Continious Glucose Monitors
Continious Glucose Monitors
com/article/2500145-print
emedicine.medscape.com
Continuous Glucose
Monitors/Hybrid Closed-Loop
Systems
Updated: May 02, 2024
Author: Satish K Garg, MD, MBBS, DM;
Overview
Since the 1990s, significant improvement has been seen in the development and implementation of insulin pumps,
continuous glucose monitors (CGMs), and hybrid closed-loop (HCL) systems.[1] The different technologies, used
independently or in combination, have improved glucose control, allowing patients to reduce their time above and below
appropriate glucose ranges.[2, 3, 4] These advances have also facilitated telehealth and virtual care worldwide, especially
during the coronavirus disease 2019 (COVID-19) pandemic.[5, 6, 7, 8, 9, 10, 11, 12, 13] However, while the adoption of
such technologies has improved in developed economies, their widespread use in emerging economies is still limited due to
barriers such as cost, knowledge gaps, and implementation challenges.[14, 15]
Indeed, although approximately 150-200 million people worldwide require insulin for optimal diabetes management, most
using multiple daily injections,[16, 17] a pre-2020 estimate indicated that less than 1% of these patients employ continuous
subcutaneous insulin infusion (CSII).[16, 18, 19]
Diabetes prevalence has increased worldwide, with more than 500 million people now living with the disease,[20] including
35-40 million in the United States.[21] This prevalence is higher in minority and socioeconomically disadvantaged
populations,[22, 23, 24, 25] but it is known that these groups do not receive the optimum diabetes care and are not offered
the same diabetes technologies.[26, 27, 28, 29] By 2017, the annual total cost of diabetes in the United States had grown
to $327 billion.[30]
Despite the development of new diabetes technologies and therapeutics,[31, 32, 33] the life expectancy of persons with
type 1 diabetes is 10-12 years below that of the general population.[34, 35, 36, 37]
Patients with diabetes previously monitored their glucose levels through urine analysis,[38] but during the 1980s, self
monitoring of blood glucose (SMBG) became available, allowing patients to test their glucose level via a finger stick.[38]
Blood was collected on a test strip and analyzed by an SMBG device, with SMBG coming to be considered the standard of
care in diabetes, especially for patients on insulin therapy.[39]
CGM accuracy has increased significantly since the end of the 20th century.[43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54,
55, 56] Indeed, most CGMs can now measure glucose values every 1-5 minutes,[57] and there have been improvements in
the precision of CGM data (with lowering of “the random noise component overlapped to the true glycemic signal”).[58]
CGM use has, in fact, seen a significant rise since 2010 due to advances in sensor size, accuracy, features, algorithms, and
connectability with insulin pumps and "smart" insulin pens.[59] (Launched in 2020, the first smart insulin pen with CGM
integration, Medtronic Diabetes’ InPen with Real-Time Guardian Connect CGM Data, allowed glucose readings and insulin
dose information to be viewed in the same app.[60] ) By 2022, there were more than 7 million CGM users worldwide.[61]
Current CGM manufacturers include Medtronic Diabetes (Northridge, CA), Dexcom Inc. (San Diego, CA), Abbott Diabetes
Care (Alameda, CA), and Senseonics Inc. (Germantown, MD).
https://emedicine.medscape.com/article/2500145-print 1/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
CGM use has been shown to increase time in range (TIR; the amount of time blood glucose levels for a person with diabetes
stay within a recommended target range), improve glucose control (A1c), and decrease episodes of diabetic ketoacidosis
and severe hypoglycemia. It can also help in the early detection of type 1 and type 2 diabetes.[62, 63, 64, 65, 66, 67]
A prospective, observational study by Charleer et al provided evidence that the use of real-time CGMs (rtCGMs) improves
glycemic control in persons with type 1 diabetes in whom such control has been insufficient. The study involved individuals
who had been using CSII therapy and then entered an rtCGM reimbursement program. The investigators found that overall
baseline HbA1c declined from 7.7% to, after participants had been in the program for 12 months, 7.4%, with the reduction
being greater in patients who had begun CGM use due to insufficient and variable glycemic control than in those who had
switched to CGMs owing to hypoglycemia or an ongoing or planned pregnancy. Moreover, while the rate at which patients
were admitted to the emergency room or hospital as a result of severe hypoglycemia and/or ketoacidosis was 16% in the
year prior to the start of rtCGM reimbursement, that figure dropped to 4% in the year after entry into the program.[68]
The first professional CGM, a Medtronic device in which glucose data were downloaded for review after 3 days, was
approved by the FDA in 1999.[69] In 2001, the GlucoWatch biographer (Cygnus Inc, Redwood City, CA) was approved. An
rtCGM, it displayed glucose data every 20 minutes for 12 hours by utilizing reverse iontophoresis, a process through which
the secretion of subcutaneous fluid could be stimulated, with the fluid's glucose being measured via an electrode.[43, 44, 70,
38]
The first transcutaneous CGM sensors were approved by the FDA to be worn for 3 days. In 2007, however, with FDA
approval of the Dexcom SEVEN, wear time was increased to 7 days.[71, 72] Transcutaneous CGM sensors can now often
be worn for 7-14 days.[72, 73]
In the past, CGMs required SMBG calibrations, but newer devices serve as standalone sources of clinical decisions.[43, 44,
45, 46, 47, 48, 49, 50, 51, 52, 55, 56] Indeed, in December 2016, the Dexcom G5 Mobile CGM system became the first
CGM to receive FDA approval for "non-adjunctive" use, which meant that the device could be used in place of, rather than
as an adjunct to, finger-stick blood glucose monitoring when decisions about insulin administration or the treatment of
hypoglycemia were being made.[74]
Approved by the FDA in 2018, Dexcom's G6 included compatibility with different insulin pumps, as well as blood glucose
monitors and other electronic diabetes management devices, becoming the first CGM to be approved for such
interoperability.[75, 76] An interoperable CGM (iCGM) designation, as outlined by the FDA, requires a lower mean absolute
relative difference (MARD), with accuracy being great in devices having reduced MARD values. In addition, paracetamol
and/or vitamin C, which can falsely increase glucose readings, must not interfere with the iCGM's measurements.[77]
CGM manufacturers are working on developing CGMs that can take continuous ketone and/or lactate measurements
through the same device as a means of reducing the likelihood of diabetic ketoacidosis.[78]
The Libre 2 includes hypoglycemic and hyperglycemic alerts,[82, 83] and the Libre 3 is an rtCGM in which the data are
displayed every minute on a smartphone only.[84] The Libre 3 is an interoperable CGM. In 2023, modified versions of the
Libre 2 and 3 were FDA-approved for use with automated insulin delivery (AID) systems (discussed below).[85]
Implantable sensors
While transcutaneous CGM sensors are inserted by the patients themselves, a minor surgical procedure is required to insert
and remove an implantable sensor, which is placed subcutaneously in the upper arm.[73]
In the early 2000s, Dexcom investigated an implantable sensor that was the size of an AA battery. However, the sensor had
high MARD values due to foreign body reactions.[46] The studies were discontinued early, and it was never filed for FDA
approval. In 2018, the Eversense CGM, the first CGM with an implantable sensor, was approved for 90-day use by the FDA.
[86, 87] In 2022, the FDA approved the Eversense E3 CGM implantable sensor for 180-day use.[88] In 2024, the FDA
approved it as the first implantable CGM to be used in an AID.[89]
Although the use of CGMs has primarily been aimed at persons with type 1 diabetes, the devices have come to increasingly
be employed in type 2 diabetes.[90] For example, the CGM device sugarBEAT (Nemaura Medical; Loughborough, UK) was
developed specifically for individuals with type 2 diabetes, being produced for such patients who are not at high risk for
hypoglycemia; it can also be utilized in persons with prediabetes. The sugarBEAT CGM uses an adhesive patch and sensor,
drawing glucose molecules from the interstitial fluid just beneath the skin’s surface for measurement. A Bluetooth connection
is employed to transmit data every 5 minutes to a smartphone app. The device is worn for 14 hours at a time during the day
and for just 2-4 days monthly. Although sugarBEAT is noninvasive, a once-daily fingerstick is still needed for calibration.
https://emedicine.medscape.com/article/2500145-print 2/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
In May 2019, it received a CE (Conformité Européenne) Mark in Europe for use as a class IIb medical device. Although still
awaiting FDA approval in the United States, permission has in the meantime been granted for this CGM to be marketed as a
“wellness” device, with sugarBEAT in this capacity producing retroactive reports for the physician and patient rather than
real-time values.[91]
In normal, healthy individuals, glucose levels are tightly regulated via continuous delivery of insulin, resulting from the ability
of beta cells in the pancreas to sense glucose values. CSII through insulin pumps attempts to mimic this physiology,[92] with
intensive insulin therapy improving glycemic control and significantly reducing microvascular and macrovascular
complications of insulin-dependent diabetes.[93]
Insulin pumps typically include the pump itself, which is about the size of a small cell phone and contains an insulin reservoir;
a small catheter inserted into the patient’s subcutaneous tissue; and tubing that delivers insulin from the pump to the
catheter. The pump can be worn in various locations, such as under the clothing or on a belt. Tubeless pumps, also known
as patch pumps, are available as well; with these, a small pump/reservoir is attached to the skin with adhesive and
connected directly to the catheter, with no tubing in between. Insulin delivery via a patch pump is programmed through a
handheld device.[94, 95]
Currently, there are three main manufacturers of insulin pumps in the United States: Medtronic Diabetes, Tandem Diabetes
Care (San Diego, CA), and Insulet Corporation (Acton, MA).
