Instrumed 3600 Operating
Instrumed 3600 Operating
System 3600
Operating manual
1
Index
2
Surgical micromotor
1. Product introduction
Description
The device is an electrical surgical product for a wide range of operations for neuro, column and
orthopedics.
Model: BJWJZ-1
Note: to simplify, angled and straight attacks are collectively called bur attachments
2. Product classification
Based on the classification of the protection against electric shock, the device is considered a
general device belonging to the application part of the Ⅰ BF class.
It cannot be used when the flammable anesthetic gas is mixed with flammable air or narcotic or
mixed with oxygen or nitrous oxide.
Operating system: continuous operation with intermittent loading. Starting from cold, operate for
2 minutes in charging condition. Please stop the continuous use, allowing the device to cool down.
After the break, you can continue the operation.
3
Spiegazione dei simboli generali usati
6 Technical specification
5
1. Button for switching on / off
2. Connection port for micro-handpiece BJ3600
3. Connection port for BJ5000 series handgun modular gun, or dedicated handpiece for cranial
drilling BJ3604
4. Connection port for BJ36012 pedal board
5. On-screen button to choose between activation by footswitch or by screen
6. Direction of rotation
7. Increase and decrease speed (indication of same on screen in rpm)
8. Blade mode (for micro-handpiece BJ3600)
9. Milling (bur) mode (for micro-handpiece BJ3600)
10. Cranial perforator mode (for dedicated BJ3604 handpiece)
6
6.2 Foot pedal BJ36012
Cable: 3 meters
Water resistant IPX6
1. Handpiece connector
2. Motor control connector
3. Handpiece
3. Cable
7
6.4 Bur attachments
Speed: 1000rpm-100000rpm
Error:±8%
Torque: 0.10 Nm
Connection type: standard round shank 2,35mm
Speed: 1000rpm-100000rpm
Error:±8%
Torque: 0.10 Nm
8
6.6 Sagittal and reciprocating attachments BJ3601 – BJ3609
Speed: 1000rpm-15000rpm
Error:±8%
Connection type for sagittal saw blades: code BJ3501
Connection type for reciprocating saw blades: code WF
9
NOTE: THIS USER MANUAL WILL FOCUS ON THE USE OF THE CONTROL UNIT WITH THE BJ3600 HIGH-SPEED
MICRO-HANDPIECE. FOR THE INSTRUCTIONS FOR USING THE BJ5000 SERIES, REFER TO THE MANUAL FOR USE OF THE
SAME SERIES.
1. Connect the electric cable to the instrument console, the light indicator indicates the power on and by pressing the
start key, the monitor indicates the icons of the different and optional operations.
2. Once activated, the instrument shows the PEDAL mode and shows the minimum operating speed; the MANUAL mode
is not shown with the icon.
3. In the pedal control mode, the screen rotation change buttons are not accessible. Press the speed button to change it,
setting the maximum reachable. By pressing the pedal, the rotary speed of the motor increases linearly from the lowest
point to the highest point that was previously set.
4. In manual control mode, press the activation button. Manual control mode is shown with the button icon. The pedal
switch is not accessible. Then, adjust the parameters on the control panel to set the speed and direction of the motor,
press the manual control mode activation button again, the motor will run in the current speed and direction.
5. When the instrument is not in use, please disconnect the power cord.
Note:
• After following the instructions, now described, please pay attention that there are no noises or imperfections of the
instrument.
• In manual control mode, do not touch the pedal control mode activation button, otherwise it will turn into the pedal
control mode.
• Depending on the different surgical needs, it is possible to select different drill attachments (straight or angled) and
insert boring bits. To lock the cutter, insert it up to the bottom with the locking system in the open position. Once the
cutter is inserted, rotate the locking system in the opposite direction. To make sure that the cutter is firm, pull it slightly.
