0 ratings0% found this document useful (0 votes) 86 views51 pagesMri Image Formation
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
cuarres 13) | EZ
Magnetic Resonance
Imaging: Advanced Image
Acquisition Methods, Artifacts,
Spectroscopy, Quality Control,
Siting, Bioeffects, and Safety
‘The essence of magnetic resonance imaging (MRI) in medicine is the acquisition,
manipulation, display, and archive of datasets that have clinical relevance in the
context of making a diagnosis or performing research for new applications and
opportunities. There are many advantages and limitations of MRI and MR spectros-
copy (MRS) as a solution to a clinical problem. Certainly, as described previously
(note that this chapter assumes a working knowledge of Chapter 12 content), the
great advantages of MR are the ability to generate images with outstanding tissue
contrast and good resolution, without resorting to ionizing radiation. Capabilities of
MR extend far beyond those basics, into fast acquisition sequences, perfusion and
diffusion imaging, MR angiography (MRA), tractography, spectroscopy, and a host of
other useful or potentially useful clinical applications. Major limitations of MR are
also noteworthy, including extended acquisition times, MR artifacts, patient claustro-
phobia, issue heating, and acoustic noise to name a few. MR safety. often ignored, 1s
also of huge concern to the safety of the patient
In this second of two MR chapters, advanced pulse sequences and fast image
acquisition methods, dedicated radiofrequency (RF) coils, methods for perfusion
diffusion, and angiography imaging, image quality metrics, common artifacts, spec-
troscopy, MR equipment and siting, as well as MR safety issues are described and
discussed with respect to the underlying phystes
The concepts of image acquisition and timing issues for standard and advanced
pulse sequences into k-space is discussed first, with several methods that can be used
to reduce acquisition times and many of the trade-offs that must be considered
33] Image Acquisition Time
A defining character of MRI is the tremendous range of acquisition time needed to
image a patient volume. Times ranging from as low as 50 ms to tens of minutes are
commonly required depending on the study, pulse sequence, number of images in
the dataset, and desired image quality. When MR was initially considered to be a
potential diagnostic imaging modality in the late 1970s, the prevailing conventional
wisdom gave no chance for widespread applicability because of the extremely long
limes required to generate a single slice from a sequentially acquired dataset, which
4ag450
Section Il ¢ Diagnostic Radiology
required several minutes or more per slice. Breakthroughs in technology, equip.
ment design, RF coils, the unique attributes of the k-space matrix, and methods of
acquiring data drastically shortened acquisition times (or effective acquisition times)
quickly, and propelled the rapid adoption of MRI in the mid-1980s, By the early
1900s, MRI established its clinical value that continues to expand today.
Acquisition Time, Two-Dimensional Fourier Transform Spin
Echo Imaging
‘The time to acquire an image is determined by the data needed to fill the fraction
of kespace that allows the image to be reconstructed by Fourier transform methods,
For a standard spin echo sequence, the relevant parameters are the TR, number of
phase encoding steps, and number of excitations (NEX) used for averaging identical
repeat cycles, as
Acquisition time = TR X # PEG Steps x NEX
Even though there may be multiple echoes as illustrated in Figure 13-1, there ts
also the same number of k-space repositories to capture the data in a specific, single
row of k-space defined by the strength of the PEG, as shown for the first echo
with proton density weighting and second echo with T2 weighting for this double
echo acquisition. Thus, effective imaging time can be reduced by producing wo
(or more) images of the same slice within the TR interval. In addition, the matrix
size that defines k-space is often not square (e.g., 256 X 256, 128 X 128), but
0
A
—r
~ a
_~ ~
cS i S
ke Vv ke Vv
echo, proton density weighted 2echo, T2 weighted
FIGURE 13-1 Standard spin echo pulse sequence is shown with two echoes per TR interval to encode
proton density contrast (short TE, first echo}, and T2 contrast (long TE, second echo). In this acquisition, two
separate images are acquited independently by storing in a designated k-space matnx accorcing to echo time.
‘Assingle PEG strength is momentarily applied to induce phase variations to encode the row to be filled in each
of the matrices (see the red PEG encoding for the last row in k-space, for instance). The full k-space matrix
requires the sequence to be repeated with incremental variations in the PEG strength until each k-space row is,
fully populated. If averaging is desired, then an identical sequence (without incrementing the PEG) is repeated
and averaged in the same row.Chapter 13 * Magnetic Resonance Imaging 451
rectangular (e.g, 256 X 192, 256 X 128) where the small matrix dimension is most
frequently along the phase encode direction to minimize the number of incremental
PEG strength applications during the acquisition. A 256 X 192 image matrix and
two averages (NEX) per phase encode step with a TR = 600 ms (for T1 weighting)
requires imaging time of 0.6 s X 192 X 2 = 230.4 s = 3.84 min for a single slice!
For a proton density and T2-weighted double echo sequence with TR = 2,500 ms
(Fig, 13-1), this increases to 16 min, although two images are created in that time
Of course, a simple first-order method would be to eliminate the number of aver-
ages (NEX), which reduces the time by a factor of 2; however, the downsides are an
increase in the statistical variability of the data, which decreases the image signal-to-
noise ratio (SNR) and makes the image appear “noisy.” Methods to reduce acquisi-
tion time and/or time per slice are crucial to making MR exam times reasonable, as
described by various methods below:
Multislice Data Acquisition
The average acquisition time per reconstructed image slice in a single-slice spin echo
sequence is clinically unacceptable. However, the average time per slice is significantly
reduced using mulkislice acquisition methods, where several slices within the tissue
volume are selectively excited in a sequential timing scheme during the TR interval
to fully utilize the dead time waiting for longitudinal recovery in an adjacent slic
as shown in Figure 13-2. This requires cycling all of the gradients and tuning the RF
excitation pulse many times during the TR interval. The total number of slices that
can be acquited simultaneously isa function of TR, TE, and machine limitations
Total Number of Slices = TRATE + C),
where C is a constant dependent on the MR equipment capabilities (computer speed,
gradient capabilities, sequence options, additional pulses, e g.. spoiling pulses in
standard SE; use of spatial saturation; and chemical shift, among others). Each slice
and each echo, if multiecho, requires its own k-space repository to store data as its
acquired. Long TR acquisitions such as proton density and T2-weighted sequences
+“_______
e— hy
gp
#lices = TR /(TE+C)
FIGURE 13-2 Multisice two-dimensional image acquisition is accomplished by discretely exciting different
slabs of tissue during the TR period; appropriate changes of the RF excitation bandwidth, SSG, PEG, anc FEG
parameters are necessary. Because cf diffuse excitation profiles, RF irradiation of acjacent slices leads to partial
saturation and loss of contrast. The number of slices (volume) that can be obtained is a function of the TR. TE,
and C, the latter representing the capabilities of the MR system and type of pulse sequence.452 Section Ii + Diagnostic Radiology
can produce a greater number of slices over a given volume than T!-weighted
sequences with a short TR. The chief trade-off is a loss of tissue contrast due to cross-
excitation of adjacent slices due to nonsquare excitation profiles, causing undesired
proton saturation as explained in Section 13.5 on artifacts.
Data Synthesis
Data “synthesis” takes advantage of the symmetry and redundant characteristics of
the frequency domain signals in k-space. The acquisition of as little as one-half the
data plus one row of k-space allows the mirroring of "complex conjugate” data to fill
the remainder of the matrix (Fig. 13-3). In the phase encode direction, “half Fourier,”
“Ya NEX,” or “phase conjugate symmetry” (vendor-specific names) techniques effec-
tively reduce the number of required TR intervals by one-half plus one line, and thus
can reduce the acquisition time by nearly one-half. In the frequency encode direc-
tion, “fractional echo" or “read conjugate symmetry” refers to reading a fraction of
the echo. While there is no scan time reduction when all the phase encode steps are
acquired, there is a significant echo time reduction, which can reduce motion-related
artifacts, such as dephasing of blood. However, the penalty for either half Fourier or
fractional echo techniques is a reduction in the SNR (caused by a reduced NEX or
Fractional NEX: Fractional Echo:
Acquired data = % matrix + 1 line minimum TE reduced
ky
zz
WAM
‘Synthesized mirror image data from Mirror’ k, Data
‘opposite quadrants image extracted
FIGURE 13-3 Fractional NEX and Fractional Echo. Left. Data synthesis uses the redundant characteristics
of the frequency domain. This is an example of phase conjugate symmetry. in which 2 of the PEG views +1
tna are acquired, and te complex conagateof he data eflectedn the symmetric quadrants. Acoust
time is thus reduced by approumately Y2 (~ 40%), although image nosse 1s increased by approximately V2
Right: Fractional echo acquisition is performed when only part of the echo is read during the application of the
FEG. Usually, the peak of the echo is centered in the middle of the readout gradient, and the echo signals prior
to the peak are identical mirror images after the peak. With fractional echo, the echo is no longer centered,
and the sampling window is shifted such that only the peak echo and the dephasing part of the echo are
sampled. As the peak of the echo is closer to the RF excitation pulse, TE can be reduced, which can improve
TI and proton density weighting contrast. A larger number of slices can also be obtained with a shorter TE in
a multislice acquisition (see Fig. 13-2)Chapter 13 * Magnetic Resonance Imaging 453,
data sampling in the volume) and the potential for artifacts if the approximations in
the complex conjugation of the signals are not accurate. Other inaccuracies result
from inhomogeneities of the magnetic field, imperfect linear gradient fields, and the
presence of magnetic susceptibility agents in the volume being imaged.
Fast Pulse Sequences
Fast Spin Echo (FSE) techniques use multiple PEG steps in conjunction with mul-
Liple 180-degree refocusing RF pulses to produce an echo train length (ETL) with
corresponding digital data acquisitions per TR interval, as illustrated in Figure 13-4
Multiple k-space rows are filled during each TR equal to the ETL, which is also
the reduction factor for acquisition time. “Effective echo time” is determined when
the central views in k-space are acquited, which are usually the first echoes, and
subsequent echoes are usually spaced apart via increased PEG strength with the
same echo spacing time. “Phase re-ordering” optimizes SNR by acquiring the low-
frequency information with the early echoes (lowest amount of T2 decay), and the
high-frequency, peripheral information with late echoes, where the impact on over-
all image SNR is lower. The FSE technique has the advantage of spin echo image
acquisition, namely immunity from external magnetic field inhomogeneities, with
4X, 8X, 10 16% faster acquisition time. However, each echo experiences difler-
ent amounts of intrinsic T2 decay, which results in image contrast differences when
compared with conventional spin echo images of similar TR and TE. Lower sig-
nal levels in the later echoes produce less SNR, and fewer images can be acquired
in the image volume during the same acquisition. A T2-weighted spin echo image
CTR = 2,000 ms, 256 phase encode steps, one average) requiresapproximately 8.5 min,
while a corresponding FSE with an ETL of 4 (Fig. 13-4) requires about 2.1 min,
PEG __
FES te
Echo a a
Bteaive
Echo train length (ETL) = 4
Effective TE = 16 ms
FIGURE 13-4 Conventional FSE uses multiple 180-degree refocusing RF pulses per TR interval with incre-
mental changes in the PEG to fil several views in k-space (the ETL). This example illustrates an ETL of four.
