0% found this document useful (0 votes)
71 views73 pages

Manual compEAct S

Service Manual Compact

Uploaded by

pramodyad5810
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
71 views73 pages

Manual compEAct S

Service Manual Compact

Uploaded by

pramodyad5810
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 73

Operating instructions

compEAct S / compEAct SMPO


Sulfur Analyzer
Manufacturer Analytik Jena AG
Konrad-Zuse-Str.1
07745 Jena  Germany
Phone + 49 3641 77 70
Fax + 49 3641 77 9279
Email info@analytik-jena.de

Service Analytik Jena AG


Konrad-Zuse-Str. 1
07745 Jena  Germany
Phone + 49 3641 77 7407
Fax + 49 3641 77 7449
Email service@analytik-jena.com

General Information http://www.analytik-jena.com

Copyrights and compEAct is a trademark of Analytik Jena AG registered in Germany.


Trademarks The identification with ® or TM is omitted in this manual.

Documentation number 11-0205-003-23


Edition 01.17
Technical documentation Analytik Jena AG
made by:

© Copyright 2017 Analytik Jena AG


compEAct S Contents

Contents
1 Basic information ............................................................................................................. 7
1.1 Notes on these operating instructions ........................................................................ 7
1.2 Intended use ............................................................................................................... 8
1.3 Warranty and liability.................................................................................................. 9

2 Safety instructions ......................................................................................................... 10


2.1 General notes ............................................................................................................ 10
2.2 Safety markings on the compEAct S .......................................................................... 10
2.3 Requirements for the operating personnel ............................................................... 11
2.4 Safety instructions – Transport and commissioning ................................................. 11
2.5 Safety instructions for operation ............................................................................... 12
2.5.1 General instructions .................................................................................................. 12
2.5.2 Safety instructions – Protection against explosion and fire ...................................... 12
2.5.3 Safety instructions – Electrical equipment ................................................................ 12
2.5.4 Safety instructions for compressed gas containers and systems ............................... 13
2.5.5 Handling of auxiliary and operating materials and samples ..................................... 13
2.5.6 Safety instructions – service and repair..................................................................... 14
2.5.7 Safety instructions for the operation with auto-injector AI and autosampler LS ...... 14
2.5.8 Behavior in case of overpressure faults (gas pressure faults) ................................... 14
2.6 Safety equipment / Behavior during emergencies .................................................... 15

3 Function and layout of the compEAct S ....................................................................... 16


3.1 Principle of operation................................................................................................ 16
3.2 Layout of the compEAct S ......................................................................................... 17
3.3 Type plate ................................................................................................................. 22
3.4 Sample supply ........................................................................................................... 22
3.4.1 Auto-injector AI......................................................................................................... 22
3.4.2 Autosamplers LS 1 and LS 2...................................................................................... 23

4 Installation and commissioning ................................................................................... 25


4.1 Installation conditions............................................................................................... 25
4.1.1 Environmental conditions ......................................................................................... 25
4.1.2 Energy supply ............................................................................................................ 25
4.1.3 Gas supply ................................................................................................................. 26
4.1.4 Device layout and space requirements ...................................................................... 26
4.2 Supply and control connections ................................................................................ 28
4.3 Installing the compEAct S with sample introduction module .................................... 30
4.3.1 Setting up and connecting the compEAct S ............................................................... 30
4.3.2 Connecting the autosampler LS ................................................................................ 31
4.3.3 Connecting the auto-injector AI ................................................................................ 32

5 Operation ........................................................................................................................ 34
5.1 Switching the compEAct S on and off ....................................................................... 34
5.2 Measuring with auto-injector AI ............................................................................... 35
5.3 Measuring with LS 1 or LS 2 ..................................................................................... 36

6 Maintenance and care ................................................................................................... 38


6.1 Maintenance intervals .............................................................................................. 38
6.2 Cleaning the compEAct S and its system components .............................................. 39
6.3 Checking system tightness ........................................................................................ 39
6.4 Adjusting the auto-injector AI ................................................................................... 40
6.5 Adjusting the autosampler LS ................................................................................... 41
6.6 Combustion furnace maintenance ............................................................................ 42
6.6.1 Removing and cleaning the combustion tube ........................................................... 42
6.6.2 Replaceing quartz wool plug ..................................................................................... 44
6.6.3 Inserting the combustion tube .................................................................................. 45

3
Contents compEAct S

6.7 Replacing the membrane dryer ................................................................................. 46


6.8 Servicing the auto-protection assembly .................................................................... 48
6.8.1 Removing and installing the auto-protection assembly............................................ 48
6.8.2 Checking and replacing the filter .............................................................................. 49
6.8.3 Replacing the pneumatic seal ................................................................................... 51
6.9 Opening and closing the side panel of the compEAct S ............................................ 52
6.10 Replacing the deuterium lamp (UVFD) ..................................................................... 53
6.11 Replacing the chemical ozone decomposer (compEAct SMPO only) ........................... 55
6.12 Removing and installing the combustion furnace ..................................................... 55
6.12.1 Removing the combustion furnace ........................................................................... 55
6.12.2 Installing the combustion furnace............................................................................. 57

7 Troubleshooting ............................................................................................................. 59
7.1 Troubleshooting according to software messages .................................................... 59
7.2 Equipment faults and analytical problems ................................................................ 63
7.2.1 General notes ............................................................................................................ 63
7.2.2 Equipment faults....................................................................................................... 64
7.2.3 Analytical problems .................................................................................................. 65

8 Transport and storage ................................................................................................... 66


8.1 Preparing the device for transport ............................................................................ 66
8.1.1 Packing the autosampler LS ...................................................................................... 66
8.1.2 Packing the auto-injector AI ..................................................................................... 67
8.1.3 Packing the compEAct S ............................................................................................ 67
8.2 Transporting the compEAct S.................................................................................... 68
8.3 Moving the compEAct S in the laboratory................................................................. 68
8.4 Storage...................................................................................................................... 69

9 Disposal ........................................................................................................................... 70

10 Specification ................................................................................................................... 71
10.1 Technical data ........................................................................................................... 71
10.1.1 Technical data of the compEAct S ............................................................................. 71
10.1.2 Technical data of the auto-injector AI....................................................................... 72
10.1.3 Technical data of the autosampler LS ....................................................................... 72
10.2 Guidelines and standards .......................................................................................... 72

4
compEAct S Figures

Figures
Fig. 1 Safety instructions on the compEAct S ............................................................. 10
Fig. 2 Principle of operation ........................................................................................ 16
Fig. 3 Main components of the compEAct S............................................................... 17
Fig. 4 Combustion tube ................................................................................................ 18
Fig. 5 Connections to the combustion tube ................................................................ 18
Fig. 6 Auto-protection assembly ................................................................................. 19
Fig. 7 Toggle switch for pneumatic seal ..................................................................... 19
Fig. 8 Membrane dryer ................................................................................................ 20
Fig. 9 Gas flow diagram of the compEAct S and compEAct SMPO .............................. 21
Fig. 10 Layout of the auto-injector AI ....................................................................... 22
Fig. 11 Autosampler LS .............................................................................................. 23
Fig. 12 Sample rack of the LS 2 ................................................................................. 23
Fig. 13 Solvent and waste containers of the LS ....................................................... 24
Fig. 14 Electrical connections of the LS .................................................................... 24
Fig. 15 Installation diagram for the compEAct S with autosampler LS .................. 27
Fig. 16 Connections on the rear of the compEAct S ................................................. 28
Fig. 17 USB port and hose bridge in the gas path behind the front door .............. 28
Fig. 18 Location of the switch on the compEAct S ................................................... 29
Fig. 19 Flow monitoring set ....................................................................................... 40
Fig. 20 compEAct S with screwed-in carrying handles ............................................ 69

5
Figures compEAct S

6
compEAct S Basic information

1 Basic information

1.1 Notes on these operating instructions

Contents The user manual describes the following two models of the compEAct series:
 compEAct S – Sulfur analyzer
 compEAct SMPO – Sulfur analyzer with MPO technology
In this manual these models are collectively called compEAct S. Any differences
between the models are explained in the relevant section.
These operating instructions provide information about the design and operation of
the compEAct S and provide operating personnel the necessary know-how for the safe
handling of the compEAct S and its components. Furthermore, the operating
instructions include information on the maintenance and servicing of the device as
well as hints on potential causes of malfunctions and their correction.

User requirements The operating instructions are intended for users who are familiar with the principles
of sulfur analysis. The user should at least be trained as a chemical laboratory assistant
or have an equivalent qualification. Knowledge of the safety guidelines for working in
a chemical laboratory and for handling the chemicals that are used are required.
Furthermore, the user must have basic knowledge of working with a computer.

Conventions Instructions for actions which occur in chronological order are numbered and
combined in action units.
Warnings are indicated by warning triangles and a signal word. The type, source and
consequences of the danger are stated together with notes on preventing the danger.
The elements of the control and analysis program are indicated as follows:
 Program terms are indicated by small caps (e.g., SYSTEM menu).
 Buttons are shown by square brackets (e.g., [OK]).
 Menu items are separated by arrows (e.g., SYSTEM  DEVICE).

Symbols and signal words The operating instructions use the following symbols and signal words to indicate
used hazards or instructions. The warnings are always placed before an action.

WARNING
Indicates a potentially hazardous situation which might cause death or very serious
injuries (deformities).

CAUTION
Indicates a potentially hazardous situation which might cause light or minor injuries.

Attention
Provides information on potential material or environmental damage.

7
Basic information compEAct S

1.2 Intended use

The compEAct S is a compact stand-alone elemental analyzer for determining sulfur


contents in liquid, gaseous and LPG samples. The detection takes place through
pyrolysis followed by thermal oxidation in accordance with national and international
standards.
The compEAct S is equipped with at least one sample introduction module. The
compEAct S is equipped with an on-board computer with touchscreen and the
operating and control software EAvolution which is used to control the device and to
analyze the measured data. There is the alternative option to operate the software
using an external keyboard, mouse and monitor or via an external PC. The software
EAvolution advanced can be accessed remotely and provides the option to control the
device remotely. For this purpose, the compEAct S must be connected to a local
network or directly to the Internet.
The compEAct S may only be used for the methods described in these operating
instructions for determining sulfur contents. Any other use is not as intended! The
operator is the sole responsible and liable for any damages that result from such
unintended use.
The following substances must not be analyzed with the compEAct S because they
pose a risk of explosion:
 highly flammable organic compounds (e.g. isopentane)
 substances tending to spontaneous decomposition (e.g. peroxides)
 explosives, explosive materials (e.g. trinitrotoluol, inorganic azides)
The following substances must not be analyzed with the compEAct S because they
could damage the analysis system:
 inorganic substances (e.g. sulfuric acid)
 substances with a high content of alkaline ions and alkaline earth ions (e.g. sodium
acetate)
 organometallic compounds (e.g. metal-organyls)
 phosphorus and organic silicon compounds or samples with a high content of these
elements (e.g. the hydraulic fluid Skydrol)
 substances or samples with a high content of fluoride ions

8
compEAct S Basic information

1.3 Warranty and liability

The warranty duration and liability comply with the legal requirements and the
provisions in the general terms and conditions of Analytik Jena.
Deviations from the intended use described in these operating instructions result in
limitations of warranty and liability in the event of a damage. Damage to wearing
parts is not included in the warranty.
Warranty and liability claims are excluded for personal injury and property damage
due to one or several of the following causes:
 use of the device other than intended
 improper commissioning, operation and servicing of the device
 modifications of the equipment without prior consultation with Analytik Jena
 operation of the device with faulty safety equipment or improperly fitted safety and
protection equipment
 inadequate monitoring of the equipment components subject to wear
 use of other than original spare parts, wearing parts or consumables
 improper repairs
 faults due to the non-observance of these operating instructions

9
Safety instructions compEAct S

2 Safety instructions

2.1 General notes

For your own safety and to ensure error-free operation of the device, please read this
chapter carefully before commissioning.
Observe all safety notes listed in these operating instructions and all messages and
notes displayed by the control and analysis software on the monitor.
Besides the safety instructions in these operating instructions and the local safety
regulations that apply to the operation of the device, the general applicable
regulations regarding accident prevention, occupational health and safety and
environmental protection have to be observed and complied with.
References to potential dangers do not replace the work protection regulations which
must be observed.

