Varian TrueBeam v4
Varian TrueBeam v4
0
4D Integrated Treatment Console 13.0
DICOM Conformance Statement
Version 7.0
L4472 Rev 0
P1064013-001-A APRIL 2023
Publication ID
P1064013-001-A
Publication Title
TrueBeam Treatment Delivery Console V4.0 DICOM Conformance Statement
Abstract
The document provides information about DICOM conformance with TrueBeam 4.0 and 4D Integrated Treatment Console
13.0.
This publication is the English-language original.
Trademarks
Varian Medical Systems, Inc. or its affiliates own the names of our products and services referenced herein. These names are
either registered trademarks (®) or trademarks (™) in the United States and/or other countries. All other trademarks are the
property of their respective owners. Any rights not expressly granted herein are reserved.
Copyright
© 2023 Varian Medical Systems, Inc. All rights reserved. Published in the United States of America.
No part of this publication may be reproduced, translated, or transmitted without the express written permission of Varian
Medical Systems, Inc.
Notice
Information in this publication is subject to change without notice and does not represent a commitment on the part of Varian.
Electronic Labeling
This symbol on the label indicates that the Instructions for Use for the corresponding product are available at
www.MyVarian.com. Access the Instructions for Use in electronic form by logging in with your assigned MyVarian user
credentials.
In compliance with Anvisa and EU regulations, Varian will send Brazil and EU customers a
free printed copy of the Instructions for Use within 7 days. Use the “Paper Publication
Request” form provided on the Varian webpage to order your copy.
Legal Manufacturer
Varian Medical Systems, Inc.
3100 Hansen Way
Palo Alto, CA 94304
United States of America
CAUTION: US Federal law restricts this device to sale by or on the order of a physician.
WHO
ICD-O codes and terms used by permission of WHO, from:
• International Classification of Diseases for Oncology, Third Edition.
ICD-10 codes and terms used by permission of WHO, from:
• International Statistical Classification of Diseases and Related Health Problems, Tenth Revision (ICD-10).
ISO 13485
Varian Medical Systems, Oncology Systems products are designed and manufactured in accordance with the requirements
specified within the ISO 13485 quality standard.
Medical Device
6.0 August 04, 2021 Treatment record only reports two control points.
7.0 April 04, 2023 Interlock sequence, Beam Gating Hold Device gating
sequence and Treatment sites support added to
TrueBeam 4.0
Figure 1: SCU Role Application Data Flow Diagram for Loading Patient 12
Figure 2: SCP Role Application Data Flow Diagram for Loading 13
Figure 3: SCU Role Application Data Flow Diagram for Saving Patient Data 14
Figure 4: Associations Used for 4DITC 17
Figure 5: 4DITC Administration Setup 32
Figure 6: AE Title/Presentation Address Mapping 33
1.1 Audience
This document is intended for the following groups of persons:
Marketing and Sales.
System Integrators of medical equipment.
Other vendors interfacing via DICOM
It is assumed that the reader is familiar with the DICOM standard.
1.2 Scope
This Conformance Statement is applicable to the following Varian treatment delivery consoles:
TrueBeam Treatment Console version 4.0
4D Console version 13.0.
The general DICOM conformance statement for Varian Medical Systems products is defined in [2]
(see section 1.7).
This Conformance Statement describes only the specials and the variations as implemented in
TrueBeam Treatment delivery console version 4.0 and 4D Integrated Treatment Console version 13.0
The scope and format of this document are defined by Part 2 of the DICOM V3.0 standard. Some
sections defined in the standard that are not applicable to the software are left out for clarity.
Query/
Load RT Retrieve
Plan SCP
(C-Move)
Load RT Query/
Treatment Retrieve
Summary SCP
Record (C-Move)
Treatment
Delivery
Console
Query/
Load RT
Retrieve
Beams
SCP
Record
(C-Find)
Query/
Retrieve
Load RT
SCP
Image
(C-Move)
DICOM
Standard
Interface
Figure 1: SCU Role Application Data Flow Diagram for Loading Patient Data
RT Plan
Store Storage
Plan SCU
RT Treatment
Store RT
Summary
Treatment Treatment Record
summary Delivery Storage SCU
Record Console
RT
Store Image
image Storage
SCU
DICOM Standard
Interface
Figure 2: SCP Role Application Data Flow Diagram for Loading Patient Data
RT Plan
Store RT
Storage
Plan
SCP
Store RT RT Beams
Treatment Treatment
Beams
Delivery Record
Treatment
Console Storage
Record
SCP
RT
Store RT Image
Image Storage
SCP
Figure 3: SCU Role Application Data Flow Diagram for Saving Patient Data
The 4DITC console also retrieves referenced RT Images. The steps below only apply to
the 4DITC console.
5. C-Move RT Image SOP class (for Reference Images if referred in the RT Plan)
6. C-Find RT Image SOP class (for query Portal Images).
7. C-Move RT Image SOP class (for Portal Images if query returned Portal Image Instance
[UIDs]).
The only requirement is to load the RT Plan first before the others. This sequence may be performed
for more than one plan at the beginning of a treatment session.
The patient record is saved in the following order to treatment delivery console:
1. C-Store RT Plan SOP class (executed if a plan is changed in treatment delivery console).
2. C-Store RT Beams Treatment Record SOP class (executed if a beam was treated).
3. C-Store RT Image SOP class (executed if a Portal Image was acquired) (4DITC only).
The only requirement is to save RT Plans first before the RT Beam Records and RT Images. This is
because the RT Beams Treatment Record and/or RT Image may refer to the new RT Plan.
Table 1: Supported SCU/SCP SOP Classes for the Treatment Delivery Console Entity
C-Store RT Plan
Management System
C-Store RT Image
C-Store RT Treatment Summary
4DITC
Record
Association
C-Echo
C-Store RT Plan
C-Store RT Beams Treatment
Record
C-Store RT Image
C-Echo
Table 3: TrueBeam Treatment delivery console list of implementation UIDs supported and used
3.1.2.1.3.2.1 4DITC and TRUEBEAM 4.0 with implementation UID ‘Varian Console Interface 4’ and ‘Varian
Console Interface 4 QA Mode’ or later
These tags are in addition to conformance rules applied for lower version of implementation UIDs
supported in older version of Varian treatment delivery console.
Varian treatment delivery consoles use the following tags from ‘Patient Setup Sequence (300A,
0180):
Table Top Vertical Setup Displacement (300A, 01d2).
Table Top Longitudinal Setup Displacement (300A, 01d4).
Table Top Lateral Setup Displacement (300A, 01d6).
Also, when SOP object has at least one table top setup displacement defined, 4DITC (only)
considers zero value for remaining undefined table top setup displacements.
3.1.2.1.3.2.2 Varian treatment delivery consoles communicating on implementation UID “Varian Console
Interface 3” and “Varian Console Interface 3 QA Mode” or later:
These tags are in addition to conformance rules applied for lower version of implementation UIDs
supported in older version of 4DITC.
Treatment Delivery Type (300A, 00CE) supports new defined Term SETUP attribute for Setup
Fields.
Accessory Code (300A, 00F9) for Wedge, Compensator, Block and Applicator.
Varian treatment delivery consoles do not accept Arc-Dynamic Beam(s), which contain gantry
rotations that are not equally spaced (e.g. 0, 10, 30, 100).
Only the first Fraction Group sequence (300A, 0070) is read. The others are discarded.
A maximum of two wedges per field are supported.
A maximum of one MLC per field is supported.
Beam energy, table parameter and rotation direction changes in the control points are not
supported.
Whenever an operator changes a dose relevant attribute from the RT Plan in 4DITC, a new RT
Plan is created. The new RT Plan will refer to the originally loaded RT Plan through the
Referenced RT Plan Sequence (300C, 0002). The Referenced RT Plan Sequence (300C,
0002) can only contain at most one item.
General Accessory Sequence (300A,0420) used for empty Trays.
RT Plan extended interface and privatization are described in section 4.
Replacement for Field Type
In Extended Interface FieldType is used to distinguish between treatment and setup fields (values:
TREATMENT, SETUP) while DICOM attribute Treatment Delivery Type (300A, 00CE) always
contains value TREATMENT. When FieldType is removed from Extended Interface, the new value
SETUP for Treatment Delivery Type (300A,00CE) is used to designate setup fields.
Replacement for CustomAddOn for Block and Compensator
For Block and Compensator the Extended Interface has been used only to transport the accessory
code. This has been added to the DICOM standard and is now available in various accessory
sequences as attribute Accessory Code (300A,00F9), RT Beams Module.
