DC Resistance or Conductance of Insulating Materials: Standard Test Methods For
DC Resistance or Conductance of Insulating Materials: Standard Test Methods For
   1
     These test methods are under the jurisdiction of ASTM Committee D09 on
                                                                                                2
Electrical and Electronic Insulating Materials and are the direct responsibility of               For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Subcommittee D09.12 on Electrical Tests.                                                     contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
   Current edition approved Sept. 1, 2005. Published October 2005. Originally                Standards volume information, refer to the standard’s Document Summary page on
approved in 1925. Last previous edition approved in 1999 as D 257 – 99.                      the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
                                                                                         1
                                                             D 257 – 99 (2005)
   3.1.4 conductivity, surface, n—the surface conductance                 4. Summary of Test Methods
multiplied by that ratio of specimen surface dimensions (dis-               4.1 The resistance or conductance of a material specimen or
tance between electrodes divided by the width of electrodes               of a capacitor is determined from a measurement of current or
defining the current path) which transforms the measured                  of voltage drop under specified conditions. By using the
conductance to that obtained if the electrodes had formed the             appropriate electrode systems, surface and volume resistance
opposite sides of a square.                                               or conductance may be measured separately. The resistivity or
   3.1.4.1 Discussion—Surface conductivity is expressed in                conductivity can then be calculated when the required speci-
siemens. It is popularly expressed as siemens/square (the size            men and electrode dimensions are known.
of the square is immaterial). Surface conductivity is the
reciprocal of surface resistivity.                                        5. Significance and Use
   3.1.5 conductivity, volume, n—the volume conductance                      5.1 Insulating materials are used to isolate components of an
multiplied by that ratio of specimen volume dimensions                    electrical system from each other and from ground, as well as
(distance between electrodes divided by the cross-sectional               to provide mechanical support for the components. For this
area of the electrodes) which transforms the measured conduc-             purpose, it is generally desirable to have the insulation resis-
tance to that conductance obtained if the electrodes had formed           tance as high as possible, consistent with acceptable mechani-
the opposite sides of a unit cube.                                        cal, chemical, and heat-resisting properties. Since insulation
   3.1.5.1 Discussion—Volume conductivity is usually ex-                  resistance or conductance combines both volume and surface
pressed in siemens/centimetre or in siemens/metre and is the              resistance or conductance, its measured value is most useful
reciprocal of volume resistivity.                                         when the test specimen and electrodes have the same form as
   3.1.6 moderately conductive, adj—describes a solid mate-               is required in actual use. Surface resistance or conductance
rial having a volume resistivity between 1 and 10 000 000                 changes rapidly with humidity, while volume resistance or
V-cm.                                                                     conductance changes slowly although the final change may
   3.1.7 resistance, insulation, (Ri), n—the ratio of the dc              eventually be greater.
voltage applied to two electrodes (on or in a specimen) to the               5.2 Resistivity or conductivity may be used to predict,
total volume and surface current between them.                            indirectly, the low-frequency dielectric breakdown and dissi-
                                                                          pation factor properties of some materials. Resistivity or
   3.1.7.1 Discussion—Insulation resistance is the reciprocal
                                                                          contivity is often used as an indirect measure of moisture
of insulation conductance.
                                                                          content, degree of cure, mechanical continuity, and deteriora-
   3.1.8 resistance, surface, (Rs), n—the ratio of the dc voltage         tion of various types. The usefulness of these indirect measure-
applied to two electrodes (on the surface of a specimen) to the           ments is dependent on the degree of correlation established by
current between them.                                                     supporting theoretical or experimental investigations. A de-
   3.1.8.1 Discussion—(Some volume resistance is unavoid-                 crease of surface resistance may result either in an increase of
ably included in the actual measurement.) Surface resistance is           the dielectric breakdown voltage because the electric field
the reciprocal of surface conductance.                                    intensity is reduced, or a decrease of the dielectric breakdown
   3.1.9 resistance, volume, (Rv), n—the ratio of the dc voltage          voltage because the area under stress is increased.
applied to two electrodes (on or in a specimen) to the current               5.3 All the dielectric resistances or conductances depend on
in the volume of the specimen between the electrodes.                     the length of time of electrification and on the value of applied
   3.1.9.1 Discussion—Volume resistance is the reciprocal of              voltage (in addition to the usual environmental variables).
volume conductance.                                                       These must be known to make the measured value of resistance
   3.1.10 resistivity, surface, (rs), n—the surface resistance            or conductance meaningful.
multiplied by that ratio of specimen surface dimensions (width               5.4 Volume resistivity or conductivity can be used as an aid
of electrodes defining the current path divided by the distance           in designing an insulator for a specific application. The change
between electrodes) which transforms the measured resistance              of resistivity or conductivity with temperature and humidity
to that obtained if the electrodes had formed the opposite sides          may be great (1, 2, 3, 4),3 and must be known when designing
of a square.                                                              for operating conditions. Volume resistivity or conductivity
   3.1.10.1 Discussion—Surface resistivity is expressed in                determinations are often used in checking the uniformity of an
ohms. It is popularly expressed also as ohms/square (the size of          insulating material, either with regard to processing or to detect
the square is immaterial). Surface resistivity is the reciprocal of       conductive impurities that affect the quality of the material and
surface conductivity.                                                     that may not be readily detectable by other methods.
                                                                             5.5 Volume resistivities above 1021 V·cm (1019 V·m), ob-
   3.1.11 resistivity, volume, (rv), n—the volume resistance
                                                                          tained on specimens under usual laboratory conditions, are of
multiplied by that ratio of specimen volume dimensions
                                                                          doubtful validity, considering the limitations of commonly
(cross-sectional area of the specimen between the electrodes
                                                                          used measuring equipment.
divided by the distance between electrodes) which transforms
                                                                             5.6 Surface resistance or conductance cannot be measured
the measured resistance to that resistance obtained if the
                                                                          accurately, only approximated, because some degree of volume
electrodes had formed the opposite sides of a unit cube.
   3.1.11.1 Discussion—Volume resistivity is usually ex-
pressed in ohm-centimetres (preferred) or in ohm-metres.                     3
                                                                               The boldface numbers in parentheses refer to the list of references appended to
Volume resistivity is the reciprocal of volume conductivity.              these test methods.
                                                                      2
                                                            D 257 – 99 (2005)
resistance or conductance is always involved in the measure-
ment. The measured value is also affected by the surface
contamination. Surface contamination, and its rate of accumu-
lation, is affected by many factors including electrostatic
charging and interfacial tension. These, in turn, may affect the
surface resistivity. Surface resistivity or conductivity can be
considered to be related to material properties when contami-
nation is involved but is not a material property in the usual
sense.
6. Electrode Systems
   6.1 The electrodes for insulating materials should be of a
material that is readily applied, allows intimate contact with the
specimen surface, and introduces no appreciable error because
of electrode resistance or contamination of the specimen (5).
The electrode material should be corrosion-resistant under the
conditions of test. For tests of fabricated specimens such as
feed-through bushings, cables, etc., the electrodes employed
are a part of the specimen or its mounting. Measurements of
insulation resistance or conductance, then, include the contami-
nating effects of electrode or mounting materials and are
generally related to the performance of the specimen in actual
use.
   6.1.1 Binding-Post and Taper-Pin Electrodes, Fig. 1 and
Fig. 2, provide a means of applying voltage to rigid insulating
materials to permit an evaluation of their resistive or conduc-
tive properties. These electrodes simulate to some degree the
actual conditions of use, such as binding posts on instrument
panels and terminal strips. In the case of laminated insulating
                                                                                           FIG. 2 Taper-Pin Electrodes
materials having high-resin-content surfaces, somewhat lower
insulation resistance values may be obtained with taper-pin
than with binding posts, due to more intimate contact with the
body of the insulating material. Resistance or conductance
values obtained are highly influenced by the individual contact
between each pin and the dielectric material, the surface
roughness of the pins, and the smoothness of the hole in the
dielectric material. Reproducibility of results on different
specimens is difficult to obtain.
   6.1.2 Metal Bars in the arrangement of Fig. 3 were prima-
rily devised to evaluate the insulation resistance or conduc-
tance of flexible tapes and thin, solid specimens as a fairly
simple and convenient means of electrical quality control. This
arrangement is somewhat more satisfactory for obtaining
                                                                     3
                                                                          D 257 – 99 (2005)
                                                                                        trodes may be obtained with a fine-bristle brush. However, for
                                                                                        circular electrodes, sharper edges can be obtained by the use of
                                                                                        a ruling compass and silver paint for drawing the outline circles
                                                                                        of the electrodes and filling in the enclosed areas by brush. A
                                                                                        narrow strip of masking tape may be used, provided the
                                                                                        pressure-sensitive adhesive used does not contaminate the
                                                                                        surface of the specimen. Clamp-on masks also may be used if
                                                                                        the electrode paint is sprayed on.
