0% found this document useful (0 votes)
26 views17 pages

Ds ct1051 en Co

Uploaded by

MohamedSafeer
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
26 views17 pages

Ds ct1051 en Co

Uploaded by

MohamedSafeer
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 17

Calibration

Precision digital pressure gauge


Model CPG1500
WIKA data sheet CT 10.51

For further approvals,


see page 7

Applications
■ Oil and gas industry
■ Maintenance and service facilities
■ Calibration service companies and service industry
■ Simple calibrations directly on-site
■ Pressure test

Special features
■ Measuring ranges between 0 ... 10,000 bar
[0 ... 150,000 psi] and also vacuum and absolute pressure
measuring ranges are available
■ Accuracy: up to 0.025 % (incl. calibration certificate)
■ Intrinsically safe version
■ Logger function with up to 50 measured values per
second Precision digital pressure gauge, model CPG1500
■ Communication with WIKA-Cal software via Bluetooth®

Description
The model CPG1500 precision digital pressure gauge The measuring rate can be user-defined to match the type
adopts the concept of an analogue pressure gauge and at of measurement required. Standard applications usually
the same time works with the precision of a digital calibration use three measured values per second. If required, this
instrument. The instrument thus combines the accuracy of measuring rate can also be set to 50 measured values per
digital measurement technology with the simplicity of an second.
analogue measuring instrument and features impressive Via an energy-saving function, the CPG1500 switches
performance and ease of use. automatically into “sleep” mode, in this mode the battery life
can be increased up to 2,500 hours.
The CPG1500 offers an accuracy of 0.1 % FS, 0.05 % FS or
0.025 % FS and is temperature-compensated in the range of Through the innovative menu navigation, simple operation
-10 … +50 °C [14 ... 122 °F]. Measurements can be displayed is ensured. The clear display, with integral bar graph display
in any of 26 pressure and 5 level units, or in custom units to and large text area, assists with the effective analysis of the
avoid intricate conversions. widest variety of measuring locations. With the Min./Max.
function, the minimum and maximum pressures can be
accessed immediately and saved automatically.

WIKA data sheet CT 10.51 ∙ 02/2024 Page 1 of 17

Data sheets showing similar products and accessories:


Hand test pumps, hydraulic; CPP series; see data sheet CT 91.07
Hand test pump, pneumatic; model CPP30; see data sheet CT 91.06
WIKA-Cal calibration software; see data sheet CT 95.10
Software
WIKA-Cal calibration software is available for the CPG1500. As free software solutions for data transfer, WIKA also offers
WIKA-Cal also enables, over and above PC-supported the WIKA-DCS PC software.
calibration, the management of the calibration and instrument
data in an SQL database.

Specifications

Basic information
Pressure types
Gauge pressure 0 ... 10,000 bar [0 ... 150,000 psi]
Absolute pressure 0 ... 0.25 bar abs. to 0 ... 40 bar abs.
[0 ... 0.35 psi abs. to 0 ... 400 psi abs.]
Vacuum and +/- measuring range -0.25 ... +0.25 mbar to -1 ... 40 bar
[-8 ... +8 psi to -14.5 ... 300 psi]
Accuracy 1) 2) ■ 0.025 % FS
■ 0.05 % FS
■ 0.1 % FS
■ 0.5 % FS
→ Also see table “Accuracy specifications”, page 3
Connection location Vertical mounting position, process connection lower mount
Adjustment Offset and span factor adjustable
Functions
Menu functions ■ Min./Max. alarm (visual)
■ Power-off function
■ Measuring rate
■ Mean value filter
■ Pressure rate
■ Mean value (via adjustable interval)
■ Level indication
■ Tare offset
■ Indication damping
Memory ■ Min./Max. memory
■ Integrated data logger

