CALIBRATION CERTIFICATE

Calibration Certificate of "Digital Pressure Calibrator" Calibration Certificate Number Calibration Certificate Date Page Number
Identification/ Sr. Number"CAL-652/8219" 2023/142/00007/74488 02/12/2023 Page 1 Of 2
Calibration Date Of Next Calibration As Per Customer Reference Of
Date Of Receipt Of Calibration
Completed On Customer's Confirmation/ Gate Pass/ Challan With Job No.
UUC To Be Calibrated Started On Date
Date Agreement Date
JI-
06/10/2023 02/12/2023 02/12/2023 02/12/2024 NA Dated 06/10/2023
74488
1.0 Name And Address Of The M/S, Oorja Technical Services Pvt. Ltd.
Customer: 315/316/317 Block MS1-B, New Siyaganj, Opp. Mall Go-Down Road, Indore - 452001, Madhya
Pradesh
2.0 Details Of The UUC To Be Calibrated
Nomenclature Of UUC Identification/ Sr. Number Make Model Number Instrument Range Resolution/ Least Count
Digital Pressure Calibrator CAL-652/8219 Polltech PSI-DPC1 (0 to 200)mmWc 0.01 mmWc
Accuracy Claimed/ Acceptable Error
±0.1%FS
3.0 Condition Of Item On Receipt: Satisfactory
4.0 Details Of The Adjustments Or Repairing Done (If Any): No
5.0 Location Of The Place Where Calibration Was Carried Out: Mechanical Calibration Laboratory(CL-02)
6.0 Environmental Conditions Of The Space Where Calibration Was Carried Out:
Temperature (°C) Relative Humidity (% RH)
20 ± 2 50 ± 10
7.0 Procedure For Calibration :CP-14
The Calibration Was Performed By Using Precision Standard Reference Pressure Indicator With A Pneumatic Pressure Source 3 Cycles With Each
In Increment & Decrement Order Is Performed Calculating Repeatability , Hysteresis & Errors Reference DKD R-6 -1
8.0 Format Number: OTS/CAL/CP-14/002
9.0 A Sticker Indicating Calibration Status Has Been Affixed On The UUC
10.0 Description Of Reference Standard Instrument Used
Description Of Standard Reference Instrument Used Sr. No. /ID No. Certificate Number Calibration Valid Upto
Ext. Module of Multifunction Pressure Calibrator 60474 / LAB-215 PSL/2268/0223/P/0409 23/02/2024
11.0 Uncertainty Of The Measurement: Refer Page 2 Of 2
12.0 Calibration Results: Refer Page 2 Of 2
13.0 Notes:
1. The reported value has been rounded off as per IS: 2-1960 wherever applicable.
2. This Certificate Refer only to the Particular Instrument submitted for the Calibration and Results are Applicable only to the Item (UUC)
mentioned Above which has been Calibrated and do not apply to the other item even though declared Identical.
3. This Certificate shall not be reproduced Except in full, if reproduced in a part or Abstract thereof has to be Specifically approved in
written from the Calibration Laboratory.
4. The results reported in this Certificate are Valid at the time of and under the stated Condition of Measurement.
5. Oorja Technical Services Pvt. Ltd. shall not be liable for any change in Calibration Data and Performance Specification on Account of
Malfunctioning of Item due to any Damage or Mishandling caused to it after the Certificate in Respect of it has been issued.
6. The standard used is traceable to National /International standards.
7. Calibration is performed without doing any adjustment or correction in it’s original condition at the time receipt.
8. The user should determine the suitability of the instrument for it’s intended use
9. This calibration certificate is valid only if the stated weights/ equipment/ UUC is used for scientific/ research and industrial purpose, if
used for commercial trading, additional recognition /approval shall be complied as required by department of legal metrology,
regulatory bodies, etc.
10. The reported expanded uncertainty is at coverage factor k=2 which corresponds to a coverage probability or level of confidence of
approximately 95% for a normal distribution.
11. The Error (UUC. Reading- Reference Standard Instrument Reading) is Positive except when Indicated otherwise; it has to be
Algebraically added to the UUC Reading in Order to get the Standard Reading.
12. ULR-CC254723000065315F
For Oorja Technical Services Pvt. Ltd.(Center For Development Of Calibration & Testing)
Calibrated By Reviewed & Approved By

Note : This Certificate/Document Is Electronically Generated And Does Not Require Any Seal/Physical Signature In Original
Calibration Certificate of "Digital Pressure Calibrator" Calibration Certificate Number Calibration Certificate Date Page Number
Identification/ Sr. Number"CAL-652/8219" 2023/142/00007/74488 02/12/2023 Page 2 Of 2
Calibration Date Of Next Calibration As Per Customer Reference Of
Date Of Receipt Of Calibration
Completed On Customer's Confirmation/ Gate Pass/ Challan With Job No.
UUC To Be Calibrated Started On Date
Date Agreement Date
JI-
06/10/2023 02/12/2023 02/12/2023 02/12/2024 NA Dated 06/10/2023
74488
12.0 Calibration Results:
Sr. No. Set Value Shown By UUC Value Shown By Reference Standard Error Expanded Uncertanity (±U)
Cycle-01 Cycle-02 Cycle-03
Average
M1(↑) M2(↓) M3(↑) M4(↓) M5(↑) M6(↓)
Unit mmWc mmWc mmWc mmWc mmWc mmWc mmWc mmWc mmWc mmWc
1 0.00 0.000 0.000 0.001 0.000 0.001 0.000 0.000 0.000 0.663
2 50.00 50.199 50.197 50.195 50.194 50.196 50.195 50.196 -0.196 0.663
3 100.00 100.197 100.195 100.194 100.193 100.192 100.191 100.194 -0.194 0.663
4 150.00 150.194 150.192 150.190 150.188 150.194 150.192 150.192 -0.192 0.663
5 199.00 198.884 198.882 198.886 198.884 198.882 198.880 198.883 0.117 0.663
Note: 1.0 Formula To Determine Repeatability =(M3-M1), (M4-M2),(M5-M3) & (M6-M4)
2.0 Formula To Determine Hysteresis= (M2-M1),(M4-M3), (M6-M5)
For Oorja Technical Services Pvt. Ltd.(Center For Development Of Calibration & Testing)
Calibrated By Reviewed & Approved By

Note : This Certificate/Document Is Electronically Generated And Does Not Require Any Seal/Physical Signature In Original

End Of The Calibration Certificate