Nominal Beam Dose Rate Set MU/Min Primary Fluence Mode High-Dose
Energy (300A, 0115) (3002,0051)* Technique Type
(300A,00C7)
(MV/MeV)
(300A,0114)
2.5x 60 STANDARD Normal
NON_STANDARD TBI
HDR
SRS
4x 5, 10, 15, 20, 30, 40, 50, 100, 150, 200, 250 STANDARD Normal
NON_STANDARD TBI
HDR
SRS
6x 5, 10, 15, 20, 40, 60, 80, 100, 200, 300, 400, 500, STANDARD Normal
600 NON_STANDARD TBI
HDR
SRS
6x FFF 400, 600, 800, 1000, 1200, 1400 STANDARD Normal
NON_STANDARD TBI
HDR
SRS
15x 20, 40, 60, 80, 100, 200, 300, 400, 500, 600 STANDARD Normal
NON_STANDARD TBI
HDR
SRS
18x 20, 40, 60, 80, 100, 200, 300, 400, 500, 600 STANDARD Normal
NON_STANDARD TBI
HDR
SRS
20x 20, 40, 60, 80, 100, 200, 300, 400, 500, 600 STANDARD Normal
NON_STANDARD TBI
HDR
SRS
4e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
6e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
2500 for HDR
9e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
2500 for HDR
12e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
15e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
16e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
18e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
20e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
22e 100, 200, 300, 400, 500, 600, 700, 800, 900, STANDARD Normal
1000 NON_STANDARD TBI
HDR
SRS
Table 5: Supported combinations for Primary Fluence Mode, Dose Rate, Energy,
and High Dose Technique
(*) Related Beam Number is a private tag added to the RT Beams Session Record module (see
section 4.1.2).
In TrueBeam 4.0 Platform release Interlock Sequence(300A,0740) and Beam Gating Hold Transition
Sequence(300C,0125) are recorded in the RT Beams Session Record module.
RT Beams Treatment Record Extended Interface and Privatizations are described in section 4.
The following SOP specific conformance rules apply to 4DITC with implementation UID ‘Varian
Console Interface 4’ or later.
3.1.5.2.2 SOP specific conformance for the RT Image Storage SOP class
The 4DITC Entity only stores images with the Image Type (0008,0008) set to PORTAL.
SOP specific Conformance for the C–Find RT Beams Treatment Record SOP Class
The C-Find RT Beams Treatment Record SOP class will be used for calculating the remaining MU.
The remaining monitor units are calculated with the Beam Meterset (300A,0086) in the RT Fraction
Scheme module (provided from the Plan IOD) minus the Delivered Primary Meterset (3008, 0036)
(provided from the Beam Record Information) -see Table 18). If the monitor units were overridden, the
Beam Record Information contains the overridden value in the Specified Meterset (3008, 0042). In this
case, the overridden Specified Meterset (3008, 0042) is used instead of the Beam Meterset
(300A,0086) to calculate the remaining monitor units.
Calculation of Remaining MU when monitor units are not overriden:
Remaining MU = Beam Meterset (300A,0086) – Delivered Primary Meterset (3008,0036)
Calculation of Remaining MU when monitor units are overriden:
Remaining MU = Specified Meterset (3008, 0042) – Delivered Primary Meterset (3008,0036)
The Treatment Record key contains the matching criteria for the RT Beams Treatment Record.
Whenever the key matches the particular RT Beams Treatment Record, the attributes defined in Table
18 (SCP column) gets filled in. Note that the Treatment Session Beam Sequence may contain one or
more item. It is up to the Management System to decide whether for each Treatment Session Beam
one Treatment Record key gets sent or if all Treatment Session Beams are sent within one Treatment
Record key.
Only the last treated fraction is the fraction of interest.
Note The unscheduled plan load feature is available only on 4DITC console.
Varian treatment delivery consoles are not an archive system. Thus DICOM IOD’s are temporarily stored
during a treatment session and deleted once the session is finished.
Note In the current release, only 0 is supported. Providing this value will
ensure safe interpretation.
Note When kV images are used as Reference Images, this value describes
the projection geometry. Therefore, this value may not coincide with the
Gantry Hardware Readout of the linear accelerator since the kV Image
radiation source may be on a different position than the linear
accelerator’ target focus.
SOP specific conformance for the RT Treatment Summary Record SOP class
RT Treatment Summary Record IOD contains additional private tags defined in section 4.1.3.
3.1.6.2 Presentation Context Acceptance Criterion
The Varian treatment delivery console entity will accept the presentation contexts listed in Table 4:
Proposed Presentation Contexts for Query/Retrieve SCU.
3.1.6.3 Transfer Syntax Selection Policies
Varian treatment delivery consoles accept both Implicit VR Little Endian and Explicit VR Little Endian
as shown in Table 5. 4DITC does not prefer one presentation context over the other.
West European (Code Page) "ISO_IR 100", or ARIA 11, and other non -
“WINDOWS 1252” Unicode supporting
Databases
Chinese Simplified (Code Page) “WINDOWS 936" ARIA 11, and other non -
Unicode supporting
Databases
Japanese (Code page) “WINDOWS 932" ARIA 11, and other non -
Unicode supporting
Databases
Unicode (UTF8) "ISO_IR 192" ARIA 13, and future Unicode
supporting Databases
TRUEBEAM 4.0 and 4DITC with implementation UID ‘Varian Console Interface 6’ or later supports:
UTF-8
Correct display of characters not in the Default Character Repertoire depends on available/configured
operating system support.
8. Security
8.1 Security Profiles
No Security Profiles are supported.
The table below summarizes the handling of RT Plan IOD Type 2 data elements in treatment delivery
consoles.
TRUEBEAM 4DITC v
Attribute Name Tag Module/Sequence
v 4.0 13.0
Id (0010, 0020) Patient Mandatory. Mandatory.
Table 10: Treatment Delivery Console handling of RT Plan IOD DICOM Type 2 attributes
Mandatory Optional
This attribute is
(300A,
Tolerance Table Label Tolerance Table being used for
0043)
unique tolerance
table identification.
Mandatory Mandatory
This attribute is
(300A, Dose Reference being used for
Dose Reference Description
0016) Sequence unique dose
reference
identification.
Table 11: Treatment Delivery Console handling of RT Plan IOD DICOM Type 3 attributes
Last Treated Fraction (3265,xx01) 1 The index of the last fraction treated SL
across the current plan and all its
former plans. In other words, this tag
will always contain the fraction, which
was last treated (completely or
partially). So it will be the same as in
the Fraction Status Summary
Sequence (3008,0240) on the last item
of the sequence in the tag Referenced
Fraction Number (3008,0223), as soon
as the sequence is not empty any
more (so after the 1st treatment has
got at least 1 MU).
In case of completing a partial
treatment, the plan’s current fraction
number is equal to Last Treated
Fraction (3265, 1001).
Additional Dose Value (3259,xx00) 3 Introduces sequence of Actual Session SQ
Sequence Dose Sequence. The sequence may
contain one or more items.
Sequence for additional Dose
Information to a Reference Point.
The ReferencePoint must be
connected to Referenced Plan
of this TreatmentSummary
Record, but also the Dose of
other Plans which have the
same Reference Point have
influence on this values.
> Actual Session Dose (3259,xx02) 1C Session Dose Delivered of Actual DS
Session
Required if Actual Session Dose
Sequence is sent.
> Daily Dose (3259,xx04) 1C Daily Dose of current day. DS
Required if Actual Session Dose
Sequence is sent.
> Life Time Total Dose (3259,xx06) 1C Total Dose of Reference Point over all DS
Plans.
Required if Actual Session Dose
Sequence is sent.
Table 16: Treatment Summary Record Module Privatization
Note In addition to the IMAGE value for the Composite Object Instance
Level, we expect the following values for querying/retrieving the
respective objects: PLAN, TREATMENTRECORD,
TREATMENTSUMMARYRECORD, TREATMENTSUMREC.
Symbol Description
U Unique Key Attribute
R Required Key Attribute
O Optional Key Attribute
B.1.1.3.1 Plan
*** not supported *** Attribute is unknown and therefore ignored by TXA. This means that the Rountrip
(load, edit and save) will result in erasing this attribute in the RT Plan IOD..
Roundtrip supported. Complete roundtrip is guaranteed. Loading, editing and saving will maintain
(unless edited) the original value in the exported RT Plan IOD.
