Manual FTC en
Manual FTC en
SIWAREX® FTC
Danger
! means that serious material damage, severe injury or even death will result if the
corresponding safety precautions are not followed carefully.
Warning
! means that serious material damage, severe injury or even death can result if the corresponding
safety precautions are not followed carefully.
Caution
! means that material damage or minor injuries can result if the corresponding safety precautions are
not followed carefully.
Caution
means that material damage can result if the corresponding safety precautions
are not followed carefully.
Attention
refers to important information on the product, handling of a product or a
corresponding segment of the documentation to which special attention should be
given.
Qualified Personnel
Installation and operation of a device may only be performed by qualified personnel. Qualified personnel in a technical
safety sense within this manual are personnel that have the qualifications for installing, grounding and identifying all
devices, systems and circuits according to the applicable technical safety standards.
Intended Utilisation
Warning
! The device may only be utilised with the replacement parts described in the catalogue and the
technical description and only with foreign or external devices and components that are approved or
suggested by Siemens.
Fault-free and safe operation of the product depends on proper transport, proper storage, assembly,
installation, operation and maintenance.
iv SIWAREX FTC
SIWAREX FTC
Contents
SIWAREX FTC v
Preface
Contents
1 PREFACE ............................................................................................................................................ 1-1
1.1 PURPOSE OF THIS MANUAL ....................................................................................................... 1-1
1.2 FUNDAMENTAL KNOWLEDGE REQUIREMENTS .......................................................................... 1-1
1.3 SCOPE OF THIS MANUAL ........................................................................................................... 1-1
1.4 FURTHER SUPPORT .................................................................................................................... 1-2
2 SCOPE OF DELIVERY ..................................................................................................................... 2-4
2.1 SCOPE OF DELIVERY ................................................................................................................. 2-4
3 PRODUCT OVERVIEW.................................................................................................................... 3-5
3.1 GENERAL INFORMATION ........................................................................................................... 3-5
3.2 BENEFITS................................................................................................................................... 3-5
3.3 RANGE OF APPLICATION ........................................................................................................... 3-6
3.4 STRUCTURE ............................................................................................................................... 3-6
3.5 FUNCTION ................................................................................................................................. 3-6
3.6 SYSTEM INTEGRATION IN SIMATIC ......................................................................................... 3-7
3.7 COMMISSIONING AND SERVICE WITH SIWATOOL FTC .......................................................... 3-8
3.8 FIRMWARE DOWNLOAD WITH SIWATOOL FTC ..................................................................... 3-9
3.9 READING THE STORED WEIGHING LOGS WITH SIWATOOL FTC .............................................. 3-9
4 HARDWARE PLANNING AND ASSEMBLY ...............................................................................4-10
4.1 PLANNING THE HARDWARE IN SIMATIC ................................................................................4-11
4.2 EMC-COMPATIBLE STRUCTURE ..............................................................................................4-11
4.2.1 Definition: EMC ..............................................................................................................4-11
4.2.2 Introduction .....................................................................................................................4-11
4.2.3 Possible Effects of Interference .......................................................................................4-12
4.2.4 Coupling Mechanisms .....................................................................................................4-12
4.2.5 Five Basic Rules for Guaranteeing EMC ........................................................................4-12
4.3 ASSEMBLY ON THE PROFILE RAIL ............................................................................................4-14
4.4 CONNECTIONS AND CABLING ...................................................................................................4-14
4.4.1 Connection areas for SIWAREX FTC..............................................................................4-14
4.4.2 Shield connection.............................................................................................................4-15
4.4.3 Connecting the 24 V power supply ..................................................................................4-16
4.4.4 Connection to front connector .........................................................................................4-17
4.4.5 Load cell connections ......................................................................................................4-17
4.4.6 Digital Inputs...................................................................................................................4-19
4.4.7 Counter Input...................................................................................................................4-21
4.4.8 Digital Outputs ................................................................................................................4-21
4.4.9 Analog Output .................................................................................................................4-23
4.4.10 RS 485 Interface ..............................................................................................................4-23
4.4.11 Connecting the Remote Display by the Siebert company ................................................4-24
4.4.12 PC Connection for SIWATOOL FTC...............................................................................4-25
4.4.13 LED Indicators ................................................................................................................4-26
4.4.14 Using the Micro Memory Card........................................................................................4-26
4.5 OPERATIONAL PREPARATION ...................................................................................................4-27
4.6 USE FOR POTENTIALLY EXPLOSIVE AREA .................................................................................4-28
5 WEIGHING FUNCTIONS................................................................................................................5-29
5.1 GENERAL INFORMATION ..........................................................................................................5-29
5.2 DR3 ADJUSTMENT PARAMETER ...............................................................................................5-30
5.2.1 DR3 – Adjustment digits 0… 4, for the zero point and adjustment weights 1… 4 ...........5-33
5.2.2 DR3 – Characteristic value range ...................................................................................5-35
5.2.3 DR3 – Filter sequence of the signal filter........................................................................5-36
vi SIWAREX FTC
SIWAREX FTC
SIWAREX FTC ix
Preface
x SIWAREX FTC
SIWAREX FTC
8.6.2 Installation and Project Planning for the Calibratable Weight Display .......................7-110
9 PROJECT PLANNING IN SIMATIC PCS 7................................................................................9-113
9.1 GENERAL INFORMATION ........................................................................................................9-113
9.2 FB FOR SIWAREX FTC........................................................................................................9-114
9.2.1 FB643 for CFC..............................................................................................................9-114
9.2.2 Function and Functionality ...........................................................................................9-114
9.2.3 Addressing and Driver Wizard ......................................................................................9-114
9.2.4 Manual/Automatic .........................................................................................................9-115
9.2.5 Data records..................................................................................................................9-115
9.2.6 Commands .....................................................................................................................9-116
9.2.7 Module error messages .................................................................................................9-116
9.2.8 Allocating message text and message class to the block parameters ............................9-117
9.2.9 Connections from SFT_FTC (without data records) .....................................................9-117
9.2.10 Calibration parameter (Data record 3): .......................................................................9-119
9.2.11 Base parameter (Data record 4): ..................................................................................9-122
9.2.12 Parameter for belt scale (Data record 5):.....................................................................9-123
9.2.13 Interface parameter (Data record 7):............................................................................9-125
9.2.14 Date/Time (Data record 8):...........................................................................................9-129
9.2.15 Application ID (Data record 9):....................................................................................9-129
9.2.16 Current belt position (angle of engagement) (Data record 14):....................................9-130
9.2.17 Tare input weight (Data record 15): .............................................................................9-130
9.2.18 Weight simulation value (Data record 16): ...................................................................9-130
9.2.19 Ext. Analog default value (Data record 17): .................................................................9-131
9.2.20 Ext. display default value (Data record 18): .................................................................9-131
9.2.21 Fill amount (Data record 21):.......................................................................................9-131
9.2.22 Process values (Data record 30):..................................................................................9-132
9.2.23 Extended process values (Data record 31):...................................................................9-134
9.2.24 Totalizers (Data record 33):..........................................................................................9-134
9.2.25 ASCII weight value (Data record 34):...........................................................................9-134
9.2.26 Encryption data (Data record 35):................................................................................9-135
9.2.27 Last log data (Data record 44):.....................................................................................9-135
9.2.28 Supplement string (Data record 45):.............................................................................9-136
9.3 EXAMPLES FOR IMAGE BLOCKS FOR SIWAREX FTC ............................................................9-136
9.3.1 Faceplate display in OS.................................................................................................9-136
9.3.2 Faceplate Creation ........................................................................................................9-139
10 COMMISSIONING USING A PC – SIWATOOL FTC...........................................................10-142
10.1 GENERAL .............................................................................................................................10-142
10.2 WINDOWS AND FUNCTIONS OF THE SIWATOOL FTC.........................................................10-142
10.3 OFFLINE PROJECT PLANNING ...............................................................................................10-142
10.4 ONLINE OPERATION .............................................................................................................10-143
10.5 ASSISTANCE .........................................................................................................................10-144
11 FIRMWARE-UPDATE WITH SIWATOOL FTC ...................................................................11-145
11.1 ADVANTAGES OF THE FIRMWARE-UPDATE ..........................................................................11-145
12 CALIBRATING APPLICATIONS ............................................................................................12-147
12.1 GENERAL NOTE ....................................................................................................................12-147
12.2 CALIBRATABLE MAIN WEIGHT DISPLAY ...............................................................................12-147
12.3 READING THE CALIBRATABLE LOGS WITH SIWATOOL FTC ..............................................12-148
13 ACCESSORIES............................................................................................................................13-149
14 TECHNICAL DATA....................................................................................................................14-152
SIWAREX FTC xi
Preface
16 ABBREVIATIONS ......................................................................................................................16-161
Images
IMAGE 3-1 CONFIGURATION SIMATIC S7/PCS7 WITH SIWAREX FTC .......................................... 3-7
IMAGE 3-2 SIWATOOL FTC OVERVIEW .......................................................................................... 3-8
IMAGE 4-1 SIWAREX FTC FRONT VIEW .........................................................................................4-15
IMAGE 4-2 SHIELD CLAMP ASSEMBLY ................................................................................................4-16
IMAGE 4-3 LOAD CELL CONNECTION IN 4-WIRE SYSTEM ....................................................................4-19
IMAGE 4-4 LOAD CELL CONNECTION IN 6-WIRE SYSTEM ....................................................................4-19
IMAGE 4-5 DIGITAL INPUTS................................................................................................................4-20
IMAGE 4-6 COUNTER INPUT ...............................................................................................................4-21
IMAGE 4-7 DIGITAL OUTPUTS ............................................................................................................4-22
IMAGE 4-8 ANALOG OUTPUT ..............................................................................................................4-23
IMAGE 4-9 RS 485 CONNECTION........................................................................................................4-24
IMAGE 4-10 S11 DISPLAY CONNECTION ..............................................................................................4-24
IMAGE 4-11 PC CONNECTION ..............................................................................................................4-25
IMAGE 5-1 ADJUSTMENT DIGITS AND WEIGHT VALUE ........................................................................5-34
IMAGE 5-2 LINEARIZATION OF THE SCALE CHARACTERISTIC CURVE ..................................................5-35
IMAGE 5-3 STEP RESPONSES OF THE DIGITAL LOW PASS FILTER .........................................................5-36
IMAGE 5-4 STANDSTILL MONITORING ................................................................................................5-40
IMAGE 5-5 DEFINING LIMIT VALUE PARAMETER ................................................................................5-46
IMAGE 8-1 FB SIWA_FTC CALL PARAMETERS...............................................................................7-105
IMAGE 8-2 CALIBRATABLE DISPLAY IN TP/OP ................................................................................7-110
IMAGE 8-3 "SECURE OUTPUT" FUNCTION IN PROTOOL ....................................................................7-111
IMAGE 8-4 PARAMETER OF THE SIWAREX FTC OCX ...................................................................7-112
IMAGE 9-1 SFT_FTC FUNCTION BLOCK IN CFC..............................................................................9-115
IMAGE 9-2 STANDARD VIEW FOR SIWAREX FTC ..........................................................................9-136
IMAGE 9-3 SERVICE VIEW, TAB CALIBRATION 1/2 ...........................................................................9-137
IMAGE 9-4 SERVICE VIEW, TAB BELT SCALE ...................................................................................9-137
IMAGE 9-5 SERVICE VIEW, TAB COMMANDS ....................................................................................9-138
IMAGE 9-6 DOSING VIEW, TAB SUM 1/2 ...........................................................................................9-138
IMAGE 9-7 DOSING VIEW, TAB OPERATION......................................................................................9-139
IMAGE 9-8 COMBO-BOX WITH SEVERAL ENTRIES .............................................................................9-140
IMAGE 9-9 COMMAND SELECTION ...................................................................................................9-140
IMAGE 10-1 SIWATOOL FTC WINDOW BELT PARAMETERS .........................................................10-143
IMAGE 11-1 DOWNLOADING FIRMWARE WITH SIWATOOL FTC...................................................11-146
IMAGE 12-1 READ THE CALIBRATABLE MEMORY WITH SIWATOOL FTC......................................12-148
Tables
TABLE 1-1 VALIDITY OF THIS MANUAL.............................................................................................. 1-1
TABLE 1-2 CHAPTER OVERVIEW ......................................................................................................... 1-2
TABLE 4-1 REQUIREMENTS FOR N SIWAREX FTC...........................................................................4-11
TABLE 4-2 POWER SUPPLY CONNECTION ...........................................................................................4-17
TABLE 4-3 LOAD CELL CONNECTION .................................................................................................4-17
TABLE 4-4 DIGITAL INPUT CONNECTIONS ..........................................................................................4-20
TABLE 4-5 PULSE ENCODER CONNECTION..........................................................................................4-21
TABLE 4-6 DIGITAL OUTPUT CONNECTIONS .......................................................................................4-22
TABLE 4-7 ANALOG OUTPUT CONNECTIONS ......................................................................................4-23
TABLE 4-8 RS 485 CONNECTIONS ......................................................................................................4-23
TABLE 4-9 PC CONNECTION ..............................................................................................................4-25
TABLE 4-10 INDICATORS (LED) ..........................................................................................................4-26
TABLE 5-1 ALLOCATION OF DR3.......................................................................................................5-32
TABLE 5-2 ALLOCATION OF DR 4......................................................................................................5-44
TABLE 5-3 ALLOCATION OF THE DR 7...............................................................................................5-57
TABLE 5-4 ALLOCATION OF DR 8......................................................................................................5-65
TABLE 5-5 ALLOCATION OF DR 9......................................................................................................5-66
TABLE 5-6 ALLOCATION OF DR 14....................................................................................................5-66
TABLE 5-7 ALLOCATION OF DR 15....................................................................................................5-67
TABLE 5-8 ALLOCATION OF DR 16....................................................................................................5-67
TABLE 5-9 ALLOCATION OF DR 17....................................................................................................5-68
TABLE 5-10 ALLOCATION OF DR 18....................................................................................................5-69
TABLE 5-11 ALLOCATION OF DR 21....................................................................................................5-69
TABLE 5-12 ALLOCATION OF DR 30....................................................................................................5-70
TABLE 5-13 DR 30 – NAWI-STATUS BITS ..........................................................................................5-71
TABLE 5-14 DR 30 – CONTI STATUS FLAGS .........................................................................................5-72
TABLE 5-15 ALLOCATION OF DR 32....................................................................................................5-75
TABLE 5-16 ALLOCATION OF DR 33....................................................................................................5-76
TABLE 5-17 ALLOCATION OF DR 34....................................................................................................5-77
TABLE 5-18 ALLOCATION OF DR 35....................................................................................................5-78
TABLE 5-19 ALLOCATION OF DR 40....................................................................................................5-78
TABLE 5-20 PROCESS VALUES FOR ALLOCATING LOG FIELDS ..............................................................5-79
TABLE 5-21 ALLOCATION OF DR 44....................................................................................................5-80
TABLE 5-22 ALLOCATION OF DR 45....................................................................................................5-80
TABLE 5-23 CONSTRUCTION OF A LOGGING ELEMENT .........................................................................5-81
TABLE 5-24 OVERVIEW OF MMC DATA ..............................................................................................5-82
TABLE 5-25 MMC LOG .......................................................................................................................5-82
TABLE 6-1 SIWAREX FTC COMMAND LIST .....................................................................................6-91
TABLE 6-2 COMMAND GROUPS OF SIWAREX FTC ..........................................................................6-92
TABLE 7-1 LIST OF DATA AND OPERATING ERRORS ........................................................................7-100
TABLE 7-2 LIST OF TECHNOLOGY MESSAGES ..................................................................................7-102
TABLE 7-3 LIST OF OPERATING MESSAGES .....................................................................................7-103
TABLE 9-1 CFC – MESSAGE TYPES .................................................................................................9-116
TABLE 9-2 CFC – MESSAGE TEXT FROM SFT_FTC ........................................................................9-117
TABLE 9-3 CFC – SFT_FTC CONNECTION WITHOUT DATA RECORDS .............................................9-119
TABLE 9-4 CFC – SFT_FTC CONNECTIONS – DR3 INPUTS .............................................................9-120
TABLE 9-5 CFC – SFT_FTC CONNECTIONS – DR3 OUTPUTS .........................................................9-122
TABLE 9-6 CFC – SFT_FTC CONNECTIONS – DR4 INPUTS .............................................................9-122
TABLE 9-7 CFC – SFT_FTC CONNECTIONS – DR4 OUTPUTS .........................................................9-123
TABLE 9-8 CFC – SFT_FTC CONNECTIONS – DR5 INPUTS .............................................................9-124
TABLE 9-9 CFC – SFT_FTC CONNECTIONS – DR5 OUTPUTS .........................................................9-125
TABLE 9-10 CFC – SFT_FTC CONNECTIONS – DR7 INPUTS .............................................................9-127
TABLE 9-11 CFC – SFT_FTC CONNECTIONS – DR7 OUTPUTS .........................................................9-129
1 Preface
All of the information required to construct and operation the SIWAREX FTC is
found in this manual.
