ISERV Information for Service ISERV

Oct 13, 1998 Page 1 of 12


©
FILE TITLE Signa Contour System IS 00017

SIGNA CONTOUR 3B

1 PRODUCT DESCRIPTION AND SERVICE FEATURES


The new Signa Contour features an improved and cost reduced operator console and new V7.42 software.
Integrated Power Supply (IPS) cabinet GPS, Patient Table, SRU and Magnet Enclosure, are same as Contour3.

1-1 PRODUCT FEATURES


• New software (V/R-7.42) with new Pulse Sequences:
Fast Recovery Fast Spin Echo (FRFSE)
Rapid Frame Imaging (RFI)
Gated 2D Time-of-Flight
Enhanced 3D Fast Gradient Echo FGRE
Diffusion Weighted Echo Planar Imaging (DWEPI)
• System Performance Test - to run SNR Head, SNR Body Flex, System Gain Cal, and automatic recon
scale factor modification with 1 button push.
• Software image filter processing on host computer. GOP MIP image filter hardware eliminated.
• Image reconstruction on host computer. MEDCAM eliminated and SBAM added.
• IPG2 Introduction
• DASM switch added on FRONT PANEL ASSY for easier software loading.
• Modem @ Install program integrated to Americas systems

1-2 DISTRIBUTION TO THE FIELD


Customer purchased (New Fixed Site, Mobile/Relocateable and Upgrade Systems)

1-3 DATE OF PRODUCT INTRODUCTION


M4 sites: 4th Qtr ‘98

1-4 KEY SERVICE DIFFERENCES


• Host computer FRUs changed. The Hard disk drive sized increased to 4GB and moved to the SCSI Bay.
• MedCam Array processor eliminated and Bulk Acquisition Memory (SBAM) added for host reconstruction.
• IPG2 does not have an LED display. A new service cable is added for remote diagnostics via laptop.
• DASM power switch added on new FRONT PANEL ASSY. Relay assembly added for the switch.
• GOP MIP board (image filter hardware) is eliminating for software filter.
• Rear connector board nest labeling improvement.
• V7.4 software provides SPT “Lite” for 1-button push to run SNR checks and automatic recon scale factor
updates to coil config file.

The system software is being shipped separately from the product. The product will not arrive with the most current version of
software loaded. Paper instructions for loading the most current version are being shipped with the system. This is due to an
export issue that is trying to be resolved.

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 2 of 12
©
FILE TITLE Signa Contour System IS 00017

1-5 TECHNICAL OVERVIEW/ SPECIFICATIONS


Operator Console :
Net Weight = 252 kg. Operator console consists of 19 inch high resolution CRT display, mouse, keyboard, ICM-
Lite, host computer, peripherals, DASM and independent physiological waveform color LCD display. Separate
AC junction box provides power connections for Operator console equipment. Breaker box houses console
breakers and fuses for mobile systems.

ICM-lite :
Contour 3 Contour 3B
TPS CPU TPS CPU
BIT3 (VME) BIT3 (VME)
MEDCAM SBAM
IPG *IPG2
CERD CERD
MODULE MODULE
TYME II TYME II

*Please Note: IPG2 will begin shipping at beginning of Feb ’99. The IPG2 does not have the 7-segment led
display. Other functionality is completely the same. A debug cable has been added to the forward production
system as a standard tool to facilitate connecting a laptop to the IPG, IPG2, Cerd, or TPS CPU. Instructions for
debugging via laptop are on the rev0 and rev15 Service CD-ROM. This cable can be ordered from Service stock
for the installed base systems at Oct ’98 end. Part number for cable is 2223407.

Smart Bulk Acquisition memory (SBAM) :


SBAM board will reside in the ICM-lite chassis as the replacement of MEDCAM. SBAM board stores Bulk
Data but doesn’t have reconstruction function. SBAM also performs the pre-process of Cine and 3D DFT
with an ADSP on it. SBAM will not work in ISE chassis system (Contour/Profile), because Host CPU
(SS20) performance is not enough for host reconstruction. SBAM must use G3 revision of CERD.

Host computer :
SUN Microsystems ULTRA SPARC2. Ultra SPARC Dual CPU processor, 64 bit, 256MB CPU memory
standard expandable to 512MB. Bit3, Aurora quad serial card included on SBUS. SUN Microsystems
genuine power supply included. Solaris 2.51 or later required.