Mean value interval 1 ... 300 seconds, adjustable


Measuring rate Max. 50/s
Data logger 3) ■ Cyclic logger
⇒ Automatic recording of up to 1,000,000 values
■ Cycle time
⇒ Selectable from 1 ... 3,600 s in 1-second steps
⇒ Selectable with the measuring rate in the following steps: 1/s, 3/s, 10/s and 50/s
Battery status display Symbol display with 4 bars indicates the battery status in 25 % steps.
Case
Rotatability The case is rotatable by 330°.
Dimensions → See dimensions in mm [in], page 10
Ingress protection IP65
Weight Incl. batteries Approx. 680 g [1.5 lbs]
With protective case cap Approx. 820 g [1.81 lbs]
1) It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
measurement uncertainty of the reference instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range during a periodic
zero point adjustment.
2) FS = Full span = end of measuring range - start of measuring range
3) The WIKA-Cal software is required to evaluate the logger data
The logger data can be downloaded as a CSV file using the demo version of WIKA-Cal. You can use the live logger for an accurate data analysis or a direct generation of certificates by
means of the logger template of the WIKA-Cal.

WIKA data sheet CT 10.51 ∙ 02/2024 Page 2 of 17


Digital display
Display
Display range -199999 ... 199999 digits
5 ½-digit 7-segment display (incl. a large matrix area for auxiliary information)
Display resolution 4 ... 5 ½ digits; adjustable; depending on the selected pressure unit
Backlighting Adjustable via key
Bar graph 0 … 100 %, 20 individual segments, which show 5 % steps
Menu languages Adjustable via menu
■ English ■ Italian
■ German ■ Russian
■ Spanish ■ Polish
■ French
Pressure units Adjustable via menu
■ mbar ■ atm ■ mmH₂O ■ kg/cm²
■ bar ■ Torr ■ mH₂O ■ kp/cm�
■ psi ■ mmHg ■ inH₂O ■ lbf/ft²
■ Pa ■ cmHg ■ inH₂O (4 °C) ■ kN/m²
■ hPa ■ inHg ■ inH₂O (20 °C) ■ micron
■ kPa ■ inHg (0 °C) ■ inH₂O (60 °F)
■ MPa ■ inHg (60 °F) ■ ftH₂O
■ 3 x user-defined unit 1)

1) The user-defined units can only be set via the WIKA-Cal and WIKA-DCS software. The CPG1500 must be fitted with Bluetooth®.

Accuracy specifications
Accuracy 1) 2)
Gauge pressure ■ 0.025 % FS 3) 4) 1 ... 1,000 bar [15 ... 15,000 psi]
■ 0.05 % FS 3)

0.1 % FS 1 ... 2,500 bar [15 ... 50,000 psi]


0.5 % FS 4,000 ... 10,000 bar [60,000 ... 150,000 psi]
Absolute pressure ■ 0.025 % FS 3) 4) 1 ... 40 bar abs. [15 ... 500 psi abs.]
■ 0.05 % FS 3)
■ 0.1 % FS

Vacuum and +/- measuring range ■ 0.025 % FS 3) 4) -1 ... 0 bar to -1 ... 40 bar and -0.6 ... +0.6 bar
■ 0.05 % FS 3) [-14.5 ... 0 psi to -14.5 ... 300 psi, and -8 ... +8 psi]
■ 0.1 % FS

Compensated range -10 ... 50 °C [14 ... 122 °F]


Reference conditions per IEC 61298-1 5)
Ambient temperature 15 ... 25 °C [59 ... 77 °F]
Atmospheric pressure 860 ... 1,060 mbar [12.5 ... 15.4 psig]
Humidity < 95 % r. h. (non-condensing)
Mounting position Vertical mounting position, process connection facing downwards
Calibration Calibrated at 23 °C [74 °F] and in a vertical mounting position, process connection facing
downwards
1) It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
measurement uncertainty of the reference instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range during a periodic
zero point adjustment.
2) FS = Full span = end of measuring range - start of measuring range
3) At a measuring rate of ≤ 3 measured values per second
4) Only with ≥ 0 ... 1 bar up to ≤ 0 ... 1,000 bar (≥ 0 ... 15 psi up to ≤ 0 ... 14,500 psi), at reference conditions 23 °C ± 3 °C