Note
This manual contains the description of all modules that are valid at the time of
publication.
We reserve the right to deliver production information along with new modules or
modules with a newer product status that contains the current information on the
module.
The layout of this manual is based on activities that must be performed in the
scope of project planning, commissioning, operation and service / maintenance.
Do you have more questions about using the SIWAREX FTC? Then please contact
your Siemens representative in the office or business location that is responsible
for your area or technical support for SIWAREX – Tel.: +49 (0)721 595 2811.
2 Scope of Delivery
A bus connector for the SIMATIC bus, the conformity details from the manufacturer
and a sheet of additional product information belong to the SIWAREX FTC scope
of delivery.
For planning your work with the SIWAREX FTC, you will need:
- The SIWAREX FTC project planning package for SIMATIC S7
or
- The SIWAREX FTC project planning package for SIMATIC PCS7
These are not components included in the scope of delivery and must be ordered
separately.
The corresponding project planning package is combined of the following
components:
- SIWATOOL FTC commissioning program for Windows
- Set-up for the installation of the module in the SIMATIC Manager hardware
catalogue
- Standard software for operating the SIWAREX FTC in SIMATIC S7
- Manual in several languages
- Set-up for PCS7 Library (Project planning package for PCS7 only)
- SIWAREX FTC OCX – AddOn for ProTool for project planning with the
calibration display (TP/OP/MP 170B/270/370) for non automatic weighing
applications only
A SIWAREX FTC "Getting started” software can be very helpful for the first
programming steps. This software can be downloaded free-of-charge over the
Internet (www.siwarex.com).
3 Product Overview
3.2 Benefits
SIWAREX FTC is the optimal solution wherever weighing technology requires high
speed and precision. Because of the high resolution (3 x 6000 d, calibratable),
scales can be built to work precisely over broad areas. With the SIWAREX FTC
weighing systems, force measurement systems and belt scales can be
constructed.
3.4 Structure
SIWAREX FTC is a function module (FM) of the SIMATIC S7-300 and can be read
directly on the SIMATIC S7-300- or ET 200M bus board. Installation / cabling
efforts for the 80 mm wide module are simplified with the profile rail assembly
(snap-in technology).
Connecting load cells, power supply and the serial interface is all done through the
40 pin standard front connector.
Operation of the SIWAREX FTC in SIMATIC guarantees complete integration of
weighing technology in the automation system.
3.5 Function
The primary task of the SIWAREX FTC consists of the precise measurement of the
current weight values in up to three measurement ranges and the exact
measurement of the amount of the uploading material.
There are different weighing procedures for which SIWAREX FTC can be
configured optimally by defining the corresponding parameters.
The following operating modes can be defined:
- Non Automatic Weighing Instrument – conforming with OIML R-76
- Force Measurement
- Belt scale
During the weighing procedure, SIWAREX FTC monitors and controls a multitude
of signals. The optimised system internal data exchange enables a direct
evaluation of weighing signals and states in the PLC program.
The weighing procedure influence on the PLC enables a flexible adjustment to suit
the changes in the system technology.
SIWAREX FTC is already adjusted in-house. Therefore, the scale can be adjusted
to theoretical settings without using any adjustment weights and modules can be
exchanged without readjusting the scale. Exchanging modules during running
operation is also possible when working with "active bus modules”.
The SIWAREX FTC has two serial interfaces. An RS 485 interface is used for
connecting digital remote display (only for the application as non automatic
weighing instrument). A PC can be connected to the RS 232 interface for setting
SIWAREX FTC parameters.
The weighing module SIWAREX FTC can also be used in explosion hazard areas
(zones 21 and 22). Load cells are supplied with inherent safety with zone 1
applications using an optional Ex-interface SIWAREX IS.
Not only does the SIWATOOL FTC support the user for program entry. Analysing
the diagnostics buffer, the contents of which can be saved together with the
parameters after reading from the module is also very helpful.
A trace mode exists in the SIWAREX FTC module for optimising weighing
progress. The recorded data can be displayed in a curve diagram using the
MS Excel.
The trace information can be used for analyzing the work of a plant during a
defined period of time.
Another feature of the SIWATOOL FTC program helps loading a new firmware
version for SIWAREX FTC on-site. It allows you to perform firmware upgrades at
any time and from anywhere.
The weighing logs are stored on an MMC (Micro Memory Card) that is inserted into
the SIWAREX FTC for the period of time defined in the weights and measures act.
If a person disagrees with the results of a certain weighing procedure, the weighing
data for that weighing procedure can be reconstructed from the MMC memory.
!
Warning Notes
This chapter contains information required for hardware planning, assembly and
preparations for operation.
The technical safety information is to be strictly adhered to.
!
Warning
Unqualified intervention in the device/system or not adhering to the warning notices
can result in serious injury or damage to equipment. Only qualified personnel are
permitted access to the operational components of this device / system.
!
Warning
The unit has been developed, manufactured, tested and documented according to
the corresponding safety standards. The device itself will not cause any danger to
equipment or personal health under normal circumstances.
!
Danger
Installation and start-up operation is not permitted until it is guaranteed that the
machine in which these components are to be integrated corresponds with the
guidelines 89/392/EWG.
Selecting the suitable SIMATIC CPU, the SIMATIC HMI (Human Machine
Interface) and communication modules does not only depend on SIWAREX FTC
requirements but also on the overall job that the automation system has to perform.
4.2.2 Introduction
Although SIWAREX FTC was developed for use in industrial environments and
meets high EMC specifications, you should do some EMC planning before
Capacitive coupling
Inductive coupling
Radiation coupling
If you follow these five basic rules, EMC can be guaranteed in most cases!
Separate the cabling into groups (high-voltage lines, power supply lines, signal
lines, ground wiring, data lines, etc.).
Run the high-voltage lines and ground wiring or data cables in separate channels
or bundles.
Run measurement lines as close to grounding surfaces as possible (e.g. support
beams, metal rails, cabinet panels).
Create a uniform reference potential and ground all electrical operational elements.
Run sufficiently dimensioned potential equalizing cabling if potential differences
exist or are to be expected between system components in the system. A potential
equalisation is mandatory for Ex applications.
When assembling the SIMATIC components and the SIWAREX FTC, the assembly
regulations (AR) for the SIMATIC S7 must be fulfilled completely.
SIWAREX FT is assembled in the following steps.
1. Check for whether the SIMATIC bus connector is connected to the left of the
SIWAREX FTC in the module group.
2. Connect the SIMATIC bus connector for the following module group in the
SIWAREX if necessary.
3. Install the shielding strip under SIWAREX.
4. Hang the SIWAREX FTC in its place.
5. Fasten the SIWAREX FTC with 2 screws in the lower area of the module.
6. Label the SIWAREX FTC corresponding to your identification system.
Special attention must be given to the shield strip for shielded lines. The
interference resistance of the system can only be guaranteed if the properly
constructed.
A cable is shielded to decrease the affects of magnetic, electrical and
electromagnetic interference on this line. Interference on cable shielding is routed
to ground through shielding rails that are conductively connected with the housing.
To ensure that this interference stream does not become a source of interference,
a low impedance connection to ground is especially important.
Use only lines with mesh-shielding. Shielding should provide at least 80%
coverage.
For fastening the meshed shielding, use only metal cable clamps. The clamps
must cover as much shielding as possible and ensure a good contact.
Shield clamps must be ordered separate to the grounding elements. The area
covered by the shielding clamp is to be chosen corresponding with the cable
diameter.
The shielding should be pressed against the shielding rail by the shielding clamp.
Note
In central operation in the S7 300 system, the SIMATIC CPU and SIWAREX FTC
is to be connected to the same 24 V power supply.
1M 24 V DC 1M Power supply M
The SIMATIC construction guidelines apply for connecting the 40 pin connector.
Flexible cables with a cross-section of 0.25 to 1.5 mm2 can be used. Remove the
insulation from the cable for 6 mm and install wire end sleeves.
Sensors equipped with strain gauges and that meet the following conditions can be
connected to the SIWAREX FTC:
- Characteristic value 1.... 4 mV/V
- Supply voltage of 10.2 V is permitted
The connection is made to the 40 pin front connector. The connection should be
made using the cable described in chapter Accessories.
2. The cable shielding is normally run on the cable guide supports of the junction
box.
In case of the danger of potential equalisation currents on the cable shielding, a
potential equalisation conductor is to be run parallel to the load cell cable or the
shield clamp is to be used in the junction box to the shield strip. Using the potential
equalisation conductor is the preferred method of dealing with EMC
(Electromagnetic Compatibility).
4. The shielding must be attached to the shielding strip on the SIWAREX FTC.
The two images below show load cell connection using 4-wire and 6-wire systems.
SIWAREX FTC has 7 digital inputs. These inputs are potentially isolated from the
module group. They have a common reference point (M).
The functionality of each input can be defined with parameters.
DI status is indicated with LEDs on the front of the SIWAREX FTC.
The counter input serves as the connection to the pulse encoder which is required
during weighing start-up for continuous operation. Twisted pair wiring should be
used for the connections.
The SIWAREX FTC has eight potential isolated digital outputs (DO) with a nominal
voltage of +24 V and an output current of maximum 0.5 A per output (max. total
current 2 A).
These outputs are potentially isolated from the module group. The outputs share
common potential. They have a common ground and a secure 24 V supply. They
are short-circuit and overload protected.