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 3 of 12
©
FILE TITLE Signa Contour System IS 00017

ULTRA2 HOST CPU FRU ASSY FOR 3 2206304 ULTRA2 HOST CPU ASSY FOR 3B 2209544
ULTRA2 HDD FRU ASSY 2206303
JUMPER PLUG FOR HDD SCSI 2195827
MOTHER BOARD N/A MOTHER BOARD N/A
HOST DIMM 2191529
ULTRA2 POWER SUPPLY 2191530 ULTRA2 POWER SUPPLY 2191530
CREATER 2D VIDEO CARD 2191528
ULTRA SPARC CPU MODULE 2191527 ULTRA SPARC CPU MODULE 2191527
BIT3 SBUS 2191531 BIT3 SBUS ASSY 2191531
SBUS PANEL 2192780 SBUS PANEL 2192780
HOST DIMM 2191529
CREATER 2D VIDEO CARD 2191528
AURORA SBUS ASSY 2191532 AURORA SBUS ASSY 2191532

HDD (FUJITSU) 2205368


SCSI CHAIN (HDD-SPA) CABLE 2207659
HOST-FDD RAINBOW COLOR CABLE 2183558 HOST-FDD RAINBOW COLOR CABLE2183558
IMAGE FILTER HARDWARE 2153366

Peripherals :
4G byte HDD, 1.44M byte FDD, 12x speed CD-ROM drive, 5.25” MOD (HITACHI) for DICOM archive, HITACHI
MOD and MAXOPTICS MOD are 100% compatible. PIONEER MOD and HITACHI MOD are NOT compatible.
5.25” MOD(PIONEER) available as a special order item or from existing upgrade. PIONEER MOD will have
Read Only feature when both the Hitachi and PIONEER are installed in the console. 4G byte HDD relocated to
SCSI bay.

DASM :
DASM-VDB (analog laser camera I/F) or DASM-LCAM (digital laser camera I/F) included. DASM is located
within SCSI tray of console. It does NOT have the case with DC power supply and the wiring harness has been
improved. A DASM power switch is added on new FRONT PANEL ASSY. DASM RELAY B’d added for new
switch.

CRT display monitor :


DOTRONIX 19 inch grayscale monitor 1280x1024, 66Hz refresh rate.
BARCO 19 inch color monitor 1280x1024, 76Hz refresh rate.
Color monitor is customer purchased option for fixed sites only.

Mouse :
Logitec PS2 3 button mouse. Same as previous Contour product.

Keyboard :
GEYMS original design keyboard. Keyboard consists of Alpha-numerical 101 key, ten key, scan control key,
intercom and volume controls. Same as previous Contour product.

Integrated Power Supply (IPS) :


IPS consists of the RF power amplifier, and Power distribution unit. Weight = approx.320 kg.(w/SDT), 290
kg.(w/o SDT). Same as previous Contour product.

RF Amplifier :

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 4 of 12
©
FILE TITLE Signa Contour System IS 00017

4 kW output quadrature drive RF power amplifier. Same as previous Contour product.

Gradient Power Supply (GPS) :


Analogic AN8280Y-TK switching GPS Half height cabinet (Same as Contour2). The GPS receives G-DATA and
clock from the TYME2 by optical fiber cables. Same as previous Contour product.

Regulated Step Down Transformer (RSDT) :


RSDT is a mandatory purchased option for Gems-Am sites. M1710CA for 60Hz and M1720DA for 50Hz. System
transformer covers input voltage 380V/415V/480V. Same as previous Contour product.

Un-interruptible Power Supply (UPS) :


UPS backup power source for operator console. System can use Exide UPS only. APC UPS is incompatible,
because power capacity is not enough. Same as previous Contour product.

2 INSTALLATION

2-1 INSTALLATION PROCESS


Installation should be done by properly trained GE Field Service Engineers or approved vendor.
Two service engineers are required to install the magnet, covers, table, cabinets and run the cables.
One service engineer is required to perform the adjustments and tests.
Special Requirements
Special installation requirements for this product are
• The console will require minimum clearance in corridor of 1000mm (39.37 inches) and door
opening of 750mm (29.53 inches).
• The cable lengths from console to IPS have been changed to 20m from the previous 30m.
• System air conditioning requirements for Equipment room are lower than previous
• System air conditioning requirements for Operator room are higher than previous
• Lifting rings have being eliminated from the VMX2 “K4” magnet. Riggers will have to use lifting
straps. Preferred vendor of straps is:
Lift-All Inc. - Atlanta, GA.
Contact Charlie Greiner
phone: 800-433-0396 ext-218 or 717-665-6821
fax: 717-665-2966
Part numbers: EET4-603 x 21ft qty: 2
F2-60 x 5.5ft qty: 2

2-3 INSTALLATION TIME


Installation duration objective is 160 man-hours plus travel. The elapsed time to install the system is 14 days
(objective).