• To connect the sagittal blade, press the button on the attachment. Keeping the same pressed, insert the blade and
make the holes of the blade correspond with the teeth of the locking system. Once the connection has been made,
release the button and check that the blade is firmly in place by pulling it slightly.
• To connect the blade to each other, rotate the locking system and align the blade inlet line. Insert the blade, as the
locking system rotates. Once the blade is inserted, release the rotating part. Check that the blade is firm, pulling it
slightly.
10
8 Cleaning and sterilization procedure
The devices must be cleaned and sterilized before each use. Use protective equipment every time even such operations
are performed.
Use a neutral pH detergent, not exceeding 10.5. Fill the sink with water and detergent, mix gently by hand. Before
cleaning the device, remove all accessories, battery connections and shells. Rinse the device under warm running water
(between 27 ° C and 44 ° C) using a soft, lint-free cloth and a soft bristle brush. Remove coarse dirt. Be careful to move
all moving parts, such as triggers, bushings and selectors under running water to remove debris. Do not use sharp
objects.
Never immerse the devices in an aqueous solution or in an ultrasonic bath. Also, do not allow moisture to remain inside
the cannulated parts, in the selectors or in the electrical connection areas. Do not rinse the handpiece from the front or
use pressurized water to avoid damaging the system.
Use a soft brush with a diameter suitable for the cannulated part to clean it. Pay attention that the brush passes through
the total length of the affected area. Move all the moving parts to clean the residual debris. Keep the device tilted, with
the head pointing downwards, under running water. Be careful to completely remove the detergent under running
water.
Visually inspect the device, including cannulations to check for traces of dirt. If present, repeat the previous steps. I
would place inclined devices on absorbent cloths to dry them. Or use compressed air (medical use).
To clean a connection cable, hold it by both ends and clean the cable using a soft cloth dampened with detergent. Finally
dry with compressed air.
11
.
When sterilizing, perform the following sterilization cycle validated by Bojin to obtain optimum performance. Steam
sterilization
Wrapping method Cycle Minimum temperature Maximum Temperature Minimum Exposure Minimum dry time
Note: The Operator must ensure that the affected parts have been sterilized, before use.
12
9 Fault handling
Maintenance
For safety reasons, electrical microsurgery devices must be inspected once a year. For
maintenance operations, please contact your dealer or contact the Bojin after-sales service directly
for maintenance, performed by the instrument's experts.
For maintenance performed by skilled and qualified figures, please turn off the power switch,
check if the fuse is blown. If it is, replace it with the one from the T2AL250V specifications.
10 Electromagnetic Requirements
Operation is subject to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Use of accessories other than those recommended may result in non-compliance with
electromagnetic compatibility and immunity standards.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
13
3600 Handpieces are intended for use in the electromagnetic environment specified below.
The customer or the user of the 3600 Handpieces should assure that they are used in such
an environment
Emissions test Complian Electromagnetic environment - guidance
RF Emissions CISPR 11 ce
Group 1 The 3600 Handpieces use RF energy only for
internal functions; therefore, RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions CISPR 11 Class A The 3600 Handpieces are intended for use by
healthcare professionals only and is suitable for
use in all establishments other than domestic
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
14
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The 3600 Handpieces are intended for use in the electromagnetic environment specified
below. The customer or the user of the 3600 Handpieces should assure that it is used in
Immunity Test
such an environment. IEC 60601 Compliance Level Electromagnetic
Test Level Environment
Electrostatic ± 6 kV contact ± 6 kV contact Floors
Guidanceshould be
discharge (ESD) IEC ± 8 kV air ± 8 kV air wood, concrete or
61000-4-2 ceramic tile. If floors
are covered with
Electrical fast ± 2 kV for power ± 2 kV for power Mains powermaterial,
synthetic quality
supply lines supply lines should
the relative humidity
should be at least
transients/bursts be that of a typical
30%.
environment.
Power frequency 3 A/m 3 A/m Power frequency
(50/60 Hz) magnetic fields
magnetic field should be at levels
IEC 61000-4-8 characteristic of a
typical location in a
typical commercial or
15 hospital
environment.