WIKA data sheet CT 10.51 ∙ 02/2024 Page 3 of 17


Measuring ranges, gauge pressure

bar psi
0 ... 0.1 0 ... 200 0 ... 5 0 ... 2,000
0 ... 0.25 0 ... 250 0 ... 10 0 ... 3,000
0 ... 0.4 0 ... 400 0 ... 15 0 ... 5,000
0 ... 0.6 0 ... 600 0 ... 20 0 ... 6,000
0 ... 1 0 ... 700 1) 0 ... 30 0 ... 8,000
0 ... 1.6 0 ... 1,000 1) 0 ... 50 0 ... 10,000 1)
0 ... 2.5 0 ... 1,600 1) 0 ... 60 0 ... 15,000 1)
0 ... 4 0 ... 2,500 1)
0 ... 100 0 ... 20,000 1)
0 ... 6 0 ... 4,000 1)
0 ... 150 0 ... 30,000 1)
0 ... 10 0 ... 5,000 1) 0 ... 160 0 ... 40,000 1)
0 ... 16 0 ... 6,000 1)
0 ... 200 0 ... 50,000 1)
0 ... 25 0 ... 7,000 1) 2) 0 ... 300 0 ... 60,000 1)
0 ... 40 0 ... 8,000 1) 2) 0 ... 500 0 ... 75,000 1)
0 ... 60 0 ... 9,000 1) 2) 0 ... 700 0 ... 100,000 1) 2)
0 ... 100 0 ... 10,000 1) 2) 0 ... 1,000 0 ... 150,000 1) 2)
0 ... 160 - 0 ... 1,500 -
1) Not possible as flush version > 600 bar [10,000 psi]
2) Measuring range not possible with CSA approval.

Measuring ranges, absolute pressure

bar abs. psi abs.


0 ... 0.25 0 ... 6 0 ... 3.5 0 ... 60
0 ... 0.4 0 ... 7 0 ... 5 0 ... 100
0 ... 0.6 0 ... 10 0 ... 10 0 ... 150
0 ... 1 0 ... 16 0 ... 15 0 ... 200
0 ... 1.6 0 ... 20 0 ... 20 0 ... 300
0 ... 2.5 0 ... 25 0 ... 30 0 ... 500
0 ... 4 0 ... 40 0 ... 50 -

Measuring ranges, vacuum and +/- measuring ranges

bar psi
-0.25 ... +0.25 -1 ... 7 -14.5 ... 0 -8 ... +8
-0.4 ... +0.4 -1 ... 9 -14.5 ... +15 -5 ... +5
-0.6 ... +0.6 -1 ... 10 -14.5 ... 40 -3 ... +3
-1 ... 0 -1 ... 15 -14.5 ... 70 -8 ... 0
-1 ... +0.6 -1 ... 24 -14.5 ... 100 -5 ... 0
-1 ... 1.5 -1 ... 25 -14.5 ... 130 -3 ... 0
-1 ... 2.5 -1 ... 39 -14.5 ... 300 -
-1 ... 3 -1 ... 40
-1 ... 5 -

Other measuring ranges on request.

WIKA data sheet CT 10.51 ∙ 02/2024 Page 4 of 17


Further details on: Measuring ranges
Overpressure limit
3 times < 25 bar < 360 psi
2 times > 25 ... ≤ 600 bar > 360 ... ≤ 8,700 psi
1.5 times > 600 ... ≤ 1,600 bar > 8,700 ... ≤ 25,000 psi
1.3 times > 1,600 ... ≤ 6,000 bar > 25,000 ... ≤ 85,000 psi
1.1 times > 6,000 bar > 85,000 psi
Display > 110 % FS or -10 % FS
Vacuum resistance Yes