2.2 Safety markings on the compEAct S

Warnings and information symbols have been attached to the compEAct S. These must
be observed under all circumstances.
Damaged or missing warnings and information symbols can cause incorrect actions
leading to personal injury or material damage! Symbol labels must not be removed or
wetted with methanol! Damaged symbol labels must be replaced without delay!

Fig. 1 Safety instructions on the compEAct S

Number Warning Position/meaning


1 On the cover plate in front of the combustion furnace +
On the cover plate above the combustion furnace (not
shown)
Warning against burns
There is a risk of burns at the combustion furnace. Allow
the furnace to cool down sufficiently before carrying out
any maintenance on or near the furnace.

10
compEAct S Safety instructions

Number Warning Position/meaning


2 and 3 On the flap inside the furnace chamber and on the
removable right side of the device
Warning against electric shock inside the device
The voltages applied within the device can be life-
threatening. Switch off the mains switch and pull the
device's mains plug from the power outlet before opening
the device.
Not shown On the cover plate above the combustion furnace
(in Fig. 1) Warning against hazard location
Do not reach into the travel path of the autosampler LS
while the sampler is moving.
Not shown On the autosampler LS
Warning against injuries caused by crushing
Do not reach into the travel path of the injector head while
the injector head is moving.
Not shown For China: The device contains controlled substances.
Analytik Jena AG guarantees that the substances will not
leak within the next 25 years and, therefore, will not cause
any environmental danger or health risk within this period
of time if the device is used as intended.

2.3 Requirements for the operating personnel

The compEAct S may only be operated by qualified specialist personnel instructed in


the use of the device. The personal instruction also includes conveying the contents of
these operating instructions. We recommend training by qualified employees of
Analytik Jena or its representatives.
The operating instructions must be accessible to the operating and service personnel.

2.4 Safety instructions – Transport and commissioning

Observe the following notes:


 Incorrect installation can create serious hazards. Therefore, the compEAct S and its
system components are always installed by the service engineers of Analytik Jena
or its authorized and trained specialist personnel. Unauthorized assembly and
installation are not permitted.
 Insufficiently secured components pose a risk of injury! During transport the
components of the equipment must be secured in accordance with the
specifications in these operating instructions. The combustion tube and the
combustion furnace must be removed before transporting the device.
 Two people are required to transport the device or to move it inside the laboratory.
They must lift and hold the device on four hand-tightened screwed-in carrying
handles. The device weighs approximately 30 kg.

11
Safety instructions compEAct S

2.5 Safety instructions for operation

2.5.1 General instructions

Each time before starting up the compEAct S, the operator must make sure that the
device and its safety equipment is sound condition. This applies in particular after each
modification or extension of the device or its repair.
Observe the following notes:
 The device may only be operated if all protective equipment (e.g. covers and doors)
are in place, properly installed and fully operational.
 The operator must have free access to the front door and the mains switch located
behind it at any moment during operation.
 The ventilation fittings at the rear and the bottom of the device must be
operational. Covered ventilation grilles or slits etc. may cause the device to break
down or may cause damage to it.
 The furnace operates with temperatures of up to 1100 °C. Hot components must
not be touched during or directly after operating the compEAct S.
 Ensure that no liquid reaches the plug connectors or enters the interior of the
device! Danger of electric shock!
 Caution when handing quartz glass and glass parts. There is the risk that glass
components may break and cause injuries!
 Keep all combustible materials away from the device.

2.5.2 Safety instructions – Protection against explosion and fire

The compEAct S must not be operated in an explosive environment.


Smoking or handling open flames are prohibited in the room in which the device is
operated!

2.5.3 Safety instructions – Electrical equipment

Contact with live components may cause death, serious injury or painful electrical
shock. Lethal voltages may occur in the compEAct S!
Observe the following notes:
 The mains plug must be connected to a proper power outlet to ensure the device's
compliance with protection class I (ground connector). The device may only be
connected to power sources whose nominal voltage is the same as that on the
rating plate of the equipment. The protective effect must not be invalidated by the
use of an extension line which does not have a protective conductor.
 The device and the system components may only be connected to the mains when
the device is switched off.
 Electrical connection cables between the device and the system components may
only be connected or disconnected when the device is switched off.

12
compEAct S Safety instructions

 Before opening the compEAct S the device must be switched off using the mains
switch and the mains plug must be disconnected from the power outlet! Work on
the electronics may only be carried out by the customer service of Analytik Jena and
specially authorized technicians.

2.5.4 Safety instructions for compressed gas containers and systems

The carrier gas (argon and oxygen) is taken from compressed gas cylinders or local
compressed gas systems. Make sure that the carrier gas has the specified purity!
Work on compressed gas cylinders and systems may only be carried out by individuals
with specialist knowledge and experience in compressed gas systems.
Observe the following notes:
 The safety instructions and guidelines for operating compressed gas cylinders or
compressed gas systems that apply at the operating location must be strictly
complied with.
 Compressed air hoses and pressure reducers may only be used for the assigned
gases.
 Pipes, hoses, screw connections and pressure reducers for oxygen must be kept free
from grease.
 All pipes, hoses and screw connections must be checked regularly for leaks and
externally visible damage. Leaks and damaged must be repaired without delay.
 The gas supply must be closed prior to inspections, service and repairs!
 After successful repair and service of the components of the compressed air
containers or system the compEAct S must be checked for sound operation prior to
recommissioning!
 Unauthorized assembly and installation are not permitted!

2.5.5 Handling of auxiliary and operating materials and samples

The operator is responsible for the selection of substances used in the process as well
as for their safe handling. This is particularly important for radioactive, infectious,
poisonous, corrosive, combustible, explosive and otherwise dangerous substances.
When handling hazardous substances the locally applicable safety instructions and/or
instructions in the EC safety data sheets provided by the manufacturer of the auxiliary
and operating materials must be complied with.
The combustion tube is filled with a quartz wool plug. Observe the following
instructions when working with quartz wool:
 Only store quartz wool in closed containers.
 When working with quartz wool avoid the formation of dust! Inhaled dust might
cause irritation to respiratory tracts.
 Wear personal protective equipment (laboratory coat, protective gloves, safety
goggles, breathing mask) when replacing the quartz wool and when cleaning the
combustion tube.

13
Safety instructions compEAct S

 Collect used quartz wool in suitable, sealed containers and dispose of the material
in accordance with applicable legal regulations. Contact the responsible waste
disposal company to organize the disposal of the waste.

2.5.6 Safety instructions – service and repair

Observe the following notes:


 The compEAct S is usually serviced by the customer service department of Analytik
Jena or its authorized and trained specialist personnel. Unauthorized servicing can
lead to maladjustment or damage of the device and its system components.
Therefore, the tasks that the operator is allowed to carry out are strictly limited to
the tasks listed in the chapter "Maintenance and care" on page 38.
 The exterior of the compEAct S may only be cleaned with a piece of cloth wetted
with water but not dripping, after the device has been switched off.
 Generally, all service and repair work on the device must be carried out in switched-
off condition (unless stated otherwise).
 Service tasks and the replacement of system components (e.g. removal of the
combustion tube) may only be carried out after a sufficiently long cooling-down
phase.
 Prior to servicing or repair the energy and gas supplies must be turned off and the
compEAct S must be vented!
 Only original accessories and original replacement parts from Analytik Jena may be
used.
 All protective equipment must be reinstalled correctly immediately after
completion of the service and repair work and be checked for operation!

2.5.7 Safety instructions for the operation with auto-injector AI and autosampler LS

Observe the following instructions when installing and operating the auto-injector AI
and the autosampler LS:
 When positioning the compEAct S with an LS autosampler on the laboratory desk
consider the movement range of the injector head during operation. Make sure
that the movement range is clear.
 There is a risk of injury within the movement range of the injector head during
operation. Keep a safe distance from the device.
 The autosampler LS and the auto-injector AI may only be opened by authorized
Analytik Jena service personnel. Disconnect the power supply before opening the
sample introduction system. Danger of electric shock!
 Only use the designated connection terminals to connect the autosampler LS and
the auto-injector to the compEAct S (9, 10 in Fig. 16).

2.5.8 Behavior in case of overpressure faults (gas pressure faults)

Greatest caution is necessary during an overpressure in the system! With incorrect


operation the operating personnel may be endangered and the analysis system will be

14
compEAct S Safety instructions

damaged. In case of an overpressure fault, the EAvolution software displays a warning


message and starts the routine to reduce the overpressure.
Observe the following notes:
 Never switch off a device that is subject to overpressure!
 Do not introduce any samples.
 Do not shut down the EAvolution software.
 Do not cut the external gas supply.
 Wait until the overpressure in the system has dropped to normal pressure. The
routine for reducing the overpressure is an automatic process.
 After the routine has ended, follow the instructions in the software.
 In case the routine does not achieve to reduce the pressure: Pierce a cannula into
the septum of the injection port so that the gas can escape via the cannula.

2.6 Safety equipment / Behavior during emergencies

Observe the following notes:


 In case of hazardous situations or accidents and if there is no immediate risk of
injury, switch off the compEAct S and the connected system components
immediately using the mains switch (behind the front door) and/or disconnect the
mains plugs from the power outlets, if possible!
 Cut the oxygen supply immediately or as soon as possible after the device is
switched off!

15
Function and layout of the compEAct S compEAct S

3 Function and layout of the compEAct S

3.1 Principle of operation

The compEAct S is a compact elemental analyzer for determining sulfur contents in


liquid samples. The separately available LPG module 2.0 allows feeding the analyzer
with pressurized liquefied gas samples (LPG). The combined GSS/LPG module allows
feeding the analyzer with both LPG samples and pressurized gaseous samples.
The samples are digested at temperatures between 1000 and 1100 °C in a two-stage
process consisting of a pyrolysis followed by thermal oxidation. During the first phase
of the digestion the sample components are vaporized or pyrolized in the Argon flow
and the generated pyrolysis gases incinerated in the oxygen flow. Next the remaining
pyrolysis products are re-incinerated in the pure oxygen flow during the second phase.
The digestion can be summed up by the following equation:
R* - S + O2 → SO2 + CO2 + H2O
R*... = residual hydrocarbons

The sample aliquot (liquid, gaseous, LPG) is metered directly into the combustion tube
within the sample introduction module (autosampler LS, auto-injector AI, LPG
module 2.0, combined GSS/LPG module) via the injection port with septum
After leaving the combustion tube the reaction gas mix is fed to the measuring gas
drying process. The measuring gas is dried in a membrane dryer. The dried measuring
gas is fed to the UV fluorescence detector (UVFD).