Study Instance (0020,000D 1 Unique identifier for the Study. Roundtrip Roundtrip
UID ) supported supported
Study Date (0008,0020) 2 Date the Study started. Roundtrip Roundtrip
supported supported
Study Time (0008,0030) 2 Time the Study started. Roundtrip Roundtrip
supported supported
Referring (0008,0090) 2 Name of the patient's referring Roundtrip Roundtrip
Physician’s physician supported supported
Name
Referring (0008,0096) 3 Identification of the patient's *** not Not
Physician referring physician. Only a supported *** supported
Identification single item shall be permitted
Sequence in this sequence.
> Person (0040,1101) 1 A coded entry which identifies *** not Not
Identification a person. The Code Meaning supported *** supported
Code Sequence attribute, though it will be
encoded with a VR of LO, may
be encoded according to the
rules of the PN VR (e.g. caret
'^' delimiters shall separate
name components), except
that a single component (i.e.
the whole name unseparated
by caret delimiters) is not
permitted. Name component
groups for use with multi-byte
character sets are permitted,
as long as they fit within the 64
characters (the length of the
LO VR). One or more Items
may be permitted in this
Sequence.
>> Code Value (0008,0100) 1C See Section 8.1. Required if a *** not Not
sequence item is present. supported *** supported
>> Coding (0008,0102) 1C See Section 8.2. Required if a *** not Not
Scheme sequence item is present. supported *** supported
Designator
>> Coding (0008,0103) 1C See Section 8.2. Required if a *** not Not
Scheme sequence item is present and supported *** supported
Version the value of Coding Scheme
Designator (0008,0102) is not
sufficient to identify the Code
Value (0008,0100)
unambiguously.
>> Code (0008,0104) 1C See Section 8.3. Required if a *** not Not
Meaning sequence item is present. supported *** supported
> Person’s (0040,1102) 3 Person's mailing address *** not Not
Address supported *** supported
> Person’s (0040,1103) 3 Person's telephone number(s) *** not Not
Telephone supported *** supported
Numbers
> Institution (0008,0080) 1C Institution or organization to *** not Not
Name which the identified individual supported *** supported
is responsible or accountable.
Shall not be present if
Institution Code Sequence
(0008,0082) is present.
> Institution (0008,0081) 3 Mailing address of the *** not Not
Address institution or organization to supported *** supported
which the identified individual
is responsible or accountable.
> Person (0040,1101) 1 A coded entry which identifies *** not Not
Identification a person. The Code Meaning supported *** supported
Code Sequence attribute, though it will be
encoded with a VR of LO, may
be encoded according to the
rules of the PN VR (e.g. caret
'^' delimiters shall separate
name components), except
that a single component (i.e.
the whole name unseparated
by caret delimiters) is not
permitted. Name component
groups for use with multi-byte
character sets are permitted,
as long as they fit within the 64
characters (the length of the
LO VR). One or more Items
may be permitted in this
Sequence.
>> Code Value (0008,0100) 1C See Section 8.1. Required if a *** not Not
sequence item is present. supported *** supported
>> Coding (0008,0102) 1C See Section 8.2. Required if a *** not Not
Scheme sequence item is present. supported *** supported
Designator
>> Coding (0008,0103) 1C See Section 8.2. Required if a *** not Not
Scheme sequence item is present and supported *** supported
Version the value of Coding Scheme
Designator (0008,0102) is not
sufficient to identify the Code
Value (0008,0100)
unambiguously.
>> Code (0008,0104) 1C See Section 8.3. Required if a *** not Not
Meaning sequence item is present. supported *** supported
> Person’s (0040,1102) 3 Person's mailing address *** not Not
Address supported *** supported
> Person’s (0040,1103) 3 Person's telephone number(s) *** not Not
Telephone supported *** supported
Numbers
> Institution (0008,0080) 1C Institution or organization to *** not Not
Name which the identified individual supported *** supported
is responsible or accountable.
Shall not be present if
Institution Code Sequence
(0008,0082) is present.
> Institution (0008,0081) 3 Mailing address of the *** not Not
Address institution or organization to supported *** supported
which the identified individual
is responsible or accountable.
> Person (0040,1101) 1 A coded entry which identifies *** not Not
Identification a person. The Code Meaning supported *** supported
Code Sequence attribute, though it will be
encoded with a VR of LO, may
be encoded according to the
rules of the PN VR (e.g. caret
'^' delimiters shall separate
name components), except
that a single component (i.e.
the whole name unseparated
by caret delimiters) is not
permitted. Name component
groups for use with multi-byte
character sets are permitted,
as long as they fit within the 64
characters (the length of the
LO VR). One or more Items
may be permitted in this
Sequence.
>> Code Value (0008,0100) 1C See Section 8.1. Required if a *** not Not
sequence item is present. supported *** supported
>> Coding (0008,0102) 1C See Section 8.2. Required if a *** not Not
Scheme sequence item is present. supported *** supported
Designator
>> Coding (0008,0103) 1C See Section 8.2. Required if a *** not Not
Scheme sequence item is present and supported *** supported
Version the value of Coding Scheme
Designator (0008,0102) is not
sufficient to identify the Code
Value (0008,0100)
unambiguously.
>> Code (0008,0104) 1C See Section 8.3. Required if a *** not Not
Meaning sequence item is present. supported *** supported
> Person’s (0040,1102) 3 Person's mailing address *** not Not
Address supported *** supported
> Person’s (0040,1103) 3 Person's telephone number(s) *** not Not
Telephone supported *** supported
Numbers
> Institution (0008,0080) 1C Institution or organization to *** not Not
Name which the identified individual supported *** supported
is responsible or accountable.
Shall not be present if
Institution Code Sequence
(0008,0082) is present.
> Institution (0008,0081) 3 Mailing address of the *** not Not
Address institution or organization to supported *** supported
which the identified individual
is responsible or accountable.
> Coding (0008,0103) 1C See Section 8.2. Required if a *** not Not
Scheme sequence item is present and supported *** supported
Version the value of Coding Scheme
Designator (0008,0102) is not
sufficient to identify the Code
Value (0008,0100)
unambiguously.
> Code (0008,0104) 1C See Section 8.3. Required if a *** not Not
Meaning sequence item is present. supported *** supported
TRUEBEAM 4DITC
Attribute Name Tag Type Attribute Description
4.0 v13.0
> Plan Integrity (3287,XX01 1 Hash value calculated from Roundtrip Roundtrip
Hash ) selected data of this RT Plan supported supported
along a specific algorith whose
version number is specified in
Plan Integrity Hash Version
(3287,xx02).
> Plan Integrity (3287,XX02 1 Version of hash algorithm used Roundtrip Roundtrip
Hash Version ) to calculate value of Plan supported supported
Integrity Hash (3287,xx01).
> Delivery (300A,0023) 3 The maximum dose (in Gy) Roundtrip Roundtrip
Maximum Dose which can be delivered to the supported supported
dose reference.
> Target (300A,0025) 3 Minimum permitted dose (in *** not Roundtrip
Minimum Dose Gy) to Dose Reference if Dose supported supported
Reference Type (300A,0020) is ***
TARGET.
> Target (300A,0026) 3 Prescribed dose (in Gy) to Roundtrip Roundtrip
Prescription Dose Reference if Dose supported supported
Dose Reference Type (300A,0020) is
TARGET.
> Target (300A,0027) 3 Maximum permitted dose (in Roundtrip Roundtrip
Maximum Dose Gy) to Dose Reference if Dose supported supported
Reference Type (300A,0020) is
TARGET.
> Referenced (3267,XX00 3 *** not Not
Patient Volume ) supported ** supported
Id
> Target (300A,0028) 3 Maximum permitted fraction (in *** not Not
Underdose percent) of Target to receive supported supported
Volume less than the Target ***
Fraction Prescription Dose if Dose
Reference Type (300A,0020) is
TARGET and Dose Reference
Structure Type (300A,0014) is
VOLUME. See C.8.8.10.1.
> Organ at Risk (300A,002A 3 Maximum dose (in Gy) to entire *** not Not
Full-volume ) Dose Reference if Dose supported supported
Dose Reference Type (300A,0020) is ***
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) is VOLUME.
> Organ at Risk (300A,002B 3 Maximum permitted dose (in Roundtrip Roundtrip
Limit Dose ) Gy) to any part of Dose supported supported
Reference if Dose Reference
Type (300A,0020) is
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) is VOLUME.
> Organ at Risk (300A,002C 3 Maximum dose (in Gy) to non- Roundtrip Roundtrip
Maximum Dose ) overdosed part of Dose supported supported
Reference if Dose Reference
Type (300A,0020) is
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) is VOLUME.
> Organ at Risk (300A,002D 3 Maximum permitted fraction (in Roundtrip Roundtrip
Overdose ) percent) of the Organ at Risk to supported supported
Volume receive more than the Organ at
Fraction Risk Maximum Dose if Dose
Reference Type (300A,0020) is
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) is VOLUME.