Process connection
Standard Thread sizes Possible measuring ranges
EN 837 ■ G¼B ≤ 1,000 bar [≤ 15,000 psi]
■ G½B
■ G⅜B
■ G ¼, female thread
G ½ B flush with O-ring (NBR) 2.5 ... 600 bar [50 ... 6,000 psi]
G 1 B flush with O-ring (NBR) 0.4 ... 16 bar [10 ... 200 psi]
ANSI/ASME ■ ½ NPT ≤ 1,000 bar [≤ 15,000 psi]
B1.20.1 ■ ¼ NPT
■ ½ NPT, female thread
■ ¼ NPT, female thread
■ ½ NPT, male thread
DIN 16288 M20 x 1.5 ≤ 1,000 bar [≤ 15,000 psi]
ISO 7 (DIN 2999) R½ ≤ 1,000 bar [≤ 15,000 psi]
DIN EN ISO 1179-2 G ½, male thread on G ⅛, female thread ≤ 400 bar [≤ 5,800 psi]
- M16 x 1.5, female thread, with sealing cone ≥ 100 bar [≥ 1,500 psi]
M20 x 1.5, female thread, with sealing cone 1,600 ... 10.000 bar [30,000 ... 150,000 psi]
9/16-18 UNF, female thread F 250-C 100 ... 7,000 bar [1,500 ... 100,000 psi]

Further details on: Process connection


Process connection adapter 1) ■ Without
■ G ½ to ¼ NPT
■ G ½ to ½ NPT

Sealing ■ NBR
■ EPDM

Other versions
Oil- and grease-free Per G93:2019 level D (< 220 mg/m²)
Oxygen, oil- and grease-free ■ Per G93:2019 level D (< 220 mg/m²)
■ Particle < 1,000 μm

Material
Case Aluminium die-casting, nickel-plated
Wetted parts -1 ... < 40 bar [-14.5 ... < 500 psi] Stainless steel 1.4404 (316L)
> 40 bar ... ≤ 1,000 bar ■ Process connection: Stainless steel 1.4404 (316L)
[> 500 ... ≤ 15,000 psi] ■ Sensor: Elgiloy® 2.4711

> 1,000 bar [>15,000 psi] Stainless steel 1.4534 (904L)


-1 ... 10,000 bar [-14,5 ... 150,000 psi] ■ Stainless steel 1.4435 (316L)
■ Hastelloy® 2.4819 (HC276)
■ Gold-plated

WIKA data sheet CT 10.51 ∙ 02/2024 Page 5 of 17


Further details on: Process connection
Internal pressure transmission Synthetic oil With measuring ranges up to ≤ 25 bar [≤ 360 psi]
medium
Halocarbon oil For oxygen version
1) Only possible for thread size G ½ B.

→ For drawings of process connections, see page 11

Radio standard
Bluetooth® 1)
Version Bluetooth® Low Energy 4.0
Bluetooth® Classic 2.1
Frequency range 2,400 ... 2,500 MHz
HF output power Max. 2 dBm (+ 2 dBi)
Number of channels Classic 79
Low Energy 40
Channel spacing Classic 1 MHz
Low Energy 2 MHz
Bandwidth 1 or 2 MHz
Max. transmission power max. 4 dBm / 2.51 mW
1) Requires a PC with at leastBluetooth®
2.1 interface.
Communication with iOS-enabled devices is supported up to and including iOS version 15.
Communication with Android-enabled mobile devices is supported to and including Android version 11.

Voltage supply and performance data


Battery 3 x 1.5 V AA batteries 1)
Battery voltage DC 4.95 V
Battery life Typically 2,000 … 2,500 h (without backlighting and with Bluetooth® not active)
1) The batteries included in the scope of delivery are approved for hazardous areas.
For hazardous areas, only approved models are permitted. These are listed separately in the operating instructions and in the additional operating instructions.