The first insulin pump was developed by Kadish in 1963, as part of a large, portable closed-loop system (as discussed
below). Owing to its size, however, the system was not practical for daily use and garnered little attention.[96] In 1974, Miles
Laboratories (Elkhart, IN) developed the Biostator, the first commercial inpatient insulin pump (also in a closed-loop system),
[97, 98] while in 1976 Kamon created the first portable autosyringe insulin pump.[96] Like Kadish's device, however, these
pumps were bulky and complex, making them infeasible for outpatient use.
In 1983, the MiniMed 502, from MiniMed Inc., was introduced.[99, 100] (The company was later acquired by Medtronic.)
Many versions of this product have since been produced, with the pumps improving glucose trend tracking and thus helping
users administer insulin corrections and understand what contributed to fluctuations in insulin values.[101]
Insulet released the first tubeless pump, the OmniPod Insulin Management System, in 2011.[102, 99] Tandem released the
t:slim Insulin Delivery System, the first touch-screen insulin pump available in the United States, in 2012.[103]
Over time, insulin pumps have evolved to receive wireless transmission of glucose values from a CGM and to have
programmable delivery patterns and remote control capabilities.
With insulin delivery devices and CGMs having come a long way, it became obvious that they could be integrated into what
are now commonly known as hybrid closed-loop (HCL) systems, or so-called artificial pancreases, using an algorithm to
automatically start or stop insulin delivery. Unliked a fully closed-loop system, an HCL still requires patients to determine the
number of carbohydrates in their food and input that data into the system, manually requesting the insulin dose needed for
meals.[104]
In 2006, the MiniMed Paradigm REAL-Time system 515 displayed glucose values on the insulin pump.[105] MiniMed
Paradigm Veo featured “low-glucose suspend” (LGS), which stopped insulin delivery if the user reached a preset low (40-110
mg/dL).[106] It was released internationally in 2009 but was not approved in the United States. In 2013, however,
Medtronic's MiniMed 530G received FDA approval. Featuring “threshold suspend” (TS), it suspended insulin delivery when
the sensor's glucose value was between 60-90 mg/dL.[107] Use of TS reduced nocturnal hypoglycemic events by 37.5%
(ASPIRE [Automation to Simulate Pancreatic Insulin REsponse] In-Home study).[108] The 640G, which was launched in
2015 but was not approved in the United States, included "predicted LGS" (PLGS).[109] The PLGS system predicted
hypoglycemia within 20 mg/dL of the preset range and stopped insulin delivery 30 minutes before the threshold.[110]
The first true HCL system was the Medtronic MiniMed 670G system, approved in the United States in 2016.[111, 112, 113]
This system also included a PLGS function. In addition, it featured AID for hyperglycemia[114] and adjusted the basal
insulin delivery based on a proportional integral derivative (PID) algorithm.[115] In a single-arm registration study, the 670G
system was shown to significantly reduce hypoglycemia in patients with type 1 diabetes.[112, 116] However, the system
required frequent calibrations (as a safety mechanism), especially when there were periods of minimum and maximum
insulin delivery. Although initially the 670G was well accepted by patients with type 1 diabetes, due to the repeated need for
calibrations and multiple alarms, many patients discontinued the system.[117]
The 780G system, commonly referred to as an advanced HCL (AHCL) system, has gained favor with patients because it
does not require frequent calibrations and is able to automatically bolus insulin every 5 minutes in the event of unanticipated
hyperglycemia.[118] It has been available in Europe since 2020 and was approved in the United States in April 2023.[119,
https://emedicine.medscape.com/article/2500145-print 3/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
120, 121, 122] The early registration studies on 780G showed significant improvement in TIR, with a reduction in
hypoglycemia, especially nocturnally;[123] many real-world studies have confirmed the registration study data.[124, 125]
Tandem’s t:slim X2 device, approved by the FDA in 2019, was the first so-called alternate controller-enabled (ACE) infusion
pump, which meant that it was designed to be interoperable with any compatible CGM, blood glucose monitor, or automated
insulin dosing system. As a result, in contrast to the MiniMed 670G HCL, in which the insulin pump is compatible only with a
Medtronic CGM using the company’s own Guardian Sensor 3, the t:slim X2 is compatible with the Dexcom G6.[126]
The t:slim X2 pump utilizes a model predictive control (MPC) algorithm, which calculates insulin delivery by reducing "the
difference between model-predicted glucose concentrations and target glucose over a pre-specified prediction time horizon."
[127, 128, 115]
The t:slim X2 has an automatic glucose suspend feature that stops insulin delivery if the sensor glucose value is less than 70
mg/dL.{ref110) Additionally, it includes a PLGS feature that interrupts delivery if glucose values are expected to fall below 80
mg/dL in 30 minutes.[129] The Control-IQ version of this pump corrects predicted hyperglycemia (>180 mg/dL) by
automatically delivering a correction bolus every hour and adjusts basal insulin delivery to maintain glucose values between
112-160 mg/dL.[130]
In 2022, Insulet's OmniPod 5 was approved by the FDA, becoming the first HCL system to feature a tubeless insulin pump.
[131, 132] This system employs an MPC algorithm similar to that of the t:slim X2. In a single-arm study, use of the OmniPod
5 was shown to produce an increase in TIR in addition to a reduction in hypoglycemia.[133, 134]
The HCL systems sold in the United States by all three of the above companies are associated with increased TIR, with
reduction in hypoglycemia.[135, 136, 137] In addition, HCL systems allow higher glucose thresholds for patients during
exercise.[111, 112, 119, 138] None of the systems have been approved for use in pregnant individuals with diabetes.[139]
However, off-label use has shown significant improvement in glycemia and maternal and fetal outcomes in pregnant patients
with type 1 diabetes.[140, 141]
Next steps
Efforts are underway to make a true artificial pancreas, which, using a combination of insulin and glucagon, would mimic
normal beta-cell function.[142] Another aim, in order to better control post-prandial glucose levels, is to develop an artificial
pancreas employing pramlintide with insulin.[143, 144, 145, 146, 147]
In 2023, researchers from the University of Cambridge reported on the successful trial of a fully closed-loop artificial
pancreas for patients with type 2 diabetes. The device used an off-the-shelf glucose monitor and insulin pump, which were
combined with an app, CamAPS HX, designed by the investigators to predict insulin requirements. In this way, the artificial
pancreas functioned independently, without needing input from the patient. Employing a glucose target range of 3.9-10.0
mmol/L, the trial demonstrated that patients using the closed-loop device remained in the target range 66.3% of the time,
compared with 32.3% of the time for patients on standard therapy.[148, 149]
Author
Satish K Garg, MD, MBBS, DM Professor, Departments of Medicine and Pediatrics, Barbara Davis Center for Childhood
Diabetes, Director of Adult Diabetes Program, Garg Endowed Clinical and Research Chairs, University of Colorado Denver
School of Medicine; Editor-in-Chief, Diabetes Technology and Therapeutics
Satish K Garg, MD, MBBS, DM is a member of the following medical societies: American Association of Clinical
Endocrinology, American Diabetes Association, Association of Physicians of India, Diabetic Association of India, Endocrine
Society, European Association for the Study of Diabetes, Geriatric Society of India, Indian Society for Atherosclerosis
Research, International Diabetes Federation, Pediatric Endocrine Society, Research Society for the Study of Diabetes in
India
Disclosure: Serve(d) as a director, officer, partner, employee, advisor, consultant or trustee for: Eli Lilly and Company, Novo-
nordisk, Medtronic, Dexcom, Lifescan, and Bayer <br/>Received research grant from: Eli Lilly, Dexcom, Dario, JDRF, T1D
Exchange, Jaeb Center, Medtronic, .
