Mu.c.p. 2037
Mu.c.p. 2037
cso@csoitalia.it | www.csoitalia.it
IFU313EN00 - 01/2021
0051
EN
1/174
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
1 INTRODUCTION ......................................................................................... 3
1.1 SYMBOLS .......................................................................................................... 3
1.1.1 Device symbols ..................................................................................................... 4
1.2 GENERAL WARNINGS ........................................................................................... 4
1.3 NORMATIVE REFERENCES ..................................................................................... 5
1.3.1 Community directives ........................................................................................... 5
1.3.2 Technical standards .............................................................................................. 5
1.3.3 Quality management systems standards ............................................................. 5
1.4 WARRANTY ....................................................................................................... 5
1.5 MANUFACTURER IDENTIFICATION ........................................................................... 7
2 SAFETY....................................................................................................... 8
2.1 SAFETY WARNINGS .............................................................................................. 8
2.2 DEVICE IDENTIFICATION ...................................................................................... 10
2.2.1 Registration data in the Medical Devices List ..................................................... 10
2.2.2 Device data plate................................................................................................ 10
2.3 INTENDED USE.................................................................................................. 11
2.4 MEDICAL DEVICES CLASSIFICATION ........................................................................ 13
2.5 MEDICAL ELECTRICAL DEVICES CLASSIFICATION ......................................................... 13
2.6 ENVIRONMENTAL CONDITIONS ............................................................................. 14
2.7 DISPOSAL AT THE END OF THE USEFUL LIFE .............................................................. 15
2.8 MANUFACTURER DECLARATIONS .......................................................................... 16
3 DEVICE DESCRIPTION .............................................................................. 17
3.1 SUPPLY DESCRIPTION ......................................................................................... 17
3.1.1 CHART PROJECTOR device .................................................................................. 19
3.1.2 Remote control ................................................................................................... 22
3.1.3 Projector supports (optional) .............................................................................. 24
3.2 TECHNICAL DATA .............................................................................................. 26
4 DEVICE USE ............................................................................................. 29
4.1 HOW TO INSTALL THE DEVICE ............................................................................... 29
4.1.1 installation on table base ................................................................................... 29
4.1.2 Installation on ophthalmic unit column .............................................................. 30
4.1.3 Wall mounting .................................................................................................... 32
4.2 HOW TO PLACE ELECTRIC CABLES .......................................................................... 34
4.3 HOW TO TURN ON THE DEVICE ............................................................................. 35
4.4 HOW TO PROGRAM THE REMOTE CONTROL IN AN ENVIRONMENT WITH SEVERAL
PROJECTORS .................................................................................................... 37
4.5 HOW TO PROGRAM THE SEQUENCES...................................................................... 38
4.5.1 How to save a sequence ..................................................................................... 38
4.5.2 How to choose a sequence ................................................................................. 40
4.5.3 How to modify a sequence ................................................................................. 41
4.5.4 How to set the early interruption of the sequence ............................................. 42
4.6 HOW TO PLACE THE PATIENT................................................................................ 43
1/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
4.7 HOW TO FOCUS THE TESTS AND START THE SEQUENCE ............................................... 45
4.8 HOW TO TURN OFF THE DEVICE ............................................................................ 46
5 ORDINARY MAINTENANCE ...................................................................... 47
5.1 SAFETY WARNINGS ........................................................................................... 47
5.2 DEVICE CLEANING............................................................................................. 47
5.3 REPLACE THE REMOTE CONTROL BATTERIES ............................................................ 48
5.4 NETWORK FUSES REPLACEMENT........................................................................... 49
5.5 SPARE PARTS AND ACCESSORIES LIST ..................................................................... 50
5.6 TROUBLESHOOTING .......................................................................................... 50
2/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
1 INTRODUCTION
The device is the result of a long research period, conducted by
experts to give the product technical innovation, quality and design.
1.1 SYMBOLS
Within the instructions for use, on the package or on the device, there
can be the following symbols:
Symbol Meaning
Caution
General obligation
Manufacturer
3/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Symbol Meaning
Fuse
Before using the device or if you don't use it since a long time, read
these instructions carefully. Read the instructions given in the
instruction manual and reported on the device.
