Specification

Dis play
Display Type: 8 inch high resolution LCD display
Sweep Speed: 25mm/sec
Information : HR, Lead/Pads, On/Off Alarm, Sp02, AED Functions and Prompts, Alarm Se-
lection and Limits, Del ivered Energy.

Defibrillator
Waveform : Biphasic
Charge Time: Less than 7 seconds with a new fully charged battery.
Energy Display: Monitor display indicates both selected and delivered energy.
Charge Prompt Type: Voice and visual prompts.
Electrode Impedance Measurement Range: 0-250 ohms.

ECG Monitoring
Patient Connection: 5- lead ECG cable, or 3-lead ECG cable, paddles.
Lead Selection: Displayed on monitor, paddles, 1, Il, Ill , AVR, AVL, AVF, V.
ECG Size: 0.25, 0.5, 1, 1.5, 2, 4 cm/mV diplay on monitor.
Heart Rate: 20-300BPM.
Heart Rate Alarm : On/Off displayed on monitor, user-selectable.
Smart Alarms: Beeper/voice prompts indicate shockable rhythm .

Recorder
Paper: 50mm thermal.
Speed: 12.5mm/sec, 25mm/sec, 50mm/sec. User-selectable 6-second delay.
Printing Method: High- resolution, thermal print head.
Print-out Modes: Manual or automatic, user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alarm activation or defibrillator charge
or defibrillator discharge.

Battary
Type: Rechargeable, Ni- MH battery, 12V.
Operating Time : For a new, fully charged battery: 60 defibrillator discharges at maximum
energy, or 3 hours minimum of continuous ECG monitoring.
Additional parameters will effect operating time with different functions.
AED Mode
AED Function: Auto analyze and charge X3 with programmable auto energy level selection,
screen prompts, and voice prompts.
Shockable Rhythms : Ventricular fibrillation with amplitude >=200UV, ventricular tachycardia
with rates >=140bpm,
and ORS complex wave duration >=140ms.
Charge Control: Control on deviee front panel, press key on paddle.
Prompts: Voice and visual prompts.

Manual Mode
Energy Selection: Selectable at 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 joules.
Synchronized Mode: Synchronizes defibrillator pulse to patient's R-wave. "SYNC" message
displayed on monito
Specification

Specification MB _pacer (Optional)


Type Wl
Pulse Type rectangular,constant current
Pulse Amplitude 0 to 180 mA± 10% or 5 mA (whichever is greater).
Variable from 30 ppm to 180 ppm ±1.5% (increments or
Pacing Rate
decrements by a value of 2 ppm)
Multi-Function Electrode (MFE)Pads multipurpose defibrillation/pacing electrodes
Pacing pulse frequency reduced by a factor of 4 when
Pause
activated
Refractory Period NC(VVI demand do not have this specification)
Pulse Width (20+ 1.5 ms)
Output Protection Fully defibrillator protected and isolated

Specification MB_EtC02 (Optional )


Type of sensor By-pass
Technical principle Non-dispersiveinfraredgas analysis NOIR
Storage condition -40 oc
to 70°C, <90% RH , non-condensing
Operating conditions 5 °C to 50 °C, 10 to 90% RH , non-condensing
Ambient pressure 55-115kPa
Power supply 5 V±S% (max ripp le 200 mVp-p)
Typical value120mA Excursion calibration typical
TOP
value280mA
Range 0-19.7% ( 0 - 150mmHg,or 0-20kPa )
Resolution 0.1mmHg
0 - 40 mmHg ± 2 mmHg
Accuracy 41 - 70 mmHg ± 5% ofreading
71 - 100 mm Hg ± 8% of reading
101 - 150 mmHg ± 10% of reading
Respiratory rate 2-150 BPM
Respiratory rate measurement
1% ±1BPM
accuracy

Specification MB_NIBP (Optional)


Measurement un ti: mmHg/kPa
Measurement Range : A duit: 10-270 mmHg/kPa
Pediatric:10-200 mmHg/kPa
Neonatal: 10-135 mmHg/kPa
Resolution 1 mmHg
Accuracy: Maximum mean error:S mmHg
Maximum standard deviation:8 mmHg
Type of sensor By-pass
Technical principle Non-dispersive infrared gas analysis NOIR
Storage condition -40 oc to 70°C, <90% RH, non-condensing
Operating conditiqns 5 oc to 50 °C, 10 to 90% RH., oon.-condensing
:·· ·.· ·.:.:·:::... ': .. .::-:.::·:.
Specification

Ambient pressure 55-115kPa


Power supply 5 V ±5% (max ripple 200 mVp-p)
TOP Typical value 120mA Excursion calibration typical value 280mA
Range 0-19.7% ( 0-150mmHg,or 0-20kPa)
Resolution 0.1mmHg
Accuracy 0- 40 mm Hg ± 2 mmHg
41 - 70 mm Hg± 5% of reading 71 - 100 mm Hg± 8% of reading
101 - 150 mm Hg ± 10% of reading
Respiratory rate 2-150 BPM
Respiratory rate
measurement accuracy 1% ±1BPM

Specification SP02 Moudle (Optional)


Measurement Range : 30 ~ 100% , ±2% between 80% ~ 90%,0thers ±5%
Alarm Range User set high limit and low limit
Alarm Accuracy ±1 0 within setting values.
Alarm Time Accuracy Less than 12 sec.

Product Feature
Battery Level lndicator Biphasic Technology

With impedance compensa-


To monitor the
tion More effective
battery real time
Less Energy and less hu rt to
t he heart

Internai Thermal Printer Data Storage

·~-.. -.·~.-, ... ,., .........---\' 65 hours of ali measured


SOmm integrated
t hermal recorder pararn eters

Paddle:

Charging and shocking


Quickly converted from adult
ca n be eas ily to pediatrie by removing t he
operated t hrough
outer su rface
according buttons