Coauthor(s)
Erika Rodriguez, MS, BA Clinical Research Assistant, Barbara Davis Center for Childhood Diabetes, University of Colorado
Denver School of Medicine
Abdulhalim Muidh Almurashi, MD Diabetes and Family Medicine Consultant, Fellow in Diabetes Technology, Barbara
Davis Center for Childhood Diabetes, University of Colorado Denver School of Medicine
Abdulhalim Muidh Almurashi, MD is a member of the following medical societies: American Diabetes Association, Saudi
https://emedicine.medscape.com/article/2500145-print 4/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
Scientific Diabetes Society
Mary L Windle, PharmD Adjunct Associate Professor, University of Nebraska Medical Center College of Pharmacy; Editor-
in-Chief, Medscape Drug Reference
References
1. Zimmerman C, Albanese-O'Neill A, Haller MJ. Advances in Type 1 Diabetes Technology Over the Last Decade. Eur Endocrinol.
2019 Aug. 15 (2):70-6. [QxMD MEDLINE Link]. [Full Text].
2. Bratke H, Margeirsdottir HD, Assmus J, Njolstad PR, Skrivarhaug T. Does Current Diabetes Technology Improve Metabolic
Control? A Cross-Sectional Study on the Use of Insulin Pumps and Continuous Glucose Monitoring Devices in a Nationwide
Pediatric Population. Diabetes Ther. 2021 Sep. 12 (9):2571-83. [QxMD MEDLINE Link]. [Full Text].
3. Eiland L, Thangavelu T, Drincic A. Has Technology Improved Diabetes Management in Relation to Age, Gender, and Ethnicity?.
Curr Diab Rep. 2019 Nov 4. 19 (11):111. [QxMD MEDLINE Link].
4. Bailey TS, Walsh J, Stone JY. Emerging Technologies for Diabetes Care. Diabetes Technol Ther. 2018 Jun. 20 (S2):S278-84.
[QxMD MEDLINE Link]. [Full Text].
5. Norgaard K. Telemedicine Consultations and Diabetes Technology During COVID-19. J Diabetes Sci Technol. 2020 Jul. 14
(4):767-8. [QxMD MEDLINE Link]. [Full Text].
6. Akturk HK, Garg S. Technological advances shaping diabetes care. Curr Opin Endocrinol Diabetes Obes. 2019 Apr. 26 (2):84-9.
[QxMD MEDLINE Link].
7. Phillip M, Bergenstal RM, Close KL, et al. The Digital/Virtual Diabetes Clinic: The Future Is Now-Recommendations from an
International Panel on Diabetes Digital Technologies Introduction. Diabetes Technol Ther. 2021 Feb. 23 (2):146-54. [QxMD
MEDLINE Link]. [Full Text].
8. Garg S, Norman GJ. Impact of COVID-19 on Health Economics and Technology of Diabetes Care: Use Cases of Real-Time
Continuous Glucose Monitoring to Transform Health Care During a Global Pandemic. Diabetes Technol Ther. 2021 Mar. 23
(S1):S15-S20. [QxMD MEDLINE Link]. [Full Text].
9. Bornstein SR, Rubino F, Khunti K, et al. Practical recommendations for the management of diabetes in patients with COVID-19.
Lancet Diabetes Endocrinol. 2020 Jun. 8 (6):546-50. [QxMD MEDLINE Link]. [Full Text].
10. Garg SK, Rodbard D, Hirsch IB, Forlenza GP. Managing New-Onset Type 1 Diabetes During the COVID-19 Pandemic:
Challenges and Opportunities. Diabetes Technol Ther. 2020 Jun. 22 (6):431-9. [QxMD MEDLINE Link].
11. Jones MS, Goley AL, Alexander BE, Keller SB, Caldwell MM, Buse JB. Inpatient Transition to Virtual Care During COVID-19
Pandemic. Diabetes Technol Ther. 2020 Jun. 22 (6):444-8. [QxMD MEDLINE Link]. [Full Text].
12. Murphy HR. Managing Diabetes in Pregnancy Before, During, and After COVID-19. Diabetes Technol Ther. 2020 Jun. 22
(6):454-61. [QxMD MEDLINE Link]. [Full Text].
13. Peters AL, Garg SK. The Silver Lining to COVID-19: Avoiding Diabetic Ketoacidosis Admissions with Telehealth. Diabetes
Technol Ther. 2020 Jun. 22 (6):449-53. [QxMD MEDLINE Link].
14. Vigersky RA. The benefits, limitations, and cost-effectiveness of advanced technologies in the management of patients with
diabetes mellitus. J Diabetes Sci Technol. 2015 Mar. 9 (2):320-30. [QxMD MEDLINE Link]. [Full Text].
15. Walker AF, Hood KK, Gurka MJ, et al. Barriers to Technology Use and Endocrinology Care for Underserved Communities With
Type 1 Diabetes. Diabetes Care. 2021 Jul. 44 (7):1480-90. [QxMD MEDLINE Link]. [Full Text].
16. Garg SK, Rewers AH, Akturk HK. Ever-Increasing Insulin-Requiring Patients Globally. Diabetes Technol Ther. 2018 Jun. 20
(S2):S21-4. [QxMD MEDLINE Link]. [Full Text].
17. Masierek M, Nabrdalik K, Janota O, Kwiendacz H, Macherski M, Gumprecht J. The Review of Insulin Pens-Past, Present, and
Look to the Future. Front Endocrinol (Lausanne). 2022. 13:827484. [QxMD MEDLINE Link]. [Full Text].
18. van den Boom L, Karges B, Auzanneau M, et al. Temporal Trends and Contemporary Use of Insulin Pump Therapy and Glucose
Monitoring Among Children, Adolescents, and Adults With Type 1 Diabetes Between 1995 and 2017. Diabetes Care. 2019 Nov.
42 (11):2050-6. [QxMD MEDLINE Link]. [Full Text].
19. Gajewska KA, Biesma R, Bennett K, Sreenan S. Barriers and facilitators to accessing insulin pump therapy by adults with type 1
diabetes mellitus: a qualitative study. Acta Diabetol. 2021 Jan. 58 (1):93-105. [QxMD MEDLINE Link].
20. International Diabetes Federation. IDF Diabetes Atlas. 10th ed. Brussels, Belgium: International Diabetes Federation; 2022. [Full
Text].
https://emedicine.medscape.com/article/2500145-print 5/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
21. US Centers for Disease Control and Prevention. National Diabetes Statistics Report 2020: Estimates of Diabetes and Its Burden
in the United States. CDC. Available at https://www.cdc.gov/diabetes/pdfs/data/statistics/national-diabetes-statistics-report.pdf.
Accessed: July 5, 2022.
22. Link CL, McKinlay JB. Disparities in the prevalence of diabetes: is it race/ethnicity or socioeconomic status? Results from the
Boston Area Community Health (BACH) survey. Ethn Dis. 2009 Summer. 19 (3):288-92. [QxMD MEDLINE Link]. [Full Text].
23. Beckles GL, Chou CF. Disparities in the Prevalence of Diagnosed Diabetes - United States, 1999-2002 and 2011-2014. MMWR
Morb Mortal Wkly Rep. 2016 Nov 18. 65 (45):1265-9. [QxMD MEDLINE Link]. [Full Text].
24. Wilf-Miron R, Peled R, Yaari E, et al. Disparities in diabetes care: role of the patient's socio-demographic characteristics. BMC
Public Health. 2010 Nov 25. 10:729. [QxMD MEDLINE Link]. [Full Text].
25. Lado JJ, Lipman TH. Racial and Ethnic Disparities in the Incidence, Treatment, and Outcomes of Youth with Type 1 Diabetes.
Endocrinol Metab Clin North Am. 2016 Jun. 45 (2):453-61. [QxMD MEDLINE Link].
26. Addala A, Auzanneau M, Miller K, et al. A Decade of Disparities in Diabetes Technology Use and HbA1c in Pediatric Type 1
Diabetes: A Transatlantic Comparison. Diabetes Care. 2021 Jan. 44 (1):133-40. [QxMD MEDLINE Link]. [Full Text].
27. Agarwal S, Schechter C, Gonzalez J, Long JA. Racial-Ethnic Disparities in Diabetes Technology use Among Young Adults with
Type 1 Diabetes. Diabetes Technol Ther. 2021 Apr. 23 (4):306-13. [QxMD MEDLINE Link]. [Full Text].
28. McKergow E, Parkin L, Barson DJ, Sharples KJ, Wheeler BJ. Demographic and regional disparities in insulin pump utilization in
a setting of universal funding: a New Zealand nationwide study. Acta Diabetol. 2017 Jan. 54 (1):63-71. [QxMD MEDLINE Link].
