Sinothinker Technology Co.

,Limited

Hematology Analyzer SK8800 Specification

Feature :
●3-part differentiation of WBC, 23 parameters, single channel counter
●Volume measurement by time ,not wrong warning
●Advanced valve technology,long life
● RS232 interface, PC connecting
● More than 35 samples test per hour
● Electrical resistance for counting and SFT method of Hemiglobin
● Low sample consumption: venous 9.8 ul, capillary 9.6 ul, prediluted 20 ul for twice
testing one time
● 8.4 color TFT, Windows interface all testing parameter displayed simultaneously
● Windows operation system graphical buttons mouse and keyboard operation
● Double convolution and intelligent fitting
● Automatic diluting , Mixing , rinsing and clog clearing
● Automatically sample probe cleanning (inside and outside)
● Large storage capacity: more than 10,000 samples +3 histograms
● Internal thermal-sensitive printer or external printer.
● RS232 interface, PC connecting

Specification :
Methodology: Electrical resistance for counting, hemiglobin cyanide method and SFT
method for hemoglobin
Parameter: 3-part differentiation of WBC; 20 parameters and 3 color histograms
(WBC, RBC, PLT)
Work mode : Single Channel + unique Hemoglobin test system
Sample volume: 9.6μL for Venous and capillary mode, 20μL for prediluted mode
Throughput: More than 35 samples per hour, operatable 24 hours a day, auto sleeping
and waking-up functions
Storage: More than 100000 sample results including histograms can be stored,
convenient for inquiry and management of history data
Operation langue: English
QC control : X-B, L-J, X , SD, CV %
Refrence Value Setting : Male , Female, Children, Neonate
Alarm : 25 test of alert messages
Input / output : RS232, parallel printer and keyboard
Print : Graphic thermal printer with various printing format, optional external printer
Temperature : 18℃ - 30℃,wet ≤ 10-90%
Power supply : 220 V ±22 VAC, 50±1Hz
Dimension : 33 CM ( L ) * 38 CM ( W ) * 43 CM ( H )
Weight : 20 KG

1 Http://www.sxk8.com Email: Perry@sxk8.com Skype : perry.sinothinker


TEL:+86-755-86238653 Cel:+86-15268292499
Add: Nanshang Building 405, Zhuguang Road ,Nanshan District ,Shenzhen ,China
Sinothinker Technology Co.,Limited

Parameters :
WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB,HCT, MCV,
MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV,PDW,PCT,L-PCR
3 –histograms : WBC,RBC and PLT

Precision :
Parameters linear range CV %
WBC( 10 9 / L ) 0.0 - 99.9 ≤2%
RBC ( 1012 / L ) 0.0 - 9.99 ≤1.5%
MCV ( fL ) 40-150 ≤0.5%
PLT ( 10 9 / L ) 0-999 ≤4.0%
HGB( g/L ) 0.0-300.0 ≤1.5%

2 Http://www.sxk8.com Email: Perry@sxk8.com Skype : perry.sinothinker


TEL:+86-755-86238653 Cel:+86-15268292499
Add: Nanshang Building 405, Zhuguang Road ,Nanshan District ,Shenzhen ,China