TrueBeam 4DITC
Attribute Name Tag Type Attribute Description
4.0 v13.0
Tolerance Table (300A,0040) 3 Introduces sequence of Roundtrip Roundtrip
Sequence tolerance tables to be used for supported supported
delivery of treatment plan. One
or more items may be included
in this sequence. See Note 1.
> Tolerance (300A,0042) 1C Identification number of the Roundtrip Roundtrip
Table Number Tolerance Table. The value of supported. supported.
Tolerance Table Number
(300A,0042) shall be unique
within the RT Plan in which it is
created. Required if Tolerance
Table Sequence (300A,0040) is
sent.
> Tolerance (300A,0043) 3 User-defined label for Roundtrip Roundtrip
Table Label Tolerance Table. supported supported.
> Gantry Angle (300A,0044) 3 Maximum permitted difference Roundtrip Roundtrip
Tolerance (in degrees) between planned supported supported.
and delivered Gantry Angle.
> Gantry Pitch (300A,014E 3 Maximum permitted difference *** not *** not
Angle Tolerance ) (in degrees) between planned supported supported
and delivered Gantry Pitch *** ***
Angle.
> Beam Limiting (300A,0046) 3 Maximum permitted difference Roundtrip Roundtrip
Device Angle (in degrees) between planned supported supported
Tolerance and delivered Beam Limiting
Device Angle.
> Beam Limiting (300A,0048) 3 Introduces sequence of beam Roundtrip Roundtrip
Device limiting device (collimator) supported supported
Tolerance tolerances. One or more items
Sequence may be included in this
sequence.
>> Fixation (300A,0192) 1C Type of Fixation Device used *** not *** not
Device Type during in Patient Setup. supported supported
Required if Fixation Device *** ***
Sequence (300A,0190) is sent.
Defined Terms:
BITEBLOCK
HEADFRAME
MASK
MOLD
CAST
HEADREST
BREAST_BOARD
BODY_FRAME
VACUUM_MOLD
WHOLE_BODY_POD
RECTAL_BALLOON
>> Fixation (300A,0194) 2C User-defined label identifier for *** not *** not
Device Label Fixation Device. Required if supported supported
Fixation Device Sequence *** ***
(300A,0190) is sent.
>> Fixation (300A,0196) 3 User-defined description of *** not *** not
Device Fixation Device. supported supported
Description *** ***
>> Fixation (300A,0198) 3 Position/Notch number of *** not *** not
Device Position Fixation Device. supported supported
*** ***
>> Fixation (300A,0199) 3 The Fixation Device Pitch *** not *** not
Device Pitch Angle, i.e. orientation of supported supported
Angle PITCHED FIXATION DEVICE *** ***
coordinate system with respect
to IEC PATIENT SUPPORT
coordinate system (degrees).
Pitching is the rotation around
IEC PATIENT SUPPORT X-
axis.
>> Fixation (300A,019A) 3 The Fixation Device Roll Angle, *** not *** not
Device Roll i.e. orientation of ROLLED supported supported
Angle FIXATION DEVICE coordinate *** ***
system with respect to IEC
PITCHED FIXATION DEVICE
coordinate system (degrees).
Rolling is the rotation around
IEC PATIENT SUPPORT Y-
axis.
>> Accessory (300A,00F9) 3 An identifier for the accessory *** not *** not
Code intended to be read by a device supported supported
such as a bar-code reader. *** ***
78 of 151 Varian Treatment Delivery
Console
IOD Details, RT Patient Setup – C.8.8.12
Attribute Name Tag Type Attribute Description TrueBeam 4DITC
4.0 v13.0
>> Setup (300A,01B6) 1C Type of Setup Device used for *** not *** not
Device Type Patient alignment. Required if supported supported
Setup Device Sequence *** ***
(300A,01B4) is sent.
Defined Terms:
LASER_POINTER
DISTANCE_METER
TABLE_HEIGHT
MECHANICAL_PTR
ARC
>> Setup (300A,01B8) 2C User-defined label for Setup *** not *** not
Device Label Device used for patient supported supported
alignment. Required if Setup *** ***
Device Sequence (300A,01B4)
is sent.
>> Setup (300A,01BA) 3 User-defined description for *** not *** not
Device Setup Device used for patient supported supported
Description alignment. *** ***
>> Setup (300A,01BC) 2C Setup Parameter for Setup *** not *** not
Device Device in appropriate IEC supported supported
Parameter 61217 coordinate system. Units *** ***
shall be mm for distances and
degrees for angles. Required if
Setup Device Sequence
(300A,011B4) is sent.
>> Setup (300A,01D0) 3 User-defined description of *** not *** not
Reference Setup Reference used for supported supported
Description patient alignment. *** ***
> Table Top (300A,01D2) 3 Vertical Displacement in IEC Roundtrip Roundtrip
Vertical Setup TABLE TOP coordinate system supported supported
Displacement (in mm) relative to initial Setup
Position, i.e. vertical offset
between patient positioning
performed using setup and
treatment position.
> Table Top (300A,01D4) 3 Longitudinal Displacement in Roundtrip Roundtrip
Longitudinal IEC TABLE TOP coordinate supported supported
Setup system (in mm) relative to initial
Displacement Setup Position, i.e. longitudinal
offset between patient
positioning performed using
setup and treatment position.
>> Referenced (300C,0051) 1C Uniquely identifies Dose *** not *** not
Dose Reference Reference specified by Dose supported supported
Number Reference Number (300A,0012) *** ***
within Dose Reference
Sequence (300A,0010) in RT
Prescription Module. Required if
Referenced Dose Reference
Sequence (300C,0050) is sent.
>> Constraint (300A,0021) 3 Relative importance of *** not *** not
Weight satisfying constraint, where high supported supported
values represent more *** ***
important constraints.
>> Delivery (300A,0022) 3 The dose (in Gy) which when *** not *** not
Warning Dose reached or exceeded should supported supported
cause some action to be taken. *** ***
>> Delivery (300A,0023) 3 The maximum dose (in Gy) *** not *** not
Maximum Dose which can be delivered to the supported supported
dose reference. *** ***
>> Target (300A,0025) 3 Minimum permitted dose (in Gy) *** not *** not
Minimum Dose to Dose Reference if Dose supported supported
Reference Type (300A,0020) of *** ***
referenced Dose Reference is
TARGET.
>> Target (300A,0026) 3 Prescribed dose (in Gy) to Dose *** not *** not
Prescription Reference if Dose Reference supported supported
Dose Type (300A,0020) of referenced *** ***
Dose Reference is TARGET.
>> Target (300A,0027) 3 Maximum permitted dose (in *** not *** not
Maximum Dose Gy) to Dose Reference if Dose supported supported
Reference Type (300A,0020) of *** ***
referenced Dose Reference is
TARGET.
>> Target (300A,0028) 3 Maximum permitted fraction (in *** not *** not
Underdose percent) of Target to receive supported supported
Volume less than the Target *** ***
Fraction Prescription Dose (300A,0027)
if Dose Reference Type
(300A,0020) of referenced
Dose Reference is TARGET
and Dose Reference Structure
Type (300A,0014) of referenced
Dose Reference is VOLUME.
>> Organ at (300A,002A) 3 Maximum dose (in Gy) to entire *** not *** not
Risk Full- Dose Reference if Dose supported supported
volume Dose Reference Type (300A,0020) of *** ***
referenced Dose Reference is
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) of referenced
Dose Reference is VOLUME.
>> Organ at (300A,002B) 3 Maximum permitted dose (in *** not *** not
Risk Limit Dose Gy) to any part of Dose supported supported
Reference if Dose Reference *** ***
Type (300A,0020) of referenced
Dose Reference is
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) of referenced
Dose Reference is VOLUME.
>> Organ at (300A,002C) 3 Maximum dose (in Gy) to non- *** not *** not
Risk Maximum overdosed part of Dose supported supported
Dose Reference if Dose Reference *** ***
Type (300A,0020) of referenced
Dose Reference is
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) of referenced
Dose Reference is VOLUME.
>> Organ at (300A,002D) 3 Maximum permitted fraction (in *** not *** not
Risk Overdose percent) of Organ at Risk to supported supported
Volume receive more than the Organ at *** ***
Fraction Risk Maximum Dose if Dose
Reference Type (300A,0020) of
referenced Dose Reference is
ORGAN_AT_RISK and Dose
Reference Structure Type
(300A,0014) of referenced
Dose Reference is VOLUME.
> Number of (300A,0078) 2 Total number of treatments Roundtrip Roundtrip
Fractions (Fractions) prescribed for supported supported
Planned current Fraction Group.