Operating conditions
Medium temperature range -10 ... +50 °C [14 ... 122 °F]
Lower temperature limit above the freezing point of the medium
Operating temperature -10 ... +50 °C [14 ... 122 °F]
Storage temperature range -20 ... +70 °C [-4 ... +158 °F]
Relative humidity, condensation < 95 % r. h. (non-condensing)
Permissible media All liquids and gases which are compatible with 316 stainless steel

WIKA data sheet CT 10.51 ∙ 02/2024 Page 6 of 17


Approvals
Logo Description Region
EU declaration of conformity European Union
EMC directive
EN 61326 emission (group 1, class B) and immunity (industrial environment)
Pressure equipment directive
PS > 200 bar, module A, pressure accessory
RED - radio equipment directive
EN 300 328 harmonised frequency range 2,400 … 2,500 MHz is used; Bluetooth® Classic,
max. transmission power 10 mW.
The instrument may be used without restrictions in the EU and in the countries of the EFTA.
RoHS directive
Radio Law Japan Japan
Radio approval

Federal Communications Commission (FCC) for US USA


Radio approval
- Innovation, Science and Economic Development (ISED) for Canada Canada
Radio approval
- ICASA South Africa
Radio certification
Radiocommunications Equipment Rules Australia and New Zealand
Radio approval
Australia - ABN 49 004 465 936
New Zealand - Company number 400909

Optional approvals

Logo Description Region


EU declaration of conformity European Union
ATEX directive
Hazardous areas
Ex i Zone 0 gas II 1G Ex ia IIC T4 Ga
Zone 1 connection to zone 0 gas II 1/2G Ex ia IIC T4 Ga/Gb
Zone 1 gas II 2G Ex ia IIC T4 Gb
T4 at -10 … +50 °C

IECEx International
Hazardous areas
Ex i Zone 0 gas Ex ia IIC T4 Ga
Zone 1 connection to zone 0 gas Ex ia IIC T4 Ga/Gb
Zone 1 gas Ex ia IIC T4 Gb
Zone 2 gas Ex ia IIC T4 Gc
T4 at -10 … +50 °C

WIKA data sheet CT 10.51 ∙ 02/2024 Page 7 of 17


Logo Description Region
CSA (only up to 6,000 bar [85,000 psi]) USA and Canada
Safety (e.g. electr. safety, overpressure, ...)
Class - 3631 06 - Electrical measurement and inspection equipment
Class - 3631 86 - Electrical equipment for measurement - certified according to US standards
Hazardous areas
Class - 2258 04 - PROCESS CONTROL EQUIPMENT - Intrinsically-safe, Entity - hazardous
area
Ex i Class I, Zone 0 Ex ia IIC T4 Ga
Class I, Division 1 Groups A, B, C and D T4
T4 at -10 … +50 °C

Hazardous areas
Class - 2258 84 - PROCESS CONTROL EQUIPMENT - Intrinsically-safe, Entity - hazardous
area - certified in accordance with US standards
AEx i Class I, Zone 0 AEx ia IIC T4 Ga
Class I, Division 1 Groups A, B, C and D T4
T4 at -10 … +50 °C

EAC Eurasian Economic


EMC directive Community
Hazardous areas
Ex i Zone 0 gas Ex ia IIC T4 Ga
Zone 1 connection to zone 0 gas Ex ia IIC T4 Ga/Gb
Zone 1 gas Ex ia IIC T4 Gb
T4 at -10 … +50 °C

- MChS Kazakhstan
Permission for commissioning
- PAC Ukraine Ukraine
Metrology, measurement technology
- CRN Canada
Safety (e.g. electr. safety, overpressure, ...)

Certificates
Certificate
Certificates ■ Without
■ 2.2 test report with NACE MR 0175 approval 2)
■ 2.2 test report with confirmation of manufactured according to state-of-the-art
technology
■ 3.1 inspection certificate with wetted parts approvals incl. chemical analysis
■ 2.2 test report with NACE MR 0175 approval together with 3.1 inspection certificate
with wetted parts approvals incl. chemical analysis 2)
Calibration 1) ■ 3.1 inspection certificate per DIN EN 10204
■ DAkkS calibration certificate (traceable and accredited in accordance with
ISO/IEC 17025)
Recommended calibration interval 1 year (dependent on conditions of use)
1) Calibrated in vertical mounting position with process connection facing downwards.
2) Only possible up to 1,000 bar [15,000 psi].

