0% found this document useful (0 votes)
12 views70 pages

Proline Fieldcheck

Uploaded by

Glady T Davianus
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
12 views70 pages

Proline Fieldcheck

Uploaded by

Glady T Davianus
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 70

Operating Instructions

Proline Fieldcheck
Flow simulator

E
+
-

BA067D/06/en/04.04
50100085
Valid as of software version
V 1.02.XX
Proline Fieldcheck Table of contents

Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 4 8.1.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . 25
8.1.4 Performance characteristics . . . . . . . . . . . . 25
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 8.1.5 Operating conditions (Environment) . . . . . . 25
1.2 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4 8.1.6 Mechanical construction . . . . . . . . . . . . . . 26
1.3 Notes on operating Fieldcheck in an environment with 8.1.7 Display and user interface . . . . . . . . . . . . . 26
strong electromagnetic interference . . . . . . . . . . . . . 4 8.1.9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . 26
1.3.1 Radiated RF noise (EN 61326-1) . . . . . . . . . 4 8.1.10 Documentation . . . . . . . . . . . . . . . . . . . . . 26
1.4 Notes on safety conventions and icons . . . . . . . . . . . 5
9 Description of device functions -
2 Identification . . . . . . . . . . . . . . . . . . . . 6
General section . . . . . . . . . . . . . . . . . . 27
2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1.1 Fieldcheck nameplate . . . . . . . . . . . . . . . . . 6 9.1 Notes on the description of device functions . . . . . . 27
2.1.2 Simubox nameplate . . . . . . . . . . . . . . . . . . . 6 9.1.1 Using the table of contents to locate a function
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 description . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.3 CE mark, Declaration of conformity . . . . . . . . . . . . . 7 9.1.2 Using the graphic of the function matrix to
2.4 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . 7 locate a function description . . . . . . . . . . . 27
9.2 Function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.3 Block USER INTERFACE . . . . . . . . . . . . . . . . . . . . 29
9.3.1 Group CONTROL . . . . . . . . . . . . . . . . . . . 29
3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.4 Block BASIC FUNCTIONS . . . . . . . . . . . . . . . . . . . 31
3.2 Connecting to Promag 10 . . . . . . . . . . . . . . . . . . . . 9 9.4.1 Group SENSOR PARAMETER . . . . . . . . . . 32
3.3 Connecting to Promag 50/53 compact version . . . 10 9.4.2 Group OUTPUTS . . . . . . . . . . . . . . . . . . . . 33
3.4 Connecting to Promag 50/53 remote version . . . . 11 9.4.3 Group PROCESS PARAMETER . . . . . . . . . 34
3.5 Connecting to Promass 40/80/83 compact version 12 9.4.4 Group READ IN PARAMETERS . . . . . . . . . 35
3.6 Connecting to Promass 80/83 remote version . . . . 13 9.5 Block SUPERVISION . . . . . . . . . . . . . . . . . . . . . . . 36
3.7 Connecting to Prowirl 72 . . . . . . . . . . . . . . . . . . . 14 9.5.1 Group RESULTS . . . . . . . . . . . . . . . . . . . . 36
3.7.1 Prowirl 72 wiring diagram . . . . . . . . . . . . . 15 9.5.2 Group VERSION INFO . . . . . . . . . . . . . . . 38
3.8 Assignment of Fieldcheck connections . . . . . . . . . . 16
3.9 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 16 10 Description of device functions -
4 Operation . . . . . . . . . . . . . . . . . . . . . . 17 Fieldcheck and electromagnetic
4.1 Display and operating elements . . . . . . . . . . . . . . . 17
measuring devices . . . . . . . . . . . . . . . . 39
4.1.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 10.1 Block FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.1.2 Display icons . . . . . . . . . . . . . . . . . . . . . . . 19 10.1.1 Group SIMULATION . . . . . . . . . . . . . . . . . 40
4.2 Brief operating instructions on the function matrix . 20 10.1.2 Group VERIFICATION . . . . . . . . . . . . . . . 44
4.2.1 General notes . . . . . . . . . . . . . . . . . . . . . . 20
4.3 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . 21 11 Description of device functions -
4.3.1 FieldTool™ operating program . . . . . . . . . 21
Fieldcheck and Coriolis measuring
5 Commissioning. . . . . . . . . . . . . . . . . . 22 systems . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 11.1 Block FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 11.1.1 Group SIMULATION . . . . . . . . . . . . . . . . . 50
5.2.1 Charging the battery . . . . . . . . . . . . . . . . . 22 11.1.2 Group VERIFICATION . . . . . . . . . . . . . . . 53

6 Maintenance. . . . . . . . . . . . . . . . . . . . 23 12 Description of device functions -


Fieldcheck and vortex measuring
7 Accessories. . . . . . . . . . . . . . . . . . . . . 24 systems . . . . . . . . . . . . . . . . . . . . . . . . 59
7.1 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . 24 12.1 Block FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . 59
12.1.1 Group SIMULATION . . . . . . . . . . . . . . . . . 60
8 Technical data . . . . . . . . . . . . . . . . . . 25 12.1.2 Group VERIFICATION . . . . . . . . . . . . . . . 63
8.1 Technical data at a glance . . . . . . . . . . . . . . . . . . . 25
8.1.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.1.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Endress+Hauser 3
Safety instructions Proline Fieldcheck

1 Safety instructions

1.1 Designated use


The Fieldcheck flow simulation system may only be used for the simulation of sensor signals with a
flowmeter of the Proline equipment generation from Endress+Hauser. The device is normally
installed by service technicians who are familiar with the function and operating method of the
equipment. The manufacturer accepts no liability for damages resulting from incorrect use.

# Warning!
Fieldcheck may not be used in hazardous areas without corresponding safety measures.

1.2 Operational safety


• The manufacturer reserves the right to modify technical data without prior notice. Please contact
your E+H distributor for information on modifications or updates of these Operating Instructions.
• The measuring system complies with the general safety requirements in accordance with EN
61010, the EMC requirements of EN 61326/A1 and NAMUR Recommendation NE 21.

1.3 Notes on operating Fieldcheck in an environment with


strong electromagnetic interference

1.3.1 Radiated RF noise (EN 61326-1)


To operate Fieldcheck with a Proline flowmeter in the simulation and verification mode, the
electronics and connection compartments must first be opened. In this mode, the radiated RF noise
will not exceed the industrial limits (EN 61326-1).
In a heavily disturbed (EMC) environment, you must expect that the operation of both Fieldcheck
and the flowmeter may be temporarily affected. We recommend that - when using Fieldcheck - you
take care to ensure effective ESD protection for the affected measuring point and the operator.
If necessary, the measuring equipment can be reset:
• Via the local controls
• By interrupting the voltage supply
• By restarting Fieldcheck.

4 Endress+Hauser
Proline Fieldcheck Safety instructions

1.4 Notes on safety conventions and icons


The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the
factory in a condition in which they are safe to operate. The devices comply with the applicable
standards and regulations in accordance with EN 61010 “Protection Measures for Electrical
Equipment for Measurement, Control, Regulation and Laboratory Procedures”. They can, however,
be a source of danger if used incorrectly or for other than the designated use.
Consequently, always pay particular attention to the safety instructions indicated in these Operating
Instructions by the following icons:

# Warning!
“Warning” indicates an action or procedure which, if not performed correctly, can result in injury
or a safety hazard. Comply strictly with the instructions and proceed with care.

" Caution!
“Caution” indicates an action or procedure which, if not performed correctly, can result in incorrect
operation or destruction of the device. Comply strictly with the instructions.

! Note!
“Note” indicates an action or procedure which, if not performed correctly, can have an indirect
effect on operation or trigger an unexpected response on the part of the device.

Endress+Hauser 5
Identification Proline Fieldcheck

2 Identification

2.1 Device designation


The Fieldcheck flow simulation system consists of the following components:
• The Fieldcheck flow simulator itself
• The Simubox for correct connection to the respective measuring device

2.1.1 Fieldcheck nameplate

ENDRESS+HAUSER
FIELD CHECK
1 Order Code: 50098801
Ser.No.: 12345678901

2 0°C(+32°F)<Tamb<+60°C(+140°F)

F06-DXC10xxx-18-00-xx-xx-000

Fig. 1: Nameplate specifications for the Fieldcheck flow simulator (example)


1 Order code/serial number
2 Permitted ambient temperature range

2.1.2 Simubox nameplate

ENDRESS+HAUSER
SIMUBOX - MID
1 Order Code: 50097102
2 Ser.No.: 12345678901

4
F06-DXC10xxx-18-00-xx-xx-001

Fig. 2: Nameplate specifications for the Simubox (example)


1 Order code
2 Serial number

6 Endress+Hauser
Proline Fieldcheck Identification

2.2 Scope of delivery


The equipment supplied with Fieldcheck comprises the following components:
• The Fieldcheck flow simulator with battery pack
• A connecting cable for the service interface
• A connecting cable for the Simubox
• Two connecting cables for current and frequency output
• An "RS 232" connecting cable for connecting the Fieldcheck to a computer (9-pin COM interface)
• A line adapter
• A softcase
• Operating Instructions
The Simuboxes for the respective flow measuring systems are available via separate order numbers.

2.3 CE mark, Declaration of conformity


The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the
factory in a condition in which they are safe to operate. The devices comply with the applicable
standards and regulations in accordance with EN 61010 “Protection Measures for Electrical
Equipment for Measurement, Control, Regulation and Laboratory Procedures”.
The measuring system described in these Operating Instructions thus complies with the statutory
requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by
affixing the CE mark.

2.4 Registered trademarks


FieldTool™, Fieldcheck™
Registration-pending or registered trademarks of Endress+Hauser Flowtec AG, Reinach,
Switzerland

Endress+Hauser 7
Wiring Proline Fieldcheck

3 Wiring
# Warning!
When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to
these Operating Instructions of the relevant measuring device. Fieldcheck may not be used in
hazardous areas. Please do not hesitate to contact your E+H representative if you have any
questions.

3.1 Overview
Fieldcheck is connected to the following transmitter connections:
• Via the Simubox to the transmitter amplifier
• To the transmitter service plug
• To the current or frequency output.
For correct synchronisation with the measuring device we recommend that you abide by the
following procedure.
1. Start the device to be tested.
2. First connect the cables to the flowmeter (→ Page 9 to Page 14).
3. Next, connect the cables to the front of Fieldcheck (→ Page 16).
! Note!
Fieldcheck does not need an external power supply during operation. Power is provided by internal
storage batteries. Make sure that the storage batteries are fully charged before beginning a
simulation.

8 Endress+Hauser
Proline Fieldcheck Wiring

3.2 Connecting to Promag 10


" Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the cover of the connection compartment.
1. Remove cover of the electronics compartment (c).
2. Remove the local display (d) from the connection compartment cover.
3. Press the side latches (e) and fold down the connection compartment cover.
4. Pull the sensor signal cable plug out of the amplifier socket (f).
5. Insert Simubox cable plug into the amplifier socket (f).
6. Connect Fieldcheck to current and pulse output (g) of the transmitter.
! Note!
The configuration of the outputs is dependent on the product type and the product version of
the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
“Wiring” chapter of the flowmeter Operating Instructions.
7. Connect FieldCheck to the service plug (h).