Fig. 2 Principle of operation

The detection is done using the UV fluorescence method. This principle is based on the
fact that SO2, which is excited by UV light, emits a characteristic fluorescence radiation

16
compEAct S Function and layout of the compEAct S

in the range of 220 to 420 nm. This fluorescence radiation is measured. The changes
in the intensity of the fluorescence are then used to calculate the SO2 concentration.

3.2 Layout of the compEAct S

Main components of the The compEAct S consists of the following main components:
compEAct S
 Electronics
 Internal computer with touchscreen
 Gas supply
 Combustion system
 Auto-protection assembly (APB)
 Measuring gas dryer
 UV fluorescence detector
 Sample introduction system
All components of the compEAct S to be operated or serviced by the user can be
reached via the front door and the opening at the top of the device.

Fig. 3 Main components of the compEAct S

1 Touchscreen 5 Combustion furnace


2 ON/OFF switch 6 Membrane dryer
3 Mains switch 7 Head of the combustion tube with injection
4 Auto-protection assembly port and gas terminals

Electrical components Electrical connections, gas connections and the connections for the sample
and gas supply introduction module are located on the rear panel of the compEAct S (→ Fig. 16,
p. 28).

17
Function and layout of the compEAct S compEAct S

The two process gases argon and oxygen are controlled via the gas box in the
compEAct S. The gas box does not require any maintenance from the user.
The compEAct S is a stand-alone device with integrated computer. The control and
analysis software EAvolution is operated via the touchscreen on the right-hand side of
the device.
Combustion system In the compEAct S a resistor-heated combustion furnace is used for digestion
temperatures between 700 °C and 1100 °C. Depending on the application, the samples
are digested inside the combustion tube at temperatures between 950 °C and 1100 °C.
A combustion tube has been inserted into the combustion furnace of the compEAct S
which is used for all standard applications. The combustion tube is made of quartz
glass. A quartz wool plug inserted in the inner tube ensures a slow and steady
vaporization of the sample. The connection between the combustion tube and the
measuring gas dryer is implemented via the auto-protection assembly.

Fig. 4 Combustion tube

1 Connection to the auto-protection assembly 3 Internal tube with quartz wool plug
2 Glass drip 4 Head with injection port and gas terminals

The injection port and the gas terminals are located on the head of the combustion
tube.

Fig. 5 Connections to the combustion tube

1 Oxygen connection (hose no. 3, blue) 3 Injection port


2 Argon connection (hose no. 4, gray)

Auto-protection assembly The auto-protection assembly is used to couple the combustion tube to the measuring
gas dryer. The assembly has an integrated replaceable filter. The filter protects the

18
compEAct S Function and layout of the compEAct S

downstream membrane dryer and the detector against soot particles and solid
pyrolysis products in case of incomplete combustion. The filter also retains condensed
water and other aerosols; only water vapor passes through.
The auto-protection assembly is mounted to a plug-in unit below the furnace.

Fig. 6 Auto-protection assembly

1 Membrane dryer 5 Holder


2 Plug-in unit for the auto-protection assembly 6 Connection to the membrane dryer
3 Gas connection for the pneumatic seal (argon) 7 Electrical connection
4 Pneumatic seal

The combustion tube is plugged into the auto-protection assembly and sealed with a
pneumatic seal. The seal is operated using argon.
A toggle switch is used to open and close the seal:
 Toggle switch down: combustion tube is sealed
 Toggle switch up: combustion tube is released

Fig. 7 Toggle switch for pneumatic seal

1 Mounted auto-protection assembly 3 Toggle switch for the pneumatic seal


2 Connection of the auto-protection assembly

19
Function and layout of the compEAct S compEAct S

Measuring gas dryer The measuring gas is dried by means of a membrane dryer which is attached to the
furnace. Oxygen is used as a purge gas for the operation of the membrane dryer. To
increase the drying effectiveness a pump is used to aspirate the rinse gas through the
membrane dryer.

Fig. 8 Membrane dryer

Detector The compEAct S uses a UV fluorescence detector. The UV radiation of a deuterium


lamp stimulates the fluorescence of the analysis gas which contains SO2. The intensity
of the fluorescence is determined with a photomultiplier (PMT).

compEAct SMPO The compEAct SMPO and its patented MPO technology were developed to determine
Detector with MPO the sulfur content without interferences in the presence of elevated nitrogen contents.
option For an optimal adaptation of the analysis process to the specific requirements of the
individual samples, it is possible to switch the MPO on and off as required using the
operating software.

20
compEAct S Function and layout of the compEAct S

Gas flow diagram

Fig. 9 Gas flow diagram of the compEAct S and compEAct SMPO

21
Function and layout of the compEAct S compEAct S

3.3 Type plate

The rating plate is located on the rear of the compEAct S next to the mains connection
and contains the following information:
 Manufacturer address / logo
 Trading name
 Technical data / electrical connection data
 CE marking
 Disposal instructions (Do not dispose of as domestic waste!)
 Serial number

3.4 Sample supply

3.4.1 Auto-injector AI

The auto-injector is used for the safe injection of individual samples up to a volume of
50 µl. For this purpose, the samples are injected directly into the combustion tube. The
AI is controlled via the software EAvolution.
Layout

Fig. 10 Layout of the auto-injector AI

1 Base element with mechanics and control 5 Syringe bracket


electronics 6 Syringe cylinder
2 Guiding device to push down the piston of the 7 Guide on the injection port
syringe in a controlled manner 8 Base plate
3 Syringe piston
4 Conical guide at the syringe

The interface on the rear panel of the device is used to control the device and to supply
it with electrical current. The AI is supplied with electrical current and controlled by the
compEAct S.

22
compEAct S Function and layout of the compEAct S

3.4.2 Autosamplers LS 1 and LS 2

The autosampler LS is an automatic sampler for liquid samples. It is available in


2 versions. The LS 1 is intended for small series with up to 18 samples. The samples
are arranged in a linear sequence on the fixed sample rack. The injector head for
taking up the samples and injecting them into the combustion tube moves on the rack
along the guide arm from left to right towards the injector port.
The LS 2 is designed for up to 120 samples. The samples are arranged in 8 rows on the
rack. The rack moves back and forth.
The injection port allots the samples that were taken up by the autosampler directly
into the combustion tube. The LS is controlled via the software EAvolution. The
maximum sample volume that can be metered is 50 µl.
Layout 1 Guide arm
2 Sample rack
3 Solvent and waste
containers
4 Injector head with
syringe drive

Fig. 11 Autosampler LS

Fig. 12 Sample rack of the LS 2

The solvent and waste containers can be removed to fill and clean the autosampler.
The solvent container has a removable lid to facilitate its cleaning. A waste hose is
attached to the waste container. This hose must be introduced to an appropriate waste
container. The solvents are absorbed and released via ports with integrated septum.

23
Function and layout of the compEAct S compEAct S

1 Solvent container
2 Waste container

Fig. 13 Solvent and waste containers of the LS

The electrical connections and interfaces of the LS are located on the rear panel of the
device. The connections for the serial interface and power supply are effected via
terminals on the rear panel of the compEAct S.
1 ON/OFF switch
2 Power connector
3 Interface to the
compEAct S
4 Functional grounding

Fig. 14 Electrical connections of the LS

24
compEAct S Installation and commissioning

4 Installation and commissioning

4.1 Installation conditions

4.1.1 Environmental conditions

Ambient conditions For requirements with regard to the ambient conditions at the installation location
refer to “Technical data” p. 71. If required, make sure that the room is temperature-
controlled.

Requirements with The installation location of the compEAct S should have the characteristics of a
regard to the installation chemical laboratory. The installation location must meet the following requirements:
location
 The laboratory atmosphere must be free from dust and sulfur compounds, nitrogen
compounds and halogenated compounds as well as corrosive gases and
hydrocarbons
 No major temperature fluctuations
 Free from caustic vapors in close proximity of the device
Connections and assemblies of the device may corrode.
 Free from draft, do not install the device close to windows or doors
 No electromagnetic sources of interference close-by
 No direct sunlight and away from radiant heaters
 The front door and air vents must not be obstructed by other equipment or
furnishings
 Keep a safety distance of at least 10 cm from the rear of the device to other
equipment or walls

4.1.2 Energy supply

WARNING
The compEAct S may only be connected to a properly grounded power outlet which
complies with the voltage specifications in the technical data on page 72.

The compEAct S is operated with single-phase alternating current. The installation of


the electrical equipment of the laboratory must comply with the standard DIN VDE
0100. After the connection point an electrical current in accordance with the standard
IEC 38 must be available.
Only use the supplied power cord with protective ground conductor to connect the
compEAct S to the power supply. The protective effect must not be invalidated by the
use of an extension line which does not have a protective conductor.

25
Installation and commissioning compEAct S

4.1.3 Gas supply

Oxygen and argon are required to operate the compEAct S. Refer to the technical data
(→ "Technical data", p. 71) for information on the required gas quality and gas
consumption.
The operator is responsible for the gas supply and the corresponding connections and
pressure reducers.
The connection hoses with outer diameter 6 mm and inner diameter 4 mm are
included with the delivery. The length is 2 m. If other lengths are preferred, please
contact the customer service department at Analytik Jena.

4.1.4 Device layout and space requirements

The compact compEAct S has a required floor space of 650 x 600 mm (including
clearance on the sides). The ventilation slits at the bottom and rear side of the device
must not be obstructed by any objects.
The auto-injector AI or the autosampler LS are mounted to the top of the compEAct S.
Therefore, the required height is the sum of the height of the compEAct S and the
height of the sample introduction module that is connected. There must be a clearance
of at least 10 cm between the composed device system and any shelve or cabinet
above the system.
Dimensions of the compEAct S and its device components:

Device Dimensions (width x height x depth) in cm


compEAct S 54 x 51 x 53
LS 1 / LS 2 54 x 41 x 34
AI 9 x 27 x 11

26
compEAct S Installation and commissioning

Fig. 15 Installation diagram for the compEAct S with autosampler LS

27
Installation and commissioning compEAct S

4.2 Supply and control connections

The electrical connections and connections for the gas supply are located on the rear
panel of the device.

Fig. 16 Connections on the rear of the compEAct S

1 Fan for furnace chamber and electronics 8 Mains connection


2 Connection for oxygen (O2) 9 Communication interface for AI and LS
3 Connection for argon (Ar) (sampler)
4 Exhaust air detector (exhaust) 10 Power connector for AI and LS (sampler)
5 Exhaust air membrane dryer (dryer exhaust) 11 Service connections
6 Carrying handles 12 Connections for internal computer
7 Device fuses

The device is equipped with a hose bridge on its right-hand side for connecting an
external flow meter (MFM) for servicing purposes. A USB port is located behind the
front door. It is provided for software updates or for backing up the database on a flash
drive.
1 USB port
2 Hose bridge (green)

Fig. 17 USB port and hose bridge in the gas path behind the front door

28
compEAct S Installation and commissioning

The mains switch is located behind the door, the ON/OFF switch for the device is
located on the right-hand side of the front skirting.

