T R A N S M I T T E R S F O R H U M I D I T Y A N D T E M P E R AT U R E
XB OEM - SERIES
THE OEM TRANSMITTER.
INTRODUCING AIRCHIP DIGITAL TECHNOLOGY.
IMPROVED PERFORMANCE IN ENVIRONMENTAL
CHAMBERS, INCUBATORS AND DRYERS
• Relative humidity and temperature measurement
• Outstanding accuracy and repeatability
• Wide choice of probes to satisfy every application
• Various analog outputs available
• Freely scalable analog signals
• Simulator mode
• Automatic humidity sensor test and drift compensation
• Pt100 RTD direct 4-wire connection option available
• 2.000 point data recording
XB OEM - SERIES
BUILD YOUR OWN OEM TRANSMITTER
XB OEM TRANSMITTER ADVANTAGES.
The ROTRONIC XB OEM-Series is the latest development in transmitters for humidity, temperature and dew point.
One outstanding feature of this innovative product is the ability to configure the transmitter to meet the requirements
of the application. Choose from a wide range of signal output types, probes, sensors and cable length.
Service-Interface
• UART (Mini-USB)
Board
• Compact size
• Easy to install
Outputs
• Two freely scalable analog outputs
• Available as 2-wire (XB20) or 3/4-wire (XB3x)
• UART service-interface
Pt100 RTD output
• Direct 4-wire connection optional available
Wide selection of probes
• PPS and steel probes with a temperature range up to
200° C available (With Hygromer HH-1R sensor up to 180 °C)
• 15mm and 25/15mm probe diameter available
• Probe lengths up to 700 mm
AirChip3000
• Compensates humidity and temperature over 30,000 reference points
• 2,000 data point memory
• Dew/frost point calculation
• Sensor auto-diagnostics and automatic correction
• Active alarming and information
• ASIC (Application Specific Integrated Circuit), micro-controller and EEPROM on one chip
XB OEM - SERIES
APPLICATIONS
The ROTRONIC XB OEM-Series of temperature and humidity transmitters is designed exclusively for the OEM market using the latest
AIRCHIP3000 technology. Digital signal processing significantly benefits humidity measurement in the following key areas:
Environmental Chambers Industrial Process Monitoring Environmental Test Drying processes
Many products are tested The maximum measurement Every product from mobile tele- Monitoring and control of any
over a variety of environmen- temperature of 200 °C and the phones to automobiles are industrial drying process can
tal conditions. The exception- robust design of ROTRONIC today subjected to thorough have significant benefits in
ally wide measuring range of probes means that even the environmental testing to en- terms of product quality, energy
the ROTRONIC XB OEM units most difficult process applica- sure reliability. Humidity and use and productivity.
means these test conditions tion can be checked with a temperature conditions are a
can be easily checked. ROTRONIC XB OEM transmitter critical part of this process.
Measurement accuracy
Digital processing of the sensor signals by the XB OEM transmitters provides more scope and greater flexibility when compensating sensor
characteristics such as linearity and temperature compensation. The ROTRONIC HYGROMER® capacitive sensors have always been the
leader in both precision and stability. With the application of digital technology, sensor performance is now further enhanced.
Calibration
Calibration and sensor data are retained permanently within each XB OEM transmitter. Software-based calibration is simple and precise;
there are no hard to reach, hard to adjust potentiometers. Multiple calibration points can be selected across the full measurement range.
Transmitters of the XB OEM-series are primarily used in environmental chamber applications that require highly accurate humidity
measurement.
Field Service made easy with the XB OEM-Series
On-site validation and maintenance of the sensors is made simple with the XB OEM transmitters.
Use a Hygropalm handheld HP22/HP23 instrument and service cable to perform the following:
• Display of the % Relative Humidity reading from the transmitter on the Hygropalm HP22/HP23
• Single point calibration of the transmitter using a reference probe attached to the HP22/HP23
• Single and multipoint calibration of the transmitter against a known reference environment
Applications
Many manufacturers of environmental chambers have chosen the ROTRONIC XB OEM transmitters. Additional applications are dryer
cabinets for clothes, complex climate controls in pasta machines, process-drying and many more.