‘with an “effective” TE equal to 16 ms. Total time of the acquisition 1s reduced by the ET factor. The reversed
polarity PEG steps reestablish coherent phase before the next gradient application. Slightly different PEG
strengths are applied to fill the center of k-space first, and then the periphery with later echoes, continuing
Until all views are recorded. As shown, data can be mirrored using conjugate symmetry to reduce the overall
time by another factor of two,asa
Section II » Diagnostic Radiology
Longer TR values allow for a greater ETL, which will offset the longer TR in terms of
overall acquisition time, and will also allow more proton density weighting Specific
FSE sequences for T2 weighting and multiecho FSE are employed with variations
in phase reordering and data acquisition. FSE is also known as “turbo spin echo” or
“RARE” (rapid acquisition with refocused echoes)
A Gradient Echo (GE) Acquisition pulse sequence is similar to a standard
spin echo sequence with a readout gradient reversal substituting for the 180-degree
pulse (Fig. 13-5). Repetition of the acquisition sequence occurs for each PEG step
and with each average. With small f_ip angles and gradient reversals, a consider-
able reduction in TR and TE is possible for fast image acquisition: however, the
ability to acquire multiple slices is compromised. A PEG rewinder pulse of oppo-
site polarity is applied to maintain phase relationships from pulse to pulse in the
coherent image acquisition. Spoiler gradients are used to eliminate persistent trans-
verse magnetization from stimulated echoes for incoherent GE (see Chapter 12,
Section 12.5)
Acquisition times are calculated in the same way as spin echo; a GE sequence for
a 256 X 192 image matrix, two averages, and a TR = 30 ms, results in an imaging,
time equal to 192 X 2 X 0.03s = 15.5 A conventional spin echo requires 3.84
min for a TR = 600 ms. Trade-offs for fast acquisition speed include SNR losses,
magnetic susceptibility anifacts, and less immunity from magnetic field inhomo-
geneities. There are several acronyms for GE sequences, including GRASS, FISP,
Spoiled GRASS, FLASH, SSFP. etc., depending on the manufacturer of the equip-
ment. Table 13-1 describes a partial list of the different GE sequences and their
method of data acquisition.
Echo Planar Image (EPI) Acquisition is a technique that provides extremely
fast imaging time. Spin Echo (SE-EP!) and Gradient Echo (GE-EPI) are two methods
used for acquiring data, and a third is a hybrid of the two, GRASE (Gradient and Spin
Echo). Single-shot (all of the image information 1s acquired within 1 TR imerval) or
multishot EPI has been implemented with these methods. For single-shot SE-EPI,
en JES
IB FIGURE 13-5 Coherent GE pulse sequence uses a smal fip angle (30 to 40 degrees) RF pulse simultaneous
to the SSG. Phase and frequency encode gradients are applied shortly afterward (with a TE of less than 3 ms
in cert
sequences). A PEG “rewinder” (reverse polanty) reestablishes the phase conditions prior to the next
pulse, smmultaneous with the extended FEG durationChapter 13 * Magnetic Resonance imaging 455,
TABLE 13-1 COMPARISON OF MANUFACTURER-NAMED ACRONYMS FOR
GE SEQUENCES
SEQUENCE GENERAL ELECTRIC PHILIPS SIEMENS TOSHIBA
Coherent GE GRASS, FGRFMPGR FFE FISP. Field echo
Incoherent GE SPGR, FSPGR THFFE Field echo
(RF spoiled)
Incoherent GE ‘MPGR FLASH Field echo
(Gradient spoiled)
Steady-state free SSFP, DE FGR T2FFE PSI
precession
‘SSFP: balanced FIESTA Balanced FFE True FISP. True SSFP
sequence / true FiSP
Note: Not ail manufacturers ae listed in this table, ner are all GE sequences. (Blank areas indicate particular
‘sequence is not performed (at time of publication)
Image acquisition typically begins with a standard 90-degree flip, then a PEG/FEG
gradient application to initiate the acquisition of data in the periphery of the k-space,
followed by a 180-degree echo-producing RF pulse. Immediately after, an oscillating,
readout gradient and phase encode gradient “blips” are continuously applied to stimu-
late echo formation and rapidly fill k-space in a stepped “zig-zag” pattern (Fig, 13-6)
The “effective” echo time occurs at a time TE, when the maximum amplitude of
the induced GEs occurs. Acquisition of the data must proceed in a period less than
12* (around 50 ms), placing high demands on the sampling rate, the gradient coils
(shielded coils are required, with low induced “eddy currents"), the RF transmitter/
receiver, and RF energy deposition limitations, For GE-EPI, a similar acquisition strat-
egy is implemented but without a 180 degrees refocusing RF pulse, allowing for faster
acquisition time. SE-EPI is generally longer, but better image quality is achieved; on
the other hand, larger RF energy deposition to the patient occurs. EPI acquisition can
FIGURE 13-6 Single shot Echo
Planar Spin Echo image (SE-EP)
‘axcquisition sequence. Data is
deposited in k=pace, initially posi
tioned by a simultaneous PEG and
FEG application to locate the initial
row and column position (in this
‘example, the upper left), followed
by phase encode gradient “blips”
simultaneous to FEG oscillations,
to fill k-space line by line by intro-
‘ducing 1-row phase changes in
‘a ng-2ag pattern. Image matrix
sizes of 64 X 64 and 128 X 64
are common.Section Il * Diagnostic Radiology
be preceded with any type of RF pulse, for instance FLAIR (EPI-FLAIR), which will
prodiice images much faster than the corresponding conventional FLAIR sequence
The GRASE (Gradient and Spin Echo) sequence combines the initial spin echo
with a series of GEs, followed by an RF rephasing (180 degrees) pulse, and the pat-
tem is repeated until k-space is filled. This hybrid sequence achieves the benefits
of both types of rephasing: the speed of the gradient and the ability of the RF pulse
to compensate for T2* effects, providing significant improvements in image quality
compared to the standard EP! methods. A trade-off isa longer acquisition ume (¢.g ,
greater than 100 ms) and much greater energy deposition from the multiple 180
degrees RF pulses.
EPI acquisitions typically have poor SNR, low resolution (matrices of 6+ X 64 or
128 X 64 are typical), and many artifacts, particularly of chemical shift and magnetic
susceptibility origin, Nevertheless, EPI offers real-time “snapshot” image capability
with 50 ms total acquisition time. EPI is emerging as a clinical tool for studying time-
dependent physiologic processes and [unctional imaging, Concems of safety with
EPI, chiefly related to the rapid switching of gradients and possible nerve stimulation
of the patient, the associated acoustic noise, image artifacts, distortion, and chemical
shift are components that will limit use for many imaging procedures
Other K-Space Filling Methods
Methods to fill k-space in a nonsequential way can enhance signal, contrast, and
achieve rapid scan times as shown in Figure 13-7. Centric k-space filling has been
discussed with FSE imaging (above), where the lower strength phase encode gradi-
Outer rows tilled last
A Centric filing Central rows filed first
Outer rows filed last
Outer rows filed first
B Keyhole hiting Contra rows ies nth
Outer rows filed first
Equal AT between points
k-space re-binning of spiral
data is required before
image reconstruction
© Spiral fing
FIGURE 13-7 Alternate methods of filing k-space. A. Centnc filing applies the lower strength PEG's frst to
‘maximize signal and contast from the eartest echoes ofa FSE or GE sequence. B. Keyhole filing applies PEG's
‘of higher strength first to fil the outer portions of k-space, and the central lines ae filed only during a certain
part of the sequence, such as with arrnal of contrast signal. C Spiral data acquisition occurs with sinusovdal
ostilation of the X and ¥ gradients 90 degrees out of phase with each other. with samples beginning in the
center of kspace and spiraling out to the periphery. interpolating the data into the k,,k, matrix is required in
order to apply 2DFT image reconstructionChapter 13 ¢ Magnetic Resonance Imaging 457
ents are applied first filling the center of k-space when the echoes have their highest
amplitude. This type of filling is also important for fast GE techniques, where the
image contrast and the SNR fall quickly with time from the initial excitation pulse.
Keyhole filling methods fil k-space similarly to centric filling, except the central
lines are filled when important events occur during the sequence, in situations such.
as contrast-enhanced angiography. Outer areas of k-space are filled first, and when
gadolinium appears in the imaging volume, the center areas are filled. At the end of
the scan, the outer and central k-space regions are meshed to produce an image with
both good contrast and resolution
Spiral filling is an alternate method of filling k-space radially, which involves
the simultaneous oscillation of equivalent encoding gradients to sample data points
during echo formation in a spiral, starting at the ongin (the center of the k-space)
and spiraling outward to the periphery in the prescribed acquisition plane. The same
contrast mechanisms are available in spiral sequences (e.g., TI, T2, proton density
weighting). and spin or GEs can be obtained. After acquisition of the signals, an addi-
onal post-processing step, re-gridding, is necessary to convert the spiral data into
the rectilinear matrix for two-dimensional Fourier transform (2DFT), Spiral scanning
is an efficient method for acquiring data and sampling information in the center of
k-space, where the bulk of image information is contained
A variant of radial sampling with enhanced filling of the center of k-space is known
generically as “blade” imaging, and commonly as propeller: Periodically Rotated
Overlapping Parallel Lines with Enhanced Reconstruction, where a rectangular block
of data is acquired and then rotated about the center of k-space. Redundant informa-
tion concentrated in the center of k-space is used for improvement of SNR or for the
identification of times during the scan in which the patient may have moved, so that
those blocks of data can be processed with a phase-shifting algorithm to eliminate the
movernent effect on the data during the reconstruction process and to mitigate motion
artifacts toa great extent. Filling of k-space for this method is shown in Figure 13-8,
Parallel Imaging
Parallel imaging is a technique that fills k-space by using the response of multiple
receive RF coils that are coupled together with independent channels, so that data
can be acquired simultaneously Specific hardware and software are necessary for the
electronic orchestration of this capability. Typically, 2, 4, 5, 7, 8, 16, 18 (or more)
colls ate arranged around the area to be imaged; if a 4-coil configuration is used,
then during each TR period, each coil acquires a view of the data as the acquisition
sequence proceeds. Lines in k-space are defined only after the processing of linear
combinations of the signals that are received by all of the coils. Since 4 views of the
data are acquired per TR interval, scan time can be decreased by a factor of 4 (known
as the reduction factor). However, the acquisition of the signals have gaps, and the
FOV in the phase direction is reduced to one-quarter of its original size This results
in a known aliasing of the information (a wrapped image—see section on Artifacts)
that i rectified by using the measured sensitivity profile of each coil to calculate from
where the signal is coming, This sensitivity profile determines the position of the sig-
nal based on ts amplitude, where the signal near the coil has a higher amplitude than
that farthest away: As a result of the process, commonly known as SENSE (SENSiti
ity Hncoding—thete are several acronyms coined by the manufacturers), the image
can be unwrapped and combined with the unwrapped images from each of the othe,
coils. A simple two-coil example is shown in Figure 13-9 for a breast image applica-
tion of SENSE, in which improved resolution is desired over reduced sean tine458 Section II + Diagnostic Radiology
Rectangular filing Propeller filling
Motion during
the acquisition
FIGURE 13-8 The propeller data acquistion compared to a rectangular filing of k-space is shown above
Instead af acquiring single lines of information to fill kspace consecutively as shown in the upper left and
middle left, a rectangular data acquisition at a specific angle (e.g., 0 degree) acquired encompassing several
lines of k-space, which represents a “blade” of mformation. The partial acquisiton is rotated about the center
of k-space at angular increments, which provides 3 dense sampling of data at the center of k-space and less in
the periphery as shown by the schematic (upper right illustration). I the patient moves during a portion of the
‘examination (lower left image). the blades in which the motion occurred can be identified, reprocessed, and
the image reconstructed without the motion artifact lower right mage)
Single coil
FIGURE 13-9 Parallel imaging with two RE
ois. Top. A single coil acquisition of a breast
(MR exam over the full FOV. Middle. indwvidual
cols with every-other row of k-space being
filed represent Ys FOV, with image overlap
caused by aliasing. Bottom. After SENSE pro-
cessing, images are combined to deliver twice
the spatial resolution in the left/nght (Phase)
direction, with the same imaging time
Right coilChapter 13 * Magn
ic Resonance Imaging 459
Parallel imaging can be used to either reduce scan times or improve resolution. It
also can be used with most pulse sequences. There are obvious benefits in terms of
scan times and/or resolution, but there isa slight loss of SNR due to the manipulation
of the signals, and chemical shift artifacts (explained in the Anifacts section) may
increase. Patient motion can also cause misalignment between the undersampled
data and the reference scans of the coils.