> Number of (300A,0079) 3 Number of digits in Fraction *** not Roundtrip
Fraction Pattern Pattern (300A,007B) used to supported supported
Digits Per Day represent one day. See Note 2. ***
> Repeat (300A,007A) 3 Number of weeks needed to *** not Roundtrip
Fraction Cycle describe treatment pattern. See supported supported
Length Note 2. ***
> Fraction (300A,007B) 3 String of 0's (no treatment) and Roundtrip Roundtrip
Pattern 1's (treatment) describing supported supported
treatment pattern. Length of
string is 7 x Number of Fraction
Pattern Digits Per Day x Repeat
Fraction Cycle Length. Pattern
shall start on a Monday. See
Note 2.
> Number of (300A,0080) 1 Number of Beams in current Import only Import only
Beams Fraction Group. If Number of
Beams is greater then zero,
Number of Brachy Application
Setups (300A,00A0) shall equal
zero.
> Referenced (300C,0004) 1C Introduces sequence of Roundtrip Roundtrip
Beam treatment beams in current supported supported
Sequence Fraction Group. Required if
Number of Beams (300A,0080)
is greater than zero. One or
more items may be included in
this sequence.
>> Referenced (300C,0006) 1C Uniquely identifies Beam Roundtrip Roundtrip
Beam Number specified by Beam Number supported supported
(300A,00C0) within Beam
Sequence (300A,00B0) in RT
Beams Module. Required if
Referenced Beam Sequence
(300C,0004) is sent.
>> Beam Dose (300A,0082) 3 Coordinates (x,y,z) of point at Roundtrip Roundtrip
Specification which Beam Dose is specified supported supported
Point in the patient based coordinate
system described in C.7.6.2.1.1
(mm). See Note 3.
>> Beam Dose (300A,0084) 3 Dose (in Gy) at Beam Dose Roundtrip Roundtrip
Specification Point (300A,0082) supported supported
due to current Beam.
>> Beam (300A,0086) 3 Machine setting to be delivered Roundtrip Roundtrip
Meterset for current Beam, specified in supported supported
Monitor Units (MU) or minutes
as defined by Primary
Dosimeter Unit (300A,00B3) (in
RT Beams Module) for
referenced Beam. See Note 4.
>> Maximum (3249,XX00) 3 The maximum treatment time, *** not Roundtrip
Treatment Time which should not be supported supported
exceeeded. This items serves ***
as a secondary safety limit
besides the Beam Meterset
value (300A,0086) itself. Units
in [min].
> Number of (300A,00A0) 1 Number of Brachy Application *** not *** not
Brachy Setups in current Fraction supported supported
Application Group. If Number of Brachy *** ***
Setups Application Setups is greater
then zero, Number of Beams
(300A,0080) shall equal zero.
> Referenced (300C,000A) 1C Introduces sequence of *** not *** not
Brachy treatment Brachy Application supported supported
Application Setups in current Fraction *** ***
Setup Group. Required if Number of
Sequence Brachy Application Setups
(300A,00A0) is greater than
zero. One or more items may
be included in this sequence.
>> Referenced (300C,000C) 1C Uniquely identifies Brachy *** not *** not
Brachy Application Setup specified by supported supported
Application Brachy Application Setup *** ***
Setup Number Number (300A,0234) within
Brachy Application Setup
Sequence (300A,0230) in RT
Brachy Application Setups
Module.
Required if Referenced Brachy
Application Setup Sequence
(300C,000A) is sent.
>> Brachy (300A,00A2) 3 Coordinates (x,y,z) of point in *** not *** not
Application the patient based coordinate supported supported
Setup Dose system described in C.7.6.2.1.1 *** ***
Specification at which Brachy Application
Point Setup Dose (300A,00A4) is
specified (mm).
>> Brachy (300A,00A4) 3 Dose (in Gy) at Brachy *** not *** not
Application Application Setup Dose supported supported
Setup Dose Specification Point (300A,00A2) *** ***
due to current Brachy
Application Setup.
> Prescribed (3269,XX00) 3 *** not *** not
Dose per supported supported
Fraction *** ***
> Prescribed (3269,XX01) 3 *** not *** not
Percentage supported supported
*** ***
TrueBeam 4DITC
Attribute Name Tag Type Attribute Description
4.0 v13.0
Beam Sequence (300A,00B0) 1 Introduces sequence of Roundtrip Roundtrip
treatment beams for current RT supported supported
Plan. One or more items may
be included in this sequence.
> Beam Number (300A,00C0) 1 Identification number of the Roundtrip Roundtrip
Beam. The value of Beam supported supported
Number (300A,00C0) shall be
unique within the RT Plan in
which it is created. See Note 1.
> Beam Name (300A,00C2) 3 User-defined name for Beam. Roundtrip Roundtrip
See Note 1. supported supported
> Beam (3243,XX09) 3 User-defined secondary name Roundtrip Roundtrip
Seconary Name for Beam. supported supported
> Beam (300A,00C3) 3 User-defined description for Roundtrip Roundtrip
Description Beam. See Note 1. supported supported
> Beam Type (300A,00C4) 1 Motion characteristic of Beam. Roundtrip Roundtrip
See Note 5. supported supported
Enumerated Values:
STATIC = All Control Point
Sequence (300A,0111)
attributes remain unchanged
between consecutive pairs of
control points with changing
Cumulative Meterset Weight
(300A,0134).
DYNAMIC = One or more
Control Point Sequence
(300A,0111) attributes change
between one or more
consecutive pairs of control
points with changing
Cumulative Meterset Weight
(300A,0134).
>> Dose Rate Set (300A,0115) 3 Dose Rate to be set on Roundtrip Roundtrip
treatment machine for segment supported supported
beginning at current control
point (e.g. MU/min).
>> Referenced (0008,1150) 1 Uniquely identifies the *** not *** not
SOP Class UID referenced SOP Class. supported supported
*** ***
>> Referenced (0008,1155) 1 Uniquely identifies the *** not *** not
SOP Instance referenced SOP Instance. supported supported
UID *** ***
Approval Status (300E,0002) 1 Approval status at the time the Roundtrip Roundtrip
SOP Instance was created. supported supported
Enumerated Values:
APPROVED = Reviewer
recorded that object met an
implied criterion
UNAPPROVED = No review of
object has been recorded
REJECTED = Reviewer
recorded that object failed to
meet an implied criterion
Review Date (300E,0004) 2C Date on which object was Roundtrip Roundtrip
reviewed. supported supported
Required if Approval Status
(300E,0002) is APPROVED or
REJECTED.
Review Time (300E,0005) 2C Time at which object was Roundtrip Roundtrip
reviewed. supported supported
Required if Approval Status
(300E,0002) is APPROVED or
REJECTED.
Reviewer Name (300E,0008) 2C Name of person who reviewed Roundtrip Roundtrip
object. supported supported
Required if Approval Status
(300E,0002) is APPROVED or
REJECTED.
Treatment Date (3008,0250) 2 Date when current fraction was Supported Supported
delivered, or Date last fraction
was delivered in case of RT
Treatment Summary Record
IOD. See Note.
Treatment Time (3008,0251) 2 Time when current fraction was Supported Supported
delivered (begun), or Time last
fraction was delivered (begun)
in case of RT Treatment
Summary Record IOD.
Referenced RT (300C,0002) 2 Reference to a RT Plan. Supported Supported
Plan Sequence One item shall be included in
this sequence.
>Include Table 10-11 “SOP Instance Reference Macro Attributes”
Referenced (3008,0030) 3 Reference to RT Treatment Supported Supported
Treatment Records to which the current
Record RT Treatment Record is
Sequence significantly related.
>Include Table 10-11 “SOP Instance Reference Macro Attributes”
>>Wedge Type (300A,00D3) 2 Type of wedge defined for delivered Supported Supported
Beam.
Defined Terms:
STANDARD standard (static)
wedge
DYNAMIC moving Beam
Limiting Device (collimator) jaw
simulating wedge
MOTORIZED single wedge that can
be removed from beam remotely
>Number of Boli (300A,00ED) 2 Number of boli used with current Supported Supported
Beam.
>>Block Tray ID (300A,00F5) 3 User-supplied identifier for block tray Supported Supported
or Electron Insert.
>>Accessory Code (300A,00F9) 3 An identifier for the Block intended to Supported Supported
be read by a device such as a bar-
code reader.
>Current Fraction (3008,0022) 2 Fraction number for this beam Supported Supported
Number administration.
> Treatment (300A,002B) 2 The reason code why the current Supported Supported
Termination Code treatment beam has terminated.