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 5 of 12
©
FILE TITLE Signa Contour System IS 00017

2-3 INSTALLATION DOCUMENTATION

DIRECTION TITLE REV CLASS DISTRIBUTION


2189673 Signa Contour 3 Pre-Installation 0 A Tab 088/491
2208849 Signa Contour Upgrade Prerequisite Procedure 0 A Tab 088/491
2220946 GE Yokogawa Hypermedia Service CD Rom 0 M Tab 494 ***New***
2219068 GE Yokogawa Hypermedia Service CD Rom for 0 C Tab 497 ***New***
Contour Only
2124628 GE Yokogawa Hypermedia Service CD Rom 15 A Tab 583; also shipped
with product.
2154654 0.5T VMX2 Magnet and Cryogens Subsystem 4 A Tab 412; shipped w/
Delivery and Installation product; Magnet CD ROM
218500-1
2154653 0.5T VMX2 Magnet and Cryogens Subsystem 4 A Tab 412; shipped w/
product; Magnet CD ROM
218500-1
2173071 SHI Operation Manual 1 A shipped w/ product; and
Magnet CD ROM 218500-1

2-4 REQUIRED INSTALLATION TOOLS

PART # DESCRIPTION DISTRIBUTION


46-260703G3 Magnet Ramping Equipment Kit Region Tool
46-260776G3 Main Coil Power Supply Region Tool
46-260777G3 Shim Power Supply (S/C shim magnet only) Region Tool
2141701 B0 Power Supply Kit Region Tool
2101360 Power Supply Calibration Kit Region Tool
2135435 Ramp Cart/Cable Kit (750 Amp) Region Tool
46-294705G1 Universal Fill Line Kit Region Tool
46-294512P1 12’ Helium Transfer Line Air Products #FS-12-2B6
46-271137G1 Cryogen Safety Equipment Kit Region Tool
46-306734G1 Helium High Pressure Regulator and Hose Kit Region Tool
46-258150P1 Non-magnetic Cylinder Cart Region Tool
46-281609G1or Mapping Fixture for Compact & AS Compact Region Tool
46-294238G1
2134428 Metric screws for attaching map fixture Spare parts
46-251865G2 Tesla Meter - Metro Lab (use #3 probe) Region Tool
46-323012G1 Voice Synthesizer Region Tool
46-301477G2or Temperature Diode Meter Kit (see Note in New Region Tool
46-317543G2 Service Tools section)
2183710 RUO Temperature Meter Kit (K4 magnet only) Region Tool
46-306801G1 Bell Gauss Meter Region Tool
n/a Laptop Computer w/CD ROM drive Standard issue
2114559-21 Service Security Key (class M) FE order
n/a Digital Volt Meter Standard Issue
n/a Oscilloscope Tektronix Standard Issue
46-194427P373 Wrist Ground Strap Standard Issue
46-320273G3 Universal Non-Magnetic Tool Kit Local Tool
46-320273G2 Comprehensive Non-Magnetic Tool Kit Local Tool
46-255837P10 (2) RF dummy loads(Bird Attenuators) Local Tool
46-306797G1 Fogg ECG Kit Region Tool

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 6 of 12
©
FILE TITLE Signa Contour System IS 00017

46-307164G3 Grafidy II Kit (0.5T compatible) Region Tool


46-317160G1 Vector Impedance Meter Region Tool

2-6 SITE PLANNING SUPPORT

GEMS-AM GEMS-E GEMS-A


Doug Marvan Alain Bureau Yoshimi Sakurai
GE Medical Systems GE Medical Systems GE Yokogawa Medical
Americas Europe Systems
Installation Support DSE/SO/ISS - Volta Bldg Architectural & Technical
Services BP 34 78530 BUC - Support
Site Planning Group France Site Engineering Dept.
tel:414-548-4569 tel:33-1-30-70-9473 tel:81-426-48-2557
fax:414-548-4575 fax:33-1-30-70-9460 fax:81-426-56-1147