Before using the device, check that there is no sign of damages due
to transport or an incorrect storage, that could compromise the
correct functioning of the device.
4/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
5/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
6/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
The client will not be refunded for damages originating from the
device halt.
7/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
2 SAFETY
2.1 SAFETY WARNINGS
DANGER
Electric shock danger. Do not let water fall on the device. Do not
immerse the device in water or other liquids.
DANGER
Electric shock danger. If the power cables are damaged, they must
be replaced in an Authorized Service Centre to prevent any risk.
DANGER
Electric shock danger. Unplug the power cable from the mains
socket before disinfecting the device and before any maintenance
operation.
CAUTION
Do not use the device if visibly damaged. Periodically inspect the
device and the connection cables to verify if there are damage
signs.
CAUTION
Always keep the device out of the reach of children.
CAUTION
Danger of device falling down. Do not leave free cables which can
represent an obstacle or a danger for the patient or the operator.
CAUTION
Danger of stumbling and falling. Do not let the power cord or the
connection cables free in a place where people could walk.
8/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
CAUTION
Electric shock risk. Do not touch the power supply cables with wet
hands.
CAUTION
Electric shock risk. Do not leave the power supply cables in contact
with sharp corners or objects. Collect and attach always the power
supply cables.
CAUTION
If you notice a wired odour or smoke coming out of the device or if
it emanates heat, turn it off immediately. Do not keep using a
damaged product or a damaged part. Danger of injuries.
CAUTION
The electrical net must have a Residual-Current Circuit Breaker
(IΔn=30mA) Thermal-Magnetic Circuit Breaker (Vn=230V) to protect
the device. Place the device in such a way that the power socket is
easily accessible.
It is forbidden to carry out any technical operation on the device that
is not recalled or described in the instructions for use.
The device does not generate and does not receive any
electromagnetic interference if it is placed near other electrical
appliances. No preventive or corrective actions are required.
9/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Pos Description
A Device data plate
10/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
11/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
12/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
13/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
CAUTION
Danger of device damages. During transport and storage, the
device can be exposed to the environmental conditions for a
maximum period of 15 weeks, only if kept in the original packaging.
14/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
The user has to consider the effects potentially dangerous for the
environment and the human health originating from an improper
disposal of the whole device or its parts.
In case the user wishes to dispose of the device used at the end of its
useful life, the Manufacturer facilitates the possibility of its reuse and
the recovery and recycling of the materials contained therein. This to
prevent the release of hazardous substances into the environment and
to promote conservation of natural resources. Before disposing of the
device, it is necessary to take into consideration the European and
national regulations that order what follows:
- not to dispose as urban waste but collect it separately and address
to a firm specialized in the disposal of electrical and electronic
equipment or to the local administration in charge for waste
collection.
15/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
16/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
3 DEVICE DESCRIPTION
3.1 SUPPLY DESCRIPTION
17/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
18/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
Pos Description
A CHART PROJECTOR device
B Blocking and tilt adjustment system
C ON/OFF switch
D Predisposition for serial interface RS232
E Supply outlet
F Fuse drawer
G Focus ferrule
H Infrared receptacle
I Projector's lens
19/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Model CP1042
Model CP 1142
20/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
Model CP 2047
Model CP 2147
21/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Pos Description
A Remote control
B Signal transmitter
C Battery compartment cover
22/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
23/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Pos Description
A Base
B Pin
24/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
Pos Description
A Support
B Pin
C Arm
Pos Description
A Support
B Pin
C Arm
25/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Model CP1042
26/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
Model CP 1142
Model CP 2047
27/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Model CP 2147
28/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
4 DEVICE USE
4.1 HOW TO INSTALL THE DEVICE
CAUTION
Danger of device falling down. The device must be installed on a
horizontal and stable surface.
The device can be installed in four different ways:
- installation on table base
- installation on ophthalmic unit column
- wall mounting
- installation on stand fitting.