DO status is indicated with LEDs on the front of the SIWAREX FTC.
When connecting inductive consumers, the digital output that is used must be
equipped with a free-wheeling diode.
One or more Siebert Type S11 displays can be connected to the RS485 interface.
When connecting a single display, ensure that the RTa and RTb bridges are wired.
When installing the S11 for operation, the same parameters that have been set by
the SIWATOOL FTC for RS 485 are to be defined.
9 pin D-
X2 RS 232 Interface
type
A SIMATIC Micro Memory Card can be used for saving the calibratable records or
for recording weighing states. Approximately 16 MB of data can be stored on the
MMC described in chapter Accessories.
!
Warning notice
After formatting the MMC with SIWAREX FTC, it can no longer be used in the
SIMATIC CPU. Therefore, the MMC should be labelled correspondingly.
Introduction At this point in the commissioning procedure, after assembling the module
group and making all connections, you should perform a partial functionality
test for the SIWAREX FTC and the connected components.
The individual steps for partial testing are to be performed in the following
sequence:
Visual check Check all previously performed steps for proper execution, i.e.:
• Is there any external damage to the module group?
• Are the modules situated in the proper positions?
• Are all fastening screws properly tightened?
• Are all connecting cables properly connected and fastened tightly?
• Is the frontal connection made properly?
• Is the shielding properly attached to the shielding conductor for all
corresponding cables?
• Is the profile rail connected to the ground conductor?
• Are all tools, materials or components that do not belong to the S7 or
SIWAREX FTC removed from the profile rail and
the module group?
LED check on the After attaching the 24V supply voltage and a short initialisation phase
SIWAREX FTC (internal testing is indicated by LED running pattern), the SIWAREX FTC
goes into operational status. The following LEDs must have the status that is
indicated below if the unit is running correctly:
Danger
! means that serious material damage, severe injury or even death will result if the
corresponding safety precautions are not followed carefully.
For use of the module for or in potentially explosive areas the certificate
conditions and regulations must be met.
DANGER EXPLOSION
Exists when the installation and setup regulations are not adhered to.
Warning
! means that serious material damage, severe injury or even death can result if the
corresponding safety precautions are not followed carefully.
Warning – Explosion Hazard – Do not disconnect while circuit is live unless area
is known to be non-hazardous.
5 Weighing Functions
The selected operating mode and the defined parameters are very important to the
behaviour of the SIWAREX FTC in the process.
The parameters are set with default values from the manufacturer. Using the "Load
default values" command, the parameter definitions defined in manufacturing are
loaded.
The default parameters are defined so that the scale is immediately ready for each
operating mode. Not all parameters must be re-entered for each operating mode.
By changing a parameter, the behaviour of the scale is changed respectively. The
advantage of this solution is that you can define how many default values should
be retained and how much that the behaviour of the scale has to be changed for
the application.
All parameters are split into data records (DR). The data records are organised
according to process steps (tasks) that you have to perform during the
commissioning phase or during the process itself.
In the following parameter description, you will find a description of the weighing
functions that are influenced by that parameter.
The parameters of a data record are shown in a table to start with. This is then
followed by the exact parameter description for the parameters of that data record.
After receiving a new parameter, SIWAREX FTC runs a plausibility check. If there
is a parameter error then the data record will not be accepted (stored) by the
SIWAREX FTC and a "synchronous" message is generated (see chapter 7
Messages and Diagnostics).
The adjustment parameter must be tested for every scale and changed if
necessary.
The scale is basically defined using the adjustment parameter and by performing
the adjustment. In verifiable operation as a non-automatic weighing machine, the
data of the DR3 cannot be changed after calibration.
Procedure:
o Check all parameters and change if required.
o Send the DR3 to the scale
o Perform scale adjustment
o Receive DR3 from the scale
Numeral step for REAL DBD090 0 Numeral step WB3 (1*10 , 2*10 , 5*10 , k ∈ N0 )
k k k
5.2.21
weighing range 3
Max. positive weight for BYTE DBB107 10 Pos. range for switch-on zero setting device [in 5.2.26
zero setting when % of WBmax]
switching on (Entry from Pos- + Neg. Zero setting value may
not exceed 20% for country code "OIML")
Maximum negative BYTE DBB108 1 Neg. Range of zero setting device [in % of 5.2.27
weight for zero setting WBmax]
(Entry from Pos- + Neg. Zero setting value may
not exceed 4% for country code "OIML")
Maximum positive weight BYTE DBB109 3 Neg. Range of zero setting device [in % of 5.2.28
for zero setting WBmax]
(Entry from Pos- + Neg. Zero setting value may
not exceed 4% for country code "OIML")
Tare max. weight T- BYTE DBB110 100 Subtractive tare device range 5.2.29
[in % of WBmax with multi-range scale]
[in % of WB1 with multi-resolution scale]
(Entry may not exceed 100% with country code
"OIML")
Scale type BYTE DBB111 0 Scale type: 5.2.30
0 NAWI-filling procedure
1 NAWI-Emptying procedure
2 Force measuring
3 Belt scale
5.2.1 DR3 – Adjustment digits 0… 4, for the zero point and adjustment weights 1… 4
The analog measurement value coming from the load cells is converted into a
digital value with an digital-digital converter. A weight value is then calculated from
this digital value. All functions of the SIWAREX FTC then uses this weight value for
executing their task.
To calculate the weight value from the digital value, the characteristic curve of the
measurement system must be determined. In the simplest case, the characteristic
curve is defined by points 0 and 1. The first operating point (point 0) is always
determined with the empty scale alone with its own construction weight. The weight
of its own construction causes the load cells to deliver a measurement voltage to
SIWAREX FTC. After the digital-digital conversion of the measurement voltage, the
digital value (adjustment digit 0) is assigned the zero weight. For determining the
average value of the adjustment point (zero point, adjustment weight) over a
predefined period of time, the determination time must be defined. On a belt scale,
the duration of a belt cycle is set normally.
If the scale is loaded with a known calibration weight (e.g. with 50 % of the
measurement range), then the second operation point can be determined. The new
digital value from the analog-digital converter is now assigned the calibration
weight.
The characteristic curve can be further adjusted using a maximum of three other
points that must lie above point 1.
Ensure that the difference between two adjustment weights is at least 5% of the
measurement range.
The adjustment sequence for the increasing adjustment weights must be retained.
Example:
The illustration shows the relationship between the adjustment digits and the
adjustment weight.
If the adjustment weights and adjustment digits of a SIWAREX FTC are known
then the adjustment procedure does not need to be performed. They are simply
sent to the SIWAREX FTC in the DR3 data record and the scale is immediately
operational (after official calibration of the scale, it is no longer possible to send
DR3).
The SIWATOOL FTC program supports you in quickly performing and adjustment.
Option 1:
After commissioning and after adjustment, all data records for the scale are read
from SIWAREX FTC and are stored as a scale file ScaleX.FTC.
Identical scales can now be put into operation immediately. Connect the PC to the
new scale and activate the "Send all data records" function. This also transfers the
adjustment weights and the adjustment digits – the characteristic curve is defined
immediately. Of course, the same applies for exchanging a SIWAREX FTC.
Option 2:
Use the SIWATOOL FTC "Theoretical adjustment" function and determine the
characteristic curve of the scale from the technical data of the load cells alone. This
case assumes proper construction of the scale.
Note
Normally, defining two operating points is sufficient to determine the characteristic
curve for the scale. Other operation points must only be defined on non-linear
systems – New digital values (adjustment digits 2, 3, 4) are assigned with other
calibration weights (e.g. 70%, 80%, 100% of the measurement range).
Depending on the characteristic value of the connected load cells, the value 1
mV/V, 2 mV/V or 4 mV/V must be selected. Since the SIWAREX FTC supplies the
load cells with approximately 10 V, the measuring input is redefined according to
the expected measurement voltage (max. 10 mV, max. 20 mV or max. 40 mV).
Changing the filter sequence can be an advantage in some cases. Normally, the
signal is first filtered by the average value filter.
Defining the limit frequency is critically important for suppressing interference. The
"speed" of the scale's reaction to the change in the measured value is determined
by specifying the limit frequency.
A value of e.g. 2 Hz leads to a relatively quick scale reaction to a weight change, a
value of e.g. 0.5 Hz makes the scale "sluggish".
The average value filter is used to settle the weight value against interference. The
weight value is based on the average of the n weight values that the SIWAREX
FTC calculates every 10 msec, e.g. if n=10, 10 values are used for calculating the
average value. Every 10 msec, the oldest value is removed and the newest is
taken into account in the calculation.
SIWAREX FTC can be defined as a scale with one, two or three different
measurement ranges. Depending on the construction type certification, the
numeral step can be defined separately for every range.
If the number of ranges is greater than one then the scale type can be defined as a
multi-range scale or a multi-resolution scale.
Information on the function of a multi-range- or multi-resolution-scale can be found
in EN 45 501.
Note
If only one range is defined then this entry is meaningless.
The scale can be set to zero automatically by switching the supply voltage on. On
verifiable scales, a weight value of +/- 10% of the max. measurement range can be
set to zero on start-up.
!
Attention
In non-verifiable operation (no OIML restrictions), a full scale can also be set to
zero after activating this function. Limiting the effectiveness is possible by entering
the max. and min. weight however. See Max. Min. weight for zero setting upon
switch-on. This function is deactivated in the "Belt scale" operating mode.
The scale can be set to zero automatically by switching the supply voltage on. If
the above function is activated, there is still the question of whether zero setting at
start-up should be run if the tare is unequal to zero.
Zero setting the scale can be performed with the "Zero setting" instruction if
required.
The automatic zero adjustment sets the scale to zero without an instruction if it
slowly drifts away from the zero point. Slow drifting is presumed if the conditions of
standard EN45501 are met. This function in the "Belt scale" operating mode will be
excecuted as dynamic zero setting function.
!
Attention
In non-verifiable operation (no OIML restrictions), after activating this function, the
scale can indicate zero after a slow drift over time if it is full. When using this
function, overloading the scale must be prevented with suitable measures.
The weight value can be used for verifiable recording with the numeral step for
measurement range 1 only above the minimum weight. The minimum weight is
defined by an adjustment or a calibration. The minimum weight depends on the
number and type of load cells used.
The value can be set to 0 on non-calibrating scales.
The weight can be used for calibrating purposes with the numeral step for
measurement range 1 only under the maximum weight. The maximum weight is
defined during adjustment.
The maximum weight depends on the number and type of load cells used.
The numeral step for weighing range 1 can correspond with standard EN 45501
(0.0001 to 50)
The weight value can only be used above the minimum weight for verifiable
recording with the numeral step for measuring range 2. The minimum weight is
The weight can only be under the maximum weight for verifiable purposes with the
numeral step for measuring range 2. The maximum weight is defined during
adjustment.
The maximum weight depends on the number and type of load cells used.
Note
If only one range is defined then this entry is meaningless.
The numeral step for the weighing range 2 can correspond with standard EN
45501 (0.0001 to 50)
If only one range is defined then this entry is meaningless.
The weighing value can only be used above the minimum weight for verifiable
recording with the numeral step for measuring range 3. The minimum weight is
defined by an adjustment or a calibration. The minimum weight depends on the
number and type of load cells used.
The value can be set to 0 on non-calibrating scales.
Note
If only one or two ranges are defined then this entry is meaningless.
The weight can only be used under the maximum weight for verifiable purposes
with the numeral step for measuring range 3. The maximum weight is defined
during adjustment.
The maximum weight depends on the number and type of load cells used.
Note
If only one or two ranges are defined then this entry is meaningless.
The numeral step for weighing range 1 can correspond with standard EN 45501
(0.0001 to 50)
Note
If only one or two ranges are defined then this entry is meaningless.
To prevent having to cancel a weighing instruction (only taring, printing and zero
setting in static operation) immediately when the stable weight of the scale is not
defined, the waiting time is given as standstill. Only the scale instruction tare,
output log or zero setting was not possible after this time has elapsed is the
corresponding technology message generated.
5.2.25 DR3 – Maximum negative weight for zero setting upon switch-on
Zero setting means that an automatic zero setting is performed for the scale when
the supply voltage is switched on.
If zero setting upon switching on the supply voltage has been activated, then the
definition will limit the affect of the function. The reference point for the
effectiveness of the limitation is not the current weight, it is the weight that was
displayed by the scale without a previous zero setting. This function has no effect
in "Belt scale" operating mode.
5.2.26 DR3 – Maximum positive weight for zero setting upon switch-on
Zero setting means that an automatic zero setting is performed for the scale when
the supply voltage is switched on.
If the zero setting when switching on the supply voltage has been activated then
the definition will limit the affect of the function. The reference point for the
effectiveness of the limitation is not the current weight, it is the weight that was
displayed by the scale without a previous zero setting. This function has no effect
in "Belt scale" operating mode.
The current gross weight of the scale is defined as zero for zero setting.
For zero setting, the specification can be used to limit the effect of the function. The
reference point for the affect of the limitation is not the current gross weight. It is
the weight that the scale would show without a previous zero setting.
For scales in calibrating operation, the limit between the negative and positive
weight for zero setting is 4% of the maximum weighing range.