3 MAINTENANCE READINESS

3-1 SPARE PARTS


Please refer to the Renewal Parts section of the GE Yokogawa Hypermedia Service CD Rom

List of Wear Parts:


Part number Quantity Req’d Description Replacement Period
U0015AK 1 Table Oring Yearly
P9325DH 1 Table ID Connector Yearly
U0047BA 1 Cradle Ultra Tape As needed
U0052AK 2 Head Coil Oring S32 Yearly
U0005AK 2 Head Coil Oring G50 Yearly
2112207 1 O2 Sensor Assy 3 years
46-294231P5 3 MRU Battery 3 years
2204436 1 Excide UPS 3 years
2130756 1 Non-Magnetic Screw kit As needed
2130757 1 Magnetic Screw kit As needed
2130758 1 Plastic Screw kit As needed
T3487MP 1 Compressor Adsorber (Leybold) 24000 hrs
46-281034P11 1 Compressor Adsorber (Balzer) 26000 hrs
T3487RP 1 Compressor Adsorber (Sumitomo) 20000 hrs

Spare Parts Correspondents


GEMS-AM GEMS-E GEMS-A
Shelly Plehn Graham Ohme Yutaka Takasaki
GPO MR Part Analyst GPO MR Part Analyst GPO MR Part Analyst
tel:414-544-3619 tel: dc*644-9786 tel:81-426-48-2037
fax:414-544-2325 dc: *451-2037
fax:81-426-48-2928

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 7 of 12
©
FILE TITLE Signa Contour System IS 00017

3-2 SERVICE TOOLS


3-2-1 New/Replacement Service Hardware Tools
PART # DESCRIPTION COST DISTRIBUTION
2191866 Service cable (for color monitor setting) - Shipped with product
2223407 Service cable (for IPG1&2, CERD, & TPS CPU - Shipped with product
debug)

Notice: The lakeshore thermometer and low cost diode box meter used to measure the VMX1 and VMX2
magnet temps contains an additional adapter cable which was introduced some time ago. A large number of the
older meters have not been upgraded. It is possible to upgrade them as follows:

Lakeshore 208 Thermometer kit 46-301477G1 + 2129919 Interface cable = 46-301477G2


Low Cost Shield Temperature Diode Box 46-317543G1 + 2136506 Interface cable = 46-317543G2
3-2-2 Hardware Tool Correspondents
GEMS-Am
Carl Wisler
8*320-2824 (Dialcom) or
(414)-548-2824 (Direct)
carl.wisler@amermsx

3-2-3 Service Software and Security Media


PART# TITLE VER CLASS DISTRIBUTION
2224819 Signa Contour/Profile Restricted Service Software 7.41 M Tab 494 (GE FEs)
(Includes InSite software)
2224820 Signa Contour/Profile Advanced Service Software 7.41 C Tab 492 (In-house Service)
2114559-21 FE key (1year reprogramable Rainbow key) required - M Tab 437 initially; order now
to use product proprietary service software. as needed.
2114561-40 In-House key (1 year reprogramable Rainbow key) - C For Licensed In-House
required to use product proprietary service software. Customers only; order
through local service.

3-2-4 Laptop Product Software Tools


To configure the optional color monitor or to debug the IPG, IPG2, CERD, and TPS CPU you will need:

Windows 3.1 Terminal Program or Windows 95 Hyper terminal program (both are standard sw with Field laptop).
Please refer to the instructions in option book on GYMS Hypermedia CD- ROM.

3-2-5 Laptop Correspondents


Refer to the latest “GEMS Contacts” list published in LIFELine monthly field laptop publication.

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 8 of 12
©
FILE TITLE Signa Contour System IS 00017

3-2-6 InSite Tools

The Contour3B program is providingModem@Install for the Americas region. A modem kit M1000NY will be shipped free of
charge for the customer with the system. If your system was ordered after Aug1st, you will not need to order a modem kit.