Voltage dips, short <5% Ut (>95% dip in <5% Ut (>95% dip in Mains power quality
interruptions and Ut) Ut) should be that of a
input linesvariations
voltage 40%
for 0.5Utcycle
(60% dip in 40%
for 0.5Utcycle
(60% dip in environment. If the
typical commercial or
IEC 61000-4-11
on power supply Ut) Ut) user of
hospital the 3600
70% Ut (30% dip in
for 5 cycles 70% Ut (30% dip in
for 5 cycles during powerrequires
Handpieces mains
Ut) Ut) interruptions, it is
continued operation
for 25 cycles for 25 cycles recommended that
the 3600
16
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (Continued)
The 3600 Handpieces are intended for use in the electromagnetic environment specified
below. The customer or the user of the 3600 Handpieces should assure that it is used in
Immunity Test
such an environment. IEC 60601 Compliance Level Electromagnetic
Test Level Environment
Recommended
Guidance
Separation Distance
Conducted RF IEC 150 kHz to 80 MHz 3 Vrms
61000-4-6RF
Radiated 80 MHz to 2.5 GHz 3 V/m d = 1.2 √√ PP 80 MHz to
IEC 61000-4-3 800 MHz
Where P is the maximum output where P is the maximum output dpower = 2.3 √rating
P 800ofMHz
the
transmitter in watts (W) according to the transmitter manufacturer and d is the
to 2.5 GHz
recommended separation distance in meters (m). Field strengths from fixed RF transmitters,
as determined by an electromagnetic site survey a, should be less than the compliance level
in each1:frequency
NOTE At 80 MHzrange b. MHz, the higher frequency range applies.
and 800
Interference
NOTE 2: These mayguidelines
occur nearmay
the not
equipment
apply inmarked with the Electromagnetic
all situations. following symbol:propagation is
affected by absorption and reflection from structures, objects, and/or people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the 3600 Handpieces is
used exceeds the applicable RF compliance level above, 3600 Handpieces should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the 3600 Handpieces.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
17
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the 3600 Handpieces @ 3Vrms
3600 Handpieces are intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of 3600 Handpieces can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the 3600
Separation Distance
Handpieces as recommended Accordingto
below, according to the
Frequency
maximumof Transmitter (meters)
output power of the
Rated Maximumequipment.
communications m
Output Power of 150 kHz to 80 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Transmitter MHz d 3.5 P d 7 P
(Watts) d 3.5 P E E
0.01 0.12 0.12 1 0.23 1
0.1 0.34
v1 0.34 0.74
1 1.7 1.7 2.3
10 3.7 3.7 7.4
100 11.7 11.7 23.3
For transmitters rated at a maximum output power not listed above, the recommended
separation distances d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts
Note 1: (W) according
At 80 MHz and to 800
the transmitter manufacturer.
MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
12 Operating conditions
normal working conditions:
ambient temperature:5 ℃~35℃
relative humidity:35%~75%
Kpa: 860hPa~1060hPa
Supply voltage:AC 110V/220V,tolerance±10%;frequency:50Hz,tolerance±1Hz
motor power:(DC)12 V
There is no conductive dust, explosive gas, and corrosive gases
13 Manufacturing date and period of use
14 Commitment
A) This device has no removable parts for non-professional personnel, in case of failure,
please contact the after-sales service.
B) If you need technical information (such as circuit diagram, list of components, illustrations,
correction details, etc.) you can request it at the Technical Assistance Center.
Shanghai Bojin Electric Instrument & Device Co., Ltd. Bojin Europe srls
Add: Room 606, No. 3388, Gonghexin Road, Shanghai City. China,200436 Via Ortana 20, Vitorchiano 01030
Web: http://www.bojin-medical.com Italy
Tel: 0086 21 66308078 eurosales@bojin-medical.com
Fax: 0086 21 66527013
19