2 1
h

g Esc

F06-DXC10xxx-04-00-xx-xx-007

Fig. 3: Connecting Fieldcheck to Promag 5X (compact version)


a Fieldcheck flow simulator
b Simubox
c Electronics compartment cover
d Local display
e Latches
f Amplifier socket
g Terminals for current/pulse output
h Service plug-in jack

Endress+Hauser 9
Wiring Proline Fieldcheck

3.3 Connecting to Promag 50/53 compact version


" Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the cover of the connection compartment.
1. Remove cover of the electronics compartment (e).
2. Pull local display (f) down.
3. Pull the sensor signal cable plug out of the amplifier socket (g).
4. Insert Simubox cable plug into the amplifier socket (g).
5. Remove the cover of the connection compartment (h) from the transmitter housing.
6. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs
22/23, 24/25 or 26/27).
! Note!
The configuration of the outputs is dependent on the product type and the product version of
the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
“Wiring” chapter of the flowmeter Operating Instructions.
7. Connect Fieldcheck to the service plug (i).

d c

h h

e e

27
b i 26
25
24
23
22
21
20

N (L-) 2
L1 (L+) 1

- + E

f a

F06-DXC10xxx-04-00-xx-xx-000

Fig. 4: Connecting Fieldcheck to Promag 5X (compact version)


a Fieldcheck flow simulator
b Simubox
c Stainless steel transmitter
d Aluminium transmitter
e Electronics compartment cover
f Local display
g Amplifier socket
h Connection compartment cover
I Service plug

10 Endress+Hauser
Proline Fieldcheck Wiring

3.4 Connecting to Promag 50/53 remote version


" Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the cover of the connection compartment.
1. Remove the cover of the connection compartment (d).
2. Completely remove the terminal block (e, terminal nos. 4-8, 36+37) from the base.
" Caution!
If the terminal block is not removed, a correct simulation/verification is not guaranteed.
3. Insert Simubox cable plug into the amplifier socket (g).
4. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs
22/23, 24/25 or 26/27).
! Note!
The configuration of the outputs is dependent on the product type and the product version of
the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
“Wiring” chapter of the flowmeter Operating Instructions.
5. Connect Fieldcheck to the service plug (f).

f e g

1 2 20 21 22 23 24 25 26 27

6 5 7 8 4 37 36

42 41

b
d

F06-DXC10xxx-04-00-xx-xx-001

Fig. 5: Connecting Fieldcheck to Promag 5X (remote version)


a Fieldcheck flow simulator
b Simubox
c Transmitter wall-mount housing
d Connection compartment cover
e Terminal block
f Service plug
g Amplifier socket

Endress+Hauser 11
Wiring Proline Fieldcheck

3.5 Connecting to Promass 40/80/83 compact version


" Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the cover of the connection compartment.
1. Remove cover of the electronics compartment (e).
2. Remove local display.
3. Disconnect the excitation current cable from the excitation current socket (f).
# Warning!
The excitation current cable must be disconnected as otherwise the measuring tube could
vibrate in the resonance frequency and become damaged!
4. Pull the sensor signal cable plug out of the amplifier socket (g).
5. Insert Simubox cable plug into the amplifier socket (g).
6. Remove the cover of the connection compartment (h) from the transmitter housing.
7. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs
22/23, 24/25 or 26/27).
! Note!
The configuration of the outputs is dependent on the product type and the product version of
the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
“Wiring” chapter of the flowmeter Operating Instructions.
8. Connect Fieldcheck to the service plug (i).

d c

h
h

e e

27
b i 26
25
24
23
22
21
20

N (L-) 2
L1 (L+) 1

f a

F06-DXC10xxx-04-00-xx-xx-004

Fig. 6: Connecting Fieldcheck to Promass 8x (compact version)


a Fieldcheck flow simulator
b Simubox
c Stainless steel transmitter
d Aluminium transmitter
e Electronics compartment cover
f Exciting current socket
g Amplifier socket
h Connection compartment cover
i Service plug

12 Endress+Hauser
Proline Fieldcheck Wiring

3.6 Connecting to Promass 80/83 remote version


" Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the cover of the connection compartment.
1. Remove the cover of the connection compartment (d).
2. Remove excitation current cable (h, terminal pair 41/42).
# Warning!
The excitation current cable must be removed as otherwise the measuring tube could vibrate
in the resonance frequency and become damaged!
3. Completely remove the terminal block (e, terminals 4-12) from the base.
" Caution!
If the terminal block is not removed, a correct simulation/verification is not guaranteed.
4. Insert Simubox cable plug into the amplifier socket (g).
5. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs
22/23, 24/25 or 26/27).
! Note!
The configuration of the outputs is dependent on the product type and the product version of
the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
“Wiring” chapter of the flowmeter Operating Instructions.
6. Connect Fieldcheck to the service plug (f).

f g e h
c
1 2 20 21 22 23 24 25 26 27

4 5 6 7 8 9 10 11 12

41 42

F06-DXC10xxx-04-00-xx-xx-005

Fig. 7: Connecting Fieldcheck to Promass 8X (remote version)


a Fieldcheck flow simulator
b Simubox
c Transmitter wall-mount housing
d Connection compartment cover
e Terminals for sensor signal cable
f Service plug
g Amplifier socket
h Terminals for excitation current cable

Endress+Hauser 13
Wiring Proline Fieldcheck

3.7 Connecting to Prowirl 72


" Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the cover of the connection compartment.
1. Remove cover of the electronics compartment (c).
2. Remove the local display module (d) from the retainer rails (e).
3. Fit the local display module (d) with its left side onto the right retainer rail (e) (this secures the
local display module).
(Continued on next page)

g
f

d e

h i

j
g

i
k

F06-DXC10xxx-04-00-xx-xx-006

Fig. 8: Connecting Fieldcheck to Prowirl 72


a Fieldcheck
b Simubox
c Electronics compartment cover
d Local display module
e Local display module retainer rails
f Fixing screw for cover of connection compartment
g Connection compartment cover
h Terminals for current output/power supply and pulse output (optional)
i Plastic cover
j Amplifier socket
k Service plug

14 Endress+Hauser
Proline Fieldcheck Wiring

4. Loosen the fixing screw (f) of the cover of the connection compartment (g) and fold down the
cover.
5. Connect Fieldcheck to the current output/power supply and to the pulse output if available.
The terminals can be disconnected for this.
" Caution!
– The current output can only be tested for accuracy during the verification if it is connected
to the current input of Fieldcheck.
– The pulse output is connected to the frequency input of Fieldcheck.
6. Fold up the plastic cover (i).
7. Disconnect the signal cable plug from the amplifier socket (j).
8. Insert Simubox cable plug into the amplifier socket (j).
9. Connect Fieldcheck to the service plug (k).

3.7.1 Prowirl 72 wiring diagram

+ - + -
1 2 3 4

F06-72xxxxxx-04-00-00-xx-000

Fig. 9: Assignment of terminals


A= Power supply/current output
B= Pulse/status output
C= Ground terminal (only relevant for remote version)

Endress+Hauser 15
Wiring Proline Fieldcheck

3.8 Assignment of Fieldcheck connections

1 2 3 4 5

F I SRV SIM DC

F06-DXC10xxx-04-00-xx-xx-003

Fig. 10: Fieldcheck connection diagram on the front panel


1 Connection to the frequency output of the device
2 Connection to the current output of the device
3 Connection to the internal service interface of the device
4 Connection to the Simubox
5 Connection to the plug power unit

3.9 Post-connection check


• Fieldcheck can only complete the start-up procedure correctly if the simulation box and the
service cable are correctly connected.
• Current and frequency outputs which are incorrectly connected or mixed up result in an incorrect
verification result.

16 Endress+Hauser
Proline Fieldcheck Operation

4 Operation

4.1 Display and operating elements


The local display enables you to read all important variables of the simulation and configure the
device using the function matrix.

1
+7.000 dm3/m
V
+7.012 dm3/m
MP1 5%

Esc
- + E
2

3 4
F06-DXC10xxx-07-xx-xx-xx-000

Fig. 11: Display and operating elements


1 Liquid crystal display
The backlit, liquid crystal display shows measured values, dialog texts, fault messages and information messages.
The HOME position (operating mode) is the name given to the display during normal measuring operation.
2 Optical operating elements for "Touch Control"
3 Plus/minus keys
– Enter numerical values, select parameters
– Select different blocks, groups and function groups within the function matrix
The following functions are activated by pressing the +/- keys (X) simultaneously:
– Exit the function matrix step by step → HOME position
– Press the +/- keys longer than 3 seconds → Direct return to the HOME position
– Cancel data entry
4 Enter key
– HOME position → Enter the function matrix
– Save numerical values entered or altered settings

Endress+Hauser 17
Operation Proline Fieldcheck

4.1.1 Display
The display area consists of three to four lines; this is where measured values are displayed, and/or
status variables (sequence information of a verification.

4 5 6

+7.000 dm3/m 1

V
+7.012 dm3/m 2

I = 17,12 mA F = 783 Hz 3

F06-DXC10xxx-07-xx-xx-xx-001

Fig. 12: Typical display for normal operating mode (HOME position)
1 Main line: shows simulation values, e.g. volume flow in [dm3/s]
2 Additional line: shows up-to-date measured values of the tested measuring device]
3 Information line: shows the current variables for current and frequency if the corresponding outputs of the device
are connected to Fieldcheck. This is only possible in the SIMULATION operating mode
4 “Information icons” field: icons representing additional information on the measured values are shown in this field.
See figure for a full list of the icons and their meanings.
5 “Measured values” field: the current measured values appear in this field.
6 “Unit of measurement” field: the units of measure and time defined for the flow values appear in this field.

18 Endress+Hauser
Proline Fieldcheck Operation

4.1.2 Display icons


The icons which appear in the field on the left make it easier to read and recognise error messages
locally.

1 2 3 4

5 6 7 8

9 10

V m

F06-DXC10xxx-07-xx-xx-xx-002

Fig. 13: Fieldcheck display icons


1 DC power supply
2 Charging status: battery fully charged
3 Charging status: battery 2/3 charged
4 Charging status: battery 1/3 charged (flashing in HOME position)
5 Simulation “continuous"
6 Simulation “step"
7 Simulation “profile"
8 Verification
9 Measured volume flow
10 Measured mass flow

Endress+Hauser 19
Operation Proline Fieldcheck

4.2 Brief operating instructions on the function matrix


! Note!
• See the general notes on Page 20.
• Description of functions → Section 9 to Section 12
1. HOME position → F → Enter the function matrix
2. Select a block (e.g. USER INTERFACE)
3. Select a group (e.g. CONTROL)
4. Select a function group (e.g. BASIC CONFIGURATION)
5. Select a function (e.g. LANGUAGE)
Change parameter / enter numerical values:
P → Select or enter release code, parameters, numerical values
F → Save your entries
6. Exit the function matrix:
– Press Esc key (X) longer than 3 seconds → HOME position
– Repeatedly press Esc key (X) → Return step by step to HOME position

F06-x3xxxxxx-19-xx-xx-xx-000

Fig. 14: Selecting functions and configuring parameters (function matrix)

4.2.1 General notes


The function matrix comprises a multiplicity of functions which, for the sake of clarity, are arranged
on a number of menu levels (blocks, groups and function groups).

20 Endress+Hauser
Proline Fieldcheck Operation

Comply with the following instructions when configuring functions:


You select functions as described on Page 20.

! Note!
A detailed description of all the functions as well as a detailed overview of the function matrix are
provided in Sections 9 to 12.

4.3 Communication
Operation via a personal computer using the “FieldTool” operating software.

4.3.1 FieldTool™ operating program


FieldTool™ is a universal service and configuration software package designed for the Proline
measuring devices. The connection is made with a commercial serial cable (9-pole DSUB, 0-
Modem) to the underside of Fieldcheck.
FieldTool™ offers the user the following fields of application:
• Configuring the equipment functions.
• Archiving the verification data.
• Printout of a verification certificate with all measuring results and I/O parameters.
You can find more information on FieldTool™ in the following E+H document:
System Information: SI 031D/06/en “FieldTool™"

Endress+Hauser 21
Commissioning Proline Fieldcheck

5 Commissioning

5.1 Function check


Make sure that Fieldcheck is correctly connected to the measuring device that is to be simulated
(→ Page 8 ff.). The measuring device must be working free of errors in order for a simulation or
verification to be started.