29/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
2 Place the pin on the arm that is on the ophthalmic unit column.
30/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
3 Place the device blocking system on the support pin and adjust
the tilt
Fig 19 - Placing the pin on the arm Fig 20 - Placing the device on the pin
3 Fasten the screw (D) with the hex wrench to block the device
on the table support.
4 Connect the power supply cable on the ophthalmic unit
column to the projector socket.
31/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Anchors and screws for wall fixing are not provided and have to be
purchased separately according to the wall characteristics. Make
sure you are using the suitable attachments.
32/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
5 Mount the arm on the support. Place the pin in the hole on the
arm
6 Place the device blocking system on the support pin.
Fig 25 - Placing the pin on the wall support Fig 26 - Placing the device on the pin
7 Adjust the tilt and fasten the screw (A) with the hex wrench to
block the device on the support.
8 Connect the power cable to the mains.
Fig 27 - Blocking system and tilt adjustment Fig 28 - Connection to the mains
33/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
CAUTION
Danger of stumbling and falling. Do not let the power cord or the
connection cables free in a place where people could walk.
CAUTION
Electric shock risk. Do not leave the power supply cables in contact
with sharp corners or objects. Collect and attach always the power
supply cables.
34/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
35/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
The distance between projector and remote control shall not exceed
150 cm.
If the LED turns on when the remote control button is pressed,
means that the signal has been emitted and the batteries are
charged.
The projector turns off automatically if it is not used for more than
10 minutes and emits an acoustic signal 30 seconds before turning
off.
36/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
2 Press the buttons (A) and (C) at the same time until the red
LED (E) turns ON.
3 Press a button between (A), (B), (C) or (D). The pressed button
will assign the remote control to the device.
4 The red LED (E) shall turn OFF.
5 If you don't press any button, after 2 seconds the red LED turns
OFF and the combination is set to 0000.
37/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Fig 35 - Buttons for the association between remote control and device
Fig 36 - Buttons 1 or 2
38/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
4 Repeat the steps 2 and 3 until you have inserted the last test.
5 Press the button NEXT after the last test.
6 Press the program button 1 or 2 for 5 seconds, until you will
obtain a beep indicating the programming end and sets the
projector in stand by again (the sequence is memorized on the
button 1 or 2 pressed at the beginning).
39/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
40/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
Fig 40 - Buttons 1 or 2
41/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
Fig 42 - Buttons 1 or 2
42/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
Fig 45 - Distance between projector and Fig 46 - Distance between patient and
projection surface projection surface
43/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
If you don't have enough space for reaching the minimum projection
distance of 290 cm, use the device in mirror projection to double the
distance.
A Projection screen
B Mirror
44/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
45/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
46/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
5 ORDINARY MAINTENANCE
5.1 SAFETY WARNINGS
DANGER
Electric shock danger. Unplug the power cable from the mains
socket before disinfecting the device and before any maintenance
operation.
CAUTION
The device does not contain any piece that requires the user's
intervention. Do not dismantle any part of the device.
47/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
48/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
49/51
INSTRUCTIONS FOR USE EYE CHARTS PROJECTOR
IFU313EN00 - 01/2021
For spare parts or accessories not included in the list, ask the
Manufacturer or the local Dealer.
5.6 TROUBLESHOOTING
Issue Cause Solution Note
If the device is
powered trough the
ophthalmic unit,
Connect the power
check the
cable of the device
Power cable not connection of the
The device does to the power supply
connected to the ophthalmic unit to
not switch ON unit Press the
socket. the power line.
ON/OFF button on
Check the
the device.
functioning of the
ophthalmic unit
fuses.
The fuses are
Replace the fuses.
blown.
Presence of dust of
Clean the surface of
grease on the
Failed image focus optical parts with a
optical parts of the
soft cloth.
device.
Contact the
The LED light does
Technical
not work
Assistance Service.
50/51
EYE CHARTS PROJECTOR INSTRUCTIONS FOR USE
IFU313EN00 - 01/2021
51/51
4/174
cso@csoitalia.it | www.csoitalia.it