                                                                                           6.1.4 Sprayed Metal, Fig. 4, Fig. 5, and Fig. 6, may be used
                                                                                        if satisfactory adhesion to the test specimen can be obtained.
                                                                                        Thin sprayed electrodes may have certain advantages in that
                                                                                        they are ready for use as soon as applied. They may be
                                                                                        sufficiently porous to allow the specimen to be conditioned, but
                                                                                        this should be verified. Narrow strips of masking tape or
                                                                                        clamp-on masks must be used to produce a gap between the
                                                                                        guarded and the guard electrodes. The tape shall be such as not
                                                                                        to contaminate the gap surface.
                                                                                           6.1.5 Evaporated Metal may be used under the same con-
                                                                                        ditions given in 6.1.4.
                                                                                           6.1.6 Metal Foil, Fig. 4, may be applied to specimen
                                                                                        surfaces as electrodes. The usual thickness of metal foil used
                                                                                        for resistance or conductance studies of dielectrics ranges from
                                                                                        6 to 80 µm. Lead or tin foil is in most common use, and is
                                                                                        usually attached to the test specimen by a minimum quantity of
 Volume Resistivity g |Ls 2t Surface Resistivity                                        petrolatum, silicone grease, oil, or other suitable material, as an
    FIG. 4 Flat Specimen for Measuring Volume and Surface
                                                                                        adhesive. Such electrodes shall be applied under a smoothing
                  Resistances or Conductances
                                                                                        pressure sufficient to eliminate all wrinkles, and to work excess
                                                                                        adhesive toward the edge of the foil where it can be wiped off
                                                                                        with a cleansing tissue. One very effective method is to use a
                                                                                        hard narrow roller (10 to 15 mm wide), and to roll outward on
                                                                                        the surface until no visible imprint can be made on the foil with
                                                                                        the roller. This technique can be used satisfactorily only on
                                                                                        specimens that have very flat surfaces. With care, the adhesive
                                                                                        film can be reduced to 2.5 µm. As this film is in series with the
                                                                                        specimen, it will always cause the measured resistance to be
                                                                                        too high. This error may become excessive for the lower-
                                                                                        resistivity specimens of thickness less than 250 µm. Also the
                                                                                        hard roller can force sharp particles into or through thin films
                                                                                        (50 µm). Foil electrodes are not porous and will not allow the
                                                                                        test specimen to condition after the electrodes have been
                                                                                        applied. The adhesive may lose its effectiveness at elevated
                                                                                        temperatures necessitating the use of flat metal back-up plates
                                                                                        under pressure. It is possible, with the aid of a suitable cutting
                                                                                        device, to cut a proper width strip from one electrode to form
                                                                                        a guarded and guard electrode. Such a three-terminal specimen
                                                                                        normally cannot be used for surface resistance or conductance
                                                                                        measurements because of the grease remaining on the gap
                                                                                        surface. It may be very difficult to clean the entire gap surface
 D0 = (D1 + D2)/2 L > 4t g |La 2t Volume Resistivity g |Ls 2t Surface Resistivity
  FIG. 5 Tubular Specimen for Measuring Volume and Surface
                                                                                        without disturbing the adjacent edges of the electrode.
                 Resistances or Conductances                                               6.1.7 Colloidal Graphite, Fig. 4, dispersed in water or other
                                                                                        suitable vehicle, may be brushed on nonporous, sheet insulat-
allow the test specimen to be conditioned after the application                         ing materials to form an air-drying electrode. Masking tapes or
of the electrodes. This is a particularly useful feature in                             clamp-on masks may be used (6.1.4). This electrode material is
studying resistance-humidity effects, as well as change with                            recommended only if all of the following conditions are met:
temperature. However, before conductive paint is used as an                                6.1.7.1 The material to be tested must accept a graphite
electrode material, it should be established that the solvent in                        coating that will not flake before testing,
the paint does not attack the material so as to change its                                 6.1.7.2 The material being tested must not absorb water
electrical properties. Reasonably smooth edges of guard elec-                           readily, and
                                                                                    4
                                                                       D 257 – 99 (2005)
                                                                                 5
                                                                           D 257 – 99 (2005)
                                                                                       constant, K, (equivalent to the A/t factor from Table 1) can be
                                                                                       derived from the following equation:
                                                                                                                    K 5 3.6 p C 5 11.3 C                           (1)
                                                                                       where:
                                                                                       K has units of centimetres, and
                                                                                       C has units of picofarads and is the capacitance of the electrode system with
                                                                                          air as the dielectric. See Test Methods D 150 for methods of measurement
                                                                                          for C.
                                                                                   6
                                                           D 257 – 99 (2005)
7. Choice of Apparatus and Test Method
   7.1 Power Supply—A source of very steady direct voltage is
required (see X1.7.3). Batteries or other stable direct voltage
supplies may be used.
   7.2 Guard Circuit—Whether measuring resistance of an
insulating material with two electrodes (no guard) or with a
three-terminal system (two electrodes plus guard), consider
how the electrical connections are made between the test
instrument and the test sample. If the test specimen is at some
distance from the test instrument, or the test specimen is tested
under humid conditions, or if a relatively high (1010 to 1015
ohms) specimen resistance is expected, spurious resistance
paths can easily exist between the test instrument and test
specimen. A guard circuit is necessary to minimize interference
from these spurious paths (see also X1.9).
   7.2.1 With Guard Electrode—Use coaxial cable, with the
core lead to the guarded electrode and the shield to the guard
electrode, to make adequate guarded connections between the
test equipment and test specimen. Coaxial cable (again with the
shield tied back to the guard) for the unguarded lead is not
mandatory here (or in 7.2.2), although its use provides some
reduction in background noise (see also Fig. 8).
   7.2.2 Without Guard Electrode—Use coaxial cable, with the
core lead to one electrode and the shield terminated about 1 cm            FIG. 9 Connections to Unguarded Electrodes for Unguarded
from the end of the core lead (see also Fig. 9).                                            Surface Measurements
   7.3 Direct Measurements—The current through a specimen
at a fixed voltage may be measured using any equipment that             nometers. Typical methods and circuits are given in Appendix
has the required sensitivity and accuracy (610 % is usually             X3. When the measuring device scale is calibrated to read
adequate). Current-measuring devices available include elec-            ohms directly no calculations are required.
trometers, d-c amplifiers with indicating meters, and galva-               7.4 Comparison Methods—A Wheatstone-bridge circuit
                                                                        may be used to compare the resistance of the specimen with
                                                                        that of a standard resistor (see Appendix X3).
                                                                           7.5 Precision and Bias Considerations:
                                                                           7.5.1 General—As a guide in the choice of apparatus, the
                                                                        pertinent considerations are summarized in Table 2, but it is not
                                                                        implied that the examples enumerated are the only ones
                                                                        applicable. This table is not intended to indicate the limits of
                                                                        sensitivity and error of the various methods per se, but rather
                                                                        is intended to indicate limits that are distinctly possible with
                                                                        modern apparatus. In any case, such limits can be achieved or
                                                                        exceeded only through careful selection and combination of the
                                                                        apparatus employed. It must be emphasized, however, that the
                                                                        errors considered are those of instrumentation only. Errors such
                                                                        as those discussed in Appendix X1 are an entirely different
                                                                        matter. In this latter connection, the last column of Table 2 lists
                                                                        the resistance that is shunted by the insulation resistance
                                                                        between the guarded electrode and the guard system for the
                                                                        various methods. In general, the lower such resistance, the less
                                                                        probability of error from undue shunting.
