TSN:101106 Approved Date:01.
2013
Autoclave Electric
S.N. Purchaser’s Specifications
Autoclave Electric
Manufacturer
Brand
Type/Model
Country of Origin
1 Description of Functions
1.1 Pressure cooker type portable steriliser unit for sterilising of surgical instruments and dressing
materials by means of steam.
2 Operational Requirements
2.1 It shall be mains electrically powered.
2.2 Shall be used with distilled water.
3 System Configurations
3.1 Pressure cooker type portable steriliser unit, 1 unit
3.2 Dressing drums made of seamless stainless steel construction, size of 275mm H x 240 mm
diameter, suitable to fit into the autoclave, 3 units
3.3 Dressing drums made of seamless stainless steel construction, size of 225mm H x 225mm
diameter, suitable to fit into the autoclave, 3 units
3.4 Floor stand, 1 unit
3.5 Graduated water measuring jug, to enable filling of steriliser with correct quantity of water, 1
unit
4 Technical Specifications
4.1 Operating temperature 121 0C – 134 0C pressure 1.1 to 2.2 kg/cm2 of steam pressure
4.2 The pressure cooker type portable steriliser shall be operated by mains electricity and shall be
used with distilled water.
4.3 Constructed of heavy-duty spun aluminium (preferably stainless steel), cylindrical shape
(Minimum Diameter 300 mm X Depth app 500 mm) and must be seamless construction to
prevent bacteria residue and dirt accumulation.
4.4 Lid have spring loaded safety valves - pressure relief (steam release) valves, dead weight type
safety valve, automatic over pressure safety valve and dial type pressure gauge (0 – 60 PSI) and
must be sealed the autoclave with joint less neoprene gasket,
4.5 Unit to include fitted spacing shelf/trivet above heating element to prevent contact of contents
with heating element. Electrical heating element to have over-temperature protection/cut out and
Maximum electrical power must not to exceed 4KW.
4.6 The life span of all items required in this document, excluding wear-and-tear parts or
consumables, shall be a minimum of 5 years from the date of successful hand-over.
5 Accessories, Spare Parts and Consumables
5.1 All standard accessories/consumables/parts required for the proper operation of the above item
shall be included in the offer. Bidders shall specify, in a separate Excel worksheet, the quantity
and details of any items included in this offer which have not been specified in this Technical
Specifications Form.
5.2 All standard Maintenance tools and cleaning /lubrication materials where applicable shall be
included. Bidders shall specify, in a separate Excel worksheet, the quantity and details of any
items included in this offer which have not been specified in this Technical Specifications
Forms.
5.3 Additional 2 spare lid gaskets
6 Operating Environment
S.N. Purchaser’s Specifications
6.1 Power supply: 220 – 240V AC, 50Hz fitted with appropriate plug type D 3pins. The power
cable must be at least 3 metres in length.
7 Standards & Safety Requirements
7.1 Must submit ISO 9001 or ISO 13485:2003/AC: 2007 AND
7.2 CE (93/42EEC Directives) or USFDA approved product certificate.
8 User Training:
8.1 The Supplier shall conduct user training for this equipment to enable operators to use the
equipment properly. The training shall include the use of all operational functions of the
equipment, as well as routine checks and maintenance expected by users.
9 Warranty
9.1 The warranty period for this item shall be 12 months after acceptance of the Goods
10 Maintenance Service During Warranty Period
10.1 Preventive and corrective maintenance services during warranty period shall be included.
11 Installation and Commissioning
11.1 It shall be installed and commissioned by the Supplier at the final destination(s),
12 Documentation
12.1 It must be supplied with detailed operating and maintenance manuals and technical information
in the English language.