For zero setting, the specification can be used to limit the effect of the function. The
reference point for the effectiveness of the limitation is not the current weight, it is
the weight that was displayed by the scale without a previous zero setting.
For scales in calibrating operation, the limit between the negative and positive
weight for zero setting is 4% of the maximum weighing range.
SIWAREX FTC accepts every weight value which is less than %-record of the
maximum weighing range with multi-range scales or weighing range 1 with a multi-
resolution scale.
On scales in verifiable application, the value is limited to 100% of the maximum
weighing range.
This parameter is used to select the suitable weighing instrument program. One of
these weighing operating modes applies depending on the task description.
This operation mode can not be used for legal trade. The functionality is the same
as in the operation mode “NAWI filling procedure“. The difference is only the plus
and minus polarity for the measuring signal.
In this operating mode SIWAREX FTC will use the parameters of DR5. The
Functionality supports belt scales.
Scales in calibrating operation are subject to certain restrictions. The "OIML" entry
activates these restrictions. To deactivate them, it is necessary to enter "----". The
entry "OIML" only makes sense if the scale is to be verifiable as a non-automatic
weighing instrument.
A 4 character string can be used as the unit of measurement e.g.: g, kg, lbs
The length unit is used for defining the belt length. Calculating the belt load refers
to this definition.
The conversion factor defines which conversion is to be used between the weight
unit and the large weight unit.
Example:
Weight unit: kg
Large weight unit: t
Conversion factor 1000,0
With a static scale (determination time = 0), the operation are executed
immediately after activating the instruction during the adjustment. Zero setting is
also executed immediately when defining a stable weight.
In continuously running scales, e.g. belt scales, a determination time can be
defined for this instruction. The scale runs the determination of the operation point
during the defined determination time as an average value of all measurements
and the average value is output after the determination time has elapsed.
The determination time applies for the following instructions: zero point valid,
adjustment weight 1...4 valid, zero setting, taring.
Example:
Determination time 10 000 ms.
Command: Zeroing
After the determination time is elapsed the average new “zero value” will be
calculated.
The basis parameters define other features of a scale, which are responsible for
the basic behavior. The parameters must be checked and sometimes have to be
changed. DR4 is not subject to write protection in calibrating operation.
Procedure:
o Check all parameters and change if necessary.
o Send the DR 4 to the scale
Limit values
Basis weight for limit BYTE DBB010 0 Bit 0 : 5.3.3
value 1 0 = Basis for limit value 1 is the gross weight
1 = Basis for limit value 1 is the net weight
Logs can be output to a connected printer or to the Micro Memory Card (MMC).
The contents of the log are defined in data records DR 40 to DR 43.
The limit values can be defined in relation to the gross weight or the net weight of
the scale. This way, a very coarse dosage (net reference) or a refilling of the scale
container (gross reference) can be defined for example.
The limit values can be defined in relation to the gross weight or the net weight of
the scale.
The limit value for the monitoring of the empty range can be defined in relation to
the gross weight or the net weight of the scale.
The value for the empty range is a limit value, that when exceeded indicates the
"empty" status to the SIWAREX FTC which gives this out as status information.
Switch-on and shut-off weights can be defined separately for a limit value. In this
way, a minimum value monitor and a maximum value monitor can be realized by
hysteresis. A maximum value monitor is realized with the following specification:
The switch-on and cut-off weight can be defined separately for the limit value. With
limit value 3, only the maximum weight monitor can be used. The basis weight for
the limit value 3 is always the gross weight.
When the limit value is undercut, the status information "limit value 3) is reset.
The basis parameter for belt scale defines other features of a scale which are
responsible for the behavior in the belt scale operating mode. The parameter must
be checked and changed in some cases.
Procedure:
o Check all parameters and change if necessary.
o Send the DR 5 to the scale
Nominal speed REAL DBD000 0 Nominal speed of the belt in LU/s (Length 5.4.1
units per second)
Time basis for belt TIME DBD004 0 Time basis for measuring the belt speed by 5.4.2
speed impulse counter input (msec)
Pulse constant REAL DBD008 0 Pulse constant of the speed sensor 5.4.3
Constant belt speed REAL DBD012 2000 Constant speed if no speed sensor is 5.4.4
connected.
0 – for connecting the speed sensor.
Minimum belt speed INT DBW016 800 Limit value for minimum belt speed in % of 5.4.5
the nominal speed.
800 corresponds with 80.0 %
Maximum belt speed INT DBW018 1020 Limit value for maximum belt speed in % of 5.4.6
the nominal speed.
1020 corresponds with 102.0 %
Reaction time for belt TIME DBD020 Reaction time for belt monitoring and belt 5.4.7
monitor cycle error message at belt start
At start
Reaction time for belt TIME DBD024 Reaction time for belt monitoring and belt 5.4.8
monitor cycle error message during belt operation
During operation
Reserve_5_28 REAL DBD028 Reserved
Belt load and flow-rate
Nominal flow-rate REAL DBB032 Nominal flow-rate in weight units / s 5.4.9
Effective belt length REAL DBB036 Effective belt length in length units LU 5.4.10
Correction factor for REAL DBB040 Correction factor for the flow rate 5.4.11
flow-rate
Minimum flow-rate INT DBD044 Limit value for minimum flow-rate in % of the 5.4.12
nominal flow-rate
Maximum flow-rate INT DBD046 Limit value for maximum flow-rate in % of the 5.4.13
nominal flow-rate
Minimum belt loading INT DBB048 Limit value for minimum belt loading in % of 5.4.14
the nominal belt loading
Maximum belt loading INT DBB050 Limit value for maximum belt loading in % of 5.4.15
the nominal belt loading
Dyn. zero setting range INT DBB052 Maximum belt loading with zero setting. 5.4.16
Minimum belt loading INT DBW054 Limit value for minimum belt loading in % of 5.4.17
for totalizing the nominal belt loading for totalizing
Delay time for TIME DBD056 Delay time after switching on the belt until 5.4.18
monitoring the belt monitoring the belt loading or flow-rate.
loading / flow-rate at
102
The nominal speed of the belt corresponds with the speed at nominal conditions,
i.e. at nominal load on the belt. This is output in LU (length units) per second e.g.
1.7 m/sec.
The belt speed is output in LU (length units) per second. The measuring time of the
pulse input can be set to another time e.g. 2 seconds. That means that the speed
value output is updated every 2 seconds, the value will still be defined in LU per
second however.
The pulse constant of the speed sensor indicates the number of pulses per LU
(length unit) of the belt. SIWAREX FTC calculates the current belt speed on this
basis.
Minimum belt speed is a limit value for minimum belt speed in % of the nominal
speed. The entry of the number e.g. 800 corresponds with 80.0 %. After this value
has been undercut, SIWAREX FTC sets a status bit (see scale status...).
Maximum belt speed is a limit value for maximum belt speed in % of the nominal
speed. The entry of the number e.g. 1020 corresponds with 102.0 %. After this
value has been exceeded, SIWAREX FTC sets a status bit (see scale status...).
Monitoring the belt speed is only activated after this delay time when the belt is
started.
When the minimum belt speed is undercut, the message or status bit is only
activated after this delay time has elapsed.
Effective belt length in length units LU. The effective belt length corresponds with
half of the distance between the belt rollers which are found before and after the
roller with the belt scale.
A material test can be performed to compare the actual required material quantity
with the total calculated by SIWAREX FTC. A correction factor can be defined for
the calculation of the flow-rate quantity for small deviations.
Minimum flow-rate is a limit value for minimum flow-rates in % of the nominal flow-
rate. The entry of the number e.g. 800 corresponds with 80.0 %. After this value
has been undercut, SIWAREX FTC sets a status bit.
Maximum flow-rate is a limit value for maximum flow-rates in % of the nominal flow-
rate. The entry of the number e.g. 1020 corresponds with 102.0 %. After this value
has been exceeded, SIWAREX FTC sets a status bit.
Minimum belt loading is a limit value for minimum belt loading in % of the nominal
belt loading. The entry of the number e.g. 800 corresponds with 80.0 %. After this
value has been undercut, SIWAREX FTC sets a status bit. In batch mode the
totalizing will be completed after the dosing signal is switched off and the belt
loadin is less than this limit value.
Maximum belt loading is a limit value for maximum belt loading in % of the nominal
belt loading. The entry of the number e.g. 1020 corresponds with 102.0 %. After
this value has been exceeded, SIWAREX FTC sets a status bit.
Maximum belt loading for zero setting is a limit value for maximum belt loading in %
of the nominal belt loading. Entering the number e.g. 20 corresponds with 2.0 %.
After this value has been exceeded, SIWAREX FTC will not execute the zero
setting instruction.
Minimum belt loading for totalizing is a limit value for minimum belt loading in % of
the nominal belt loading. Entering the number e.g. 20 corresponds with 2.0 %.
After the actual value is less than the limit, SIWAREX FTC will not execute the
totalizing function.
5.4.18 DR 5 – Delay time for the monitoring of the belt loading after the start
After the belt is switched on, undercutting the minimum belt loading or the
minimum flow-rate is only indicated after this time has elapsed.
5.4.19 DR 5 – Delay time for monitoring the belt loading in continuous operation
After the minimum belt load has been undercut or the minimum flow-rate is
undercut, the respective message or status bit is output only after this time has
elapsed.
The totalizing memory 1...4 is run with the weight unit indicated in data record 3.
The user can define the smallest step for creating a total with the numeral step.
E.g. weight unit 1 kg, numeral step for creating the total can be 5 kg.
Totalizing memory 4...8 is run with large weight unit indicated in data record 3. The
user can define the smallest step for creating a total with the numeral step.
E.g. Weight unit 1 kg, numeral step for totalizing can be 1 t.
The material quantity per pulse 1 indicates, after which quantity increase of
totalizing memory 1.....4 should output a pulse through the defined digital output.
For the external quantity counter, the pulse duration can be defined so that the
device can recognize the pulse for sure. The pulse duration is not permitted to be
selected so long that stopping the minimum pause duration will make the next
pulse late.
For the external quantity counter, the pulse duration can be defined so that the
device can recognize the pulse for sure. The pulse duration is not permitted to be
selected so long that stopping the minimum pause duration will make the next
pulse late.
The material quantity per pulse 2 indicates, after which quantity increase of
totalizing memory 1.....4 1 should output a pulse through the defined digital output.
For the external quantity counter, the pulse duration can be defined so that the
device can recognize the pulse for sure. The pulse duration is not permitted to be
selected so long that stopping the minimum pause duration will make the next
pulse late.
For the external quantity counter, the pulse duration can be defined so that the
device can recognize the pulse for sure. The pulse duration is not permitted to be
selected so long that stopping the minimum pause duration will make the next
pulse late.
5.5 DR 7 Interfaces
The parameters that define the behavior of the SIWAREX FT on the interfaces are
defined in DR7. If an interface is not used the default values can be left alone.
o Change parameter if required.
o Send the DR 7 to the scale
Process value 2 for fast BYTE DBB006 30 Various process values are available for selection 5.5.5
output to the SIMATIC after listing.
CPU
S7-Alarm
Definition of the process WORD DBW010 0 Value 0 – No process alarm generation 5.5.6
alarm 1
Value range 1 – 0FFh
Number of the technical error
Definition of the process WORD DBW012 0 Value 0 – No process alarm generation 5.5.6
alarm 2
Value range 1 – 0FFh
Number of the technical error
Definition of the process WORD DBW014 0 Value 0 – No process alarm generation 5.5.6
alarm 3
Value range 1 – 0FFh
Number of the technical error
Definition of the process WORD DBW016 0 Value 0 – No process alarm generation 5.5.6
alarm 4
Value range 1 – 0FFh
Number of the technical error
Definition of the process WORD DBW018 0 Value 0 – No process alarm generation 5.5.6
alarm 5
Definition of the process WORD DBW020 0 Value 0 – No process alarm generation 5.5.6
alarm 6
Value range 1 – 0FFh
Number of the technical error
Definition of the process WORD DBW022 0 Value 0 – No process alarm generation 5.5.6
alarm 7
Value range 1 – 0FFh
Number of the technical error
S7 – FB Life bit TIME DBD024 0 0 = S7- Life bit monitor switched off 5.5.7
monitoring time 1 to n = Monitoring time [msec]
Analog output 0
Weight for zero point REAL DBD028 0 Value output with the 0 or 4 mA 5.5.8
(0 or 4 mA)
Reference for end value REAL DBD032 0 Value output with the 20 mA. 5.5.9
(20 mA)
Replacement value for REAL DBD036 0 Replacement value which is output with an active 5.5.10
the analog output at OD OD signal if the parameters are defined
respectively
Source for the analog BYTE DBB040 0 0 = Control signals-SIMATIC 5.5.11
output 1 = ext. default value through DR 17
2 = Gross
3 = Net
Coarse/fine default values
Other definitions not permitted.
Current range for the BYTE DBB041 0 Bit 0: 5.5.12
analog output 0: 0 .. 20mA
1: 4 .. 20mA
Bit 1 – 7 unused
RS232 0
RS232-printer baud rate BYTE DBB042 3 0 = 1200 Bit/sec. 5.5.13
1 = 2400 Bit/sec.