                                                                          NOTE 2—No matter what measurement method is employed, the
                                                                        highest precisions are achieved only with careful evaluation of all sources
                                                                        of error. It is possible either to set up any of these methods from the
                                                                        component parts, or to acquire a completely integrated apparatus. In
                                                                        general, the methods using high-sensitivity galvanometers require a more
                                                                        permanent installation than those using indicating meters or recorders. The
                                                                        methods using indicating devices such as voltmeters, galvanometers, d-c
                                                                        amplifiers, and electrometers require the minimum of manual adjustment
FIG. 8 Connections to Guarded Electrode for Volume and Surface          and are easy to read but the operator is required to make the reading at a
  Resistivity Measurements (Volume Resistance hook-up shown)            particular time. The Wheatstone bridge (Fig. X1.4) and the potentiometer
                                                                    7
                                                                      D 257 – 99 (2005)
                                                    TABLE 2 Apparatus and Conditions for Use
                                                                                                                                     Ohms Shunted by
                                                   Reference                      Maximum Ohms   Maximum Ohms                      Insulation Resistance
                                                                                                                     Type of
               Method                                                               Detectable    Measurable to                        from Guard to
                                                                                                                   Measurement
                                         Section             Figure                  at 500 V    66 % at 500 V                            Guarded
                                                                                                                                          Electrode
Voltmeter-ammeter (galvanometer)          X3.1         X1                          1012               1011        deflection           10 to 105
Comparison (galvanometer)                 X3.4         X3                          1012               1011        deflection           10 to 105
Voltmeter-ammeter (dc amplifica-          X3.2         X2(a)                                                      deflection           102 to 109
  tion, electrometer)                                  (Position 1)                1015               1013
                                                       X2(a)                                                      deflection           102 to 103
                                                       (Position 2)                1015               1013        deflection           103 to 1011
                                                       X2(b)                       1017               1015        null                 0 (effective)
                                                       X2(b)                       1017               1015
Comparison (Wheatstone bridge)           X3.5          X4                          1015               1014        null                 105 to 106
Voltage rate-of-change                   X3.3          X5                          ;100 MV·F                      deflection           unguarded
Megohmmeter (typical)                   commercial instruments                     1015               1014        direct-reading       104 to 1010
method (Fig. X1.2 (b)) require the undivided attention of the operator in          available. If 10-mV input to the amplifier or electrometer gives
keeping a balance, but allow the setting at a particular time to be read at        full-scale deflection with an error not greater than 2 % of full
leisure.
                                                                                   scale, with 500 V applied, a resistance of 5000 TV can be
   7.5.2 Direct Measurements:                                                      measured with a maximum error of 6 % when the voltmeter
   7.5.2.1 Galvanometer-Voltmeter—The maximum percent-                             reads full scale, and 10 % when it reads 1⁄3 scale.
age error in the measurement of resistance by the                                     7.5.2.3 Comparison-Galvanometer—The maximum per-
galvanometer-voltmeter method is the sum of the percentage                         centage error in the computed resistance or conductance is
errors of galvanometer indication, galvanometer readability,                       given by the sum of the percentage errors in Rs, the galvanom-
and voltmeter indication. As an example: a galvanometer                            eter deflections or amplifier readings, and the assumption that
having a sensitivity of 500 pA/scale division will be deflected                    the current sensitivities are independent of the deflections. The
25 divisions with 500 V applied to a resistance of 40 GV                           latter assumption is correct to well within 62 % over the useful
(conductance of 25 pS). If the deflection can be read to the                       range (above 1⁄10 full-scale deflection) of a good, modern
nearest 0.5 division, and the calibration error (including Ayrton                  galvanometer (probably 1⁄3 scale deflection for a dc current
Shunt error) is 62 % of the observed value, the resultant                          amplifier). The error in Rs depends on the type of resistor used,
galvanometer error will not exceed 64 %. If the voltmeter has
                                                                                   but resistances of 1 MV with a limit of error as low as 0.1 %
an error of 62 % of full scale, this resistance can be measured
                                                                                   are available. With a galvanometer or d-c current amplifier
with a maximum error of 66 % when the voltmeter reads full
                                                                                   having a sensitivity of 10 nA for full-scale deflection, 500 V
scale, and 610 % when it reads one-third full scale. The
                                                                                   applied to a resistance of 5 TV will produce a 1 % deflection.
desirability of readings near full scale are readily apparent.
                                                                                   At this voltage, with the preceding noted standard resistor, and
   7.5.2.2 Voltmeter-Ammeter—The maximum percentage er-
                                                                                   with Fs = 105, ds would be about half of full-scale deflection,
ror in the computed value is the sum of the percentage errors
                                                                                   with a readability error not more than 61 %. If dx is approxi-
in the voltages, Vx and Vs, and the resistance, Rs. The errors in
                                                                                   mately 1⁄4 of full-scale deflection, the readability error would
Vs and Rs are generally dependent more on the characteristics
                                                                                   not exceed 64 %, and a resistance of the order of 200 GV
of the apparatus used than on the particular method. The most
                                                                                   could be measured with a maximum error of 651⁄2 %.
significant factors that determine the errors in Vs are indicator
errors, amplifier zero drift, and amplifier gain stability. With                      7.5.2.4 Voltage Rate-of-Change—The accuracy of the mea-
modern, well-designed amplifiers or electrometers, gain stabil-                    surement is directly proportional to the accuracy of the
ity is usually not a matter of concern. With existing techniques,                  measurement of applied voltage and time rate of change of the
the zero drift of direct voltage amplifiers or electrometers                       electrometer reading. The length of time that the electrometer
cannot be eliminated but it can be made slow enough to be                          switch is open and the scale used should be such that the time
relatively insignificant for these measurements. The zero drift                    can be measured accurately and a full-scale reading obtained.
is virtually nonexistent for carefully designed converter-type                     Under these conditions, the accuracy will be comparable with
amplifiers. Consequently, the null method of Fig. X1.2 (b) is                      that of the other methods of measuring current.
theoretically less subject to error than those methods employ-                        7.5.2.5 Comparison Bridge—When the detector has ad-
ing an indicating instrument, provided, however, that the                          equate sensitivity, the maximum percentage error in the com-
potentiometer voltage is accurately known. The error in Rs is to                   puter resistance is the sum of the percentage errors in the arms,
some extent dependent on the amplifier sensitivity. For mea-                       A, B, and N. With a detector sensitivity of 1 mV/scale division,
surement of a given current, the higher the amplifier sensitivity,                 500 V applied to the bridge, and RN = 1 GV, a resistance of
the greater likelihood that lower valued, highly precise wire-                     1000 TV will produce a detector deflection of one scale
wound standard resistors can be used. Such amplifiers can be                       division. Assuming negligible errors in RA and RB, with RN = 1
obtained. Standard resistances of 100 GV known to 62 %, are                        GV known to within 62 % and with the bridge balanced to one
                                                                              8
                                                            D 257 – 99 (2005)
detector-scale division, a resistance of 100 TV can be mea-              the measurement (this is particularly important for high-input-
sured with a maximum error of 66 %.                                      impedance instruments, such as electrometers). If the gap is
                                                                         made equal to twice the specimen thickness, as suggested in
8. Sampling                                                              9.3.3, so that the specimen can be used also for surface
   8.1 Refer to applicable materials specifications for sam-             resistance or conductance determinations, the effective area of
pling instructions.                                                      electrode No. 1 can be taken, usually with sufficient accuracy,
                                                                         as extending to the center of the gap. If, under special
9. Test Specimens                                                        conditions, it becomes desirable to determine a more accurate
   9.1 Insulation Resistance or Conductance Determination:               value for the effective area of electrode No. 1, the correction
   9.1.1 The measurement is of greatest value when the speci-            for the gap width can be obtained from Appendix X2. Elec-
men has the form, electrodes, and mounting required in actual            trode No. 3 may have any shape provided that it extends at all
use. Bushings, cables, and capacitors are typical examples for           points beyond the inner edge of electrode No. 2 by at least
which the test electrodes are a part of the specimen and its             twice the specimen thickness.
normal mounting means.                                                      9.2.4 For tubular specimens, electrode No. 1 should encircle
   9.1.2 For solid materials, the test specimen may be of any            the outside of the specimen and its axial length should be at
practical form. The specimen forms most commonly used are                least four times the specimen wall thickness. Considerations
flat plates, tapes, rods, and tubes. The electrode arrangements          regarding the gap width are the same as those given in 9.2.3.
