Electrochemical O3 Sensor
(Model: ME2-O3-16×15)
Manual
Version: 1.1
Valid Date: 2016-05-22
Zhengzhou Winsen Electronics Technology Co., Ltd
Statement
This manual copyright belongs to Zhengzhou Winsen Electronics Technology Co., LTD. Without the
written permission, any part of this manual shall not be copied, translated, stored in database or
retrieval system, also can’t spread through electronic, copying, record ways.
Thanks for purchasing our product. In order to let customers use it better and reduce the faults
caused by misuse, please read the manual carefully and operate it correctly in accordance with the
instructions. If users disobey the terms or remove, disassemble, change the components inside of
the sensor, we shall not be responsible for the loss.
The specific such as color, appearance, sizes &etc, please in kind prevail.
We are devoting ourselves to products development and technical innovation, so we reserve the
right to improve the products without notice. Please confirm it is the valid version before using this
manual. At the same time, users’ comments on optimized using way are welcome.
Please keep the manual properly, in order to get help if you have questions during the usage in the
future.
Zhengzhou Winsen Electronics Technology CO., LTD
Zhengzhou Winsen Electronics Technology Co., Ltd www.winsen-sensor.com
ME2- O3- 16x15 O3 gas sensor
ME2-O3-16x15 electrochemical gas sensor is fuel cell type, which utilizes the electrochemical oxidation process
of O3 gas on the working electrode. The current produced in electrochemical reaction of the target gas are in
direct proportion with its concentration while following Faraday law, then concentration of the gas could be get
by measuring value of current.
Features
* Low consumption
* High precision
* High sensitivity
* Wide linear range
* Good anti-interference ability
* Excellent repeatability and stability
Application
Widely used in the detection of ozone gas concentration for civilian and environmental protection fields
Technical Parameter
Detection gas O3
Measurement Range 0~10/0~100ppm
Max detecting concentration 200ppm
Sensitivity ≥30nA/ppm
Resolution ratio ≤0.02ppm
Response time(T90) ≤120S
Load resistance 200Ω(recommended)
Repeatability <2﹪output value
Output Linearity linear
Zero drift(-20℃~40℃) ≤0.2ppm
Storage temperature -20℃~50℃
Storage Humidity 15%~90%RH
(no condensation)
Pressure range Standard atmosphere
pressure ±10%
Anticipated using life 2 years(In air)
Tel: 86-371-67169097 Fax: 86-371-60932988 Email: sales@winsensor.com
Leading gas sensing solutions supplier in China
Zhengzhou Winsen Electronics Technology Co., Ltd www.winsen-sensor.com
Testing Circuit
Characterization
Features of Sensitivity, response and resume Linearity features
Output upon variable temperature Zero point output upon Variable Temperature
Tel: 86-371-67169097 Fax: 86-371-60932988 Email: sales@winsensor.com
Leading gas sensing solutions supplier in China
Zhengzhou Winsen Electronics Technology Co., Ltd www.winsen-sensor.com
There are two ways to fix the ME2-O3-16×15 sensor
Fix the sensor in PCB or instrument shells through fixing hole by using nut
Fix the sensor through the fixing slot
Fixing Hole
Fixing Slot
Cross interference gas:
ME2-O3-16x15 sensor also responds to other gases besides target gas. Below are the response characteristics of
interferential gases
Gas Concentration ME2-O3
CL2 3ppm 1ppm
H2S 15ppm <-1ppm
CO 300ppm 0ppm
SO2 5ppm 0ppm
HCN 10ppm 0ppm
HCL 5ppm 0ppm
NO 35ppm 0ppm
NO2 5ppm 2ppm
H2 500ppm 0ppm
C2H4 100ppm 0ppm
8. Application Notes:
Do not break and bend the PCB terminals;
To preheat for no less than 48hs before using;
Do not take apart the sensor, otherwise electrolyte leakage can cause sensor damage;
Sensor shall avoid organic solvent, coatings, medicine, oil and high concentration gases;
All electrochemical sensors shall not be encapsulated completely by resin materials, and shall not immerse
in pure oxygen environment, otherwise, it will damage the function of sensor;
All ME sensors shall not be applied in corrosive gas environment, or the sensor will be damaged;
Please test the zero point of gas sensors in clean atmosphere;
Tel: 86-371-67169097 Fax: 86-371-60932988 Email: sales@winsensor.com
Leading gas sensing solutions supplier in China
Zhengzhou Winsen Electronics Technology Co., Ltd www.winsen-sensor.com
Sensors shall be avoided to face the gas, which flow directly from front side;
Blowhole of the sensor should not be blocked and polluted, which will cause the sensitivity decrease;
Do not open and damage the above waterproof breathable membrane of the sensor;
Excessive impact or vibration should be avoided;
Do not use the sensor when the shell is damaged;
It takes some time for the sensor to return to normal state after applied in high concentration gas;
Working electrode and reference electrode of the sensor shall be in short circuit when stored;
Please do not encapsulate the sensor using melt adhesive and sealant, whose curing temperature is higher
than 80°C;
Please do not expose and use the sensor in high concentration alkaline gas for a long time.
Note: To keep continual product development, we reserve right to change design features without prior notice!
Zhengzhou Winsen Electronics Technology Co., Ltd
Add.: NO.299 Jin Suo Road, National Hi-Tech Zone,
Zhengzhou, 450001 China
Tel.: 0086-371-67169097
Fax:0371-60932988
Site: www.winsen-sensor.com
E-mail:sales@winsensor.com
Tel: 86-371-67169097 Fax: 86-371-60932988 Email: sales@winsensor.com
Leading gas sensing solutions supplier in China