2 = 4800 Bit/s
3 = 9600 Bit/s
Bit 3 – 7 unused
RS485 0
Log selection BYTE DBB044 0 0: No device 5.5.15
1: SIEBERT- Display S11
Decimal place for the BYTE DBB045 0 0 to 4 5.5.16
remote display
Other definitions not permitted.
RS485-Baudrate BYTE DBB046 3 0 = 1200 Bit/sec. 5.5.17
1 = 2400 Bit/sec.
2 = 4800 Bit/s
3 = 9600 Bit/s
4 = 19200 Bit/s
Other definitions not permitted.
RS485-character parity BYTE DBB047 0 Bit 0: 5.5.18
0: even
1: odd
RS485-Stop-Bits 0 Bit 2:
0: 1 stop bit
1: 2 stop bit
Digital inputs 0
Definition of digital input 1 BYTE DBB060 0 0 = No command 5.5.23
1- 254 = Command code
Level definition for digital BYTE DBB067 0 Bit 0: Digital input 1 -> 0 = high active: 1 = low 5.5.24
inputs 1...7 active
....
Bit 6: Digital input 7 -> 0 = high active: 1 = low
active
Bit 7: Not used
Reserved TIME DBD068 0 Reserved
Memory segment for BYTE DBB077 50 Up to 100% but the total for trace function and log 5.5.26
trace function may not exceed 100%
Memory segment for logs BYTE DBB078 50 Up to 100% but the total for trace function and log 5.5.27
may not exceed 100%
80
Instead of the actual weight determination, a weight simulation can be activated for
test purposes. The simulated weight value can be defined through the SIMATIC
interface or through the RS232 interface (SIWATOOL FTC). The weight simulation
makes commissioning a scale much easier in certain situations.
5.5.2 DR 7 – Decade for rounding the decimal places for the process values
This parameter can be used to define to how many decimal places that the
(weight) process values should be rounded. This entry decouples the display of
weight values which trigger the restrictions of the calibration requirements and the
values that are used in the control software.
After activating this function, the digital outputs can be forced independent of their
parameter definitions. Force control is only possible in service operation and only
with the SIMATIC interface.
!
Warning notice
The current process values can be read as data packages (data records in the
SIMATIC CPU using function block FB43. Transferring a large data record can last
for several CPU cycles and puts a load on system performance.
If a process value should be transferred very quickly to the SIMATIC CPU
immediately after its origination in the SIWAREX FTC, the peripheral interface of
the SIWAREX FTC exists. The data is read by the FB43 cyclically and provided for
the user as an output variable. In the belt scale operating mode, normally the
current belt load (selection code 45) is of some importance.
The selection is made from the following list Table 5-20 Process values for
allocating log fields.
Process value 2 can be defined just the same as process value 1 (see above).
Normally the current Conti status (selection code 30) is of interest in the belt scale
operating mode.
The process alarms can be defined with this parameter. The module group can
generate process alarms in the SIMATIC CPU. The control program can react
immediately to an event this way. Defining a process alarm is explained with the
example of process alarm 0. The explanation applies for process alarms 1 to 7 as
well.
If parameter value 0 is defined for process alarm 0 then no process alarms are
generated.
By entering a value unequal to zero, a process alarm is allocated according to the
following procedures:
Value range 1 to 255
Number of the technology message. If the defined technology error occurs then
process alarm 0 is generated.
Value range 256 to 319
Number of the bit from the NAWI- or Conti status (see DR 30). If the defined status
occurs then process alarm 0 is generated.
256+ 0... 31 Bit-No. NAWI status bit coming
288+ 0... 31 Bit-No. Conti status bit coming
Value range 512 to 575
Number of the bit from the NAWI- or Conti status (see DR 30). If the defined status
goes then process alarm 0 is generated.
A weight for the range start can be defined when defining the range parameters for
the analog output.
When defining the parameters for the range for the analog output, a weight can be
defined for the range end.
A weight can be defined for the analog output, for which a corresponding signal is
output if the OD signal (Output Disable) exists.
Normally this corresponds with the SIMATIC CPU status – STOP.
The source for the analog output can be defined with this parameter. The weight
values from the defined source are output as current signals corresponding with
the parameter definitions for start and end ranges.
The following sources are available:
The analog output can be operated with current range 0 ... 20 mA or 4 ... 20 mA.
If a printer is connected to the RS232 interface, the following baud rates can be
defined.
1200 Bit/sec
2400 Bit/sec
4800 Bit/sec
9600 bps
The transfer control parameters for the RS232 interfaces can be defined as
follows:
If a display from the Siebert company is connected to the RS 485 interface, the
decimal place for the display of non-verifiable process values can be defined.
5.5.17 DR 7 – RS 485-Baudrate
0 = 1200 Bit/sec.
1 = 2400 Bit/sec.
2 = 4800 Bit/s
3 = 9600 Bit/s
4 = 19200 Bit/s
Parity:
0: even
1: odd
The definition of the digital outputs can be defined with this parameter. Assigning a
bit from the pool of status bits for the scale causes the output to be active (active
can be signal level 0 or 1 – see below) if the bit is set.
When assigning a number between 0 and 63, the digital output is assigned to a
status bit of the scale and when assigning a number > 63, the output remains
inactive.
0... 31 Bit-No. NAWI status flag
31 + 0... 31 Bit-No. Conti status flag
40h....Pulse output with steps, like totalizing memory 1
41h... Pulse output with steps, like totalizing memory 5
FFh Output always active
Value range 42h – FEh not allowed
After assigning the definitions for digital outputs to a status bit, the signal that the
output should have when it is active can be determined.
E.g. Digital output 1 (DO1) is determined with bit 0:
Normally, the outputs are reset with a STOP of the SIMATIC CPU. This behavior
corresponds with the default settings.
If it makes sense to set an output with a SIMATIC CPU STOP, you can define it
with this parameter.
E.g. Digital output 1 (DO1) is determined with bit 0:
If bit 0 is 0 then DO1 is also 0 with OD signal.
If bit 0 is 1 then DO1 is 1 with OD signal.
!
Warning notice
If an output is to be set with the SIMATIC CPU STOP then you must ensure that it
does not lead to a hazardous situation.
Normally, the outputs are reset with a module fault (operating error). This behavior
corresponds with the default settings.
If it makes sense to set an output with a fault, you can define it with this parameter.
E.g. Digital output 1 (DO1) is determined with bit 0:
If bit 0 is 0 then DO1 is also 0 with fault.
If bit 0 is 1 then DO1 is 1 with fault.
etc.
Warning
If an output is to be set with the fault (operating error) then you must ensure that it
does not lead to a hazardous situation.
The definition of the digital inputs can be defined with this parameter. This can
happen by allocating an instruction.
Commands:
If the signal 1 appears at the defined input then the assigned command is
executed. To allocate an instruction to an input, its instruction number (1 to 255)
must be entered (see Command list). By entering 0, the input is assigned as
unused.
After assigning the definition for digital inputs to a command or a step-on condition,
you can define which signal that the should be interpreted as active on the input.
E.g. Digital input 1 (DI1) is determined with bit 0:
If bit 0 is 0 then DI1 is high active and if bit 0 is 1 then DI1 is low active,
If bit 1 is 0 then DI2 is high active and if bit 1 is 1 then DI2 is low active,
etc.
5.5.25 DR 7 – MMC log overflow, MMC Trace overflow, Target memory for trace function
Bit 0 is used for defining how the logging should be performed in the verifiable
MMC memory when the memory is full.
0: with full memory, the logging procedure is stopped
1: with full memory, the oldest entries are overwritten
Bit 1 is used for defining how the logging of trace data should continue if memory is
full.
0: No overwriting of trace data is possible with full MMC memory
1: The oldest trace data is overwritten with full MMC memory
The description of the trace function is found in 5.5.28
Bit 2 defines which memory is used for recording the trace data.
0: Trace data is stored in RAM
The MMC memory can be used for recording the trace data and for recording the
verifiable scale logs.
The percentage of MMC memory that is available for the trace function is defined
with this parameter.
Up to 100% is permitted in this definition but the total for trace function and
verifiable log may not exceed 100%
The description of the trace function is found in 5.5.28
The MMC memory can be used for recording the trace data and for recording the
verifiable scale logs.
The percentage of MMC memory that is available for the verifiable logs is defined
with this parameter.
Up to 100% is permitted in this definition but the total for trace function and
verifiable log may not exceed 100%
The description of the trace function is found in 5.5.28
SIWAREX FTC has its own hardware clock. The actual date and time can be read
through DR8. The date and time can be set on the module group using this data
record if required.
Procedure:
o Setting the date and time.
o Sending the DR 8 to the scale
No entries can be made in DR9. The data record uses the information on the inner
life of the module.
The information on the SIWAREX FTC module is for the purpose of identifying the
module in the manufacturer's plant (e.g. for repairs). This information has no
influence on user operation.
DR 14 is used for an external definition of the tilt angle for the belt scale.
Procedure:
o Enter tilt angle
o Send DR 14 to the scale
DR 14 is used for an external definition of the belt tilt angle. If the belt is mounted
horizontally, the definition is 0. The belt can be tilted to a maximum of 30°.
Always ensure that the definition corresponds with the current tilt angle, otherwise
the belt load will be calculated incorrectly.
DR 15 is used for external specification of the tare weight. Once specified with DR
15, the tare value is not yet activated. The transfer to tare memory of the
SIWAREX FTC is then performed with instruction "Accept tare entry" (see
command code 24)
If the DR 16 was defined as the source for the weight simulation (see DR 7 Source
for the weight simulation), defining a weight value through DR 16 will deactivate the
measurement input of the SIWAREX FTC and the defined value is "simulated" as
the weight value.
Procedure:
o Check in DR 7, whether DR 16 has been defined as the source for the weight
simulation
o Enter a value to be simulated
o Send DR 16 to the scale
If the DR 16 has been defined as the source for the weight simulation (see DR 7
Source for the weight simulation), defining a weight value through DR 16
deactivates the measurement input of the SIWAREX FTC and the defined value is
"simulated" as the weight value. This way, plant components (including filter
settings) can be tested even though the scale is not set up yet.
If DR17 has been defined as the source for the analog output (see DR 7 Source for
the analog output) then sending a weight value will output a corresponding output
stream on the analog output.
Procedure:
o Check in DR 7, whether DR 17 was defined as the source for the analog output.
o Check the parameter definitions of the analog output (see DR 7 Weight for zero
point and following parameters).
o Enter a value in DR 17
o Send DR 17 to the scale
If the DR 17 has been defined as the source for controlling the analog output (see
DR 7 – Source for the analog output) then the analog output is controlled with the
defined value with a definition of a weight value on DR 17. This way, the analog
output can be set continuously from the SIMATIC control program.
A value can be defined through DR18 which is to be shown on the remote display.
This way, the SIEBERT remote display can be used for displaying values that are
calculated in the SIMATIC.
Procedure:
o Enter a value in DR 18
o Send DR 18 to the scale
In batch operation, the entire quantity of the material to be loaded is loaded. The
scale runs the loading according to the predefined total quantity. Procedures:
o Define set weight
o Send DR 21 to the scale
18
The total quantity for loading operation can be set using this parameter. The belt is
stopped when the total quantity is achieved in totalizing memory 1 and the
totalizing is ended. For loading the predefined quantity again, totalizing memory 1
must be deleted.
The maximum loading time is started as well when loading is started. If the total
quantity is not achieved before this time has elapsed, SIWAREX FTC sends out a
technology message.
The trailing weight is the material quantity that must be weighed in addition after
the set value has been reached..
After the total quantity has been reached, a log can be created automatically. One
of four log text messages can be defined with this parameter.
The process values and the expanded process values can be used to monitor the
current states and data in the scale.
Observing the selected data is very helpful in test operation for optimizing
parameters. The status can also be monitored if the SIWAREX FTC is controlled
by the SIMATIC CPU.
Belt load in % INT DBW038 Belt load in relation with the nominal belt load 5.14.11
Nominal belt load REAL DBD040 Nominal belt load [WU/LU] calculated from 5.14.12
nominal flowrate and nominal speed
Belt load REAL DBD044 Current belt load [WU/LU] 5.14.13
Flowrate 1 REAL DBD048 Flowrate 1 [WU./s] 5.14.14
Flowrate 2 REAL DBD052 Flowrate 2 [WU/h] 5.14.15
Flowrate 3 REAL DBD056 Flowrate 3 [LWU/h] 5.14.16
Flowrate in % INT DBW060 In reference to the nominal flowrate 5.14.17
Operating hours counter INT DBW062 (is not reset, begins after overflow at 0) 5.14.18
Pulse counter value DINT DBD064 Pulse counter value in reference to the set 5.14.19
measuring time
68
5.14.1 DR 30 – NAWI-Statusbits
10 Dyn. zero setting aborted 1: If dynamic zero setting could not be completed or was
ended early because of leaving the zero setting range
11
12
13
14 Limit value undercut Limit value has undercut minimum flowrate (if indicated with a
minimum flowrate delay in cases)
15 Limit value exceeds Limit value has exceeded maximum flowrate (is indicated
maximum flowrate with a delay in cases)
- Pulse1 external Pulses for external totalizer 1
- Pulse 2 external Pulses for external totalizer 2
16
17
18
19
20
21
22
23
24
25
26 Dosing 1:Switch on material discharge organ
27 Batch activeted Batch mode is active
28 Batch finished 1: Batch is finished
Reserved
The current weight value, which is used for the main display. Dimension: Weight
Unit [WU].