29. Agarwal S, Simmonds I, Myers AK. The Use of Diabetes Technology to Address Inequity in Health Outcomes: Limitations and
Opportunities. Curr Diab Rep. 2022 Jul. 22 (7):275-281. [QxMD MEDLINE Link]. [Full Text].
30. American Diabetes Association. Economic Costs of Diabetes in the U.S. in 2017. Diabetes Care. 2018 May. 41 (5):917-28.
[QxMD MEDLINE Link]. [Full Text].
31. Deeb LC. Diabetes Technology During the Past 30 Years: A Lot of Changes and Mostly for the Bette. Diabetes Spectrum. 2008.
21 (2):[Full Text].
32. Distiller LA. Why do some patients with type 1 diabetes live so long?. World J Diabetes. 2014 Jun 15. 5 (3):282-7. [QxMD
MEDLINE Link]. [Full Text].
33. Sussman M, Benner J, Haller MJ, Rewers M, Griffiths R. Estimated Lifetime Economic Burden of Type 1 Diabetes. Diabetes
Technol Ther. 2020 Feb. 22 (2):121-30. [QxMD MEDLINE Link]. [Full Text].
34. Rawshani A, Rawshani A, Franzen S, et al. Mortality and Cardiovascular Disease in Type 1 and Type 2 Diabetes. N Engl J Med.
2017 Apr 13. 376 (15):1407-18. [QxMD MEDLINE Link]. [Full Text].
35. Huo L, Harding JL, Peeters A, Shaw JE, Magliano DJ. Life expectancy of type 1 diabetic patients during 1997-2010: a national
Australian registry-based cohort study. Diabetologia. 2016 Jun. 59 (6):1177-85. [QxMD MEDLINE Link]. [Full Text].
36. Heald AH, Stedman M, Davies M, et al. Estimating life years lost to diabetes: outcomes from analysis of National Diabetes Audit
and Office of National Statistics data. Cardiovasc Endocrinol Metab. 2020 Dec. 9 (4):183-5. [QxMD MEDLINE Link]. [Full Text].
37. Tran-Duy A, Knight J, Clarke PM, Svensson AM, Eliasson B, Palmer AJ. Development of a life expectancy table for individuals
with type 1 diabetes. Diabetologia. 2021 Oct. 64 (10):2228-36. [QxMD MEDLINE Link]. [Full Text].
38. Hirsch IB. Introduction: History of Glucose Monitoring. Hirsch IB, Battelino T, Peters AL, Chamberlain JJ, Aleppo G, Bergenstal
RM. Role of Continuous Glucose Monitoring in Diabetes Treatment. Arlington, Va: American Diabetes Association; 2018. [Full
Text].
39. Schnell O, Barnard K, Bergenstal R, et al. Clinical Utility of SMBG: Recommendations on the Use and Reporting of SMBG in
Clinical Research. Diabetes Care. 2015 Sep. 38 (9):1627-33. [QxMD MEDLINE Link]. [Full Text].
40. Sherwood JS, Russell SJ, Putman MS. New and Emerging Technologies in Type 1 Diabetes. Endocrinol Metab Clin North Am.
2020 Dec. 49 (4):667-78. [QxMD MEDLINE Link]. [Full Text].
41. Rodbard D. Continuous Glucose Monitoring: A Review of Recent Studies Demonstrating Improved Glycemic Outcomes.
Diabetes Technol Ther. 2017 Jun. 19 (S3):S25-S37. [QxMD MEDLINE Link]. [Full Text].
42. Garg SK, Liljenquist D, Bode B, et al. Evaluation of Accuracy and Safety of the Next-Generation Up to 180-Day Long-Term
Implantable Eversense Continuous Glucose Monitoring System: The PROMISE Study. Diabetes Technol Ther. 2022 Feb. 24
(2):84-92. [QxMD MEDLINE Link]. [Full Text].
43. Garg SK, Potts RO, Ackerman NR, Fermi SJ, Tamada JA, Chase HP. Correlation of fingerstick blood glucose measurements
with GlucoWatch biographer glucose results in young subjects with type 1 diabetes. Diabetes Care. 1999 Oct. 22 (10):1708-14.
[QxMD MEDLINE Link]. [Full Text].
44. Tamada JA, Garg S, Jovanovic L, Pitzer KR, Fermi S, Potts RO. Noninvasive glucose monitoring: comprehensive clinical results.
Cygnus Research Team. JAMA. 1999 Nov 17. 282 (19):1839-44. [QxMD MEDLINE Link]. [Full Text].
45. Chase HP, Roberts MD, Wightman C, et al. Use of the GlucoWatch biographer in children with type 1 diabetes. Pediatrics. 2003
Apr. 111 (4 Pt 1):790-4. [QxMD MEDLINE Link].
https://emedicine.medscape.com/article/2500145-print 6/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
46. Garg SK, Schwartz S, Edelman SV. Improved glucose excursions using an implantable real-time continuous glucose sensor in
adults with type 1 diabetes. Diabetes Care. 2004 Mar. 27 (3):734-8. [QxMD MEDLINE Link]. [Full Text].
47. Garg S, Zisser H, Schwartz S, et al. Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose
sensor: a randomized controlled trial. Diabetes Care. 2006 Jan. 29 (1):44-50. [QxMD MEDLINE Link]. [Full Text].
48. Garg S, Jovanovic L. Relationship of fasting and hourly blood glucose levels to HbA1c values: safety, accuracy, and
improvements in glucose profiles obtained using a 7-day continuous glucose sensor. Diabetes Care. 2006 Dec. 29 (12):2644-9.
[QxMD MEDLINE Link]. [Full Text].
49. Ellis SL, Bookout T, Garg SK, Izuora KE. Use of continuous glucose monitoring to improve diabetes mellitus management.
Endocrinol Metab Clin North Am. 2007 Dec. 36 Suppl 2:46-68. [QxMD MEDLINE Link].
50. Garg SK, Voelmle MK, Gottlieb P. Feasibility of 10-day use of a continuous glucose-monitoring system in adults with type 1
diabetes. Diabetes Care. 2009 Mar. 32 (3):436-8. [QxMD MEDLINE Link]. [Full Text].
51. Blevins TC, Bode BW, Garg SK, et al. Statement by the American Association of Clinical Endocrinologists Consensus Panel on
continuous glucose monitoring. Endocr Pract. 2010 Sep-Oct. 16 (5):730-45. [QxMD MEDLINE Link].
52. Garg SK, Voelmle MK, Beatson CR, et al. Use of continuous glucose monitoring in subjects with type 1 diabetes on multiple daily
injections versus continuous subcutaneous insulin infusion therapy: a prospective 6-month study. Diabetes Care. 2011 Mar. 34
(3):574-9. [QxMD MEDLINE Link]. [Full Text].
53. Moser EG, Morris AA, Garg SK. Emerging diabetes therapies and technologies. Diabetes Res Clin Pract. 2012 Jul. 97 (1):16-26.
[QxMD MEDLINE Link]. [Full Text].
54. Garg SK. Glucose monitoring: an important tool for improving glucose control and reducing hypoglycemia. Diabetes Technol
Ther. 2008 Jun. 10 (Suppl 1):S1-4.
55. Garg SK. The future of continuous glucose monitoring. Diabetes Technol Ther. 2009 Jun. 11 Suppl 1:S1-3. [QxMD MEDLINE
Link].
56. Rodbard D, Bailey T, Jovanovic L, Zisser H, Kaplan R, Garg SK. Improved quality of glycemic control and reduced glycemic
variability with use of continuous glucose monitoring. Diabetes Technol Ther. 2009 Nov. 11 (11):717-23. [QxMD MEDLINE Link].
[Full Text].
57. Reddy N, Verma N, Dungan K, et al. Monitoring Technologies- Continuous Glucose Monitoring, Mobile Technology, Biomarkers
of Glycemic Control. Endotext. 2020 Aug 16. [QxMD MEDLINE Link]. [Full Text].
58. Facchinetti A. Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges. Sensors (Basel). 2016
Dec 9. 16 (12):[QxMD MEDLINE Link]. [Full Text].
59. Smith MB, Albanese-O'Neill A, Macieira TGR, et al. Human Factors Associated with Continuous Glucose Monitor Use in Patients
with Diabetes: A Systematic Review. Diabetes Technol Ther. 2019 Oct. 21 (10):589-601. [QxMD MEDLINE Link].
60. Tucker ME. 'Smart' Insulin Pen With CGM First to Launch in Emerging Field. Medscape Medical News. 2020 Nov 17. [Full Text].
62. Lind M, Polonsky W, Hirsch IB, et al. Continuous Glucose Monitoring vs Conventional Therapy for Glycemic Control in Adults
With Type 1 Diabetes Treated With Multiple Daily Insulin Injections: The GOLD Randomized Clinical Trial. JAMA. 2017 Jan 24.