Three-Dimensional Fourier Transform Image Acquisition
Three-dimensional image acquisition (volume imaging) requires the use of a broad-
band, nonselective, or “slab-selective” RF pulse to excite a large volume of pro-
tons simultaneously. Two phase gradients are discretely applied in the slice encode
and phase encode directions, prior to the frequency encode (readout) gradient
(Fig, 13-10). The image acquisition time is equal to
‘TR X # Phase Encode Steps (z-axis) X # Phase Encode Steps
(y-axis) X # Signal Averages
A three-dimensional Founer transform (three one-dimensional Fourier transforms)
is applied for each column, row, and depth axis in the image matrix “cube.” Volumes
obtained can be either isotropic, the same size in all three ditections, oF anisotropic,
where at least one dimension ts different in size. The advantage of the former is
equal resolution in all directions; reformations of images from the volume do not
suller from degradations of larger sample size from other directions. After the spatial
domain data are obtained, individual two-dimensional slices in any arbitrary plane
are extracted by interpolation of the cube data
When using a standard TR of 600 ms with one average for a T1-weighted exam, a
128 X 128 X 128 cube requires 163 min or about 2.7 h! Obviously, this is unaccept-
able for standard clinical imaging. GE pulse sequences with TR of 50 ms acquire the
same image volume in about 15 min. Another shortcut is with anisotropic voxels, where
the phase encode steps in one dimension are reduced, albeit with a loss of resolution
A major benefit to isotropic three-dimensional acquisition is the uniform resolution
Isotropic
Slice Encode
(phase #1)
e
‘pres #2)
Frequency Encode PM Enas®
FIGURE 13-10 Three-dimensional image acquisition requires the application of a broadband RF pulse to
excite all of the protons in the volume simultaneousy, followed by a phase encode gradient along the sice
encode direction, a phase encode gradient along the phase encode direction, and a frequency encode gra:
dient in the readout direction. Spatial location ss decoded sequentially by the Fourier transtorm along each
encode path, storing intermechate results inthe three-dimensional k-space matrix.460
Section Il * Diagnostic Radiology
in all directions when extractin,
: ig any two-dimensional image from the ma
tition, high SNRs achieved compared to asm twccimenstnal mage allowing
feconsirutin of very thin slices with good detail les paral volume averaging) and
ig A downside is the increased probability of motion artifacts and
computer hardware requirements for data andl, and storage nl
132| MR Image Characteristics
Spatial Resolution and Contrast Sensitivity
Spatial resolution, contrast sensitivity, and SNR parameters form the basis for evalu-
aling the MR image characteristics. The spatial resolution is dependent on the FOV,
which determines pixel size, the gradient field strength, which determines the FOV.
the receiver coil characteristics (head coil, body coil, and various surface coil designs),
the sampling bandwidth, and the image matrix. Common image matrix sizes are 128
128, 256 X 128, 256 X 192, and 256 x 256, with 512 x 256, 512 x 512, and
1,024 x 512 becoming prevalent. In general, MR provides spatial resolution approx-
imately equivalent to that of CT, with pixel dimensions on the order of 0.5 to 1.0
mm for a high-contrast object and a reasonably large FOV (greater than 250 mm)
‘4.250 mm FOV and a 256 X 256 matrix will have a pixel size on the order of 1
mm. In small FOV acquisitions with high gradient strengths and with surface coil
receivers, the effective pixel size can be smaller than 0.1 to 0.2 mm (of course, with
4 limited FOV of 25 to 50 mm. Slice thickness in MRI is usually 5 to 10 mm and
11 that produces the most partial volume averaging,
higher field strength magnets due toa larger
SNR. which allows thinner slice acquisition, and/or higher sampling rates (smaller pix-
cls) for a given acquisition, However, with higher B,. increased RF absorption, artifact
production, and a lengthening of TI relaxation occur. The later decreases TI contrast
Sensitivity because of increased saturation of the longitudinal magnetization
‘Contrast sensitivity is the major attribute of MR. The spectacular contrast sensitiv”
ity of MR enables the exquisite discrimination of soft issues and contrast due to blood
few This sensitivity is achieved through dillerences inthe T!, T2, proton density and
flow velocity characteristics. Contrast, which is dependent upon these parameter,
sueved through the proper application of pulse sequences, as discussed previously:
SiR contrast materials, usually suscepubihty agents that disrupt the local magnetic field
to enhance T2 decay or provide a relaxation mechanism [or shorter TI recovery time
(eg. bound water in hydration layers), are becoming import enhancement agents
fonnlifferentiation of normal and diseased tissues. The absolute contrat sensitivity of
the MR image is ultimately limited by the SNR and presence of image artifacts
represents the dimensiot
‘Spatial resolution can be improved with
Signal-to-Noise Ratio, SNR
ultimate SNR achievable by the MR system
numerous dependencies on the
Te bibede signal tmensity based on TH, T2, and proton density parameters has
been discussed; to summarize, the TR, TE, and flip angle will have an impact on the
magnitude of the signal generated in the image While there are many mitigaing
fears, a long TR increases the longitudinal magnelzst 0 recovery aa ten
the SNR. a Jong TE increases the transverse magne a0 decay and reduces th
SNR: a smaller flip angle (reduced from 90 degrees) reduces the SN .
‘ ith large flip angle, long TR, short TE, coarse mans
spin echo pulse sequences WChapter 13 * Magnetic Resonance Imaging 461
large FOV, thick slices, and many averages will generate the best SNR: however, the
resultant image may not be clinically relevant or desirable. While SNR is important,
its not everything
For a given pulse sequence (TR, TE, flip angle), the SNR of the MR image is
dependent on a number of variables, as shown in the equation below for a (wor
dimensional image acquisition:
ANEX
SNR 1>¢voxel,,. XSF X (QE) X (B)X face gap) x freconstruction)
where Is the intrinsic signal intensity based on pulse sequence, voxel... 1s the voxel
volume, determined by FOV, image matrix, and slice thickness, NEX is the number
of excitations, determined by the number (or fractional number) of repeated signal
acquisitions into the same voxels, BW is the frequency bandwidth of the RE receiver,
{QP is the function of the coil quality factor parameter (tuning the coil), (B) is the
function of magnetic field strength, B, {(slice gap) is the function of interslice gap
effects, and {(reconstruction) is the function of the reconstruction algorithm
Other factors in the above equation are explained briefly below:
Voxel Volume
The voxel volume is equal to
FOV, FOV,
— FOV, _,__FOM, _. stice thickness,
No. of pixels, x \ No.of pixels, y © t Uuckness. 2
Volume
SNRs linearly proportional to the voxel volume. Thus, by reducing the image matrix
size from 256 X 256 to 256 X 128 over the same FOV, the effective voxel size
increases by a factor of wo, and therefore increases the SNR by a factor of two for the
same image acquisition time (e.g., 256 phase encodes with one average versus 128
phase encodes with two averages).
Signal Averages
Signal averaging (also known as number of excitations, NEX) is achieved by averag-
ing sets of data acquired using an identical pulse sequence (same PEG strength). The
SNR is proportional to the square root of the number of signal averages. A 2-NEX
acquisition requires a doubling (100% increase) of the acquisition time for a 40%
imerease in the SNR (V2 =1.4) Doubling the SNR requires 4 NEX. In some cases,
less than | average (e.g,, Y4 or %4 NEX) can be selected. Here, the number of phase
encode steps is reduced by % or %, and the missing data are synthesized in the
k-space matrix. maging time is therefore reduced by a similar amount: however, a
loss of SNR accompanies the shorter imaging times by the same square root factor.
RF Bandwidth
The receiver bandwidth defines the range of frequencies to which the detector is
tuned during the application of the readout gradient. A narrow bandwidth (a nar-
row spread of frequencies around the center frequency) provides a higher SNR,
1
proportional to YBW . A twofold reduction in RF bandwidth—from 8 to 4 kHz,462
Section Il * Diagnostic Radiology
for instance—increases the SNR by 1.4 X (40% increase). This is mainly related
to the fact that the white noise, which is relatively constant across the bandwidth,
does not change, while the signal distribution changes with bandwidth. In the spa-
tial domain, bandwidth is inversely proportional to the sample dwell ume, AT to
sample the signal: BW = 1/AT. Therefore, a narrow bandwidth has a longer dwell
time, which incteases the signal height (Fig. 13-11), compared to the shorter dwell
lime for the broad bandwidth signal, thus spreading the signal over a larger range
of frequencies. The SNR is reduced by the square root of the dwell time. However,
any decrease in RF bandwidth must be coupled with a decrease in gradient strength
to maintain the sampling across the FOV, which might be unacceptable if chemt-
cal shift artifacis are of concern (see Anifacts, below). Narrower bandwidths also
require a longer time for sampling, and therefore affect the minimum TE time that
is possible for an imaging sequence. Clinical situations that can use narrow band-
widths are with T2-weighted images and long TEs that allow the echo to evolve over
an extended period, particularly in situations where fat saturation pulses are used to
reduce the effects of chemical shift in the acquired images. Use of broad bandwidth
settings is necessary when very short TEs are required, such as in fast GE imaging
10 reduce the sampling time
RF Coil Quality Factor
The coil quality factor is an indication of RF coil sensitivity to induced currents in
response to signals emanating from the patient. Coil losses that lead to lower SNR are
caused by patient “loading” effects and eddy currents, among other factor. Patient load-
ing refers o the electric impedance characteristics of the body, which to a certain extent
acts like an antenna, This effect causes a variation in the magnetic field that is different
‘Signal
broad (16 kHz) oe
“I \\
“aT \
(KHz) ——~Il\v -
sal |
narrow (4 kHz)
RF bandwidth =
1/ dwell time =
1/aT
‘Sample dwell time
MEFIGURE 13-11 RF Receiver Bandwidth is determined by the FEG strength, the FOV, and sampling rate. This
figure illustrates the spatial domam ew of SNR and corresponding sample dwell time. Evolution of the echo
in the broad bandwidth situation occurs rapidly with minimal dwell time, which might be needed in situations
where very short TE is required, even though the SNR is reduced. On the other hand, in T2 weighted images
requiring a long TE, narrow bandwidth can improve SNRChapter 13 ¢ Magnetic Resonance Imaging 463
for each patient, and must be measured and corrected for. Consequently, tuning the
receiver coil to the resonance frequency is mandatory before image acquisition. Eddy
currents ate signals that are opposite of the induced current produced by transverse
‘magnetization in the RF coil, and reduce the overall signal. Quadrature coils increase the
SNR as two coils are used in the reception of the signal; phased array coils increase the
SNR even more when the data from several coils are added together (see Paralle! Imag-
ing, Section 13.1). The proximity of the receiver coil to the volume of interest affects the
coll quality factor, but there are trade-offs with image uniformity, Positioning of the coil
with respect to the direction of the main magnetic field is also an issue that occurs with
air core (horizontal B.) to solid core (vertical B,) magnets. Body receiver coils positioned
in the bore of the magnet have a moderate quality factor, wheteas surface coils have a
high quality factor. With the body coil, the signal is relatively uniform across the FOV;
however, with surface coils, the signal falls off abruptly near the edges of the field, lmit-
ing the useful imaging depth and resulting in nonuniform brightness across the image
Magnetic Field Strength
Magnetic field strength influences the SNR of the image by a factor of BI" to B!*
Thus, one would expect a three- to fivefold improvement in SNR with a 1.5 T magnet
over 0.5 T magnet, Although the gains in the SNR are real, other considerations
mitigate the SNR improvement in the clinical environment, including longer TL
relaxation times and greater RF absorption, as discussed previously:
Cross-Excitation
Cross-excitation occurs from the nonrectangular RF excitation profiles in the spatial
domain and the resultant overlap of adjacent slices in muhtislice image acquisition
sequences. This saturates the protons and reduces contrast and the contrast-to-noise
ratio. To avoid cross-excitation, interslice gaps or interleaving procedures are neces-
sary (see Artifacts section, below)
Image Acquisition and Reconstruction Algorithms
Image acquisition and reconstruction algorithms have a profound effect on SNR. The
various acquisition/reconstruction methods that have been used in the past and those
used today are, in order of increasing SNR, point acquisition methods, line acquisition
methods, two-dimensional Fourier transform acquisition methods, and three-dimen-
sional Fourier transform volume acquisition methods. In each of these techniques,
the volume of tissue that is excited is the major contributing factor to improving the
SNR and image quality. Reconstruction filters and image processing algorithms will
also affect the SNR. High-pass filtration methods that increase edge definition will
‘generally decrease the SNR, while low-pass filtration methods that smooth the image
data will generally increase the SNR at the cost of reduced resolution
Summary, Image Quality
The best possible image quality is always desirable, but not always achievable because
of the trade-off between SNR, scan speed, and spatial resolution. To increase one of
these three components of image quality involves the consideration of reducing one
or both of the other two. Itis thus a balancing act that is chosen by the operator, the
protocol, and the patient in order to acquire images with the best diagnostic yield464 Section II + Diagnostic Radiology
MR parameters that may be changed include TR, TE, TI, ETL, Matrix Size, Slice
Thickness, Field of view, and NEX. Working with these parameters in the optimiza-
tion of acquisition protocols to achieve high image quality is essential
B Signal from Flow
The appearance of moving fluid (vascular and cerebrospinal fluid [CSF]) in MR images
is complicated by many factors, including flow velocity, vessel orientation, laminar
versus turbulent flow patterns, pulse sequences, and image acquisition modes. Flow-
related mechanisms combine with image acquisition parameters to alter contrast.