The current weight value in increased resolution, which is used for the main
display. Dimension: Weight Unit [WU].
5.14.8 DR 30 – Tare
The current weight value tare (numeral step from DR3). Dimension: Weight Unit
[WU].
The current belt speed defined in length units (e.g. m) per second.
The current belt load is output as the relationship of the current belt load to the
nominal belt load.
From the nominal flowrate (defined in DR5) and nominal speed (defined in DR5)
calculated nominal belt load in the dimension weight unit/length unit [WU/LU].
The current belt load is output as the relationship of the currently measured weight
to the effective belt length. The dimension is: Weight unit/Length unit [WU/LU].
5.14.14 DR 30 – Flowrate 1
The current flowrate 1 is output in the dimension Weight unit / sec [WU/s].
5.14.15 DR 30 – Flowrate 2
5.14.16 DR 30 – Flowrate 3
The current flowrate 3 is output in the dimension Large Weight Unit / h [LWU/h].
The operating hours counter records the number of hours that the SIWAREX FTC
is supplied with 24V voltage.
Current counter value on the pulse input in relation to the defined measurement
time.
The user or the service technician can retrieve information on the current status of
the scale in the expanded process values.
Operating error DWORD DBD012 Error word Operating error (for test dept., 5.15.3
Bit-No. = Error code –1)
curr. Date/Time DATE_A DBD016 curr. Date/Time 5.15.4
ND_TIME
Current temperature SHORT DBW024 Current inner-temperature of the module [0.1°C]. 5.15.5
If no temperature value is available: -100.0°C.
DE status BYTE DBB026 Current status of the digital inputs 5.15.6
Reserve_31_27 BYTE DBB027 Reserve
32
The current number of digits from the AD converter is output without being filtered.
The current number of digits from the AD converter is output after the filtering.
The current existing operating errors (fault) are output as double word.
The current date and time from the internal clock of the SIWAREX FTC.
5.15.5 DR 31 – Temperature
The current signal level to digital inputs 1...6. The counter input signal will not be
shown in this value.
The total resistance of the load cell(s) which is measured upon commissioning.
Measurement is initiated with an instruction.
Current measured total resistance of the load cell(s). Measurement is initiated with
an instruction.
Totalizing is performed at the same time for every totalizing memory location. The
user can delete a certain location at any time which defines the totalizing period.
Content of the totalizing memory. Dimension in weight units [WU]. Can be reset
with instruction "Delete totalizing memory 1" or "Log and delete totalizing memory
1".
Content of the totalizing memory. Dimension in weight units [WU]. Can be reset
with the instruction "Delete totalizing memory 2" or "Log and delete totalizing
memory 2".
Content of the totalizing memory. Dimension in weight units [WU]. Can be reset
with the instruction "Delete totalizing memory 3" or "Log and delete totalizing
memory 3".
Content of the totalizing memory. Dimension in weight units [WU]. Can be reset
with the instruction "Delete totalizing memory 4" or "Log and delete totalizing
memory 4".
Content of the totalizing memory. Dimension in Large weight unit [LWU]. Can be
reset with instruction "Delete totalizing memory 5" or "Log and delete totalizing
memory 5".
Content of the totalizing memory. Dimension in Large weight unit [LWU]. Can be
reset with instruction "Delete totalizing memory 6" or "Log and delete totalizing
memory 6".
Content of the totalizing memory. Dimension in Large weight unit [LWU]. Can be
reset with instruction "Delete totalizing memory 7" or "Log and delete totalizing
memory 7".
Content of totalizing memory. Dimension in Large weight unit [LWU]. Can be reset
with instruction "Delete totalizing memory 8" or "Log and delete totalizing memory
8".
The ASCII weight value corresponds with the value on the main display of the
scale.
The data record is displayed in a separate window in SIWATOOL FTC by
activating function "Actual value".
18
SIWAREX FTC controls the content of DR 34 with instructions 25, 26, 128...144.
The content of the DR35 corresponds with the DR34, but is encoded. The
verifiable display is generated on the SIMATIC OP/TP from the contents of DR 35.
The logs can be defined in data records DR40 to DR 43. A log text can either be
printed automatically or on command or it can be saved in the verifiable MMC
memory.
Log text entries 1 to 4 are composed of fixed segments such as title and label and
of variable fields.
The variable fields can contain various values from the SIWAREX FTC.
In addition, 4 designation variables can be defined in addition to the process
variables. These designations are 4*16 bytes long and can be specified at any time
by the SIMATIC CPU. DR 40 to 43 cannot be defined through the SIMATIC CPU.
NAWI-Felder (nn)
0 NAWI-Status flags j n - -
1 Grossweight_Process j j 14 •12345.678•kg•
2 Netweight_Process j j 14 •12345.678•kg•
3 Tareweight_Process j j 14 •12345.678•kg•
4 G/N weight verifiable j j 18 <N1•12345.678•kg•>
<B1•-12345.67•kg•>
5 G/N weight_x10 j n - -
The log data up to the next logging procedure are stored in DR 44. If required, the
contents can be output again.
Last log data STRING[160] DBB012 STRING[160] = 0H Log text for last logging procedure
174
5.20.1 DR 44 – MMC-ID
5.20.2 DR 44 – Log-ID
Log ID is a single time log number which can be used for identifying the log. The
number is incremented with every logging procedure.
The last log data that was output can be read until the next logging procedure.
5.21 DR 45 String
The string is test that can be inserted as variables in the log. The content of the
string can be defined through SIWATOOL or from control program of the SIMATIC
CPU.
72
The trace function is used for logging measurement values and current states of
the scale in RAM memory of the SIWAREX FTC or in the MMC card.
When logging to RAM, SIWATOOL FTC reads the trace elements with data record
120 and when logging to the MMC it is done through data record 121.
Logging is started with instruction "Start recording" 70 and can be ended with the
instruction "End running recording" 71.
Length
Name Type Range of Values / Meaning Reference
bytes
Timestamp1 DATE&TIME 8
Conti- Status flag DINT 4 status bit
NAWI status flags DINT 4 status bit
Unfiltered ADC value DINT 4 Unfiltered digit value from ADC
Filtered ADC value DINT 4 Filtered digit value
Net_Process REAL 4 Net weight
Belt load REAL 4 Current belt load WU/m
Pulse counter value DINT 4 Pulse counter value with reference to the
measurement time
Flowrate 1 REAL 4 Flowrate [Weight/Time unit]
Belt speed m/sec REAL 4 Belt speed
Status Digital Inputs BYTE 1 Actual statos of digital inputs
Reserve BYTE 1 Reserve
46
The evaluation of the collected data can be handled using SIWATOOL FTC and
MS Excel. All of the logged elements that have been recorded are stored in the
Excel table and can be read in a graph over time.
The progress of the weighing procedure can be analyzed and optimized.
In the event of sporadic events in the system, the recording can be activated as
circulating memory and can be ended with a condition in the SIMATIC control
program at any time. The subsequent evaluation can help to explain a sporadic
event in the system. A simultaneous evaluation of the message buffer, in which the
last 100 events are stored (evaluation with SIWATOOL FTC) offers a good basis
for an analysis or remote analysis.
An element requires 46 bytes. If the recording is run e.g. once per 50 msec, it
means memory requirements of 920 bytes per second.
In the simplest case, the RAM memory (approx. 2 MB) of the SIWAREX FTC can
be used for the recording.
By reading the DR 122, SIWATOOL FTC can be used to determine which data is
stored on the Micro Memory Card. Based on this information, the user can read
targeted records and logs.
MMC
The ID is incremented with every log output (not
with repeat") and is stored and managed in MMC
memory. This number can only be reset by
deleting or formatting the card.
MMC ID BYTE[5] 5 1 byte manufacturer ID and 4 bytes serial number
Reserve 1 BYTE 1 Reserve 1
Reserve 2 INT 2 Reserve 2
MMC memory capacity DINT 4 MMC – total memory capacity in bytes
Available MMC capacity DINT 4 Capacity of the MMC for log data [Bytes]
for log data
Available capacity for DINT 4 Display depending on the defined trace target
trace data RAM or MMC in the interface parameters [Bytes]
48
By reading the DR 122, SIWATOOL FTC can be used to determine which log data
is stored on the MMC.
A log can then be read for a defined log ID.
6 Commands
The SIWAREX commands are divided into groups. The combining of the
commands into a group is done according to their functionalities.
Every command has a unique number. A command can be sent via various
interfaces (SIMATIC, SIWATOOL FTC, digital input).
Every time that a command is sent to SIWAREX FTC, a check must be made for
whether the command has been executed correctly. The data and operating errors
(synchronous errors) that are generated provide information on why the command
could not be executed.
The group Service- and adjustment commands is used during the
commissioning of the scale. As long as the scale is not adjusted, it can only be
used in service operation. Switching to service operation is indicated in the status
bits.
The group of Scale commands NAWI contains all commands that affect the
handling of a statistic scale. Normally, commands that have anything to do with the
operation of the SIWAREX FTC as a Non-Automatic Weighing Instrument (e.g.
zero setting, taring).
The extensive group of Belt Scale commands is used for controlling the belt
scale. The basic behaviour of the SIWAREX FTC is defined one time by defining
parameters.
The group of Log commands contains commands that control the log output to a
printer or the calibratable storage of the logs to the MMC.
Use of the Micro Memory Card MMC is controlled with commands from the
command group Micro Memory Commands.
After receiving a command, SIWAREX FTC checks for whether the command can
be executed. If the check results are negative, the user is informed of the cause by
an output of a "synchronous" message (see chapter 7 Messages and
Diagnostics).
Scale commands
22 Tare
The current weight is set to zero and the weight
display is designated as "Net" and "Tare" at the
same time.
23 Delete tare
The tare is deleted. The current weight is
displayed and designations of "Net" are changed
to "Gross", the designation "Tare" or "Preset-
Tare" is reset.
Log commands
40 Reserve
41 Reserve
Scale commands
127
The commands in the table above can be activated over all interfaces.
Command Description
group
1... 199 Commands are passed on to the module group without reading or writing from or to
data records (scale, weighing, logging commands). The meanings of these
commands correspond with the settings in table XX.
203... 245...399 Read a data record 3 ... 45. The numbers 246 ... 399 are reserved for expansions.
403... 445...599 Write a data record 3 ... 45. The numbers 446 ... 599 are reserved for expansions.
601... 699 Range for combined commands. The function block SIWA_FTC (FB44) can
transfer several data records in sequence.
649 Read all data records in the SIWAREX FTC (DR3, DR4, DR7, DR5, DR8, DR9,
DR14, DR15, DR16, DR17, DR18, DR20, DR21, DR30, DR31, DR32, DR34,
DR35, DR44, DR45).
699 Write data blocks DR3, DR4, DR5, DR7, DR8, DR14, DR15, DR18, DR21, DR45 to
SIWAREX FTC.
Table 6-2 Command groups of SIWAREX FTC
More information on command transfers from the control program using the SIMATIC interface
can be found in chapter 8 Programming in SIMATIC STEP 7.
The status displays are not messages in this respect. The status display
describes the status of the scale in normal operation and can be monitored or
evaluated at any time.
The SIWAREX FTC messages are passed on to the user by various paths. During
project planning, it is important to choose the right path for routing and processing
the messages.
A message buffer is integrated in the module and holds up to 100 entries. If the
number of messages in the message buffer reaches 100 then a new message will
immediately delete the oldest message. The message buffer can be read at any
time with SIWATOOL FTC (menu point "Read all data records") and can be saved
together with the scale parameters. This helps in recognising, analysing and
solving problems in the system.
All of the messages of the SIWAREX module can be recognised and processed in
the controller using the FB SIWA_FTC. Additional errors in the processing of the
FB SWIA_FTC are put out through the FB_ERR output variable (see chapter 8
Programming in SIMATIC STEP 7.)
Operating messages (hardware faults) are recognised in the SIMATIC CPU with
diagnostic alarms. For more information, see chapter 8 Programming in SIMATIC
STEP 7.
66 Logging not possible since Logging is not possible since the MMC is full. Switch
the MMC is full MMCs.
67 Logging task cannot be Logging is not possible since the MMC is defective.
executed, since MMC is Switch MMCs.
defective or non-existent
68 Log output not possible since Repeating the last log print-out is not possible since
no data is available. there has been no previous print-out or the saved data
is defective.
70 Logging output not possible Log output is not possible since the content of the text
since the log ID is not in the does not contain the log ID. Define the log ID in the
print text text. This entry is necessary if the verifiable weight is
contained in the log text in verifiable operation
((Technical regulation code OIML in DR3).
76 Value range Adjustment digits The definition for the adjustment digits in DR 3 is too
exceeded large. Enter a smaller value. The highest permitted
value is 16 777 215.
77 Regulation code not permitted The regulation code is incorrect. Enter the proper
code. The permitted entries are ---- or OIML.