5.2 Commissioning
Fieldcheck is powered via an adapter or via the storage battery included in the delivery. Before the
simulation, make sure that the storage batteries are fully charged. To ensure correct commissioning
you are advised to abide by the processes as they are described on Page 8 ff.
Fieldcheck switches itself on automatically as soon as one of the connecting cables is inserted.
Fieldcheck performs a number of self-tests after power-on. As this procedure progresses, diverse
messages which have to be acknowledged appear on the local display.
You know that Fieldcheck is ready for operation when the HOME position appears on the display.
If batteries are being used, the charge status of the battery pack also appears (Page 19 ff.).

5.2.1 Charging the battery


The battery is charged automatically as soon as the power unit or the car-adapter cable is connected
to Fieldcheck. The charging time is less than 3 hours if there is no simulation in progress. When the
battery is fully charged it is switched to conservation charging. The charging time is considerably
longer during a simulation.
The charge status is displayed in the HOME position via the charge status icon (Page 19 ff.).

22 Endress+Hauser
Proline Fieldcheck Maintenance

6 Maintenance
Fieldcheck requires no special maintenance.

" Caution!
To prevent a low discharge of the battery in case of not using Fieldcheck for a longer period it is
recommended to charge the device every quarter of a year.
Exterior cleaning
When cleaning the exterior of measuring devices, always use cleaning agents that do not attack the
surface of the housing.

Endress+Hauser 23
Accessories Proline Fieldcheck

7 Accessories
Various accessories can be supplied for Fieldcheck and they can be ordered separately from
Endress+Hauser. The E+H service organisation can provide detailed information on the order code
of your choice.

Accessory Description Order code

FieldTool™ Configuration and service software for flowmeters in the DXS10 - * * * * *


field:
– Commissioning, maintenance analysis
– Configuring flowmeters
– Service functions
– Visualisation of process data
– Troubleshooting
– Controlling the “Fieldcheck™” tester/simulator
Contact your E+H representative for more information.

Simubox MID for FieldCheck Serves communication between FieldCheck and 50097102
electromagnetic flowmeters.

Simubox Coriolis for FieldCheck Serves communication between FieldCheck and Coriolis 50097101
mass flowmeters.
Plug-in power unit 85260 VAC Power unit for charging the battery and the operation of 50095659
FieldCheck at the mains

Car adapter 12 VDC Adapter for connecting to the car battery. 50095660

NiMH AccuPack Additional battery 50095661

Softcase Practical storage bag 50095662

Simubox/FieldCheck connecting cable For connecting the Simubox to FieldCheck 50095663

Proline measuring device/FieldCheck For connecting FieldCheck to the internal service 50097100
interface cable interface of the flowmeter

FieldCheckcurrent or frequency For connecting FieldCheck to the current or frequency 50095664


output measuring cable output of the flowmeter

FC/Tooling interface cable For connecting FieldCheck to the Tooling computer 50098800

7.1 Software history


Software version / date Changes to software Changes to documentation

V 1.00.00 - -

V 1.00.01/11.2001 Device release -

V 1.00.02/08.2002 Software adaptation -

V 1.01.00/07.2003 Software extension for Coriolis flow Operating Instructions


measuring systems

V 1.02.00/04.2004 Software extension for Prowirl 72 Operating Instructions

24 Endress+Hauser
Proline Fieldcheck Technical data

8 Technical data

8.1 Technical data at a glance

8.1.1 Input

Measured variable Current input, frequency input

Measuring range 0...25 mA, 0...15000 Hz

8.1.2 Output

Output signal Electromagnetic: simulation voltage


Coriolis: 2 phase-shifted sinusoidal signals

Galvanic isolation Yes

8.1.3 Power supply

Electrical connections Input for power unit: 100-240 V AC (47-63 Hz) Prim., 12 V DC (2.5 A) Sec.

Supply voltage 12 V DC, internal battery 7.2 V, 3800 mAh

Power consumption 140 mA


Charging current: max. 2A (max. 350 mA for simultaneous operation)

8.1.4 Performance characteristics

Reference operating +22 °C ± 2K


conditions

Max. measured error for Electromagnetic measuring device


outputs Voltage ±0.2 % of full scale (0.150 ... 20 mV)
Coriolis measuring device:
Phase difference ±0.2 % of full scale

Max. measured error for Current


inputs ± 5 µA (0 ... 25 mA)
Frequency
< 0.5 Hz (0 ...15 kHz)

8.1.5 Operating conditions (Environment)

Ambient temperature range 0 ... 60°C

Electromagnetic compatibility In accordance with EN 61326/A1 and NAMUR Recommendation NE 21


(EMC)

Endress+Hauser 25
Technical data Proline Fieldcheck

8.1.6 Mechanical construction

Design, dimensions Fieldcheck


270 x 130 x 60 mm (length x width x height)
Simubox (without cable)
120 x 60 x 30 mm (L x W x H)
Softcase
400 x 180 x 300 mm (L x W x H)

Weight 1.6 kg

8.1.7 Display and user interface

Display elements Liquid crystal display: illuminated,


for presenting different measured value and status variables

Operating elements Local operation with three optical keys (O, S, F)

Computer interface FieldTool™


• Permanent storage of verification data
• Printing of certificates

CE mark The measuring system is in conformity with the statutory requirements of the EC Directives.
Endress+Hauser confirms successful testing of the device by affixing the CE mark.

Other standards and EN 60529:


guidelines Degrees of protection by housing (IP code)
EN 61010:
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory
Procedures
EN 61326/A1 (IEC 1326):
Electromagnetic compatibility (EMC requirements)

8.1.8 Ordering information

The E+H service organisation can provide detailed ordering information and information on the
order codes on request.

8.1.9 Accessories
Various accessories can be supplied for Fieldcheck and they can be ordered separately from
Endress+Hauser (see → Page 24). The E+H service organisation can provide detailed information
on the order codes of your choice.

8.1.10 Documentation
• System Information SI 030D/06/en
• Technical Information TI 059D/06/en

26 Endress+Hauser
Proline Fieldcheck Description of device functions - General section

9 Description of device functions - General


section

9.1 Notes on the description of device functions


The description of device functions is split into the following sections:
• Description of device functions - General section (→ Page 27 ff.)
This introduces you to the function matrix and provides you with an overview of the device
functions and a description of the device functions which apply to all applications.
• Description of device functions - Fieldcheck and electromagnetic measuring devices
(→ Page 39 ff.)
This provides you with a description of the device functions which only apply to the simulation
and verification of electromagnetic flow measurement systems.
• Description of device functions - Fieldcheck and Coriolis measuring systems
(→ Page 49 ff.)
This provides you with a description of the device functions which only apply to the simulation
and verification of Coriolis mass flow measuring systems.
• Description of device functions - Fieldcheck and vortex measuring systems
(→ Page 59 ff.)
This provides you with a description of the device functions which only apply to the simulation
and verification of vortex measuring systems.
There are various ways of locating the description of a function of your choice in the manual:

9.1.1 Using the table of contents to locate a function description


The designations of all the cells in the function matrix are listed in the table of contents. You can
use these unambiguous designations (such as USER INTERFACE, INPUTS, OUTPUTS, etc.) to
choose whichever functions are applicable to a particular set of conditions. The page references
show you exactly where to find the detailed descriptions of the functions in question. The table of
contents can be found on Page 3.

9.1.2 Using the graphic of the function matrix to locate a function


description
This step-by-step, top-down approach starts with the blocks, the highest level, and works down
through the matrix to the description of the function you need:
1. All blocks available, and their related groups are illustrated on Page 28. Select the block (or the
group within the block) which you need for your application and use the page reference to
locate the information corresponding to the next level.
2. The page in question contains a graphic illustration of the block with all its subordinate groups,
function groups and functions.Select the function which you need for your application and use
the page reference to locate the detailed function description.

Endress+Hauser 27
Description of device functions - General section Proline Fieldcheck

9.2 Function matrix

BLOCKS GROUPS FUNCTIONS

FUNCTION → SIMULATION
Proline Promag: Page 39 ff.
Proline Promass: Page 49 ff. Electromagnetic measuring devices (Proline → Page 40 ff.
Promag)
Proline Vortex: Page 59 ff.
Coriolis measuring devices (Proline Promass) → Page 50 ff.

Vortex measuring devices (Proline Prowirl) → Page 60 ff.

↓ VERIFICATION

Electromagnetic measuring devices (Proline → Page 44 ff.


Promag)

Coriolis measuring devices (Proline Promass) → Page 53 ff.

Vortex measuring devices (Proline Prowirl) → Page 63 ff.

USER INTERFACE → CONTROL → Page 29 ff.

BASIC FUNCTIONS → Sensor data → Page 32 ff.

OUTPUTS → Page 33 ff.


↓ PROCESS PARAMETER → Page 34

READ IN PARAMETERS → Page 35

SUPERVISION → RESULTS → Page 36 ff.

VERSION INFO → Page 38 ff.

28 Endress+Hauser
Proline Fieldcheck Description of device functions - General section

9.3 Block USER INTERFACE

Block Group Function group Function

CONTROL LANGUAGE CONTRAST FORMAT DATE


USER BACK LIGHT
⇒ ⇒ LCD DATE
INTERFACE P. 29
P. 29 P. 29 P. 29 P. 29 P. 29

FORMAT TIME TIME

P. 29 P. 30

9.3.1 Group CONTROL

Function description
USER INTERFACE → CONTROL

LANGUAGE Use this function to select the language for all texts, parameters and messages shown on
the local display.

Options:
ENGLISH
DEUTSCH

CONTRAST LCD Use this function to optimise display contrast to suit


local operating conditions.

User input:
10...100%
Factory setting:
50%

BACK LIGHT Options:


ON(AUTOMATIC)
OFF

Factory setting:
ON(AUTOMATIC)

! Note!
If the ON(AUTOMATIC) setting is selected, the lighting of the display is automatically
switched off if the equipment is not used for 10 minutes.

FORMAT DATE Use this function to select a date format.

Options:
DD.MM.YYYY
MM/DD/YYYY

Factory setting:
DD.MM.YYYY

DATE Use this function to enter the date.

User input:
DD.MM.YYYY (MM/DD/YYYY)

FORMAT TIME Options:


24H
12H (AM/PM)

Factory setting:
24H

Endress+Hauser 29
Description of device functions - General section Proline Fieldcheck

Function description
USER INTERFACE → CONTROL

TIME Use this function to enter the time.

User input:
HH:MM

30 Endress+Hauser
Proline Fieldcheck Description of device functions - General section

9.4 Block BASIC FUNCTIONS

Block Group Function group Function

NOMINAL
SENSOR K-FACTOR ZERO POINT FLOW SERIAL
BASIC DIAMETER
⇒ PARAMETER ⇒ POS/NEG DIRECTION NUMBER
FUNCTIONS
P. 32 P. 32 P. 32 P. 32 P. 32
P. 32

TAG NAME.
⇓⇑ P. 32

OUTPUTS OUTPUT CURRENT VALUE 20 mA


ASSIGN VALUE 0_4 mA
⇒ RANGE
P. 33 P. 33
P. 33 P. 33 P. 33 P. 33

OPERATION START VALUE END VALUE FLOW


ASSIGN
MODE FREQUENCY FREQUENCY
P. 33
P. 33 P. 33 P. 33 P. 33

⇓⇑ VALUE FHIGH OUTPUT


PULSE VALUE
SIGNAL
P. 34
P. 34 P. 34

DENSITY
PROCESS STANDARD OPERATING REFERENCE
VALUE APPLICATION
PARAMETER ⇒ DENSITY DENSITY DENSITY
P. 34
P. 34 P. 34 P. 34 P. 34
P. 34
TYPE UNIT
MEAS.
⇓⇑ P. 34

READ IN READ IN
PARAMETERS ⇒ PARAMETERS
P. 35 P. 35

Endress+Hauser 31
Description of device functions - General section Proline Fieldcheck

9.4.1 Group SENSOR PARAMETER


Use this function group to display the characteristic data of the connected sensor, such as calibration
factor, nominal diameter, serial number, etc.