317 (4):379-87. [QxMD MEDLINE Link]. [Full Text].
63. Beck RW, Riddlesworth TD, Ruedy K, et al. Continuous Glucose Monitoring Versus Usual Care in Patients With Type 2 Diabetes
Receiving Multiple Daily Insulin Injections: A Randomized Trial. Ann Intern Med. 2017 Sep 19. 167 (6):365-74. [QxMD MEDLINE
Link].
64. Martens T, Beck RW, Bailey R, et al. Effect of Continuous Glucose Monitoring on Glycemic Control in Patients With Type 2
Diabetes Treated With Basal Insulin: A Randomized Clinical Trial. JAMA. 2021 Jun 8. 325 (22):2262-72. [QxMD MEDLINE Link].
[Full Text].
65. Soliman A, DeSanctis V, Yassin M, Elalaily R, Eldarsy NE. Continuous glucose monitoring system and new era of early diagnosis
of diabetes in high risk groups. Indian J Endocrinol Metab. 2014 May. 18 (3):274-82. [QxMD MEDLINE Link]. [Full Text].
66. Steck AK, Dong F, Geno Rasmussen C, et al. CGM Metrics Predict Imminent Progression to Type 1 Diabetes: Autoimmunity
Screening for Kids (ASK) Study. Diabetes Care. 2022 Feb 1. 45 (2):365-71. [QxMD MEDLINE Link].
67. Steck AK, Dong F, Taki I, et al. Continuous Glucose Monitoring Predicts Progression to Diabetes in Autoantibody Positive
Children. J Clin Endocrinol Metab. 2019 Aug 1. 104 (8):3337-44. [QxMD MEDLINE Link]. [Full Text].
68. Charleer S, Mathieu C, Nobels F, et al. Effect of Continuous Glucose Monitoring on Glycemic Control, Acute Admissions, and
Quality of Life: A Real-World Study. J Clin Endocrinol Metab. 2018 Mar 1. 103 (3):1224-32. [QxMD MEDLINE Link]. [Full Text].
69. Gross TM, Bode BW, Einhorn D, et al. Performance evaluation of the MiniMed continuous glucose monitoring system during
patient home use. Diabetes Technol Ther. 2000 Spring. 2 (1):49-56. [QxMD MEDLINE Link].
70. Tierney MJ, Tamada JA, Potts RO, Jovanovic L, Garg S, Cygnus Research Team. Clinical evaluation of the GlucoWatch
biographer: a continual, non-invasive glucose monitor for patients with diabetes. Biosens Bioelectron. 2001 Dec. 16 (9-12):621-9.
https://emedicine.medscape.com/article/2500145-print 7/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
[QxMD MEDLINE Link].
71. Zisser HC, Bailey TS, Schwartz S, Ratner RE, Wise J. Accuracy of the SEVEN continuous glucose monitoring system:
comparison with frequently sampled venous glucose measurements. J Diabetes Sci Technol. 2009 Sep 1. 3 (5):1146-54. [QxMD
MEDLINE Link]. [Full Text].
72. Didyuk O, Econom N, Guardia A, Livingston K, Klueh U. Continuous Glucose Monitoring Devices: Past, Present, and Future
Focus on the History and Evolution of Technological Innovation. J Diabetes Sci Technol. 2021 May. 15 (3):676-83. [QxMD
MEDLINE Link]. [Full Text].
73. Boscari F, Vettoretti M, Cavallin F, et al. Implantable and transcutaneous continuous glucose monitoring system: a randomized
cross over trial comparing accuracy, efficacy and acceptance. J Endocrinol Invest. 2022 Jan. 45 (1):115-24. [QxMD MEDLINE
Link]. [Full Text].
74. Coverage and Coding of Continuous Glucose Monitors. Medical Minute MD. Available at
https://www.cgsmedicare.com/jc/education/video/mm/GSM.pdf. March 15, 2017; Accessed: August 3, 2022.
75. FDA authorizes first fully interoperable continuous glucose monitoring system, streamlines review pathway for similar devices.
US Food and Drug Administration. Available at https://www.fda.gov/news-events/press-announcements/fda-authorizes-first-fully-
interoperable-continuous-glucose-monitoring-system-streamlines-review. March 27, 2018; Accessed: June 6, 2022.
76. Tucker ME. FDA Okays New Interoperable Continuous Glucose Monitor. Medscape Medical News. 2018 Mar 27. [Full Text].
77. De Novo request for classification of the Dexcom G6 Continuous Glucose Monitoring System: Results. FDA. Available at
https://www.accessdata.fda.gov/cdrh_docs/pdf17/DEN170088.pdf. March 27, 2018; Accessed: July 6, 2022.
78. Teymourian H, Moonla C, Tehrani F, et al. Microneedle-Based Detection of Ketone Bodies along with Glucose and Lactate:
Toward Real-Time Continuous Interstitial Fluid Monitoring of Diabetic Ketosis and Ketoacidosis. Anal Chem. 2020 Jan 21. 92
(2):2291-300. [QxMD MEDLINE Link].
79. Blum A. Freestyle Libre Glucose Monitoring System. Clin Diabetes. 2018 Apr. 36 (2):203-4. [QxMD MEDLINE Link]. [Full Text].
80. Crabtree TSJ, Choudhary P, Kar P, Wilmot EG. Flash glucose monitoring: the story so far and the journey ahead. BMJ
Innovations. 2022 Apr 25. [Full Text].
81. Flash Glucose Monitors (FreeStyle Libre) and Continuous Glucose Monitors (CGM). Diabetes UK. Available at
https://www.diabetes.org.uk/guide-to-diabetes/diabetes-technology/flash-glucose-monitors-and-continuous-glucose-monitors.
Accessed: August 19, 2022.
82. Wysham CH, Kruger DF. Practical Considerations for Initiating and Utilizing Flash Continuous Glucose Monitoring in Clinical
Practice. J Endocr Soc. 2021 Sep 1. 5 (9):bvab064. [QxMD MEDLINE Link]. [Full Text].
83. Alva S, Bailey T, Brazg R, et al. Accuracy of a 14-Day Factory-Calibrated Continuous Glucose Monitoring System With
Advanced Algorithm in Pediatric and Adult Population With Diabetes. J Diabetes Sci Technol. 2022 Jan. 16 (1):70-7. [QxMD
MEDLINE Link]. [Full Text].
84. Whooley S. FDA clears Abbott’s next-gen FreeStyle Libre 3 14-day CGM. Drug Delivery Business News. Available at
https://www.drugdeliverybusiness.com/fda-abbott-next-gen-freestyle-libre-3-cgm/. May 31, 2022; Accessed: July 6, 2022.
85. FDA Clears Integration of FreeStyle Sensors and Automated Insulin Delivery Systems. FDA News. Available at
https://www.fdanews.com/articles/211401-fda-clears-integration-of-freestyle-sensors-and-automated-insulin-delivery-systems.
March 8, 2023; Accessed: May 2, 2024.
86. Tweden KS, Deiss D, Rastogi R, Addaguduru S, Kaufman FR. Longitudinal Analysis of Real-World Performance of an
Implantable Continuous Glucose Sensor over Multiple Sensor Insertion and Removal Cycles. Diabetes Technol Ther. 2020 May.
22 (5):422-7. [QxMD MEDLINE Link]. [Full Text].
87. Tucker ME. FDA Approves First Implantable Continuous Glucose Monitor. Medscape Medical News. 2018 Jun 21. [Full Text].
88. Eversense E3 Continuous Glucose Monitoring System - P160048/S016. US Food and Drug Administration. Available at
https://www.fda.gov/medical-devices/eversense-e3-continuous-glucose-monitoring-system-p160048s016. Accessed: August 18,
2022.
89. Tucker ME. FDA Allows Implantable CGM to Integrate With Insulin Pumps. Medscape Medical News. 2024 May 1. [Full Text].
90. Kompala T, Neinstein A. A new era: increasing continuous glucose monitoring use in type 2 diabetes. Am J Manag Care. 2019
Mar. 25 (4 Spec No.):SP123-6. [QxMD MEDLINE Link]. [Full Text].
91. Tucker ME. Noninvasive, Low Cost CGM for Type 2 Diabetes Coming in US and EU. Medscape Medical News. 2020 Dec 2. [Full
Text].
92. Hartman I. Insulin analogs: impact on treatment success, satisfaction, quality of life, and adherence. Clin Med Res. 2008 Sep. 6
(2):54-67. [QxMD MEDLINE Link]. [Full Text].
93. Diabetes Control and Complications Trial Research Group, Nathan DM, Genuth S, et al. The effect of intensive treatment of
diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med.