Contour implements Proactive Diags to provide dialout capability should the system monitor detect a failure.
The call will be
routed to the closest Online Support Center for evaluation.
Due to the sensitive nature of the cryocooler, a dedicated phone line
for InSite must be provided upon magnet installation.
• Domestic Customer owned Modem Kit (includes US Robotics) 2211403
• International Modem Kit (no modem included) 46-301708G13
• InSite software is activated with MR Signa Contour Restricted Service Software 2224819
• Refer to direction 2195902 or 2220946 for Signa Contour/Profile InSite & Prodiag Installation
3-2-7 InSite Correspondents
GEMS-AM
Online Center
800-321-7937

3-3 APPLICATION FEATURES

3-3-1 Application Software


• Signa Contour/Profile Application Software v7.42 and later

PART# TITLE VER CLASS DISTRIBUTION


2224817 Auto Install FD 7.41 B Shipped w/product
2210914-2 Auto Label FD 7.41 B Shipped w/product
2224818 Applications CD Rom 7.41 B Shipped w/product
2227910 Patch CD Rom 7.42 B Shipped w/product
M1795SW Solaris Operating Software 2.5.1 B Shipped w/product

3-3-2 Generic Product Codes

Base Hardware - Refer to Price pages for more info and upgrade options

NAME Cat No.


Signa Contour K4 w/ B&W Monitor for FIXED site S7504KA
Signa Contour K4 w/ B&W Monitor and Phased Array for FIXED site S7504KB
Signa Contour K4 w/ Color Monitor for FIXED site S7504KC
Signa Contour K4 w/ B&W Monitor and Phased Array for FIXED site S7504KD
Signa Contour K4 Relocateable w/ B&W Monitor S7504KE
Signa Contour K4 Relocateable w/ B&W Monitor and Phased Array S7504KF
Signa Contour K4 Relocateable w/ Color Monitor S7504KG
Signa Contour K4 Relocateable w/ Color Monitor and Phased Array S7504KH
Signa Contour Upgrade Std. Magnet for Fixed site with B&W Monitor S7504SA
Signa Contour Upgrade Std. Magnet for Fixed site with B&W Monitor S7504SB
and Phased Array

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 9 of 12
©
FILE TITLE Signa Contour System IS 00017

Base Hardware Continued - Refer to Price pages for more info and upgrade options

NAME Cat No.


Signa Contour Upgrade Std. Magnet for Mobile/Relocateable with B&W S7504SC
Monitor
Signa Contour Upgrade Std. Magnet for Mobile/Relocateable with B&W S7504SD
Monitor and Phased Array
Signa Contour Upgrade Compact Magnet for Fixed site with B&W S7504SE
Monitor
Signa Contour Upgrade Compact Magnet for Fixed site with B&W S7504SF
Monitor and Phased Array
Signa Contour Upgrade Compact Magnet for Relocateable/ISO Box with S7504SG
B&W Monitor
Signa Contour Upgrade Compact Magnet for Relocateable/ISO Box with S7504SH
B&W Monitor and Phased Array

3-5 SERVICE DOCUMENTATION


Documentation List
DIRECTION TITLE REV CLASS AVAIL DISTRIBUTION
2220946 GE Yokogawa Hypermedia Proprietary Service 0 M Now Tab 494 (GE FE’s)
CD Rom **New**
2219068 GE Yokogawa Hypermedia Service CD Rom for 0 C Now Tab 497 (for in-house
Contour Only use) **New**
2124628 GE Yokogawa Hypermedia Non-Proprietary 15 A Now Tab 583; also shipped
Service CD Rom with product
2180500-3 GE Magnet Subsystems CD-ROM 1 M 4Q ‘98 Tab 181
**New**
2154653 GE 0.5T K4 Magnet & Cryogens Subsystem 4 A Now Tab 412 and shipped w/
product
2154654 GE 0.5T K4 Magnet & Cryogens Subsystem 4 A Now Tab 412 and shipped w/
Delivery and Installation product
2173071 SHI Compressor/Coldhead Operation Manual 1 A Now Shipped with product
- Contour/Profile Basic Service FMI and Service - A Now Tab 491
Note
2195902 Signa Contour/Profile InSite & Prodiag Installation 0 M Now Tab 494 (GE FE’s)

Proprietary Documentation Ordering Process


GEMS-AM: From MSX MAIL “COMMON FORMS”, enter a “TECHNICAL PUBLICATN ORDER” (Class A) or a
“proprietary publ request” (Class C, D, or M). The proprietary tracking group will review and authorize Class C,
D, and M orders.