Patents, property rights


Patent number Description
US D 803.082 S Design patent

→ For approvals and certificates, see website

WIKA data sheet CT 10.51 ∙ 02/2024 Page 8 of 17


Safety-related characteristic values (Ex)

Permissible temperature ranges


Parameters
Ambient temperature range (Ta) -10 ≤ Ta ≤ +50 °C
Max. medium temperature range (Tm) -10 ≤ Tm ≤ +50 °C

Battery supply
Parameters
Rated voltage (3 x AA-sized batteries) DC 4.5 V
Max. voltage (sparking) DC 4.95 V

High-frequency output with Bluetooth®


Parameters
Output frequency 2,400 ... 2,500 MHz
Nominal output power 10 mW
Max. output power (under fault conditions for “ia”) 490 mW

WIKA data sheet CT 10.51 ∙ 02/2024 Page 9 of 17


Dimensions in mm [in]

CPG1500 without protective case cap

Ø100,7 [3.96] 58,9 [2.32]

14562303
Ø110,1 [4.33]
Ø114,9 [4.53]
27

G 1/2 34,7 [1.37]

Ø36 [1.42]
Ø44,4 [1.75]

CPG1500 with protective case cap

65,3 [2.57]
14546316

Ø108,4 [4.27] 5 [0.2]

.39]
Ø10 [0
120,8 [4.75]
116 [4.57]

27

G 1/2 36,8 [1.45]


Ø36 [1.42]
Ø44,4 [1.75]

WIKA data sheet CT 10.51 ∙ 02/2024 Page 10 of 17


Process connections

Threaded connection G ½ Threaded connection G ¼

14562604

14562604
13 [0.51]
25.5 [1.00]
27.5 [1.05]
32.5 [1.28]

2 [0.08]
35.5 [1.40]
3 [0.12]
20 [0.79]
G 1/4
G 1/2 Ø36 [1.42]
Ø36 [1.42] Ø44,4 [1.75]
Ø44,4 [1.75]

Threaded connection ½ NPT Threaded connection ¼ NPT

14562604

14562604
13 [0.51]
25,5 [1.00]
31,5 [1.24]
19 [0.75]

1/4 NPT
1/2 NPT
Ø36 [1.42]
Ø36 [1.42]
Ø44,4 [1.75]
Ø44,4 [1.75]

Threaded connection M16 Threaded connection M20


14562604

14562604
39,2 [1.55]

M16 M20 39,2 [1.55]


Ø36 [1.42] Ø36 [1.42]
Ø44,4 [1.75] Ø44,4 [1.75]

Threaded connection 9/16 - 18 UNF


14562604
39,2 [1.55]

9/16-18 UNF
Ø36 [1.42]
Ø44,4 [1.75]

WIKA data sheet CT 10.51 ∙ 02/2024 Page 11 of 17


Front foil
1
3 Battery status
13
2
4 Pressure indication
12 5 Matrix field
Serves as menu and secondary display
11 3 6 On/Off button
7 MENU button
4
10 Calling up the menu
8 LIGHT button
5
Turning the backlighting on and off
9 ZERO button
9 6
The current pressure value will be set to “0” (gauge) or
reference pressure (abs.)
8 7
10 Bar graph
Indicates the current pressure graphically
11 Lock symbol
1 Warning notice for: Menu button or ZERO button locked/unlocked
■ Exceeding or dropping below the pressure range 12 Bluetooth® symbol
■ Exceeding or dropping below the temperature range Bluetooth® active/inactive
■ Logger memory is more than 90 % full 13 Tare
■ Instrument error or battery status < 10 % TARE active/inactive
2 Log
Logger function active/inactive

WIKA data sheet CT 10.51 ∙ 02/2024 Page 12 of 17


WIKA-Cal calibration software
Easy and fast creation of a high-quality calibration certificate
The WIKA-Cal calibration software is used for generating
calibration certificates or logger protocols for pressure
measuring instruments and is available as a demo version for
a cost-free download.
To switch from the demo version to a licenced version, a USB
dongle with a valid licence must be purchased.
The preinstalled demo version changes automatically to
the selected version when plugging in the USB dongle and
remains available as long as the USB dongle is connected to
the PC.