Fig. 18 Location of the switch on the compEAct S

1 Mains switch for emergencies or servicing 2 ON/OFF switch

Switch functions The mains switch is used to disconnect or to connect the compEAct S from the electric
mains. In normal operation the mains switch remains switched on. It only needs to be
switched off for maintenance works or in case of emergencies (→ "Safety equipment /
Behavior during emergencies", p. 15).
The ON/OFF switch is used to switch the compEAct S on and off. The following device
functions are activated when the device is switched on:
 The software EAvolution is started.
 The detector is switched on and the detector starts warming up.
 The method that was activated before switching the device off is loaded.
 The gas flows are regulated when the gas supply to the device is established.
 The combustion furnace is heated to the temperature that is specified in the
method.
When the device is switched off using the ON/OFF switch or the respective command
in the user interface of the EAvolution software the system is shut down:
 The combustion furnace heating is switched off.
 The detector is switched off.
 The EAvolution software is shut down.

29
Installation and commissioning compEAct S

4.3 Installing the compEAct S with sample introduction module

4.3.1 Setting up and connecting the compEAct S

WARNING
The compEAct S may only be set up, assembled and installed by the customer service
department of Analytik Jena or by specialist personnel authorized and trained by
Analytik Jena!
Any unauthorized intervention in the compEAct S can endanger the user and the
operational safety of the equipment and limits or completely invalidates any warranty
claims.

Attention
Retain the transport packaging! Return transport for service must be in the original
packaging. This alone prevents transport damage.

1. Carefully remove the compEAct S and its components from the transport
packaging. Do not damage the transport packaging!
2. Use the provided packing list to verify that the delivery is complete.
3. Place the compEAct S on its intended location.
4. Unscrew and detach the carrying handles. Store the carrying handles by clamping
them in the bracket on the rear panel of the device.
5. Attach the black skirting:
Push the skirting on the device from the front side. The guide pins on the skirting
must protrude into the openings on the front and the back of the compEAct S.
6. Attach the gas supply hoses to the connections on the rear panel of the device.
7. Install the combustion furnace (→ "Removing and installing the combustion
furnace", p. 55).
8. Install the auto-protection assembly (→ "Removing and installing the auto-
protection assembly", p. 48).
9. Connect the membrane dryer (→ "Replacing the membrane dryer", p. 46).
10. Install the combustion tube (→ "Inserting the combustion tube", p. 45).
11. Connect the autosampler LS or the auto-injector AI.
12. Plug in the mains plug to the connection at the rear of the device.
13. Plug in the mains cable to the mains power outlet.
The installation of the LPG module 2.0 and the combined GSS/LPG module is
described in the respective user manuals of these accessories.

30
compEAct S Installation and commissioning

4.3.2 Connecting the autosampler LS

Attention
Short circuits during connection can damage the LS
The autosampler LS may only be connected after the compEAct S was switched off
using the mains switch and the mains plug was detached from its connection on the
compEAct S!

1. Put the cover onto the opening above the combustion tube of the compEAct S.
2. Screw on the attachment angles provided with the autosampler LS to the
compEAct S.
3. Place the autosampler LS onto the compEAct S. Ensure that the rearward and the
left edge of the LS are flush fitting with the edges of the compEAct S.
4. Attach the autosampler LS to the attachment angles using the
plastic brackets. To do this, insert the nuts to the groove of the
plastic bracket. Do not yet tighten the connection. It must still be
possible to slightly move the autosampler back and forth within the
groove.

5. Establish the electrical connections between the autosampler LS


and the compEAct S:
 Power supply (1)
 Communication interface (2)
 Functional grounding (3)
6. Plug in the mains cable to the compEAct S (4).
7. Switch on the ON/OFF switch of the LS (5).

31
Installation and commissioning compEAct S

8. Insert the syringe to the injector head:


 Push the cannula through the cannula guide (3)
 Insert the ball knob of the piston into the bracket (1)
 Slightly pull out the knob and attach the syringe (2). The spring
on the knob pushes the syringe into the bracket.

9. Fill the solvent container with a maximum of 30 ml of solvent and


place it on the bracket of the LS together with the waste container.

Attention
Proceed with care when inserting the filled solvent container and do
not tilt the container too far to avoid spilling the solvent.
10. Introduce the waste hose to an appropriate waste container. Route
the hose in a constant downward slope.
11. Insert the sample rack.
12. Switch on the compEAct S and adjust the autosampler LS using the
EAvolution software (→ Adjusting the autosampler LS, p. 41)
13. Use the two screw connections to fix the autosampler LS in the
adjusted position.

 The installation of the compEAct S and the autosampler LS is now complete.

4.3.3 Connecting the auto-injector AI

Attention
The auto-injector AI may only be connected after the compEAct S was switched off
using the mains switch and the compEAct S is not connected to the mains!
To move the auto-injector AI do not hold it at the red front cover.

1. Remove the cover above the combustion tube of the compEAct S.


2. Place the auto-injector AI onto the opening at the top of the compEAct S. When
doing so, move the base plate until the ring-shaped seating fits onto the injection
port of the combustion tube.
The AI is correctly inserted when the ground plate completely sits in the opening
and is flush fitting with the surface of the compEAct S.

32
compEAct S Installation and commissioning

3. Establish the electrical connections between the auto-


injector AI and the compEAct S:
 Power supply (1)
 Communication interface (2)
4. Plug in the mains cable to the compEAct S (3).

 The installation of the compEAct S and the auto-injector AI is now complete.

33
Operation compEAct S

5 Operation

5.1 Switching the compEAct S on and off

Switching on via the Note: In normal operation the mains switch remains switched on. The compEAct S
mains switch only needs to be disconnected from the mains by actuating the mains switch in case of
emergency, for transport or during maintenance and servicing.
1. Open the external gas supply.
2. Switch on the mains switch behind the front door.
3. Push the ON/OFF switch on the skirting after 30 seconds.
 The device system starts up and the EAvolution software is started. The
method that was activated before switching the device off is loaded.

Switching on using the The mains switch is already switched on.


ON/OFF switch
1. Open the external gas supply.
2. Switch on the compEAct S using the ON/OFF switch.
 The device system starts up and the EAvolution software is started. The
method that was activated before switching the device off is loaded.

Switching off using the 1. Switch off the compEAct S using the ON/OFF switch. Alternatively, select the item
ON/OFF switch SYSTEM ▸ POWER OFF in the EAvolution menu.
 The combustion furnace heating is switched off. After 2 minutes, the gas
flows within the device are shut off and the detector is switched off.
2. Wait for 2 minutes, then cut the external gas supply.
 The fans keep running for 30 minutes to ensure that the compEAct S is
properly cooled down. During this time, the LED in the front door flashes.
After that, the LED is turned off and the device system is shut down.

Switching off the device For all maintenance and service work:
using the mains switch
1. First, switch off the device using the ON/OFF switch (see above).
2. Wait until the LED in the front door stops flashing.
3. Open the door and switch off the mains switch.
 The compEAct S is now completely switched off.

Attention
Switching off the compEAct S using the mains switch before shutting down the device
with the ON/OFF switch and before the LED in the front door has permanently
stopped flashing can cause permanent damage to the compEAct S due to overheating.
Furthermore, this may cause malfunctioning of the installed software.
Therefore, only switch off the device prematurely via the mains switch in case of
emergency!

34
compEAct S Operation

5.2 Measuring with auto-injector AI

CAUTION
Risk of injury when handling the cannula
The cannula of the syringe has a very sharp tip. Do not touch the syringe by the
cannula. This could result in injuries or contamination of the cannula.

Attention
The auto-injector AI may only be fitted with the syringes included in the scope of
delivery. The syringes are coded.
Ensure that when inserting the syringe into the auto-injector AI dosing does not start
prematurely.

1. Draw the sample into the AI syringe up to the required volume marking.
To do this, immerse the cannula into the sample solution, draw in and eject sample
solution several times to rinse the cannula before drawing the required volume.
Carefully wipe the outer face of the cannula with cellulose. Otherwise, sample
residues could contaminate the septum.
2. Start the measurement in EAvolution.
3. Remove the cover from the AI.

Attention
When inserting the syringe into the AI, do not
touch the piston and do not press it down!
4. Hold the syringe by the white plastic cone (2)
and pierce the cannula through the cannula
guidance (4) of the AI and through the
septum of the combustion tube.
5. Insert the cannula so far into the combustion
tube that the cone of the syringe fully rests on
the holder of the AI (3).
The positioning pin must sit in the cavity if the
syringe is in the right position.
6. Replace the cover.

7. Start measurement.
 The software controls the injection of the sample into the combustion tube
while the ram moves downward (1). The sample is analyzed.
8. Remove the syringe from the AI after the analysis.

35
Operation compEAct S

5.3 Measuring with LS 1 or LS 2

CAUTION
Risk of injury when handling the cannula! The cannula of the syringe has a very sharp
tip. Do not touch the syringe by the cannula. This could result in injuries or
contamination of the cannula.
There is a risk of injury within the movement range of the injector head during
operation. Keep a safe distance from the device.

Attention
The autosampler LS must be adjusted before starting the measurement (→ "Adjusting
the autosampler LS", p. 41).

1. Remove the injector head of the autosampler LS from the magnetic attachment.
2. Insert the syringe to the injector head:
 Push the cannula through the
cannula guide (3)
 Insert the ball knob of the piston into
the bracket (1)
 Slightly pull out the knob and rotate
into horizontal position (2). The
spring on the knob pushes the
syringe into the bracket and retains it
in this position.

3. Remove the solvent and waste


containers from the LS.

36
compEAct S Operation

4. Fill the solvent container (2) with a


maximum of 30 ml of solvent.
5. Empty and clean the waste container (1).
6. Assemble the containers and attach
them to the bracket of the LS.

Attention
Proceed with care when inserting the
filled solvent container and do not tilt the
container too far to avoid spilling the
solvent.
7. Route the hose from the waste container
in a constant downward slope to a
receptacle.
8. Insert the sample containers into the
rack.
9. Start the measurement in EAvolution.

37
Maintenance and care compEAct S

6 Maintenance and care

6.1 Maintenance intervals

Analyzer Maintenance task Maintenance interval


Check system tightness If the EAvolution software reports any leaks
Clean and maintain the compEAct S Weekly
Check hose connections for tight fit Weekly
Replace membrane dryer In case of soiling or damage
Replace deuterium lamp (UVFD) If the intensity of radiation is no longer
sufficient to generate satisfactory measuring
results
If the measurement sensitivity is no longer
sufficient
compEAct SMPO: Chemical ozone decomposer Annually
Or more often in continuous operation if an
ozone scent can be smelled at the "Exhaust"
gas outlet

Combustion tube Maintenance task Maintenance interval


Check injection port with septum for damage Daily and in case of abnormal measurement
and soiling results (wide scattering, carry-over,
excessively high measured values)
Quartz wool plug: check correct position Daily and after maintenance work on the
injection port
Replace quartz wool plug If the plug is damaged or soiled
In case of abnormal measured values (wide
scattering, carry-over, elevated blank values)
Check the combustion tube for cracks and Monthly and each time it is cleaned
damages
Replace combustion tube In case of devitrification, cracks or other
damages and soiling that cannot be
removed
Check the FAST connectors on the Monthly and in case of gas leaks
combustion tube inlet for tight fit, cracks or
other damages
Replace septum If leaking, damaged or contaminated

Auto-protection assembly Maintenance task Maintenance interval


Check filter Monthly and in case of abnormal
measurement results, in case of incomplete
combustion of the samples.
Replace filter In case of soiling and in case of abnormal
measurement results
Replace the pneumatic seal If leaking or damaged

38
compEAct S Maintenance and care

Autosampler LS 1 Maintenance task Maintenance interval


Autosampler LS 2 Adjust autosampler  When starting up the device
 After the LS was removed from the
device
 After replacing the combustion tube
Clean and maintain the device Weekly
Clean or replace syringe As required, but no later than after
12 months
Replace cannula If clogged or damaged

Auto-injector AI Maintenance task Maintenance interval


Clean and maintain the device Weekly
Clean or replace syringe As required, but no later than after
12 months
Replace cannula If clogged or damaged

The maintenance of the LPG module 2.0 and the combined GSS/LPG module is
described in the respective user manuals of these accessories.