of Fig. 2 may be used for flat plates, rods, or rigid tubes whose        Electrode No. 2 consists of an encircling electrode at each end
inner diameter is about 20 mm or more. The electrode                     of the tube, the two parts being electrically connected by
arrangement of Fig. 3 may be used for strips of sheet material           external means. The axial length of each of these parts should
or for flexible tape. For rigid strip specimens the metal support        be at least twice the wall thickness of the specimen. Electrode
may not be required. The electrode arrangements of Fig. 6 may            No. 3 must cover the inside surface of the specimen for an axial
be used for flat plates, rods, or tubes. Comparison of materials         length extending beyond the outside gap edges by at least twice
when using different electrode arrangements is frequently                the wall thickness. The tubular specimen (Fig. 5) may take the
inconclusive and should be avoided.                                      form of an insulated wire or cable. If the length of electrode is
   9.2 Volume Resistance or Conductance Determination:                   more than 100 times the thickness of the insulation, the effects
   9.2.1 The test specimen may have any practical form that              of the ends of the guarded electrode become negligible, and
allows the use of a third electrode, when necessary, to guard            careful spacing of the guard electrodes is not required. Thus,
against error from surface effects. Test specimens may be in the         the gap between electrodes No. 1 and No. 2 may be several
form of flat plates, tapes, or tubes. Fig. 4 and Fig. 7 illustrate       centimetres to permit sufficient surface resistance between
the application and arrangement of electrodes for plate or sheet         these electrodes when water is used as electrode No. 1. In this
specimens. Fig. 5 is a diametral cross section of three elec-            case, no correction is made for the gap width.
trodes applied to a tubular specimen, in which electrode No. 1
                                                                            9.3 Surface Resistance or Conductance Determination:
is the guarded electrode, electrode No. 2 is a guard electrode
consisting of a ring at each end of electrode No. 1, and                    9.3.1 The test specimen may be of any practical form
electrode No. 3 is the unguarded electrode (7, 8). For materials         consistent with the particular objective, such as flat plates,
that have negligible surface leakage, the guard rings may be             tapes, or tubes.
omitted. Convenient and generally suitable dimensions appli-                9.3.2 The arrangements of Fig. 2 and Fig. 3 were devised for
cable to Fig. 4 in the case of test specimens that are 3 mm in           those cases where the volume resistance is known to be high
thickness are as follows: D3 = 100 mm, D2 = 88 mm, and                   relative to that of the surface (2). However, the combination of
D1 = 76 mm, or alternatively, D3 = 50 mm, D2 = 38 mm, and                molded and machined surfaces makes the result obtained
D1 = 25 mm. For a given sensitivity, the larger specimen                 generally inconclusive for rigid strip specimens. The arrange-
allows more accurate measurements on materials of higher                 ment of Fig. 3 is somewhat more satisfactory when applied to
resistivity.                                                             specimens for which the width is much greater than the
   9.2.2 Measure the average thickness of the specimens in               thickness, the cut edge effect thus tending to become relatively
accordance with one of the methods in Test Methods D 374                 small. Hence, this arrangement is more suitable for testing thin
pertaining to the material being tested. The actual points of            specimens such as tape, than for testing relatively thicker
measurement shall be uniformly distributed over the area to be           specimens. The arrangements of Fig. 2 and Fig. 3 should never
covered by the measuring electrodes.                                     be used for surface resistance or conductance determinations
   9.2.3 It is not necessary that the electrodes have the circular       without due considerations of the limitations noted previously.
symmetry shown in Fig. 4 although this is generally conve-                  9.3.3 The three electrode arrangements of Fig. 4, Fig. 6 and
nient. The guarded electrode (No. 1) may be circular, square, or         Fig. 7 may be used for purposes of material comparison. The
rectangular, allowing ready computation of the guarded elec-             resistance or conductance of the surface gap between elec-
trode area for volume resistivity or conductivity determination          trodes No. 1 and No. 2 is determined directly by using
when such is desired. The diameter of a circular electrode, the          electrode No. 1 as the guarded electrode, electrode No. 3 as the
side of a square, or the shortest side of a rectangular electrode,       guard electrode, and electrode No. 2 as the unguarded electrode
should be at least four times the specimen thickness. The gap            (7, 8). The resistance or conductance so determined is actually
width should be great enough so that the surface leakage                 the resultant of the surface resistance or conductance between
between electrodes No. 1 and No. 2 does not cause an error in            electrodes No. 1 and No. 2 in parallel with some volume
                                                                     9
                                                           D 257 – 99 (2005)
resistance or conductance between the same two electrodes.                  12.3.1 Measure the electrode dimensions and the distance
For this arrangement the surface gap width, g, should be                 between the electrodes, g. Measure the surface resistance or
approximately twice the specimen thickness, t, except for thin           conductance between electrodes No. 1 and 2 with a suitable
specimens, where g may be much greater than twice the                    device having the required sensitivity and accuracy. Unless
material thickness.                                                      otherwise specified, the time of electrification shall be 60 s, and
   9.3.4 Special techniques and electrode dimensions may be              the applied direct voltage shall be 500 6 5 V.
required for very thin specimens having such a low volume                   12.3.2 When the electrode arrangement of Fig. 3 is used, P
resistivity that the resultant low resistance between the guarded        is taken as the perimeter of the cross section of the specimen.
electrode and the guard system would cause excessive error.              For thin specimens, such as tapes, this perimeter effectively
   9.4 Liquid Insulation Resistance—The sampling of liquid               reduces to twice the specimen width.
insulating materials, the test cells employed, and the methods              12.3.3 When the electrode arrangements of Fig. 6 are used
of cleaning the cells shall be in accordance with Test Method            (and the volume resistance is known to be high compared to the
D 1169.                                                                  surface resistance), P is taken to be the length of the electrodes
                                                                         or circumference of the cylinder.
10. Specimen Mounting
                                                                         13. Calculation
   10.1 In mounting the specimens for measurements, it is                  13.1 Calculate the volume resistivity, rv, and the volume
important that there shall be no conductive paths between the            conductivity, gv, using the equations in Table 1.
electrodes or between the measuring electrodes and ground that             13.2 Calculate the surface resistivity, rs, and the surface
will have a significant effect on the reading of the measuring           conductivity, gs, using the equations in Table 1.
instrument (9). Insulating surfaces should not be handled with
bare fingers (acetate rayon gloves are recommended). For                 14. Report
referee tests of volume resistivity or conductivity, the surfaces           14.1 Report the following information:
should be cleaned with a suitable solvent before conditioning.              14.1.1 A description and identification of the material
When surface resistance is to be measured, the surfaces should           (name, grade, color, manufacturer, etc.),
be cleaned or not cleaned as specified or agreed upon.                      14.1.2 Shape and dimensions of the test specimen,
                                                                            14.1.3 Type and dimensions of electrodes,
11. Conditioning                                                            14.1.4 Conditioning of the specimen (cleaning, predrying,
  11.1 The specimens shall be conditioned in accordance with             hours at humidity and temperature, etc.),
Practice D 618.                                                             14.1.5 Test conditions (specimen temperature, relative hu-
  11.2 Circulating-air environmental chambers or the methods             midity, etc., at time of measurement),
described in Practices E 104 or D 5032 may be used for                      14.1.6 Method of measurement (see Appendix X3),
controlling the relative humidity.                                          14.1.7 Applied voltage,
                                                                            14.1.8 Time of electrification of measurement,
12. Procedure                                                               14.1.9 Measured values of the appropriate resistances in
                                                                         ohms or conductances in siemens,
  12.1 Insulation Resistance or Conductance—Properly
                                                                            14.1.10 Computed values when required, of volume resis-
mount the specimen in the test chamber. If the test chamber and
                                                                         tivity in ohm-centimetres, volume conductivity in siemens per
the conditioning chamber are the same (recommended proce-
                                                                         centimetre, surface resistivity in ohms (per square), or surface
dure), the specimens should be mounted before the condition-
                                                                         conductivity in siemens (per square), and
ing is started. Make the measurement with a suitable device
                                                                            14.1.11 Statement as to whether the reported values are
having the required sensitivity and accuracy (see Appendix ).
                                                                         “apparent” or “steady-state.”
Unless otherwise specified, the time of electrification shall be
60 s and the applied direct voltage shall be 500 6 5 V.                  15. Precision and Bias
  12.2 Volume Resistivity or Conductivity—Measure the di-                  15.1 Precision and bias are inherently affected by the choice
mensions of the electrodes and width of guard gap, g. Make the           of method, apparatus, and specimen. For analysis and details
measurement with a suitable device having the required                   see Sections 7 and 9, and particularly 7.5.1-7.5.2.5.
sensitivity and accuracy. Unless otherwise specified, the time
of electrification shall be 60 s, and the applied direct voltage         16. Keywords
shall be 500 6 5 V.                                                        16.1 DC resistance; insulation resistance; surface resistance;
  12.3 Surface Resistance or Conductance:                                surface resistivity; volume resistance; volume resistivity
                                                                    10
                                                           D 257 – 99 (2005)
                                                           APPENDIXES
(Nonmandatory Information)
   X1.1 Inherent Variation in Materials—Because of the                      NOTE X1.1—The resistance of an electrical insulating material may be
variability of the resistance of a given specimen under similar          affected by the time of temperature exposure. Therefore, equivalent
test conditions and the nonuniformity of the same material               temperature conditioning periods are essential for comparative measure-
                                                                         ments.
from specimen to specimen, determinations are usually not                   NOTE X1.2—If the insulating material shows signs of deterioration after
reproducible to closer than 10 % and often are even more                 conditioning at elevated temperatures, this information must be included
widely divergent (a range of values from 10 to 1 may be                  with the test data.
obtained under apparently identical conditions).
