TMR-300 E4012c
TMR-300 E4012c
Strain gauge
DC voltage
Thermocouple
(by using Thermocouple adapter TA-01KT)
Ethernet hub
LAN LAN
Data
Measured data are saved in recording media of the personal computer such as HDD
by using Real time data acquisition software RD-7300 (standard accessory) or RD-7640 (option)
Carrier type strain unit, which is less affected by noise, is available in TMR-300 series. Carrier wave bridge excitation has the
advantage of not being influenced by low frequency noise such as thermal electromotive force and commercial power
noise. It also shows high SN ratio and excellent stability. The carrier type strain unit enables highly accurate measurement
even in a site where induction noise or commercial power noise is expected. The number of measurement points is
expandable up to 80.
Waveform affected by noise Noise reduced by the use of carrier type strain unit
Distributed connection
For measurement of automobiles
Measurement units are scattered by extending the
cable between the control unit and measurement
unit, and/or between adjacent measurement units
Control cable
CR-6490 (3cm) standard accessory
CR-6491 (1m)
CR-6493 (3m)
CR-6495 (5m)
80 points at
maximum
80 points at
LAN maximum
80 points at
Synchronization cable maximum
(100m at maximum) 80 points at
maximum
Measurement is possible for multiple points in remote places (320 points at maximum, When using
Dynamic measurement software TMR-7630)
Distribution
adapter
10 units at maximum
Compact size, anti-vibration and DC drive; suited to Data saving in momentary power failure and
vehicle onboard measurement automatic restart after power recovery
Due to its compact size, light weight and vibration tolerance, multi-recorder is Multi recorder has an UPS (uninterruptible power supply) circuit. If a momentary
suited to vehicle onboard measurement. The control unit TMR-311 is driven by a power failure occurs unexpectedly, measurement is stopped and data are saved in
DC power supply, and the power for each measurement unit is supplied from the the SD card automatically before shutting down. When CONTINUE or FREE RUN is
control unit. Supply voltage range is DC 10 to 30 V. selected as the trigger mode, measurement is started again after power recovery.
Tablet PC
(Setting, monitoring,
Carrier type strain unit TMR-323
Recorder measurement result, analysis
Strain gauge
and processing, etc.)
DC voltage *2: Built-in wireless LAN is not available for overseas model.
Unit extension 100m at the maximum (star type connection); Each measurement unit
Measured data are stored in SD card. SD card up to 32GB can be used to enable
long-time data recording. It is also possible to perform recording in high speed
Data Editing Software RD-7300-E
mode of 0.01ms.
Data editing software RD-7300-E can edit data file which is collected by the dynamic
Recording time of 16GB SD card (standard accessory) measurement software RD-7300. Its function includes merging of files, calculation,
data thinning and chart display. In addition, by converting the data into text file of
Automatic recording mode Free-run
CSV format, it can be processed by our FFT analysis software DFA-7610.
Sampling 1ms
Number of channels Recording time 7-segment LED of 3-digit to indicate the
8 (1 unit) Approx. 277 hours instrument status
80 (10 units) Approx. 27 hours
Status including IP address setting and error code are indicated by the 3-digit
7-segment LED display on the front side of the TMR-311. The status of this
USB, LAN and wireless LAN are provided for instrument is easily checked.