Signal due to flow covers the entire gray scale of MR signal intensities, from “black
blood’ to “bright blood’ levels, and flow can be a source of artifacts, The signal from
flow can also be exploited to produce MR angiographic images
Low signal intensities (low voids) are often a result of high-velocity signal loss
(HSL), in which protons in the flowing blood move out of the slice during echo ref-
ormation, causing a lower signal. Flow turbulence can also cause flow voids, by causing
a dephasing of protons in the blood with a resulting loss of the tissue magetization in
the area of turbulence. With HVSL, the amount of signal loss depends on the velocity
of the moving fluid. Pulse sequences to produce “black blood’ in images can be very
useful in cardiac and vascular imaging. A typical black blood pulse sequence uses a
“double inversion recovery” method, whereby a nonselective 180-degree RF pulse
1s initially applied, inverting all protons in the body, and is followed by a selective
180-degree RF pulse that restores the magnetization in the selected slice. During the
inversion time, blood outside of the excited slice with inverted protons flows into the
slice, producing no signal; therefore, the blood appears dark
Flow-Related Enhancement
Flow-related enhancement is a process that causes increased signal intensity due to
lowing protons; it occurs during imaging of a volume of tissues. Even-echo rephas-
ing is a phenomenon that causes flow to exhibit increased signal on even echoes in
a multiple-echo image acquisition. Flowing protons that experience two subsequent
180-degree pulses (even echoes) generate higher signal intensity due to a construc-
live rephasing of protons during echo formation. This effect is prominent in slow
Jaminar flow (e.g., veins show up as bright structures on even-echo images),
Flow enhancement in GE images is pronounced for both venous and arterial
structures, as well as CSF The high intensity is caused by the wash-in (between
subsequent RF excitations) of fully unsaturated protons into a volume of partially
saturated protons due to the short TR used with gradient imaging. During the next
excitation, the signal amplitude resulting from the moving unsaturated protons is,
about 10 times greater than that of the nonmoving saturated protons. With GE tech-
niques, the degree of enhancement depends on the velocity of the blood, the slice
or volume thickness, and the TR. As blood velocity increases, unsaturated blood
exhibits the greatest signal, Similarly, a thinner slice or decreased repetition time
results in higher flow enhancement. In arterial imaging of high-velocity flow, it is
possible to have bright blood throughout the imaging volume of a three-dimensional
acquisition if unsaturated blood can penetrate into the volume prior to experiencing
an RF pulse,
Signal from blood is dependent on the relative saturation of the surrounding
tissues and the incoming blood flow in the vasculature. In a multislice volume,Chapter 13 + Magnetic Resonance Imaging 465
FIGURE 13-12 The repeated RF excitation within
an imaging volume produces partial saturation of
the tissue magnetization (top figure, gray area)
Unsaturated protons flowing into the volume gener-
ate a large signal cifference that 's bright relate to
the surrounding tissues. Bright blood effects can be
reduced by applying pre-saturation RF pulses adjacent
to the imaging volume, so that protons in infiowing
blood will have a similar partial saturation (bottom
figure; note ne blood signal)
Flow-Related Enhancement
Pre-saturated spins: equal signal
Flow presaturation
repeated excitation of the tissues and blood causes a pantial saturation of the protons,
dependent on the T1 characteristics and the TR of the pulse sequence. Blood out-
side of the imaged volume does not interact with the RF excitations, and therefore
these unsaturated protons may enter the imaged volume and produce a large signal
compared to the blood within the volume, This is known as llow-related enhance-
ment. As the pulse sequence continues, the unsaturated blood becomes partially
saturated and the protons of the blood produce a similar signal to the tissues in the
inner slices of the volume (Fig. 13-12). In some situations, flow-related enhance
ment is undesirable and is eliminated with the use of “presaturation” pulses applied
to volumes just above and below the imaging volume. These same saturation pulses
are also helpful in reducing motion artifacts caused by adjacent tissues outside the
imaging volume
MR Angiography
Exploitation of blood flow enhancement isthe basis for MRA, Two techniques to create
images of vascular anatomy include time-of-flight and phase contrast angiography
Time-of-Flight Angiography
The time-of-flight technique relies on the tagging of blood in one region of the body
and detecting it in another. This differentiates moving blood from the surround station-
ary tissues. Tagging is accomplished by proton saturation, inversion, of relaxation to
change the longitudinal magnetization of moving blood. The penetration of the tagged
blood into a volume depends on the T1, velocity, and direction af the blood. Since the
detectable range is limited by the eventual saturation of the tagged blood, long vessels
are difficult to visualize simultaneously in a three-dimensional volume. For these rea-
sons, a two-dimensional stack of slices is typically acquired, where even slowly mov-
ing blood can penetrate the region of RF excitation in thin slices (Fig. 13-13). Each
slice is acquired separately, and blood moving in one direction (north or south, €..
aneries versus veins) can be selected by delivering a presaturation pulse on an adja-
cent slab superior or inferior to the slab of data acquisition. Thin slices are also help-
ful in preserving resolution of the flow pattern. Often used for the two-dimensional466 Section Il » Diagnostic Radiology
FIGURE 13-13 The time of fight MRA acquisition collects each slice separately with a sequence to enhance
biood flow. Explcitation of blood flow is achieved by detecting unsaturated protons moving into the volume,
producing a bright signal. A coherent GE image acquisition pulse sequence 1s shown, TR = 24 ms, TE = 3.1
ims, Flip Angle = 20 degrees. Every 10th image in the stack 1s displayed above, from left to right and top to
bottom,
image acquisition is a “GRASS” or “FISP” GE technique that produces relatively poor
anatomic contrast, yet provides a high-contrast “bright blood” signal. Magnetization
transfer contrast sequences (see below) are also employed to increase the contrast of
the signals due to blood by reducing the background anatomic contrast
Two-dimensional TOF MRA images are obtained by projecting the content of the
stack of slices at a specific angle through the volume. A maximum intensity projec-
tion (MIP) algorithm detects the largest signal along a given ray through the volume
and places this value in the image (Fig. 13-14). The superimposition of residual
MLFIGURE 13-14 simple illustration shows Projections are cast through the image stack (volume)
how the MIP algorithm extracts the highest ‘The maximum signal along each line is projected
(maximum) signals nthe two-dimensional
stack of images along a specific direction in
the volume, and produces projection images
with maximum intensity variations as a func MIP images
tion of angleChapter 13 * Magnetic Resonance Imaging 467
stationary anatomy often requires further data manipulation to suppress undesirable
signals. This is achieved in a vanety of ways, the simplest of which is setting a win-
dow threshold. Another method is to acquire a dataset without contrast, and subtract
the noncontrast MIP from the contrast MIP to reduce background signals. Clini-
cal MRA images show the three-dimensional characteristics of the vascular anatomy
from several angles around the volume stack (Fig. 13-15) with some residual signals
from the stationary anatomy Time-ol-flight angiography often produces variation in
vessel intensity dependent on orientation with respect to the image plane, a situation
that is less than optimal
Phase Contrast Angiography
Phase contrast imaging relies on the phase change that occurs in moving protons
such as bload. One method of inducing a phase change is dependent on the applica
tion of a bipolar gradient (one gradient with positive polarity followed by a second
gradient with negative polarity, separated by a delay time AT). In a second acquisi-
tion of the same view of the data (same PEG), the polarity of the bipolar gradients is
reversed, and moving protons are encoded with negative phase, while the stationary
protons exhibit no phase change (Fig. 13-16). Subtracting the second excitation from
the first cancels the magnetization due to stationary protons but enhances magne-
tization due to moving protons. Alternating the bipolar gradient polarity for each
subsequent excitation during the acquisition provides phase contrast image informa-
tion. The degree of phase shift is directly related to the velocity encoding (VENC)
ime, AT, between the positive and negative lobes of the bipolar gradients and the
velocity of the protons within the excited volume. Proper selection of the VENC time
2D Projection
‘Angiograms from MIP
FIGURE 13-15 A volume stack of bright blood images (left) is used with MIP processing to create a series
of projection angiograms at regular intervals, the three-dimensional perspective is appreciated in a stack wew,
with virtual rotation of the vasculature468 Section Il + Diagnostic Radiology
bipolar gradient encoding
T T
Excitation <1 Excitation | I-—1
| a #2 ‘
Excitation #1 - Excitation #2 = netresidual phase
- ( Vd secre
net phase shift = 0
Moving spins, low
velocity, forward direction:
|
le. 7 __talphaee siti amet
~ = Backward direction, high
7 | eccrmeemne
| ene: axpoene
IEFIGURE 13-16 Phase Contrast Angiography uses consecutive excitations that have a bipolar gradient
‘encoding withthe polarity reversed between the fist and second excitation, as shown in the top row. Mag
netization vectors (lower two rows) illustrate the effect of the bppolar gracients an stationary and moving
spins forthe frst and second excitations. Subtracting the two will cancel stationary tissue magnetization and
enhance phase differences caused by the velocity of moving blood.