                                                                            X1.3 Temperature and Humidity—The insulation resis-
  X1.2 Temperature—The resistance of electrical insulating               tance of solid dielectric materials decreases both with increas-
materials is known to change with temperature, and the                   ing temperature as described in X1.2 and with increasing
variation often can be represented by a function of the form:            humidity (1, 2, 3, 4). Volume resistance is particularly sensitive
(18)                                                                     to temperature changes, while surface resistance changes
                                                                         widely and very rapidly with humidity changes (2, 3). In both
                            R 5 Bem/T                      (X1.1)
                                                                         cases the change is exponential. For some materials a change
 where:                                                                  from 25 to 100°C may change insulation resistance or conduc-
 R = resistance (or resistivity) of an insulating material or            tance by a factor of 100 000, often due to the combined effects
        system,                                                          of temperature and moisture content change; the effect of
 B = proportionality constant,                                           temperature change alone is usually much smaller. A change
 m = activation constant, and                                            from 25 to 90 % relative humidity may change insulation
 T = absolute temperature in kelvin (K).                                 resistance or conductance by as much as a factor of 1 000 000
   This equation is a simplified form of the Arrhenius equation          or more. Insulation resistance or conductance is a function of
relating the activation energy of a chemical reaction to the             both the volume and surface resistance or conductance of the
absolute temperature; and the Boltzmann principle, a general             specimen, and surface resistance changes almost instanta-
law dealing with the statistical distribution of energy among            neously with change of relative humidity. It is, therefore,
large numbers of minute particles subject to thermal agitation.          absolutely essential to maintain both temperature and relative
The activation constant, m, has a value that is characteristic of        humidity within close limits during the conditioning period and
a particular energy absorption process. Several such processes           to make the insulation resistance or conductance measurements
may exist within the material, each with a different effective           in the specified conditioning environment. Another point not to
temperature range, so that several values of m would be needed           be overlooked is that at relative humidities above 90 %, surface
to fully characterize the material. These values of m can be             condensation may result from inadvertant fluctuations in hu-
determined experimentally by plotting the natural logarithm of           midity or temperature produced by the conditioning system.
resistance against the reciprocal of the absolute temperature.           This problem can be avoided by the use of equivalent absolute
The desired values of m are obtained from such a plot by                 humidity at a slightly higher temperature, as equilibrium
measuring the slopes of the straight-line sections of the plot.          moisture content remains nearly the same for a small tempera-
This derives from (Eq X1.1), for it follows that by taking the           ture change. In determining the effect of humidity on volume
natural logarithm of both sides:                                         resistance or conductance, extended periods of conditioning
                                        1                                are required, since the absorption of water into the body of the
                        1n R 5 ln B 1 m T                  (X1.2)        dielectric is a relatively slow process (10). Some specimens
                                                                         require months to come to equilibrium. When such long
  The change in resistance (or resistivity) corresponding to a
                                                                         periods of conditioning are prohibitive, use of thinner speci-
change in absolute temperature from T1 to T2, based on Eq
                                                                         mens or comparative measurements near equilibrium may be
X1.1, and expressed in logarithmic form, is:
                                                                         reasonable alternatives, but the details must be included in the
                            S1   1
                                        D S D
                                        DT
              ln ~R2/R1! 5 m T 2 T 5 m T T
                              2   1     1 2
                                                           (X1.3)        test report.
   These equations are valid over a temperature range only if               X1.4 Time of Electrification—Measurement of a dielectric
the material does not undergo a transition within this tempera-          material is not fundamentally different from that of a conductor
ture range. Extrapolations are seldom safe since transitions are         except that an additional parameter, time of electrification, (and
seldom obvious or predictable. As a corollary, deviation of a            in some cases the voltage gradient) is involved. The relation-
plot of the logarithm of R against 1/T from a straight line is           ship between the applied voltage and the current is involved in
evidence that a transition is occurring. Furthermore, in making          both cases. For dielectric materials, the standard resistance
comparisons between materials, it is essential that measure-             placed in series with the unknown resistance must have a
ments be made over the entire range of interest for all                  relatively low value, so that essentially full voltage will be
materials.                                                               applied across the unknown resistance. When a potential
                                                                    11
                                                             D 257 – 99 (2005)
difference is applied to a specimen, the current through it                  X1.6 Contour of Specimen:
generally decreases asymptotically toward a limiting value
                                                                              X1.6.1 The measured value of the insulation resistance or
which may be less than 0.01 of the current observed at the end
                                                                           conductance of a specimen results from the composite effect of
of 1 min (9, 11). This decrease of current with time is due to
                                                                           its volume and surface resistances or conductances. Since the
dielectric absorption (interfacial polarization, volume charge,
                                                                           relative values of the components vary from material to
etc.) and the sweep of mobile ions to the electrodes. In general,
                                                                           material, comparison of different materials by the use of the
the relation of current and time is of the form I(t) = At −m, after
                                                                           electrode systems of Fig. 1, Fig. 2, and Fig. 3 is generally
the initial charge is completed and until the true leakage current
                                                                           inconclusive. There is no assurance that, if material A has a
becomes a significant factor (12, 13). In this relation A is a
                                                                           higher insulation resistance than material B as measured by the
constant, numerically the current at unit time, and m usually,
                                                                           use of one of these electrode systems, it will also have a higher
but not always, has a value between 0 and 1. Depending upon
the characteristics of the specimen material, the time required            resistance than B in the application for which it is intended.
for the current to decrease to within 1 % of this minimum value               X1.6.2 It is possible to devise specimen and electrode
may be from a few seconds to many hours. Thus, in order to                 configurations suitable for the separate evaluation of the
ensure that measurements on a given material will be compa-                volume resistance or conductance and the approximate surface
rable, it is necessary to specify the time of electrification. The         resistance or conductance of the same specimen. In general,
conventional arbitrary time of electrification has been 1 min.             this requires at least three electrodes so arranged that one may
For some materials, misleading conclusions may be drawn                    select electrode pairs for which the resistance or conductance
from the test results obtained at this arbitrary time. A                   measured is primarily that of either a volume current path or a
resistance-time or conductance-time curve should be obtained               surface current path, not both (7).
under the conditions of test for a given material as a basis for
selection of a suitable time of electrification, which must be               X1.7 Deficiencies in the Measuring Circuit:
specified in the test method for that material, or such curves                X1.7.1 The insulation resistance of many solid dielectric
should be used for comparative purposes. Occasionally, a                   specimens is extremely high at standard laboratory conditions,
material will be found for which the current increases with                approaching or exceeding the maximum measurable limits
time. In this case either the time curves must be used or a                given in Table 2. Unless extreme care is taken with the
special study undertaken, and arbitrary decisions made as to               insulation of the measuring circuit, the values obtained are
the time of electrification.                                               more a measure of apparatus limitations than of the material
                                                                           itself. Thus errors in the measurement of the specimen may
  X1.5 Magnitude of Voltage:                                               arise from undue shunting of the specimen, reference resistors,
   X1.5.1 Both volume and surface resistance or conductance                or the current-measuring device, by leakage resistances or
of a specimen may be voltage-sensitive (4). In that case, it is            conductances of unknown, and possibly variable, magnitude.