connection to a personal computer Contents of indication by 7-segment LED
IP address
Control unit TMR-311 is equipped with two interfaces USB(2.0) and LAN. In State of charge of UPS
addition, built-in wireless LAN is provided to perform setting, monitoring and Wireless LAN status
measurement by wireless using a tablet PC with the supplied software TMR-7300 Serial number
installed. *2 SD card information
Power drop
*2: Built-in wireless LAN is not available for overseas model. Error
MULTI RECORDER TMR-300 Series
Input unit for strain gauge type transducer
Strain Full Bridge Unit TMR-321 and DC voltage
8 measurement points per one unit
Balancing
within ±3×10 -6 strain
accuracy
Stability on zero ±1×10 -6 strain/°C (at maximum sensitivity)
Stability on
±0.05%/°C (at maximum sensitivity)
sensitivity
[Voltage measurement] (when using optional cable CR-4010)
Measuring range ±20 V
Measuring ±20/10/5V range: ±0.2%FS (at 23±5°C)
accuracy ±2V range: ±0.3%FS (at 23±5°C)
Settable range ±20/10/5V range (1mV resolution)
±2V range (0.1mV resolution)
INPUT Strain gauge type transducer DC Voltage Frequency response DC ~ 10kHz
Low pass filter
Load cell Acceleration Cutoff frequency Digital filter
transducer etc. 1Hz ~ 1kHz (settable by every 1Hz)
and PASS (analog filter 10kHz)
-3dB ± 1dB
Specifications TMR-321 Cutoff 1Hz ~ 1kHz: -12dB/oct or - 48dB/oct
characteristics Butterworth filter or Bessel filter
Number of measuring PASS (10kHz): -12dB/oct Bessel filter
8
points High pass filter
Input Strain, Voltage (when using optional cable CR-4010)
Cutoff frequency Digital filter
[Strain measurement] 0.2Hz, 1Hz and OFF
Applicable gauge A/D converter 24bit
120 ~ 1000Ω
resistance Indicator Channel LED (open, over, etc.)
Bridge excitation DC 0.5V, 2V Unit number setting switch
Measuring range ±20000×10 -6 strain (bridge excitation DC 2V) Power supply DC 10 ~ 30V, 0.2A at maximum (12V)(supplied from
±80000×10 -6 strain (bridge excitation DC 0.5V) TMR-311)
Measuring ±20000/10000/5000×10 -6 strain range Environment 0 ~ +50°C, 85%RH or less (no condensation)
accuracy ±0.1%FS (at 23±5°C) Vibration tolerance 29.4m/s2 (10 ~ 55Hz), 3 directions
±2000×10 -6 strain range
±0.2%FS (at 23±5°C) External dimensions 200(W) × 25(H) × 100(D)mm (excluding projected parts)
Settable range ±20000/10000/5000/2000×10 -6 strain range Weight Approx. 550g (including rubber protectors)
Resolution ±20000/10000/5000×10 -6 strain range
1×10 -6 strain (bridge excitation 2V) Standard accessories
4×10 -6 strain (bridge excitation 0.5V)
±2000×10 -6 strain range Operation manual (A3 folded in one-eighth) … …………… 1 copy
0.1×10 -6 strain (bridge excitation 2V) Control cable CR-6490……………………………………………1 pc.
0.4×10 -6 strain (bridge excitation 0.5V) Sensor input cable CR-6186… ……………………………… 8 pcs.
Balancing method Electronic automatic
Balancing range ±10000×10 -6 strain
Carrier type Strain Unit TMR-323 Most suited to measurement on site where
induc tion noise or commercial power
noise is expected
Frequency response DC ~ 2.5kHz
Low pass filter
Cutoff frequency Digital filter
5Hz ~ 1kHz (settable by every 1Hz)
and PASS (2.5kHz)
-3dB ± 1dB
Cutoff 5Hz ~ 1kHz: - 48dB/oct
characteristics Butterworth filter or Bessel filter
PASS (2.5kHz): Butterworth filter
High pass filter
INPUT Strain gauge type transducer
Cutoff frequency Digital filter
0.2Hz, 1Hz and PASS
Load cell Acceleration
transducer etc. A/D converter 18 bit
Indicator Channel LED (open, over, etc.)
Unit number setting switch
Specifications TMR-323 Power supply DC 10 ~ 30V, 0.3A at maximum (12V)(supplied from
TMR-311)
Number of measuring Environment 0 ~ +50°C, 85%RH or less (no condensation)
8
points Vibration tolerance 29.4m/s2 (10 ~ 55Hz), 3 directions
Input Strain External dimensions 200(W) × 25(H) × 100(D)mm (excluding projected parts)
[Strain measurement] Weight Approx. 660g (including rubber protectors)
Applicable gauge
120 ~ 350Ω
resistance
Bridge excitation 0.5Vrms, 2Vrms 5kHz Standard accessories
Measuring range ±20000×10 -6 strain (bridge excitation 2Vrms) Operation manual (A3 folded in one-eighth) … …………… 1 copy
±80000×10 -6 strain (bridge excitation 0.5Vrms) Control cable CR-6490……………………………………………1 pc.