Function description
BASIC FUNCTIONS → SENSOR PARAMETER

K-FACTOR Use this function to view the positive/negative K factor


POSITIVE/NEGATIVE

ZERO POINT Use this function to view the zero-point correction value for the sensor.

! Note!
This function is not displayed when testing vortex measuring devices.

NOMINAL DIAMETER Use this function to view the nominal diameter of the sensor.

DIRECTION OF FLOW Use this function to view the set mounting direction of the sensor.

! Note!
This function is not displayed when testing vortex measuring devices.

SERIAL NUMBER Use this function to view the serial number of the connected measuring device.

TAG NAME Use this function to view the device tag name of the connected measuring device.

32 Endress+Hauser
Proline Fieldcheck Description of device functions - General section

9.4.2 Group OUTPUTS


Use this function group to view the configuration and settings of the current and frequency outputs.

Function description
BASIC FUNCTION → OUTPUTS

OUTPUT Use this function to select the output

Options:
CURR. OUTPUT
FREQUENCY OUTPUT

ASSIGN Use this function to view the physical measured variable which is assigned to the current
output of the measuring device (e.g. volume flow, mass flow, temperature)

! Note!
This function is only available if CURRENT OUTPUT was selected in the OUTPUT
function.

CURRENT RANGE Use this function to view the set current range in the measuring device.

! Note!
This function is only available if CURRENT OUTPUT was selected in the OUTPUT
function.

VALUE 0_4 mA Use this function to view the flow value, which is assigned to the 0_4 mA current in the
measuring device.

! Note!
This function is only available if CURRENT OUTPUT was selected in the OUTPUT
function.

VALUE 20 mA Use this function to view the flow value, which is assigned to the 20 mA current in the
measuring device.

! Note!
This function is only available if CURRENT OUTPUT was selected in the OUTPUT
function.

OPERATION MODE Use this function to view the set operating mode of the frequency output in the
measuring device.

! Note!
This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT
function.

ASSIGN Use this function to view the physical measured variable which is assigned to the
frequency output of the measuring device (e.g. volume flow, mass flow, temperature)

! Note!
This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT
function.

START FREQUENCY Use this function to view the set start frequency in the measuring device.

! Note!
This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT
function.

END FREQUENCY Use this function to view the set end frequency in the measuring device.

! Note!
This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT
function.

VALUE f-LOW Use this function to view the flow value, which is assigned to the start frequency in the
measuring device.

! Note!
This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT
function.

Endress+Hauser 33
Description of device functions - General section Proline Fieldcheck

Function description
BASIC FUNCTION → OUTPUTS

VALUE f-HIGH Use this function to view the flow value, which is assigned to the end frequency in the
measuring device.

! Note!
This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT
(6900) function.

PULSE VALUE Use this function to view the set pulse value in the measuring device.

! Note!
This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT
(6900) function and the operating mode in the measuring device is set to PULSE.

OUTPUT SIGNAL Use this function to view the output signal settings (ACTIVE/PASSIVE,
POSITIVE/NEGATIVE).

9.4.3 Group PROCESS PARAMETER

Function description
BASIC FUNCTION → PROCESS PARAMETER

DENSITY VALUE ! Note!


This function only appears if a Promag 53-type electromagnetic flow measurement system
is being tested.
Use this function to view the density factor entered in the measuring device. With this
value, the volume flow is converted to a mass flow during simulation.

STANDARD DENSITY ! Note!


This function only appears if a Coriolis flow measurement system is tested.
Use this function to view the standard density entered in the measuring device. With this
value, the mass flow is converted to the corrected volume flow during simulation.

OPERATING DENSITY ! Note!


This function only appears if a vortex flow measurement system is tested.
Use this function to view the density at process conditions entered in the measuring
device. With this value, the volume flow is converted to a mass flow or corrected volume
flow during simulation.

REFERENCE DENSITY ! Note!


This function only appears if a vortex flow measurement system is tested.
Use this function to view the density at reference conditions entered in the measuring
device. With this value, the volume flow is converted to a mass flow or corrected volume
flow during simulation.

APPLICATION ! Note!
This function only appears if a vortex flow measurement system is tested.
Use this function to view the set application (liquid or gas/steam) in the measuring device.

TYPE UNIT
MEASURAND
! Note!
This function only appears if a vortex flow measurement system is tested.
Use this function to view the set unit type (volume flow, calculated mass flow or calculated
corrected volume flow) in the measuring device.

34 Endress+Hauser
Proline Fieldcheck Description of device functions - General section

9.4.4 Group READ IN PARAMETERS

Function description
BASIC FUNCTION → READ IN PARAMETERS

READ IN PARAMETERS Use this function to transfer the complete data record from the measuring device to
Fieldcheck, if, for example, you have changed parameters in the connected measuring
device,

Options:
START
CANCEL

Endress+Hauser 35
Description of device functions - General section Proline Fieldcheck

9.5 Block SUPERVISION

Block Group Function group Function

SUPERVISION RESULTS ⇒ STORAGE INFO RESULTS VER. RESULTS VER. RESULTS VER.
LOC. VER. STORAGE MID COR. VOR.

P. 36 P. 36 LOC. P. 36 P. 37 P. 37
P. 36

DELETE STORAGE
STORAGE LOCATION
⇓⇑ P. 37 P. 37

VERSION INFO FIELDCHECK HW VERSION SW VERSION SERIAL CALIBRATION


P. 38 ⇒ ⇒ NUMBER DATE
P. 38 P. 38 P. 38 P. 38 P. 38

SIMUBOX TYPE HW VERSION SW VERSION SERIAL CALIBRATION


⇒ NUMBER DATE
P. 38 P. 38 P. 38 P. 38 P. 38 P. 38

MANUFACT./
TEST DATE
P. 38

9.5.1 Group RESULTS

Function description
SUPERVISION → RESULTS

STORAGE LOCATION Use this function to define the storage location of the verification to be displayed.
VERIFICATION
User input:
1...20

INFO STORAGE Use this function to view information about the measuring device and the verification
LOCATION saved in the storage location.

Display:
Main line: measuring device type
Additional line: serial number
Information line: date of verification

RESULTS Use this function to recall the results of the verification. The result (PASSED, FAILED) is
VERIFICATION displayed in the information line.
MAGMETER If the results of a linearity check are recalled, the list of deviation values also appears when
confirming the selection.

Options:
FINAL RESULT
ZERO POINT (with list of deviation values)
AMPLIFIER (with list of deviation values)
CURRENT OUTPUT (with list of deviation values)
FREQ. OUT. (with list of deviation values)
PULSE MEASUREMENT (with list of deviation values)
SENSOR

36 Endress+Hauser
Proline Fieldcheck Description of device functions - General section

Function description
SUPERVISION → RESULTS

RESULTS Use this function to recall the results of the verification. The result (PASSED, FAILED) is
VERIFICATION displayed in the information line.
CORIOLIS If the results of a linearity check are recalled, the list of deviation values also appears when
confirming the selection.

Options:
FINAL RESULT
ZERO POINT (with list of deviation values)
AMPLIFIER (with list of deviation values)
DENSITY (with list of deviation values)
TEMPERATURE (with list of deviation values)
CURRENT OUTPUT (with list of deviation values)
FREQUENCY OUTPUT (with list of deviation values)
PULSE MEASUREMENT (with list of deviation values)
SENSOR

RESULTSVERIFICATI Use this function to recall the results of the verification. The result (PASSED, FAILED) is
ON VORTEX displayed in the information line.
If the results of a linearity check are recalled, the list of deviation values also appears when
confirming the selection.

Options:
FINAL RESULT
ZERO POINT (with measured value for the current output)
AMPLIFIER (with list of deviation values)
CURRENT OUTPUT (with list of deviation values)
PULSE MEASUREMENT (with list of deviation values)
SENSOR

DELETE STORAGE Use this function to delete the content of the memory.

Options:
NO
STORAGE LOCATION
ALL

STORAGE LOCATION Use this function to specify the storage location whose contents should be deleted. User
input :1...20. This function does not appear unless the STORAGE LOCATION parameter
was selected in the DELETE STORAGE function.

Endress+Hauser 37
Description of device functions - General section Proline Fieldcheck

9.5.2 Group VERSION INFO

Function group FIELDCHECK

Function description
SUPERVISION → VERSION → INFO FIELDCHECK

HW VERSION Use this function to view the hardware version number of Fieldcheck.

SW VERSION Use this function to view the software version number of Fieldcheck.

SERIAL NUMBER Use this function to view the manufacturer serial number of Fieldcheck.

CALIBRATION DATE Use this function to view the last calibration date of Fieldcheck.

Function group SIMUBOX

Function description
SUPERVISION → VERSION-INFO → SIMUBOX

TYPE Use this function to view the measuring principle of the Simubox

HW VERSION Use this function to view the hardware version number of the Simubox.

SW VERSION Use this function to view the software version number of the Simubox.

SERIAL NUMBER Use this function to view the manufacturer serial number of the Simubox.

CALIBRATION DATE Use this function to view the last calibration date of the Simubox -(electromagnetic and
Coriolis Simuboxes).

MANUFACTURING/TE Use this function to view the date of manufacture of the Simubox (vortex and ultrasonic
ST DATE Simuboxes).
or
Use this function to view the last test date (vortex and ultrasonic Simuboxes ).

38 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and electromagnetic measuring devices

10 Description of device functions - Fieldcheck and


electromagnetic measuring devices

10.1 Block FUNCTION

Block Group Function group Function

SIMULATION CONFIGURATI OPERATION


SIMUL SIGNAL FLOW 1 FLOW 2 SIMUL. TIME 1
FUNCTION ⇒ ⇒ ON ⇒ MODE
P. 41 P. 41 P. 41 P. 41
P. 40 P. 40 P. 40

PROFILE TIME PROFILE TIME


SIMUL. TIME 2 1 2
⇓⇑ ⇓⇑
P. 42
P. 42 P. 42

OPERATION SIMULATE

P. 43 P. 43

VERIFICATION OUTPUTS OUTPUTS


⇒ ⇒
P. 44 P. 45 P. 45
⇓⇑

DEVIATION DEV.CURR.OU DEV.FREQU.O


LIMIT VALUES
⇒ FLOW TPUT UTPUT
P. 46
P. 46 P. 46 P. 46

⇓⇑

PARAMETER MEAS. POINT


FLOW 100%
⇒ 1/2
P. 47
P. 47 P. 47

⇓⇑

OPERATION STORAGE
VERIFICATION VERIFICATE
⇒ LOCATION
P. 48 P. 48
P. 48 P. 48

Endress+Hauser 39
Description of device functions - Fieldcheck and electromagnetic measuring devices Proline Fieldcheck

10.1.1 Group SIMULATION

Function group CONFIGURATION

Function description
FUNCTION → SIMULATION → CONFIGURATION

OPERATION MODE Use this function to select the operation mode for the simulation.
There are two freely adjustable flow signals of -12.5 m/s to + 12.5 m/s and the assigned
duration of simulation is 1 second to 999 seconds.

Options:

CONTINUOUS
The flow behaviour in the process has low dynamics.

Q
Q1

START STOP t
F06-DXC10xxx-05-xx-xx-xx-000

Q1 = Simulation value 1

STEP
The flow rates in the process are very dynamic and change very quickly. This operation
mode is also suitable for the simulation of flows outside the tolerance limits (Error
simulation). Here very dynamic or disrupted conditions can also be simulated.

Q T1
Q1

Q2
T2
t
START STOP
F06-DXC10xxx-05-xx-xx-xx-001

Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1,


T2 = Duration of simulation 2

PROFILE
For the depiction of real sequences in the course of a process or of a filling procedure.
The controlled rise or fall in one flow value to the next can be depicted.

Q T1 T3 T2 T4
Q2

Q1

t
START STOP
F06-DXC10xxx-05-xx-xx-xx-002.eps

Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1,


T2 = Duration of simulation 2, T3 = Profile time 1, T4 = Profile time 2

Factory setting:
CONTINUOUS

40 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and electromagnetic measuring devices

Function description
FUNCTION → SIMULATION → CONFIGURATION

SIMULATION SIGNAL Use this function to select which measurement is to be simulated.