necessary that the same voltage gradient be used if measure-                  X1.7.2 Electrolytic, contact, or thermal emf’s may exist in
ments on similar specimens are to be comparable. Also, the                 the measuring circuit itself; or spurious emf’s may be caused
applied voltage should be within at least 5 % of the specified             by leakage from external sources. Thermal emf’s are normally
voltage. This is a separate requirement from that given in                 insignificant except in the low resistance circuit of a galva-
X1.7.3, which discusses voltage regulation and stability where             nometer and shunt. When thermal emf’s are present, random
appreciable specimen capacitance is involved.                              drifts in the galvanometer zero occur. Slow drifts due to air
   X1.5.2 Commonly specified test voltages to be applied to                currents may be troublesome. Electrolytic emf’s are usually
the complete specimen are 100, 250, 500, 1000, 2500, 5000,                 associated with moist specimens and dissimilar metals, but
10 000 and 15 000 V. Of these, the most frequently used are                emf’s of 20 mV or more can be obtained in the guard circuit of
100 and 500 V. The higher voltages are used either to study the            a high-resistance detector when pieces of the same metal are in
voltage-resistance or voltage-conductance characteristics of               contact with moist specimens. If a voltage is applied between
materials (to make tests at or near the operating voltage                  the guard and the guarded electrodes a polarization emf may
gradients), or to increase the sensitivity of measurement.                 remain after the voltage is removed. True contact emf’s can be
   X1.5.3 Specimen resistance or conductance of some mate-                 detected only with an electrometer and are not a source of
rials may, depending upon the moisture content, be affected by             error. The term “spurious emf’’ is sometimes applied to
the polarity of the applied voltage. This effect, caused by                electrolytic emf’s. To ensure the absence of spurious emf’s of
electrolysis or ionic migration, or both, particularly in the              whatever origin, the deflection of the detecting device should
presence of nonuniform fields, may be particularly noticeable              be observed before the application of voltage to the specimen
in insulation configurations such as those found in cables                 and after the voltage has been removed. If the two deflections
where the test-voltage gradient is greater at the inner conductor          are the same, or nearly the same, a correction can be made to
than at the outer surface. Where electrolysis or ionic migration           the measured resistance or conductance, provided the correc-
does exist in specimens, the electrical resistance will be lower           tion is small. If the deflections differ widely, or approach the
when the smaller test electrode is made negative with respect              deflection of the measurement, it will be necessary to find and
to the larger. In such cases, the polarity of the applied voltage          eliminate the source of the spurious emf (5). Capacitance
shall be specified according to the requirements of the speci-             changes in the connecting shielded cables can cause serious
men under test.                                                            difficulties.
                                                                      12
                                                            D 257 – 99 (2005)
   X1.7.3 Where appreciable specimen capacitance is in-                   should be connected together until the measurement is to be
volved, both the regulation and transient stability of the applied        made to prevent any build-up of charge from the surroundings.
voltage should be such that resistance or conductance measure-
ments can be made to prescribed accuracy. Short-time tran-                   X1.9 Guarding:
sients, as well as relatively long-time drifts in the applied                X1.9.1 Guarding depends on interposing, in all critical
voltage may cause spurious capacitive charge and discharge                insulated paths, guard conductors which intercept all stray
currents which can significantly affect the accuracy of mea-              currents that might otherwise cause errors. The guard conduc-
surement. In the case of current-measuring methods particu-               tors are connected together, constituting the guard system and
larly, this can be a serious problem. The current in the                  forming, with the measuring terminals, a three-terminal net-
measuring instrument due to a voltage transient is I0 = CxdV/dt.          work. When suitable connections are made, stray currents from
The amplitude and rate of pointer excursions depend upon the              spurious external voltages are shunted away from the measur-
following factors:                                                        ing circuit by the guard system.
   X1.7.3.1 The capacitance of the specimen,                                 X1.9.2 Proper use of the guard system for the methods
   X1.7.3.2 The magnitude of the current being measured,                  involving current measurement is illustrated in Figs. X1.1-
   X1.7.3.3 The magnitude and duration of the incoming                    X1.3, inclusive, where the guard system is shown connected to
voltage transient, and its rate of change,                                the junction of the voltage source and current-measuring
   X1.7.3.4 The ability of the stabilizing circuit used to pro-           instrument or standard resistor. In Fig. X1.4 for the
vide a constant voltage with incoming transients of various               Wheatstone-bridge method, the guard system is shown con-
characteristics, and                                                      nected to the junction of the two lower-valued-resistance arms.
   X1.7.3.5 The time-constant of the complete test circuit as             In all cases, to be effective, guarding must be complete, and
compared to the period and damping of the current-measuring               must include any controls operated by the observer in making
instrument.                                                               the measurement. The guard system is generally maintained at
   X1.7.4 Changes of range of a current-measuring instrument              a potential close to that of the guarded terminal, but insulated
may introduce a current transient. When Rm [Lt ] Rx and Cm[Lt             from it. This is because, among other things, the resistance of
]Cx, the equation of this transient is                                    many insulating materials is voltage-dependent. Otherwise, the
                                                                          direct resistances or conductances of a three-terminal network
                     I 5 ~V0 / Rx!@I 2 e2t/RmCx#            (X1.4)
                                                                          are independent of the electrode potentials. It is usual to ground
where:                                                                    the guard system and hence one side of the voltage source and
V0 = applied voltage,                                                     current-measuring device. This places both terminals of the
Rx = apparent resistance of the specimen,                                 specimen above ground. Sometimes, one terminal of the
Rm = effective input resistance of the measuring instru-                  specimen is permanently grounded. The current-measuring
        ment,                                                             device usually is then connected to this terminal, requiring that
Cx = capacitance of the specimen at 1000 Hz,                              the voltage source be well insulated from ground.
Cm = input capacitance of the measuring instrument, and                      X1.9.3 Errors in current measurements may result from the
t    = time after Rm is switched into the circuit.                        fact that the current-measuring device is shunted by the
  For not more than 5 % error due to this transient,                      resistance or conductance between the guarded terminal and
                            RmCx # t/3                      (X1.5)        the guard system. This resistance should be at least 10 to 100
                                                                          times the input resistance of the current measuring device. In
   Microammeters employing feedback are usually free of this
                                                                          some bridge techniques, the guard and measuring terminals are
source of error as the actual input resistance is divided,
                                                                          brought to nearly the same potentials, but a standard resistor in
effectively, by the amount of feedback, usually at least by 1000.
                                                                          the bridge is shunted between the unguarded terminal and the
                                                                          guard system. This resistance should be at least 1000 times that
   X1.8 Residual Charge—In X1.4 it was pointed out that the
                                                                          of the reference resistor.
current continues for a long time after the application of a
potential difference to the electrodes. Conversely, current will
continue for a long time after the electrodes of a charged
specimen are connected together. It should be established that
the test specimen is completely discharged before attempting
the first measurement, a repeat measurement, a measurement of
volume resistance following a measurement of surface resis-
tance, or a measurement with reversed voltage (9). The time of
discharge before making a measurement should be at least four
times any previous charging time. The specimen electrodes                   FIG. X1.1 Voltmeter-Ammeter Method Using a Galvanometer
                                                                     13
                                                            D 257 – 99 (2005)
                                                                     14
                                                            D 257 – 99 (2005)
   where B is the fraction of the gap width to be added to the                    0.4                 0.85               2.0              0.41
diameter of circular electrodes or to the dimensions of rectan-                   0.5                 0.81               2.5              0.34
                                                                                  0.6                 0.77               3.0              0.29
gular or cylindrical electrodes.                                                  0.8                 0.71
   X2.2.2 Laminated materials, however, are somewhat aniso-
tropic after volume absorption of moisture. Volume resistivity              NOTE X2.1—The symbol “ln” designates logarithm to the base
                                                                          e = 2.718. ... When g is approximately equal to 2t, d is determined with
parallel to the laminations is then lower than that in the                sufficient approximation by the equation:
perpendicular direction, and the fringing effect is increased.
With such moist laminates, d approaches zero, and the guarded                                              d 5 0.586t                          (X2.4)
electrode effectively extends to the center of the gap between               NOTE X2.2—For tests on thin films when t << g, or when a guard
                                                                          electrode is not used and one electrode extends beyond the other by a
guarded and unguarded electrodes (14).
                                                                          distance which is large compared with t, 0.883t should be added to the
   X2.2.3 The fraction of the gap width g to be added to the              diameter of circular electrodes or to the dimensions of rectangular
diameter of circular electrodes or to the electrode dimensions            electrodes.
of rectangular or cylindrical electrodes, B, as determined by the            NOTE X2.3—During the transition between complete dryness and
preceding equation for d, is as follows:                                  subsequent relatively uniform volume distribution of moisture, a laminate
       g/t               B                g/t             B               is neither homogeneous nor isotropic. Volume resistivity is of questionable
       0.1              0.96              1.0            0.64             significance during this transition and accurate equations are neither
       0.2              0.92              1.2            0.59             possible nor justified, calculations within an order of magnitude being
       0.3              0.88              1.5            0.51             more than sufficient.