Measuring
±0.3%FS (at 23±5°C) Sensor input cable CR-6186… ……………………………… 8 pcs.
accuracy
Resolution 1×10 -6 strain (bridge excitation 2Vrms)
4×10 -6 strain (bridge excitation 0.5Vrms)
Balancing Resistance ±10000×10 -6 strain
range Capacity 3000pF
Balancing method Software method
Stability on zero within ±0.1×10 -6 strain/°C
Stability on
within ±0.05%FS/°C
sensitivity
FEATURES
Measurement units can be distributed in star-type
connection Up to 10 measurement units are connected
The connection between the distribution unit TMR-371 and each measurement unit Ten measurement units can be connected at the maximum including measurement
(distribution adapter TMR-371-1) is made by STP cable (100 m at the maximum). units directly connected to the control unit.
Synchronized measurement of sensors scattered in a large area can be easily
performed.
Measurement unit Measurement unit Measurement unit LAN Measurement unit Measurement unit Measurement unit
TMR-300
Standard accessories
Operation manual (A3 folded in one-eighth) … …………… 1 copy
Control cable CR-6490……………………………………………1 pc.
Standard accessories
Operation manual (A3 folded in one-eighth) … …………… 1 copy
Option
Bridge Box SB-120T / SB-350T Handles
These are connected to the strain 1G2G4G unit TMR-322 and used to These are attached to the upper sides of the
connect strain gauges in quarter bridge 3-wire or half bridge method. control unit and used for carrying and/or fixing
the combined control unit and measurement
Eight pieces of SB-120T or SB-350T are
units. (Screws for attaching the handles are
supplied with the TMR-322 as standard included.)
accessories.
Brackets
Number of measuring point 1 These are attached to the lower sides of the
bottom unit and used for the installation of the
Applicable gauge resistance 120Ω (SB-120T)
combined control unit and measurement units.
350Ω (SB-350T)
(Screws for attaching the brackets are included.)
Connection method Quarter bridge 3-wire, Half bridge
Environment 0 ~ +50°C, 85%RH or less (no condensation)
External dimensions 20(W) × 14.5(H) × 25(D) mm (excluding
projected parts)
Extension cable for distribution adapter (STP cable)
Weight Approx. 10g This is a STP (Shielded Twisted Pair) cable used for connecting between the
distribution unit TMR-371 and the distribution adapter TMR-371-1. The
maximum available extension distance is 100 meters.
Attenuator cable CR-4010
Type Cable length
This is used for voltage measurement with TMR-321.
CR-8805 5m
CR-8810 10m
CR-8820 20m
Measuring range of voltage : ±20V CR-8850 50m
CR-8899 CR-8899 100m
BNC connector
Measurement software
Dynamic measurement software TMR-7300, RD-7300 and RD-7300E, which are TMR-311 as standard accessories. Optional software programs with expanded
capable of measuring up to 80 channels using one control unit, are supplied to the functions are also available.
Applicable software
Standard
Optional software Dynamic measurement software RD-7300 for
software real time data acquisition (standard software)
Dynamic measurement TMR-7630
software TMR-7300 TMR-7630-H (Frequency analysis) The software RD-7300 directly collects the data measured by TMR-300 series into a
TMR-7630-M (Video applicable) personal computer and records them. Long-time and large-capacity recording is
Real time data possible without depending on the capacity of the TMR-311 data memory or a SD
RD-7300 RD-7640
acquisition software card. Data processing is possible by the software RD-7300-E which is also supplied as
Waveform view standard accessory.