THE OUPUTS OF THE XB OEM TRANSMITTER SERIES
The high flexibility of the XB OEM transmitter allows customer using ROTRONIC HW4 software to choose the output signal (0…1V, 0…5V,
0…10V, 0…20 mA, 4…20mA) and freely scale it with the HW4 ROTRONIC software, to any value up to the probe limits. Dew-/frost point,
Pt100 RTD direct outputs optional available.
XB20-Series XB3-Series
T+V H+V
K2: PWR K1: OUT1 K2: PWR K1: OUT1
T_OUT H_OUT K1: OUT2 K1: OUT2
~
=
K2: GND K1: GND K2: GND K1: GND
2-wire, loop powered transmitter 3-wire transmitter voltage signal 3-wire transmitter current signal
XB OEM- SERIES
TECHNICAL INFORMATION.
Specification XB20 XB3
Power supply / Connections
Circuit type 2-wire loop powered 3-wire
Supply voltage VDC 10…28 VDC , V min = 10 V + (0.02 x Load*) 15…40 VDC or 12…28 VAC
*Load resistance in ohms
Output signal type 4...20 mA 0...1V, 0...5V, 0...10V,
Dimension of the board 0...20 mA, 4...20 mA
90.05 Nominal current consumption 2 x 20 mA < 50 mA
81.5
77.5 Electrical connections Terminal block
Polarity protection Protective diode on V+
PT_S
Humidity measurement
PT_AS
Sensor Hygromer® IN-1
56
58
64
PT_AR
H_OUT
H+V
T_OUT
T+V
PT_R
Measurement range 0…100 %rh
Measurement accuracy at 23 °C ±1.0 %rh
Repeatability 0.3 %rh
Long term stability < 1 %rh / year
Temperature measurement
Sensor (Standard) Pt100 RTD, IEC 751 1/3 class B
Probe dimension Measuring range -100…200 °C (see also environmental limits)
PPS probe Ø15mm Measurement accuracy at 23 °C ±0.2 °C
Repeatability 0.05 °C
Long term stability < 0.1 °C / year
Calculated parameters
16 100/250/400/550/700
Psychrometric calculations Dew or frost point (user configurable option)
39
Environmental limits
Storage and transit -50…+70 °C, 0…100 %rh, non condensing
15
Operating limits at electronics -40… +85 °C, 0…100 %rh, non condensing
Maxi. humidity /temperature See humidity/temperature window of sensor
PPS probe Ø25/15mm at sensor
Maximum air velocity at probe 40 m/s (7,870 ft /min) – depends on dust filter type
Digital interface (service connector)
150 250/400/550/700
Interface type UART (Universal Asynchronous Receiver Transmitter)
39
Maximum service cable length 5 m (16.4 ft)
Firmware update Via service connector, requires HW4 software
15
25
Start-up time and data refresh rate
Steel probe Ø15mm Start-up time 3.4s (typical) 1.9s (typical)
Data refresh rate 3.4s (typical) 1.7s (typical)
Conformity with standards
120/270/420/570 CE / EMC immunity EMC Directive 2004/108/EG:
39 EN 61000-6-1: 2001 EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11
Solder type Lead free (RoHS directive)
15
FDA / GAMP directives compatible
Change over now to the HygroClip2 probes with integrated AirChip3000 technology for your
temperature and humidity measurements. You can find detailed information about our probes
at www.rotronic-humidity.com. We would also be happy to advise you in person or on the phone.
59023E-2010/10
ROTRONIC AG, Grindelstrasse 6, CH - 8303 Bassersdorf, Tel. +41 44 838 11 11, Fax +41 44 836 44 24, www.rotronic.ch
ROTRONIC Instruments (UK) Ltd, Crompton Fields, Crompton Way, Crawley, West Sussex, RH10 9EE, UK, Phone +44 (0)1293 571000, Fax +44 (0)1293 571008, www.rotronic.co.uk
ROTRONIC Instrument Corp, 135 Engineers Road, Hauppauge, NY 11788, USA , Phone, +1 631 427-3898, Fax, +1 631 427-3902, www.rotronic-usa.com