1993 Sep 30. 329 (14):977-86. [QxMD MEDLINE Link]. [Full Text].
https://emedicine.medscape.com/article/2500145-print 8/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
94. Berget C, Messer LH, Forlenza GP. A Clinical Overview of Insulin Pump Therapy for the Management of Diabetes: Past,
Present, and Future of Intensive Therapy. Diabetes Spectr. 2019 Aug. 32 (3):194-204. [QxMD MEDLINE Link]. [Full Text].
96. Alsaleh FM, Smith FJ, Keady S, Taylor KM. Insulin pumps: from inception to the present and toward the future. J Clin Pharm
Ther. 2010 Apr. 35 (2):127-38. [QxMD MEDLINE Link]. [Full Text].
97. Fogt EJ, Dodd LM, Jenning EM, Clemens AH. Development and evaluation of a glucose analyzer for a glucose controlled insulin
infusion system ((Biostator). Clin Chem. 1978 Aug. 24 (8):1366-72. [QxMD MEDLINE Link].
98. Alsaleh FM, Smith FJ, Keady S, Taylor KM. Insulin pumps: from inception to the present and toward the future. J Clin Pharm
Ther. 2010 Apr. 35 (2):127-38. [QxMD MEDLINE Link]. [Full Text].
99. Kesavadev J, Saboo B, Krishna MB, Krishnan G. Evolution of Insulin Delivery Devices: From Syringes, Pens, and Pumps to DIY
Artificial Pancreas. Diabetes Ther. 2020 Jun. 11 (6):1251-69. [QxMD MEDLINE Link]. [Full Text].
100. Weaver KW, Hirsch IB. The Hybrid Closed-Loop System: Evolution and Practical Applications. Diabetes Technol Ther. 2018 Jun.
20 (S2):S216-23. [QxMD MEDLINE Link]. [Full Text].
101. Allen N, Gupta A. Current Diabetes Technology: Striving for the Artificial Pancreas. Diagnostics (Basel). 2019 Mar 15. 9 (1):
[QxMD MEDLINE Link]. [Full Text].
102. Zisser HC. The OmniPod Insulin Management System: the latest innovation in insulin pump therapy. Diabetes Ther. 2010 Aug. 1
(1):10-24. [QxMD MEDLINE Link]. [Full Text].
103. Schaeffer NE. Human factors research applied: the development of a personal touch screen insulin pump and users' perceptions
of actual use. Diabetes Technol Ther. 2013 Oct. 15 (10):845-54. [QxMD MEDLINE Link].
104. FDA approves first automated insulin delivery device for type 1 diabetes. US Food and Drug Administration. Available at
https://www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/ucm522974.htm. September 28, 2016; Accessed: August 19,
2022.
105. Mastrototaro J, Shin J, Marcus A, Sulur G, STAR 1 Clinical Trial Investigators. The accuracy and efficacy of real-time continuous
glucose monitoring sensor in patients with type 1 diabetes. Diabetes Technol Ther. 2008 Oct. 10 (5):385-90. [QxMD MEDLINE
Link].
106. Agrawal P, Welsh JB, Kannard B, Askari S, Yang Q, Kaufman FR. Usage and effectiveness of the low glucose suspend feature
of the Medtronic Paradigm Veo insulin pump. J Diabetes Sci Technol. 2011 Sep 1. 5 (5):1137-41. [QxMD MEDLINE Link]. [Full
Text].
107. Garg S, Brazg RL, Bailey TS, et al. Reduction in duration of hypoglycemia by automatic suspension of insulin delivery: the in-
clinic ASPIRE study. Diabetes Technol Ther. 2012 Mar. 14 (3):205-9. [QxMD MEDLINE Link].
108. Bergenstal RM, Klonoff DC, Garg SK, et al. Threshold-based insulin-pump interruption for reduction of hypoglycemia. N Engl J
Med. 2013 Jul 18. 369 (3):224-32. [QxMD MEDLINE Link]. [Full Text].
109. Zhong A, Choudhary P, McMahon C, et al. Effectiveness of Automated Insulin Management Features of the MiniMed® 640G
Sensor-Augmented Insulin Pump. Diabetes Technol Ther. 2016 Oct. 18 (10):657-63. [QxMD MEDLINE Link]. [Full Text].
110. Buckingham BA, Bailey TS, Christiansen M, et al. Evaluation of a Predictive Low-Glucose Management System In-Clinic.
Diabetes Technol Ther. 2017 May. 19 (5):288-92. [QxMD MEDLINE Link].
111. Bergenstal RM, Garg S, Weinzimer SA, et al. Safety of a Hybrid Closed-Loop Insulin Delivery System in Patients With Type 1
Diabetes. JAMA. 2016 Oct 4. 316 (13):1407-8. [QxMD MEDLINE Link].
112. Garg SK, Weinzimer SA, Tamborlane WV, et al. Glucose Outcomes with the In-Home Use of a Hybrid Closed-Loop Insulin
Delivery System in Adolescents and Adults with Type 1 Diabetes. Diabetes Technol Ther. 2017 Mar. 19 (3):155-63. [QxMD
MEDLINE Link]. [Full Text].
113. FDA Approves Hybrid Closed Loop System for People with Type 1 Diabetes. Endocrine News. Available at
https://endocrinenews.endocrine.org/medtronic-receives-fda-approval-for-hybrid-closed-loop-system-for-people-with-type-1-
diabetes/. September 2016; Accessed: August 5, 2022.
114. Messer LH, Berget C, Forlenza GP. A Clinical Guide to Advanced Diabetes Devices and Closed-Loop Systems Using the
CARES Paradigm. Diabetes Technol Ther. 2019 Aug. 21 (8):462-9. [QxMD MEDLINE Link]. [Full Text].
115. Boughton CK, Hovorka R. New closed-loop insulin systems. Diabetologia. 2021 May. 64 (5):1007-15. [QxMD MEDLINE Link].
[Full Text].
116. Forlenza GP, Pinhas-Hamiel O, Liljenquist DR, et al. Safety Evaluation of the MiniMed 670G System in Children 7-13 Years of
Age with Type 1 Diabetes. Diabetes Technol Ther. 2019 Jan. 21 (1):11-9. [QxMD MEDLINE Link]. [Full Text].
117. Knebel T, Neumiller JJ. Medtronic MiniMed 670G Hybrid Closed-Loop System. Clin Diabetes. 2019 Jan. 37 (1):94-5. [QxMD
MEDLINE Link]. [Full Text].
https://emedicine.medscape.com/article/2500145-print 9/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
118. Nimri R, Grosman B, Roy A, et al. Feasibility Study of a Hybrid Closed-Loop System with Automated Insulin Correction Boluses.
Diabetes Technol Ther. 2021 Apr. 23 (4):268-76. [QxMD MEDLINE Link].
119. Carlson AL, Sherr JL, Shulman DI, et al. Safety and Glycemic Outcomes During the MiniMed™ Advanced Hybrid Closed-Loop
System Pivotal Trial in Adolescents and Adults with Type 1 Diabetes. Diabetes Technol Ther. 2022 Mar. 24 (3):178-89. [QxMD
MEDLINE Link]. [Full Text].
120. FDA Roundup: April 21, 2023. US Food & Drug Administration. Available at https://www.fda.gov/news-events/press-
announcements/fda-roundup-april-21-2023. April 21, 2023; Accessed: April 27, 2023.
121. Peters AL. Why the Approval of MiniMed 780G Is a 'Quantum Leap' Forward. Medscape Diabetes & Endocrinology. Available at
https://www.medscape.com/viewarticle/991154?src=. April 26, 2023; Accessed: April 27, 2023.
122. McCall B. UK Launch for Most Advanced 'Artificial Pancreas', MiniMed 780G. Medscape Medical News. 2021 Jan 27. [Full Text].
123. Bergenstal RM, Nimri R, Beck RW, et al. A comparison of two hybrid closed-loop systems in adolescents and young adults with
type 1 diabetes (FLAIR): a multicentre, randomised, crossover trial. Lancet. 2021 Jan 16. 397 (10270):208-19. [QxMD MEDLINE
Link]. [Full Text].
124. Silva JD, Lepore G, Battelino T, et al. Real-World Performance of the MiniMed™ 780G System: First Report of Outcomes from
4120 Users. Diabetes Technol Ther. 2022 Feb. 24 (2):113-9. [QxMD MEDLINE Link]. [Full Text].
125. Collyns OJ, Meier RA, Betts ZL, et al. Improved Glycemic Outcomes With Medtronic MiniMed Advanced Hybrid Closed-Loop
Delivery: Results From a Randomized Crossover Trial Comparing Automated Insulin Delivery With Predictive Low Glucose
Suspend in People With Type 1 Diabetes. Diabetes Care. 2021 Apr. 44 (4):969-75. [QxMD MEDLINE Link]. [Full Text].