■ The user is guided through the calibration or logger


process
■ Management of calibration data and instrument data
■ Intelligent preselection via SQL database
■ Menu languages: German, English, Italian, French,
Dutch, Polish, Portuguese, Romanian, Spanish, Swedish,
Russian, Greek, Japanese, Chinese
More languages are due with software updates
■ Customer-specific complete solutions possible
■ Maximum degree of automation in connection with our
CPx series

The supported instruments are continuously expanded and


even customer-specific adaptations are possible.

→ For further information, see data sheet CT 95.10

There are two licences available for WIKA-Cal in conjunction with a precision pressure measuring instrument of the
CPx series
The WIKA-Cal calibration software is available both for reading the logger data stored in the precision digital pressure gauge
as well as for online calibrations together with a PC. The scope of software functions depends on the selected licence.
Several licences can be combined on one USB dongle.

Cal-Template (demo version) Cal-Template (light version) Log-Template (full version)


Fully automatic calibration Semi-automatic calibration ■ Live measured value recording for a
certain period of time with selectable
Limitation to two measuring points No limitation of the measuring points
interval, duration and start time
approached
■ Readout of the integrated data logger of
■ Creation of 3.1 inspection certificates per DIN EN 10204 the precision digital pressure gauge
■ Calibration data can be exported to Excel® template or XML file ■ Creation of logger protocols with graphic
■ Calibration of pressure measuring instruments and/or tabular representation of the
measuring results in PDF format
■ Possibility of exporting measuring results
as CSV file
Ordering information for your enquiry for a single licence:
Is available for a cost-free download WIKA-CAL-LZ-Z-Z WIKA-CAL-ZZ-L-Z
WIKA-CAL-LZ-L-Z

WIKA data sheet CT 10.51 ∙ 02/2024 Page 13 of 17


WIKA-DCS configuration software
As a supplement to the model CPG1200 and CPG1500
digital pressure gauges, in addition to the WIKA-Cal
calibration software, the WIKA-DCS configuration software is
also available.

With the help of the software, the instruments can be easily


configured, either via the USB or Bluetooth® connection.
The configuration includes, for example, setting the indication
unit, displaying the temperature, min./max values and other
menu settings.

The software also enables the configuration, control and


storage of logger operations.
■ Configuring the instrument
■ Configuring logger sequences with the CPG1200/
CPG1500
■ Carrying out logger sequences with the CPG1200/
CPG1500
■ Downloading log files from the CPG1200/CPG1500
■ Download the logger protocols as a CSV file
■ Menu languages: German, English, French and Spanish

Minimum system requirements


Processor Intel® Pentium® 4 or AMD Athlon® 64
Software ■ Microsoft® Windows® XP with Service Pack 3
■ Windows® 7 with Service Pack 1
■ Windows® 8
■ Windows® 10
Storage 1 GB RAM and 1 GB free hard disc space (no installation possible on portable flash storage media)
Display 1024 x 768 pixel screen resolution (1280 x 800 pixel recommended) with 16-bit colour depth and 256 MB
VRAM

WIKA data sheet CT 10.51 ∙ 02/2024 Page 14 of 17


Complete test cases
The available test case is individually equipped according to your needs.