Expected if the Treatment
Termination Status (3008,002A) is not
NORMAL.
Enumerated Values:
OPERATOR operator terminated
treatment
FAULT machine terminated
treatment
UNKNOWN status at termination
unknown
*** not
>>Interlock User-readable description of the
(300A,0742) 1 Supported supported
Description interlock.
***
*** not
Text that conveys the meaning of the
>>>Code Meaning (0008,0104) 1C Supported supported
Coded Entry.
***
*** not
Text that conveys the meaning of the
>>>Code Meaning (0008,0104) 1C Supported supported
Coded Entry.
***
*** not
>>>>Code Text that conveys the meaning of the
(0008,0104) 1 Supported supported
Meaning Coded Entry.
***
>>Beam Hold (300C,0127) 1 DateTime of the beam hold transition. *** not
Transition Supported supported
DateTime ***
*** not
>>>Long Device User-defined description for this
(0050,0021) 3 Supported supported
Description device.
***
*** not
>>>Device An identifier intended to be read by a
(3010,001B) 2 Supported supported
Alternate Identifier device such as a bar code reader.
***
*** not
>>>Device Label (3010,002D) 1 User-defined label for this device Supported supported
***
*** not
>>>>Code Text that conveys the meaning of the
(0008,0104) 1 Supported supported
Meaning Coded Entry.
***
*** not
>>> Manufacturer’s An identifier issued by the
(3040,0043) 2 Supported supported
Device Identifier manufacturer.
***
>Control Point (3008,0040) 1 Sequence of beam control points for Supported Supported
Delivery Sequence current treatment beam.
The number of Items in this Sequence
shall equal the value of Number of
Control Points (300A,0110).
>>Dose Rate Set (300A,0115) 2 Dose Rate set on treatment machine Supported Supported
for segment beginning at current
control point (Meterset/min).
>>Dose Rate (3008,0048) 2 Dose Rate actually delivered for Supported Supported
Delivered segment beginning at current control
point (Meterset/min).
>>Nominal Beam (300A,0015) 1C Units used for Nominal Beam Energy Supported Supported
Energy Unit (300A,0114). Required if Nominal
Beam Energy (300A,0114) is present.
Defined Terms:
MV Megavolt
MEV Mega electron-Volt
If Radiation Type (300A,00C6) is
PHOTON, Nominal Beam Energy Unit
(300A,0015) shall be MV. If Radiation
Type (300A,00C6) is ELECTRON,
Nominal Beam Energy Unit
126 of 151 Varian Treatment Delivery
Console
IOD Details, RT Beams Session Record – C.8.8.21
Attribute Name Tag Type Attribute Description TrueBeam 4DITC
4.0 v13.0
(300A,0015) shall be MEV.
>>>RT Beam (300A,00B8) 1 Type of beam limiting device. The Supported Supported
Limiting Device value of this Attribute shall
Type correspond to RT Beam Limiting
Device Type (300A,00B8) defined in
an element of Beam Limiting Device
Leaf Pairs Sequence (3008,00A0).
Enumerated Values:
X symmetric jaw pair in IEC X
direction
Y symmetric jaw pair in IEC Y
direction
ASYMX asymmetric jaw pair in IEC X
direction
ASYMY asymmetric jaw pair in IEC Y
direction
MLCX multileaf (multi-element) jaw
pair in IEC X direction
MLCY multileaf (multi-element) jaw
pair in IEC Y direction
>>Gantry Angle (300A,011E) 1C Treatment machine gantry angle, i.e., Supported Supported
orientation of IEC GANTRY
coordinate system with respect to IEC
FIXED REFERENCE coordinate
system (degrees). Required for
Control Point 0 of Control Point
Delivery Sequence (3008,0040) or if
Gantry Angle changes during beam
administration.
>>Patient Support (300A,0122) 1C Patient Support angle, i.e., orientation Supported Supported
Angle of IEC PATIENT SUPPORT
(turntable) coordinate system with
respect to IEC FIXED REFERENCE
coordinate system (degrees).
Required for Control Point 0 of
Control Point Delivery Sequence
(3008,0040) or if Patient Support
Angle changes during beam
administration.
>>Table Top (300A,0124) 3 Distance (positive) from the IEC Supported Supported
Eccentric Axis PATIENT SUPPORT vertical axis to
Distance the IEC TABLE TOP ECCENTRIC
vertical axis (mm).
>>Table Top (300A,0125) 1C Table Top (non-isocentric) angle, i.e., Supported Supported
Eccentric Angle orientation of IEC TABLE TOP
ECCENTRIC coordinate system with
respect to IEC PATIENT SUPPORT
coordinate system (degrees).
Required for Control Point 0 of
Control Point Delivery Sequence
(3008,0040) or if Table Top Eccentric
Angle changes during beam
administration.
>>Table Top Pitch (300A,0140) 1C Table Top Pitch Angle, i.e., the Supported Supported
Angle rotation of the IEC TABLE TOP
coordinate system about the X-axis of
the IEC TABLE TOP coordinate
system (degrees). If required by
treatment delivery device, shall be
present for first Item of Control Point
Sequence. If required by treatment
delivery device and if Table Top Pitch
Angle changes during Beam, shall be
present in all subsequent Items of
Control Point Sequence.
>>Table Top Pitch (300A,0142) 1C Direction of Table Top Pitch Rotation Supported Supported
Rotation Direction when viewing the table along the
positive X-axis of the IEC TABLE
TOP coordinate system, for segment
following Control Point. If required by
treatment delivery device, shall be
present for first Item of Control Point
Sequence. If required by treatment
delivery device and if Table Top Pitch
Rotation Direction changes during
Beam, shall be present in all
subsequent Items of Control Point
Sequence.
Enumerated Values:
CW clockwise
CC counter-clockwise
NONE no rotation
>>Table Top Roll (300A,0144) 1C Table Top Roll Angle, i.e., the rotation Supported Supported
Angle of the IEC TABLE TOP coordinate
system about the IEC Y-axis of the
IEC TABLE TOP coordinate system
(degrees). If required by treatment
delivery device, shall be present for
first Item of Control Point Sequence. If
required by treatment delivery device
and if Table Top Roll Angle changes
during Beam, shall be present in all
subsequent Items of Control Point
>>Table Top Roll (300A,0146) 1C Direction of Table Top Roll Rotation Supported Supported
Rotation Direction when viewing the table along the
positive Y-axis of the IEC TABLE
TOP coordinate system, for segment
following Control Point. If required by
treatment delivery device, shall be
present for first Item of Control Point
Sequence. If required by treatment
delivery device and if Table Top Roll
Rotation Direction changes during
Beam, shall be present in all
subsequent Items of Control Point
Sequence.
Enumerated Values:
CW clockwise
CC counter-clockwise
NONE no rotation
>>Table Top (300A,0128) 2C Table Top Vertical position in IEC Supported Supported
Vertical Position TABLE TOP coordinate system (mm).
This value is interpreted as an
absolute, rather than relative, Table
setting. Required for Control Point 0
of Control Point Delivery Sequence
(3008,0040) or if Table Top Vertical
Position changes during beam
administration.
>>Table Top (300A,0129) 2C Table Top Longitudinal position in IEC Supported Supported
Longitudinal TABLE TOP coordinate system (mm).
Position This value is interpreted as an
absolute, rather than relative, Table
setting. Required for Control Point 0
of Control Point Delivery Sequence
(3008,0040) or if Table Top
Longitudinal Position changes during
beam administration.
>>Table Top (300A,012A) 2C Table Top Lateral position in IEC Supported Supported
Lateral Position TABLE TOP coordinate system (mm).
This value is interpreted as an
absolute, rather than relative, Table
setting. Required for Control Point 0
of Control Point Delivery Sequence
(3008,0040) or if Table Top Lateral
Position changes during beam
administration.
>>>Override (3008,0062) 2 Contains the Data Element Tag of the Supported Supported
Parameter Pointer Attribute that was overridden.
SOP Class UID (0008,0016) 1 Uniquely identifies the SOP Roundtrip Roundtrip
Class. See C.12.1.1.1 for supported supported
further explanation. See also
PS 3.4.
SOP Instance (0008,0018) 1 Uniquely identifies the SOP Roundtrip Roundtrip
UID Instance. See C.12.1.1.1 for supported supported
further explanation. See also
PS 3.4.
Specific (0008,0005) 1C Character Set that expands or Roundtrip Roundtrip
Character Set replaces the Basic Graphic Set. supported supported
Required if an expanded or
replacement character set is
used. See C.12.1.1.2 for
Defined Terms.