126. Tucker ME. FDA Approves First Interoperable Insulin Pump. Medscape Medical News. 2019 Feb 19. [Full Text].
127. Messer LH, Campbell K, Pyle L, Forlenza GP. Basal-IQ technology in the real world: satisfaction and reduction of diabetes
burden in individuals with type 1 diabetes. Diabet Med. 2021 Jun. 38 (6):e14381. [QxMD MEDLINE Link].
128. Pinsker JE, Muller L, Constantin A, et al. Real-World Patient-Reported Outcomes and Glycemic Results with Initiation of Control-
IQ Technology. Diabetes Technol Ther. 2021 Feb. 23 (2):120-7. [QxMD MEDLINE Link]. [Full Text].
129. Forlenza GP, Li Z, Buckingham BA, et al. Predictive Low-Glucose Suspend Reduces Hypoglycemia in Adults, Adolescents, and
Children With Type 1 Diabetes in an At-Home Randomized Crossover Study: Results of the PROLOG Trial. Diabetes Care. 2018
Oct. 41 (10):2155-61. [QxMD MEDLINE Link]. [Full Text].
130. Breton MD, Kovatchev BP. One Year Real-World Use of the Control-IQ Advanced Hybrid Closed-Loop Technology. Diabetes
Technol Ther. 2021 Sep. 23 (9):601-8. [QxMD MEDLINE Link]. [Full Text].
131. Brown SA, Forlenza GP, Bode BW, et al. Multicenter Trial of a Tubeless, On-Body Automated Insulin Delivery System With
Customizable Glycemic Targets in Pediatric and Adult Participants With Type 1 Diabetes. Diabetes Care. 2021 Jul. 44 (7):1630-
40. [QxMD MEDLINE Link]. [Full Text].
132. Buckingham BA, Forlenza GP, Pinsker JE, et al. Safety and Feasibility of the OmniPod Hybrid Closed-Loop System in Adult,
Adolescent, and Pediatric Patients with Type 1 Diabetes Using a Personalized Model Predictive Control Algorithm. Diabetes
Technol Ther. 2018 Apr. 20 (4):257-62. [QxMD MEDLINE Link]. [Full Text].
133. Sherr JL, Bode BW, Forlenza GP, et al. Safety and Glycemic Outcomes With a Tubeless Automated Insulin Delivery System in
Very Young Children With Type 1 Diabetes: A Single-Arm Multicenter Clinical Trial. Diabetes Care. 2022 Aug 1. 45 (8):1907-10.
[QxMD MEDLINE Link]. [Full Text].
134. Tucker ME. FDA Okays First Tubing-Free 'Artificial Pancreas' Omnipod 5. Medscape Medical News. 2022 Jan 28. [Full Text].
135. Petrovski G, Al Khalaf F, Campbell J, et al. One-year experience of hybrid closed-loop system in children and adolescents with
type 1 diabetes previously treated with multiple daily injections: drivers to successful outcomes. Acta Diabetol. 2021 Feb. 58
(2):207-13. [QxMD MEDLINE Link]. [Full Text].
136. Tornese G, Buzzurro F, Carletti C, Faleschini E, Barbi E. Six-Month Effectiveness of Advanced vs. Standard Hybrid Closed-Loop
System in Children and Adolescents With Type 1 Diabetes Mellitus. Front Endocrinol (Lausanne). 2021. 12:766314. [QxMD
MEDLINE Link]. [Full Text].
137. Janez A, Battelino T, Klupa T, et al. Hybrid Closed-Loop Systems for the Treatment of Type 1 Diabetes: A Collaborative, Expert
Group Position Statement for Clinical Use in Central and Eastern Europe. Diabetes Ther. 2021 Dec. 12 (12):3107-35. [QxMD
MEDLINE Link]. [Full Text].
138. Brown SA, Kovatchev BP, Raghinaru D, et al. Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1
Diabetes. N Engl J Med. 2019 Oct 31. 381 (18):1707-17. [QxMD MEDLINE Link]. [Full Text].
139. Moreno-Fernandez J, García-Seco JA. Commercialized Hybrid Closed-Loop System (Minimed Medtronic 670G) Results During
Pregnancy. AACE Clin Case Rep. 2021 May-Jun. 7 (3):177-9. [QxMD MEDLINE Link]. [Full Text].
140. Feig DS, Donovan LE, Corcoy R, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT): a
multicentre international randomised controlled trial. Lancet. 2017 Nov 25. 390 (10110):2347-59. [QxMD MEDLINE Link]. [Full
Text].
https://emedicine.medscape.com/article/2500145-print 10/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
141. Feig DS, Corcoy R, Donovan LE, et al. Pumps or Multiple Daily Injections in Pregnancy Involving Type 1 Diabetes: A
Prespecified Analysis of the CONCEPTT Randomized Trial. Diabetes Care. 2018 Dec. 41 (12):2471-9. [QxMD MEDLINE Link].
[Full Text].
142. Wilson LM, Jacobs PG, Ramsey KL, et al. Dual-Hormone Closed-Loop System Using a Liquid Stable Glucagon Formulation
Versus Insulin-Only Closed-Loop System Compared With a Predictive Low Glucose Suspend System: An Open-Label,
Outpatient, Single-Center, Crossover, Randomized Controlled Trial. Diabetes Care. 2020 Nov. 43 (11):2721-9. [QxMD MEDLINE
Link]. [Full Text].
143. Tsoukas MA, Cohen E, Legault L, et al. Alleviating carbohydrate counting with a FiASP-plus-pramlintide closed-loop delivery
system (artificial pancreas): Feasibility and pilot studies. Diabetes Obes Metab. 2021 Sep. 23 (9):2090-8. [QxMD MEDLINE
Link].
144. Weinzimer SA, Sherr JL, Cengiz E, et al. Effect of pramlintide on prandial glycemic excursions during closed-loop control in
adolescents and young adults with type 1 diabetes. Diabetes Care. 2012 Oct. 35 (10):1994-9. [QxMD MEDLINE Link]. [Full Text].
145. Wilson LM, Jacobs PG, Castle JR. Role of Glucagon in Automated Insulin Delivery. Endocrinol Metab Clin North Am. 2020 Mar.
49 (1):179-202. [QxMD MEDLINE Link]. [Full Text].
146. El-Khatib FH, Balliro C, Hillard MA, et al. Home use of a bihormonal bionic pancreas versus insulin pump therapy in adults with
type 1 diabetes: a multicentre randomised crossover trial. Lancet. 2017 Jan 28. 389 (10067):369-80. [QxMD MEDLINE Link].
[Full Text].
147. Haidar A, Tsoukas MA, Bernier-Twardy S, et al. A Novel Dual-Hormone Insulin-and-Pramlintide Artificial Pancreas for Type 1
Diabetes: A Randomized Controlled Crossover Trial. Diabetes Care. 2020 Mar. 43 (3):597-606. [QxMD MEDLINE Link]. [Full
Text].
148. Daly AB, Boughton CK, Nwokolo M, et al. Fully automated closed-loop insulin delivery in adults with type 2 diabetes: an open-
label, single-center, randomized crossover trial. Nat Med. 2023 Jan 11. [QxMD MEDLINE Link]. [Full Text].
149. [Guideline] University of Cambridge. Artificial Pancreas Developed That Can Help Maintain Healthy Glucose Levels in Type 2
Diabetes Patients. SciTechDaily. Available at https://scitechdaily.com/artificial-pancreas-developed-that-can-help-maintain-
healthy-glucose-levels-in-type-2-diabetes-patients/. January 11, 2023; Accessed: January 13, 2023.
150. Randelovic S, Bipat R. A Review of Coumarins and Coumarin-Related Compounds for Their Potential Antidiabetic Effect. Clin
Med Insights Endocrinol Diabetes. 2021. 14:11795514211042023. [QxMD MEDLINE Link]. [Full Text].
151. Garg SK, Hoff HK, Chase HP. The role of continuous glucose sensors in diabetes care. Endocrinol Metab Clin North Am. 2004
Mar. 33 (1):163-73, x-xi. [QxMD MEDLINE Link].
152. Lal RA, Basina M, Maahs DM, Hood K, Buckingham B, Wilson DM. One Year Clinical Experience of the First Commercial Hybrid
Closed-Loop System. Diabetes Care. 2019 Dec. 42 (12):2190-6. [QxMD MEDLINE Link]. [Full Text].
153. Christiansen MP, Garg SK, Brazg R, et al. Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor.
Diabetes Technol Ther. 2017 Aug. 19 (8):446-56. [QxMD MEDLINE Link]. [Full Text].