78 Zero setting range > 4% or The zero definition range or zero setting range is too
zero setting range exceeded large. This is the case if you have defined a zero
setting range of > 4% of the maximum weighing range,
or > 20% of the zero setting range in DR3 in verifiable
operation (technical regulation code OIML in DR3).
The sum of the negative and positive values apply in
this instance. Reduce the zero definition range or zero
setting range.
79 Number of weight ranges not The number of the weighing range is incorrect. Enter
allowed 1, 2 or 3.
81 Relationship between The definition for weighing ranges is incorrect in DR 3
weighing ranges implausible (Max- and Min-values).
On a multi-resolution scale, the Min-value of the next
higher range must be equal to the Max-value of the
range underneath. The Max-value must be greater
than the Min-value.
97
98 Totalizing memory already The Delete totalizing memory command could not be
deleted executed since the totalizing memory is already zero
99 Totalizing memory could not An error has occurred while logging the totalizing
be logged memory. Check the output device. Check the
additional technology errors for indications concerning
the cause of the error. A technology fault is defined
with the cause.
100 Weight outside zero definition Zero definition cannot be executed since the current
range zero point lies outside the zero definition range defined
in DR 3 or the current gross value lies above the
largest defined weighing range value. For the belt
scale the limit is defined in DR5.
101 Scale command cannot be The weighing command (Taring, Zero definition,
executed since not at Logging) cannot be executed since the Standstill 1 is
standstill not within the defined standstill time in DR 3.
102 Adjustment weight missing Adjustment command cannot be executed because
the respective adjustment weight is not entered in DR
3.
105 Weighing operating mode not Incorrect weighing operating mode in DR 3. Enter a
allowed or unknown correct weighing operating mode.
106 Limit values implausible The defined limit values are implausible in DR 4. Limit
value 3 may only be used as a maximum limit value
(On ≥ Off). Correct limit value 3.
111 Parameter implausible for At maximum flowrate, the pulse for the external
external totalizing memory summator cannot no longer be output.
112 Nominal speed implausible. Nominal speed implausible
(must be larger than 0).
113 Pulse constant implausible. Pulse constant implausible
(must be larger than 0).
114
115 Allocation of log text No. for The log text number is incorrect in DR21. Correct the
autom. logging not permitted number (0..4).
116 Belt length implausible Effective belt length implausible
(must be larger than 0).
119 Tilt angle implausible Tilt angle of the belt implausible
(must be larger than 0).
143 Load set point too low Load set point too low
150 Command impossible since Command impossible since SIMATIC CPU is in STOP
SIMATIC CPU is in STOP
166 Date Time implausible The definition for date or time is not plausible. Correct
the entry.
The following are allowed:
Year: 0..99
Month: 1..12
Day: 1..28/29/30/31
(depending on the month)
Hour: 0..23
169 Negative default value not Negative default value not permitted Change the
permitted definition.
170 Number range for weighing The numeral range for a definition is invalid. The
values exceeded values must be greater than -100 000 000 and less
than 100 000 000.
16 Logging is not possible since there is no Logging is not possible since no log ID is
log ID in the print text defined in the log string (verifiable application,
code in DR3 – OIML).
17 Logging is not possible because the Logging is not possible because the printer has
printer is not ready not indicated that it is ready. Check the printer.
18 Logging is not possible because the Logging is not possible because the MMC is
MMC is defective or is not inserted defective or it has been removed. Change or
insert the MMC.
19 Logging is not possible because the Logging is not possible because the
SIWATOOL-driver is active SIWATOOL program is active on the interface.
Disconnect the connection to the PC and
connect a printer.
20 Reserved
21-40 Error in weighing function
21 Scale command aborted, not at standstill Taring/Zero definition or logging not possible
because no standstill has been achieved within
the standstill waiting time 1 defined in DR 3.
22 Zero setting not possible The switch-on weight lies outside of the value
range defined in DR 3 for zero setting with the
maximum positive and negative weight.
23 Zero setting / taring not possible Zero definition or zero setting not possible
because the gross is outside the zero definition
range or taring is not possible because the tare
maximum load has been exceeded.
41-100 Error in continuous operation
41 Belt speed too small Belt speed too small. Test the belt.
42 Belt speed too large Belt speed too large. Check the drive control.
43 Belt load too small Belt load too small, check the material feed.
44 Belt load too large Belt load too large, check the material feed.
45 Flowrate too small Flowrate too small, check the material feed.
46 Flowrate too large Flowrate too large, check the material feed.
47 Dynamic zero setting or adjustment Dynamic zero setting or adjustment has been
aborted aborted.
48 Batch time exceeded The time for one batch (DR21) is exceeded.
49 Impulse output error Overrun of the impulse output buffer.
50 Weighing stopped by SIMATIC-CPU- Weighing stopped by SIMATIC-CPU-stop
stop
Table 7-2 List of Technology Messages
If an error generates an operating message, the red SF LED on the front of the
assembly is illuminated. The operating messages are generated coming and going.
(K) – coming, (G) – going
During project planning for the hardware configuration in the SIMATIC manager,
the basic properties of the module are defined:
- The peripheral address of the module
- Diagnostic alarm enable
- Process alarm enable
- Behaviour for CPU stop
SIWAREX FTC communicates with the SIMATIC CPU with the function block FB
SIWA_FTC. While programming the call, an instance data block is created for the
FB SIWA_FTC. Besides the instance data block, a scale DB is required for every
Scale SIWAREX FTC, in which the scale parameters are stored. The UDT that
comes with it can also be used for creating the scale DB.
The vector DB must also be loaded in the SIMATIC CPU. A vector DB can be
used by more than one SIWAREX FTC.
The function block FB SIWA_FTC and the data block are found on the CD for the
project planning package SIWAREX FTC for SIMATIC S7 in the S7_Software
directory.
The function block FB SIWA_FTC is called one time for each scale in the
application program cyclically in a program level (e.g. in OB1) and provided with
call parameters.
CALL "SIWA_FTC" , DB 15 (
ADDR := 256,
DB_SCALE := 15,
DB_VECTOR := 14,
CMD_IN := "DB_SCALE".i_CMD_INPUT,
SIM_VAL := "DB_SCALE".r_SIM_VALUE,
ANA_OUT := "DB_SCALE".r_ANALOG_OUT_VALUE,
DO_FORCE := "DB_SCALE".b_DIG_OUTPUT_FORCE,
TRANSITION := “DB_SCALE”.b_Reserve
CMD_INPR := "DB_SCALE".bo_CMD_IN_PROGRESS,
CMD_FOK := "DB_SCALE".bo_CMD_FINISHED_OK,
CMD_ERR := "DB_SCALE".bo_CMD_ERR,
CMD_ERR_C := "DB_SCALE".b_CMD_ERR_CODE,
REF_COUNT := "DB_SCALE".b_INFO_REFRESH_COUNT,
PROC_VAL1 := "DB_SCALE".r_PROCESS_VALUE1,
PROC_VAL2 := "DB_SCALE".dw_PROCESS_VALUE2,
SC_STATUS := "DB_SCALE".dw_SCALE_STATUS,
ERR_MSG := "DB_SCALE".bo_ERR_MSG,
ERR_MSG_TYPE := "DB_SCALE".b_ERR_MSG_TYPE,
ERR_MSG_C := "DB_SCALE".b_ERR_MSG_CODE,
FB_ERR := "DB_SCALE".bo_FB_ERR,
FB_ERR_C := "DB_SCALE".b_FB_ERR_CODE,
START_UP := "DB_SCALE".bo_START_UP_IN_PROGRESS,
CMD_EN := "DB_SCALE".bo_CMD_ENABLE,
ERR_MSG_Q := "DB_SCALE".bo_ERR_MSG_QUIT);
The call parameters of the FB SIWA_FTC are described in the following section.
The call parameters are defined as variables in the scales DB when delivered. It is
possible to define the call parameters with other variables of the same type.
While calling the FB SIWA_FTC, the number of the instance DB to be generated
must be defined.
SIWAREX FTC requires 16 bytes in the input and output range of the SIMATIC
CPU for operation. The ADDR parameter must correspond wit the definition in the
HW configuration.
A scale DB must be defined for every scale, in which the parameter of the
SIWAREX FTC and the current actual value are found. The number of the DB can
be selected as desired. DB15 has been defined as the scale DB in the project
planning package. In addition, the UDT15 is also provided as a template for
creating block modules.
The content of the vector DB may not be modified by the user. It must be loaded
one time only per SIMATIC CPU, independent of the number of SIWAREX FTC
modules that are used. The number of the DB can be selected as desired.
The user controls all command using this input variable, whether for transferring a
data record or executing a weighing task. The commands are described in chapter
6. The user prepares the command number using this variable and triggers the
command using variable CMD_EN:= "DB_SCALE".bo_CMD_ENABLE (see
chapter 8.4.22). The FB SIWA_FTC does not delete the command number, it
resets the trigger variable CMD_EN:= "DB_SCALE".bo_CMD_ENABLE after the
command is finished.
If the simulation has been enabled (see chapter 5.5.1), the value to be simulated
can be defined on this input. The value should be found in the weighing range of
the weighing instrument.
If control of the analog outputs has been enabled (see chapter 5.5.11), the value to
be controlled can be defined on this input. The value should be found in the
defined range of the weighing instrument.
If forced control of the digital outputs has been enabled (see chapter 5.5.3), the
value to be controlled can be defined on this input. Bit 0 corresponds with digital
output 0, bit 1 corresponds with digital output 1, etc.
This bit informs the user that a command is being processed at the moment.
This bit informs the user that a command has been executed successfully
(Command complete without errors).
This bit informs the user that a command has not been executed. The bit is set for
one cycle (edge) only. The cause can be evaluated in the same cycle with variable
CMD_ERR_C:= "DB_SCALE".b_CMD_ERR_CODE. The number is decoded in
the table "Data and Operating Errors" in chapter 7.6. If no error code is defined, the
error must be evaluated in "DB_SCALE".b_FB_ERR_CODE.
If a command is not executed (completed with error) the error code number is
output here. The number that is output is decoded in the table "Data and Operating
Errors" in chapter 7.6. The value remains in the output until the next command is
triggered. The evaluation is to be performed when the set bit CMD_ERR:=
"DB_SCALE".bo_CMD_ERR appears. If an error code is defined, the error must be
evaluated in "DB_SCALE".b_FB_ERR_CODE.
The current output values which are prepared as output variables of the FB
SIWA_FTC are read cyclically by the FB through the peripheral range. SIWAREX
FTC updates the values internally in a 10 msec rhythm. Every update is assigned a
number which can be used in the SIMATIC CPU like a time stamp.
The selected process value can be output using this variable (See chapter 5.5.4).
The gross weight or the net weight of the scale is normally output here.
The selected process value is output using this variable (See chapter 5.5.5). The
status of the automatic weighing instrument AWI (see chapter 5.14.2 ) is normally
output here.
The status of the non-automatic weighing instrument NAWI (see chapter 5.14.1) is
always output through this variable.
All messages that are prepared by SIWAREX FTC are prepared in an output buffer
on the module. If a new message appears, this bit is set. The user can evaluate the
meaning using variables RR_MSG_TYPE:= "DB_SCALE".b_ERR_MSG_TYPE
and ERR_MSG_C:= "DB_SCALE".b_ERR_MSG_CODE. After the message is
The user is informed of the message number with this variable (suitable for
ERR_MSG_TYPE:= "DB_SCALE".b_ERR_MSG_TYPE). After the evaluation of
the message, the user indicates to the FB that the message has been evaluated
(with variable ERR_MSG_Q:= "DB_SCALE".bo_ERR_MSG_QUIT) and the FB
SIWA_FTC is ready to output the next message if one exists.
If an error has occurred during the processing of the function block itself, it is
indicated through this variable.
!
Warning
If a processing error occurs for FB SIWA_FTC, we have to assume that the
variables that have been output do not correspond with the actual status in the
module.
Bit 7 – Reserved
!
Warning
If a processing error occurs for FB SIWA_FTC, we have to assume that the
variables that have been output do not correspond with the actual status in the
module.
The contents of the scale DB are data of data records witch are available for
SIMATIC-CPU. All components of the SIMATIC STEP7 software for SIWAREX
FTC are in the English language.
For displaying the calibratable weight value, the same device that the operator
uses for running the system can be used.
The value for the calibratable weight display is generated internally by the
SIWAREX FTC, it is encoded and then provided in the data record DR 35 for the
user. The DR 35 data record is read by the FB SIWA_FTC and is put in the scale
DB just like every other data record in a defined address. The content of the DR 35
can be sent through bus connections to other SIMATIC CPUs or can be evaluated
locally.
An AddOn for ProTool is used for evaluating – a special function which is in the
position to decode the contents of the DR 35 and to display it in a special output
field. This AddOn is called "Secured output" and must be installed separately to
ProTool.
8.6.2 Installation and Project Planning for the Calibratable Weight Display
First, the OCX must be linked into the project planning environment of ProTool for
the secured output. This is possible with version 6.0 SP2 ProTool(Pro) or higher.
The setup.bat program which can be used for performing the installation is found in
the OCX_DISPLAY directory of the project planning package.