RD-7300-E WF-7630
software System
OS Windows Vista(SP2), 7(SP1), 8, 8.1, 10
Dynamic measurement software TMR-7300 Computer Model recommended by the above OS with dual or more
(standard software) core CPU is recommended
Interface LAN(100BASE-TX)
The dynamic measurement software TMR-7300 controls one TMR-311 for on-line and Memory capacity 4GB or more is recommended
off-line measurement. Monitoring, acquisition, edition (listing and chart drawing) HDD capacity Free space of 5GB or more
and processing of data, and data calculation using expanded channels are possible. Basic specifications
In off-line measurement, free-run, data trigger and program measurement can be Applicable
executed. TMR-311 Maximum number of connection: 1
instrument
System Number of
80 channels at maximum
measuring points
OS Windows Vista(SP2), 7(SP1), 8, 8.1, 10
Sampling clock Setting is possible within the range of 0.1 to 0.9ms (by every
Computer Model recommended by the above OS with dual or more
0.1ms) and 1 to 1000ms (by every 1ms)
core CPU is recommended If the number of used channels is 41 or more for one
Interface Wireless LAN *1, LAN(100BASE-TX), USB instrument, the fastest sampling clock is 0.2ms
Memory capacity 4GB or more is recommended Expanded channel 1000 channels at maximum (four arithmetic operation,
HDD capacity Free space of 5GB or more various functions and rosette analysis)
Basic specifications Measurement Monitor measurement, Manual measurement, Data trigger
measurement, Interval measurement
Applicable
TMR-311, TMR-211 Maximum number of connection: 1 Display Value monitor, T-Y monitor, T-Y graph, Bar monitor,
instrument
Spectrum, Dial scale monitor, Vector monitor, Arrow monitor
Number of
80 channels at maximum
measuring points
Expanded channel 1000 channels at maximum (four arithmetic operation, Data editing Software RD-7300-E
various functions and rosette analysis) (standard software)
On-line Balance, Monitor, Manual, Interval, Data comparater, Free
measurement run, Data trigger, Program measurement, Alarm output This software performs post-processing of data files collected by the RD-7300 such
Off-line as file management, batch processing of two or more files and chart creation.
Free run, Data trigger, Program measurement
measurement
System
Display Value monitor, T-Y monitor, T-Y graph, Bar monitor,
Spectrum Applicable data file *.hed / *.dat (DADiSP compatible format)
OS Windows Vista(SP2), 7(SP1), 8, 8.1, 10
Data format DADiSP format
CPU Conforms to the system requirements of the above OS
Conversion to text file (CSV format) possible
Memory Conforms to the system requirements of the above OS
Data processing Display and print of T-Y graph, Display of value list
Disc capacity Free space of 5GB or more
*1: Built-in wireless LAN is not available for overseas model of TMR-311. Basic specifications
Processings below are applied to optionally selected two or
Data file more files
management File display, File rename, File move, Text conversion,
Merging files
Re-setting of channel setting
Data file Setting of expanded channel and re-calculation
processing Searching maximum/minimum values, Cutting out, Thinning
out, Text conversion
T-Y graph, X-Y graph, Spectrum, Label, Saving,
Graph display
Text saving, Copy of graph, Saving pictures
TMR-300
Measurement software
Visual LOG® Dynamic measurement software TMR-7630 (option)
The sof t ware TMR-7630 is for multi- channel dynamic measurement and data
processing using multi-recorder TMR series. Simultaneous control of 320 channels at
maximum is possible by connecting four control units TMR-311.
Measurement is possible on-line and off-line. In on-line measurement, calculation
using expanded channels and monitoring measurement are available. In off-line
measurement, free-run, data trigger and program measurement can be executed.