Options:
VOLUME FLOW
MASS FLOW

! Note!
• "Calculated mass flow" simulation is only possible with "Promag 53" type devices.
• The value stored in the function DENSITY VALUE (DENSITY PARAMETER function
group of the device) is used for the calculation.
• If a current input is available in the device and the density is assigned to it, mass flow
simulation is not possible
Factory setting:
VOLUME FLOW

FLOW 1 Use this function to enter the flow value which is to be simulated.

User input:
0.000...999999 (6-digit floating-point number, with sign)

! Note!
The unit of the connected measuring device is adopted. The corresponding flow velocity
must lie within the value range - 12.5 m/s to + 12.5 m/s.
Unit:
The unit configured in the measuring device is adopted

Factory setting:
0.000

FLOW 2 Use this function to enter the second flow value which is to be simulated.

! Note!
• This function is only available if the STEP or PROFILE setting was selected in the
OPERATION MODE function.
• More information is provided in the OPERATION MODE function.
User input:
0.000...999999 (6-digit floating-point number, with sign)

! Note!
The unit of the connected measuring device is adopted. The corresponding flow velocity
must lie within the value range - 12.5 m/s to + 12.5 m/s.
Unit:
The unit configured in the measuring device is adopted

Factory setting:
0.000

DURATION OF Use this function to enter the duration of simulation signal 1 (flow).
SIMULATION 1
! Note!
• This function is only available if the STEP or PROFILE setting was selected in the
OPERATION MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 999 (max. 3-digit number )

Unit:
s ( seconds )

Factory setting:
10

Endress+Hauser 41
Description of device functions - Fieldcheck and electromagnetic measuring devices Proline Fieldcheck

Function description
FUNCTION → SIMULATION → CONFIGURATION

DURATION OF Use this function to enter the duration of simulation signal 2 (flow).
SIMULATION 2
! Note!
• This function is only available if the STEP or PROFILE setting was selected in the
OPERATION MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 999 (max. 3-digit number )

Unit:
s ( seconds )

Factory setting:
10

PROFILE TIME 1 Use this function to enter the period of time for the change from simulation signal 1 to
simulation signal 2.

! Note!
• This function is only available if the PROFILE setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 99 (max. 2-digit number)

Unit:
s ( seconds )

Factory setting:
5

PROFILE TIME 2 Use this function to enter the period of time for the change from simulation signal 2 to
simulation signal 1.

! Note!
• This function is only available if the PROFILE setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 99 (max. 2-digit number)

Unit:
s ( seconds )

Factory setting:
5

42 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and electromagnetic measuring devices

Function group OPERATION

Function description
FUNCTION → SIMULATION → OPERATION

SIMULATE Use this function to control the simulation.

Options:
START
STOP

Factory setting:
START

Display during the simulation:


Main line
Symbol of the selected operating mode and the simulated value (with unit).
Additional line
The measured value sent back by the measuring device.
Information line
In the SIMULATION operating mode, the current values of the current and frequency
output of the tested device are displayed (multimeter function). Pulses are displayed as
frequency if there is more than 1 pulse/second at the frequency input of Fieldcheck.

In the STOP status the simulation signal value “0” is always given. Affected settings for
the simulation, however, are maintained.

Endress+Hauser 43
Description of device functions - Fieldcheck and electromagnetic measuring devices Proline Fieldcheck

10.1.2 Group VERIFICATION


In the VERIFICATION group, a measuring device can be checked for the following characteristic
values.
• Functionality and deviation in the flow measurement (based on the set point preset by Fieldcheck)
• Functionality and deviation of the current and frequency outputs if these outputs are available in
the measuring device. The specified set points correspond in the gradation to the measuring
points when determining flow.
Data bus outputs such as PROFIBUS or FOUNDATION FIELDBUS cannot be directly checked by
Fieldcheck.
• Static and dynamic characteristic values of the magnetic field and electrodes.

Verification of the transmitter


Checking the linearity of the amplifier:
Simulated flow at four measuring points which are determined via the 100% full scale value (can be
selected for a presumed fluid velocity between 4 m/s and 8 m/s).
Checking a current output:
Each current output can be verified in the configuration in the measuring device (CURRENT
RANGE function (0-20 mA, 4-20 mA) in the same gradation (5% ... 100%) as the amplifier. For this
purpose, the current output must be directly connected to the current input of the Fieldcheck. If
more than one current output is available, verification takes place in several, menu-guided steps.
Checking a frequency output
(FREQUENCY operating mode, (not for "Promag 10" type devices):
Each frequency output can be verified in the configuration in the measuring device (START VALUE
FREQUENCY, END VALUE FREQUENCY, OUTPUT SIGNAL functions) in the same gradation ( 5%
... 100% ) as the amplifier. For this purpose, the frequency output must be directly connected to the
frequency input of the Fieldcheck. If more than one frequency output is available, verification takes
place in several, menu-guided steps.
A frequency output cannot be checked on "Promag 10" type devices.
Checking a frequency output (PULSE operating mode):
Each pulse output available can be verified in the configuration in the measuring device (PULSE
VALUE, PULSE WIDTH, OUTPUT SIGNAL functions). For this purpose, the flow specified by
Fieldcheck must trigger at least one pulse within 30 s. Otherwise, 0 PULSE is specified. In this
instance, the output is not completely tested in its components.

Verification of the sensor


Coil current test:
For testing the sensor (for the electromagnetic measuring principle) the rise time of the coil current
and further magnetic field-relevant characteristic variables are determined and compared with the
corresponding set points. Exceeding the limit values leads to a FAILED message in the results
presentation.

" Achtung!
Electrode integrity test (not for "Promag 10" type devices):
When testing the sensor the electrode integrity test is carried out at the end of the verification. For
this the measuring tube must be filled and the sensor signal cable must be inserted into the
measuring amplifier socket. A corresponding message appears in the display and you are given the
option to CONTINUE or CANCEL.

44 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and electromagnetic measuring devices

Evaluation of the results


During verification, the current verification test section is displayed in the 'HOME position'.The
results can be shown on the display once verification is complete. They can be filed in Fieldcheck
under a storage location number of your choice (up to max. 20) together with the measuring device
serial no. and the measuring device tag name.
Via a connection with the FieldTool software, a protocol of the results can later be printed out and
the data from the verification can be stored in an archive file system.

Function group OUTPUTS


During verification, the function and accuracy of a current output and a frequency output can be
checked. Fieldcheck determines the current and frequency outputs available in the measuring
device.

Function description
FUNCTION → VERIFICATION → OUTPUTS

OUTPUTS Use this function to display the outputs determined in the flowmeter on the local display
with the related terminals and to select the outputs to be tested. The outputs selected
are displayed inversely.
1. Use the Pkeys to select the outputs which should be tested during the verification
process.
2. Confirm the option selected with the F key.

! Note!
• It is not possible to select two outputs of the same type.
• Start the verification process in the VERIFICATE function (→ Page 48).
Two outputs of the same type are tested in two separate runs. After the 1st run, the
message MORE OUTPUTS TO BE TESTED? appears on the display.
If you want to test other outputs, Fieldcheck returns to the output selection area.
1. Select the output to be tested.

! Note!
Having selected an output, make sure that the related terminal pair of the flowmeter
is connected to the -Fieldcheck input before the verification process is started.
The following combinations of current and frequency outputs are possible for PROline
Promag:
• I
• I+I
• I+F
• I+I+F
• I+F+F
• I+P (only Promag 10)

! Note!
• Fieldbus interfaces cannot be tested.
• With this software-version current outputs run in passive mode cannot be tested.

Endress+Hauser 45
Description of device functions - Fieldcheck and electromagnetic measuring devices Proline Fieldcheck

Function group LIMIT VALUES


You can specify the limit values for measured variables in this function group. If the deviation of a
verification step (see OPERATION, P. 38) lies within the configured limit values the result is
PASSED; if it lies outside of this range the result is FAILED. For a quick and unambiguous
identification, the FAILED results are presented inversely on the display.

! Note!
For a frequency output operated in the pulse mode the fixed error limit is: ± 1 pulse ± 0.1% of the
pulse set number.

Function description
FUNCTION → VERIFICATION → LIMIT VALUES

DEVIATION FLOW User input:


0.50 ... 10.0

Unit:
%

Factory setting:
0.5

DEVIATION CURRENT This function is only available if a current output was selected in the OUTPUTS function
OUTPUT (→ Page 45).

User input:
0.02...10.0

Unit:
mA

Factory setting:
0.05

DEVIATION This function is only available if a frequency output was selected in the OUTPUTS
FREQUENCY OUTPUT function (→ Page 45).

! Note!
This function is not available on "Promag 10" type devices.
User input:
1.0...10.0

Unit:
Hz

Factory setting:
2.00

46 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and electromagnetic measuring devices

Function group PARAMETER


Verification of the amplifier takes place along four measuring points MP1...MP4. Use this function
group to enter the full scale value (100%) of the verification flow and the two selectable measuring
points MP2 and MP3 (in % of the full scale value).

! Note!
The value of MP1 is predetermined (5% of the full scale value MP4).

Function description
FUNCTION → VERIFICATION → PARAMETER

FLOW 100% Use this function to enter the full scale value of the verification flow.
User input:
max 6-digit floating point number (without sign)

! Note!
Only values between 4 and 8 m/s are permissible. The input limits MIN and MAX are
displayed beyond the input line.
Unit:
The unit configured in the measuring device is adopted

Factory setting:
according 6m/s

MEASURING POINT Use this function to enter the measuring points MP2 and MP3 (in % of the full scale
2/3 value).

User input:
MP2: integer between 10 and 90
MP3: integer between 15 and 95

! Note!
• The value of MP2 must be less than the value of MP3.
• Both values are stored permanently.

Unit:
%

Factory setting:
MP2: 10 %
MP3: 50%

Endress+Hauser 47
Description of device functions - Fieldcheck and electromagnetic measuring devices Proline Fieldcheck

Function group OPERATION


Use this function group to verify the measuring point.

Function description
FUNCTION → VERIFICATION → OPERATION

VERIFICATION Use this function to define the scope of the verification.

! Note!
• A verification of the current and frequency outputs of the transmitter can only be
achieved with the TRANSMITTER or TRANSM. + SENSOR option.
• If SENSOR or TRANSM. + SENSOR is selected, the electrode integrity test is carried out
at the end of the verification (→ Page 44).
Options:
TRANSMITTER
SENSOR
TRANSM. + SENSOR

Factory setting:
TRANSMITTER

VERIFICATE Use this function to control the verification. After completion of the verification, the
question STORE DATA? appears in this function.

Options:
START
The verification is started and the message VERIFICATION ACTIVE appears on the
display.At the end of the verification, the message RESULT: PASSED/FAILED - STORE
DATA? appears on the display.
STOP
Via the STOP option the active verification can be aborted.
STORE DATA?
The results of the verification can be saved using the STORE DATA? option.

Factory setting:
START

STORAGE LOCATION Use this function to define where the results of the verification are to be stored.

User input:
1...20

Factory setting:
First available storage location

! Note!
If there is no storage location available or the selected storage location is occupied, a
corresponding warning is given before saving (“Overwrite storage location? Y/N”).