Instance (0008,0012) 3 Date the SOP Instance was Roundtrip Roundtrip
Creation Date created. supported supported
Related General (0008,001A) 3 Uniquely identifies a Related *** not *** not
SOP Class UID General SOP Class for the SOP supported supported
Class of this Instance. See PS *** ***
3.4.
Original (0008,001B) 3 The SOP Class in which the *** not *** not
Specialized SOP Instance was originally supported supported
Class UID encoded, but which has been *** ***
replaced during a fall-back
conversion to the current
Related General SOP Class.
See PS 3.4.
Coding Scheme (0008,0110) 3 Sequence of items that map *** not *** not
Identification values of Coding Scheme supported supported
Sequence Designator (0008,0102) to an *** ***
external coding system
registration, or to a private or
local coding scheme. One or
more items may be present in
the sequence.
> Coding (0008,0102) 1 The value of a Coding Scheme *** not *** not
Scheme Designator, used in this SOP supported supported
Designator Instance, which is being *** ***
mapped.
> Coding (0008,0112) 1C The name of the external *** not *** not
Scheme Registry registry where further definition supported supported
of the identified coding scheme *** ***
may be obtained. Required if
coding scheme is registered.
Defined term:
HL7
> Coding (0008,010C) 1C The coding scheme UID *** not *** not
Scheme UID identifier. supported supported
Required if coding scheme is *** ***
identified by an ISO 8824 object
identifier compatible with the UI
VR.
> Coding (0008,0114) 2C The coding scheme identifier as *** not *** not
Scheme External defined in an external registry. supported supported
ID Required if coding scheme is *** ***
registered and Coding Scheme
UID (0008,010C) is not present.
> Coding (0008,0115) 3 The coding scheme full *** not *** not
Scheme Name common name supported supported
*** ***
> Coding (0008,0103) 3 The coding scheme version *** not *** not
Scheme Version associated with the Coding supported supported
Scheme Designator *** ***
(0008,0102).
> Responsible (0008,0116) 3 Name of the organization *** not *** not
Organization responsible for the Coding supported supported
Scheme. May include *** ***
organizational contact
information.
Timezone Offset (0008,0201) 3 Contains the offset from UTC to *** not *** not
From UTC the timezone for all DA and TM supported supported
Attributes present in this SOP *** ***
Instance. Encoded as an ASCII
string in the format "&ZZZZ".
The components of this string,
from left to right, are & = "+" or
"-", and ZZZZ = Hours and
Minutes of offset.
Notes:
1. This encoding is the same as
described in PS 3.5 for the DT
Value Representation.
2. This Attribute does not apply
to values with a DT Value
Representation, which may
contain an explicitly encoded
timezone.
3. The corrected time may
cross a 24 hour boundary. For
example, if Local Time = 1.00
a.m. and Offset = +0200, then
UTC = 11.00 p.m. (23.00) the
day before.
4. The "+" sign may not be
omitted. Time earlier than UTC
is expressed as a negative
offset.
Note:
For example: UTC = 5.00 a.m.
Local Time = 3.00 a.m. Offset =
-0200 The local timezone offset
is undefined if this Attribute is
absent.
Contributing (0018,A001) 3 Sequence of Items containing *** not *** not
Equipment descriptive attributes of related supported supported
Sequence equipment which has *** ***
contributed to the acquisition,
creation or modification of the
composite instance. One or
more Items may be included in
this Sequence. See C.12.1.1.5
for further explanation.
> Purpose of (0040,A170) 1 Describes the purpose for *** not *** not
Reference Code which the related equipment is supported supported
Sequence being reference. Only a single *** ***
Item shall be permitted in this
sequence. See C.12.1.1.5 for
further explanation.
>> Code Value (0008,0100) 1C See Section 8.1. Required if a *** not *** not
sequence item is present. supported supported
*** ***
Treatment Delivery Console 135 of 151
IOD Details, SOP Common – C.12.1
Attribute Name Tag Type Attribute Description TrueBeam 4DITC
4.0 v13.0
>> Coding (0008,0102) 1C See Section 8.2. Required if a *** not *** not
Scheme sequence item is present. supported supported
Designator *** ***
>> Coding (0008,0103) 1C See Section 8.2. Required if a *** not *** not
Scheme Version sequence item is present and supported supported
the value of Coding Scheme *** ***
Designator (0008,0102) is not
sufficient to identify the Code
Value (0008,0100)
unambiguously.
>> Code (0008,0104) 1C See Section 8.3. Required if a *** not *** not
Meaning sequence item is present. supported supported
*** ***
> Manufacturer (0008,0070) 1 Manufacturer of the equipment *** not *** not
that contributed to the supported supported
composite instance. *** ***
> Institution (0008,0080) 3 Institution where the equipment *** not *** not
Name that contributed to the supported supported
composite instance is located. *** ***
> Institution (0008,0081) 3 Address of the institution where *** not *** not
Address the equipment that contributed supported supported
to the composite instance is *** ***
located.
> Station Name (0008,1010) 3 User defined name identifying *** not *** not
the machine that contributed to supported supported
the composite instance. *** ***
> Institutional (0008,1040) 3 Department in the institution *** not *** not
Department where the equipment that supported supported
Name contributed to the composite *** ***
instance is located.
> Manufacturer’s (0008,1090) 3 Manufacturer's model name of *** not *** not
Model Name the equipment that contributed supported supported
to the composite instance. *** ***
> Device Serial (0018,1000) 3 Manufacturer's serial number of *** not *** not
Number the equipment that contributed supported supported
to the composite instance. *** ***
> Software (0018,1020) 3 Manufacturer's designation of *** not *** not
Version(s) the software version of the supported supported
equipment that contributed to *** ***
the composite instance.
> Spatial (0018,1050) 3 The inherent limiting resolution *** not *** not
Resolution in mm of the acquisition supported supported
equipment for high contrast *** ***
objects for the data gathering
and reconstruction technique
chosen. If variable across the
images of the series, the value
at the image center.
> Date of Last (0018,1200) 3 Date when the image *** not *** not
Calibration acquisition device calibration supported supported
was last changed in any way. *** ***
Multiple entries may be used for
additional calibrations at other
times. See C.7.5.1.1.1 for
further explanation.
> Time of Last (0018,1201) 3 Time when the image *** not *** not
Calibration acquisition device calibration supported supported
was last changed in any way. *** ***
Multiple entries may be used.
See C.7.5.1.1.1 for further
explanation.
> Contribution (0018,A002) 3 The Date & Time when the *** not *** not
Date Time equipment contributed to the supported supported
composite instance. *** ***
> Contribution (0018,A003) 3 Description of the contribution *** not *** not
Description the equipment made to the supported supported
composite instance. *** ***
Instance Number (0020,0013) 3 A number that identifies this *** not *** not
Composite object instance. supported supported
*** ***
SOP Instance (0100,0410) 3 A flag that indicates the storage *** not *** not
Status status of the SOP Instance. Not supported supported
Specified (NS) implies that this *** ***
SOP Instance has no special
storage status, and hence no
special actions need be taken.
Original (OR) implies that this is
the primary SOP instance for
the purpose of storage, but that
it has not yet been authorized
for diagnostic use. Authorized
Original (AO) implies that this is
the primary SOP instance for
the purpose of storage, which
has been authorized for
diagnostic use. Any copies of
an Authorized Original should
be given the status of
Authorized Copy. Authorized
Copy (AC) implies that this is a
copy of an Authorized Original
SOP Instance.
Enumerated Values:
NS, OR, AO, AC
Note:
Proper use of these flags is
specified in Security Profiles.
Implementations that do not
conform to such Security
Profiles may not necessarily
handle these flags properly.
SOP (0100,0420) 3 The date and time when the *** not *** not
Authorization SOP Instance Status supported supported
Date and Time (0100,0410) was set to AO. *** ***
SOP (0100,0424) 3 Any comments associated with *** not *** not
Authorization the setting of the SOP Instance supported supported
Comment Status (0100,0410) to AO. *** ***
Authorization (0100,0426) 3 The certification number issued *** not *** not
Equipment to the Application Entity that set supported supported
Certification the SOP Instance Status *** ***
Number (0100,0410) to AO.
MAC Parameters (4FFE,0001) 3 A sequence of one or more *** not *** not
Sequence items that describe the supported supported
parameters used to calculate a *** ***
MAC for use in Digital
Signatures.