System
OS Windows Vista(SP2), 7(SP1), 8, 8.1, 10
Computer Model recommended by the above OS with dual or more
core CPU is recommended
Interface Wireless LAN *1, LAN(100BASE-TX), USB
Memory capacity 4GB or more is recommended
HDD capacity Free space of 10GB or more is recommended
Protect key USB dongle
Basic specifications
Applicable
TMR-311, TMR-211 Maximum number of connection: 4
instrument
Number of
320 channels at maximum
measuring points
Expanded channel 1000 channels at maximum (four arithmetic operation,
various functions and rosette analysis)
On-line Balance, Monitor, Manual, Interval, Data comparater, Free [Option]
measurement run, Data trigger, Program measurement,Alarm output TMR-7630-H Performs frequency analysis of measured dynamic wave
Off-line form in post-processing. Frequency analysis and S-N
Free run, Data trigger, Program measurement
measurement
analysis of expanded channels are also possible.
Display Value monitor, T-Y monitor, T-Y graph, Bar monitor,
Spectrum
TMR-7630-M Videos taken by a camera conforming to DirectX are
Data format DADiSP format
Conversion to text file (CSV format) possible saved linking with the measurement. The saved data are
Data processing Display and print of T-Y graph, Display of value list reproduced in synchronization with the video.
*1: Built-in wireless LAN is not available for overseas model of TMR-311.
System
OS Windows Vista(SP2), 7(SP1), 8, 8.1, 10
Computer Model recommended by the above OS with CPU of Intel
Core i5 3.0GHz or higher is recommended
(excluding Turbo Boost)
Interface LAN(100BASE-TX)
Memory capacity 4GB or more is recommended
HDD capacity Free space of 5GB or more
Protect key USB dongle
Basic specifications
Applicable TMR-311 Maximum number of connection: 4
instrument In addition, this software is applicable to DS-50A and TFM-104
Sampling clock Setting is possible within the range of 0.1 to 0.9ms (by every
(when using TMR- 0.1ms) and 1 to 1000ms (by every 1ms)
311) If the number of used channels is 41 or more for one
instrument, the fastest sampling clock is 0.2ms
Expanded channel 1000 channels at maximum (four arithmetic operation,
various functions and rosette analysis)
Measurement Monitor measurement, Manual measurement, Data trigger
measurement, Interval measurement
Display Value monitor, T-Y monitor, T-Y graph, Bar monitor,
Spectrum, Dial scale monitor, Vector monitor, Arrow monitor
Visual LOG® Waveform view software WF-7630 (option)
The software WF-7630 is for viewing DADiSP format data as data list and waveform.
DADiSP format data outputted by our instrument TMR-311/TMR-211 or software
RD-7640/TMR-7300/TMR-7630 and so on are acceptable. It is possible to execute
re-calculation of data, merging, cutting out, thinning out and CSV conversion of data
files, searching of maximum/minimum values, FFT analysis, and calculation and chart
drawing (X-Y, T-Y, spectrum) of expanded channels.
System
*.hed / *.dat (DADiSP compatible format)
DADiSP file of Integer format or ASCII format outputted
from instrument TMR-311/TMR-211/DC-204/DC-104/DH-
14A, or dynamic measurement software RD-7640/TMR-
Applicable data file 7630/TMR-7300/TMR-7200/DS-750/DC-7630/DRA-7630/
DC-7004P (below referred to as data file)
Note) If GPS data and/or frequency data are included in
measurement data recorded by TMR-211,
the measurement data cannot be read by this software.
OS Windows Vista(SP2), 7(SP1), 8, 8.1, 10
CPU Conforms to the system requirements of the above OS
Memory Conforms to the system requirements of the above OS
Disc capacity Free space of 5GB or more
Basic specifications
Maximum number of channels: 1000
Data file
Number of expanded channels: 1000
Processings below are applied to optionally selected two or
Data file more data files
management File display, File rename, File move, Text conversion,
Merging files
Re-setting of channel setting
Setting of expanded channel and re-calculation
Data file processing
Searching maximum/minimum values, Cutting out, Thinning
out, Text conversion
T-Y graph, X-Y graph, Spectrum, Label, Saving,
Graph display
Text saving, Copying graph, Saving pictures
Visual LOG is a registered trade mark of Tokyo Sokki Kenkyujo Co., Ltd.
The contents of this catalog are subject to change without prior notice.
The contents of this catalog are as of February 2018.