is necessary to avoid phase wrap error (exceeding 180-degree phase change) and to
ensure an optimal phase shift range for the velocities encountered. Intensity varia-
tions are dependent on the amount of phase shift, where the brightest pixel values
represent the largest forward (or backward) velocity, a mid-scale value represents 0
velocity, and the darkest pixel values represent the largest backward (or forward)
velocity Figure 13-17 shows a representative magnitude and phase contrast image of
the cardiac vasculature. Unlike the time-of-flight methods, the phase contrast image
Magnitude Image Phase Image
EFIGURE 12-17 Magnitude (left) and phase (ight) images provide contrast of flowing blood. Magnitude
images are sensitive to flow, but not to direction, phase images provide cirection and veloaty information
‘The blood flow from the heart shows forward flow in the ascending aorta (dark atea) and forward flow in the
descending aorta at this point in the heart cycle forthe phase image. Some bright ow pattems inthe ascend
ing aorta represent backward flow to the coronary arteries. Grayscale amplitude is propartional to velocity,
where intermediate grayscale s 0 velocity.Chapter 13 * Magnetic Resonance Imaging 469.
is inherently quantitative, and when calibrated carefully, provides an estimate of the
mean blood flow velocity and direction, Two- and three-dimensional phase contrast
image acquisitions for MRA are possible
Gradient Moment Nulling
In spin echo and gradient recalled echo imaging, the slice select and readout gradients are
balanced, so that the uniform dephasing with the initial gradient application is rephased
by an opposite polarity gradient of equal area, However, when moving protons are sub-
jected to the gradients, the amount of phase dispersion is not compensated (Fig, 13-18).
This phase dispersal can cause ghosting (laint, displaced copies of the anatomy) in images.
Its possible, however, to rephase the protons by a gradient moment nulling technique.
With constant velocity flow (first-order motion), all protons can be rephased using the
application of a gradient tnplet. In this technique, an initial positive gradient of unit area
is followed by a negative gradient of twice the area, which creates phase changes that are
compensated by a third positive gradient of unit area. The velocity compensated gradi-
cent (right graph in Fig. 13-18) depicts the evolution of the proton phase back to zero
for both stationary and moving protons. Note that the overall applied gradient has a net
area of zero—equal to the sum of the positive and negative areas. Higher order cortec-
tions such as second- of third-order moment nulling to correct for acceleration and other
‘motions are possible, but these techniques require more complicated gradient switching,
Gradient moment nulling can be applied to both the slice select and readout gradients to
correct for problems such as motion ghosting as elicited by CSF pulsatile flow:
-— Perfusion and Diffusion Contrast Imaging
Perfusion of tissues via the capillary bed permits the delivery of oxygen and nutrients
to the cells and removal of waste (¢.g,, CO,) from the cells. Conventional perfusion
measurements are based on the uptake and wash-out of radioactive tracers or other
‘exogenous tracers that can be quantified from image sequences using well-characterized
imaging equipment and calibration procedures. For MR perfusion images, exogenous
RF pulse N\
ff} Stationary
Poa ¢ ! Stationary
“a andmaving
spins spins
FIGURE 13-18 Left. Phase dispersion of stationary and moving spins under the influence of an applied
gradient (no flow compensation) as the gradient is inverted is shown. The stationary spins return to the onginal
phase state, whereas the moving spins do not. Right. Gradient moment nulling of first order linear velocity
(flow compensation) requires a doubling of the negative gradient amplitude followed by a positive gradient
such that the total summed area is equal to zero. This will retum both the stabonary spins and the moving
spins back to their original phase state.and endogenous tracer methods are used. Freely diffusible tracers using nuclei such
as ‘H (deuterium), ‘He, "O, and “F are employed in experimental procedures to
produce differential contrast in the tissues, More clinically relevant are intravascular
blood pool agents such as gadolinium-diethylenetriaminepentaacetic acid, which
‘modify the relaxation of protons in the blood in addition to producing a shorter T2*
This produces signal changes that can be visualized in pre and post contrast images
(Fig. 13-19). Endogenous tracer methods do not use externally added agents, but
instead depend on the ability to generate contrast from specific excitation or diffusion
mechanisms. For example, labeling of inflowing protons (black blood” perfusion)
uses protons in the blood as the contrast agent. Tagged (labeled) protons outside of
the imaging volume perfuse into the tissues, resulting in a drop in signal intensity, a
time course of events that can be monitored by quantitative measurements
Functional MRI (fMRI) is based on the increase in blood flow to the local vascu-
lature that accompanies neural activity in the specific areas of the brain, resulting in a
local reduction of deoxyhemoglobin because the increase in blood flow occurs without
an increase in oxygen extraction, As deoxyhemoglobin isa paramagnetic agent, it alters
the T2*-weighted MRI image signal. Thus, this endogenous contrast enhancing agent
serves as the signal for [MRI Area voxels (represented by x-y coordinates and 2 slice
thickness) of high metabolic activity resulting from a task-induced stimulus produce
correlated signal for Blood Oxygen Level Dependent (BOLD) acquisition techniques. A.
BOLD sequence produces multiple T2*-weighted images of the head before the appli-
cation of the stimulus. The patient is repeatedly subjected to the stimulus and multiple
BOLD images are acquired. Because the BOLD sequence produces images that are
highly dependent on blood oxygen levels, areas of high metabolic activity will dem-
onstrate a change in signal when the prestimulus image data set is subtracted, voxel
by voxel, from post-stimulus image data set. Voxel locations defined by significant
rE Pre {top row) and post (bottom row) gadolinium contrast T!-weighted MR axial images
of the brain ilustrate the signal change that occurs with the appearance of gadolinium by shortening the T1
time of the perfused tssuesChapter 13 Magnetic Resonance Imaging 471
signal changes indicate regions of the brain activated by a specific task. Stimuli in (MRI
experiments can be physical (finger movement), sensory (light flashes of sounds), or
cognitive (repetition of “good” or “bad” word sequences, complex problem solving)
among others. To improve the SNR in the {MRI images, a stimulus is typically applied
ina repetitive, periodic sequence, and BOLD images are acquired continuously, tagged
with the timing of the stimulus. Regions in the brain that demonstrate time-dependent
activity and correlate with the time-dependent application of the stimulus are sta-
Uistically analyzed, and coded using a color scale, while voxels that do not show a
significant intensity change are not colored. The resultant color map is overlaid onto a
grayscale image of the brain for anatomic reference, as shown in Figure 13-20
High-speed imaging and 12* weighting necessary for {MRI is typically achieved
with EPL acquisition techniques that can be acquired in as little as 50 ms for a 64
64 acquisition matrix, Gradient Recalled Echo acquisitions using standard sequences
are also employed with multiple contiguous slices (e.g., 16 slices, slice thickness 5
mm, TR = 3s, TE = 60 ms, 90-degree flip angle) at 1.5 T, with 25 to 30 complete
head acquisitions. The latter acquisition techniques provide better spatial resolution
but rely on very cooperative subjects and a much longer exam time
Diffusion Weighted Ima
ig
Diffusion relates to the random motion of water molecules in tissues. Interaction of
the local cellular structure with the movement of water molecules produces aniso-
tropic, directionally dependent diffusion (e.g,, in the white matter of brain tissues)
Bilateral finger tapping paradigm 1-0.728
11
ovo}
FIGURE 13-20 Functional MR image bilateral finger tapping paradigm shows the areas of the brain acti-
vated by this repeated activity. The paradigm was a right finger tap alternated by a left finger tap (time
sequence on the right side of the figure) and the correlated BOLD signals (black traces) derived from the echo
planar image sequence. A voxel-by-voxel correlation of the periodic stimulus and MR signal is performed, and
When exceeding a correlation threshold, a color overlay is added to the grayscale image. In this example, red
inccates the right finger tap thatexctes the left motar cortex, which appears onthe nght seo the image
and blue the left finger tap. (Figure compliments of MH Buonacore, MD, PhD University of California Davis472 Section Il + Diagnostic Radiology
FIGURE 13-21 The basic elements of @ DWI
pulse sequence are shown. The diffusion weight- re—h h
ing gradients are of amplitude G, duration of the
gradients & and time between gradients is A $86 7 *—_———_
pw Ls !
gradients i =
a
PEG
eG —— $$ r
S.qnai —>
Diffusion sequences use strong MR gradients applied symmetrically about the relo-
cusing pulse to produce signal differences based on the mobility and directionality of
water diffusion, as shown in Figure 13-21. Tissues with more water mobility (normal)
have a greater signal loss than those of lesser water mobility (injury) under the influ-
ence of the diffusion weighted imaging (DWI) gradients.
The in vivo structural integrity of certain tissues (healthy, diseased, or injured) can
be measured by the use of DWI, in which water dilfusion characteristics are deter-
mined through the generation of apparent diffusion coefficient (ADC) maps. This,
equires two or more acquisitions with different DWI parameters. A low ADC cor
responds to high signal intensity on a calculated image, which represents restricted
diffusion (Fig, 13-22). ADC maps of the brain and the spinal cord have shown prom-
1 FIGURE 13-22 Left Diffusion weighted image. Right Calculated ADC image. shownng an area of increased
brightness related to restricted mobulty of water molecules.Chapter 13 * Magnetic Resonance Imaging 473
ise in predicting and evaluating pathophysiology before it is visible on conventional
T1- or T2-weighted images. DWI is also a sensitive indicator for early detection
of ischemic injury Areas of acute stroke show a drastic reduction in the diffusion
coefficient compared with nonischemic tissues. Diagnosis of multiple sclerosis is a
potential application, based on the measurements of the diffusion coefficients in three-
dimensional space
Various acquisition techniques are used to generate diffuston-weighted contrast
Standard spin echo and EPI pulse sequences with applied diffusion gradients of high
strength are used. Challenges for DWI are the extreme sensitivity to motion of the
head and brain, which is chiefly caused by the large pulsed gradients required for the
diffusion preparation. Eddy currents are also an issue, which reduce the effectiveness
of the gradient fields, so compensated gradient cails are necessary, Several strategies
hhave been devised to overcome the motion sensitivity problem, including common
electrocardiographic gating and motion compensation methods
134] Magnetization Transfer Contrast
Magnetization transfer contrast is the result of selective observation of the interaction,
between protons in free water molecules and protons contained in the macromolecules
of a protein, Magnetization exchange occurs between the two proton groups because
of coupling or chemical exchange, Because the protons exist in slightly different mag-
netic environments, the selective saturation of the protons in the macromolecule can
be excited separately from the bulk water by using narrow-band RF pulses (because the
Larmor frequencies are different). A transfer of the magnetization from the protons in
the macromolecule partially saturates the protons in bulk water, even though these pro-
tons have not experienced an RF exeitation pulse (Fig, 13-23). Reduced signal from the
Hydration Layer
ny
Hy e
fon ng
Mn
RaPay Se
Macromolecule He an Bulk Water
Sone
Sen? ye
Tags" “8 NS ne
Oe
Meta “ote
Putonchoma | )
Rr Excitaon: _ 4 Protons of
(off-resonance pulse) Ve bulk water
~1.5 kHz shift Frequency Spectrum
IEFIGURE 13-23 Magnetization transfer contrast is implemented with an off-resonance RF pulse of about
1,500 He from the Larmor frequency. Excitation of hydrogen atoms on macromolecules is transferred via the
hydration layer to agjacent "free-water” hydrogen atoms. A partial saturation of the tissues reduces the sig-
nals that would otherwise compete with signals from blood flow, making this useful for time-of-flight MRA,47a
Section | * Diagnostic Radiology
adjacent free water protons by the saturation “label” affects only those protons having a
‘chemical exchange with the macromolecules and improves local image contrastin many
situations by decreasing the otherwise large signal generated by the protons in the bulk
water. This technique is used for anatomic MRI of the heart, the eye, multiple sclerosis,
knee cartilage, and general MRA. Tissue characterization is also possible, because the
image contrast in partis caused by the surface chemistry of the macromolecule and the
tissue-specific factors that affect the magnetization transfer characteristics.