6.2 Cleaning the compEAct S and its system components

WARNING
Risk of short circuit
Before cleaning the device, use the mains switch to switch off the device.
Do not use any dripping wet cloths for cleaning. No liquids are allowed to ingress the
device interior.

 Avoid contamination by handling sample substances carefully.


 Immediately wipe up spilled sample or reagents with blotting paper.
 Use a soft, clean cloth to wipe away any contaminations from the device. A
commercially available neutral cleaning agent (dishwashing liquid) can be used for
maintaining the casing.

6.3 Checking system tightness

The system's tightness is monitored automatically. If the system is leaking, the


message "device leaky" appears in the DEVICE STATE window. Starting a measurement is
not possible in this case.
Smaller leaks can be detected if the control flow in the system drops when applying
additional counter pressure. The flow monitoring set which is provided with the
compEAct S can be helpful for finding leaks. The set is assembled as follows (→Fig.
19):

39
Maintenance and care compEAct S

1. Plug the Luer adapter (d) onto the filter (c).


2. Connect the two short pieces of hose (b, e) to the Luer adapter and the filter
outlet.
3. Push the long thin hose (a) into the short hose (b).

Fig. 19 Flow monitoring set

Proceed as follows to start the device and the gas supply:


1. Open the gas supply to the compEAct S.
2. Switch on the compEAct S (→ "Switching the compEAct S on and off" p. 34).
3. Make sure that the pneumatic seal on the auto-protection assembly is closed. To
do this, move the toggle switch (1) down.
 After starting up the system, the status window of the EAvolution displays
the specified gas flows in the active methods:
Proceed as follows to check for leaks before the detector:
1. Read the control flow under SYSTEM  DEVICE  MAINTENANCE  CONTROL FLOW.
2. Detach the hose bridge (hose no. 7, 2 in Fig. 17) from one of the two
FAST connectors.
3. Attach the loose end of hose a (→Fig. 19) to the FAST connector.
4. Plug the other end of the flow monitoring set into the loose end of hose no. 7
using the hose e.
5. Re-read the control flow.
If the impact of the filter causes the control flow to reduce by more than 5 ml/min
there is a leak before the detector. Determine and correct possible causes. Contact the
customer service department at Analytik Jena if this is not successful.
The control flow may deviate from the indicated gas inflow (sum of Main O2 and argon
flow) because the control flow sensor responds differently to the different gases.

6.4 Adjusting the auto-injector AI

During the software-supported adjustment of the auto-injector the system adjusts the
number of steps that the stepper motor pushes down the syringe piston to the
geometry of the syringe that is used.
1. Select the command SYSTEM  DEVICE  MAINTENANCE  ADJUST AUTOINJECTOR.

2. Start the adjustment with .

40
compEAct S Maintenance and care

3. When being prompted, pull out the syringe piston up to the maximum volume
(50 µl) and confirm with OK.
 The auto-injector pushes the syringe down to 25 µl.

4. Read the actual value on the syringe piston and enter the value in the input field.
Confirm with OK.
 The auto-injector is now adjusted.

6.5 Adjusting the autosampler LS

The autosampler LS is mounted to the compEAct S and connected (→ "Connecting the


autosampler LS", p. 31).
1. Switch on the compEAct S and switch on the autosampler LS using the ON/OFF
switch on the rear panel of the LS.
2. Remove the cover of the injector head of the autosampler LS from the magnetic
attachment to facilitate the view on the syringe.
3. Put a closed sample container to position 1 of the sample rack.
4. Successively select the following commands in EAvolution: SYSTEM  DEVICE 
MAINTENANCE  ADJUST SAMPLER.
5. From the list REFERENCE POSITIONS select the 1ST RACK POSITION and tap on the
adjustment icon .
6. Use LEFT – RIGHT to align the cannula at the center of the septum in the sample
cover. Use UP – DOWN to lower the cannula to between 1 and 2 mm above the
bottom of the container.
Only for LS 2: Use BACKWARD –FORWARD to align the depth of the sample rack
position.

7. Select the sample position FURNACE and tap on the adjustment icon .
8. Use LEFT – RIGHT and UP – DOWN to position the cannula just above the injection port
of the combustion tube.
If the depth position of the needle is not yet centered in the combustion tube
loosen the LS attachment at the rear of the compEAct S and move the LS to the
required depth position. After that, retighten the screw connection (→ "Connecting
the autosampler LS", p. 31).

41
Maintenance and care compEAct S

9. Use UP – DOWN to lower the injector head until the cannula screw
connection of the metering syringe (1) is located in the cannula
guidance (2).
Only lower the injector head up to a level that allows moving the
cannula guidance up and down by hand by approximately 1 or
2 mm.

10. After the adjustment is completed, check the correct positioning in the adjustment
positions 1st RACK POSITION and FURNACE by tapping the icon to move to these
positions.

6.6 Combustion furnace maintenance

6.6.1 Removing and cleaning the combustion tube

CAUTION
Risk of burns on the combustion tube
Only remove the combustion tube when the device is cold or allow the device to cool
down sufficiently!

CAUTION
Use personal protective equipment (laboratory coat, protective gloves, safety goggles)
when cleaning the combustion tube and replacing the quartz wool. Wear a respiratory
mask or work underneath an exhaust vent because quartz wool dust irritates the
respiratory tracts.

1. First switch off the compEAct S from the ON/OFF switch and then, after the system
has been shut down completely, from the mains switch (→ "Switching the
compEAct S on and off" p. 34).
2. Clear the opening at the top of the compEAct S.
Autosampler LS: Slide the injector head over the rack. The injector head must only
be moved by hand when the compEAct S is switched off from the mains switch!
Remove the cover from the opening.
Auto-injector AI: Remove the Autoinjector AI from the compEAct S and put it aside.
Do not tear off the connecting pipes!

42
compEAct S Maintenance and care

3. Open the pneumatic seal on the auto-protection assembly. Move


the toggle switch up.
 This releases the combustion tube and it can now be pulled
out of the auto-protection valve assembly.

4. Pull hose 3 (a, blue), and hose 4 (b, gray) out of the FAST
connectors.

5. Open the two-part heat guard above the opening of the


combustion furnace.
6. Carefully pull the combustion tube out of the furnace.

7. Check the combustion tube for devitrification, cracks and blown out
spots.
Note: Only reuse intact combustion tubes.

8. If contaminated:
 Remove the screw cap with septum and the 2 angled FAST connectors.
 Remove the quartz wool plug from the combustion tube.
 Clean the inside of the combustion tube with a suitable solvent and a cotton
bud or a bottle brush. Clean the side arms with pipe cleaners. Rinse with
distilled water if the solvent mixes with water. Otherwise rinse with ethanol.
 Dry the combustion tube (e.g. by blowing it through with an inert gas).
 Push new quartz wool loosely up to the positioning mandrel in the inner tube
using a glass rod (→ "Replaceing quartz wool plug" p. 44).
Note: Product of incomplete combustion, such as soot, or solid pyrolysis residue can
deposit in the combustion tube. These deposits can also be removed by burning off
in the muffle furnace at 750 – 900 °C or with a suitable burner flame, e.g. propane
gas burner.
 The combustion furnace is cleaned.

43
Maintenance and care compEAct S

6.6.2 Replaceing quartz wool plug

ATTENTION
Only use the pure quartz wool supplied by Analytik Jena. Contaminated quartz wool
may damage the combustion tube and clog the filter.
Make sure that the quartz wool plug is correctly positioned. When the plug is not in
correct position the sample does not vaporize evenly.

1. Remove the combustion tube (→ "Removing and cleaning the


combustion tube", p. 42).
2. Roll a small amount of quartz wool into a loose plug approx. 2 to
4 cm long.

3. Insert the quartz wool plug with the glass rod supplied into the
inner tube of the combustion tube.

Attention
Clean all tools prior to use.
4. Push the quartz wool plug into the tube until the positioning
mandrel is at the center of the plug.

Attention
The plug must not seal the bottom slot in the inner tube.
The plug should cover the entire cross section of the inner tube.
When viewed from above the slot must no longer be visible in the
tube.

After replacing the quartz wool the analysis system must be cleaned through a
minimum of 3 measurements with pure solvent (e.g. isooctane, toluene, xylene).
Repeat the blind measurements until the nitrogen blind value is below 20 AU/µl
solvent (AU/µl: area unit per microliter of solvent). The blank value may vary
dependent on the solvent that is used.

44
compEAct S Maintenance and care

6.6.3 Inserting the combustion tube

WARNING
Risk of explosion or carbonization!
The gas hoses on the combustion tube must not be mixed up.

CAUTION
Risk of burns and possible damage of the auto-protection assembly!
Only install when the device is cold.

Attention
Alkaline salts (hand perspiration) cause devitrifications in the quartz glass when
heating the combustion furnace. These reduce the service life of the combustion tube.
Do not touch the combustion tube with your bare hands during installation. Wear
protective gloves. Wipe the combustion tube externally by with ethanol and cellulose
prior to installing it in the combustion furnace.

1. Slide the angled FAST connectors onto the gas connections of the combustion
tube.
2. Close the injection port of the combustion tube with septum and screw cap.
3. Wipe the combustion tube on the outside with a cellulose wetted with ethanol.
4. Open the two-part heat guard at top of the combustion furnace.

5. Slide the auto-protection assembly into the mount below the


furnace.
6. Insert the combustion tube into the combustion furnace.
7. When inserting the tube, change the position of the auto-protection
assembly until the combustion tube slides easily into the opening of
the assembly.

Attention
The combustion tube is fragile. If pressure is applied to the tube it
might break.

45
Maintenance and care compEAct S

8. In the gap between the combustion furnace and the auto-


protection assembly check that the combustion tube sits correctly in
the assembly.
9. Fit the heat guard in front of the auto-protection assembly. Hook
the heat guard into the knurled head screws at the combustion
furnace.

10. Slide hose 3 (a, blue), and hose 4 (b, gray) into the FAST
connectors at the combustion tube.

WARNING
Risk of explosion and carbonization! The hoses at the combustion
tube must not be mixed up!

11. Open the gas supply at the pressure reducer.


12. Close the pneumatic seal on the auto-protection assembly. Move
the toggle switch down.

13. Assemble the sampler module:


Auto-injector AI: Place the auto-injector AI on the combustion tube.
Autosampler LS: Place the cover over the opening to the combustion tube.
 The compEAct S is ready for operation and can be switched on from the
mains switch.