3-6 TRAINING
Field Engineer Training
Early Readiness training not required due to the small differences from previous product release. Advanced
notice of the product release was provided to the ETC+ members in Spring and Fall sessions of ’98. Product
differences are integrated into forward training.

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 10 of 12
©
FILE TITLE Signa Contour System IS 00017

Field Engineer Training Correspondents


Contact the following individuals to sign up for service training classes:
GEMS-AM GEMS-A
Lynn Ross (GE-EC) Akiko Kakiuchi (ATI)
8*320-8824 (Dialcom) or Tel: 81 426 48 2625
(414)-574-8824 (Direct) Fax: 81 426 48 2920

Customer Training Correspondents


Contact the following individuals to sign up for customer training classes:
GEMS-AM
Betty Paschke (GE-EC)
8*320-8815(Dialcom) or
(414)574-8815 (Direct)

Applications Training Correspondents


Contact the following individuals to sign up for applications training classes:
GEMS-AM
Vicki Hansen (GE-EC)
8*320-6954 (Dialcom) or
(414)-521-6954 (Direct)

3-7 Traceability
Contact the following individuals for Product Locator information:
GEMS-AM
800-548-3366 option 3
then option 4
Mail code W523
fax 414-548-2886

4 MAINTENANCE DELIVERY
This section describes maintenance service (in and out of warranty).

4-1 Planned Maintenance


Planned maintenance (PMs) are proactive service visits intended to reduce the probability of failure during the
year. PMs instructions (procedures and frequencies) are based on engineering recommendations to meet safety,
regulatory, and image quality requirements. For the Contour product, the following PM goals have been set:
• Number of PMs/year/fixed site unit = 3
• Number of PMs/year/mobile unit = 12
• Hours of PM per visit for fixed site unit (without travel allowance) = 8 hours
• Hours of PM per visit for mobile unit (without travel allowance) = 2.5 hours
• Number of people required = 1 person

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 11 of 12
©
FILE TITLE Signa Contour System IS 00017

Planned Maintenance Continued:


• Parts to be changed during PMs:
1. The adsorber in the Sumitomo compressor should be replaced every 20,000 hours. See
wear parts section for details on Leybold and Balzer.
2. The coldhead should be used until failure.
3. The UPS should be replaced every 3 years.
4. Table oring, id connector and cradle tape should be replaced yearly.
5. Head coil orings should be replaced yearly
6. O2 sensor should be replaced every 3 years
7. MRU batteries should be replaced every 3 years
• For Proprietary PMs procedures, refer to Direction 2124629, GE Yokogawa Hypermedia Service
CD-ROM.

5 FURTHER INFORMATION
An electronic copy of this ISERV permanently resides on globe in globepro/k4magnet/iserv_Con3B.doc. For any
information regarding this ISERV, please contact:
Lead Program Integrator - GEYMS MR Programs MKE
3200 N. Grandview Blvd.
W808 MR Bldg.
Waukesha, WI 53188

office: 414-521-6141 dial comm: 320-6141


fax: 414-521-6081

©
Copyright 1997 by General Electric Company
ISERV Information for Service ISERV
Oct 13, 1998 Page 12 of 12
©
FILE TITLE Signa Contour System IS 00017

APPENDIX 1: PILOT SITES

GEMS-AM GEMS-E GEMS-A


Pilot Site MaryAnn Camacho Christian Beheuliere Hiroyuki Nogawa
Correspondent tel.:8* 320.6141 (Dialcom) tel.: 8*644.4387(Dialcom) tel.: 81.426.48.2602
tel.:414.521.6141 tel.: 33.1.30.70.43.87 fax.:81.426.56.0002
fax.:414.521.6081 fax.: 33.1.30.70.99.95 nogawa@mr.geyms.med.
camacho.mary@amermsx christian.beheuliere@med ge.com
.ge.com
Pilot Site: none none Yamanaka Ityou Hospital;
MIE Prefecture
Pilot Site: none none Nisi-Harima Hospital;
Hougo Prefecture

IMPORTANT!
Installation and site FE’s should enter labor hours and parts in the service data base immediately
after pilot site installation, customer turnover, and repairs in order to facilitate pilot tracking.

MaryAnn Camacho Christian Beheuliere Hiroyuki Nogawa


GEMS-AM GEMS-E GEMS-A

MR SERVICE ENGINEERING MR SERVICE ENGINEERING MR SERVICE ENGINEERING

©
Copyright 1997 by General Electric Company