Magnetization transfer contrast pulse sequences are often used in conjunction
with MRA time-of-flight methods. Hydrogen atoms constitute a large fraction of mac-
romolecules in proteins, ate tightly bound to these macromolecules, and have a very
short T2 decay with a broad range of resonance frequencies compared to protons in
fice water. Selective excitation of these protons is achieved with an off-resonance RF
pulse of approximately 1,500 Hz from the Larmor frequency, causing their satura-
tion. The protons in the hydration layer bound to these molecules are affected by
the magnetization and become partially saturated themselves. MR signals from these
tissues are suppressed, with an impact of reducing the contrast variation of the anat-
omy As a result, the differential contrast of the flow-enhanced signals is increased.
with overall better image quality angiographic sequence
MR Artifacts
Artifacts manifest as positive or negative signal intensities that do not accurately rep-
resent the imaged anatomy. Alhough some artifacts are relatively insignificant and
are easily identified, others can limit the diagnostic potential of the exam by obscur-
ing or mimicking pathologic processes or anatomy: One must realize the impact
of MR acquisition protocols and understand the etiology of artifact production 10
exploit the information they convey.
‘To minimize the impact of MR artifacts, a working knowledge of MR physics as
well as image acquisition techniques is requited. On the one hand. there are many
variables and options available that complicate the decision-making process for MR
image acquisition, On the other, the wealth of choices enhances the goal of achieving
diagnostically accurate images. MR artifacts are classified into three broad areas—
those based on the machine, on the patient, and on signal processing,
Machine-Dependent Artifacts
Magnetic field inhomogeneities are either global or focal field perturbations that lead
to the mismapping of tissues within the image, and cause more rapid T2 relaxation
Distortion or misplacement of anatomy occurs when the magnetic field 1s not com-
pletely homogeneous, Proper site planning, sell-shielded magnets, automatic shim-
ming, and preventive maintenance procedures help to reduce inhomogeneities
Focal field inhomogeneities arise from many causes. Ferromagnetic objects in oF
oon the patient (e.g, makeup, metallic implants, prostheses, surgical clips. dentures)
produce field distortions and cause protons to precess at frequencies different from
the Larmor frequency in the local area. Incorrect proton mapping, displacement, and
appearance as a signal void with a peripherally enhanced rim of increased signal ate
Common findings Geometric distortion of surrounding tissue ts also usually evi-
dent, Even nonferromagnetic conducting materials (e.g.. aluminum) produce field
distortions that distur the local magnetic environment, Partial compensation by
the spin echo (180-degree RF) pulse sequence reduces these artifacts; on the otherChapter 13 * Magnetic Resonance Imaging 475,
hand, the gradient-refocused echo sequence accentuates distortions, since the protons
always experience the same direction of the focal magnetic inhomogencities within
the patient
Susceptibility Artifacts
Magnetic suscepubility is the ratio of the induced internal magnetization in a tissue
to the external magnetic field. As long as the magnetic suscepuilty of the tissues
being imaged is relatively unchanged across the ficld of view, then the magnetic field
will remain uniform. Any drastic changes in the magnetic susceptibility will distort
the magnetic field, The most common susceptibility changes occur at tissue-air anter=
faces (e.g, lungs and sinuses), which cause a signal loss due to more rapid dephasing
(12+) at the tissue-air interface (Fig, 13-24). Any metal (ferrous or not) may have
a significant elfect on the adjacent local ussues due to changes in suscepubility and
the resultant magnetic field distortions. Paramagnetic agents exhibit a weak magne-
tization and increase the local magnetic field causing an artifactual reduction in the
surrounding T2* relaxation
Magnetic susceptibility can be quite helpful in some diagnoses. Most notable is the
ability to diagnose the age of a hemorrhage based on the signal characteristics of the
blood degradation products, which are different in the acute, subacute, and chronic
phases. Some of the iron-containing compounds (deoxyhemoglobin, methemoglobin,
hemosiderin, and ferritin) can dramatically shorten T1 and T2 relaxation of nearby pro-
tons. The amount of associated free water, the type and structure of the iron-containing
molecules, the distribution (intracellular versus extracellular), and the magnetic field
strength all influence the degree of relaxation effect that may be seen. For example,
in the acute stage, T2 shortening occurs due to the paramagnetic susceptibility of the
organized deoxyhemoglobin in the local area, without any large effect on the T1 relax-
ation time. When red blood cells lyse during the subacute stage, the hemoglobin
altered into methemoglobin, and spin-lattice relaxation is enhanced with the formation
of a hydration layer, which shortens TI relaxation, leading to a much stronger signal
BE FIGURE 13-24 Suscepnbaity artifacts due to dental filings are shown in the same axial image sce. Left.
Axial T2-weighted fast spin echo image illustrates significant suppression of susceptibility artifacts with
180-degree refocusing pulse. Right. Axial T2*-weighted gracient echo image ilstrates significant image vord
exacerbated by the gradient echo, where external inhomogeneities are not canceled inthe reformed echo476
Section Il + Diagnostic Radiology
on Tl-weighted images. Increased signal intensity on T1-weighted images not found
in the acute stage of hemorrhage identifies the subacute stage. In the chronic stage,
hemosiderin, found in the phagocytic cells in sites of previous hemorrhage, disrupts
the local magnetic homogeneity, causes loss of signal intensity, and leads to signal void,
producing a characteristic dark rim around the hemorthage site
Gadolinium-based contrast agents (paramagnetic characteristics shorten T2 and
hydration layer interactions shorten TL) are widely used in MRI. Tissues that uptake
gadolinium contrast agents exhibit shortened T1 relaxation and demonstrate increased
signal on T1-weighted images. Although focal inhomogeneities are generally consid-
eted problematic, there are certain physiologic and anatomic manifestations that can
be idemtfied and diagnostic information obtained
Gradient Field Artifacts
Magnetic field gradients spatially encode the location of the signals emanating from
excited protons within the volume being imaged. Proper reconstruction requires lin-
car, matched, and properly sequenced gradients. The slice select gradient defines the
volume (slice), Phase and frequency encoding gradients provide the spatial informa-
tion in the other wo dimensions
ince the reconstruction algorithm assumes ideal, linear gradients, any deviation
or temporal instability will be represented as a distortion. Gradient strength has a
tendency to fall off at the periphery of the FOV. Consequently, anatomic compres-
sion occurs, especially pronounced on coronal and sagittal images having a large
FOV, typically greater than 35 cm (Fig, 13-25), Minimizing the spatial distortion
entails either reducing the FOV by lowering the gradient field strength or by hold-
ing the gradient field strength and number of samples constant while decreasing the
frequency bandwidth. Of course, gradient calibration is part of a continuous quality
control (QC) checklist, and geometric accuracy must be periodically verified.
Anatomie proportions may simulate abnormalities, so venfication of pixel dimen-
sions in the PEG and FEG directions are necessary. I the strength of the FEG and the
strength ofthe largest PEG are different, the height or width of the pixels can become dis-
torted and produce inaccurate measurements, Ideally, the phase and frequency encod
ing gradients should be assigned to the smaller and larger dimensions of the object
respectively, to preserve spatial resolution while limiting the number of phase encoding
steps. In practice, this is not always possible, because motion artifacts or high-intensity
FIGURE 13-25 Gradient nonlinearity causes image distortions by mis-mapping anatomy. In the above
‘examples, the strength of the gradient at the periphery is iess than the iceal (orange line versus black line). This
results in a compression of the imaged anatomy, with inaccurate geometry (images with orange border). For
‘comparison, images acquired with linear corrections are shown above.Chapter 13 * Magnetic Resonance Imaging 477
signals that need to be displaced away from imporuant ateas of interest after an initial
scan might require swapping the frequency and phase encode gradient directions
RF Coil Artifacts
RF surface coils produce variations in uniformity across the image caused by RF
excitation variability, attenuation, mismatching, and sensitivity falloff with distance
Proximal to the surface coil, receive signals are intense, and with distance, signal
Intensity 1s attenuated, resulting in grayscale shading and loss of brightness in the
image, Nonuniform image intensities are the all-too-frequent result. Also, compensa-
tion for the disturbance of the magnetic field by the patient is typically compensated
by an automanie shimming calibration. When this is not performed, or performed
snadequately, a significant negative impact on image quality occurs. Examples of vari-
able response are shown in Figure 13-26.
Other common artifacts from RF coils occur with RF quadrature coils (coils that
simultaneously measure the signal from orthogonal views) that have two separate ampli-
fer and gain controls. If the amplifiers are imbalanced, a bright spot in the center of the
image. known as a center point artifact, arises as a “O frequency” direct current offset
Vanations in gain between the quadrature coils can cause ghosting of objects diagonally
in the image. The bottom line for all RF coils is the need for continuous measurement
and consistent calibration of their response, so that artifacts are minimized.
RF Artifacts
RF pulses and precessional frequencies of MRI instruments occupy the same fre-
quencies of common RF sources, such as TV and radio broadcasts, electric motors,
fluorescent lights, and computers, Stray RF signals that propagate to the MRI antenna
can produce various artifacts in the image. Narrow-band noise creates noise pat-
tems perpendicular to the frequency encoding direction. The exact position and
spatial extent depends on the resonant frequency of the imager, applied gradient
field strength, and bandwidth of the noise. A narrow band pattern of black/white
~
Coil close to skin Inadequate shimming for fat saturation
recewve cols are 100 close to the skin, as exern-
FIGURE 13-26 Signal intensity vanations occur when surface RF recewe coi i
plified by the MR breast imace on the left. With inadequate shimming calibration, saturation pulses for adipose
tissue in the breast 1s uneven, causing a significant variation in the unvformity of the reconstructed image. From
Hendrick RE, Breast MRI. fundementals and technical aspects. New York, NY: Springer, 2007. By permission478
Section Il * Diagnostic Radiology
alternating noise produces a “zipper” artifact, Broadband RF noise disrupts the image
over a much larger area of the reconstructed image with diffuse, contrast-reducing
“herringbone” artifacts. Appropriate site planning and the use of properly installed RF
shielding materials (e.g., a Faraday cage) reduce stray RF interfetence to an accept-
ably low level. An example RF zipper artifact is shown in Figure 13-44
RF energy received by adjacent slices during a multislice acquisition excite and
saturate protons in adjacent slices, chiefly due to RF pulses without sharp off/on/
off transitions. This is known as cross-excitation. On T2-weighted images, the slice-
to-slice interference degrades the SNR, on T1-weighted images, the extra partial
saturation reduces image contrast by reducing longitudinal recovery during the TR
interval, A typical truncated “sinc” RF profile and overlap areas in adjacent slices are
shown in Figure 13-27. Interslice gaps reduce the overlap of the profile tails, and
pseudo-rectangular RF pulse profiles reduce the spatial extent of the tails. Important
anatomic findings could exist within the gaps, so slice interleaving is a technique to
mitigate cross-excitation by reordering slices into two groups with gaps. During the
first half of the TR, the first slices are acquited (slices 1 to 5), followed by the second
group of slices that are positioned in the gap of the first group (slices 6 to 10). This
method reduces cross-excitation by separating the adjacent slice excitations in time
The most effective method is to acquire two independent sets of gapped multislice
images, but the image time is doubled. The most appropriate solution is to devise RF
pulses that approximate a rectangular profile; however, the additional time necessary
for producing such an RF pulse can be prohibitive
K-Space Errors
Errors in k-space encoding affect all areas of the reconstructed image, and cause
the artifactual superimposition of wave patterns across the FOV. Each individual
pixel value in k-space contributes to all pixel values in image space as a frequency
harmonic with a signal amplitude. One bad pixel introduces a significant artifact,
rendering the image suboptimal, as shown in Figure 13-28.