Attention
When using the autosampler LS: After servicing the combustion tube, the adjustment
of the injector head must be checked (→ "Adjusting the autosampler LS", p. 41).

6.7 Replacing the membrane dryer

CAUTION
Risk of burns on the combustion furnace
Only remove the membrane dryer when the device is cold or allow the device to cool
down sufficiently!

46
compEAct S Maintenance and care

Attention
Damage to the membrane dryer
The membrane dryer is equipped with a sensitive membrane for exchanging water
vapor. Squeezing or twisting the hose of the membrane dryer will damage the
membrane. The connectors are particularly sensitive. Make sure that the connectors
are not twisted during installation.

1. First switch off the compEAct S from the ON/OFF switch and then, after the system
has been shut down completely, from the mains switch (→ "Switching the
compEAct S on and off" p. 34).
2. Switch off the gas supply.
3. Open the pneumatic seal on the auto-protection assembly. Move the toggle switch
up.
4. Allow the device to cool down sufficiently.

5. At the top of the membrane dryer unscrew the hose olives at the
top and side connection and pull the hose adapter for the hoses 5
(a) and 12 (b) out of the connectors.

6. Remove the heat guard in front of the auto-protection assembly.

7. At the bottom of the membrane dryer unscrew the hose olives at


hose 13 (c) and at the connection of the auto-protection assembly.

47
Maintenance and care compEAct S

8. Remove the old membrane heater from the mount.


9. Carefully insert and clamp the new membrane dryer in the mount.
The gas connection at the top end must point to the right and the
gas connection at the bottom end to the left.
10. Screw the gas connections back on in reverse order. Whilst doing so,
hold the connections at the membrane dryer from the yellow shrink
tubes to prevent them from twisting.

Attention
The connections at the membrane dryer must never be twisted. The
sensitive internal membrane could be damaged and the membrane
dryer rendered useless.
 The membrane dryer is installed and ready for operation.

6.8 Servicing the auto-protection assembly

The following maintenance tasks must be performed at the auto-protection assembly:


 Check the filter for contamination: during maintenance of the combustion tube,
after incomplete combustion or carbonization of the system, in case of
overpressure faults and analytical problems.
 Replace the filter
 Replace the pneumatic seal: If the seal no longer seals.

6.8.1 Removing and installing the auto-protection assembly

CAUTION
Risk of burns on the combustion furnace
Only remove the auto-protection assembly (APB) when the device is cold or allow the
device to cool down sufficiently!

1. First switch off the compEAct S from the ON/OFF switch and then, after the system
has been shut down completely, from the mains switch (→ "Switching the
compEAct S on and off" p. 34).
2. Switch off the gas supply. Allow the device to cool down sufficiently.
3. Open the pneumatic seal on the auto-protection assembly. Move the
toggle switch up.
4. Remove the combustion tube from the combustion furnace (→
"Removing and cleaning the combustion tube" p. 42).

48
compEAct S Maintenance and care

5. Remove the heat guard in front of the auto-protection assembly.

6. Unscrew the union nut from the gas connection between the
membrane dryer and the auto-protection assembly. Whilst doing so,
hold the connections at the membrane dryer from the yellow shrink
tubes to prevent it from twisting.

Attention
The connections at the membrane dryer must never be twisted. The
sensitive internal membrane could be damaged and the membrane
dryer rendered useless.
7. Unplug the connector of the auto-protection assembly from the
electrical connection in the right side panel.
8. Pull the auto-protection assembly out of the mount under the
combustion furnace.

9. Press the ring of the connector of hose no. 11 down and pull the
hose out of the connection.
10. The installation of the auto-protection assembly is in reverse order.

6.8.2 Checking and replacing the filter

1. Remove the combustion tube (→ "Removing and cleaning the combustion tube",
p. 42).
2. Remove the auto-protection assembly (→ "Removing and installing the auto-
protection assembly", p. 48).

49
Maintenance and care compEAct S

3. Look into the opening of the seal and inspect the filter for
carbonization, other contamination or cracks.
If the filter is OK, re-install the auto-protection assembly.
If the filter needs replacing, follow the remaining instructions.

4. Unscrew the 4 screws attaching the pneumatic seal to the auto-


protection assembly.
5. Remove the seal from the assembly.

6. Remove the intermediate ring.

7. Remove the filter from the intermediate ring. If required, carefully


wipe the intermediate ring or the Teflon base element of the APB
to clean it.

8. The filter consists of a PTFE filter and a carrier fabric which are
pressed against one another. Insert the new filter to the assembly
with the carrier fabric facing downwards.
9. Insert the intermediate ring.
10. Reinsert the pneumatic seal and attach it using the 4 screws.
 The filter has been replaced. The auto-protection assembly
may now be reinstalled underneath the furnace (→
"Removing and installing the auto-protection assembly",
p. 48).

50
compEAct S Maintenance and care

6.8.3 Replacing the pneumatic seal

1. Remove the combustion tube (→ "Removing and cleaning the combustion tube",
p. 42).
2. Remove the auto-protection assembly (→ "Removing and installing the auto-
protection assembly", p. 48).
3. Unscrew the 4 screws attaching the pneumatic seal to the auto-
protection valve assembly.
4. Remove the housing of the pneumatic seal from the auto-
protection assembly.

5. Remove the gas connection. Insert a hexagon socket wrench into


the connection and unscrew the connection.

6. Remove the seal from the housing.


7. Pull the PTFE intermediate films from the seal.

8. Separate the seal from the ring.


9. Insert a new seal into the ring.

10. Insert the PTFE intermediate film into the housing.

51
Maintenance and care compEAct S

11. Insert the seal into the housing. The openings for the gas
connection in the sealing ring and housing must be above each
other.

12. Screw in the gas connection. Tighten with the hexagon socket
wrench.

13. Place the PTFE intermediate film onto the intermediate ring in the
auto-protection assembly.

14. Insert the pneumatic seal into the auto-protection assembly and
attach it with 4 screws.
 The pneumatic seal has been replaced. The auto-protection
assembly may now be reinstalled underneath the furnace (→
"Removing and installing the auto-protection assembly",
p. 48).

6.9 Opening and closing the side panel of the compEAct S

WARNING
Danger of electric shock!
Before opening the side panels, switch off the compEAct S using the mains switch
behind the front door and pull the mains plug from the connection on the rear panel of
the compEAct S.

The side panel must be opened for some maintenance tasks at the compEAct S.

52
compEAct S Maintenance and care

1. Unscrew the 3 hexagon nuts securing the side panel to the rear
panel of the device and remove the washers.

2. Push the side panel back until it slips out of the attachment.
3. Pull the functional grounding off the side panel. Place the side
panel aside.

4. To close the side panel first plug the functional grounding into the
housing and side panel.
5. First place the side panel against the housing in a way that the
clamping plate at the bottom of the panel sits behind the housing
frame and the top clamping plates offset with the top attachments
of the housing. The panel then sits flat against the housing.
6. Push the panel forward. The front clamping plate of the panel then
moves behind the housing frame. The screw pins at the rear of the
compEAct S protrude into the corresponding slots in the panel rail.
7. Push the washers onto the screw pin and re-attach the panel with 3 hexagon nuts
(see above).
 The compEAct S is closed again.

6.10 Replacing the deuterium lamp (UVFD)

WARNING
Danger of electric shock!
Before replacing the deuterium lamp, switch off the compEAct S using the mains
switch behind the front door and pull the mains plug from the connection on the rear
panel of the compEAct S.

Attention
Soiling of the lamp body reduces the intensity of radiation of the lamp!
Never touch the lamp body with bare hands. Wipe off dirt and fingerprints with a piece
of cellulose moistened with ethanol and allow the lamp to dry afterwards.

53
Maintenance and care compEAct S

1. First switch off the compEAct S from the ON/OFF switch and then, after the system
has been shut down completely, from the mains switch (→ "Switching the
compEAct S on and off" p. 34). Pull the mains plug at the rear of the device out of
the connection.
2. Open the side panel, pull the functional grounding off the side panel. Safely put
the side panel aside (→ "Opening and closing the side panel of the compEAct S" p.
52).
3. Pull the connector of the deuterium lamp out of the connection on
the circuit board.

4. Unscrew the screws for attaching the deuterium lamp.


5. Pull the lamp out of the socket.

6. Place a new deuterium lamp into the socket. The guide pin must
protrude into the groove of the lamp base. The lamp sits correctly in
the socket if the edge of the base sits level on the socket.

Attention
Never touch the lamp body with bare hands. Wipe off dirt and
fingerprints with a piece of cellulose moistened with ethanol and
allow the lamp to dry afterwards.

7. Attach the deuterium lamp with screws.


8. Plug the lamp connector into the interface on the circuit board.
9. Plug the functional grounding into the side panel. Close the side panel.
 The deuterium lamp has been replaced and the compEAct S is ready for
operation again.

54
compEAct S Maintenance and care

6.11 Replacing the chemical ozone decomposer (compEAct SMPO only)

The chemical ozone decomposer is part of the MPO option. It must be replaced no
later than when an ozone odor is noticeable at the "exhaust" gas outlet
1. First switch off the compEAct S from the ON/OFF switch and then,
after the system has been shut down completely, from the mains
switch (→ "Switching the compEAct S on and off" p. 34). Pull the
mains plug at the rear of the device out of the connection.
2. Open the front door and the inspection flap.
3. Pull the cartridge of the ozone decomposer out of the mount in the
inspection flap.
4. Pull the top and bottom FAST connectors off the cartridge.
5. Attach the hoses with FAST connectors to the new ozone
decomposer: top - hose no. 43, bottom - hose no. 44.
6. Close the inspection flap.

 The chemical ozone decomposer has been replaced. The compEAct S is ready
for operation again.

6.12 Removing and installing the combustion furnace

WARNING
Danger of electric shock!
Before removing and installing the combustion furnace, switch off the compEAct S
using the mains switch behind the front door and pull the mains plug from the
connection on the rear panel of the compEAct S.

CAUTION
Risk of burns on the furnace! Only remove when the device is cold!

6.12.1 Removing the combustion furnace

1. First switch off the compEAct S from the ON/OFF switch and then, after the system
has been shut down completely, from the mains switch (→ "Switching the
compEAct S on and off" p. 34). Disconnect the mains plug from the mains socket.
2. Open and unhook the front door.
3. Remove the combustion tube (→ "Removing and cleaning the combustion tube",
p. 42).

55
Maintenance and care compEAct S

4. Remove the auto-protection assembly (→ "Removing and installing the auto-


protection assembly", p. 48).
5. Release the top screw connections of the top connections at the
membrane dryer and pull out the gas hoses.
6. Unscrew the bottom screw connection for the rinse flow of the
membrane dryer.

7. Unscrew the 4 screws for holding the heat guard in front of the
furnace and remove the heat guard together with the membrane
dryer.

8. Open the inspection flap.


9. Unscrew the knurled head screw for furnace attachment to the
inside wall of the electronics chamber.

If the knurled head screw is difficult to reach through the inspection


flap:
 Open the side panel, pull the functional grounding off the side
panel and safely put the side panel aside (→ "Opening and closing
the side panel of the compEAct S" p. 52).
 Unscrew the knurled head screw.