> MAC ID (0400,0005) 1 A number used to identify this *** not *** not
Number MAC Parameters Sequence supported supported
item. *** ***
> MAC (0400,0010) 1 The Transfer Syntax UID used *** not *** not
Calculation to encode the values of the supported supported
Transfer Syntax Data Elements included in the *** ***
UID MAC calculation. Only Transfer
Syntaxes that explicitly include
the VR and use Little Endian
encoding shall be used. Notes:
Certain Transfer Syntaxes,
particularly those that are used
with compressed data, allow the
fragmentation of the pixel data
to change. If such
fragmentation changes, Digital
Signatures generated with such
Transfer Syntaxes could
become invalid.
> MAC Algorithm (0400,0015) 1 The algorithm used in *** not *** not
generating the MAC to be supported supported
encrypted to form the Digital *** ***
Signature. Defined Terms:
RIPEMD160
MD5
SHA1. Note: Digital Signature
Security Profiles (see PS 3.15)
may require the use of a
restricted subset of these terms.
> Data Elements (0400,0020) 1 A list of Data Element Tags in *** not *** not
Signed the order they appear in the supported supported
Data Set which identify the Data *** ***
Elements used in creating the
MAC for the Digital Signature.
See Section C.12.1.1.3.1.1.
Digital (FFFA,FFFA) 3 Sequence holding one or more *** not *** not
Signatures Digital Signatures. supported supported
Sequence *** ***
> MAC ID (0400,0005) 1 A number used to identify which *** not *** not
Number MAC Parameters Sequence supported supported
item was used in the calculation *** ***
of this Digital Signature.
> Digital (0400,0100) 1 A UID that can be used to *** not *** not
Signature UID uniquely reference this supported supported
signature. *** ***
> Digital (0400,0105) 1 The date and time the Digital *** not *** not
Signature Signature was created. The supported supported
DateTime time shall include an offset (i.e., *** ***
time zone indication) from
Coordinated Universal Time.
Note: This is not a certified
timestamp, and hence is not
completely verifiable. An
application can compare this
date and time with those of
other signatures and the validity
date of the certificate to gain
confidence in the veracity of this
date and time.
> Certificate (0400,0110) 1 The type of certificate used in *** not *** not
Type (0400,0115). Defined Term: supported supported
X509_1993_SIG Note: Digital *** ***
Signature Security Profiles (see
PS 3.15) may require the use of
a restricted subset of these
terms.
> Certificate of (0400,0115) 1 A certificate that holds the *** not *** not
Signer identity of the entity producing supported supported
this Digital Signature, that *** ***
entity's public key or key
identifier, and the algorithm and
associated parameters with
which that public key is to be
used. Algorithms allowed are
specified in Digital Signature
Security Profiles (see PS 3.15).
Notes:
1. As technology advances,
additional encryption algorithms
may be allowed in future
versions. Implementations
should take this possibility into
account.
2. When symmetric encryption
is used, the certificate merely
identifies which key was used
by which entity, but not the
actual key itself. Some other
means (e.g., a trusted third
party) must be used to obtain
the key.
> Signature (0400,0120) 1 The MAC generated as *** not *** not
described in Section 12.2.1.1 supported supported
and encrypted using the *** ***
algorithm, parameters, and
private key associated with the
Certificate of the Signer
(0400,0115). See Section
C.12.1.1.3.1.2.
> Certified (0400,0305) 1C The type of certified timestamp *** not *** not
Timestamp Type used in the Certified Timestamp supported supported
(0400,0310) Attribute. Required *** ***
if Certified Timestamp
(0400,0310) is present. Defined
Terms:
CMS_TSP û Internet X.509
Public Key Infrastructure Time
Stamp Protocol Note: Digital
Signature Security Profiles (see
PS 3.15) may require the use of
a restricted subset of these
terms.
> Certified (0400,0310) 3 A certified timestamp of the *** not *** not
Timestamp Digital Signature (0400,0120) supported supported
Attribute Value, which shall be *** ***
obtained when the Digital
Signature is created. See
Section C.12.1.1.3.1.3.
Encrypted (0400,0500) 1C Sequence of Items containing *** not *** not
Attributes encrypted DICOM data. One or supported supported
Sequence more Items shall be present. *** ***
Required if application level
confidentiality is needed and
certain recipients are allowed to
decrypt all or portions of the
Encrypted Attributes Data Set.
See C.12.1.1.4.1.
> Encrypted (0400,0510) 1 Transfer Syntax used to encode *** not *** not
Content Transfer the encrypted content. Only supported supported
Syntax UID Transfer Syntaxes that explicitly *** ***
include the VR and use Little
Endian encoding shall be used.
> Encrypted (0400,0520) 1 Encrypted data. See *** not *** not
Content C.12.1.1.4.2. supported supported
*** ***
HL7 Structured (0040,A390) 1C Sequence of items defining *** not *** not
Document mapping and/or access supported supported
Reference mechanism for HL7 Structured *** ***
Sequence Documents referenced from the
current SOP Instance. One or
more Items may be included in
this sequence. See C.12.1.1.6.
Required if HL7 Structured
Documents are referenced
within the Instance.
> Referenced (0008,1150) 1 Unique identifier for the class of *** not *** not
SOP Class UID HL7 Structured Document. supported supported
*** ***
> Referenced (0008,1155) 1 Unique identifier for the HL7 *** not *** not
SOP Instance Structured Document as used supported supported
UID in DICOM instance references. *** ***
> HL7 Instance (0040,E001) 1 Instance Identifier of the *** not *** not
Identifier referenced HL7 Structured supported supported
Document, encoded as a UID *** ***
(OID or UUID), concatenated
with a caret ("^") and Extension
value (if Extension is present in
Instance Identifier).
> Retrieve URI (0040,E010) 3 Retrieval access path to HL7 *** not *** not
Structured Document. Includes supported supported
fully specified scheme, *** ***
authority, path, and query in
accordance with RFC 2396
D.1 RT Plan
The XML data structure for RT Plan IODs is contained in private Extended Interface
Module which is optional within RT Plan IOD.
The structure of the XML data structure is described in the following table.
Column “Operational Mode” indicates the presence for certain XML nodes depending of
the applications Operational Mode.
Legend for column “Operational Mode”:
✓ Attribute supported
Attribute not supported
1) Present for Block, Compensator and empty Trays
1)* Present for empty Trays
Attributes not supported anymore in the Extended Interface XML data are covered by
standard DICOM attributes.
Extended Interface, Extended Interface – Private
8.8 / QA
6.5
ExtendedVAPlanInterface 0 1 ✓ ✓
> Beams 0 1 ✓ ✓
>> Beam 0 ∞ ✓ ✓
>>>>> CustomAddOn 0 4 1)
>>>>>> CustomAddOnType 1 1 string Three Add On Type to choose from: 1)
BLOCK
COMPENSATOR
TRAY
>>>>>> CustomCode 0 1 int Codes for accessory which may be custom 1)
made or factory made (in that case the
8.8 / QA
6.5
internal code in equal the custom code.)
>>>>>> ReferencedAddOnNumber 0 1 int Reference to the identification number of the 1)
Block or Compensator.
>>>>>> TraySpecification 0 1 In case the CustomAddOn is a tray, it gets 1)*
reported back through this node
>>>>>>> TrayID 1 1 string User-supplied identifier for tray. Max length 1)*
is 16 characters.
>>>>>>> SourceTrayDistance 0 1 int The distance from source to slot. This 1)*
information is needed to distinguish the slot
> ToleranceTables 0 1 ✓ ✓
>> DoseReference 0 ∞ ✓ ✓
8.8 / QA
6.5
>>>>> Breakpoint 0 ∞ Breakpoint extension since DICOM has just ✓ ✓
one DeliveryWarningDose without a
description.
>>>>>> BreakpointDose 1 1 decimal Unit is [Gy] ✓ ✓
>>>>>> BreakpointWarning 1 1 string Text which describes the breakpoint. Max ✓ ✓
length is 254 characters
Min Max
8.8 / QA
6.5
RTBeamSessionRecordExtendedIF 0 1 ✓ ✓
> DoseOverrides 0 1 ✓ ✓
8.8 / QA
6.5
DAILYDOSE: Daily dose limit overridden
SESSIONDOSE: Session dose limit
overridden
TOTALDOSE:
>>> Dose 0 1 decimal Contains the specific overridden/authorized ✓ ✓
dose for DoseOverrideType ‘BREAKPOINT’
at the Reference Point referenced by Dose
Reference Number. Unit is [Gy]
>>> DoseReferenceNumber 1 1 non- References the Reference Point on which ✓ ✓
negative the dose was applied.
int
>>> OperatorsName 1 1 Operator Name who overrode/approved the ✓ ✓
specific doses.
>>>> FamilyName 0 1 string Max length is 64 characters. ✓ ✓
> TraySpecifications 0 1 ✓