FIGURE 13-27 Top. Poor pulse pro-
files are caused by truncated RF pulses,
and resulting profile overlap causes
‘unwanted partial saturation in adjacent
slices. with a loss of SNR and CNR. Opti-
mized pulses are produced by consider-
ing the tradeotf of pulse duration versus
excitation profile. Bottom. Reduction of
cross-excitation is achieved with inter-
slice gaps, but anatomy at the cap loca-
tion might be missed. An interleaving
technique acquires the first half of the
images with an intersice gap, and the
second half of the images are positioned
in the gaps of the fist images. The sep-
aration in time reduces the amount of
contrast reducing saturation of the adja-
cent slices
“sinc
profile
overlapped
Interslice gap
Interleaving
“computer-optimized” “rectangular”
profile profile
a
Nig Agno, is J ieChapter 13 * Magnetic Resonance Imaging 479
Bad pixel in k-space Resultant image
IB FIGURE 13-28 A single bad pixel in k-space causes a significant artifact in the reconstructed image. The
bad pixel is located at k, = 2, k, = 3, which produces a superimposed sinusoidal wave on the spatial domain
image as shown above
Motion Artifacts
The most ubiquitous and noticeable artifacts in MRI arise with patient motion, including
voluntary and involuntary movement, and flow (blood, CSF). Although motion artifacts
are not unique to MRI, the long acquisition time of certain MRI sequences increases the
probability of motion blurring and contrast resolution losses. Motion anifacts occur
mostly along the phase encode direction, as adjacent phase encoding measurements in
k-space are separated by a TR interval that can last 3,000 ms or longer. Even very slight
motion can cause a change in the recorded phase variation across the FOV throughout
the MR acquisition sequence. Examples of motion artifacts are shown in Figure 13-29
The frequency encode direction is less aflected, especially by peiodic motion, since the
evolution of the echo signal, frequency encoding, and sampling occur simultaneously
over sevetal milliseconds, Ghost images, which are faint copics of the image displaced
along the phase encode direction, are the visual result of patient motion
Several techniques can compensate for motion-related artifacts, The simplest
technique transposes the PEG and FEG to relocate the motion artifacts out of the
ins, are most always displayed in the phase encode
le ghost images of the anatomy, since the variation
im FIGURE 13-29 Motion arifacts, particularly of flow patt
‘gradient direction. Slight changes in phase produce multi
in phase caused by motion can be substantial between excitations480 Section Il * Diagnostic Radiology
region of diagnostic interest with the same puilse sequences. This does not reduce the
magnitude of the artifacts, however, and often there is a mismatch when placing the
PEG along the long axis of a rectangular FOV (c.g. an exam of the thoracic spine)
in terms of longer examination times of a significant loss of spatial resolution or of
SNR.
There are other motion compensation methods:
1, Cardiac and respiratory gating—signal acquisition at a particular cyclic location
synchronizes the phase changes applied across the anatomy (Fig. 13-30)
2. Respiratory ordering of the phase encoding projections based on location within the
respiratory cycle, Mechanical or video devices provide signals to monitor the cycle
3, Signal averaging to reduce artifacts of random motion by making displaced sig-
nals less conspicuous relative to stationary anatomy.
4. Short TE spin echo sequences (limited to proton density, Tl weighted scans, frac-
tional echo acquisition, Fig. 13-3). Note: Long TE scans (T2 weighting) are more
susceptible to motion.
5. Gradient moment nulling (additional gradient pulses for flow compensation) to
help rephase protons that are dephased due to motion, Most often, these techniques
require a longer TE and are more useful for T2-weighted scans (Fig, 13-18).
6. Presaturation pulses applied outside the imaging region to reduce flow artifacts
{rom inflowing protons, as well as other patient motions that occur in the periph-
ery (Fig. 13-12)
7. Muluple redundant sampling in the center of k-space (¢-g., propeller) to identify
and remove those sequences contnbuting to motion, without deletenously alfect-
ing the image (Fig 13-8)
Chemical Shift Artifacts of the First Kind
There are two types of chemical shift artifacts that affect the display of anatomy due
to the precessional frequency differences of protons in fat versus protons in water.
Chemical shift refers to the resonance frequency variations resulting {rom intrinsic
Non Gated
ECG Signal |
Velocity i 13 i
i
[eis Nias ANETTISZ
ECG Gated
cesta | |
Velocity i 4
on | Like Le, LA
EFIGURE 13-30 Motion artifacts occur when data is acquired without consideration of physiologic periodic-
ity. Top. The electrocardiogram measures the R-wave at each heartbeat, but data acquisition proceeds in a
linear fashion without regard to reproducibility. The result is a set of images degraded with motion artifact,
with diagnostic usefulness marginal, at best. Bottom. Acquis'ton of images proceeds with the detection of
the R-wave signal and synchronization of the collection of image data in a stepwise fashion over the period
between Rewaves. A reduced number of mages or extended accuisition time ts required to collet the data,Chapter 13 * Magnetic Resonance Imaging 481
magnetic shielding of anatomic structures. Molecular structure and electron orbital
characteristics produce fields that shield the main magnetic field and give rise to dis-
tinct peaks in the MR spectrum. In the case of proton spectra, peaks correspond to
water and fat, and in the case of breast imaging, silicone material is another material
to consider. Lower frequencies of about 3.5 parts per million for protons in fat and
5.0 parts per million for protons in silicone occur, compared to the resonance frequency
of protons in water (Fig, 13-31). Since resonance frequency increases linearly with field
strength, the absolute difference between the fat and water resonance also increases,
making high field strength magnets more susceptible to chemical shift artifact.
Data acquisition methods cannot directly discriminate a frequency shift due to
the application of a frequency encode gradient or a chemical shift artifact, Water and
{at differences therefore cannot be distinguished by the frequency difference induced
by the gradient. The protons in fat resonate at a slightly lower Irequency than the
corresponding protons in water, and cause a shift in the anatomy (misregtstration of
water and fat moieties) along the frequency encode gradient direction,
‘A sample calctilation in the example below demonstrates frequency variations in
fat and water for two different magnetic field and gradient field strengths.
‘Chemical shift artifact numerical calculation for field strength, with a 3.5-ppm
(3.5 X 10-®) variation in resonance frequency between fat and water results in the
following frequency differences:
L5 T: 63.8 X 10°Hz X 3.5 x 10°
30 T: 127.7 X 10°Hz x 3.5 X 10° = 447 He
223 Hz
‘Thus, the chemical shift is more severe for higher field strength magnets
Chemical shift artifact numerical calculation for gradient strength results in the
following numerical calculations for a 25-cm (0.25 m) FOV, 256 X 256 matrix:
Low gradient strength: 2.5 mT/m % 0.25 m = 0.000625 T variation, gives frequency
range of 0.000625 T X 42.58 MH2/T = 26.6 kHz actoss FOV and 26.6 kH2/256
pixels = 104 Hz/pixel
High gradient strength: 10 m
ange of 0.0025 T X 42.58 M
pixels = 416 Hz/pixel
\T/m X 0.25 m = 0.0025 T variation, gives frequency
H2/T = 106.5 kHz actoss FOV and 106.5 kH2/256
‘Thus, a chemical shift occurrence is more severe for lower gradient strengths, since
displacement will occur over a large number of pixels. With a higher gradient
Strength, water and fat are more closely contained within the broader pixel boundary
bandwidths. Normal and low bandwidth images are illustrated in Figure 13-32
MEFIGURE 13-31 Chemical shift refers to
ssa .
ss ; water Chemical sen the slightly different precessional frequen-
- ifferent materials oF tis-
tat Ges of protons in i
NRT 24 pm renee ess po ae eters 0
-_ " water for fat and silicone. Fat chernical shift
‘ artifacts are represented by a shitt of water:
‘and fat in the images of anatomical struc
+ * ‘ture, mainly in the frequency encode gract
ent direction Swapping the PEG and the
FEG wall cause a shift ofthe fat and water
= a ‘components ofthe tissues in the image.
—_ —
Fi
PEG «6Section I « Diagnostic Radiology
High Bandwidth
FIGURE 13-32 MR images of the breast, containing glandular and adipose tissue, are acquired under a high
bandwidth (32 kHz) and a low bandwidth (4 kH2),dlustrating the more severe chemical shift with low readout
gradient steength and bandwidth (Reprinted by permission, Hendrick RE. Breast MRI: fundamentals and tech-
ical aspects. New York, NY: Springer, 2007.)
RF bandwidth and gradient strength considerations can mitigate chemical shift ari-
facts. While higher gradient strength can confine the chemical shift of fat within the pixel
bandwidth boundaries, a significant SNR penalty occurs with the broad RF bandwidth
required to achieve a given slice thickness. A more widely used method is to use lower
gradient strengths and narrow bandwidths in combination with off-resonance “chemi-
cal presaturation” RF pulses to minimize the fat (or the silicone) signal in the image (Fig
13-33). Another alternative is to use STIR techniques to eliminate the signals due to fat
at the “bounce point.” Swapping the phase and frequency gradient directions or chan;
ing the polarity of the frequency encode gradient can displace chemical shift artifacts
from a specific image region, even though the chemical shift displacement still exists
Identification of fat in a specific anatomic region is easily discerned from the chemical
shift artlact displacement caused by changes in FEG/PEG gradient directions
Chemical Shift Artifacts of the Second Kind
(Chemical shift anifacts ofthe second kind occur with GE images. resulting from the rephas
ing and dephasing of the echo in the same direction relative to the main magnetic field
Signal appearance is dependent on the selection of TE. This happens because of construc
tive (in phase) or destructive (out of phase) transverse magnetization events that occur
periodically due to the difference in precessional frequencies. At 1.5 T, the chemical shift
IE FIGURE 13-33 The left image of the lumbar
spine is Spin Echo T1 weighted, TR = 450 ms,
TE = 14 ms. The night image is T1 weighted
with chemical fat saturation pulses, TR = 667
ims, TE = 8 ms, In both images, the FEG ts verti=
cally orientedChapter 13 * Magnetic Resonance Imaging 483
Water & Fat Out-of-Phase at 22, 6.6, 11.0 ms,
FIGURE 13-34 For GE image sequences, signal intensity of the transverse magnetization vector due to
the 220 Hz lower precessional frequency of fat protons, where in-phase magnetization occurs every 4.4 ms
(172205), and out-of-phase magnetization occurs every 4.4 ms shifted by Ys cycle (2.2 ms). Signal intensity is
dependent on the selection of TE, as shown above.
is 220 Hz, and the periodicity of each peak (in phase) between water and fat occurs at 0.