48 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and Coriolis measuring systems

11 Description of device functions - Fieldcheck and


Coriolis measuring systems

11.1 Block FUNCTION

Block Group Function group Function

SIMULATION CONFIGURATI OPERATION


SIMUL SIGNAL FLOW 1 FLOW 2 SIMUL. TIME 1
FUNCTION ⇒ ⇒ ON ⇒ MODE
P. 50 P. 51 P. 51 P. 51
P. 50 P. 50 P. 50

SIMUL. TIME 2
⇓⇑ ⇓⇑
P. 51

OPERATION SIMULATE

P. 52 P. 52

VERIFICATION APPLICAT./OU TYPE UNIT OUTPUTS


FLOW RANGE
⇒ TP. ⇒ FLOW
P. 54
P. 53 P. 54 P. 54 P. 55

⇓⇑

DEVIATION DEVIATION DEVIATION


DEVIATION DEVIATION
LIMIT VALUES DENSITY CURRENT FREQU.
⇒ FLOW TEMPERATURE
P. 56 OUTPUT OUTPUT
P. 56 P. 56
P. 56 P. 56 P. 56
⇓⇑

PARAMETER MEAS. POINT


FLOW 100%
⇒ 1/2
P. 57
P. 57 P. 57

⇓⇑

OPERATION STORAGE
VERIFICATION VERIFICATE
⇒ LOCATION
P. 58 P. 58
P. 58 P. 58

Endress+Hauser 49
Description of device functions - Fieldcheck and Coriolis measuring systems Proline Fieldcheck

11.1.1 Group SIMULATION

Function group CONFIGURATION

Function description
FUNCTION → SIMULATION → CONFIGURATION

OPERATION MODE Use this function to select the operation mode for the simulation.
There are two freely adjustable flow signals of -12.5 m/s to + 12.5 m/s and the assigned
duration of simulation is 1 second to 999 seconds.

Options:

CONTINUOUS
The flow behaviour in the process has low dynamics.

Q
Q1

START STOP t
F06-DXC10xxx-05-xx-xx-xx-000

Q1 = Simulation value 1

STEP
The flow rates in the process are very dynamic and change very quickly. This operation
mode is also suitable for the simulation of flows outside the tolerance limits (Error
simulation). Here very dynamic or disrupted conditions can also be simulated.

Q T1
Q1

Q2
T2
t
START STOP
F06-DXC10xxx-05-xx-xx-xx-001

Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1,


T2 = Duration of simulation 2

Factory setting:
CONTINUOUS

SIMULATION SIGNAL Use this function to select which measurement is to be simulated.

Options:
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW (not for Promass 40)

! Note!
• The option CORRECTED VOLUME FLOW is only available if the
FIXED REFERENCE DENSITY option was selected in the CORRECTED VOLUME
CALCULATION function in the tested device.
• For the simulation of corrected volume flow, the system uses the values specified by
the flowmeter in the REFERENCE DENSITY and UNIT REFERENCE DENSITY
functions.
Factory setting:
MASS FLOW

50 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and Coriolis measuring systems

Function description
FUNCTION → SIMULATION → CONFIGURATION

FLOW 1 Use this function to enter the flow value which is to be simulated.

User input:
0.000...999999 (6-digit floating-point number, with sign)

! Note!
The unit of the connected measuring device is adopted.
Unit:
The unit configured in the measuring device is adopted

Factory setting:
0.000

FLOW 2 Use this function to enter the second flow value which is to be simulated.

! Note!
• This function is only available if the STEP setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
0.000...999999 (6-digit floating-point number, with sign)

! Note!
The unit of the connected measuring device is adopted.
Unit:
The unit configured in the measuring device is adopted

Factory setting:
0.000

DURATION OF Use this function to enter the duration of simulation signal 1 (flow).
SIMULATION 1
! Note!
• This function is only available if the STEP setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 999 (max. 3-digit number )

Unit:
s ( seconds )

Factory setting:
10

DURATION OF Use this function to enter the duration of simulation signal 2 (flow).
SIMULATION 2
! Note!
• This function is only available if the STEP setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 999 (max. 3-digit number )

Unit:
s ( seconds )

Factory setting:
10

Endress+Hauser 51
Description of device functions - Fieldcheck and Coriolis measuring systems Proline Fieldcheck

Function group OPERATION

Function description
FUNCTION → SIMULATION → OPERATION

SIMULATE Use this function to control the simulation.

Options:
START
STOP

Factory setting:
START

Display during the simulation:


Main line
Symbol of the selected operating mode and the simulated value (with unit).
Additional line
The measured value sent back by the measuring device.
Information line
In the SIMULATION operating mode, the current values of the current and frequency
output of the tested device are displayed (multimeter function). Pulses are displayed as
frequency if there is more than 1 pulse/second at the frequency input of Fieldcheck.

In the STOP status the simulation signal value “0” is always given. Affected settings for
the simulation, however, are maintained.

52 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and Coriolis measuring systems

11.1.2 Group VERIFICATION


In the VERIFICATION group, a measuring device can be checked for the following characteristic
values.
• Functionality and deviation in the flow measurement (based on the set point preset by Fieldcheck)
• Functionality and deviation of the current and frequency outputs if these outputs are available in
the measuring device. The specified set points correspond in the gradation to the measuring
points when determining flow.
Data bus outputs such as PROFIBUS or FOUNDATION FIELDBUS cannot be directly checked by
FieldCheck.
• Characteristic values from internal tests for the sensor part.

Verification of the transmitter


Checking the linearity of the amplifier:
Simulated flow at four measuring points which are determined via the 100% full scale value. The
mass flow is verified using a constant density of 1.0kg/l. The presumed fluid velocity is 4...8m/s
for the standard range and 0.4...0.8m/s for the low range.
Checking a current output:
Each current output can be verified in the configuration in the measuring device (CURRENT
RANGE function ( 0-20 mA, 4-20 mA) in the same gradation (5% ... 100%) as the amplifier. For this
purpose, the current output must be directly connected to the current input of the Fieldcheck. If
more than one current output is available, verification takes place in several, menu-guided steps.
Checking a frequency output (FREQUENCY operating mode) :
Each frequency output can be verified in the configuration in the measuring device (START VALUE
FREQUENCY, END VALUE FREQUENCY, OUTPUT SIGNAL functions) in the same gradation ( 5%
... 100% ) as the amplifier. For this purpose, the frequency output must be directly connected to the
frequency input of the Fieldcheck. If more than one frequency output is available, verification takes
place in several, menu-guided steps.
Checking a frequency output (PULSE operating mode):
Each pulse output available can be verified in the configuration in the measuring device (PULSE
VALUE, PULSE WIDTH, OUTPUT SIGNAL functions). For this purpose, the flow specified by
Fieldcheck must trigger at least one pulse within 30 s. Otherwise, 0 PULSE is specified. In this
instance, the output is not completely tested in its components.

Verification of the sensor circuits


A series of internal tests are run through for checking the sensor connections and the closed-control
loops of excitation current and frequency with plugs connected for the excitation current and
amplifier. A PASSED or FAILED is output as a result.
The test series are combined into four groups for the result output:
• Excitation circuit
• Frequency parameters
• Temperature measuring points
• Pick-up coils

! Note!
One of the following options must be selected in the VERIFICATION function (→ Page 58) to carry
out the test series:
• TRANSMITTER + SENSOR
• SENSOR

Evaluation of the results


During verification, the current verification test section is displayed in the 'HOME position'. The
results can be shown on the display once verification is complete. They can be filed in Fieldcheck
under a storage location number of your choice (up to max. 20) together with the measuring device
serial no. and the measuring device tag name.

Endress+Hauser 53
Description of device functions - Fieldcheck and Coriolis measuring systems Proline Fieldcheck

Via a connection with the FieldTool software, a protocol of the results can later be printed out and
the data from the verification can be stored in an archive file system.

Function group APPLICATION/OUTPUTS


During verification, the function and accuracy of a current output and a frequency output can be
checked. Fieldcheck determines the current and frequency outputs available in the measuring
device.

Function description
FUNCTION → VERIFICATION → APPLICATION/OUTPUTS

FLOW RANGE Use this function to select the amplifier verification range suitable for the application in
question.

Options:
STANDARD RANGE:
Flow range of the original factory calibration.
LOW RANGE:
Compared to the standard range, the flow range has dropped by a factor of 10.

" Caution!
If the LOW RANGE option is selected, the smallest permissible limit value (error limit)
in the DEVIATION FLOW function (→ Page 56) is higher than in the STANDARD
RANGE option.

Factory setting:
STANDARD RANGE

TYPE UNIT FLOW Use this function to select the unit type for the verification of the amplifier.

Options:
MASS FLOW
CORRECTED VOLUME FLOW (not available for Promass 40)

! Note!
The option CORRECTED VOLUME FLOW is only available if the FIXED REFERENCE
DENSITY option was selected in the CORRECTED VOLUME CALCULATION function
in the tested device.
Factory setting:
MASS FLOW

54 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and Coriolis measuring systems

Function description
FUNCTION → VERIFICATION → APPLICATION/OUTPUTS

OUTPUTS Use this function to display the outputs determined in the flowmeter on the local display
with the related terminals and to select the outputs to be tested.
1. Use the Pkeys to select the outputs which should be tested during the verification
process.
2. Confirm the option selected with the F key.

! Note!
• It is not possible to select two outputs of the same type.
• Start the verification process in the VERIFICATE function (→ Page 58).
Two outputs of the same type are tested in two separate runs. After the 1st run, the
message MORE OUTPUTS TO BE TESTED? appears on the display.
If you want to test other outputs, Fieldcheck returns to the output selection area.
1. Select the output to be tested.

! Note!
Having selected an output, make sure that the related terminal pair of the flowmeter
is connected to the -Fieldcheck input before the verification process is started.
The following combinations of current and frequency outputs are possible for Proline
Promass:
• I
• I+I
• I+I+I
• I+F
• I+I+F
• I+F+F

! Note!
• Fieldbus interfaces cannot be tested.
• With this software-version current outputs run in passive mode cannot be tested.

Endress+Hauser 55
Description of device functions - Fieldcheck and Coriolis measuring systems Proline Fieldcheck

Function group LIMIT VALUES


You can specify the limit values for measured variables in this function group. If the deviation of a
verification step (see OPERATION, P. 38) lies within the configured limit values the result is
PASSED; if it lies outside of this range the result is FAILED. For a quick and unambiguous
identification, the FAILED results are presented inversely on the display.

Function description
FUNCTION → VERIFICATION → LIMIT VALUES

DEVIATION FLOW User input:


0.30...10.00 (for Promass 80/83 with sensors A, F, M)
0.50...10.00 (for Promass 80/83 with sensors E, H, I and Promass 40)
5.00...10.00 (if LOW RANGE is selected in the FLOW RANGE function, see Page 54)

Unit:
%

Factory setting:
0.5 or 5.0

DEVIATION DENSITY User input:


0.50 ... 10.00

Unit:
%

Factory setting:
0.5

DEVIATION User input:


TEMPERATURE 1.00...10.00 (for the units °C and K)
2.00...10.00 (for the units °F and R)

Unit:
The unit configured in the measuring device is adopted

Factory setting:
2.0 (for the units °C and K)
4.0 (for the units °F and R)

DEVIATION CURRENT This function is only available if a current output was selected in the OUTPUTS function
OUTPUT (→ Page 55).

User input:
0.02...10.0

Unit:
mA

Factory setting:
0.05

DEVIATION This function is only available if a frequency output was selected in the OUTPUTS
FREQUENCY OUTPUT function (→ Page 55).

User input:
1.0...10.0

Unit:
Hz

Factory setting:
2.00

56 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and Coriolis measuring systems

Function group PARAMETER


Verification of the amplifier takes place along four measuring points MP1...MP4. Use this function
group to enter the full scale value (100%) of the verification flow and the two selectable measuring
points MP2 and MP3 (in % of the full scale value).

! Note!
The value of MP1 is predetermined (5% of the full scale value MP4).

Function description
FUNCTION → VERIFICATION → PARAMETER

FLOW 100% Use this function to enter the full scale value of the verification flow.
User input:
max 6-digit floating point number (without sign)

! Note!
The permitted value range depends on the setting in the FLOW RANGE function
(→ Page 54).
The input limits MIN and MAX are displayed beyond the input line.
Unit:
The unit configured in the measuring device is adopted

Factory setting:
according 6m/s

MEASURING POINT Use this function to enter the measuring points MP2 and MP3 (in % of the full scale
2/3 value).