If the ProTool is not on drive C: or D: then the setup.bat program must be edited in
an editor and rewritten to correspond with the drive that it is on.
After executing the setup.bat program, the Prootool.ini file must be changed
according to the note in setup.bat.
If ProTool is called again afterwards, a new function appears in the tool-bar (see
arrow) – "Secure output", which is projected just as all other functions.
During planning, the address of the data record DR 35 must be defined. The
variable is BYTE and has a length of 32 bytes.
The refresh rate lies somewhere around 200 to 300 msec which is comfortable to
the eye.
Note
After converting a project to another target device, the secured output must be
deleted in the images and projected again.
In the first step, SIWAREX FTC must be added to the hardware catalogue by
running the SETUP_FOR_SIMATIC_S7.
While planning the hardware configuration in the SIMATIC Manager, the basic
features of the module are defined:
- The peripheral address of the module
- Enabling the diagnostic alarms
- Enabling the process alarms
- Behaviour in the case of a CPU-Stop
Other scale specific parameters that are also changed while the control program is
running can be defined in three different ways.
- Using the SIWATOOL FTC parameter definition tool
- Internally by making the definition in FB641 and then transferring to SIWAREX
FTC
- In the OS using the Faceplate.
Here first the SIWAREX FB is described , then the Face Plate from the operator's
viewpoint and finally there are notes on project planning with the Faceplate.
The SFT_FTC function block is integrated into an interrupt alarm-OB e.g. OB32.
The block must also be integrated in the following OBs in the run sequence (done
automatically in CFC):
The block is used for controlling a SIWAREX FTC module group. Data is
transferred cyclically through the peripheral interface and the various data records
are read from the modules or transferred to the modules acyclically. The message
queue of the modules is read continuously and respective WinCC messages are
generated.
Note:
In the definition of the S7 interface in DR7, the values for PROCESS_VALUE_1
(5.5.4) and PROCESS_VALUE_2 (5.5.5) must be assigned as follows:
PROCESS_VALUE_1 = 2 (Flow 1)
PROCESS_VALUE_2 = 30 (Conti Status)
The IO addresses for the SIWAREX FTC modules must be completely within the
process image of the CPU. The LADDR input has to be connected with the base
address of the SIWAREX FTC modules. Procedures:
Select input -> right mouse key -> switch to operand ... -> entry of e. g. EW512.
The PCS7 driver wizard then automatically creates all required driver blocks. The
Function block inputs MODF, PERAF and RACKF are automaticly connected by
the driver wizard, the inputs SUBN1_ID, SUBN2_ID, RACK_NO, SLOT_NO,
BASADR and DADDR are automaticly defined according to the data in the HW
configuration.
9.2.4 Manual/Automatic
Switching between the two operating modes is either done using the OS
operations via AUT_ON_OP (LIOP_SEL = 0) or by switching the input AUT_L
(LIOP_SEL = 1). When selecting with the OS system, the corresponding
AUTOP_EN and MANOP_EN enables are required. The defined operating mode is
displayed on output QMAN_AUT (1: Auto, 0: Manual).
Manual operation: The commands are sent from the operator to the module
through input MAN_CMD. Each change in the command code on this input is
recognized as a new command. The source of data records to be sent to the
modules are the manual inputs (extension ‘_M’).
Automatic operation: The function block takes its commands on a positive edge
on input AUTCMDEN from the switchable AUT_CMD input.
The source of data records to be sent to the modules are – as long as they exist –
the automatic inputs (extension ‘_A’), and otherwise the manual inputs (extension
‘_M’).
A command sequence (e.g. read all data records) will only be interrupted by a new
command code after the current command is completely executed.
All data records that the S7 controller can access exist as individual parameters of
the function block. The parameters of the data blocks that can be read have the
extension ‘_O’ for Output. The parameters of the data blocks that can be written to
have the extension ‘_M’ for Manual and are transferred to WinCC for visualization.
9.2.6 Commands
The error message buffer on the SIWAREX FTC module group is read
continuously by the block. If a message is read then the ERR_MSG output is set to
"TRUE” for one cycle. Outputs ERR_MSG_TYPE and ERR_MSG_C contain the
type of error and the error code of the respective message.
ERR_MSG_TYPE Meaning
16#01 Operating messages (Fault)
16#02 Technological fault
16#04 Data or Operating Error
sent in addition with operating message
16#80
(fault).
Table 9-1 CFC – Message types
The meaning of the error number codes can be found in the message lists.
Corresponding with the type of error, WinCC sets messages with text,
technological error, data/operating error, internal or external error with the error
code attached. These messages always have either a received or sent status. The
error code of the last read error message is always present. The most important
operating error messages are generated individually.
9.2.8 Allocating message text and message class to the block parameters
Outputs:
Outputs:
Outputs:
Outputs:
Outputs:
The example faceplate for the SIWAREX FTC modules was created with the
Faceplate Designer from PCS7 version 6.0. The WinCC images and scripts that
are created can be modified according to individual requirements.
The Faceplate should be called with the dynamic wizard “Picture selection via
measurement point”. The example faceplate includes the following views:
Here the operator can view the current Flow of the scale and some selected status
bits. The Manual/Automatic operating modes can also be switched.
In this view the operator can view and change important calibration data. Pressing
the button “Send Data” or “Receive Data” will send and receive the hole data
record 3 respectively, also the data that are not shown in these views.
Here the operator can execute all important commands for calibration and
operation of the belt scale further on he can watch selected important status
information. !! Pressing the button “Send Data” or “Receive Data” will send and
receive the all data records of the Siwarex FTC respectively.!!
In this view the operator can watch, delete- and delete and protocol all the
totalizers. Deleting means that the totalizer left hand to the button will be set to
zero. “Delete and Prot.” Means that the toalizer will be printed with log text 1 (The
totalizer has to be in log text 1) and after that set to zero. The button “Delete all”
will cause a set to zero of all totalizers, in tab “Sum 1” and “Sum 2”. The “Actualize”
button will read the data record 33 one time, all totalizers will be updated, only after
pressing this button the display will be updated.
Mainly, those standard items that are described in the documentation on the
Faceplate Designer and that have been delivered with the Faceplate Designer are
used. This description concentrates on the features that have been implemented
for the SIWAREX FTC Faceplate.
Tabs
To help clarify matters, two Faceplate view with up to 4 tabs have been shown in
different images. Switching between the tabs is done using function
"SH6_ChangeView_tab.fct". Each tab must have the name of the image that it is
calling.
Operating authorization
In every view, an element with the name "Level5_MODE" or "Level6_MODE" is
found. These elements do not only the enable operating authorization from the
User Administrator but they also deny operating authorization in Automatic
operating mode. This is done with function "SH6_CheckPermission_Plus.fct",
which is called when the image is loaded and when the operating mode is
changed. Passing the operating mode to the individual elements is performed
through direct connections.
While selecting a command with the mouse, the command code is written into the
IO-field "Value" of the "@PG_SFT_FTC_SCROLL_WEIGHING.pdl" image. If the
Faceplate Views
In all Faceplate views, the invisible objects in the WinCC runtime are also
displayed. The switching of the individual elements can also be handled in the
WinCC images.
10.1 General
The scale can be put into operation using the SIWAREX FTC program
independent of commissioning the SIMATIC automation system.
The program is included with the delivery of the project planning package.
The program (Catalogue SIWATOOL_FTC) must be installed in the first step.
Memory requirements on the hard disk are less than 30 MB.
!
Warning
In online operation, all data in the module can be edited. The changes are not
automatically fed into the respective scales-data records. You, as the user, must
decide whether the data adjustment is necessary and whether it should be
performed or not.
After clicking on a data record in the left-hand program window, the "Info" card can
be selected from the right-hand program window. A description on a white area
indicates the affect that the data record has on the scale's behaviour.
After selecting a card, a section of the respective data record is displayed as input
and output fields. Besides the parameter identification, Tool Tips (text is displayed
when the mouse is over the field) help to describe the parameter.
Clicking on the menu point "Help" opens the "SIWAREX FTC" manual. The
Acrobat Reader program must be installed in order to read the manual.
If you want to utilize the firmware update offer on the SIWAREX internet pages
(www.siwarex.com / Support), you are able to download the latest firmware cost-
free and using the SIWATOOL FTC, you can transfer it to the modules.
The firmware is located in Flash memory. If required, new firmware can be
transferred to the modules.
The new firmware can slightly vary from the earlier version – this is the case if the
data structures of SIWAREX FTC parameters are not changed. In this case,
loading the new firmware does not change the actual data.
If function expansion in the new firmware results in new internal data structures,
new data records or changes within existing data records then the SIWAREX FTC
assigns the parameters with default values after the download. Therefore, the
original parameter status is read using the SIWATOOL FTC ("Communication
menu", "Retrieve all data records") and stored in a file.
Loading the firmware onto the SIWAREX FTC module is performed over several
steps:
1. Switch the SIMATIC CPU to STOP.
2. Log SIWATOOL FTC in (online)
3. Select Firmware Download, set download mode
4. Select Firmware file
5. Activate download mode with the checkmark
6. Start the transfer
The transfer can take up to several minutes.
!
Warning notice
During the firmware download, SIWAREX FTC do not respond to communication
requests from the SIMATIC CPU. The CUP indicates the peripheral access error
and if no respective organization block (OB) is programmed then the CPU can go
to STOP.
Therefore, loading the firmware should only be done when the CPU is in STOP
status.
12 Calibrating Applications
Attention
Calibratable scales may only be certified by a calibration expert or qualified
! representative.
Preparations The following preparations must be made by the scale users before the
actual certification by the calibration expert:
Calibration sticker The calibration sticker is found in the calibration set which you can get as an
accessory.
Certifying the The certification of the calibrated scale is done by a calibration expert from a
SIWAREX FTC recognised position.
Stamp of approval on After activating the write protection (WRP switch on the front), the calibration
SIWAREX FTC expert attaches the certification stamp and the calibration mark.
The calibratable main display for SIWAREX FTC can either be done on a
calibratable S11 display (company Siebert) or with certain standard SIMATIC
13 Accessories
There are necessary and optional accessories for the SIWAREX FTC.
The required accessories are indicated in the following table.
• Manual on CD-ROM
SIWAREX FTC Project Planning Package for PCS7 V6.0 on CD-ROM 7MH4900-3AK61
• Manual on CD-ROM
SIWAREX FTC "Getting started”
STEP 7-Software application for SIWAREX FTC.
• Length 2 m 7MH4702-8CA
• Length 5 m 7MH4702-8CB
- Scale connection
- RS 485 interface
- RS 232 interface
• S7 profile rail
The digital remote display model Siebert S11 can be connected directly through
an RS 485 interface to the SIWAREX FTC
Cables (optional)
Rated voltage DC 24 V
Static upper / lower limits DC 20.4 V/28.8V
Dynamic upper / lower limits DC 18.5 V/30.2 V
Non-periodic over-voltages DC 35 V for 500 msec
with a recovery time of 50 sec.
Max. current consumption 500 mA
Typical assembly power loss 7.5 W
Table 14-1 Data: 24 V power supply
Current = 184 mA
Permitted load cell resistance without Ex-i-Interface > 56 Ω
< 4010 Ω
Permitted load cell resistance with Ex-i-Interface > 87 Ù
< 4010 Ù
Monitor for sensing inputs Typical = 5 V
Hysteresis 120 mV
Response time for sense line monitor = 1 sec
Common mode rejection CMRR @50 Hz Typical 120 dB
Measuring value filtering for low pass filter 0.05...20 Hz
Measuring value filtering for average value filter 2...250 values
Impedance measurement for load cells Measuring range 56 Ω...4010 Ω
Precision ±5%
Repeat accuracy <1 %
Potential separation 500 V
Number 1
Rated voltage DC 24 V
Potential separation 500 V
conforming with IEC 1131, UL 508,
CSA C22.2 NO. 142
Voltage range for H signal DC 9 V to 30 V
Voltage range for L signal DC -3 V to 5 V
Input current (15 to 30 V) 2 to 15 mA
Switching frequency max. 10 kHz
Potential separation 500 V
Table 14-6 Data: Counter input CI
EMC For EMC, guidelines according to NAMUR NE21 Part1, as well as the
European guidelines 90/384/EWG for non-automatic weighing instruments
and 89/336/EWG concerning the emission and sensitivity of electromagnetic
disturbance are taken into account.
For use in extreme operating conditions (e.g. heavy dust, acidic moisture or
gasses etc.), additional measures are to be taken such as B. Encapsulation.
Climatic requirements
Remarks Environmental conditions Application ranges
Operating temperature: The S7-300 standard module
Vertical installation in S7-300 -10 to +60 °C groups may not be operated
Horizontal installation in S7- -10 to +40 °C under 0°C
300 -10 to +40 °C
Calibrated operation
Storage and transport -40 to +70 °C
temperature
Relative humidity 5 to 95 % Without condensation,
corresponding with relative
humidity (RH) – Exposure
level 2 according to DIN IEC
1131-2
Pollutant concentration SO2: < 0,5 ppm; RH < 60% no condensation
H2S: < 0.1 ppm;
Table 14-13 Data: Climatic requirements
16 Abbreviations