISPLAY appears the error message
"Error 01".
Turn off the unit with the mains switch, refill bath fluid and turn the unit on
again!
Caution:
For refill always use the same bath fluid type that is already in the bath.
36
Notice:
Check the low liquid level protection device at least twice a year!
• To execute a functional test, drain the liquid until the alarm for low liquid level is
triggered. Refill liquid afterwards.
9. Troubleshooting guide / Error messages
Whenever the microprocessor electronics registers a failure, a complete
shutdown of the compressor and circulating pump is performed. The alarm
• The recirculating cooler is operated without bath fluid, or the liquid level is
insufficient.
Replenish the bath tank with the bath fluid.
• Tube breakage has occured (insufficient filling level due to excessive bath
fluid pumped out). Replace the tubing and replenish the bath tank with the
bath fluid.
Cable of the working temperature sensor interrupted or short-circuited.
37
The return temperature is above the switch-off value of the high temperature
protection (85°C). Check dimensioning of application.
Use a stronger recirculating cooler if necessary.
After eliminating the malfunction, press the mains power switch off and on
again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized service
station.
38
10. Electrical connections
Notice: Use shielded cables only.
The shield of the connecting cable is electrically connected to the plug housing.
Use a null modem cable.
The unit ensures safe operation if connecting cables with a maximum length of 3 m are
used. The use of longer cables does not affect proper performance of the unit, however
external interferences may have a negative impact on safe operation.
Pin assignments:
5 1 Pin 2 RxD Receive Data
Pin 3 TxD Transmit Data
9 6 Pin 5 0V Signal GND
RS232C
Pin 7 RTS Request to send
Pin 8 CTS Clear to send
39
Alarm output
Potential-free change-over contact for external alarm signal.
Type RS232
Baudrate 4800 bauds
Parity even
Handshake hardware handshake
Data bits 7
Stop bit 1
40
11.2. Communication with a PC or a superordinated data system
If the recirculating cooler is put into remote control mode the MULTI-DISPLAY
(LED) will read „R -OFF-„ = REMOTE STOP. The recirculating cooler is now
operated via the computer.
In general, the computer (master) sends commands to the recirculating cooler
(slave). The recirculating cooler sends data (including error messages) only
when the computer sends a query.
41
Important times for a command transmission:
To ensure a safe data transfer, the time gap between two commands should
be at least 250 ms.
The recirculating cooler automatically responds to an in command with a data
string followed by a LF (Line Feed). The next command should only be sent
after 10 ms.
42
11.3. List of commands
43
11.4. Status messages
Status messages Description
44
Error messages Description
-11 VALUE TOO LARGE Entered value too large.
-12 TEMPERATURE MEASUREMENT Error in A/D converter.
ALARM
-14 EXCESS TEMPERATURE The return temperature is above the switch-off value of
PROTECTOR ALARM the high temperature warning function of 85 °C.
Check dimensioning of application.
Use a stronger recirculating cooler if necessary.
Notice:
Risk of injury for hands when mounting the venting grid.
45
To maintain the full cooling performance, clean the condenser from time to time.
Cleaning:
Clean the outside of the unit using a wet cloth and low surface tension water.
The recirculating cooler is designed for continuous operation under normal
conditions. Periodic maintenance is not required.
The tank should be filled only with a bath fluid recommended by JULABO. To
avoid contamination, it is essential to change the bath fluid from time to time.
Repairs:
Before asking for a service technician or returning a JULABO instrument for
repair, please contact an authorized JULABO service station.
46
When returning the unit:
• Clean the unit in order to avoid any harm to the service personnel
• Attach a short fault description.
• When returning a unit, take care of careful and adequate packing.
• JULABO is not responsible for damages that might occur from insufficient
packing.
JULABO reserves the right to carry out technical modifications with repairs for
providing improved performance of a unit.
47
14. Draining
Notice:
Store and dispose the used bath fluid according to the laws for environmental
protection.
48
15. Warranty conditions
JULABO GmbH warrants its products against defects in material or in workmanship, when used
under appropriate conditions and in accordance with appropriate operating instructions
for a period of ONE YEAR.
With the ‘1PLUS warranty’ the user receives a free of charge extension to the warranty of up to 24
months, limited to a maximum of 10 000 working hours.
To apply for this extended warranty the user must register the unit on the JULABO web site
www.julabo.com, indicating the serial no. The extended warranty will apply from the date of JULABO
GmbH’s original invoice.
JULABO GmbH reserves the right to decide the validity of any warranty claim. In case of faults
arising either due to faulty materials or workmanship, parts will be repaired or replaced free of
charge, or a new replacement unit will be supplied.
49
JULABO GmbH Tel. +49 (0) 7823 / 51-0 info.de@julabo.com
77960 Seelbach / Germany Fax +49 (0) 7823 / 24 91 www.julabo.com