User input:
MP2: integer between 10 and 90
MP3: integer between 15 and 95

! Note!
• The value of MP2 must be less than the value of MP3.
• Both values are stored permanently.

Unit:
%

Factory setting:
MP2: 10 %
MP3: 50%

Endress+Hauser 57
Description of device functions - Fieldcheck and Coriolis measuring systems Proline Fieldcheck

Function group OPERATION


Use this function group to specify the scope of the verification and to start/stop the verification.

Function description
FUNCTION → VERIFICATION → OPERATION

VERIFICATION Use this function to define the scope of the test.

! Note!
A verification of the current and frequency outputs of the transmitter can only be achieved
with the TRANSMITTER or TRANSM. + SENSOR option.
Options:
TRANSMITTER:
The meter electronics and the electronics of the current and frequency outputs selected
are tested. Excitation current cable and sensor signal cable are plugged out.

SENSOR
The temperature measuring points, the closed-control loops of the excitation current and
frequency and the pick-up coils are tested. Excitation current cable and sensor signal cable
are plugged in.

TRANSM. + SENSOR
All test groups are run through. At the end of the transmitter test, you are requested to
reconnect the excitation current cable and sensor signal cable.

Factory setting:
TRANSMITTER

VERIFICATE Use this function to control the verification. After completion of the verification, the
question STORE DATA? appears in this function.

Options:
START
The verification is started and the message VERIFICATION ACTIVE appears on the
display.At the end of the verification, the message RESULT: PASSED/FAILED - STORE
DATA? appears on the display.
STOP
Via the STOP option the active verification can be aborted.
STORE DATA?
The results of the verification can be saved using the STORE DATA? option.

Factory setting:
START

STORAGE LOCATION Use this function to define where the results of the verification are to be stored.

User input:
1...20

Factory setting:
First available storage location

! Note!
If there is no storage location available or the selected storage location is occupied, a
corresponding warning is given before saving (“Overwrite storage location? Y/N”).

58 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and vortex measuring systems

12 Description of device functions - Fieldcheck and


vortex measuring systems

12.1 Block FUNCTION

Block Group Function group Function

OPERATION
SIMULATION CONFIGURATI SIMUL.
MODE FLOW 1 FLOW 2 SIMUL. TIME 1
FUNCTION ⇒ ⇒ ON ⇒ SIGNAL
P. 61 P. 61 P. 61
P. 60 P. 60 P. 60
P. 60

SIMUL. TIME 2
⇓⇑ ⇓⇑
P. 61

CONFIGURATI SIMULATE
ON
P. 62 P. 61

VERIFICATION OUTPUTS OUTPUTS

P. 63 P. 64 P. 64

⇓⇑
DEVIATION DEV.CURR.OU
LIMIT
FLOW TPUT
VALUESP. 65
P. 65 P. 65

⇓⇑

PARAMETER MEAS. POINT


FLOW 100%
1/2
P. 66
P. 66 P. 66

⇓⇑

OPERATION VERIFICATION STORAGE


VERIFICATE
LOCATION
P. 67
P. 67 P. 67 P. 67

Endress+Hauser 59
Description of device functions - Fieldcheck and vortex measuring systems Proline Fieldcheck

12.1.1 Group SIMULATION


Please note the following:
• In contrast to electromagnetic or Coriolis measuring devices, a gap-free flow range 0 - Qmax
cannot be simulated with vortex flowmeters. Below a certain fluid speed (approx. 0.15 m/s for
water), the amplitude of the vortex signal becomes so small that it is no longer possible to record
the measured value. For this reason, when setting the simulated flow, the values 0, the smallest
flow Qmin and the largest flow Qmax are displayed.
• The flow value actually simulated can deviate slightly from the value entered in the FLOW 1/2
function. The value actually simulated is displayed on the first line in the HOME position of the
Fieldcheck local display.

Function group CONFIGURATION

Function description
FUNCTION → SIMULATION → CONFIGURATION

OPERATION MODE Use this function to select the operation mode for the simulation.
The following signals can be freely configured:
• Operation mode GAS/STEAM: 0, Qmin...70 m/s
• Operation mode LIQUID: 0, Qmin...9 m/s
Options:

CONTINUOUS
The flow behaviour in the process has low dynamics.

Q
Q1

START STOP t
F06-DXC10xxx-05-xx-xx-xx-000

Q1 = Simulation value 1

STEP
The flow rates in the process are very dynamic and change very quickly. This operation
mode is also suitable for the simulation of flows outside the tolerance limits (Error
simulation). Here very dynamic or disrupted conditions can also be simulated.

Q T1
Q1

Q2
T2
t
START STOP
F06-DXC10xxx-05-xx-xx-xx-001

Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1,


T2 = Duration of simulation 2

SIMULATION SIGNAL Use this function to view the set simulation signal in the measuring device.

Display:
VOLUME FLOW
CALCULATED MASS FLOW
CALCULATED CORRECTED VOLUME FLOW

! Note!
• For the simulation of a calculated mass flow, the system uses the value specified by
the flowmeter in the OPERATING DENSITY function.
• For the simulation of corrected volume flow, the system uses the values specified by
the flowmeter in the OPERATING DENSITY and REFERENCE DENSITY functions.

60 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and vortex measuring systems

Function description
FUNCTION → SIMULATION → CONFIGURATION

FLOW 1 Use this function to specify the first flow value which is to be simulated.

User input:
0.000...99999 (6-digit floating-point number)

! Note!
The flow limits are adopted depending on the option selected in the APPLICATION
function of the flowmeter. Values smaller than Qmin cannot be entered (→ Page 60).
They have the following values:
• GAS/STEAM: 0, Qmin...70 m/s
• LIQUID: 0, Qmin...9 m/s

Unit:
The unit configured in the measuring device is adopted

Factory setting:
0.000

FLOW 2 Use this function to enter the second flow value which is to be simulated.

! Note!
• This function is only available if the STEP setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
• The flow limits are adopted depending on the option selected in the APPLICATION
function of the flowmeter. Values smaller than Qmin cannot be entered
(→ Page 60). They have the following values:
– GAS/STEAM: 0, Qmin...70 m/s
– LIQUID: 0, Qmin...9 m/s
User input:
0.000...99999 (6-digit floating-point number)

Unit:
The unit configured in the measuring device is adopted

Factory setting:
0.000

DURATION OF Use this function to enter the duration of simulation signal 1 (flow).
SIMULATION 1
! Note!
• This function is only available if the STEP setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 999 (max. 3-digit number )

Unit:
s ( seconds )

Factory setting:
10

DURATION OF Use this function to enter the duration of simulation signal 2 (flow).
SIMULATION 2
! Note!
• This function is only available if the STEP setting was selected in the OPERATION
MODE function.
• More information is provided in the OPERATION MODE function.
User input:
1 ... 999 (max. 3-digit number )

Unit:
s ( seconds )

Factory setting:
10

Endress+Hauser 61
Description of device functions - Fieldcheck and vortex measuring systems Proline Fieldcheck

Function group OPERATION

Function description
FUNCTION → SIMULATION → OPERATION

SIMULATE Use this function to control the simulation.

Options:
START
STOP

Factory setting:
START

Display during the simulation:


Main line
Symbol of the selected operating mode and the simulated value (with unit).
Additional line
The measured value sent back by the measuring device.
Information line
In the SIMULATION operating mode, the current values of the current and pulse output
of the tested device are displayed (multimeter function). Pulses are displayed as
frequency if there is more than 1 pulse/second at the frequency input of Fieldcheck.

In the STOP status the simulation signal value “0” is always given. Affected settings for
the simulation, however, are maintained.

62 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and vortex measuring systems

12.1.2 Group VERIFICATION


In the VERIFICATION group, a measuring device can be checked for the following characteristic
values.
• Functionality and deviation in the flow measurement (based on the set point preset by Fieldcheck)
• Functionality and deviation of the current output and optional pulse output if these outputs are
available in the measuring device. The specified set points correspond in the gradation to the
measuring points when determining flow.
Fieldbus outputs such as PROFIBUS or FOUNDATION FIELDBUS cannot be directly checked by
Fieldcheck.
! Note!
The device can be tested at its mounting location; power is supplied by the switching unit
connected to the current output. In order to test the current output for accuracy at the mounting
location, it must be disconnected from the switching unit and connected to the Fieldcheck
current input.

Verification of the transmitter


Checking the linearity of the amplifier:
Simulated flow at four measuring points which are determined via the 100% full scale value (for a
presumed fluid velocity of 8 m/s for liquids and 60 m/s for gases/steam).
Checking the current output:
The current output can be verified in the configuration in the measuring device (CURRENT RANGE
function (0-20 mA, 4-20 mA) in the same gradation (5% ... 100%) as the amplifier. For this purpose,
the current output must be directly connected to the current input of the Fieldcheck.
Checking the pulse output:
An existing pulse output can be verified in the configuration in the measuring device (PULSE
VALUE, PULSE WIDTH, OUTPUT SIGNAL functions). For this purpose, the pulse output must be
directly connected to the frequency input of the Fieldcheck.

Evaluation of the results


During verification, the current verification test section is displayed in the 'HOME position'. The
results can be shown on the display once verification is complete. They can be filed in FieldCheck
under a storage location number of your choice (up to max. 20) together with the measuring device
serial no. and the measuring device tag name.
Via a connection with the FieldTool software, a protocol of the results can later be printed out and
the data from the verification can be stored in an archive file system.

Endress+Hauser 63
Description of device functions - Fieldcheck and vortex measuring systems Proline Fieldcheck

Function group OUTPUTS


During verification, the function and accuracy of the current output and the optional pulse output
can be checked. Fieldcheck determines the outputs available in the measuring device.

Function description
FUNCTION → VERIFICATION → OUTPUTS

OUTPUTS Use this function to display the outputs determined in the flowmeter on the local display
with the related terminals and to select the outputs to be tested.
1. Use the Pkeys to select the outputs which should be tested during the verification
process.
2. Confirm the option selected with the Fkey.

! Note!
Start the verification process in the VERIFICATE function (→ Page 67).
The following combinations are possible for Proline Prowirl:
• I
• I+P

! Note!
Fieldbus interfaces cannot be tested.

64 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and vortex measuring systems

Function group LIMIT VALUES


You can specify the limit values for measured variables in this function group. If the deviation of a
verification step (see OPERATION, P. 38) lies within the configured limit values the result is
PASSED; if it lies outside of this range the result is FAILED. For a quick and unambiguous
identification, the FAILED results are presented inversely on the display.

Function description
FUNCTION → VERIFICATION → LIMIT VALUES

DEVIATION FLOW User input:


(when measuring 1.50...10.00
liquids)
Unit:
%

Factory setting:
1.5

DEVIATION FLOW User input:


(when measuring gas or 2.0...10.00
steam)
Unit:
%

Factory setting:
2.0

DEVIATION CURRENT This function is only available if the current output was selected in the OUTPUTS
OUTPUT function (→ Page 64).

User input:
0.02...1.00

Unit:
mA

Factory setting:
0.05

Endress+Hauser 65
Description of device functions - Fieldcheck and vortex measuring systems Proline Fieldcheck

Function group PARAMETER


Verification of the amplifier takes place along four measuring points MP1...MP4. Use this function
group to enter the two selectable measuring points MP2 and MP3 (in % of the full scale value).

! Note!
The value of MP1 is predetermined (5% of the full scale value MP4).

Function description
FUNCTION → VERIFICATION → PARAMETER

FLOW 100% Use this function to display the full scale value of the verification flow.

In liquid applications, the 100% value usually corresponds to a fluid velocity of 8 m/s
and a velocity of 60 m/s in gas/steam applications.

MEASURING POINT Use this function to enter the measuring points MP2 and MP3 (in % of the full scale
2/3 value).