4.5, 9.0, 13.5, ms, and each valley (out of phase) at 2.25, 6.75, 11.0, .... ms, as shown
in Figure 13-34 Thus, selection of TE at 9 ms will lead to a constructive addition of water
and fat, and TE at 7 ms will lead to a destructive addition of water and fat. The in-phase
timing will lead to a conventional chemical shuft image ofthe first kind, while the out-of-
pphase timing will lead to chemical shift image of the second kind, manifesting a dark rim
around heterogeneous water and fat anatomical structures, shown in Figure 13-35.
Ringing Artifacts
Ringing anifact (also known as Gibbs phenomenon) occurs near sharp boundaries,
and high-contrast transitions in the image, and appears as multiple, regularly spaced
parallel bands of alternating bright and dark signal that slowly fades with distance.
TE=9ms TE =7ms
Water & Fat in-phase Water & Fat out-of-phase
FIGURE 13-35 Breast MAI images show the effect of selecting a specific TE for a GE acquisition. On the left,
chemical shift ofthe “Tirst kind” is shown wath TE = 9 ms and water and fat in phase for tansverse magnet
zation, shifted only due to the intrinsic chemical shift differences of fat and water. On the right, chemical shift
Of the second kind is additionally manifested with TE = 7 ms, due to fat and water being out of phase, creating
‘a lower signal at all fat-water interfaces, and resuiting in reduced intensity (Reprinted by permission, Hendnck
RE. Breast MRt: fundamentals and technical aspects. New York, NY: Springer, 2007.)Section Il + Diagnostic Radiology
The cause is the insufficient sampling of high frequencies inherent at sharp disconti-
rites in the signal, Images of objects can be reconstructed from a summation of sinu-
soidal waveforms of specific amplitudes and frequencies, as shown in Figure 13-36 for
a simple rectangular object. In the figure, the summation of frequency harmonics, each
with a particular amplitude and phase, approximates the distribution of the object, but
initially does very poorly, particularly at the sharp edges. As the number of higher fre-
quency harmonics increase, a better estimate is achieved, although an infinite number
of frequencies are theoretically necessary to reconstruct the sharp edge perfectly.
In the MR acquisition, the number of frequency samples is determined by the
number of pixels (frequency, k,, or phase, k,, increments) across the k-space matrix.
For 256 pixels, 128 discrete frequencies are depicted, and for 128 pixels, 64 discrete
frequencies are specified (the k-space matrix is symmetric in quadrants and dupli-
cated about its center). A lack of high-frequency signals causes the “ringing” at sharp
transitions described as a diminishing hyper- and hypointense signal oscillation from
the transition. Ringing artifacts are thus more likely for smaller digital matrix sizes
(Fig, 13-37, 256 versus 128 matrix). Ringing artifact commonly occurs at skull/brain
interfaces, where there is a large transition in signal amplitude
A 7 8.
Frequency |’ ‘pati
harmonics | domain |
1 rectangie | '|
: estimate |
1+ Bed
eaten |. —-
esteem a7). —
FIGURE 13-36 The syrthesis of a spa-_C. Sharptransition in MR image:
‘val object occurs by the summation of
frequency harmonics in the MR image
‘A. Left column: frequency harmonics that
estimate a rectangle function wth progres-
sively higher frequencies and lower ampt- “Ringing”
tudes are shown. B. Middle column. As NX
higher frequency harmonics are included,
the summed result more faithfully rep:
resents the object shape, in this example
a rectangle with two vertical edges. The
‘number of frequencies encoded in the MR
image is dependent on the matrix size
Right column: A sharp transition bourd-
ay i an MR image is represented with 256 72256
samples better than with128 samples ({re- samples
quency harmonics in k-space). The amount
fringing caused by insutficent samping is
reduced with a larger number of samples.Chapter 13 * Magnetic Resonance Imaging 485,
256 (vertical) x 128 (horizontal) 256 x 256 FIGURE 13-37 Example of ninging
artifacts caused by a sharp signal tran-
sition at the skull in a brain image for
a 256 X 128 matrix (eft) along the
short (horizontal axis, and the elimi:
nation of the artifact in a 256 X 256
matrix (right). The short axis defines
the PEG direction
Wraparound Artifacts
The wraparound artifact is a result of the mismapping of anatomy that lies out-
side of the FOV but within the slice volume. The anatomy is usually displaced to
the opposite side of the image. It 1s caused by nonlinear gradients or by under-
sampling of the frequencies contained within the returned signal envelope. For
the latter, the sampling rate must be twice the maximal frequency that occurs in
the object (the Nyquist sampling limit). Otherwise, the Fourier transform can-
not distinguish frequencies that are present in the data above the Nyquist [re-
quency limit, and instead assigns a lower frequency value to them (Fig. 13-38)
Frequency signals will “wraparound” to the opposite side of the image, masquer-
ading as low-frequency (aliased) signals.
In the frequency encode direction, a low-pass filter can be applied to the acquired
tume domain signal to eliminate frequencies beyond the Nyquist frequency. In the
phase encode direction, aliasing anifacts can be reduced by increasing the number
of phase encode steps (the trade-oll is increased image time). Another approach is to
‘move the region of anatomic interest to the center of the imaging volume to avoid the
‘Sampling rate
FIGURE 13-38 Left. Wraparound artifacts are caused by aliasing. Shown isa fixed sampling rate and net pre-
Cessional frequencies accurnng at poston A and positon B within the FOV that have identical frequencies but it
ferent phase. f sgnal from position Cis at twice the frequency of B and insutfiiently sampled, the same frequency
and phase wall be assigned to C as that assigned to A, and therefore will appear at that location. Right. A wrap-
around artfact example csplaces anatomy from one side of the image (or outude of the FOV} to the other side486
Section Il * Diagnostic Radiology
overlapping anatomy, which usually occurs at the periphery of the FOV. An “antialias-
ing” saturation pulse just outside of the FOV is yet another method of eliminating
high-frequency signals that would otherwise be aliased into the lower frequency spec-
trum. This example of wrap-around artifact is easy to interpret. In some cases, the
antfact is not as well delineated (e.g., the top of the skull wrapping into the brain)
Partial Volume Artifacts
Partial volume artifacts anse from the finite size of the voxel over which the signal is
averaged. This results in a loss of detail and spatial resolution. Reduction of partial
volume artifacts is accomplished by using a smaller pixel size and/or a smaller slice
thickness. With a smaller voxel, the SNR is reduced for a similar imaging time, result-
ing ina noisier signal with less low-contrast sensitivity. Of course, with a greater NEX
(averages), the SNR can be maintained, at the cost of longer imaging time.
36 Magnetic Resonance Spectroscopy
Magnetic resonance spectroscopy (MRS) is a method to measure tissue chemistry (an
“electronic” biopsy) by recording and evaluating signals from metabolites by iden-
tilying metabolic peaks caused by frequency shifts (in parts per million, ppm) rela-
live to a frequency standard, In vivo MRS can be performed with 'H (proton), Na
Godium), and "P (phosphorus) nuclei, but proton spectroscopy provides a much
higher SNR and can be included in a conventional MRI protocol with about 1010 15
min extra exam time. Uses of MRS include serial evaluation of biochemical changes
in tumors, analyzing metabolic disorders, infections and diseases, as well as evalua-
tion of therapeutic oncology treatments for tumor recurrence versus radiation dam-
age. Early applications were dedicated to brain disorders, but now breast, liver, and
prostate MRS are also performed. Correlation of spectroscopy results and MR images
are always advised before making a final diagnosis
In MRS, signals are derived from the amplitude of proton metabolites in targeted
tissues In these metabolites, chemical shifts occur due to electron cloud shielding
of the nuclei, causing slightly different resonance frequencies, which exist in a fre-
quency range between water and fat. The very small signal amplitudes of the metabo-
lites require suppression of the extremely large (~10,000 times higher) amplitudes
due to bulk water and fat protons, as shown in Figure 13-39. This is achieved by
using specific chemical saturation techniques, such as CHESS (Chemical Shift-Selec-
tive) or STIR (see Chaptet 12). In many cases, the areas evaluated are away from fat
structures, and only bulk water signal suppression is necessary; however, in organs
such as the liver and the breast, suppression of both fat and water are required. Once
the water and fat signals are suppressed, localization of the targeted area volume is
achieved by either a single voxel or multivoxel technique
Single voxel MRS sampling areas, covering a volume of about 1 cm? are delin-
eated by a STEAM (Stimulated echo acquisition mode) or a PRESS (Point Resolved
Spectroscopy) sequence. The STEAM method uses a 90-degree excitation pulse and
90-degree refocusing pulse to collect the signal in conjunction with gradients to
define each dimension of the voxel. The PRESS sequence uses a 90-clegree excitation
and 180-degree refocusing pulse in each direction. STEAM achieves shorter echo
times and superior voxel boundanes, but with lower SNR_ After the voxel data are col-
lected, a Fourier transform is applied to separate the composite signal into individual
frequencies, which are plotted as a trace for a normal brain spectrum (Fig. 13-40)
The resulting line widths are based on homogeneity of the main magnetic field asChapter 13 * Magnetic Resonance imaging 487
‘Water peak FIGURE 13-39 MRS metabolites of
interest in comparison to the water and
fat peaks commonly used for imaging,
Jn order to isolate the very small signals,
chemical saturation of the water (and fat
7 when present) signal is essential.
Fat peak
Relative Amplitude:
5 4 3 2 4 0
Frequency shift (ppm)
well as the magnetic field strength. Higher field strengths (¢.g.. 3.0 T) will improve
resolution of the peaks and corresponding SNR.
Mubivoxel MRS uses a CSI (Chemical Shift Imaging) technique to delineate multiple
voxels of approximately 1 cm’ volume in 1, 2, or 3 planes over a rectangular block of sev-
eral centimeters, achievable with more sophisticated equipment and longer scan times
This is followed by MRSI (Magnetic Resonance Spectroscopic Imaging) where the signal
intensity of a single metabolite in each voxel is color encoded for each voxel according
to concentration and the generated parameter maps superimposed on the anatomical
MR image. In practice, the single voxel technique is used to make the initial diagnosis
because the SNR is high and all metabolites are represented in the MRS trace. Then, a
multivoxel acquisition to assess the distribution of a specific metabolite is performed.
Proton MRS can be performed with short (20 to 40 ms), intermediate (135 to
145 ms), or long (270 to 290 ms) echo times. For short TE, numerous resonances
from metabolites of lesser importance (with shorter T2) can make the spectra more
difficult to interpret, and with long echo times, SNR losses are too severe. Therefore,
most MRS acquisitions use a TE of approximately 135 ms at 1.5 T. Metabolites of
interest for brain spectroscopy are listed in Table 13-2
Applications of MRS are achieved through the interpretation of the spectra that
are oblained from the lesion, from its surroundings, and presumably healthy tissue in
RactoGrapics 2006: 265173-5180
MR Spectrum trom anaplastic cligoastrocytoma Choline / Creatine ratio map
BE FIGURE 13-40 Left: intermediate echo (TE=135 ms) single voxel spectrum s shown, positioned over an anaplastic
‘ligoastrocytoma brain lesion. Note the elevated Choline peak and lowered Creatine and NAA peas. Right: Mult
‘voxel spectrum is color codes to the Choline / Creatine rato. ilstrating the regional variaton of the metabolites cor-
responding to tumor. From Al-Okail RN, Kreza J, Wang S, Woo JH, Meihem ER. Advanced MR imaging Techniques
in the Diagnosis of itraaxial Brain Tumors in Adults. Radiographics 2006; 26. $173-5189. By permission.