154. Whooley S. Medtronic wins CE mark for smart insulin pen, Guardian 4 sensor. Drug Delivery Business News. Available at
https://www.drugdeliverybusiness.com/medtronic-wins-ce-mark-for-smart-insulin-pen-cgm-system/. May 26, 2021; Accessed:
July 6, 2022.
155. Whooley S. Medtronic CEO confirms FDA warning could affect approval timing for next-gen diabetes tech. Drug Delivery
Business News. Available at https://www.drugdeliverybusiness.com/medtronic-ceo-confirms-fda-warning-affects-approval-timing-
for-next-gen-diabetes-tech-updates-on-pipeline/. January 10, 2022; Accessed: July 6, 2022.
156. Nakamura K, Balo A. The Accuracy and Efficacy of the Dexcom G4 Platinum Continuous Glucose Monitoring System. J
Diabetes Sci Technol. 2015 Mar 23. 9 (5):1021-6. [QxMD MEDLINE Link]. [Full Text].
157. Link M, Kamecke U, Waldenmaier D, Pleus S, Garcia A, Haug C, et al. Comparative Accuracy Analysis of a Real-time and an
Intermittent-Scanning Continuous Glucose Monitoring System. J Diabetes Sci Technol. 2021 Mar. 15 (2):287-93. [QxMD
MEDLINE Link]. [Full Text].
158. Garg SK, Kipnes M, Castorino K, et al. Accuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with
Diabetes. Diabetes Technol Ther. 2022 Jun. 24 (6):373-80. [QxMD MEDLINE Link]. [Full Text].
159. Weinstein RL, Schwartz SL, Brazg RL, Bugler JR, Peyser TA, McGarraugh GV. Accuracy of the 5-day FreeStyle Navigator
Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements. Diabetes Care. 2007
May. 30 (5):1125-30. [QxMD MEDLINE Link]. [Full Text].
160. Rao H, Fakourfar N, Sun C, Lewis J. The Use of Continuous Glucose Monitoring in Older People With Type 2 Diabetes. Sr Care
Pharm. 2021 Nov 1. 36 (11):556-67. [QxMD MEDLINE Link].
161. Burckhardt MA, Fried L, Bebbington K, et al. Use of remote monitoring with continuous glucose monitoring in young children with
Type 1 diabetes: the parents' perspective. Diabet Med. 2019 Nov. 36 (11):1453-9. [QxMD MEDLINE Link].
162. Agarwal S, Mathew J, Davis GM, et al. Continuous Glucose Monitoring in the Intensive Care Unit During the COVID-19
Pandemic. Diabetes Care. 2021 Mar. 44 (3):847-9. [QxMD MEDLINE Link]. [Full Text].
https://emedicine.medscape.com/article/2500145-print 11/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
163. Andres E, Meyer L, Zulfiqar AA, et al. Telemonitoring in diabetes: evolution of concepts and technologies, with a focus on results
of the more recent studies. J Med Life. 2019 Jul-Sep. 12 (3):203-14. [QxMD MEDLINE Link]. [Full Text].
165. Animas Corporation to Close Operations and Exit Insulin Pump Market. Johnson & Johnson. Available at
https://www.jnj.com/media-center/press-releases/animas-corporation-to-close-operations-and-exit-insulin-pump-market. October
5, 2017; Accessed: July 7, 2022.
166. Hoskins M. Roche Hands Over Remaining U.S. Insulin Pump Customers to Medtronic. Healthline. Available at
https://www.healthline.com/diabetesmine/roche-hands-off-remaining-pump-customers-medtronic#1. June 1, 2018; Accessed:
July 7, 2022.
169. Zhou K, Isaacs D. Closed-Loop Artificial Pancreas Therapy for Type 1 Diabetes. Curr Cardiol Rep. 2022 Jun 21. [QxMD
MEDLINE Link].
170. Amadou C, Franc S, Benhamou PY, et al. Diabeloop DBLG1 Closed-Loop System Enables Patients With Type 1 Diabetes to
Significantly Improve Their Glycemic Control in Real-Life Situations Without Serious Adverse Events: 6-Month Follow-up.
Diabetes Care. 2021 Mar. 44 (3):844-6. [QxMD MEDLINE Link]. [Full Text].
171. Lewis DM. Do-It-Yourself Artificial Pancreas System and the OpenAPS Movement. Endocrinol Metab Clin North Am. 2020 Mar.
49 (1):203-13. [QxMD MEDLINE Link].
172. Tauschmann M, Thabit H, Bally L, et al. Closed-loop insulin delivery in suboptimally controlled type 1 diabetes: a multicentre, 12-
week randomised trial. Lancet. 2018 Oct 13. 392 (10155):1321-9. [QxMD MEDLINE Link]. [Full Text].
173. Lilly and Ypsomed collaborate to advance an automated insulin delivery system for people with diabetes. Lilly Investors.
Available at https://investor.lilly.com/news-releases/news-release-details/lilly-and-ypsomed-collaborate-advance-automated-
insulin-delivery. November 19, 2020; Accessed: July 11, 2022.
174. Younk LM, Mikeladze M, Tate D, Davis SN. Exercise-related hypoglycemia in diabetes mellitus. Expert Rev Endocrinol Metab.
2011 Jan 1. 6 (1):93-108. [QxMD MEDLINE Link]. [Full Text].
175. Chung ST, Haymond MW. Minimizing morbidity of hypoglycemia in diabetes: a review of mini-dose glucagon. J Diabetes Sci
Technol. 2015 Jan. 9 (1):44-51. [QxMD MEDLINE Link]. [Full Text].
176. Basu R, Johnson ML, Kudva YC, Basu A. Exercise, hypoglycemia, and type 1 diabetes. Diabetes Technol Ther. 2014 Jun. 16
(6):331-7. [QxMD MEDLINE Link]. [Full Text].
177. Klonoff DC, Kerr D. Smart Pens Will Improve Insulin Therapy. J Diabetes Sci Technol. 2018 May. 12 (3):551-3. [QxMD MEDLINE
Link]. [Full Text].
178. Galindo RJ, Ramos C, Cardona S, et al. Efficacy of a Smart Insulin Pen Cap for the Management of Patients with Uncontrolled
Type 2 Diabetes: A Randomized Cross-Over Trial. J Diabetes Sci Technol. 2021 Jul 22. 19322968211033837. [QxMD MEDLINE
Link]. [Full Text].
179. Companion Medical InPen Receives FDA Clearance for Expanded Pediatric Use. Cision PR Newswire. Available at
https://www.prnewswire.com/news-releases/companion-medical-inpen-receives-fda-clearance-for-expanded-pediatric-use-
301081939.html. June 23, 2020; Accessed: July 11, 2022.
180. Sangave NA, Aungst TD, Patel DK. Smart Connected Insulin Pens, Caps, and Attachments: A Review of the Future of Diabetes
Technology. Diabetes Spectr. 2019 Nov. 32 (4):378-84. [QxMD MEDLINE Link]. [Full Text].
181. Rodbard D, Garg SK. Standardizing Reporting of Glucose and Insulin Data for Patients on Multiple Daily Injections Using
Connected Insulin Pens and Continuous Glucose Monitoring. Diabetes Technol Ther. 2021 Mar. 23 (3):221-6. [QxMD MEDLINE
Link]. [Full Text].
182. Bergenstal RM, Ahmann AJ, Bailey T, et al. Recommendations for standardizing glucose reporting and analysis to optimize
clinical decision making in diabetes: the Ambulatory Glucose Profile (AGP). Diabetes Technol Ther. 2013 Mar. 15 (3):198-211.
[QxMD MEDLINE Link].
183. Shah VN, Garg SK. Standardized Hybrid Closed-Loop System Reporting. Diabetes Technol Ther. 2021 May. 23 (5):323-31.
[QxMD MEDLINE Link]. [Full Text].
184. Al-Badri M, Hamdy O. Diabetes clinic reinvented: will technology change the future of diabetes care?. Ther Adv Endocrinol
Metab. 2021. 12:2042018821995368. [QxMD MEDLINE Link]. [Full Text].
185. de Kreutzenberg SV. Telemedicine for the Clinical Management of Diabetes; Implications and Considerations After COVID-19
Experience. High Blood Press Cardiovasc Prev. 2022 Jul. 29 (4):319-26. [QxMD MEDLINE Link]. [Full Text].
186. Chamberlain JJ. Continuous Glucose Monitoring Systems: Categories and Features. Role of Continuous Glucose Monitoring in
Diabetes Treatment. 2018 Aug. [QxMD MEDLINE Link]. [Full Text].
https://emedicine.medscape.com/article/2500145-print 12/13
11/4/24, 6:55 PM https://emedicine.medscape.com/article/2500145-print
https://emedicine.medscape.com/article/2500145-print 13/13