56
compEAct S Maintenance and care

10. Unscrew furnace from the mount in the furnace space (2 screws).

11. Pull the connectors for the furnace heating and thermocoupler
(orange cable) from the connections under the furnace at the rear
of the furnace space.

12. Lift the furnace out of the furnace space and package it.
13. Close the inspection flap or plug the functional grounding into the side panel and
close the side panel. Hook the front door back in and close it.
 The furnace is completely removed.

6.12.2 Installing the combustion furnace

1. Check that the compEAct S is not connected to the mains!


2. Open and unhook the front door.
3. Insert the furnace into the furnace space.

4. Plug the connectors for the furnace heating and thermocoupler


(orange cable) into the connections under the furnace at the rear of
the furnace space.

57
Maintenance and care compEAct S

5. Open the inspection flap.


6. At the inside wall of the electronics chamber attach the furnace
with the knurled head screw. Do not tighten the screw yet.

If the screw connection is difficult to reach through the inspection flap:


 Open the side panel, pull the functional grounding off the side panel
and safely put the side panel aside (→ "Opening and closing the side
panel of the compEAct S" p. 52).
 At the inside wall of the electronics chamber attach the furnace with
the knurled head screw. Do not tighten the screw yet.

7. Screw the furnace with 2 Allen screws to the mount in the furnace
space.
8. Evenly tighten all 3 fastening screws (knurled head screw and Allen
screws).

9. Attach the heat guard with membrane dryer to the front of the
furnace with 4 Allen screws.
 The combustion furnace is fully installed.

58
compEAct S Troubleshooting

7 Troubleshooting

7.1 Troubleshooting according to software messages

Error message
100002 Communication with companion chip is disturbed.
Cause Remedy
The communication with the companion chip is Inform the customer service department.
disturbed.
101001 Main mfm defective
Cause Remedy
The flow meter is defective. Inform the customer service department.
The membrane dryer is defective. Replace the membrane dryer (→ "Replacing the
membrane dryer", p. 46).
101200 Control flow too high
Cause Remedy
The measured control flow is too low. Inform the customer service department.

101201 Device leaky


Cause Remedy
The device has a leak.  Check whether the pneumatic seal on the APB is
closed, also refer to chapter "Equipment fault",
p. 64.
 Use the flow monitoring set to check the tightness
of the system. Remove the leak (→ "Checking
system tightness", p. 39).
 Inform the customer service department.
101501 Communication with gasbox is disturbed.
Cause Remedy
The communication with the gas box is disturbed. Inform the customer service department.

101703 Hardware error in furnace temperature


Cause Remedy
A hardware error has occurred in the controller for Inform the customer service department.
the furnace temperature.
101723 Hardware error in furnace temperature
Cause Remedy
The measured temperature of temperature controller Inform the customer service department.
furnace temperature is out of specified sensor range.
101743 Hardware error in furnace temperature
Cause Remedy
The requested temperature could not be achieved by Inform the customer service department.
temperature controller furnace temperature.

59
Troubleshooting compEAct S

101760 Gas flow in Main-O₂ insufficient


Cause Remedy
The required gas flow could not be achieved with the  Check gas supply and connections.
Main-O₂ flow controller.  Inform the customer service department.
101764 Hardware error in flow sensor Main-O₂
Cause Remedy
A hardware error has occurred in the Main-O₂ flow  Inform the customer service department.
sensor.
101768 Hardware error in flow valve Main-O₂
Cause Remedy
A hardware error has occurred in the Main-O₂ flow  Inform the customer service department.
valve.
10176C Gas pressure error
Cause Remedy
Overpressure was detected in the gas path.  Wait until the routine for reducing the
overpressure is completed. Follow the instructions
of the software.
 Pierce a cannula into the septum of the injection
port so that the gas can escape via the cannula.
 Cut the gas supply.
 Search for cause for overpressure, correct the
defect.
 Reopen the gas supply and start the device.
102002 No connection to sampler LS I
Cause Remedy
It is not possible to establish a connection to the  Check the connections.
autosampler LS 1.  Inform the customer service department.
102003 No connection to autoinjector
Cause Remedy
It is not possible to establish a connection with the  Check the connections.
auto-injector AI.  Inform the customer service department.
102004 No connection to sampler LS II
Cause Remedy
It is not possible to establish a connection to the  Check the connections.
autosampler LS 2.  Inform the customer service department.
102100 No connection to LPG 2.0
Cause Remedy
It is not possible to establish a connection to the  Check the connections.
sample introduction module LPG 2.0.  Inform the customer service department.
102101 No connection to GSS/LPG
Cause Remedy
It is not possible to establish a connection to the  Check the connections.
combined GSS/LPG module.  Inform the customer service department.
102200 Drive of autoinjector is defective.
Cause Remedy
The drive or drive sensor of the AI have failed. Inform the customer service department.

60
compEAct S Troubleshooting

102210 Sampler error


Cause Remedy
A general sampler error has occurred (zero). Inform the customer service department.

102211 Error in sampler x-axis


Cause Remedy
The drive of the sampler on the X axis does not move Inform the customer service department.
or the sensor for the zero point has failed.
102220 LPG 2.0: argon system pressure too low.
Cause Remedy
LPG 2.0: The primary argon pressure is insufficient to Check gas supply and connections.
run the module.
102221 LPG 2.0: cooling defective
Cause Remedy
LPG 2.0: The cooling of the dosing valve has failed. Inform the customer service department.
102222 LPG 2.0: cooling temperature sensor defective
Cause Remedy
LPG 2.0: The temperature sensor of the cooling has Inform the customer service department.
failed.
102223 LPG 2.0: heating defective
Cause Remedy
LPG 2.0: The heating of the expansion chamber has Inform the customer service department.
failed.
102224 LPG 2.0: heating temperature sensor defective
Cause Remedy
LPG 2.0: The temperature sensor of the heating has Inform the customer service department.
failed.
102225 LPG 2.0: housing fan defective
Cause Remedy
LPG 2.0: The housing fan has failed. Inform the customer service department.
102230 GSS/LPG: argon system pressure too low.
Cause Remedy
GSS/LPG: The primary argon pressure is insufficient Check gas supply and connections.
to run the module.
102231 GSS/LPG: cooling defective
Cause Remedy
GSS/LPG: The cooling of the dosing valve has failed. Inform the customer service department.
102232 GSS/LPG: cooling temperature sensor defective
Cause Remedy
GSS/LPG: The temperature sensor of the cooling has Inform the customer service department.
failed.
102233 GSS/LPG: heating defective
Cause Remedy
GSS/LPG: The heating of the expansion chamber has Inform the customer service department.
failed.

61
Troubleshooting compEAct S

102234 GSS/LPG: heating temperature sensor defective


Cause Remedy
GSS/LPG: The temperature sensor of the heating has Inform the customer service department.
failed.
102235 GSS/LPG: heating defective
Cause Remedy
GSS/LPG: The housing fan has failed. Inform the customer service department.
105201 Communication with S-UVFD is disturbed.
Cause Remedy
The communication with the S-UVFD is disturbed. Inform the customer service department.

105202 Communication with S-UVFD+ is disturbed.


Cause Remedy
For compEAct SMPO: The communication with the S- Inform the customer service department.
UVFD+ is disturbed.
105230 S-UVFD: deuterium lamp defective
Cause Remedy
The deuterium lamp is defective or not connected  Check the connection of the connector of the
correctly. deuterium lamp on the circuit board.
 Replace the deuterium lamp (→ "Replacing the
deuterium lamp (UVFD)", p. 53).
 Inform the customer service department.
105231 S-UVFD: deuterium lamp not ignited
Cause Remedy
The deuterium lamp was not ignited. Inform the customer service department.

105232 S-UVFD: deuterium lamp temperature too high


Cause Remedy
The temperature of the deuterium lamp is too high. Inform the customer service department.

105233 S-UVFD: temperature sensor of deuterium lamp defective


Cause Remedy
The temperature sensor of the deuterium lamp has Inform the customer service department.
failed.
105234 S-UVFD: MPO module defective
Cause Remedy
For compEAct SMPO:The MPO module is defective. It Inform the customer service department.
cannot be switched on or off.
201100 Last active method not available.
Cause Remedy
The method that was active before switching off the Create a new method.
device is no longer available and cannot be activated.
201101 Transferring method parameters failed
Cause Remedy
The method parameters could not be transferred to Wait until the device is ready for measurements.
the device. The device may not be ready to carry out
the measurement.

62
compEAct S Troubleshooting

201200 Service septum in 300 injections


Cause Remedy
The maintenance interval for the septum is due after Replace the injection port with septum in due course.
300 additional injections.
203001 Automatic backup success
Cause Remedy
The automatic backup was completed successfully. None. The message is for informational purpose only.

203002 Automatic backup failure


Cause Remedy
The automatic backup has failed due to a system Restart the device. Repeat the backup process.
error.
204000 Calibration point import failed
Cause Remedy
The automatic import of a measured calibration point Repeat the process or inform the customer service
has failed. department, if required.
206003 Sequence canceled with errors
Cause Remedy
A sequence was aborted with errors. Check the sequence for errors. Resume the
measurement.
207001 Simulation activated
Cause Remedy
The simulation mode was activated. None. The message is for informational purpose only.

207002 Simulation deactivated


Cause Remedy
The simulation mode was deactivated. None. The message is for informational purpose only.

7.2 Equipment faults and analytical problems

7.2.1 General notes

Other problems not detected by the system monitoring can also occur. Starting a
measurement is possible. Such errors are usually detected on the basis of implausible
measuring results (analytical problems) or are clearly visible in the equipment
technology.
If the suggested solutions are not successful, inform the customer service department
of Analytik Jena.