Case variants and existing recesses Test case


Recesses for standard components
Precision digital pressure gauge model CPG1500 or digital pressure gauge mode CPG1200 x
Sensor connection cable 1.1 m [3.3 ft] x
Sealing set x
Pneumatic hand test pump, model CPP10-H, CPP30 or CPP40 x
Hydraulic hand test pump, model CPP700-H or CPP1000-H x
Ordering information for enquiry DTK1X00
→ For further information, see data sheet CT 93.03

Recommended pressure generation


Hand test pump Pneumatic Hydraulic
Model CPP30 Model CPP700-H/CPP1000-H

Pressure range -0.95 ... +35 bar [-28 inHg ... +500 psi] ■ 0 ... 700 bar [0 ... 10,000 psi]
■ 0 ... 1,000 bar [0 ... 14,500 psi]

→ For further information, see data sheet CT 91.06 CT 91.07

The recommended hand test pumps and service cases are not approved for use in hazardous areas!

Test case incl. pneumatic hand test pump Test case incl. hydraulic hand test pump

WIKA data sheet CT 10.51 ∙ 02/2024 Page 15 of 17


Accessories and spare parts

Description 1) Order code


CPG-A-15-Z
Batteries - Ex-approved -C-
3 x 1.5 V AA

Protective case cap - Ex-approved -G-


For CPG1500 case

Only approved protective case caps may be used in Ex areas!


You can find further information on this in the instructions.

Sealing set
Consisting of: -D-
■ 4 x G ½ USIT seals
■ 2 x G ¼ USIT seals
■ Plastic box

- Dips (Diaphragm Impact Protection System) -O-


For M16 x 1.5
Pressure range: 1,600 ... 2,500 bar
For M16 x 1.5 -P-
Pressure range: > 4,000 bar
For M20 x 1.5 -Q-
Pressure range: 1,600 ... 2,500 bar
For M20 x 1.5 -R-
Pressure range: > 4,000 bar
For 9/16-18 UNF -S-
Pressure range: 1,600 ... 2,500 bar
For 9/16-18 UNF -T-
Pressure range: > 4,000 bar
Instrument mounting bracket 90° -1-
Material: Aluminium
Screws: stainless steel

Not approved in hazardous area for Zone 0!


Only approved in hazardous area for Zone 1 and Zone 2!

Plastic case -5-


For 1 x CPG1500 for storage and transport

The plastic case is not permitted to be used in hazardous areas!

Plastic case -6-


For 3 x CPG1500 for storage and transport

The plastic case is not permitted to be used in hazardous areas!

Plastic case
For 1 x digital pressure gauge, 1 x CPP700-H hydraulic hand test pump -N-

The plastic case is not permitted to be used in hazardous areas!

For 1 x digital pressure gauge, 1 x CPP7-H or CPP30 pneumatic hand test pump -L-

The plastic case is not permitted to be used in hazardous areas!

WIKA data sheet CT 10.51 ∙ 02/2024 Page 16 of 17


Description 1) Order code
CPG-A-15-Z
Bluetooth® USB stick -2-

May not be used in Ex areas!

Ordering information for your enquiry:


1. Order code: CPG-A-15-Z ⇓
2. Option: [ ]
1) The figures are an example and may change depending on the state of the art in design, material composition and representation

Scope of delivery
■ Precision digital pressure gauge model CPG1500
■ 3 x 1.5 V AA batteries
■ Operating instructions
■ Calibration certificate

Precision digital pressure gauge with protective case


cap, model CPG1500

Ordering information
CPG1500 / Version / Explosion protection / Instrument variant / Unit / Pressure type / Measuring range / Process connection /
Wetted parts / Version for specific media / Accuracy / Type of certificate / Communication / Data logger / protective case cap /
Software / Carrying case / Further approvals / Certificates / Additional accessories / Additional ordering information

Microsoft® and Windows® are registered trademarks of Microsoft Corporation in the United States and other countries.

The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such trademarks by WIKA is
under licence. Other brands and trademarks are those of their respective owners.

© 04/2016 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specifications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modifications to the specifications and materials.

WIKA data sheet CT 10.51 ∙ 02/2024 Page 17 of 17


02/2024 EN

WIKA Alexander Wiegand SE & Co. KG


Alexander-Wiegand-Straße 30
63911 Klingenberg/Germany
Tel. +49 9372 132-0
info@wika.de
www.wika.de

You might also like