   X3.1 Voltmeter-Ammeter Method Using a Galvanometer:                    electrometer are connected to the voltage source and the
   X3.1.1 A dc voltmeter and a galvanometer with a suitable               specimen as illustrated in Fig. X1.2. The applied voltage is
shunt are connected to the voltage source and to the test                 measured by a dc voltmeter having the same characteristics as
specimen as shown in Fig. X1.1. The applied voltage is                    prescribed in X3.1.1. The current is measured in terms of the
measured by a dc voltmeter, having a range and accuracy that              voltage drop across a standard resistance, Rs.
will give minimum error in voltage indication. In no case shall              X3.2.2 In the circuit shown in Fig. X1.2(a) the specimen
a voltmeter be used that has an error greater than 62 % of full           current, Ix, produces across the standard resistance, Rs, a
scale, nor a range such that the deflection is less than one third        voltage drop which is amplified by the dc amplifier, and read
of full scale (for a pivot-type instrument). The current is               on an indicating meter or galvanometer. The net gain of the
measured by a galvanometer having a high current sensitivity              amplifier usually is stabilized by means of a feedback resis-
(a scale length of 0.5 m is assumed, as shorter scale lengths will        tance, Rf, from the output of the amplifier. The indicating meter
lead to proportionately higher errors) and provided with a                can be calibrated to read directly in terms of the feedback
precision Ayrton universal shunt for so adjusting its deflection          voltage,Vf, which is determined from the known value of the
that the readability error does not, in general, exceed 62 % of           resistance of Rf, and the feedback current passing through it.
the observed value. The galvanometer should be calibrated to              When the amplifier has sufficient intrinsic gain, the feedback
within 62 %. Current can be read directly if the galvanometer             voltage, Vs, differs from the voltage, IxRs, by a negligible
is provided with an additional suitable fixed shunt.                      amount. As shown in Fig. X1.2(a) the return lead from the
   X3.1.2 The unknown resistance, Rx, or conductance, Gx, is              voltage source, Vx, can be connected to either end of the
calculated as follows:                                                    feedback resistor, Rf. With the connection made to the junction
                    Rx 5 1/Gx 5 Vx/Ix 5 Vx/KdF              (X3.1)        of Rs andRf (switch in dotted position l), the entire resistance of
                                                                          Rs is placed in the measuring circuit and any alternating
where:                                                                    voltage appearing across the specimen resistance is amplified
K    = galvanometer sensitivity, in amperes per scale divi-
                                                                          only as much as the direct voltage Ix Rs , across Rs. With the
       sion,
                                                                          connection made to the other end of Rf (switch position 2), the
d    = deflection in scale divisions,
F    = ratio of the total current, Ix, to the galvanometer                apparent resistance placed in the measuring circuit is Rs times
       current, and                                                       the ratio of the degenerated gain to the intrinsic gain of the
Vx = applied voltage.                                                     amplifier; any alternating voltage appearing across the speci-
                                                                          men resistance is then amplified by the intrinsic amplifier gain.
   X3.2 Voltmeter-Ammeter Method Using DC Amplification                      X3.2.3 In the circuit shown in Fig. X1.2(b), the specimen
or Electrometer:                                                          current, Ix, produces a voltage drop across the standard
   X3.2.1 The voltmeter-ammeter method can be extended to                 resistance, Rs which may or may not be balanced out by
measure higher resistances by using dc amplification or an                adjustment of an opposing voltage, Vs, from a calibrated
electrometer to increase the sensitivity of the current measuring         potentiometer. If no opposing voltage is used, the voltage drop
device (6, 15, 16). Generally, but not necessarily, this is               across the standard resistance, Rs, is amplified by the dc
achieved only with some sacrifice in precision, depending on              amplifier or electrometer and read on an indicating meter or
the apparatus used. The dc voltmeter and the dc amplifier or              galvanometer. This produces a voltage drop between the
                                                                     15
                                                            D 257 – 99 (2005)
measuring electrode and the guard electrode which may cause               s usually specified). Alternatively, the voltage, DV, appearing
an error in the current measurement unless the resistance                 on the electrometer in a time, Dt, can be used. Since this gives
between the measuring electrode and the guard electrode is at             an average of the rate-of-change of voltage during Dt, the time
least 10 to 100 times that of Rs. If an opposing voltage, Vs, is          Dt should be centered at the specified electrification time (time
used, the dc amplifier or electrometer serves only as a very              since closing S2).
sensitive, high-resistance null detector. The return lead from               X3.3.2 If the input resistance of the electrometer is greater
the voltage source, Vx, is connected as shown, to include the             than the apparent specimen resistance and the input capaci-
potentiometer in the measuring circuit. When connections are              tance is 0.01 or less of that of the specimen, the apparent
made in this manner, no resistance is placed in the measuring             resistance at the time at which dV/dt or DV/Dt is determined is
circuit at balance and thus no voltage drop appears between the
                                                                                         Rx 5 V0/Ix 5 V0dt/C0dVm or, V0Dt/C0DVm        (X3.3)
measuring electrode and the guard electrode. However, a
steeply increasing fraction of Rs is included in the measuring               depending on whether or not a recorder is used. When the
circuit, as the potentiometer is moved off balance. Any                   electrometer input resistance or capacitance cannot be ignored
alternating voltage appearing across the specimen resistance is           or when Vmis more than a small fraction of V0 the complete
amplified by the net amplifier gain. The amplifier may be either          equation should be used.
a direct voltage amplifier or an alternating voltage amplifier                        Rs 5 $V0 [~Rx 1 Rm!/Rm#Vm% / ~C0 1 Cm!dVm/dt     (X3.4)
provided with input and output converters. Induced alternating
voltages across the specimen often are sufficiently troublesome           where:
that a resistance-capacitance filter preceding the amplifier is           C0 =      capacitance of Cx at 1000 Hz,
required. The input resistance of this filter should be at least          Rm =      input resistance of the electrometer,
100 times greater than the effect resistance that is placed in the        Cm =      input capacitance of the electrometer,
measurement circuit by resistance Rs.                                     V0   =    applied voltage, and
   X3.2.4 The resistance Rx, or the conductance,G x, is calcu-            Vm =      electrometer reading = voltage decrease on Cx.
lated as follows:
                                                                           X3.4 Comparison Method Using a Galvanometer or DC
                   Rx 5 1/Gx 5 Vx/Ix 5 ~Vx/Vs!Rs            (X3.2)
                                                                          Amplifier (1):
where:                                                                       X3.4.1 A standard resistance, Rs, and a galvanometer or dc
Vx =     applied voltage,                                                 amplifier are connected to the voltage source and to the test
Ix   =   specimen current,                                                specimen as shown in Fig. X3.1 . The galvanometer and its
Rs =     standard resistance, and                                         associated Ayrton shunt is the same as described in X3.1.1. An
Vs =     voltage drop across Rs, indicated by the amplifier               amplifier of equivalent direct current sensitivity with an
         output meter, the electrometer or the calibrated                 appropriate indicator may be used in place of the galvanometer.
         potentiometer.                                                   It is convenient, but not necessary, and not desirable if batteries
   X3.3 Voltage Rate-of-Change Method:                                    are used as the voltage source (unless a high-input resistance
                                                                          voltmeter is used), to connect a voltmeter across the source for
   X3.3.1 If the specimen capacitance is relatively large, or             a continuous check of its voltage. The switch is provided for
capacitors are to be measured, the apparent resistant, Rx, can be         shorting the unknown resistance in the process of measure-
determined from the charging voltage, V0, the specimen                    ment. Sometimes provision is made to short either the un-
capacitance value, C0 (capacitance of Cx at 1000 Hz), and the             known or standard resistance but not both at the same time.
rate-of-change of voltage, dV/dt, using the circuit of Fig. X3.1
(17). To make a measurement the specimen is charged by                       X3.4.2 In general, it is preferable to leave the standard
closing S2, with the electrometer shorting switch S1 closed.              resistance in the circuit at all times to prevent damage to the
When S1 is subsequently opened, the voltage across the                    current measuring instrument in case of specimen failure. With
specimen will fall because the leakage and absorption currents            the shunt set to the least sensitive position and with the switch
must then be supplied by the capacitance C0 rather than by V0.            open, the voltage is applied. The Ayrton shunt is then adjusted
The drop in voltage across the specimen will be shown by the              to give as near maximum scale reading as possible. At the end
electrometer. If a recorder is connected to the output of the             of the electrification time the deflection, dx, and the shunt ratio,
electrometer, the rate of change of voltage, dV/dt, can be read           Fx, are noted. The shunt is then set to the least sensitive
from the recorder trace at any desired time after S2 is closed (60        position and the switch is closed to short the unknown
                                                                          resistance. Again the shunt is adjusted to give as near maxi-
                                                                          mum scale reading as possible and the galvanometer or meter
                                                                          deflection, ds, and the shunt ratio, Fs, are noted. It is assumed
                                                                          that the current sensitivities of the galvanometer or amplifier
                                                                          are equal for nearly equal deflections dxand ds.