63
Troubleshooting compEAct S

7.2.2 Equipment faults

Error Possible cause Remedy


Furnace does not heat Temperature set incorrectly in the Check temperature configuration in the
software method
No method loaded Load method
Malfunction in power supply  Switching the device on
 Check the internal fuse
Malfunction in the internal electronics Inform customer service department
Furnace temperature is Temperature controller faulty Inform customer service department
outside tolerance limits or
Malfunction in electronics
target temperature is not
reached
Process gases (inlet flow) Gas supply not connected / not opened Connect / open gas supply
are not supplied
Primary gas pressure too low Set primary gas pressure at the delivery point
to between 4 and 6 bar
Gas supply leaking Check gas supply
No method loaded Load method
Gas box faulty Inform customer service department
Target flow at the outlet to Connection between hose, FAST Check connection and ensure correct fit at the
the detector too low / connector and combustion tube connection points
message "Device leaky" incorrect
Pneumatic seal in the auto-protection  Check Ar supply
assembly does not seal combustion tube  Move toggle switch for pneumatic seal
down
Septum sits incorrectly in the injection Check position of the septum, insert new
port or is leaking septum if necessary
Connection of membrane dryer or Check connections (do not jam thread, tighten
transfer line to the auto-protection finger-tight)
assembly leaking
Gas escapes from Hose no. 11 connector loose Press hose no. 11 tightly into the quick-
pneumatic seal (audible release connector
hiss)
Pneumatic seal defective Replace pneumatic seal (→ "Replacing the
pneumatic seal", p. 51)
Auto-protection assembly Connector not connected Connect the APB connector (→ "Removing and
does not heat up installing the auto-protection assembly",
p. 48)
Heating of the APB defective Inform customer service department
Temperature controller faulty Inform customer service department
Door LED flashes, ozone Warm-up phase not yet completed Wait until warm-up phase is completed
generator off (approx. 30 min)
Device in standby Initialize the device
Process gases switched off Switch on gases
No method loaded Load method
Device state Wait until the device is ready for
TRANSFERRING METHOD PARAMETERS measurements

64
compEAct S Troubleshooting

Error Possible cause Remedy


Only for compEAct SMPO: Ozone decomposer not effective Replace ozone decomposer (→ "Replacing the
Ozone smell chemical ozone decomposer (compEAct SMPO
only)", p. 55).
Device leaking or defective Inform customer service department
Device operation via System crash Shut down system and switch off mains
touchscreen not possible or switch; wait for 30 s; switch on device again
restricted

7.2.3 Analytical problems

Error Possible cause Remedy


Combustion at the cannula Argon and oxygen connection at the Connect process gases correctly (→ "Inserting
combustion tube swapped the combustion tube", p. 45)
Septum damaged Replace septum in injection port
Lower results Metering fault Check metering
System leaking Check system tightness
Temperature set too low Check temperature configuration in method
Incorrect or unsuitable calibration Check calibration, if necessary recalibrate
Sample loss due to vaporization or Keep liquid samples sealed
spillage Check sampler function for solids
Post-combustion period insufficient Increase the post-combustion time
Decrease in lamp intensity Carry out a new calibration,
if necessary, replace the deuterium lamp
Excessively high results Incomplete combustion Clean or replace contaminated parts
TN interferences Analyze the sample with compEAct SMPO
Carryover Insufficient rinsing of the metering Rinse metering syringes adequately prior to
syringes in AI or LS sampling
Combustion tube not rinsed Rinse combustion tube adequately with clean
adequately solvent, i.e. blank measurements until values
are constant
Metering faulty Check metering
Contamination of injection port or Wipe or replace septum, clean combustion
combustion tube tube (→ "Removing and cleaning the
combustion tube", p. 42)
Sample contains inorganic sulfur Clean or replace contaminated parts
compounds
H2SO4 / H2SO3 condensation products Clean or replace affected assemblies
in the system (combustion tube, APB,
membrane dryer)
Incomplete combustion Clean or replace contaminated parts
Scattering measured values Metering faulty Check metering
Combustion tube contaminated or Clean or replace combustion tube
severely crystallized
Incomplete combustion Clean or replace contaminated parts
Clogged cannula in autosampler LS or Use the wire provided for this purpose to clean
auto-injector AI the cannula or replace it

65
Transport and storage compEAct S

8 Transport and storage

8.1 Preparing the device for transport

WARNING
Risk of damage to health due to improper decontamination!
Perform a professional and documented decontamination of the device before
returning it to Analytik Jena. The decontamination report is available from the
customer service department when registering the return. Analytik Jena must refuse
acceptance of contaminated devices. The sender may be liable for any damage caused
by inadequate decontamination of the device.

CAUTION
Risk of burning at the hot furnace! Only remove the combustion furnace when the
device is cold or allow the device to cool down sufficiently!

CAUTION
When removing the glass components there is a risk of injury from glass breakage!
Remove all glass components carefully from the analyzer!

8.1.1 Packing the autosampler LS

Attention
The autosampler LS may only be disassembled after the compEAct S was switched off
using the mains switch and the mains plug was detached from its connection on the
compEAct S!

1. Remove, empty and dry the solvent and waste containers.


2. Remove the sample rack and the sample containers.
3. Remove the metering syringe from the injector head and pack it separately.
4. Detach the electrical connections on the rear panel of the compEAct S and the
autosampler LS.
5. Put the autosampler LS into its original packaging.

66
compEAct S Transport and storage

8.1.2 Packing the auto-injector AI

Attention
The auto-injector AI may only be disassembled after the compEAct S was switched off
using the mains switch and the mains plug was detached from its connection on the
compEAct S!
To move the auto-injector AI do not hold it at the red front cover.

1. Detach the electrical connections on the rear panel of the compEAct S and the
auto-injector AI.
2. Remove the auto-injector AI from the compEAct S.
3. Pack the metering syringe separately.
4. Put the auto-injector AI into its original packaging.

8.1.3 Packing the compEAct S

CAUTION
There is a risk of burning on the combustion furnace!
Only remove the combustion tube when the device is cold! Allow the compEAct S to
cool down before packing it!

1. Shut down the compEAct S using the ON/OFF switch.


2. Open the pneumatic seal on the auto-protection assembly. Move the toggle switch
up.
3. Wait 30 min. until the fans no longer rotate and the door LED goes out. Then
switch off the mains switch behind the door. Allow the device to cool down.
4. Cut the external gas supply.
5. Remove auto-injector AI or autosampler LS from the compEAct S and pack them.
6. Detach all connections on the rear panel of the compEAct S.
7. Remove the combustion tube (→ "Removing and cleaning the combustion tube",
p. 42).
8. Remove the auto-protection assembly (→ "Removing and installing the auto-
protection assembly", p. 48).
9. Remove the membrane dryer (→ "Replacing the membrane dryer", p. 46).
10. Remove the combustion furnace (→ "Removing the combustion furnace", p. 55).
11. Pack open hose ends in protective bags and attach them with adhesive tape.
12. Close the door of the compEAct S.
13. Pull off the black skirting towards the front of the device.
14. Take the carrying handles from the bracket on the rear panel of the compEAct S
and screw them hand-tight into the lateral openings at the base of the device.

67
Transport and storage compEAct S

15. Place the top cover and secure it with adhesive tape.
16. Carefully pack the combustion tube and additional accessories in their original
packaging. Especially the glass components must be packed safely against
breakage.

8.2 Transporting the compEAct S

Observe the safety instructions in chapter “Safety instructions – Transport and


commissioning” p. 11. Avoid the following during transport:
 Impacts and vibrations
Risk of damage due to shock, impacts or vibrations!
 Major temperature fluctuations
Risk of condensate build-up!

8.3 Moving the compEAct S in the laboratory

CAUTION
Dropping the device poses a risk of injury and damage to the device!
Move the compEAct S extremely carefully! Two persons are required to lift and carry
the device. The device may only be lifted by the carrying handles.

1. Shut down the compEAct S using the ON/OFF switch. Wait 30 min. until the fans
no longer rotate and the door LED goes out. Then switch off the mains switch
behind the door. Allow the device to cool down.
2. Close the door.
3. Switch off the gas supply.
4. Remove any loose parts:
Autosampler LS: Remove the solvent and waste containers and the sample rack.
Auto-injector AI: Remove the AI and unplug the connections on the rear panel of
the compEAct S.
5. Detach all connections on the rear panel of the compEAct S.
6. Pull off the black skirting towards the front of the device.
7. Take the carrying handles from the bracket on the rear panel of the compEAct S
and screw them into the lateral openings at the base of the device.
8. Move the device with 2 people who must lift and hold the device by its carrying
handles.
The following must be observed for the compEAct S:
 Observe the guide values and adhere to the legally mandated limits for lifting and
carrying without auxiliary means!

68
compEAct S Transport and storage

 For installation at the new location observe the notes in section “Installation
conditions” p. 25.

Fig. 20 compEAct S with screwed-in carrying handles

8.4 Storage

Attention
Environmental influences and condensate formation can destroy individual
components of the device!
The device may only be stored in air-conditioned rooms. The atmosphere must be low
in dust and free from aggressive vapors.

If the device is not installed immediately after delivery or not required for prolonged
periods, it should be stored in its original packaging. A suitable desiccant should be
added to the equipment to prevent damage from moisture.
Ambient conditions For requirements with regard to the ambient conditions at the storage location refer to
“Technical data” p. 71.

69
Disposal compEAct S

9 Disposal
At the end of its service life the compEAct S and all its electronic components must be
disposed of as electronic waste in accordance with the applicable regulations.
The chemical ozone decomposer contains metal oxides. Used cartridges must be
disposed of in accordance with local regulations.

70
compEAct S Specification

10 Specification

10.1 Technical data

10.1.1 Technical data of the compEAct S

General characteristics Designation/type compEAct S / compEAct SMPO


Dimensions (width x height x depth) 54 x 51 x 53 cm
Mass approx. 30 kg

Procedural data Digestion principle Pyrolysis followed by thermal oxidation


Digestion temperature 700 to 1100 °C
Analysis parameter Total sulfur TS
Detection principle UV fluorescence of SO2
Sample volume 50 µl

Gas supply Oxygen Quality 4.5


Pressure 4 to 6 bar
Consumption 1000 ml/min
Argon Quality 4.6
Pressure 4 to 6 bar
Consumption 100 to 250 ml/min (depending on the method)

Electrical variables Connection 100 to 240 V ~, 50/60 Hz


Fuses 12 A T
Power consumption Max. 1100 VA
Operating system of the internal computer Linux

Ambient conditions Temperature range 21 – 35 °C


during operation Humidity Max. 90% at +30 °C
Air pressure 0.7 – 1.06 bar

Ambient conditions Temperature range 15 – 55 °C


during storage Humidity 10 to 30 % (use desiccant)

71
Specification compEAct S

10.1.2 Technical data of the auto-injector AI

General characteristics Designation/type Auto-injector AI (type AI-SC)


Dimensions (width x height x depth) 9 x 27 x 11 cm
Mass 1.5 kg
Sample volume 50 µl

Electrical variables Connection with 24 V  2.0 A

10.1.3 Technical data of the autosampler LS

General characteristics Designation/type LS 1 LS 2


Dimensions (width x height x 54 x 41 x 34 cm
depth)
Mass 4.5 kg 5 kg
Number of samples 18 120
Capacity of the sample containers 2 ml

Electrical variables Connection with 24 V  2.0 A

10.2 Guidelines and standards

Safety class and safety The compEAct S is classified as protection class I. The casing is protection type IP 20.
type

Device safety The compEAct S complies with the safety standards


 EN 61010-1
 EN 61010-2-081
 EN 61010-2-010

EMC compatibility The compEAct S has been checked for interference emission and noise immunity.
 It meets the requirement for transient emissions according to EN 61326-1
(EN 55011, group 1, class B)
 It meets the immunity requirements according to EN 61326-1 (requirements for
use in a basic EMC environment)

Environmental The compEAct S has been tested for environmental compatibility and meets the
compatibility requirements of
 ISO 9022-3
 ISO 9022-2

72
compEAct S Specification

EU directives The compEAct S is designed and tested in accordance with standards meeting the
requirements of EU directives 2014/35/EU, 2014/30/EU and 2011/65/EU. The
device leaves the factory in a sound condition as far as technical safety is concerned.
To maintain this condition and to ensure safe operation, the operator must strictly
observe the safety and operating instructions contained in this operating manual. For
accessories delivered with the device and system components from other
manufacturers, the information in the respective operating instructions have priority.

Guidelines for China The device contains substances subject to regulation (according to directive
"Management Methods for the Restriction of the Use of Hazardous Substances in
Electrical and Electronic Products"). Analytik Jena guarantees that the substances will
not leak within the next 25 years and, therefore, will not cause any environmental
danger or health risk within this period of time if the device is used as intended.

73

You might also like