User input:
MP2: integer between 10 and 90
MP3: integer between 15 and 95

! Note!
• The value of MP2 must be less than the value of MP3.
• Both values are stored permanently.
• In demanding measuring applications, the usable measuring range of the vortex
measuring device can be restricted to such an extent that the lower MP1 verification
measuring point cannot achieve the 5% value. In this instance, the minimum which
can be achieved of the measurable flow is determined and calculated as a % value of
the maximum value MP 4 and displayed. MP2 and MP3 are then determined again
as default values and displayed. However, they are not saved when the work with
Fieldcheck is completed.
Unit:
%

Factory setting:
MP2: 10 %
MP3: 50%

66 Endress+Hauser
Proline Fieldcheck Description of device functions - Fieldcheck and vortex measuring systems

Function group OPERATION


Use this function group to specify the scope of the verification and to start/stop the verification.

Function description
FUNCTION → VERIFICATION → OPERATION

VERIFICATION Use this function to define the scope of the test.

Options:
TRANSMITTER
SENSOR
TRANSMITTER + SENSOR

Factory setting:
TRANSMITTER

VERIFICATE Use this function to control the verification. After completion of the verification, the
question STORE DATA? appears in this function.

Options:
START
The verification is started and the message VERIFICATION ACTIVE appears on the
display.At the end of the verification, the message RESULT: PASSED/FAILED - STORE
DATA? appears on the display.
STOP
Via the STOP option the active verification can be aborted.
STORE DATA?
The results of the verification can be saved using the STORE DATA? option.

Factory setting:
START

STORAGE LOCATION Use this function to define where the results of the verification are to be stored.

User input:
1...20

Factory setting:
First available storage location

! Note!
If there is no storage location available or the selected storage location is occupied, a
corresponding warning is given before saving (“Overwrite storage location? Y/N”).

Endress+Hauser 67
Proline Fieldcheck Index

Index
A DEVIATION FREQUENCY OUTPUT
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . 25 Electromagnetic measuring devices . . . . . . . . . . . . . . . 46
ASSIGN (current output) . . . . . . . . . . . . . . . . . . . . . . . . . 33 DEVIATION MASS FLOW
ASSIGN (frequency output) . . . . . . . . . . . . . . . . . . . . . . . 33 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56
DEVIATION TEMPERATURE
B Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56
BACK LIGHT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
BASIC FUNCTIONS (Block) . . . . . . . . . . . . . . . . . . . . . . . 31 Device functions
Block General section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
BASIC FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Test Coriolis measuring devices . . . . . . . . . . . . . . . . . . 49
FUNCTION Test electromagnetic measuring devices. . . . . . . . . . . . 39
Coriolis measuring devices. . . . . . . . . . . . . . . . . . . 49 Test vortex measuring devices . . . . . . . . . . . . . . . . . . . 59
Electromagnetic measuring devices . . . . . . . . . . . . 39 dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Vortex measuring devices . . . . . . . . . . . . . . . . . . . 59 DIRECTION OF FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . 32
SUPERVISION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, 26
USER INTERFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Display and operating elements . . . . . . . . . . . . . . . . . . . . . 17
Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Display elements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Display icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
C Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
CALIBRATION DATE DURATION OF SIMULATION 1
Fieldcheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 51
Simubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Electromagnetic measuring devices . . . . . . . . . . . . . . . 41
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7, 26 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61
Charging the battery DURATION OF SIMULATION 2
Battery charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 51
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Electromagnetic measuring devices . . . . . . . . . . . . . . . 42
Communication Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61
FieldTool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
CONFIGURATION (Function group) E
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 50 Electrical connection
Electromagnetic measuring devices . . . . . . . . . . . . . . . 40 Assignment of Fieldcheck connections . . . . . . . . . . . . . 16
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 60 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connection Promag 50/53 (compact version) . . . . . . . . . . . . . . . . 10
see Electrical connection Promag 50/53 (remote version) . . . . . . . . . . . . . . . . . 11
CONTRAST LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Promass 80/83 (compact version) . . . . . . . . . . . . . . . . 12
CONTROL (Group) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Promass 80/83 (remote version) . . . . . . . . . . . . . . . . . 13
CURRENT RANGE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Prowirl 72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Electrode integrity test Electromagnetic measuring devices 44
D Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . 25
DATE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 EMC - operation in environment with strong interference . . 4
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 7 EN 61010 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
DEL. STORAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 END FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
DENSITY VALUE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Evaluation of the results
Description of functions Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 53
Sections 9 to 12 Electromagnetic measuring devices . . . . . . . . . . . . . . . 45
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 63
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 External cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
DEVIATION CURRENT OUTPUT
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56
Electromagnetic measuring devices . . . . . . . . . . . . . . . 46
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 65
DEVIATION DENSITY
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56
DEVIATION FLOW
Electromagnetic measuring devices . . . . . . . . . . . . . . . 46
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 65

68 Endress+Hauser
Proline Fieldcheck Index

F SENSOR PARAMETER. . . . . . . . . . . . . . . . . . . . . . . . . 32
FIELDCHECK (Function group). . . . . . . . . . . . . . . . . . . . . 38 SIMULATION
FLOW 1 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 50
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 51 Electromagnetic measuring devices . . . . . . . . . . . . . 40
Electromagnetic measuring devices . . . . . . . . . . . . . . . 41 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 60
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61 VERIFICATION
FLOW 100% . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 53
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 57 Electromagnetic measuring devices . . . . . . . . . . . . . 44
Electromagnetic measuring devices . . . . . . . . . . . . . . . 47 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 63
FLOW 2 VERSION INFO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 51 Groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Electromagnetic measuring devices . . . . . . . . . . . . . . . 41 Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61
FORMAT DATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 H
FORMAT TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 History (software) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
FUNCTION (Block) HOME position (display operating mode) . . . . . . . . . . . . . . 17
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 49 HW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Electromagnetic measuring devices . . . . . . . . . . . . . . . 39
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 59
I
Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Function group
INFO STORAGE LOCATION. . . . . . . . . . . . . . . . . . . . . . . 36
CONFIGURATION
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 50
Electromagnetic measuring devices. . . . . . . . . . . . . 40 K
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 60
K-FACTOR POSITIVE/NEGATIVE . . . . . . . . . . . . . . . . . . 32
FIELDCHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
LIMIT VALUES L
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 56 LANGUAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Electromagnetic measuring devices. . . . . . . . . . . . . 46 LIMIT VALUES (Function group)
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 65 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 56
OPERATION (Group SIMULATION) Electromagnetic measuring devices. . . . . . . . . . . . . . . . 46
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 52 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 65
Electromagnetic measuring devices. . . . . . . . . . . . . 43 Local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 62
OPERATION (Group VERIFICATION) M
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 58 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Electromagnetic measuring devices. . . . . . . . . . . . . 48 MANUFACTURING/TEST . . . . . . . . . . . . . . . . . . . . . . . . 38
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 67 Measured error (inputs) . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
OUTPUTS Measured error (outputs) . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 54 Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Electromagnetic measuring devices. . . . . . . . . . . . . 45 MEASURING POINT 2/3
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 64
PARAMETER Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 57
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 57 Electromagnetic measuring devices. . . . . . . . . . . . . . . . 47
Electromagnetic measuring devices. . . . . . . . . . . . . 47 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 66
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 66 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
SIMUBOX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Function groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 N
Brief operating instructions . . . . . . . . . . . . . . . . . . . . . 20 Nameplate
Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Fieldcheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Simubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
G Nameplate specifications
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Fieldcheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Group Simubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 NOMINAL DIAMETER . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
OUTPUTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
PROCESS PARAMETER. . . . . . . . . . . . . . . . . . . . . . . . 34
READ IN PARAMETERS . . . . . . . . . . . . . . . . . . . . . . . 35
RESULTS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

Endress+Hauser 69
Proline Fieldcheck Index

O R
Operating conditions (Environment) . . . . . . . . . . . . . . . . . 25 Radiated RF noise (EN 61326-1) . . . . . . . . . . . . . . . . . . . . . 4
OPERATING DENSITY . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 READ IN PARAMETERS (Group) . . . . . . . . . . . . . . . . . . . 35
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 REFERENCE DENSITY . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Reference operating conditions . . . . . . . . . . . . . . . . . . . . . 25
OPERATION (Group SIMULATION) Registered trademarks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 52 Remote configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Electromagnetic measuring devices . . . . . . . . . . . . . . . 43 RESULTS (Group). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 62 RESULTS VERIFICATION CORIOLIS . . . . . . . . . . . . . . . . 37
OPERATION (Group VERIFICATION) RESULTS VERIFICATION MAGMETER . . . . . . . . . . . . . . 36
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 58 RESULTS VERIFICATION VORTEX . . . . . . . . . . . . . . . . . 37
Electromagnetic measuring devices . . . . . . . . . . . . . . . 48
Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 67 S
OPERATION MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Safety conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 50 Safety instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Electromagnetic measuring devices . . . . . . . . . . . . . . . 40 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 60 Sensor circuit verification (Coriolis measuring devices) . . . 53
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 SENSOR PARAMETER (Group) . . . . . . . . . . . . . . . . . . . . 32
Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Sensor verification
Other standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Electromagnetic measuring devices . . . . . . . . . . . . . . . 44
OUTPUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 SERIAL NUMBER
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Fieldcheck. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
OUTPUT SIGNAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Simubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 SERIAL NUMBER (Function)
OUTPUTS (Function group) Measuring device . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 54 SIMUBOX (Function group) . . . . . . . . . . . . . . . . . . . . . . . 38
Electromagnetic measuring devices . . . . . . . . . . . . . . . 45 SIMULATE
Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 64 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 52
OUTPUTS (Group) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Electromagnetic measuring devices . . . . . . . . . . . . . . . 43
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 62
P SIMULATION (Group)
PARAMETER (Function group) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 50
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 57 Electromagnetic measuring devices . . . . . . . . . . . . . . . 40
Electromagnetic measuring devices . . . . . . . . . . . . . . . 47 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 60
Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 66 SIMULATION SIGNAL
Performance characteristics . . . . . . . . . . . . . . . . . . . . . . . . 25 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 50
Power consumption. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Electromagnetic measuring devices . . . . . . . . . . . . . . . 41
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 60
PROCESS PARAMETER (Group) . . . . . . . . . . . . . . . . . . . . 34 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
PROFILE TIME 1 STANDARD DENSITY . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Electromagnetic measuring devices . . . . . . . . . . . . . . . 42 START FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
PROFILE TIME 2 STORAGE LOCATION . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Electromagnetic measuring devices . . . . . . . . . . . . . . . 42 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 58
Promag 50/53 (compact version) Electromagnetic measuring devices . . . . . . . . . . . . . . . 48
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 67
Promag 50/53 (remote version) STORAGE LOCATION VERIFICATION . . . . . . . . . . . . . . 36
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 11 SUPERVISION (Block) . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Promag 80/83 (compact version) Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 12 SW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Promag 80/83 (remote version)
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 13 T
Prowirl 72 TAG NAME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Technical data at a glance . . . . . . . . . . . . . . . . . . . . . . . . . 25
PULSE VALUE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Transmitter verification
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 53
Electromagnetic measuring devices . . . . . . . . . . . . . . . 44
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 63
TYPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

70 Endress+Hauser
Proline Fieldcheck Index

U
User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
USER INTERFACE (Block). . . . . . . . . . . . . . . . . . . . . . . . . 29

V
VALUE 0_4 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
VALUE 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
VALUE f- . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33–34
VERIFICATE (Function)
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 58
Electromagnetic measuring devices . . . . . . . . . . . . . . . 48
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 67
VERIFICATION (Function)
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 58
Electromagnetic measuring devices . . . . . . . . . . . . . . . 48
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 67
VERIFICATION (Group)
Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 53
Electromagnetic measuring devices . . . . . . . . . . . . . . . 44
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 63
VERSION INFO (Group) . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . . . . 66

W
Wiring
see Electrical connection

Z
ZERO POINT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Endress+Hauser 71

You might also like