                                                                             X3.4.3 The unknown resistance, Rx, or conductance, Gx, is
                                                                          calculated as follows:
                                                                                             R 5 1/Gx 5 Rs@~dsFs / dxFx! – 1]          (X3.5)
                                                                          where:
            FIG. X3.1 Voltage Rate-of-Change Method
                                                                     16
                                                                     D 257 – 99 (2005)
Fxand Fs     = ratios of the total current to the galvanometer                      indicating meter of Fig. X1.2(a) can be replaced by a recording
                or dc amplifier with Rx in the circuit, and                         milliammeter or millivoltmeter as appropriate for the amplifier
                shorted, respectively.                                              used. The recorder may be either the deflection type or the
   X3.4.4 In case Rs is shorted when Rx is in the circuit or the                    null-balance type, the latter usually having a smaller error.
ratio of Fs to Fx is greater than 100, the value of Rx or Gx is                     Null-balance-type recorders also can be employed to perform
computed as follows:                                                                the function of automatically adjusting the potentiometer
                        Rx 5 1/Gx 5 R ~dsFs/dxFx!                    (X3.6)         shown in Fig. X1.2(b) and thereby indicating and recording the
                                                                                    quantity under measurement. The characteristics of amplifier,
  X3.5 Comparison Methods Using a Wheatstone Bridge (2):                            recorder balancing mechanism, and potentiometer can be made
   X3.5.1 The test specimen is connected into one arm of a                          such as to constitute a well integrated, stable, electromechani-
Wheatstone bridge as shown in Fig. X1.4. The three known                            cal, feedback system of high sensitivity and low error. Such
arms shall be of as high resistance as practicable, limited by the                  systems also can be arranged with the potentiometer fed from
errors inherent in such resistors. Usually, the lowest resistance,                  the same source of stable voltage as the specimen, thereby
RA, is used for convenient balance adjustment, with either RBor                     eliminating the voltmeter error, and allowing a sensitivity and
RNbeing changed in decade steps. The detector shall be a dc                         precision comparable with those of the Wheatstone-Bridge
amplifier, with an input resistance high compared to any of                         Method (X3.5).
these arms.
   X3.5.2 The unknown resistance, Rx, or conductance, Gx, is                           X3.7 Direct-Reading Instruments—There are available,
calculated as follows:                                                              and in general use, instruments that indicate resistance directly,
                          Rx 5 1/Gx 5 RBRN / RA                      (X3.7)         by a determination of the ratio of voltage and current in bridge
                                                                                    methods or related modes. Some units incorporate various
  where RA, RB, and RN are as shown in Fig. X1.4. When arm
                                                                                    advanced features and refinements such as digital readout.
A is a rheostat, its dial can be calibrated to read directly in
                                                                                    Most direct reading instruments are self-contained, portable,
megohms after multiplying by the factor RBRN which for
                                                                                    and comprise a stable dc power supply with multi-test voltage
convenience can be varied in decade steps.
                                                                                    capability, a null detector or an indicator, and all relevant
   X3.6 Recordings—It is possible to record continuously                            auxiliaries. Measurement accuracies vary somewhat with type
against time the values of the unknown resistance or the                            of equipment and range of resistances covered; for the more
corresponding value of current at a known voltage. Generally,                       elaborate instruments accuracies are comparable to those
this is accomplished by an adaptation of the voltmeter-ammeter                      obtained with the voltmeter-ammeter method using a galva-
method, using dc amplification (X3.2). The zero drift of direct                     nometer (X3.1). The direct-reading instruments do not neces-
coupled dc amplifiers, while slow enough for the measure-                           sarily supplant any of the other typical measurement methods
ments of X3.2, may be too fast for continuous recording. This                       described in this Appendix, but do offer simplicity and conve-
problem can be resolved by periodic checks of the zero, or by                       nience for both routine and investigative resistance
using an ac amplifier with input and output converter. The                          measurements.
REFERENCES
(1) Curtis, H. L., “Insulating Properties of Solid Dielectric,” Bulletin,           (8) Witt, R. K., Chapman, J. J., and Raskin, B. L.,“ Measuring of Surface
    National Institute of Standards and Technology, Vol II, 1915, Scientific            and Volume Resistance,” Modern Plastics, Vol 24, No. 8, April 1947,
    Paper No. 234, pp. 369–417.                                                         p. 152.
(2) Field, R. F., “How Humidity Affects Insulation, Part I, D-C Phenom-             (9) Scott, A. H.,“ Insulation Resistance Measurements,” Fourth Electrical
    ena,” General Radio Experimenter, Vol 20, Nos. 2 and 3, July–August                 Insulation Conference, Washington, DC, February 19–22, 1962.
    1945.                                                                           (10) Kline, G. M., Martin, A. R., and Crouse, W. A.,“ Sorption of Water by
(3) Field, R. F., “The Formation of Ionized Water Films on Dielectrics                   Plastics,” Proceedings, American Society for Testing and Materials,
    Under Conditions of High Humidity,” Journal of Applied Physics, Vol                  Vol 40, 1940, pp. 1273–1282.
    5, May 1946.                                                                    (11) Greenfield, E. W.,“ Insulation Resistance Measurements,” Electrical
(4) Herou, R., and LaCoste, R., “Sur La Mésure Des Resistivities et                     Engineering, Vol 66, July 1947, pp. 698–703.
    L’Etude de Conditionnement des Isolantes en Feuilles,” Report IEC               (12) Cole, K. S., and Cole, R. H., “Dispersion and Absorption in
    15-GT2(France) April 4, 1963.                                                        Dielectrics, II Direct Current Characteristics,” Journal of Chemical
(5) Thompson, B. H., and Mathes, K. N., “Electrolytic Corrosion—                         Physics, Vol 10, 1942.
    Methods of Evaluating Materials Used in Tropical Service,” Transac-             (13) Field, R. F., “Interpretation of Current-Time Curves as Applied to
    tions, American Institute of Electrical Engineers, Vol 64, June 1945, p.             Insulation Testing,” AIEE Boston District Meeting, April 19–20,
    287.                                                                                 1944.
(6) Scott, A. H., “Anomalous Conductance Behavior in Polymers,” Report              (14) Lauritzen, J. I., “The Effective Area of a Guarded Electrode,” Annual
    of the 1965 Conference on Electrical Insulation, NRC-NAS.                            Report, Conference on Electrical Insulation. NAS-NRC Publication
(7) Amey, W. G., and Hamberger, F., Jr., “A Method for Evaluating the                    1141, 1963.
    Surface and Volume Resistance Characteristics of Solid Dielectric               (15) Turner, E. F., Brancato, E. L., and Price, W., “The Measurement of
    Materials,” Proceedings, American Society for Testing and Materials,                 Insulation Conductivity,” NRL Report 5060, Naval Research Labo-
    Vol 49, 1949, pp. 1079–1091.                                                         ratory, Feb. 25, 1958.
                                                                               17
                                                                          D 257 – 99 (2005)
(16) Dorcas, D. S., and Scott, R. N., “Instrumentation for Measuring the                     Impregnated Paper as Insulation for HV-DC Cables,” IEEE Transac-
     D-C Conductivity of Very High Resistivity Materials,” Review of                         tions on Power Apparatus and Systems, Vol PAS-86, No. 3, March
     Scientific Instruments, Vol 35, No. 9, September 1964.                                  1967.
(17) Endicott, H. S., “Insulation Resistance, Absorption, and Their Mea-                (19) Endicott, H. S., “Guard-Gap Correction for Guarded-Electrode Mea-
     surement,” Annual Report, Conference on Electrical Insulation,                          surements and Exact Equations for the Two-Fluid Method of Mea-
     NAS-NRC Publication, 1958.                                                              suring Permittivity and Loss,” Journal of Testing and Evaluation, Vol
(18) Occhini, E., and Maschio, G., “Electrical Characteristics of Oil-                       4, No. 3, May 1976, pp. 188–195.
                    ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
                in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
                of infringement of such rights, are entirely their own responsibility.
                    This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
                if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
                and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
                responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
                make your views known to the ASTM Committee on Standards, at the address shown below.
                  This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
                United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
                address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
                (www.astm.org).
18